duced in any form or by any means
(including electronic storage and
retrieval or translation into a foreign
language) without prior agreement and
written consent from Keysight Technologies as governed by United States and
international copyright laws.
Manual Part Number
N1912-90002
Edition
Edition 18, September 29, 2017
Printed in:
Printed in Malaysia
Published by:
Keysight Technologies
Bayan Lepas Free Industrial Zone,
11900 Penang, Malaysia
Technology Licenses
The hardware and/or software
described in this document are furnished under a license and may be
used or copied only in accordance with
the terms of such license.
Declaration of Conformity
Declarations of Conformity for this
product and for other Keysight products may be downloaded from the
Web.
U.S. Government Rights
The Software is “commercial computer
software,” as defined by Federal Acquisition Regulation (“FAR”) 2.101. Pursuant to FAR 12.212 and 27.405-3 and
Department of Defense FAR Supplement (“DFARS”) 227.7202, the U.S.
government acquires commercial computer software under the same terms
by which the software is customarily
provided to the public. Accordingly,
Keysight provides the Software to U.S.
government customers under its standard commercial license, which is
embodied in its End User License
Agreement (EULA). The license set
forth in the EULA represents the
exclusive authority by which the U.S.
government may use, modify,
distribute, or disclose the Soft-ware.
The EULA and the license set forth
therein, does not require or per-mit,
among other things, that Keysight: (1)
Furnish technical information related to
commercial computer
commercial computer software
documentation that is not customarily
provided to the public; or (2) Relinquish
to,
or otherwise provide, the government rights in excess of these rights
customarily provided to the public to
use, modify, reproduce, release, per-form,
display, or
computer software
com-puter software documentation.
No additional government
requirements beyond those set forth in
the EULA shall apply, except to the
extent
licenses are explicitly required from all
providers of commercial computer
software pursu-ant to the FAR and the
DFARS and are set forth specifically in
writing
Keysight shall be under no obligation to
update, revise or otherwise modify the
Software. With respect to any
technical data as defined by FAR
2.101, pursuant to FAR 12.211 and
27.404.2 and DFARS 227.7102, the
U.S. government acquires no greater
than Limited Rights as defined in FAR
27.401 or DFAR 227.7103-5 (c), as
applicable in any technical data.
disclose commercial
that those terms, rights, or
else-where in the EULA.
soft-ware or
or commercial
Warranty
THE MATERIAL CONTAINED IN THIS
DOCUMENT IS PROVIDED “AS IS,”
AND IS SUBJECT TO BEING
CHANGED, WITHOUT NOTICE, IN
FUTURE EDITIONS. FURTHER, TO THE
MAXIMUM EXTENT PERMITTED BY
APPLICABLE LAW, KEYSIGHT DISCLAIMS ALL WARRANTIES, EITHER
EXPRESS OR IMPLIED, WITH REGARD
TO THIS MANUAL AND ANY INFORMATION CONTAINED HEREIN, INCLUDING BUT NOT LIMITED TO THE
IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A
PARTICULAR PURPOSE. KEYSIGHT
SHALL NOT BE LIABLE FOR ERRORS
OR FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES IN CONNECTION
WITH THE FURNISHING, USE, OR
PERFORMANCE OF THIS DOCUMENT
OR OF ANY INFORMATION CONTAINED HEREIN. SHOULD KEYSIGHT
AND THE USER HAVE A SEPARATE
WRITTEN AGREEMENT WITH WARRANTY TERMS COVERING THE MATERIAL IN THIS DOCUMENT THAT
CONFLICT WITH THESE TERMS, THE
WARRANTY TERMS IN THE SEPARATE
AGREEMENT SHALL CONTROL.
Safety Information
A CAUTION notice denotes a hazard. It
calls attention to an operating procedure, practice, or the like that, if not
correctly performed or adhered to,
could result in damage to the product
or loss of important data. Do not proceed beyond a CAUTION notice until
the indicated conditions are fully
understood and met.
A WARNING notice denotes a hazard. It
calls attention to an operating procedure, practice, or the like that, if not
correctly performed or adhered to,
could result in personal injury or death.
Do not proceed beyond a WARNING
notice until the indicated conditions are
fully understood and met.
2Keysight N1911A/1912A User’s Guide
Certification
Keysight Technologies certifies that this product met its published specifications
at the time of shipment. Keysight further certifies that its calibration
measurements are traceable to the United States National Institute of Standard
and Technology (formerly National Bureau of Standards), to the extent allowed by
that organization’s calibration facility, and to the calibration facilities of other
International Standards Organization members.
General Warranty
The material contained in this document is provided “as is,” and is subject to
being changed, without notice, in future editions. Further, to the maximum extent
permitted by applicable law, Keysight disclaims all warranties, either express or
implied with regard to this manual and any information contained herein,
including but not limited to the implied warranties of merchantability and fitness
for a particular purpose. Keysight shall not be liable for errors or for incidental or
consequential damages in connection with the furnishing, use, or performance of
this document or any information contained herein. Should Keysight and the user
have a separate written agreement with warranty terms covering the material in
this document that conflict with these terms, the warranty terms in the separate
agreement shall control. Duration and conditions of warranty for this product may
be superseded when the product is integrated into (becomes a part of) other
Keysight products. During the warranty period, Keysight will, at its option, either
repair or replace products which prove to be defective. The warranty period begins
on the date of delivery or on the date of installation if installed by Keysight.
Warranty Service
For warranty service or repair, this product must be returned to a service facility
designated by Keysight. For products returned to Keysight for warranty service,
the Buyer shall prepay shipping charges to Keysight and Keysight shall pay
shipping charges to return the product to the Buyer. However, the Buyer shall pay
all shipping charges, duties, and taxes for products returned to Keysight from
another country.
Keysight N1911A/1912A User’s Guide3
Limitation of Warranty
The foregoing warranty shall not apply to defects resulting from improper or
inadequate maintenance by the Buyer, Buyer-supplied products or interfacing,
unauthorized modification or misuse, operation outside of the environmental
specifications for the product, or improper site preparation or maintenance.
The design and implementation of any circuit on this product is the sole
responsibility of the Buyer. Keysight does not warrant the Buyer’s circuitry or
malfunctions of Keysight products that result from the Buyer’s circuitry. In
addition, Keysight does not warrant any damage that occurs as a result of the
Buyer’s circuit or any defects that result from Buyer-supplied products.
To the extent allowed by local law, Keysight makes no other warranty, expressed
or implied, whether written or oral with respect to this product and specifically
disclaims any implied warranty or condition of merchantability, fitness for a
particular purpose or satisfactory quality.
Exclusive Remedies
To the extent allowed by local law, the remedies provided herein are the Buyer’s
sole and exclusive remedies. Keysight shall not be liable for any direct, indirect,
special, incidental, or consequential damages (including lost profit or data),
whether based on warranty, contract, tort, or any other legal theory.
4Keysight N1911A/1912A User’s Guide
Restricted Rights Legend
The Software and Documentation have been developed entirely at private
expense. They are delivered and licensed as “commercial computer software” as
defined in DFARS 252.227-7013 (Oct 1988), DFARS 252.211-7015 (May 1991), or
DFARS 252.227-7014 (Jun 1995), as a “commercial item” as defined in FAR
2.101(a), or as “restricted computer software” as defined in FAR 52.227-19 (Jun
1987) (or any equivalent agency regulation or contract clause), whichever is
applicable. You have only those rights provided for such Software and
Documentation by the applicable FAR or DFARS clause or the Keysight standard
software agreement for the product involved.
Technology Licenses
The hardware and/or software described in this document are furnished under a
license and may be used or copied only in accordance with the terms of such
license.
Keysight N1911A/1912A User’s Guide5
Safety Symbols
The following symbols on the instrument and in the documentation indicate
precautions which must be taken to maintain safe operation of the instrument.
Direct current (DC).Alternating current (AC).
Off (mains supply).On (mains supply).
Caution, risk of electric shock.
Earth (ground) terminal.Frame or chassis (ground) terminal.
Protective earth (ground) terminal.
Caution, risk of danger (refer to this
manual for specific Warning or Caution
information).
Equipment protected throughout by
double insulation or reinforced
insulation.
Both direct and alternating current.
Caution, hot surface.In position of a bi-stable push control.
Three-phase alternating current.Equipotentiality.
This symbol indicates the operating
switch for ‘Stand-by’ mode. Note, the
instrument is NOT isolated from the
mains when the switch is pressed. To
isolate the instrument, the mains
coupler (mains input cord) should be
removed from the power supply.
Out position of a bi-stable push
control.
6Keysight N1911A/1912A User’s Guide
Safety Considerations
Read the information below before using this instrument.
The following general safety precautions must be observed during all phases of
operation, service, and repair of this instrument. Failure to comply with these
precautions or with specific warnings elsewhere in this manual violates safety
standards for design, manufacture, and intended use of the instrument. Keysight
Technologies assumes no liability for the customer’s failure to comply with these
requirements.
Recommended Calibration Interval
Keysight Technologies recommends a two- years calibration cycle for the
N1911A/1912A P-Series power meter.
Keysight N1911A/1912A User’s Guide7
General Safety Information
WARNING
This is a Safety Class I instrument (provided with a protective earthing ground,
incorporated in the power cord). The mains plug shall only be inserted in a socket
outlet provided with a protective earth contact. Any interruption of the protective
conductor inside or outside of the instrument is likely to make the instrument
dangerous. Intentional interruption is prohibited.
– DO NOT operate the product in an explosive atmosphere or in the
presence of flammable gases or fumes.
– DO NOT use repaired fuses or short-circuited fuseholders: For continued
protection against fire, replace the line fuse(s) only with fuse(s) of the
same voltage and current rating and type.
– DO NOT perform procedures involving cover or shield removal unless you
are qualified to do so: Operating personnel must not remove equipment
covers or shields. Procedures involving the removal of covers and shields
are for use by service-trained personnel only.
– DO NOT service or adjust alone: Under certain conditions, dangerous
voltages may exist even with the equipment switched off. To avoid
dangerous electrical shock, service personnel must not attempt internal
service or adjustment unless another person, capable of rendering first
aid and resuscitation, is present.
– DO NOT operate damaged equipment: Whenever it is possible that the
safety protection features built into this product have been impaired,
either through physical damage, excessive moisture, or any other reason,
REMOVE POWER and do not use the product until safe operation can be
verified by service-trained personnel. If necessary, return the product to a
Keysight Technologies Sales and Service Office for service and repair to
ensure the safety features are maintained.
– DO NOT substitute parts or modify equipment: Because of the danger of
introducing add itional hazards, do not install substitute parts or perform
any unauthorized modification to the product. Return the product to a
Keysight Technologies Sales and Service Office for service and repair to
ensure the safety features are maintained.
8Keysight N1911A/1912A User’s Guide
Environmental Conditions
The N1911A/1912A is designed for indoor use and in an area with low
condensation. The table below shows the general environmental requirements for
this instrument.
Environmental conditionRequirement
Temperature
Humidity
AltitudeUp to 3000 m
Pollution degree2
Operating condition
– 0 °C to 55 °C
Storage condition
– –40 °C to 70 °C
Operating condition
– Up to 95% RH at 40 °C (non-condensing)
Storage condition
– Up to 90% RH at 65 °C (non-condensing)
Keysight N1911A/1912A User’s Guide9
Regulatory Information
The N1911A/1912A complies with the following safety and Electromagnetic
Compatibility (EMC) compliances:
Safety compliance
– IEC 61010-1:2010/EN 61010-1:2010 (3rd Edition)
– Canada: CAN/CSA-C22.2 No. 61010-1-12
– USA: ANSI/UL 61010-1:2012
EMC compliance
– IEC 61326-1:2005/EN 61326-1:2006
– CISPR11:2003/EN 55011:1998+A1:1999+A2:2002, Group 1 Class A
– Canada: ICES/NMB-001:Issue 4, June 2006
– Australia/New Zealand: AS/NZS CISPR 11:2004
10Keysight N1911A/1912A User’s Guide
Regulatory Markings
The CE mark is a registered trademark
of the European Community. This CE
mark shows that the product complies
with all the relevant European Legal
Directives.
ICES/NMB-001 indicates that this ISM
device complies with the
Canadian ICES-001.
Cet appareil ISM est conforme a la
norme NMB-001 du Canada.
ISM GRP.1 Class A indicates that this
is an Industrial Scientific and Medical
Group 1 Class A product.
This symbol is a South Korean Class A
EMC Declaration. This is a Class A
instrument suitable for professional
use and in electromagnetic
environment outside of the home.
This symbol indicates the time period
during which no hazardous or toxic
substance elements are expected to
leak or deteriorate during normal use.
Forty years is the expected useful life
of the product.
The CSA mark is a registered
trademark of the Canadian
Standards Association.
The RCM mark is a registered
trademark of the Australian
Communications and Media Authority.
This instrument complies with the
WEEE Directive (2002/96/EC) marking
requirement. This affixed product label
indicates that you must not discard
this electrical or electronic product in
domestic household waste.
Keysight N1911A/1912A User’s Guide11
Waste Electrical and Electronic Equipment (WEEE) Directive 2002/
96/EC
This instrument complies with the WEEE Directive (2002/96/EC) marking
requirement. This affixed product label indicates that you must not discard this
electrical or electronic product in domestic household waste.
Product category:
With reference to the equipment types in the WEEE directive Annex 1, this
instrument is classified as a “Monitoring and Control Instrument” product.
Keysight N1911A/1912A P-Series Power Meters
User’s Guide
1Introduction
LXI Class-C Compliant Power Meter 30
Rack Mounting 30
Power Meter and Sensor Capability 31
Conventions Used in this Guide 32
Front Panel Keys and Connections 33
The Display Layout 38
Window Symbols and Pop-ups 46
Using the Instrument Web Browser 50
This chapter introduces you to the front panel display and instrument Web
browser of the P-Series power meter.
29
1Introduction
LXI Class-C Compliant Power Meter
P-Series power meter is a LXI Class C compliant
instrument, developed using LXI Technology. LXI, an
acronym for LAN eXtension for Instrumentation, is an
instrument standard for devices that use the Ethernet
(LAN) as their primary communication interface.
Hence, it is an easy-to-use instrument especially with the
usage of an integrated Web browser that provides a convenient way to configure
the instrument’s functionality.
Rack Mounting
The N1911A/1912A can be mounted in a standard 19-inch rack. Rack mount kits
are available as listed below. Support rails are also required for rack mounting.
These are normally supplied with the rack and are not included with the rack
mount options.
If you are installing an instrument on top of the N1911A/1912A, ensure that the
instrument does not obstruct the ventilation holes at the top of the N1911A/
1912A. If required, use a filler panel above the N1911A/1912A to ensure adequate
space for air circulation.
OptionDescription
N1911A Option 908Rack mount kit for one instrument
N1911A Option 909Rack mount kit for two instruments
N1912A Option 908Rack mount kit for one instrument
N1912A Option 909Rack mount kit for two instruments
30Keysight N1911A/1912A User’s Guide
Power Meter and Sensor Capability
NOTE
Your P-Series power meter is compatible with Keysight P-Series, E-Series E9320,
E-Series E9300, E-Series E4410, 8480 Series and the N8480 Series power
sensors. However, not all sensor and meter combinations have the same features
or capabilities. The main differences are as below:
Introduction1
Features
Average power of CW signal******
Average power of modulated signal*****
Peak power**
Cal factors stored on EEPROM****
>200 readings/sec****
Peak/burst average power**
Time gated measurements**
Rising edge trigger**
Falling edge trigger**
[a] Not applicable for N8480 Series power sensors with Option CFT
P-Series
N1920
E-Series
E9320
E-Series
E9300
E-Series
E4410
8480 Series
– The E-Series, 8480 Series and N8480 Series power sensors require N1917A/
B/C/D cables when connected to the P-Series power meters.
– The N8480 Series power sensors used in this user’s guide is referring to all
sensors from the N8480 Series unless otherwise stated.
N8480
Series
[a]
*
Specifications
The specifications for the power meter are listed in Chapter 11, "Specifications
and Characteristics".
Keysight N1911A/1912A User’s Guide31
1Introduction
Softkey
Channel
Channel A
Channel B
Softkey
Conventions Used in this Guide
The following conventions are used throughout this guide.
This symbol and text represents a labeled key on the power meter front panel.
This symbol and text represents a labeled softkey and is used to indicate you should
press the unmarked key beside the displayed text.
MessageThis text represents a displayed message.
ParameterThis is used to represent a parameter, value, or title.
This User’s Guide describes the operation for both the single channel and the dual
channel power meter. To identify channels on a dual channel meter a
“Channel”
softkey on an N1911A meter becomes and an
N1912A.
When you are asked to press “the channel” in a procedure, make sure
you select the relevant channel.
32Keysight N1911A/1912A User’s Guide
Front Panel Keys and Connections
This section briefly describes the functions of the front panel keys and connectors.
These keys are located to the left of the display.
KeyFunction
Press this key to preset the power meter. An option list of pre-installed
measurement configurations is presented. Select an option and press the confirm
key.
Introduction1
Press this key to control the powermeter from the front panel when it is operating
via the remote interfaces (when Local Lock Out is not enabled).
Press this key to select the upper or lower measurement window. The selected
window is highlighted by a blue line on the right side of the window. Any
measurement setup you create is performed in the selected window.
Press this key to choose windowed, expanded, or full-screen display of a numeric
measurement. It also provides quick access to the Gate Control screen and menu
when a trace window is selected.
Press this key to switch the meter between on and standby. When power is
supplied, the background LED is red. Pressing the key, switches the power meter on
and the background LED is green.
Keysight N1911A/1912A User’s Guide33
1Introduction
These keys are located along the lower edge of the display.
KeyFunction
Press this key to access general configuration-menus, such as GPIB address. You
can also access the Self-Test and Table entry menus. The measurement screen
remains visible.
Press this key to access the channel configuration menus. Channel parameters such
as averaging and offsets are configured from this menu.
Press this key to access the triggering menu. Unless a P-Series or an E-Series
E9320A Sensor is connected, all the menu keys are disabled (grayed out).
Press this key to setup relative measurements or set display offsets.
Use this key to configure the selected measurement.
Press this key to access the measurement display menu. You can choose the
displayed measurement resolution, units and display format.
Use this key together with to configure measurement displays.
34Keysight N1911A/1912A User’s Guide
Introduction1
These keys are all associated with the menu labels and data entry. They are
located to the right of the display.
KeyFunction
Press this key to return to the previous screen. This key also cancels pop-up entry.
These unmarked keys are called ‘softkeys’ and are referred to by the text on the
display next to them.
For example, during a Preset, you are given an option to confirm the command.
Press to continue, that is, press the softkey beside the displayed word ‘confirm’.
The lowest of the unmarked softkeys is used when there is a two page menu to be
displayed. For example, a 1 of 2 is displayed beside the key indicating the first page
of a two page menu. Press the key to access the next page or second page. (A 2 of 2
is displayed).
Keysight N1911A/1912A User’s Guide35
1Introduction
These keys and connectors are associated with the measurement channels and
are located on the right-hand side of the front panel.
KeyFunction
The arrow keys are used for navigation around the parameter entry screens.
The up and down arrows are used for selecting values from a pop-up list, they
can also be used for stepping two values, X Scale and Y Scale. They are also
used to enter text, for example, table names.
Press this key to select a highlighted field to allow data entry, check a
checkbox and terminate entry of a popup list.
Press this key to access the zero and calibration menus.
36Keysight N1911A/1912A User’s Guide
Press this key to start and stop the acquisition of the measurement.
Press these keys to enter numeric values in the pop-up fields, for example, the
offset values. To complete the entry use the softkey.
ConnectorFunction
The power reference is a 1 mW (0 dBm) 50 MHz signal available
from a 50 W type-N connector. It is used for calibrating an 8480,
E-Series or N8480 Series power sensor and meter system. If the
meter is configured with Option 003, the connector is fitted to the
rear panel. The Green LED beside the connector is lit when the
calibrator is turned on.
Introduction1
The sensor input connectors (N1912A shown, the N1911A has one
input). If the meter is configured with Option 003, the connectors
are fitted to the rear panel.
Keysight N1911A/1912A User’s Guide37
1Introduction
Disp Type
1
3
4
6
9
2
8
5
7
The Display Layout
Figure 1-1 shows the display layout when two windows are configured in dual
numeric mode.
Pressing , the other display formats are available.
Figure 1-1Dual numeric display
1 The status reporting line displays messages and the control status of the
power meter. For example, the status can be either RMT (remote, GPIB, USB
or LAN operation) or LCL (local, front panel operation).
The message fields indicate ERR for any error conditions that occur or
informing you to Please Zero the power sensor.
2 The blue highlight on the right hand side of the window shows it is the
currently selected measurement display line. This measurement line is the
Upper Window/Upper Measurement.
3 The measured channel is shown. With a P-Series or an E-Series E9320 power
sensor connected, and channel in trigger mode, the associated gate number is
shown.
4 The associated measurement type is shown below the channel and gate
number.
5 This displays the measurement units, either dBm, dB, Watts, or percent (%).
38Keysight N1911A/1912A User’s Guide
Introduction1
NOTE
With a P-Series or an E-Series E9320 power sensor connected, a measurement
result of –270 dBm indicates the input power level is beyond the sensitivity of
the sensor.
6 With a P-Series or an E-Series E9320 power sensor connected, you can make
combined measurements with a single channel meter. A dual channel meter
extends this feature across both channels.
7 This displays the number of pages in the current menu. For example, 1 of 2
indicates that there are two pages in the menu and the first page is currently
displayed. Pressing the softkey displays the next page, indicated by 2 of 2.
(Press the softkey to display the previous menu page.)
8 The available softkey labels are displayed in these four fields. Additionally,
settings associated with the labeled function are displayed under the label.
9 This field displays the menu title. For example, Channel Setup or press
and the Zero/Cal menu is displayed.
Keysight N1911A/1912A User’s Guide39
1Introduction
14
13
11
12
15
10
Figure 1-2Single numeric and analog display
10 Figure 1-2 shows the default display mode of two measurement windows. The
upper measurement window has a blue highlight on the right hand side of the
window showing it has been selected.
Using the , , or keys you can change the selection of the
measurement window.
Using the key on numeric measurement results window you can
choose either two rectangular windows, a single enlarged window, or a full
screen display by pressing. The display style is applied to the currently
selected window or measurement line.
11 The upper window is configured to show a single numeric display.
12 The lower window is configured to show an analog meter which displays the
measurement result and the meter scaling.
13 With a P-Series or an E-Series E9320 power sensor connected, the symbol
shows the trigger state (Free Run), (Stopped), (Negative Slope), or
(Positive Slope).
14 The channel measurement frequency.
15 This displays the connected sensor, the offset value, and the acquisition mode
on the channel. On dual channel models it shows for both channels.
40Keysight N1911A/1912A User’s Guide
Introduction1
21
20 19
18
16
17
NOTE
Figure 1-3Full screen numeric display
16 Figure 1-3 shows a single numeric full screen displaying a relative result.
This field displays Rel if relative mode is on.
17 This field indicates the measurement result is beyond the configured upper or
lower limit. If the measurement is within the limits this field is empty. If the
measurement result is less than the minimum limit set, Undr Lmt is displayed.
If the measurement result is more than the maximum limit set, Over Lmt is
displayed.
18 This field displays Ofs if an offset is set.
19 This field displays Rng Hld if a range is selected.
20 This field displays Dty Cyc if a duty cycle is set.
21 The information in this field is displayed on two lines and depends on the
sensor type, sensor calibration table, frequency dependent offset table
currently selected, and the measurement frequency.
The following trace displays are only available when a P-Series or an E-Series
E9320 power sensor is connected.
Keysight N1911A/1912A User’s Guide41
1Introduction
23
22
NOTE
Sing Trig
Cont Trig
Trigger
Trace
Free Run
Figure 1-4Trace display in both windows
22 Figure 1-4 shows both windows configured to trace display mode. This is only
available with an P-Series or an E-Series E9320 power sensor is connected.
The captured trace, channel, and scaling are displayed. The lower window is
the selected window.
23 The indicates the point on the trace where trigger event occurs.
Single or continuous triggering ( or ) must be selected from
the Acqn menu to view a trace window. The Acqn menu is accessed by pressing
, . is disabled when is selected.
42Keysight N1911A/1912A User’s Guide
Introduction1
28
26
27
25
24
29
Gate
Marker
NOTE
Figure 1-5Trace display in single enlarged window in markers mode
Figure 1-5 shows a single enlarged window with the Gate Ctrl menu and
associated tables and markers. The Gate Control is the default display when using
trace display.
24 Pressing scrolls through the 4 gates available for each channel.
25 The markers, 1 and 2, indicate the start and end points of the selected gate.
Pressing toggles between the two markers. You can use the and
keys to move the active marker along the trace.
26 The highlighted table, with the green border, shows the time (Time:) and the
instantaneous power level (Pow:) of the active marker at the configured point.
A negative time value indicates a measurement before the trigger point.
Gate timing parameters are all related to your chosen trigger point. This may be
different from the timing of the triggering event if you have configured a trigger
delay. Refer to item No 30 for more information.
27 This table shows the time (Time:) and the instantaneous power level (Pow:) of
the inactive marker at the configured point.
28 The highlighted marker shows the active marker at the configured point.
29 This table shows the width of the gate
average, peak, and peak-to-average ratio power measurements within the
ΔT: (time between the markers), the
active gate.
Keysight N1911A/1912A User’s Guide43
1Introduction
31
32
30
Select TgDel
Figure 1-6Trace display in single enlarged window in trigger delay mode
Figure 1-6 shows a single enlarged window with the Gate Ctrl menu and
associated tables and trigger delay.
30 Pressing removes the gate markers and displays the trigger
marker(s). The indicates when the trigger event occurs, whilst shows the
delayed trigger point. When the two points coincide only the trigger is
shown.
In the example shown in Figure 1-6, the appears ahead of the as a
trigger delay of –5.00 µs has been configured, placing the measurement
trigger before the trigger event. You can configure the trigger delay by
pressing and entering a numeric value.
To indicate an off-screen trigger event, is displayed. To indicate an
off-screen trigger point, is displayed.
31 The selected channel and gate number is displayed.
32 The trigger setting and slope is displayed. Examples shown in Figure 1-6 are
Internal Channel A and a Positive Slope.
44Keysight N1911A/1912A User’s Guide
Introduction1
33
34
35
Figure 1-7Trace display in single enlarged window in trace control mode
Figure 1-7 shows a single enlarged window with the Trace Ctrl menu with
information on the pulse and associated X and Y control.
33 This is the X and Y trace setup fields. Using the or keys you can
highlight the item and change its value.
34 The current settings of the X and Y scale are displayed in this reporting line.
35 This table shows 8 automatic time measurements performed on the first
complete captured pulse after the trigger. The 8 measurements are rise time
, fall time , time to positive occurrence , time to negative
occurrence , pulse period, pulse width, duty cycle 50.0 %
and pulse repetitive frequency 12.5 kHz.
Keysight N1911A/1912A User’s Guide45
1Introduction
Window Symbols and Pop-ups
There are several different graphic symbols and pop-up windows that can occur
on the power meter display. These can occur for a variety of reasons, for example:
– An error or warning occurs
– You are required to wait while the power meter carries out a procedure
– You are required to select an entry from a list
– You are required to enter a numeric value
There are three different colors used to signify the pop-up status:
– Green - used to allow data entry
– Orange - used to display information
– Red - used to display an error
Warning symbol pop-up
The warning symbol is displayed either in a pop-up window or directly in the
measurement window when such an event occurs. A pop-up window is displayed
for approximately two seconds. The text in the pop-up window gives details of the
warning type, for example, to indicate that a power sensor has insufficient
bandwidth or a previous entered frequency value in a table. Depending on the
severity of the warning, the pop-up may be displayed in Orange or Red.
46Keysight N1911A/1912A User’s Guide
Wait symbol pop-up
The wait symbol is displayed when the power meter is carrying out a procedure
and no action is required from you. The symbol appears in a pop-up window. It
may appear, for example, during a calibration.
Confirm symbol pop-up
This type of pop-up window is displayed when you are required to press Confirm
to verify your previous selection. For example, prior to a Save being carried out.
Introduction1
Numeric entry pop-up
This type of pop-up window is displayed when you need to modify numeric data.
The numeric keys allow you to enter the value.
Keysight N1911A/1912A User’s Guide47
1Introduction
Text entry pop-up
This type of pop-up window is displayed when you need to modify alphanumeric
data, for example, table names. The up/down arrow keys increment and
decrement the alphanumeric digit that the cursor is currently positioned. The left/
right arrow keys move the cursor to another alphanumeric digit.
Trace scaling pop-up
This pop-up window is displayed when you are in the Trace Control menu and you
can either use the up/down arrow keys for increment and decrement, or the
numeric keys to set, the values of the X scale and Y scale.
48Keysight N1911A/1912A User’s Guide
List pop-up
This pop-up window is displayed when you are required to select an entry from a
list. Use the up/down arrow keys to highlight your choice. Press select to
complete the entry.
Introduction1
Keysight N1911A/1912A User’s Guide49
1Introduction
NOTE
Using the Instrument Web Browser
P-Series power meters network settings can be configured using its Web-based
interface (Web browser). The instrument’s Web browser can be opened from
Keysight Connection Expert as shown in Figure 1-8.
Alternatively, the instrument’s Web-based interface can also be opened directly
from a Web browser by entering the instrument’s IP address or hostname in the
browser’s ‘address’ window.
Figure 1-8Opening the instrument Web interface.
50Keysight N1911A/1912A User’s Guide
An example of the Web browser is shown in Figure 1-9.
Remote Interfaces
Introduction1
Keysight N1911A/1912A User’s Guide51
Figure 1-9N1912A P-Series power meter Web browser (Welcome Page)
Instrument on the network can be physically identified through the Status
message on Remote Interfaces screen by clicking Toggle ID within the Web
browser. This changes the status of the Instrument ID to ON or OFF.
Before the instrument identification starts, press
, to
view the Status message.
When Instrument ID is toggled to ON, this causes the Status message
“LAN identify start” in Remote Interface screen on the front panel to blink 3 times.
When the Toggle ID is selected again, the default Status message “Running” is
returned. See Figure 1-10 and Figure 1-11.
1Introduction
Status message
“LAN identity start” to
blink 3 times when
Toggle ID is selected.
Remote Interfaces
Figure 1-10Status message before Toggle ID is selected.
52Keysight N1911A/1912A User’s Guide
Figure 1-11Status message when Toggle ID is selected.
When power meter is configured to LAN, the Lan Status on
screen will show the LAN error condition and status of the LAN configuration
connection. There are six types of Lan Status messages that may occur. See
Table 1-1. See also Figure 1-12 for the example of Lan Status message.
Table 1-1Type of Lan Status message
Type of messageDescription
Introduction1
Lan: No Fault
Status: Initialized
Lan: No Fault
Status: Running
Lan: Fault
Status: Initialization failed
Lan: Fault
Status: Disconnected
Lan: –
Status: Restarting
Lan: DHCP Not Available
Status: Running
– A valid IP address is successfully obtained using selected LAN
configuration and the network state is initialized.
– A valid IP address is successfully obtained using selected LAN
configuration, while network is running.
– IP conflict occurs, or
– IP address failed to be obtained using selected LAN
configuration, or
– No LAN configuration is selected.
– LAN cable is unplugged.
– Restart the network and try to obtain an IP address using
selected LAN configuration.
– Unable to obtain IP address from DHCP server (if user select
DHCP configuration)
– IP address obtained from Auto-IP or manual configuration.
Figure 1-12Example of Lan Status message
Keysight N1911A/1912A User’s Guide53
1Introduction
NOTE
For more details on remote interface configuration, refer to P-Series Power
Meters Installation Guide.
Editing the Instrument’s LAN Settings
Once communication path to the instrument has been established, the
instrument’s LAN configuration can be viewed and modified using the Web
browser.
On the Welcome Page, click View and Modify Configuration. This opens the
configuration window shown in Figure 1-13.
Figure 1-13Viewing LAN configuration settings from the Web interface.
54Keysight N1911A/1912A User’s Guide
Introduction1
NOTE
Remote Interfaces
1 of 2
LAN Reset
NOTE
LAN Reset
NOTE
To edit parameters shown, click Mod ify Configuration. The Enter Password
dialog box appears as shown in Figure 1-14.
Figure 1-14Password security dialog box.
Click the Submit (accept the default password) and the window opens as shown
in Figure 1-15. The default password is “keysight”.
For P-Series power meters with latest firmware upgrade to be LXI-C compliant,
a LAN reset needs to be performed to ensure that the password is reset to
default. See LAN reset procedure as below.
Procedure:
Keysight N1911A/1912A User’s Guide55
1 Press , to display the Remote Interfaces screen.
2 Press softkey to display the second page of the Remote I/F menu.
3 Press softkey to reset the LAN settings to default.
When softkey is pressed, it will also change the GPIB address to
default.
You can change the password from the Configuring your N1912A Power Meter
window as shown in Figure 1-15. Scroll down the Parameter column until you
locate the Change Password parameter.
1Introduction
Figure 1-15Changing the instrument LAN interface configuration
56Keysight N1911A/1912A User’s Guide
General Power Meter Functions2
2General Power Meter
Functions
Setting the Units of Measurement 58
Setting the Measurement Frequency 60
Setting the Resolution 61
Making Relative Measurements 62
Setting the Pulse Reference Levels 64
Setting Offsets 67
Setting Measurement Averaging 78
Step Detection 81
Setting the Video Averaging 82
Setting the Video Bandwidth 84
Setting Measurement Channel Gates 87
Setting up the Channel Trace 92
Setting the Trigger 96
Setting External Trigger for Normal and Average Power Measurement 103
Setting Measurement Limits 118
Setting the Measurement Display 122
Scaling the Analog Display 129
Recorder Output 131
Saving and Recalling Power Meter States 137
Zeroing and Calibrating the P-Series Sensor 140
Presetting the Power Meter 145
Setting the Cable Short/Long 147
This chapter describes the general operation of the P-Series power meter.
Keysight N1911A/1912A User’s Guide57
2General Power Meter Functions
Units
dBmWdB
%
NOTE
Setting the Units of Measurement
The Units menu is used to select the measurement units for the currently selected
window. These can either be logarithmic (dBm or dB) or linear (Watt or %) units.
Presetting () the power meter sets the measurement units to dBm
(logarithmic units). Tab le 2-1 and Table 2-2 show units that are applicable to each
measurement mode.
Press , . Select the unit of measurement from , , , and
. Softkeys which cannot be selected in your particular mode of operation are
grayed out.
When the measurement unit is set to Watt (W), it is possible that negative power
results are displayed when measuring low power levels.
Tab le 2-1Measurement units - Single channel meters
Measurement ModeRelative Mode OffRelative Mode On
LogdBmdB
LinearWatt%
Tab le 2-2Measurement units - Dual channel meters
Measurement ModeRelative Mode OffRelative Mode On
Ratio
DifferenceLogdBmdB
LogdBdB
Linear%%
LinearWatt%
58Keysight N1911A/1912A User’s Guide
General Power Meter Functions2
mW
μW
nW
2 of 2
W
Increment
multiplier
Decrement
multiplier
Cancel
Increment Multiplier
Decrement Multiplier
W
W
NOTE
Selecting units of measurement from the softkeys
In some menus, for example, units in the Trace Setup menu, you are required to
enter the units of measurement for power. In some cases, due to the availability of
wide power range, the following menu is displayed:
Keysight N1911A/1912A User’s Guide59
Pressing or increases or decreases the
multiplier shown in front of .Pressing after the correct multiplier has been
selected confirms the entry.
Invalid softkeys are grayed out so the value cannot be entered.
2General Power Meter Functions
MHz
GHz
Setting the Measurement Frequency
Entering the frequency of the RF signal you are measuring optimizes the accuracy
and minimizes measurement uncertainty, especially when making comparative
measurements between signals.
Procedure
Set the measurement frequency as follows:
1 Press . On dual channel meters select the required channel.
2 Use the and keys to highlight the Frequency value field and press
to display the Frequency pop-up. Use the numeric keypad to enter the
required value in the Frequency pop-up window.
Figure 2-1Frequency pop-up
3 Confirm your choice by pressing or .
4 Press key to close the Channel Setup screen.
60Keysight N1911A/1912A User’s Guide
Setting the Resolution
Resolution
Resolution
The resolution of each of the power meter’s numeric type windows can be set to
four different levels (1, 2, 3 or 4).
These four levels represent:
1, 0.1, 0.01, 0.001 dB respectively if the measurement suffix is dBm or dB.
1, 2, 3 or 4 significant digits respectively if the measurement suffix is W or %.
The default value is 0.01 dB (3 digits).
To set the resolution on the currently selected window:
1 Press . The current setting of the resolution is highlighted under the
2 To change this setting, press repeatedly until the required
resolution setting is highlighted.
General Power Meter Functions2
softkey.
Keysight N1911A/1912A User’s Guide61
2General Power Meter Functions
Meas Select
Relative setting
field
Rel field
Original result
Selected
Window/
Measuremen
Result field
Making Relative Measurements
Relative mode enables comparison of a measurement result to a reference value.
The relative reading, or difference, can be displayed in either dB or % terms. When
the measurement result is displayed in % a prefix multiplier may be shown.
Procedure
1 Press to display the Measurement Setup menu.
Figure 2-2 shows a Measurement Setup display and the relative measurement
items labeled.
2 Select the window you wish to set a reference value on by pressing the
key. The currently selected window/measurement is displayed.
3 Use the and keys to highlight the Relative setting field.
62Keysight N1911A/1912A User’s Guide
Figure 2-2Typical relative measurement display
4 Press to check the Relative setting field.
5 Confirm that the power meter is measuring the signal you want to use as the
reference. This is displayed under the Result field.
General Power Meter Functions2
NOTE
Rel indicator
NOTE
6 Press the , , key to highlight the Rel field.
7 Press and the original result value will appear on the right hand side of
the Rel check field.
8 The relative value displayed under Result field will change as the measured
signal varies.
If you return the power meter to display the numeric display, a Rel symbol will
be displayed in the measurement window it is applied to.
Figure 2-3Numeric display
The Rel symbol is not displayed when the associated measurement is displayed
in Dual Numeric or Analog format.
Keysight N1911A/1912A User’s Guide63
2General Power Meter Functions
1 of 2
Meas Config
Channel A B
NOTE
NOTE
Setting the Pulse Reference Levels
– Press and proceed to the second page of Sys/Inputs menu by
pressing the softkey.
– Press softkey to display the Measurement Configuration screen.
– Press the softkey to select the desired channel for a dual channel
power meter (N1912A).
Figure 2-4Measurement configuration screen
Trace Ref Lvl 1 and Trace Ref Lvl 2 are used in calculation of transition durations
and occurrences.
Figure 2-5Pulse duration measurement between two reference levels
1 Use the and keys to highlight the Trace Ref Lvl 1 field.
Keysight N1911A/1912A User’s Guide65
2 Press to display the Trace Ref Lvl 1 pop-up, and use the numeric
keypad to enter the intended value in the Trace Ref Lvl 1 pop-up window.
Figure 2-6Trace Ref Lvl 1 pop-up
3 Complete the entry by pressing the softkey.
4 Use the and keys to highlight the Trace Ref Lvl 2 field.
2General Power Meter Functions
%
%
NOTE
5 Press to display the Trace Ref Lvl 2 pop-up, and use the numeric
keypad to enter the intended value in the Trace Ref Lvl 2 pop-up window.
Figure 2-7Trace Ref Lvl 2 pop-up
6 Complete the entry by pressing the softkey.
7 Use the and keys to highlight the Pul se Duration Ref Lvl field.
8 Press to display the Pul se Duration Ref Lvl pop-up, and use the
numeric keypad to enter the intended value in the Pulse Duration Ref Lvl
pop-up.
Figure 2-8Pulse Duration Ref Lvl pop-up
9 Complete the entry by pressing the softkey.
By default, the Trace Ref Lvl 1, Trace Ref Lvl 2 and Pulse Duration Ref Lvl
settings are set to 10 %, 90 % and 50 % respectively.
66Keysight N1911A/1912A User’s Guide
Setting Offsets
Channel functions
Window functions
Channel A
Cal
Factor
Freq Dep
Offset
Channel
Offset
Channel B
Cal
Factor
Freq Dep
Offset
Channel
Offset
Measurement Highway
Upper Window
Lower Window
Display
Offset
Display
Offset
Relative
Relative
Display
Offset
Display
Offset
Relative
Relative
The power meter can be configured to compensate for a signal loss or gain in your
test setup. The power meter allows you to apply offsets at three different points in
the measurement path.
General Power Meter Functions2
Figure 2-9Simplified measurement path
Figure 2-9 shows that how you can apply a Channel Offset or a
Frequency Dependent Offset prior to any mathematical functions. These allow
you to compensate each channel individually. An overall offset can be applied if
required using the Display Offset.
Keysight N1911A/1912A User’s Guide67
2General Power Meter Functions
Offsets
Offset setting field
Offset value field
dB
Setting channel offsets
This gain or loss is applied to the measured power before any mathematical
functions, display offsets or relative functions are included.
Offsets are entered in dB and the range is –100 dB to +100 dB.
A positive value compensates for a loss and a negative value compensates for a
gain.
Procedure
To enter a channel offset:
1 Press to display the Channel Setup screen. Confirm the channel
requiring setup is displayed.
2 Press to display the Offsets Setup.
3 Use the and keys to highlight the Offset setting field.
68Keysight N1911A/1912A User’s Guide
4 Press to check the Offset setting field.
Figure 2-10Typical channel offset display
5 Press to highlight the Offset value field and press to display the
Offset pop-up. Use the numeric keypad to enter the required value in the
Offset pop-up window.
6 Confirm your choice by pressing .
General Power Meter Functions2
Offset indicator
NOTE
7 Press key to complete the offset entry. If either a channel or a display
offset is set, the Ofs indicator is displayed.
Figure 2-11Channel offset indicator
The Ofs symbol is not displayed when the associated measurement is displayed
in Dual Numeric or Analog format.
Keysight N1911A/1912A User’s Guide69
2General Power Meter Functions
Meas Select
Offset setting field
Offset value field
Selected
Window/
Measurement
Setting display offsets
This gain or loss is applied to the measured power after any channel offsets or
mathematical functions have been included.
Offsets are entered in dB and the range of values is –100 dB to +100 dB. A
positive value compensates for a loss, and a negative value compensates for a
gain.
Procedure
Enter a display offset on the currently selected window:
1 Press to display the Measurement Setup screen.
2 Select the window you wish to set the offset value on by pressing the
key. The currently selected window/measurement is displayed.
3 Use the and keys to highlight the Offset setting field.
70Keysight N1911A/1912A User’s Guide
4 Press to check the Offset setting field.
Figure 2-12Typical display offset display
5 Press to highlight the Offset value field and press to display the
dB
NOTE
Display Offset pop-up. Use the numeric keypad to enter the required value in
the Offset pop-up window.
6 Confirm your choice by pressing . Press key to complete the
offset entry.
The Ofs indicator is displayed if Display Offset is selected.
Setting frequency dependent offsets
Frequency dependent offset tables provide a quick and convenient method of
compensating for frequency related changes in the response of your test system.
Note that when selected, frequency dependent offset corrections are applied
IN ADDITION to any correction for sensor frequency response.
The power meter is capable of storing 10 frequency dependent offset tables with a
maximum of 80 frequency points each.
General Power Meter Functions2
To use frequency dependent offset tables:
1 Select the table to be applied to a channel. Refer to “Setting frequency
dependent offsets” on page 71 for further information. If you require to edit the
table refer to “Editing frequency dependent offset tables” on page 74 for
further information.
2 If using an 8480 Series, N8480 Series or an E-Series sensor, zero and calibrate
the power meter. The reference calibration factor used during the calibration is
automatically set by the power meter from the sensor calibration table (if
selected).
3 Specify the frequency of the signal you want to measure. The calibration
factor/offset is automatically set by the power meter from the sensor
calibration table (if selected) and the frequency dependent offset table. Refer
to “Procedure” on page 72 for further information.
4 Make the measurement.
Keysight N1911A/1912A User’s Guide71
2General Power Meter Functions
Tab les
Freq. Dep. Offset
NOTE
Offsets
Tab les
Freq. Dep. Offset
Offsets
Selecting a frequency dependent offset table
You can select a frequency dependent offset table from the key menu
followed by and .
The State column indicates if any frequency dependent offset tables are currently
selected. The Offset Tables screen is shown in Figure 2-13.
You can also view which FDO table is being used by pressing the ,
and use the and keys to highlight the FDO Table setting field and
press to display the table.
Procedure
Select an offset table as follows:
1 Press, either:
72Keysight N1911A/1912A User’s Guide
a, , .
b, and use the and keys to highlight the FDO Table
2 Use the and keys to highlight one of the 10 table titles and press
to highlight .
When no data is contained in the highlighted table, the key is disabled
(grayed out).
Keysight N1911A/1912A User’s Guide73
3 Press to complete the selection of the offset table.
4 Press again to display the measurement screen. Figure 2-14 shows
which offset table is selected.
Figure 2-14Frequency dependent offset indicator
2General Power Meter Functions
NOTE
Tab les
Freq. Dep. Offset
5 To change the frequency, press and use the and keys to
highlight the Frequency field.
6 Press to display the Frequency pop-up window. Use the numeric
keypad to enter the required value in the Frequency pop-up window.
7 To confirm your choice, press the appropriate unit softkey.
8 Connect the power sensor to the signal to be measured.
9 The measurement result, including offset, is now displayed.
If the measurement frequency does not correspond directly to a frequency in the
sensor calibration table (if selected) and the frequency dependent offset table
being used, the power meter calculates the calibration factor and offset using
linear interpolation.
If you enter a frequency outside the frequency range defined in the sensor
calibration table or the frequency dependent offset table, the power meter uses
the highest or lowest frequency point in the appropriate table to set the
calibration factor and offset.
74Keysight N1911A/1912A User’s Guide
Editing frequency dependent offset tables
There are ten frequency dependent offset tables named CUSTOM_A through
CUSTOM_J. They are empty of any data when the power meter is shipped from
the factory.
You cannot delete any of the 10 existing frequency dependent offset tables or
create any additional tables. However, you can enter values into the 10 existing
tables. Each frequency dependent offset table can contain a maximum of 80
frequency points.
To view the frequency dependent offset tables currently stored in the power
meter, press , , . The Frequency Dependent Offset Tables screen is displayed as shown in Figure 2-13.
General Power Meter Functions2
NOTE
Tab les
Freq. Dep. Offset
Edit Table
Editing frequency dependent offset tables requires the following steps:
1 Identify and select the table you want to edit
2 Rename the table
3 Enter the frequency and offset pairs
4 Save the table
Procedure
First select the table you want to edit as follows:
A frequency in the range of 0.001 MHz to 1000.0 GHz can be entered. A
calibration factor in the range of 1% to 150% can be entered.
1 Press , , to display the Offset Tbls screen.
2 Choose the table you want to edit using the and keys. Press
to display the Ed it Offset screen as shown in Figure 2-15.
Keysight N1911A/1912A User’s Guide75
Figure 2-15“Edit Offset” display with data added
2General Power Meter Functions
Change
Insert Char
Delete Char
Enter
NOTE
3 Highlight the table title using the and keys. Press and use the
, , and keys to select and change the characters in the Tab le
Name pop-up to create the name you want to use.
Figure 2-16Edit table title pop-up
Pressing adds a new character to the right of the selected character.
Pressing removes the selected character.
4 Press to complete the entry.
The following rules apply to naming sensor calibration tables:
– The name must consist of no more than 12 characters.
– All characters must be upper or lower case alphabetic characters, or numeric
(0-9), or an underscore (_).
– No other characters are allowed.
– No spaces are allowed in the name.
76Keysight N1911A/1912A User’s Guide
General Power Meter Functions2
Insert
Change
GHz
MHz
Change
%
Done
NOTE
Enter (or edit) the frequency and offset pairs as follows:
1 Press to add a new frequency value (or press to edit). Use the
numeric keypad to enter the required value in the Frequency pop-up window.
Complete the entry by pressing the , keys.
2 Enter the new offset value (or press to edit). Use the numeric keypad
to enter the required value in the Offset pop-up window. Complete the entry by
pressing the key.
3 Continue adding/editing values until you have entered all the data you require.
4 When you have finished editing the table press to save the table.
If you measure a signal with a frequency outside the frequency range defined in
the frequency dependent offset table, the power meter uses the highest or
lowest frequency point in the frequency dependent offset table to calculate the
offset.
Keysight N1911A/1912A User’s Guide77
2General Power Meter Functions
Minimum sensor power
Resolution setting
Maximum sensor power
Power sensor dynamic range
Number of averages
10 dB
10 dB
10 dB
10 dB
1
1
1
8
2
2
1
1
1
1
1
1
3
4
16
16
128
128
256
32
8
1
1
1
Setting Measurement Averaging
The power meter uses a digital filter to average power readings. The number of
readings averaged can range from 1 to 1024. This filter is used to reduce noise,
obtain the desired resolution and to reduce the jitter in the measurement results.
Increasing the value of the measurement average reduces measurement noise.
However, the measurement time is increased. You can manually select the
measurement average or you can set the power meter to auto measurement
average mode. The default is AUTO.
When the auto measurement average mode is enabled, the power meter
automatically sets the number of readings averaged together to satisfy the
filtering requirements for most power measurements. The number of readings
averaged together depends on the resolution setting and the power level
currently being measured.
Figure 2-17Typical averaged readings
Figure 2-17 shows the typical number of averages for each range and resolution
when the power meter is in auto filter mode and is set to normal speed mode.
P-Series power meters recognize different sensor types when they are connected,
and configure suitable averaging automatically.
78Keysight N1911A/1912A User’s Guide
Resolution is a measurement display function and not a channel function. In the
case where a channel is set up in both the upper and lower window and the
resolution settings are different, the highest resolution setting is taken to calculate
the averaging number.
These four resolution levels represent:
Procedure
Set measurement averaging as follows:
1 Press . On dual channel meters select the required channel. The
2 Use the and keys to select the Filter setting field.
General Power Meter Functions2
– 1, 0.1, 0.01, 0.001 dB respectively if the measurement suffix is dBm or dB.
– 1, 2, 3 or 4 significant digits respectively if the measurement suffix is
Wor%.
current setting is shown in the Meas Avg field (
Channel Setup screen. The default is
AUTO.
AUTO, MAN, or OFF) on the
3 Press and use the and to step through the available settings.
If you have selected AUTO or OFF proceed at step 7. If you have selected MAN
proceed as follows:
4 Use the key to select the Meas Avg: value field.
5 Press to display the Meas Avg Count pop-up.
Figure 2-18Meas Avg Count pop-up
Keysight N1911A/1912A User’s Guide79
2General Power Meter Functions
Enter
6 Use the numeric keys to enter the required value and press .
7 Press key to close the Channel Setup screen.
80Keysight N1911A/1912A User’s Guide
Step Detection
To reduce the filter settling time after a significant step in the measured power the
filter can be set to re-initialize upon detection of a step increase or decrease in the
measured power. Step detection can be set in both manual and automatic
measurement average modes.
Procedure
Set step detection as follows:
1 Press . On dual channel meters select the required channel.
2 Use the and keys to select the Step Detect setting field.
3 Press to check the step detection to on or off.
4 Press key to close the Channel Setup screen.
General Power Meter Functions2
Keysight N1911A/1912A User’s Guide81
2General Power Meter Functions
NOTE
Setting the Video Averaging
This feature is only available when a P-Series or an E-Series E9320 power sensor
is connected.
Video averaging uses a digital filter to average repetitions of a triggered signal.
The number of acquisitions averaged can range from 1 to 256, in multiples of 2
With video averaging the average of a number of acquisitions is calculated to
smooth the displayed trace and reduce apparent noise. The measurement
requires a continuously repeating signal. Increasing the value of this filter reduces
noise but increases the time required to make the measurement.
Procedure
Set the video average as follows:
1 Press . On dual channel meters select the required channel.
n
.
2 Use the and keys to select the Video Avg setting field.
3 Press to check the video averaging to on or off.
4 Use the key to select the Video Avg: value field.
5 Press to display the Video Avg Count pop-up.
82Keysight N1911A/1912A User’s Guide
General Power Meter Functions2
Figure 2-19Video Avg Count pop-up
6 Use the and keys to highlight the required Video Avg Count value and
press .
7 Press key to close the Channel Setup screen.
Keysight N1911A/1912A User’s Guide83
2General Power Meter Functions
NOTE
Setting the Video Bandwidth
This feature is only available when a P-Series or an E-Series E9320 power sensor
is connected.
Selecting a bandwidth value close to or slightly greater than required by the
modulating signal bandwidth can help reduce noise and improve accuracy on
peak measurements (see Table 2-3). It can however, reduce the processing speed
for long acquisition times.
Tab le 2-3P-Series Sensor Video Bandwid th Setting
Low: Medium: High: Off
5.0 MHz15.0 MHz30.0 MHz>30.0 MHz
The Low, Medium, and High pass band shapes achieved by the video bandwidth
settings provide flat filter responses with very sharp cut-off points by applying
digital signal processing techniques to ensure accurate power measurement
within the specified band.
84Keysight N1911A/1912A User’s Guide
General Power Meter Functions2
AmplitudeP-Series sensors
Frequency
Approx
-3.5 dB
0
5 MHz 15 MHz
30 MHz
Low
Med
High
Off
Meter B/W
setting
Figure 2-20Bandwidth filter shapes
When video bandwidth is set to Off, it removes all digital signal conditioning. This
provides approximately 3 dB roll off at the maximum sensor’s bandwidth and is
best suited for accurate trace capture, minimizing overshoot, and removing any
ringing effects caused by the sharp cut-off filters used in the Low, Med and High
settings. Figure 2-20 shows the filter shapes associated with the P-Series
Procedure
wideband power sensors.
Set the video bandwidth as follows:
1 Press . On dual channel meters select the required channel.
2 Use the and keys to select the Video B/W setting field.
Keysight N1911A/1912A User’s Guide85
2General Power Meter Functions
3 Press to display the Video B/W pop-up.
Figure 2-21Video bandwidth pop-up
4 Use the and keys to highlight the required Video Bandwid th setting
and press .
5 Press key to close the Channel Setup screen.
86Keysight N1911A/1912A User’s Guide
Setting Measurement Channel Gates
NOTE
Gate 1
Gate 4
Gate 3
Gate 2
Trigger
This feature is only available when a P-Series or an E-Series E9320 power sensor
is connected.
A system of gates, controlled by and referenced to a trigger point, is used to
obtain measurement data from a captured trace. The trace data within each gate
period is subsequently used for the individual measurement calculations. Up to 4
gates can be set up for each channel. Figure 2-22 shows an example of 4 gates
setup to perform the following measurements simultaneously:
– Average power level of the pulse:
Gate 1, average measurement
– Average “off” power level ahead of the pulse:
Gate 2, average measurement
– Peak to average ratio:
Gate 1, peak-to-average measurement
General Power Meter Functions2
–Pulse drop:
Gate 3, average measurement, minus Gate 4, average measurement
Keysight N1911A/1912A User’s Guide87
Figure 2-22Example of measurement gates
2General Power Meter Functions
NOTE
NOTE
Gates Setup
NOTE
Figure 2-22 shows measurement results using this example of the gates
placement.
Procedure
This procedure uses the Gates Setup under the Channel Setup menu.
Alternatively, you can use a more visual method, when in the Graphical Trace
Mode (Gate Control Menu), to set the channel’s Gate Control. (See “Setting the
Trace Display” on page 152).
– Press . The Channel Gate Setup screen is displayed.
Figure 2-23Gate Setup screen
1 Highlight the Gate Start you want to configure using the , , and
keys.
The gate start time is relative to the trigger event. Positive values set a
measurement gate, to a maximum time of 1 second, after the trigger. Negative
time gate start values set a measurement gate, to a maximum time of 1 second,
before the trigger.
88Keysight N1911A/1912A User’s Guide
General Power Meter Functions2
smsμs
ns
smsμs
ns
2 Press and use the numeric keypad to enter the required value in the
Time Gating Start pop-up window.
Figure 2-24Time Gating Start pop-up
3 Complete the entry by pressing the required second, millisecond, microsecond
or nanosecond ( , , or ) softkey.
4 Highlight the Gate Length you want to configure using the , , , and
keys.
Figure 2-25Time Gating Length pop-up
5 Complete the entry by pressing the required second, millisecond, or
microsecond or nanosecond ( , , or ) softkey.
6 Repeat this process until you have setup all the required gates.
Keysight N1911A/1912A User’s Guide89
2General Power Meter Functions
Auto Gate
Auto Gate Settings
NOTE
Gate 1 Ref 1 0.0%
Auto Gate
This feature will automatically set the gates (start and stop times of the user
defined time interval) for each measurement relative to the rising and falling
edges’ trigger point.
– Proceed to the second page of Gate Ctrl menu and press softkey.
–To change the Auto Gate setup:
1 Press softkey.
90Keysight N1911A/1912A User’s Guide
Figure 2-26Auto Gate Settings display
Auto Gate Settings enables the placing of markers (gates) at different positions.
2 Press softkey. The Marker 1 Ref Percentage pop-up is
displayed.
Figure 2-27Marker 1 Ref Percentage pop-up
General Power Meter Functions2
%
Gate 1 Ref 2 0.0%
%
Ref 1 x PD
Ref 2 x PD
Pulse Duration (PD)
NOTE
Perpetual Off On
3 Use the numeric keys to complete the entry and press softkey.
4 Press softkey. The Marker 2 Ref Percentage pop-up is
displayed.
Figure 2-28Marker 2 Ref Percentage pop-up
5 Use the numeric keys to complete the entry and press softkey.
Keysight N1911A/1912A User’s Guide91
Figure 2-29The placement of gates (markers) in Auto Gating mode
Enabling the softkey allows the gates to be repositioned
automatically according to the changes of pulse duration.
2General Power Meter Functions
NOTE
NOTE
NOTE
Trace Setup
Setting up the Channel Trace
This feature is only available when a P-Series or an E-Series E9320 power sensor
is connected.
The Trace Setup is the only location where you can change the Y-axis units from
dBm to Watts.
Procedure
This procedure uses the Trace Setup under the Channel Setup menu.
Alternatively, you can use a more visual method, when in the Graphical Trace
Mode (Trace Control Menu), to set the channel’s Trace Control. (See “Setting the
Trace Display” on page 152).
1 Press , to display the Trace Setup menu.
Figure 2-30Trace setup display
92Keysight N1911A/1912A User’s Guide
General Power Meter Functions2
smsμs
ns
smsμs
ns
2 Use the and keys to highlight the X Start field.
3 Press to display the Trace Start Time pop-up and use the numeric
keypad to enter the required value in the Trace Start Time pop-up window.
Figure 2-31Trace Start Time pop-up
4 Complete the entry by pressing the required second, millisecond, microsecond
or nanosecond ( , , or ) softkey.
5 Use the and keys to highlight the X Scale field.
6 Press to display the X Scale/Division pop-up.
a Use the numeric keypad to enter the required value in the X Scale/Division
pop-up window.
b Use the and keys to increase or decrease the value.
Figure 2-32X Scale/Division pop-up
7 Complete the entry by pressing:
a The required second, millisecond, microsecond or nanosecond ( , ,
or ) softkey.
Keysight N1911A/1912A User’s Guide93
2General Power Meter Functions
Enter
dBm
b The softkey.
8 Use the and keys to highlight the Units field.
If you want to view the trace in linear scale, use this field. Otherwise the
default scale is in logarithmic.
9 Press to display the Units pop-up, and use the and to highlight
Watt or dBm.
Figure 2-33Trace Units pop-up
10 Press to complete the entry.
11 Use the and keys to highlight the Y Max field.
12 Press to display the Trace Maximum pop-up, and use the numeric
keypad to enter the required value in the Trace Maximum pop-up window.
Figure 2-34Y scale maximum value pop-up
13 Complete the entry by pressing the or the linear value.
14 Use the and keys to highlight the Y Scale field.
94Keysight N1911A/1912A User’s Guide
General Power Meter Functions2
dB
Trace Control
Auto Scale
NOTE
15 Press to display the Y Scale/Division pop-up.
a Use the numeric keypad to enter the required value in the Y Scale/Division
pop-up window.
b Use the and keys to increase or decrease the value.
Figure 2-35Y Scale/Division pop-up
16 Complete the entry by pressing the or the linear value.
17 Press key to complete the setup and display the measurements
results.
Auto Scale
Auto Scale automatically configures the window’s X-axis and Y-axis to best
display the input trace.
1 Press softkey under Gate Ctrl menu.
2 softkey is pressed to accomplish auto scaling.
This feature will work reliably with well behaved pulses. Modulated pulses may
cause the meter to auto-scale on the details within the pulse envelope.
Keysight N1911A/1912A User’s Guide95
2General Power Meter Functions
NOTE
Acqn
Trigger status
NOTE
Setting the Trigger
This feature is only available when a P-Series power sensor is connected.
The trigger can be taken from a rising or falling measured power level or
controlled externally using the Ext Trig input. Additional control features such as;
hold-off, hysteresis, and delay are provided to help you achieve a stable and
reliable trigger. To use the measurement gates, the power meter must be
triggered.
Press . The Trigger menu is displayed.
The trigger status is displayed below the label in the Trigger menu.
Figure 2-36 shows the power meter display in Free Run mode. In this mode the
meter is not synchronized to any modulated RF signal at the sensor input.
Consequently, power levels within the configured time gates are random and the
displayed measurement results are not valid.
Figure 2-36Trigger menu - free run mode
The symbol in the upper window indicates the power meter is in free run
mode.
96Keysight N1911A/1912A User’s Guide
Procedure
Acqn
Sing Trig
Cont Trig
Sing Trig
Cont Trig
Settings
Rising edge source
Channel A is
trigger source
1 Press softkey to configure a trigger.
2 Select either or .
3 Press to configure the remaining trigger parameters. The trigger
General Power Meter Functions2
– is a single shot mode. After triggering, the measurement is
halted, the symbol is displayed. You can start another measurement by
pressing the key.
– is a continuos trigger mode. The symbol or is displayed.
Settings menu have two pages. Figure 2-37 shows page one and Figure 2-40
shows page two.
Keysight N1911A/1912A User’s Guide97
Figure 2-37Trigger Setting Menu 1 of 2
To help you quickly check the triggering setup, the current settings for all the
trigger parameters are shown with their respective softkey labels.
2General Power Meter Functions
Source
Source
Ext
Int
Ext
NOTE
Source
Ext
Ch.A
Ch.B
Mode
Source Int
Mode
Norm
AutoLvl
Norm
AutoLvl
Level
Norm
Level
Level
Source
The current setting is shown below the label.
To change the setting press and choose or .
The trigger source is shown beside the trigger symbol when a measurement
window is configured in single numeric mode. When an external trigger () is
selected, the power meter can be triggered via the Ext Trig input.
For N1912A, by pressing , you can choose your source through the three
different softkeys: , , and .
Mode
The key is only available when trigger is selected.
To change the setting press and choose or .
The current setting is displayed below the label. When is selected you can
choose the RF power level transition used as the trigger. The power meter
automatically finds a triggering power level transition when is selected.
Level
is only available when trigger mode has been selected. The current
value is shown below the label. The minimum power level you can enter is
limited to 40 dB below the maximum sensor power. To change the setting press
and use the numeric keypad to enter the required value in the Trigger Level
pop-up window to enter the new value.
98Keysight N1911A/1912A User’s Guide
General Power Meter Functions2
dBm
Delay
Delay
smsμs
ns
NOTE
1 of 2
Figure 2-38Trigger Level pop-up
Complete the entry by pressing .
Delay
The current setting is displayed below the label. The delay time is applied
between the trigger event and all the gate start times. This allows you to
time-shift all the gates by the same amount with one setting change. A delay of
up to 1 second can be entered. To enter or change the setting press and
use the numeric keypad to enter the required value in the Trigger Delay pop-up
window to enter the new value.
Figure 2-39Trigger Delay pop-up
Complete the entry by pressing the required second, millisecond, microsecond or
nanosecond ( , , or ) softkey.
The trigger delay time can also be entered or changed in the Trace Display mode
when the Gate Control menu is displayed. See “Setting the Trace Display” on
page 152 for further detail.
Press to display the second menu page.
Keysight N1911A/1912A User’s Guide99
2General Power Meter Functions
Slope
+
–
Slope
+
–
Holdoff
Holdoff
Figure 2-40Trigger setting menu 2 of 2
Slope
The current setting is displayed below the label and the or symbol is
displayed in single numeric display mode. (and ) is used to generate the
trigger from an increasing power level. Similarly, (and ) is used to generate
the trigger from a decreasing power level. To change the setting, press to
highlight or as required.
Holdoff
The current setting is displayed below the label. After a trigger event
occurs, the trigger mechanism is disabled for the configured time period. This
allows stable triggering to be achieved even when a signal has multiple edges, for
example, a TDMA signal with non-constant amplitude modulation. Values up to
400 ms can be configured.
To change the setting press and use the numeric keypad to enter the
required value in the Trigger Holdoff pop-up window to enter the new value.
100Keysight N1911A/1912A User’s Guide
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.