GlucCellTM SYSTEM USER’S GUIDE
ver 2.3
Intended Use
The GlucCellTM Cell Culture Glucose Monitoring System
(The GlucCellTM System) is designed to quantitatively
measure the concentration of glucose during cell culture.
It is for use by laboratory researchers or bioreactor
professionals to obtain a quantitative measurement of
glucose in cell culture.
About the System
The GlucCellTM System uses the latest technology in Cell
Culture Glucose monitoring to provide you with easy and
reliable testing. The system requires only a 1.5μl of
medium sample to complete the testing in only 15
seconds.
The GlucCellTM System consists of 1) the GlucCellTM
Glucose Meter; 2) GlucCellTM Glucose Test Strips, and 3)
the GlucCellTM code key and program key. These products
are intended to be used together to get accurate Cell
Culture Glucose test results. Do not use other commercial
test strips with your GlucCell meter.
Important Information
The GlucCell
TM
System is intended for in vitro analysis
use with cell culture medium. The system should not be
used for diagnosis of diabetes.
Caution
A sample with large amounts of reducing agent such as
ascorbic acid and uric acid may cause result slightly
higher than the actual glucose level.
High altitudes above than 2,750 meter (8,800 feet) may
affect the test results.
Temperatures outside the range of 10°C to 40°C (50°F to
104°F) may affect the test results.
CELL CULTURE GLUCOSE METER

DISPLAY: The large, easy to read display shows test
results, messages, cell culture glucose results stored
in memory, time and date.
S BUTTON: Press S button to enter the time setting,
and to decrease values in the setting mode.
M BUTTON: Press M button to enter memory mode to
recall the information stored in meter’s memory and to
increase values in the setting mode.
STRIP SLOT : Hold a GlucCellTM Glucose Test Strip in
place when you perform Cell Culture Glucose test or
perform check test.
DATA PORT : Allow you to transfer the information
stored in the meter’s memory to a computer to view,
analyze and print.
BATTERY COMPARTMENT : Hold one 3v Lithium
battery. The battery is not yet installed into meter when
new purchasing. Before using the meter, please install
the battery first.
TEST RESULT
BATTERY SYMBOL : Appear when battery is weak.
THERMOMETER SYMBOL : Appears when ambient
temperature is above or below the acceptable range
needed for testing.
DROP SYMBOL : Appear when the test strip is
inserted and meter is ready for testing.
STRIP SYMBOL
M188 : Memory capacity is 180 entries. The display
shows M188 to ensure that all digits are working
properly.
mg/dL mmol/L : The meter has two units of
measurement (mg/dL or mmol/L) available to choose
from.
DAY AVG : Appear when the meter is in the memory
mode while recalling 1/7/14/21/28 day test averages.
DATE : Month/date.
TIME : The meter is programmed for a 12 hour period
format.

3
CELL CULTURE GLUCOSE TEST
STRIP
1: Aperture: apply the sample to either side of strip.
2: Test Confirmation Window: make sure the chamber
is filled up with test sample to ensure the correct result.
For example:
3. Gray Electrode End: insert this end of test strip into
meter.
IMPORTANT TEST STRIP INFORMATION
GlucCellTM System measures the amount of glucose in
cell culture medium. Sample can be applied both to the
right or left side of the test strip’s apertures and is
automatically drawn to the test confirmation window
where the reaction takes place.
GlucCellTM Glucose Test Strips are intended for in vitro
analysis use with cell culture medium. Results will not be
accurate if used with samples containing no electrolyte.
Do not use test strips beyond the expiration date indicated
on the strip vial label.
The discard date for test strips is 90 days after first
opening the vial. Record the discard date on the vial,
when you open a new vial of test strips.
GlucCellTM Glucose Test Strip could be damaged by heat
and light. Keep them sealed in the original vial.
Store the vial in a cool, dry place below to 400C/1040F and
above 100C/500F. Do not refrigerate.
Do not use damaged test strips or abnormal test strips.
Use the test strip immediately once after removing it from
the vial; replace the vial cap and close it tightly.
Do not transfer test strips to a new vial. Always carry test
strips in their original vial.
Do not place test strips in direct heat or sunlight.
Do not carry loose test strips in your carrying case.
Test strips are for single use only.

The GlucCellTM Check Strip
can be used in 2 ways:
To check that the meter is
operating properly. Use when
performing quality control
check on your meter.
To change the measuring unit
of your meter.
1. How to check meter by check strip
Step 1: Insert the check strip into strip slot with label side
up as above.
Step 2: You should obtain an acceptable “OK“ reading
within 10 seconds, which mean your meter is working
properly.
Step 3: Remove the check strip to exit. Meter will
automatically off.
NOTE: IF YOU DO NOT GET “OK” READING BUT
APPEAR OTHER ERROR MESSAGE, PLEASE
FOLLOW BELOW PROCEDURES:
1. TURN OFF THE METER BY PRESSING M BUTTON
FOR 3 SECONDS AND REMOVE THE CHECK STRIP
FROM THE METER.
2. INSERT CHECK STRIP TO REPEAT TEST. IF THE
SECOND RESULT PERSISTS, CONSULT YOUR
LOCAL DISTRIBUTOR FOR HELP.
2. How to change measuring unit
The meter has two measuring units to choose from,
mg/dL and mmol/L. The default unit is mg/dL if you do not
change.
Step 1: Insert check strip into strip slot with label side up.
Step 2: After “OK“ displayed, press S button until
flashing mg/dL shows up.
Step 3: Press S button to switch from mg/dL to mmol/L.
Wait until new setting flash for 3 times to validate the
setting. Meter will display “OK” before off, and the unit is
been changed successfully.
Step 4: Remove the check strip from the meter.

5
SETTING METER PARAMETERS
SETTING TIME & DATE
When you first time install the battery into the meter or
every time when replace a battery, the meter will
automatically enter to the setting mode. Please set
correct time and date before you begin testing.
How to set time and date
Press S BUTTON for 3 seconds to turn on the meter. The
meter will enter to the setting mode automatically.
At the beep sound, the year will flash - for example for the
year of 2009, number ” 09 “will appear.
Press S or M button to obtain the desired year.
Wait the new setting flash 3 times to confirm and shifts to
the next setting.
Repeat to set the month, day, hour and minute. After
minute is set, the meter will display “OK” before turn off.
NOTE: WHILE SETTING THE TIME AND DATE, YOU
CAN EXIT THE SETTING MODE ANYTIME BY PRESS
M BUTTON FOR 3 SECONDS.