No part of this manual may be reproduced
in any form or by any means (including
electronic storage and retrieval or translation into a foreign language) without prior
agreement and written consent from Agilent Technologies, Inc. as governed by
United States and international copyright
laws.
Manual Part Number
E4418-90032
Edition
Eighth Edition, April 5, 2013
Printed in Malaysia
Agilent Technologies, Inc.
3501 Stevens Creek Blvd.
Santa Clara, CA 95052 USA
Warranty
The material contained in this document is provided “as is,” and is subject to being changed, without notice,
in future editions. Further, to the maximum extent permitted by applicable
law, Agilent disclaims all warranties,
either express or implied, with regard
to this manual and any information
contained herein, including but not
limited to the implied warranties of
merchantability and fitness for a particular purpose. Agilent shall not be
liable for errors or for incidental or
consequential damages in connection with the furnishing, use, or performance of this document or of any
information contained herein. Should
Agilent and the user have a separate
written agreement with warranty
terms covering the material in this
document that conflict with these
terms, the warranty terms in the separate agreement shall control.
Technology Licenses
The hardware and/or software described in
this document are furnished under a
license and may be used or copied only in
accordance with the terms of such license.
Restricted Rights Legend
U.S. Government Restricted Rights. Software and technical data rights granted to
the federal government include only those
rights customarily provided to end user customers. Agilent provides this customary
commercial license in Software and technical data pursuant to FAR 12.211 (Technical
Data) and 12.212 (Computer Software) and,
for the Department of Defense, DFARS
252.227-7015 (Technical Data - Commercial
Items) and DFARS 227.7202-3 (Rights in
Commercial Computer Software or Computer Software Documentation).
Safety Notices
A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like
that, if not correctly performed or
adhered to, could result in damage
to the product or loss of important
data. Do not proceed beyond a
CAUTION notice until the indicated conditions are fully understood and met.
A WARNING notice denotes a
hazard. It calls attention to an
operating procedure, practice, or
the like that, if not correctly performed or adhered to, could result
in personal injury or death. Do not
proceed beyond a WARNING
notice until the indicated conditions are fully understood and
met.
ii E4418B Power Meter User’s Guide
Certification
Warranty
Agilent Technologies certifies that this product met its published
specifications at the time of shipment from the factory. Agilent
Technologies further certifies that its calibration measurements are
traceable to the United States National Institute of Standards and
Technology, to the extent allowed by the Institute’s calibration facility and
to the calibration facilities of other International Standards Organization
members.
This Agilent Technologies instrument product is warranted against defects
in material and workmanship for a period of 3 years from date of shipment.
During the warranty period, Agilent Technologies will at its option, either
repair or replace products which prove to be defective. For warranty
service or repair, this product must be returned to a service facility
designated by Agilent Technologies. Buyer shall prepay shipping charges
to Agilent Technologies and Agilent Technologies shall pay shipping
charges, duties and taxes for products returned to Aglent Technologies
from another country. Agilent Technologies warrants that its software and
firmware designated by Agilent Technologies for use with an instrument
will execute its programming instructions when properly installed on that
instrument. Agilent Technologies does not warrant that the operation of
the instrument or firmware will be uninterrupted or error free.
Limitation of Warranty
The foregoing warranty shall not apply to defects resulting from improper
or inadequate maintenance by Buyer, Buyer-supplied software or
interfacing, unauthorized modification or misuse, operation outside of the
environmental specifications for the product or improper site preparation
or maintenance. NO OTHER WARRANTY IS EXPRESSED OR IMPLIED.
AGILENT TECHNOLOGIES SPECIFICALLY DISCLAIMS THE IMPLIED
WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR
PURPOSE.
E4418B Power Meter User’s Guide iii
Exclusive Remedies
WARNING
CAUTION
Warnings and Cautions
THE REMEDIES PROVIDED HEREIN ARE BUYER’S SOLE AND EXCLUSIVE
REMEDIES. AGILENT TECHNOLOGIES SHALL NOT BE LIABLE FOR ANY
DIRECT, INDIRECT, SPECIAL, INCIDENTAL or CONSEQUENTIAL
DAMAGES, WHETHER BASED ON CONTRACT, TORT or ANY OTHER
LEGAL THEORY.
This guide uses warnings and cautions to denote hazards.
A warning calls attention to a procedure, practice or the like, which, if
not correctly performed or adhered to, could result in injury or the loss of
life. Do not proceed beyond a warning until the indicated conditions are
fully understood and met.
A caution calls attention to a procedure, practice or the like which, if not
correctly performed or adhered to, could result in damage to or the
destruction of part or all of the equipment. Do not proceed beyond a
caution until the indicated conditions are fully understood and met.
Recommended Calibration Interval
iv E4418B Power Meter User’s Guide
Agilent Technologies recommends a two- years calibration cycle for the
E4418B EPM Series power meter.
Safety Symbols
The following symbol on the instrument and in the documentation
indicates precautions that must be taken to maintain safe operation of the
instrument.
Caution, risk of danger.
The
Instruction Documentation Symbol. The product is marked with this
symbol when it is necessary for the user to refer to the instructions in the
supplied documentation.
Alternating current (AC).
This symbol indicates the operating switch for ‘Stand-by’ mode. Note, the
strument is NOT isolated from the mains when the switch is pressed.
in
To isolate the instrument, the mains coupler (mains input cord) should be
removed from the power supply.
This symbol indicates the operating switch for ‘On’ mode).
External Protective Earth
While this is a Class I product, provided with a protective earthing
conductor in a power cord, an external protective earthing terminal has
also been provided. This terminal is for use where the earthing cannot be
assured. At least an 18AWG earthing conductor should be used in such
an instance, to ground the instrument to an assured earth terminal.
E4418B Power Meter User’s Guide v
Terminal.
Personal Safety Considerations
WARNING
WARNING
CAUTION
This is a Safety Class I product (provided with a protective earthing ground
incorporated in the power cord). The mains plug shall only be inserted in a
socket outlet provided with a protective earth contact. Any interruption of
the protective conductor, inside or outside the instrument is likely to make
the instrument dangerous. Intentional interruption is prohibited
If this instrument is not used as specified, the protection provided by the
equipment could be impaired. This instrument must be used in a normal
condition (in which all means of protection are intact) only.
No operator serviceable parts inside. Refer servicing to qualified
personnel. To prevent electrical shock, do not remove covers.
For continued protection against fire hazard, replace the line fuse(s)
only with fuses of the same type and rating (for example, normal blow,
time delay, etc.). The use of other fuses or material is prohibited.
General Safety Considerations
Before this instrument is switched on, make sure it has been properly
grounded through the protective conductor of the ac power cable to a
socket outlet provided with protective earth contact.
Any interruption of the protective (grounding) conductor, inside or
outside the instrument or disconnection of the protective earth terminal
can result in personal injury.
vi E4418B Power Meter User’s Guide
Any adjustments or service procedures that require operation of the
instrument with protective covers removed should be performed only by
trained service personnel.
Regulatory Markings
The CE mark shows that the product complies with all the relevant
European legal Directives (if accompanied by a year, it signifies when the
design was proven.
This ISM device complies with Canadian ICES-001.
t appareil ISM est conforme à la norme NMB-001 du Canada.
Ce
This is the symbol of an Industrial Scientific and Medical Group 1 Class A
t.
produc
The CSA mark is a registered trademark of the Canadian Standards
ation.
Associ
E4418B Power Meter User’s Guide vii
IEC 1010-1 Compliance
This instrument has been designed and tested in accordance with IEC
Publication 61010-1 +A1:1992 Safety Requirements for Electrical
Equipment for Measurement, Control and Laboratory Use and has been
supplied in a safe condition. The instruction documentation contains
information and warnings which must be followed by the user to ensure
safe operation and to maintain the instrument in a safe condition.
Statement of Compliance
This product has been designed and tested for compliance with IEC 60529
(1989) Degrees of Protection Provided by Enclosures (IP Code). Level IPx4
is attained if, and only if, the carry case( Agilent Technologies part number
34141A) is fitted.
User Environment
This product is designed for use in a sheltered environment (avoiding
extreme weather conditions) in accordance with Pollution Degree 3
defined in IEC 60664-1, with the carry case (Agilent Technologies part
number 34141A) fitted over the instrument.
The product is suitable for indoor use on
fitted.
ly, when this carry case is not
Installation Instructions
To avoid unnecessary over-temperature conditions, while this carry case is
fitted do not apply an ac mains supply voltage, only operate your Agilent
Technologies E4418B from the battery pack.
viii E4418B Power Meter User’s Guide
Regulatory Information
N279
Sound Emission
Herstellerbescheinigung
Diese Information steht im Zusammenhang mit den Anforderungen der
Maschinenlarminformationsverordnung vom 18 Januar 1991.
• Sound Pressure LpA < 70 dB.
• Am Arbeitsplatz.
• Normaler Betrieb.
• Nach DIN 45635 T. 19 (Typprufung).
Manufacturers Declaration
This statement is provided to comply with the requirements of the German
Sound DIN 45635 T. 19 (Typprufung).
• Sound Pressure LpA < 70 dB.
• At operator position.
• Normal operation.
• According to ISO 7779 (Type Test).
Australian EMC Regulations
The C-Tick mark is a registered trademark of the Spectrum Management
Agency of Australia. This signifies compliance with the Australian EMC
Framework Regulations under the terms of the Radiocommunications Act
of 1992.
E4418B Power Meter User’s Guide ix
Declaration of Conformity (DoC)
NOTE
The Declaration of Conformity (DoC) for this instrument is available on the
Web site. You can search the DoC by its product model or description.
http://regulations.corporate.a
If you are unable to search for the respective DoC, please contact your
local Agilent representative.
gilent.com/DoC/search.htm
x E4418B Power Meter User’s Guide
List of Related Publications
The Agilent E4418B is also available in the following
languages:
• English Language User’s Guide - Option ABA
• German Language User’s Guide - Option ABD
• Spanish Language User’s Guide - Option ABE
• French Language User’s Guide - Option ABF
• Italian Language User’s Guide - Option ABZ
• Japanese Language User’s Guide - Option ABJ
Agilent Technologies E4418B/E4419B Programming Guide (Engli
language only) is also supplied with the above options except Option
ABA. For Option ABA, the Programming Guide is available only by
ordering Option 0BF.
Agilent Technologies E4418B/E4419B Service Guide is av
ordering Option 915.
Agilent Technologies E4418B/E4419B CLIPs (Component Location and
formation Pack) is available by ordering E4418-90031.
In
Useful information on SCPI (Standard Commands for
Instruments) can be found in:
Beginner’s Guide to SCPI, which is available by ordering Agilent
• A
Technologies Part Number 5010-7166.
• The SCPI reference manuals which are available from:
SCPI Consortium,
8380 Hercules Drive, Suite P3,
La Mesa, CA 91942, USA.
Telephone: 619-697-4301
Fax: 619-697-5955
Programmable
sh
ailable by
E4418B Power Meter User’s Guide xi
Agilent Technologies E4418B Options
The Agilent E4418B power meters have the following options available:
• Option 001, supplies an internal rec
instrument functionality when access to an ac power outlet is not
available.
• Option 002, supplies parallel rear p
reference oscillator output is on the front panel.
• Option 003, supplies parallel rear p
reference oscillator output is also on the rear panel.
• Option 004, deletes the Agilent 11730
• Option 0BO, deletes manual set.
• Option 908, provides rackmount kit for one instrument.
• Option 909, provides rackmount kit for two instruments.
• Option 915, provides the Agilent E4418B/E4419B Service Guide.
Figure 2-20 Test Setup for Recording Swept Measurements 86
Figure 2-21
Figure 2-22 How Measurements are Calculated 90
Figure 2-23 Re
Figure 4-1
Battery Removal/Replacement 28
“Sensor Tbls” Screen 45
“Edit” Screen 56
Pulsed Signal 68
Limits Checking Results 71
Pass/Fail Limit Indicators 75
Analog Display 76
Digital and Analog Display 76
Interface Overview Examples 84
“Save/Recall” Screen 89
placing the fuse 102
Error Annunciator Position 146
E4418B Power Meter User’s Guide xxi
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.
xxii E4418B Power Meter User’s Guide
List of Tables
Ta b l e 1-1 Relevant Softkeys in Various Mode of Operation 10
Ta b l e 1-2
Ta b l e 2-1
Ta b l e 2-2 TTL Inputs Control Logic 35
Ta b l e 2-3 TTL Input Timing Diagram 1 36
Ta b l e 2-4 TTL Inputs Timing Diagram 2 37
Ta b l e 2-5 Measurement Units 59
Ta b l e 2-6 Range of Values for Window Limits 72
Ta b l e 3-1 Combination of Sensor Type and Correction Applied to Current Measurement 121
Ta b l e 5-1 Zero Set Specifications 158
Ta b l e 5-2 Noise Multiplier 162
Ta b l e 5-3 Power Sensor Specifications± 162
Ta b l e 5-4 Settling Time 165
Ta b l e 5-5 Settling Time 166
Ta b l e 5-6 Settling Time 168
List of Possible Combinations for the Upper and Lower Display Lines 10
Power Sensor Connection Requirements 33
E4418B Power Meter User’s Guide xxiii
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.
xxiv E4418B Power Meter User’s Guide
E4418B Power Meter
User’s Guide
1
Getting Started
Introduction 2
Turning On the Power Meter 3
The Front Panel at a Glance 5
The Display Layout 8
Window Symbols 14
The Rear Panel at a Glance 16
Adjusting the Carrying Handle 18
Rack Mounting the Power Meter 19
Agilent Technologies
1
1Getting Started
NOTE
Introduction
One of the first things you will want to do with your power meter is to switch
it on and become acquainted with its front panel. The sections in this chapter
prepare the power meter for use and help you get familiar with some of the
front panel operations.
The front panel consists of both keys and softkeys which allow you to select
us functions and operations. When some keys are selected, the
vario
corresponding softkey labels are displayed on the power meter display.
If you are using the power meter remotely, refer to the AgTechnologies E4418B/4419B Programming Guide
details.
The N8480 Series power sensors used in this user’s guide is refering to all sensors from
the N8480 Series unless otherwise stated.
for remote operating
ilent
2E4418B Power Meter User’s Guide
Turning On the Power Meter
CAUTION
CAUTION
NOTE
The following steps show you how to turn on the power meter and verify that it
is operating correctly.
1 C
onnect the power cord and turn on the power meter.
The front panel display and the green power LED light up when the power
met
er is switched on. The power meter performs its power on self test. If
the self test is not successful, the error annunciator turns on. If this occurs,
contact your Agilent Technologies Sales and Service office for instructions
on returning the power meter to Agilent Technologies for service.
This instrument is designed for use in Installation Category II and Pollution Degree 2 per
IEC 1010 and 664 respectively.
This instrument has an autoranging line voltage input, be sure the supply voltage is within
the range of 85 to 264 Vac.
Getting Started1
E4418B Power Meter User’s Guide3
If the power meter has been stored at extremely low temperatures out of the operating
range of the power meter, the display may take a few minutes to operate.
1Getting Started
2 Set t
3 Connect
4 M
he display contrast if required.
The display contrast is adjusted by pressing and . If these
softkeys are not displayed, press repeatedly until they appear.
a power sensor.
Connect one end of the sensor cable to the power meter’s channel input and
th
e other end to the power sensor.
aking a measurement.
A minimum warm- up time of 30 minutes is recommended before accurate
measurements can
Prior to making your first measurement, you must zero and calibrate the
sen
sor and meter combination. Refer to Chapter 2 for further
information if you are not familiar wit
measurements with a power meter.
be made.
h zeroing, calibrating or making
4E4418B Power Meter User’s Guide
The Front Panel at a Glance
1
2
3
4
6
7
8
9101112
5
Getting Started1
1
This key allows you to preset the power meter if you are currently working
in
local mode (that is, front panel operation). In local mode, a confirmation
pop- up window is displayed prior to a preset being carried out. However, if
you are in remote mode (that is, GPIB, RS232, or RS422 operation), then
pressing this key places the power meter in local mode provided local lock
out (LLO) is not enabled.
eys relating to the display layout.
2 K
This key allows you to select the upper or lower measurement
E4418B Power Meter User’s Guide5
window on the power meter’s display. The
highlighted by a shadowed box. Any measurement setup you create is
implemented in the selected window.
This key allows you to choose either a one or a two window display.
window which is selected is
1Getting Started
3
This key switches the power meter between on and standby. When the
wer meter is switched to standby (that is, when this key has not been
po
selected but the line power is connected to the instrument) the red LED is
lit. When the power meter is switched on, the green LED is lit.
Option 001 Battery: On standby with the battery installed and the AC
wer source disconnected, the red LED is off.
po
4 “Syst
5
6 Dedic
7 Dedicat
8 Channel Input.
9 PO
em/Inputs” key with softkey menu.
The key allows access to softkey menus which affect the general
power meter system setup, (for example t
softkey menus which effect the setup of the channel inputs. Refer to
Chapter 3 for further information about this key and its softkey menu.
This key is the only one that is completely dedicated to the control of the
po
wer meter as a system. The only other key which affects system
parameters is the key. Refer to Chapter 3 for further
information about this key and its softkey menu.
ated “Window” keys with softkey menus.
, ,
These keys allow access to softkey menus which affect the setup of the
measurement windo
about these keys and their softkey menus.
ed “Channel” keys with softkey menus.
,
These keys allow access to softkey menus which af
channel. Refer to Chapter 3 for further information about these keys
and their softkey menus.
The AGILENT E4418B has one sensor input. Power meters configured with
Options
panel.
The power reference output is a 50 Ω type-
1 mW at 50 MHz is used for calibrating the sensor and meter combination.
ower meters configured with Option 003 have the power reference on the
P
rear panel.
002 or 003 have the sensor inputs on the rear panel and the front
WER REF Output.
ws. Refer to Chapter 3 for further information
he GPIB address) and also to
fect the measurement
N connector. The output signal of
6E4418B Power Meter User’s Guide
Getting Started1
row keys.
10 Ar
, , , and keys allow you to move the position of the
cursor, select fields for editing and edit alphanumeric characters. Refer
to
Chapter 3 for further information.
11 Menu r
bottom right of the power meter display indicates the number of pages in
t
he menu. For example, if “1 of 2” is displayed, pressing moves you
to “2 of 2”. Pressing again moves you back to “1 of 2”.
Repeatedly pressing accesses a menu which allows you to increase
and decrease the display contrast.
12 Sof
These four keys are used to make a selection from the menus.
elated keys.
This key allows you to move through all pages of a menu. The
This key allows you to move back one level in the softkey menu.
tkeys.
E4418B Power Meter User’s Guide7
1Getting Started
6
7
8
9101112
13
1
2
14
3
5
4
The Display Layout
The following figure details the display layout when two
measurement windows are displayed, one analog and one digital.
However, it is possible using the key to display just one
measurement window.
1 The st
8E4418B Power Meter User’s Guide
2 The me
atus reporting line displays five fields, three associated
with either GPIB, RS232, or RS422 status and two associated
with error and warning conditions. The first field displays either
“RMT” (remote, GPIB, RS232, or RS422 operation) or
“LCL” (local, front panel operation).
For GPIB operation, the second field displays “TLK” if the power
er is addressed to talk or “LSN” if it is addressed to listen.
met
The third field indicates an “SRQ” (service request).
For RS232 and RS422 operation, the second field displays “RX”
n data is being received. The third field displays “TX” when
whe
the power meter is transmitting data.
The fourth field indicates “ERR” for any error conditions. The
last f
rectangular windows depending on the setting of . Pressing
ield is used to report error and warning messages.
asurement data is displayed in either one or two
allows you to toggle between a one or two window display.
Getting Started1
When two windows are displayed and this k
displayed is the one which was previously highlighted with the shadowed
box. On the two window display, the measurement setup menus work on the
window which is shadowed.
3 This is t
4 This f
5 This w
measurement result and the meter scaling.
6 Thi
example, “1 of 2” indicates that there are two pages of softkeys and you are
on the first page. Pressing moves you to page “2 of 2”.
7 An
8 Thi
meter is initially switched on, the “Contrast” menu is displayed and if you
press “Zero/Cal” is displayed.
9 This f
limits set. If the measurement is within the limits, this field is empty. If the
measurement result is less than the minimum limit set, “Undr Lmt” is
displayed. If the measurement result is more than the maximum limit set,
“Over Lmt” is displayed. Refer to “Setting Measurement Limits” on
page 70 for fur ther information.
10 This
Relative Measurements” on page 61 for further information.
11 This f
page 63 for fur ther information.
12 T
his field displays “Rng Hld” if a range is selected. Refer to “Setting the
Range” on page 79 for further information.
13 This f
measure the power of a pulsed signal. Refer to “Measuring Pulsed
Signals” on page 68 for further information.
14 The i
combination of sensor type, sensor calibration table and frequency
dependent offset table that are currently selected. Table 1- 1 shows all
the possible combinations for the two lines of the display. Find the
able entry which matches your display and use the reference number
t
in the left- hand column to look up Table 1-2 for the combination of
sensor type and correction being appl
he measurement result field.
ield displays the units of measurement either dBm, dB, Watts, or %.
indow is configured to show an analog meter which displays the
s field displays the number of pages in the current softkey menu. For
y softkeys available are displayed in these four fields.
s field displays the title of the menu. For example, when the power
ield indicates if the measurement result is out of the upper or lower
field displays “Rel” if relative mode is on. Refer to “Making
ield displays “Ofs” if an offset is set. Refer to “Setting Offsets” on
ield displays “Dty Cyc” if a duty cycle is set. This allows you to
nformation in this field is displayed on two lines and depends on the
ey is pressed, the single window
ied to the current measurement.
E4418B Power Meter User’s Guide9
1Getting Started
For example, the display shows:
50MHz
(10,C)
This is equivalent to reference number 4 in Table 1- 1 and when looked
up in Table 1- 2 shows that:
he sensor type is 8480 series
• t
• a sensor calibration ta
• a frequency
dependent offset table is selected (C).
ble is selected (10)
Tab l e 1- 1 R
Reference
Number
1CF:xxx.x%
2CF:xxx.x%xxx.xyHz(a)
3xxx.xyHz(nn)
4xxx.xyHz(nn,a)
5xxx.xyHz
6xxx.xyHz(a)
Where “y” is the frequency multipl
sensor calibration table number and “a” is the frequency
dependent offset table letter.
elevant Softkeys in Various Mode of Operation
Upper
Display Line
Lower
Display Line
ier (M or G), “nn” is the
Tab l e 1- 2 List of Possible Combinations for the Upper and Lower Display Lines
Reference
Number
18480 Series
2From offset table
3Frequency dependent -
4From offset table
5E-Series sensors/
6From offset table
Sensor
Series
sens
ors/ N8480
Series sensors
with Option CFT
N8480 Series
sensors (excluding
Option CFT)
Sensor
Correction
Directly entered
Calibration Factor
from sel
ected sensor
calibration table
Frequency dependent downloaded directly from
sensor
Frequency Dependent
Offset Correction
None
None
None
10E4418B Power Meter User’s Guide
Getting Started1
, Display Format
Upper
Window
Lower
Window
Display
One Window
Two Windows
Digital
Display
Analog
Display
Digital
Display
Analog
Display
Digital
Display
Analog
Display
Select using
Select using
Select using
Select using
, Display Format
Selecting Your Display Layout
Your power meter display is extremely flexible. It can be used to display a
variety of different measurements and windows depending on your needs. The
following diagram details the various options available to you.
E4418B Power Meter User’s Guide11
1Getting Started
Confirm
Display Format
Meter Dgtl Anlg
Display Tutorial
If you wish to experiment with the display layout before you move on to make
measurements, the following procedure guides you through a few of the
display setups you can choose.
1 P
ress , .
Notice that the upper window (which has a
dig
ital display) is the one which is highlighted
by the dark box.
ress . The display remains the same
2 P
apart from the fact that it is now the lower
w (which is an analog display) that is
windo
highlighted by the dark box.
3 Press
one window. This is the analog window which,
in t
4 Press , ,
The display shows a digital window.
5 Press
digital windows.
. The display now only shows the
he previous step was selected with the
key.
(Dgtl should be highlighted).
. The display now shows two
12E4418B Power Meter User’s Guide
Getting Started1
Meter Dgtl Anlg
6 Press . The upper window is now the
selected window and is highlighted by the black
x.
bo
7 Press (Anlg should be
highlighted). The upper window now displays
an analog
Select the digital display using .
8
meter.
E4418B Power Meter User’s Guide13
1Getting Started
Confirm
Window Symbols
There are a number of different graphic symbols and pop- up windows that can
occur on the power meter display. These can occur for a variety of reasons
such as when:
• an er
• a
• y
• y
• y
Warning Symbol
The warning symbol is displayed either directly in the
measurement window or in a pop- up window when such an
event occurs. A pop- up window is displayed for
approximately two seconds. The text in the pop- up window
gives details of the warning type. This symbol may also
appear on a measurement window, for example, to indicate
that a power sensor is not connected.
ror or warning occurs.
confirmation is required.
ou are required to wait while the power meter carries out a procedure.
ou are required to select an entry from a list.
ou are required to enter an alphanumeric value.
Confirmation Window
This pop- up window is displayed when
you are required to press to
verify your previous selection. For
ple, prior to a preset being carried
exam
out.
14E4418B Power Meter User’s Guide
Getting Started1
Wait Symbol
The wait symbol is displayed when the power meter is carrying out a
procedure, but no action is required from you. The symbol may
appear directly in the measurement window or in a pop- up
window. It may appear, for example, during zeroing or
calibration.
One of N Entry Window
This pop- up window is displayed when you
are required to select an entry using
and from the list.
Numeric or Alphanumeric Entry Window
This pop- up window is displayed when you
are required to modify numeric or
alphanumeric data. The and
keys move the position of the cursor.
The and keys increment and
decrement the alphanumeric digit on
which t
E4418B Power Meter User’s Guide15
he cursor is currently positioned.
1Getting Started
123
4
5
6
7
98
10
N279
The Rear Panel at a Glance
1 Channel A (Option 002 or 003 only)
2 Po
wer Ref (Option 003 only)
The power reference output is a 50 Ω typeused for calibrating the sensor meter combination.
3 Rec
order Output
This output produces a DC voltage that corresponds to the power level of
t
he channel input. Refer to “Recorder Output” on page 86 for further
information.
4 Po
wer socket
This power meter has an auto configuring power supply. This allows it to
ope
rate over a range of voltages without manually being set to a certain
voltage.
5 Fu
se
An F3.15AH fuse is installed f
6 GP
IB
This connector allows the power meter to be controlled remotely using the
General Purpo
7 RS232/422
This connector allows the power meter to be controlled remotely using
eit
her the RS232 or RS422 serial interface standards.
se Interface Bus.
or all voltage supplies.
N connector. The output signal is
16E4418B Power Meter User’s Guide
Getting Started1
ial Label
8 Ser
Each power meter has its own individual identif
“Instrument serial numbers” on page 104 for further information.
9 Rm
t I/O
This connector is an RJ- 45 series shielded
provides a TTL logic level output when a measurement exceeds a
predetermined limit. TTL inputs are also provided to initiate zero and
calibration cycles.
10 Ground Connect
Binding post, accepts 4 mm plug or
or
bare wire connection.
ication number. Refer to
modular jack assembly. It
E4418B Power Meter User’s Guide17
1Getting Started
Bench top viewing positionsCarrying position
Adjusting the Carrying Handle
To adjust the position, grasp the handle by the sides and pull outward. Rotate
the handle to the desired position.
18E4418B Power Meter User’s Guide
Rack Mounting the Power Meter
Front
Rear (bottom view)
You can mount the power meter in a standard 19- inch rack cabinet using one
of three optional kits. Instructions and mounting hardware are included with
each rack mounting kit. Any Agilent System II instrument can be rack
mounted besides the Agilent E4418B power meter.
To rack mount the power meter:
1 Remo
2 Re
ve the handle by rotating it to the vertical position and pulling the
ends outward.
move the rubber bumper by stretching a corner and sliding it off.
Getting Started1
To rack mount a single instrument or
E4418B Power Meter User’s Guide19
der Option 908 or adapter kit 5063- 9240.
1Getting Started
To rack mount two instruments side by side or
5061- 9694 and flange kit 5063- 9212.
To install one or two instruments in a sliding support shelf order shelf
5063-
9255 and slide kit 1494- 0015 (for a single instrument, also order filler
panel 5002- 3999).
der Option 909 or lock- link kit
20E4418B Power Meter User’s Guide
E4418B Power Meter
User’s Guide
2
Power Meter Operation
Introduction 22
Battery Operation (Option 001) 24
Zeroing and Calibrating the Power Meter 29
Making Measurements with the Agilent E-Series Power Sensors and
N8480 Series Power Sensors (excluding Option CFT) 39
Making Measurements with the Agilent 8480 Series Power Sensors and
N8480 Series Power Sensors with Option CFT 41
Making Measurements using Sensor Calibration Tables 44
Making Measurements using Frequency Dependent Offset Tables 53
Setting the Units of Measurement 59
Selecting Units of Measurement from the Softkeys 60
Making Relative Measurements 61
Setting the Resolution 62
Setting Offsets 63
Setting Averaging 65
Measuring Pulsed Signals 68
Setting Measurement Limits 70
Selecting a Digital or Analog Display 76
Setting the Range 79
Configuring the Remote Interface 80
Recorder Output 86
Saving and Recalling Power Meter Configurations 88
How Measurements are Calculated 90
Presetting the Power Meter 91
Self Test 94
Operator Maintenance 101
Contacting Agilent Technologies 103
Agilent Technologies
21
2Power Meter Operation
Introduction
This chapter describes the parameters which configure the power meter to
make measurements and help you determine settings to optimize
performance. This chapter contains the following sections:
• “Battery Operation (Option 001)” on page 24
• “Zeroing and Calibrating the Power Meter” on page 29.
• “Calibrating the Power Meter” on page 30.
• “Making Measurements with the Agilent E- Series Power Sensors and
N8480 Series Power Sensors (ex
• “Making Measurements with the Agilent 8480 Series Power Sensors and
N8480 Series Power Sensors
• “Making Measurements using Sensor Calibration Tables” on page 44.
• “Making Measurements using Frequency Dependent Offset Tables” on
page 53
• “Setting the Units of Measurement” on page 59.
• “Selecting Units of Measurem
• “Making Relative Measurements” on page 61.
• “Setting the Resolution” on page 62.
• “Setting Offsets” on page 63.
• “Setting Averaging” on page 65.
• “Measuring Pulsed Signals” on page 68.
• “Setting Measurement Limits” on page 70.
• “Selecting a Digital or Analog Display” on page 76.
• “Setting the Range” on page 79.
• “Configuring the Remote Interface” on page 80.
• “Recorder Output” on page 86.
• “Saving and Recalling Power Meter Configurations” on page 88.
• “How Measurements are Calculated” on page 90
• “Presetting the Power Meter” on page 91.
• “Self Test” on page 94.
• “Operator Maintenance” on page 101.
cluding Option CFT)” on page 39.
with Option CFT” on page 41.
ent from the Softkeys” on page 60.
22E4418B Power Meter User’s Guide
• “Contacting Agilent Technologies” on page 103.
Power Meter Operation2
E4418B Power Meter User’s Guide23
2Power Meter Operation
Continue
CAUTION
Battery Operation (Option 001)
The battery Option 001 allows the power meter to be used in operating
environments where there is no convenient access to an AC power source.
General Information
With the battery option installed and the power meter connected to an AC
power source, the meter operates from the AC source and the battery runs in a
controlled charge mode.
If the meter is powered up under battery po
connected to an AC source, a pop-up window displays the message
“Running Under Battery Power”. When this screen is visible, the backlight
mode (see “Backlight” on page 26) is over- ridden and the backlight is on
continuously. Pressing the softkey returns the display to the
previous screen.
wer or if AC power is lost while
Carry Case
A soft carry or operating case is available which makes it easy to transport
and operate your power meter in installation and maintenance environments.
To obtain a carry case, order Agilent part number 34141A.
Do not attempt to recharge the power meter or operate the power meter from an AC power
source while it is contained in the carry case.
24E4418B Power Meter User’s Guide
Power Meter Operation2
Service
Battery
Running Time
A fully charged battery will typically allow the power meter to be used
continuously for up to three hours with the backlight off and two hours with
the backlight on.
Battery Status
You can check the status of the battery by pressing , ,
, , . The Battery Status display (Figure 2- 1 )
shows the charge condition of the battery represented on an analog meter
scale. Belo
meter running time under battery operation with the present charge level.
w the analog scale is a message indicating the estimated power
Running time can be increased by opera
backlight turned off (see “Backlight” on page 26 ).
Figure 2-1 Ba
When the power meter is running on battery power and there are less than 10
nutes of run- time remaining, the message “Battery Low” appears at the top
mi
of the screen. In addition, a pop- up window appears every minute with the
message “Battery Power Low”.
ttery Status
ting the power meter with the display
E4418B Power Meter User’s Guide25
2Power Meter Operation
More
Service
Battery
On
Off
Timed
NOTE
Charging Times
The battery charges automatically when the power meter is connected to an
AC source. From an empty state, the battery is fully charged in less than two
hours. After approximately 50 minutes, the battery has enough charge to
enable one hour of operation with the backlight on. After approximately 35
minutes, the battery has enough charge to enable one hour of operation with
the backlight off.
Backlight
When you are operating the power meter under battery power, you can turn
the display backlight off or on or set it to timed mode. In timed mode, the
display backlight turns off 10 minutes after the last key press the display can
be turned on again by pressing any key.
Working with the backlight off reduces the drain on the battery and increases
he operating time available by approximately 50%. The display can be
t
comfortably read in ambient daylight with the backlight off.
To access the backlight menu, press , , , . Use
the softkeys on the menu to select either , , or .
If the power meter is connected to an AC power source the backlight menu is greyed out
and the backlight is permanently on.
26E4418B Power Meter User’s Guide
Power Meter Operation2
WARNING
CAUTION
Battery Removal/Replacement
The battery unit is easily removed and replaced. Follow the instructions given
in Figure 2- 2. To obtain a replacement battery unit, order Agilent part
number E9287A (this can only be used in power meters with Option 001
alled).
inst
This product uses a Nickel Metal Hydride battery.
Do not short circuit the battery terminals.
Do not subject the battery to excessive heat.
Do not dispose of by burning.
Refer to your local country regulatory requir
Hydride batteries.
ements on the disposal of Nickel Metal
E4418B Power Meter User’s Guide27
Observe static precautions while removing and replacing the battery module.
2Power Meter Operation
1 Using a flat headed screwdriver, turn the two spring catches one quarter
turn anticlockwise.
2 Holding the catches, pull the battery assembly clear of its slot in the power
meter.
3 Support the closure assembly on a flat surface.
4 Pinch the release tab between finger and thumb while -
5 lightly pressing down on the battery module.
1
2
Release Tab
Battery
Module
Closure
Assembly
3
4
5
Replacing the battery is basically a reversal of the removal procedure.
Figure 2-2 Battery Removal/Replacement
28E4418B Power Meter User’s Guide
Zeroing and Calibrating the Power Meter
Zero
This section describes how to zero and calibrate the power meter. You should
always zero the power meter prior to calibrating it.
Zeroing the Power Meter
Zeroing adjusts the power meter for a zero power reading with no power
applied to the power sensor. During zeroing, which takes approximately 10
seconds, the wait symbol is displayed.
To zero the power meter:
1 Press , . During zeroing, the wait symbol is displayed.
When to Zero?
Zeroing of the power meter is recommended:
• when a 5 °C change in t
• when y
• e
• prior t
specified power for your power sensor.
ou change the power sensor.
very 24 hours.
o measuring low level signals. For example, 10 dB above the lowest
emperature occurs.
Power Meter Operation2
Zero/Cal Lockout
The Zero/Cal Lockout facility provides a mean of ensuring that a measurement
cannot be taken until the connected sensor has been zeroed and calibrated. If
the Zero/Cal Lockout facility is enabled and a sensor is connected which have
not been zeroed and calibrated, then the display window for the sensor will
display the message “Please Zero + Cal ChA”.
If you zero the sensor before calibrating it, the message changes to “Please Cal
.
ChA”
If you calibrate the sensor before zeroing it, t
Zero ChA”.
E4418B Power Meter User’s Guide29
he message changes to “Please
2Power Meter Operation
Must Cal
Off
On
Must Cal
Off
On
NOTE
The Zero/Cal Lockout facility can be enabled or disabled through either the
em Inputs menu or the Zero Cal menu as follows:
Syst
Press , , , , or .
Press , , , , or .
Calibrating the Power Meter
Calibration sets the gain of the power meter using a 50 MHz 1 mW calibrator
as a traceable power reference. The power meter’s POWER REF output or a
suitable external reference is used as the signal source for calibration. An
essential part of calibrating is setting the correct reference calibration factor
for the power sensor you are using. The Agilent Technologies E4418B Series
ower Meter User’s Guide require you to set the reference calibration factor.
P
The Agilent E-Series power sensors and N8480 Series power sensors
cluding Option CFT) set the reference calibration factor automatically.
(ex
During calibration, the wait symbol is displayed. Offset, relative, and duty
cycle settings are ignored during calibration.
During calibration, the power meter automatically switches the power reference calibrator
on (if it is not already on), then after calibration, it switches it to the state it was in prior to
the calibration.
Calibration Procedure Using Agilent E-Series Power Sensors
and N8480 Series Power Sensors (excluding Option CFT)
The following procedure describes how you calibrate the power meter with an
Agilent E- Series power sensor or N8480 Series power sensor (excluding
Option CFT). Since
Agilent E- Series power sensor or N8480 Series power sensor (excluding
Option CFT)’s c
calibration factor. The power meter identifies that an Agilent E- Series power
sor or N8480 Series power sensor (excluding Option CFT) is connected and
sen
will not allow you to select certain softkeys. The text on these softkeys appears
grayed out.
30E4418B Power Meter User’s Guide
the power meter automatically downloads the
alibration table, there is no requirement to enter the reference
Power Meter Operation2
Cal
Cal
NOTE
Tables
Linearity
ATyp
DTyp
1 Press .
2 Connect
3 Press to calibrate the power meter. During calibration the wait symbol
is displayed. (The power meter automatically turns on the POWER REF
output.)
the power sensor to the POWER REF output.
Example
To calibrate the power meter with an Agilent E-Series power sensor.
• Press .
• Connect
• Press .
the power sensor to the POWER REF output.
Calibration Procedure using Agilent 8480 Series power
sensors and N8480 Series Power Sensors with Option CFT
The following procedure describes how you calibrate the power meter with the
Agilent 8480 Series power sensors and N8480 Series power sensors with
Option CFT.
V8486A and W8486A sensors
For most 8480 Series sensors, the
automatically selected. The linearity-type field on the measurement screen is grayed out.
Under the gray-out condition, the selected linearity correction table that is displayed on
the screen is irrelevant to the correct linearity correction table that is automatically
selected by power meter firmware.
However, for the V8486A and
automatic selection must be overridden by selecting the D type linearity. Subsequent
connection of another A type sensor while the D type is selected will result in a warning
message stating that “Linearity Override May be Required”.
To select the linearity type to be applied:
Press , , , , or .
E4418B Power Meter User’s Guide31
correct (A type or D type) linearity correction table is
W8486A sensors (V8486A and W8486A sensors only,) the
2Power Meter Operation
Ref CF
Ref CF
%
Cal
Ref CF
%
Cal
There are a variety of methods to connect the power sensors to the power
er depending on the model of power sensor you are using. Refer to
met
Table 2- 1 on
models.
1 Press .
2 V
erify the reference calibration factor of your power sensor with that
displayed under . The value shown is obtained from the sensor
calibration table if one is selected, otherwise it is the last value set or the
default of 100%.
displays the reference calibration factor in a popreference calibration factor (see below) as desired.
• U
positioned.
• Use or to move to other digits.
3 T
o confirm your choice, press .
4 Connect
5 Press to calibrate the power meter. During calibration, the wait symbol
is displayed. (The power meter automatically turns on the POWER REF
output.)
page 33 for details on connecting different power sensor
If the value is not correct, press . The power meter
up window. Modify this
se or to modify the digit on which the cursor is currently
the power sensor to the POWER REF output.
Example
To calibrate the power meter with a power sensor whose reference calibration
factor is 99.8%.
• Press .
• Press . Use the , , , and keys to enter 99.8. Press
These power sensors connect directly to the reference calibrator.
Prior to the power meter being cali
reference attenuator should be connected between the power sensor and
the reference calibrator. Remove this attenuator from the power sensor
input before making measurements.
(1250-0597) to connect to the reference calibrator. Remove this adapter
from the power sensor input before making measurements.
The waveguide power sensors have two conne
connector is the one which is used to calibrate the power meter.
These power sensors are configured wi
power meter being calibrated, this attenuator must be removed. The
attenuator must be reconnected prior to making measurements.
This power sensor requires an APC 3.5 (f) to 50 Ω (m) N(08485-60005) to connect to the reference calibrator. Remove this adapter
before making measurements.
th an attenuator. Prior to the
Type adapter
E4418B Power Meter User’s Guide33
2Power Meter Operation
Sensor ModelConnection Requirements
Agilent 8485D Prior to the power meter being zeroed and calibrated, an Agilent 11708A
30 dB reference attenuator and
(08485-60005) should be connected between the power sensor and the
reference calibrator. Remove this attenuator from the power sensor input
before making measurements.
Agilent 8487A
Agilent N8487A
Agilent N8488A
Agilent 8487DPrior to the power meter being zeroed and calibrated, an Agilent 11708A
This power sensor requires an APC 2.4 (f) to 50 Ω (m) N(08487-60001) to connect to the power meter. Remove this attenuator
before making measurements.
30 dB reference attenuator and
(08487-60001) should be connected between the power sensor and the
reference calibrator. Remove this attenuator from the power sensor input
before making measurements.
an APC 3.5 (f) to 50 Ω (m) N-Type adapter
Type adapter
an APC 2.4 (f) to 50 Ω (m) N-Type adapter
34E4418B Power Meter User’s Guide
Power Meter Operation2
Pin NumberConnection
1none
2Ground
3Upper Window TTL Output
4Lower Window TTL Output
5TTL Input 1
6TTL Input 2
7Ground
8Ground
1 2 3 4 5 6 7 8
Zeroing and Calibrating Using TTL Inputs
You can use the TTL inputs on the rear panel Rmt I/O port to initiate
zero and calibration cycles on the power meter. The connector is an RJ- 45
Series shielded modular jack with the TTL input pins connected as shown
in Figure 2- 3.
Figure 2-3 Rm
The TTL inputs are active low and control t
as shown in Table 2- 2
Tab l e 2- 2 TTL Inputs Control Logic
Input 1Input 2Operation
11None
10CAL
01ZERO
00CAL
t I/O Port TTL Inputs
he zero and calibration functions
E4418B Power Meter User’s Guide35
2Power Meter Operation
T1
T2
T3
T4
Input
Operation
Effective control of the zero and calibration cycles using the TTL inputs
depends on cor
rect timing of the input signals as shown in Table 2- 3 and
Table 2- 4.
Tab l e 2 -3 TTL
Timing of zero/cal inputs for conditions “01” and “10”.
TimeDescriptionVal ue
T1Minimum width of input300 ms
T2Time between input detection and start of zero/cal
T3Maximum width of input. Longer inputs may cause a
Input Timing Diagram 1
cy
cle. This is determined by the number of averages
x sample rate or if there is an existing zero/cal
operation in progress, the time for this current
operation to complete. Note that the worst case is
1024 averages x 50 ms = 51.2 s. For front panel
operation (in free run mode) the time is 1 x 50 ms.
subseque
current one is complete.
nt zero/cal operation some time after the
Max: 50 ms (typical)
Min: 0 ms
4 s
36E4418B Power Meter User’s Guide
Power Meter Operation2
Inputs
Operation
T5
T6
T7
T8
Tab l e 2- 3 TTL Input Timing Diagram 1
T4Time for zero/cal operation to complete.Zero: 10 s (8480 Series)
12 s (E-Series)
20 s (N8480 Series exclude
Option C
FT)
8 s (N8480 Series with Option
CFT)
Cal: 6 s (8480 Series)
7 s (E-Series)
9 s (N8480 Series exclude
Option C
FT)
6 s (N8480 Series with Option
CFT)
All timings based on 100 ms firmware polling.
Tab l e 2- 4 TTL Inputs Timing Diagram 2
Timing of zero/cal inputs for condition “00”.
TimeDescriptionValue
T5Maximum time between inputs going low.100 ms
T6Minimum overlap of low inputs.200 ms
E4418B Power Meter User’s Guide37
2Power Meter Operation
Tab l e 2- 4 TTL Inputs Timing Diagram 2
T7Time between input detection and start of zero/cal
cycle. This is determined by the number of averages
x sample rate or if there is an existing zero/cal
operation in progress, the time for this current
operation to complete. Note that the worst case is
1024 averages x 50 ms = 51.2 s. For front panel
operation (in free run mode) the time is 1 x 50 ms.
T8Time for cal operation to complete.Cal: 6 s (8480 Series)
All timings based on 100 ms firmware polling.
If both TTL inputs are simultaneously low under any circumstances othe
above, the operation is undefined.
4 s
7 s (E-Series)
9 s (N8480 Series exclude
Option CFT)
6 s (N8480 Series with Option
CFT)
r than those shown
38E4418B Power Meter User’s Guide
Power Meter Operation2
Zero
Cal
Freq
Freq
Making Measurements with the Agilent E-Series Power Sensors
and N8480 Series Power Sensors (excluding Option CFT)
This section describes how to make continuous wave measurements using the
Agilent E- Series power sensors and N8480 Series
Option CFT) . The Agilent E- Series power sensors and N8480
sensors (excluding Option CFT) have their sensor calibration tables stored in
EEPROM. This allows the frequency and calibration data to be downloaded to
the power meter automatically.
To make measurements, perform the following steps:
o and calibrate the power meter.
1 Zer
2 Set t
3 Tak
he frequency for the signal you want to measure.
e a reading.
Procedure
power sensors (excluding
Series power
The following procedure details how to make a measurement on channel A of
the power meter.
1 Ensure t
2 Press .
3 Press . During zeroing, which takes appr
symbol is displayed.
4 Connect
5 Press to calibrate the power meter. During calibration, the wait symbol
is displayed. (The power meter automatically turns on the POWER REF
output.)
6 Press . The current setting of the frequency is displayed
under the softkey.
7 T
o change this setting, press . The power meter displays the frequency
in a pop- up window. Modify this frequency (see belo
•
E4418B Power Meter User’s Guide39
hat no power is applied to the power sensor.
oximately 10 seconds, the wait
the power sensor to the POWER REF output.
w) as desired.
Use or to modify the digit on which the cursor is currently
positioned.
2Power Meter Operation
Zero
Cal
Freq
MHz
• Use or to move to other digits.
o confirm your choice, press the appropriate frequency units.
8 T
9 C
onnect the power sensor to the signal to be measured.
10 The me
asurement result is now displayed.
Example
To make a measurement using an Agilent E- Series power sensor. The
frequency of the signal to be measured is 100 MHz.
• Dis
connect the power sensor from any source.
• Press .
• Press .
• Connect
• Press .
• Press , . Use the , , , and keys to
enter 100. Press .
• C
onnect the power sensor to the signal to be measured.
• The me
the power sensor to the POWER REF output.
asurement result is now displayed.
40E4418B Power Meter User’s Guide
Power Meter Operation2
Zero
Making Measurements with the Agilent 8480 Series Power
Sensors and N8480 Series Power Sensors with Option CFT
This section applies to all Agilent 8480 Series power sensors and N8480 Series
power sensors with Option CFT. It does not apply to the Agilent E- Series
wer sensors and N8480 Series power sensors (excluding Option CFT) .
po
For the Agilent 8480 Series power sensors and N8480 Series power sensors
h Option CFT, there are two methods of providing correction data to the
wit
power meter:
a in
putting the individual calibration factor for a frequency prior to
making the measurement, or
b us
ing sensor calibration tables.
This section describes how to make measurements without using sensor
calibration
frequency prior to making the measurement. This method is advantageous if
you are only measuring at one frequency, as entering all the calibration data in
this instance is not required.
tables, that is inputting the individual calibration factor for a
To make measurements without sensor c
following steps:
1 Zer
o and calibrate the power meter. Before carrying out the calibration, you
must set the reference calibration factor of the power sensor.
2 Set t
3 Tak
he calibration factor value for the frequency of the signal you want to
measure.
e a reading.
alibration tables, perform the
Procedure
The following procedure details how to make a measurement on channel A.
1 Ensure t
2 Press .
3 Press . During zeroing, which takes appr
symbol is displayed.
4 Connect
E4418B Power Meter User’s Guide41
hat no power is applied to the power sensor.
oximately 10 seconds, the wait
the power sensor to the POWER REF output.
2Power Meter Operation
Ref CF
Ref CF
%
Cal
Cal Fac
Cal Fac
%
5 The cur
the softkey. To change this setting, press . The power meter
displays the reference calibration factor in a popreference calibration factor (see below) as desired.
• U
• Use or to move to other digits.
6 T
7 Press to calibrate the power meter. During calibration, the wait symbol
is displayed. (The power meter automatically turns on the POWER REF
output.)
8 Press . The current setting of the calibration factor is
displayed under the softkey. To change this setting, press .
The power meter displays the calibration factor in a pop- up window. Modify
t
his calibration factor (see below) as desired.
• U
• Use or to move to other digits.
9 T
10 C
11 The me
rent setting of the reference calibration factor is displayed under
up window. Modify this
se or to modify the digit on which the cursor is currently
positioned.
o confirm your choice, press .
se or to modify the digit on which the cursor is currently
positioned.
o confirm your choice, press .
onnect the power sensor to the signal to be measured.
asurement result is now displayed.
42E4418B Power Meter User’s Guide
Power Meter Operation2
Zero
Ref CF
%
Cal
Cal Fac
%
Example
To make a measurement on channel A with a power sensor which has a
reference calibration factor of 99.8% and a calibration factor of 97.8% at the
measurement frequency.
• Dis
connect the power sensor from any source.
• Press .
• Press .
• Press . Use the , , , and keys to enter 99.8. Press
.
• Connect
• Press .
• Press , . Use the , , , and keys to
enter 97.8. Press .
• C
onnect the power sensor to the signal to be measured.
• The me
the power sensor to the POWER REF output.
asurement result is now displayed.
E4418B Power Meter User’s Guide43
2Power Meter Operation
Making Measurements using Sensor Calibration Tables
This section applies to all Agilent 8480 Series power sensors and N8480 Series
power sensors with Option CFT. It does not apply to the Agilent E- Series
wer sensors and N8480 Series power sensors (excluding Option CFT) .
po
For the Agilent 8480 Series power sensors and N8480 Series power sensors
h Option CFT, there are two methods of providing correction data to the
wit
power meter:
a in
putting the individual calibration factor for a frequency prior to
making the measurement, or
b us
ing sensor calibration tables.
This section describes how to use sensor calibration tables. Sensor calibration
ables are used to store the measurement calibration factors, supplied with
t
each power sensor in the power meter. These calibration factors are used to
correct measurement results.
Using sensor calibration tables provides y
method for making power measurements over a range of frequencies using one
or more power sensors. The power meter is capable of storing 20 sensor
calibration tables of 80 frequency points each.
To use sensor calibration tables you:
1 Select
Calibration Table” on page 45 for further information. If you are
required to edit the table, refer to “Editing Sensor Calibration
Tables” on page 46 for further information.
2 Ze
during the calibration is automatically set by the power meter from the
sensor calibration table.
3 Spe
factor is automatically set by the power meter from the sensor calibration
table. Refer to “Making the Measurement” on page 45 for further
information.
4 Mak
the table to work on a channel. Refer to “Selecting a Sensor
ro and calibrate the power meter. The reference calibration factor used
cify the frequency of the signal you want to measure. The calibration
e the measurement.
ou with a quick and convenient
44E4418B Power Meter User’s Guide
Power Meter Operation2
Tables
Sensor Cal Tables
Table Off On
Zero
Ref CF
Selecting a Sensor Calibration Table
You can select a sensor calibration table for use by pressing ,
, , . The “State” field indicates if
any sensor calibration tables are currently selected. The “Sensor Tbls”
een is displayed as shown in Figure 2- 4.
scr
Figure 2-4 “Senso
r Tbls” Screen
Making the Measurement
To make the power measurement, you zero and calibrate the power meter,
then set it for the frequency of the signal you want to measure. The power
meter automatically selects the calibration factor from the sensor calibration
table.
1 Press .
2 Press . During zeroing, which takes appr
symbol is displayed.
3 The settin
calibration table and displayed under the softkey.
4 Connect
E4418B Power Meter User’s Guide45
g of the reference calibration factor is obtained from the sensor
the power sensor to the POWER REF output.
oximately 10 seconds, the wait
2Power Meter Operation
Cal
Freq
Freq
NOTE
Cal Fac
Tables
Sensor Cal Tables
5 Press to calibrate the power meter. During calibration, the wait symbol
is displayed. (The power meter automatically turns on the POWER REF
output.)
6 Press . The current setting of the frequency is displayed
under the softkey.
o change this setting, press . The power meter displays the frequency
7 T
in a pop- up window. Modify this fr
• U
se or to modify the digit on which the cursor is currently
positioned.
• Use or to move to other digits.
8 T
o confirm your choice, press the appropriate frequency unit.
9 C
onnect the power sensor to the signal to be measured.
10 The me
If the measurement frequency does not correspond directly to a frequency in the sensor
calibration table, the power meter calculates the calibration factor using linear
interpolation.
If you enter a frequency outside the frequency range
table, the power meter uses the highest or lowest frequency point in the sensor calibration
table to set the calibration factor.
The value of the calibration factor being used by the power meter to make a measurement
splayed under the softkey.
is di
asurement result is now displayed.
equency (see below) as desired.
defined in the sensor calibration
46E4418B Power Meter User’s Guide
Editing Sensor Calibration Tables
You can edit sensor calibration tables using the “Edit Cal” menu as shown
in Figure 2- 5. To view the sensor calibration tables currently stored in the
power meter, press , , . The “Sensor
Tbls” screen is displayed as shown in Figure 2- 4.
Power Meter Operation2
Figure 2-5 “Edit Cal S
creen”
The power meter is shipped with a set of predefined sensor calibration tables.
The dat
a in these tables is based on statistical averages for a range of Agilent
Technologies power sensors.
These power sensors are:
1
• DEF
AULT
• HP 8481A
• HP 8482A
2
• HP 8483A
• HP 8481D
• HP 8485A
• HP R8486A
• HP Q8486A
• HP R8486D
• HP 8487A
There are also ten sensor calibration tables named CUSTOM_0 through
CUSTOM_9 which do not cont
ain any data when the power meter is shipped
from the factory.
1 DEFAULT is a sensor calibration table in which the reference calibration factor and calibration factors are 100%.
This sensor calibration table can be used during the performance testing of the power meter.
2 The Agilent 8482B and Agilent 8482H power sensors use the sam
e data as the Agilent 8482A.
E4418B Power Meter User’s Guide47
2Power Meter Operation
NOTE
Edit Table
Change
Insert
Predefined sensor calibration factor table stored in power meter is not applicable for
N8480 Series power sensors with Option CFT. Therefore, users are required to
Agilent
create a new sensor calibration table for the sensors when a sensor calibration table is
needed.
You cannot delete any of the 20 sensor calibration tables. Each sensor
calibration
want a new sensor calibration table, you must edit an existing one.
To edit a table:
1 Se
tables. When the table you want to edit is highlighted, press .
The “Edit Cal” screen is displayed as shown in Figure 2- 5 on page 47.
In this screen, you can modify, add, and remove frequency and
calib
and calibration factors.
A frequency in the range of 0.001 MHz to 999.999 GHz can be entered.
table can contain a maximum of 80 frequency points each. If you
lect one using the and keys to scroll through the displayed
ration factors and also edit the table name. Use the , ,
, and keys to move between the table name and its frequency
A calibration factor in the range of 1% to 150% can be entered.
The following rules apply to naming sensor calibration tables:
• The name
• All characters must be upper or lower case alphabetic characters or
numeric (0- 9) or an underscore (_).
• No other characters are allowed.
• No spaces are allowed in the name.
2 T
o edit the parameter you currently have selected, press . The
power meter displays the data in a popbelow) as desired.
• Use
currently positioned.
• Use or to move to other characters.
3 T
o confirm your choice, press the appropriate softkey.
4 T
o add a new sensor calibration table entry, press . You are prompted
for the frequency and calibration factor. The entry is sorted by frequency.
48E4418B Power Meter User’s Guide
must consist of no more than 12 characters.
up window. Modify this data (see
or to modify the character on which the cursor is
Power Meter Operation2
Delete
NOTE
Done
NOTE
o remove a sensor calibration table entry, use the , , , and
5 T
keys to select the entry, press . If you delete the frequency, the
equivalent calibration factor is
You can only edit the table name, you cannot delete it.
also removed and vice versa.
6 Whe
Ensure that the frequency points you use cover the frequency range of the signals you
want to measure. If you measure a signal with a frequency outside the frequency range
defined in the sensor calibration table, the power meter uses the highest or lowest
frequency point in the sensor calibration table to calculate the calibration factor.
n you have completed editing the sensor calibration table, press
to return to the “Sensor Tables” screen.
E4418B Power Meter User’s Guide49
2Power Meter Operation
The following lists detail the contents of the predefined sensor calibration
Making Measurements using Frequency Dependent Offset
Ta bl e s
This section describes how to use frequency dependent offset tables.
Frequency dependent offset tables pro
method of compensating for your external test setup over a range of
frequencies. Note that when selected, frequency dependent offset correction is
IN ADDITION to any correction applied for sensor frequency response.
The power meter is capable of storing 10 frequency dependent offset tables of
frequency points each.
80
To use frequency dependent offset tables you:
1 Select
Dependent Offset Table” on page 54 for further information. If you are
required to edit the table, refer to “Editing Frequency Dependent Offset
Tables” on page 56 for further information.
2 Ze
during the calibration is automatically set by the power meter from the
sensor calibration table (if selected).
3 Spe
factor or offset is automatically set by the power meter from the sensor
calibration table (if selected) and the frequency dependent offset table.
Refer to “Making the Measurement” on page 45 for further information.
4 Mak
the table to work on a channel. Refer to “Selecting a Frequency
ro and calibrate the power meter. The reference calibration factor used
cify the frequency of the signal you want to measure. The calibration
e the measurement.
vide you with a quick and convenient
E4418B Power Meter User’s Guide53
2Power Meter Operation
Tables
Freq Dep Offset
Table Off On
Selecting a Frequency Dependent Offset Table
You can select a frequency dependent offset table for use by pressing
, , , . The “State” field
indicates if any frequency dependent offset tables are currently selected.
The “Of
fset Tbls” screen is displayed as shown in Figure 2- 6.
Figure 2-6 “Offset Tbls”
54E4418B Power Meter User’s Guide
Screen
Power Meter Operation2
Zero
Ref CF
Cal
Freq
Freq
Making the Measurement
To make the power measurement, you zero and calibrate the power meter,
then set it for the frequency of the signal you want to measure. The power
meter automatically selects the calibration factor from the sensor
calibration table (if selected) and the offset from the frequency dependent
offset table.
1 Press .
2 Press . During zeroing, which takes approximately 10 seconds, the
wait symbol is displayed.
3 The setting of t
sensor calibration table (if selected) and displayed under the
softkey.
4 Connect
5 Press to calibrate the power meter. During calibration, the wait
symbol is displayed. (The power meter automatically turns on the
PO
6 Press . The current setting of the frequency is displayed
under the softkey.
the power sensor to the POWER REF output.
WER REF output.)
he reference calibration factor is obtained from the
7 T
o change this setting, press . The power meter displays the
frequency in a pop- up window. Modify this frequency (see below) as
desired.
• Use
currently positioned.
• Use or to move to other digits.
8 T
o confirm your choice, press the appropriate frequency unit.
9 C
onnect the power sensor to the signal to be measured.
10 The me
E4418B Power Meter User’s Guide55
or to modify the digit on which the cursor is
asurement result is now displayed.
2Power Meter Operation
NOTE
Cal Fac
Tables
Freq Dep Offset
If the measurement frequency does not correspond directly to a frequency in the sensor
calibration table (if selected) and the frequency dependent offset table being used, the
power meter calculates the calibration factor and offset using linear interpolation.
If you enter a frequency outside the frequency range defined in the sensor calibration table
or the frequency
frequency point in the appropriate table to set the calibration factor and offset.
The value of the calibration factor being used by the power meter to make a measurement
splayed under the softkey.
is di
dependent offset table, the power meter uses the highest or lowest
Editing Frequency Dependent Offset Tables
You can edit frequency dependent offset tables using the “Edit” menu as
shown in Figure 2- 7.
To view the frequency dependent offset tables currently stored in the
wer meter, press , , . The “Offset Tbls”
po
screen is displayed as shown in Figure 2- 6.
56E4418B Power Meter User’s Guide
Figure 2-7 “Edit” Screen
Power Meter Operation2
Edit Table
Change
Insert
Delete
There are ten frequency dependent offset tables named CUSTOM_A
hrough CUSTOM_J which do not contain any data when the power meter
t
is shipped from the factory.
You cannot delete any of the 10 existing frequency depen
or create any additional tables. However, you can edit the 10 existing
tables. Each frequency dependent offset table can contain a maximum of
80 frequency points.
To edit a table:
elect one using the or keys to scroll through the displayed
1 S
tables. When the table you want to edit is highlighted, press .
The “Edit Offset” screen is displayed as shown in Figure 2- 7 on
page 56. In this screen, you can modify, add, and remove frequencies
and offset and also edit the table name. Use the , , , and
keys to move between the table name and its frequencies and
offsets.
A frequency in the range of 0.001 MHz to 999.999 GHz can be entered.
An offset in the range of 1% t
• The f
• The name must consist of no more than 12 characters.
• All characters must be upper or lower case alphabetic characters or
• No other characters are allowed.
• No spaces are allowed in the name.
2 T
power meter displays the data in a popbelow) as desired.
• Use
• Use or to move to other characters.
3 T
4 T
prompted for the frequency and offset. The entry is sorted by frequency.
5 T
frequency, the equivalent offset is also removed and vice versa.
ollowing rules apply to naming frequency dependent offset tables:
numeric (0- 9) or an underscore (_).
o edit the parameter you currently have selected, press . The
or to modify the character on which the cursor is
currently positioned.
o confirm your choice, press the appropriate softkey.
o add a new frequency dependent offset table entry, press . You are
o remove a frequency dependent offset table entry, use the , ,
, and keys to select the entry, press . If you delete the
o 150% can be entered.
up window. Modify this data (see
dent offset tables
E4418B Power Meter User’s Guide57
2Power Meter Operation
NOTE
Done
NOTE
You can only edit the table name, you cannot delete it.
6 When y
press to return to the “Offset Tbls” screen.
Ensure that the frequency points you use cover the frequency range of the signals you
want to measure. If you measure a signal with a frequency outside the frequency range
defined in the frequency dependent offset table, the power meter uses the highest or
lowest frequency point in the frequency dependent offset table to calculate the offset.
ou have completed editing the frequency dependent offset table,
58E4418B Power Meter User’s Guide
Setting the Units of Measurement
dBm
W
dB
%
NOTE
Rel Off On
The “dBm/W” menu is used to select the measurement units on the
currently selected window. These can either be logarithmic (dBm or dB) or
linear (Watt or %) units. sets the measurement units to dBm
(logarithmic units). Table 2- 5 shows which units are applicable to the
individual measurement modes.
Press , then select the unit of measurement from , , ,
or . Softkeys which cannot be selected in your particular mode of
operation appear with their text grayed out.
When the unit of measurement is set to Watt, it is possible that negative power
measurements may be displayed when measuring low power levels.
Power Meter Operation2
Tab l e 2- 5 Me
Measurement ModeRelative Mode OffRelative Mode On
Single ChannelWat tdBm%dB
1 When relative mode is on (that is, , is “On”), the measurement is com-
pared to a reference value.
asurement Units
1
LinearLogLinearLog
E4418B Power Meter User’s Guide59
2Power Meter Operation
Increment
Decrement
Multiplier
Multiplier
Cancel
mW
Cancel
mW
nW
W
NOTE
Increment Multiplier
Decrement Multiplier
W
W
Selecting Units of Measurement from the Softkeys
In various softkey menu structures, you are required to enter the units of
measurement for power. In some cases, due to the wide power range
available, the following menu is displayed:
60E4418B Power Meter User’s Guide
Some softkeys may be grayed out so that an invalid value cannot be entered.
Pressing or increases or
decreases the multiplier shown in front of . Pressing after the
correct multiplier has been selected confirms the entry.
Making Relative Measurements
dB
%
Rel
Rel Off On
Rel
Rel Off On
Relative mode allows you to compare any measurement result to a
reference value. The relative reading can be displayed in either dB or % by
pressing and selecting either or . When the measurement
result is displayed in %, a pref
Power Meter Operation2
ix multiplier may be shown.
Relative mode is window dependent and
the window to which it’s applied.
when enabled, displays “Rel” in
Procedure
To set a reference value on the currently selected window:
1 Press .
2 Press to use the current reading as the reference value. This allows
you to compare any measurement resul
3 is automatically set to “On”, when is pressed.
Successive measurements are now displa
value. The relative mode can be disabled and re- enabled simply by
pressing .
t in dB or percentage (%).
yed relative to the reference
E4418B Power Meter User’s Guide61
2Power Meter Operation
Resolution 1 2 3
Resolution 1 2 3 4
Setting the Resolution
The resolution of each of the power meter’s windows can be set to four
different levels (1, 2, 3, or 4).
These four levels represent:
• 1, 0.1,
dB.
• 1, 2, 3, or 4 signif
W or %.
The default value is 0.01 dB (3 digits).
0.01, 0.001 dB respectively if the measurement suffix is dBm or
icant digits respectively if the measurement suffix is
To set the resolution on the cur
1 Press . The current setting of the resolu
softkey.
2 T
o change this setting, press until the required
resolution setting is highlighted.
rently selected window:
tion is highlighted on the
62E4418B Power Meter User’s Guide
Setting Offsets
Input Settings
Offset
dB
Offset Off On
Offset
Offset Off On
Setting Channel Offsets
The power meter can be configured to compensate for a signal loss or gain
in your test setup (for example, to compensate for the loss of a 10 dB
enuator). This gain or loss, referred to as a channel offset, is added to
att
the measured power before the result is displayed.
Offsets are entered in dB. The allowable range of values is –100 dB to
+100 dB. A positive value compensates for a loss and a negative value
mpensates for a gain.
co
If either a channel or a display offset is set, then “Ofs” is displayed.
To enter a channel offset which compensates for a loss or gain:
1 Press .
2 Press .
3 Press . The power meter displays the offset value in a pop- up
window. Modify this offset
• Use
• Use or to move to other digits.
4 T
5 is automatically set to “On” when a value is entered
using .
Power Meter Operation2
(see below) as desired.
or to modify the digit on which the cursor is
currently positioned.
o confirm your choice, press .
E4418B Power Meter User’s Guide63
Channel offsets can be disabled and
.
re- enabled simply by pressing
2Power Meter Operation
Offset
dB
Offset Off On
Offset
Offset Off On
Ch Input with
Ch Input
Ch
Offset
†
Display
Offset
††
READING
ON POWER
METER
DISPLAY
Ch Input with Ch Offset and
Ch Offset
Display Offset
Input Settings
Offset
Offset
Setting Display Offsets
The display offset function provides a method for entering display offset
values.
If either a channel or a display offset is set, then “Ofs” is displayed.
To enter a display offset on the currently selected window:
1 Press , .
2 The po
this offset (see below) as desired.
• Use
• Use or to move to other digits.
3 T
4 is automatically set to “On”, when a value is entered
using .
Display offsets can be disabled and re- enabled simply by pressing
wer meter displays the offset value in a pop- up window. Modify
or to modify the digit on which the cursor is
currently positioned.
o confirm your choice, press .
.
64E4418B Power Meter User’s Guide
The display offset is a function of the window. That is, if you have
ed a two window display, each can have its own offset.
select
Figure 2- 8 displays the effects of the channel and display offsets on the
power meter’s displayed reading.
Figure 2-8 Eff
† Channel Offset entered using , , .
†† Display Offset entered using , .
ect of Offsets on a Channel Measurement
Setting Averaging
The power meter uses a digital filter to average power readings. The
number of readings averaged can range from 1 to 1024. This filter is used
to reduce noise, obtain the desired resolution and to reduce the jitter in
the measurement results. Increasing the value of the filter length reduces
measurement noise. However, the time to take the measurement is
increased. You can select the filter length or you can set the power meter
to auto filter mode. The default is “AUTO”.
When the auto filter mode is enabled, the power meter automatically sets
he number of readings averaged together to satisfy the filtering
t
requirements for most power measurements. The number of readings
averaged together depends on the resolution and the power level currently
being measured. Figure 2- 9 lists the number of readings averaged for each
range and resolution when the power meter is in auto filter mode and is
o normal speed mode (refer to the Agilent EPM- 4418B/4419B
set t
ogramming Guide for details of the readings averaged in the other
Pr
speed modes).
Power Meter Operation2
Resolution is a window function and not a
where a channel is set up in both the upper and lower window and the
resolution settings are different, the highest resolution setting is taken to
calculate the averaging number. For example, if the upper window has a
resolution setting of two and is measuring channel A and the lower
window has a resolution of four and is also measuring channel A. In this
instance, channel A averaging is calculated with a resolution of four.
E4418B Power Meter User’s Guide65
channel function. In the case,
2Power Meter Operation
10 dB
10 dB
Minimum Sensor Power
Maximum Sensor Power
Power Sensor
Dynamic Range
10 dB
10 dB
1234
18128128
Resolution Setting
Number of Averages
1116256
11232
11116
1118
Input Settings
Filter
Length
Length
Enter
Filter Off On
Figure 2-9 A
These four resolution levels represent:
• 1, 0.1,
dB.
• 1, 2, 3, or 4 signif
W or %.
To set averaging:
1 Press , , . Press the softkey to
access the filter menu.
2 The cur
To change this setting, press . A pop- up window appears. Use
66E4418B Power Meter User’s Guide
the , , , or keys to set your filter length.
3 T
o confirm your choice, press .
The filter can be disabled and re- enabled simply by pressing
veraged Readings
0.01, 0.001 dB respectively if the measurement suffix is dBm or
icant digits respectively if the measurement suffix is
rent setting of averaging is displayed under the softkey.
.
Power Meter Operation2
Input Settings
Filter
Step Det Off On
Step Detection
To reduce the filter settling time after a significant step in the measured
power, the filter can be set to re- initialise upon detection of a step
increase or decrease in the measured power. Step detection can be set in
both manual and automatic filter modes.
To set step detection:
1 Press , , .
2 Press t
3 Use t
he softkey to access the filter menu.
he softkey to enable or disable step detection.
E4418B Power Meter User’s Guide67
2Power Meter Operation
NOTE
Power
Time
B
A
Duty Cycle = A
B
Measuring Pulsed Signals
The power meter can be used to measure the power of a pulsed signal.
The measurement result is a mathematical representation of the pulse
power rather than an actual measurement (assumes constant peak power).
The power meter measures the average power of the pulsed input signal
and then divides the measurement result by the duty cycle value to obtain
the pulse power reading. The allowable range of values is 0.001% to 100%.
The default value is 1.000%.
If duty cycle is enabled, then “Dty Cyc” is displayed.
Pulse measurements are not recommended using Agilent E4412A and E4413A power
sensors.
An example of a pulsed signal is shown in Figure 2-10.
Figure 2-10 Pu
68E4418B Power Meter User’s Guide
lsed Signal
Power Meter Operation2
Input Settings
Duty Cycle
Duty Cycle
%
Duty Cycle Off On
Duty Cycle
Duty Cycle Off On
NOTE
To set the duty cycle:
1 Press , , . The current setting of the duty
cycle is displayed under the softkey.
2 T
o change this setting, press . The power meter displays the
duty cycle in a pop-up window. Modify this value (see below) until the
desired duty
• Use
currently positioned.
• Use or to move to other digits.
3 T
o confirm your choice, press .
4 is automatically set to “On”, when a value is entered
using .
Duty cycle can be disabled and re- enabled simply by pressing
Pulse power averages out any aberrations in the pulse such as overshooting or ringing. For
this reason, it is called pulse power and not peak power or peak pulse power.
In order to ensure accurate pulse
rectangular pulse. Other pulse shapes (such as triangle, chirp, or Gaussian) will cause
erroneous results.
The pulse power on or off ratio must be much greater than the duty cycle ratio.
cycle is displayed.
or to modify the digit on which the cursor is
.
power readings, the input signal must be pulsed with a
E4418B Power Meter User’s Guide69
2Power Meter Operation
NOTE
Input Settings
Limits
Max
Min
Limits Off On
Setting Measurement Limits
You can configure the power meter to detect when a measurement is
outwith a predefined upper and/or lower limit value.
There are two types of measurement limits you can set:
• Channel Limits
measurements only.
• Windo
Only one set of limits can be on at a time, that is, Channel, OR Window.
w Limits - are windows based (upper and lower) and can be
applied to power, ratio, or difference measurements. In addition, the
window based limits can be set to output a TTL logic level at the rear
panel Rmt I/O port when the predefined limits are exceeded.
Setting Channel Limits
- are applied to the input channel and are for power
70E4418B Power Meter User’s Guide
The power meter can be configured to verify the power being measured
against an upper and/or lower limit value. The range of values that can be
set for upper and lower limits is –150 dBm to 230 dBm. The default
upper
To set the limits:
1 Press , , . The current setting of the
2 T
3 T
Limits can be disabled and re- enabled simply by pressing .
limit is 90.00 dBm and the default lower limit is –90.00 dBm.
maximum and minimum limits are displayed under the and
softkeys respectively.
o change either of these settings, press the appropriate softkey. The
power meter displays the current value in a pop- up window. Modify
this value (see below) until the desired value is displayed.
• Use
currently positioned.
• Use or to move to other digits.
o confirm your choice, press the appropriate measurement units.
or to modify the digit on which the cursor is
Power Meter Operation2
Power Meter
Swept Source
CHANNEL A INPUT
OUT
IN
OUT
Device
Under Test
+4 dBm
+10 dBm
Amplitude
Frequency
o
o
o
o
o
o
o
Fail
Fail
A typical application for this feature is shown in Figure 2- 11.
Figure 2-11 Limits Checking Application
Figure 2-12 Limi
In this application, a swept frequency s
Device Under Test. The power meter measures the output power. The
limits have been set at +4 dBm and +10 dBm. A fail occurs each time the
output po
E4418B Power Meter User’s Guide71
ts Checking Results
ignal is applied to the input of the
wer is outside these limits as shown in Figure 2- 12.
2Power Meter Operation
Setting Window Limits
The power meter can be configured to verify the current measurement in
either window against predefined upper and/or lower limit values. The
range of values that can be set for the upper and lower limits and the
default values depends on the measurement units in the currently selected
window, see Table 2-6.
Tab l e 2- 6 R
Window UnitsMaxMinDefault
dB+200 dB–180 dB60 dB–120 dB
dBm+230 dBm–150 dBm90 dBm–90 dBm
%10.0 Z%100.0 a%100.0 M%100.0 p%
W100.000 XW1.000 aW1.000 MW1.000 pW
ange of Values for Window Limits
MaxMin
The window based limits can also be set t
o output a TTL logic level at the
rear panel Rmt I/O port when the predefined limits are exceeded. You can
switch the rear panel TTL outputs, on, or off; set the TTL output level to
active high or low; and determine whether the TTL output represents an
over limit condition, under limit condition or both.
The TTL connector is an RJ- 45 Series shielded modula
r jack assembly
with the TTL output pins connected as shown in Figure 2- 13.
72E4418B Power Meter User’s Guide
Power Meter Operation2
Pin NumberConnection
1none
2Ground
3Upper Window TTL Output
4Lower Window TTL Output
5TTL Input 1
6TTL Input 2
7Ground
8Ground
1 2 3 4 5 6 7 8
Limits
Max
Min
Limits Off On
Figure 2-13 Remote
I/O TTL Outputs
To set the limits:
1 Press , . The current setting of the maximum and
minimum limits are displayed under the and softkeys
respectively.
2 T
o change either of these settings, press the appropriate softkey. The
power meter displays the current value in a pop- up window. Modify
this value (see below) until the desired value is displayed.
or to modify the digit on which the cursor is
• Use
currently positioned.
• Use or to move to other digits.
3 T
o confirm your choice, press the appropriate measurement units.
Limits can be disabled and re- enabled simply by pressing .
E4418B Power Meter User’s Guide73
2Power Meter Operation
Limits
TTL Output
TTL Output
Limits
Fail O/P
Limits
OVER
UNDER
EITHER
Fail O/P
High
Low
TTL Output Off On
NOTE
To set the TTL output:
1 Press , , . The current settings for the TTL
output are displayed under the , , and
softkeys.
o select whether the TTL output reflects an overlimit condition, under
2 T
limit condition or both, press . Use any of the four arrow keys
to make a selection between , , and on the
pop- up menu that appears.
o select whether a high or low level TTL output represents a limits
3 T
failure, use the softkey to toggle between and
TTL outputs can be disabled and re-
.
enabled simply by pressing
Checking for Limit Failures
Limit failures are displayed in the appropriate field in the measurement
window on the power meter’s display as shown in Figure 2- 14.
The same limit fail field is used by both channel and window limits.
74E4418B Power Meter User’s Guide
Power Meter Operation2
This measurement
has passed. This is
indicated by the
limit field being
empty.
This measurement
has failed as the
result is less than
the minimum limit
level set of
–55 dBm. This is
indicated with the
text “Undr Lmt”.
This measurement
has failed as the
result is more than
the maximum limit
level set of
–60 dBm. This is
indicated with the
text “Over Lmt”.
Figure 2-14 Pass
E4418B Power Meter User’s Guide75
/Fail Limit Indicators
2Power Meter Operation
Selecting a Digital or Analog Display
The measurement windows can display the result in either a digital or
analog format or both as shown in the following figures.
Figure 2-15 Digital Disp
Figure 2-16 An
Figure 2-17 Di
alog Display
gital and Analog Display
lay
76E4418B Power Meter User’s Guide
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.