Vaccixcell GlucCell User Manual

1
GlucCellTM SYSTEM USERS GUIDE
ver 2.3
Intended Use
The GlucCellTM Cell Culture Glucose Monitoring System (The GlucCellTM System) is designed to quantitatively measure the concentration of glucose during cell culture. It is for use by laboratory researchers or bioreactor professionals to obtain a quantitative measurement of glucose in cell culture.
About the System
The GlucCellTM System uses the latest technology in Cell Culture Glucose monitoring to provide you with easy and reliable testing. The system requires only a 1.5μl of medium sample to complete the testing in only 15 seconds.
The GlucCellTM System consists of 1) the GlucCellTM Glucose Meter; 2) GlucCellTM Glucose Test Strips, and 3)
the GlucCellTM code key and program key. These products are intended to be used together to get accurate Cell
Culture Glucose test results. Do not use other commercial test strips with your GlucCell meter.
Important Information
The GlucCell
System is intended for in vitro analysis use with cell culture medium. The system should not be used for diagnosis of diabetes.
Caution
A sample with large amounts of reducing agent such as ascorbic acid and uric acid may cause result slightly higher than the actual glucose level. High altitudes above than 2,750 meter (8,800 feet) may affect the test results. Temperatures outside the range of 10°C to 40°C (50°F to 104°F) may affect the test results.
CELL CULTURE GLUCOSE METER
2
DISPLAY: The large, easy to read display shows test results, messages, cell culture glucose results stored in memory, time and date. S BUTTON: Press S button to enter the time setting, and to decrease values in the setting mode. M BUTTON: Press M button to enter memory mode to
recall the information stored in meter’s memory and to
increase values in the setting mode. STRIP SLOT : Hold a GlucCellTM Glucose Test Strip in place when you perform Cell Culture Glucose test or perform check test. DATA PORT : Allow you to transfer the information stored in the meter’s memory to a computer to view, analyze and print. BATTERY COMPARTMENT : Hold one 3v Lithium battery. The battery is not yet installed into meter when new purchasing. Before using the meter, please install the battery first.
TEST RESULT BATTERY SYMBOL : Appear when battery is weak. THERMOMETER SYMBOL : Appears when ambient
temperature is above or below the acceptable range needed for testing. DROP SYMBOL : Appear when the test strip is inserted and meter is ready for testing.
STRIP SYMBOL M188 : Memory capacity is 180 entries. The display
shows M188 to ensure that all digits are working properly. mg/dL mmol/L : The meter has two units of measurement (mg/dL or mmol/L) available to choose from. DAY AVG : Appear when the meter is in the memory mode while recalling 1/7/14/21/28 day test averages.
DATE : Month/date. TIME : The meter is programmed for a 12 hour period format.
3
CELL CULTURE GLUCOSE TEST STRIP
1: Aperture: apply the sample to either side of strip. 2: Test Confirmation Window: make sure the chamber is filled up with test sample to ensure the correct result. For example:
3. Gray Electrode End: insert this end of test strip into meter.
IMPORTANT TEST STRIP INFORMATION
GlucCellTM System measures the amount of glucose in cell culture medium. Sample can be applied both to the
right or left side of the test strip’s apertures and is
automatically drawn to the test confirmation window where the reaction takes place. GlucCellTM Glucose Test Strips are intended for in vitro analysis use with cell culture medium. Results will not be
accurate if used with samples containing no electrolyte. Do not use test strips beyond the expiration date indicated on the strip vial label. The discard date for test strips is 90 days after first opening the vial. Record the discard date on the vial, when you open a new vial of test strips. GlucCellTM Glucose Test Strip could be damaged by heat and light. Keep them sealed in the original vial. Store the vial in a cool, dry place below to 400C/1040F and above 100C/500F. Do not refrigerate. Do not use damaged test strips or abnormal test strips. Use the test strip immediately once after removing it from the vial; replace the vial cap and close it tightly. Do not transfer test strips to a new vial. Always carry test strips in their original vial. Do not place test strips in direct heat or sunlight. Do not carry loose test strips in your carrying case.
Test strips are for single use only.
4
CHECK STRIP
The GlucCellTM Check Strip can be used in 2 ways: To check that the meter is operating properly. Use when performing quality control check on your meter. To change the measuring unit of your meter.
1. How to check meter by check strip Step 1: Insert the check strip into strip slot with label side
up as above.
Step 2: You should obtain an acceptable “OK“ reading
within 10 seconds, which mean your meter is working properly.
Step 3: Remove the check strip to exit. Meter will
automatically off.
NOTE: IF YOU DO NOT GET OK READING BUT APPEAR OTHER ERROR MESSAGE, PLEASE FOLLOW BELOW PROCEDURES:
1. TURN OFF THE METER BY PRESSING M BUTTON FOR 3 SECONDS AND REMOVE THE CHECK STRIP FROM THE METER.
2. INSERT CHECK STRIP TO REPEAT TEST. IF THE SECOND RESULT PERSISTS, CONSULT YOUR
LOCAL DISTRIBUTOR FOR HELP.
2. How to change measuring unit
The meter has two measuring units to choose from, mg/dL and mmol/L. The default unit is mg/dL if you do not change.
Step 1: Insert check strip into strip slot with label side up. Step 2: After “OK“ displayed, press S button until
flashing mg/dL shows up.
Step 3: Press S button to switch from mg/dL to mmol/L.
Wait until new setting flash for 3 times to validate the setting. Meter will display OK before off, and the unit is been changed successfully.
Step 4: Remove the check strip from the meter.
5
SETTING METER PARAMETERS
SETTING TIME & DATE
When you first time install the battery into the meter or every time when replace a battery, the meter will automatically enter to the setting mode. Please set correct time and date before you begin testing.
How to set time and date
Press S BUTTON for 3 seconds to turn on the meter. The meter will enter to the setting mode automatically. At the beep sound, the year will flash - for example for the year of 2009, number 09 will appear. Press S or M button to obtain the desired year. Wait the new setting flash 3 times to confirm and shifts to the next setting. Repeat to set the month, day, hour and minute. After minute is set, the meter will display OK before turn off.
NOTE: WHILE SETTING THE TIME AND DATE, YOU CAN EXIT THE SETTING MODE ANYTIME BY PRESS M BUTTON FOR 3 SECONDS.