Varian Cary 5000 User Manual

1"
"
User%guide%for%Cary%5000%absorption%spectrometer%
with%external%DRA%1800%attachment.%
(last&updated&03/17/2016)&
This"guide"is"for"use"of"the"Cary"5000"with"the"DRA"only.""For"use"without"the"DRA"as"a" standard"transmission"spectrometer"see"the"User"guide"for"Cary"5000"in"absorption" mode."
Important%warnings!%
Do"not"unplug"or"plug"in"the"external"DRA"attachment"when"the"instrument"is" on.""This"will"ruin"the"detectors"of"the"integrating"sphere"
Do"not"put"white"light"into"the"DRA"with"the"DRA"electrically"connected"to"the" Cary."
Minimize"the"amount"of"time"that"the"integrating"sphere"is"open."Long-term"light" exposure"hurt"the"detector"over"time."""
Wear"clean"gloves"while"using"the"instrument"
Do"not"use"liquid"or"powder"samples"in"the"integrating"sphere."
Quick%Start%for%using%DRA%
1. Turn"on"computer""
2. Check"lens"and"mirror"are"correct"for"your"measurement."
a. Install"Small"Sample"Kit"mirror"(M3"SSK)"or"regular"M3"mirror"if"needed"
b. If"using"SSK"install"correct"lens"L2."
3. Make"sure"cover"for"the"DRA"is"fully"closed"and"turn"Cary"on."
4. Click"on"the"Cary"icon"on"the"taskbar"
5. Wait"20"minutes"for"the"lamps"to"warm"up."
6. Select"Setup"
a. set"wavelength"range,"%T"or"%R"
b. Under"Options"tab:"set"slit"width"and"height,"SBW"to"2"nm,"Detector"and"
Grating"change"wavelengths"
c. Under"Baseline"tab:"chose"zero/baseline"
d. Under"Storage"tab:"Set"your"filename"
7. Make"sure"that"all"exits"of"the"integrating"sphere"are"blocked"with"full"reflectors."
8. Baseline%and"run"100%"T"scan,"then"block"the"sample"beam"at"the"entrance"to"the" integrating"sphere"for"the"0%"T"scan.""If"using"an"aperture"holder"for"reflection" measurements,"please"see"the"“Aperture"kit"for"small"samples”"section"on"page"4."""
9. Mount"your"sample"
10. Run"Spectra"
11. When"done"remove"sample"and"remount"reflectors."
12. Turn"off"Cary"
2"
"
13. Shut"down"computer."
"
Introduction%and%Help%
For"a"complete"guide"to"the"Cary"hardware"and" open"the"Cary"“Scan”"program"and"click"on"the"
Help"tab."
For"the"basic"theory"of"diffuse"reflectance" measurements,"useful"diagrams"of"the"diffuse" reflectance"adaptor"(DRA),"alignment"tips"and" procedures,"and"descriptions"of"all"the"DRA" accessories"(VATH,"Small"Spot"Kit,"and"center" mount"holders)"go"to"Help">"Help"topics">" Accessories">"Cary"4000/5000/6000i">"Solid" Sample">"external"DRA""
To"learn"more"about"the"software"and"taking" measurements"in"the"Scan"menu,"go"to"Help">" Help"topics">"Software"Applications">"Scan"
Turning%the%instrument%on:%
Open"the"Cary"Scan"program"on"the"taskbar"or"Desktop"
The"DRA"must"to"be"fully"closed"in"order"for"the"instrument"to"initialize."Turn"on"the" instrument"(0/I"switch"on"the"lower"left"side"of"the"front"of"the"instrument).""In"the"Scan" program,"watch"the"lower"left"corner"to"see"the"status"of"the"instrument."When"the" traffic"light"is"green"(top"center"of"screen),"then"a"scan"can"be"run.""
Basics%configuration%and%alignment%
There"are"four"positions"in"which"the"sample"can"be"placed"for"use"with"the"integrating" sphere"(see"Figure"1"andFigure"2)."
1. The"entrance"slot"to"the"integrating" sphere"is"for"normal"incidence"diffuse" transmission"
2. The"rear"slot"is"for"normal"incidence" diffuse"reflection"
3. The"center"mount"holder"allows"for" variable"angle"diffuse"reflectance"and" transflectance"measurements"
4. In"addition,"the"VATH"(variable"angle" transmission"holder)"accessory"can"be" put"in"front"of"the"integrating"sphere"for" variable"angle"specular"transmission" measurements."If"you"would"like"to"use"
Figure&1&Integrating&sphere&entrance&slot.
!
Figure&2&Integrating&sphere&exit&slots.!
3"
"
the"VATH"accessory,"please"talk"to"one"of"the"GLAs"for"the"instrument"first."""
For"the"normal%or%large%sample%setup"there"are"three"positions"for"the"large"standard" M3"mirror"(see"Figure"3"andFigure"4)that"correspond"to"the"three"sample"positions" (Transmission,"Center,"and"Reflection).""Each"of"the"mirror"positions"will"focus"the"light" onto"the"position"of"the"sample.""The"three"different"mirror"positions"are"marked"T" (transmittance"slot),"C"(center"mount),"and"R"(rear"reflectance"slot)"in"Figure"3."Thus," the"mirror"needs"to"be"moved"to"the"right"position"depending"on"the"placement"of"your" sample.""The"cable"for"the"DRA"can"get"in"the"way"of"moving"this"mirror.""Turn%off%the% Cary%first,%if%you%need%to%unplug%the%DRA%to%move%the%mirror.""
For"the"small"sample"setup"(see"small"sample"discussion"below)"the"M3"mirror"is" replaced"by"the"small"sample"kit"mirror"(M3"SSK)"in"the"T"position.""Lens"(L2)"mounts"in" front"of"the"sample"(see"Figure"4).""There"are"three"lenses"that"can"be"used"again"for"the" three"positons"marked"as"T,"C"or"R.""
Each"time"the"sample"configuration"is"changed,"the"optics"need"to"be"realigned,"such" that"the"sample"and"reference"beams"are"not"clipped"at"the"entrance"to"the"integrating" sphere"and"that"the"sample"beam"is"centered"on"the"sample.""
1. With"the"top"door"to"the"DRA"closed,"under"the"commands"menu"click"on"Go%to," and"then"type"in"550"nm."When"you"darken"the"room"lights,"you"should"see"a" green"beam,"which"you"can"use"to"align"the"beam"onto"the"sample"(Alternatively," if"you"have"trouble"seeing"the"beam:"Block"the"sample"and"reference"beams"from" entering"the"integrating"sphere."Under"the"commands"menu"go"to"align"to" switch"the"gratings"to"zero"order"to"send"white"light"into"the"DRA.""The"white" light"should"not"enter"the"integrating"sphere)"""
2. Most"of"the"mirrors"have"two"knobs,"which"you"can"rotate"to"adjust"the"vertical" and"horizontal"positions"of"the"beam.""Start"with"mirror"M2"and"adjust"so"the"
Figure&3&Diffuse&Reflectance&Accessory&with &beam &p ath.&
4"
"
beam"hits"the"center"of"mirror"M3"then"use"M3"so"the"beam"hits"the"center"of"the" sample.""Some"of"the"mirrors"are"more"complicated"to"adjust"(although"these"can" often"be"left"alone).""For"help"with"alignment,"talk"to"one"of"the"GLAs"and"read" the"alignment"instructions,"which"can"be"found"under"Help."
Small%samples%
Most"typical"lab"samples"are"smaller"in"size"than"what"the"instrument"was"designed"to" measure.""For"all"three"types"of"measurements"(T,"C,"and"R),"it"is"best"to"under"fill"the" sample"area.""
Spot%size"-"There"are"two"ways"of" reducing"the"beam"spot"size:"
1)"Slit%width:"The"simplest"way"is"
to"reduce"beam"size"is"by"the"slit" width.""Click"on"the"Setup"the"
options"tab.""Under"slit"height,"
switch"to"Reduced.""It"takes"about"a" minute"for"the"slit"to"change."
2)"Small%spot%kit:"There"are"also"a"
series"of"lenses"that"can"be"put"in" the"optical"path"to"focus"the"beam" into"a"smaller"spot"on"the"sample." Again"there"are"three"lenses"with" different"focal"lengths"and"marked" T,"C,"and"R"for"the"three"positions"of" the"integrating"sphere.""Also,"the" large"mirror"(M3)"is"switched"out" for"a"one"with"a"different"curvature;" this"mirror"is"always"kept"in"the"T" position,"regardless"of"the"type"of" measurement"being"made."""
3)%Aperture%kit%for%reflection% measurements%
For"normal"incidence"reflection"measurements,"the"open"circle"in"the"rear"of"the" integrating"sphere"is"several"inches"in"diameter.""There"are"a"series"of"apertures," which"have"the"white"Spectralon"material"(same"roughened"Teflon"that"is"inside"the" integrating"sphere),"that"can"slide"onto"the"slot"on"the"back"of"the"integrating"sphere."" The"spring-loaded"mount"that"holds"the"sample"against"the"integrating"sphere"first" needs"to"be"taken"off.""Slide"on"the"appropriate"aperture"and"then"either"tape"your" sample"onto"the"back"of"the"aperture,"or"put"back"on"the"spring-loaded"sample" mount."
If"the"beam"spot"overfills"the"aperture,"you"can"correct"for"this"“stray"light”"by" performing"a"Zero/baseline"Correction.""Normally,"the"0%TBaseline"scan"is"run"by" blocking"the"beam"before"it"enters"the"integrating"sphere,"which"accounts"for"
Figure&4&Photo&of&diffuse&reflectance&accessory&with&
beam&path.&
5"
"
electronic"noise.""In"this"case,"do"not"block"the"beam"so"that"the"0%TBaseline" accounts"for"the"light"reflecting"on"the"edges"of"the"aperture.""
Performing!a!Zero/Baseline!Correction:"Select"this"option"to"apply"a"100%T" baseline"correction"(100%TBaseline)"and"a"zero-line"correction"(0%TBaseline)"to" the"sample"scan"(S)."This"correction"is"performed"with"the"raw"transmission"data" and"is"thus:"
𝑆 0%𝑇𝐵𝑎𝑠𝑒𝑙𝑖𝑛𝑒
100%𝑇𝐵𝑎𝑠𝑒𝑙𝑖𝑛 0%𝑇𝐵𝑎𝑠𝑒𝑙𝑖𝑛𝑒
"
With"this"option,"the"Cary"will"prompt"you"to"perform"a"100%TBaseline"first," followed"by"a"0%TBaseline"when"you"perform"a"baseline"collection."
Basics%of%measurement%acquisition%
The"wavelength"range"is"1800"nm"to"250"nm."""Scans"normally"go"from"long"to"short" wavelengths.""There"are"two"detectors,"a"PMT"and"an"InGaAs.""The"wavelength"at"which" the"system"changes"from"one"to"the"other"(accompanied"by"a"grating"change)"can"be" adjusted.""Usually"750"to"800"nm"is"a"good"value"for"this"change"over.""You"can"see"the" change"in"the"spectrum"when"you"take"a"baseline"(%"Transmission"or"Reflection).""If" you"see"a"step"in"the"spectrum,"it"may"mean"your"baseline"is"bad"or"that"your"sample" has"a"polarization"dependency.""
Setup%–"The"setup"button"located"in"the"upper"left-hand"corner"defines"the"scan"
parameters.""The"following"useful"tabs"can"be"found"under"Setup.""
Cary!tab!!
Here"you"can"set"the"units"of"the"x-and"y"axes"(nm,"cm-1,"etc."and"A,"%T,"%R,"etc.)," as"well"as"the"range"for"the"axes.""Under"scan"controls,"the"average"time"per"data" point"and"the"data"interval"can"be"set"can"be"controlled.""Note"that"you"only"have" independent"control"over"two"of"these"values."""
Options!tab!under!Setup!
Here"you"control"the"Beam"Mode,"Slit"Band"Width,"and"Slit"Height"(see&Appendix& A&of&this&guide&for&more&info).""The"Double"beam"mode"needs"to"be"used"for"the"
integrating"sphere"measurements"and"the"SBW"is"typically"set"to"2nm.""
As"mentioned"earlier,"the"Slit"Height"can"be"used"to"decrease"the"beam"spot"size""
Typically,"the"UV-Vis"button"will"be"selected.""If"you"do"not"need"to"go"to"shorter" wavelengths"than"400"nm,"you"can"select"the"Vis"button,"which"will"save"on"the" lifetime"of"the"UV"lamp.""""
You"can"also"set"the"wavelength"at"which"the"Detector"and"grating"change.""Both" of"these"are"typically"set"at"800"nm,"if"you"have"an"important"spectroscopic" feature"at"800"nm,"you"can"move"the"change"over"+/-"50"nm"
Baseline!tab!!
This"tab"enables"you"to"choose"the"type"of"baseline(s)"that"you"want"to"use"in"the" Scan"run."The"different"types"of"baseline"corrections"are"particularly"useful"with"
6"
"
some"of"the"Varian"accessories."You"can"collect"a"baseline"immediately"before" scanning"a"sample"or"use"a"stored"baseline."
The"accuracy"of"your"baseline"determines"the"accuracy"of"your"measurement."" You"should"perform"a"zero/baseline"correction"before"running"your"samples" after"the"instrument"has"warmed"up"for"~20"minutes"using"the"same"scan" parameters"that"you"will"use"for"your"sample.""
The"Cary"WinUV"system"collects"a"new"baseline"when"you"select"any"type"of" baseline"correction"in"the"Baseline"tab"of"the"Setup"dialog"box"and"then"click"
Baseline"button"in"the"application"window."Depending"on"the"type"of"correction"
you"have"selected,"the"Cary"will"prompt"you"to"perform"either"one"or"two" baseline"scans."The"first"is"a"100%Tbaseline"(for"the"DRA"with"the"100%" reference"material"in"the"sample"positions)."The"second,"if"required,"is"a" 0%Tbaseline"(with"the"sample"beam"blocked)."
Once"the"baselines"have"been"collected,"they"will"be"automatically"applied"to" each"collected"sample"data"point"as"it"is"displayed"(indicated"by"the"red"display)." If"there"is"not"a"corresponding"data"point"in"the"current"baseline,"then"the"file" will"be"interpolated"to"provide"the"correction."The"correction"applied"to"each" point"is"dependent"on"the"selected"baseline"type"in"the"Baseline"tab.""
A"baseline"cannot"be"applied"to"a"sample"where"the"abscissa"range"of"the"sample" exceeds"that"of"the"baseline."If"you"change"your"abscissa"range"to"one"outside" that"of"the"baseline,"the"Cary"will"display"an"error"message."Also"the"red"text"in" the"ordinate"display"which"indicates"that"baseline"correction"mode"is"activated" will"disappear."
For"reflection"measurements"using"the"aperture"kit,"if"the"beam"spot"overfills"the" aperture,"you"can"correct"for"this"“stray"light”"by"performing"a"Zero/baseline" Correction.""Normally,"the"0%TBaseline"scan"is"run"by"blocking"the"beam"before" it"enters"the"integrating"sphere,"which"accounts"for"electronic"noise."In"this"case," do"not"block"the"beam,"but"instead"remove"the"reflectance"port"cap"to"account" for"the"light"reflecting"on"the"edges"of"the"aperture"(the"remainder"of"the"light" should"pass"through"the"integrating"sphere"and"not"be"detected)."""
Performing&a&Zero/Baseline&Correction:"Select"this"option"to"apply"a"100%T" baseline"correction"(100%TBaseline)"and"a"zero-line"correction"(0%TBaseline)" to"the"sample"scan"(S)."This"correction"is"performed"with"the"raw"transmission" data"and"is"thus:"
(𝑆 − (0%𝑇𝐵𝑎𝑠𝑒𝑙𝑖𝑛𝑒))
(100%𝑇𝐵𝑎𝑠𝑒𝑙𝑖𝑛𝑒 0%𝑇𝐵𝑎𝑠𝑒𝑙𝑖𝑛𝑒)
"
With"this"option,"the"Cary"will"prompt"you"to"perform"a"100%TBaseline"first," followed"by"a"0%TBaseline"when"you"perform"a"baseline"collection."
This"baseline"option"corrects"for"any"inherent"variations"in"the"electronic"zero" line"of"the"instrument."You"should"use"this"baseline"option"if"your"samples"have" areas"of"very"low"transmission"(high"absorbance)"as"any"variations"in"the"
Loading...
+ 13 hidden pages