UWE DM-3000, DM-6000, DM-15K, DM-30K, DM-1100 Operation Manual

...
Page 1
DM SERIES
ELECTRONIC WEIGHING SCALE
OPERATION MANUAL
PLEASE READ THIS MANUAL VERY CAREFULLY BEFORE
ATTEMPT TO OPERATE THE SCALE
FEBRUARY 2002
Specifications subject to change without prior notice
CONTENTS
CAUTIONS
1. INSTALLATION
2. SPECIFICATIONS
3. KEYBOARD LAYOUT AND DESCRIPTION
4. INITIAL SETUP
5. INSTRUCTION FOR USE
6. TROUBLE SHOOTING
7. DAILY CARE AND MAINTENANCE
Page 2
CAUTIONS
1. This unit is legal for trade only when it is sealed (and/or stamped) and bearing a serial number. Do not attempt to break the seal (or stamp) affixed to this scale or remove the serial number. Contact your dealer for more information and after sales service.
2. For most accurate weighing result, do not use this scale in where or when the environment condition falls beyond as those listed on SPECIFICATIONS.
3. Do not attempt to open this unit or conduct any trouble shootings other than those listed on TROUBLE SHOOTING.
4. Do not submerge scale into water.
5. Do not use pressurized water nozzle for cleaning.
Do not use this scale for chemical substances or in any environment that contains chemical vapor.
6. Precision measuring instrument, handle with care.
7. Always apply the rubber cap (FIG 2 E) when not in charging.
8. Do not attempt to open scale.
Page 3
1. INSTALLATION
1.1 Check and make sure that the following accessories are included: This operation manual Scale x 1 Plastic platter x 1 Stainless steel cover x 1 Power adaptor x 1
Contact your dealer if any item missed.
FIG 1
FIG 2
1.2 Locks the plastic platter (FIG 1 A) to the platter holder, with the screw (FIG 1 B) included, and then apply the rubber cap (FIG 1 C).
1.3 Place the stainless steel platter (FIG 1 D) onto the plastic platter.
Page 4
1.4 Power on the scale a. Before plugging the power adaptor into
electricity outlet, check and make sure that the input voltage of power adaptor matches with the output voltage of the outlet. If not, do not plug in and contact your dealer immediately.
b. Before first time use, plug the adaptor
into the wall outlet and charge the scale for at least 8 hours.
c. After battery is fully recharged, remove
adaptor from scale. Apply the rubber cap (FIG 2 E) to seal the DC input jack of scale.
2. SPECIFICATIONS
Model No. Capacity Division Accuracy Min.Load DM-3000 3000g 1g 1/3000 20g DM-6000 6000g 2g 1/3000 40g DM-15K 15kg 0.005kg 1/3000 0.1g DM-30K 30kg 0.01kg 1/3000 0.2g DM-1100 1100g 0.1g 1/11000 2g DM-11K 11kg 1g 1/11000 20g Tare Range Full tare range (Subtractive) Display 5 x 24mm LCD display Operation Environment
0o~40oC (32o~104oF), Non-condensed. R.H.¡Ø85%
Power source By Built-
in rechargeable battery or power
adaptor
Specifications subject to change without notice
NOTE
To comply with the law of certain countries and certain approval requirements, the imperial weight units may be disabled. Contact your dealer for more information.
Page 5
3. KEYBOARD LAYOUT AND DESCRIPTION
1. ON/ZERO KEY
Press this key to: (a) turn scale on or reset display value to zero during normal operation, (b) return to normal operation during function setup
2. TARE KEY
Press this key to tare off the weight of a container during normal operation.
3. MODE KEY
Press this key to shift among different weight units and display parameters.
4. OFF KEY
Press this key to turn scale off.
5. ZERO INDICATOR
This indicator appears to indicate a zero weight status.
6. TARE INDICATOR
This indicator appears to indicate the tare function is in operation and weight displayed is net weight.
7. WEIGHT UNIT INDICATOR
This indicator appears to indicate the weighing unit being selected is at metric.
8. LO-BAT INDICATOR
When this indicator appears, recharge the scale immediately. Failure to do so may cause unrecoverable damage to the rechargeable battery inside scale.
9. IN-CHARGE INDICATOR
Red color: Recharging battery Green color: Charging completed
Page 6
10. WEIGHT UNIT INDICATOR
This indicator appears to indicate the weighing unit being selected. Two avoirdupois weight units are available and this indicator means the weighing unit is at decimal pound (e.g. 5.62 means 5.62 lb).
11. WEIGHT UNIT INDICATOR
This indicator appears to indicate the weighing unit being selected is at pound/ounce (e.g. 2.15.1 means 2 lb and 15.1 ounce).
4. INITIAL SETUP
4.1 PLACING THE SCALE
In order to obtain an accurate weighing result, the scale must be placed on a strong and level surface.
If required, adjust the adjustable feet (FIG 2 F) underneath scale to obtain a level position. Refer to the bubble indicator for leveling condition.
4.2 INTERNAL SETTING
4.2.1 Select Weight Unit and Display Resolution
The scale is equipped with metric and avoirdupois weight units. Default setting = metric weight unit only. Follow the below steps to enable/disable the avoirdupois units.
Turn scale off a. Press and hold TARE, then press ON/ZERO b. Scale displays F.1 c. Press TARE twice d. Scale displays F.3
Page 7
e. Press and hold MODE for 3 second and then
release.
f. Press MODE to shift between desired
configurations appears.
g. Press ON/ZERO to confirm and return to
normal operation.
NOTE
To comply with the law of certain countries and certain approval requirements, the avoirdupois weight units may be disabled. Contact your dealer for more information.
4.2.2 Select Auto Power Off Setting
The scale is equipped with Auto Power Off Function. The scale will be turned off
automatically if it is not used for 4 minutes. Follow the below steps to enable/disable Auto Power Off Function
a. Turn scale off b. Press and hold TARE, then press ON/ZERO c. Scale displays F.1 d. Press TARE three times
e. Scale displays F.4 f. Press MODE to shift between 0._OFF and
4._OFF.
g. To enable Auto Power Off Function, press
ON/ZERO when 4._OFF appears.
h. To disable Auto Power Off Function, press
ON/ZERO when 0._OFF appears.
i. Press ON/ZERO to confirm and return to
normal operation.
4.3 SELECT BAUD RATE SETTING
This scale equipped with 6 RS232C output baud rate settings for user to choose the desire print transmission speed. Default setting =
9600. Follow the below steps to select baud rate.
a. Turn scale off b. Press and hold TARE, then press ON/ZERO c. Scale displays F.1 d. Press TARE four times e. Scale displays F.5 f. Press MODE to select baud rate of 1200,
2400,4800, 9600, 19200, 38400.
Page 8
g. Press ON/ZERO to confirm and return to
normal operation.
Note:
No transmission will be activated if weight detected is less than 20e.
The default data output format of the scale is
Baud Rate = 9600 Data Bit = 8 Stop Bit = 1 Parity = None
5. INSTRUCTION FOR USE
5.1 BEFORE WEIGHING
Make sure that: a. The plastic platter is screwed firmly and
the stainless steel platter cover is placed on the plastic platter.
b. The scale is placed on a level and strong
surface.
c. The scale is turned on. d. The ZERO INDICATOR is on. If not, press
ZERO/ON to set display to zero.
5.2 WEIGHING
a. Always place an object onto the platter
gently. Excessive force applied to platter may cause damages to sensor inside scale.
b. The weight of the object would be displayed
automatically.
c. It is a good practice to remove all loads
from platter after weighing. It would prolong the lifetime of scale.
Page 9
5.3 SELECT WEIGHT UNIT
Press MODE to shift among kg à decimal lb à lb/oz. à kg weight units. Refer to 4.2.1 for internal setting procedures.
The weight unit used before power off would be employed when the scale is turned on again.
5.4 SMART POWER SAVING FUNCTION
When this function is enable, the backlight will be turned off when constant/zero weight is detected and remains stable for about 10/3 seconds. Backlight will be automatically turned on again when a weight is detected.
5.5 SET DISPLAYED VALUE TO ZERO WHEN UNLOADED
Pressing the ON/ZERO can eliminate a positive or negative value. After the ON/ZERO is pressed, the ZERO INDICATOR appears.
Refer to SPECIFICATIONS for maximum zero range.
5.6 TARE OFF THE WEIGHT OF A CONTAINER
When a container is used, follow the below steps to tare off the weight of the container and get the net weight. a. Remove all loads away from platter b. Make sure that the ZERO INDICATOR is on. c. Place the container on the platter. d. Press TARE.
After TARE is pressed, the TARE INDICATOR would appear.
Refer to 5.2 for weighing procedures. Weight displayed under is the net weight of the subject matter. Refer to SPECIFICATIONS for maximum tare range.
5.7 CLEAR THE TARE FUNCTION
Remove all loads from platter then press TARE. After TARE is pressed, the TARE INDICATOR would disappear.
Page 10
5.8 RECHARGE THE SCALE
The scale is equipped with a built-in rechargeable battery. When the LO-BAT INDICATOR appears, recharge the scale immediately. Failure to do so may cause unrecoverable damage to the rechargeable battery. Charging is completed when the color of the
IN-CHARGE INDICATOR turns to green.
6. TROUBLE SHOOTING
Syndrome Scale cannot be turned on Check: Is the scale charged? Action: Recharge the scale for at least 8
hours before first time use or plug in the power adaptor before power on the scale.
Check Is the power adaptor inserted
properly into both the electricity outlet and the DC inlet of scale.
Action: Secure both ends of the power adaptor
and try again.
Syndrome Scale turned off automatically Check Is Auto Power Off Function employed? Action Refer to 4.2.2 to disable Auto Power
Off Function. Check Is the LO-BAT INDICATOR on? Action Apply the power adaptor.
Page 11
Syndrome Rated capacity cannot be reached Check Is the TARE INDICATOR on? Action: Turn the scale off. Remove all loads
from platter and turn on again.
Check Is there anything obstructing the
platter?
Action Remove all obstacles.
Syndrome Weighing result is not accurate Check Is the scale placed in a level
condition?
Action Adjust the adjustable feet to a level
condition.
Check Is the scale affected by airflow,
vibration or RFI?
Action: Place the scale away from all
disturbances.
Check Is the scale calibrated correctly? Action Contact your dealer. Check Is the correct weight unit employed? Action Refer to 4.2.1 and 5.3 for weight
unit setting and usage.
Syndrome Scale displays
00000
Check Is the scale overloaded? Action Remove some loads from platter.
Syndrome Scale displays
00000
Check Is the stainless steel platter cover
placed on the platter? Action Place the stainless steel platter
cover onto the platter.
Page 12
7. DAILY CARE AND MAINTENANCE
7.1 Clean the scale with a soft and damp cloth. If necessary, apply a mild detergent.
7.2 Do not use any harsh, abrasive material, acetone, volatile solvent, thinner or alcohol for cleaning.
7.3 Verify the accuracy of scale periodically. Re-calibrate the scale if necessary. Note: In some countries, calibration is restricted to be done by an authorized/qualified agent only. Contact your dealer for more information.
7.4 It is a good practice to apply the dust cover when operating the scale.
7.5 The scale must be placed horizontally during transportation or long time storage.
7.6 Remove platter from scale before transportation or long time storage.
7.7 Store scale in a dry and clean place.