Used Agilent Agilent Used N9320B Manual

Page 1
Keysight N9320B Spectrum Analyzer
Notice: This document contains references to Agilent. Please note that Agilent’s Test and M
easurement business has become Keysight
Technologies.
User’s Guide
Page 2

Notices

© Keysight Technologies, Inc. 2006-2014
No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval or translation into a foreign language) without prior agreement and written consent from Keysight Technologies, Inc. as governed by United States and international copyright laws.
Trademark Acknowledgments
Manual Part Number
N9320-90007
Edition
Edition 4, June 2014
Printed in China
Published by: Keysight Technologies
No 116 Tianfu 4th street Chiengdu, 610041 China
Warranty
THE MATERIAL CONTAINED IN THIS DOCUMENT IS PROVIDED “AS IS,” AND IS SUBJECT TO BEING CHANGED, WITHOUT NOTICE, IN FUTURE EDITIONS. FURTHER, TO THE MAXIMUM EXTENT PERMITTED BY APPLICABLE LAW, KEYSIGHT DISCLAIMS ALL WARRANTIES, EITHER EXPRESS OR IMPLIED WITH REGARD TO THIS MANUAL AND ANY INFORMATION CONTAINED HEREIN, INCLUDING BUT NOT LIMITED TO THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. KEYSIGHT SHALL NOT BE LIABLE FOR ERRORS OR FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES IN CONNECTION WITH THE FURNISHING, USE, OR PERFORMANCE OF THIS DOCUMENT OR ANY INFORMATION CONTAINED HEREIN. SHOULD KEYSIGHT AND THE USER HAVE A SEPARATE WRITTEN AGREEMENT WITH WARRANTY TERMS
COVERING THE MATERIAL IN THIS DOCUMENT THAT CONFLICT WITH THESE TERMS, THE WARRANTY TERMS IN THE SEPARATE AGREEMENT WILL CONTROL.
Technology Licenses
The hardware and/or software described in this document are furnished under a license and may be used or copied only in accordance with the terms of such license.
U.S. Government Rights
The Software is “commercial computer software,” as defined by Federal Acquisition Regulation (“FAR”) 2.101. Pursuant
12.212 and 27.405-3 and Department of Defense FAR Supplement (“DFARS”) 227.7202, the U.S. government acquires commercial computer software under the same terms by which the software is customarily provided to the public. Accordingly, Keysight provides the Software to U.S. government customers under its standard commercial license, which is embodied in its End User License Agreement (EULA). The license set forth in the EULA represents the exclusive authority by which the U.S. government may use, modify, distribute, or disclose the Software. The EULA and the license set forth therein, does not require or permit, among other things, that Keysight: (1) Furnish technical information related to commercial computer software or commercial computer software documentation that is not customarily provided to the public; or (2) Relinquish to, or otherwise provide, the government rights in excess of these rights customarily provided to the public to use, modify, reproduce, release, perform, display, or disclose commercial computer software computer software
to FAR
or commercial
documentation. No additional government requirements beyond those set forth in the EULA shall apply, except to the extent that those terms, rights, or licenses are explicitly required from all providers of commercial computer software pursuant to the FAR and the DFARS and are set forth specifically in writing elsewhere in the EULA. Keysight shall be under no obligation to update, revise or otherwise modify the Software. With respect to any technical data as defined by FAR 2.101, pursuant to FAR 12.211 and 27.404.2 and DFARS 227.7102, the U.S. government acquires no greater than Limited Rights as defined in FAR 27.401 or DFAR 227.7103-5 (c), as applicable in any technical data.
Safety Notices
A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.
A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met.
Page 3
1 Overview 1
Keysight N9320B at a Glance 2
Front Panel Overview 4
Rear Panel Overview 9
Front and rear panel symbols 10
2 Getting Started 11
Check the Shipment and Order List 12
Safety Notice 14
Power Requirements 15
Power On and Check 17
Environmental Requirements 19
South Korea Class A EMC Declaration 22
Helpful Tips 23

Contents

Perform a Self Alignment 23
Perform a Time Base Calibration 23
Using an External Reference 24
Enable an Options 24
Firmware Upgrade 25
IO Configuration 25
Power Preset Last 26
3 Functions and Measurements 27
Making a Basic Measurement 28
Using the Front Panel 28
Presetting the Spectrum Analyzer 29
1
Page 4
Contents
Viewing a Signal 30
Measuring Multiple Signals 32
Comparing Signals on the Same Screen Using Marker Delta 32
Comparing Signals not on the Same Screen Using Marker Delta 34
Resolving Signals of Equal Amplitude 36
Resolving Small Signals Hidden by Large Signals 39
Measuring a Low-Level Signal 41
Reducing Input Attenuation 41
Decreasing the Resolution Bandwidth 43
Trace Averaging 44
Improving Frequency Resolution and Accuracy 46
Tracking Drifting Signals 48
Making Distortion Measurements 50
Identifying Analyzer Generated Distortion 50
Third-Order Intermodulation Distortion 53
Measuring Phase Noise 56
Stimulus Response Transmission 57
Measuring Stop Band Attenuation of a Lowpass Filter 60
Making a Reflection Calibration Measurement 63
Measuring Return Loss Using the Reflection Calibration Routine 66
Making an Average Power Measurement 67
Demodulate the AM/FM signal 71
Demodulating an AM Signal 71
Demodulating an FM Signal 72
2
Page 5
Analysis the Modulated Signals 74
AM/FM Modulation Analysis 74
ASK/FSK Modulation Analysis 77
Measuring Channel Power 80
Viewing Catalogs and Saving Files 83
Locating and Viewing Files in the Catalog 83
Saving a File 84
Loading a File 86
Copying a File 87
Deleting a File 87
4 Key Reference 89
Amplitude 90
Auto Tune 94
Contents
Back <- 95
BW/Avg 96
Det/Display 100
Enter 106
File 107
Frequency 113
Marker 115
Marker-> 121
Meas 122
Channel Power 122
Occupied BW 124
ACP 127
3
Page 6
Contents
Intermod (TOI) 130
Spectrum Emission Mask 132
MODE 139
Tracking Generator 139
Power Meter 143
AM/FM Modulation Analysis 146
ASK/FSK Modulation Analysis 150
Peak Search 154
Preset/System 158
SPAN 164
Sweep/Trig 166
Trace 168
5 Instrument Messages 171
Overview 172
Command Errors 173
Execution Conflict 175
Device-Specific Errors 177
6 Troubleshooting 181
Check the basics 182
Contact Keysight Technologies 184
Index 207
4
Page 7
7 Menu Maps 185
Amplitude Menu 186
BW/Avg Menu 187
Det/Display Menu 188
File Menu (1 of 2) 189
File Menu (2 of 2) 190
Frequency Menu 191
Marker Menu 192
Marker-> Menu 193
Measure Menu (1 of 2) 194
Measure Menu (2 of 2) 195
MODE - Tracking Generator 196
MODE - Power Meter 197
Contents
MODE - AM/FM Modulation Analysis 198
MODE - ASK/FSK Modulation Analysis 199
Name Editor Menu 200
Peak Search Menu 201
Preset/System Menu 202
SPAN Menu 203
Sweep/Trig Menu 204
Trace Menu 205
5
Page 8
Contents
6
Page 9

Overview

1 Overview
Keysight N9320B at a Glance 2
Front Panel Overview 4
Rear Panel Overview 9
Front and rear panel symbols 10
This chapter provides a description of the Keysight N9320B spectrum analyzer and an introduction to the buttons, features, and functions of the front and rear instrument panels.
1
Page 10
Overview Keysight N9320B at a Glance

Keysight N9320B at a Glance

The Keysight N9320B spectrum analyzer is a portable, swept spectrum can be a fundamental component of an automated system. It can also be widely used in an electronic manufacturing environment and in functional/final/QA test systems.

Features

The Keysight N9320B spectrum analyzer primary features and functions are described below:
• High Sensitive Measurement
The spectrum analyzer includes an optional pre-amplifier for signals in the frequency range up to 3 GHz, enabling more sensitive measurements. This feature is a great help in analysis of weaker signals.
• High Accuracy Power Measurement
analyzer with a frequency range of 9 kHz to 3.0 GHz. It
The N9320B supports Keysight U2000 series power sensors for high accuracy power measurement as a power meter.
• Power Measurement Suite
The built-in one-button power measurement suite offers channel power, ACP, OBW, and TOI measurements.
• Spectrum Emission Mask
provides a Pass/Fail testing capability with a mask for out-of-channel emissions measurement.
2
Page 11
Overview
Keysight N9320B at a Glance
• Modulation Analysis Function
provides optional AM/FM and ASK/FSK modulation analysis function. (AM/FM: Option-AMA ASK/FSK: Option-DMA)
• Tracking Generator
provides an RF source for scalar network analysis (Option-TG3).
3
Page 12
Overview
Local
Save
PROBE POWER
RF IN 50
50VDC MAX 30dBm 1W MAX
TG SOURCE CAL OUT
50MHz 10dBm
Remot e
Standby On
7
·
4
102
596
3
Back
Frequency
SPAN
Marker
Peak Search
Marker
Auto Tune
Det/ Display
File/ Print
BW/ Avg
View/ Trace
MeasMODE
Preset/ System
Amplitude
Enter
Sweep/ Trig
8
N9320B 9 kHz - 3.0 GHz
SPECTRUM ANALYZER
1
8
9
10
11
12
18
17
16
15
14
7
5
6
4
3
2
13

Front Panel Overview

Front Panel Overview
1Screen 6.5 inch TFT color screen shows information of the
current function, including the signal traces, status indicators, and instrument messages. Labels for softkeys are located on the right-hand side of the screen.
2 Softkeys are the unlabeled keys next to the screen. They activate
functions displayed to the left of each key.
3 Amplitude activates the reference level function and accesses
the amplitude softkeys, with which you set functions that affect data on the vertical axis.
4 SPAN sets the frequency range symmetrically about the center
frequency. The frequency-span readout describes the total displayed frequency range.
5 Frequency activates the center-frequency function, and accesses
the menu of frequency functions.
6 Function Keys relate directly to the following main functions:
4
Preset/System (Local) accesses the softkeys to reset the
analyzer to a known state, if the analyzer is in the remote mode, pressing this key returns the analyzer to the local mode and enables front-panel control.
Page 13
Overview
Front Panel Overview
Auto Tune searches the signal automatically and locates the
signal to the center of the graticule. see
page 94
BW/Avg activates the resolution bandwidth function and
accesses the softkeys that control the bandwidth functions and averaging.
Sweep/Trig accesses the softkeys that allow you to set the
sweep time, select the sweep mode and trigger mode of the analyzer.
View/Trace accesses the softkeys that allow you to store and
manipulate trace information.
Det/Display accesses the softkeys that allow you to configure
detector functions and control what is displayed on the analyzer, including the display line, graticule and annotation, as well as the testing of trace data against entered limits.
MODE selects the measurement mode of your analyzer.
Meas accesses the softkeys that let you make transmitter
power measurements such as adjacent channel power, occupied bandwidth, and harmonic distortion, etc.
Marker accesses the marker control keys that select the type
and number of markers and turns them on and off.
Marker accesses the marker function softkeys that help you
with the measurement.
Peak Search places a marker on the highest peak.
File accesses the softkeys that allow you to configure the file
system of the analyzer.
.
"Auto Tune" on
7 Arrow Keys The up and down arrow keys shift the selected item
when you press [MODE] hardkey; you can also change the mode by rotating the knob.
8 Knob The front panel knob increases or decreases a value, a
numeric digit, or scrolls up and down to select an item in a list.
5
Page 14
Overview
CAUTION
Front Panel Overview
9 Data Control Keys includes the numeric keypad, ENTER key and
backspace key, change the numeric value of an active function such as center frequency, start frequency, resolution bandwidth, and marker position.
10 RF IN connector is the signal input for the analyzer. The
maximum damage level is average continuous power +40 dBm, DC voltage 50 VDC or max pulse voltage 125 V. The impedance is 50  (N-type female).
11 PROBE POWER connector provides power for high-impedance
AC probes or other accessories (+15 V, –12 V, 150 mA maximum).
12 CAL OUT connector provides an amplitude reference signal
output of 50 MHz at –10 dBm (BNC female).
13 TG SOURCE connector N-type female, is the source output for
the built-in tracking generator. The impedance is 50 . (for Option TG3)
If the tracking generator output power is higher than the maximum power that the device under test can tolerate, it may damage the device under test. Do not exceed the maximum power of DUT.
6
14 Standby Switch switches on all functions of the analyzer. To
switch the analyzer off, press the switch for at least 2 seconds. This deactivates all the functions but retains power to internal circuits so long as the analyzer is connected to line power.
15 On LED (green) lights when the analyzer is switched on.
16 Standby LED (orange) lights when the analyzer is connected to
the line power.
17 Remote LED lights when the analyzer is remotely controlled by a
PC via the USB host interface on the rear panel.
18 USB Device Connector provides a connection between external
USB devices and the analyzer, such as a USB memory device.
Page 15

Display Annotations

12
11
10
8
5
13
14
15
3
2
1
26
25
22
24
17
16
23
21
20
6
9
7
1819
4
Overview
Front Panel Overview
Item Description Notes (Associated function key)
1 Amplitude scale [Amplitude] > Scale Type
2 Detector mode [Det/Display] > Detector
3 Reference level [Amplitude]
4 Active function block The function in use
5 Time and date display [Preset/System] > Time/Date
6 RF attenuation [Amplitude] > Attenuation Auto
7 Marker frequency [Marker] or
> Ref Level
[Marker] > Function > Frequency
Counter
8 Uncal indicator The readout of amplitude is
uncalibrated.
9 Marker amplitude [Marker]
10 External reference An external frequency reference is in use.
7
Page 16
Overview Front Panel Overview
11 Remote mode The analyzer is in remote mode
12 Key menu title Dependent on key selection.
13 Softkey menu Refer to
Dependent on current function key
selection.
14 Frequency span [SPAN]
15 Sweep time [Sweep/Trig] > Sweep Time
16 Video bandwidth [Bw/Avg] > Video BW
17 Display status line Display status and instrument messages.
18 Resolution bandwidth [Bw/Avg] > Res BW
19 Trigger/Sweep
F - free run trigger L - line trigger V - video trigger E - external trigger C - continuous sweep S - single sweep
20 Continuous peak [Peak Search] > Continuous pk
21 Signal track [SPAN] > Span
22 Internal preamp [Amplitude] > Int Preamp
23 Key menu title Dependent on key selection
24 Trace mode
W - clear write M - maximum hold m - minimum hold V - view S - store blank
25 Average
VAvg - video average PAvg - power average
26 Display line [Det/Display] > display Line On Off
[Sweep/Trig]
[Trace]
[Bw/Avg] > Average On Off
"Key Reference" for details.
8
Page 17

Rear Panel Overview

REF IN
REF OUT
TRIG IN
USB
LAN
VGA OUT
T T L
~100-240
V
50-60
Hz
100
W MAX
SERIAL LABEL ATTACH HERE
HIPOT LABEL ATTACH HERE
10MHz
10MHz
GPIB
1
2
9
4
5
6
7
8
3
10
1 REF OUT connector provides a frequency of 10 MHz, amplitude of
2 REF IN connector accepts an external timebase with a frequency
3 Kensington Lock lock the instrument and keep its safety. 4 LAN port A TCP/IP Interface that is used for remote analyzer
5 EXT TRG IN (TTL) connector accepts an external voltage input,
6 Power switch isolates the analyzer from the AC line power. After
7 AC Power Receptacle accepts a three-pin line power plug.
Overview
Rear Panel Overview
–10 dBm reference output. (BNC female)
of 10 MHz, amplitude of –5 to +10 dBm. (BNC female)
operation.
the positive edge of which triggers the analyzer sweep function. (BNC female)
switch on, the analyzer enters into standby mode and the orange standby LED on the front panel lights.
8 VGA connector provides the video output signal to an external
monitor or projector. (D-sub 15-pin female)
9 USB Host connector provides a connection between the analyzer
and an PC for remote control.
10 GPIB connector (Option G01) is an optional interface. GPIB
supports remote analyzer operation.
9
Page 18
Overview
ISM1-A
C
US
ICES/NMB-001

Front and rear panel symbols

Front and rear panel symbols
Instruction manual symbol: indicates that the user must refer to specific instructions in the manual.
CE mark: a registered trademark of the European Community.
Shows that this is an Industrial Scientific and Medical Group 1 Class A product. (CISPR 11, Clause 4)
The CSA mark: a registered trademark of the Canadian Standards Association International.
The ISM device complies with Canadian Interference-Causing Equipment Standard-001. Cet appareil ISM est conforme à la norme NMB-001 du Canada.
All Level 1, 2 or 3 electrical equipment offered for sale in Australia and New Zealand by Responsible Suppliers must be marked with the Regulatory Compliance Mark (RCM Mark).
marks the “on/standby” position of the switch.
10
indicates that the instrument requires AC power input.
indicates this product complies with the WEEE Directive(2002/96/EC) marking requirements and you must not discard this equipment in domestic household waste. Do not dispose in domestic household waste.
indicates the time period during which no hazardous or toxic substance elements are expected to leak or deteriorate during normal use. Forty years is the expected useful life of the product.
KC mark: as Korea Certification. This equipment is Class A suitable for professional use and is for use in electromagnetic environments outside of the home.
Page 19

Getting Started

2 Getting Started
Check the Shipment and Order List 12
Safety Notice 14
Power Requirements 15
Power On and Check 17
Environmental Requirements 19
South Korea Class A EMC Declaration 22
Helpful Tips 23
This chapter helps you in preparing the spectrum analyzer for use and provides the information to start using the spectrum analyzer correctly.
11
Page 20
Getting Started Check the Shipment and Order List

Check the Shipment and Order List

After receiving the shipment, first check the shipment and your order list.
Inspect the shipping container and the cushioning material for signs of stress. Retain the shipping materials for future use, as you may wish to ship the analyzer to another location or to Keysight Technologies for service. Verify that the contents of the shipping container are complete.
Each spectrum analyzer includes the following accessories as standard:
Item Description
USB cable (A-B) Connection for remote control
N-BNC adapter and BNC cable Connection for alignment
Three-pin power cord Specific to shipping location
Help kit CD-ROM HSA and BSA PC Software
Keysight IO libraries suite N9320B documentation
12
Shipping Problems?
If the shipping materials are damaged or the contents of the container are incomplete:
• Contact the nearest Keysight Technologies office to arrange for repair or replacement. You will not need to wait for a claim settlement.
• Keep the shipping materials for the carrier’s inspection.
• If you must return an analyzer to Keysight Technologies, use the original (or comparable) shipping materials.
For any questions about your shipment, please contact Keysight Technologies for consulting and service.
Page 21

Options

Getting Started
Check the Shipment and Order List
Verify if that the shipment includes your ordered options by checking the option label on the shipping container:
Option Description Part number
PA3 3 GHz Preamplifier N9320B-PA3 G01 GPIB interface N9320B-G01 AMA AM/FM demodulation N9320B-AMA DMA ASK/FSK demodulation N9320B-DMA TG3 3 GHz tracking generator N9320B-TG3 EMF EMI filter N9320B-EMF G01 GPIB interface N9320B-G01 TR1 RF training kit N9320B-TR1 1HB Handle and bumpers N9320B-1HB 1CM Rackmount kit N9320B-1CM 1TC Hard transit case N9320B-1TC UK6 Commercial calibration certificate
with testing data
N9320B-UK6
Unless specified otherwise, all options are available when you order a spectrum analyzer; some options are also available as kits that you can order and install/activate after you receive the analyzer. Order kits through your local Keysight Sales and Service Office.
At the time of analyzer purchase, options can be ordered using your product number and the number of the option you are ordering. For example, if you are ordering option TG3, you would order N9320B-TG3.
If you are ordering an option after the purchase of your analyzer, you will need to add a K to the product number and then specify which option you are ordering. For example, N9320BK-TG3.
13
Page 22
Getting Started
WARNING
WARNING
WARNING
CAUTION
CAUTION

Safety Notice

Safety Notice
Read the following warnings and cautions carefully before powering on the spectrum analyzer to ensure personal and instrument safety.
Always use a well-grounded, three-pin AC power cord to connect to power source. Personal injury may occur if there is any interruption of the AC power cord. Intentional interruption is prohibited. If this product is to be energized via an external auto transformer for voltage reduction, make sure that its common terminal is connected to a neutral (earthed pole) of the power supply.
Personal injury may result if the spectrum analyzer covers are removed. There are no operator-serviceable parts inside. To avoid electrical shock, refer servicing to qualified personnel.
14
Electrical shock may result if the spectrum analyzer is connected to the power supply while cleaning. Do not attempt to clean internally.
Prevent damage to the instrument and ensure protection of the input mixer by limiting average continuous power input to +33 dBm, DC voltage to 50 VDC with > Instrument damage may result if these precautions are not followed.
To install the spectrum analyzer in other racks, note that they may promote shock hazards, overheating, dusting contamination, and inferior system performance. Consult your Keysight customer engineer about installation, warranty, and support details.
10 dB input attenuation.
Page 23

Power Requirements

AC Power Cord

Getting Started
Power Requirements
The spectrum analyzer has an auto-ranging line voltage input. The AC power supply must meet the following requirements:
Voltage: 100 to 240 VAC (90 to 264 VAC)
Frequency: 50 to 60 Hz
Power: Maximum 100 W
The analyzer is equipped with a three-wire power cord, in accordance with international safety standards. This cable grounds the analyzer cabinet when connected to an appropriate power line outlet. The cable appropriate to the original shipping location is included with the analyzer.
Various AC power cables are available that are unique to specific geographic areas. You can order additional AC power cables for use in different areas. The table AC Power Cords lists the available AC power cables, the plug configurations, and identifies the geographic area in which each cable is appropriate.
The detachable power cord is the product disconnecting device. It disconnects the mains circuits from the mains supply before other parts of the product. The front panel switch is only a standby switch and do not disconnect instrument from LINE power.
15
Page 24
Getting Started
250V 10A
250V 10A
250V 16A
230V 15A
250V 16A
Power Requirements
AC Power Cords
Plug Type Cable Part
Number
8121-1703 BS 1363/A Option 900
8120-0696 AS
8120-1692 IEC 83 C4 Option 902
8120-1521 CNS 10917-2
125V 10A
8120-2296 SEV 1011 Option 906
250V 10A
8120-4600 SABS 164-1 Option 917
8120-4754 JIS C8303 Option 918
125V 15A
a
Plug Description
3112:2000
/NEMA 5-15P
For use in Country & Region
United Kingdom, Hong Kong, Singapore, Malaysia
Option 901
Australia, New Zealand
Continenta l Europe, Korea, Indonesia, Italy, Russia
Option 903
United States, Canada, Tai w an, M e xico
Switzerland
South Africa, India
Japan
8120-5181 SI 32 Option 919
Israel
8120-8377 GB 1002 Option 922
China
250V 10A
a. Plug description describes the plug only. The part number is for the complete cable assembly.
16
Page 25

Power On and Check

NOTE
1 Connect the AC power cord into the instrument. Insert the plug
2 Press the AC line switch on the rear panel. The standby LED on
Getting Started
Power On and Check
into a power socket provided with a protective ground. Set the tilt adjustors for your preference.
Figure 1 Plug in and angle adjustment
the front panel will light and the spectrum analyzer is in standby mode (AC power applied).
3 Press the standby switch on the front panel. The On LED will
light, and the spectrum analyzer boots up.
Self-initialization takes about 25 seconds; the spectrum analyzer then defaults to the menu mode. After power on, let the spectrum analyzer warm up for 45 minutes for stabilization.
The front panel switch is a standby switch only; it is not a power switch. To completely disconnect the spectrum analyzer from the AC line power, shut off the power switch on the rear panel.
17
Page 26
Getting Started
Preset/ System
Preset/ System
Preset/ System
Power On and Check

Running Internal Alignments

Check for Instrument Messages

To meet the instrument performance specifications, the analyzer must periodically be manually aligned.
1 Connect a BNC cable with the N-BNC adapter between the CAL
OUT and RF IN front panel connectors.
2 After 15-minute warm up, press > Alignment > Align > All
The self alignment takes about 5 minutes.
When an alignment is being run, there will be an audible clicking sound as the attenuator settings are changed. This sound is not an indication of a problem.
Please refer to
"Alignment" on page 159 for further information.
The spectrum analyzer has two categories of instrument messages: error and warning messages. A error message is triggered by operation errors, for example, parameter setting conflicts or data input that is out of the range of a parameter. An warning message may be triggered by hardware defects which could result in damage to instrument.
Here are some tips to check the instrument messages.
1 Check the display to see if any messages display in the status bar.
Press > More > Show Errors to review each message. Refer to
Chapter 5, "Instrument Messages" for detailed system message
descriptions.
2 When you have reviewed and resolved all of the error messages,
press > More > Show Errors > Clear error queue to delete the messages.
3 Cycle the power to the analyzer and re-check to see if the
instrument messages are still there.
4 If the error messages cannot be resolved, please contact the
Keysight Customer Contact Center for assistance or service.
18
Page 27

Environmental Requirements

Keysight Technologies has designed this product for use in Installation Category II, Pollution Degree 2, per IEC 61010-1. Keysight has designed the spectrum analyzer for use under the following conditions:
• Indoor use
• Altitude < 3,000 meters
• Operating temperature range: +5 to +45
Storage temperature range: –20 to +70
• Relative humidity range 15% to 95% at 40

Ventilation

Ventilation holes are located on the rear panel and one side of the spectrum analyzer cover. Do not allow these holes to be obstructed, as they allow air flow through the spectrum analyzer.
When installing the spectrum analyzer in a cabinet, do not restrict the convection of the analyzer. The ambient temperature outside the cabinet must be less than the maximum operating temperature of the spectrum analyzer by 4 within the cabinet.
Getting Started
Environmental Requirements
o
C;
o
C
o
C
o
C for every 100 watts dissipated

Cleaning Tips

To prevent electrical shock, disconnect the spectrum analyzer from line power before cleaning. Use a dry cloth or one slightly dampened with water to clean the external case parts. Do not attempt to clean internally.
19
Page 28
Getting Started Environmental Requirements

Rack Mount

It is recommended to use the Keysight rackmount kit (option 1CM) to install the spectrum analyzer into a rack.
Do not attempt to rack mount the spectrum analyzer by the front panel handles only. This rackmount kit will allow mounting of the spectrum analyzer with or without handles.
Refer to the following instructions when installing the rackmount kit on the spectrum analyzer.
1 Remove feet, key-locks and tilt stands.
2 Remove side trim strips and one middle screw per side.
20
Page 29
Getting Started
CAUTION
Environmental Requirements
3 Attach rackmount flange and front handle assembly with 3 screws
per side.
4 Attach the spectrum analyzer to the rack using the rackmount
flanges with two dress screws per side.
Installing the spectrum analyzers into other racks may promote shock hazards, overheating, dust contamination, and inferior system performance. Consult your Keysight customer engineer about installation, warranty, and support details.
21
Page 30
Getting Started South Korea Class A EMC Declaration

South Korea Class A EMC Declaration

This equipment is Class A suitable for professional use and is for use in electromagnetic environments outside of the home.
22
Page 31

Helpful Tips

Preset/ System

Perform a Self Alignment

Getting Started
Helpful Tips
The following contains information to help in using and maintaining the instrument for optimum operation, including alignment, external reference, firmware update and option activation.
The N9320B provides a self alignment function to manually align the amplitude based on the time base. The analyzer should warm up for approximately 30 minutes before self alignment.
When the self aligment function is triggered, the current measurement is interrupted and a gauge displays on the LCD. The gauge simply indicates alignment is in process. When the aligment is finished, the interrupted measurement restarts.
Please refer to the operation procedures as below:
1 Connect a BNC cable with a N-BNC adapter between the CAL OUT
and RF IN front panel connectors.
2 Press > Alignment > Alignment > Align (Ext Cable) to initiate
the alignment.

Perform a Time Base Calibration

The N9320B provides a manual calibration function to calibrate the time base. The analyzer should warm up for approximately 30 minutes before calibration.
When the calibration function is triggered, the current measurement is interrupted and a gauge displays on the LCD. The gauge simply indicates calibration action rather than calibration course.
23
Page 32
Getting Started
Preset/ System
Preset/ System
Helpful Tips
Please refer to the operation procedures as below:
1 Connect a 10 MHz reference TTL signal from external signal
generator to EXT TRG IN connectors on rear panel.
2 Press > Alignment > Timebase to initiate a calibration.
When the calibration is finished, the interrupted measurement restarts.

Using an External Reference

To use an external 10 MHz source as the reference frequency, connect the external reference source to the REF IN connector on the rear panel. An EXT REF indicator will display in the upper bar of the display. The signal level must be in the range of –5 to +10 dBm. When the reference signal is ready, the instrument switches the reference from time base to the external reference.

Enable an Options

24
Option license key information is required to enable product options. Contact your nearest Keysight office for purchasing a license. Refer to the procedures below to activate the options you have purchased.
1 Press > More > More > Licensing > Option
2 Enter the option number to be enabled. Press [Enter] to confirm
your input.
3 Press License key, enter the license key in the onscreen window.
Press [Enter] to confirm your input and terminate the license key input.
4 Press Activate License. The option
will be enabled immediately.
The analyzer provides the trial license with limited time usage of the interested options.
Page 33

Firmware Upgrade

Preset/ System
Preset/ System
CAUTION
Preset/ System
Preset/ System
Press > More > Show software to view the firmware revision of your analyzer. If you call Keysight Technologies regarding your analyzer, it is helpful to have this revision and the analyzer serial number available.
Follow this procedure to finish the firmware update:
1 Download the firmware package from web. Extract and copy the
file version and folder “n9320b” into the root directory of a USB stick.
2 Plug the USB stick into the connector. Press > Upgrade
The analyzer will perform the update process automatically. The upgrade procedure will take several minutes. Once the upgrade is completed, please follow the instruction to reboot the instrument.
Any interruption during the update process will result in update failure and system data lost. Do not remove the USB storage device until the update is finished.
Getting Started
Helpful Tips

IO Configuration

The N9320B spectrum analyzer provides three types of IO connnection: USB, LAN and optional GPIB interface. Press > {IO Configure} to select the corresponding interface as your need.
USB
Select USB to enable the USB connection for remote control.
LAN
The N9320B supports LAN port connection for remote control. Press > {More} > {More} > {IP Admin} to set the IP parameters for the network connection.
25
Page 34
Getting Started
Preset/ System
Preset/ System
Preset/ System
Preset/ System
Helpful Tips
Press {IP Select Static} to manually set the IP address, gateway and subnet mask with the proper LAN information, or just press {IP Select DCHP} to set the IP address in LAN dynamically according DCHP.
Press {Apply} to enable all the parameters you set.
Press > {IO Configure} > {LAN} to enable the LAN connection according to the IP configuration you set.
GPIB
Press > {More} > {More} > {GPIB Address} to set the GPIB address for the analyzer.
Press > {IO Configure} > {GPIB} to enable the GPIB connection. The GPIB softkey is only vaild after the option G01 installed.
26

Power Preset Last

The option C20 provides the capability to save the last state of the analyzer before power off/power failed (essentially uplugging, replugging).
Press > {Pwr On/Preset} > {Power On Last} to activate this function. For the standard N9320B, this operation only save the last state if the analyzer is turned off by the front panel power switch.
Page 35

Functions and Measurements

3 Functions and Measurements
Making a Basic Measurement 28
Measuring Multiple Signals 32
Measuring a Low-Level Signal 41
Improving Frequency Resolution and Accuracy 46
Tracking Drifting Signals 48
Making Distortion Measurements 50
Measuring Phase Noise 56
Stimulus Response Transmission 57
Measuring Stop Band Attenuation of a Lowpass Filter 60
Making a Reflection Calibration Measurement 63
Measuring Return Loss with the Reflection Calibration Routine 66
Making an Average Power Measurement 67
Demodulate the AM/FM signal 71
Analysis the Modulated Signals 74
Measuring Channel Power 80
Viewing Catalogs and Saving Files 83
This chapter provides information on the analyzer functions and specific measurements capabilities of the spectrum analyzer.
27
Page 36
Functions and Measurements Making a Basic Measurement

Making a Basic Measurement

In this guide, the keys labeled with [ ], for example, [Preset/System] refer to front-panel hardkeys. Pressing many of the hardkeys accesses softkey menus that are displayed along the right side of the screen. The softkey menu labels are aligned so that they are located next to the softkeys at the right side of the display screen. For example, Preset is a softkey menu selection when first pressing [Preset/System].

Using the Front Panel

This section provides you with the information on using the front panel of the spectrum analyzer.
Entering Data
When setting the measurement parameters, there are several ways to enter or modify the value of the active function:
Knob Increments or decrements the current value.
Arrow Keys Increments or decrements the current value by a step unit.
28
Numeric Keys Enters a specific value. Then press the desired terminator (either a
unit softkey, or [Enter] hardkey).
Unit Softkeys Terminate (enter) a value with a unit softkey from the menu.
Enter Key Terminates an entry when no unit of measure is required, or the
instrument uses the default unit.
Back Key To delete
the current input digit prior to entering the value.
Using Softkeys
Softkeys are used to modify the analyzer function parameter settings. Some examples of softkey types are:
Toggle Turn on or off an instrument state.
Submenu Displays a secondary menu of softkeys, {More}.
Choice Selecting from a list of standard values or filenames.
Adjust Highlights the softkey and sets the active function.
Page 37
Functions and Measurements
Preset/ System
Preset/ System
Preset/ System
Preset/ System
Making a Basic Measurement

Presetting the Spectrum Analyzer

Preset function provides a known instrument status for making measurements. There are two types of presets, factory and user:
Factory Preset When this preset type is selected, it restores the analyzer to its
factory-defined state. A set of known instrument parameter settings defined by the factory. Refer to “Factory Default Preset
State” on page 159 for details.
User Preset Restores the analyzer to a user-defined state. A set of user defined
instrument parameter settings saved for assisting the user in quickly returning to known a instrument measurement setup.
Press > Pwr on/Preset > Preset Type to select the preset type.
When Preset Type is set to Factory, pressing > Preset triggers a factory preset condition. The instrument will immediately return to the factory default instrument parameter setting.
When Preset Type is set to User, pressing > Preset displays both User Preset and Factory Preset softkeys. The user may then select the preset desired from the softkey menu selections.
Setting up a User Preset
To quickly return to instrument settings that are user defined, perform the following steps to save the instrument state as the user-defined preset:
1 Set the instrument parameters to the values and settings
necessary for the user preset state. This would include the frequency, span, amplitude, BW, and measurement type and any other setup details desired.
2 Press > Pwr on/Preset > Save User Preset, to save the
current instrument settings as the ‘user preset’ state. The user preset will not affect the default factory preset settings. User preset settings can be changed and saved at any time.
29
Page 38
Functions and Measurements
Preset/ System
Preset/ System
Amplitud
Frequenc
SPAN
NOTE
Peak Search
Making a Basic Measurement

Viewing a Signal

1 Press > Pow on/Preset > Preset Type > Factory to enable the
2 Press > Preset to restore the analyzer to its factory-defined
3 Connect the 10 MHz REF OUT on the rear panel to the front-panel
1 Press > 10 > dBm to set 10 dBm reference level.
2 Press > 30 > MHz to set 30 MHz to center frequency.
Refer to the procedures below to view a signal.
factory-defined preset state.
state.
RF IN.
Setting the Reference Level and Center Frequency
Setting Frequency Span
Press > 50 > MHz to set 50 MHz frequency span.
Changing the reference level changes the amplitude value of the top graticule line. Changing the center frequency changes the horizontal placement of the signal on the display. Increasing the span will increase the frequency range that appears horizontally across the display.
Reading Frequency and Amplitude
1 Press to place a marker (labeled 1) on the 10 MHz peak.
30
Page 39
Functions and Measurements
Marker
10.000000 MHz
0.43 dBm
Active function block
Marker Annotation
Amplitude
Marker
NOTE
Making a Basic Measurement
Note that the frequency and amplitude of the marker appear both in the active function block, and in the upper-right corner of the screen.
Figure 3-1 10 MHz Internal Reference Signal
2 Use the knob, the arrow keys, or the softkeys in the Peak Search
menu to move the marker. The marker information will be displayed in the upper-right corner of the screen.
Changing Reference Level
1 Press and note that reference level (Ref Level) is now the
active function.
2 Press
Changing the reference level changes the amplitude value of the top graticule line.
> Mkr-> Ref Lvl.
31
Page 40
Functions and Measurements
Preset/ System
Frequenc
SPAN
Amplitud
Peak Search
Marker

Measuring Multiple Signals

Measuring Multiple Signals
This section provides the information on how to measure multiple signals.

Comparing Signals on the Same Screen Using Marker Delta

The delta marker function allows the user to compare two signals when both appear on the screen at the same time.
In the following example, harmonics of the 10 MHz reference signal available are used to measure frequency and amplitude differences between two signals on the same screen. Delta marker is used to demonstrate this comparison.
1 Preset the analyzer:
Press > Preset (With Preset Type of Factory)
2 Connect the rear panel REF OUT to the front panel RF IN.
3 Set the analyzer center frequency, span and reference level to view
the 10 MHz signal and its harmonics up to 50 MHz:
32
Press > 30 > MHz
Press > 50 > MHz
Press > 10 > dBm
4 Place a marker at the highest peak on the display (10 MHz):
Press
The marker should be on the 10 MHz reference signal. Use the Next Pk Right and Next Pk Left softkeys to move the marker from peak to peak.
5 Anchor the first marker and activate a second marker:
Press
> Delta > Delta (On)
The label on the first marker now reads 1R, indicating that it is marking the reference point.
Page 41
Functions and Measurements
Peak Search
Peak Search
NOTE
Measuring Multiple Signals
6 Move the second marker to another signal peak using the
front-panel knob or by using Peak Search.
Press > Next Peak or
Press > Next Pk Right or Next Pk Left. Continue pressing the Next Pk softkeys until the marker is on the correct signal peak.
The amplitude and frequency differences between the markers are displayed in the active function block.
Figure 3-2 Delta pair marker with signals on the same screen
To increase the resolution of the marker readings, turn on the frequency count function. For more information, refer to "Improving
Frequency Resolution and Accuracy" on page 46.
33
Page 42
Functions and Measurements
Preset/ System
Frequenc
SPAN
Amplitud
Peak Search
Marker
Marker
Frequenc
FM
Measuring Multiple Signals

Comparing Signals not on the Same Screen Using Marker Delta

1 Preset the analyzer:
2 Connect the rear panel REF OUT to the front panel RF IN.
3 Set the center frequency, span and reference level to view only the
The analyzer will compare the frequency and amplitude differences between two signals which are not displayed on the screen at the same time. (This technique is useful for harmonic distortion tests.)
In this example, the analyzer’s 10 MHz signal is used to measure frequency and amplitude differences between a signal on screen, and another signal off screen. Delta marker is used to demonstrate this comparison.
Press > Preset (With Preset Type set to Factory)
50 MHz signal:
Press > Center Freq > 50 > MHz
Press > Span > 25 > MHz
34
Press > Ref Level > 10 > dBm
4 Place a marker on the 50 MHz peak and then set the center
frequency step size equal to the marker frequency (10 MHz):
Press
Press
> Mkr -> CF Step
5 Activate the marker delta function:
Press
> Delta > Delta (On)
6 Increase the center frequency by 10 MHz:
Press > Center Freq,
The first marker moves to the left edge of the screen, at the amplitude of the first signal peak.
Page 43
Functions and Measurements
Marker
Measuring Multiple Signals
Figure 3-3
shows the reference annotation for the delta marker (1R) at the left side of the display, indicating that the marker set at the 50 MHz reference signal is at a lower frequency than the frequency range currently displayed.
Figure 3-3 Delta Marker with Reference Signal Off-Screen
The delta marker appears on the peak of the 100 MHz component. The delta marker annotation displays the amplitude and frequency difference between the 50 and 100 MHz signal peaks.
7 Press
> Off to turn the markers off.
35
Page 44
Functions and Measurements
Remote
Standby On
TG SOURCE CAL OUT
50MHz 10dBm
7
·
4
102
5896
3
Back
Enter
Marker
Peak
Search
Marker
Auto Tune
Det/ Display
File/ Print
BW/ Avg
View/ Trace
Meas
MODE
Sweep/ Trig
Local
Save
N9320A SPECTRUM ANALYZER 9 kHz - 3.0 GHz
PROBE POWER
RF IN 50
50VDC MAX 30dBm 1W MAX
CAT
Frequency
Enter
7
MOD On/Off
RF
4
102
9
6
3
On/Off
Amplitude FM
Utility
LF Out
Preset
Local
AM I/Q
File
TriggerPulseM
·
Sweep
8
5
Remote
Standby
On
N9310A RF Signal Generator 9 kHz - 3.0 GHz
REVERSE PWR 4W MAX 30VDC
LF OUT RF OUT 50
FUNCTIONS
Frequency
Enter
7
MOD On/Off
RF
4
102
9
6
3
On/Off
Amplitude FM
Utility
LF Out
Preset
Local
AM I/Q
File
TriggerPulseM
·
Sweep
8
5
Remote
Standby
On
N9310A RF Signal Generator 9 kHz - 3.0 GHz
REVERSE PWR 4W MAX 30VDC
LF OUT RF OUT 50
FUNCTION S
Signal Generator
Signal Generator
Spectrum Analyzer
Directional Coupler
RF OUT
RF IN
RF OUT
Preset/ System
Frequenc
BW/ Avg
SPAN
Measuring Multiple Signals

Resolving Signals of Equal Amplitude

In this example a decrease in the resolution bandwidth (RBW) is used in combination with a decrease in video bandwidth (VBW) to resolve two signals of equal amplitude with a frequency separation of 100 kHz.
Figure 3-4 Setup for obtaining two signals
36
Notice that the final RBW selection to resolve the signals is the same width as the signal separation while the VBW is slightly narrower than the RBW.
1 Connect two sources to the analyzer input as shown above.
2 Set one source to 300 MHz. Set the frequency of the other source
300.1 MHz. Set both source amplitudes to –20 dBm.
to
3 Setup the analyzer to view the signals:
Press > Preset (With Preset Type of Factory)
Press > 300 > MHz
Press > 300 > kHz
Press > 2 > MHz
A single signal peak is visible. See
Figure 3-5 for example.
Page 45
Functions and Measurements
SPAN
Peak Search
Frequenc
SPAN
Frequenc
BW/ Avg
BW/ Avg
Measuring Multiple Signals
If the signal peak is not present on the display, increase the frequency span out to 20 MHz, turn signal tracking on, decrease the span back to 2 MHz and turn signal tracking off:
Press > Span > 20 > MHz
Press
Press > Signal Track (On)
Press > 2 > MHz
Press > Signal Track (Off)
Figure 3-5 Unresolved Signals of Equal Amplitude
4 Change the resolution bandwidth (RBW) to 100 kHz so that the
RBW setting is less than or equal to the frequency separation of the two signals:
Press > 100 > kHz
Notice that the peak of the signal has become flattened indicating that two signals are present.
5 Decrease the video bandwidth to 3 kHz:
Press > Video BW > 3 > kHz
Two signals are now visible as shown in Figure 3-6. Use the front-panel knob or arrow keys to further reduce the resolution bandwidth and better resolve the signals.
37
Page 46
Functions and Measurements Measuring Multiple Signals
Decreasing the resolution bandwidth improves the resolution of the individual signals and increases the sweep time.
Figure 3-6 Resolving Signals of Equal Amplitude
For fastest measurement times, use the widest possible resolution bandwidth. Under factory preset conditions, the resolution bandwidth is coupled to the span.
38
Page 47
Functions and Measurements
Preset/ System
Frequenc
BW/ Avg
SPAN
Peak Search
Marker
NOTE
Measuring Multiple Signals

Resolving Small Signals Hidden by Large Signals

This example uses narrow resolution bandwidths to resolve two RF signals that have a frequency separation of 50 kHz and an amplitude difference of 60 dB.
1 Connect two sources to the RF IN as shown in Figure 3-4.
2 Set one source to 300 MHz at –10 dBm. Set the other source to
300.05 MHz at –70 dBm.
3 Set the analyzer as follows:
Press > Preset. (With Preset Type of Factory)
Press > 300 > MHz
Press > 30 > kHz
Press > 500 > kHz
4 Set the 300 MHz signal to the reference level:
Press
Press
> Mkr -> Ref Lvl
The 30 kHz filter shape factor of 15:1 has a bandwidth of 450 kHz at the 60 dB point. The half-bandwidth (225 kHz) is NOT narrower than the frequency separation of 50 kHz, so the input signals can not be resolved.
Figure 3-7 Unresolved small signal from large signal
39
Page 48
Functions and Measurements
BW/ Avg
Peak Search
Marker
NOTE
Measuring Multiple Signals
5 Reduce the resolution bandwidth filter to view the smaller signal.
The smaller signal will be hidden by the larger signal when the bandwidth settings are wider, as in Figure 7. Reducing the RBW setting will allow less of the larger signal to pass through the analyzer and the smaller signals peak will then rise out of the noise floor. Place a delta marker on the smaller signal:
Press > 1 > kHz
Press
Press
> Delta > Delta (On)
Press 50 > kHz
Figure 3-8 Resolved small signal from large signal
The 1 kHz filter shape factor of 15:1 has a bandwidth of 15 kHz at the 60 dB point. The half-bandwidth (7.5 kHz) is narrower than the frequency separation of 50 kHz, so the input signals can be resolved.
40
Page 49

Measuring a Low-Level Signal

CAUTION
Preset/ System
Remote
Standby On
TG SOURCE CAL O UT
50MHz 10dBm
7
·
4
102
5896
3
Back
Enter
Marker
Peak Search
Marker
Auto Tune
Det/ Display
File/ Print
BW/ Avg
View/ Trace
Meas
MODE
Sweep/ Trig
Local
Save
N9320A SPECTRUM ANALY ZER 9 kHz - 3 .0 GHz
PROBE POWER
RF IN 50
50VDC MAX 30dBm 1W MAX
CAT
Frequency
Enter
7
MOD On/Off
RF
4
102
9
6
3
On/Off
Amplitude FM
Utility
LF Out
Preset
Local
AM I/Q
File
TriggerPulseM
·
Sweep
8
5
Remote
Standby
On
N9310A RF Signal Generator 9 kHz - 3.0 GHz
REVERSE PWR 4W MAX 30VDC
LF OUT RF OUT 50
FUNCTION S
Signal Generator
Spectrum Analyzer
RF OUT
RF IN
This section provides information on measuring low-level signals and distinguishing them from spectrum noise.

Reducing Input Attenuation

The ability to measure a low-level signal is limited by internally generated noise of the spectrum analyzer. The analyzers input attenuator affects the level of a signal passing through the analyzer. If a signal power level is close to the noise floor, reducing the analyzer input attenuation will help raise the signal so that it can be seen rising out of the noise.
Ensure that the total power of all input signals at the analyzer RF input does not exceed +30 dBm (1 Watt).
1 Preset the analyzer
Press > Preset (With Preset Type of Factory)
Functions and Measurements
Measuring a Low-Level Signal
2 Set the source frequency to 300 MHz, amplitude to –70 dBm.
Connect the source RF OUT to the analyzer RF IN.
Figure 3-9 Setup for obtaining one signal
41
Page 50
Functions and Measurements
Frequenc
SPAN
Amplitud
Peak Search
Marker
SPAN
Amplitud
Measuring a Low-Level Signal
3 Set the center frequency, span and reference level:
4 Move the desired peak to the center of the display:
5 Reduce the span to 500 kHz, if necessary re-center the peak:
6 Set the attenuation to 20 dB:
Press > Center Freq > 300 > MHz
Press > Span > 2 > MHz
Press > Ref Level > 40 > –dBm.
Press
Press
> Mkr -> CF
Press > 500 > kHz
Press > Attenuation > 20 > dB
Figure 3-10 A signal closer to the noise level
Note that increasing the attenuation moves the noise floor closer to the signal level.
7 To allow more of the signal power to pass through the analyzer,
decrease the attenuation to 0 dB.
42
Page 51
Functions and Measurements
Amplitud
BW/ Avg
FM
Measuring a Low-Level Signal
A lower attenuation value will mean that more of the signal strength will be displayed on screen:
Press > Attenuation > 0 > dB
Figure 3-11 Measuring a low-level signal using 0 dB Attenuation

Decreasing the Resolution Bandwidth

Resolution bandwidth settings affect the level of internal noise but have little affect on the displayed level of continuous wave (CW) signals. Decreasing the RBW by a decade (factor of ten) reduces the noise floor by 10 dB.
1 Refer to the procedure "Reducing Input Attenuation" on page 41
and follow steps 1, 2 and 3.
2 Decrease the resolution bandwidth:
Press and
The low-level signal appears more clearly due to the noise level being reduced by the decrease in RBW (see
Figure 3-12).
43
Page 52
Functions and Measurements
NOTE
Measuring a Low-Level Signal
Figure 3-12 Decreasing Resolution Bandwidth
A “#” mark appears next to the Res BW annotation in the lower left corner of the screen, indicating that the resolution bandwidth is uncoupled.
Uncoupled indicates that the function is in manual control mode, not auto control mode. Manual control mode allows the user to change the parameter value for that function without affecting any other settings.
The analyzer allows you to change the RBW in a 1-3-10 sequence by the data control keys. The RBWs below 1 kHz are digital and have a selectivity ratio of 5:1 while RBWs at 1 kHz and higher have a 15:1 selectivity ratio. The maximum RBW is 3 MHz and minimum is 10 Hz.

Trace Averaging

Averaging is a digital process in which each sweep of the trace returns measurement values for each point in the trace. These values are then mathematically averaged with the previous sweep trace data which has been stored in the analyzer. The amount of averaging is selected by choosing the number of trace sweeps to
44
Page 53
Functions and Measurements
NOTE
BW/ Avg
Enter
Measuring a Low-Level Signal
be included in the process. The averaging function uses the most recent trace sweep values so that the display shows any signal changes. Selecting the detection mode to Sample, smoothing the displayed noise level.
This is a trace processing function and is not the same as using the average detector (as described on page 44).
1 Refer to the procedure "Reducing Input Attenuation" on page 41
of this chapter and follow steps 1, 2 and 3.
2 To turn averaging on, toggle the softkey menu labeled Average:
Press > Average (On)
3 Set the number of averages to 20, using the number keypad, up
and down arrows, or the knob:
Press 20,
The averaging process smooths the viewed trace, low level signals become more visible (see Figure 3-13). Changes to the average number will restart the averaging process.
averaging, when the analyzer is auto-coupled, changes
Figure 3-13 Trace Averaging
45
Page 54
Functions and Measurements
Preset/ System
Preset/ System
Auto Tune
Marker
Marker
50.032500 MHz – 49.30 dBm
NOTE

Improving Frequency Resolution and Accuracy

Improving Frequency Resolution and Accuracy
This section provides information on using the frequency counter function to improve frequency resolution and accuracy.
1 Press > Preset (With Preset Type of Factory)
2 Connect a cable from the front panel CAL OUT to RF IN;
Press > Alignment > Align > CAL OUT ON to toggle on and enable the 50 MHz amplitude reference signal.
3 Press hardkey.
The analyzer will detect the signal peak and locate it to the
the display screen (Refer to “Auto Tune” on page 94).
of
center
4 Turn the frequency counter on:
Press
> Function > Freq Counter > Freq Counter (On).
5 Move the marker, with the front-panel knob, half-way down the
skirt of the signal response.
Figure 3-14 Using Frequency Counter
Counted result
The frequency and amplitude of the marker appears in the active function area (this is not the counted result). The counted result appears in the upper-right corner of the display to the right-side of
Cntr1.
46
Page 55
Functions and Measurements
NOTE
Marker
Marker
NOTE
Improving Frequency Resolution and Accuracy
The marker readout in the active frequency function changes while the counted frequency result (upper-right corner of display) does not. For an accurate count, the marker need not be placed at the exact peak of the signal response.
The Frequency counter properly functions only on stable, CW signals or discrete spectral components. The marker power level must be greater than 40 dB above the displayed noise level.
6 To change the counter resolution:
Press
> Function > Freq Counter > Resolution
The frequency-counter resolution range is from 1 Hz to 1 kHz, and may be set to Auto or Manual.
7 To turn off the marker counter:
Press
> Function > Freq Counter > Freq Counter (Off).
When using the frequency counter function, the ratio of the resolution bandwidth to the span must be greater than 0.02.
47
Page 56
Functions and Measurements
Preset/ System
Frequenc
SPAN
Amplitud
Peak Search
Frequenc
SPAN

Tracking Drifting Signals

Tracking Drifting Signals
This section provides information on measuring and tracking signals that drift in frequency.

Measuring a Source’s Frequency Drift

The analyzer will measure source stability. The maximum amplitude level and the frequency drift of an input signal trace can be displayed and held by using the maximum hold function. Using the maximum hold function you can measure and determine how much of the frequency spectrum a signal occupies. For more information, refer to “Max Hold” on page 168.
Use signal tracking to return a signal drifting in frequency to the center of the display. The drifting is captured by the analyzer using the maximum hold function.
1 Connect the signal generator to the analyzer RF IN.
2 Output a signal with the frequency of 300 MHz and amplitude of
–20 dBm.
3 Set the analyzer center frequency, span and reference level.
Press > Preset. (With Preset Type of Factory)
Press > Center Freq > 300 > MHz
Press > Span > 10 > MHz
Press > Ref Level > –10 > dBm
4 Place a marker on the peak of the signal and turn signal tracking
on:
Press
Press > Signal Track (On)
Press > 1 > MHz
Notice that this holds the signal in the center of the display.
48
Page 57
Functions and Measurements
Frequenc
View/ Trace
View/ Trace
Tracking Drifting Signals
5 Turn off the signal track function:
Press > Signal Track (Off)
6 Measure the excursion of the signal with maximum hold:
Press > Max Hold
As the signal varies, maximum hold maintains the maximum responses of the input signal. Annotation on the left side of the screen indicates the trace mode. For example, M1 S2 S3 S4, indicates trace 1 is in maximum-hold mode, trace 2, trace 3, and trace 4 are in store-blank mode.
7 Activate trace 2 and change the mode to continuous sweep:
Press > Select Trace > Trace 2
Press Clear Write
Trace 1 remains in maximum hold mode to show any drift in the signal.
8 Slowly increase the frequency of the signal generator. Your
analyzer display should look similar to Figure 3-15.
Figure 3-15 Viewing a Drifting Signal With Max Hold and Clear Write
49
Page 58
Functions and Measurements
Preset/ System
Frequenc
SPAN

Making Distortion Measurements

Making Distortion Measurements
This section provides information on measuring and identifying signal distortion.

Identifying Analyzer Generated Distortion

High-level input signals may cause analyzer distortion products that could mask the real distortion measured on the input signal. Use trace and the RF input attenuator to determine which signals, if any, are internally generated distortion products.
In this example, we use the RF output of a signal generator to determine whether the harmonic distortion products are internally generated by the analyzer.
1 Connect the signal generator to the analyzer RF IN. 2 Set the source frequency to 200 MHz, amplitude to 0 dBm. 3 Set the analyzer center frequency and span:
Press > Preset
Press > 400 > MHz
Press > 500 > MHz
Figure 3-16 Harmonic Distortion
50
Page 59
Functions and Measurements
Peak Search
Marker
SPAN
Peak Search
Marker
Amplitud
View/ Trace
View/ Trace
Peak Search
Marker
Amplitud
Making Distortion Measurements
The signal produces harmonic products (spaced 200 MHz from the original 200 MHz signal) in the analyzer input mixer as shown in
Figure 3-16.
4 Change the center frequency to the value of the first harmonic:
Press > Next Peak
Press
5 Change the span to 50 MHz and re-center the signal:
Press > 50 > MHz
Press
Press
6 Set the attenuation to 0 dB:
Press > Attenuation > 0 > dB
7 To determine whether the analyzer generates harmonic distortion
products, first display the trace data in trace 2 as follows:
Press > Select Trace > Trace 2
Press Clear Write
> Mkr -> CF
> Mkr -> CF
8 Allow trace 2 to update (minimum two sweeps), then store the data
from trace 2 and place a delta marker on the harmonic of trace 2:
Press > View
Press
Press
> Delta > Delta (On)
The analyzer display shows the stored data in trace 2 and the measured data in trace 1. The Markerindicator reads the difference in amplitude between the reference and active markers.
9 Increase the RF attenuation to 10 dB:
Press > Attenuation > 10 > dB
51
Page 60
Functions and Measurements
Making Distortion Measurements
Notice the MarkerDamplitude readout. This is the difference of the distortion product amplitude between 0 dB and 10 dB input attenuation settings. If the Markerabsolute amplitude is approximately 1 dB for an input attenuator change, the analyzer is generating, at least in part, the distortion.
The Markeramplitude readout comes from two sources:
1) Increased input attenuation causes poorer signal-to-noise ratio. This causes the Markerto be positive.
2) The reduced contribution of the analyzer circuits to the harmonic measurement causes the Markerto be negative.
Large Markerreadout indicates significant measurement errors. Set the input attenuator to minimize the absolute value of Marker.
52
Page 61
Functions and Measurements
Frequenc y
Enter
7
MOD On/Off
RF
4
102
9
6
3
On/Off
Amplitude FM
Utility
LF Out
Preset
Local
AM I/Q
File
TriggerPulseM
·
Sweep
8
FUNCTIONS
LF OUT RF OUT 50
Remote
Standby On
N9310A RF Signal Generator 9 kHz - 3.0 GHz
5
REVERSE PWR 4W MAX 30VDC
Frequenc y
Enter
7
MOD On/Off
RF
4
102
9
6
3
On/Off
Amplitude FM
Utility
LF Out
Preset
Local
AM I/Q
File
TriggerPulseM
·
Sweep
8
FUNCTIONS
LF OUT RF OUT 50
Remote
Standby On
N9310A RF Signal Generator 9 kHz - 3.0 GHz
5
REVERSE PWR 4W MAX 30VDC
Remote
Standby On
TG SOURCE CAL OUT
50MHz 10dBm
7
·
4
102
5896
3
Back
Enter
Marker
Peak
Search
Marker
Auto Tune
Det/ Display
File/ Print
BW/ Avg
View/ Trace
Meas
MODE
Sweep/ Trig
Local
Save
N9320A SPECTRUM ANALYZER 9 kHz - 3.0 GHz
PROBE POWER
RF IN 50
50VDC MAX 30dBm 1W MAX
CAT
Signal Generator
Signal Generator
Spectrum Analyzer
Directional Coupler
RF OUT
RF IN
300 MHz LOW PASS FILTER
300 MHz LOW PASS FILTER
Making Distortion Measurements

Third-Order Intermodulation Distortion

Two-tone, third-order intermodulation distortion is a common specification in communication systems. When two signals are present in a non-linear system, they may interact and create third-order intermodulation distortion (TOI) products that are located close to the original signals. System components such as amplifiers and mixers contribute to the generation of these distortion products.
For an example of the quick setup of TOI measurement, refer to
“Intermod (TOI)” on page 130.
This example will test a device for third-order intermodulation through the use of markers. Two sources are used, one set to 300 MHz and the other to 301 MHz.
1 Connect the equipment as shown in figure below. This combination
of signal generators, low pass filters, and directional coupler (used as a combiner) results in a two-tone source with very low intermodulation distortion. Although the distortion from this setup may be better than the specified performance of the analyzer, it is useful for determining the TOI performance of the source/analyzer combination. After the performance of the source/analyzer combination has been verified, the device-under-test (DUT) (for example, an amplifier) would be inserted between the directional coupler output and the analyzer input.
53
Page 62
Functions and Measurements
NOTE
Preset/ System
Frequenc
SPAN
BW/ Avg
FM
Peak Search
Marker
BW/ Avg
FM
Marker
Peak Search
Marker
Peak Search
Making Distortion Measurements
2 Set one source (signal generator) to 300 MHz and the other source
3 Set the analyzer center frequency and span:
4 Reduce the RBW until the distortion products are visible:
5 Move the signal to the reference level:
The coupler should have a high degree of isolation between the two input ports so the sources do not intermodulate.
to 301 MHz, for a frequency separation of 1 MHz. Set the sources equal in amplitude as measured by the analyzer (in this example, they are set to –5 dBm).
Press > Preset (With Preset Type of Factory)
Press > Center Freq > 300.5 > MHz
Press > 5 > MHz
Press and
Press
54
Press
> Mkr -> Ref Lvl
6 Reduce the RBW until the distortion products are visible:
Press and
7 Activate the second marker and place it on the peak of the
distortion product (beside the test signal) using the Next PeaK:
Press
> Delta > Delta (On)
Press > Next Peak
8 Measure the other distortion product:
Press
> Normal
Press > Next Peak
Page 63
Functions and Measurements
Marker
Peak Search
Making Distortion Measurements
9 Measure the difference between this test signal and the second
distortion product (see Figure 3-17):
Press
Press > Next Peak
Figure 3-17 Measuring the Distortion Product
> Delta > Delta (On)
55
Page 64
Functions and Measurements
Preset/ System
Preset/ System
Auto Tune
Peak Search
Marker
Marker
Marker D
Noise -97.91 dBc/Hz
20.000 kHz

Measuring Phase Noise

Measuring Phase Noise
Phase noise is a frequency domain measure of stability. Phase noise is specified as single-sideband power in relation to the fundamental RF output frequency. It is measured at various offset frequencies from the carrier, normalized to a 1 Hz measuring bandwidth.
In this example the 50 MHz amplitude reference signal is used as the fundamental signal.
1 Press > Preset (With Preset Type of Factory)
2 Connect a cable from the front panel CAL OUT to RF IN.
Press > Alignment > CAL OUT On
3 Press to auto-tune the signal to the center of display.
4 Press to place a marker on the peak of the signal.
5 Turn on the phase noise function to view the results:
56
Press
> Function > Phase Noise > Phase Noise (On)
6 Change the offset:
Press
> Function > Phase Noise > Offset Manual
Then select the offset by pressing the corresponding softkeys under the Phase Noise menu.
Page 65

Stimulus Response Transmission

Remote
Standby
On
TG SOURCE CAL OUT
50MHz 10dBm
7
·
4
102
5896
3
Back
Enter
Marker
Peak Search
Marker
Auto Tune
Det/ Display
File/ Print
BW/ Avg
View/ Trace
MeasMODE
Sweep/ Trig
Local
Save
N9320A SPECTRUM ANALYZER 9 kHz - 3.0 GHz
PROBE POWER
RF IN 50
50VDC MAX 30dBm 1W MAX
CAT
DUT
Preset/ System
Frequenc
Frequency
BW/ Avg
MODE
CAUTION
The procedure below describes using the built-in tracking generator of the analyzer to measure the rejection of a low pass filter. A type of transmission measurement.
1 To measure the rejection of a low pass filter, connect the
equipment as shown in Figure 3-18. A 370 MHz low pass filter as the DUT.
Figure 3-18 Transmission Measurement Test Setup
Functions and Measurements
Stimulus Response Transmission
2 Press > Preset. (With Preset Type of Factory)
3 Set the start and stop frequencies and resolution bandwidth:
Press > Start Freq > 100 > MHz Press > Stop Freq > 1 > GHz
Press > Res Bw > 3 > MHz
4 Turn on the tracking generator and if necessary, set the output
power to -10 dBm:
Press > Tracking Generator > Amplitude TG (On) > -10 > dBm
Excessive signal input may damage the DUT. Do not exceed the maximum power that the device under test can tolerate.
57
Page 66
Functions and Measurements
NOTE
BW/ Avg
BW/ Avg
MODE
MODE
MODE
Stimulus Response Transmission
5 To increase measurement sensitivity and smooth the noise:
To reduce ripples caused by source return loss, use 6 dB or greater output attenuation. Tracking generator output attenuation is normally a function of the source power selected. However, the output attenuation may be controlled in the Tracking Generator menu.
Press > Res Bw > 30 > KHz
Press > Video Bw > 300 > KHz
A decrease in displayed amplitude is caused by tracking error
.
6 Use peak tracking to correct the frequency offset:
Press > Tracking Generator > More > Tracking Peak
Tracking error occurs when the output frequency of the tracking generator is not matched to the input frequency of the analyzer. The amplitude should return to the value that was displayed prior to the decrease in resolution bandwidth.
7 Connect the cable from the tracking generator output to the
analyzer input. Store the frequency response in trace 3 and normalize:
Press > Tracking Generator > More > Normalize > Store Ref (1
3) > Normalize (On)
8 Reconnect the DUT to the analyzer and change the normalized
reference position:
Press > Tracking Generator > More > Normalize > Norm Ref
Posn > 8
58
Page 67
Functions and Measurements
Marker
Stimulus Response Transmission
9 Measure the rejection of the low pass filter:
Press
The marker readout displays the rejection of the filter at 160 MHz above the cutoff frequency of the low pass filter. See as below.
Figure 3-19 Measure the Rejection Range
> Normal > 370 > MHz, Delta > 160> MHz
Figure 3-19
59
Page 68
Functions and Measurements
Remote
Standby On
TG SOURCE CAL OUT
50MHz 10 dBm
7
·
4
102
5896
3
Back
Enter
Marker
Peak Search
Marker
Auto Tune
Det/ Display
File/ Print
BW/ Avg
View/ Trace
Meas
MODE
Sweep/ Trig
Local
Save
N9320A SPECTRUM ANALYZER 9 kHz - 3.0 GHz
PROBE POWER
RF IN 50
50VDC MAX 30dBm 1W MAX
CAT
DUT
Preset/ System
Frequenc
Frequency
Amplitud

Measuring Stop Band Attenuation of a Lowpass Filter

Measuring Stop Band Attenuation of a Lowpass Filter
When measuring filter performance, it is useful to look at the stimulus response over a wide frequency range. Setting the analyzer y-axis (amplitude) to Log scale type presents the measurement data in a usable format for determining filter characteristics.
The following example uses the tracking generator to measure the stop band attenuation of a 370 MHz low pass filter.
1 To measure the response of a low pass filter, connect the
equipment as shown in 370 MHz low pass filter.
Figure 3-20 Transmission Measurement Test Setup
Figure 3-20. This example uses a
60
2 Press > Preset. (With Preset Type of Factory)
3 Set the start and stop frequencies and resolution bandwidth:
Press > Start Freq > 100 > MHz
Press > Stop Freq > 1 > GHz Press > Scale Type > Log
Page 69
Functions and Measurements
BW/ Avg
CAUTION
MODE
MODE
MODE
Marker
Marker
Measuring Stop Band Attenuation of a Lowpass Filter
4 Set the resolution bandwidth to 10 kHz:
Press > Res Bw > 10 > KHz
Excessive signal input may damage the DUT. Do not exceed the maximum power that the device under test can tolerate.
5 Turn on the tracking generator and, if necessary, set the output
power to -10 dBm
Press > Tracking Generator > Amplitude TG (On) > -10 > dBm
6 Connect the test cable (but not the DUT) from the tracking
generator output to the analyzer input. Store the frequency response into trace 3 and normalize:
Press > Tracking Generator > More > Normalize > Store Ref (1
3) > Normalize (On)
7 Reconnect the DUT to the analyzer. Note that the units of the
reference level have changed to dB, indicating that this is now a relative measurement.
:
8 To change the normalized reference position:
Press > Tracking Generator > More > Normalize > Norm Ref
Posn > 8
9 Place the reference marker at the specified cutoff frequency:
Press
> Delta Pair (Ref) > 370 > MHz
10 Place the second marker at 37 MHz:
Press
> Delta Pair (Delta) > 37 > MHz
61
Page 70
Functions and Measurements Measuring Stop Band Attenuation of a Lowpass Filter
11 In this example, the attenuation over this frequency range is -17.98
dB/octave (one octave above the cutoff frequency).
12 Use the front-panel knob to place the marker at the highest peak in
the stop band to determine the minimum stop band attenuation. In this example, the peak occurs with Delta of 563 MHz. The attenuation is -60.93 dB. See
Figure 3-21 Minimum Stop Band Attenuation
Figure 3-21.
62
Page 71
Functions and Measurements
Remote
Standby On
TG SOURCE CAL OUT
50MHz
10dBm
7
·
4
102
5896
3
Back
Enter
Marker
Peak Search
Marker
Auto Tune
Det/ Display
File/ Print
BW/ Avg
View/ Trace
MeasMODE
Sweep/ Trig
Local
Save
N9320A SPECTRUM ANALYZER 9 kHz - 3.0 GHz
PROBE POWER
RF IN 50
50VDC MAX 30dBm 1W MAX
CAT
Short Circuit
DUT
Reference Plane
or
Coupled Port
Directional
Bridge/Coupler
Test Port
NOTE

Making a Reflection Calibration Measurement

Making a Reflection Calibration Measurement
The following procedure makes a reflection measurement using a coupler or directional bridge to measure the return loss of a filter. This example uses a 370 MHz low pass filter as the DUT.
The calibration standard for reflection measurements is usually a short circuit connected at the reference plane (the point at which the device under test (DUT) is connected.) See Figure 3-22, a short circuit has a reflection coefficient of 1 (0 dB return loss). It reflects all incident power and provides a convenient 0 dB reference.
1 Connect the DUT to the directional bridge or coupler as shown in
Figure 3-22. Terminate the unconnected port of the DUT.
Figure 3-22 Reflection Measurement Short Calibration Test Setup
For greatest accuracy, use a coupler or bridge with the correct test port connector. Any additional adapters or components placed between the test port and DUT degrades coupler/bridge directivity and system source match.
For best results, use the same adapter for the calibration and the measurement. Be sure to terminate the second port of a two port device.
63
Page 72
Functions and Measurements
Preset/ System
MODE
CAUTION
Frequenc
Frequency
BW/ Avg
MODE
NOTE
Making a Reflection Calibration Measurement
2 Connect the tracking generator output of the analyzer to the
directional bridge or coupler.
3 Connect the analyzer RF input to the coupled port of the
directional bridge or coupler.
4 Press > Preset. (With Preset Type of Factory)
5 Turn on the tracking generator and, if necessary, set the output
power to -10 dBm
Press > Tracking Generator > Amplitude TG (On) > -10 > dBm
Excessive signal input may damage the DUT. Do not exceed the maximum power that the device under test can tolerate.
6 Set the start and stop frequencies and resolution bandwidth:
Press > Start Freq > 100 > MHz Press > Stop Freq > 1 > GHz
:
64
Press > Res Bw > 3 > MHz
7 Replace the DUT with a short circuit.
8 Normalize the trace:
Press > Tracking Generator > More > Normalize > Store Ref (1
3) > Normalize (On)
This activates the trace 1 minus trace 3 function and displays the results in trace 1 (Figure 3-23). The normalized trace or flat line represents 0 dB return loss. Normalization occurs each sweep. Replace the short circuit with the DUT.
The reference trace is stored in trace 3, changing trace 3 to Clear Write will invalidate the normalization.
Page 73
Making a Reflection Calibration Measurement
Figure 3-23 Short Circuit Normalized
Functions and Measurements
65
Page 74
Functions and Measurements Measuring Return Loss Using the Reflection Calibration Routine

Measuring Return Loss Using the Reflection Calibration Routine

This procedure uses the reflection calibration routine in the proceeding procedure "Making a Reflection Calibration
Measurement" on page 63 to calculate the return loss of the 370
MHz low pass filter.
1 After calibrating the system with the above procedure, reconnect
the filter (DUT) in place of the short circuit used for calibration. Do not change any of the analyzer settings or the measurement setup may be invalidated.
2 Use the marker to read return loss. Position the marker with the
front-panel knob to read the return loss at that frequency.
Rotate the knob to find the highest peak and the readout is the maximum return loss. See Figure 3-24.
Figure 3-24 Measuring the Return Loss of the Filter
66
Page 75

Making an Average Power Measurement

CAUTION
Making an Average Power Measurement
Average power measurements provide a key metric in transmitter performance.
Base station transmit power must be set accurately to achieve optimal coverage in wireless networks. If the transmit power is set too high due to inaccurate power measurements, undesired interference can occur. If the transmit power is set too low, coverage gaps or holes may occur. Either case affects system capacity and may translate into decreased revenue for service providers. Average power can be measured for the channel of interest while the base station is active. All other channels should be inactive. Average power is a broadband measurement. If other signals are present the analyzer will also measure their power contributions.
The maximum power for the RF IN port and the RF OUT port of the analyzer is +30 dBm (1 W). The maximum power for the U2000 Series Power Sensor port is +25 dBm (316 mW). When directly coupled to a base station, the test set can be damaged by excessive power applied to any of these three ports.
To prevent damage in most situations when directly coupling an analyzer to a base station, use a high power attenuator between the analyzer and the BTS.
Functions and Measurements
The N9320B spectrum analyzer supports the U2000 Series USB power sensors.
The U2000 Series USB power sensors do not need manual calibration and zero routines performed. Calibration and zeroing are performed without removing the power sensor from the source, through internal zeroing. With internal zeroing of U2000 Series USB power sensors, there is no need
67
Page 76
Functions and Measurements
NOTE
NOTE
Making an Average Power Measurement
to disconnect the sensor or power-off the DUT. The U2000 Series do not require 50 MHz reference signal calibration, allowing the factory calibration to ensure measurement accuracy. For best accuracy, users are recommended to perform external zeroing for input signals below -30 dBm for best accuracy.
If you suspect other signals may be present, it is recommended that you turn off all the other channels and measure average power only on the signal of interest. Another option is to measure channel power (which is less accurate), that filters out all other channels (signals).
Connect the power meter as close as possible to the power amplifier/duplexer output. Do not use a coupled port. Sensors may not be as accurate at the power levels provided by coupled ports.

Making a Basic Average Power Measurement

To make an average power measurement, you connect the power sensor and cable, zero and calibrate the meter, before making a measurement. Zeroing of the power meter will occur automatically:
• Every time you use the Power Meter.
• When a 5 degree C. change in temperature occurs.
• Whenever you change the power sensor.
• Every 24 hours.
• Before measuring low level signals—for example, 10 dB above
the lowest specified power for your power sensor.
Calibrating the power meter every time you cycle the power on and off.
In most situations, you can press {Zero} to complete the two steps (zero and cal) together.
68
Page 77
Functions and Measurements
Preset/ System
MODE
Making an Average Power Measurement
To Make a Basic Average Power Measurement
Follow the steps below to make a basic average power measurement.
1 Press > Preset. (With Preset Type of Factory)
2 Press , rotate the knob or press the arrow hardkey to highlight
Power Meter, then press [ENTER] to turn on the power meter.
3 Zero and calibrate the meter. Press {Zero} to make a Zero operation
of the power sensor followed by a calibration operation.
4 Connect the power sensor to the power Ref 50 MHz port. The
analyzer supports the U2000 Series power sensors.
5 Connect the external attenuation, if required, with the power
sensor used.
6 Connect the analyzer to the base station. Connect the power
sensor to the signal to be measured. A typical BTS connection is after the output of the power amplifier and duplexer. The connection requires a 2 ft. cable and two Type-N barrel connectors, in addition to the power sensor and external attenuator.
Figure 3-25 Connection with base station
69
Page 78
Functions and Measurements
Preset/ System
MODE
Preset/ System
MODE
Making an Average Power Measurement
Setting Power Meter Resolution
You can choose from four levels of Power Meter resolution. Higher resolutions provide more accuracy but slow the measurement speed.
1 Press > Preset. (With Preset Type of Factory)
2 Press , rotate the knob or press the arrow hardkey to highlight
Power Meter, then press [ENTER] to turn on the power meter.
3 Press {Meas Disp} > {Resolution}. Select 1,2,3 or 4. The four options
represent different resolution as follow:
• [1] = 1
• [2] = 0.1
• [3] = 0.01
• [4] = 0.001
Setting the Power Meter.s Top and Bottom End-Points
Setting the end points (Disk Range) close to expected measurement value changes the sensitivity of the Power Meter scale resolution. This is an advantage in viewing small changes in power. However, this will not affect the overall range of the sensor.
1 Press > Preset. (With Preset Type of Factory)
2 Press , rotate the knob or press the arrow hardkey to highlight
Power Meter, then press [ENTER] to turn on the power meter.
3 Press {Meas Disp} > {Dis Range} to access the end-points menu.
4 Press {Top} and enter the maximum scale value using the numeric
keypad. Then press {dBm} to complete the setup.
5 Press {Bottom} and enter the minimum scale value desired using
the numeric keypad. Then press {dBm} to complete the setup.
70
Page 79

Demodulate the AM/FM signal

SPAN
Preset/ System
Frequenc
Peak Search
Peak Search
BW/ Avg
SPAN
Det/ Display
SPAN
The N9320B provide you the AM/FM demodulation function.

Demodulating an AM Signal

The demodulation functions listed in the menu under > {Demod} allow you to demodulate and hear signal information displayed on the analyzer.
Simply place a marker on a signal of interest, set the analyzer in zero span, activate AM demodulation, turn the speaker on, and then listen.
1 Press > Preset. (With Preset Type of Factory)
2 Connect an antenna to the analyzer input.
3 Select a frequency range on the analyzer, such as the range for AM
radio broadcasts. For example, the frequency range for AM broadcasts in the United States is 550 kHz to 1650 kHz:
Functions and Measurements
Demodulate the AM/FM signal
Press >{Start Freq}> 550 > {kHz} > {Stop Freq} > 1650>{kHz}.
4 Place a marker on the signal of interest:
Press > {Next Pk Right} or {Next Pk Left}
5 Set the frequency of interest to center frequency:
Press > {Next Pk Right} or {Next Pk Left} (as necessary).
6 Change the resolution bandwidth to 100 kHz:
Press > 100 > {kHz}.
7 Press > {Zero Span}.
8 Set the detector type and turn on AM demodulation:
Press > {Detector} > {Sample}.
Press > {Demod} > {AM ON}.
71
Page 80
Functions and Measurements
Preset/ System
NOTE
Frequenc
SPAN
Sweep/ Tri g
Peak Search
BW/ Avg
Amplitud
Demodulate the AM/FM signal

Demodulating an FM Signal

1 Press > Preset. (With Preset Type of Factory)
2 Use a signal source or an antenna for an FM signal to analyze. In
3 Set the center frequency to the center of the FM signal (in this case
This section demonstrates how to demodulate and listen to an FM signal using the built-in FM demodulator.
Using the built in FM demodulator you can tune to an FM signal and view the results of the detector output as displayed in the time-domain.
this example the signal source is used transmitting at 300 MHz with FM deviation of 10 kHz and FM rate of 1 kHz.
If you are using a broadcast FM signal in the United States, for example, the FM channels are broadcasting between 87.7 MHz to
107.7 MHz.
300 MHz):
72
Press > 300 > {MHz}.
4 Set the analyzer to zero span for time-domain analysis:
Press > {Zero Span}.
Press > {Sweep Time} > 5 > {s}.
Press > {Next Pk Right} or {Next Pk Left}
5 Set the resolution bandwidth to capture the full bandwidth of the
FM signal.
Press > 100 > {kHz}.
6 Turn on the pre-amplifier to observe the small signal:
Press > {Int Preamp On}.
Page 81
Functions and Measurements
SPAN
SPAN
Demodulate the AM/FM signal
7 Turn on the FM demodulator:
Press >{Demod} > {FM ON}.
Listen to the FM signal.
Press >{Demod} > {Speaker Vol}, rotate the knob to adjust the speaker volume.
The headphone jack is availabe for the audio signal output alternatively.
73
Page 82
Functions and Measurements Analysis the Modulated Signals

Analysis the Modulated Signals

The N9320B provides the optional AM/FM (Option AMA) and ASK/FSK (Option-DMA) modulation analysis function for analysis the modulated signals. These functions can be activated by license key.

AM/FM Modulation Analysis

Optional AM/FM demodulation analysis provides modulation metrics, including carrier power, modulation rate, AM depth/FM deviation, SINAD and carrier frequency offset. User definable limits provide Pass/Fail indicators of four types:
Maximum carrier power, Maximum AM modulation index or FM deviation, Minimum AM modulation index or FM deviation, and Minimum carrier frequency offset.
The user can save the waveforms with metrics for reporting as well as the set-up parameters for future measurements or analysis. In the following example, you can use any source available (such as AM/FM radio) as long as it has AM/FM signal generation capability.
74
AM Modulation Analysis
Please refer to the procedures below for the AM modulation analysis:
1 Connect the signal generator to RF IN connector and turn on
the signal generator’s AM and RF output.
2 Turn on the AM demodulation analysis function.
Press [MODE], Rotating the knob to highlight AM modulation Analysis, press [ENTER].
3 Press {Carries Freq} > [500] > {MHz} to set the carrier frequency.
Page 83
Functions and Measurements
Analysis the Modulated Signals
4 Set the attenuation state to Auto (default):
Press {More} > {More} > {Attenuation Auto} > {More}
5 Press {Y Scale} > {Auto Scale} > {Return} to set the Y scale.
6 Set the AM depth limit and turn on the limit function:
Press {More} > {Limits} > {AM Depth UP} > [76] > {%} > {AM Depth Low} > [70] > {%} > {Limits on}
The red value in the result indicates the demodulated AM depth exceeds the limits set as above.
Figure 3-26 AM demodulation analysis
FM Modulation Analysis
Please refer to the procedures below for the FM modulation analysis:
1 Connect the signal generator to RF IN connector and turn on
the signal generator’s FM and RF output.
2 Turn on the FM demodulation analysis function.
Press [MODE], rotating the knob to highlight FM modulation Analysis, press [ENTER].
3 Press {Carries Freq} > [1.5] > {GHz} to set the carrier frequency.
75
Page 84
Functions and Measurements Analysis the Modulated Signals
4 Set the attenuation state to Auto (default):
Press {More} > {More} > {Attenuation Auto} > {More}
5 Press {Y Scale} > {Auto Scale} > {Return} to set the Y scale.
6 Set the FM deviation limit and turn on the limit function:
Press {More} > {Limits} > {FreqDev UP} > [60] > {%} > {FreqDev Low} > [30] > {%} > {Limits on} > {Return}
The purple value in the result indicates the demodulated FM deviation exceeds the limits set as above.
Figure 3-27 FM modulation analysis
76
Page 85

ASK/FSK Modulation Analysis

The ASK/FSK modulation analysis function supports four display modes:
•Symbol
• Waveform
• ASK/FSK Error
•Eye Diagram.
User definable limits provide Pass/Fail indicators of four types:
• Maximum carrier power
• Maximum ASK modulation depth/FSK frequency deviation
• Minimum ASK modulation depth/FSK frequency deviation
• Maximum carrier frequency offset.
The metrics includes carrier power, ASK/FSK error, ASK depth/FSK frequency deviation, and ASK index etc. The waveform with metrics and setup parameters can be saved for reports and future Measurements.
Functions and Measurements
Analysis the Modulated Signals
In the following example, you can use any source available (such as your remote key of the car) as long as it has ASK/FSK generation capability. For ASK/FSK signal, its Filter is Nyquist and Alpha is
0.35.
ASK Modulation Analysis
Please refer to the procedures below to analysis the ASK demodulation.
1 Connect the signal generator to RF IN connector and turn on
the signal generator’s ASK and RF output.
2 Turn on the ASK demodulation analysis function.
Press [MODE], rotating the knob to highlight ASK modulation Analysis, press [ENTER].
3 Press {Carries Freq} > [100] > {MHz} to set the carrier frequency.
4 Press {Symbol Rate} > [10] > {MHz} to set the symbol rate.
77
Page 86
Functions and Measurements Analysis the Modulated Signals
5 Press {Filter Setup} > {Ref Filter} > {Nyquist} > {Return} to set the
filter type.
6 Set the attenuation state to Auto (default):
Press {More} > {More} > {Attenuation Auto} > {More}
7 Press {Y Scale} > {Auto Scale} > {Return} to set the Y scale.
8 Setting the ASK depth limit and turn on the limit function:
Press {More} > {Limits} > {ASK Depth UP} > [71] > {%} > {ASK Depth Low} > [70] > {%} > {Limits on} > {Return}
9 Press {More} > {More} > {View} > {Eye Diagram} to view the eye
diagram.
FSK Demodulation Analysis
Please refer to the procedures below to analysis the FSK demodulation.
1 Connect the signal generator to RF IN connector and turn on
the signal generator’s FSK and RF output.
2 Turn on the ASK demodulation analysis function.
Press [MODE], Rotating the knob to highlight FSK modulation Analysis, press [ENTER].
3 Press {Carries Freq} > [1.5] > {GHz} to set the carrier frequency.
4 Press {Symbol Rate} > [10] > {ksps} to set the symbol rate.
5 Press {Filter Setup} > {Ref Filter} > {Nyquist} > {Return} to set the
filter type.
6 Set the attenuation state to Auto (default):
Press {More} > {More} > {Attenuation Auto} > {More}
7 Press {Y Scale} > {Auto Scale} > {Return} to set the Y scale.
78
Page 87
Functions and Measurements
Analysis the Modulated Signals
8 Setting the FSK deviation limit and turn on the limit function:
Press {More} > {Limits} > {FreqDev UP} > [85] > {kHz} > {FreqDev
Low} > [80] > {kHz} > {Limits on}
The figure below indicates the FSK EYE Diagram as a measurement resul t.
Figure 3-28 FSK Eye Diagram
79
Page 88
Functions and Measurements
RBW
kspan
n
----------------------------=
Meas

Measuring Channel Power

Measuring Channel Power
Channel Power measures the power and power spectral density in the specified channel bandwidth. A white frame on the display indicates the edge limits of the channel bandwidth. The user must specify the center frequency, reference level, and channel bandwidth to be used for the measurement.
The power calculation method used to determine the channel power is a traditional method known as the integration bandwidth (IBW) method. A swept spectrum is used as the basis for this calculation. Therefore, it is important to set the resolution bandwidth correctly before making this measurement using the following formula:
Please refer to the procedures below to set a channal power measurement.
80
1 Press
> Channel Power to select the channel power
measurement mode.
Page 89
Functions and Measurements
Measuring Channel Power
2 Press Standard Customize in the submenu to choose the radio
stadard.
3 There are four wireless standard options: WCDMA, CDMA, LTE,
GSM. Press WCDMA as an example.
Once the standard is set, the integration bandwidth and span will be automatically set according to the standard.
81
Page 90
Functions and Measurements Measuring Channel Power
4 Press Center Freq to change the center frequency of the current
signal channel. Adjust the center frequency to measure other channel power in the same standard.
Please follow the similar procedure as above to measure the channel power in the CDMA and LTE standard.
82
To measure the channel power in GSM standard, the default setting is 10 MHz span, 300 kHz RBW, and the continuous peak search is enabled. The instrument will measure the channel power after the center frequency is changed.
Page 91

Viewing Catalogs and Saving Files

NOTE
File/ Print
The analyzer will store and retrieve data files similar to using a PC. The analyzer allows you to view and save files in the internal storage of the analyzer or to a USB storage device. This section provides information on how to locate catalogs (file types) and to save files.

Locating and Viewing Files in the Catalog

The analyzer has four types of files:
State stores analyzer settings (file extension: .STA).
Trace stores trace information (file extension: .TRA).
Screen stores screen captures of the display (file extension: .JPG).
Limits which stores data sets to determine whether a trace has exceeded
preset specifications (file extension: .LIM). Trace files are not bitmap files. They cannot be imported into other
applications or converted. If you require the use of other applications, be sure to save as .JPG formats using Screen for your file type.
Functions and Measurements
Viewing Catalogs and Saving Files
Press > Catalog to bring up a screen display below.
83
Page 92
Functions and Measurements
File/ Print
Viewing Catalogs and Saving Files
1 The entry fields show the parameters for the files viewed:
Name states the filename.
Type displays the type of files viewed.
Path displays the location of the files (Local or USB).
2 There are four column headings in this area of the display:
Name displays the filename.
Typedisplays the filename extension (TRA, STA, TRC or JPG).
Sizedisplays the size of the file in bytes.
Modified displays the date and time the change occurred.
3 The softkey menu has four options:
Type selects the type of file you wish to view.
Sort sorts and views the files according to a file attribute.
Local accesses the internal storage to view files.
USB accesses the external USB storage device to view files.
84
4 The drive space line shows the number of bytes used on the drive
and the number of bytes still free on the drive.

Saving a File

The analyzer can save files (states, traces, limits or screens) to a USB storage device (USB/), or internal storage (Local/). The following is an example of how to save a trace to internal storage (Local/).
1 Select the save destination for the file. This example will select the
internal storage (Local/):
Press > Catalog > Local
2 Press Return > Save, to access the Save menu.
3 Select the type of file. In this example we will save a trace. The
filename extension for the trace file type is .TRA:
Press Type > Tr a ce
Page 93
Functions and Measurements
Enter
CAUTION
Viewing Catalogs and Saving Files
4 Select the trace to save (1, 2, 3, 4 or all traces). For this example,
we will select trace 3:
Press Source > Trace 3
5 Enter a filename by pressing Name. Assume you name the file
TEST1”. The alpha portion of the filename is selected using the softkey menus. The numeric keypad is used for the numeric portion of the filename. Press to accept the filename and to terminate this process and return to the Save menu.
The filename can be composed of alpha (A-Z) or numeric (0-9) characters in any combination. The analyzer will not overwrite an existing file. If the filename already exists, the status bar will display the message: “File already exists”. If a filename is not edited, the analyzer will automatically generate a name based on the type of file you are saving (State: STATE, Trace: TRACE, Limits: LIMIT, Screen: SCREN). It also generates a three digit integer, starting at 000 and extending through 999, which it adds to the name, for example: TRACE056.TRC. The analyzer will remember the previously saved value even through a power cycle.
6 Press Save Now to save the file.
Saving Now” is displayed during this operation. In this example, the status bar displays “Local/TEST1.TRC file saved”.
When saving a file to a USB storage device, do not disconnect the USB device during the save operation. To do so may cause unexpected data loss.
85
Page 94
Functions and Measurements
File/ Print
Viewing Catalogs and Saving Files

Loading a File

This feature will recall (load) onto the displayed screen a previously saved filename and type compatible with this analyzer. The file types are described above.
1 Select the directory where your file is located. In this example,
select the internal storage (Local/): Press > Catalog > Local
2 Press Return > Load to access the Load menu
3 Select the type of file. Assume you need to load a trace file:
Press Type > Tr a ce
4 Using the knob or arrow keys to highlight the file you need. And
then select the trace into which you wish to load the file. Assume you need to load file TEST1.TRC into Trace 2:
Press Destination > Trace 2.
5 Press Load Now to load the specified file. The status bar displays:
Local/TEST1.TRC file loaded”. Screen files are in .jpg format and cannot be loaded or recalled into the analyzer display, they can only be downloaded and used in PC applications.
86
Page 95

Copying a File

File/ Print
File/ Print
To copy a file from internal storage to a USB storage device, follow this procedure:
1 Press > Catalog > Local to select the path. 2 Connect a USB storage device to the USB connector (DEV). 3 Select the type of file. Assume you need to copy a trace file:
Press Return > Copy > Type > Trace
4 Using the knob or arrow keys to highlight the file name you need. In
this example, highlight the file TEST1.TRC.
5 Press Copy Now to execute the operation. A “Copying file” indicator
displays. Then “/USB/TEST1.TRC file copied” appears on the status bar. You have now copied the file TEST1.TRC from the internal storage to the USB storage device.

Deleting a File

Perform the following process to delete a specific file. Note that once a file is deleted from local memory, it cannot be recovered.
1 Press > Catalog > Local to select the path 2 Press Return > Delete to access the Delete menu 3 Select the type of file. Assume you need to delete a trace file press
Typ e > Tra c e
4 Using the knob or arrow keys to highlight the file name you want to
load. In this example, selecting the file TEST1.TRC.
5 Press Delete Now. “Deleting file” indicates the operation is in
process. Then, “Local/TEST1.TRC file deleted” shown on the status bar indicates the file has been deleted.
Functions and Measurements
Viewing Catalogs and Saving Files
87
Page 96
Functions and Measurements Viewing Catalogs and Saving Files
88
Page 97

Key Reference

4Key Reference
This chapter describes each front panel hardkey and their associated softkeys (menus). The front panel key descriptions are organized alphabetically. The lower level softkeys are listed as they appear in the softkey menus.
89
Page 98
Key Reference
Amplitud
Amplitud
Amplitud
Amplitud
Amplitud
Amplitud

Amplitude

Amplitude
Press hardkey to activate the reference level function and access the amplitude softkey menu.
The softkey menu selections are described as follows:
Ref Level Sets the reference level. This function is activated when is
pressed. The reference level is the amplitude power or voltage represented by the top line of the graticule on the display. Setting the reference level changes the absolute amplitude level (in the selected amplitude units) of the top graticule line.
Key access:
Attenuation
Auto Man
Toggles the input attenuation between Auto and Man. The attenuation level can be set in as little as 1 dB increments when Man is selected. The analyzer input attenuator is normally coupled to the reference level control, it reduces the power level of the input signal at the input mixer. Selecting Auto re-couples the attenuator. Change the attenuation by using the arrow keys, the knob, or the numeric keypad.
Key access:
Scale/Div Sets the logarithmic units per vertical graticule division on the
display. The Scale/Div function is available only when the Scale Ty pe key is set to Log.
Key access:
Scale Type
Log Lin
Scales the vertical graticule divisions in logarithmic units when Log is selected. The valid logarithmic units are 1, 2, 5 and 10 dB per division. When Lin is underlined, the vertical divisions are linearly scaled with default amplitude units of volts. The top of the graticule represents the reference-level value; the bottom of the graticule represents zero volts. (Each division is one-tenth of the reference level.)
Key access:
90
Page 99
Key Reference
Amplitud
Amplitud
Amplitud
Amplitud
Amplitude
Scale
On Off
Toggles the amplitude Scale for each graticule. The vertical divisions are linearly scaled with default amplitude units of volts. The default unit is dBm. This softkey is only valid when the Scale Ty pe is set to Lin.
Key access:
Int Preamp
On Off
(Option PA3) Toggles the internal pre-amplifier function between on and off. Pressing Int Preamp (On) results in a correction being applied to compensate for the gain of the preamp. Amplitude readings then represent the signal value at the input connector. When the preamp is on, a “PA” indication appears on the left side of the display. The preamp operates over a range of 100 kHz to 3 GHz.
Key access:
Y Axis Accesses the softkeys that change the amplitude units (logarithmic
or linear). Change the amplitude units by pressing dBm, dBmV, dBuV, dBuA, Watts, Volts, and Amps.
Key access: > More
Ref Lvl Offset Adds an offset value to the displayed reference level.
Reference-level offsets are only entered by using the numeric keypad. Entering an offset does not affect the trace or the attenuation value. Reference-level offsets are used when gain or loss occurs between a device under test and the analyzer input. Thus, the signal level measured by the analyzer may be referred to as the level at the input of an external amplitude-conversion device. When an amplitude offset is entered, its value appears on the left side of the display under Offst (as opposed to frequency offsets which appear at the bottom of the display). To eliminate an offset, press Ref Lvl Offset > 0 > dB.
Key access: > More
91
Page 100
Key Reference
Amplitud
Amplitud
Amplitud
Amplitude
Att limit
On Off
Toggles the attenuation limit function between On and Off. When this softkey is set to On, the reference level will not exceed the maximum value = (attenuation + 20) dBm.
Key access: > More
Corrections Accesses the corrections submenu to enable the corrections
function for amplitude compensation. The amplitude correction is used to compensate the gain/attenuation from antenna/cable. Please refer to the procedures below to use the amplitude correction function.
1 Press [Amplitude] > {More} > {Corrections} to acess the
correction function.
2 Press {Cable/Antenna/Other/User} to select the corrction type.
3 Press {Edit} to add the point with frequency and amplitude offset
value.
4 When this editing is completed, press {Corrections On} to load
the correction data. The status will be shown on the corresponding softkey and set the {Apply Corrections} On, the amplitude compensation will be applied to the displayed trace.
Press {Apply Corrections off} to turn off all the corrections.
Key access: > More
Ext Amp Gain Adds a positive or negative preamplifier gain value, which is
subtracted from the displayed signal. (Use negative values for gain and positive values for loss.) The function is similar to the Ref Lvl Offset function, however with the Ext Amp Gain function, the attenuation may be changed depending on the preamplifier gain entered. A preamplifier gain offset is used for measurements that require an external preamplifier or long cables. The offset is subtracted from the amplitude readout so that the displayed signal level represents the signal level at the input of the preamplifier. The preamplifier gain offset is displayed at the top of the screen and is removed by entering zero. The preamplifier gain offset can only be entered using the numeric keypad. The preamplifier gain value is not affected by an instrument preset or a power cycle.
Key access: > More
92
Loading...