any form or by any means (including electronic
storage and retrieval or translation into a
foreign language) without prior agreement and
written consent from Keysight Technologies,
Inc. as governed by United States and
international copyright laws.
Part Number
N9342-90002
Edition
Fifth Edition, February 2015
Printed in China
Keysight Technologies, Inc.
No. 116 Tian Fu 4th Street HiTech Industrial Zone (South)
Chengdu 610041, China
Software Revision
This guide is valid for A.03.40 revisions of the
N9342C/43C/44C firmware or later.
CAUTION
A CAUTION notice denotes a hazard. It
calls attention to an operating procedure,
practice, or the like that, if not correctly
performed or adhered to, could result in
damage to the product or loss of important data. Do not proceed beyond a CAU-TION notice until the indicated conditions
are fully understood and met.
WARNING
A WARNING notice denotes a hazard. It
calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could
result in personal injury or death. Do not
proceed beyond a WARNING notice until
the indicated conditions are fully understood and met.
Battery Marking
Keysight Technologies, through Rechargeable
Battery Recycling Corporation (RBRC), offers free
and convenient battery recycling options in the
U.S. and Canada.
Warranty
The material contained in this document is
provided “as is,” and is subject to being
changed, without notice, in future editions.
Further, to the maximum extent permitted by
applicable law, Keysight disclaims all
warranties, either express or implied, with
regard to this manual and any information
contained herein, including but not limited to
the implied warranties of merchantability
and fitness for a particular purpose. Keysight
shall not be liable for errors or for incidental
or consequential damages in connection with
the furnishing, use, or performance of this
document or of any information contained
herein. Should Keysight and the user have a
separate written agreement with warranty
terms covering the material in this document
that conflict with these terms, the warranty
terms in the separate agreement shall
control.
Technology Licenses
The hardware and/or software described in
this document are furnished under a license
and may be used or copied only in accordance
with the terms of such license.
Restricted Rights Legend
If software is for use in the performance of a
U.S. Government prime contract or
subcontract, Software is delivered and
licensed as “Commercial computer software”
as defined in DFAR 252.227-7014 (June 1995),
or as a “commercial item” as defined in FAR
2.101(a) or as “Restricted computer software”
as defined in FAR 52.227-19 (June 1987) or any
equivalent agency regulation or contract
clause. Use, duplication or disclosure of
Software is subject to Keysight Technologies’
standard commercial license terms, and
non-DOD Departments and Agencies of the
U.S. Government will receive no greater than
Restricted Rights as defined in FAR
52.227-19(c)(1-2) (June 1987). U.S.
Government users will receive no greater than
Limited Rights as defined in FAR 52.227-14
(June 1987) or DFAR 252.227-7015 (b)(2)
(November 1995), as applicable in any
technical data.
1Overview
Introduction 2
Front Panel Overview 5
Display Annotations 6
Top Panel Overview 8
Instrument Markings 9
2Getting Started
Checking Shipment and Order List 12
Power Requirements 13
AC Power Cords 14
Safety Considerations 15
Working with Batteries 18
Powering the Analyzer on for the First Time 20
Preparation for Use 21
Power On and Preset Settings 21
Factory Default Settings 22
Visual and Audio Adjustment 23
General System Settings 23
Timed Power On/Off 24
IP configuration 24
Ext Input 24
Show System 25
Adding an Option 26
Show Error 26
Perform Calibration 26
Data Securities 28
Upgrading Firmware 28
Probe Power Output 29
HSA and BSA PC software 30
Making Basic Measurements 31
Contents
3Functions and Measurements
Measuring Multiple Signals 34
Measuring a Low-Level Signal 39
Improving Frequency Resolution and Accuracy 44
Making Distortion Measurements 45
Making a Stimulus Response Transmission Measurement 51
Measuring Stop Band Attenuation of a Low-pass Filter 53
Making a Reflection Calibration 55
Measuring Return Loss Using the Reflection Calibration Routine 57
Making a Power Measurement with USB Power Sensor 58
The Keysight N934XC is a series of handheld
spectrum analyzer with a frequency range from
100 kHz to 20.0 GHz.
N9342C: 100 kHz - 7 GHz
N9343C: 1 MHz - 13.6 GHz
N9344C: 1 MHz - 20 GHz
It provides good usability and exceptional
performance for installation and maintenance,
spectrum monitoring, and on- site repair tasks. It
provides several measurement modes for different
applications. Each mode offers a set of automatic
measurements that pre- configure the analyzer
settings for ease of use.
1
Overview
Introduction
Introduction
The analyzer provides ultimate measurement
flexibility in a package that is ruggedized for field
environments and convenient for mobile
applications.
Functionality and Feature
The analyzer provides you with a comprehensive
functionality set and measurement convenience,
including:
• Power Measurement
provides power measurement functionality on
(Occupied Bandwidth), channel power, and
OBW
(Adjacent Channel Power Ratio).
ACPR
• Spectrum Emission Mask
Provides a Pass/Fail testing capability with a
mask for out- of- channel emissions measurement.
• Tracking Generator (Option TG7)
Provides an RF source for scalar network
analysis (exclusive for N9342C).
• Spectrum Monitor (Option SIM)
Provides
a signal over time. The analyzer can be used to
monitor the signal capturing performance or
intermittent
15Softkey menuSee key label description in the Key
16Softkey menu titleRefers to the current activated
17Remote annunciator
and shift annunciator
18Marker information[Marker]
19GPS information[System] > {More} > {GPS}
20Attenuation[Amplitude] > {Attenuation}
[BW] > {VBW} or [Freq] > {Freq
Offset}
[Span] or [Freq] > {Stop Freq}
Reference for more information.
function
Indicates the remote mode and shift
key mode
7
Overview
Ext Pow er
Charging
PC
Ext Trig /
RF Out 50
Antenna
GPS
Probe
Power Ext Ref
12-16
VDC
55
W MAX
RF Input 50
50 VDC MAX
30dBm (1W) M AX
50 VDC MAX
REV
PWR
30dBm (1W) MAX
LAN
1
2
10
3
4
5
7
8
12
6
9
11
Top Panel Overview
Top Panel Overview
ItemsFunction
1External DC power
connector
2 LED indicator (Charging) Lights (On) when the battery is charging
3 LED indicatorLights (On) when external DC power is
4 USB interface (Device)Connects to a PC
5 USB interface (Host)Connects to a USB memory stick or disk
6 HeadphoneConnects to a headphone
7 LAN InterfaceConnects to a PC for SCPI remote control
8 RF OUT ConnectorThe output for the built-in tracking generator.
9 Probe power connectorProvides power for high- impedance AC probes
10 EXT TRIG IN/REF IN
(BNC, Female)
11 GPS antenna connectorConnects an GPS Antenna (option GPA) for GPS
12 RF IN Connector (50 Ω)Accepts an external signal input.
Provides input for the DC power source via an
AC-DC adapter, or Automotive type DC adapter.
connected.
Enabled with Option TG7.
or other accessories (+15 V, –12 V, 150 mA
maximum).
Connects to an external TTL signal or a 10 MHz
reference signal. TheTTL signal is used to
trigger the analyzer’s internal sweep
application.
8
Instrument Markings
ISM1-A
ICES/NMB-001
The CE mark shows that the product complies with all
relevant European Legal Directives.
The CSA mark is a registered trademark of the
Canadian Standards Association.
All Level 1, 2 or 3 electrical equipment offered for sale
in Australia and New Zealand by Responsible Suppliers
must be marked with the Regulatory Compliance Mark.
This symbol is an Industrial Scientific and Medical
Group 1 Class A product (CISPR 11, Clause 4)
The ISM device complies with Canadian
Interference- Causing Equipment Standard- 001.
Cet appareil ISM est conforme à la norme NMB- 001 du
Canada.
The instruction manual symbol: indicates that the user
must refer to specific instructions in the manual.
The standby symbol is used to mark a position of the
instrument power switch.
Indicates the time period during which no hazardous or
toxic substance elements are expected to leak or
deteriorate during normal use. Forty years is the
expected useful life of the product.
Korea Certification indicates this equipment is Class A
suitable for professional use and is for use in
electromagnetic environments outside of the home.
Instrument Markings
Overview
The symbol indicates this product complies with the
WEEE Directive (2002/96/EC) marking requirements and
you must not discard this equipment in domestic
household waste.
9
Overview
Instrument Markings
10
Getting Started
2Getting Started
Information on checking the analyzer when
received, preparation for use, basic instrument use,
familiarity with controls, defining preset
conditions, updating firmware, and contacting
Keysight Technologies.
11
Getting Started
Checking Shipment and Order List
Checking Shipment and Order List
Check the shipment and order list when you
receive the shipment.
• Inspect the shipping container for damages.
Signs of damage may include a dented or torn
shipping container or cushioning material that
might indicate signs of unusual stress or compacting.
• Carefully remove the contents from the shipping
container, and verify if the standard
and your ordered options are included in the
shipment.
accessories
12
Power Requirements
Getting Started
Power Requirements
The AC power supplied must meet the following
requirements
Voltage :100 to 240 VAC
Frequency:50/60 Hz
Power:Maximum 80 W
The AC/DC power supply charger adapter supplied
with the analyzer is equipped with a three- wire
power cord, in accordance with international safety
standards. This power cord grounds the analyzer
cabinet when it is connected to an appropriate
power line outlet. The power cord appropriate to
the original shipping location is included with the
analyzer.
Various AC power cables are available from
Keysight that are unique to specific geographic
areas. You can order additional AC power cords
that are appropriate for use in different areas. The
AC Power Cord table provides a lists of the
available AC power cords, the plug configurations,
and identifies the geographic area in which each
cable is typically used.
The detachable power cord is the product
disconnecting device. It disconnects the main AC
circuits from the DC supply. The front- panel
switch is only a standby switch and does not
disconnect the instrument from the AC LINE
power.
:
13
Getting Started
250V 10A
125V 10A
230V 15A
250V 16A
AC Power Cords
AC Power Cords
Plug TypeCable Part
Number
8121-1703BS 1363/A Option 900
8120-0696AS 3112:2000 Option 901
250V 10A
8120-1692IEC 83 C4Option 902
250V 16A
8120-1521CNS 10917-2
8120-2296SEV 1011Option 906
250V 10A
8120-4600SABS 164-1 Option 917
8120-4754JIS C8303Option 918
125V 15A
8120-5181SI 32Option 919
a
Plug
Description
A 5-15P
/NEM
For use in
Country & Region
United Kingdom, Hong
ong, Singapore, Malaysia
K
Australia, New Zealand
Continental Europe, Korea,
nesia, Italy, Russia
Indo
Option 903
Unite States, Canada,
iwan , M e x i co
Ta
Switzerland
South Africa, India
Japan
Israel
8120-8377GB 1002Option 922
China
250V 10A
a. Plug description describes the p lug only. The part number is for the complete cable assembly.
14
Safety Considerations
WARNING
WARNING
WARNING
WARNING
Keysight has designed and tested the N934xC
handheld spectrum analyzer for measurement,
control and laboratory use in accordance with
Safety Requirements IEC 61010- 1. The tester is
supplied in a safe condition. The N934xC is also
designed for use in Installation Category II and
Pollution Degree 2 per IEC 61010- 1.
Read the following safety notices carefully before
you start to use a N934xC handheld spectrum
analyzer to ensure safe operation and to maintain
the product in a safe condition.
Personal injury may result if the analyzer’s cover is
removed. There are no operator-serviceable parts inside.
Always contact Keysight qualified personnel for service.
Disconnect the product from all voltage sources while it
is being opened.
This product is a Safety Class I analyzer. The main plug
should be inserted in a power socket outlet only if provided
with a protective earth contact. Any interruption of the
protective conductor inside or outside of the product is likely
to make the product dangerous. Intentional interruption is
prohibited.
Getting Started
Safety Considerations
Electrical shock may result when cleaning the analyzer
with the power supply connected. Do not attempt to
clean internally.
case only.
Use a dry soft cloth to clean the outside
Always use the three-pin AC power cord supplied with
this product. Failure to ensure adequate earth grounding
by not using this cord may cause personal injury and
product damage.
15
Getting Started
WARNING
CAU-CAUTION
CAU-CAUTION
CAU-CAUTION
CAU-CAUTION
Safety Considerations
Danger of explosion if the battery is incorrectly replaced.
Replace only with the same type battery recommended.
Do NOT dispose of batteries in a fire.
Do NOT place batteries in the trash. Batteries must be
recycled or disposed of properly.
Recharge the battery only in the analyzer. If left unused, a
fully charged battery will discharge itself over time.
Temperature extremes will affect the ability of the battery
to charge. Allow the battery to cool down or warm up as
necessary before use or charging.
Storing a battery in extreme hot or cold temperatures will
reduce the capacity and lifetime of a battery. Battery
storage is recommended at a temperature of less than
o
25
C.
Never use a damaged or worn-out adapter or battery.
Charging the batteries internally, even while the analyzer
is powered off, the analyzer may keep warm. To avoid
overheating, always disconnect the analyzer from the AC
adapter before storing the analyzer into the soft carrying
case.
Connect the automotive adapter to the power output
connector for IT equipment, when charging the battery on
your automotive.
16
The VxWorks operating system requires full conformity to
USB 1.1 or USB 2.0 standards from a USB disk. Not all the
USB disk are built that way. If you have problems
connecting a particular USB disk, please reboot the
analyzer before inserting another USB stick.
The analyzer cannot be used in the standard soft carrying
case for more than 1 hours if the ambient temperature is
higher than 35
o
C.
Getting Started
Safety Considerations
Environmental Requirements
The N934xC is designed for use under the
following conditions:
• Operating temperature:
o
C to 40oC (using AC- DC adapter)
0
o
C to +50oC (using battery)
–10
• Storage temperature: –40
• Battery temperature: 0
• Humidity: < 95% (40
o
C to 45oC
o
C)
o
C to +70oC
Electrical Requirements
The analyzer allows the use of either a lithium
battery pack (internal), AC- DC adapter shipped
with the analyzer, or optional automotive +12 VDC
adapter for its power supply.
Electrostatic Discharge (ESD) Precautions
This analyzer was constructed in an ESD protected
environment. This is because most of the
semiconductor devices used in this analyzer are
susceptible to damage by static discharge.
Depending on the magnitude of the charge, device
substrates can be punctured or destroyed by
contact or proximity of a static charge. The result
can cause degradation of device performance, early
failure, or immediate destruction.
These charges are generated in numerous ways,
such as simple contact, separation of materials,
and normal motions of persons working with static
sensitive devices.
When handling or servicing equipment containing
static sensitive devices, adequate precautions must
be taken to prevent device damage or destruction.
Only those who are thoroughly familiar with
industry accepted techniques for handling static
sensitive devices should attempt to service circuitry
with these devices.
17
Getting Started
Working with Batteries
Working with Batteries
The battery provides you approximately 3 hours of
operating time for your long time measurement in
field test.
Installing a Battery
StepNotes
1 Open the battery cover Use a phillips type screwdriver,
2 Insert the batteryObserve correct battery polarity
3 Close the battery cover Push the cover closed, then
loosen the retaining screw, then
pull the battery cover open.
orientation when installing.
re-fasten the cover with the
retaining screw.
Viewing the Battery Status
Determine the battery status using either of the
following methods:
• Check the battery icon in the lower- right corner
of the front- panel screen: it indicates the
approximate level of charge.
• Press[System] > {System Info} > {Show System} >
{Page down} to check the current battery
information.
18
Getting Started
CAU-CAUTION
Working with Batteries
Charging a Battery
You may charge the battery both in the tester and
in the external battery charger (option BCG).
Connect the automotive adapter to the IT power outlet of your
automobile (with option 1DC) for battery recharging.
1 Insert the battery in the analyzer.
2 Plug in the AC- DC adapter and switch on the
external power.
3 The charge indicator lights, indicating that the
battery is charging. When the battery is fully
charged, the green charging indicator turns off.
During charging and discharging, the battery
voltage, current, and temperature are monitored. If
any of the monitored conditions exceed their safety
limits, the battery will terminate any further
charging or discharging until the error condition is
corrected.
The charging time for a fully depleted battery, is
approximately four hours.
19
Getting Started
CAU-CAUTION
Install battery
Press Power Switch
Powering the Analyzer on for the First Time
Powering the Analyzer on for the First Time
Insert the battery into the analyzer or connect the
analyzer to an external power supply via the
AC- DC adapter, then press the power switch on
the front panel of your N934xC to power on the
analyzer.
Use only the original AC-DC adapter or originally supplied
battery for the power source.
The maximum RF input level of an average continuous
power is 30 dBm (or +
connecting a signal into the analyzer that exceeds the
maximum level.
Allow the analyzer to warm-up for 30 minutes
before making a calibrated measurement. To meet
its specifications, the analyzer must meet operating
temperature conditions.
50 VDC signal input). Avoid
20
Preparation for Use
This section provides the basic system
configuration which is frequently used before or
after the measurement operation.
Power On and Preset Settings
Selecting a preset type
Press [SYS] > {PwrOn/Off Preset} > {Preset Type} to
choose the preset types. The analyzer has three
types of preset setting for you to choose from:
DFT Restores the analyzer to its factory- defined
settings. The factory default settings can be
found, “Factory Default Settings” on page 22.
User Restores the analyzer to a user- defined
setting. Refer to the descriptions as below.
Restores the analyzer to the last setting.
Last
Saving a User-defined Preset
If you frequently use system settings that are not
the factory defaults, refer to the following steps to
create a user- defined system settings that can be
easily recalled:
1 Set the analyzer parameters using the knob, the
arrow keys, or the numeric keypad.
2 Press [SYS] > {PwrOn/Off Preset} > {Save User} to
save the current parameters as the user preset
setting.
3 Press [SYS] > {PwrOn/Off Preset}>{Preset Typ e U se r }
to set the preset mode to user defined system
setting.
4 Press [Preset]. The instrument will be set to the
state you previously saved.
Getting Started
Preparation for Use
21
Getting Started
Preparation for Use
Factory Default Settings
ParameterDefault Setting
Center FrequencySpecific to Product
Start Frequency 0.0 Hz
Stop FrequencySpecific to Product
SpanSpecific to Product
Reference Level0.0 dBm
Attenuation Auto (20 dB)
Scale/DIV10 dB/DIV
Scale TypeLog
RBWAuto (3 MHz)
VBWAuto (3 MHz)
Average TypeLog Power
Sweep timeAuto
Sweep ModeNormal
Probe PowerOff
Trace 1Clear write
Trace 2Blan k
Trace 3Blan k
Trace 4Blan k
Trace 1 D e t e c tio n Pos Peak
Trace 2 Detection Pos Peak
Trace 3 Detection Pos Peak
Trace 4 Detection Pos Peak
Trace AverageAll Off
MarkerAll Off
ModeSpectrum Analyzer
22
Getting Started
Preparation for Use
Visual and Audio Adjustment
Display Adjustment
Press [System] > {Brightness} > {Brightness} to toggle
the screen brightness between Auto and Man. When
it is set to Auto, the brightness adjusts according to
the environment automatically with the built-in
light sensor. When it is set to Man, you can set a
fixed brightness value manually.
Setting Button Backlight
Press [System] > {Keypad Setting} > {BackLight} to tog-
gle the backlight button between Auto and Man. You
can select the backlight brightness and the auto- off
idle time in manual mode.
Setting Key Beep
Press [System] > {Key Settings} >{Beeper} to activate
the key beep function as an indicator of key
operation.
General System Settings
Provides the following system setting options:
Time/Date
Press [System] > {Time/Date} to set the date and time
of the analyzer.
The allowed input for the time is HHMMSS format,
and YYYYMMDD format for the date.
Power Saving
Press [System] > {Screen Setting} > {Power Saving} to
select a power saving mode which turns off the
LCD display after a user- defined idle time. Press
any key to re- activate the LCD display after the
LCD display power- saving mode has been triggered.
23
Getting Started
Preparation for Use
Timed Power On/Off
Pressing [System] > {Power On/Off Preset} > {Timed Pwr
On} or {Timed Pwr Off} sets the time switch to power
on/off the N934xC in a user- defined time and date.
This function requires the power supply to be
connected or charged battery installed.
Press {Repeat Mode Once/Everyday} to set the N934xC
boot up/off in the pre- saved time everyday. The
pre- saved date is invalid in this mode.
IP configuration
The N934xC supports LAN port connection for
data transfer. Press [System] > {Port Setting} > {IP Admin} > {IP Address Static} to manually set the IP
address, gateway and subnet mask with the proper
LAN information. Or, just press [System] > {Setting} > {IP Admin} > {IP Address DCHP} to get the IP address
in LAN dynamically according DCHP.
Press {Apply} to enable all the configurations you
set.
Ext Input
Toggles the channel for external input between Ref
and Tr i g. Ref refers to a 10 MHz reference signal; Tri g
refers to a TTL signal.
24
External Reference (Ref)
Use the external reference function as follows:
1 Input a 10 MHz signal into the EXT TRIG IN/REF IN
2 Press [SPAN] > {Zero Span} to activate the Tri gg e r
function.
3 Access the associated softkeys to select the
rising edge (Ext Rise) or the falling edge (Ext Fall)
as the trigger threshold.
The trace will halt in external trigger mode until the trigger
threshold is met or the free run function is activated.
Show System
Pressing [System] > {System Info} > {Show system}
displays the following hardware, software, and
battery information of the analyzer:
Machine ModelBattery Info
MCU Firmware VersionName
DSP Firmware VersionSerial NO.
FPGA Firmware VersionCapacity
RF Firmware VersionTemperature
RF Module S/NCharge Cycles
KeyBoard Module S/NVoltage
This Run TimeCurrent
TemperatureCharge Status
Source VoltageRemain Time
Power SourceHost ID
25
Getting Started
Preparation for Use
Adding an Option
Pressing [System] > {More} > {Service} > {AddOption}
brings up a dialog box for entering the option
license code. Use the numeric keypad to input the
option license code and then use the [ENTER] key as
a terminator. If the analyzer recognizes the option
license code, a message “Option activated successfully” will appear in the status line.
Otherwise, a message “Invalid option licence” will
appear in the status line. Press [System] > {System
Info} > {Installed Options} to view the options.
Show Error
Pressing [System] > {System Info} > {Error history}
accesses a list of the 30 most recent error
messages. The most recent error will appear at the
bottom of the list. If the error list is longer than 30
entries, the analyzer reports an error message
[–350, Query overflow]. For more information,
refer to “Error Messages” on page 155.
Perform Calibration
The N934xC provides three manual calibration
function to calibrate the time base and amplitude.
The analyzer should warm up for 30 minutes
before calibration.
26
Time Base Calibration
Perform a time base calibration to guarantee the
frequency accuracy.
When the calibration function is triggered, the
current measurement is interrupted and a gauge
displays on the LCD. A message will display on the
LCD which indicates the calibration is finished,
and the interrupted measurement restarts.
The analyzer provides the GPS time base
calibration function (Option GPS is required).
Locate the analyzer on an open ground to receive
the GPS signal from satellites. Then press [System]
> {More} > {Service} > {Calibration} > {Time Base by GPS}
to perform a GPS time base calibration.
Time base calibration takes only a short time when the inner
temperature is stable. When the inner temperature is
increasing, calibration takes a long period of time or will fail. If
the input reference signal is abnormal, the calibration cycle
will take a long and unpredictable time to exit, and the LCD
displays an error message.
Amplitude Calibration
The analyzer provides the internal amplitude
calibration function. Please refer to the procedures
below to perform an amplitude calibration:
The allowed amplitude range is from –2 dBm to
2 dBm. Then press [Enter] to continue.
3 Input the amplitude number of the 50 MHz
signal in the pop- up window and press [Enter] as
a terminator.
The analyzer will perform a calibration according
to the input amplitude value. Press {Clear data} to
set to the factory- preset status with default
amplitude calibration data. The amplitude
calibration function is only available with the
firmware revision A.02.08 or later.
27
Getting Started
CAU-CAUTION
CAU-CAUTION
Preparation for Use
Data Securities
The N934xC offers the optional memory erase
function for data security. Press [System] > {More} >
{Securities} > {Erase Memory} to erase all the user
data in internal memory. Press Enter as a
terminator to start the erase process immediately.
The memory erase process takes 15 minutes
approximately. During the erase process, there must be a
constant power supply to ensure the successful erase. If
the erase process is interrupted, please reboot the
instrument and erase memory again.
Upgrading Firmware
Please follow the steps below to update the
firmware:
1 Download the latest N934xC firmware from
2 Extract files to the root directory of a USB stick.
You will see a folder named “N934xDATA” with
file Bappupgrade.hy.
3 Insert the USB stick into the top panel USB
connector.
4 Press [System] > {More} > {Service} > {Upgrade
Firmware} to activate the updating procedure.
Press Enter to upgrade the firmware. The
analyzer will perform the update automatically.
5 Unplug the USB stick and restart the analyzer
when message “All modules have been upgraded,
please restart” is displayed.
6 Press [System] > {System Info} > {Show System} to
find the updated MCU firmware version.
In updating process, there must be a constant power supply to
for at least 15 minutes. If power fails during the updating
process it can cause damage to the instrument.
28
Getting Started
Preparation for Use
Probe Power Output
The Probe Power provides power for
high- impedance AC probes or other accessories
(+15 V, –12V, 150 mA maximum).
The Probe power is set to off as default. Press
[System] > {More} {Port Setting} > {Probe Power On} to
switch on the probe power output.
29
Getting Started
HSA and BSA PC software
HSA and BSA PC software
Keysight HSA and BSA PC software is an
easy- to- use, PC- based remote control tool for the
N9342C/43C/44C HSA handheld spectrum analyzer.
It is able to be discretely used as a spectrum
monitor to display and control the trace scans
simultaneously with the analyzer, or a file manager
to send/get files between the analyzer and PC. It
also provides some data analysis function for your
further use.
For the further description of the HSA PC
software, please refer to the online help embedded
in this software.
30
Making Basic Measurements
This section provides information on basic analyzer
operations. It assumes that you are familiar with
the front and top panel buttons and keys, and
display annotations of your analyzer. If you are
not, please refer to “Front Panel Overview” on page 5,
“Top Panel Overview” on page 8, and “Instrument
Markings” on page 9.
For more details on making measurements, please
refer to“Functions and Measurements” on page 33”.
Entering Data
When setting measurement parameters, there are
several ways to enter or modify active function
values:
1 Using the Front Panel Knob
Increases or decreases the current value.
2 Using the Arrow Keys
Increases or decreases the current value by the
step unit defined.
Press [Freq] > {CF Step} to set the frequency by an
auto- coupled step (Step = Span/10, when {CF Step}
mode is set to Auto).
3 Using the Numeric Keypad
Enters a specific value. Press a terminator key
(either a specified unit softkey or [ENTER]) to
confirm input.
4 Using the Shift Hardkey
Press the blue shift key first, then press the
function hardkeys to select the upper alternative
function.
and select DFT to toggle the preset setting to the
factory- defined status.
3 Press the green [Preset] key to restore the
analyzer to its factory- defined setting.
4 Connect the generator’s RF OUT connector to the
analyzer’s RF IN connector.
5 Press [Freq] > 1 > {GHz} to set the analyzer center
frequency to 1 GHz.
6 Press [Span] > 5 > {MHz} to set the analyzer
frequency span to 5 MHz.
7 Press [Peak] to place a marker (M1) at the high-
est peak (1 GHz) on the display.
The Marker amplitude and frequency values appear
in the function block and in the upper- right corner
of the screen.
Use the front- panel knob, arrow keys, or the
softkeys in the Peak Search menu to move the
marker and show the value of both frequency and
amplitude displayed on the screen.
Figure 2-1View a signal (1 GHz, 0 dBm)
32
Functions and Measurements
3Functions and Measurements
33
Functions and Measurements
Measuring Multiple Signal s
Measuring Multiple Signals
This section provides information on measuring
multiple signals.
Comparing Signals on the Same Screen
The N934xC can easily compare frequency and
amplitude signal differences, for example,
measuring radio or television signal spectra. The
Delta Marker function allows two signals to be
compared when both appear on the screen at the
same time.
In the following example, a 50 MHz signal is used
to measure frequency and amplitude differences
between two signals on the same screen. The Delta
Marker function is demonstrated in this example.
1 Press [Preset] to set the analyzer to the factory
default setting.
2 Input a signal (0 dB, 50 MHz) to the RF IN
connector of the analyzer.
3 Set the analyzer start frequency, stop frequency,
and reference level to view the 50 MHz signal
and its harmonics up to 100 MHz:
• Press [FREQ] > 40 > {MHz}
• Press [FREQ] > 110 > {MHz}
• Press [AMPTD] > 0 > {dBm}
4 Press [PEAK]to place a marker on the highest
peak on the display (50 MHz).
The {Next Left PK} and {Next Right PK} softkeys are
available to move the marker from peak to peak.
5 Press [Marker] > {Delta} to anchor the first marker
(labeled as M1) and activate a delta marker.
The label on the first marker now reads 1R,
indicating that it is the reference point.
34
Functions and Measurements
NOTE
Measuring Multiple Signals
6 Move the second marker to another signal peak
using the front panel knob. In this example the
next peak is 100 MHz, a harmonic of the 50 MHz
signal:
Press [Peak] > {Next Right PK} or {Next Left PK}.
To increase the resolution of the marker readings, turn on
the frequency count function. For more information, please
refer to
“Improving Frequency Resolution and
Accuracy” on page 44
.
Figure 3 Del ta pair marker with signals (same screen)
35
Functions and Measurements
Frequency
Enter
7
MOD
On/Off
RF
4
102
9
6
3
On/Off
Amplitude FM
Utility
LF Out
Preset
Local
AM I/Q
File
Trigger
PulseM
·
Sweep
8
5
Remote
Standby
On
N9310A RF Signal Generator 9 kHz - 3.0 GHz
REVERSE PWR
4W MAX 30VDC
LF OUT RF OUT 50
FUNCTIONS
Frequency
Enter
7
MOD
On/Off
RF
4
102
9
6
3
On/Off
Amplitude FM
Utility
LF Out
Preset
Local
AM I/Q
File
TriggerPulseM
·
Sweep
8
5
Remote
Standby
On
N9310A RF Signal Generator 9 kHz - 3.0 GHz
REVERSE PWR
4W MAX 30VDC
LF OUT RF OUT 50
FUNCTIONS
Directional
coupler
Signal generator
Signal generator
Measuring Multiple Signal s
Resolving Signals of Equal Amplitude
In this example a decrease in resolution bandwidth
is used in combination with a decrease in video
bandwidth to resolve two signals of equal
amplitude with a frequency separation of 100 kHz.
Notice that the final RBW selected is the same
width as the signal separation, while the VBW is
slightly narrower than the RBW.
1 Connect two sources to the analyzer input as
shown below.
Figure 4 Setup for obtaining two signals
36
2 Set on
e source to 300 MHz. Set the frequency of
the other source to 300.1 MHz. Set both source
amplitudes to –20 dBm.
3 Setup the analyzer to view the signals:
• Press [PRESET]
• Press [FREQ] > 300.05 > {MHz}
• Press [SPAN] > 2 > {MHz}
• Press [BW] > 30 > {kHz}
Functions and Measurements
Measuring Multiple Signals
Use the knob or the arrow keys to further reduce
the resolution bandwidth and better resolve the
signals.
As you decrease the resolution bandwidth, you
improve the resolution of the individual signals and
it also increases the sweep timing. For fastest
measurement times, use the widest possible
resolution bandwidth.
Under factory preset conditions, the resolution
bandwidth is coupled to the span.
Figure 5 Resol ving signals of equal amplitude
37
Functions and Measurements
Measuring Multiple Signal s
Resolving Small Signals Hidden by Large
Signals
This example uses narrow resolution bandwidths to
resolve two input signals with a frequency
separation of 50 kHz and an amplitude difference
of 60 dB.
1 Connect two sources to the analyzer input
connector as shown in Figure 4 on page 36.
2 Set one source to 300 MHz at –10 dBm. Set the
other source to 300.05 MHz at –70 dBm.
3 Set the analyzer as follows:
• Press [PRESET]
• Press [FREQ] > 300.05 > {MHz}
• Press [SPAN] > 500 > {kHz}
• Press [BW] > 300 > {Hz}
4 Reduce the resolution bandwidth filter to view
the smaller hidden signal. Place a delta marker
on the smaller signal:
• Press [Peak]
• Press [MARKER] > {Delta}
• Press [Peak] > {Next Right PK} or {Next Left PK}
38
Figure 6 Resol ving a small signal hidden by a larger signal
Functions and Measurements
Measuring a Low-Level Signal
Measuring a Low-Level Signal
This section provides information on measuring
low- level signals and distinguishing them from
spectrum noise. There are four techniques used to
measure low- level signals.
Reducing Input Attenuation
The ability to measure a low- level signal is limited
by internally generated noise in the spectrum
analyzer.
The input attenuator affects the level of a signal
passing through the analyzer. If a signal is very
close to the noise floor, reducing input attenuation
will bring the signal out of the noise.
1 Preset the analyzer:
2 Input a signal (1 GHz, –80 dBm) to RF IN.
3 Set the CF, span and reference level:
• Press [FREQ] > 1 > {GHz}
• Press [SPAN] > 5 > {MHz}
• Press [AMPTD] > –40 > {dBm}
4 Move the desired peak (1 GHz) to the center of
the display:
• Press [Peak]
• Press [MARKER] > {Marker To} > {To Center}
Figure 7 A signal closer to the noise level (Atten: 10 dB)
39
Functions and Measurements
Measuring a Low-Level Signal
5 Reduce the span to 1 MHz and if necessary
re- center the peak.
• Press [SPAN] > 1 > {MHz}
6 Set the attenuation to 20 dB. Note that increas-
ing the attenuation moves the noise floor closer
to the signal level.
• Press [AMPTD] > {Attenuation} > 20 > {dB}
Figure 8 A signal closer to the noise level (Atten: 20 dB)
40
7 Press [AMPTD] >
attenuation to 0 dB.
Figure 9 A signal closer to the noise level (Atten: 0 dB)
{Attenuation} > 0 > {dB} to set the
Functions and Measurements
Measuring a Low-Level Signal
Decreasing the Resolution Bandwidth
Resolution bandwidth settings affect the level of
internal noise without affecting the amplitude level
of continuous wave (CW) signals. Decreasing the
RBW by a decade reduces the noise floor by 10 dB.
1 Refer to “Reducing Input Attenuation” on page 39,
and follow steps 1, 2 and 3.
2 Decrease the resolution bandwidth:
• Press [BW], and toggle RBW setting to Man
(manual), then decrease the resolution
bandwidth using the knob, the arrow keys or
the numeric keypad.
The low level signal appears more clearly because
the noise level is reduced.
Figure 10 Decreasing resolution band width
41
Functions and Measurements
Measuring a Low-Level Signal
Using the Average Detector and
Increased Sweep Time
The analyzer’s noise floor response may mask
low- level signals. Selecting the instruments
averaging detector and increasing the sweep time
will smooth the noise and improve the signal’s
visibility. Slower sweep times are necessary to
average noise variations.
1 Refer to “Reducing Input Attenuation” on page 39,
and follow steps 1, 2, and 3.
2 Press [TRACE] > {More} > {Detector} > {Average} to
select the average detector.
3 Press [Sweep] > {Sweep Time} to set the sweep
time to 500 ms.
Note how the noise appears to smooth out. The
analyzer has more time to average the values for
each of the displayed data points.
4 Press [BW] > {Avg Type} to change the average
type.
Figure 11 Using the average detector
42
Functions and Measurements
NOTE
Measuring a Low-Level Signal
Tra ce Av era gin g
Averaging is a digital process in which each trace
point is averaged with the previous sweep’s data
average for the same trace point.
Selecting averaging, when the analyzer is auto
coupled, changes the detection mode to sample,
smoothing the displayed noise level.
This is a trace processing function and is not the same as
using the average detector (as described on
1 Refer to the first procedure “Reducing Input
Attenuation” on page 39, and follow steps 1, 2,
and 3.
2 Press[TRACE] > {Average} (On) to turn average on.
3 Press50 > [ENTER] to set the average number to
50.
As the averaging routine smooths the trace, low
level signals become more visible.
Figure 12 Trace averaging
page 42).
43
Functions and Measurements
NOTE
Improving Frequency Resolution and Accuracy
Improving Frequency Resolution and Accuracy
This section provides information on using the
frequency counter to improve frequency resolution
and accuracy.
Marker count properly functions only on CW signals or discrete spectral components. The marker must be > 40 dB
above the displayed noise level.
1 Press [PRESET] (factory preset)
2 Input a signal (1 GHz, –30 dBm)to the ana-
lyzer’s RF IN connector.
3 Set the center frequency to 1 GHz and the span
MHz.
to 5
4 Press [MARKER] > {Function} > {Counter} to turn the
frequency counter on.
5 Move the marker by rotating the knob to a point
half- way down the skirt of the signal response.
6 Press [MARKER] > {Function} > {Normal} to turn off
the marker counter.
Figure 13 Using the frequency counter
44
Functions and Measurements
Making Distortion Measurements
Making Distortion Measurements
This section provides information on measuring
and identifying signal distortion.
Identifying Analyzer Generated Distortion
High level input signals may cause analyzer
distortion products that could mask the real
distortion present on the measured signal. Use
trace and the RF attenuator to determine which
signals, if any, may be internally generated
distortion products.
In this example, a signal from a signal generator is
used to determine whether the harmonic distortion
products are generated by the analyzer.
1 Input a signal (200 MHz, –10 dBm) to the
analyzer RF IN connector.
2 Set the analyzer center frequency and span:
• Press [Preset] (factory preset)
• Press [Freq] > 400 > {MHz}
• Press [Span] > 700 > {MHz}
The signal produces harmonic distortion products
200MHzfromthe original 200 MHz signal).
(spaced
Figure 14 Harmonic d istortion
45
Functions and Measurements
Making Distortion Measurements
3 Change the center frequency to the value of the
second (400 MHz) harmonic:
• Press [Peak]
• Press [Marker] > {Marker To} > {To Center}
4 Change the span to 50 MHz and re- center the
signal:
• Press [Span] > 50 > {MHz}
• Press [Peak]
5 Set the attenuation to 0 dB:
• Press [Amptd] > {Attenuation} > 0 > {dB}
• Press [Marker] > {Marker To} > {To Ref}
6 To determine whether the harmonic distortion
products are generated by the analyzer, first save
the trace data in trace 2 as follows:
• Press [Trace] > {Trace (2)}
• Press [Trace] > {Clear Write}
7 Allow trace 2 to update (minimum two sweeps),
then store the data from trace 2 and place a
delta marker on the harmonic of trace 2:
• Press [Trace] > {View}
• Press [Peak]
• Press [Marker] > {Delta}
The Figure 15 shows the stored data in trace 2 and
the measured data in trace 1. The Marker Delta
indicator reads the difference in amplitude between
the reference and active trace markers.
46
Functions and Measurements
Making Distortion Measurements
Figure 15 Identifying Analyzer Distortion (O dB atten)
ess [AMPTD] > {Attenuation} > 10 > {dB} to
8 Pr
increase the RF attenuation to 10 dB.
Figure 16 Identifying Analyzer Distortion (10 dB atten)
The marker readout comes from two sources:
eased input attenuation causes poorer
• Incr
signal- to- noise ratio. This causes the marker
delta value to be positive.
• Reduced contribution of the analyzer circuits
to the harmonic measurement causes the
marker to be negative.
A large marker delta value readout indicates
significant measurement errors. Set the input
attenuator at a level to minimize the absolute value
of marker delta.
47
Functions and Measurements
Frequency
Enter
7
MOD
On/Off
RF
4
102
9
6
3
On/Off
Amplitude FM
Utility
LF Out
Preset
Local
AM I/Q
File
TriggerPulseM
·
Sweep
8
5
Remote
Standby
On
N9310A RF Signal Generator 9 kHz - 3.0 GHz
REVERSE PWR
4W MAX 30VDC
LF OUT RF OUT 50
FUNCTIONS
Frequency
Enter
7
MOD
On/Off
RF
4
102
9
6
3
On/Off
Amplitude FM
Utility
LF Out
Preset
Local
AM I / Q
File
TriggerPulseM
·
Sweep
8
5
Remote
Standby
On
N9310A RF Signal Generator 9 kHz - 3.0 GHz
REVERSE PWR
4W MAX 30VDC
LF OUT RF OUT 50
FUNCTIONS
Signal generator
Signal generator
Directional
coupler
Making Distortion Measurements
Third-Order Intermodulation Distortion
Two- tone, third- order intermodulation (TOI)
distortion is a common test in communication
systems. When two signals are present in a
non- linear system, they may interact and create
third- order intermodulation distortion products
that are located close to the original signals.
System components such as amplifiers and mixers
generate these distortion products.
In this example we test a device for third-order
intermodulation using markers. Two sources are
used, one is set to 300 MHz and the other to
301 MHz.
1 Connect the equipment as shown in figure below.
48
This combination of signal generators and
directional coupler (used as a combiner) results in
a two- tone source with very low intermodulation
distortion.
Although the distortion from this setup may be
better than the specified performance of the
analyzer, it is useful for determining the TOI
performance of the source/analyzer combination.
Functions and Measurements
NOTE
Making Distortion Measurements
After the performance of the source/analyzer
combination has been verified, the DUT (device
under test, for example, an amplifier) would be
inserted between the directional coupler output
and the analyzer input.
The coupler used should have a high isolation between
the two input ports to limit the sources intermodulation.
2 Set one source (signal generator) to 300 MHz
and the other source to 301 MHz. This will
define the frequency separation at 1 MHz. Set
both sources equal in amplitude, as measured by
the analyzer. In this example, they are both set
to –5 dBm.
3 Set the analyzer center frequency and span:
• Press [PRESET] (Factory preset)
• Press [FREQ] > 300.5 > {MHz}
• Press [SPAN] > 5 > {MHz}
4 Reduce the RBW until the distortion products
are visible:
• Press [BW] > {RBW}, and reduce the RBW using
the knob, the arrow keys or the numeric keypad.
5 Move the signal to the reference level:
• Press [Peak]
• Press [MARKER] > {Marker To} > {To Ref}
6 Reduce the RBW until the distortion products
are visible:
• Press [BW] > {RBW}, and reduce the RBW using
the knob, the arrow keys or the numeric keypad.
7 Activate the second marker and place it on the
peak of the distortion product (beside the test
signal) using the Next Peak:
• Press [MARKER] > {Delta}
• Press [Peak] > {Next Left (Right) PK}
49
Functions and Measurements
Making Distortion Measurements
8 Measure the other distortion product:
• Press [MARKER] > {Normal}
• Press [Peak] > {Next Left (Right) Peak}
9 Measure the difference between this test signal
and the second distortion product.
• Press [MARKER] > {Normal}
• Press [Peak] > {Next Left/Right Peak}
Figure 17 TOI test screen
50
Making a Stimulus Response Transmission Measurement
N9340A
HANDHELD SPECTRUM ANALYZER
100kHz - 3.0GHz
PRESET
ENTER
FREQ SPANAMPTD
BW/
SWP
SYS MO DE MEAS TRACE
ESC/CLR
2DEF 3GHI1ABC
5MNO4JKL
6
PQR
8
VWX7STU9YZ_
0SAVE
LIMIT
MARKER
DUT
CAU-CAUTION
Functions and Measurements
Making a Stimulus Response Transmission
Measurement
The procedure below describes how to use a
built-in tracking generator to measure the rejection
of a low pass filter, a type of transmission
measurement.
1 To measure the rejection of a low pass filter,
connect the equipment as shown below.
A 370 MHz low- pass filter is used as a DUT in
this example.
Figure 18 Transmission Measurement Test Setup
2 Press [Pres
et] to perform a factory preset.
3 Set the start and stop frequencies and resolution
bandwidth:
• Press [FREQ] > {Start Freq} > 100 > {MHz}
• Press [FREQ] > {Stop Freq} > 1 > {GHz}
• Press [BW] > {RBW} > 1 > {MHz}
4 Turn on the tracking generator and if necessary,
The marker readout displays the rejection of the
filter at 130 MHz above the cutoff frequency of
the low- pass filter.
52
Figure 19 Measure the Rejection Range
Measuring Stop Band Attenuation of a Low-pass Filter
CAU-CAUTION
Functions and Measurements
Measuring Stop Band Attenuation of a
Low-pass Filter
When measuring filter characteristics, it is useful
to look at the stimulus response over a wide
frequency range. Setting the analyzer x- axis
(frequency) to display logarithmically provides this
function. The following example uses the tracking
generator to measure the stop band attenuation of
a 370 MHz low pass filter.
1 Connect the DUT as shown in Figure 18 on
page 51. This example uses a 370 MHz low pass
filter.
2 Press [Preset] to perform a factory preset.
3 Set the start and stop frequencies:
• Press [FREQ] > {Start Freq} > 100 > {MHz}
• Press [FREQ] > {Stop Freq} > 1 > {GHz}
• Press [AMPTD] > {Scale Type} > {Log}
4 Press [BW] > 10 > {kHz} to set the resolution
bandwidth to 10 kHz.
Excessivesignalinput maydamagetheDUT. Donot exceed
the maximum power that the device under test can
tolerate.
5 Turn on the tracking generator and if necessary,
frequency range is 19.16 dB/octave (one octave
above the cutoff frequency).
13Use the front- panel knob to place the marker at
the highest peak in the stop band to determine
the minimum stop band attenuation. In this
example, the peak occurs at 600 MHz. The
attenuation is 51.94 dB.
Figure 20 Minimum Stop Band Attenuation
54
Functions and Measurements
N9340A
HANDHELD SPECTRUM ANALYZER
100kHz - 3.0GHz
PRESET
ENTER
FREQ SPANAMPTD
BW/
SWP
SYS MODE MEAS TRACE
ESC/CLR
2DEF 3GHI1ABC
5MNO4JKL
6
PQR
8VWX7ST U 9 YZ_
0SAVE
LIMIT
MARKER
Coupled Port
Short
Circuit
DUT
or
NOTE
Making a Reflection Calibration
Making a Reflection Calibration
The following procedure makes a reflection
calibration using a coupler or directional bridge to
measure the return loss of a filter. This example
uses a 370 MHz low- pass filter as the DUT. The
tracking generator (option TG7) is needed for this
measurement. For N9342C handheld spectrum
analyzer with option CA7 or CAU, please refer to
the “Measuring Cable Reflection” on page 84 to make
a reflection measurement.
The calibration standard for reflection calibration
is usually a short circuit connected at the reference
plane (the point at which the DUT is connected). A
short circuit has a reflection coefficient of 1 (0 dB
return loss). It reflects all incident power and
provides a convenient 0 dB reference.
1 Connect the DUT to the directional bridge or
coupler as shown below. Terminate the
unconnected port of the DUT.
Figure 21
Reflection Measurement Short Calibration Test Setup
If possible, use a coupler or bridge with the correct test port
connector types for both calibrating and measuring. For the
best results, use the same adapter for the calibration and the
measurement. Terminate the second port of a two port device.
2 Connect the tracking generator output of the
analyzer to the directional bridge or coupler.
55
Functions and Measurements
CAU-CAUTION
NOTE
Making a Reflection Calibration
3 Connect the analyzer input to the coupled port
of the directional bridge or coupler.
4 Press [Preset] to perform a factory preset.
5 Turn on the tracking generator and if necessary,
Excessive signal input may damage the DUT. Do not exceed
the maximum power that the device under test can tolerate.
6 Set the start and stop frequencies and RBW:
• Press [FREQ] > {Start Freq} > 100 > {MHz}
• Press [FREQ] > {Stop Freq} > 1 > {GHz}
• Press [BW] > 1 > MHz
7 Replace the DUT with a short circuit.
8 Press [MEAS] > {Normalize} > {Store Ref (1 → 4)} >
{Normalize (On)} to normalize the trace.
It activates the trace 1 minus trace 4 function
and displays the results in trace 1. The
normalized trace or flat line represents 0 dB
return loss. Normalization occurs in each sweep.
Replace the short (cal device) with the DUT.
Since the reference trace is stored in trace 4, changing trace 4
to Clear Write invalidates the normalization.
Figure 22 Short Circuit Normalized
56
Measuring Return Loss Using the Reflection Calibration Routine
Functions and Measurements
Measuring Return Loss Using the Reflection
Calibration Routine
This procedure uses the reflection calibration
routine in the previous procedure “Making a
Reflection Calibration” on page 55, to calculate the
return loss of the 370 MHz low- pass filter.
1 After calibrating the system with the above
procedure, reconnect the filter in place of the
short (cal device) without changing any analyzer
settings.
2 Use the marker to read return loss. Position the
marker with the front-panel knob to read the
return loss at that frequency.
Rotate the knob to find the highest peak and
the readout is the maximum return loss.
Figure 23 Measuring the Return Loss of the Filter
57
Functions and Measurements
CAU-CAUTION
Making a Power Measurement with USB Power Sensor
Making a Power Measurement with USB
Power Sensor
Average power measurements provide a key metric
in transmitter performance.
Base station transmit power must be set accurately
to achieve optimal coverage in wireless networks. If
the transmit power is set too high due to
inaccurate power measurements, undesired
interference can occur. If the transmit power is set
too low, coverage gaps or holes may occur. Either
case may affect system capacity and may translate
into decreased revenue for service providers.
Figure 24 Connection with base station
Average power can be measured for the channel of
erest while the base station is active. All other
int
channels should be inactive. Average power is a
broadband measurement. If other signals are
present the analyzer will also measure their power
contributions.
58
The maximum power for the RF IN port and the RF OUT port of
the analyzer is +20 dBm. The maximum power for the Power
Sensor port is +24 dBm. When directly coupled to a base station, the test set can be damaged by excessive power applied
to any of these three ports.
To prevent damage in most situations when directly coupling
an analyzer to a base station, use a high power attenuator
between the analyzer and the BTS.
Making a Power Measurement with USB Power Sensor
NOTE
NOTE
Functions and Measurements
The N934xC spectrum analyzer supports the U2000
and U2020 series USB power sensors.
The U2000 Series USB power sensors do not need
manual calibration and zero routines performed.
Calibration and zeroing are performed without
removing the power sensor from the source,
through internal zeroing. With internal zeroing of
U2000 Series USB power sensors, there is no need
to disconnect the sensor or power- off the DUT. The
U2000 Series do not require 50 MHz reference
signal calibration, allowing the factory calibration
to ensure measurement accuracy. For best
accuracy, users are recommended to perform
external zeroing for input signals below - 30 dBm
for best accuracy
If you suspect other signals may be present, it is
recommended that you turn off all the other channels and
.
measure average power only on the signal of interest.
Another option is to measure channel power (which is less
accurate), that filters out all other channels (signals). You can
measure channel power for CDMA using the CDMA Analyzer
or CDMA Over Air tool. For other modulation formats, use
their respective analyzers (that is, GSM, 1xEV-DO, or
W-CDMA) or measure channel power using either the
spectrum analyzer or the Channel Scanner tool.
Connect the power meter as close as possible to the power
amplifier/duplexer output. Do not use a coupled port. Sensors
may not be as accurate at the power levels provided by
coupled ports.
59
Functions and Measurements
CAU-CAUTION
Making a Power Measurement with USB Power Sensor
Making an Average Power Measurement
To make an average power measurement, connect
the power sensor and cable, zero and calibrate the
meter, before making a measurement.
Zeroing of the Power Meter will occur
automatically:
• Every time the Power Meter function is used.
• When a 5 degree C. change in instrument
temperature occurs.
• Whenever the power sensor is changed.
• Every 24 hours (min.).
• Before measuring low level signals -for example,
10 dB above the lowest specified power the
power sensor is capable of.
Calibrate the Power Meter every time you cycle the
power on and off.
In most situations, you can press {Zero} to complete
the two steps (zero and cal) together.
To Make a Basic Average Power Measurement
You can follow the steps below to make a basic
average power measurement.
1 Press [Preset] to perform a factory preset.
2 Press [MODE]> {Power Meter} to access the power
meter mode.
3 Zero and calibrate the meter. Press {Zeroing} to
make a zero operation of the power sensor
followed by a calibration operation.
4 Connect the power sensor to the RF Input
50 MHz port. The analyzer supports the U2000
series power sensors.
To prevent damage to your sensor, do not exceed the power
levels below.
U2000/1/2A +25 dBm (320 mW) average, 20 VDC
+33 dBm (2 W) peak, <10 μs
For more information on the maximum power of U2000 series
power sensors, please refer to the Operating and Service Guide of Keysight U2000 Series USB Power Sensors.
60
Making a Power Measurement with USB Power Sensor
5 Connect the external attenuation, if required,
with the power sensor used.
6 Connect the power sensor to the signal to be
measured. A typical base transceiver station
connection is after the output of the power
amplifier and duplexer. The connection requires
a 2 ft. cable and two Type- N barrel connectors,
in addition to the power sensor and external
attenuator.
7 Press Meas Setup > Disp Range to set the display
range of the measurement. when it is set to On,
the display range will be adjusted automatically
according to the signal power dynamic range.
When it is set to Off, press To p and Bottom to set
the display range manually.
8 Press General Setup > Freq to set the power fre-
quency for the DUT. The current frequency will
display in the bottom line.
9 Press Power Range > Auto Power to toggles the
measurement range between On and Off. When it
is set to On, the power range will be adjusted
automatically. Use auto range when you are not
sure of the power level you are about to measure. When it is set to Off, press Range to select
the upper range or lower range. The lower range
uses the more sensitive path and the upper
range uses the attenuated path.
Table 1 Power Sensor Range
Functions and Measurements
Sensor Lower range Upper range
U2000/1/2/4A–60 dBm to –7 dBm–7 dBm to +20 dBm
U2000/1/2H–50 dBm to +3 dBm+3 dBm to +30 dBm
U2000/1B–30 dBm to +23 dBm +23 dBm to +44 dBm
61
Functions and Measurements
Making a Power Measurement with USB Power Sensor
10Press Meas > Meas Seup > Disp Mode to set the
display in meter. chart or trace display. the measurement result will displays on the screen as
shown.
Figure 25 Power Measurement
62
Making a Power Measurement with USB Power Sensor
NOTE
Functions and Measurements
Making a Peak Power Measurement
The Keysight U2020 X-series USB peak and average
power sensors enable peak power measurements
that have the same accuracy as measurements
obtained using traditional peak power meters,
while providing the advantages of being a very
compact and portable form factor.
Follow the steps below to make a peak power
measurement.
1 Press Mode > Power Meter to access the power
meter mode.
2 Connect the power sensor to the USB port of
N9322C. The sensor will automatically initialize,
perform zeroing (While remaining connected to
the device under test), and calibrate without
connecting to an external source.
You can manually perform zeroing and calibration by pressing
Zeroing > Zeroing and Calibration at any time after the
sensor is connected and initialized.
3 Press General Setup > Freq to set the frequency of
the signal under test.
As a wideband device, the power sensor has a
frequency response. The power sensor utilizes a
complex correction algorithm to remove power
measurement error due to linearity error,
frequency response, and temperature drift to
generate a more accurate measurement.
The analyzer provide the customer preset function
with U2020 X- series power sensor. Press Meas > Preset and the predefined protocol list will pop up.
Select the wanted item and press Preset as a
terminator. The related settings will be set
automatically.
63
Functions and Measurements
Making a Power Measurement with USB Power Sensor
Figure 26 Customized Preset the U2020 Power Sensor
Marker Readings
64
4 Press Mea
measurement results as Figure 27 below.
Figure 27 Peak Power Measurement
s Disp > Disp Mode > Trace to find the
Gate Results
Pulse Parameter Results
Making a Power Measurement with USB Power Sensor
Functions and Measurements
Power Meter Settings
Please find the sections below to find the general
power meter measurement settings.
Setting Power Meter Resolution
You can choose from four levels of Power Meter
resolution. Higher resolutions provide more
accuracy but slow the measurement speed.
1 Press [Preset] to perform a factory preset.
2 Press [MODE]> {Power Meter} to access the power
meter mode.
3 Press {Meas Disp} > {Resolution}. Select 1,2,3 or 4.
The four options represent different resolution as
follow:
• [1] = 1
• [2] = 0.1
• [3] = 0.01
• [4] = 0.001
SettingthePowerMeter’sTop and Bottom End-Points
Setting the end points (Disk Range) close to the
expected measurement value changes the
sensitivity of the Power Meter scale resolution. This
is an advantage in viewing small changes in power.
However, this will not affect the overall range of
the sensor.
1 Press [Preset] to perform a factory preset.
2 Press [MODE]> {Power Meter} to access the power
meter mode.
3 Press {Meas Disp} > {Disp Range} to access the
end- points menu.
4 Press {Top} and enter the maximum scale value
desired using the numeric keypad. Then press
{dBm} to complete the setup.
5 Press {Bottom} and enter the minimum scale value
desired using the numeric keypad. Then press
{dBm} to complete the setup.
65
Functions and Measurements
Making a Power Measurement with USB Power Sensor
Setting the Power Meter’s Upper and Lower Limits
The internal Power Meter can be configured to
detect when a measurement has failed a user
predefined upper and lower limits.
1 Press [Preset] to perform a factory preset.
2 Press [MODE]> {Power Meter} to access the power
meter mode.
3 Press {Meas Setup} > {Limits} to access the limits
menu.
4 Press {Limits} to activate the limits function. Each
time the softkey is pressed, the selected option
changes.
5 Press {Upper Limits} and enter the high limit value
using the numeric keypad. Then press {dBm} to
complete the setup.
Press {Lower Limits} and enter the low limit value
using the numeric keypad. Then press {dBm} to
complete the setup.
Set the Step Detection
Press Meas > General Setup > Average > Step Detect to
set the step Detect On and Off. When it is set to
on, the filter will re- initialize upon detection of a
step increase or decrease in the measured power
to reduce the filter settling time after a significant
step in the measured power, Step detection can be
set in both manual and automatic filter modes.
66
Set the Measurement Gate
Press Meas > Peak Setup > Gate Setup to set the start
time and gate length for a measurement gate. A
measurement gate allows measurements to be
performed on particular sections of the input
signal. The gate is defined by a start time relative
to the trigger instant and a duration. Signal
samples acquired during the time interval specified
by the gate are used for the measurements in that
gate. A system of up to four independent gates is
provided.
Making a Power Measurement with USB Power Sensor
Functions and Measurements
Set the Trigger
Press Meas > Peak Setup > Trig/Acq to set the trigger
parameters listed below for the measurement gate.
Acqn Set the single, free run or continuous
trigger mode. Select to enable trace and auto
trigger delay for the single and continuous trigger
modes. For the free run mode, only auto trigger
delay can be enabled.
Source Set the trigger source to internal channel
or an external source.
Mode Enable auto level or manually set the
trigger level if the internal trigger source is
selected.
Level sets the trigger level manually when the
Mode is set to manual.
Delay Set the delay time to be applied between
the trigger event and all the gate start times. This
allows you to time- shift all the gates by the same
amount with one setting change.
Slope Select the positive or negative slope type to
determine if the trigger event is recognized on the
rising or falling edge of a signal respectively.
Holdoff Set the holdoff time to disable the trigger
mechanism after a trigger event occurs.
Hysteresis Set the hysteresis to help generate a
more stable trigger by preventing triggering unless
the RF power level achieves the trigger level and
the additional hysteresis value. It can be applied to
both rising and falling edge trigger generation.
Hysteresis is only available for the internal trigger
source and manual trigger level.
67
Functions and Measurements
Spectrum Monitor
Spectrum Monitor
The scrolling three- dimensional display is noted for
its ability to track the frequency and power
behavior over the time, particularly intermittent
signals. Spectrum Monitor may be used to analyze
the stability of a signal over the time, or to identify
intermittent interference signals in communications
systems.
The X- axis represents frequency, and amplitude is
represented by color; red for a strong signal and
blue for noise floor. The Y-axis represents time,
with the trace from the newest sweep displayed at
the bottom of the screen. Earlier traces move up
toward the top with each new sweep. Two coupled
markers allow the user to place a marker on any
trace in the Spectrum Monitor and view the
normal spectrum for the time of that sweep. The
time interval between sweeps can be adjusted, and
up to 1500 traces can be displayed and saved.
Figure 28 Spectrum Monitor submenu screenshot
68
Please refer to the procedures below to perform a
ectrum Monitor measurement.
Sp
1 Connect the Omni Antenna to RF IN connector.
Functions and Measurements
2 Set the center frequency and span.
Press [FREQ]>[1.8052]>{GHz}, [SPAN]>[5]>{MHz}.
3 Turn on the Spectrum Monitor function.
Press [MEAS]>{Spectrum Monitor}.
4 Set the update interval time between two frames.
Press {Update Int. On}, rotating the knob or
pressing numeric keypads to set the update
interval time.
7 Press [Meas]>{File Logging}>{Start Save} to save
Spectrogram data. After pressing {Stop Save}. The
spectrogram will be saved into a trace or CSV
data file for your future use. With option GPS,
you can convert the CSV file into KML file in
HSA PC software for review the test GPS
location in Google Earth on PC.
Please refer to “Loading a file” on page 93 to load
the saved spectrogram trace file.
Spectrum Monitor
Spectrum Monitor Audio Alert
The spectrum monitor provides an audio alert with
a frequency dependant on the different control
parameters. This alert needs two markers to
specify the measure channel. The Band Power of the
channel specified by two markers is measured
without the audio alert. The Alert Limit is set to
specify the lowest band power that makes the
buzzer beep.
Please refer to the procedures below to turn on the
audio alert.
1 Press [Meas] > {Spectrum Monitor} to access into
the spectrogram mode.
2 Press [Marker] > {More} > {Audio Alert} to enable
the audio alert.
69
Functions and Measurements
NOTE
Demodulating an FM Signal
Demodulating an FM Signal
This section demonstrates how to demodulate and
listen to an FM signal using the built- in FM
demodulator.
Using the built- in FM demodulator you can tune to
an FM signal and view the results of
output as displayed in the time- domain.
Alternatively, the demodulated signal is also
available as an audio output (to the speaker or
headphone jack).
1 Perform a factory preset:
Press [Preset] (factory preset).
2 Use a signal source or an antenna for an FM
signal to analyze. In this example the signal
source is used transmitting at 300 MHz with FM
deviation of 10 kHz and FM rate of 1 kHz.
If you are using a broadcast FM signal in the United States,
for example, the FM channels are broadcasting between
87.7 MHz to 107.7 MHz. The optional preamplifier is essential
for the broadcast FM signal demodulation.
3 Set the center frequency to the center of the FM
6 Set the resolution bandwidth to capture the full
bandwidth of the FM signal. To calculate the
required bandwidth use
RBW =((2 x Frequency Deviation)+(2 x Modulation
Rate))
In our case the RBW should be: (2 x 10 kHz) +
(2 x 1 kHz) = 22 kHz With 1- 3- 10 sequence
RBW selections, choose the next highest RBW of
30 kHz:
Press [BW] > 30 > {kHz}.
7 Turn on the FM demodulator:
Press [Marker] > {Function} > {Demod}
{Demod Setting} > {Demod Type FM}.
Listen to the FM audio signal.
Press {Speaker Vol}, rotate the knob.
Alternatively, you can also use the headphone
jack.
71
Functions and Measurements
Modulation Analysis
Modulation Analysis
The analyzer provides the optional AM/FM (Option
AMA) and ASK/FSK (Option- DMA) modulation
analysis function for analysis the modulated
signals.
AM/FM Modulation Analysis
Optional AM/FM demodulation analysis provides
modulation metrics, including carrier power,
modulation rate, AM depth/FM deviation, SINAD
and carrier frequency offset. User definable limits
provide Pass/Fail indicators of four types:
Maximum carrier power, Maximum AM modulation
index or FM deviation, Minimum AM modulation
index or FM deviation, and Minimum carrier
frequency offset.
The user can save the waveforms with metrics for
reporting as well as the set- up parameters for
future measurements or analysis. In the following
example, you can use any source available (such as
AM/FM radio) as long as it has AM/FM signal
generation capability.
AM Modulation Analysis
Please refer to the procedures below for the AM
demodulation analysis:
1 Connect the signal generator to RF IN connector
and turn on the signal generator’s AM and RF
output.
2 Turn on the AM modulation analysis function.
Press [MODE] > {Modulation Analysis} > {AM}.
3 Press [Freq] > {Carrier Freq} > [100] > {MHz} to set
the carrier frequency.
4 Set the attenuation state to Auto (default):
Press {More} > {Attenuation Auto}
5 Press [Disp] > {Y Scale}> {Scale Coupling On} to set
the Y scale auto coupling.
72
Functions and Measurements
6 Set the AM depth limit and turn on the limit
function:
Press [Limits] > {AMod Depth UP} > [51] > {%} >
{AMod Depth Low} > [49] > {%} > {Limits on}
The red value in the result indicates the
modulated AM depth exceeds the limits set as
above.
Figure 29 AM mod ulation analysis
Modulation Analysis
FM Modulation Analysis
Please refer to the procedures below for the FM
modulation analysis:
1 Connect the signal generator to RF IN connector
and turn on the signal generator’s FM and RF
output.
2 Turn on the FM modulation analysis function.
Press [MODE] > {Modulation Analysis} > {FM}.
3 Press [Freq] > [Carrier Freq] > [100] > {MHz} to set
the carrier frequency.
4 Set the attenuation state to Auto (default):
Press {Attenuation Auto}
5 Press [Disp] > {Y Scale}> {Scale Coupling On} to set
the Y scale auto coupling.
73
Functions and Measurements
Modulation Analysis
6 Set the FM deviation limit and turn on the limit
The red value in the result indicates the
modulated FM deviation exceeds the limits set
as above.
Figure 30 FM mod ulation analysis
74
Functions and Measurements
Modulation Analysis
ASK/FSK Modulation Analysis
The ASK/FSK modulation analysis function
supports four display modes:
• Symbol
• Waveform
• ASK/FSK Error
• Eye Diagram.
User definable limits provide Pass/Fail indicators
of four types:
• Maximum carrier power
• Maximum ASK modulation depth/FSK frequency
deviation
• Minimum ASK modulation depth/FSK frequency
deviation
• Maximum carrier frequency offset.
The metrics includes carrier power, ASK/FSK error,
ASK depth/FSK frequency deviation, and ASK
index etc. The waveform with metrics and setup
parameters can be saved for reports and future
Measurements.
In the following example, you can use any source
available (such as your remote key of the car) as
long as it has ASK/FSK generation capability. For
ASK/FSK signal, its Filter is Nyquist and Alpha is
0.35.
ASK Modulation Analysis
Please refer to the procedures below to analysis
the ASK modulation.
1 Connect the signal generator to RF IN connector
and turn on the signal generator’s ASK and RF
output.
2 Turn on the ASK modulation analysis function.
Press [MODE], {Modulation Analysis}, {ASK}.
3 Press [Freq] > [100] > {MHz} to set the carrier
frequency.
75
Functions and Measurements
Modulation Analysis
4 Press [Meas] >{Symbol Rate} > [10] > {MHz} to set
The channel scanner - option SCN is a tool for
coverage test, band clearance, and spectrum
monitoring. It measures the channel power of
multiple transmitted channels. The power is
displayed via either the graphic bar chart or time
bar chart. It allows you to set maximum 20
channels for your test. You can continuously log
the scanned channel data as CSV file for post
processing or KML file for further use in Google
Earth application on your PC.
The CSV file has smaller size compared with KML file. It is recommended to save CSV file in analyzer and convert the CSV
file into KML file in HSA PC software. Both the CSV and KML
files record the channel information and GPS location (option
GPS is needed).
Top/Bottom N Channel Scanner
Top/Bottom N channel scan is used for searching
the top/bottom channels from specified channel
range according channel power.
Figure 33 Top N Channel Scanner Measurement
Please refer to the procedures below to perform a
78
Functions and Measurements
Top N channel scanner measurement.
1 Connect the antenna to RF IN connector.
2 Turn on the channel scanner function.
Press [Meas] > {Channel Scanner}
3 Press [Meas] > {Scan Mode} > {Top N} to set the
scan mode to TOP N channel mode. Input the
channel number you need to monitoring. Press
{Enter} as a terminator.
4 Press {Range Edit} to access the submenu to edit
the scan range.
Togg le {Edit Method} to underline Std. It allows
you to set the range according to the embedded
communication standard, such as CDMA and
GSM which is used specifically in different
countries. Press {Signal Std} to select the
standard. Press {Recall} as a terminator. Press
{Begin Chn} and {End Chn} to set the begin and end
channel for scanning. Press {Apply} to apply all
the range settings.
If you toggle {Edit Method} to underline Custom, it
allows you to set the customer- defined scan as
your need.
Press {Start Freq} to set the scan start frequency.
Press {Freq Step Size} to set the frequency interval
for adjacent channels.
Press {Bandwidth} to set the bandwidth of each
frequency channel which is used to calculate the
channel power.
Press {Number of Chn} to set the number of
frequency channels which will be scanned.
Press {Apply} as a terminator to apply all the
settings above.
5 Press {Scan Start} to initiate the channel scan
function.
6 Press {Logging Start} to start recording the scan
process which includes the channel power and
related channel bandwidth value to a CSV/KML
file. Press {Logging Stop} to end the logging. The
CSV/KML file will be saved immediately.
Channel Scanner
79
Functions and Measurements
Channel Scanner
List N Channel Scanner
The List N channel scan is used for listing the
frequency channels from specific channel standard
or customized frequency bandwidth.
Please refer to the procedures below to perform a
List N channel scanner measurement.
1 Connect the antenna to RF IN connector.
2 Turn on the channel scanner function.
Press [Meas] > {Channel Scanner}
3 Press [Meas] > {Scan Mode} > {List} to set the scan
mode to List N channel mode. Input the channel
number you need to monitoring. Press {Enter} as
a terminator.
4 Press {List Edit} to access the submenu to edit the
scan range.
Press {Insert} to insert a channel and access the
channel setting submenu to customize the
channel parameters.
Press {Edit Method Std}. It allows you to set the
range according to the embedded communication
standard, such as CDMA and GSM which is used
specifically in different countries.
Press {Signal Std} to load a standard file as your
need. Please refer to “Loading a file” on page 93
for details.
Press {Chn ID} to select the specific channel
according to the signal standard.
Or press {Edit Method Custom}. It allows you to set
the customer- defined scan as your need.
Press {Freq} to set the center frequency of the
current channel.
Press {Bandwidth} to set the bandwidth of each
frequency channel which is used to calculate the
channel power.
The first channel editing is completed. Press
{Return} > {Insert} to add more channels as your
need.
80
Functions and Measurements
NOTE
Channel Scanner
If there is a need to further edit the channel set
previously, press {Index}, rotate the knob to
highlight the channel as your need for further
editing.
Press {Apply} as a terminator to apply all the
settings above.
5 Press {Scan Start} to initiate the channel scan
function.
6 Press {Logging Start} to start recording the scan
process which includes the channel power and
related channel bandwidth value to a CSV/KML
file. Press {Logging Stop} to end the logging. The
CSV/KML file will be saved immediately.
Please make sure the file catalog is set to INT or USB unless
the logging file can not be saved successfully. When the
logging file exceeds the maximum volume of the storage
catalog, the logging will automatically pause and the file will
be saved immediately.
Display
In channel scanner mode, the [Disp] hardkey is
used to adjust the channel scanner display to
review the customized scan result.
Press {Freeze} to pause the display refresh but the
scan is still running.
You can sort the channel by ID or power, in
ascending or descending order. Or you can press
{Disp Mode} to toggle the display mode between bar
mode and time mode.
Figure 34 Time d isplay mode in List N channel scanner
81
Functions and Measurements
NOTE
Channel Scanner
Channel Scanner Setup
To customized the channel scan process, you can
press [Meas] > {Channel Scanner} > {Setup} to set the
measurement interval and rule in the submenu.
The Meas Interval is a value of time/distance during
which the every two scan process is triggered.
When it is set to off, the channel will be scanned
with default settings.
The Interval Type is toggled between time and
distance. When it is set to Time, the minimum
interval time is 100 ms. When it is set to Distance,
it allows you to measure the channel with the
specific distance on the road test. The minimum
interval distance is 1 meter. Please refer to the
Figure 39 on page 91 for the distance interval
channel scan.
The option GPS is required to set the Interval Type to
Distance. Press [System] > {More} > {GPS} > {GPS On}
The Measure Rule and Search Rule is toggled between
Speed and Accuracy.
For each scan process, the analyzer will first
search and find out the available channels
according to the selected scan mode. Then, a
precise measurement will perform to get the scan
result.
The accuracy mode in measurement rule is set as
default to get the precise measurement for the
searching- out channels.
The speed mode will speed the measurement but
reduce the measurement accuracy. It is set to
default in search rule to search out the available
channels approximately.
The Top Cy c l e number is the searching times for
each measurement. It allows you to measure the
specific searching- out channels continuously.
82
Cable & Antenna Test
Press [MODE] > {Cable&Antenna Test} to access the
cable and antenna measurement submenu. There
are three different measurement mode for your
selection:
• Reflection measurement
• One port cable loss
• Distance to fault
Preparation
Before making a measurement in a specific mode,
the HSA requires a calibration to make sure the
measurement accuracy. A new calibration must be
performed if any components used in the
calibration setup are changed. For example, a short
coaxial cable used to connect the tester to the DUT
Press [Freq] and [Span] to set the measurement
frequency range. It is best to keep the calibration
frequency range as close as possible to the actual
sweep frequencies necessary for the measurements.
Calibrating over a larger frequency range than the
actual measurement range will reduce the accuracy
of the test results.
Calibrators
The calibrator are available with the option CA7:
T-combo calibrator This is a T-shaped mechanical
calibrator. It consists of a calibrated open, short
and 50 Ω load. Using the T- combo calibrator will
result in a more accurate calibration than when
using the an electronic calibrator.
Functions and Measurements
Cable & Antenna Test
83
Functions and Measurements
Cable & Antenna Test
Calibrate with a T-combo calibrator
Using a T- combo calibrator is a three step process
requiring the user to manually change the
open- short- load components. The default order of
the calibration process is open- short- load.
1 Press [Freq] > {Calibrate} to bring up the dialog
box for calibration instruction. Follow the
on- screen instruction to connect the Open
component of the T- combo calibrator to the RF
OUT connector of the analyzer. Press [ENTER] to
start the open calibration.
2 Follow the instruction to connect the short
component of the T- combo calibrator to the RF
OUT connector of the tester. Press [ENTER] to
start the short calibration.
3 Connect the Load component of the T- combo
calibrator to the RF OUT connector of the tester.
Press [ENTER] to start the load calibration.
A Calibrated indicator is shown on the upper left
corner of the screen when the calibration is
completed. Then, the analyzer is ready for your
further cable measurement.
Measuring Cable Reflection
Press [Mode] > {Cable&Antenna Test} > {Reflection
Measurement} to select the Reflection Measurement
Mode. It allows you to set the Y axis of the
graticule as Return Loss, VSWR or Reflection Coefficient
value in Disp Mode.
84
Perform a Basic Reflection Measurement
Please refer to the procedure below to make a
reflection measurement over a frequency range of 5
MHz to 7 GHz.
1. Press [MODE] > {Cable&Antenna Test} > Reflection Measurement} to access the reflection measurement
mode.
Functions and Measurements
2. Press {Disp Mode} > {Return Loss} to choose the
display mode.
3. Press [Freq] > {Start Freq} > 5 > {MHz} to set the
start frequency to 5 MHz.
4. Press {Stop Freq} > 7 > {GHz} to set the stop
frequency to 7 GHz.
5. Press {Calibrate} to calibrate the anaylzer with
the T- combo calibrator. A Calibrated indicator is
shown on the upper left corner of the screen when
the calibration is completed.
6. Remove the calibrator and connect the cable to
the RF Output connector. The return loss
measurement starts and the results are shown as
figure below.
Figure 35 Reflection Measurement - Return Loss
Cable & Antenna Test
85
Functions and Measurements
Cable & Antenna Test
Measuring Distance to Fault
For this measurement, an internally generated
signal emits from the RF OUT port of the tester
and is transmitted to the cable and antenna system
(DUT). The tester receives and processes the signal
power reflected from faults and imperfections. In
performing a DTF measurement, the tester uses
frequency domain reflectometry. The transmitted
and reflected signals contain information about the
distance to the faults. This information is used to
determine the physical distance to the faults. The
tester displays the physical distance to the
probable faults, corrected for cable loss and the
velocity propagation factor of the cable.
Perform a Distance-To-Fault measurement
Perform the following steps to make a
Distance- To- Fault measurement over a frequency
range of 5 MHz to 7 GHz:
1 Press the [MODE] > {Cable&Antenna Test}>{Distance
3 Press {Stop Freq} > 7 > {GHz} > [ENTER] to set the
stop frequency to 7 GHz.
4 Press {Calibrate} to bring up the calibration
process guidance. Follow the instructions to
perform a calibration. For more information
about calibration, refer to “Preparation” on
page 83.
5 Remove the calibrator and connect the cable
under test (DUT) to the RF OUT connector on
the top panel of the tester.
6 Press {Meas} > {More} > {Cable Specification} to set
the cable specification. Press {Cable Type} to select
the default cable type. Press {Select Cable} to
choose the standard cable types with fixed cable
attenuation and velocity factor. Or press {Cable Typ e } > {C u s to m } to set the cable attenuation by
86
Functions and Measurements
{Cable Atten} and velocity factor by {Vel Factor}. In
this example, the cable attenuation is 0.48
dB/meter and the velocity factor is 66%.
The DTF measurement starts and the results are
displayed on the screen. A peak in the horizontal
trace indicates a defective location.
Press [Meas] > {Freq Domain On} to enable the split
screen with cable loss trace in frequency domain.
Figure 36 Frequency Domain Split Screen In DTF Test
Cable & Antenna Test
Measurement Functions
The cable and Antenna test mode provides the
following measurement functions to help you with
the complicated field test tasks.
Trac e M a th
Trace math provide a trace comparison function
for your test. Press [Trace] > {Data->Mem} to save
the current trace in yellow to the internal memor y
in blue. Press {Data}, {Memory} or {Data&Memory} to
display the corresponding traces on the screen.
Press {Trace Math} to choose the calculation types
for the trace comparison: {Data+Mem}, {Data-Mem}
87
Functions and Measurements
Cable & Antenna Test
and {Data/Mem}. The result is shown on the screen
in logarithmic power.
Press {Data&Memory} to show the trace math result
in yellow and the saved trace in blue as Figure 37.
Figure 37 Trace Math In Distance To Fault Test
88
Press {A
the current trace in yellow.
verage On} to turn on the trace average for
Interference Immunity
Interference immunity function is provided to filter
the interference signal. This function key is
available in all the measurement modes, including
Reflection Measurement, One Port Cable Loss and
Distance To Fault. Press {IIM On/Off} to toggle this
function on and off.
Pass & Fail Test with Limit and Limit line
Press [Shift] > [Limit/Trace] to use the limit function
for pass & fail test. There are two limits
configuration which allows you to set the upper
limit and lower limit for the trace. Press {Limit 1/2} > {Type} to toggle the limit direction between upper
and lower. Press {Edit Limits} to set the limit points
one by one. Then, press {Return} > {Limit on/off} to
Functions and Measurements
turn on the customized limit function. Then, the
pass and fail indicator is displayed on the screen
according to the customized criteria.
Press {Limit 1/2} > {Limit Line} to set a single limit
line. Input the limit value and then press {Limit Line On} to enable the limit line as Figure 38.
Press {Limit Beep} to turn on the beep indicator. If
the trace is failed to pass the limit line, there will
be continuous beep indicating the failure.
Figure 38 Limit Line in Reflection Measurement
Cable & Antenna Test
89
Functions and Measurements
File Operation
File Operation
Pressing [File] accesses a menu that allows you to
manage file saving and loading.
Viewing a file list
Refer to the following two steps to view a file list:
1 Select the directory to view.
Press [File] > {Directory} > {Media Type} to toggle
the displayed file list from internal memory,
external USB memory stick or system memory.
2 Select the file type you wish to view.
Press [File] > {View FileType} to select a file type.
File Types
Each file type has a specific purpose as defined
below:
• Trace ( * . D AT)
A trace file records trace data and controls.
• Screen (*.JPG)
A screen file records graphic information of the
current screen.
• State (*.STA)
A state file records the current controls and
settings of the analyzer. Use this file type for
saving test parameters for future recall, such as
Frequency, Amplitude and BW settings.
• Limit (*.LIM)
A limit file records the limit settings. You can
edit the limit file and transfer this file between
the N934xC and PC software.
• CSV (*.CSV)
A CSV file records the trace data and available
for review on PC. With option GPS, you can
convert the CSV file into KML file in HSA PC
software for review the GPS location in Google
Earth on PC.
90
Functions and Measurements
• Task Planner (*.TPF)
A task planner file contains the user- defined
measurement tasks in sequence which make the
test task easier in field test.
• Setup (*.SET)
A setup file records the system setting
information, such as language, date/time, and
power saving mode. This file type is not used for
test parameter setup information.
• Standard (*.STD)
A standard file contains the frequency range and
corresponding parameters for a specific
communication standard. It allows you to recall
those default settings for the singal searching
and testing in that range.
• Amplitude Correction (*.COR)
The amplitude correction file is used for trace
amplitude offset compensation. Please refer to
“Correction” on page 98 to edit and save the
correction file on HSA PC software.
• KML (*.KML)
A KML file is available for Google Earth on PC.
It is specified for channel scanner measurement
and the GPS location is only available when the
option GPS is activated.
Figure 39 Review the KML file in Google Earth
File Operation
91
Functions and Measurements
NOTE
File Operation
Saving a file
Refer to the following three steps to save a file:
1 Press [File] > {Directory} > {Media Type} to select the
file directory.
2 Press [File] > {Save as}, The pop- up window
displays for your further setting.
3 Edit a file name.
A file name can consist of letters and digits. A
sing le key stroke on the numerical keypad inputs
a digital number; and consecutive key stroke
selects and inputs a letter.
4 Rotate the knob to choose the file type you need.
5 Press [Enter] to save the file.
When a file saving completes, the saved file will
display in the file list..
When a file is selected, you can edit the file name by pressing
the numeric and alphabetic hardkeys on the right side of the
analyzer’s front panel.
Quick Saving a File
Press [Save] to quickly save a trace or a screen to
either the local memory or an external USB device,
depending upon the setup of the save path.
Refer to the steps for saving a trace/screen:
1 Press [File] > {Setup} > {Qsave Type} to set the quick
save file type.
2 Press [File] > {Directory} > {Media Type} to select the
file directory.
3 Press [Return], [Save] to save a trace or a screen
copy to the pre- defined memory. The trace or
screen will be automatically assigned with a
default file name. A trace will be assigned with a
file name, such as HYTRACE, HYTRACE_1, and
consecutive decimal numbers appended to the
alphabetic file name, such as HYSCREEN, HYSCREEN_1 and consecutive decimal numbers
appended to the alphabetic file name.
92
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.