No part of this manual may be reproduced
in any form or by any means (including
electronic storage and retrieval or translation into a foreign language) without prior
agreement and written consent from Keysight Technologies as governed by United
States and international copyright laws.
Manual Printing History
The manual’s printing date and part number indicate its current edition. The printing
date changes when a new edition is printed
(minor corrections and updates that are
incorporated at reprint do not cause the
date to change). The manual part number
changes when extensive technical
changes are incorporated.
October 2010 First Edition
September 2011 Second Edition
August 2013 Third Edition
November 2014 Fourth Edition
.
Manual Part Number
E5061-90017
Edition
Edition 4, November 2014
Printed in Malaysia
Keysight Technologies
1-3-3 Higashikawasaki-cho, Chuo-ku Kobe
Hyogo Japan
Warranty
The material contained in this document is provided “as is,” and is subject to being changed, without notice,
in future editions. Further, to the maximum extent permitted by applicable
law, Keysight disclaims all warranties, either express or implied, with
regard to this manual and any information contained herein, including
but not limited to the implied warranties of merchantability and fitness for
a particular purpose. Keysight shall
not be liable for errors or for incidental or consequential damages in connection with the furnishing, use, or
performance of this document or of
any information contained herein.
Should Keysight and the user have a
separate written agreement with
warranty terms covering the material
in this document that conflict with
these terms, the warranty terms in
the separate agreement shall control.
licensed as “Commercial computer software” as defined in DFAR 252.227-7014
(June 1995), or as a “commercial item” as
defined in FAR 2.101(a) or as “Restricted
computer software” as defined in FAR
52.227-19 (June 1987) or any equivalent
agency regulation or contract clause. Use,
duplication or disclosure of Software is
subject to Keysight Technologies’ standard
commercial license terms, and non-DOD
Departments and Agencies of the U.S. Government will receive no greater than
Restricted Rights as defined in FAR
52.227-19(c)(1-2) (June 1987). U.S. Government users will receive no greater than
Limited Rights as defined in FAR 52.227-14
(June 1987) or DFAR 252.227-7015 (b)(2)
(November 1995), as applicable in any
technical data.
Safety Notices
A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like
that, if not correctly performed or
adhered to, could result in damage
to the product or loss of important
data. Do not proceed beyond a
CAUTION notice until the indicated
conditions are fully understood and
met.
Technology Licenses
The hardware and/or software
described in this document are
furnished under a license and
may be used or copied only in
accordance with the terms of
such license
Restricted Rights Legend
If software is for use in the performance of
a U.S. Government prime contract or subcontract, Software is delivered and
A WARNING notice denotes a
hazard. It calls attention to an
operating procedure, practice, or
the like that, if not correctly performed or adhered to, could result
in personal injury or death. Do not
proceed beyond a WARNING
notice until the indicated conditions are fully understood and met.
Caution
Do not exceed the operating input power, voltage,
and current level and signal type appropriate for the
instrument being used, refer to your instrument's Function
Reference.
Electrostatic discharge (ESD) can damage the highly
sensitive microcircuits in your instrument. ESD damage is
most likely to occur as the test fixtures are being connected
or disconnected. Protect them from ESD damage by wearing
a grounding strap that provides a high resistance path to
ground. Alternatively, ground yourself to discharge any static
charge built-up by touching the outer shell of any grounded
instrument chassis before touching the test port connectors.
Network Analyzer3
Safety Summary
When you notice any of the unusual conditions listed below,
immediately terminate operation and disconnect the power
cable.
Contact your local Keysight Technologies sales representative
or authorized service company for repair of the instrument.
If you continue to operate without repairing the instrument,
there is a potential fire or shock hazard to the operator.
• Instrument operates abnormally.
• Instrument emits abnormal noise, smell, smoke or a
spark-like light during operation.
• Instrument generates high temperature or electrical shock
during operation.
• Power cable, plug, or receptacle on instrument is
damaged.
• Foreign substance or liquid has fallen into the instrument.
4Network Analyzer
Manufacturer’s Declaration
Herstellerbescheinigung
GERA- USCHEMISSION
Manufacturer's Declaration
ACOUSTIC NOISE EMISSION
LpA < 70 dB
am Arbeitsplatz
normaler Betrieb
nach DIN 45635 T. 19
LpA < 70 dB
operator position
normal operation
per ISO 7779
Network Analyzer5
Regulatory Compliance Information
This product complies with the essential requirements of the
following applicable European Directives, and carries the CE
marking accordingly:
• The Low Voltage Directive 2006/95/EC
• The EMC Directive 2004/108/EEC
To obtain Declaration of Conformity, please contact your
local Keysight Technologies sales office, agent or distributor.
6Network Analyzer
Safety Notice Supplement
• This equipment complies with EN/IEC61010-1:2001.
• This equipment is of MEASUREMENT CATEGORY I (CAT
I). Do not use for CAT II, III, or IV.
• Do not connect the measuring terminals to mains.
• This equipment is a POLLUTION DEGREE 2, INDOOR
USE product.
• This equipment is tested in stand-alone condition and in
combination with the accessories supplied by Keysight
Technologies against the requirement of the standards
described in the Declaration of Conformity. If it is used
as a system component, compliance of related regulations
and safety requirements are to be confirmed by the
builder of the system.
Network Analyzer7
General Safety Precautions
NOTE
NOTE
The following general safety precautions must be observed
during all phases of operation, service, and repair of this
instrument. Failure to comply with these precautions or with
specific WARNINGS elsewhere in this manual may impair the
protection provided by the equipment. Such noncompliance
would also violate safety standards of design, manufacture,
and intended use of the instrument. Keysight Technologies
assumes no liability for the customer’s failure to comply
with these precautions.
The E5061B complies with INSTALLATION CATEGORY II as well as
POLLUTION DEGREE 2 in IEC61010-1. The E5061B is an INDOOR USE
product.
The LEDs in the E5061B are Class 1 in accordance with IEC60825-1,
CLASS 1 LED PRODUCT.
• Ground the Instrument
To avoid electric shock, the instrument chassis and
cabinet must be grounded with the supplied power cable’s
grounding prong.
• DO NOT Operate in an Explosive Atmosphere
Do not operate the instrument in the presence of
inflammable gasses or fumes. Operation of any electrical
instrument in such an environment clearly constitutes a
safety hazard.
• Keep Away from Live Circuits
Operators must not remove instrument covers. Component
replacement and internal adjustments must be made by
qualified maintenance personnel. Do not replace
components with the power cable connected. Under
certain conditions, dangerous voltage levels may remain
even after the power cable has been disconnected. To
avoid injuries, always disconnect the power and discharge
circuits before touching them.
• DO NOT Service or Adjust the Instrument Alone
Do not attempt internal service or adjustment unless
another person, capable of rendering first aid and
resuscitation, is present.
• DO NOT Substitute Parts or Modify the Instrument
8Network Analyzer
To avoid the danger of introducing additional hazards, do
WARNING
not install substitute parts or perform unauthorized
modifications to the instrument. Return the instrument to
a Keysight Technologies Sales and Service Office for
service and repair to ensure that safety features are
maintained in operational condition.
• Dangerous Procedure Warnings
Warnings, such as the example below, precede potentially
dangerous procedures throughout this manual.
Instructions contained in the warnings must be followed.
Dangerous voltage levels, capable of causing death, are present in
this instrument. Use extreme caution when handling, testing, and
adjusting this instrument.
Network Analyzer9
Safety Symbols
WARNING
CAUTION
NOTE
General definitions of safety symbols used on the instrument
or in manuals are listed below.
Instruction Manual symbol: the product is marked
with this symbol when it is necessary for the user
to refer to the instrument manual.
Alternating current.
Direct current.
On (Supply).
Off (Supply).
In-position of push-button switch.
Out-position of push-button switch.
A chassis terminal; a connection to the
instrument’s chassis, which includes all exposed
metal structure.
Standby.
This warning sign denotes a hazard. It calls attention to a
procedure, practice, or condition that, if not correctly performed or
adhered to, could result in injury or death to personnel.
This Caution sign denotes a hazard. It calls attention to a procedure,
practice, or condition that, if not correctly performed or adhered to,
could result in damage to or destruction of part or all of the
instrument.
This Note sign denotes important information. It calls attention to a
procedure, practice, or condition that is essential for the user to
understand.
10Network Analyzer
Certification
Keysight Technologies certifies that this product met its
published specifications at the time of shipment from the
factory. Keysight Technologies further certifies that its
calibration measurements are traceable to the United States
National Institute of Standards and Technology, to the extent
allowed by the Institution’s calibration facility or by the
calibration facilities of other International Standards
Organization members.
Documentation Warranty
The material contained in this document is provided "as is,"
and is subject to being changed, without notice, in future
editions. Further, to the maximum extent permitted by
applicable law, Keysight disclaims all warranties, either
express or implied with regard to this manual and any
information contained herein, including but not limited to
the implied warranties of merchantability and fitness for a
particular purpose. Keysight shall not be liable for errors or
for incidental or consequential damages in connection with
the furnishing, use, or performance of this document or any
information contained herein. Should Keysight and the user
have a separate written agreement with warranty terms
covering the material in this document that conflict with
these terms, the warranty terms in the separate agreement
will control.
Exclusive Remedies
The remedies provided herein are Buyer’s sole and exclusive
remedies. Keysight Technologies shall not be liable for any
direct, indirect, special, incidental, or consequential damages,
whether based on contract, tort, or any other legal theory.
Assistance
Product maintenance agreements and other customer
assistance agreements are available for Keysight Technologies
products.
For any assistance, contact your nearest Keysight
Technologies Sales and Service Office. Addresses are
provided at the back of this manual.
Network Analyzer11
Manuals for E5061B
Keysight provides the following three manuals for E5061B.
Installation Guide
The installation guide (this manual) provides start up setup
information when you use the E5061B for the first time and
troubleshooting information when the Windows cannot be
boot up. See this manual first when you use the E5061B for
the first time.
Online Help
The online help provides the information about the quick
start, measurement operation, programming, built-in VBA,
I/O interface This is pre-installed in the E5061B. Press
[Help] hard key on the front panel to open. Quick Start
helps in understanding the E5061B operation quickly.
The latest version of online help is available at:
The online help has context sensitive help, which is a great
feature of the E5061B help. It allows you to get information
about the selected softkey by pressing the Help key in the
E5061B or by pressing F1 in a keyboard attached to the
E5061B or by clicking the help button in a dialog box. With
context sensitive help, users can receive information quickly
about the area the user is working in the firmware of the
E5061B. It provides information relevant to the task that
needs to be accomplished and reduces the time to search
relevant information required to complete a task.
Service Guide
The service manual provides information about the parts,
troubleshooting, performance test, adjustment and service
menu.
Network Analyzer13
In This Guide...
The following shows the contents of this manual.
Chapter 1, "Installation"
This chapter provides information about how to set up the
Keysight E5061B.
Chapter 2, "Troubleshooting"
This chapter describes the troubleshooting during start up
and the procedure of the operating system (OS) recovery
when the Windows XP OS has been damaged.
14Network Analyzer
Contents
Regulatory Compliance Information6
Safety Notice Supplement7
1Installation
Checking the Shipment18
Environmental Requirements20
Operating Environments20
Ventilation Requirement20
Protection Against Electrostatic Discharge (ESD)21
Ensuring Adequate Free Space around Analyzer for Immediate
Disconnection of Power Cable in Case of Emergency22
Installing Front Handles/Rack Mounting Flanges23
How to Install the Handle Kit (Option 1CN)24
How to Install the Rack-mount Kit (Option 1CM)25
How to Install the Rack-mount and Handle Kit (Option 1CP)25
Connecting the Accessories27
Connecting the BNC Adaptor (Option 1E5)27
Connecting Mouse and Keyboard27
Power Supply28
Verification of the Power Supply28
Verification and Connection of Power Cable28
Blown Fuses29
Starting the E5061B30
Turning the Power ON and OFF30
Disconnection from Supply Source31
Initial Registration of E5061B33
Calibration of the Touch Screen46
2Troubleshooting
Troubleshooting during Startup50
System Recovery51
Network Analyzer15
Notes on executing the factory recovery function51
Procedure to execute the factory recovery51
Checking the Shipment 18
Environmental Requirements 20
Installing Front Handles/Rack Mounting Flanges 23
Connecting the Accessories 27
Power Supply 28
Blown Fuses 29
Starting the E5061B 30
Initial Registration of E5061B 33
Calibration of the Touch Screen 46
This chapter provides information about how to set up the
Keysight E5061B.
1Installation
WARNING
NOTE
Checking the Shipment
After you receive the analyzer, carry out checks during
unpacking according to the following procedure.
When unpacking the analyzer, if the exterior of the analyzer (such
as the cover, front/rear panel, LCD screen, power switch, and port
connectors) appear to be damaged during transport, do not turn
on the power switch. Otherwise, you may get an electric shock.
Step 1. Check that the packing box or shock-absorbing
material used to package the analyzer has not been
damaged.
If the packing box or shock-absorbing material has been damaged, leave the
packing box and shock-absorbing material as is until other inspection items
are checked as follows.
Step 2. Check the packaged items supplied with the
analyzer for any damage or defects.
Step 3. By referring to the furnished contents list, check
that all packaged items supplied with the analyzer have been
received as per the specified options.
Step 4. After checking, if one of the following applies,
contact your nearest Keysight Technologies sales and service
office.
1 The packing box or shock-absorbing material used to
package the analyzer has been damaged or the
shock-absorbing material has traces where extreme force
has been applied.
2 A packaged item supplied with the analyzer has
mechanical damage or defects.
3 A packaged item supplied with the analyzer is missing.
4 A fault has been detected in the subsequent operation
check of the analyzer.
If an abnormality is detected in step 1, contact the
company that transported the analyzer as well as your
nearest Keysight Technologies sales and service office. For
18Network Analyzer
Installation1
inspection by the transport company, save the packing
box, shock-absorbing material, and packaged items as you
received them.
Network Analyzer19
1Installation
NOTE
Environmental Requirements
Set up the E5061B where the following environmental
requirements are met.
Operating Environments
Ensure that the operating environment meets the following
requirements.
Temperature 5 °C to 40 °C
Temperature range at
the error-correction
Humidity20% to 80% at wet bulb temperature < +29 °C
Altitude0 to 2,000 m (0 to 6,561 feet)
Vibration0.21 G maximum, 5 Hz to 500 Hz
Above environmental requirements are NOT for the specifications and
measurement accuracy of the analyzer, but for the operating environment of
the analyzer.
Ventilation Requirement
To ensure safety requirements, the specifications and
measurement accuracy of the analyzer, you must keep the
environmental temperature within the specified range by
providing appropriate cooling clearance around the analyzer
or, for the rack mount type, by forcefully air-cooling inside
the rack housing. For more information on environmental
temperature to satisfy the specifications and measurement
accuracy of the analyzer, see the specification in the E5061B Online Help.
23 °C ± 5 °C (< 1 °C deviation from the temperature
when performing the error-correction)
(non-condensation)
When the environmental temperature around the analyzer is
kept within the temperature range of the operating
environment specification (See Section “Operating
Environments" on page 20), the analyzer conforms to the
requirements of the safety standard. Furthermore, under that
temperature requirement, the analyzer still conforms to the
requirements of the safety standard even when placing the
analyzer with cooling clearance as follows:
20Network Analyzer
Requirements
NOTE
Back≤ 180 mm
Sides≤ 60mm (both right and left)
Installation1
OO
50
Avoid Static
1M! +26 dBm RF
42 V DC Max CAT
50! +26 dBm RF
7 V DC Max
ELECTROSTATIC DISCHARGE
OR EXCESSIVE INPUT VOLTAGE
DC Bias
1M
!
/50
!
MAY DAMAGE CIRCUITS.
Ι
CAT
Ι
±42V Peak Max Output
OO
e5061b060
Figure 1Ventilation space at the installation
Place the E5061B in a proper position as show in Figure 1.
Protection Against Electrostatic Discharge (ESD)
OO
DC Bias
!
+20 dBm RF 7 V DC Max CAT
50
!
Ι
Set up a static-safe work-station to protect the electronic
components from the damage by the electrostatic discharge
(ESD) as shown in Figure 2.
Network Analyzer21
1Installation
Figure 2Example of the static-safe work station
Ensuring Adequate Free Space around Analyzer for Immediate
Disconnection of Power Cable in Case of Emergency
As described in “Disconnection from Supply Source" on page 31,
the power supply is disconnected by removing the power cable’s
connector plug from either the AC outlet or the E5061B unit.
When installing the E5061B, ensure that there is sufficient free
space around the unit to permit quick disconnection of the plug
(from AC outlet or E5061B unit) in case of emergency.
22Network Analyzer
Installing Front Handles/Rack Mounting Flanges
The E5061B can be installed on a workbench or in a rack.
This section describes how to install the front handles
(Option 1CN) used for moving or transporting the
instrument, and how to install the analyzer in an equipment
rack as part of a measurement system (Option 1CM: without
the handles, Option 1CP: with the handles).
Tab l e 1Keysight E5061B handles/rack mounting options
OptionNameKeysight Part Number
1CNHandle Kit5063-9229
1CMRack-mount Kit5063-9216
1CPRack-mount and Handle Kit5188-4430
Installation1
Tab l e 2Contents of each option
OptionContentsQuantity
1CNFront Handles2
Screws6
Trim S t rips2
1CMRack-mounting flanges (locking side plate)2
Screws6
1CPRack-mounting flanges (locking side plate)2
Front Handles2
Screws8
Network Analyzer23
1Installation
Figure 3Installing front handle/rack-mount kits
How to Install the Handle Kit (Option 1CN)
The handle kit is used for transport and relocation of the
E5061B. While referring to Figure 3 install the handle kit by
following these steps.
Step 1. Remove the adhesive-backed trim strip (1) from
each side of the outer frame of the E5061B front panel.
Step 2. Use the screws provided screws to mount the front
handles (2) on each side of the E5061B front panel frame.
Step 3. Attach the modified trim strip (3) provided to each
front handle in order to cover the front panel locking
screws.
24Network Analyzer
WARNING
If the installed front handle becomes damaged, replace it with a
CAUTION
new one immediately. A damaged handle can break while moving
or lifting the instrument and cause personal injury or damage to
the instrument.
How to Install the Rack-mount Kit (Option 1CM)
The rack-mount kit includes two flanges (locking side plates)
for mounting the E5061B on a rack (482.6 mm/19 inches)
conforming to the EIA Standard. While referring to Figure 3,
install the rack-mount kit by following these steps.
Step 1. Remove the adhesive-backed trim strip (1) from
each side of the outer frame of the E5061B front panel.
Step 2. Use the screws provided to mount a rack-mounting
flange (4) on each side of the E5061B front panel frame.
Installation1
Step 3. Remove the four bottom feet of the E5061B (lift the
bar marked TAB on the inner side of the foot and slide the
foot toward the bar).
Step 4. Mount the E5061B on the rack.
How to Install the Rack-mount and Handle Kit (Option 1CP)
The rack-mount and handle kit includes both the
rack-mounting flanges (locking side plates) and front
handles. While referring to Figure 3, install the rack-mount kit
by following these steps.
Step 1. Remove the adhesive-backed trim strip (1) from
each side of the outer frame of the E5061B front panel.
Step 2. Use the screws provided to mount a front handle
(2) and rack-mounting flange (4) on each side of the E5061B
front panel frame.
Use both a front handle and a rack-mounting flange in the same time
certainly. Do not attempt to install flanges or handles separately
with hardware provided. Serious electrical damage will occur to the
instrument.
Network Analyzer25
1Installation
Step 3. Remove the four bottom feet of the E5061B (lift the
bar marked TAB on the inner side of the foot and slide the
foot toward the bar).
Step 4. Mount the E5061B on the rack.
26Network Analyzer
Connecting the Accessories
BNC adaptor
You need to connect some accessories before use.
Connecting the BNC Adaptor (Option 1E5)
When the E5061B is equipped with Option 1E5
(high-stability frequency standard), connect the BNC adaptor
from the Ref Oven terminal to the Ref In terminal on the
rear panel as shown in Figure 4. The BNC adaptor is included in
Option 1E5. When Option 1E5 is installed, the frequency
accuracy and stability of the E5061B improves further.
Installation1
TEST SET I/F
GPIB
VIDEO
e5061b059
Figure 4Connecting the BNC adaptor
Connecting Mouse and Keyboard
The E5061B allows you to connect Mouse and/or keyboard
through USB. USB mouse and keyboard can be connected
with the USB ports on the front or rear panels.
Initial registration of the E5061B requires the mouse and
keyboard before turning on the power.
WARNING:
Serial Label
No operator serviceable parts inside.
Refer servicing to qualified personnel.
E5061B
ICES/NMB-001
4
10 MHz
Windows Label
115 V
50/60 Hz 300 VA Max
LINE
Network Analyzer27
1Installation
WARNING
WARNING
Power Supply
Verification of the Power Supply
Before turning on the E5061B power, check the following.
Confirm that the power supplied to the E5061B meets the
following requirements:
Requirements
Voltage90 to 132 VAC or 198 to 264 VAC
Frequency47 to 63 Hz
*1
Maximum power
consumption
* 1. Switched automatically by the E5061B in conformity to the voltage used.
300 VA
Verification and Connection of Power Cable
The three-wire power cable attached to the E5061B has one
wire serving as a ground. Using this power cable allows the
E5061B to be grounded, thereby protecting you against
electrical shock from the power outlet.
Step 1. Confirm that the power cable is not damaged.
NEVER use a power cable showing any sign of damage. Faulty
cables can cause electrical shock.
Step 2. Use the supplied cable to connect between the
power cable receptacle (Figure 5 on page 31) on the rear panel
of the E5061B and a three-wire power outlet with the grounding
prong firmly connected in the ground slot.
Use the supplied power cable with grounding wire to securely
ground the E5061B.
Power cord list, 16000-99101 shows the power cable options.
28Network Analyzer
Blown Fuses
WARNING
Installation1
If the fuse appears to have blown during operation, this
instrument may be subject to failure and must be repaired.
For any assistance, contact Keysight Technologies Customer
contact center listed at the end of this guide.
The product uses the following fuse types:
UL/CSA Type, Slow-Blo, 10 A-250 Vac.
DO NOT replace the fuse yourself; doing this may expose you to
electrical shock.
Network Analyzer29
1Installation
Starting the E5061B
Turning the Power ON and OFF
This section describes how to turn on/off the E5061B power
and how to cut off the power supply in an emergency.
The standby switch can turn/off the E5061B. The color on
the button shows the status, as shown below:
Indicator ColorDescription
GreenNormal power on status
OrangeStandby status
RedIllegal power on status
Perform the following steps to turn the power ON or OFF.
Turning the Power ON
Step 1. Confirm if the Line Switch on the rear panel is on.
The switch should always be turned on.
Step 2. Press the standby switch once and it changes to
green color.
This operation turns ON the power, and the E5061B starts
the self-test.
Step 3. Confirm that the self-test indicates normal
operation.
Normal operation is confirmed by the self-test if no error
message appears.
Turning the Power OFF
Step 1. To turn off the power of the E5061B, first, press
this standby switch or send a shutdown command from the
external controller to activate the shutdown process (the
process of software and hardware necessary to turn off the
power supply). This puts the E5061B into the standby state
and the button changes to orange color.
30Network Analyzer
Installation1
NOTE
Step 2. Next, if necessary, turn off power supply to the
power cable receptacle (Line Switch) on the rear panel.
Under normal use, never directly interrupt the power supply to the power
cable receptacle on the rear panel when the power supply is on. Always
keep the Line Switch ON and never turn it.
If you directly interrupt the power supply to the power cable receptacle
when the power supply is on, or turn off the Line Switch, the shutdown
process will not work. This could damage the software and hardware of
the E5061B and lead to device failure.
Turning on the power supply after a faulty shutdown may cause the
system to start up in a condition called "safe mode." If this occurs, first
shut down the system to set it to the standby state and then turn on the
power supply again to start up the system in normal mode.
GPIB
e5061b061
Figure 5Line switch (Always ON) and power cable receptacle
Disconnection from Supply Source
The power supply of the E5061B is cut off by disconnecting
the plug of the power cable (on either AC outlet side or
E5061B side). When it is necessary to disconnect the power
supply in order to avoid shock hazards, etc., pull out the
power cable plug from either the AC outlet side or the
E5061B side.
TEST SET I/F
Power Cable Receptacle
VIDEO
Serial Label
E5061B
10 MHz
Line Switch: Always ON ( | )
115 V
50/60 Hz 300 VA Max
LINE
4
Windows Label
WARNING:
No operator serviceable parts inside.
Refer servicing to qualified personnel.
ICES/NMB-001
Network Analyzer31
1Installation
NOTE
To allow this operation to be performed smoothly, be sure to follow the
guidelines in “Ensuring Adequate Free Space around Analyzer for
Immediate Disconnection of Power Cable in Case of Emergency" on
page 22.
When turning the power OFF under normal circumstances, always follow
the methods described in “Turning the Power OFF" on page 30.
32Network Analyzer
Initial Registration of E5061B
NOTE
NOTE
When you start up the E5061B for the first time, you need
to perform the initial registration of the Windows XP
operating system of the E5061B.
There are two kind of windows licenses for E5061B:
- Windows Vista Business
- Windows XP Pro for Embedded Systems
The license is shown at the Windows Lincese label on the E5061B rear
panel.
You have different initial registration procedure depending on the windows
license.
As for Windows XP Pro for Embedded Systems, you have different initial
registration procedure depending on the hard disk drive (HDD revision).
Installation1
If you perform the following procedure incorrectly, a message will prompt
to give you option to return to the previous registration screen and perform
the registration. In such case, follow the instruction to return to the
previous registration screen.
Network Analyzer33
1Installation
NOTE
Windows XP Pro for Embedded Systems (HDD Rev KY22x and
above)
Step 1. Turn on the E5061B. Do not touch any key on the
front panel during boot up.
Step 2. The screen as shown in the Figure 6 appears. Select
Windows XP Professional and press [Enter].
Figure 6System start up screen in Windows XP Pro for Embedded
Systems (1/2)
After several seconds, the next screen appears automatically even if you
do not press any key.
Step 3. The screen as shown in the Figure 7 appears. Press
[Enter].
34Network Analyzer
Installation1
Figure 7System start up screen in Windows XP Pro for Embedded
Systems (2/2)
Step 4. The windows start up screen as shown in Figure 8
appears.
Network Analyzer35
1Installation
Figure 8Windows XP Pro for Embedded Systems start up screen
Step 5. In the Keysight Tech. Software End User License
Agreement dialog box, select the Agree button. Figure 9 shows
the license agreement dialog box in Windows XP Pro for
Embedded Systems (for HDD Rev KY22x and above).
36Network Analyzer
Installation1
Figure 9Windows XP Pro for Embedded Systems license agreement
dialog box
Step 6. After a while, the E5061B restarts automatically.
Windows XP Pro for Embedded Systems (HDD Rev KY21x and
below)
Step 1-4. Follow the steps of Windows XP Pro for
Embedded Systems (HDD Rev KY22x and above).
Step 5. In the License agreement dialog box, select the I accept
this agreement box and click the Next >
button. Figure 10
shows the license agreement dialog box in Windows XP Pro for
Embedded Systems (for HDD Rev KY21x and below).
Network Analyzer37
1Installation
Figure 10 Windows XP Pro for Embedded Systems license agreement
dialog box
38Network Analyzer
Installation1
Step 6. In the next dialog box, set the date and time for
your Windows. Then, click the Next> button (Figure 11).
Figure 11 Date and Time Setting dialog box
Step 7. After a while, the E5061B restarts automatically.
Windows Vista Business
Step 1. Follow the steps of Windows XP Pro for Embedded
Systems (HDD Rev KY22x and above).
Network Analyzer39
1Installation
NOTE
Step 2. The screen as shown in the Figure 12 appears. Select
Microsoft Windows XP Professional and press [Enter].
Figure 12 System start up screen in Windows Vista Business (1/2)
After several seconds, the next screen appears automatically even if you
do not press any key.
Step 3. The screen as shown in the Figure 13 appears. Press
[Enter].
40Network Analyzer
Installation1
Figure 13 System start up screen in Windows Vista Business (2/2)
Step 4. The Windows XP Professional Setup wizard appears.
Click the Next> button (Figure 14).
Network Analyzer41
1Installation
Figure 14 Windows XP Professional Setup wizard
42Network Analyzer
Installation1
Step 5. In the License agreement dialog box, select the I
accept this agreement box and click the Next >
button.
Figure 15 shows the license agreement dialog box in Windows
Vista Business.
Figure 15 Windows Vista Business license agreement dialog box
Network Analyzer43
1Installation
Step 6. In the next dialog box, set the date and time for
your Windows. Then, click the Next> button (Figure 16).
Figure 16 Date and Time Setting dialog box
Step 7. After a while, the E5061B restarts automatically.
The Keysight Recovery Image Setup dialog box as shown in
the Figure 17 appears. Click the OK button.
Figure 17 Keysight Recovery Image Setup
44Network Analyzer
Installation1
Step 8. Once the recovery image setup is completed, the
E5061B restarts again.
Network Analyzer45
1Installation
Calibration of the Touch Screen
When E5061B measurement screen appears, you have to
calibrate the touch screen. Follow the procedure described
below to calibrate the touch screen.
Step 1.Press .
Step 2. Press Service Menu.
Step 3. Press Test Menu.
Step 4. Press Adjust Touch Screen.
The touch screen calibration screen (Figure 18) appears.
Figure 18 Touch Panel Calibration Screen
Step 5. Touch the symbol on the upper left of the screen
with your finger. The symbols also appears on the upper
right, lower left, and lower right of the screen. Touch all the
symbols in that order with your finger.
Touching all the four locations described above with your
finger automatically concludes the touch screen calibration.
46Network Analyzer
Installation1
NOTE
If you do not start to perform the touch screen calibration within several
seconds, the previous measurement screen re-appears automatically.
Troubleshooting during Startup 50
System Recovery 51
This chapter describes the troubleshooting process during
start up and the procedure of the operating system (OS)
recovery when the Windows XP OS has been damaged.
2Troubleshooting
Troubleshooting during Startup
When you encounter problems during start up, see Table 3.
System recovery saves most of problems.
Tab l e 3Troubleshooting during startup
SymptomSolution
Turning on (|) the standby switch does
not start up the system.
The system starts up, but it automatically
shuts down immediately.
The system starts up, but it enters the
service mode (The status bar at the lower
right of the screen displays SVC in red).
The measurement screen appears after
start up, but the power-on test fails
with Error Message 241 prompting on a
red background in the instrument’s
message/warning area at the lower left
of the screen.
Confirm that the power cable is
properly plugged in.
Confirm that the line switch at the
rear panel is turned on.
Execute the system recovery.
50Network Analyzer
System Recovery
CAUTION
NOTE
Notes on executing the factory recovery function
Troubleshooting2
By executing system recovery, you can return the E5061B’s
system (the Windows operating system and the firmware) to
the factory state (at the time of purchase
*1
).
Strictly follow the steps described below. If you do any operation
other than the following steps, the system may not be recovered.
Executing the factory recovery function causes the following
conditions:
• In addition to the Windows operating system and the
firmware, the following settings of the E5061B are
returned to the factory state.
• Network setting
• GPIB setting
• Printer setting
• The driver for the supported printer which is installed
after purchase is deleted.
• You need to execute initial registration again.
Files that you have created using the save function (files in
the D drive) are not affected. However, Keysight recommends
backing them up before executing system recovery for
precautionary purposes. For more information on backup,
refer to “Backing Up the Data” as described in E5061B Online Help.
Procedure to execute the factory recovery
This section describes how to return the contents of the C
drive to the factory state.
You need a keyboard for this operation.
* 1. The default setting of the hard disk is recovered during the system recovery. For the
hard disk that had failed earlier and had been replaced (after purchase), its default setting
may not be the same as the factory state.
Network Analyzer51
2Troubleshooting
NOTE
Step 1. Shut down the E5061B.
Step 2. Disconnect all of the USB device from the USB
ports.
Step 3. Connect the keyboard to the E5061B.
Step 4. Press the standby switch of the E5061B to turn it
on.
Windows Vista Business
Step 5. When the screen as shown in the figure below
appears, select Keysight Recovery System and press
[Enter].
Figure 19 Keysight Recovery System selection screen
After several seconds, the next screen appears automatically even if you
do not press any key.
52Network Analyzer
Troubleshooting2
Step 6. Type 2 to select Recover Factory Backup Image
option and press [Enter].
Figure 20 System recovery selection screen
Step 7. A confirmation message appears. Type 1 and press
[Enter] to continue the process.
Figure 21 System recovery confirmation screen (1/2)
Network Analyzer53
2Troubleshooting
NOTE
Step 8. Another confirmation message appears. It also
displays the approximate run time for the system
configuration process. Type 1 and press [Enter] to continue
the process. The recovery takes a few minutes depending on
the amount of data.
Figure 22 System recovery confirmation screen (2/2)
Step 9. Once the recovery process is completed, completion
message prompts. Press any key to continue. The E5061B
restarts automatically.
Figure 23 Recovery Option confirmation dialog box
After several seconds, the E5061B restarts automatically even if you do
not press any key.
54Network Analyzer
Troubleshooting2
NOTE
Windows XP Pro for Embedded Systems
Step 5. When the screen as shown in the figure below
appears, select Keysight Recovery System and press
[Enter].
Figure 24 Keysight Recovery System selection screen
After several seconds, the next screen appears automatically even if you
do not press any key.
Step 6. Type 2 to select Recover Factory Backup Image
option and press [Enter].
Figure 25 System recovery selection screen
Network Analyzer55
2Troubleshooting
Step 7. A confirmation message appears. Click OK to
continue the process.
Figure 26 System recovery confirmation screen
Step 8. The progress of the system configuration is
displayed on the screen. The recovery takes a few minutes
depending on the amount of data.
Figure 27 System configuration progress screen
56Network Analyzer
Troubleshooting2
CAUTION
Step 9. Once the recovery process is completed, completion
message prompts. Click OK. The E5061B restarts
automatically.
Figure 28 Recovery Option confirmation dialog box
Step 10. Type 5 to select Exit and Restart the instrument
and press [Enter].
Step 11. After the restart, execute initial registration. For
information on the execution procedure, refer to “Initial
Registration of E5061B" on page 33.
Step 12. Execute the calibration of the touch screen. For
information on the execution procedure, refer to “Calibration of
the Touch Screen" on page 46.
Never turn off the power during the system recovery because doing
so may cause serious damage to the E5061B.
Network Analyzer57
2Troubleshooting
58Network Analyzer
Index
A
accessories
connecting accessories,27
assistance,11
C
connection
connecting accessories,27
connecting BNC adaptor,27
connecting mouse and keyboard,27