Overview of Internal Configuration ......................................................................................... 5
Scan Functions Control MS Processes ...................................................................................................................... 6
Steps to Create a Mass Spectrum .............................................................................................................................. 6
Ion Storage ........................................................................................................................................................... 7
Ion Preparation ..................................................................................................................................................... 7
Ion Analysis .......................................................................................................................................................... 7
About Internal Electron Ionization ............................................................................................................................... 7
Ion Preparation Options ....................................................................................................................................... 8
Scanning Ions to Collect Mass Spectra ............................................................................................................... 8
Library Searching and Data Handling .................................................................................................................. 9
About Internal Chemical Ionization ............................................................................................................................. 9
Chemical Ionization Reactions in PCI ................................................................................................................ 10
Ion Preparation Options ..................................................................................................................................... 10
Scanning Ions to Collect Mass Spectra ............................................................................................................. 11
Type of Matrix .................................................................................................................................................... 11
Using EI and PCI to Get More Information ........................................................................................................ 11
Conversion from External to Internal Configuration ................................................................................................. 12
Conversion from Hybrid to Internal Configuration .................................................................................................... 12
Effects of Hardware Configuration Changes ...................................................................................................... 12
Starting the Instrument ............................................................................................................................................. 13
Initial Pump down ............................................................................................................................................... 13
Check the Vacuum Status .................................................................................................................................. 13
Setting System Temperatures ........................................................................................................................... 15
System Bakeout........................................................................................................................................... 15
Startup and Shutdown ........................................................................................................................................ 16
Starting the System ..................................................................................................................................... 16
Shutting Down the System .......................................................................................................................... 16
Adjustment and Tuning of the MS ............................................................................................................................ 16
Checks and Adjustments ................................................................................................................................... 16
Adjusting the RF Ramp ............................................................................................................................... 17
Adjusting the RF Ramp ............................................................................................................................... 17
Adjusting the Calibration Gas ...................................................................................................................... 17
CI Gas Adjustment ...................................................................................................................................... 18
Setting Up CI Reagents ............................................................................................................................................ 18
Installing a Liquid Reagent ................................................................................................................................. 19
Installing a Gaseous Reagent ............................................................................................................................ 20
Auto Tune ........................................................................................................................................................... 20
Integrator Zero ............................................................................................................................................. 20
Set Electron Multiplier .................................................................................................................................. 20
Electron Lens Tuning ................................................................................................................................... 20
Ion Lens Tuning ........................................................................................................................................... 21
RF Full Scale Adjust .................................................................................................................................... 21
Mass Calibration .......................................................................................................................................... 21
Trap Frequency Calibration ......................................................................................................................... 21
Trap DC Offset Voltage ............................................................................................................................... 21
Preparing an MS Method for Data Acquisition ......................................................................................................... 22
Building a New MS Method ................................................................................................................................ 22
Name the Method ........................................................................................................................................ 24
Set the 4000 MS Instrument Configuration ................................................................................................. 24
Select the Acquisition Data Type ................................................................................................................. 25
Edit Chromatographic Time Segments ....................................................................................................... 25
Edit the 4000 MS Method Segments ................................................................................................................. 25
Scan Function Settings ................................................................................................................................ 26
General Parameters Tab ............................................................................................................................. 26
Ionization Control ......................................................................................................................................... 27
EI Ionization Parameters ............................................................................................................................. 27
Setting Parameters for Chemical Ionization ................................................................................................ 28
Full Scan Parameters .................................................................................................................................. 29
Setting Parameters for an MS/MS Method Segment .................................................................................. 30
Viewing Method Parameters in Manual Control................................................................................................. 31
Activating a Method ..................................................................................................................................... 31
Editing a Method in Manual Control ............................................................................................................ 33
Saving a Method .......................................................................................................................................... 35
Acquiring Data .......................................................................................................................................................... 35
Status and Control ....................................................................................................................................... 36
Activating a Method ............................................................................................................................................ 37
Injecting a Single Sample ................................................................................................................................... 37
Injecting Using a SampleList .............................................................................................................................. 38
Monitoring the Status of Runs ............................................................................................................................ 39
4
Page 5
Overview of Internal Configuration
The Internal configuration is one of the three operational configurations of the
4000 GC/MS system. In this configuration, it is possible to collect either electron
ionization or positive chemical ionization (EI or PCI) data. Unlike the External
Ionization configuration, it is possible to collect EI or PCI data during different
time segments of a single chromatographic acquisition. However, it is not
possible to collect Negative Chemical (NCI) data in Internal configuration. In
common with the other configurations, it is possible to perform ion preparation
techniques including Selected Ion Storage (SIS) or with optional software and
equipment, Tandem Mass Spectrometry: (Automated Methods Development
(AMD), MS/MS, MS
GC/MS Software Operation Manual for on-line help information on setting up
these experiments.
The 4000 GC/MS analyzes the gas-phase ions formed from a sample in terms of
their mass-to-charge (m/z) ratios and their relative abundances in the resulting
spectra. The mass spectrum is a graphical representation of the ion intensities
versus the mass-to charge ratio.
The 4000 GC/MS creates the mass spectrum with an ion trap analyzer. The ion
trap confines the ions within a single region where time-dependent
electromagnetic fields control ion processes.
n
, and Multiple Reaction Monitoring (MRM). Refer to the 4000
Filament
Endcap
Filament
Electron Gate
Transfer
Line Tip
5
E.M.
Endcap
Electron Multiplier
H.E.D.
Ring Electrode
Page 6
Scan Functions Control MS Processes
Scan functions are programmable time sequences for applying voltages to the
external source, lens, and ion trap electrodes. The scan function controls the
variations in time of the RF potential applied to the ring electrode as well as any
supplementary waveforms applied to the endcap electrodes. The 4000 MS
module has two techniques, Electron Ionization (EI) and Chemical Ionization (CI),
for generating sample ions in the external source. Consequently, there are two
basic scan functions to perform electron ionization and chemical ionization (either
positive or negative). The diagram below illustrates the RF voltage profile of a
typical EI scan function. These scan functions are discussed in more detail in the
4000 GC/MS Software Operation Manual.
RF Voltage
Steps to Create a Mass Spectrum
Sample analysis with an ion trap analyzer may be divided into several steps:
Sample Introduction
Sample components are introduced from the transfer line to the ion trap analyzer
through the direct-coupled capillary column.
Ionization
The sample is ionized in the mass spectrometer by one of the following method:
•Electron Ionization (EI): The sample molecules are struck with energetic
electrons removing an electron from a molecular orbital, to create
molecular ion.
•Positive Chemical Ionization (PCI): A reagent gas is introduced in the ion
trap and EI is performed on that gas to form reagent ions. The reagent
ions then undergo ion-molecule reactions with the sample molecules to
create ions of the sample molecules and their fragments.
6
Page 7
Fragmentation
A
Depending on the structure of the molecular ion and the excess internal energy
remaining after electron impact, there may be further unimolecular decomposition
of some of the molecular ions to form various fragment ions and neutrals.
Unimolecular decomposition happens in picoseconds, the time scale of a few
molecular vibrations, effectively occurring at the same time as ionization.
Ion Storage
The molecular and fragment ions are stored and stabilized in the ion trap cavity,
where they travel in defined orbits governed by voltages applied between the ring
electrode and the endcap electrodes. Helium buffer gas is also present in the
cavity. It buffers the ions into more compact orbits that produce well-resolved
mass peaks as they are scanned out. Helium is used as a buffer gas because it
is light and does not degrade MS resolution. While helium ions are created in the
trap, they are ejected electronically as soon as they are formed.
Ion Preparation
Once ions are stored in the trap they can be manipulated, if desired. Examples of
ion preparation techniques are tandem mass spectrometry (MS/MS) and
selected ion storage (SIS). Advantages associated with ion preparation methods
are similar to those of other sample preparation methods, e.g., reduction of
background noise and increased selectivity.
Ion Analysis
The stored ions are analyzed by ramping the RF voltage applied to the ring
electrode to a high value, during which time ions from low to high mass are
successively destabilized and ejected from the trap. Supplemental dipole and
quadrupole voltages applied to the endcap electrodes improve the mass
resolution of the process. After ejection, the ions strike a conversion dynode,
initiating a signal multiplication process at the electron multiplier.
Each of the areas above is discussed in general in this guide, and in detail in the
4000 GC/MS Software Operation Manual.
About Internal Electron Ionization
Internal electron ionization is the most common mode of operation of the 4000
MS. All of the sample ionization, fragmentation, storage, and scanning steps
occur within the confines of the ion trap. Electron ionization generates fragments
that are specific to the compound, providing fingerprint spectra. It is, however,
not selective and will show ions for all compounds present. EI is sensitive but the
sensitivity can be improved by reducing the interfering ions using ion preparation
techniques.
Forming Ions
In the EI mode, electrons are gated into the ion trap cavity during the ionization
period. These electrons collide with the neutral sample (or Analyte) molecules A,
removing an additional electron to create energetically-excited molecular ions
∗+
. Some of the excited molecular ions equilibrate through collisions with
7
Page 8
helium but others undergo unimolecular decomposition to create various
+
−∗+
−
+
eAe
A
2
∗
+
A
(
)
3
3
fragment ions, f
. This set of fragment ions, characteristic of the sample
i
molecule, makes up the mass spectrum.
→+
Ionization
is excited
He
+∗+
De-excitation
→
AA
∗+
+
A
→
→
→
Equilibration
nf
+
11
+
nf
+
22
+
Fragmentation
nf+
The ion trap has a maximum storage capacity beyond which mass resolution and
spectral quality deteriorate. The number of ions created depends on the
ionization time. As the ionization time increases, more ions are created.
Automatic Gain Control (AGC) controls the ionization time to always create an
optimum number of ions in the trap.
The AGC scan-function consists of a prescan and up to six analytical scan
segments. The number of ions detected in the prescan is used to calculate the
ionization time for the analytical scan.
Ion Preparation Options
The 4000 MS can use a combination of waveforms applied to ion trap electrodes
to isolate or remove specific ions once they have been formed and are stored in
the trap.
Options like Selected Ion Storage (SIS) and Tandem Mass Spectrometry
(MS/MS) can be performed on the ions stored in the ion trap before mass
analysis takes place. In SIS, resonant waveforms are applied to eject unwanted
ions within the stored mass range and fill the trap only with ions in the mass
range(s) of interest. In MS/MS, a parent ion is isolated and then dissociated by
energetic collisions with helium buffer gas to form product ions.
n
The Internal configuration can have SIS, MS/MS, MS
Monitoring (MRM) as ion preparation options. SIS is included with all
instruments, while MS/MS, MS
n
, and MRM are available with the MS/MS option
, and Multiple Reaction
installed.
Scanning Ions to Collect Mass Spectra
After ionization, trapping, and ion preparation steps, ions are scanned out of the
trap to the conversion dynode and electron multiplier. The ions are analyzed by
applying a radio frequency (RF) voltage to the ring electrode encircling the trap
cavity. As the voltage increases on the ring electrode, ions are sequentially
ejected from the trap according to their mass-to-charge ratio. Supplemental
waveform voltages are applied to the endcap electrodes during the analysis to
improve mass resolution and mass axis stability. Ions strike the conversion
dynode and then electrons are ejected from the conversion dynode, held at
-10 kV, and are repelled to the electron multiplier. The signal is amplified by ~
by the electron multiplier and sent through an integrator to collect an intensity for
each m/z. MS data are stored as sets of ion-intensity pairs for each m/z over the
acquired mass range. A complete mass spectrum is stored for each analytical
10
5
8
Page 9
scan. There are actually two types of mass scanning in internal EI. First there is a
prescan to count the number of ions formed in a short fixed ionization time. After
a calculation based on the prescan ion count, ions are formed for the ionization
time recommended by the AGC prescan algorithm and the analytical scan is
carried out. The analytical scan can be broken up into up to six segments and the
relative ionization times for the segments can be adjusted to meet tuning
requirements for methods such as those of the US EPA for the compounds
DFTPP and BFB.
Library Searching and Data Handling
The mass spectra collected from the MS are analyzed through Varian MS Data
Review. The identity of most compounds is determined by the comparison of the
collected spectrum with a reference library. The mass and intensity listing is
compared to results collected on other instruments. Such listings include the
NIST library, the Wiley MS library, and the PMW library. Each library has a
different focus, from pharmaceutical to environmental analysis. Custom libraries
can also be generated from results collected on the 4000 MS system.
About Internal Chemical Ionization
Chemical ionization (CI) provides mass spectral data that complement electron
ionization (EI) data for the analysis of complex compounds. The generation of
ions in chemical ionization is a two-step process. A CI reagent gas is introduced
into the ion trap analyzer. The reagent gas is ionized by EI. Then, the sample
molecules are ionized by ion-molecule reactions with the reagent gas ions.
CI is a softer ionization technique than EI. That is, CI imparts less energy to the
sample molecules than does EI. Thus, the ionized sample molecule undergoes
less fragmentation, and an ion indicative of the molecular weight is more likely to
be observed. In addition to molecular weight confirmation, CI mass spectra often
provide other significant structural information that may not be available from EI
mass spectra.
Forming Reagent Ions
In the first step, reagent gas ions are formed as the reagent gas is ionized by
interaction with electrons emitted by the filament. The ion trap operates in a
pulsed mode. The supply of reagent ions is created during the ionization pulse
and consumed during the reaction period to form analyte ions.
In internal mode, the CI reagent can be either liquid or gas. The most common
reagents are methane, methanol, acetonitrile, and isobutane.
Reagent ion formation can be a complex process. For example, when methane is
used as the reagent gas, reagent gas ions are formed as follows:
First, methane is ionized to form two primary ions:
−+•−
+→+e2CHeCH
)(
44
−−+−
These primary ions then react very rapidly to form predominantly the secondary
ions,
++
HCandCH
525
34
:
++→+HeCHeCH
9
Page 10
•++•
+→+
CHCHCHCH )(
3544
++
HHCCHCH+→+
25243
Chemical Ionization Reactions in PCI
In the second step, the reagent gas ions react with sample molecules in the ion
trap to form sample ions. The four principal reactions between reagent gas ions
and sample molecules are as follows:
+
(A) Proton transfer:
(B) Hydride abstraction:
(C) Association:
(D) Charge transfer:
where R
+
is the secondary reagent gas ion and M is the neutral sample molecule.
)RH(+ M → (MH)
+
R
+ M → [M-H]+ + RH
+
R
+ M → (MR)+
+
R
+ M →
M
For methane CI, proton transfer (A) is the major reaction, and hydride abstraction
(B) is the next most often observed reaction. In both cases the resulting evenelectron ions are often relatively stable, and the observation of strong (M+1) or
(M-1) ions is possible even if the EI spectrum of the same component shows no
molecular ion. The exothermicity of the reactions determines the amount of
energy deposited; therefore the degree of fragmentation can be controlled by the
choice of a suitable CI reagent gas. The proton affinities of some common
reagent gases of this type, known as proton transfer agents or Bronsted acids,
range from 130 kcal/mol to 200 kcal/mol in the following order: methane, water,
isobutane, and ammonia (with ammonia resulting in the “softest” ionization).
Among the common liquid CI reagents, methanol has a proton affinity of 180.3
kcal/mol, while acetonitrile is 1862 kcal/mol. By choosing a suitable reagent gas,
you can obtain high specificity (i.e., less efficient detection of background or
matrix interferences compared to the analyte) as well as molecular weight
information for the compounds of interest.
+
+ R
+
+ R
Association reactions (C) typically have very low reaction rates, and the reaction
products require rapid collisional stabilization. The products of these association
reactions are called adduct ions because the reagent ion has been added to the
analyte. They are typically seen on the 4000 GC/MS in Internal configuration at
much lower abundance then the (M+1) ion, but when (M+29) and (M+41) adduct
ions are observed using methane, they are useful for verifying the molecular
weight.
The charge transfer reaction (D) produces a radical molecular ion (i.e., an ion
with an odd number of electrons) that dissociates quickly, giving EI-like spectra.
However, the energy deposited in the molecular ion and the resulting
fragmentation pattern does not depend on the electron energy of the ionizing
electrons.
Ion Preparation Options
The 4000 MS uses a combination of waveforms and RF to isolate, or remove,
specific ions after they are formed and are stored in the trap. Selected Ion
Storage (SIS) and Tandem Mass Spec (MS/MS) can be performed on the ions
stored in the ion trap before mass analysis takes place.
10
Page 11
Scanning Ions to Collect Mass Spectra
The scanning process for chemical ionization is similar to that for electron
ionization. After ionization, trapping, and ion preparation steps, ions are scanned
out to the conversion dynode and electron multiplier. Mass scanning is
implemented by increasing the RF voltage on the ring electrode; the mass
spectrum is collected in order from low to high mass over the user-designated
scan range. In positive modes, electrons are ejected from the conversion
dynode, held at -10,000 V, and repelled to the electron multiplier. In negative
mode, positive ions are ejected from the conversion dynode, held at +10,000 V,
and repelled toward the electron multiplier. The signal is amplified by ~10
electron multiplier and sent through an integrator to collect an intensity for each
m/z. MS data are stored as sets of ion-intensity pairs for each m/z over the
acquired mass range. A complete mass spectrum is stored for each analytical
scan. There is no prescan in internal PCI; the ionization time is calculated based
on the base peak intensity of the previous scan. Ions are formed for this
ionization time and the analytical scan is carried out.
5
by the
Library Searching
The 4000 MS has small CI libraries generated using internal configuration ion
trap GC/MS systems. The libraries are organized by the CI reagent, which
include methane, methanol, and isobutane.
Selectivity Considerations
Type of Matrix
An advantage of chemical ionization is selectivity. In PCI, hydrocarbons have a
poor response in methane CI. It is easier to locate target compounds in a sample
contaminated with hydrocarbons using methane PCI than using EI. Because of
these selectivity considerations, the time spent to develop a method using the
different ionization and ion preparation options is well spent.
Using EI and PCI to Get More Information
Generating EI and CI data on a single sample gives both the ion-intensity
fingerprint information allowing library searching as well as the molecular weight
information to confirm species identification. Fatty acid methyl ester (FAME)
analysis is an example. Under EI conditions, the FAME fragmentation is
extensive and molecular ion intensity is weak. Using CI, the molecular weight is
obvious and the most intense ion is M+1.
Because the 4000 MS is able to switch between EI and CI in a single run, the
best analytical conditions can be used for a given compound.
NOTE: Wait several seconds to switch between EI and CI. The CI segment
should not be less than 60 seconds wide and the CI peak should be at least 20
seconds into the segment to allow CI reagent stabilization.
11
Page 12
Conversion from External to Internal Configuration
Converting the 4000 MS from External to Internal configuration involves changing
both the ion source and the column position. The External ion source assembly is
removed from the trap assembly and replaced with the Internal ion source
assembly.
See the 4000 GC/MS Hardware Operation Manual for details.
Checklist
• Remove the analyzer from the MS manifold.
• Change the ion source to internal.
• Move the heat shield to the forward position.
• Add the filament adapter and connect the flex cable.
• Change the transfer line orientation.
• Change the transfer line tip.
• Cut the column to 7 mm.
• Change the transfer line switch to Internal/Hybrid.
• Replace the analyzer in the MS manifold.
Conversion from Hybrid to Internal Configuration
Changing from the Hybrid to Internal configuration only requires changing the ion
source. The column position and transfer line are already directed to the trap and
do not need to be adjusted.
Outline
1. Change the ion source.
2. Move the heat shield to the forward position.
3. Add the filament adapter and connect the flex cable.
Effects of Hardware Configuration Changes
After changing the configuration, for example from External to Internal
configuration, the following occurs when System Control is restarted.
1. System Control compares the current configuration stored in the current
Module Attributes with the configuration reported by the hardware.
2. If these do not match, the Module Attributes are updated (preset) to the
appropriate configuration. A similar process occurs for the default method
(Default.mth).
3. After making the hardware configuration change, new methods will have the
appropriate instrument configuration by default.
NOTE: Presetting the Module Attributes requires performing all Auto Tune
routines, because the prior Auto Tune results are invalid.
12
Page 13
Starting the Instrument
Initial Pump down
Check the following:
• Check the Vacuum connections.
• Make sure the transfer line is in.
• Make sure the vent valve is fully closed.
• Make sure the column is not broken.
Turn on the power at the main power switch; the roughing pump should stop
gurgling after about 10 to 20 seconds.
If the pump continues to gurgle, then do the following.
1. Check that the analyzer assembly is seated properly on the manifold (there
should be no gaps).
2. Check that the transfer line is in.
3. Check that the vent valve is sealed.
Open System Control and the Startup/Shutdown page opens.
Check the Vacuum Status
The vacuum readings tell a lot about the state of the MS after pump down (and
during operation). Typical operating ranges for the 4000 MS in Internal mode are:
• Speed 100%
• Current 200 – 300 mAmps
• Power 9 – 13 Watts
• Ion Gauge Pressure < 20 μTorr
• Roughing Line < 50 mTorr
If the Pump Spin Speed does not steadily increase, there may be a leak in the
system. Large leaks are indicated by a turbo speed less than 100%. Small leaks
are indicated by an increase in the pump current after 100% is reached, or in the
ion gauge pressure, see “Diagnostics”. Small leaks are diagnosed by changes in
the ion gauge reading and can be pinpointed using the leak check section in the
method Service.mth. For more detail on troubleshooting leaks, see the
Troubleshooting section in the 4000 GC/MS Hardware Operation Manual.
13
Page 14
Diagnostic Tests
Monitor the current state of the instrument using the Monitoring tab. Monitor the
vacuum system, the electron multiplier, the waveform system, temperatures, and
the ion source.
Perform hardware checks on the 4000 MS using the Diagnostics tab. For more
details on the diagnostic tests, seethe Diagnostics section in the 4000 MS Ion Trap Software Operation Manual.
14
Page 15
Setting System Temperatures
Analysis Temperatures
Analysis temperatures are sample dependant. The stability of the compounds
and their volatility affect the temperature choice. For species that are more
fragile, decrease the trap temperature. For example, to obtain molecular ion
information on saturated hydrocarbons, which have unstable molecular ions, set
the trap temperature to 80 °C. However for semi-volatile analysis set the trap
temperature to 220 °C so the heavy PAHs (benzo[ghi]perylene,
dibenz[a,h]anthracene and indeno[1,2,3-cd]pyrene) do not tail.
After changing the temperature setting, allow about one hour for the trap to reach
the new setting. After the setting is reached, allow about two hours for the trap to
stabilize. The temperature of the trap affects mass calibration, and therefore the
trap frequency calibration.
Set the transfer line temperature so there are no cold spot between the GC
column oven and the MS. This is typically about the maximum column
temperature (± 20 °C).
The manifold temperature (typically 50 °C) reduces the effect room temperature
variation may have on the system.
System Bakeout
To remove water adsorbed on the manifold while the 4000 MS was vented,
perform a Bakeout from the Temperatures tab in System Control.
Perform a system bakeout to remove chemical background after running
samples. After doing a bakeout to remove contamination, decrease the
temperature to the analytical temperature.
Typical bake-out settings are:
• Trap Temperature 230 °C
• Manifold Temperature 110 °C
• Transfer Line Temperature 280 °C
• Bakeout time 12 hr
NOTE: You can use shorter bakeout times.
15
Page 16
NOTE: The transfer line temperature should not exceed the maximum isothermal
temperature limit of the column.
Startup and Shutdown
Starting the System
When the system is first turned on, System Control only operates in
Startup/Shutdown mode. During system startup, observe the increase in Pump
Spin Speed in the Operating Conditions field. The software is locked in the
Startup/Shutdown mode until the speed reaches 100%. You will also see the
temperature readings for heated zones begin to increase in the Operating
Conditions field.
NOTE: Failure to reach 100% pump speed in a reasonable time indicates a
vacuum leak and corrective action should be taken. For details, see the
appropriate Troubleshooting section in the 4000 GC/MS Hardware Operatio n Manual.
Shutting Down the System
To shut down the 4000 MS, click the Shut Down button in the upper left corner
of the Startup/Shutdown dialog. The heaters will be turned off and the speed of
the turbo pump will be gradually reduced to 35% of full speed.
After the temperature zones have cooled below 80 °C, turn off the main power by
placing the switch at the rear of system into the OFF (down) position. Manually
vent the system for at least 5 minutes by opening the vent valve on the front
panel one full turn counterclockwise.
NOTE: Retract the transfer line before lifting the analyzer assembly from the
vacuum manifold. Failure to retract the transfer line can cause damage to the
transfer line tip and to the trap assembly.
Adjustment and Tuning of the MS
Checks and Adjustments
RF Tune
Adjust the RF tuning in the Checks and Adjustments tab dialog of Manual Control
after performing one of the following:
16
• Performing MS maintenance.
Page 17
• Changing the analyzer assembly.
• Changing the MS configuration.
Adjusting the RF Ramp
Adjusting the RF Ramp
To adjust the RF ramp, do the following:
1. Click RF Ramp Adjustment in the Checks and Adjustments tab.
2. Click Start.
3. Use a flathead screwdriver to turn the RF Adjustment screw, inside the front
door of the 4000 MS, either clockwise or counterclockwise until the tuning
display shows a straight line and the intensity is at a minimum. The Status
Bar in the Adjustment Results field should be just below OK.
Adjusting the Calibration Gas
To adjust the calibration gas, do the following:
1. Check the flow of perfluorotributylamine (PFTBA or FC-43) calibration gas
before doing Auto Tune procedures.
2. Click Cal Gas Adjustment in the Checks and Adjustments tab in Manual
Control.
3. Turn the Cal Gas valve inside the front door of the 4000 MS either clockwise
to decrease the flow or counterclockwise to increase the flow. Adjust the flow
so that the status bar in the Adjustment Results field reads OK.
17
Page 18
CI Gas Adjustment
Before performing chemical ionization, adjust the reagent gas pressure. The next
section explains the set up of the most common reagent gases.
Air/Water Check
Too high a pressure of air or water in the system because of an air leak or a
need to bake out the system will result in poor performance. This checking
routine gives advice on the levels of air and water.
NOTE: The Air/Water Check uses the electron multiplier voltage giving a 10
gain, not the manual setting. After replacing the electron multiplier, auto tune of
Electron Multiplier before performing the Air/Water Check.
Setting Up CI Reagents
Before running CI experiments, adjust the CI gas pressure.
To adjust the CI gas pressure, do the following:
1. Click the Checks and Adjustments tab of the Manual Control window.
5
2. Select CI Gas Adjustment. Select your CI reagent from the drop down box
in the Parameters section.
3. Click Start below the adjustments selection.
4. The criteria for adjusting the flow of CI reagent into the instrument are shown.
5. After completing this, you must perform a mass calibration and trap function
calibration before beginning your experiments.
18
Page 19
Installing a Liquid Reagent
To connect a liquid reagent reservoir to the liquid reagent inlet block, do the
following:
1. Connect a liquid reagent reservoir containing the chosen liquid to the
liquid reagent inlet block.
2. Open the CI needle valve 6-7 turns clockwise.
3. Click CI Gas Adjustment to open the CI gas valve solenoids.
4. Allow the vapor flow from the reservoir to equilibrate. If, after several
minutes, there is not enough CI gas entering the trap, open the needle
valve (clockwise).
5. While observing the spectrum, use the Adjust CI Gas, and turn the CI
needle valve to increase or decrease the amount of reagent entering the
trap until the resolution between M and M+1 just starts to degrade.
Details for the three most common reagents can be found under the following
sections in the 4000 GC/MS Software Operation Manual: PCI with CH
CN, and PCI with CH3OH.
CH
3
, PCI with
4
Properly Adjusted Methanol Reagent Spectrum
Properly Adjusted Acetonitrile Reagent Spectrum
NOTE: For best results with acetonitrile, use a filament emission current of at
least 20 μA and maintain at least 50% valley between m/z 41 and m/z 42.
19
Page 20
Installing a Gaseous Reagent
To install a gaseous reagent, do the following:
1. Connect the regulator of the gas cylinder to the back of the instrument using
a 4 mL/min restrictor.
2. See the 4000 MS Ion Trap Hardware Operation Manual for more details.
Auto Tune
Depending on your configuration and settings, you may not see all the available
Auto Tune routings. Perform auto tune after the instrument is first set up and
whenever significant maintenance operations are performed. Also, perform Mass
Calibration and Trap Frequency Calibration whenever the temperature or RF
adjustment is changed.
Auto Tune works the same way in either EI or Hybrid CI modes; you do not need
to run a different automatic setup, tuning, and calibration program for Hybrid CI.
Integrator Zero
Integrator Zero obtains the average value of the signal level from the integrator
circuitry when the filament is off. When the filament is off, the major source of
signal coming is electronic noise. The integrator zero is adjusted so electronic
noise does not create artificial ions, which may create a measurable signal.
Set Electron Multiplier
Set Electron Multiplier determines two settings, the multiplier voltage needed to
achieve a multiplier gain of approximately 10
boost for optimum peak intensity and resolution.
5
, and the Electron Multiplier voltage
Electron Lens Tuning
Electron lens tuning involves measuring the transient behavior of the emission
current immediately after the lenses have been switched on or off. If the lenses
are unbalanced, the emission current will change in time and be proportional to
20
Page 21
the imbalance. If the balance is outside the range of 200 to 300 µA, the algorithm
will search the optimal values by changing values of four variables one at a time.
If it fails to find the best voltage setting for lens tuning, auto tune will generate an
error message, and restore the last values in the instrument.
When the Electron Lens Tuning Box is clicked, an additional “Turn on CI gas flow
during tune” option appears. For CI methods in Hybrid mode, the electron
/repeller lens must be tuned with the CI plunger (CI volume) in place and the CI
gas turned on. The user should adjust the CI gas flow in Manual Control before
this tune function is done.
Ion Lens Tuning
The Ion Lens system consists of three lenses (Lens 1, 2 & 3). These lenses are
tuned using Cal Gas ions at m/z 131 and 414. Optimum voltages are determined
based on weighted intensities of the two ions. This is an iterative process in
which transmission of both low and high mass ions is monitored as a function of
lens voltages.
RF Full Scale Adjust
RF Full Scale Adjust sets the full scale adjust potentiometer to give the correct
mass assignment for high mass ions in the calibration gas spectrum. This routine
should always be run before Mass Calibration and Trap Frequency Calibration.
Mass Calibration
Mass Calibration locates and correctly assigns the masses of the PFTBA
calibration gas ions at m/z 69, 131, 264, 414, 464, and 614.
Ion trap temperature changes can shift the mass calibration axis; do not run this
procedure until the ion trap temperature has stabilized for at least two
hours. There could also be subtle effects on mass assignments after ion source
temperature are changed. Mass calibration does not have to be performed again
after the auxiliary Helium buffer gas flow rate is changed.
Trap Frequency Calibration
After the mass calibration has been completed, the Trap Frequency Calibration
must be performed. This calibration determines parameters that are required for
the operation of ion preparation methods such as MS/MS and SIS. These
parameters also help to isolate the range of ions to be acquired in full scan
acquisitions. The routine takes several minutes.
NOTE: Trap Frequency Calibration should always be run after Mass Calibration
is done.
Trap DC Offset Voltage
The trap DC offset is adjusted by this routine to optimize the ion signal for m/z
414 in the calibration gas.
21
Page 22
Preparing an MS Method for Data Acquisition
Building a New MS Method
From the Workstation Toolbar, open the Method Builder .
The ‘Create/Open Method File’ dialog will be displayed. You may choose to
create a new method or open an existing method. If you do not need to see this
dialog again, check the box labeled ‘Do not display this dialog at startup.’
Select the instrument. You can build an offline method by selecting the Custom
configuration and selecting the desired modules.
22
Page 23
Select which detectors (MS and/or GC) will have data handling.
Select what data analysis sections are required; Standard MS Reports and/or MS
Data Handling.
23
Page 24
The method will contain the following sections: 3800 GC Control, 4000 MS
Control, Standard MS Reports and MS Data Handling.
Name the Method
Use the command File… Save As. Enter a name for the method and save it,
either in the default folder or another folder you select.
Set the 4000 MS Instrument Configuration
The GC and the MS will be set to the configuration of the instrument connected
to the Varian MS Workstation. For the 4000 MS Internal configuration, both EI
and CI can be performed during the same run and so there is only one internal
setting.
24
Page 25
Select the Acquisition Data Type
Centroid data is the default acquisition data type, as data handling, library
searching, and spectral comparison can only be done from this type of data.
Centroid data is reported as ion/intensity pairs to one described point.
Profile data is typically used mainly for diagnostic purposes. Profile files are also
approximately 10 times larger than centroid files, but they can be converted to
centroid after acquisition.
Profile data is collected at 10 pts/mz and is displayed as peaks similar to a
chromatogram. The display allows you to observe the true dispersion of the
response and determine if adequate resolution has been obtained.
Edit Chromatographic Time Segments
When the 4000 MS is in Internal Configuration, it can do EI, CI, EI/MS/MS, and
CI/MS/MS in a single run. The Chromatographic Time Segments table allows you
to time-program analysis conditions to get the best results for each segment in
the analysis. Up to 250 time segments can be created for runs up to 650 minutes
in length. By default, there is a Filament/Multiplier Delay segment at the start of
the run so that the system will not be stressed during the elution of the
chromatographic solvent. Following this segment, you could just acquire the
mass spectra in full-scan with a single analysis segment. However, you can tailor
variables such as acquired mass range, insert MS/MS segments for individual
analytes, and otherwise set up the instrument to acquire the best data for each
analyte.
Chromatographic Time Segments Table
Adding or inserting a segment copies all parameters from the previous segment
to the newly created segment. Double-click on the required field to edit the
Segment Description, Start Time, or End time of a segment.
Edit the 4000 MS Method Segments
For advice on performing Internal CI, see the appropriate section in the 4000
GC/MS Software Operation Manual.
25
Page 26
Scan Function Settings
The 4000 MS method segments control what MS scan function is performed. The
different scan functions in the Internal configuration are EI and CI, full scan,
MS/MS, Multiple Reaction Monitoring (MRM), or Selected Ion Storage (SIS).
Scan Function selection
The 4000 MS has three scan modes - normal speed (5,000 u/sec), fast speed
(10,000 u/sec), or fastest (10,000 u/sec without a prescan). The Fastest speed
only allows acquisition with single-segment mass ranges.
Mass Range Tune
50 – 1000 1 Segment 3 0.76 sec 0.47 sec 0.41 sec
50 – 1000 4 Segment 3 1.08 sec 0.79 sec N/A
50 – 400 DFTPP 3 0.70 sec 0.59 sec N/A
μScans
Normal Fast Fastest
General Parameters Tab
Scan Time, Scans Averaged, and Data Rate are all linked. The number of
scans averaged is updated when the scan time is adjusted and vice versa. The
best way to set the scan time is to set the mass range from the Scan Parameters
tab and then change the scans averaged to three. Three scans averaged gives
the best compromise between a high chromatographic data rate and good
spectral averaging.
The Mass Defect allows for a systematic correction of the difference between
the nominal mass of an atom (or ion) and its exact mass. Its importance arises
26
Page 27
from the fact that the NIST library reports molecular weights to the nearest
integer mass unit only. The Varian MS Workstation software must decide to
which mass to assign measured intensity. If the exact mass of an ion happens to
fall close to the dividing line between integer masses, the software may make an
incorrect mass assignment. This scenario is more likely for molecules with higher
molecular weights, since the mass defects for several atoms may add together to
produce a sizable mass defect. For example, the exact mass for the lightest
isotope form of C
is 497.51002, which could be assigned as either 497 or
2Br6
498.
The Multiplier Offset adjusts the EM voltage as by much as ±300V relative to
the current multiplier setting in the Module Attributes tab dialog in Manual Control
(this is usually the 10
5
gain value from Auto Tune). Sometimes better sensitivity
is achieved, particularly in techniques such as MS/MS, when the multiplier
voltage is increased. NOTE that this adjustment can be made on a segment-bysegment basis.
The Count Threshold is normally 1; a value of 2-3 counts will reduce the
number of low-level ions reported in the mass spectrum. This approach may
improve library searches and reduce data file size at the cost of somewhat less
detailed information in the mass spectra. The count threshold is shown only if the
Customize button is active.
Ionization Control
The Target Total Ion Current, or TIC is specified here. The Automatic Gain
Control (AGC) algorithm uses the ion count from a prescan at fixed ion time,
along with this target value, to calculate an ion time necessary to fill the ion trap
with the target number of ions during the analytical scan. The objective is to fill
the trap with an optimal number of ions during each analytical scan. The Target
TIC is usually not set below 10,000 for full scan acquisitions, but it should also
not be set too high or spectral distortions due to space charge may result (loss of
MS resolution and/or shift in mass assignments for strong chromatographic
peaks). Typically, a Target TIC between 20,000 and 40,000 counts gives the best
results.
EI Ionization Parameters
The Emission Current is the current of electrons produced by the filament and it
controls the number of electrons that enter the trap. By increasing the emission
current so that the ionization time of the baseline is approximately the maximum
ionization time, the maximum number of ions will be present. In certain cases, a
high emission current can be used to increase the sensitivity of the analysis. An
example of this is in EI/MS/MS where the MS/MS isolation step eliminates the
higher background that would normally arise from a high emission current.
27
Page 28
Setting Parameters for Chemical Ionization
After selecting a standard reagent (methane, isobutane, acetonitrile, or
methanol), the CI parameters for the selected reagent are set automatically so
you do not need to change the default values of the CI parameters. However, the
remaining CI parameters (Reaction Storage Level, Ejection Amplitude (V), and
Max Reaction Time) can be optimized for sensitivity. After adjusting these
parameters, click Save. You will be prompted for a name under which to save the
modified values.
Reagent Low Mass is the lowest mass stored in the trap during ionization.
Reagent High Mass is the highest mass stored in the trap during ionization. All
masses above this mass are ejected during the ionization step. This reduces the
number of EI generated ions that are in the trap.
The combination of Reagent Low and Reagent High Mass and Reaction Storage Level allows for the selection of a specific reagent ion in the CI process.
For example, only mass 29 of methane could be used as the reagent ion by
ejecting mass 41 during ionization and by reacting at a mass greater than 19 in
the reaction step.
NOTE: When a reagent ion that is formed through a complex process, such as
+
HC formed by the reaction of
52
within the High and Low mass range.
+
CH with neutral4CH , both ions must be
4
Reaction Storage Level is the value of the smallest mass stored in the ion trap
during the reaction period. The optimum Reaction Storage Level will depend on
28
Page 29
the molecular ion of the analyte. Generally, one should use higher Reaction
Storage Levels for higher molecular ions, without raising the storage level to
cause ejection of CI reagent ions. For example, the molecular ion of the analyte
is 352 m/z. Using acetonitrile, raising the RF storage level to 25 m/z may give
better sensitivity than the default of 19 m/z. If the CI storage level is raised, the
Ejection Amplitude will have to be increased.
Ejection Amplitude (V) is a voltage that corresponds to a low mass ejection
cutoff that is slightly higher than the Reagent High Mass. This voltage actively
ejects unwanted ions (i.e., not reagent ions) that are produced during ionization.
Generally, higher CI storage values require higher ejection voltages. The voltage
should not be set so high as to cause the ejection of CI reagent ions.
Max Reaction Time is the maximum time that reagent gas ions are allowed to
react with sample molecules to form ions. The maximum reaction time can be set
to any value from 1 to 2000 milliseconds. The typical reaction time is 100
milliseconds
For details on how to optimize these parameters, or on how to set up for nonstandard reagents, refer to the 4000 GC/MS Software Operation Manual.
NOTE: Be aware that if you do change a CI parameter and set it incorrectly, you
may not be able to operate properly in CI.
Full Scan Parameters
Each MS scan type has different parameters that need to be specified. The
following are examples of the two most common scan types used in the Internal
configuration; Full Scan and MS/MS. Detailed information on these, and the other
scan function parameters can be found in the 4000 GC/MS Software Operation Manual.
Full Scan data acquisition is used for general-purpose GC/MS analysis. In the
Mass Range area (upper left), enter Low Mass and High Mass values to specify
the full scan mass range. This is the most common scan type for the 4000 MS.
The mass range also determines the scan range of the AGC calculation.
The Tune field specifies how the mass range is scanned. There are three
specified tune types; Auto, DFTPP, and BFB. Under the Auto tune type each EI
scan is divided by default into four mass segments: 10 to 99 m/z, 100 to 249 m/z,
250 to 399 m/z and 400 to 1000 m/z. Under these conditions, the RF Storage Level (m/z) and the Ionization Time Factor (%) can be adjusted on a mass
segment basis.
29
Page 30
When DFTPP and BFB tune types are selected, mass segments and ion time
factors, which will be good starting points for meeting US EPA semivolatile and
volatile tuning requirements, are displayed in the mass segment table.
Each mass segment has its own RF Storage Level. This is the RF voltage used
to hold ions in the trap during the ionization period and is specified in mass units.
It affects ion storage in two ways C the storage efficiency of higher mass ions
increases as the level increases, while lower mass ions are not stored if their
mass falls below the cutoff. With AGC on, the default storage level is set to
35 m/z, causing all ions above 35 m/z to be stored. This value gives good
storage efficiency for ions up to 650 m/z. For masses up to 1000 m/z, a storage
level of 45 m/z may be required.
The Ion Time Factor is a number that is multiplied by the calculated ionization
time (determined by the AGC pre-scan calculation) to give the actual ionization
time for each segment of the mass range. The default value is 100%. Adjust this
factor to increase or decrease the relative intensity of any segment in the
acquisition mass range. For example, adjusting four or five segments
appropriately allows the system to pass DFTPP or BFB tune requirements for US
EPA environmental methods.
Setting Parameters for an MS/MS Method Segment
Tandem mass spectrometry, or MS/MS, uses ion preparation steps after the
ionization step and before mass analysis. MS/MS may be performed after either
electron or chemical ionization. Briefly, all ions are eliminated from the stored
mass range except at the m/z of a precursor ion. The precursor ions are then
excited by waveforms applied to the ion trap. When enough energy is deposited
in this way, collisions of precursor ions with helium buffer gas cause dissociation
of the precursor ions to lower mass product ions. The remaining ions are then
scanned to collect an MS/MS spectrum.
When properly designed, an MS/MS method will:
•Fill the ion trap with only the selected precursor ions, so that trap capacity is
used so that in many cases, co-eluting interfering compounds are excluded
from the trap.
•Create product ions via a unique dissociation pathway, eliminating chemical
noise.
NOTE: MS/MS is useful only when the target compounds of an analysis are
known. It is not useful for general qualitative analysis of unknowns except to the
degree one is determining a set of isomers of a given class such as PCBs or
Dioxins.
30
Page 31
The Precursor Ion is usually an intense ion in the full scan mass spectrum.
Usually, the Isolation Window is the parent ion mass ±1.0 (3.0 mass units
wide). Waveform Type is either Resonant or Non-resonant. The Excitation Storage Level is the lowest mass stored during collision-assisted dissociation. A
good value can be calculated using the “q” Calculator at the bottom of the
window. The “q” Calculator sets arbitrary limits to the Excitation Storage Level so
you should be aware that it may calculate a value of 300 when the Precursor Ion
m/z is large. The excitation amplitude needed to dissociate the precursor ion
must be determined experimentally; e.g., using several runs with different ranges
of excitation amplitudes. Using the AMD (Automated Method Development)
mode is the easiest way to determine this voltage. The Product Ion Mass range
during method development encompasses the range from Excitation Storage Level to the Precursor Ion mass. For more detail on MS/MS methods, seethe
Tandem Mass Spectrometry section in the 4000 GC/MS Software Operation
Manual.
Viewing Method Parameters in Manual Control
After a method is created in Method Builder, it can be previewed in Manual
Control. All MS parameters can be edited and previewed before a run. However,
the number of segments, or the start and end times of existing segments, can
only be changed in Method Builder.
Activating a Method
To activate a different method, do the following:
1. Click the File menu.
2. Click Activate Method.
3. Select a method by doing one of the following:
31
• Click Recent Files to display the eight most recent methods.
• Click Open after selecting a method from a folder.
Page 32
4. The active method is displayed in the toolbar.
Displaying Ions
To view ions do the following:
1. Select an ionization segment in which the ionization is on. You cannot turn
on the ion trap in a segment where ionization is OFF as in the Fil/Mul Delay
segment #1. Change to an ionization segment:
→
2. Click the Trap check box to turn on the ion trap.
3. Select the method segment to view. Turn on the calibration gas or CI gas by
32
selecting the check box.
Page 33
Viewing Method Parameters
In the following figure, the Method Segment tab dialog is shown with method
related controls in the lower pane.
The information in the top row of the Method Segment tab indicates if the trap is
on, the Scan Type, Ionization mode, and Scan Mode.
Editing a Method in Manual Control
Examine and edit all the parameters in the active MS method and observe the
changes on the mass spectra being acquired. The exact set of tab dialogs
depends on the ionization and ion preparation modes in the current method
segment.
33
Page 34
After editing a parameter, implement the change by clicking the Activate
Changes button.
The changes are reflected in the spectrum. The example here is a change in the
Start Mass from 50 m/z to 100 m/z:
34
Page 35
Saving a Method
To save changes to the method, click the Upload MS Method button above the
Ion Trap icon. Click Edit Method to open the Method Builder and make other
changes.
If you do not upload changes, the method is still checked to see if changes are
made when the segment is changed or when you leave Manual Control or the Method Segment. If changes have been made, you will be given the option to
save these changes or discard them.
Acquiring Data
NOTE: If you leave System Control by starting automation or choosing Inject
Single Sample, you are prompted to save the method.
Click Yes, the injection operation is cancelled, and you are returned to Manual
Control so you can save the method.
Click No, the last saved copy of the method is used to acquire the data file.
Acquisition
Click StartAcquisition to start a run. If you start an analysis while the
instrument is in another mode, the software automatically shifts the MS module
into Acquisition mode.
35
Page 36
If the GC is not ready, you will see a Not Ready message at the top of the
screen. After the GC and AutoSampler come to a ready state, the Not Ready
message will change to Ready. To determine the individual ready states of the
components, you can see the top pull down menu under Windows and see the
states for the 4000, 3800, and Combi PAL modules. After all components are
ready, you can start an analysis.
An analysis can be run as a single sample or through an automated sequence.
•To run a single sample, see “Injecting a Single Sample” on page 36.
To run in automation mode, see “Injecting Using a SampleList” on page
•
38.
Status and Control
Before an acquisition has been started, the Status and Control field is similar to
the following figure.
• The Run Time is 0.00 minutes.
• The End Time is the run length specified in the active method.
• The Ready and No Fault lights are green.
Click the Start Acquisition button to override automation and start a run before
the system becomes Ready. However, the file name of a run started in this way
is named as 4000.x.sms, not the file name specified for automation runs.
Click the Edit Method button to open the Method Builder and modify the method.
You are prompted to re-activate the method after saving changes and are
returned to System Control.
NOTE: A change in the End Time for the MS module does not change the GC
End Time. You must access the GC module from the Windows pull-down menu
and change the GC End Time separately.
36
Page 37
Activating a Method
To activate a GC/MS method, do the following:
1. Click the File menu.
2. Click Activate Method.
3. Select a method by doing one of the following:
• Click Recent Files to display the eight most recent methods.
• Click Open, after selecting a method from a folder.
Injecting a Single Sample
To inject a single sample from System Control using Inject Single Sample do the
following:
1. Click Inject Single Sample, from the Inject menu.
2. After the Inject Single Sample window opens, do the following:
a. Type a sample name.
b. Enter the vial number of the sample vial if an autosampler is configured.
c. Check that the injection volume and injector used are correct.
d. Click Defaults, to change the default values for any parameter.
e. Click Data Files to create a name that includes more information such as
date and time, or to change the directory for data file storage.
f. Click Inject to acquire the data.
37
Page 38
•If the MS is not in Acquisition mode, it changes to that mode
automatically.
•If an AutoSampler is doing the injection, it begins after the instrument
modules are Ready.
• If you are doing a manual injection, wait until the System Control title bar
reads “Waiting for Injection of Sample” and there is a blinking yellow
Waiting light on the right of the System Control toolbar. Then inject the
sample.
Injecting Using a SampleList
You can create and edit a SampleList in the Automation File Editor or in System
Control.
To edit a SampleList and inject multiple samples from System Control do the
following:
1. Click either New SampleList or Open SampleList from the File menu.
2. The SampleList window for the open SampleList opens. It contains fields that
are specific to the configured autosampler. See the following figure.
•Change the size of the spreadsheet columns by dragging their
border with the left button of the mouse.
•Right-click a column header for formatting options. When the table is
scrolled to the right, the Sample Name column does not scroll so you
can easily tell for which sample you are entering additional
parameters.
• Click Add to add additional samples. Enter the name, sample type,
and vial number for all samples.
3. Click the Begin button in the lower left corner to start analyzing the samples
on the SampleList.
38
Page 39
Monitoring the Status of Runs
Monitor the status of the run in the instrument window. The Status and Control
window and the Toolbar show the run status.
Monitor the chromatogram and spectra in System Control, or click the far right
button in the Chromatogram toolbar to open MS Data Review, where you can
perform operations like library searching while the data file is being acquired.
For more information on data acquisition features, see the Acquiring GC/MS
Data section in the 4000 GC/MS Software Operation Manual.
39
Loading...