5"
"
electronic"noise.""In"this"case,"do"not"block"the"beam"so"that"the"0%TBaseline"
accounts"for"the"light"reflecting"on"the"edges"of"the"aperture.""
Performing!a!Zero/Baseline!Correction:"Select"this"option"to"apply"a"100%T"
baseline"correction"(100%TBaseline)"and"a"zero-line"correction"(0%TBaseline)"to"
the"sample"scan"(S)."This"correction"is"performed"with"the"raw"transmission"data"
and"is"thus:"
𝑆 − 0%𝑇𝐵𝑎𝑠𝑒𝑙𝑖𝑛𝑒
100%𝑇𝐵𝑎𝑠𝑒𝑙𝑖𝑛 − 0%𝑇𝐵𝑎𝑠𝑒𝑙𝑖𝑛𝑒
"
With"this"option,"the"Cary"will"prompt"you"to"perform"a"100%TBaseline"first,"
followed"by"a"0%TBaseline"when"you"perform"a"baseline"collection."
Basics%of%measurement%acquisition%
The"wavelength"range"is"1800"nm"to"250"nm."""Scans"normally"go"from"long"to"short"
wavelengths.""There"are"two"detectors,"a"PMT"and"an"InGaAs.""The"wavelength"at"which"
the"system"changes"from"one"to"the"other"(accompanied"by"a"grating"change)"can"be"
adjusted.""Usually"750"to"800"nm"is"a"good"value"for"this"change"over.""You"can"see"the"
change"in"the"spectrum"when"you"take"a"baseline"(%"Transmission"or"Reflection).""If"
you"see"a"step"in"the"spectrum,"it"may"mean"your"baseline"is"bad"or"that"your"sample"
has"a"polarization"dependency.""
Setup%–"The"setup"button"located"in"the"upper"left-hand"corner"defines"the"scan"
parameters.""The"following"useful"tabs"can"be"found"under"Setup.""
Cary!tab!!
Here"you"can"set"the"units"of"the"x-and"y"axes"(nm,"cm-1,"etc."and"A,"%T,"%R,"etc.),"
as"well"as"the"range"for"the"axes.""Under"scan"controls,"the"average"time"per"data"
point"and"the"data"interval"can"be"set"can"be"controlled.""Note"that"you"only"have"
independent"control"over"two"of"these"values."""
Options!tab!under!Setup!
Here"you"control"the"Beam"Mode,"Slit"Band"Width,"and"Slit"Height"(see&Appendix&
A&of&this&guide&for&more&info).""The"Double"beam"mode"needs"to"be"used"for"the"
integrating"sphere"measurements"and"the"SBW"is"typically"set"to"2nm.""
As"mentioned"earlier,"the"Slit"Height"can"be"used"to"decrease"the"beam"spot"size""
Typically,"the"UV-Vis"button"will"be"selected.""If"you"do"not"need"to"go"to"shorter"
wavelengths"than"400"nm,"you"can"select"the"Vis"button,"which"will"save"on"the"
lifetime"of"the"UV"lamp.""""
You"can"also"set"the"wavelength"at"which"the"Detector"and"grating"change.""Both"
of"these"are"typically"set"at"800"nm,"if"you"have"an"important"spectroscopic"
feature"at"800"nm,"you"can"move"the"change"over"+/-"50"nm"
Baseline!tab!!
This"tab"enables"you"to"choose"the"type"of"baseline(s)"that"you"want"to"use"in"the"
Scan"run."The"different"types"of"baseline"corrections"are"particularly"useful"with"