duced in any form or by any means
(including electronic storage and
retrieval or translation into a foreign
language) without prior agreement and
written consent from Keysight Technologies as governed by United States and
international copyright laws.
Trademarks
Microsoft and Windows are U.S.
registered trademarks of Microsoft
Corporation.
Manual Part Number
53220-90010
Edition
Edition 8, March 2018
Printed in:
Printed in Malaysia
Published by:
Keysight Technologies
Bayan Lepas Free Industrial Zone,
11900 Penang, Malaysia
Technology Licenses
The hardware and/or software
described in this document are furnished under a license and may be
used or copied only in accordance with
the terms of such license.
Declaration of Conformity
U.S. Government Rights
The Software is “commercial computer
software,” as defined by Federal Acquisition Regulation (“FAR”) 2.101. Pursuant to FAR 12.212 and 27.405-3 and
Department of Defense FAR Supplement (“DFARS”) 227.7202, the U.S.
government acquires commercial computer software under the same terms
by which the software is customarily
provided to the public. Accordingly,
Keysight provides the Software to U.S.
government customers under its standard commercial license, which is
embodied in its End User License
Agreement (EULA). The license set
forth in the EULA represents the
exclusive authority by which the U.S.
government may use, modify,
distribute, or disclose the Soft-ware.
The EULA and the license set forth
therein, does not require or per-mit,
among other things, that Keysight: (1)
Furnish technical information related to
commercial computer
commercial computer software
documentation that is not customarily
provided to the public; or (2) Relinquish
to,
or otherwise provide, the government rights in excess of these rights
customarily provided to the public to
use, modify, reproduce, release, per-form,
display, or
computer software
com-puter software documentation.
No additional government
requirements beyond those set forth in
the EULA shall apply, except to the
extent
licenses are explicitly required from all
providers of commercial computer
software pursu-ant to the FAR and the
DFARS and are set forth specifically in
writing
Keysight shall be under no obligation to
update, revise or otherwise modify the
Software. With respect to any
technical data as defined by FAR
2.101, pursuant to FAR 12.211 and
27.404.2 and DFARS 227.7102, the
U.S. government acquires no greater
than Limited Rights as defined in FAR
27.401 or DFAR 227.7103-5 (c), as
applicable in any technical data.
disclose commercial
that those terms, rights, or
else-where in the EULA.
soft-ware or
or commercial
Warranty
THE MATERIAL CONTAINED IN THIS
DOCUMENT IS PROVIDED “AS IS,”
AND IS SUBJECT TO BEING
CHANGED, WITHOUT NOTICE, IN
FUTURE EDITIONS. FURTHER, TO THE
MAXIMUM EXTENT PERMITTED BY
APPLICABLE LAW, KEYSIGHT DISCLAIMS ALL WARRANTIES, EITHER
EXPRESS OR IMPLIED, WITH REGARD
TO THIS MANUAL AND ANY INFORMATION CONTAINED HEREIN, INCLUDING BUT NOT LIMITED TO THE
IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A
PARTICULAR PURPOSE. KEYSIGHT
SHALL NOT BE LIABLE FOR ERRORS
OR FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES IN CONNECTION
WITH THE FURNISHING, USE, OR
PERFORMANCE OF THIS DOCUMENT
OR OF ANY INFORMATION CONTAINED HEREIN. SHOULD KEYSIGHT
AND THE USER HAVE A SEPARATE
WRITTEN AGREEMENT WITH WARRANTY TERMS COVERING THE MATERIAL IN THIS DOCUMENT THAT
CONFLICT WITH THESE TERMS, THE
WARRANTY TERMS IN THE SEPARATE
AGREEMENT SHALL CONTROL.
Safety Information
A CAUTION notice denotes a hazard. It
calls attention to an operating procedure, practice, or the like that, if not
correctly performed or adhered to,
could result in damage to the product
or loss of important data. Do not proceed beyond a CAUTION notice until
the indicated conditions are fully
understood and met.
A WARNING notice denotes a hazard. It
calls attention to an operating procedure, practice, or the like that, if not
correctly performed or adhered to,
could result in personal injury or death.
Do not proceed beyond a WARNING
notice until the indicated conditions are
fully understood and met.
2Keysight 53210A/53220A/53230A Assembly Level Service Guide
Software Updates/Licenses
Periodically, Keysight releases software updates to fix known defects and
incorporate product enhancements.
A portion of the software in this product is licensed under terms of the General
Public License Version 2 (“GPLv2”).
This product utilizes Microsoft Windows CE. Keysight highly recommends that all
Windows-based computers connected to Windows CE instruments utilize current
anti-virus software.
Keysight 53210A/53220A/53230A Assembly Level Service Guide3
Assistance
This product comes with the standard product warranty. Warranty options,
extended support contacts, product maintenance agreements and customer
assistance agreements are also available. Contact your nearest Keysight
Technologies Sales and Service office for further information on Keysight
Technologies' full line of Support Programs.
Certification
Keysight Technologies certifies that this product met its published specifications
at time of shipment from the factory. Keysight Technologies further certifies that
its calibration measurements are traceable to the United States National Institute
of Standards and Technology, to the extent allowed by the Institute's calibration
facility, and to the calibration facilities of other International Standards
Organization members.
Exclusive Remedies
The remedies provided herein are the customer's sole and exclusive remedies.
Keysight Technologies shall not be liable for any direct, indirect, special,
incidental, or consequential damages, whether based on contract, tort, or any
other legal theory.
4Keysight 53210A/53220A/53230A Assembly Level Service Guide
Lithium Battery Recycling
CAT I
The 53210A/53220A/53230A counters contain a 3 V “coin cell” lithium battery.
Keysight recommends that this battery be replaced every year at the instrument’s
1-year calibration interval. Replacement procedures are provided in the 53210A/
53220A/53230A Service Guide.
Option 300 provides battery operation of the 53210A/53220A/53230A using a 12
lithium battery. Performance of this battery will degrade over time as a function of
the number of battery charge/discharge cycles.
Refer to local, state, or Federal regulations when disposing of, or recycling either
battery.
Safety Symbols
The following symbols on the instrument and in the documentation indicate
precautions which must be taken to maintain safe operation of the instrument.
Standby Power. Unit is not completely
disconnected from AC mains when
power switch is in standby position.
Caution, risk of danger (refer to this
manual for specific Warning or Caution
information)
Frame or chassis (ground) terminal
IEC Measurement Category I. Do NOT
connect inputs to AC mains or to
circuits derived from AC mains.
Keysight 53210A/53220A/53230A Assembly Level Service Guide5
Safety Information
The following general safety precautions must be observed during all phases of
operation of this instrument. Failure to comply with these precautions or with
specific warnings or operating instructions in the product manuals violates safety
standards of design, manufacture, and intended use of the instrument. Keysight
Technologies assumes no liability for the customer's failure to comply with these
requirements.
General
Do not use this product in any manner not specified by the manufacturer. The
protective features of this product may be impaired if it is used in a manner not
specified in the operation instructions.
Before applying power
Verify that all safety precautions are taken. Note the instrument's external
markings described under "Safety Symbols".
Ground the instrument
The Keysight 53210A/53220A/53230A is provided with a grounding-type power
plug. The instrument chassis and cover must be connected to an electrical ground
to minimize shock hazard. The ground pin must be firmly connected to an
electrical ground (safety ground) terminal at the power outlet. Any interruption of
the protective (grounding) conductor or disconnection of the protective earth
terminal will cause a potential shock hazard that could result in personal injury.
Fuses
The Keysight 53210A/53220A/53230A is provided with an internal line fuse
appropriate for the line voltages listed on the instrument. This fuse is not user
accessible.
6Keysight 53210A/53220A/53230A Assembly Level Service Guide
Do not operate in an explosive atmosphere
Do not operate the instrument in the presence of flammable gases or fumes.
Do not operate near flammable liquids
Do not operate the instrument in the presence of flammable liquids or near
containers of such liquids.
Do not remove the instrument cover
Only qualified, service-trained personnel who are aware of the hazards involved
should remove instrument covers. Always disconnect the power cable and any
external circuits before removing the instrument cover.
Do not modify the instrument
Do not install substitute parts or perform any unauthorized modification to the
product. Return the product to a Keysight Sales and Service Office for service and
repair to ensure that safety features are maintained.
In case of damage
Instruments that appear damaged or defective should be made inoperative and
secured against unintended operation until they can be repaired by qualified
service personnel.
Cleaning the instrument
Clean the outside of the instrument with a soft, lint-free, slightly-dampened cloth.
Do not use detergents or chemical solvents.
Keysight 53210A/53220A/53230A Assembly Level Service Guide7
WARNING
Do not connect the input channels of the 53210A/53220A/53230A to AC
WARNING
1 Watt Max into 50W
!
Channel / Setup
WARNING
WARNING
line-voltage mains or to circuits derived from AC mains. The instrument
must be used in CAT I (isolated from mains) applications only. Do not use in
other IEC Measurement Category (CAT II, CAT III, or CAT IV) applications.
Failure to observe these precautions may result in electric shock and serious
personal injury.
The BNC shells of the input terminals are connected to the instrument
chassis. Verify signal polarity before making any connections to the input
terminals.
Protection limits
The Keysight 53210A/53220A/53230A 350 MHz Universal Frequency Counter/
Timer provides protection circuitry to prevent damage to the instrument and to
protect against the danger of electric shock, provided the Protection Limits are
not exceeded and the instrument is properly grounded. To ensure safe operation
of the instrument, do not exceed the Protection Limits defined on the front panel:
During battery operation, the maximum measured signal supplied by the
user is +42 V.
Product Options 201/202 add parallel Channel 1 and Channel 2 inputs to the
rear panel of the instrument. Signals on the center conductor of either
panel’s channel BNCs are also present on the corresponding center
conductor of the BNC on the opposite panel.
8Keysight 53210A/53220A/53230A Assembly Level Service Guide
Installing the instrument
WARNING
WARNING
The Keysight 53210A/53220A/53230A operates in the following line-voltage
ranges:
100 V to 240 V, 50 Hz to 60 Hz
100 V to 127 V, 440 Hz
90 VA Max
Instrument ventilation is through the sides and rear. Do not obstruct the
ventilation holes in any of these locations.
Battery operation
When operating the 53210A or 53220A under battery power (Option 300), failure
to observe the following warnings may result in damage to the instrument,
electric shock, and serious personal injury:
Connect the instrument chassis to earth ground during battery operation to
minimize shock hazard. Any interruption or disconnection of the protective
earth terminal will cause a potential shock hazard that could result in
personal injury.
Keysight 53210A/53220A/53230A Assembly Level Service Guide9
Under battery power, the instrument chassis may float to the potential of the
measured signal supplied by the user.
Waste Electrical and Electronic Equipment (WEEE) Directive
This instrument complies with the WEEE Directive marking requirement. This
affixed product label indicates that you must not discard this electrical or
electronic product in domestic household waste.
Product category:
With reference to the equipment types in the WEEE directive Annex 1, this
instrument is classified as a “Monitoring and Control Instrument” product.
The affixed product label is as shown below.
Do not dispose in domestic household waste.
10Keysight 53210A/53220A/53230A Assembly Level Service Guide
Keysight 53210A/53220A/53230A Performance Test Record (Tests 1 to 7)
90
Verifying specifications.
23
1Performance Tests
Introduction
This chapter provides procedures to test the electrical performance of
the Keysight 53210A/53220A/53230A Frequency Counters, using the
specifications provided on the CD-ROM (“Keysight 53210A/53220A/53230A
Specifications”) that came with your counter. Successful completion of all the
performance tests verifies the counter is meeting all factory specifications.
– Keysight 53210A/53220A/53230A Operational Verification and Performance
Test Record
– Keysight 53210A/53220A/53230A Performance Test Record (Tests 1 to 7)
24Keysight 53210A/53220A/53230A Assembly Level Service Guide
Chapter overview
Power-on test
When the counter is first turned ON, a limited turn-on test is performed to verify
the counter is sufficiently functional to perform basic operations.
Self-test
Running the counter Self-Test initiates a series of checks against internal
standards. Successful completion of the Self-Test provides a high confidence level
that the counter is completely functional. No external test equipment is required
for this operator-initiated test.
Operational verification
The operational verification is an abbreviated check that may be performed to give
a strong degree of confidence that the instrument is capable of making basic
measurements accurately without running the complete performance tests. An
operational verification is useful for incoming inspection and routine preventative
maintenance.
Complete performance verification tests
Performance Tests1
The complete performance verification tests verify the specifications listed on the
Keysight 53210A/53220A/53230A Product Reference CD-ROM as “Keysight
53210A/53220A/53230A Specifications.” All tests can be performed without
accessing the inside of the instrument. The complete performance tests are
recommended as acceptance tests upon initial receipt of your counter. They will
also help you to become familiar with its performance and operation. Complete
performance tests should be run at least once a year to ensure the counter is
meeting all specifications.
Quick performance checkA combination of selected performance tests
(marked ‘Q’) plus the counter Self-Test provides a simple method to achieve a
high confidence in the counter’s ability to both functionally operate and to meet
specifications. These tests represent the absolute minimum set of performance
checks recommended following any service activity.
Keysight 53210A/53220A/53230A Assembly Level Service Guide25
1Performance Tests
Auditing the counter’s performance for the quick check points (designated by a
‘Q’) verifies performance for normal accuracy drift mechanisms.
This test does not check for abnormal component failures. To perform the quick
performance check, do the following:
– Perform the counter Self-Test as shown on page 31.
– Perform only the performance verification tests indicated with the letter Q at
the beginning of the test.
– If the counter fails the quick performance check, repair may be required. Refer
to “Troubleshooting the Counter” on page 160.
Remote testing of the counter
All of the counter tests may be run using an automated test system and remote
programming commands. Three interfaces are provided in each counter to allow
operation over an interface from a remote computer: GPIB, LAN, and USB. Refer
to the SCPI Programmer’s Reference for information on the remote commands
available for controlling the counter.
Performance test record
The results of the operational verification, and performance tests should be
recorded on a copy of the Performance Test Record, located at the end of the
complete performance tests section in this chapter.
Test considerations
For optimum performance, all performance tests should comply with the following
recommendations:
– Ensure that the ambient temperature for running the Performance Tests is
stable and between 18 degrees C and 28 degrees C. Ideally, the tests should
o
be performed at 23
– Ensure that ambient relative humidity is less than 80%.
– See page 37 for warm-up considerations.
– Keep the instrument cables as short as possible, consistent with the
impedance requirements. Use only RG-58 or equivalent 50 Ω cable for
measurements.
26Keysight 53210A/53220A/53230A Assembly Level Service Guide
C ±1o C.
– To verify the total number of accuracy places for frequency measurements, it
NOTE
may be necessary to turn off Auto-Digit control. Use this set of key depressions
to turn AutoDigits OFF:
[Digits] / (AutoDigits) / <Off>
– The softkey menus shown in the test procedures represent the functionality of
the 53230A. The softkey menus for the 53210A and 53220A may be different
depending on the function selected.
Recommended calibration cycle
The counter requires periodic verification of operation. Depending on the use and
environmental conditions, aging, and measurement accuracy required, the
counter should be checked using the operational verification procedure at least
once every year. The recommended calibration cycle for specification accuracy is
also one year.
Internal auto-calibration
An Internal Auto-Calibration feature is available (see Chapter 2, "Service") that
uses internal instrument standards to refine the accuracy of the counter. It is
recommended that Auto-Cal be performed regularly (such as at the beginning of a
shift change, starting a contemporary series of measurements, after a large
ambient temperature change, etc.) for optimum counter performance.
Performance Tests1
Auto-calibration is also a subset of the counter sel f-test.
Keysight 53210A/53220A/53230A Assembly Level Service Guide27
1Performance Tests
Equipment required
Tab le 1-1Recommended test equipment for operational verification and
InstrumentRequired characteristicsRecommended model
Microwave signal generator
performance tests
100 MHz to 15 GHz; Freq Accy at
6 GHz and 15 GHz:
PW Accy: ±10%
±10ppb;
Keysight E8257D-520-UNW or
equivalent
Pulse generator
RF signal generator
Primary frequency standard
(10 MHz house standard)
DC power supplyAdjustable to 5.000VKeysight 6234A or equivalent
Digital multimeter (DMM)Microvolt accuracyKeysight 34410A or equivalent
Stanford Research Systems
FS725 Rubidium Timebase or
equivalent
Keysight E9637A or Newark
10M4161
28Keysight 53210A/53220A/53230A Assembly Level Service Guide
Power-On Test
NOTE
NOTE
1 Inspect the counter for any damage. Make sure no cables are connected to the
2 Connect the power cord to the counter, and connect the other end of the
It is normal for the fan in the counter to run at different speeds as necessary
(depending upon internal temperature) when the counter is in standby mode.
Power to the timebase is continuous to maintain long term measurement
reliability, and the fan helps maintain timebase temperature stability.
3 Verify the following condition for the Power LED with power off (lower L/H
4 Press the POWER key (lower left-hand corner of the front panel) to turn on the
Performance Tests1
input channels (rear and front panels) of the counter.
power cord to the primary power source.
corner of the front panel).
– No OCXO or OCXO in Off mode - LED is off when power is off.
– OCXO present, heated by AC power - LED is on solid when power
is off.
– OCXO present, heated by battery power - LED blinks when power
is off.
counter. Verify the following results:
– The front panel display turns white as all pixels are illuminated.
– The LED under the power switch is on solid.
– After several seconds, a screen is briefly displayed that shows the status
and ID of the IP Address and the GPIB address. The factory default GPIB
address is 3. (“Disabled” means the GPIB is turned Off under [Utility] / (I/O
Config).
– The counter is now ready to measure frequency of a signal applied to
CHANNEL 1 input as indicated by illumination of the Freq softkey and
Channel 1 annunciator.
An auto-cal is performed at power-on. (See “To run auto calibration:” on
page 148 for more information.)
Keysight 53210A/53220A/53230A Assembly Level Service Guide29
1Performance Tests
NOTE
NOTE
If a fail message is displayed at power-on, refer to “Troubleshooting the
Counter” on page 160 of this service guide.
5 Mark Pass or Fail in the Keysight 53210A/53220A/53230A Power-ON Test
Record, Test 1, at the end of this chapter.
The power-on hardware initialization activities check the following circuits for
operation:
– Front panel display driver
–Power manager
– Battery (if installed)
–GPIB
– Measurement board hardware
30Keysight 53210A/53220A/53230A Assembly Level Service Guide
Self-Test (Q)
NOTE
NOTE
Performance Tests1
A brief power-on test occurs automatically whenever you turn on the counter. This
limited test assures that the counter is functional.
It is recommended that the counter Self-Test be performed to ensure that the
instrument is completely functional before running the Operational Verification
or Performance Tests.
SELF-TEST and the remote command *TST perform the same internal tests.
An auto-calibration is performed as a subset of the counter Self-Test.
(See “Internal auto-calibration” on page 147 for more information.)
After turn-on, or at any time a reasonable operational status needs to be verified,
the internal Self-Test can be run.
Run internal self-test
1 Disconnect any input signal(s) from the counter.
2 Press the Utility key and then the Instr Setup softkey.
Keysight 53210A/53220A/53230A Assembly Level Service Guide31
1Performance Tests
NOTE
The Instrument Setup menu softkeys are displayed:
3 Press the Sel f Test softkey. The display shows:
– “Performing Self-Test. Please wait...” (displayed for several seconds.)
– If the self-test passes, “Self-Test Passed.” is momentarily displayed.
If “Self-Test Failed” is displayed or any other failure is indicated, refer to the
troubleshooting section in Chapter 2, "Service" on page 99.
4 Mark Pass or Fail in the “Keysight 53210A/53220A/53230A Performance Test
Record (Tests 1 to 7)” on page 90.
32Keysight 53210A/53220A/53230A Assembly Level Service Guide
The operational verification consists of checks that may be performed at any time
to give a high degree of confidence that the instrument is operating properly,
without running the complete performance tests.
These checks represent the minimum set of tests recommended following any
option retrofit installations (Refer to Chapter 4, "Retrofitting Options").
If you are unfamiliar with the operation of the counter, you can review the
“Keysight 53210A/53220A/53230A Quick Start Tutorial.” However, the
procedures in this chapter are written so that little operational experience is
necessary. The procedures should be followed in the order in which they appear.
Frequency checks
1 Connect the counter’s rear-panel Int Ref Out as shown in
Figure 1-1 to the front panel Channel 1 input connector.
Figure 1-1Operational verification setup for Keysight 53210A/220A/230A
If the counter has Option 201 (rear panel input connectors) installed, connect
the Int Ref Out to the rear panel Channel 1 input connector and NOT the front
panel Channel 1 input connector.
Keysight 53210A/53220A/53230A Assembly Level Service Guide33
1Performance Tests
NOTE
NOTE
2 Press the Preset key to preset the counter.
3 Verify the display shows a frequency reading of approximately 10,000,000 Hz.
4 Mark Pass or Fail in the “Keysight 53210A/53220A/53230A Performance Test
Record (Tests 1 to 7)” on page 90.
The remaining steps apply only to the Keysight 53220A and Keysight
53230A53230A since the Keysight 53210A does not have a second RF frequency
input.
5 In the same way as for Channel 1, connect the Int Ref Out to the Channel 2 input
connector (front or rear panel).
6 Press the Channel 2 button.
If necessary, press the Freq/Period key to read the Channel 2 frequency.
–Channel 1 annunciator is not lit. Channel 2 annunciator is lit.
7 Verify the display shows a frequency reading of approximately 10,000,000 Hz.
8 Mark Pass or Fail in the “Keysight 53210A/53220A/53230A Performance Test
Record (Tests 1 to 7)” on page 90.
9 Disconnect the test setup.
34Keysight 53210A/53220A/53230A Assembly Level Service Guide
Termination checks (optional)
NOTE
DMM
Ch 1
HI
LO
Once the termination checks have been successfully performed, it is usually not
necessary to run them again unless the counter’s motherboard has been
replaced.
1 Press the Preset key to preset the counter.
2 Connect a digital multimeter (DMM) to CHANNEL 1 as shown in Figure 1-2, and
set it to measure ohms (set DMM to appropriate Ohms range). On the DMM,
use a dual-banana plug-to-BNC connector.
Performance Tests1
Figure 1-2Termination Check Setup
3 Press the Channel 1 key.
– The LED turns on and the Channel 1 Setup softkeys are displayed:
Keysight 53210A/53220A/53230A Assembly Level Service Guide35
1Performance Tests
NOTE
4 Select the following:
– Coupling: DC
–The DC annunciator indicates DC input coupling for Channel 1.
– Impedance: 50 Ω
–The 50 Ω annunciator turns on, indicating a 50 Ω input impedance for
5 Verify the DMM reads approximately 50 Ohms.
6 Mark Pass or Fail in the “Keysight 53210A/53220A/53230A Performance Test
Record (Tests 1 to 7)” on page 90.
7 Select the following:
– Impedance: 1 MΩ.
–The 1 MΩ annunciator turns on, indicating a 1 MΩ input impedance for
Channel 1.
8 Verify the DMM reads approximately 1.000 MOhm.
9 Mark Pass or Fail in the “Keysight 53210A/53220A/53230A Performance Test
Record (Tests 1 to 7)” on page 90.
10 Repeat steps 2 through 9 for RF Channel 2 for the Keysight 53220A/53230A
counters ONLY, since the Keysight 53210A counter does not have a standard
RF Channel 2.
Channel 1.
11 Mark Pass or Fail in the Keysight 53220A/53230A Operational Verification Test
Record, Tests 6a and 6b for Chan 2, starting on page 90.
12 Disconnect the test setup.
This completes the operational verification testing for the Keysight 53210A/
53220A/53230A.
36Keysight 53210A/53220A/53230A Assembly Level Service Guide
The following seven performance tests, when successful, verify that the counter
meets the instrument specifications as provided on the CD ROM that came with
your Keysight 53210A/53220A/53230A Counter. See also the information on
running a ‘QUICK’ performance test on page 25. Before running these tests,
review the “Test considerations” on page 26.
Record the results of the performance tests in the appropriate place on the
“Keysight 53210A/53220A/53230A Performance Test Record (Tests 1 to 7)” on
page 90.
To verify the total number of accuracy places for frequency measurements, it may
be necessary to turn off Auto-Digit control. Use this set of key depressions to turn
AutoDigits OFF:
[Digits] / (AutoDigits) / <Off>
If the instrument is currently located in the calibration environment with the
power removed, turn on the instrument and allow a 45 minute warm-up period
before starting the Performance Test procedures. Upon receipt and delivery to
the calibration environment, turn on the instrument and allow a 2½ hour
warm-up/stabilization period before starting the procedures.
If it is known that the instrument has been in storage or decommissioned
(unused) for several months, turn on the instrument and allow it to stabilize for
72 hours. Periodically monitor the aging (drift) rate relative to the counter’s
24-hour specification. If the aging rate exceeds the 24-hour specification, allow
an additional stabilization period and continue to monitor the rate relative to the
specification. The counter has stabilized when aging remains within the 24-hour
rate. Otherwise, the 45 minute or 2½ warm-up period applies.
An instrument that has not or will not be used for several months should be
stored in the same ambient environment and orientation in which it will
eventually be used.
Keysight 53210A/53220A/53230A Assembly Level Service Guide37
1Performance Tests
Tab le 1-2The Keysight 53210A/53220A/53230A performance tests
[a]
Test description
Test 1: Absolute time base
Test 2: Frequency accuracy - channels 1 and 2
Test 3: Single period
Test 4: Pulse width
Test 5: Time interval
Test 6: Frequency accuracy - microwave channel options 106 and 115
Test 7: Pulse burst frequency - 53230A channel 3 option 150
[a] Other counter measurement functions are mathematically derived by the processor from the
parameters verified by these performance tests.
Test 1: (Q) Absolute time base
This test uses the Keysight 53210A/53220A/53230A Frequency Counters to
measure the absolute frequency of the internal 10 MHz clock.
Equipment
– FS725 Rubidium timebase
– CG635 signal generator (E8257D-520 is a valid substitute)
– Keysight 10503 BNC cables (2)
CG635 signal generator setup
1 Set Output Channel: Q
2 Set Output Level: +7 dBm
3 Set Frequency: 10 MHz
38Keysight 53210A/53220A/53230A Assembly Level Service Guide
Performance Tests1
FS725
Rubidium
Timebase
CG635
Signal
Generator
Ch 1
Counter setup
1 Connect the equipment as shown in Figure 1-3.
Figure 1-3Absolute timebase test setup
2 Press the Preset key on the front panel to preset the counter.
–Display shows:
– Instrument has been preset.
– Channel 1 annunciator is lit.
3 Press the Trigger key.
– The following softkeys are displayed:
4 Press and set:
– Trigger Source: Manual
Keysight 53210A/53220A/53230A Assembly Level Service Guide39
– Wait For Trig is displayed.
– Rdgs/Trigger: 100.
1Performance Tests
*
*Opt 150
*Opt 150
*
NOTE
5 Press the Freq Period key.
– The following softkeys are displayed:
6 Press the Freq softkey.
– The following softkeys are displayed:
7 Press and set:
– Noise Rejection: Off
8 Press the Frequency/Period key.
– The following softkeys are displayed:
–If Ad vanced is appears, press Auto to select Auto measurement mode.
The 53210A does not have an Auto measurement mode.
40Keysight 53210A/53220A/53230A Assembly Level Service Guide
9 Press the Channel 1 input key.
– The following softkeys are displayed:
10 Press and set:
– Coupling: DC
– Input Impedance: 50 Ω
– Range: 5 V
– Level Setup / Auto Level: Off
– Level Setup / Input Level: 0.000 V
–BW Limit: Off
–Probe: None
11 Press the Math key.
Performance Tests1
–The Math softkeys are displayed:
12 Press to set:
–All Math: On
Keysight 53210A/53220A/53230A Assembly Level Service Guide41
1Performance Tests
13 Press the Statistics softkey:
–The Statistics softkeys are displayed:
14 Press to set:
– Statistics: On
– The display shows:
Absolute timebase test procedure
1 Press the Trigger key to manually trigger the counter.
– Counter takes 100 readings and fills in Math functions.
Keysight 53210A/53220A/53230A Assembly Level Service Guide55
1Performance Tests
Opt 010 UOCXO
Line
100-240V, 50-60 Hz
100-127V, 400 Hz
90VA Max
USB LAN
Int Ref Out Trig In
Ext Re f In Gate In/Ou t
Ch 1 Opt 201 Ch 2 GP-IB
Opt 106/115
IECS / NMB-001
N10149
ISM 1-A
C US
Opt 106
100 MHz -6 G Hz
+27 dBm Max
!
Opt 300 Battery
Opt 115
300 MHz - 15 GHz
+27 dBm Max
!
FS725
Rubidium
Timebase
Ext Ref In
81134A
Pulse
Generator
Ch 1
Counter setup
1 Connect the equipment as shown in Figure 1-6.
Figure 1-6Pulse wid th test setup
2 Press the Preset key to preset the counter.
– Display shows:
– Instrument has been preset.
– Ext Ref is shown in the upper right-hand corner of the display.
3 Press the Time Interval key.
– The following softkeys are displayed:
56Keysight 53210A/53220A/53230A Assembly Level Service Guide
4 Press the Pulse Wid th softkey.
– The following softkeys are displayed:
5 Press and set:
–Set width: pos
–auto level: Off
– Input level: +0.00 V
– Set noise rejection: Off
6 Press the Trigger key.
Front panel display shows:
Performance Tests1
7 Press and set:
–Source: Manual
– Wait For Trig is displayed
– Rdgs/Trigger: 100
–Trig Delay: 0 Seconds
Keysight 53210A/53220A/53230A Assembly Level Service Guide57
1Performance Tests
8 Press the Channel 1 key:
– The Channel 1 Setup softkeys are displayed:
9 Press and set:
– Coupling: DC
– Input Impedance: 50 Ω
– Range: 5 V
–BW Limit: OFF
10 Press the Math key.
– The following softkeys are displayed:
11 Press and set:
–All math: On
12 Press the Statistics softkey:
–The Statistics softkeys are displayed:
58Keysight 53210A/53220A/53230A Assembly Level Service Guide
13 Press to set:
NOTE
– Statistics: On
– The display shows:
Pulse wid th test procedure
1 Press the Trigger key.
– Counter takes 100 readings and fills in Math functions.
2 Verify the ‘Mean’ is:
– 5 ± 0.4 nSec (53220A)
– 5 ± 0.2 nSec (53230A)
Performance Tests1
3 Verify the ‘Std Dev’ is less than:
– 0.140 nSec (53220A)
– 0.028 nSec (53230A)
4 Record the Mean and Standard Deviation for Channel 1, Test 4a, in the
“Keysight 53210A/53220A/53230A Performance Test Record (Tests 1 to 7)” on
page 90.
5 Repeat steps 1 to 3 for Channel 2 (Test 4b). Be sure to do the Channel 2 setup
first. (See steps 8, 9 in the Counter setup above.)
Selecting Channel 2 automatically resets the statistics.
Keysight 53210A/53220A/53230A Assembly Level Service Guide59
1Performance Tests
NOTE
NOTE
6 Record the Mean and Standard Deviation for Channel 2, Test 4b, in the
“Keysight 53210A/53220A/53230A Performance Test Record (Tests 1 to 7)” on
page 90.
7 Proceed to Test #5 or disconnect the test setup.
Test 5: (Q) Time interval
This test verifies the time interval specifications between Channel 1 and
Channel 2.
Test #5 applies to the 53220A and 53230A only.
Equipment
– FS725 Rubidium timebase
– Keysight 81134A pulse generator
– Keysight 10503 BNC cables (2)
– Keysight 11500E (3.5) mm cables (2)
– Keysight 1250-1200 BNC(f)-to-SMA(m) adapter
The 3.5 mm cables must be the exact same length or erroneous
measurements will occur.
81134A setup
– Output Level: 1 Vpp
– Period: 66.6 nSec
– Pulse Width Ch1: 5 nSec
– Pulse Width Ch2: 5 nSec
– Delay Ch2: 49.9 nSec
60Keysight 53210A/53220A/53230A Assembly Level Service Guide
Counter setup
Opt 010 UOCXO
Line
100-240V, 50-60 Hz
100-127V, 400 Hz
90VA Max
USB L AN
Int Re f Ou t Trig In
Ext Re f In G ate In /Ou t
Ch 1 Op t 201 Ch 2 GP-IB
Opt 106/115
IECS / NMB-001
N10149
ISM 1-A
C US
Opt 106
100 MHz - 6 GHz
+27 dBm Max
!
Opt 300 Battery
Opt 115
300 MHz - 15 GHz
+27 dBm Max
!
FS725
Rubidium
Timebase
Ext
Ref
In
81134A
Pulse
Generator
Ch 1Ch 2
1 Connect the equipment as shown in Figure 1-7.
Performance Tests1
Figure 1-7Time interval (channel 1 to 2) test setup
2 Press the Preset key to preset the counter.
–Display shows:
– Instrument has been preset.
– Ext Ref is shown in the upper right-hand corner of the display.
3 Press the Time Interval key.
– The following softkeys are displayed:
4 Press the Time Interval softkey.
Keysight 53210A/53220A/53230A Assembly Level Service Guide61
1Performance Tests
NOTE
–The Time Interval softkeys are displayed:
5 Press to set Time Int: 1-2.
Time Int 2-1 could also be tested, but only one test is required. If Time Int 2-1 is
used, set the 81134A to delay Channel 1 by 49.9 nSec instead of Channel 2.
6 Press the Start Ch Level softkey.
7 The following softkeys are displayed:
8 Press and set:
–Slope: Pos
–Auto Level: Off
– Level: 0 V
–Noise Rej: Off
9 Press the Back key.
10 Press the Stop Ch Level softkey.
11 Press and set:
–Slope: Pos
–Auto Level: Off
– Level: 0 V
–Noise Rej: Off
62Keysight 53210A/53220A/53230A Assembly Level Service Guide
12 Press the Trigger key.
– Front panel display shows:
13 Press to set:
–Source: Manual
– Wait For Trig is displayed
– Rdgs/Trigger: 100
–Trig Delay: 0 seconds
14 Press the Math key.
– The following softkeys are displayed:
Performance Tests1
15 Press and set:
–All Math: On
Keysight 53210A/53220A/53230A Assembly Level Service Guide63
1Performance Tests
16 Press the Statistics softkey:
–The Statistics softkeys are displayed:
17 Press to set:
– Statistics: On
The display shows:
Time interval test procedure
1 Press the Trigger key.
– The counter takes 100 readings and fills in the math functions.
2 Verify the ‘Mean’ is:
– 49.9 ± 0.5 nSec (53220A)
– 49.9 ± 0.25 nSec (53230A)
3 Verify the ‘Std Dev’ is less than:
– 0.140 nSec (53220A)
– 0.028 nSec (53230A)
4 Record the Median and Standard Deviation, Test 5, in the “Keysight 53210A/
53220A/53230A Performance Test Record (Tests 1 to 7)” on page 90.
5 Proceed to Test #6 or disconnect the test setup.
64Keysight 53210A/53220A/53230A Assembly Level Service Guide
Performance Tests1
NOTE
Test 6a: (Q) Frequency accuracy - channel 2/3, option 106
(6 GHz microwave channel)
This test verifies the frequency accuracy of Option 106 Channel 2 (53210A) or
Channel 3 (53220A/53230A) at two frequencies (100 MHz and 5.88 GHz).
Equipment
– FS725 Rubidium timebase
– Keysight E8257D-520 microwave signal generator
– Keysight 10503A BNC cable (2)
– Keysight 11500E (3.5) mm cable
– Keysight 1250-1250 N(M) to SMA(F)
E8257D-520 setup
– Output Frequency: 100 MHz
– Output Power: -27 dBm
The input power MUST be -27 ±0.5 dBm at the channel 3 input connector. If
necessary, use a 3.5 mm cable with certified loss at 100 MHz or a power meter
to verify input power.
Keysight 53210A/53220A/53230A Assembly Level Service Guide65
For the Keysight 53210A Counter, Option 106 is installed as Channel 2. For the
Keysight 53220A and 53230A Counters, Option 106 is installed as Channel 3. For
this procedure, “Channel 3” is used to refer to the microwave input channel.
66Keysight 53210A/53220A/53230A Assembly Level Service Guide
Performance Tests1
NOTE
3 Press the Channel 3 input key.
– Channel 3 is selected for test input.
There are no parameters to set for Channel 3 for this test.
4 Press the Gate key.
–The Gate softkeys are displayed:
5 Press and set:
– Gate Src: Timed
– Gate Out: On
–Gate Time: 10 mSec
6 Press the Trigger key.
–The Trigger softkeys are displayed:
Keysight 53210A/53220A/53230A Assembly Level Service Guide67
1Performance Tests
7 Press to set:
–Source: Manual
– Wait For Trig is displayed
–Rdgs/Trigger: 100
–Trig Delay: 0 second s
8 Press the Math key.
–The Math softkeys are displayed:
9 Press to set:
–All Math: On
10 Press the Statistics softkey:
–The Statistics softkeys are displayed:
68Keysight 53210A/53220A/53230A Assembly Level Service Guide
Performance Tests1
NOTE
11 Press to set:
– Statistics: On
– The display shows:
Channel 3 (option 106) frequency accuracy test procedure
1 Press the Trigger key.
– The counter takes 100 readings and fills in the math functions.
2 Verify the ‘Mean’ is:
– 100 MHz ± 1.000 Hz (53210A)
– 100 MHz ± 100 mHz (53220A/53230A)
3 Verify the ‘Std Dev’ is less than:
– 1.400 Hz (53210A)
– 700 mHz (53220A)
– 140 mHz (53230A)
4 Record the Mean and Standard Deviation for 100 MHz, Test 6a,
in the Keysight 53210A/53220A/53230A Performance Test Record.
5 Press the Reset Stats softkey.
6 Set the E8257D-520 to 5.88 GHz at -27 dBm.
The input power MUST be -27 ±0.5 dBm at the channel 3 input connector. If
necessary, use a 3.5 mm cable with certified loss at 5.88 GHz or a power meter
to verify input power.
Keysight 53210A/53220A/53230A Assembly Level Service Guide69
1Performance Tests
7 Press the Trigger key.
– The counter takes 100 readings and fills in the math functions.
8 Verify the ‘Mean’ is:
– 5.88 GHz ± 58.800 Hz (53210A)
– 5.88 GHz ± 5.880 Hz (53220A/53230A)
9 Verify the ‘Std Dev’ is less than:
– 82.300 Hz (53210A)
– 41.100 Hz (53220A)
– 8.230 Hz (53230A)
10 Record the Mean and Standard Deviation for 5.88 GHz, Test 6a,
in the Keysight 53210A/53220A/53230A Performance Test Record.
11 Press the Reset Stats softkey.
12 If Opt 150 is installed, proceed to Test 7a. Otherwise, disconnect the test
setup.
70Keysight 53210A/53220A/53230A Assembly Level Service Guide
Performance Tests1
NOTE
Test 6b: (Q) Frequency accuracy - channel 2/3, option 115
(15 GHz microwave channel)
This test verifies the frequency accuracy of Option 115 Channel 2 (53210A) or
Channel 3 (53220A/53230A) at three frequencies (300 MHz, 3.1 GHz, and 15.0
GHz).
Equipment
– FS725 Rubidium timebase
– Keysight E8257D-520 microwave signal generator
– Keysight 10503A BNC cable (2)
– Keysight 11500E (3.5) mm cable
– Keysight 1250-1250 N(M) to SMA(F)
E8257D-520 setup
– Output Frequency: 300 MHz
– Output Power: -23 dBm
The input power MUST be -23 ±0.5 dBm at the channel 3 input connector. If
necessary, use a 3.5 mm cable with certified loss at 300 MHz or a power meter
to verify input power.
Keysight 53210A/53220A/53230A Assembly Level Service Guide71
For the Keysight 53210A Counter, Option 115 is installed as Channel 2. For the
Keysight 53220A and 53230A Counters, Option 115 is installed as Channel 3. For
this procedure, “Channel 3” is used to refer to the microwave input channel.
72Keysight 53210A/53220A/53230A Assembly Level Service Guide
Performance Tests1
NOTE
3 Press the Channel 3 input key.
– Channel 3 is selected for test input.
There are no parameters to set for Channel 3 for this test.
4 Press the Gate key.
–The Gate softkeys are displayed:
5 Press and set:
– Gate Src: Timed
– Gate Out: On
–Gate Time: 10 mSec
6 Press the Trigger key.
–The Trigger softkeys are displayed:
Keysight 53210A/53220A/53230A Assembly Level Service Guide73
1Performance Tests
7 Press to set:
–Source: Manual
– Wait For Trig is displayed
–Rdgs/Trigger: 100
–Trig Delay: 0 second s
8 Press the Math key.
–The Math softkeys are displayed:
9 Press to set:
–All Math: On
10 Press the Statistics softkey:
–The Statistics softkey menu is displayed:
11 Press to set:
– Statistics: On
– The display shows:
74Keysight 53210A/53220A/53230A Assembly Level Service Guide
Performance Tests1
NOTE
Channel 3 (option 115) frequency accuracy test procedure
1 Press the Trigger key.
– The counter takes 100 readings and fills in the math functions.
2 Verify the ‘Mean’ is:
– 300 MHz ± 3.0 Hz (53210A)
– 300 MHz ± 0.3 Hz (53220A/53230A)
3 Verify the ‘Std Dev’ is less than:
– 4.2 Hz (53210A)
– 2.1 Hz (53220A)
– 0.42 Hz (53230A)
4 Record the Mean and Standard Deviation for 300 MHz, Test 6b, in the
Keysight 53210A/53220A/53230A Performance Test Record.
5 Press the Reset Stats softkey.
6 Set the E8257D-520 to 3.1 GHz at -27 dBm.
The input power MUST be -27 ±0.5 dBm at the channel 3 input connector. If
necessary, use a 3.5 mm cable with certified loss at 3.1 GHz or a power meter to
verify input power.
7 Press the Trigger key.
– The counter takes 100 readings and fills in the math functions.
Keysight 53210A/53220A/53230A Assembly Level Service Guide75
1Performance Tests
NOTE
8 Verify the ‘Mean’ is:
– 3.1 GHz ± 31 Hz (53210A)
– 3.1 GHz ± 3.1 Hz (53220A/53230A)
9 Verify the ‘Std Dev’ is less than:
– 43.4 Hz (53210A)
– 21.7 Hz (53220A)
– 4.34 Hz (53230A)
10 Record the Mean and Standard Deviation for 3.1 GHz, Test 6b,
in the Keysight 53210A/53220A/53230A Performance Test Record.
11 Press the Reset Stats softkey.
12 Set the E8257D-520 to 15.0 GHz at -4 dBm.
The input power MUST be -4 ±0.5 dBm at the channel 3 input connector. If
necessary, use a 3.5 mm cable with certified loss at 15 GHz or a power meter to
verify input power.
13 Press the Trigger key.
– The counter takes 100 readings and fills in the math functions.
14 Verify the ‘Mean’ is:
– 15.0 GHz ± 150 Hz (53210A)
– 15.0 GHz ± 15 Hz (53220A/53230A)
15 Verify the ‘Std Dev’ is less than:
– 210 Hz (53210A)
– 105 Hz (53220A)
– 21.0 Hz (53230A)
16 Record the Mean and Standard Deviation for 15.0 GHz, Test 6b,
in the Keysight 53210A/53220A/53230A Performance Test Record.
17 Press the Reset Stats softkey.
18 If Opt 150 is installed, proceed to Test 7c. Otherwise, disconnect the test
setup.
76Keysight 53210A/53220A/53230A Assembly Level Service Guide
Performance Tests1
NOTE
NOTE
Test 7a: (Q) Pulse burst frequency - 53230A channel 3 option 150 with
option 106
This test uses the 53230A to measure both narrow and wide pulse outputs of a
6 GHz carrier.
This test procedure applies only to the 53230A Counter with Option 106 and
Option 150 installed. Option 106 is installed as microwave channel 3.
The Quick Verification (Q) requires only the Narrow Pulse Wid th Burst Frequency
Test be performed.
Equipment
– FS725 Rubidium timebase
– Keysight E8257D-520-UNW microwave signal generator
– Keysight 10503A BNC cable (2)
– Keysight 11500E (3.5) mm cable
– Keysight 1250-1250 N(M) to SMA(F)
E8257D-520-UNW setup
– Output Frequency: 6 GHz
– Output Power: 0 dBm
–Auto-Level Control: Off
–Pulse Mode: On
– Pulse Repetition Interval Period: 1 mSec
– Pulse Width (Narrow): 200 nSec
Keysight 53210A/53220A/53230A Assembly Level Service Guide77
1Performance Tests
NOTE
Opt 010 UOCXO
Line
100-240V, 50-60 Hz
100-127V, 400 Hz
90VA Max
USB L AN
Int Re f Ou t Trig In
Ext R ef In Gate In/O ut
Ch 1 Op t 201 C h 2 GP -IB
Opt 106/115
IECS / NMB-001
N10149
ISM 1-A
C US
Opt 106
100 MHz -6 G Hz
+27 dBm Max
!
Opt 300 Battery
Opt 115
300 MHz - 15 GHz
+27 dBm Max
!
FS725
Rubidium
Timebase
Ext Ref In
E8257D-520-UNW
Microwave
Signal
Generator
Ch 3
The input power MUST be 0 ± 0.5 dBm at the channel 3 input connector. If
necessary, use a 3.5 mm cable with certified loss at 6 GHz or a power meter to
verify input power.
Counter setup
1 Connect the equipment as shown in Figure 1-10.
78Keysight 53210A/53220A/53230A Assembly Level Service Guide
Figure 1-10Option 106/150 microwave channel 3 test setup
2 Press the Preset key to preset the counter.
3 If needed, perform an Auto-Calibration. (Refer to “Internal auto-calibration” on
page 27.)
Performance Tests1
NOTE
4 Press the Channel 3 input key.
– Channel 3 is selected for test input.
There are no parameters to set for Channel 3 for this test.
5 Press the Freq period key.
–The Freq/Period softkeys are displayed:
6 Press the Pulse/Burst softkey.
–The Pulse/Burst menu softkeys are displayed:
Keysight 53210A/53220A/53230A Assembly Level Service Guide79
1Performance Tests
7 Press the Burst Freq softkey.
–The Burst Freq menu softkeys are displayed:
8 Press to select:
–dB Level: -12
–Narrow: On
– Gate Auto: On
9 Press the Trigger key.
–The Trigger softkeys are displayed:
10 Press to set:
–Source: Manual
– Wait For Trig is displayed
–Rdgs/Trigger: 100
–Trig Delay: 0 second s
11 Press the Math key.
–The Math softkeys are displayed:
12 Press to set:
–All Math: On
80Keysight 53210A/53220A/53230A Assembly Level Service Guide
13 Press the Statistics softkey:
–The Statistics softkeys are displayed:
14 Press to set:
– Statistics: On
– The display shows:
Performance Tests1
Channel 3 pulse burst frequency test procedure
1 Press the Trigger key.
– The counter takes 100 readings and fills in the math functions.
Keysight 53210A/53220A/53230A Assembly Level Service Guide81
1Performance Tests
NOTE
2 Verify the ‘Mean’ is:
– 6 GHz ± 6 MHz
3 Verify the ‘Std Dev’ is less than:
–3 MHz
4 Record the Mean and Standard Deviation for Test 7a, in the “Keysight 53210A/
53220A/53230A Performance Test Record (Tests 1 to 7)” on page 90.
5 Press the Reset Stats softkey.
For Quick Verification (Q), only the narrow pulse wid th burst frequency test
needs to be performed.
6 Set the E8257D-520 to a pulse width of 13 uSec (Wide Pulse).
Keysight 53210A/53220A/53230A Assembly Level Service Guide93
1Performance Tests
Test
number
4a
(Q)
Test descriptionMinimumActual readingMaximum
Pulse Width Test
Ch 1 (53220A)
Ch 1 (53230A)
Complete performance tests
4.600 000 000 nSec
4.800 000 000 nSec
Record: Mean
__________________
Record: Std Dev
__________________
Record: Mean
__________________
Record: Std Dev
__________________
5.400 000 000 nSec
0.140 000 000 nSec
5.200 000 000 nSec
0.028 000 000 nSec
4b
(Q)
Pulse Width Test
Ch 2 (53220A)
Ch 2 (53230A)
4.600 000 000 nSec
4.800 000 000 nSec
Record: Mean
__________________
Record: Std Dev
__________________
Record: Mean
__________________
Record: Std Dev
__________________
5.400 000 000 nSec
0.140 000 000 nSec
5.200 000 000 nSec
0.028 000 000 nSec
94Keysight 53210A/53220A/53230A Assembly Level Service Guide
Test
number
5
(Q)
6a
(Option 106)
(Q)
Performance Tests1
Complete performance tests
Test descriptionMinimumActual readingMaximum
Time Interval
Test
53220A
53230A
53220A
53230A
Frequency Accuracy
Channel 3
100 MHz
53210A
53220A
53230A
49.400 000 000 nSecs
49.650 000 000 nSecs
99.999 999 000 MHz
99.999 999 900 MHz
99.999 999 900 MHz
Record: Mean
__________________
__________________
Record: Std Dev
__________________
__________________
Record: Mean
__________________
__________________
__________________
50.400 000 000 nSec
50.150 000 000 nSec
0.140 000 000 nSec
0.028 000 000 nSec
100.000 001 000 MHz
100.000 000 100 MHz
100.000 000 100 MHz
53210A
53220A
53230A
Record: Std Dev
__________________
__________________
__________________
1 400.000 000 mHz
700.000 000 mHz
140.000 000 mHz
Keysight 53210A/53220A/53230A Assembly Level Service Guide95
1Performance Tests
Test
number
6a
(Option 106)
(Q)
Test descriptionMinimumActual readingMaximum
Frequency Accuracy
Channel 3
5.88 GHz
53210A
53220A
53230A
Complete performance tests
5.879 999 941 200 GHz
5.879 999 994 120 GHz
5.879 999 994 120 GHz
Record: Mean
__________________
__________________
__________________
5.880 000 058 800 GHz
5.880 000 005 880 GHz
5.880 000 005 880 GHz
6b
(Option 115)
(Q)
53210A
53220A
53230A
Frequency Accuracy
Channel 3
300 MHz
53210A
53220A
53230A
53210A
53220A
53230A
299.999 997 MHz
299.999 999 7 MHz
299.999 999 7 MHz
Record: Std Dev
__________________
__________________
__________________
Record: Mean
__________________
__________________
__________________
Record: Std Dev
__________________
__________________
__________________
82.300 000 000 Hz
41.100 000 000 Hz
8.230 000 000 Hz
300.000 003 MHz
300.000 000 3 MHz
300.000 000 3 MHz
4.2 Hz
2.1 Hz
0.42 Hz
96Keysight 53210A/53220A/53230A Assembly Level Service Guide
Test
number
6b
(Option 115)
(Q)
Performance Tests1
Complete performance tests
Test descriptionMinimumActual readingMaximum
Frequency Accuracy
Channel 3
3.1 GHz
53210A
53220A
53230A
3.099 999 969 GHz
3.099 999 996 9 GHz
3.099 999 996 9 GHz
Record: Mean
__________________
__________________
__________________
3.100 000 031 GHz
3.100 000 003 1 GHz
3.100 000 003 1 GHz
6b
(Option 115)
(Q)
7a
(Q)
Option 106
Option 150
53230A
53210A
53220A
53230A
Frequency Accuracy
Channel 3
15.0 GHz
53210A
53220A
53230A
53210A
53220A
53230A
Pulse Burst
Frequency
(6 GHz)
Narrow Pulse Wid th
(200 nSec)
14.999 999 850 GHz
14.999 999 985 GHz
14.999 999 985 GHz
5.994 000 000 GHz
Record: Std Dev
__________________
__________________
__________________
Record: Mean
__________________
__________________
__________________
Record: Std Dev
__________________
__________________
__________________
Record: Mean
__________________
Record: Std Dev
_____________________
43.4 Hz
21.7 Hz
4.34 Hz
15.000 000 150 GHz
15.000 000 015 GHz
15.000 000 015 GHz
210 Hz
105 Hz
21 Hz
6.006 000 000 GHz
3.000 MHz
Keysight 53210A/53220A/53230A Assembly Level Service Guide97
1Performance Tests
Test
number
7b
Option 106
Option 150
53230A
7c
(Q)
Option 115
Option 150
53230A
Test descriptionMinimumActual readingMaximum
Pulse Burst
Frequency
(6 GHz)
Wide Pulse Wid th
(13 uSec)
Pulse Burst
Frequency
(1.35 GHz)
Narrow Pulse Wid th
(400 nSec)
Complete performance tests
5.999 954 000 GHz
1.348 865 000 GHz
Record: Mean
__________________
Record: Std Dev
_____________________
Record: Mean
__________________
Record: Std Dev
_____________________
6.000 046 000 GHz
18.500 kHz
1.351 350 000 GHz
338 kHz
7d
Option 115
Option 150
53230A
Pulse Burst
Frequency
(1.35 GHz)
Wide Pulse Wid th
(13 uSec)
1.349 979 200 GHz
Record: Mean
__________________
Record: Std Dev
_____________________
1.350 020 800 GHz
7.79 kHz
98Keysight 53210A/53220A/53230A Assembly Level Service Guide
Keysight 53210A/53220A/53230A 350 MHz Universal Frequency
Counter/Timer
Assembly Level Service Guide
2Service
Introduction 100
Returning the Instrument to Keysight Technologies for Service 104
About the Keysight 53210A/53220A/53230A Calibration Menu 109
The Keysight 53210A/53220A/53230A Calibration Procedures 118
Pre-Troubleshooting Information 152
Troubleshooting the Counter 160
Error Messages 174
99
2Service
NOTE
Introduction
Returning the counter to Keysight Technologies for service
About the Keysight 53210A/53220A/53230A calibration menu
Unless otherwise indicated, this manual applies to all serial numbers for the
Keysight 53210A/53220A/53230A Counters.
This chapter provides service information for your Keysight 53210A, 53220A, and
53230A Counters. It is divided into five major sections:
This section provides you with instructions on how to return the counter to
Keysight for service.
This section contains reference information that gives an overview of the
calibration menu for the Keysight 53210A, 53220A, and 53230A Counters.
The Keysight 53210A/53220A/53230A calibration procedures
This section provides step-by-step procedures for calibrating the
Keysight 53210A, 53220A, and 53230A Counters.
Pre-troubleshooting information
This section provides you with pertinent information such as safety considerations,
recommended test equipment, repair and after-service calibration considerations,
and assembly identification and location.
Troubleshooting information
This section provides you with troubleshooting procedures that can assist you to
isolate an instrument failure to the faulty assembly or module. A final section on
“Error Messages” can be used to help isolate a failed assembly for replacement.
100Keysight 53210A/53220A/53230A Assembly Level Service Guide