Used Agilent Agilent Used N1912A Manual

Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com
Keysight N1911A/1912A P-Series Power Meters
User’s Guide
Notices
CAUTION
WARNING
Copyright Notice
© Keysight Technologies 2006-2017 No part of this manual may be repro-
duced in any form or by any means (including electronic storage and retrieval or translation into a foreign language) without prior agreement and written consent from Keysight Technol­ogies as governed by United States and international copyright laws.
Manual Part Number
N1912-90002
Edition
Edition 18, September 29, 2017
Printed in:
Printed in Malaysia
Published by:
Keysight Technologies Bayan Lepas Free Industrial Zone, 11900 Penang, Malaysia
Technology Licenses
The hardware and/or software described in this document are fur­nished under a license and may be used or copied only in accordance with the terms of such license.
Declaration of Conformity
Declarations of Conformity for this product and for other Keysight prod­ucts may be downloaded from the Web.
U.S. Government Rights
The Software is “commercial computer software,” as defined by Federal Acqui­sition Regulation (“FAR”) 2.101. Pursu­ant to FAR 12.212 and 27.405-3 and Department of Defense FAR Supple­ment (“DFARS”) 227.7202, the U.S. government acquires commercial com­puter software under the same terms by which the software is customarily provided to the public. Accordingly, Keysight provides the Software to U.S. government customers under its stan­dard commercial license, which is embodied in its End User License Agreement (EULA). The license set forth in the EULA represents the exclusive authority by which the U.S. government may use, modify, distribute, or disclose the Soft-ware. The EULA and the license set forth therein, does not require or per-mit, among other things, that Keysight: (1) Furnish technical information related to commercial computer commercial computer software documentation that is not customarily provided to the public; or (2) Relinquish to,
or otherwise provide, the govern­ment rights in excess of these rights customarily provided to the public to use, modify, reproduce, release, per-form, display, or computer software com-puter software documentation. No additional government requirements beyond those set forth in the EULA shall apply, except to the extent licenses are explicitly required from all providers of commercial computer software pursu-ant to the FAR and the DFARS and are set forth specifically in writing Keysight shall be under no obligation to update, revise or otherwise modify the Software. With respect to any technical data as defined by FAR
2.101, pursuant to FAR 12.211 and
27.404.2 and DFARS 227.7102, the U.S. government acquires no greater than Limited Rights as defined in FAR
27.401 or DFAR 227.7103-5 (c), as applicable in any technical data.
disclose commercial
that those terms, rights, or
else-where in the EULA.
soft-ware or
or commercial
Warranty
THE MATERIAL CONTAINED IN THIS DOCUMENT IS PROVIDED “AS IS,” AND IS SUBJECT TO BEING CHANGED, WITHOUT NOTICE, IN FUTURE EDITIONS. FURTHER, TO THE MAXIMUM EXTENT PERMITTED BY APPLICABLE LAW, KEYSIGHT DIS­CLAIMS ALL WARRANTIES, EITHER EXPRESS OR IMPLIED, WITH REGARD TO THIS MANUAL AND ANY INFORMA­TION CONTAINED HEREIN, INCLUD­ING BUT NOT LIMITED TO THE IMPLIED WARRANTIES OF MER­CHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. KEYSIGHT SHALL NOT BE LIABLE FOR ERRORS OR FOR INCIDENTAL OR CONSE­QUENTIAL DAMAGES IN CONNECTION WITH THE FURNISHING, USE, OR PERFORMANCE OF THIS DOCUMENT OR OF ANY INFORMATION CON­TAINED HEREIN. SHOULD KEYSIGHT AND THE USER HAVE A SEPARATE WRITTEN AGREEMENT WITH WAR­RANTY TERMS COVERING THE MATE­RIAL IN THIS DOCUMENT THAT CONFLICT WITH THESE TERMS, THE WARRANTY TERMS IN THE SEPARATE AGREEMENT SHALL CONTROL.
Safety Information
A CAUTION notice denotes a hazard. It calls attention to an operating proce­dure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not pro­ceed beyond a CAUTION notice until the indicated conditions are fully understood and met.
A WARNING notice denotes a hazard. It calls attention to an operating proce­dure, practice, or the like that, if not correctly performed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met.
2 Keysight N1911A/1912A User’s Guide

Certification

Keysight Technologies certifies that this product met its published specifications at the time of shipment. Keysight further certifies that its calibration measurements are traceable to the United States National Institute of Standard and Technology (formerly National Bureau of Standards), to the extent allowed by that organization’s calibration facility, and to the calibration facilities of other International Standards Organization members.

General Warranty

The material contained in this document is provided “as is,” and is subject to being changed, without notice, in future editions. Further, to the maximum extent permitted by applicable law, Keysight disclaims all warranties, either express or implied with regard to this manual and any information contained herein, including but not limited to the implied warranties of merchantability and fitness for a particular purpose. Keysight shall not be liable for errors or for incidental or consequential damages in connection with the furnishing, use, or performance of this document or any information contained herein. Should Keysight and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms in the separate agreement shall control. Duration and conditions of warranty for this product may be superseded when the product is integrated into (becomes a part of) other Keysight products. During the warranty period, Keysight will, at its option, either repair or replace products which prove to be defective. The warranty period begins on the date of delivery or on the date of installation if installed by Keysight.

Warranty Service

For warranty service or repair, this product must be returned to a service facility designated by Keysight. For products returned to Keysight for warranty service, the Buyer shall prepay shipping charges to Keysight and Keysight shall pay shipping charges to return the product to the Buyer. However, the Buyer shall pay all shipping charges, duties, and taxes for products returned to Keysight from another country.
Keysight N1911A/1912A User’s Guide 3

Limitation of Warranty

The foregoing warranty shall not apply to defects resulting from improper or inadequate maintenance by the Buyer, Buyer-supplied products or interfacing, unauthorized modification or misuse, operation outside of the environmental specifications for the product, or improper site preparation or maintenance.
The design and implementation of any circuit on this product is the sole responsibility of the Buyer. Keysight does not warrant the Buyer’s circuitry or malfunctions of Keysight products that result from the Buyer’s circuitry. In addition, Keysight does not warrant any damage that occurs as a result of the Buyer’s circuit or any defects that result from Buyer-supplied products.
To the extent allowed by local law, Keysight makes no other warranty, expressed or implied, whether written or oral with respect to this product and specifically disclaims any implied warranty or condition of merchantability, fitness for a particular purpose or satisfactory quality.

Exclusive Remedies

To the extent allowed by local law, the remedies provided herein are the Buyer’s sole and exclusive remedies. Keysight shall not be liable for any direct, indirect, special, incidental, or consequential damages (including lost profit or data), whether based on warranty, contract, tort, or any other legal theory.
4 Keysight N1911A/1912A User’s Guide

Restricted Rights Legend

The Software and Documentation have been developed entirely at private expense. They are delivered and licensed as “commercial computer software” as defined in DFARS 252.227-7013 (Oct 1988), DFARS 252.211-7015 (May 1991), or DFARS 252.227-7014 (Jun 1995), as a “commercial item” as defined in FAR
2.101(a), or as “restricted computer software” as defined in FAR 52.227-19 (Jun
1987) (or any equivalent agency regulation or contract clause), whichever is applicable. You have only those rights provided for such Software and Documentation by the applicable FAR or DFARS clause or the Keysight standard software agreement for the product involved.

Technology Licenses

The hardware and/or software described in this document are furnished under a license and may be used or copied only in accordance with the terms of such license.
Keysight N1911A/1912A User’s Guide 5

Safety Symbols

The following symbols on the instrument and in the documentation indicate precautions which must be taken to maintain safe operation of the instrument.
Direct current (DC). Alternating current (AC).
Off (mains supply). On (mains supply).
Caution, risk of electric shock.
Earth (ground) terminal. Frame or chassis (ground) terminal.
Protective earth (ground) terminal.
Caution, risk of danger (refer to this manual for specific Warning or Caution information).
Equipment protected throughout by double insulation or reinforced insulation.
Both direct and alternating current.
Caution, hot surface. In position of a bi-stable push control.
Three-phase alternating current. Equipotentiality.
This symbol indicates the operating switch for ‘Stand-by’ mode. Note, the instrument is NOT isolated from the mains when the switch is pressed. To isolate the instrument, the mains coupler (mains input cord) should be removed from the power supply.
Out position of a bi-stable push control.
6 Keysight N1911A/1912A User’s Guide

Safety Considerations

Read the information below before using this instrument.
The following general safety precautions must be observed during all phases of operation, service, and repair of this instrument. Failure to comply with these precautions or with specific warnings elsewhere in this manual violates safety standards for design, manufacture, and intended use of the instrument. Keysight Technologies assumes no liability for the customer’s failure to comply with these requirements.

Recommended Calibration Interval

Keysight Technologies recommends a two- years calibration cycle for the N1911A/1912A P-Series power meter.
Keysight N1911A/1912A User’s Guide 7

General Safety Information

WARNING
This is a Safety Class I instrument (provided with a protective earthing ground, incorporated in the power cord). The mains plug shall only be inserted in a socket outlet provided with a protective earth contact. Any interruption of the protective conductor inside or outside of the instrument is likely to make the instrument dangerous. Intentional interruption is prohibited.
– DO NOT operate the product in an explosive atmosphere or in the
presence of flammable gases or fumes.
– DO NOT use repaired fuses or short-circuited fuseholders: For continued
protection against fire, replace the line fuse(s) only with fuse(s) of the same voltage and current rating and type.
– DO NOT perform procedures involving cover or shield removal unless you
are qualified to do so: Operating personnel must not remove equipment covers or shields. Procedures involving the removal of covers and shields are for use by service-trained personnel only.
– DO NOT service or adjust alone: Under certain conditions, dangerous
voltages may exist even with the equipment switched off. To avoid dangerous electrical shock, service personnel must not attempt internal service or adjustment unless another person, capable of rendering first aid and resuscitation, is present.
– DO NOT operate damaged equipment: Whenever it is possible that the
safety protection features built into this product have been impaired, either through physical damage, excessive moisture, or any other reason, REMOVE POWER and do not use the product until safe operation can be verified by service-trained personnel. If necessary, return the product to a Keysight Technologies Sales and Service Office for service and repair to ensure the safety features are maintained.
– DO NOT substitute parts or modify equipment: Because of the danger of
introducing add itional hazards, do not install substitute parts or perform any unauthorized modification to the product. Return the product to a Keysight Technologies Sales and Service Office for service and repair to ensure the safety features are maintained.
8 Keysight N1911A/1912A User’s Guide

Environmental Conditions

The N1911A/1912A is designed for indoor use and in an area with low condensation. The table below shows the general environmental requirements for this instrument.
Environmental condition Requirement
Temperature
Humidity
Altitude Up to 3000 m
Pollution degree 2
Operating condition – 0 °C to 55 °C Storage condition – –40 °C to 70 °C
Operating condition – Up to 95% RH at 40 °C (non-condensing) Storage condition – Up to 90% RH at 65 °C (non-condensing)
Keysight N1911A/1912A User’s Guide 9

Regulatory Information

The N1911A/1912A complies with the following safety and Electromagnetic Compatibility (EMC) compliances:

Safety compliance

– IEC 61010-1:2010/EN 61010-1:2010 (3rd Edition)
– Canada: CAN/CSA-C22.2 No. 61010-1-12
– USA: ANSI/UL 61010-1:2012

EMC compliance

– IEC 61326-1:2005/EN 61326-1:2006
– CISPR11:2003/EN 55011:1998+A1:1999+A2:2002, Group 1 Class A
– Canada: ICES/NMB-001:Issue 4, June 2006
– Australia/New Zealand: AS/NZS CISPR 11:2004
10 Keysight N1911A/1912A User’s Guide

Regulatory Markings

The CE mark is a registered trademark of the European Community. This CE mark shows that the product complies with all the relevant European Legal Directives.
ICES/NMB-001 indicates that this ISM device complies with the Canadian ICES-001. Cet appareil ISM est conforme a la norme NMB-001 du Canada.
ISM GRP.1 Class A indicates that this is an Industrial Scientific and Medical Group 1 Class A product.
This symbol is a South Korean Class A EMC Declaration. This is a Class A instrument suitable for professional use and in electromagnetic environment outside of the home.
This symbol indicates the time period during which no hazardous or toxic substance elements are expected to leak or deteriorate during normal use. Forty years is the expected useful life of the product.
The CSA mark is a registered trademark of the Canadian Standards Association.
The RCM mark is a registered trademark of the Australian Communications and Media Authority.
This instrument complies with the WEEE Directive (2002/96/EC) marking requirement. This affixed product label indicates that you must not discard this electrical or electronic product in domestic household waste.
Keysight N1911A/1912A User’s Guide 11

Waste Electrical and Electronic Equipment (WEEE) Directive 2002/ 96/EC

This instrument complies with the WEEE Directive (2002/96/EC) marking requirement. This affixed product label indicates that you must not discard this electrical or electronic product in domestic household waste.

Product category:

With reference to the equipment types in the WEEE directive Annex 1, this instrument is classified as a “Monitoring and Control Instrument” product.
The affixed product label is as shown below.
Do not dispose in domestic household waste.
12 Keysight N1911A/1912A User’s Guide

Table of Contents

Certification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
General Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Warranty Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
Limitation of Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
Exclusive Remedies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
Restricted Rights Legend . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
Technology Licenses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
Safety Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
Safety Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
Recommended Calibration Interval . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
General Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8
Environmental Conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
Regulatory Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10
Safety compliance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10
EMC compliance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10
Regulatory Markings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11
Waste Electrical and Electronic Equipment (WEEE) Directive 2002/96/
EC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Product category: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12
Sales and Technical Support . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12
1Introduction
LXI Class-C Compliant Power Meter . . . . . . . . . . . . . . . . . . . . . . . . . . .30
Rack Mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .30
Power Meter and Sensor Capability . . . . . . . . . . . . . . . . . . . . . . . . . . . .31
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .31
Conventions Used in this Guide . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .32
Front Panel Keys and Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . .33
The Display Layout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .38
Keysight N1911A/1912A User’s Guide 13
Window Symbols and Pop-ups . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Warning symbol pop-up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Wait symbol pop-up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
Confirm symbol pop-up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
Numeric entry pop-up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
Text entry pop-up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
Trace scaling pop-up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
List pop-up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
Using the Instrument Web Browser . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
Editing the Instrument’s LAN Settings . . . . . . . . . . . . . . . . . . . . . . . 54
2 General Power Meter Functions
Setting the Units of Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
Selecting units of measurement from the softkeys . . . . . . . . . . . . . 59
Setting the Measurement Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . 60
Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
Setting the Resolution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Making Relative Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
Setting the Pulse Reference Levels . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
Setting Offsets . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
Setting channel offsets . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
Setting display offsets . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
Setting frequency dependent offsets . . . . . . . . . . . . . . . . . . . . . . . . 71
Editing frequency dependent offset tables . . . . . . . . . . . . . . . . . . . . 74
Setting Measurement Averaging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
Step Detection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
Setting the Video Averaging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82
Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82
14 Keysight N1911A/1912A User’s Guide
Setting the Video Bandwidth . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .84
Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .85
Setting Measurement Channel Gates . . . . . . . . . . . . . . . . . . . . . . . . . .87
Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .88
Auto Gate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
Setting up the Channel Trace . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .92
Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .92
Auto Scale . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .95
Setting the Trigger . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .96
Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .97
Setting External Trigger for Normal and Average Power Measurement . .
103
Power sweep mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .104
Frequency sweep mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .112
Determine the right step to be set . . . . . . . . . . . . . . . . . . . . . . . . .117
Setting Measurement Limits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .118
Setting limits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .119
Checking for limit failures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .121
Setting the Measurement Display . . . . . . . . . . . . . . . . . . . . . . . . . . . .122
Numeric format . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .124
Single function measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . .125
Combined Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .127
Scaling the Analog Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .129
Recorder Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .131
Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .131
N1911A/N1912A Option H01 . . . . . . . . . . . . . . . . . . . . . . . . . . . . .135
Saving and Recalling Power Meter States . . . . . . . . . . . . . . . . . . . . . .137
Editing a register’s name . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .138
Recalling a measurement setup . . . . . . . . . . . . . . . . . . . . . . . . . . .139
Zeroing and Calibrating the P-Series Sensor . . . . . . . . . . . . . . . . . . .140
Zeroing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .140
Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .141
Keysight N1911A/1912A User’s Guide 15
Zero/Cal Lockout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 143
Presetting the Power Meter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 145
Setting the Cable Short/Long . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 147
Short . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 147
Long . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 147
3 Using P-Series Power Sensor
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 150
Configuring a Power Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . 151
Setting the Trace Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 152
Step 1. Connecting the power sensor . . . . . . . . . . . . . . . . . . . . . . 153
Step 2. Setting the channel frequency . . . . . . . . . . . . . . . . . . . . . . 153
Step 3. Setting the trigger . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 154
Step 4. Setting the display to trace . . . . . . . . . . . . . . . . . . . . . . . . 154
Step 5. Expanding the trace display . . . . . . . . . . . . . . . . . . . . . . . . 155
Gate Control in a Single Enlarged Window . . . . . . . . . . . . . . . . . . 156
Viewing the results in a full screen window . . . . . . . . . . . . . . . . . . 160
Improving the accuracy with additional settings . . . . . . . . . . . . . . 162
4 Using E9320 E-Series Power Sensors
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 166
Power Meter Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 168
Default Channel Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 168
Configuring a Peak Power Measurement . . . . . . . . . . . . . . . . . . . . . . 169
Measurement Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 170
Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 170
Setting the Trace Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 172
Step 1. Connecting the power sensor . . . . . . . . . . . . . . . . . . . . . . 173
Step 2. Setting the channel frequency . . . . . . . . . . . . . . . . . . . . . . 173
Step 3. Setting the trigger . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 173
Step 4. Setting the display to trace . . . . . . . . . . . . . . . . . . . . . . . . 174
Step 5. Expanding the trace display . . . . . . . . . . . . . . . . . . . . . . . . 175
Gate Control in a Single Enlarged Window . . . . . . . . . . . . . . . . . . 176
Viewing the results in a full screen window . . . . . . . . . . . . . . . . . . 180
16 Keysight N1911A/1912A User’s Guide
Improving the accuracy with additional settings . . . . . . . . . . . . . .181
5 Using E9300 E-Series Power Sensors
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .186
Power Meter Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .187
Default Channel Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .188
Measurement Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .189
Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .190
Measuring Spread Spectrum and Multitone Signals . . . . . . . . . . . . . .192
CDMA signal measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .193
Multitone signal measurements . . . . . . . . . . . . . . . . . . . . . . . . . . .194
Measuring TDMA Signals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .195
Power meter and sensor operation . . . . . . . . . . . . . . . . . . . . . . . . .195
Achieving stable results with TDMA signals . . . . . . . . . . . . . . . . . .195
Achieving stable results with GSM signals . . . . . . . . . . . . . . . . . . .196
Electromagnetic Compatibility (EMC) Measurements . . . . . . . . . . . . .197
Measurement Accuracy and Speed . . . . . . . . . . . . . . . . . . . . . . . . . . .198
Setting the range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .198
Measurement considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . .199
6 Using E4410 E-Series Power Sensors
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .202
Power Meter Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .203
Default Channel Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .204
Measurement Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .205
Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .205
7 Using 8480 Series Power Sensors
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .208
Power Meter Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .209
Default Channel Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .210
8480 Series sensors connection requirements . . . . . . . . . . . . . . . .210
Measurement Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .212
Keysight N1911A/1912A User’s Guide 17
Frequency Specific Calibration Factors . . . . . . . . . . . . . . . . . . . . . . . . 213
Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 213
Sensor Calibration Tables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 217
Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 218
Editing/Generating sensor calibration tables . . . . . . . . . . . . . . . . . 221
Pre-installed calibration table contents . . . . . . . . . . . . . . . . . . . . . 225
8 Using N8480 Series Power Sensors
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 230
Power Meter Configuration Changes . . . . . . . . . . . . . . . . . . . . . . . . . 232
Default Channel Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 233
N8480 Series Sensors Connection Requirements . . . . . . . . . . . . . . . 234
N8480 Series Power Sensors (excluding Option CFT) . . . . . . . . . . . . 235
Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 235
N8480 Series Power Sensors with Option CFT . . . . . . . . . . . . . . . . . . 237
Frequency specific calibration factors . . . . . . . . . . . . . . . . . . . . . . 237
Sensor calibration tables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 242
Selecting a sensor calibration table . . . . . . . . . . . . . . . . . . . . . . . . 243
Editing/Generating sensor calibration tables . . . . . . . . . . . . . . . . . 246
9 Complementary Cumulative Distribution Function (CCDF)
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 252
Setting the CCDF Table . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 253
Windowed CCDF table . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 253
Expanded CCDF table . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 254
Full Screen CCDF table . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 255
Setting the CCDF Trace . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 256
10 Maintenance
Self Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 262
Power on self test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 262
Front panel selection of Self Tests . . . . . . . . . . . . . . . . . . . . . . . . . 262
Instrument Self Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 263
18 Keysight N1911A/1912A User’s Guide
Remote testing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .264
Test descriptions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .264
Error Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .266
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .266
Error message list . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .268
Operator Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .275
Replacing the power line fuse . . . . . . . . . . . . . . . . . . . . . . . . . . . . .275
Contacting Keysight Technologies . . . . . . . . . . . . . . . . . . . . . . . . . . . .277
Before calling Keysight Technologies . . . . . . . . . . . . . . . . . . . . . . .277
Check the basics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .277
Instrument serial numbers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .278
Erasing Memory Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .280
Returning Your Power Meter for Service . . . . . . . . . . . . . . . . . . . . . . .281
Packaging the power meter for shipment . . . . . . . . . . . . . . . . . . . .281
11 Specifications and Characteristics
Keysight N1911A/1912A User’s Guide 19
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.
20 Keysight N1911A/1912A User’s Guide

List of Figures

Figure 1-1 Dual numeric display . . . . . . . . . . . . . . . . . . . . . . . . . .38
Figure 1-2 Single numeric and analog display . . . . . . . . . . . . . . .40
Figure 1-3 Full screen numeric display . . . . . . . . . . . . . . . . . . . . .41
Figure 1-4 Trace display in both windows . . . . . . . . . . . . . . . . . . .42
Figure 1-5 Trace display in single enlarged window in markers
Figure 1-6 Trace display in single enlarged window in trigger delay
Figure 1-7 Trace display in single enlarged window in trace control
Figure 1-8 Opening the instrument Web interface. . . . . . . . . . . . 50
Figure 1-9 N1912A P-Series power meter Web browser (Welcome
Figure 1-10 Status message before Toggle ID is selected. . . . . . .52
Figure 1-11 Status message when Toggle ID is selected. . . . . . . .52
Figure 1-12 Example of Lan Status message . . . . . . . . . . . . . . . . .53
Figure 1-13 Viewing LAN configuration settings from the Web
Figure 1-14 Password security dialog box. . . . . . . . . . . . . . . . . . . . 55
Figure 1-15 Changing the instrument LAN interface configuration 56
Figure 2-1 Frequency pop-up . . . . . . . . . . . . . . . . . . . . . . . . . . . .60
Figure 2-2 Typical relative measurement display . . . . . . . . . . . . .62
Figure 2-3 Numeric display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .63
Figure 2-4 Measurement configuration screen . . . . . . . . . . . . . . .64
Figure 2-5 Pulse duration measurement between two reference
Figure 2-6 Trace Ref Lvl 1 pop-up . . . . . . . . . . . . . . . . . . . . . . . . .65
Figure 2-7 Trace Ref Lvl 2 pop-up . . . . . . . . . . . . . . . . . . . . . . . . .66
Figure 2-8 Pulse Duration Ref Lvl pop-up . . . . . . . . . . . . . . . . . . .66
Figure 2-9 Simplified measurement path . . . . . . . . . . . . . . . . . . .67
Figure 2-10 Typical channel offset display . . . . . . . . . . . . . . . . . . .68
Figure 2-11 Channel offset indicator . . . . . . . . . . . . . . . . . . . . . . . .69
Figure 2-12 Typical display offset display . . . . . . . . . . . . . . . . . . . .70
Figure 2-13 Frequency dependent offset tables display . . . . . . . . .73
mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .43
mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .44
mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .45
Page) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .51
interface. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .54
levels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .65
Keysight N1911A/1912A User’s Guide 21
Figure 2-14 Frequency dependent offset indicator . . . . . . . . . . . . 73
Figure 2-15 “Edit Offset” display with data added . . . . . . . . . . . . . 75
Figure 2-16 Edit table title pop-up . . . . . . . . . . . . . . . . . . . . . . . . . 76
Figure 2-17 Typical averaged readings . . . . . . . . . . . . . . . . . . . . . . 78
Figure 2-18 Meas Avg Count pop-up . . . . . . . . . . . . . . . . . . . . . . . 79
Figure 2-19 Video Avg Count pop-up . . . . . . . . . . . . . . . . . . . . . . . 83
Figure 2-20 Bandwidth filter shapes . . . . . . . . . . . . . . . . . . . . . . . . 85
Figure 2-21 Video bandwidth pop-up . . . . . . . . . . . . . . . . . . . . . . . 86
Figure 2-22 Example of measurement gates . . . . . . . . . . . . . . . . . 87
Figure 2-23 Gate Setup screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . 88
Figure 2-24 Time Gating Start pop-up . . . . . . . . . . . . . . . . . . . . . . 89
Figure 2-25 Time Gating Length pop-up . . . . . . . . . . . . . . . . . . . . 89
Figure 2-26 Auto Gate Settings display . . . . . . . . . . . . . . . . . . . . . 90
Figure 2-27 Marker 1 Ref Percentage pop-up . . . . . . . . . . . . . . . . 90
Figure 2-28 Marker 2 Ref Percentage pop-up . . . . . . . . . . . . . . . . 91
Figure 2-29 The placement of gates (markers) in Auto Gating mode
91
Figure 2-30 Trace setup display . . . . . . . . . . . . . . . . . . . . . . . . . . . 92
Figure 2-31 Trace Start Time pop-up . . . . . . . . . . . . . . . . . . . . . . . 93
Figure 2-32 X Scale/Division pop-up . . . . . . . . . . . . . . . . . . . . . . . 93
Figure 2-33 Trace Units pop-up . . . . . . . . . . . . . . . . . . . . . . . . . . . 94
Figure 2-34 Y scale maximum value pop-up . . . . . . . . . . . . . . . . . 94
Figure 2-35 Y Scale/Division pop-up . . . . . . . . . . . . . . . . . . . . . . . 95
Figure 2-36 Trigger menu - free run mode . . . . . . . . . . . . . . . . . . . 96
Figure 2-37 Trigger Setting Menu 1 of 2 . . . . . . . . . . . . . . . . . . . . . 97
Figure 2-38 Trigger Level pop-up . . . . . . . . . . . . . . . . . . . . . . . . . . 99
Figure 2-39 Trigger Delay pop-up . . . . . . . . . . . . . . . . . . . . . . . . . 99
Figure 2-40 Trigger setting menu 2 of 2 . . . . . . . . . . . . . . . . . . . . 100
Figure 2-41 Trigger Holdoff pop-up . . . . . . . . . . . . . . . . . . . . . . . 101
Figure 2-42 Trigger Hysteresis pop-up . . . . . . . . . . . . . . . . . . . . . 102
Figure 2-43 TRIG IN and TRIG OUT connection diagram between power
meter and power source. . . . . . . . . . . . . . . . . . . . 105
Figure 2-44 Channel Setup display . . . . . . . . . . . . . . . . . . . . . . . . 106
Figure 2-45 Sensor Mode pop-up. . . . . . . . . . . . . . . . . . . . . . . . . 107
Figure 2-46 Trigger setting menu 1 of 2 . . . . . . . . . . . . . . . . . . . . 108
Figure 2-47 Trigger setting menu 2 of 2 . . . . . . . . . . . . . . . . . . . . 108
Figure 2-48 Limits checking applications . . . . . . . . . . . . . . . . . . . 118
22 Keysight N1911A/1912A User’s Guide
Figure 2-49 Limits checking results . . . . . . . . . . . . . . . . . . . . . . . .118
Figure 2-50 Minimum limit pop-up . . . . . . . . . . . . . . . . . . . . . . . .120
Figure 2-51 Limit failures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .121
Figure 2-52 Sixteen measurements per channel . . . . . . . . . . . . .122
Figure 2-53 Measurement feed example . . . . . . . . . . . . . . . . . . . .123
Figure 2-54 Measurement Setup showing single configuration . .124
Figure 2-55 Function pop-up . . . . . . . . . . . . . . . . . . . . . . . . . . . . .125
Figure 2-56 Measurement type pop-up . . . . . . . . . . . . . . . . . . . .126
Figure 2-57 Gating Number pop-up . . . . . . . . . . . . . . . . . . . . . . .126
Figure 2-58 Measurement Setup showing combined configuration .
127
Figure 2-59 Measurement example display . . . . . . . . . . . . . . . . .128
Figure 2-60 Maximum limit pop-up . . . . . . . . . . . . . . . . . . . . . . . .129
Figure 2-61 Minimum limit pop-up . . . . . . . . . . . . . . . . . . . . . . . .130
Figure 2-62 Increment and decrement multiplier . . . . . . . . . . . . .130
Figure 2-63 Recorder Minimum pop-up . . . . . . . . . . . . . . . . . . . .132
Figure 2-64 Recorder Maximum pop-up . . . . . . . . . . . . . . . . . . . .132
Figure 2-65 Increment and decrement multiplier . . . . . . . . . . . . .134
Figure 2-66 Measurement example for Option H01 . . . . . . . . . . .135
Figure 2-67 Example screenshot of the RF Multitone waveform .136
Figure 2-68 Save/Recall screen . . . . . . . . . . . . . . . . . . . . . . . . . . .137
Figure 2-69 Save confirm pop-up . . . . . . . . . . . . . . . . . . . . . . . . .138
Figure 2-70 File name pop-up . . . . . . . . . . . . . . . . . . . . . . . . . . . .138
Figure 2-71 Recall pop-up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .139
Figure 2-72 Zeroing pop-up . . . . . . . . . . . . . . . . . . . . . . . . . . . . .140
Figure 2-73 Calibration wait pop-up . . . . . . . . . . . . . . . . . . . . . . .142
Figure 2-74 Please zero and calibrate window . . . . . . . . . . . . . . .143
Figure 2-75 Preset display options . . . . . . . . . . . . . . . . . . . . . . . .145
Figure 2-76 Short/long cable option . . . . . . . . . . . . . . . . . . . . . . .147
Figure 3-1 P-Series sensor default channel setup . . . . . . . . . . .153
Figure 3-2 Trace display in dual channel power meter . . . . . . . .154
Figure 3-3 Trace display in single enlarged window . . . . . . . . . .155
Figure 3-4 Trace display with gate control menu . . . . . . . . . . . .156
Figure 3-5 Trace display with trigger delay menu . . . . . . . . . . . .158
Figure 3-6 Trace display with trace control menu . . . . . . . . . . . 159
Figure 3-7 Trace display in full screen . . . . . . . . . . . . . . . . . . . . .161
Figure 3-8 P-Series power sensor default channel setup . . . . . .162
Keysight N1911A/1912A User’s Guide 23
Figure 3-9 Channel offset display . . . . . . . . . . . . . . . . . . . . . . . . 163
Figure 4-1 Bandwidth Filter Shapes . . . . . . . . . . . . . . . . . . . . . . 167
Figure 4-2 E–Series E9320 power sensor default channel setup . . .
168
Figure 4-3 Frequency pop-up . . . . . . . . . . . . . . . . . . . . . . . . . . . 171
Figure 4-4 E-Series E9320 power sensor channel setup when trigger
is selected . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 174
Figure 4-5 Trace display in dual channel power meter . . . . . . . 174
Figure 4-6 Trace display in single enlarged window . . . . . . . . . 175
Figure 4-7 Trace display with gate control menu . . . . . . . . . . . . 176
Figure 4-8 Trace display with trigger delay menu . . . . . . . . . . . 177
Figure 4-9 Trace display with trace control menu . . . . . . . . . . . 179
Figure 4-10 Trace display in full screen . . . . . . . . . . . . . . . . . . . . 180
Figure 4-11 E-Series E9320 power sensor default channel setup 181
Figure 4-12 Channel offset display . . . . . . . . . . . . . . . . . . . . . . . . 182
Figure 5-1 E9300 E-Series auto-averaging settings . . . . . . . . . 187
Figure 5-2 . . . E9300 E-Series sensor default channel setup 188
Figure 5-3 Frequency pop-up . . . . . . . . . . . . . . . . . . . . . . . . . . . 191
Figure 5-4 Spread spectrum signal . . . . . . . . . . . . . . . . . . . . . . . 192
Figure 5-5 Wideband CDMA error of E-Series E9300 power sensor
versus corrected CW sensor. . . . . . . . . . . . . . . . . 193
Figure 5-6 CDMA (IS-95A): 9Ch Fwd . . . . . . . . . . . . . . . . . . . . . 193
Figure 5-7 Calibration factors versus frequency . . . . . . . . . . . . . 194
Figure 6-1 E-Series CW sensor auto-averaging settings . . . . . . 203
Figure 6-2 E-Series E4410 sensor default channel setup . . . . . 204
Figure 6-3 Frequency pop-up . . . . . . . . . . . . . . . . . . . . . . . . . . . 206
Figure 7-1 8480 Series auto-averaging settings . . . . . . . . . . . . 209
Figure 7-2 8480 Series sensor default channel setup . . . . . . . . 210
Figure 7-3 Reference Calibration Factor pop-up window . . . . . 214
Figure 7-4 Calibration factor pop-up window . . . . . . . . . . . . . . 215
Figure 7-5 Calibration factor display . . . . . . . . . . . . . . . . . . . . . . 215
Figure 7-6 Sensor table selected . . . . . . . . . . . . . . . . . . . . . . . . 218
Figure 7-7 Frequency dependent offset indicator . . . . . . . . . . . 219
Figure 7-8 Frequency/calibration table display . . . . . . . . . . . . . 220
Figure 7-9 “Sensor Tbls” screen . . . . . . . . . . . . . . . . . . . . . . . . . 222
Figure 7-10 “Edit Cal” display . . . . . . . . . . . . . . . . . . . . . . . . . . . . 223
Figure 7-11 Edit table title pop-up . . . . . . . . . . . . . . . . . . . . . . . . 223
24 Keysight N1911A/1912A User’s Guide
Figure 8-1 Auto-averaging settings . . . . . . . . . . . . . . . . . . . . . . .232
Figure 8-2 N8480 Series sensor (excluding Option CFT) default chan-
nel setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .233
Figure 8-3 N8480 Series sensor with Option CFT default channel
setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .233
Figure 8-4 Frequency pop-up . . . . . . . . . . . . . . . . . . . . . . . . . . .236
Figure 8-5 Reference calibration factor pop-up window . . . . . .238
Figure 8-6 Calibration factor pop-up window . . . . . . . . . . . . . . .239
Figure 8-7 Calibration factor displayed . . . . . . . . . . . . . . . . . . . .240
Figure 8-8 Sensor table selected . . . . . . . . . . . . . . . . . . . . . . . . .244
Figure 8-9 Frequency dependent offset indicator . . . . . . . . . . . .244
Figure 8-10 Frequency/calibration table display . . . . . . . . . . . . .245
Figure 8-11 “Sensor Tbls” screen . . . . . . . . . . . . . . . . . . . . . . . . .247
Figure 8-12 “Edit Cal” display . . . . . . . . . . . . . . . . . . . . . . . . . . . .248
Figure 8-13 Edit table title pop-up . . . . . . . . . . . . . . . . . . . . . . . .248
Figure 9-1 CCDF table in windowed display . . . . . . . . . . . . . . . .253
Figure 9-2 CCDF table in expanded display . . . . . . . . . . . . . . . .254
Figure 9-3 CCDF table in full-screen display . . . . . . . . . . . . . . .255
Figure 9-4 Trace Display Select control menu . . . . . . . . . . . . . .256
Figure 9-5 CCDF trace display with Marker Trace menu . . . . . .257
Figure 9-6 Scale/Div pop-up . . . . . . . . . . . . . . . . . . . . . . . . . . . .259
Figure 9-7 Selection of CCDF trace to be stored as a reference
trace . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .259
Figure 10-1 Self test complete . . . . . . . . . . . . . . . . . . . . . . . . . . .263
Figure 10-2 Error indicator position . . . . . . . . . . . . . . . . . . . . . . . .266
Figure 10-3 Replacing the fuse . . . . . . . . . . . . . . . . . . . . . . . . . . .276
Figure 10-4 Secure erase status pop-up . . . . . . . . . . . . . . . . . . . .280
Keysight N1911A/1912A User’s Guide 25
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.
26 Keysight N1911A/1912A User’s Guide

List of Tables

Table 1-1 Type of Lan Status message . . . . . . . . . . . . . . . . . . . .53
Table 2-1 Measurement units - Single channel meters . . . . . . . 58
Table 2-2 Measurement units - Dual channel meters . . . . . . . . .58
Table 2-3 P-Series Sensor Video Bandwidth Setting . . . . . . . . .84
Table 2-4 Power and frequency sweep mode compatibility table .
103 Table 2-5 Power and frequency sweep measurement functions
table . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .104
Table 2-6 Range of values for window limits . . . . . . . . . . . . . . .119
Table 2-7 Ranges of recorder output setting . . . . . . . . . . . . . . .133
Table 2-8 Preset compatibility matrix for E-Series E9320 power
sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .146
Table 4-1 Sensor Bandwidth . . . . . . . . . . . . . . . . . . . . . . . . . . .166
Table 5-1 Power sensor connection requirements . . . . . . . . . .189
Table 7-1 8480 Series connection requirements . . . . . . . . . . . .210
Table 7-2 Installed power sensor models . . . . . . . . . . . . . . . . .221
Table 8-1 Power range in the Range setting . . . . . . . . . . . . . . .230
Table 8-2 N8480 Series connection requirements . . . . . . . . . .234
Table 8-3 Installed power sensor models . . . . . . . . . . . . . . . . .246
Keysight N1911A/1912A User’s Guide 27
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.
28 Keysight N1911A/1912A User’s Guide
Keysight N1911A/1912A P-Series Power Meters User’s Guide

1 Introduction

LXI Class-C Compliant Power Meter 30 Rack Mounting 30 Power Meter and Sensor Capability 31 Conventions Used in this Guide 32 Front Panel Keys and Connections 33 The Display Layout 38 Window Symbols and Pop-ups 46 Using the Instrument Web Browser 50
This chapter introduces you to the front panel display and instrument Web browser of the P-Series power meter.
29
1Introduction

LXI Class-C Compliant Power Meter

P-Series power meter is a LXI Class C compliant instrument, developed using LXI Technology. LXI, an acronym for LAN eXtension for Instrumentation, is an instrument standard for devices that use the Ethernet (LAN) as their primary communication interface.
Hence, it is an easy-to-use instrument especially with the usage of an integrated Web browser that provides a convenient way to configure the instrument’s functionality.

Rack Mounting

The N1911A/1912A can be mounted in a standard 19-inch rack. Rack mount kits are available as listed below. Support rails are also required for rack mounting. These are normally supplied with the rack and are not included with the rack mount options.
If you are installing an instrument on top of the N1911A/1912A, ensure that the instrument does not obstruct the ventilation holes at the top of the N1911A/ 1912A. If required, use a filler panel above the N1911A/1912A to ensure adequate space for air circulation.
Option Description
N1911A Option 908 Rack mount kit for one instrument
N1911A Option 909 Rack mount kit for two instruments
N1912A Option 908 Rack mount kit for one instrument
N1912A Option 909 Rack mount kit for two instruments
30 Keysight N1911A/1912A User’s Guide

Power Meter and Sensor Capability

NOTE
Your P-Series power meter is compatible with Keysight P-Series, E-Series E9320, E-Series E9300, E-Series E4410, 8480 Series and the N8480 Series power sensors. However, not all sensor and meter combinations have the same features or capabilities. The main differences are as below:
Introduction 1
Features
Average power of CW signal * * * * * *
Average power of modulated signal * * * * *
Peak power * *
Cal factors stored on EEPROM * * * *
>200 readings/sec * * * *
Peak/burst average power * *
Time gated measurements * *
Rising edge trigger * *
Falling edge trigger * *
[a] Not applicable for N8480 Series power sensors with Option CFT
P-Series
N1920
E-Series
E9320
E-Series
E9300
E-Series
E4410
8480 Series
– The E-Series, 8480 Series and N8480 Series power sensors require N1917A/
B/C/D cables when connected to the P-Series power meters.
– The N8480 Series power sensors used in this user’s guide is referring to all
sensors from the N8480 Series unless otherwise stated.
N8480
Series
[a]
*

Specifications

The specifications for the power meter are listed in Chapter 11, "Specifications
and Characteristics".
Keysight N1911A/1912A User’s Guide 31
1Introduction
Softkey
Channel
Channel A
Channel B
Softkey

Conventions Used in this Guide

The following conventions are used throughout this guide.
This symbol and text represents a labeled key on the power meter front panel.
This symbol and text represents a labeled softkey and is used to indicate you should press the unmarked key beside the displayed text.
Message This text represents a displayed message.
Parameter This is used to represent a parameter, value, or title.
This User’s Guide describes the operation for both the single channel and the dual channel power meter. To identify channels on a dual channel meter a
“Channel”
softkey on an N1911A meter becomes and an N1912A.
When you are asked to press “the channel” in a procedure, make sure you select the relevant channel.
32 Keysight N1911A/1912A User’s Guide

Front Panel Keys and Connections

This section briefly describes the functions of the front panel keys and connectors.
These keys are located to the left of the display.
Key Function
Press this key to preset the power meter. An option list of pre-installed measurement configurations is presented. Select an option and press the confirm key.
Introduction 1
Press this key to control the power meter from the front panel when it is operating via the remote interfaces (when Local Lock Out is not enabled).
Press this key to select the upper or lower measurement window. The selected window is highlighted by a blue line on the right side of the window. Any measurement setup you create is performed in the selected window.
Press this key to choose windowed, expanded, or full-screen display of a numeric measurement. It also provides quick access to the Gate Control screen and menu when a trace window is selected.
Press this key to switch the meter between on and standby. When power is supplied, the background LED is red. Pressing the key, switches the power meter on and the background LED is green.
Keysight N1911A/1912A User’s Guide 33
1Introduction
These keys are located along the lower edge of the display.
Key Function
Press this key to access general configuration-menus, such as GPIB address. You can also access the Self-Test and Table entry menus. The measurement screen remains visible.
Press this key to access the channel configuration menus. Channel parameters such as averaging and offsets are configured from this menu.
Press this key to access the triggering menu. Unless a P-Series or an E-Series E9320A Sensor is connected, all the menu keys are disabled (grayed out).
Press this key to setup relative measurements or set display offsets. Use this key to configure the selected measurement.
Press this key to access the measurement display menu. You can choose the displayed measurement resolution, units and display format.
Use this key together with to configure measurement displays.
34 Keysight N1911A/1912A User’s Guide
Introduction 1
These keys are all associated with the menu labels and data entry. They are located to the right of the display.
Key Function
Press this key to return to the previous screen. This key also cancels pop-up entry.
These unmarked keys are called ‘softkeys’ and are referred to by the text on the display next to them.
For example, during a Preset, you are given an option to confirm the command. Press to continue, that is, press the softkey beside the displayed word ‘confirm’.
The lowest of the unmarked softkeys is used when there is a two page menu to be displayed. For example, a 1 of 2 is displayed beside the key indicating the first page of a two page menu. Press the key to access the next page or second page. (A 2 of 2 is displayed).
Keysight N1911A/1912A User’s Guide 35
1Introduction
These keys and connectors are associated with the measurement channels and are located on the right-hand side of the front panel.
Key Function
The arrow keys are used for navigation around the parameter entry screens. The up and down arrows are used for selecting values from a pop-up list, they can also be used for stepping two values, X Scale and Y Scale. They are also used to enter text, for example, table names.
Press this key to select a highlighted field to allow data entry, check a checkbox and terminate entry of a popup list.
Press this key to access the zero and calibration menus.
36 Keysight N1911A/1912A User’s Guide
Press this key to start and stop the acquisition of the measurement.
Press these keys to enter numeric values in the pop-up fields, for example, the offset values. To complete the entry use the softkey.
Connector Function
The power reference is a 1 mW (0 dBm) 50 MHz signal available from a 50 W type-N connector. It is used for calibrating an 8480, E-Series or N8480 Series power sensor and meter system. If the meter is configured with Option 003, the connector is fitted to the rear panel. The Green LED beside the connector is lit when the calibrator is turned on.
Introduction 1
The sensor input connectors (N1912A shown, the N1911A has one input). If the meter is configured with Option 003, the connectors are fitted to the rear panel.
Keysight N1911A/1912A User’s Guide 37
1Introduction
Disp Type
1
3
4
6
9
2
8
5
7

The Display Layout

Figure 1-1 shows the display layout when two windows are configured in dual
numeric mode.
Pressing , the other display formats are available.
Figure 1-1 Dual numeric display
1 The status reporting line displays messages and the control status of the
power meter. For example, the status can be either RMT (remote, GPIB, USB or LAN operation) or LCL (local, front panel operation).
The message fields indicate ERR for any error conditions that occur or informing you to Please Zero the power sensor.
2 The blue highlight on the right hand side of the window shows it is the
currently selected measurement display line. This measurement line is the Upper Window/Upper Measurement.
3 The measured channel is shown. With a P-Series or an E-Series E9320 power
sensor connected, and channel in trigger mode, the associated gate number is shown.
4 The associated measurement type is shown below the channel and gate
number.
5 This displays the measurement units, either dBm, dB, Watts, or percent (%).
38 Keysight N1911A/1912A User’s Guide
Introduction 1
NOTE
With a P-Series or an E-Series E9320 power sensor connected, a measurement result of –270 dBm indicates the input power level is beyond the sensitivity of the sensor.
6 With a P-Series or an E-Series E9320 power sensor connected, you can make
combined measurements with a single channel meter. A dual channel meter extends this feature across both channels.
7 This displays the number of pages in the current menu. For example, 1 of 2
indicates that there are two pages in the menu and the first page is currently displayed. Pressing the softkey displays the next page, indicated by 2 of 2. (Press the softkey to display the previous menu page.)
8 The available softkey labels are displayed in these four fields. Additionally,
settings associated with the labeled function are displayed under the label.
9 This field displays the menu title. For example, Channel Setup or press
and the Zero/Cal menu is displayed.
Keysight N1911A/1912A User’s Guide 39
1Introduction
14 13
11
12
15
10
Figure 1-2 Single numeric and analog display
10 Figure 1-2 shows the default display mode of two measurement windows. The
upper measurement window has a blue highlight on the right hand side of the window showing it has been selected.
Using the , , or keys you can change the selection of the
measurement window.
Using the key on numeric measurement results window you can
choose either two rectangular windows, a single enlarged window, or a full screen display by pressing. The display style is applied to the currently selected window or measurement line.
11 The upper window is configured to show a single numeric display.
12 The lower window is configured to show an analog meter which displays the
measurement result and the meter scaling.
13 With a P-Series or an E-Series E9320 power sensor connected, the symbol
shows the trigger state (Free Run), (Stopped), (Negative Slope), or (Positive Slope).
14 The channel measurement frequency.
15 This displays the connected sensor, the offset value, and the acquisition mode
on the channel. On dual channel models it shows for both channels.
40 Keysight N1911A/1912A User’s Guide
Introduction 1
21
20 19
18
16
17
NOTE
Figure 1-3 Full screen numeric display
16 Figure 1-3 shows a single numeric full screen displaying a relative result.
This field displays Rel if relative mode is on.
17 This field indicates the measurement result is beyond the configured upper or
lower limit. If the measurement is within the limits this field is empty. If the measurement result is less than the minimum limit set, Undr Lmt is displayed. If the measurement result is more than the maximum limit set, Over Lmt is displayed.
18 This field displays Ofs if an offset is set.
19 This field displays Rng Hld if a range is selected.
20 This field displays Dty Cyc if a duty cycle is set.
21 The information in this field is displayed on two lines and depends on the
sensor type, sensor calibration table, frequency dependent offset table currently selected, and the measurement frequency.
The following trace displays are only available when a P-Series or an E-Series E9320 power sensor is connected.
Keysight N1911A/1912A User’s Guide 41
1Introduction
23
22
NOTE
Sing Trig
Cont Trig
Trigger
Trace
Free Run
Figure 1-4 Trace display in both windows
22 Figure 1-4 shows both windows configured to trace display mode. This is only
available with an P-Series or an E-Series E9320 power sensor is connected. The captured trace, channel, and scaling are displayed. The lower window is the selected window.
23 The indicates the point on the trace where trigger event occurs.
Single or continuous triggering ( or ) must be selected from the Acqn menu to view a trace window. The Acqn menu is accessed by pressing
, . is disabled when is selected.
42 Keysight N1911A/1912A User’s Guide
Introduction 1
28
26
27
25
24
29
Gate
Marker
NOTE
Figure 1-5 Trace display in single enlarged window in markers mode
Figure 1-5 shows a single enlarged window with the Gate Ctrl menu and
associated tables and markers. The Gate Control is the default display when using trace display.
24 Pressing scrolls through the 4 gates available for each channel.
25 The markers, 1 and 2, indicate the start and end points of the selected gate.
Pressing toggles between the two markers. You can use the and
keys to move the active marker along the trace.
26 The highlighted table, with the green border, shows the time (Time:) and the
instantaneous power level (Pow:) of the active marker at the configured point. A negative time value indicates a measurement before the trigger point.
Gate timing parameters are all related to your chosen trigger point. This may be different from the timing of the triggering event if you have configured a trigger delay. Refer to item No 30 for more information.
27 This table shows the time (Time:) and the instantaneous power level (Pow:) of
the inactive marker at the configured point.
28 The highlighted marker shows the active marker at the configured point.
29 This table shows the width of the gate
average, peak, and peak-to-average ratio power measurements within the
ΔT: (time between the markers), the
active gate.
Keysight N1911A/1912A User’s Guide 43
1Introduction
31
32
30
Select TgDel
Figure 1-6 Trace display in single enlarged window in trigger delay mode
Figure 1-6 shows a single enlarged window with the Gate Ctrl menu and
associated tables and trigger delay.
30 Pressing removes the gate markers and displays the trigger
marker(s). The indicates when the trigger event occurs, whilst shows the delayed trigger point. When the two points coincide only the trigger is
shown.
In the example shown in Figure 1-6, the appears ahead of the as a trigger delay of –5.00 µs has been configured, placing the measurement trigger before the trigger event. You can configure the trigger delay by pressing and entering a numeric value.
To indicate an off-screen trigger event, is displayed. To indicate an off-screen trigger point, is displayed.
31 The selected channel and gate number is displayed.
32 The trigger setting and slope is displayed. Examples shown in Figure 1-6 are
Internal Channel A and a Positive Slope.
44 Keysight N1911A/1912A User’s Guide
Introduction 1
33
34
35
Figure 1-7 Trace display in single enlarged window in trace control mode
Figure 1-7 shows a single enlarged window with the Trace Ctrl menu with
information on the pulse and associated X and Y control.
33 This is the X and Y trace setup fields. Using the or keys you can
highlight the item and change its value.
34 The current settings of the X and Y scale are displayed in this reporting line.
35 This table shows 8 automatic time measurements performed on the first
complete captured pulse after the trigger. The 8 measurements are rise time
, fall time , time to positive occurrence , time to negative
occurrence , pulse period , pulse width , duty cycle 50.0 % and pulse repetitive frequency 12.5 kHz.
Keysight N1911A/1912A User’s Guide 45
1Introduction

Window Symbols and Pop-ups

There are several different graphic symbols and pop-up windows that can occur on the power meter display. These can occur for a variety of reasons, for example:
– An error or warning occurs
– You are required to wait while the power meter carries out a procedure
– You are required to select an entry from a list
– You are required to enter a numeric value
There are three different colors used to signify the pop-up status:
– Green - used to allow data entry
– Orange - used to display information
– Red - used to display an error

Warning symbol pop-up

The warning symbol is displayed either in a pop-up window or directly in the measurement window when such an event occurs. A pop-up window is displayed for approximately two seconds. The text in the pop-up window gives details of the warning type, for example, to indicate that a power sensor has insufficient bandwidth or a previous entered frequency value in a table. Depending on the severity of the warning, the pop-up may be displayed in Orange or Red.
46 Keysight N1911A/1912A User’s Guide

Wait symbol pop-up

The wait symbol is displayed when the power meter is carrying out a procedure and no action is required from you. The symbol appears in a pop-up window. It may appear, for example, during a calibration.

Confirm symbol pop-up

This type of pop-up window is displayed when you are required to press Confirm to verify your previous selection. For example, prior to a Save being carried out.
Introduction 1

Numeric entry pop-up

This type of pop-up window is displayed when you need to modify numeric data. The numeric keys allow you to enter the value.
Keysight N1911A/1912A User’s Guide 47
1Introduction

Text entry pop-up

This type of pop-up window is displayed when you need to modify alphanumeric data, for example, table names. The up/down arrow keys increment and decrement the alphanumeric digit that the cursor is currently positioned. The left/ right arrow keys move the cursor to another alphanumeric digit.

Trace scaling pop-up

This pop-up window is displayed when you are in the Trace Control menu and you can either use the up/down arrow keys for increment and decrement, or the numeric keys to set, the values of the X scale and Y scale.
48 Keysight N1911A/1912A User’s Guide

List pop-up

This pop-up window is displayed when you are required to select an entry from a list. Use the up/down arrow keys to highlight your choice. Press select to complete the entry.
Introduction 1
Keysight N1911A/1912A User’s Guide 49
1Introduction
NOTE

Using the Instrument Web Browser

P-Series power meters network settings can be configured using its Web-based interface (Web browser). The instrument’s Web browser can be opened from Keysight Connection Expert as shown in Figure 1-8.
Alternatively, the instrument’s Web-based interface can also be opened directly from a Web browser by entering the instrument’s IP address or hostname in the browser’s ‘address’ window.
Figure 1-8 Opening the instrument Web interface.
50 Keysight N1911A/1912A User’s Guide
An example of the Web browser is shown in Figure 1-9.
Remote Interfaces
Introduction 1
Keysight N1911A/1912A User’s Guide 51
Figure 1-9 N1912A P-Series power meter Web browser (Welcome Page)
Instrument on the network can be physically identified through the Status message on Remote Interfaces screen by clicking Toggle ID within the Web browser. This changes the status of the Instrument ID to ON or OFF.
Before the instrument identification starts, press
, to
view the Status message.
When Instrument ID is toggled to ON, this causes the Status message “LAN identify start” in Remote Interface screen on the front panel to blink 3 times. When the Toggle ID is selected again, the default Status message “Running” is returned. See Figure 1-10 and Figure 1-11.
1Introduction
Status message “LAN identity start” to blink 3 times when Toggle ID is selected.
Remote Interfaces
Figure 1-10 Status message before Toggle ID is selected.
52 Keysight N1911A/1912A User’s Guide
Figure 1-11 Status message when Toggle ID is selected.
When power meter is configured to LAN, the Lan Status on screen will show the LAN error condition and status of the LAN configuration connection. There are six types of Lan Status messages that may occur. See
Table 1-1. See also Figure 1-12 for the example of Lan Status message.
Table 1-1 Type of Lan Status message
Type of message Description
Introduction 1
Lan: No Fault Status: Initialized
Lan: No Fault Status: Running
Lan: Fault Status: Initialization failed
Lan: Fault Status: Disconnected
Lan: – Status: Restarting
Lan: DHCP Not Available Status: Running
– A valid IP address is successfully obtained using selected LAN
configuration and the network state is initialized.
– A valid IP address is successfully obtained using selected LAN
configuration, while network is running.
– IP conflict occurs, or – IP address failed to be obtained using selected LAN
configuration, or
– No LAN configuration is selected.
– LAN cable is unplugged.
– Restart the network and try to obtain an IP address using
selected LAN configuration.
– Unable to obtain IP address from DHCP server (if user select
DHCP configuration)
– IP address obtained from Auto-IP or manual configuration.
Figure 1-12 Example of Lan Status message
Keysight N1911A/1912A User’s Guide 53
1Introduction
NOTE
For more details on remote interface configuration, refer to P-Series Power Meters Installation Guide.

Editing the Instrument’s LAN Settings

Once communication path to the instrument has been established, the instrument’s LAN configuration can be viewed and modified using the Web browser.
On the Welcome Page, click View and Modify Configuration. This opens the configuration window shown in Figure 1-13.
Figure 1-13 Viewing LAN configuration settings from the Web interface.
54 Keysight N1911A/1912A User’s Guide
Introduction 1
NOTE
Remote Interfaces
1 of 2
LAN Reset
NOTE
LAN Reset
NOTE
To edit parameters shown, click Mod ify Configuration. The Enter Password dialog box appears as shown in Figure 1-14.
Figure 1-14 Password security dialog box.
Click the Submit (accept the default password) and the window opens as shown in Figure 1-15. The default password is “keysight”.
For P-Series power meters with latest firmware upgrade to be LXI-C compliant, a LAN reset needs to be performed to ensure that the password is reset to default. See LAN reset procedure as below.
Procedure:
Keysight N1911A/1912A User’s Guide 55
1 Press , to display the Remote Interfaces screen.
2 Press softkey to display the second page of the Remote I/F menu.
3 Press softkey to reset the LAN settings to default.
When softkey is pressed, it will also change the GPIB address to default.
You can change the password from the Configuring your N1912A Power Meter window as shown in Figure 1-15. Scroll down the Parameter column until you locate the Change Password parameter.
1Introduction
Figure 1-15 Changing the instrument LAN interface configuration
56 Keysight N1911A/1912A User’s Guide

General Power Meter Functions 2

2 General Power Meter
Functions
Setting the Units of Measurement 58 Setting the Measurement Frequency 60 Setting the Resolution 61 Making Relative Measurements 62 Setting the Pulse Reference Levels 64 Setting Offsets 67 Setting Measurement Averaging 78 Step Detection 81 Setting the Video Averaging 82 Setting the Video Bandwidth 84 Setting Measurement Channel Gates 87 Setting up the Channel Trace 92 Setting the Trigger 96 Setting External Trigger for Normal and Average Power Measurement 103 Setting Measurement Limits 118 Setting the Measurement Display 122 Scaling the Analog Display 129 Recorder Output 131 Saving and Recalling Power Meter States 137 Zeroing and Calibrating the P-Series Sensor 140 Presetting the Power Meter 145 Setting the Cable Short/Long 147
This chapter describes the general operation of the P-Series power meter.
Keysight N1911A/1912A User’s Guide 57
2 General Power Meter Functions
Units
dBmWdB
%
NOTE

Setting the Units of Measurement

The Units menu is used to select the measurement units for the currently selected window. These can either be logarithmic (dBm or dB) or linear (Watt or %) units.
Presetting ( ) the power meter sets the measurement units to dBm (logarithmic units). Tab le 2-1 and Table 2-2 show units that are applicable to each
measurement mode.
Press , . Select the unit of measurement from , , , and
. Softkeys which cannot be selected in your particular mode of operation are
grayed out.
When the measurement unit is set to Watt (W), it is possible that negative power results are displayed when measuring low power levels.
Tab le 2-1 Measurement units - Single channel meters
Measurement Mode Relative Mode Off Relative Mode On
Log dBm dB
Linear Watt %
Tab le 2-2 Measurement units - Dual channel meters
Measurement Mode Relative Mode Off Relative Mode On
Ratio
Difference Log dBm dB
Log dB dB
Linear % %
Linear Watt %
58 Keysight N1911A/1912A User’s Guide
General Power Meter Functions 2
mW
μW
nW
2 of 2
W
Increment multiplier
Decrement multiplier
Cancel
Increment Multiplier
Decrement Multiplier
W
W
NOTE

Selecting units of measurement from the softkeys

In some menus, for example, units in the Trace Setup menu, you are required to enter the units of measurement for power. In some cases, due to the availability of wide power range, the following menu is displayed:
Keysight N1911A/1912A User’s Guide 59
Pressing or increases or decreases the multiplier shown in front of .Pressing after the correct multiplier has been
selected confirms the entry.
Invalid softkeys are grayed out so the value cannot be entered.
2 General Power Meter Functions
MHz
GHz

Setting the Measurement Frequency

Entering the frequency of the RF signal you are measuring optimizes the accuracy and minimizes measurement uncertainty, especially when making comparative measurements between signals.

Procedure

Set the measurement frequency as follows:
1 Press . On dual channel meters select the required channel.
2 Use the and keys to highlight the Frequency value field and press
to display the Frequency pop-up. Use the numeric keypad to enter the
required value in the Frequency pop-up window.
Figure 2-1 Frequency pop-up
3 Confirm your choice by pressing or .
4 Press key to close the Channel Setup screen.
60 Keysight N1911A/1912A User’s Guide

Setting the Resolution

Resolution
Resolution
The resolution of each of the power meter’s numeric type windows can be set to four different levels (1, 2, 3 or 4).
These four levels represent:
1, 0.1, 0.01, 0.001 dB respectively if the measurement suffix is dBm or dB.
1, 2, 3 or 4 significant digits respectively if the measurement suffix is W or %.
The default value is 0.01 dB (3 digits).
To set the resolution on the currently selected window:
1 Press . The current setting of the resolution is highlighted under the
2 To change this setting, press repeatedly until the required
resolution setting is highlighted.
General Power Meter Functions 2
softkey.
Keysight N1911A/1912A User’s Guide 61
2 General Power Meter Functions
Meas Select
Relative setting
field
Rel field
Original result
Selected Window/
Measuremen
Result field

Making Relative Measurements

Relative mode enables comparison of a measurement result to a reference value. The relative reading, or difference, can be displayed in either dB or % terms. When the measurement result is displayed in % a prefix multiplier may be shown.

Procedure

1 Press to display the Measurement Setup menu.
Figure 2-2 shows a Measurement Setup display and the relative measurement
items labeled.
2 Select the window you wish to set a reference value on by pressing the
key. The currently selected window/measurement is displayed.
3 Use the and keys to highlight the Relative setting field.
62 Keysight N1911A/1912A User’s Guide
Figure 2-2 Typical relative measurement display
4 Press to check the Relative setting field.
5 Confirm that the power meter is measuring the signal you want to use as the
reference. This is displayed under the Result field.
General Power Meter Functions 2
NOTE
Rel indicator
NOTE
6 Press the , , key to highlight the Rel field.
7 Press and the original result value will appear on the right hand side of
the Rel check field.
8 The relative value displayed under Result field will change as the measured
signal varies.
If you return the power meter to display the numeric display, a Rel symbol will be displayed in the measurement window it is applied to.
Figure 2-3 Numeric display
The Rel symbol is not displayed when the associated measurement is displayed in Dual Numeric or Analog format.
Keysight N1911A/1912A User’s Guide 63
2 General Power Meter Functions
1 of 2
Meas Config
Channel A B
NOTE
NOTE

Setting the Pulse Reference Levels

– Press and proceed to the second page of Sys/Inputs menu by
pressing the softkey.
– Press softkey to display the Measurement Configuration screen.
– Press the softkey to select the desired channel for a dual channel
power meter (N1912A).
Figure 2-4 Measurement configuration screen
Trace Ref Lvl 1 and Trace Ref Lvl 2 are used in calculation of transition durations and occurrences.
Pulse Duration Ref Lvl allows pulse duration measurements between non-standard reference levels.
64 Keysight N1911A/1912A User’s Guide
General Power Meter Functions 2
Trace Ref Lvl 2 = 90%
Pulse Duration Ref Lvl = 50%
Trace Ref Lvl 1 = 10%
Measured pulse duration
%
Figure 2-5 Pulse duration measurement between two reference levels
1 Use the and keys to highlight the Trace Ref Lvl 1 field.
Keysight N1911A/1912A User’s Guide 65
2 Press to display the Trace Ref Lvl 1 pop-up, and use the numeric
keypad to enter the intended value in the Trace Ref Lvl 1 pop-up window.
Figure 2-6 Trace Ref Lvl 1 pop-up
3 Complete the entry by pressing the softkey.
4 Use the and keys to highlight the Trace Ref Lvl 2 field.
2 General Power Meter Functions
%
%
NOTE
5 Press to display the Trace Ref Lvl 2 pop-up, and use the numeric
keypad to enter the intended value in the Trace Ref Lvl 2 pop-up window.
Figure 2-7 Trace Ref Lvl 2 pop-up
6 Complete the entry by pressing the softkey.
7 Use the and keys to highlight the Pul se Duration Ref Lvl field.
8 Press to display the Pul se Duration Ref Lvl pop-up, and use the
numeric keypad to enter the intended value in the Pulse Duration Ref Lvl pop-up.
Figure 2-8 Pulse Duration Ref Lvl pop-up
9 Complete the entry by pressing the softkey.
By default, the Trace Ref Lvl 1, Trace Ref Lvl 2 and Pulse Duration Ref Lvl settings are set to 10 %, 90 % and 50 % respectively.
66 Keysight N1911A/1912A User’s Guide

Setting Offsets

Channel functions
Window functions
Channel A
Cal Factor
Freq Dep Offset
Channel Offset
Channel B
Cal Factor
Freq Dep Offset
Channel Offset
Measurement Highway
Upper Window
Lower Window
Display Offset
Display Offset
Relative
Relative
Display Offset
Display Offset
Relative
Relative
The power meter can be configured to compensate for a signal loss or gain in your test setup. The power meter allows you to apply offsets at three different points in the measurement path.
General Power Meter Functions 2
Figure 2-9 Simplified measurement path
Figure 2-9 shows that how you can apply a Channel Offset or a
Frequency Dependent Offset prior to any mathematical functions. These allow
you to compensate each channel individually. An overall offset can be applied if required using the Display Offset.
Keysight N1911A/1912A User’s Guide 67
2 General Power Meter Functions
Offsets
Offset setting field
Offset value field
dB

Setting channel offsets

This gain or loss is applied to the measured power before any mathematical functions, display offsets or relative functions are included.
Offsets are entered in dB and the range is –100 dB to +100 dB.
A positive value compensates for a loss and a negative value compensates for a gain.

Procedure

To enter a channel offset:
1 Press to display the Channel Setup screen. Confirm the channel
requiring setup is displayed.
2 Press to display the Offsets Setup.
3 Use the and keys to highlight the Offset setting field.
68 Keysight N1911A/1912A User’s Guide
4 Press to check the Offset setting field.
Figure 2-10 Typical channel offset display
5 Press to highlight the Offset value field and press to display the
Offset pop-up. Use the numeric keypad to enter the required value in the Offset pop-up window.
6 Confirm your choice by pressing .
General Power Meter Functions 2
Offset indicator
NOTE
7 Press key to complete the offset entry. If either a channel or a display
offset is set, the Ofs indicator is displayed.
Figure 2-11 Channel offset indicator
The Ofs symbol is not displayed when the associated measurement is displayed in Dual Numeric or Analog format.
Keysight N1911A/1912A User’s Guide 69
2 General Power Meter Functions
Meas Select
Offset setting field
Offset value field
Selected Window/
Measurement

Setting display offsets

This gain or loss is applied to the measured power after any channel offsets or mathematical functions have been included.
Offsets are entered in dB and the range of values is –100 dB to +100 dB. A positive value compensates for a loss, and a negative value compensates for a gain.

Procedure

Enter a display offset on the currently selected window:
1 Press to display the Measurement Setup screen.
2 Select the window you wish to set the offset value on by pressing the
key. The currently selected window/measurement is displayed.
3 Use the and keys to highlight the Offset setting field.
70 Keysight N1911A/1912A User’s Guide
4 Press to check the Offset setting field.
Figure 2-12 Typical display offset display
5 Press to highlight the Offset value field and press to display the
dB
NOTE
Display Offset pop-up. Use the numeric keypad to enter the required value in the Offset pop-up window.
6 Confirm your choice by pressing . Press key to complete the
offset entry.
The Ofs indicator is displayed if Display Offset is selected.

Setting frequency dependent offsets

Frequency dependent offset tables provide a quick and convenient method of compensating for frequency related changes in the response of your test system. Note that when selected, frequency dependent offset corrections are applied IN ADDITION to any correction for sensor frequency response.
The power meter is capable of storing 10 frequency dependent offset tables with a maximum of 80 frequency points each.
General Power Meter Functions 2
To use frequency dependent offset tables:
1 Select the table to be applied to a channel. Refer to “Setting frequency
dependent offsets” on page 71 for further information. If you require to edit the
table refer to “Editing frequency dependent offset tables” on page 74 for further information.
2 If using an 8480 Series, N8480 Series or an E-Series sensor, zero and calibrate
the power meter. The reference calibration factor used during the calibration is automatically set by the power meter from the sensor calibration table (if selected).
3 Specify the frequency of the signal you want to measure. The calibration
factor/offset is automatically set by the power meter from the sensor calibration table (if selected) and the frequency dependent offset table. Refer to “Procedure” on page 72 for further information.
4 Make the measurement.
Keysight N1911A/1912A User’s Guide 71
2 General Power Meter Functions
Tab les
Freq. Dep. Offset
NOTE
Offsets
Tab les
Freq. Dep. Offset
Offsets
Selecting a frequency dependent offset table
You can select a frequency dependent offset table from the key menu followed by and .
The State column indicates if any frequency dependent offset tables are currently selected. The Offset Tables screen is shown in Figure 2-13.
You can also view which FDO table is being used by pressing the ,
and use the and keys to highlight the FDO Table setting field and
press to display the table.
Procedure
Select an offset table as follows:
1 Press, either:
72 Keysight N1911A/1912A User’s Guide
a , , .
b , and use the and keys to highlight the FDO Table
setting field and press to display the table.
General Power Meter Functions 2
Tab le
On
NOTE
Tab le
Done
Done
Offset Table A Selected
Figure 2-13 Frequency dependent offset tables display
2 Use the and keys to highlight one of the 10 table titles and press
to highlight .
When no data is contained in the highlighted table, the key is disabled (grayed out).
Keysight N1911A/1912A User’s Guide 73
3 Press to complete the selection of the offset table.
4 Press again to display the measurement screen. Figure 2-14 shows
which offset table is selected.
Figure 2-14 Frequency dependent offset indicator
2 General Power Meter Functions
NOTE
Tab les
Freq. Dep. Offset
5 To change the frequency, press and use the and keys to
highlight the Frequency field.
6 Press to display the Frequency pop-up window. Use the numeric
keypad to enter the required value in the Frequency pop-up window.
7 To confirm your choice, press the appropriate unit softkey.
8 Connect the power sensor to the signal to be measured.
9 The measurement result, including offset, is now displayed.
If the measurement frequency does not correspond directly to a frequency in the sensor calibration table (if selected) and the frequency dependent offset table being used, the power meter calculates the calibration factor and offset using linear interpolation.
If you enter a frequency outside the frequency range defined in the sensor calibration table or the frequency dependent offset table, the power meter uses the highest or lowest frequency point in the appropriate table to set the calibration factor and offset.
74 Keysight N1911A/1912A User’s Guide

Editing frequency dependent offset tables

There are ten frequency dependent offset tables named CUSTOM_A through CUSTOM_J. They are empty of any data when the power meter is shipped from the factory.
You cannot delete any of the 10 existing frequency dependent offset tables or create any additional tables. However, you can enter values into the 10 existing tables. Each frequency dependent offset table can contain a maximum of 80 frequency points.
To view the frequency dependent offset tables currently stored in the power
meter, press , , . The Frequency Dependent Offset Tables screen is displayed as shown in Figure 2-13.
General Power Meter Functions 2
NOTE
Tab les
Freq. Dep. Offset
Edit Table
Editing frequency dependent offset tables requires the following steps:
1 Identify and select the table you want to edit
2 Rename the table
3 Enter the frequency and offset pairs
4 Save the table
Procedure
First select the table you want to edit as follows:
A frequency in the range of 0.001 MHz to 1000.0 GHz can be entered. A calibration factor in the range of 1% to 150% can be entered.
1 Press , , to display the Offset Tbls screen.
2 Choose the table you want to edit using the and keys. Press
to display the Ed it Offset screen as shown in Figure 2-15.
Keysight N1911A/1912A User’s Guide 75
Figure 2-15 “Edit Offset” display with data added
2 General Power Meter Functions
Change
Insert Char
Delete Char
Enter
NOTE
3 Highlight the table title using the and keys. Press and use the
, , and keys to select and change the characters in the Tab le
Name pop-up to create the name you want to use.
Figure 2-16 Edit table title pop-up
Pressing adds a new character to the right of the selected character.
Pressing removes the selected character.
4 Press to complete the entry.
The following rules apply to naming sensor calibration tables:
– The name must consist of no more than 12 characters.
– All characters must be upper or lower case alphabetic characters, or numeric
(0-9), or an underscore (_).
– No other characters are allowed.
– No spaces are allowed in the name.
76 Keysight N1911A/1912A User’s Guide
General Power Meter Functions 2
Insert
Change
GHz
MHz
Change
%
Done
NOTE
Enter (or edit) the frequency and offset pairs as follows:
1 Press to add a new frequency value (or press to edit). Use the
numeric keypad to enter the required value in the Frequency pop-up window. Complete the entry by pressing the , keys.
2 Enter the new offset value (or press to edit). Use the numeric keypad
to enter the required value in the Offset pop-up window. Complete the entry by pressing the key.
3 Continue adding/editing values until you have entered all the data you require.
4 When you have finished editing the table press to save the table.
If you measure a signal with a frequency outside the frequency range defined in the frequency dependent offset table, the power meter uses the highest or lowest frequency point in the frequency dependent offset table to calculate the offset.
Keysight N1911A/1912A User’s Guide 77
2 General Power Meter Functions
Minimum sensor power
Resolution setting
Maximum sensor power
Power sensor dynamic range
Number of averages
10 dB
10 dB
10 dB
10 dB
1
1
1
8
2
2
1
1
1
1
1
1
3
4
16
16
128
128
256
32
8
1
1
1

Setting Measurement Averaging

The power meter uses a digital filter to average power readings. The number of readings averaged can range from 1 to 1024. This filter is used to reduce noise, obtain the desired resolution and to reduce the jitter in the measurement results. Increasing the value of the measurement average reduces measurement noise. However, the measurement time is increased. You can manually select the measurement average or you can set the power meter to auto measurement average mode. The default is AUTO.
When the auto measurement average mode is enabled, the power meter automatically sets the number of readings averaged together to satisfy the filtering requirements for most power measurements. The number of readings averaged together depends on the resolution setting and the power level currently being measured.
Figure 2-17 Typical averaged readings
Figure 2-17 shows the typical number of averages for each range and resolution
when the power meter is in auto filter mode and is set to normal speed mode. P-Series power meters recognize different sensor types when they are connected, and configure suitable averaging automatically.
78 Keysight N1911A/1912A User’s Guide
Resolution is a measurement display function and not a channel function. In the case where a channel is set up in both the upper and lower window and the resolution settings are different, the highest resolution setting is taken to calculate the averaging number.
These four resolution levels represent:

Procedure

Set measurement averaging as follows:
1 Press . On dual channel meters select the required channel. The
2 Use the and keys to select the Filter setting field.
General Power Meter Functions 2
– 1, 0.1, 0.01, 0.001 dB respectively if the measurement suffix is dBm or dB.
– 1, 2, 3 or 4 significant digits respectively if the measurement suffix is
Wor%.
current setting is shown in the Meas Avg field (
Channel Setup screen. The default is
AUTO.
AUTO, MAN, or OFF) on the
3 Press and use the and to step through the available settings.
If you have selected AUTO or OFF proceed at step 7. If you have selected MAN proceed as follows:
4 Use the key to select the Meas Avg: value field.
5 Press to display the Meas Avg Count pop-up.
Figure 2-18 Meas Avg Count pop-up
Keysight N1911A/1912A User’s Guide 79
2 General Power Meter Functions
Enter
6 Use the numeric keys to enter the required value and press .
7 Press key to close the Channel Setup screen.
80 Keysight N1911A/1912A User’s Guide

Step Detection

To reduce the filter settling time after a significant step in the measured power the filter can be set to re-initialize upon detection of a step increase or decrease in the measured power. Step detection can be set in both manual and automatic measurement average modes.

Procedure

Set step detection as follows:
1 Press . On dual channel meters select the required channel.
2 Use the and keys to select the Step Detect setting field.
3 Press to check the step detection to on or off.
4 Press key to close the Channel Setup screen.
General Power Meter Functions 2
Keysight N1911A/1912A User’s Guide 81
2 General Power Meter Functions
NOTE

Setting the Video Averaging

This feature is only available when a P-Series or an E-Series E9320 power sensor is connected.
Video averaging uses a digital filter to average repetitions of a triggered signal. The number of acquisitions averaged can range from 1 to 256, in multiples of 2
With video averaging the average of a number of acquisitions is calculated to smooth the displayed trace and reduce apparent noise. The measurement requires a continuously repeating signal. Increasing the value of this filter reduces noise but increases the time required to make the measurement.

Procedure

Set the video average as follows:
1 Press . On dual channel meters select the required channel.
n
.
2 Use the and keys to select the Video Avg setting field.
3 Press to check the video averaging to on or off.
4 Use the key to select the Video Avg: value field.
5 Press to display the Video Avg Count pop-up.
82 Keysight N1911A/1912A User’s Guide
General Power Meter Functions 2
Figure 2-19 Video Avg Count pop-up
6 Use the and keys to highlight the required Video Avg Count value and
press .
7 Press key to close the Channel Setup screen.
Keysight N1911A/1912A User’s Guide 83
2 General Power Meter Functions
NOTE

Setting the Video Bandwidth

This feature is only available when a P-Series or an E-Series E9320 power sensor is connected.
Selecting a bandwidth value close to or slightly greater than required by the modulating signal bandwidth can help reduce noise and improve accuracy on peak measurements (see Table 2-3). It can however, reduce the processing speed for long acquisition times.
Tab le 2-3 P-Series Sensor Video Bandwid th Setting
Low: Medium: High: Off
5.0 MHz 15.0 MHz 30.0 MHz >30.0 MHz
The Low, Medium, and High pass band shapes achieved by the video bandwidth settings provide flat filter responses with very sharp cut-off points by applying digital signal processing techniques to ensure accurate power measurement within the specified band.
84 Keysight N1911A/1912A User’s Guide
General Power Meter Functions 2
Amplitude P-Series sensors
Frequency
Approx
-3.5 dB
0
5 MHz 15 MHz
30 MHz
Low
Med
High
Off
Meter B/W setting
Figure 2-20 Bandwidth filter shapes
When video bandwidth is set to Off, it removes all digital signal conditioning. This provides approximately 3 dB roll off at the maximum sensor’s bandwidth and is best suited for accurate trace capture, minimizing overshoot, and removing any ringing effects caused by the sharp cut-off filters used in the Low, Med and High settings. Figure 2-20 shows the filter shapes associated with the P-Series

Procedure

wideband power sensors.
Set the video bandwidth as follows:
1 Press . On dual channel meters select the required channel.
2 Use the and keys to select the Video B/W setting field.
Keysight N1911A/1912A User’s Guide 85
2 General Power Meter Functions
3 Press to display the Video B/W pop-up.
Figure 2-21 Video bandwidth pop-up
4 Use the and keys to highlight the required Video Bandwid th setting
and press .
5 Press key to close the Channel Setup screen.
86 Keysight N1911A/1912A User’s Guide

Setting Measurement Channel Gates

NOTE
Gate 1
Gate 4
Gate 3
Gate 2
Trigger
This feature is only available when a P-Series or an E-Series E9320 power sensor is connected.
A system of gates, controlled by and referenced to a trigger point, is used to obtain measurement data from a captured trace. The trace data within each gate period is subsequently used for the individual measurement calculations. Up to 4 gates can be set up for each channel. Figure 2-22 shows an example of 4 gates setup to perform the following measurements simultaneously:
– Average power level of the pulse:
Gate 1, average measurement
– Average “off” power level ahead of the pulse:
Gate 2, average measurement
– Peak to average ratio:
Gate 1, peak-to-average measurement
General Power Meter Functions 2
–Pulse drop:
Gate 3, average measurement, minus Gate 4, average measurement
Keysight N1911A/1912A User’s Guide 87
Figure 2-22 Example of measurement gates
2 General Power Meter Functions
NOTE
NOTE
Gates Setup
NOTE
Figure 2-22 shows measurement results using this example of the gates
placement.

Procedure

This procedure uses the Gates Setup under the Channel Setup menu. Alternatively, you can use a more visual method, when in the Graphical Trace Mode (Gate Control Menu), to set the channel’s Gate Control. (See “Setting the
Trace Display” on page 152).
– Press . The Channel Gate Setup screen is displayed.
Figure 2-23 Gate Setup screen
1 Highlight the Gate Start you want to configure using the , , and
keys.
The gate start time is relative to the trigger event. Positive values set a measurement gate, to a maximum time of 1 second, after the trigger. Negative time gate start values set a measurement gate, to a maximum time of 1 second, before the trigger.
88 Keysight N1911A/1912A User’s Guide
General Power Meter Functions 2
smsμs
ns
smsμs
ns
2 Press and use the numeric keypad to enter the required value in the
Time Gating Start pop-up window.
Figure 2-24 Time Gating Start pop-up
3 Complete the entry by pressing the required second, millisecond, microsecond
or nanosecond ( , , or ) softkey.
4 Highlight the Gate Length you want to configure using the , , , and
keys.
Figure 2-25 Time Gating Length pop-up
5 Complete the entry by pressing the required second, millisecond, or
microsecond or nanosecond ( , , or ) softkey.
6 Repeat this process until you have setup all the required gates.
Keysight N1911A/1912A User’s Guide 89
2 General Power Meter Functions

Auto Gate

Auto Gate Settings
NOTE
Gate 1 Ref 1 0.0%
Auto Gate
This feature will automatically set the gates (start and stop times of the user defined time interval) for each measurement relative to the rising and falling edges’ trigger point.
– Proceed to the second page of Gate Ctrl menu and press softkey.
–To change the Auto Gate setup:
1 Press softkey.
90 Keysight N1911A/1912A User’s Guide
Figure 2-26 Auto Gate Settings display
Auto Gate Settings enables the placing of markers (gates) at different positions.
2 Press softkey. The Marker 1 Ref Percentage pop-up is
displayed.
Figure 2-27 Marker 1 Ref Percentage pop-up
General Power Meter Functions 2
%
Gate 1 Ref 2 0.0%
%
Ref 1 x PD
Ref 2 x PD
Pulse Duration (PD)
NOTE
Perpetual Off On
3 Use the numeric keys to complete the entry and press softkey.
4 Press softkey. The Marker 2 Ref Percentage pop-up is
displayed.
Figure 2-28 Marker 2 Ref Percentage pop-up
5 Use the numeric keys to complete the entry and press softkey.
Keysight N1911A/1912A User’s Guide 91
Figure 2-29 The placement of gates (markers) in Auto Gating mode
Enabling the softkey allows the gates to be repositioned automatically according to the changes of pulse duration.
2 General Power Meter Functions
NOTE
NOTE
NOTE
Trace Setup

Setting up the Channel Trace

This feature is only available when a P-Series or an E-Series E9320 power sensor is connected.
The Trace Setup is the only location where you can change the Y-axis units from dBm to Watts.

Procedure

This procedure uses the Trace Setup under the Channel Setup menu. Alternatively, you can use a more visual method, when in the Graphical Trace Mode (Trace Control Menu), to set the channel’s Trace Control. (See “Setting the
Trace Display” on page 152).
1 Press , to display the Trace Setup menu.
Figure 2-30 Trace setup display
92 Keysight N1911A/1912A User’s Guide
General Power Meter Functions 2
smsμs
ns
smsμs
ns
2 Use the and keys to highlight the X Start field.
3 Press to display the Trace Start Time pop-up and use the numeric
keypad to enter the required value in the Trace Start Time pop-up window.
Figure 2-31 Trace Start Time pop-up
4 Complete the entry by pressing the required second, millisecond, microsecond
or nanosecond ( , , or ) softkey.
5 Use the and keys to highlight the X Scale field.
6 Press to display the X Scale/Division pop-up.
a Use the numeric keypad to enter the required value in the X Scale/Division
pop-up window.
b Use the and keys to increase or decrease the value.
Figure 2-32 X Scale/Division pop-up
7 Complete the entry by pressing:
a The required second, millisecond, microsecond or nanosecond ( , ,
or ) softkey.
Keysight N1911A/1912A User’s Guide 93
2 General Power Meter Functions
Enter
dBm
b The softkey.
8 Use the and keys to highlight the Units field.
If you want to view the trace in linear scale, use this field. Otherwise the default scale is in logarithmic.
9 Press to display the Units pop-up, and use the and to highlight
Watt or dBm.
Figure 2-33 Trace Units pop-up
10 Press to complete the entry.
11 Use the and keys to highlight the Y Max field.
12 Press to display the Trace Maximum pop-up, and use the numeric
keypad to enter the required value in the Trace Maximum pop-up window.
Figure 2-34 Y scale maximum value pop-up
13 Complete the entry by pressing the or the linear value.
14 Use the and keys to highlight the Y Scale field.
94 Keysight N1911A/1912A User’s Guide
General Power Meter Functions 2
dB
Trace Control

Auto Scale

NOTE
15 Press to display the Y Scale/Division pop-up.
a Use the numeric keypad to enter the required value in the Y Scale/Division
pop-up window.
b Use the and keys to increase or decrease the value.
Figure 2-35 Y Scale/Division pop-up
16 Complete the entry by pressing the or the linear value.
17 Press key to complete the setup and display the measurements
results.
Auto Scale
Auto Scale automatically configures the window’s X-axis and Y-axis to best display the input trace.
1 Press softkey under Gate Ctrl menu.
2 softkey is pressed to accomplish auto scaling.
This feature will work reliably with well behaved pulses. Modulated pulses may cause the meter to auto-scale on the details within the pulse envelope.
Keysight N1911A/1912A User’s Guide 95
2 General Power Meter Functions
NOTE
Acqn
Trigger status
NOTE

Setting the Trigger

This feature is only available when a P-Series power sensor is connected.
The trigger can be taken from a rising or falling measured power level or controlled externally using the Ext Trig input. Additional control features such as; hold-off, hysteresis, and delay are provided to help you achieve a stable and reliable trigger. To use the measurement gates, the power meter must be triggered.
Press . The Trigger menu is displayed.
The trigger status is displayed below the label in the Trigger menu.
Figure 2-36 shows the power meter display in Free Run mode. In this mode the
meter is not synchronized to any modulated RF signal at the sensor input. Consequently, power levels within the configured time gates are random and the displayed measurement results are not valid.
Figure 2-36 Trigger menu - free run mode
The symbol in the upper window indicates the power meter is in free run mode.
96 Keysight N1911A/1912A User’s Guide

Procedure

Acqn
Sing Trig
Cont Trig
Sing Trig
Cont Trig
Settings
Rising edge source
Channel A is trigger source
1 Press softkey to configure a trigger.
2 Select either or .
3 Press to configure the remaining trigger parameters. The trigger
General Power Meter Functions 2
is a single shot mode. After triggering, the measurement is
halted, the symbol is displayed. You can start another measurement by
pressing the key.
is a continuos trigger mode. The symbol or is displayed.
Settings menu have two pages. Figure 2-37 shows page one and Figure 2-40 shows page two.
Keysight N1911A/1912A User’s Guide 97
Figure 2-37 Trigger Setting Menu 1 of 2
To help you quickly check the triggering setup, the current settings for all the trigger parameters are shown with their respective softkey labels.
2 General Power Meter Functions
Source
Source
Ext
Int
Ext
NOTE
Source
Ext
Ch.A
Ch.B
Mode
Source Int
Mode
Norm
AutoLvl
Norm
AutoLvl
Level
Norm
Level
Level
Source
The current setting is shown below the label.
To change the setting press and choose or .
The trigger source is shown beside the trigger symbol when a measurement window is configured in single numeric mode. When an external trigger ( ) is
selected, the power meter can be triggered via the Ext Trig input.
For N1912A, by pressing , you can choose your source through the three different softkeys: , , and .
Mode
The key is only available when trigger is selected.
To change the setting press and choose or .
The current setting is displayed below the label. When is selected you can choose the RF power level transition used as the trigger. The power meter
automatically finds a triggering power level transition when is selected.
Level
is only available when trigger mode has been selected. The current
value is shown below the label. The minimum power level you can enter is limited to 40 dB below the maximum sensor power. To change the setting press
and use the numeric keypad to enter the required value in the Trigger Level
pop-up window to enter the new value.
98 Keysight N1911A/1912A User’s Guide
General Power Meter Functions 2
dBm
Delay
Delay
smsμs
ns
NOTE
1 of 2
Figure 2-38 Trigger Level pop-up
Complete the entry by pressing .
Delay
The current setting is displayed below the label. The delay time is applied between the trigger event and all the gate start times. This allows you to time-shift all the gates by the same amount with one setting change. A delay of
up to 1 second can be entered. To enter or change the setting press and use the numeric keypad to enter the required value in the Trigger Delay pop-up window to enter the new value.
Figure 2-39 Trigger Delay pop-up
Complete the entry by pressing the required second, millisecond, microsecond or nanosecond ( , , or ) softkey.
The trigger delay time can also be entered or changed in the Trace Display mode when the Gate Control menu is displayed. See “Setting the Trace Display” on page 152 for further detail.
Press to display the second menu page.
Keysight N1911A/1912A User’s Guide 99
2 General Power Meter Functions
Slope
+
Slope
+
Holdoff
Holdoff
Figure 2-40 Trigger setting menu 2 of 2
Slope
The current setting is displayed below the label and the or symbol is displayed in single numeric display mode. (and ) is used to generate the trigger from an increasing power level. Similarly, (and ) is used to generate the trigger from a decreasing power level. To change the setting, press to highlight or as required.
Holdoff
The current setting is displayed below the label. After a trigger event occurs, the trigger mechanism is disabled for the configured time period. This allows stable triggering to be achieved even when a signal has multiple edges, for example, a TDMA signal with non-constant amplitude modulation. Values up to 400 ms can be configured.
To change the setting press and use the numeric keypad to enter the required value in the Trigger Holdoff pop-up window to enter the new value.
100 Keysight N1911A/1912A User’s Guide
Loading...