Varex 4336Wv4 User Manual

®

PaxScan 4336Wv4 X-ray Detector

Before using the X-ray detector, be sure to read this manual thoroughly along with any other manuals for the software and other system components. Keep this manual where it is easily accessible.
Before You Begin
To avoid personal injury or product damage, read the manual and all accompanying papers
The X-ray detector is intended for use by qualified professional personnel who are trained
Install the X-ray Detector horizontally on a flat, stable surface. It is required to use a bucky or
The user is responsible for using and maintaining the detector according to prescribed
X-ray imaging, processing, image acquisition, and data storage must be performed in
In no event is the X-ray detector manufacturer liable for direct, indirect, or consequential
USER MANUAL
carefully before operating the PaxScan 4336Wv4 system.
and knowledgeable in the use of X-ray detectors, X-ray systems, and electrical equipment.
a wall-mount for vertical or tilted positions. The X-ray Detector may cause an injury if it falls or is dropped.
installation, usage, maintenance, handling, and storage specifications. To keep the detector and its accessories in safe and proper condition, only trained and qualified person(s) shall be in charge of maintenance.
accordance with all applicable laws. The user is also responsible for compliance to laws pertaining to the privacy of image data.
injury, damage, or loss of equipment operation time or image data arising from the use of the X-ray detector, its components, and/or accessories.
Protection Against Ionizing Radiation
Exposure of any part of the human body to X-ray radiation may be harmful to health.
Whenever X-ray equipment, ionizing, or radioactive sources are in use, appropriate safety precautions and measures shall be instituted, and all regulatory requirements must be met. It is the responsibility of the X-ray system installer, operator, and user to comply with applicable requirements.
The X-ray detector does not contain a primary barrier for X-rays or Gamma rays. The X-ray
system installer or X-ray system manufacturer must provide the necessary protection based on the X-ray system’s intended use.
For portable applications, the X-ray system installer or X-ray system manufacturer must
provide the necessary training for operators to protect themselves, patients, and surrounding persons.
ii www.vareximaging.com
4336Wv4 X-ray Detector
For Your Safety
To avoid personal injury or product damage, read this manual and all accompanying information carefully before handling, installing, or using the X-ray detector. Follow all instructions, warnings, and cautions in this manual and all warnings and cautions printed on the warning label. Ignoring instructions, warnings, or cautions in the handling, installing, or use of the detector may result in personal injury, death, or product damage. Keep this manual for future reference.
Meaning of Alerts and Notes
USER MANUAL
DANGER
WARNING
Caution
Note
This indicates a potentially hazardous situation which, if ignored, will result in severe personal injury, death, or substantial product damage.
This indicates a potentially hazardous situation which, if ignored, may result in severe personal injury, death, or substantial product damage.
This indicates a potential hazardous situation which, if ignored, may result in minor or moderate personal injury or damage to the product.
This emphasizes or supplements important information about the main text.
www.vareximaging.com iii
Installation and Environment of Use
USER MANUAL
WARNING
WARNING
WARNING
WARNING
The X-ray detector is intended to be installed, maintained, and used by qualified professional personnel who are trained and qualified in the installation, maintenance, and use of X-ray equipment. All parts of the detector are suitable for use within the patient environment. The proximity of the detector to the patient is dependent upon the application by the system integrator.
Do not operate the X-ray detector in or around flammable gases, gas mixtures, liquids, chemicals, or other substances. Ignoring this warning may result in explosion, fire, or electric shock, which may result in severe personal injury, death, or substantial product damage.
Do not connect the X-ray detector to any component or accessory other than the manufacturer’s specified components and accessories. Ignoring this warning may result in explosion, fire, or electric shock, which may result in severe personal injury, death, or substantial product damage.
Do not modify or alter the X-ray detector, its components, or accessories. Ignoring this warning may result in explosion, fire, or electric shock, which may result in severe personal injury, death, or substantial product damage.
WARNING
WARNING
WARNING
WARNING
WARNING
Caution
The detector is not designed to control X-ray dose. The system integrator is responsible for controlling the X-ray radiation.
The detector is not intended to be used as a primary barrier to X-rays. The user is responsible for ensuring the safety of the operator, bystanders, and the subjects being radiographed.
The detector is not suitable for use in the presence of or in combination with active implanted devices such as defibrillators and pacemakers. Doing so may prevent normal operation of these peripherals.
For portable applications, the operator and end-user must take precautions to protect themselves against dangerous X-ray exposure when using the detector in the X-ray beam path.
Portable RF communications equipment should be used no closer than 30cm from the product or degradation of performance could result.
It is important that the detector is not directly connected to the installed network. Connection of the detector with the installed computer network may disturb the IT environment.
iv www.vareximaging.com
4336Wv4 X-ray Detector
USER MANUAL
Caution
Caution
Do not operate the X-ray detector in a location with the following conditions.
Close to fluid or places where fluid is used
Close to heat sources, such as a heater
High temperature environment
High humidity environment
High condensation environment
Extreme cold environment
Dusty environment
Salty or sulfurous environment
Near a vibrating environment
Environment where there is insufficient air circulation to enable the X-ray
detector and power supply to dissipate heat
Ignoring this caution may result in personal injury or damage to the product.
The detector is an applied part (patient contact device) and the surface shall not exceed 42°C internal temperature sensor data is provided in the diagnostic data attached to each image. These temperature measurements are well correlated with the panel external surface temperature. It is advisable to monitor this diagnostic data as an additional safety precaution, see document FP2032 - VSP Software Interface.
Caution
Caution
Monitor the temperature of the X-ray detector, and allow the detector to cool down when the temperature of the detector gets hot. The device is not designed to supply heat to a patient. Ignoring this caution may result in personal injury or damage to the product.
An Image Quality Test or Customer Acceptance test should be performed before the detector is used for the first time with patients.
www.vareximaging.com v
Service Cable
Handling
USER MANUAL
WARNING
WARNING
WARNING
Caution
Disconnect the Service Cable by pulling on the connector and not the cable itself. Ignoring this warning may cause substantial product damage.
Do not modify the Service Cable or subject the cable to external stress or damage. Avoid placing anything heavy, including the detector, on the cable, stepping on the cable, pulling the cable, or subjecting the cable to excessive bending or bundling. Ignoring this warning may cause cable failure resulting in substantial product damage.
Never disassemble, modify, or alter the X-ray detector, its components, battery pack, battery charger, or accessories. Ignoring this warning may cause electrical shock and/or unknown hazards, which may result in severe personal injury, death, or substantial product damage.
Place the X-ray detector horizontally on a flat, stable surface. If the X-ray detector is placed vertically or in any tilted position, the X-ray detector must be securely placed in the Bucky tray or securely fastened to the X-ray detector enclosure or support structure. Ignoring this caution may result in personal injury or damage to the product.
Caution
Caution
Caution
Caution
Do not exceed the maximum uniform load weight of 150 kg distributed across the surface of the X-ray detector.
Do not exceed the maximum load weight of 100 kg distributed on an area of 40 mm in a diameter of the X-ray detector surface.
Do not drop the X-ray detector. If the X-ray detector is dropped, remove the X-ray detector from service, and immediately ask your establishment’s safety representative to verify or re-validate the proper function of the X-ray detector prior to resuming use of the detector. Further use under abnormal conditions may result in severe personal injury, death, or substantial product damage.
Do not expose the backside of the detector with X-rays. Always use the top side of the detector for examinations.
vi www.vareximaging.com
4336Wv4 X-ray Detector
Battery
USER MANUAL
WARNING
WARNING
WARNING
Do not use the battery pack if the casing is broken or if it emits an unusual odor, smoke, or excessive heat, or if it leaks any substance. Avoid contact with any substance seeping from the battery pack. If any fluid touches your skin or eyes, wash the affected area with clean, running water and immediately seek medical attention.
The cells within the battery pack contain toxic substances. Do not attempt to open the battery packs. Do not insert any object into the battery pack or use any device to pry at the battery pack casing. Attempting to open the battery pack casing will damage the casing, which could cause the battery pack to release toxic and harmful substances causing injuries such as electric shock or burns, or cause a fire, and will render the pack unusable.
Be sure to remove the battery before servicing, maintaining, connecting, or disconnecting the accessories.
Do not touch the battery pack, X-ray detector, cable, connector, or any other electrical component or equipment with wet hands. Ignoring this warning may cause electrical shock, which may result in severe personal injury, death, or substantial product damage.
WARNING
WARNING
WARNING
WARNING
WARNING
WARNING
WARNING
Do not insert the battery into the X-ray detector when condensation is on the X-ray detector or any of its components or accessories. Ignoring this warning may cause electrical shock, which may result in severe personal injury, death, or substantial product damage.
Observe and follow all safety information in this manual and on the warning label found on the battery pack. Ignoring a warning may result in personal injury or damage to the product.
Use only charging devices approved by device manufacturer, and never attempt to bypass or override their charging protection circuits.
Keep out of reach of children.
Remove the battery pack from the detector if the X-ray detector is not likely to be used for some time.
Do not submerge the battery pack in water or other liquid.
Do not charge the battery pack near flammable materials.
www.vareximaging.com vii
WLAN
USER MANUAL
WARNING
WARNING
Caution
WARNING
WARNING
WARNING
Do not connect the battery pack to an electrical outlet directly, or to any other electrical source not described in the manual.
Do not drop or hit the battery against hard objects since this may cause damage to the battery pack and risk release of the battery toxic and harmful substances, causing injuries such as electric shock or burns or causing a fire, and will render the battery pack unusable.
There is a risk of explosion, personal injury, or damage to product if the battery pack is replaced by non-OEM approved components.
Do not obstruct the detector antenna. If it is obstructed by metal, wood, or a human body, the wireless communication can be slowed down or disconnected.
Do not route Ethernet cables near or with power cables, degradation of performance could result.
Follow the laws and regulations for each country, and select the regional code accordingly.
WARNING
WARNING
Caution
Caution
Caution
Do not use the detector in aircraft because there is a potential affect to aviation systems.
Do not modify or alter the detector as this can violate the certification of the Radio Law.
Use WLAN access point devices to get the best communication performance.
The electromagnetic emission of the detector may influence implantable medical devices like pacemakers. Check the information for these devices.
Use a Wi-Fi friendly environment and avoid Bluetooth devices, mobile phones, and other Wi-Fi devices close to the detector or router.
viii www.vareximaging.com
4336Wv4 X-ray Detector
vTrigger Mode
USER MANUAL
WARNING
WARNING
WARNING
WARNING
The vTrigger mode requires a sufficient X-ray dose rate to the detector surface to trigger image acquisition. The required dose rate can vary at different use conditions. The system integrator must evaluate the vTrigger operation with the complete X-ray setup in order to secure the proper image acquisition in vTrigger mode.
The external sync mode is the default acquisition mode, and the detector needs to switch into the vTrigger mode to use. An image may only be acquired in vTrigger mode once the software has informed the user that the system is ready.
The vTrigger on-time is limited and should be turned off directly after acquiring the image. Staying longer in vTrigger mode will reduce the battery duty cycle.
Do not apply any, handling, loading, mechanical shock, or electronic noise to the detector while it is in vTrigger mode.
If a Problem Occurs
WARNING
WARNING
If any abnormal condition, such as smoke, fumes, or strange sounds, is evident, remove the battery from the X-ray detector, and immediately ask your establishment’s safety representative to contact your dealer, distributor, or device manufacturer.
Further use under abnormal conditions may result in severe personal injury, death, or substantial product damage.
When liquid has been spilled into or on any part of the X-ray detector or battery, or when the X-ray detector, its component, or accessory is dropped, immediately ask your establishment’s safety representative to contact your dealer, distributor, or device manufacturer.
Further use under abnormal conditions may result in severe personal injury, death, or substantial product damage.
www.vareximaging.com ix
Maintenance and Inspection
USER MANUAL
WARNING
WARNING
WARNING
Caution
Turn off the power of the X-ray detector when the inspections indicated in this manual are going to be performed. Ignoring this warning may result in electric shock, which may result in severe personal injury, death, or substantial product damage.
When the X-ray detector system is going to be cleaned, remove the battery pack. Never use thinner, benzine, acetone, or other flammable cleaning agents. Ignoring this warning may result in explosion, fire, or electric shock, which may result in severe personal injury, death, or substantial product damage.
The X-ray detector must be repaired by X-ray detector manufacturer-authorized personnel only. Ignoring this warning may result in explosion, fire, electric shock, or unknown hazards, which may result in severe personal injury, death, or substantial product damage.
Follow the manufacturer’s recommendations for inspecting the X-ray detector before use.
x www.vareximaging.com

Table of Contents

1.0 Scope . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
2.0 Contraindication . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
3.0 Intended Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
4.0 Audience. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
5.0 Abbreviations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
6.0 Definition of Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
7.0 Standards and Regulations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
8.0 Description of the PaxScan 4336Wv4 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
8.1 Shipment Contents. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
8.1.1 Supporting Files. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
8.2 Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
8.3 Detector Surfaces and Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
8.4 Detector Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
8.5 Detector Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
8.6 Environmental Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
8.7 Battery Installation and Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
8.7.1 Battery Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
8.7.2 Battery Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
8.7.3 Battery Charge Level. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
8.8 Power on Sequence . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
USER MANUAL
9.0 Modes of Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
9.1 Radiography Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
9.2 vTrigger Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
9.3 Operational States . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
10.0 System Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
10.1 X-ray Generator Interface Signals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
10.2 VSP_COMM (SDK) Files. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
10.2.1 Bonjour Installation Files . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
10.2.2 Sample Code . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
10.2.3 Utility Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
10.3 Detector Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
www.vareximaging.com xi
Table of Contents 4336Wv4 X-ray Detector
11.0 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
USER MANUAL
10.3.1 Offset Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
10.3.2 Offset Calibration Steps. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
10.3.3 Gain Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
10.3.4 Radiographic Mode Gain Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
10.3.5 Radiography Mode Gain Calibration Steps . . . . . . . . . . . . . . . . . . . . . . . 31
10.4 Link to the Detector. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
10.5 ViVA Mode Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
10.6 Image Acquisition . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
10.6.1 Radiography Mode. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
10.6.2 Radiographic Acquisition Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
10.7 VSP Control Panel. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
10.7.1 Access the VSP Control Panel. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
10.7.2 Modify Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
10.7.3 Change Password . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
10.7.4 Firmware Update . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
11.1 Cleaning and Disinfecting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
11.1.1 Chemicals for Cleaning and Disinfecting . . . . . . . . . . . . . . . . . . . . . . . . . 50
11.1.2 Cleaning the X-ray Detector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
11.1.3 Disinfecting the X-ray Detector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
12.0 Trouble-Shooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
12.1 After-sales Service for Varex Imaging Products. . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
13.0 Disposing of the X-ray Detector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
14.0 Safety. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
14.1 Electromagnetic Interference. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
14.1.1 Electromagnetic Emissions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
14.1.2 Electromagnetic Immunity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
14.1.3 RF Immunity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
14.2 Radio Frequency (RF) Compliance Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
14.2.1 FCC/IC Compliance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
14.2.2 Restrictions for 2.4GHz Use in Europe . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
15.0 Regulatory. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
15.1 Industry Canada Notice . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
15.2 Declaration of Conformity for European Union. . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
15.3 International Compliance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
15.3.1 Argentina Compliance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
15.3.2 Mexico Compliance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
15.3.3 Nigeria Compliance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
xii www.vareximaging.com
4336Wv4 X-ray Detector Table of Contents
15.3.4 Philippines Compliance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
15.3.5 UAE Compliance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
15.3.6 Singapore Compliance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
15.3.7 Thailand Statement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
15.3.8 Taiwan NCC Warning Statement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
USER MANUAL
www.vareximaging.com xiii
Table of Contents 4336Wv4 X-ray Detector
This page intentionally left blank.
xiv www.vareximaging.com

List of Tables

Table 1 Abbreviations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Table 2 Explanation of Symbols. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Table 3 Standards and Regulations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Table 4 Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Table 5 Detector Surfaces and Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Table 6 X-ray Detector Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Table 7 Environmental Conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Table 8 LED Status Details . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Table 9 Operational Mode Example . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Table 10 Gain Calibration - All Modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
USER MANUAL
Table 11 Types of Images Detector will Acquire in Radiographic and vTrigger Modes . . 39
Table 12 Problems and Solutions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Table 13 Radiated/Conducted Emissions, Harmonics, Voltage, Fluctuations & Flicker . . 58
Table 14 ESD, Transient/Burst, Surge, Voltage Variation, Magnetic Fields. . . . . . . . . . . . . 59
Table 15 Conducted/Radiated RF. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
Table 16 Europe - Restrictions for Use of 2.4GHZ Frequency in European Community . . 63 Table 17 Europe - Declaration of Conformity in Languages of European Community . . . 66
www.vareximaging.com xv
List of Tables 4336Wv4 X-ray Detector
This page intentionally left blank.
xvi www.vareximaging.com

List of Figures

Figure 1 Typical Detector Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Figure 2 Imaging System Overview - without Router . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Figure 3 Imaging System Overview - with Router . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Figure 4 Imaging System Overview – Service Cable Connection. . . . . . . . . . . . . . . . . . . . . . . 7
Figure 5 Detector Surfaces and Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Figure 6 Detector Dimensions (measurements are in mm) . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Figure 7 Representation of Battery Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Figure 8 Unlatch Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Figure 9 Battery Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
USER MANUAL
Figure 10 Battery Charge Level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Figure 11 LED Status Indicator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Figure 12 Exposure: Corrected Frames. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Figure 13 Exposure: Raw Frames . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Figure 14 Mode Drop Down Menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Figure 15 Initiate Offset Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Figure 16 Calibration Setting window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Figure 17 Offset Calibration Progress. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Figure 18 Finish Offset Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Figure 19 Mode Drop Down Menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Figure 20 Initiate Gain Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Figure 21 Gain Calibration Progress window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Figure 22 Gain Calibration Updating . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Figure 23 Gain Calibration Complete . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Figure 24 The ViVA Icon. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Figure 25 Detector Linked in ViVA. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Figure 26 Mode Drop Down Menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Figure 27 Hardware Handshaking Not Selected . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Figure 28 Hardware Handshaking Selected . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Figure 29 Open System Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Figure 30 System Settings window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
www.vareximaging.com xvii
List of Figures 4336Wv4 X-ray Detector
Figure 31 Mode Drop Down Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Figure 32 Acquire Image Button. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Figure 33 Acquire Image . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Figure 34 Start Acquisition . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Figure 35 Image Acquisition Progress . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Figure 36 Acquisition Type Drop Down Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Figure 37 Radiographic Acquisition Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Figure 38 Radiographic Acquisition Settings window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Figure 39 VSP Control Panel Access Authentication. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
Figure 40 VSP Control Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Figure 41 Modify VSP Control Panel Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Figure 42 Change Password . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
Figure 43 Updating the Firmware . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
Figure 44 Firmware Updated . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
USER MANUAL
www.vareximaging.com xviii
4336Wv4 X-ray Detector

1.0 Scope

This document describes design elements and respective interfaces for the PaxScan 4336Wv4 X-ray detector (Version 2). All applicable mechanical, electronic, and software interfaces are described.

2.0 Contraindication

There are no contraindication situations.

3.0 Intended Use

Varex Imaging digital X-ray detectors and their accessories are components designed to be integrated into products by X-ray system manufacturers. Manufacturers are responsible for qualifying, validating, and certifying their products for their intended uses and meeting all applicable regulatory requirements.
Final application and intended use is determined by the X-ray system manufacturer and is based on the completed X-ray system design. It is the responsibility of the X-ray system manufacturer to confirm safety, efficacy and compliance of the X-ray system intended for use, inclusive of the detector. The Digital Radiography Software referred to in this manual is intended to be used for development, testing, and maintenance purposes only.
USER MANUAL
Note
The system integrator is responsible for obtaining FDA clearance for medical use.

4.0 Audience

This document is for users of the X-ray detector and for X-ray system manufacturers and X-ray system installers who are responsible for installing the X-ray detector into an X-ray system.
www.vareximaging.com 1

5.0 Abbreviations

Table 1 Abbreviations
Abbreviations Descriptions
API Application Program Interface
a-si Amorphous Silicon
fps Frames per second
hPa Hectopascal
OEM Original Equipment Manufacturer
PREP Prepare
REQ Request
SDK Software Developer Kit
ViVA Varex Imaging and Viewing Application
VSP Varex Smart Panel
4336Wv4 X-ray Detector
USER MANUAL
vTrigger Automatic Exposure Detection
WLAN Wireless Local Area Network

6.0 Definition of Symbols

Table 2 Explanation of Symbols
Symbol Description
On (power connection)
Off (power disconnection)
Handle with Care
Direct Current
Authorized Representative in the European Community/European Union
IP54 Protected from limited dust ingress.
Protected from water spray in any direction.
2 www.vareximaging.com
4336Wv4 X-ray Detector
Table 2 Explanation of Symbols (Continued)
Symbol Description
Consult Instruction for Use
Load Weight Restriction
Do Not Discard with Domestic Waste
European Union Mark of Conformity to Applicable European Directives
Manufacturer
Non-ionizing radiation
USER MANUAL
Type B Applied Part
Underwriters Laboratory Safety Mark
China RoHS environmentally friendly for 10 years
www.vareximaging.com 3
4336Wv4 X-ray Detector

7.0 Standards and Regulations

The assembly of the X-ray Detector PaxScan 4336Wv4 is designed to be compliant with the requirements detailed in Table 3 below. All regulatory certificates are valid only if the original accessories are used. All regulatory certificates are rendered invalid if any modifications to the Product are made, or any portion thereof, without obtaining the prior written authorization of Varex Imaging.
.
Table 3 Standards and Regulations
Detector Standard Description
UL 60601-1 Medical Electrical Equipment, Part 1: General Requirements for Safety 1st ed.
IEC 60601-1 Medical Electrical Equipment Part 1: General Requirements for Safety 2nd ed.
EN/IEC 60601-1 2006+A1:2013/ 2005+A1:2012 (ED. 3.1)
ANSI/AAMI ES60601-1 (2005)
CSA-C22.2 No 60601-1 (2008)
USER MANUAL
CAN/CSA-C22.2 No
601.1-M90, 2005
EN/IEC 60601-1-2 Medical Electrical Equipment Part 1-2: General Requirements for Basic Safety
93/42/EEC European Union Medical Device Directive
2014/30/EC European Union Electromagnetic Compatibility Directive
2014/53/EU European Union Radio Equipment Directive
FCC Part 15 Subpart C and Subpart E
EN 300 328 Wideband Transmission Systems: 2.4GHz
EN 301 893 Wideband Transmission Systems: 5GHz
EN 301 489-1 Radio Equipment - Electromagnetic Compatibility
EN 301 489-17 Broadband Data Transmission Systems
Medical Electrical Equipment – Part 1: General Requirements for Basic Safety and Essential Performance.
Medical Electrical Equipment, Part 1 General Requirements for Basic Safety and Essential Performance.
Medical Electrical Equipment, Part 1 General Requirements for Safety.
and Essential Performance Collateral Standard: Electromagnetic Compatibility 4th ed.
Telecommunication - Intentional Radiators and Unlicensed Devices
MDD Class IIa
A Declaration of Conformity has been filed for this product and available upon request by contacting Varex Imaging.
4 www.vareximaging.com
4336Wv4 X-ray Detector

8.0 Description of the PaxScan 4336Wv4

The PaxScan 4336Wv4 detector is a light-weight, wireless detector designed for medical and veterinary use. The detector together with the Varex Smart Panel (VSP) software is designed for integration into an X-ray system. The VSP performs all the interface functions with the detector; such as control, image acquisition and calibration.
The PaxScan 4336Wv4 detector fits standard bucky trays and the wireless communication enables easy migration between table, above the table, chest stand, and mobile cart physical applications. The wireless access point is the interface between the detector and the imaging system and may be mounted in an equipment enclosure, or it may also be wall or ceiling mounted to maximize wireless signal strength. During operation, the detector is often draped or bagged to ensure cleanliness and sterilization, and is manipulated such that the detector’s input window is located near, but on the opposite side of the patient, from the X-ray source. An additional cable is supplied with the detector to allow for set-up of the wireless interface and to retrieve images from the detector in case of failed wireless transmission.
Figure 1 shows the configuration of the detector in the context of the typical overall imaging
system.
USER MANUAL
Figure 1 Typical Detector Configuration
www.vareximaging.com 5
4336Wv4 X-ray Detector
USER MANUAL
Note
Note
For proper operation, detector antennas and access point antennas should be unobstructed.
Access point should not be installed next to power supply or generator equipment.
Figure 2 Imaging System Overview - without Router
6 www.vareximaging.com
4336Wv4 X-ray Detector
USER MANUAL
Figure 3 Imaging System Overview - with Router
Figure 4 Imaging System Overview – Service Cable Connection
www.vareximaging.com 7
4336Wv4 X-ray Detector

8.1 Shipment Contents

Upon receipt, inspect the shipment and its contents against the Delivery Note enclosed with the shipment for evidence of damage or missing components. Save all shipping containers in case a return is warranted. If there is any discrepancy, See “After-sales Service for Varex Imaging Prod­ucts” on page 56.
USER MANUAL
Flat Panel Detector Assembly (includes a service cable for image recovery)
Ethernet Service Cable, 6 ft., see Table 4 for a list of accessories.
Note
Note
There is one (1) cable connection for the detector which is the Service Cable. This cable functions as an interface between the detector and the workstation by providing a 100T Ethernet connection for set-up of the wireless interface and as a last resort for retrieval of images from the detector in case of wireless transmission failure.
The Service Cable is not for image acquisition.

8.1.1 Supporting Files

This section describes the supporting files that are delivered with every detector.
User Manual
X-ray Detector Files (files specific to the detector in the shipment)
Software Package
Note
Supporting files may be physically or electronically delivered. If no physical files are present in the shipment, please contact your Varex representative for instructions to download these files.
8 www.vareximaging.com
4336Wv4 X-ray Detector

8.2 Accessories

The PaxScan 4336Wv4 shall only be used with its approved Varex Imaging accessories and replacement parts. Product certification and warranty are void if any modifications to the product is made, or any instruction, warning, or caution is not followed.
Table 4 Accessories
Description
Varex 1-bay Battery Charger
Varex 1-bay Battery Charger Power Supply, 65W, 19V (3.3 ft/1m)
1-bay Battery Charger Mounting Plate
Varex 3-bay Battery Charger
Varex 3-bay Battery Charger Power Supply 19V, 85W (3.3 ft/1m)
3-bay Battery Charger Mounting Plate
Varex Wireless Battery
Battery Overlay LED Display (Silver, Pd Silver, Black, Pure White, Var101 White)
USER MANUAL
Mains Hospital Grade 110V Cable (6.5 ft/2.0m)
Mains European Cable (8.2 ft/2.5m)
Mains Chinese Cable (8.2 ft/2.5)
Upon receipt, inspect the shipment and its contents against the Delivery Note enclosed with the shipment for evidence of damage or missing components. Save all shipping containers in case a return is warranted. If there is any discrepancy, See “After-sales Service for Varex Imaging Prod­ucts” on page 56.
www.vareximaging.com 9

8.3 Detector Surfaces and Features

4336Wv4 X-ray Detector
USER MANUAL
Figure 5 Detector Surfaces and Features
Table 5 Detector Surfaces and Features
1Handles
2 Battery Alignment Marker
3 Battery and Battery Well
Number Description
4 Battery Latch
5 Patient Contact Surfaces
6 Antenna
7 Antenna
10 www.vareximaging.com
4336Wv4 X-ray Detector
Table 5 Detector Surfaces and Features (Continued)
Number Description
8 LED Status Indicator
9 Service Cable Connection
10 Orientation Mark

8.4 Detector Dimensions

USER MANUAL
Figure 6 Detector Dimensions (measurements are in mm)
www.vareximaging.com 11

8.5 Detector Specifications

Table 6 X-ray Detector Specifications
Sensor
Detector Amorphous Silicon active TFT/PIN diode Technology
Scintillator Direct-Deposit CSI or DRZ+
Pixel Matrix 3072 (v) x 2476 (h)
Pixel Pitch 139 µm
Active Area 3062 (v) x 2466 (h) DRZ+, 3052 (v) x 2456 (h) CSI
Electronics
4336Wv4 X-ray Detector
Minimum Computer Requirements
Power Consumption
USER MANUAL
Battery Lithium-ion
Battery Charger 1 or 3 Bay
ADC 16-bit
Image Acquisition Time @550ms 5.6 Seconds (MSR2, RCT)
X-ray Window 350-3500ms
Mechanical
Housing Aluminum with Carbon Fiber entrance window
Weight (with Battery) DRZ+ 2.9 kg (6.39 lbs), CSI 3.0 kg (6.61 lbs)
Load Support 100 kg over diameter 40 mm at center, 150 kg entire surface
Surface Temperature Rated to not exceed 42°C
RAM - 2.00 GB
CPU - 1GHz or faster processor (32-bit or 64-bit)
Idle - 4.3 watts
Acquisition - 8.1 watts
Image Transfer - 10.0 watts
Wireless Communication
Signal Strength Require > -80 dBm or no image will be acquired
Standard 802.11 ac/b/g/n, 2 x 2 MIMO
Chipset Realtek RTL8812AU
Bus Interface USB
12 www.vareximaging.com
4336Wv4 X-ray Detector
Table 6 X-ray Detector Specifications (Continued)
USER MANUAL
Security
WEP
Wi-Fi Protected Access (WPA)
WPA2
TKIP
AES
Operating Voltage 5V +
Radio
Antenna 2 x U.FL
Frequencies
5%
2.412 - 2.484 GHz (Subject to Local Regulations)
5.15 - 5.35 GHz
5.47 - 5.825 GHz
Modulation
802.11b: DSSS (DBPSK, DQPSK, CCK)
802.11a/g: OFDM (BPSK, QPSK, 16-QAM, 64-QAM)
802.11n: OFDM (BPSK, QPSK, 16-QAM, 64-QAM)
802.11ac: OFDM (BPSK, QPSK, 16-QAM, 64-QAM,
256-QAM)
Transmit Power
Receive Sensitivity
802.11a/g: 14 + 1.5dBm@54Mbps
802.11b: 17 + 1.5dBm@11Mbps
802.11gn HT20: 14 + 1.5dBm@MCS7
802.11gn HT40: 14 + 1.5dBM@MCS7
802.11an HT20: 14 + 1.5dBM@MCS7
802.11an HT40: 14 + 1.5dBM@MCS7
802.11ac HT20: 12 + 1.5dBM@MCS8
802.11ac HT40: 12 + 1.5dBM@MCS9
802.11ac HT80: 12 + 1.5dBM@MCS9
802.11b: -72dBm + 2dBm@11Mbps
802.11a/g: -65dBm + 2dBm@54Mbps
802.11n: -64dBm + 2dBm@HT20 MCS7
802.11n: -61dBm + 2dBm@HT40 MCS7
802.11ac: -57dBm + 2dBm@HT20 MCS8
802.11ac: -54dBm + 2dBm@HT40 MCS9
802.11ac: -51dBm + 2dBm@HT80 MCS9
www.vareximaging.com 13
Table 6 X-ray Detector Specifications (Continued)
Image Performance
4336Wv4 X-ray Detector
MTF
DQE
Max Linear Dose DRZ+ 84 µGy, CSI 58 µGy
NED DRZ+ 0.56 µGy, CSI 0.36 µGy
USER MANUAL
Limiting Resolution 3.6 lp/mm
Radiographic Lag 0.5% Typical
Energy Range 40kVp to 150kVp
Sensitivity DRZ+ 0.60 LSB/nGy, CSI 0.86 LSB/nGy
DRZ+ 56%, CSI 57% (1 lp/mm)
DRZ+ 24%, CSI 28%(2 lp/mm)
DRZ+ 12%, CSI 16% (3 lp/mm)
DRZ+ 7%, CSI 11% (Nyquist frequency)
DRZ+ 39%, CSI 78% (0 lp/mm)
DRZ+ 28%, CSI 58% (1 lp/mm)
DRZ+ 18%, CSI 42% (2 lp/mm)
DRZ+ 8%, CSI 24% (3 lp/mm)
DRZ+ 4%, CSI 14% (Nyquist Frequency)
Pixel Noise (1000ms) DRZ+ 9.2 LSB, CSI 8.7 LSB
Memory Effect DRZ+ 0.002 (@60 sec), CSI 0.006 (@ 60 sec)
14 www.vareximaging.com
4336Wv4 X-ray Detector

8.6 Environmental Considerations

Environments outside the specification reduce the lifetime and may irreparably damage the Product.
Table 7 Environmental Conditions
Category Limits
Storage & Transport (ambient) -20º C to +70º C
Storage Humidity Range (non-condensing) Detector: 10% to 90%
USER MANUAL
Normal Operation Temperature (measured at the center of the back cover)
Operation Humidity (non-condensing) 10% to 90%
Atmospheric Pressure Range 700hPa to 1060hPa
Detector Altitude
10º C to 35º C
Operates at 3000m
www.vareximaging.com 15

8.7 Battery Installation and Removal

This section describes installing and removing the battery.
4336Wv4 X-ray Detector

8.7.1 Battery Installation

1 Insert battery at a slight angle so that the side with contacts sits over the adjoining contacts in
USER MANUAL
2 Lay the battery down, one side will be slightly lifted. 3 Press down on the lifted side of battery snapping it into place in the battery compartment. 4 The detector is now ready for user.
Note
the battery compartment.
Note
For additional information about the Varex Imaging Wireless Battery and Chargers, visit www.vareximaging.com.
When inserting the battery the angle of the battery should not be more than 20 degrees, inserting a battery at a larger angle could cause damage to the battery contact pins.
Figure 7 Representation of Battery Installation
16 www.vareximaging.com
4336Wv4 X-ray Detector

8.7.2 Battery Removal

1 Slide the battery latch button to the right until there is an audible click and red is visible.
Doing this will cause one side of the battery to lift out of the battery compartment.
2 Grab the lifted side of the battery and finish removing.
Figure 8 Unlatch Battery
USER MANUAL
Figure 9 Battery Removal
www.vareximaging.com 17
4336Wv4 X-ray Detector

8.7.3 Battery Charge Level

The battery charge level is indicated in the Radiographic Acquisition Progress window.
Figure 10 Battery Charge Level
USER MANUAL
18 www.vareximaging.com
4336Wv4 X-ray Detector

8.8 Power on Sequence

1 Insert the battery into the charger to wake-up the battery from sleep mode (only applies to
batteries that are new).
2 Place battery into detector making sure the battery latches into place. The detector will
automatically power on when battery is inserted.
USER MANUAL
Note
When a battery is inserted into the detector, the detector will boot into
Standby mode and the LED Status Indicator will blink at 1 second intervals.
Figure 11 LED Status Indicator
Table 8 LED Status Details
LED Behavior Status
Green and Yellow Solid Booting
Green Blinking (100000) Not associated
Green Blinking (1100) Service Cable Connected
Green Blinking (101010) Associated to the Access Point
Green Solid Linked to VSP
Yellow Solid Panel Error
Note
The blinking behavior occurs based on a 4Hz clock. Each digit for the blinking pattern represents 1/4s. 0 - LED is off, 1 - LED on.
www.vareximaging.com 19
4336Wv4 X-ray Detector
USER MANUAL
Note
WARNING
WARNING
WARNING
Note
The Service Cable does not provide power to the detector when tethered. Before servicing, ensure that a fully charged battery is inserted.
Changes or modifications not expressly approved by the manufacturer could void the user’s authority to operate the equipment.
Precautions should be taken to not open detector module. Depending upon the type of scintillator used, opening the detector module may expose the user to potentially toxic materials.
IP54
Detector able to withstand water spray from any direction.
Accessory or optional equipment connected to the analog and digital interfaces must be certified to the respective IEC standards (i.e., IEC 60601-1 for medical equipment). Furthermore, all configurations shall comply with the system standard IEC 60601-1-1. Anyone connecting additional or optional equipment to the signal inputs or signal outputs as part of a configuration for medical equipment is therefore responsible for compliance with the equipment standard IEC 60601-1. If in doubt, consult technical support personnel.
20 www.vareximaging.com
4336Wv4 X-ray Detector

9.0 Modes of Operation

The purpose of each mode is to configure the detector to achieve optimal performance during specific imaging procedures. Modes are defined by a combination of factors, such as; cycle time, integration time, field of view, noise, and pixel binning. Mode definitions are specified by a configuration files. The configuration file is stored in the detector’s non-volatile memory. Each mode requires a unique set of calibration files.
The PaxScan 4336Wv4 supports the Radiographic and vTrigger modes of operation see Table 9 for a typical mode example. The sensitivity of the detector is optimized to match the X-ray dose used in each mode.
USER MANUAL
Note
Table 9 Operational Mode Example
Mode Radiography - Full Resolution
Cycle Time 5.6 Seconds
Pixel Binning 1 x 1
Integration Time 550ms
X-ray Window Time 0.35 to 3.5 seconds
Image Area Full Field
Frame Size 2476 x 3072
Acquisition Type Accumulation
The OEM should work with Varex technical support for configuration of the mode(s) which best suit the customers intended application.
Characteristic Specification
www.vareximaging.com 21
4336Wv4 X-ray Detector

9.1 Radiography Mode

Within one exposure, multiple readouts are initiated by Prepare and followed by Trigger. Before readouts, the detector has an integration period. The integration time is defined by the configuration file for the current mode. In order to interface with the X-ray generator, two events, exposure on and exposure off, are delivered to the workstation. The X-ray exposure must happen between these two events.
One exposed frame is readout immediately after the integration window. The one or two post-offset frame(s) are readout. The frames are transferred to the workstation as soon as possible. When all the frames for the exposure are transferred to the workstation, a Transfer End event is delivered to the workstation.
USER MANUAL
Figure 12 Exposure: Corrected Frames
Figure 13 Exposure: Raw Frames
22 www.vareximaging.com
4336Wv4 X-ray Detector

9.2 vTrigger Mode

The vTrigger mode gives the OEM the benefit of not requiring any hardware synchronization with the generator. The vTrigger acquisition mode utilizes the detector’s hardware to sense when X-rays are presented to the detector. Once X-rays have been detected, the detector automatically begins a fixed integration time followed by the image readout.

9.3 Operational States

The operational states of the detector can be categorized as follows:
Radiography Acquisition: (Radiography-type)
Offset Calibration: (OEM-initiated)
Gain Calibration: (always-OEM initiated)
Each operating mode employs all types of calibration. For example, in radiography-type acquisitions, the PaxScan 4336Wv4 will acquire one frame with its respective offset.
USER MANUAL
www.vareximaging.com 23
4336Wv4 X-ray Detector

10.0 System Software

The PaxScan 4336Wv4 detector deploys the Varex Smart Panel (VSP) architecture. Detector software is embedded within the detector and is composed of two parts:
Software necessary to capture, process, and correct X-ray images.
The VSP COMM workstation libraries comprise the VSP SDK.
Because the main software is embedded within the detector, the only software required on the workstation is a small set of DLLs that should be copied from the software package to the workstation.
The VSP COMM libraries are copied to the workstation. Software interfacing to this detector will make API calls to these SDK libraries to control the image acquisition process. These libraries manage connections to the detector and the transfer of files from the detector to the workstation.
USER MANUAL
24 www.vareximaging.com
4336Wv4 X-ray Detector

10.1 X-ray Generator Interface Signals

Normally a set of hardware signals are used to interface between the detector and the X-ray generator and hand switch. For example, PREP and REQ are two hand switch signals that are typically used to indicate the user request for an X-ray exposure. Typically PREP (Prepare) indicates the user’s input that X-ray exposure is imminent, while REQ (Request) indicates the user’s input for immediate X-ray exposure. In response to the REQ signal, the detector will indicate the proper time for the X-ray exposure with the EXP_OK (Exposure OK) signal.
To accomplish this task of passing hardware signals to-and-from the detector, the VSP provides a software version of these signals in the API. For detectors, the workstation software will have to provide the hardware interface to the X-ray Generator and hand switch and relay those signals to the various API calls in the VSP library.

10.2 VSP_COMM (SDK) Files

libvsp.dll
libvsp-zf.dll
libwinpthread.dll
USER MANUAL
vspcs.dll
vsp.h
This set of files should be copied to the workstation. For customers working with C/C++, a vsp.h header file is included along with the libraries. For customers working with C#, the vspcs.dll provides the C# wrapper interface.
Note
Refer to FP2032_VarexSmartPanel_SoftwareInterfaceSpecification.pdf which provides API documentation for the software.

10.2.1 Bonjour Installation Files

Bonjour.msi
Bonjour64.msi
Bonjour is an optional installation and is required if you use the List()/vsp_list() API function and ViVA software.

10.2.2 Sample Code

vsp-example.cs – C# sample code project
vsp-example.c– C/C++sample code project
www.vareximaging.com 25

10.2.3 Utility Software

vsp-file.exe
This file utility is used to transfer a configuration file to the detector.
4336Wv4 X-ray Detector

10.3 Detector Calibration

Detector calibration files are stored inside the detector non-volatile memory. There are three calibration files per mode: offset, gain, and defects. The calibration files are used for corrections during image acquisition.
USER MANUAL

10.3.1 Offset Calibration

Offset calibration compensates for fixed pattern pixel intensity variations in the image associated with the dark current and electronic offsets. The Offset reference image is an average of a series of frames acquired without X-ray and referred to as dark fields. Prior to acquiring images, an offset calibration must be performed in each mode. This offset calibration is used for the preview image.
Note
Note
For interfaces connection, synchronization and timing diagrams information please reference the Software Interface Specification
FP2032_VarexSmartPanel_SoftwareInterfaceSpecification.pdf.
After a calibration, ensure the new calibration files are used by closing and re-opening the link to the detector.
Offset calibration should not be performed during X-ray.
The X-ray to digital conversion factor does not change because of calibration.
26 www.vareximaging.com
4336Wv4 X-ray Detector

10.3.2 Offset Calibration Steps

1 Choose the desired mode from the Mode Drop Down Menu.
Figure 14 Mode Drop Down Menu
2
Click the Offset Calibration Button or click Offset Calibration from the menu bar under Acquisition.
USER MANUAL
Figure 15 Initiate Offset Calibration
3
A Calibration Setting window appears.
4 Enter the number of frames desired and click OK.
Figure 16 Calibration Setting window
www.vareximaging.com 27
5 An Offset Calibration Progress window appears.
Figure 17 Offset Calibration Progress
4336Wv4 X-ray Detector
USER MANUAL
6
7 Click Close.
Once all frames are acquired, the detector is updated with the averaged offset calibration frame for the current mode. The updated message will show on the progress window and calibration process is complete.
Figure 18 Finish Offset Calibration
28 www.vareximaging.com
4336Wv4 X-ray Detector

10.3.3 Gain Calibration

To compensate for non-uniformities in the detector, a gain reference image (flat field) is used by the Corrections module as required to correct all images. The flat field image must be captured by the Varex Smart Panel (VSP) prior to acquiring images. The process of capturing the flat field image is known as Gain Calibration.
USER MANUAL
Note
Gain calibration is based upon the linear response of the detector to dose. Normalization is achieved by applying the flat field image acquired during the Gain calibration to all images corrected by the VSP. Normalization will fail with pixels that are responding to dose in a non-linear manner. Pixels responding to dose in a non-linear manner are usually caused by the saturation of the detector, or a low signal-to-noise ratio.
Note
Flat field images acquired near or exceeding the saturation point will cause normalization failures with all images acquired until a Gain calibration with the correct dose is performed.
Note
Every time a gain calibration is run, an offset calibration is required beforehand; this will ensure that the detector is properly calibrated.
It is critical to acquire the flat field image within a range that is large enough to be higher than the background noise created by the X-ray source and readout electronics of the detector, but lower than the saturation point of the detector.
It is recommended that flat field images be acquired with a median count of approximately 13000 - 14000.
This range will ensure that Gain calibration will meet both the upper and lower dose requirements under all modes of operation. Dose requirements are determined by the settings of the generator X-ray source.
To reduce the effects of noise, the average of each pixel in the flat field image is calculated by accumulating a number of frames into an internal memory buffer, then dividing the sum of each pixel by the number of frames acquired.
Note
Using larger numbers of calibration frames to capture the flat field image will result in more accurate calibration.
www.vareximaging.com 29
4336Wv4 X-ray Detector
The number of calibration frames used during Gain and Offset calibrations can be adjusted under the Mode Settings pull down menu.
The actual number of calibration frames used must be determined solely by the system integrator depending upon their specific performance requirements.
The general procedure for Gain calibration for all modes is described in Table 10. Detailed instructions on performing gain calibrations are covered in the ViVA Online help documentation.
USER MANUAL
Table 10 Gain Calibration - All Modes
Note
Gain calibration requires the production of X-rays and therefore certain precautions must be taken by the human operator.
Note
We recommend accumulating 32 frames for gain calibration and 8 frames for offset calibration for optimal image quality.
Action Results
Power On Power On the detector by inserting a battery. Once the detector is
powered on the detector is in low-power state. Wait 5 minutes prior to Gain calibration.
Offset Calibration Software performs a new Offset calibration referred to as dark field
acquisition.
Note: X-rays must not be used for this part of the calibration.
X-ray Radiation A uniform flat field with no obstructions in the path of the X-ray beam.
The radiation should ideally be at a level and technique representative of the typical radiation dose for the keeping in mind the general consideration outlined above.
Repeat The above procedure must be repeated for each of the stored imaging
modes.
detector during typical procedures,
30 www.vareximaging.com
4336Wv4 X-ray Detector

10.3.4 Radiographic Mode Gain Calibration

Radiography Gain calibration requires an Offset calibration performed prior to collecting the Flat Field image. Therefore, an Offset Calibration process must be run prior to the Gain Calibration. X-ray illuminated frames are then offset-corrected and accumulated in the VSP. A series of accumulated frames equals one radiographic X-ray exposure. Exposures are averaged to obtain the Flat Field image used by the VSP. The number of exposures acquired can be selected from the Settings window.
USER MANUAL
Note
ViVA provides the convenience of automatically running Offset Calibration as part of the Gain Calibration process.
However, API driven Gain Calibrations do not automatically run Offset Calibration. OEM Applications should be sure to run Offset Calibration prior to Gain Calibration.

10.3.5 Radiography Mode Gain Calibration Steps

1 Choose the desired imaging mode from the Mode Drop Down Menu.
Figure 19 Mode Drop Down Menu
www.vareximaging.com 31
4336Wv4 X-ray Detector
2 Under the Acquisition menu, click Gain Calibration. This invokes hardware handshaking
for the dark field calibration.
Figure 20 Initiate Gain Calibration
USER MANUAL
3
WARNING
A Gain Calibration Progress window will appear. Use operator control to perform an exposure.
In order to acquire an image with X-ray exposure in vTrigger mode, Hardware Handshaking must be enabled, see Figure 28, or an image without exposure will be returned.
Figure 21 Gain Calibration Progress window
32 www.vareximaging.com
4336Wv4 X-ray Detector
4 Once all X-ray frames have been acquired, the detector will update itself with the averaged
gain calibration for the current mode.
USER MANUAL
Figure 22 Gain Calibration Updating
5
After the detector has finished updating, an updated message and successfully finished calibration message will appear. The calibration process is complete. Click Close.
Figure 23 Gain Calibration Complete
Note
Operator Control is user supplied equipment.
www.vareximaging.com 33
4336Wv4 X-ray Detector
USER MANUAL
Note
Note
Note
Gain calibration should be performed at regular intervals, typically once every six (6) months, or whenever the central beam of the X-ray source has been moved relative to the detector.
Replacement of the X-ray tube will require a new gain calibration to be performed.
If the detector is dropped, a new gain calibration is required.
Varex recommends accumulating 32 frames for gain calibration for optimal image quality. However, the actual number of calibration frames used must be determined solely by the system integrator depending upon their specific performance requirements.
For additional assistance operating ViVA™, use the ViVA Online help documentation.
34 www.vareximaging.com
4336Wv4 X-ray Detector

10.4 Link to the Detector

1 Place a battery into the battery slot on the detector and latch into place. The detector LED will
begin to blink Green. See Table 8.
2 To link to the detector, click the ViVA icon, to launch the application.
Figure 24 The ViVA Icon
3
The detector will link automatically in ViVA. When the detector is linked, the Mode Drop Down Menu will become populated.
USER MANUAL
Figure 25 Detector Linked in ViVA
www.vareximaging.com 35

10.5 ViVA Mode Settings

This section describes the ways to view or change calibration and system settings that are currently used in ViVA.
1 Choose the desired imaging mode from the Mode Drop Down Menu.
Figure 26 Mode Drop Down Menu
2
USER MANUAL
4336Wv4 X-ray Detector
To enable or disable Hardware Handshaking, click Acquisition under the Menu Bar and select or de-select Hardware Handshaking. ViVA will remember your preferences for future launches.
Figure 27 Hardware Handshaking Not Selected
36 www.vareximaging.com
4336Wv4 X-ray Detector
USER MANUAL
Figure 28 Hardware Handshaking Selected
3
To see System Settings, click Acquisition under the Menu Bar.
www.vareximaging.com 37
4336Wv4 X-ray Detector
4 Click on System Settings. A System Settings window will appear. Image and calibration
settings may be turned On or Off in this window.
USER MANUAL
Figure 29 Open System Settings
Figure 30 System Settings window
38 www.vareximaging.com
4336Wv4 X-ray Detector

10.6 Image Acquisition

After Offset and Gain Calibration have been performed the detector is ready to acquire images. This section describes how to acquire images.
Table 11 Types of Images Detector will Acquire in Radiographic and vTrigger Modes
Typ e Images transferred to Workstation
Raw Exposed Frame and Post-Offset Frame
Corrected A single fully corrected frame
Offset Corrected A single offset corrected frame, without gain and defect
correction
Gain Corrected A single gain corrected frame, without defect correction
USER MANUAL
Note
When acquiring Correct, Offset Corrected, and Gain corrected images; if preview is enabled and additional preview frame will be delivered to the workstation prior to any additional frames.
www.vareximaging.com 39
4336Wv4 X-ray Detector

10.6.1 Radiography Mode

The Radiography mode provides superior single-shot, high-resolution images for diagnosis.
1 Choose the desired imaging mode from the Mode Drop Down Menu.
Figure 31 Mode Drop Down Menu
USER MANUAL
2
Note
Make sure Hardware Handshaking is selected, see Figure 28.
Note
The OEM workstation should have the ability to employ the PREP and Expose_OK signals.
In order to acquire an image with X-ray exposure in vTrigger mode, Hardware Handshaking must be enabled, see Figure 28, or an image without exposure will be returned.
40 www.vareximaging.com
4336Wv4 X-ray Detector
3 Click Acquire Image button or click Acquire Image in the Acquisition menu under the Menu
Bar to begin acquiring images.
Figure 32 Acquire Image Button
USER MANUAL
Figure 33 Acquire Image
www.vareximaging.com 41
4336Wv4 X-ray Detector
4 An Acquisition Progress window will appear. Click Start Acquisition and initiate X-ray
exposure.
Figure 34 Start Acquisition
USER MANUAL
5
Figure 35 Image Acquisition Progress
6
The Radiographic Acquisition Progress window will begin to show acquisition process with a blue status bar. Once the image has appeared, another image may be taken or the window may be closed by clicking Close.
The acquired image can be saved in the desired file format by clicking File/Save As.
42 www.vareximaging.com
4336Wv4 X-ray Detector

10.6.2 Radiographic Acquisition Settings

In the Radiographic Acquisition Progress window there are several settings that can be changed.
1 Click the Acquisition Type Drop Down Menu to switch between Software or Hardware
Acquisition.
USER MANUAL
Figure 36 Acquisition Type Drop Down Menu
www.vareximaging.com 43
2 Click on Settings, the Acquisition Settings Menu will appear.
Figure 37 Radiographic Acquisition Settings
3
USER MANUAL
4336Wv4 X-ray Detector
In the Acquisition Settings window; the type of corrected image, the number of loop acquisitions, auto-save, and Debug Verbosity may be selected. Click OK.
Figure 38 Radiographic Acquisition Settings window
44 www.vareximaging.com
4336Wv4 X-ray Detector

10.7 VSP Control Panel

The VSP Control Panel allows the user to control and modify settings of the detector function.

10.7.1 Access the VSP Control Panel

To access the Control Panel:
1 Open a browser and navigate to the IP Address of the detector, example:
http://192.168.2.31
2 Authentication is required to make any modifications to the settings in Control Panel. The
first time any of the links are clicked on the left side of the control panel, you will be prompted with a dialog box which allows you to log in.
USER MANUAL
Figure 39 VSP Control Panel Access Authentication
Note
Enter the default user and password:
3
The Authentication Required Dialog Box will look different in each browser, but require the same User Name and Password.
User Name: admin
Password: password
4 Click Log In.
Note
Once the detector is configured, the default password should be changed.
www.vareximaging.com 45
4336Wv4 X-ray Detector
5 The main VSP Control Panel screen displays links in the left pane. When a link is clicked the
content will show in the right pane.
Figure 40 VSP Control Panel
USER MANUAL

10.7.2 Modify Settings

To modify the settings:
1 Click Settings. 2 The Settings content will load and any settings that are able to be modified. 3 After inputting desired settings, click Update. The detector will reboot and the new settings
Figure 41 Modify VSP Control Panel Settings
will updated on the detector.
46 www.vareximaging.com
4336Wv4 X-ray Detector

10.7.3 Change Password

To change the password after logging in:
1 Click Change Password. 2 Enter a new Password. 3 Once the new password is entered along with its matching confirmation, click Change to
confirm the new password.
4 To clear the passwords entered in the forms, click Reset.
USER MANUAL
Figure 42 Change Password
Note
WARNING
Passwords must be at least 8 characters in length.
There is no back door available to the User or Varex. If the password is mismanaged, it will result in the return of the detector for service.
www.vareximaging.com 47
4336Wv4 X-ray Detector

10.7.4 Firmware Update

To update the VSP Firmware:
1 Click Update Firmware. This option is used to update all detector software
(embedded software, firmware, etc).
2 Select the Varex provided firmware image by clicking Browse. 3 After selecting the desired firmware, click Update Firmware to send the image to the VSP
and start the upgrade process.
USER MANUAL
Figure 43 Updating the Firmware
48 www.vareximaging.com
4336Wv4 X-ray Detector
4 During the update, another page will display showing the file upload was successful. At this
point the detector will apply the updated firmware.
Figure 44 Firmware Updated
USER MANUAL
WARNING
WARNING
Do not remove the Service Cable until the detector’s LED is blinking Green.
A fully charged battery should be used during firmware update. Do not remove the battery during firmware update.
www.vareximaging.com 49
4336Wv4 X-ray Detector

11.0 Maintenance

In principle, the detector assembly is maintenance-free, however; it is important that all correction files are regularly updated and used for image processing. In some applications it is advantageous to update the offset correction file directly before each image.
Although the a-Si detectors are resistant to X-rays they can exhibit degradation over time when exposed to high X-ray dose environments. Sensitivity and uniformity may change depending on the weekly exposure duration and X-ray dose. Therefore the pixel correction files should be checked and compared regularly to determine changes in bad pixels and clusters.

11.1 Cleaning and Disinfecting

The detector is likely to become contaminated during use. The specific material most likely to become contaminated is the X-ray grade carbon fiber input window and housing.
USER MANUAL

11.1.1 Chemicals for Cleaning and Disinfecting

All chemicals listed below are approved for use can be used for cleaning and disinfecting detector surfaces. To achieve full cleaning and disinfecting results, perform the procedures in
Section 11.1.2 and Section 11.1.3 for each cleaning product.
A list of chemicals is presented below:
CAVI-Wipes from Metrex
Isopropyl Alcohol
Mild Soap and Water
Ethanol 70%
WARNING
When the detector system is going to be cleaned or disinfected, be sure to remove the battery pack. Ignoring this warning may result in explosion, fire, or electrical shock, which may result in severe personal injury, death, or substantial product damage.
70% aqueous solution
Chlorine Bleach, 3% aqueous solution
Quaternary Ammonium Components
Steris Coverage Plus NPD
1 part Coverage Plus NPD to 255 parts Water
50 www.vareximaging.com
4336Wv4 X-ray Detector

11.1.2 Cleaning the X-ray Detector

To clean the X-ray detector:
1 Remove the battery from the X-ray detector. 2 Locate and read the cleaning instructions specified on the product label. Follow the product
instructions for cleaning.
USER MANUAL
Note
3 Dispense wipe from container. 4 Visually inspect the device. If contamination is present, use a wipe to remove the
contamination from the device surface. Use multiple wipes, as applicable, to remove the visible contamination.
5 Take care to remove contamination from seams, joints, and other difficult-to-reach areas. 6 Dispose of each wipe used for cleaning. 7 If visible residue from the cleaning product is evident after air-drying, remove the residue
with a general Isopropyl Alcohol wipe, and let air-dry.
If you are using a disinfectant other than those specified, we recommend you consult a specialist for the procedure for disinfection.
www.vareximaging.com 51
4336Wv4 X-ray Detector

11.1.3 Disinfecting the X-ray Detector

To disinfect the X-ray detector:
1 Remove the battery from the X-ray detector. 2 Locate and read the cleaning instructions specified on the product label. Follow the product
instructions for cleaning.
3 Wipe the entire surface of the target areas of the system until they are visibly wet with the
4 Take care to wet seams, joints, and other difficult-to-reach areas. 5 Ensure that the surfaces remain visibly wet for the maximum necessary disinfection time
USER MANUAL
Note
If you are using a disinfectant other than those specified, we recommend you consult a specialist for the procedure for disinfection.
chemical solution. Target areas include any surface that was cleaned and/or any other surfaces potentially contaminated during system use.
specified on the product label.
a If a disinfection time is not specified on a chemical label for the concentration used,
ensure that the surfaces remain visibly wet for a minimum of 10 minutes.
b Use additional fresh wipes as needed to ensure continuous wet contact time during the
specified disinfection period.
c Let air-dry. d If visible residue is evident after air-drying, remove the residue with a general Isopropyl
Alcohol wipe, and let air dry.
e Do not re-use wipes. Discard wipes and gloves into the correct waste container.
WARNING
Do not use flowing liquid or immersion on the detector, battery, battery compartment, or battery charger.
52 www.vareximaging.com
4336Wv4 X-ray Detector

12.0 Trouble-Shooting

Table 12 Problems and Solutions
Problem Solution
Detector fails to link wirelessly 1. Ensure that the detector is associated to the access point refer to
Table 8 for LED indicator behavior.
2. Remove the battery, re-insert the battery, and try to associate with the access point again.
3. If the detector will still not link, power cycle the access point.
4. If the detector will still not link, check the connections and the settings on the access point. Check SSID and password to ensure they match.
5. If problem persists, consult the manufacturer or field service technician for help.
USER MANUAL
Detector causes Electro-Magnetic Interference
Acquired image is Non-linear or the pixel are past the saturation point of dose
Acquired image is completely dark 1. Remove the battery, re-insert the battery, re-establish the link
Acquired image is noisy 1. Ensure that surrounding equipment is not interfering and
1. Reorient or relocate the receiving device.
2. Increase the separation between the equipment.
3. Consult the manufacturer or field service technician for help.
1. Remove the battery, re-insert the battery, re-establish the link.
2. Ensure that the tube dose is set to the correct settings.
3. Acquire another image at the recommended median count as stated in Section 10.3.3.
4. If problem persists, consult the manufacturer or field service technician for help.
and acquire another image.
2. If problem persists, consult the manufacturer or field service technician for help.
re-position the detector further away from equipment.
2. Remove the battery, re-insert the battery, re-establish the link and acquire another image.
3. If problem persists, consult the manufacturer or field service technician for help.
www.vareximaging.com 53
Table 12 Problems and Solutions (Continued)
4336Wv4 X-ray Detector
Acquired image shows a white or a black channel
Acquired images show defective pixels
USER MANUAL
Acquired image does not respond to exposure or the subject that is being X-rayed does not show up in the acquired images
1. Remove the battery, re-insert the battery, re-establish the link.
2. Ensure that the tube dose is set to the correct settings.
3. Acquire another image at the recommended median count as stated in Section 10.3.3.
4. If problem persists, consult the manufacturer or field service technician for help.
1. Remove the battery, re-insert the battery, re-establish the link.
2. Ensure that the tube dose is set to the correct settings.
3. Acquire another image at the recommended median count as stated in Section 10.3.3.
4. If the problem persists, add the defective pixels to the defect map and redo a gain calibration.
5. Acquire another image.
6. If problem persists, consult the manufacturer or field service technician for help.
1. Remove the battery, re-insert the battery, re-establish the link.
2. Ensure that the tube dose is set to the correct settings and that the detector is in the path of the X-ray beam.
3. Acquire another image at the recommended median count as stated in Section 10.3.3.
4. If problem persists, consult the manufacturer or field service technician for help.
Acquired image shows vertical or horizontal lines
Residual X-ray image from previous acquired image shows in current acquired image
Out of virtual memory Close some of the windows that are currently open.
1. Remove the battery, re-insert the battery, re-establish the link.
2. Ensure that the tube dose is set to the correct settings.
3. Acquire another image at the recommended median count as stated in Section 10.3.3.
4. If the problem persists, add the defects to the defect map and redo a gain calibration.
5. Acquire another image.
6. If problem persists, consult the manufacturer or field service technician for help.
Charge on the sensor pixels from an exposure beyond the saturation point of the pixels may cause a residual image. It can be erased by increasing time between exposures or taking multiple images until the artifact fades away.
54 www.vareximaging.com
4336Wv4 X-ray Detector
Table 12 Problems and Solutions (Continued)
ViVA software freezes Restart the computer and re-launch the ViVA software.
ViVA error message 1. Email the error log file generated to:
USER MANUAL
flatpanel.warranty@vareximaging.com found at C:\users\{username}\AppData\Local\
crashdumps\viva.log
Detector dropped 1. Remove battery from detector and inspect for damage.
2. If the battery does not appear damaged, place into battery charger to see if battery charger reports an error.
3. Inspect the detector for any physical damage.
4. Insert a charged battery into the detector and see if it powers on.
Note: It is best to use a different battery than the one that was in
the detector when it was dropped.
5. Re-establish a link.
6. Acquire an image from the detector and inspect for regions of missing data.
Note: If the detector has been dropped and is not returned to
Varex, a new gain calibration must be performed.
. This log file is normally
www.vareximaging.com 55
4336Wv4 X-ray Detector

12.1 After-sales Service for Varex Imaging Products

For product returns, contact your distributor or device manufacturer for shipping and packaging instructions. Do not return products to Varex Imaging for repair or service without advance notification. Include all required papers in the shipment.
We request that you obtain an RMA number for in-warranty and out-of-warranty returns of products.
Contact: flatpanel.warranty@vareximaging.com
For warranty and returns, please visit:
https://www.vareximaging.com/flat-panel-service-report
To ship the battery pack, follow the local and regional requirements for proper packaging and shipping of Lithium Batteries.
.
USER MANUAL

13.0 Disposing of the X-ray Detector

Contact you supplier or distributor, and check the terms of conditions of the purchase contract. This product should not be mixed with other commercial waste for disposal. Follow the local radiation protection regulations.
The Varex Imaging product may be attached as part of a component to other manufacturers’ systems. These other manufacturers are directly responsible for the collection and processing of their own waste products under the terms of the WEEE Directive. Contact these producers directly before discarding any of their products.
WARNING
WARNING
High energy X-rays and gamma rays greater than 5.5 MeV and/or neutrons may cause activation above regulatory limits for shipping. Varex cannot accept any radioactive shipments for activation above the local regulatory limits.
If the X-ray detector or accessories have been contaminated with potentially harmful substances, they cannot be accepted without written evidence of decontamination. See Section 11.1.2 and Section 11.1.3 for procedure of decontamination.
WARNING
56 www.vareximaging.com
Precautions should be taken to not open the detector module. Depending upon the type of scintillator used, opening the detector module may expose the user to potentially toxic materials.
4336Wv4 X-ray Detector

14.0 Safety

This section describes electromagnetic safety information.

14.1 Electromagnetic Interference

This equipment generates, uses and can radiate radio frequency (RF) energy and, if not installed and used in accordance with the instructions, may cause harmful interference to other devices or may be affected by other equipment in the vicinity. If this equipment does cause harmful interference to other devices or is affected by other equipment, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the measures listed in Section 12.0.
This equipment uses wireless LAN (WLAN) radios for transferring images. The WLAN power levels and antenna configurations have been tested and certified compliant through specific absorption rate (SAR) limits set by FCC/IC Canada (Less than 1.6W/kG) with separations as small as 0 cm between the panel antennas and human tissue. While compliant, it is still recommended to reduce exposure when possible by:
Positioning subject to be X-rayed away from the antennas (this also helps reduce image
transfer time).
USER MANUAL
Removing the detector panel promptly when X-ray exposure is complete.
www.vareximaging.com 57
4336Wv4 X-ray Detector

14.1.1 Electromagnetic Emissions

Table 13 Radiated/Conducted Emissions, Harmonics, Voltage, Fluctuations & Flicker
RF conducted emis­sions
EN55011/CISPR11
RF radiated emissions
EN55011/CISPR 11
Harmonic emissions
EN/IEC 61000-3-2
USER MANUAL
Voltage fluctuations / flicker emissions
IEC 61000-3-3
Emissions test
IEC 60601-1-2
test level
Group 1, Class B, 150 kHz – 30 MHz
Group 1, Class B, 30 MHz – 1 GHz
Class A N/A
Complies N/A
N/A
Battery power equip­ment not connected to mains
Group 1, Class B, 30 MHz – 1 GHz
Battery power equip­ment not connected to mains.
Battery power equip­ment not connected to mains
Compliance Electromagnetic environment
The detector uses RF energy for its internal function. Nearby electronic equipment may be affected.
The detector uses RF energy for its internal function. Nearby electronic equipment may be affected.
The detector is suitable for use in all estab­lishments other than domestic and those directly connected to the low voltage power supply network that supplies buildings used for domestic pur­poses.
The detector is suitable for use in all estab­lishments other than domestic and those directly connected to the low voltage power supply network that supplies buildings used for domestic pur­poses.
58 www.vareximaging.com
4336Wv4 X-ray Detector

14.1.2 Electromagnetic Immunity

Table 14 ESD, Transient/Burst, Surge, Voltage Variation, Magnetic Fields
USER MANUAL
Immunity test
Electrostatic discharge (ESD)
IEC 61000-4-2
Electrical fast transient/burst
IEC 61000-4-4
Surge IEC 61000-4-5
Voltage dips, short interruptions and voltage variations on power supply input lines
IEC 61000-4-11
IEC 60601-1-2
test level
± 8 kV contact discharge
± 2, 4, 8, 15 kV air discharge
±2 kV AC Mains
±2 kV I/O Lines
±0.5 kV, ±1 kV Line to Line
±0.5 kV, ±1 kV, ±2 kV Line to Ground
Voltage dips: 0 % UT
(100 % dip in UT) for 0,5 cycle at 0°, 45°, 90°, 135°, 180°, 225°, 270°, 315°
0 % UT (100 % dip in UT) for 1 cycle at 0°
70 % UT (30 % dip in UT) for 25/30 cycles at 0°
Voltage Interruptions: 0 % UT
(100 % dip in UT) for 250/300 cycle
Compliance Electromagnetic environment
± 8 kV contact discharge
± 2, 4, 8, 15 kV air discharge
N/A Battery power
equipment not connected to mains
N/A Battery power
equipment not connected to mains
N/A Battery power
equipment not connected to mains
Floors should be wood, concrete, or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%.
Mains power quality should be that of a typical commercial or hospital environment.
Mains power quality should be that of a typical commercial or hospital environment.
Mains power quality should be that of a typical commercial or hospital environment. If the user of the detector requires continued operation during power mains interruptions, it is recommended that the system be powered from an uninterruptible power supply or a battery.
Power frequency (50/60 Hz) magnetic field
IEC 61000-4-8
30 A/m 30 A/m Magnetic field should be that of a typical
location in a typical commercial or hospital environment.
www.vareximaging.com 59

14.1.3 RF Immunity

d
3.5
V
1
-------
P=
d
3.5
E
1
-------
P=
d
7
E
1
------
P=
Table 15 Conducted/Radiated RF
4336Wv4 X-ray Detector
Immunity test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
USER MANUAL
IEC 60601-1-2
test level
3 Vrms 150 kHz to 80 MHz
10 V/m 80 MHz to 2.7 GHz
Compliance Electromagnetic environment
N/A
Battery power equipment not connected to mains
10 V/m 80 MHz to 2.7 GHz
Portable and mobile RF communications equipment should be used no closer to any part of the detector than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
Recommended separation distance:
80 MHz to 800 MHz:
800 MHz to 2.5 GHz:
Where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as
a
Note
determined by an electromagnetic site survey should be less than the compliance level in each
frequency range
Interference may occur in the vicinity of equipment marked with the following symbol:
Some degradation of video & increase in video noise possible.
.
b
At 80 MHz and 800 MHz, the higher frequency range applies.
60 www.vareximaging.com
4336Wv4 X-ray Detector
USER MANUAL
Note
Note
Note
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the detector is used exceeds the applicable RF compliance level above, the detector should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the detector.
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than [V
] V/m.
1
www.vareximaging.com 61
4336Wv4 X-ray Detector

14.2 Radio Frequency (RF) Compliance Information

This section describes RF Compliance information.

14.2.1 FCC/IC Compliance

This device complies with Part 15 of the FCC Rules and RSS-Gen (RSS-210, etc.) of IC Rules. Its operation is subject to the following two conditions:
This device may not cause harmful interference.
This device must accept any interference received, including interference that may cause
undesired operation.
Transmission Band Frequencies: 2.4GHz, 5.2GHz, and 5.8Ghz band ranges
USER MANUAL
If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures noted below:
Reorient or relocate the receiving antenna.
Increase the separation between the equipment and receiver.
Connect the equipment into an outlet on a circuit different from the one the receiver is
Note
connected to.
This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to part 15 of the FCC Rules and Canadian ICES-003. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense.
Consult the dealer or an experienced radio/TV technician for help.
The user may find the following booklet prepared by the Federal Communications Commission helpful: The Interference Handbook
This booklet is available from the U.S. Government Printing Office, Washington, D.C. 20402. Stock No. 004-000-00345-4.
62 www.vareximaging.com
4336Wv4 X-ray Detector
USER MANUAL
Note
Modifications not expressly approved by the manufacturer could void the user's authority to operate the equipment under FCC rules.
Note
In the 5150 to 5250 MHz frequency range this transmitter is restricted to indoor use only.

14.2.2 Restrictions for 2.4GHz Use in Europe

Table 16 Europe - Restrictions for Use of 2.4GHZ Frequency in European Community
Language Statement
België/
Belgique
For private usage outside buildings across public grounds over less than 300m no special registration with IBPT/BIPT is required. Registration to IBPT/BIPT is required for private usage outside buildings across public grounds over more than 300m. For registration and license please contact IBPT/BIPT.
Voor privé-gebruik buiten gebouw over publieke groud over afstand kleiner dan 300m geen registratie bij BIPT/IBPT nodig; voor gebruik over afstand groter dan 300m is wel registratie bij BIPT/IBPT nodig. Voor registratie of licentie kunt u contact opnemen met BIPT.
Dans le cas d’une utilisation privée, à l’extérieur d’un bâtiment, au-dessus d’un espace public, aucun enregistrement n’est nécessaire pour une distance de moins de 300m. Pour une distance supérieure à 300m un enregistrement auprès de I’IBPT est requise. Pour les enregistrements et licences, veuillez contacter I’IBPT.
Deutschland License required for outdoor installations. Check with reseller for procedure to follow
Anmeldung im Outdoor-Bereich notwendig, aber nicht genehmigungspflichtig.Bitte mit Händler die Vorgehensweise abstimmen.
France Restricted frequency band: only channels 1 to 7 (2400 MHz and 2454 MHz respectively)
may be used outdoors in France.
Bande de fréquence restreinte : seuls les canaux 1- 7 (2400 et 2454 MHz respectivement) doivent être utilisés endroits extérieur en France. Vous pouvez contacter I’Autorité de Régulation des Télécommuniations (http://www.art-telecom.fr) pour la procédure à suivre.
Italia License required for indoor use. Use with outdoor installations not allowed.
E’necessaria la concessione ministeriale anche per l’uso interno.
Verificare con i rivenditori la procedura da seguire.
www.vareximaging.com 63
4336Wv4 X-ray Detector
Table 16 Europe - Restrictions for Use of 2.4GHZ Frequency in European Community
Nederland License required for outdoor installations. Check with reseller for procedure to follow.
Licentie verplicht voor gebruik met buitenantennes. Neem contact op met verkoper voor juiste procedure.
All EU member states and EFTA countries
Note: To remain in conformance with European spectrum usage laws for Wireless LAN operation, the
USER MANUAL
This device may only be used indoors in the frequency bands 5150 – 5250 MHz and 5250 – 5350 MHz.
above 2.4GHz channel limitations apply for outdoor usage. The user should use the wireless LAN utility to check the current channel of operation. If operation is occurring outside of the allowable frequencies for outdoor use, as listed above, the user must contact the applicable national spectrum regulator to request a license for outdoor operation.
64 www.vareximaging.com
4336Wv4 X-ray Detector

15.0 Regulatory

This section includes the manufacturers’ declaration of standards and regulations for which the detectors complies with.

15.1 Industry Canada Notice

To prevent radio interference to the licensed service, this device is intended to be operated indoors and away from windows to provide maximum shielding. Equipment (or its transmitting antenna) that is installed outdoors is subject to licensing. The installer of this radio equipment must ensure that the antenna is located or pointed such that it does not emit RF field in excess of Health Canada limits for the general population; consult Safety Code 6, obtainable from Health Canada's web site www.hc-sc.gc.ca/rpb.
Cet appareil numérique de la classe A est conforme à la norme NMB-003 du Canada Avis de Conformité à la Réglementation d'Industrie Canada:
Pour empêcher toute interférence aux services faisant l'objet d'une licence, cet appareil doit être utilisé à l'intérieur seulement et devrait être placé loin des fenêtres afin de fournir un écran de blindage maximal. L'installateur du présent matériel radio doit s'assurer que l'antenne est située ou pointée de manière à ce que cette dernière n'émette pas de champs radioélectriques supérieurs aux limites specifées par Santé Canada pour le grand public; consulter le Code de sécurité 6, disponible sur le site Web de Santé Canada, à l'adresse suivante: www.hc-sc.gc.ca/rpb.
USER MANUAL
This equipment complies with FCC RF radiation and RSS 102 exposure limits set forth for an uncontrolled environment. Body-worn operation and use near the head this device has been tested and meets both FCC/ IC RF exposure guidelines when used within this product guideline. The maximum SAR Value (Head) is 1.34W/kg. The maximum SAR Value (Body) is
1.37W/kg.
Cet équipement est conforme aux rayonnements RF de la FCC et RSS 102 limites d'exposition définies pour un environnement non contrôlé. Opération Porté au corps et utiliser près de la tête de ce dispositif a été testé et répond aux consignes d'exposition à la fois FCC / IC RF lorsqu'il est utilisé dans ce produit directive. La valeur maximale SAR (Head) est 1.34W / kg. Le maximum Valeur SAR (Body) est 1.37W/kg.
www.vareximaging.com 65
4336Wv4 X-ray Detector

15.2 Declaration of Conformity for European Union

Table 17 Europe - Declaration of Conformity in Languages of European Community
Language Statement
Česky [Czech] Varex Imaging, Inc. tímto prohlašuje, že tento Radiolan je ve shodě se základními
požadavky a dalšími příslušnými ustanoveními směrnice 1999/5/ES.
Dansk [Danish] Undertegnede Varex Imaging, Inc. erklærer herved, at følgende udstyr Radiolan
overholder de væsentlige krav og øvrige relevante krav i direktiv 1999/5/EF.
Deutsch [German] Hiermit erklärt Varex Imaging, Inc., dass sich das Gerät Radiolan in
Übereinstimmung mit den grundlegenden Anforderungen und den übrigen einschlägigen Bestimmungen der Richtlinie 1999/5/EG befindet.
Eesti [Estonian] Käesolevaga kinnitab Varex Imaging, Inc. seadme Radiolan vastavust direktiivi
1999/5/EÜ põhinõuetele ja nimetatud direktiivist tulenevatele teistele asjakohastele sätetele.
English Hereby, Varex Imaging, declares that this Radiolan is in compliance with the
essential requirements and other relevant provisions of Directive 1999/5/EC.
USER MANUAL
Nederlands [Dutch] Hierbij verklaart Varex Imaging dat het toestel Radiolan in overeenstemming is
Español [Spanish] Por medio de la presente Varex Imaging. declara que el Radiolan cumple con los
requisitos esenciales y cualesquiera otras disposiciones aplicables o exigibles de la Directiva 1999/5/CE.
Ελληνική [Greek] ΜΕ ΤΗΝ ΠΑΡΟΥΣΑ Varex Imaging. ΔΗΛΩΝΕΙ ΟΤΙ Radiolan
ΣΥΜΜΟΡΦΩΝΕΤΑΙ ΠΡΟΣ ΤΙΣ ΟΥΣΙΩΔΕΙΣ ΑΠΑΙΤΗΣΕΙΣ ΚΑΙ ΤΙΣ ΛΟΙΠΕΣ ΣΧΕΤΙΚΕΣ ΔΙΑΤΑΞΕΙΣ ΤΗΣ ΟΔΗΓΙΑΣ 1999/5/ΕΚ.
Français [French] Par la présente Varex Imaging déclare que l'appareil Radiolan est conforme aux
exigences essentielles et aux autres dispositions pertinentes de la directive 1999/5/CE.
Italiano [Italian] Con la presente Varex Imaging. dichiara che questo Radiolan è conforme ai
requisiti essenziali ed alle altre disposizioni pertinenti stabilite dalla direttiva 1999/5/CE.
Latviski [Latvian] Ar šo Varex Imaging deklarē, ka Radiolan atbilst Direktīvas 1999/5/EK
būtiskajām prasībām un citiem ar to saistītajiem noteikumiem.
Lietuvių
[Lithuanian]
Šiuo Varex Imaging deklaruoja, kad šis Radiolan atitinka esminius reikalavimus ir kitas 1999/5/EB Direktyvos nuostatas.
met de essentiële eisen en de andere relevante bepalingen van richtlijn 1999/5/EG.
Malti [Maltese] Hawnhekk, Varex Imaging, jiddikjara li dan Radiolan jikkonforma mal-ħtiġijiet
essenzjali u ma provvedimenti oħrajn relevanti li hemm fid-Dirrettiva 1999/5/EC.
Magyar [Hungarian] Alulírott, Varex Imaging nyilatkozom, hogy a Radiolan megfelel a vonatkozó
alapvetõ követelményeknek és az 1999/5/EC irányelv egyéb elõírásainak.
66 www.vareximaging.com
4336Wv4 X-ray Detector
Table 17 Europe - Declaration of Conformity in Languages of European Community (Continued)
Language Statement
Polski [Polish] Niniejszym Varex Imaging oświadcza, że Radiolan jest zgodny z zasadniczymi
wymogami oraz pozostałymi stosownymi postanowieniami Dyrektywy 1999/5/EC.
USER MANUAL
Português
[Portuguese]
Slovensko
[Slovenian]
Slovensky [Slovak] Varex Imaging týmto vyhlasuje, že Radiolan spĺňa základné požiadavky a všetky
Suomi [Finnish] Varex Imaging vakuuttaa täten että Radiolan tyyppinen laite on direktiivin
Svenska [Swedish] Härmed intygar Varex Imaging att denna Radiolan står I överensstämmelse med
Varex Imaging declara que este Radiolan está conforme com os requisitos essenciais e outras disposições da Directiva 1999/5/CE.
Varex Imaging izjavlja, da je ta Radiolan v skladu z bistvenimi zahtevami in ostalimi relevantnimi določili direktive 1999/5/ES.
príslušné ustanovenia Smernice 1999/5/ES.
1999/5/EY oleellisten vaatimusten ja sitä koskevien direktiivin muiden ehtojen mukainen.
de väsentliga egenskapskrav och övriga relevanta bestämmelser som framgår av direktiv 1999/5/EG.
www.vareximaging.com 67
4336Wv4 X-ray Detector

15.3 International Compliance

This section describes international compliance.

15.3.1 Argentina Compliance

Certificate Number: C-15871

15.3.2 Mexico Compliance

The operation of this equipment is subject to the following two conditions: (1) it is possible that this equipment or device does not cause harmful interference and (2) this equipment or device must accept any interference, including interference that may cause undesired operation.
USER MANUAL

15.3.3 Nigeria Compliance

15.3.4 Philippines Compliance

15.3.5 UAE Compliance

68 www.vareximaging.com
4336Wv4 X-ray Detector

15.3.6 Singapore Compliance

15.3.7 Thailand Statement

This telecommunication equipment conforms to the requirements of the National Telecommuni­cations Commission.
USER MANUAL

15.3.8 Taiwan NCC Warning Statement

Based on low-power radio waves radiated motor management approach:
Article 12 Type certified low power radio frequency motor, unlicensed, company,
firm or user shall not alter the frequency, increase the power or change the characteristics of the original design and function.
Article XIV using low power radio frequency motors shall not affect flight safety and
interference lawful communication; when there is interference by the discovery should be immediately suspended, and to improve without interfering may continue to be used. Legal communications in the preceding paragraph refers to the wireless telecommunication operations in accordance with the provisions of telecommunications. Low-power radio communication interference motor must endure radiation and Electronic Equipment legal or industrial, scientific and medical radio.
This module after getting certified in accordance with the provisions of the Flag tag in the examination and certification body modules and require manufacturers to label platform to platform.
The equipment belongs to the module certification, applicable to a variety of platforms.
Transmitter module includes: EW5270UM
Certificate #: CCAJ16LP0440T3
MPE (Maximum Permission Exposure) statement:
Electromagnetic exposure amount MPE standard value 1mW / cm2, send test products measured value: 0.9846 mW / cm2.
www.vareximaging.com 69
4336Wv4 X-ray Detector
NCC 警語
低功率電波輻性電機管理辦法
第十二條經型式認證合格之低功率射頻電機,非經許可,公司、商號或使用者均不得擅自變更頻率、加大 功率或變更原設計之特性及功能。第十四條低功率射頻電機之使用不得影響飛航安全及干擾合法通信;經 發現有干擾現象時,應立即停用,並改善至無干擾時方得繼續使用。前項合法通信,指依電信規定作業之 無線電信。低功率射頻電機須忍受合法通信或工業、科學及醫療用電波輻射性電機設備之干擾
本模組於取得認證後,將依規定於模組本體標示審驗合格標籤,並要求平台廠商於平台上標示
本器材屬於模組認證, 可適用於各種平台
變送器模組包括
證書編號: CCAJ16LP0440T3
MPE
電磁接觸量
USER MANUAL
: EW5270UM
(最大允許曝光) 聲明
MPE 標準1mW / cm2, 發送測試產品測量值: 0.9846 mW / cm2
70 www.vareximaging.com

Index

USER MANUAL
A
Abbreviations 2 Accessories 9 Acquire Image 39 Alerts and Notes, Meaning of iii
B
Battery
Installation 16 Removal 16
Battery Charge Level 18
C
Calibration Procedures 26 Calibration, Gain
Counts 29 Definition 29 Frames 30 Frequency 34 Steps 31
Calibration, Offset
Definition 26 Steps 27
Change Password 47 Cleaning and Disinfecting
Chemicals to Use 50 Detector 51
Country Code 45
D
Detector
Features 10 Intended Use 1
Detector Settings 45
Modify 46
Disposal 56
E
Electromagnetic Interference 57 Energy Range 14
F
Factory Restore 45 Features, Detector 10 Firmware 45 Firmware Update 48
G
Gain Calibration 29 Gain Calibration Steps 31
H
How To
Access Control Panel 45 Acquire Image 39 Gain Calibration 31 Link 35 Offset Calibration 27 Power On 19
I
Image Acquisition
Auto-Save 44 Corrected Image 44 Debug Verbosity 44 Loop 44
Time 12 Imaging Mode 40 Install Battery 16 Intended Use 1 Ionizing Radiation, Protection Against ii
L
Lag 14 Latch, Battery 17 LED Status 19 Load Support 12 Log File 45 Log In 45
M
Maintenance 50 Meaning of Alerts and Notes iii MTF 14
N
Notes and Alerts, Meaning of iii
O
Offset Calibration 26 Optional Parts 9
P
Password 45
Change 47
Recovery 47
Requirements 47 Protection Against Ionizing Radiation ii
R
Radioactive Shipments 56 Remove Battery 16
S
Safety Note
Battery vii
www.vareximaging.com 71
Index 4336Wv4 X-ray Detector
Handling vi If a Problem Occurs ix Installation and Environment of Use iv Maintenance and Inspection x Service Cable vi vTrigger Mode ix
WLAN viii Safety, For Your iii Sensitivity 14 Service
Change Password 47
Firmware Update 48
In-Warranty 56
Log In 45
Maintenance 50
Obtain RMA Number 56
Out-of-Warranty 56
Repair 56
Trouble-Shooting 53 Specifications
Active Area 12
Energy Range 14
Environmental Considerations 15
Lag 14
Load Support 12
USER MANUAL
MTF 14
Power Consumption 12
Sensitivity 14
Signal Strength 12
Weigh t 12 Standards and Regulations 4 Standards, Regulatory Requirements 4
Radiographic Acquisition Progress Window 42 Save Image 42 System Settings Window 38
W
Weigh t 12 Wireless 12
X
X-ray Tube 34
T
Trouble-Shooting 53
U
Update Country Code 45 Updates 45 User Name 45
V
Varex Smart Panel (VSP) 5 ViVA
Acquire Image Button 41
Acquisition Settings 43
Acquisition Type Drop Down Menu 43
Calibration Settings Window 27
Change Imaging Mode 40
Error 55
Gain Calibration Progress Window 32
Hardware Handshaking 36
Image Acquisition 39
Launch 35
Minimum Computer Requirements 12
Mode Drop Down Menu 27
Mode Settings 36
Offset Calibration Button 27
Offset Calibration Progress window 28
, 44
www.vareximaging.com 72
4336Wv4 X-ray Detector
This page intentionally left blank.
73
4336Wv4 X-ray Detector
®
USA Varex Imaging Corporation 1678 So. Pioneer Rd. Salt Lake City, UT 84104 United States +1 801 972 5000 Phone +1 801 973 5023 Fax
Arazy Group GmbH Am Flughafen, The Squaire 12 60549 Frankfurt am Main Germany
www.vareximaging.com
For a complete listing of our global offices, visit www.vareximaging.com ©2019 Varex Imaging All rights reserved. The Varex Imaging logo and design are registered trademarks of Varex Imaging or its subsidiaries, in the United States and other countries. All other trademarks not owned by Varex Imaging or its subsidiaries that are depicted herein are the property of their respective owners. Varex Imaging reserves the right to change this document at any time without notice and disclaims liability for editorial, pictorial or typographical errors.
140836 D
www.vareximaging.com 74
Loading...