Vapotherm Precision Flow Plus Nurse Call, Precision Flow Plus EMR User Manual

Precision Flow Plus Nurse Call and EMR Installation Instructions
Precision Flow Plus Nurse Call and EMR Installation Instructions Revision B
Document Number: 3100970 Page 2 of 12
Table of Contents
Nurse Call Installation Instructions ....................................................................................................3
1 Introduction ...........................................................................................................................3
2 Scope .......................................................................................................................................3
3 Hardware Interface Description ........................................................................................3
3.1 Nurse Call Interface ................................................................................................................................ 3
4 Wiring Guide: Connecting the Nurse Call Cable ...........................................................3
4.1 Figure 1: Nurse Call & EMR Cable Schematic ................................................................................ 4
4.2 Figure 1A: 3.5mm Stereo Audio Jack Male Pinout ....................................................................... 4
4.3 Figure 2: System Schematic ................................................................................................................. 5
5 Installation Instructions ......................................................................................................5
6 Alarms .....................................................................................................................................5
6.1 Table 1: Alarms that signal Nurse Call ............................................................................................. 6
7 Installation Verification Procedure ..................................................................................7
EMR Installation Instructions ............................................................................................................8
8 Introduction ...........................................................................................................................8
9 Scope .......................................................................................................................................8
10 Hardware Interface Description ........................................................................................8
10.1 Hardware ................................................................................................................................................... 8
10.2 Serial Port Configuration / Parameters .......................................................................................... 9
11 Communication Interface Description .............................................................................9
11.1 Introduction .............................................................................................................................................. 9
11.2 EMR Data Acquisition General Description ................................................................................... 9
11.3 Null Byte in message ........................................................................................................................... 10
11.4 EMR Data Format Specifics ............................................................................................................... 10
12 Installation Instructions ....................................................................................................11
13 Installation Verification Procedure ................................................................................12
Precision Flow Plus Nurse Call and EMR Installation Instructions Revision B
Document Number: 3100970 Page 3 of 12
Nurse Call Installation Instructions
1 Introduction
This document describes the Nurse Call communication interface of the Precision Flow Plus device. This document is intended for use by hospital IT, biomedical engineering or other experts who wish to connect the Precision Flow Plus into a Nurse Call system.
2 Scope
This document applies to the current design and embedded firmware version of the Precision Flow Plus unit. The current software version is 3.12.0. This document does not cover any earlier versions of Precision Flow Plus embedded firmware and will require a review and update for future versions of the firmware.
3 Hardware Interface Description
3.1 Nurse Call Interface
The Nurse Call / EMR Communication Cable (Part No. 3100897) contains a 3 pin 3.5 mm Stereo Audio Jack (See P2 in Figure 1) for indicating an alarm condition on a hospital nurse call system and a DB9 female Data Jack for interfacing Electronic Medical Record capable technologies.
An adapter or re-termination of the cable’s 1/8 in. (3.5 mm) stereo audio jack connection will be necessary and shall be supplied by the user to connect the Nurse Call Interface Adapter to the hospital’s nurse call system.
4 Wiring Guide: Connecting the Nurse Call Cable
The cable’s 3.5mm stereo jack supports “Normally Open” and/or “Normally Closed” Nurse Call system connections.The signals on the 3 connector contacts are (see P2 in Figure 1 and Figure 1A):
• Normally Closed (pin 18 of HDMI connector) Ring or Right Channel of 3.5mm audio Jack (See Figure 1A)
• Normally Open (pin 14 of HDI connector) Tip or Left channel of 3.5mm audio Jack (See Figure 1A)
• Common (pin 16 of HDMI connector) Sleeve or common channel of 3.5mm audio Jack (See Figure 1A)
Precision Flow Plus Nurse Call and EMR Installation Instructions Revision B
Document Number: 3100970 Page 4 of 12
4.1 Figure 1: Nurse Call & EMR Cable Schematic
4.2 Figure 1A: 3.5mm Stereo Audio Jack Male Pinout