No part of this manual may be reproduced in any
form or by any means (including electronic
storage and retrieval or translation into a foreign
language) without prior agreement and written
consent from Keysight Technologies as governed
by United States and international copyright laws.
Manual Part Number
81160-91020
Edition
Edition 2.0, August 2014
Printed in Germany
Keysight Technologies, Deutschland GmbH
Herrenberger Str. 130
71034 Böblingen, Germany
For Assistance and Support
Warranty
The material contained in this document is
provided “as is,” and is subject to being
changed, without notice, in future editions.
Further, to the maximum extent permitted by
applicable law, Keysight disclaims all
warranties, either express or implied, with
regard to this manual and any information
contained herein, including but not limited to the
implied warranties of merchantability and
fitness for a particular purpose. Keysight shall
not be liable for errors or for incidental or
consequential damages in connection with the
furnishing, use, or performance of this document
or of any information contained herein. Should
Keysight and the user have a separate written
agreement with warranty terms covering the
material in this document that conflict with
these terms, the warranty terms in the separate
agreement shall control.
Technology Licenses
The hardware and/or software described in this
document are furnished under a license and may
be used or copied only in accordance with the
terms of such license.
Restricted Rights Legend
If software is for use in the performance of a U.S.
Government prime contract or subcontract,
Software is delivered and licensed as
“Commercial computer software” as defined in
DFAR 252.227-7014 (June 1995), or as a
“commercial item” as defined in FAR 2.101(a) or
as “Restricted computer software” as defined in
FAR 52.227-19 (June 1987) or any equivalent
agency regulation or contract clause. Use,
duplication or disclosure of Software is subject to
Keysight Technologies’ standard commercial
license terms, and non-DOD Departments and
Agencies of the U.S. Government will receive no
greater than Restricted Rights as defined in FAR
52.227-19(c)(1-2) (June 1987). U.S. Government
users will receive no greater than Limited Rights
as defined in FAR 52.227-14 (June 1987) or DFAR
252.227-7015 (b)(2) (November 1995), as
applicable in any technical data.
Safety Notices
A CAUTION notice denotes a hazard. It
calls attention to an operating
procedure, practice, or the like that, if
not correctly performed or adhered to,
could result in damage to the product
or loss of important data. Do not
proceed beyond a CAUTION notice until
the indicated conditions are fully
understood and met.
A WARNING notice denotes a hazard.
It calls attention to an operating
procedure, practice, or the like that, if
not correctly performed or adhered to,
could result in personal injury or
death. Do not proceed beyond a
WARNING notice until the indicated
conditions are fully understood and
met.
The following general safety precautions
must be observed during all phases of
operation of this instrument. Failure to
comply with these precautions or with
specific warnings elsewhere in this manual
violates safety standards of design,
manufacture, and intended use of the
instrument.
Keysight Technologies assumes no liability
for the customer's failure to comply with
these requirements.
Before operation, review the instrument and
manual for safety markings and instructions.
You must follow these to ensure safe
operation and to maintain the instrument in
safe condition.
This product is a Safety Class 1 instrument
(provided with a protective earth terminal).
The protective features of this product may
be impaired if it is used in a manner not
specified in the operation instructions.
All Light Emitting Diodes (LEDs) used in this
product are Class 1 LEDs as per IEC 60825-1.
This instrument is intended for indoor use in
an installation category II, pollution degree 2
environment. It is designed to operate at a
maximum relative humidity of 95% and at
altitudes of up to 2000 meters.
Refer to the specifications tables for the ac
mains voltage requirements and ambient
operating temperature range.
Verify that all safety precautions are taken.
The power cable inlet of the instrument
serves as a device to disconnect from the
mains in case of hazard. The instrument
must be positioned so that the operator can
easily access the power cable inlet. When
the instrument is rack mounted the rack
must be provided with an easily accessible
mains switch.
To minimize shock hazard, the instrument
chassis and cover must be connected to an
electrical protective earth ground. The
instrument must be connected to the ac
power mains through a grounded power
cable, with the ground wire firmly connected
to an electrical ground (safety ground) at the
power outlet. Any interruption of the
protective (grounding) conductor or
disconnection of the protective earth
terminal will cause a potential shock hazard
that could result in personal injury.
Do not operate the instrument in the
presence of flammable gases or fumes.
Operating personnel must not remove
instrument covers. Component replacement
and internal adjustments must be made only
by qualified personnel.
Instruments that appear damaged or
defective should be made inoperative and
secured against unintended operation until
they can be repaired by qualified service
personnel.
4
Indicates warning or caution. If you see this
This product complies with the WEEE
Directive (2002/96/EC) marketing
requirements. The affixed label indicates
that you must not discard this
electrical/electronic product in domestic
household waste.
Product category: With reference to the
equipment types in the WEEE Directive
Annexure I, this product is classed as a
“Monitoring and Control instrumentation”
product.
Do not dispose in domestic household
waste.
symbol on a product, you must refer to the
manuals for specific Warning or Caution
information to avoid personal injury or damage to
the product.
Notice for European Community: This product
complies with the relevant European legal
Directives: EMC Directive 89/336/EEC and Low
Voltage Directive 73/23/EEC.
General Recycling Mark for plastic parts used in
the product.
Safety requirements for electrical equipment for
measurement, control, and laboratory use
CAN/CSA C22.2 No. 1010.1 (1993) UL 3101, 3111
(First Editions). This equipment has also been
evaluated to IEC 61010 edition 1 including
amendments 1 and 2.
Conformity Mark of the Australian ACA for EMC
compliance.
7.15 Coupling between Channels ............................................................................................... 651
A Appendix ............................................................................................................................................. 653
A.1 Coupled Parameters when channel coupling is on .......................................................... 653
A.4 Preparing a USB Flash Drive using Windows Vista® ....................................................... 668
Introduction
Keysight
Technologies’
81150A / 81160A
Pulse Function
Arbitrary Noise
Generator
The Keysight Technologies 81150A and 81160A is a Pulse Pattern and
Function Arbitrary Noise Generator with built-in arbitrary waveform and
pulse capabilities
Its combination of bench-top and system features makes this Pulse Function
Arbitrary Noise Generator a versatile solution for your testing requirements
now and in the future.
Features and
Benefits
81150A: 1 Hz-120 MHz pulse generation with variable rise/fall time
81150A: 1 Hz-240 MHz sine waveform outputs
81160A: 1 Hz-330 MHz pulse generation with variable rise/fall time
81160A: 1 Hz-500 MHz sine waveform outputs
Pulse, sine, square, ramp, noise and arbitrary waveforms
FM, AM, PM, FSK, PWM modulation capability
One or two channels
81150A: 14-bit, 2GSa/s, 512 KSa deep arbitrary waveform memory
per channel
81160A: 14-bit, 2.5GSa/s, up to 256 KSa deep arbitrary waveform
memory per channel
USB, GPIB and LAN connectivity.
Glitch free change of timing parameters delay, frequency, transition
time, width, duty cycle
LXI class C compliant
Benchtop Testing
The 81150A / 81160A features a graphic display showing all pulse
parameters at a glance. The cursor keys and the modify knob allow fast and
simple operation.
The user interface is designed to minimize the time invested in getting
familiar with the instrument. After familiarization, the instrument supports
quick setups of signals. This leaves you free to concentrate on the
measurement task and testing of the DUT.
1 Introduction
81150Aand81160AUser’sGuide13
Appendix
What’s inside this
Manual
This manual provides detailed information about the following:
Front-Panel Menu Operation
Features and Functions
Remote Programming Reference
Error Messages
Application Programs
Tutorial
Purpose of this
Manual
The purpose of this manual is to enable you to install, initialize, and start the
81150A / 81160A and to understand the front-panel menu features of the
81150A / 81160A.
Who should read
this Manual
This manual is intended for testers and Engineers who will be using the
81150A / 81160A to test other devices.
How this document
is organized
This section provides information on the chapters, and their content.
Topic
What information does it contain?
Introduction
Introduces the 81150A / 81160A, defines the purpose and
intended audience of this manual; explains how information is
organized in this manual.
Front-Panel Menu Operation
Introduces you to the Front-Panel Menu and describes some of
the menu features of the 81150A / 81160A Pulse Pattern and
Function Arbitrary Noise Generator.
Features and Functions
Gives a detailed description of the 81150A / 81160A’s
capabilities and operation. You will find this section useful when
you are operating the 81150A / 81160A from the front panel or
over the remote interface.
Navigating this manual
14
Introduction
Topic
What information does it contain?
Remote Programming Reference
Contains reference information to help you program the 81150A
/ 81160A over the remote interface.
Error Messages
Describes the error reporting model that is used by the 81150A /
81160A.
Application Programs
Describes the various types of programming examples available
for the 81150A / 81160A and where to find them.
Tutorial
Gives an overview of the internal operations of the 81150A /
81160A.
Terms and
conventions used in
this manual
The following table lists the terms and conventions used in this manual:
The icon...
Indicates…
A note or important information.
A tip
A caution or warning
Notes within a table
Acronyms used in
this manual
The following table lists the acronyms and abbreviations used in this manual:
Acronym
Explanation
DHCP
Dynamic Host Configuration Protocol
DNS
Domain Name Service
DUT
Device Under Test
Conventions
Acronyms used in this Document
81150A and 81160A User’s Guide15
Appendix
References
The Getting Started Guide along with this manual forms a part of the 81150A
and 81160A product documentation suite.
16
Front-Panel Menu Operation
Introduction
This section introduces the front-panel menu and describes the menu
features of the 81150A / 81160A Pulse Pattern and Function Arbitrary Noise
Generator.
2 Front-Panel Menu Operation
81150Aand81160AUser’sGuide17
Appendix
What’s inside this
Chapter
The following topics are discussed in this section:
The Front Panel
Help is Available
The Front-Panel Display at a Glance
Menu Mode
Graph Mode
The Front-Panel Number Entry
The Rear Panel
Preparing the 81150A / 81160A for Use
Using the Built-in Help System
Selecting the Mode of Operation
Selecting Trigger Mode
Selecting the Waveform
Selecting the Advanced Mode
Setting the Output Frequency
Setting the Output Amplitude
Selecting Delay
Selecting DC Volts
Setting a DC Offset Voltage
Setting the Duty Cycle of a Square Wave
Setting the High-Level and Low-Level Values
Configuring a Pulse Waveform
Setting up a Pattern
Viewing a Waveform Graph
Outputting a Stored Arbitrary Waveform
Selecting the Output Termination
Outputting a Modulated Waveform
Outputting an FSK Waveform
Outputting a PWM Waveform
Outputting a Frequency Sweep
Outputting a Burst Waveform
Triggering a Sweep or Burst
Storing the Instrument State
Configuring the Remote Interface
Resetting the 81150A / 81160A
18
Front-Panel Menu Operation
Introduction
The instrument is mainly operated from the front panel, when used for
benchtop testing.
This section explains the Keys, Functions, Inputs/Outputs and Controls,
seen on the Front Panel of the 81150A / 81160A.
Ch 1
Ch 2
Coupling
Graph
Cancel
789
4
56
123
0.+/-
Out 2Out 2Out 1Out 1
Man Cont Pulse Square M od
Store/
Recall
Utility
Help
Sweep
BurstArb
Ramp
Noise
SineTrig
Gat ed
max.±15V
Trigger Out 2 Strobe Out 2External In
Trigger Out 1 Strobe Out 1
81150A
Pulse Function Arbitrary Generator
120 MHz
Local
Power SwitchMenu SoftkeysCancel
Graph / Local
Numeric KeypadInputs / Outputs
USB Host
Channel 1 Selection
Channel Coupling
Channel 2 Selection
Navigation
Keys
Rotary Knob
Cursor Keys
Trigger
Modes
Waveform
Type
Advanced
Modes
Special
Function Keys
Ch 1
Ch 2
Coupli ng
Gr aph
Canc el
L oc a l
m a x . ± 1 0 V
Ex t e r na l I nSy nc Ou t A Sy nc Ou t B
7
8
9
456
123
0.
+/-
Out2Out 2Out1Out 1
ManContPulse Square Mod
Store/
Recall
Utility
Help
Sweep
BurstArb
Ramp
Noise
SineTrig
Gated
Channel 1 Selection
Channel Coupling
Channel 2 Selection
USB Host
Power Switch
Menu Softkeys Cancel
Graph / Local
Numeric KeypadInputs / Outputs
Special
Function Keys
Advanced
Modes
Waveform
Type
Trigger
Modes
Rotary Knob
Cursor Keys
Navigation
Keys
2.1 The Front Panel
Front Panel of the 81150A
81150A and 81160A User’s Guide19
Front Panel of the 81160A
Appendix
Power Switch
The front panel switch is used to switch on and off the instrument.
When the front panel switch is off, the instrument is in standby mode. The
instrument is disconnected from the AC line power only by disconnecting
the power cord.
USB
The Front Panel contains a USB host connector. It is intended to connect
USB drives to store instrument states and waveforms on an external
memory.
Menu Softkeys
The five keys below the main display screen are called softkeys (softwarecontrolled keys). The current function of each softkey is indicated in the
corresponding box on the display.
Some softkeys hold additional symbols to indicate that they provide
extended functionality.
Yellow Triangles: These are visible on the five softkeys on the
Front Panel. The yellow triangles indicate that there are more choices
further, and keep pressing to view the available options.
White Rectangles: Indicate that by pressing them, you can
enter into another screen.
The following examples explain the above-mentioned points:
Pressing this softkey will toggle between
Frequency and Period representation.
Pressing this softkey will open the output
configuration screen.
More Key
Pressing the MORE key switches to the next layer of softkeys on the current
screen.
20
Front-Panel Menu Operation
Pressing the Cancel key cancels the selection/input and helps you exit out
of a screen. But the functionality of exiting out of a screen is limited to a
few screens, like selecting an Arbitrary waveform.
The 81150A / 81160A consists of two channels and operates in two
different modes of operation:
Coupling off: The two channels operate entirely independent. Frequency
generation for both channels is based on same clock reference.
Coupling on: The frequency, trigger mode, waveform type and advanced
mode are identical for both channels. The delay between the channels is
specified.
A modulation input (for AM, FM, PM, FSK, PWM) for each channel is
provided on the back-panel. In the two channel version each channel can
modulate the other channel.
Refer to the Appendix for a list of all coupled parameters.
Press the Graph key to view a graphical representation of the waveform.
Press it again to exit from that mode.
Not every screen has a graphical representation, e.g. like the Trigger Mode
Screen.
Numeric Keypad,
Cursor Keys,
Rotary Knob
These keys are used to select and modify parameters when operating the
instrument.
Use the numeric keypad to enter numbers and the menu softkeys to select
units.
Use the Rotary knob and cursor keys to modify the displayed number.
81150A and 81160A User’s Guide21
Appendix
Inputs/Outputs
The major inputs and outputs of the instrument are available at the front
panel:
The external input (EXTERNAL IN) can be used to connect an
external arming source (triggered or gated modes).
The trigger signal (TRIGGER OUT) marks the start of the pulse period
or of parts of a pattern (see Mode/Trigger Screen.). You can set the
output levels according to the used technology (TTL and ECL) or
enter test-specific values. Trigger Out is available as a physical BNC
connector at the 81150A. For the 81160A, a ‘Logical Trigger Signal’ is
generated for channel 1 and channel 2 internally and is routed to the
physical BNC connector Sync Out A and/or Sync Out B using a
configurable switch matrix.
The strobe signal (STROBE OUT) marks beginning and end of a burst
in Burst mode. Strobe Out is available as a physical BNC connector
at the 81150A. For the 81160A, a ‘Logical Strobe Signal’ is generated
for channel one and channel 2 internally and is routed to the physical
BNC connector Sync Out A and/or Sync Out B using a configurable
switch matrix.
81160A only: The sync signal (SYNC OUT) outputs the ‘Logical
Trigger Signal’ and/or the ‘Logical Strobe Signal’ at the front panel
BNC connector. The instrument offers full flexibility how ‘Logical Trigger Signal’ and/or ‘Logical Strobe Signal’ is routed to Sync Out A
and/or Sync Out B.
The OUTPUT connectors provide the signal output (normal and
inverted) and the indicators show the current state of the output (on
or off).
Special Function
Keys
The instrument provides the following special function keys:
Man
Store/Recall
Help
The function of each of these keys is explained below.
In triggered or gated mode, the MAN key can be used to manually arm
and/or trigger the instrument.
22
Front-Panel Menu Operation
Trigger Modes
The trigger modes are explained below in subsequent sections:
See Selecting Trigger Mode section for more details.
Waveform Types
The standard waveforms include: Pulse, Sine, Noise, Square, Ramp and
Arbitrary.
The predefined arbitrary waveforms include: Exponential rise, exponential
fall, sin(x)/x, cardiac, gaussian, haversine, negative ramp, and DC.
Advanced modes of
Operation
There are three advanced modes of operation available:
Modulation: Selects the modulation type from AM, FM, PM, FSK,
PWM.
Sweep: For frequency sweeps.
Burst: Repeats selected waveform n times.
The following sections provide more details on the modulation types.
The Store/Recall key can be used to store to/recall from 1 to 4 individual
settings in the instrument memory. In the internal memory location 0 there
is a default setting stored.
The Utility key enables you to enable/disable DC mode, change the Output
Setup, and also contains information about the I/O Interfaces and the
system settings of the 81150A / 81160A.
The Helpkey provides access to the instrument’s integrated help or in
warning or error state, access to Warning/Error Report screen.
81150A and 81160A User’s Guide23
Appendix
Introduction
Whenever you are in doubt or the instrument signals warnings or errors,
press the Help key.
Pressing the Help key opens the ‘Main Help Page’.
This main help page is the table of contents of the integrated help system
which lists all the help topics. You can obtain information by selecting the
corresponding link and then by pressing the Follow Link softkey.
Warnings and
Errors
If there are warnings or errors pending (indicated by a flushing W or E on
the screen), pressing the Help key displays a list of the current messages.
Using the Error Queue and Warning softkeys, you can toggle between both
lists. For more information on warnings and errors, see Warnings and Errors.
Exit Help
To exit Help, press the Help key again, or press any other parameter screen
key, e.g. Pulse, Sine, etc.
2.2 Help is Available
24
Front-Panel Menu Operation
Introduction
This section explains the Menu and the Graph mode as seen on the Front
Panel of the 81150A / 81160A.
Introduction
This section explains the Menu as seen on the Front Panel of the 81150A /
81160A.
Channel Information
Units
Channel 1 Information
(Channel addition)
Channel 2 Information
(Frequency Coupling)
Softkey Labels
Trigger Information for
Channel 1 and 2
Numeric Readout
2.3 The Front-Panel Display at a Glance
2.3.1 Menu Mode
81150A and 81160A User’s Guide 25
Appendix
Introduction
To enter the Graph mode, press the key.
To exit, press the key again.
Not all screens have a graphical representation.
The trigger mode screen will always be in textual mode, even if graph
mode is enabled.
2.3.2 Graph Mode
26
Front-Panel Menu Operation
Entering/modifying
numbers
You can enter numbers or modify the displayed number from the front-panel
using one of the following two methods:
Use the numeric keypad and menu softkeys to select the units.
The left/right arrows below the Rotary Knob are used to move left
and right to select the digit to be modified on a given screen.
When setting the cursor to the ‘exponent field’ the exponent can be
changed via the Rotary Knob (only if the resulting number does fall
within the allowed range)
The cursor position is remembered when the cursor is placed on the
leftmost digit and the value is decreased from 1 to 0. In this case the
cursor changes it’s color to green for some seconds and the cursor
position will be set back to the previous one when incrementing the
value again. For example, select the frequency to be edited and set
the cursor to the leftmost digit and then decrement the value by
turning the Rotary Knob counter clockwise. The cursor will move one
digit to the right when the digit would go from 1 to 0 and changes it’s
color to green. When incrementing while the cursor is green, it will
jump back to the initial digit when crossing the 0 to 1 border.
The left arrow key can be used to delete the digit left to the input
cursor when entering values with the numeric keypad.
2.4 The Front-Panel Number Entry
81150A and 81160A User’s Guide27
Appendix
Introduction
The rear panel contains:
GP-IB connector
USB device connector
LAN connector
These three are used for remote control of the instrument.
Channel 1 Modulation In
Channel 2 Modulation In
10 MHz Clock Ref-In
10 MHz Clock Ref-Out
A USB Host Connector is used to connect external USB storage device for
storing instrument settings or software updates.
456789213
2.5 The Rear Panel
Rear panel of the 81150A
1 USB Interface Connector (Host type for external mass memory)
2 USB Interface Connector (device type for remote programming)
3 LAN Interface Connector
4 Channel 1 External Modulation Input Terminal
5 Channel 2 External Modulation Input Terminal
6 External 10 MHz Reference Input Terminal
7 10 MHz Reference output Terminal
8 GPIB Interface Connector
9 Power
28
Front-Panel Menu Operation
G PI B
L AN
M o d u l a t i o n I n 1M o d u l a t i o n I n 21 0 M H z Re f I n1 0 M H z Re f O u t
1V
pp
2 0 0 - 240VAC50- 6 0 H z 1 8 0 W m a x .
1 0 0 - 1 27VAC5 0 - 4 0 0 H z 1 8 0 W m a x .
10Vpk10V
pk
42V
pk
m a x . 5 V
r m s
± 1 0 V
m a x .
m a x . 5 V
r m s
567123
4
89
Rear panel of the 81160A
1 10 MHz Reference Output Terminal
2 External 10 MHz Reference Input Terminal
3 Channel 1 External Modulation Input Terminal
4 Channel 2 External Modulation Input Terminal
5 USB Interface Connector (Host type for external mass memory)
6 USB Interface Connector (device type for remote programming)
7 LAN Interface Connector
8 GPIB Interface Connector
9 Power
81150A and 81160A User’s Guide29
Appendix
Check the List of
supplied items
Before preparing the 81150A / 81160A for use, check the list of supplied
items, given below:
One Power Cord
USB Cable
Product CD
This User’s Guide (if ordered in printed version)
Getting Started Guide (if ordered in printed version)
Certificate of Calibration
Keysight Automation Ready CD
Connect the Power
Cord and turn on
the 81150A /
81160A
The instrument runs a short power-on self test, which takes about 45-50
seconds. The 81150A / 81160A powers up in the sine wave function at 1
MHz with an amplitude of 1 Vpp (into a 50 termination) or the power-down
setting. At power-on, the Output connector is disabled. To enable the Output
connector, press the output key.
If the 81150A /
81160A does not
turn on….
Steps:
Verify that the power cord is firmly connected to the power
receptacle on the rear panel (the power-line voltage is automatically
sensed at power-on).
You should also make sure that the 81150A / 81160A is connected to
a power source that is energized.
Then, verify that the 81150A / 81160A is turned on.
If the power-on tests fail, the instrument automatically switches to the
diagnostics screen and displays the power on messages.
2.6 Preparing the 81150A / 81160A for Use
30
Loading...
+ 642 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.