Varian 900-LC Series Getting Started Manual

Page 1
Varian 900-LC Series Liquid Chromatograph
Getting Started Manual
Installation Category II Pollution Degree 2 Safety Class 1 (EN 61010-1)
8510243600 December 2007
Page 2
Varian 900-LC Series Liquid Chromatograph
Varian, Inc. offices
Varian, Inc. has offices in most countries. The major offices for liquid chromatography products are listed below:
Varian Australia Pty Ltd (Manufacturing site) 679 Springvale Road Mulgrave, Victoria 3170 Australia International telephone: + 61 3 9560 7133 International fax: + 61 3 9560 7950
Varian Instruments 2700 Mitchell Dr Walnut Creek, CA 94598 Telephone: 1 800 926 3000 International telephone: +1.925.939.2400 International fax: +1.925.945.2102
Varian BV Herculesweg 8 4338 PL, Middelburg, Netherlands International telephone: +31 118 67 1000 International fax: +31 118 62 3193
Internet
The Varian, Inc. Internet home page can be found at: www.varianinc.com
Varian Australia Pty Ltd is the owner of copyright on this document and any associated software. Under law, the written permission of Varian Australia Pty Ltd must be obtained before either the documentation or the software is copied, reproduced, translated or converted to electronic or other machine-readable form, in whole, or in part.
First published December 2007 in Australia. Comments about this manual should be directed to the Marketing Communications Manager, Varian Australia at the address above or by e-mail to: webmaster@osi.varianinc.com
Varian Australia is ISO9001:2000 certified.
©2007 Varian Australia Pty Ltd (ABN 67 004 559 540) Galaxie, Varian and the Varian logo are trademarks or registered trademarks of Varian, in the U.S. and other countries.
All rights reserved.
2 Publication date: December 2007
Page 3
Varian 900-LC Series Liquid Chromatograph
Declaration of Conformity
We hereby declare that the equipment listed below complies with the requirements of:
The Low Voltage Directive 73/23/EEC (93/68/EEC)
The EMC Directive 89/336/EEC (92/31/EEC and 93/68/EEC)
Applicable Standards LVD BS EN 61010-1:1993 EMC BS EN 61326:2003 EN 55011:1999 EN 61000-4-2:1995 EN 61000-4-3:2003 EN 61000-4-4:1995 EN 61000-4-5:1995 EN 61000-4-6:1996
Equipment Model Number Varian Model Instrument Type
Authorized Representative in the EU
Name: G. A. Wassink Company Name Varian BV
Address Herculesweg 8
4330 EA Middelburg
Signed:
The Netherlands Position: Quality Manager Telephone +31 (0) 118 671 000 Date: 1 Month 20XX Facsimile +31 (0) 118 633 118
Manufacturer
Name: Phil Binns Company Name Varian Australia Pty Ltd
Address 679 Springvale Road
Mulgrave VIC 3170 Signed:
Australia Position: Director of Operations Telephone +61 (0) 3 9560 7133
Date: 1 Month 20XX Facsimile +61 (0) 3 9560 7950
Publication number 8510243600 Month 20XX
Publication date: December 2007 3
Page 4
Varian 900-LC Series Liquid Chromatograph
This page is intentionally left blank.
4 Publication date: December 2007
Page 5
Varian 900-LC Series Liquid Chromatograph
Contents
Safety Practices and Hazards................................................................9
General 9 Electrical Hazards 10 Compressed Gas Cylinders 11 High Pressure Hazards 12 Ultraviolet Radiation 13 Flash Chromatography 13 Solvent Hazards 13
Ignition of Flammable Chemicals 14
Inflammation or Injury due to Toxic, Corrosive or Stimulative
Solvent 14
Explosion of Vapor from Flammable Chemicals 14 Other Precautions 15 Warning and Caution Messages 15 Warning Symbols 16 Color Coding 17 US FCC Advisory Statement 17 Spurious and Harmonic Radiation Compliance 18 CE Compliance 18
1. Introduction ................................................................................ 19
1.1 Pre-installation Requirements 19
1.2 User Documentation 19
1.2.1 Conventions 19
1.2.2 Notes and Hot Tips 20
1.3 Specifications 20
1.3.1 Environmental Conditions 20
Publication date: December 2007 5
Page 6
Varian 900-LC Series Liquid Chromatograph
1.3.2 Electrical Power Supplies 21
1.3.3 Gas Supplies 23
1.3.4 Waste Container 24
1.3.5 Laboratory Conditions 24
1.3.6 Instrument Cooling Air Supply 24
1.3.7 Weights and Dimensions 25
1.4 Personal Computer Requirements 26
2. Installation .................................................................................. 29
2.1 900-LC Series LC System to PC Communications 29
2.2 Installing the Galaxie™ Chromatography Data System 32
2.3 About the Varian Integrated HPLC Online Help 40
2.4 Hardware Installation 41
2.5 Installation Qualification Tests 41
2.6 Customer Training 41
2.7 Using the Online Help 42
2.8 Instrument Maintenance Software (IMS) 42
2.8.1 How to launch IMS 42
2.8.2 About IMS 43
3. Operation..................................................................................... 45
3.1 Analysis Checklist 45
3.2 Turning on the Instrument and Software 45
3.2.1 Connecting to a System 46
3.3 Preparing the System for Use 46
3.4 Configuring the Device Settings and System Parameters 47
3.4.1 Flows 48
3.4.2 Detectors 48
3.4.3 Column 48
3.4.4 Sample Introduction 49
3.4.5 System Parameters 49
3.5 Creating a Method 50
6 Publication date: December 2007
Page 7
Varian 900-LC Series Liquid Chromatograph
Starting an Acquisition 52
3.6
3.6.1 Sequence 52
3.6.2 Quick Start 52
3.7 Printing a report 53
3.8 Saving Your Files 54
4. Maintenance................................................................................ 55
4.1 Routine Maintenance 56
4.2 Cleaning 57
4.3 Spare Parts 57
5. Troubleshooting .......................................................................... 59
Publication date: December 2007 7
Page 8
Varian 900-LC Series Liquid Chromatograph
This page is intentionally left blank.
8 Publication date: December 2007
Page 9
Varian 900-LC Series Liquid Chromatograph
Safety Practices and Hazards
Your Varian instrument and accessories have been carefully designed so that when used properly you have an accurate, fast, flexible and safe LC system.
Information about safety practices appears throughout the documentation (both hard copy and online) provided with your instrument. Before using the instrument, you must thoroughly read these safety practices.
Observe all relevant safety practices at all times.
This instrument is designed for chromatographic analysis of appropriately prepared samples. It must be operated using appropriate solvents and within specified maximum ranges for pressure, flows, and temperatures as described in this manual. If the equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired.
It is your responsibility to inform Varian representatives if the instrument has been used for the analysis of hazardous biological, radioactive, or toxic samples, prior to any instrument service being performed or when an instrument is being returned to the service center for repair, or is being recycled at end-of-life.

General

Operation of a Varian 900-LC Series Liquid Chromatograph involves the use of pressurized liquid, ultraviolet radiation, and hazardous materials including corrosive fluids and flammable liquids. Careless, improper or unskilled use of this liquid chromatograph can cause death or serious injury to personnel, and/or severe damage to equipment and property.
The 900-LC Series Liquid Chromatograph incorporates covers that are designed to prevent inadvertent contact with any potential hazards. If the instrument is used in any manner not specified by
Publication date: December 2007 9
Page 10
Varian 900-LC Series Liquid Chromatograph
Varian, this protection may be impaired. It is essential that no cover is bypassed, damaged or removed.
The safety practices described below are provided to help the user operate the instrument safely. Read each safety topic thoroughly before attempting to operate the instrument and ALWAYS operate the liquid chromatograph in accordance with these safety practices.
Follow these safety practices to ensure safe equipment operation.
Perform periodic leak checks on all supply lines and pneumatic plumbing.
Do not allow liquid lines to become kinked or punctured. Place lines away from foot traffic and extreme heat or cold.
Store organic solvents in fireproof, vented and clearly labeled cabinets so they are easily identified as toxic and/or flammable materials.
Keep the waste vessel in view and empty as required.
Do not accumulate waste solvents. Dispose of such materials
through a regulated disposal program and not through municipal sewage lines.
This instrument has been tested per applicable requirements of EMC Directive as required to carry the European Union CE Mark. This equipment may be susceptible to radiation/interference levels or frequencies, which are not within the tested limits.

Electrical Hazards

Exposure to high voltages and UV energy can cause severe skin damage and cataracts of the eyes, while close contact with the electrical components can result in severe heat burns to the skin, and an electrical discharge which may cause death, severe electric shock or sub-surface skin burns.
The 900-LC Series Liquid Chromatograph has been carefully designed to operate safely and effectively when using components that conform to Varian’s design criteria. Use of non-approved components in the LC system may render the system inoperative and/or hazardous. It may also invalidate the warranty on the instrument. Use only related components supplied or authorized by Varian.
Disconnect the instrument from all power sources before
10 Publication date: December 2007
Page 11
Varian 900-LC Series Liquid Chromatograph
removing protective panels to avoid exposure to potentially dangerous voltages. Panels or covers which are retained by screws on the LC system may be opened ONLY by Varian­trained, Varian-qualified, or Varian-approved Representatives. Consult the manuals or product labels supplied with your PC to determine which parts are operator-accessible.
When it is necessary to use a non-original power cord plug, make sure the replacement cord adheres to the color coding and polarity described in the manual and all local building safety codes.
Good grounding/earthing is essential to avoid a potentially serious electric shock hazard. Consult the manuals or product labels supplied with your PC for the relevant grounding requirements.
Replace blown fuses with fuses of the size and rating shown on the fuse panel or in the manual.
Replace faulty or frayed power cords immediately with the same type and rating.
Make sure that voltage sources and line voltage match the value for which the instrument is wired.
Avoid using power supplies from a source that may be subject to electrical or RF interface from other services (for example, large electrical motors, elevators and welders).

Compressed Gas Cylinders

Unless your system is equipped with a helium sparging unit or an ELS detector, there are no gas requirements for the 900-LC Series LC systems. For gas requirements on these devices, consult the Operation Manual or other documentation enclosed with the particular device. Compressed gas cylinders contain highly pressurized gas. If storage conditions are outside of the recommended suppliers safety codes the cylinders can explode or rapidly release gas into the environment. This may result in injury or death.
Store and handle compressed gases carefully and in strict adherence to safety codes.
Secure cylinders to an immovable structure or wall.
Store and move cylinders in an upright, vertical position.
Before transport, remove regulators and install cylinder cap.
Publication date: December 2007 11
Page 12
Varian 900-LC Series Liquid Chromatograph
Store cylinders in a well ventilated area away from heat, direct sunshine, freezing temperatures, and ignition sources.
Mark cylinders clearly so there is no doubt as to their contents.
Use only approved regulators and connections.
Use only connector tubing that is chromatographically clean
(Varian part number 0391832600) and has a pressure rating significantly greater than the highest outlet pressure from the regulator.

Storage Cylinder Instructions

Cylinders containing gas under pressure should be firmly secured to a rigid structure, and the storage area must be adequately ventilated.
Never locate gas cylinders near a source of ignition, or in a position that is subject to direct heat. Gas storage cylinders often incorporate a pressure release device which will discharge the gas at a predetermined temperature, usually around 52 °C (125 °F).
If gases are to be plumbed from a remote storage area to the instrument site, ensure that the local outlets are fitted with stop valves, pressure gauges and suitable regulators which are easily accessible to the instrument operator. The gas outlets must be provided within 1.5 metres (5 feet) of the instrument

High Pressure Hazards

If a line ruptures, a relief device opens, or a valve opens accidentally under pressure, potentially hazardous high liquid pressures can be generated by the pump causing a high velocity stream of volatile and/or toxic liquids.
Wear personal protective equipment when you inject samples or perform routine maintenance.
Never open a solvent line or valve under pressure. Stop the pump first and let the pressure drop to zero.
Read and adhere to all NOTES, CAUTIONS, and WARNINGS in the manual.
Always keep the doors and covers closed during operation.
12 Publication date: December 2007
Page 13

Ultraviolet Radiation

900-LC Series Liquid Chromatograph systems that use an ultraviolet light source have shielding to prevent radiation exposure to personnel.
For continued protection:
Ensure that any protective lamp covers are in place during operation.
Do not look directly into detector fluid cells or at the UV light source. When inspecting the light source or fluid cell, always use protective eye covering such as borosilicate glass or polystyrene.
Ozone can be generated by radiation from source lamps. Exposure to ozone can result in severe irritation to the skin, eyes, and upper respiratory system. The maximum permissible exposure level is 0.1 ppm (0.2 mg/m ventilate the area surrounding the LC system such that the concentration of ozone does not exceed the maximum permissible level. All venting must be to outside air, never within the building.
Varian 900-LC Series Liquid Chromatograph
3
). ALWAYS

Flash Chromatography

The operator should be familiar with the physico-chemical properties of the components of the mobile phase.
Only use solvents compatible with the LC system tubing and fittings as certain solvents may cause weakening and leaks with possible bursting.
Employ static measuring and static discharge devices to safeguard against the buildup of static electricity.

Solvent Hazards

All solvents can create a hazard if they leak into the atmosphere. Even small leaks in solvent supply systems can be dangerous. Any leak (except that of air) can result in an oxygen-deficient atmosphere, which can cause asphyxiation. The area in which solvents are stored and the area surrounding the instrument must be adequately ventilated to prevent such accumulations.
Publication date: December 2007 13
Page 14
Varian 900-LC Series Liquid Chromatograph

Ignition of Flammable Chemicals

This instrument is not explosion-proof. In unattended operation, do not use organic solvents having an ignition point below 70 °C.
Beware of ignition hazard when using flammable chemicals such as organic solvents.
Do not bring a heat or flame source near the instrument.
Well-ventilate the laboratory room where the instrument is used.
Always check the following conditions. If an abnormality is found stop operation immediately.
Leakage of solvent or waste solution.
Leakage of solvent inside the instrument.
When using flammable chemicals, be careful about possible ignition due to static electricity. To prevent the build-up of static electricity, use a conductive container for waste.

Inflammation or Injury due to Toxic, Corrosive or Stimulative Solvent

When using a toxic, corrosive or stimulative solvent, be careful not to incur a physical inflammation or injury. For details of the properties of each solvent and how to handle it, refer to the relevant Material Safety Data Sheets (MSDS). Be sure to handle each solvent properly.
Wear proper protective clothes (e.g., safety goggles) so that a solvent will not come into direct contact with the skin.
Ventilate the laboratory room adequately to prevent accidental inhalation of harmful solvent vapor.

Explosion of Vapor from Flammable Chemicals

If a flammable chemical such as organic solvent leaks from the flow path of the instrument and its vapor concentration reaches the explosion limit, it could cause spontaneous combustion with dangerously explosive results.
When using a flammable and readily volatile chemical, be sure to check for leakage from the instrument flow path and ventilate the laboratory room adequately.
Use only approved regulator and hose connectors (refer to the
14 Publication date: December 2007
Page 15
supplier’s instructions). Keep solvents cool and properly labeled. Ensure that you have the correct solvent before connecting it to the instrument.

Other Precautions

Use of the 900-LC Series Liquid Chromatograph and accessories may involve materials, solvents and solutions, which are flammable, corrosive, toxic or otherwise hazardous. Careless, improper or unskilled use of such materials, solvents and solutions can create explosion hazards, fire hazards, toxicity and other hazards that can result in death, serious personal injury or damage to equipment.
Always ensure that laboratory safety practices governing the use, handling and disposal of hazardous materials are strictly observed. These safety practices should include wearing appropriate safety clothing and safety glasses.
Airflow to the cooling fans of the liquid chromatograph must be unobstructed. Do not block the ventilation grills on the liquid chromatograph and accessories. Consult the manuals supplied with your PC, monitor and for their specific ventilation requirements.
Varian 900-LC Series Liquid Chromatograph
Great care should be taken when working with glass or quartz parts to prevent breakage and cuts.
The 900-LC Series Liquid Chromatograph weighs approximately 140 kg (309 lb). To avoid injury to personnel or damage to the instrument or property, always use appropriate lifting procedures to move the instrument.
Use only Varian-supplied spares with your instrument.

Warning and Caution Messages

A Warning message is used in the text when failure to observe instructions or precautions could result in death or injury. The list of symbols that appear in conjunction with warnings are detailed in the next section.
A Caution message is used when failure to observe instructions could result in damage to equipment (Varian supplied and/or other associated equipment).
Publication date: December 2007 15
Page 16
Varian 900-LC Series Liquid Chromatograph

Warning Symbols

The following is a list of symbols that appear in conjunction with warnings in this manual or on the 900-LC Series Liquid Chromatograph. The hazard they describe is also shown.
A triangular symbol indicates a warning. The meanings of the symbols that may appear alongside warnings in the documentation or on the instrument itself are as follows:
Corrosive liquid
Eye hazard
Heavy weight
(danger to hands)
Radiation source
The following symbol may be used on warning labels attached to the instrument. When you see this symbol, refer to the relevant operation or service manual for the correct procedure referred to by that warning label.
Electrical shock
Fire hazard
Hot surface
Explosion hazard
Heavy weight
(danger to feet)
Moving parts
The following symbols appear on the instrument for your information.
16 Publication date: December 2007
I
Mains power on
Page 17
Varian 900-LC Series Liquid Chromatograph

Color Coding

The various indicator lights appearing on Varian instruments and associated accessories are color coded to represent the status of the instrument or accessory.
A green light indicates the instrument is in normal/standby
0
mode.
Mains power off
Fuse
Single phase alternating current
Direct current
When attached to the rear of the instrument, indicates that the product complies with the requirements of one or more EU directives.
When attached to the rear of the product, indicates that the product has been certified (evaluated) to CSA 1010.1 and UL 3101-1.
An orange light indicates that a potential hazard is present.
A blue light indicates that operator intervention is required.
A red light warns of danger or an emergency.

US FCC Advisory Statement

This equipment generates, uses and can radiate radio frequency energy, and if not installed and operated in accordance with the instruction manual may cause interference to radio communications. It has been tested and found to comply with the limits for a Class A computing device pursuant to Subpart J of Part 15 of Federal Communications Commission (FCC) Rules, which are designed to provide reasonable protection against such interference when operated in a commercial environment. Operation of this equipment in a residential area may cause interference, in which case the user will be required to take whatever measures may be necessary to correct the interference at his or her expense.
Publication date: December 2007 17
Page 18
Varian 900-LC Series Liquid Chromatograph

Spurious and Harmonic Radiation Compliance

Your Varian 900-LC Series Liquid Chromatograph complies with the requirements of FCC rules and Regulations part 18, subpart H—Miscellaneous Equipment, at the date of manufacture.

CE Compliance

Your Varian 900-LC Series Liquid Chromatograph has been designed to comply with the requirements of the Electromagnetic Compatibility (EMC) Directive and the Low Voltage (electrical safety) Directive (commonly referred to as the LVD) of the European Union. Varian has confirmed that each product complies with the relevant Directives by testing a prototype against the prescribed EN (European Norm) standards.
Proof that a product complies with these directives is indicated by:
The CE Marking appearing on the rear of the product, and
The documentation package that accompanies the product
containing a copy of the Declaration of Conformity. The Declaration of Conformity is the legal declaration by Varian that the product complies with the directives listed above, and shows the EN standards to which the product was tested to demonstrate compliance. It is also signed by Varian’s Authorized Representative in the EU, and by the representative of the manufacturing plant.
18 Publication date: December 2007
Page 19
Varian 900-LC Series Liquid Chromatograph

1. Introduction

1.1 Pre-installation Requirements

Prior to receiving your instrument, you will have been provided with a 900-LC Series Liquid Chromatograph Pre-installation Manual (publication number 8510243900), which describes the environmental and operating requirements of the LC system. You must prepare your laboratory according to these instructions before the LC system can be installed. You should keep the pre­installation manual for future reference. If you have misplaced your copy, you can obtain a free replacement from your local Varian office. Alternatively, download a PDF from the Varian, Inc. Web site, www.varianinc.com

1.2 User Documentation

You have been provided with the following documentation to help you set up and operate your Varian Integrated Liquid Chromatograph:
This Getting Started manual, with safety practices and hazards
information, where to find information about installing and maintaining instrument components and a brief operation overview.
An extensive online Help system containing context-sensitive
help, step-by-step instructions for frequently performed operations and instructions for using any accessories you ordered.

1.2.1 Conventions

The following conventions have been used throughout the documentation:
Publication date: December 2007 19
Page 20
Varian 900-LC Series Liquid Chromatograph
Menus, menu options and field names (e.g. select Copy from the Edit menu) have been typed in bold. Bold is also used to signify the pushbuttons appearing throughout the software (e.g. click OK).
ALL CAPITALS indicate keyboard and mouse commands (e.g. press the F2 key) and text you must type in from the keyboard (e.g. type SETUP at the prompt).

1.2.2 Notes and Hot Tips

A Note is used to give advice or information. A Note is denoted by the following symbol:
A Hot Tip is used to give practical hints to help you achieve the best possible performance from your instrument. A Hot Tip is denoted by the following symbol: ✰✰✰

1.3 Specifications

The instrument is suitable for indoor use only and is classified
suitable under the following categories (EN 61010-1):
Installation category II Pollution degree 2 Safety class 1

1.3.1 Environmental Conditions

Condition Altitude Temp
(°C)
Non-operating (Transport)
Non-operating & meeting dielectric strength tests
Operating within specifications
Table 1. Environmental conditions
20 Publication date: December 2007
0–3 050 m (0–10 000 ft)
Sea level 40 90–95
0–2 000 m (0–6 560 ft)
5–45 20–80
10–35 8–80
Humidity (%RH) non-condensing
Page 21
Varian 900-LC Series Liquid Chromatograph
Note For optimum analytical performance, it is recommended that the
ambient temperature of the laboratory be between 20 and 25 °C and be held constant to within ±2 °C throughout the entire working day.

1.3.2 Electrical Power Supplies

All power supplies should be single phase AC, 3 wire system (active, neutral, ground or two active and ground) and should be terminated at an appropriate connection receptacle that is within reach of the system power cable.
1.3.2.1
900-LC Series HPLC System Power Connections
Power supply phase
Power supply, current rating and overload protection
Australia
USA
Canada
Europe
Plug supplied Required wall socket
Complies with AS3112 To comply with AS3112
Complies with Nema 5-15P To comply with Nema 5-15R
Complies with Nema 5-15P
Complies with CS 22.2 No. 42
Complies with CEE 7 Sheet VII
Single
Between 3 and 5 A (for 220-240 V countries) Between 6 and 12 A (for 110-120 V countries)
HPM 787 or Clipsal 2015 (250 V, 10 A)
Hubbell 5262 (125 V, 15 A)
To comply with Nema 5-15R
Hubbell 5262 (125 V, 15 A)
To comply with CEE 7 Sheet VII (250 V, 10/16 A)
Table 2. Power connections
1.3.2.2
Publication date: December 2007 21
Fuses
1FS1 & 1FS2
T10 A H250 V, 3AB SLO-BLO (for 220-240 V countries)
Page 22
Varian 900-LC Series Liquid Chromatograph
T20 A H250 V, 3AB SLO-BLO (for 100-120 V countries)
FS3
Table 3. Fuses
Note For safety reasons, any other internal fuse or circuit breaker is not
operator accessible, and should only be replaced by Varian-authorized personnel.
Fuse information on the rear of the instrument is the most up to date.
1.3.2.3
Electrical Requirements
The installation of electrical power supplies must comply with the rules and/or regulations imposed by the local authorities responsible for the use of electrical energy in the workplace.
All power supplies for the integrated liquid chromatograph and accessories should be single phase, AC, 3 wire systems (active, neutral, ground or two active and ground). Each connection should be terminated at an appropriate receptacle within reach of each assembly’s power cable. Use of power boards or extension cables
is not recommended.
It is recommended that separate mains circuits, (individually protected by fuses or circuit breakers) are used for each component in the system such as the integrated liquid chromatograph, PC and printer.
T2.5 A H250 V, 5x20 mm (for both 100-120 V and 220-240 V countries)
Avoid using power supplies from a source that may be subject to electrical interference from other services (such as large electric motors, elevators, welders and air conditioning units).
The Varian 900-LC Series Liquid Chromatographs are supplied with a 2 meter long (6 ft 6 in), mains supply cable.
System unit Required supply voltage Nominal rating
920-LC Analytical Liquid Chromatograph 100-120, 220-240 V AC, 50-60 Hz 1000 VA
940-LC Semi-Preparative Liquid Chromatograph
940-LC Preparative Liquid Chromatograph 100-120, 220-240 V AC, 50-60 Hz 1200 VA
22 Publication date: December 2007
100-120, 220-240 V AC, 50-60 Hz 800 VA
Page 23
Varian 900-LC Series Liquid Chromatograph
Fraction Collector
ELSD Detector 115-230 V AC, 50-60 Hz 480 VA
External RI Detector 115-230 V AC, 50-60 Hz 150 VA
Ethernet Switch 115-230 V AC, 50-60 Hz 7 VA
Personal computer (Typical) 115, 120, 220, 240 VAC 300 VA
Printer (Typical) 115, 120, 220, 240 VAC 85 VA
Table 4. Power requirements
115-230 V AC, 50-60 Hz 264 VA

1.3.3 Gas Supplies

The installation of compressed or liquid gas supplies must comply with the rules and/or regulations imposed by the local authorities responsible for such use in the workplace.
The main gas supply requirement is nitrogen for supply to ELS Detector. Nitrogen from a cylinder or nitrogen generator may be used with Varian 900-LC Series Liquid Chromatograph ELS Detector. Helium may also be required to if you are using a helium sparging system to degas solvents. Consult the documentation that came with your helium sparging system for gas requirements.
Nitrogen Gas
Purity
Permissible pressure range
Maximum pressure
Required flow rates
1.3.3.1
Publication date: December 2007 23
Storage Cylinder Instructions
Cylinders containing gas under pressure should be firmly secured to a rigid structure, and the storage area must be adequately ventilated.
Never locate gas cylinders near a source of ignition, or in a position that is subject to direct heat. Gas storage cylinders often incorporate a pressure release device which will discharge the gas at a predetermined temperature, usually around 52 °C (125 °F).
>98% filtered to 0.2 μm
60 to 100 psi (4–6.7 bar)
100 psi (6.7 bar) regulated
Up to 3.25 L/min at 60 psi at 25 °C
Page 24
Varian 900-LC Series Liquid Chromatograph
If gases are to be plumbed from a remote storage area to the instrument site, ensure that the local outlets are fitted with stop valves, pressure gauges and suitable regulators which are easily accessible to the instrument operator. The gas outlets must be provided within 1.5 metres (5 feet) of the instrument.

1.3.4 Waste Container

The Varian 900-LC Series Liquid Chromatograph needs a drain vessel for disposal of fluids and vapors from the system. Suitable tubing is supplied with the LC system for use with most HPLC solvents. Always check that your LC system tubing is compatible with your solvents.
A chemically inert container, not glass or of a narrow-necked style, to hold a minimum of 2 liters (4 pints) of waste must be provided by the user. It should be placed where it is protected by the bench and in full view of the operator.

1.3.5 Laboratory Conditions

The area selected for the operation of a Varian LC system must be free from drafts, corrosive atmospheres and vibration.
Sample preparation areas and materials storage facilities should be located in a separate room.
The area should be a dust-free, low humidity environment. Air-conditioning is strongly recommended for control of the environment. For optimum analytical performance it is recommended that the ambient temperature of the laboratory is between 20 and 25 °C (68 and 77 °F).

1.3.6 Instrument Cooling Air Supply

The Varian 900-LC Series Liquid Chromatograph requires clean, dry, non-corrosive air for cooling purposes. This is supplied to
the instrument through an air supply vent located at the rear of the instrument.
The air supply is used to cool the electronics of the instrument. Several of these assemblies contain parts prone to corrosion. The introduction of cooling air contaminated with high levels of acid, vapors or other corrosive substances may cause damage to the instrument.
Due to the corrosive nature of some work, it is recommended that
24 Publication date: December 2007
Page 25
Varian 900-LC Series Liquid Chromatograph
in applications demanding high usage of corrosive materials, an
external cooling air supply system is provided. It is strongly recommended that the cooling air be supplied from an
environmentally controlled area that is away from the instrument exhaust and any other area where corrosive materials are stored or used.
The cooling air system with flue, fan, ducting and supply cowl, must provide a minimum positive flow of 6 cubic meters per minute (200 cfm). The ducting should be corrosion-resistant and fireproof.

1.3.7 Weights and Dimensions

System unit Width Depth Height Weight
Varian 920-LC Analytical Liquid Chromatograph
Varian 940-LC Semi-Preparative Liquid Chromatograph
Varian 940-LC Preparative Liquid Chromatograph
Shipping dimensions for all LC systems
IBM® compatible (typical)
Laser (typical)
External Options
Fraction Collector
Shipping dimensions 760 mm
Refractive Index Detector
750 mm 30 in
750 mm 30 in
750 mm 30 in
950 mm 38 in
450 mm 18 in
500 mm 18 in
400 mm 16 in
30 in
296 mm 12 in
725 mm 29 in
725 mm 29 in
725 mm 29 in
895 mm 36 in
770 mm 30 in
650 mm 30 in
600 mm 24 in
500 mm 20 in
385 mm 15 in
665 mm 27 in
665 mm 27 in
665 mm 27 in
1165 mm 46 in
520 mm 20 in
200 mm 20 in
480 mm 19 in
600 mm 24 in
212 mm 9 in
133 kg 294 lb
139 kg 307 lb
137 kg 302 lb
170 kg 375 lb
(n/a)
(n/a)
18 kg 34 lb
31 kg 69 lb
11 kg 25 lb
Shipping dimensions 775 mm
31 in
Evaporative Light Scattering Detector
Publication date: December 2007 25
200 mm 360 mm 415 mm 11 kg
475 mm 19 in
460 mm 18 in
14 kg 31 lb
Page 26
Varian 900-LC Series Liquid Chromatograph
(ELSD)
Shipping dimensions 700 mm
Table 5. 900-LC Series Liquid Chromatograph and options dimensions
Warning
The LC system weighs approximately 140 kg (309 lb). To avoid injury to personnel or damage to equipment, always use a forklift or other suitable lifting device when moving the instrument.
8 in 14 in 17 in 25 lb
450 mm
28 in
18 in
1.4 Personal Computer Requirements
The recommended and minimum PC specifications are listed in the following table. They can also be found on Varian’s Web site at www.varianinc.com. The minimum configuration represents the absolute minimum that is required to run the Galaxie™ software. This PC configuration may be out of manufacture, but you may want to use a PC you already have. The recommended configuration should be followed when buying a new PC.
Minimum Recommended
600 mm 24 in
18 kg 40 lb
Pentium 4 processor, 2.8 GHz or higher Core 2 Duo, 2.13 GHz or higher
1 GB of RAM 2 GB of RAM
160 GB hard drive 2 x 200 GB hard drives
16 speed CD-ROM or DVD-RW 16 speed CD-ROM or DVD-RW
Super VGA monitor with high color (16 bit) display, 1024 x 768 resolution, minimum 17” monitor
For the use of PDA features, the graphic card should be in the list below:
INTEL integrated video card
AM Radeon AGP
ATI Mobility 128 AGP
ATI Rage Mobility 128 AGP
ATI mobility M3
ATI 3D Rage PRO AGP 2X
26 Publication date: December 2007
Super VGA monitor with high color (16 bit) display, 1024 x 768 resolution, minimum 17” monitor
For the use of PDA features, the graphic card should be in the list below:
INTEL integrated video card
AM Radeon AGP
ATI Mobility 128 AGP
ATI Rage Mobility 128 AGP
ATI mobility M3
ATI 3D Rage PRO AGP 2X
Page 27
Varian 900-LC Series Liquid Chromatograph
ATI Rage 128 Pro Ultra GL AGP
ATI Rage XL PCI (B41)
ATI Rage 128 Ultra (16MB)
Neomagic Magicgraph 128XD
NeoMagic MagicMedia256ZX
Mga-mil: (chip: MGA-2064WB4R2;
CNA: TI TVP3026 220MHZ)
Nvidia Riva TNT (with last driver version: chip: NV4; memory: 16Mo; Files: nv4_mini.sys, nv4_disp.dll; Version : 4.00,4.0.0)
NVIDIA GeForce4 MX 420,
DDR
NVidia GEForce4 440 MX
Matrox G200 AGP
64MB
ATI Rage 128 Pro Ultra GL AGP
ATI Rage XL PCI (B41)
ATI Rage 128 Ultra (16MB)
Neomagic Magicgraph 128XD
NeoMagic MagicMedia256ZX
Mga-mil: (chip: MGA-2064WB4R2;
CNA: TI TVP3026 220MHZ)
Nvidia Riva TNT (with last driver version: chip: NV4; memory: 16Mo; Files: nv4_mini.sys, nv4_disp.dll; Version : 4.00,4.0.0)
NVIDIA GeForce4 MX 420,
DDR
NVidia GEForce4 440 MX
Matrox G200 AGP
64MB
16 bit sound card 16 bit sound card
Windows® compatible keyboard Windows® compatible keyboard
Microsoft® or compatible mouse Microsoft® or compatible mouse
Two PCI-compatible slot for I/O card Two PCI-compatible slot for I/O card
Windows® 2000 Windows® XP SP 1
Microsoft® Internet Explorer® v6.0 Microsoft® Internet Explorer® v6.0
Table 6. Recommended PC specifications
*For Galaxie™ CDS server specifications please contact your local Varian sales representative.
PCs supplied with Letter of Credit orders will be an international brand and will be the ‘Recommended’ configuration or better.
Higher rated PC components can be substituted for those listed above e.g., processor type, amount of memory, screen size and resolution, operating system version etc.
For the latest PC recommendations please see the Varian, Inc. website at www.varianinc.com
Publication date: December 2007 27
Page 28
Varian 900-LC Series Liquid Chromatograph
This page is intentionally left blank.
28 Publication date: December 2007
Page 29
Varian 900-LC Series Liquid Chromatograph

2. Installation

The Varian 900-LC Series Liquid Chromatograph must be installed by a Varian-trained, Varian-qualified or Varian-authorized representative.
You should have completed and returned the form in the pre-installation manual stating that you have prepared the laboratory in accordance with the requirements detailed in that manual. A Varian representative will then arrange a suitable installation date with you.
Details for unpacking the instrument and what to do in case it has been damaged in transit are also outlined in the pre-installation manual.
Basic installation involves:
Setting up the PC to instrument communications
Installing the Galaxie™ software
Installing the Online Help
Installing the hardware
Running the Installation Qualification tests
Basic customer training

2.1 900-LC Series LC System to PC Communications

The company network must allow TCP/IP communications between the client, acquisition server and HPLC systems. However a desktop switch is provided with the instrument, to allow local networking. The following are suggested network connections using the provided switch:
Publication date: December 2007 29
Page 30
Varian 900-LC Series Liquid Chromatograph
Figure 1. Single 900-LC Series LC System in standalone configuration
Figure 2. Multiple 900-LC Series LC Systems in standalone configuration
Up to four systems can be controlled by one Galaxie Acquisition Server. For a client server system there will need to be a PC with the Galaxie Acquisition Server running for every four HPLC
30 Publication date: December 2007
Page 31
Varian 900-LC Series Liquid Chromatograph
systems.
Figure 3. Networked multiple 900-LC Series LC System configuration option 1
Figure 4. Networked multiple 900-LC Series LC System configuration option 2
You or your network administrator must provide any hardware, adapters, cables, and configuration and setup work required to
Publication date: December 2007 31
Page 32
Varian 900-LC Series Liquid Chromatograph
convert from the company network wiring type to 10/100 BaseT wiring. This hardware must comply with the IEEE 802.3 standard.
Choose the appropriate set up for your laboratory. Use standard Ethernet cables. Your 900-LC Series HPLC system came with one Ethernet cable. You will need to provide additional cables based on your LC system and PC configuration.
To connect the Ethernet cables:
1. Connect an Ethernet cable from the Network/PC port on the side of the 900-LC Series HPLC system to the provided Ethernet switch. Consult the instructions provided with the switch for more information.
Figure 5. Side of 900-LC Series HPLC
This image has been flipped to show horizontal. The connections are vertically oriented on the side of the LC system.
2. Connect an Ethernet cable from the Ethernet switch to the Ethernet port on the PC(s) and/or company network.

2.2 Installing the Galaxie™ Chromatography Data System

Varian recommends that you purchase a PC as part of the Varian 900-LC Series Liquid Chromatograph package. The PC included in the package will come with the appropriate operating system as loaded by the PC supplier.
For instructions on installing the Windows® operating system, please refer to the appropriate manuals supplied with the software. It is your responsibility to ensure that the operating system software has been installed and is functional if you are supplying the PC or the operating system.
Note Varian will not assume responsibility for loss of data.
Your Varian representative will connect the PC to the LC system and any factory-approved accessories purchased at installation. Initial instrument software installation is also included as part of the
32 Publication date: December 2007
Page 33
Varian 900-LC Series Liquid Chromatograph
system installation.
Although the Varian-trained, Varian-qualified or Varian-authorized representative will install the Galaxie™ software for you during the installation process, you may need to install the software yourself at some later stage, for example if you change the PC.
The Galaxie LC Drivers Serial Number will also work for the 900­LC. If you are running both types of Varian LC systems, the same serial number will need to be entered twice, both when installing LC Drivers and when installing the 900-LC driver. You will only need to purchase the one serial number. For subsequent installs, the serial number is recognized.

2.2.1 Client/Server Installation Information

The client/server installation involves using up to three PCs:
Main server – where data and methods are stored.
Acquisition server – communicates with the hardware of the
900-LC Series HPLC system. Up to four 900-LC systems can be run by one acquisition server. The system is configured on the acquisition server.
Client – methods are created, data viewed and reprocessed.
Note The 900-LC Series driver must be installed on all three PCs. The
online help must be installed on every PC where it may need to be viewed on.

2.2.2 Stand-alone Installation Information

When using a stand-alone configuration the software for the main server, acquisition server and client are installed on one PC. This PC can then be used to control up to four 900-LC Series HPLC systems.

2.2.3 Galaxie Installation Procedure

The 900-LC Series Driver installation has several steps:
1. Install the Galaxie core software.
2. Install LC or GC drivers if required.
3. Install the 900-LC Series HPLC driver.
Publication date: December 2007 33
Page 34
Varian 900-LC Series Liquid Chromatograph
4. Install the 900-LC Series HPLC Online Help.
5. Configure the 900-LC Series HPLC system(s).
Note Before installation you must have the LC system set up and turned
on. Set up includes having the network cabling and hardware plugged in.
Contact your IT department to determine if you will be setting up the instrument as a dynamic or static IP and if you will be setting up a client/server system or a standalone system.
Note The dynamic IP address must be leased to the instrument and the
static or dynamic IP address of the LC system must be on the same gateway/subnet as the acquisition server or stand-alone PC. In addition, the LC systems must be connected to the provided Ethernet switch.
Note Disable the Windows® firewall on the PC(s) to be used with
Galaxie software. This will help prevent installation and data collection problems.
Prior to installation check the following:
Disable “Hyperthreading in the Bios. Consult your PC manual for instructions.
Check that power management of your Ethernet network card is set to never allow the card to be turned off. Consult your PC manual for instructions.
In the Windows® Control Panel > Display > Screen Saver set the power scheme for your monitor and hard drive to Never.
Turn off the screen saver.
Turn the Windows® XP Pro sp2 firewall off.
Disable any virus scans.
Turn off Automatic Updates in the Windows® Control Panel.
2.2.3.1
Note The 900-LC Series HPLC Driver requires Galaxie™ software
34 Publication date: December 2007
Installing the Galaxie Core Software
To install the Galaxie™ core software:
1. Please see the Galaxie Chromatography Data System Installation
Guide, p/n 03-914948-02 for installation instructions.
version 1.9 sp1.
Page 35
Varian 900-LC Series Liquid Chromatograph
2.2.3.2
2.2.3.3
Note Before installation you must have the LC system set up and turned
Installing Other LC or GC Device Drivers
The drivers work with the Galaxie core software to provide instrument control and data acquisition. Go to the next section, Installing the 900-LC Series HPLC Driver, if this section is not required.
To install other LC or GC device drivers:
1. Please see the Galaxie Chromatography Data System Installation Guide, p/n 03-914948-02 for installation instructions.
Installing the 900-LC Series HPLC Driver
The drivers work with the Galaxie core software to provide instrument control and data acquisition. Install this driver on every Main Server, Acquisition Server and Client PC that will be working with the 900­LC Series HPLC system.
on. Set up includes having the network cabling and hardware plugged in.
To install the 900-LC Series HPLC Driver:
1. Insert the 900-LC Series Galaxie Driver CD into your CD-ROM drive. The software will automatically start
2. Click Next on the Welcome to Install Shield Wizard for Varian Series 900-LC Drivers window.
3. Click Yes to accept to the License agreement. Clicking no will terminate the program.
4. Enter in your name, the company name, and the serial number of the 900-LC Series HPLC Driver Disk. The serial number is provided on your driver serial number card and is different from the Galaxie serial number.
The installer begins to install the driver files onto the PC.
5. Click Yes to install the Flash Player.
1
.
Note If you click No, you will not get the full functionality of the 900-LC
Series HPLC driver; however, Flash is not required to run the 900-LC.
1
If you have turned off the auto-start capability of your CD-ROM drive, you will need to run the setup.exe file on the
CD-ROM.
Publication date: December 2007 35
Page 36
Varian 900-LC Series Liquid Chromatograph
6. Click Close on the Flash Player 9 Installer Setup: Completed
window to continue the driver installation.
7. Click Yes to install Adobe Acrobat Reader.
Note If you click No, installing Adobe Acrobat Reader will be skipped.
However, you may not be able to view any accompanying PDF files in the help.
If you click Yes and the Repair/Modify window appears, Adobe Acrobat Reader is already installed on your PC. Click Cancel to stop the Adobe Acrobat Reader installation or click Repair to re-install.
8. Click Finish to complete the Adobe Acrobat Reader installation.
2.2.3.4
2.2.3.5
Installing the 900-LC Series HPLC Help
This step is a continuation from the previous section, Installing the 900-LC Series HPLC Driver.
To install the 900-LC Series HPLC Help:
1. Remove the driver CD and insert the 900-LC Series HPLC Help
CD when instructed. Click OK. If you click Cancel, a message will appear confirming your choice not to install the Online Help. Click No to exit the installation or click Yes to continue installing the Online Help.
2. Click OK to continue the installation.
3. Click Next on the Welcome to the Varian 900-LC Series HPLC
Help Setup Wizard window.
4. Click OK to accept the default installation location or choose a
different location to install a shortcut to the Online Help.
5. Click on Finished.
Configuring the 900-LC Series HPLC System(s)
This step is a continuation from the previous section, Installing the 900-LC Series HPLC Help. To manually configure an LC system, please see the Galaxie Configuration Manager online help or access this set up program by navigating to C:\Galaxie\Server\Devices\SetupWizard.exe. Double click SetupWizard.exe to launch the setup program.
To access the Galaxie Configuration Manager Online Help: Click the Windows® Start button and choose Programs >
Galaxie™ (Chromatography) > Help > Galaxie™ Configuration Manager.
36 Publication date: December 2007
Page 37
Varian 900-LC Series Liquid Chromatograph
Systems should only be configured on an acquisition server or a stand-alone PC configuration.
To configure the 900-LC Series HPLC System(s)
1. Click Yes to configure a system if you are installing on an Acquisition server or on a Stand-Alone system. If you choose No, the installer will exit.
2. Click Yes or No to the question ‘Do you want to give the right “Allow to run instrument maintenance software’ to ALL of the Galaxie user profiles?’. This will provide access to the Instrument Maintenance Software (IMS) for all profiles. See page
42 for
more information on IMS.
Note To configure IMS access for individual profiles use Galaxie
Configuration Manager.
3. If Yes was clicked in step 2, click OK at the user profile changed notification.
4. The 900-LC Series Setup Wizard window appears. Enter in the name of the system.
Note Ensure you have typed in the correct system name. To modify a
system name later you must first delete that LC system and then use this SetupWizard.exe to recreate the LC system.
5. Click on the magnifying glass icon
to start the system
discovery process. This can take up to 5-10 seconds.
6. Click the down arrow button in the Instrument field to select the LC system.
Note Identify the system by the MAC address which can be found on
the back of the LC system and in your documentation that came with the LC system.
Once the system is chosen, IP address information will be automatically entered into the IP address, Subnet mask, and Default gateway fields. This information will be what the instrument was shipped with and may not be appropriate for your use.
7. Choose Obtain an IP address automatically (DHCP) or Use a static IP address.
(a) When Obtain IP address automatically (DHCP) is chosen,
the system will search for an IP address automatically. If one is not found, click Retry. The Switching to DHCP mode window pops up. Click Initialize instrument. This will
Publication date: December 2007 37
Page 38
Varian 900-LC Series Liquid Chromatograph
reboot the LC system communications card. Click Close
when the initialization is complete.
Note If an IP address still does not appear, contact your IT department
to check your network connections. The LC systems must be on the same gateway/subnet as the acquisition server or stand-alone PC configuration and be connected to the provided Ethernet switch.
(b) When Use a static IP address is chosen, enter in the IP
address, Subnet mask, and Default gateway for the LC system. The LC systems must be on the same gateway/subnet as the acquisition server or stand-alone PC configuration and be connected to the provided Ethernet switch.
8. Click Create.
9. Click OK when the message confirming the system creation
comes up.
10. Create up to three more systems or click Close to exit the
installer.

2.2.4 Delete a System

If the system needs to be renamed or the IP address changed you will need to delete the existing system and recreate it.
To delete a system:
1. Open Galaxie Configuration Manager.
2. Click the Windows® Start button and choose Programs > Galaxie™ > Galaxie™ Configuration Manager.
3. Log in to Galaxie Configuration Manager.
4. Click Systems in the upper left side of the window.
5. Right click the system to be deleted and choose Delete.
6. Click Communication engine configuration on the lower left side of the window.
7. Highlight the first line with the system name to be deleted and then click Remove on the upper right side of the window.
8. Repeat the previous step for all lines with the system name to be deleted.
9. Click Apply.
38 Publication date: December 2007
Page 39
Varian 900-LC Series Liquid Chromatograph
10. Click Interface configuration on the bottom left side of the
window.
11. Highlight the first line with the system name to be deleted and
then click Remove on the upper right side of the window.
12. Repeat the previous step for all lines with the system name to be deleted.
13. Click Apply.
14. Exit Galaxie Configuration Manager.
2.2.5 Change the 900-LC Series HPLC System Name or IP
Address
You cannot change the name or IP address of an LC system. You must delete the system first and then re-create it using the SetupWizard.exe.
To change the name or IP address of a system:
1. Delete the LC system. See page 38 for instructions.
2. Re-create the LC system. See page 36 for instructions.

2.2.6 Update Galaxie to Version 1.9 SP1

The 900-LC Series HPLC Driver requires Galaxie version 1.9 sp1 to run. The ordering information for the CD is part number 03-950000­98 (Galaxie Software Service Pack 1 (SP1) CD).
Note Contact your local Varian representative to update Galaxie version
1.8 to 1.9 SP1. This CD will not update Galaxie version 1.8 to 1.9
SP1.
To update to version 1.9 sp1:
1. Insert the CD into your CD-ROM drive.
2. Click Next on the Welcome screen.
The updater will install and update the required files.
3. Click Yes to restart your PC.
4. Click Finish to restart the PC.
Publication date: December 2007 39
Page 40
Varian 900-LC Series Liquid Chromatograph
When the restart has completed a window will appear asking for the Main server name.
5. Enter the main server name into the field.
✰✰✰ Hot Tip To familiarize yourself with the
click the Windows® Start button and choose Programs > Galaxie™ > Galaxie™ Help.
Galaxie™ software, after installation,

2.3 About the Varian Integrated HPLC Online Help

Throughout this manual references are made to the Varian 900-LC Series LC online help or Galaxie™ Chromatography Data Systems help.
The Varian LC system online help is provided on a separate CD to the Galaxie™ software and is installed on your PC during the Galaxie™ software installation.
The LC system help includes information and instructions on:
Safety
LC system standard and optional hardware
Installation of the LC system options and components
including the:
hardware
LC system driver for the Galaxie™ software
Creating methods
Viewing run details
Saving the run
Maintenance
Troubleshooting
Spares and options
40 Publication date: December 2007
Page 41

2.4 Hardware Installation

Your LC system should be ready to operate after the Varian representative has installed it. However, over time you may need to replace some hardware components such as detector lamps, autosampler needles, etc or may need to install a new device to your LC system. For instructions on how to do this, please refer to the online help. Specifically:
1. Click the Windows® Start button and choose Programs > Galaxie™ > Help > Varian Integrated Help.
2. When the Integrated Help appears, click the How to > Installing Components link to access step-by-step instructions on how to
remove and install the various instrument components.
For information on your instrument connections including connecting the network and communication cables please refer to the online help.
1. Click the Windows® Start button and choose Programs > Galaxie™ > Help > Varian Integrated Help.
2. When the Integrated Help appears, click About Your Instrument link to access information about the front, side and back connections and components of your LC system.
Varian 900-LC Series Liquid Chromatograph

2.5 Installation Qualification Tests

Your Varian customer service representative will perform the Installation Qualification tests during the installation procedure.
You will receive a copy of the results of the tests. It is recommended you save those results as a record of your instrument performance at installation.

2.6 Customer Training

During the installation process you will receive basic training on how to:
set up the hardware
use the Galaxie™ software
use the Online Help
use the IMS software.
Publication date: December 2007 41
Page 42
Varian 900-LC Series Liquid Chromatograph

2.7 Using the Online Help

The online help is context sensitive. Pressing the F1 key on any screen in the device control section of the Galaxie™ software will bring up help that relates to that screen.
Alternatively, you can access the online help by:
1. Click the Windows® Start button.
2. Click Programs.
3. Click Galaxie™.
4. Click Help.
5. Click Varian Integrated Help.

2.8 Instrument Maintenance Software (IMS)

IMS software is used to keep track of any maintenance performed on the LC system and usage over time of components within each device in your LC system.
IMS has several features:
Counters which record component usage
Actions which lists all procedures that can be tracked
The Log Book lists all actions and counter information and can
be saved in a PDF file
Service tests which can test the performance of devices in the LC system.

2.8.1 How to Access IMS

Access the Instrument Maintenance Software from the link on the right side of the Status Overview window in the Galaxie™ software. The Instrument Maintenance software can be run with the Galaxie™ version 1.9 SP1 or later. When the logon window appears, enter a User identification, and choose the group and project in which the system to be maintained is defined in Galaxie. Type the corresponding password. The password is case sensitive (abc ABC).
To be able to log into the Instrument Maintenance Software, 'allow to run instrument maintenance software' must be checked in the Acquisition/system properties part of the user profile. If the user
42 Publication date: December 2007
Page 43
does not have the right to run maintenance software an error message is displayed.

2.8.2 About IMS

The software has the menu bar at the top which includes:
File
Varian 900-LC Series Liquid Chromatograph
Open Audit trail
Save Audit trail
Preview Audit trail
Print Audit trail
Quit
View
Actions
Maintenance mode ON
Maintenance mode OFF
About
For detailed information about the IMS software please see the 900-LC Series HPLC online help.
To access the online help:
1. Click the Windows® Start button.
2. Click Programs.
3. Click Galaxie™.
4. Click Help.
5. Click Varian Integrated Help.
6. Click Instrument Maintenance Software in the online help
table of contents.
For information on audit trails:
1. Click the Windows® Start button.
2. Click Programs.
3. Click Galaxie™.
Publication date: December 2007 43
Page 44
Varian 900-LC Series Liquid Chromatograph
4. Click Help.
5. Click Varian Integrated Help.
6. Click the plus sign next to Instrument Maintenance Software in the online help table of contents.
7. Click Audit Trails.
44 Publication date: December 2007
Page 45
Varian 900-LC Series Liquid Chromatograph

3. Operation

You will find step-by-step instructions for common operations in the Integrated HPLC Help. To access this information:
1. Select the Windows® Start button and choose Programs > Galaxie™ > Help > Varian Integrated Help.
2. When the Integrated HPLC Help appears, click the How to link to view the available step-by-step instructions.
Below is a quick guide to getting the instrument set up and running samples.

3.1 Analysis Checklist

You need to complete the following steps in turn to measure a sample/s. You will find details on each step in this chapter.
Turn on the instrument and software
Prepare the instrument for use
Configure the Device Settings and System Parameters
Create a method
Run samples
View or Print a report

3.2 Turning on the Instrument and Software

Before starting the system, carefully read the Safety practices and hazards laboratory is set up according to the details specified in the Pre-installation manual.
To turn on the instrument and software:
1. Switch on the computer, monitor and printer.
Publication date: December 2007 45
section at the front of this manual and ensure that the
Page 46
Varian 900-LC Series Liquid Chromatograph
2. Turn on the LC system.
3. To start the Galaxie™ software, click the Windows® Start button and then choose Programs > Galaxie™ > Galaxie™.
4. Enter in your login details and click OK.

3.2.1 Connecting to a System

Depending on the configuration you purchased, you may have up to four LC systems connected to one PC. Due to this flexibility in the software, you must connect to a system when you open the Galaxie™ program.
To connect to a system:
1. Click the Systems tab at the bottom left of the main Galaxie™ window.
2. Select the empty box next to your system.
Your Galaxie™ software will automatically begin communicating with your LC system.
Note There will be a delay between choosing your system and the system
information appearing on your screen. The Galaxie™ software is communicating with the LC system.

3.3 Preparing the System for Use

For consistent and reproducible results with each use of the iHPLC it is recommended that you follow this procedure to prepare your system to run samples.
To prepare the system for use:
1. Check that all tubing to the column(s), pumps, and waste is correctly connected and in the proper position.
2. Place all samples into the autosampler tray. Cool the tray prior to placing samples into it if necessary.
3. Close the LC system doors.
4. Check that all tubes are placed in the fraction collector (if necessary).
5. Set the flow rate manually to remove any air bubbles or old solvent from the system.
46 Publication date: December 2007
Page 47
Varian 900-LC Series Liquid Chromatograph
6. Purge the pumps if new solvent is being used.
7. Purge any air bubbles from the autosampler injection valve if necessary.
8. Turn on the column heater and allow the column(s) to come to the correct temperature if necessary.
Caution
Caution
It is imperative that whenever the column heater is turned on, the pump should be running with a non-absorbing solvent flowing through the column to avoid column degradation.
9. Turn the detector lamp(s) on.
It is imperative that whenever the Varian detector lamp is turned on, the pump should be running with a non-absorbing solvent flowing through the cell, with no air bubbles present.
For more information on getting your system prepared for use please see the 900-LC Series HPLC Help.
1. Click the Windows® Start button.
2. Click Programs.
3. Click Galaxie™.
4. Click Help.
5. Click Varian Integrated Help.
6. Click the plus sign next to Getting Started.
7. Click Preparing the LC System For Use.

3.4 Configuring the Device Settings and System Parameters

The Device Settings allow you to program settings for each component within your iHPLC as well as control flow rates, purge the pumps, turn the detector lamps on, heat the column oven and move the autosampler needle and fraction collector probe.
The System Parameters are settings that will download directly
Publication date: December 2007 47
Page 48
Varian 900-LC Series Liquid Chromatograph
your method. They include information such as the pressure units to be displayed on screen, autosampler tray size, fraction collector rack sizes, leak detection actions, pump head size, sample loop size, and enabling the cooling on the autosampler trays.
For additional information on your system components please see the 900-LC Series HPLC Online Help.

3.4.1 Flows

In Flows you can:
manually set and turn on your flow rate and solvent composition
purge the pumps
set the bottle management parameters
select the solvents for the scale up module valves (if present
in your system)

3.4.2 Detectors

Depending on the detector you purchased, in the detector Device Settings you can:
choose wavelengths
autozero the detectors
turn your detector lamp(s) on
choose the data collection parameters
set the detector temperature conditions.

3.4.3 Column

Depending on the column options you purchased you can:
set the column heater temperature
turn the heater on
choose the scale up module column
choose between two columns if a column switching valve is
present
view the column properties of your column if you have
48 Publication date: December 2007
Page 49
downloaded your method to the LC system.

3.4.4 Sample Introduction

Sample introduction can occur by using an autosampler or manual injector if you are running an analytical or analytical to semi­preparative LC system or via a manual injector or pump if you are running a preparative scale LC system.
Autosampler
If you purchased an analytical or analytical to semi preparative system you may have an autosampler.
In the Autosampler Device Settings you can move the needle to the:
wash port
injection port.
You can also flush the syringe barrel to replace solvent or remove bubbles by clicking the Flush Syringe Barrel button.
If you have the cooling option on your autosampler you can change the temperature.
Varian 900-LC Series Liquid Chromatograph
Manual Injector
Depending on the options you chose, you may have a manual injector. This allows you to make single injections. The Device Settings window lets you make an injection and shows you the state of the injector.
Injection Pump
The preparative scale LC system may have an injection pump. In the injection pump method you can program the injection volume and flow rate.

3.4.5 System Parameters

System Parameters are settings that affect new methods. If you change the System Parameters while a method is already open, the changes will not affect that method.
Depending on the system you purchased you can:
set the units preferred for the pressure and the cell ratio if using a semi prep or prep cell
Publication date: December 2007 49
Page 50
Varian 900-LC Series Liquid Chromatograph
set the option to turn the lamps on as the detector(s) power up
select normalization for the PDA detector
synchronize with other modules can be selected to have the
system wait for when a contact closure is to be used to start data acquisition
choose the tray type and cooling options in the autosampler section
edit the manual injector options including the reset time, analytical (SUM) loop size and preparative loop size
edit the reset time for a manual injector which is used to switch the valve back into the idle position
choose which leak detector option will be performed when a leak is detected
choose solvents to be used in the valve module section if you have a Scale Up Module
with semi-prep and prep systems you can choose the pump head sizes, sample delivery pump head size and pressure units
in the fraction collector section you must choose your rack type, tube volume and probe depth
save your settings

3.5 Creating a Method

To create a new method:
1. Click File > New > Method in the upper left corner of the
Galaxie™ software.
2. Choose the Project Name and System name from the drop-
down menus and then click Next.
3. Enter the Method Name and Description and then click OK.
Note Galaxie™ will open to the main method screen on the Integration
Events tab.
4. In the lower left corner click on control.
Note Galaxie™ will show the Device Select window. You should see a list of
your system components just to the left of the Device Select window.
50 Publication date: December 2007
Page 51
Varian 900-LC Series Liquid Chromatograph
5. Choose the devices needed for your run by selecting or deselecting the options in the Device Select window.
6. On each device window set your parameters.
Note Your System Parameters will automatically appear in your method.
All other settings will need to be checked.
7. Click acquisition in the lower left corner of the Galaxie™
software.
8. Click on the button with the three dots next to the File prefix field to open up variable options for your run name. Double click the variables to be used in your run name or type in a
name for your run and then click OK.
9. Set the identifier number if something other than 1 is required to name the run.
10. In the Acquisition Parameters section enter in the Vial and Rack number if an autosampler is to be used, the Acquisition length and Injection Volume.
Note For information about the additional fields in the Acquisition window
please see your Galaxie™ Chromatography Data System software Online Help.
11. Click File > Save > Save Method. The method will now be
saved in the default data folder.
For more information on creating a method please see the 900-LC Series HPLC Help.
1. Click the Windows® Start button.
2. Click Programs.
3. Click Galaxie™.
4. Click Help.
5. Click Varian Integrated Help.
6. Click the plus sign next to Software.
7. Click Galaxie software tutorial for a basic overview of the
main Galaxie windows.
8. Alternatively, click the plus sign next to How To…
9. Click Program a method from start to finish for a step by
step tutorial on how to program a method.
Publication date: December 2007 51
Page 52
Varian 900-LC Series Liquid Chromatograph

3.6 Starting an Acquisition

There are two ways to start an acquisition. You can have multiple injections by creating a sequence or single injections using Quick Start.

3.6.1 Sequence

1. From the File menu click New > New Sequence.
2. Choose the Project name and System name to be associated with
this sequence and then click Next.
3. Enter the number of lines (number of samples) in the sequence
and then click Next.
4. Enter the new sequence name and a description if needed and
then click OK.
Note A window appears with the number of lines you chose in step 3. You
can auto-fill the columns by entering or choosing a value in the first line and then right clicking the heading of that column. Choose Fill Block and then follow the prompts. Alternatively you can manually fill in each field to customize your sequence.
5. Enter in all required fields.
6. Click File > Save > Save Sequence.
7. On the Sequence page, click the green arrow button just above and left of the lines.
8. To view your runs click the Systems tab on the bottom left of the Galaxie™ software.
Note For further information about starting an acquisition and
interpreting the results, refer to the Galaxie™ Chromatography Data System Online Help.

3.6.2 Quick Start

To start a Quick Start run:
1. Click Acquisition on the top menu bar.
52 Publication date: December 2007
Page 53
Varian 900-LC Series Liquid Chromatograph
2. Choose Quick Start.
3. Select the Project name, System name and Method name and
then click OK.
4. Click on the button with the three dots next to the File prefix field to open up variable options for your run name. Double click the variables to be used in your run name or type in a
name for your run and then click OK.
5. Set the identifier number if something other than 1 is required to name the run.
6. In the Acquisition Parameters section enter in the Vial and Rack number if an autosampler is to be used, the Acquisition length and Injection Volume.
7. Click the Start button.
8. To view your run click the Systems tab on the bottom left of
the Galaxie™ software.
Note For further information about running samples and interpreting the
results, refer to the Galaxie™ Chromatography Data System Online Help.

3.7 Printing a report

To print a report:
1. Click Report Style in the lower left corner of the Galaxie™ software.
Note When printing a report, you can simply select Print from the File
menu. However, if you do so, a report will be generated based upon the last used Report Settings. You will not be able to select the data to be included in the printout.
2. Choose a report style or edit a style.
Note For more information on how to edit report styles please see the
Galaxie™ Chromatography Data System Online Help.
3. Click File > Print to generate a report as specified.
4. The Print dialog box will open. You can select to print all of the report, or a range of pages. Reports are printed to your default
Publication date: December 2007 53
Page 54
Varian 900-LC Series Liquid Chromatograph
printer, unless you specify otherwise. You can specify your printer options in the Print Setup dialog box, accessible from the File menu.
As the Print dialog is a standard Windows® dialog, you can obtain “What’s this” help on an item by clicking the question mark icon in the top right corner and then clicking the item of interest.

3.8 Saving Your Files

There are two ways to save a file.
Save a file with the same name (overwriting the previous one)
by using the menu File > Save / Save file type
Save the file with another name by using the menu File > Save As / Save file type As.
In both cases, the Galaxie™ Software prompts as default to save the file type corresponding to the active file.
The Save file window displays the list of the files already created in the directory associated with the group (when the user is logged on to All projects) or with the project (when the user is logged on to a specific project).
Access the sub-directories by double-clicking on the corresponding folder icon. Once inside the sub-directory, the path is displayed at the top of the window.
54 Publication date: December 2007
Page 55
Varian 900-LC Series Liquid Chromatograph

4. Maintenance

This chapter includes information on how to access the Varian 900-LC Series HPLC system maintenance requirements that may be carried out by an operator. Any maintenance procedures not specifically mentioned should be carried out only by Varian-trained, Varian-qualified or Varian-authorized representatives.
Warning
The UV-Vis, PDA and Fluorescence detector lamps are intense light sources. Direct viewing of the light source will cause eye damage. Operators and other unauthorized personnel must NEVER remove the main covers.
Warning
This instrument contains electrical circuits, devices and components operating at dangerous voltages. Contact with these circuits, devices and components can result in death, cause serious injury, or painful electrical shock.
Warning
Allow hot parts to cool before proceeding with any maintenance procedure.
Publication date: December 2007 55
Page 56
Varian 900-LC Series Liquid Chromatograph
Warning
There is a danger of ignition when using flammable chemicals such as organic solvent. Aqueous solvents or organic solvents having an ignition point above 70 °C can be used. Do not use organic solvents having an ignition point below 70 °C.
Be careful about ignition due to static electricity when using flammable chemicals. Use a conductive vessel and carry out grounding correctly, especially when using a non-conductive chemical. This instrument is not explosion-proof.
Warning
Always check the flow path for leakage of solvent or waste solution during use and leakage of solvent inside the instrument. If an abnormality is found, stop the operation immediately.
Provide sufficient ventilation in the room.
Use of the LC system and accessories may involve materials, solvents and solutions which are flammable, corrosive, toxic or otherwise hazardous. Careless, improper or unskilled use of such materials, solvents and solutions can create explosion hazards, fire hazards, toxicity and other hazards that can result in death, serious personal injury or damage to equipment.
Note This section refers only to maintenance procedures for the instrument.
You should refer to your PC and printer manuals for their maintenance procedures and to the 900-LC Series HPLC Help for the maintenance procedures for any accessories you ordered.

4.1 Routine Maintenance

Routine maintenance information is provided in the 900-LC Series HPLC online help. To access these instructions:
1. Click the Windows® Start button.
2. Click Programs.
3. Click Galaxie™.
4. Click Help.
56 Publication date: December 2007
Page 57
5. Click Varian Integrated Help.
6. Click the Maintenance link.
7. Click Checklist.

4.2 Cleaning

Any spills in the sample compartment or on the LC system should be wiped up immediately.
The exterior surfaces of the instrument should be kept clean. All cleaning should be done with a soft cloth. If necessary, this cloth can be dampened with water or a mild detergent. Do not use organic solvents or abrasive cleaning agents.

4.3 Spare Parts

For spare parts and consumables ordering information, refer to the Varian, Inc. Web site:
www.varianinc.com
Varian 900-LC Series Liquid Chromatograph
Publication date: December 2007 57
Page 58
Varian 900-LC Series Liquid Chromatograph
This page is intentionally left blank.
58 Publication date: December 2007
Page 59
Varian 900-LC Series Liquid Chromatograph

5. Troubleshooting

For troubleshooting help please refer to the Varian 900-LC Series HPLC online help.
To access the online help troubleshooting section:
1. Click the Windows® Start button.
2. Click Programs.
3. Click Galaxie™.
4. Click Help.
5. Click Varian Integrated Help.
6. Click the Troubleshooting link for general LC system
troubleshooting or click the plus sign next to Troubleshooting to access the device troubleshooting information.
Publication date: December 2007 59
Loading...