Used Agilent Agilent Used 53181A Manual

Page 1
Agilent 53181A 225 MHz Frequency Counter
Operating Guide
Manual Part Number 53181- 90001
Page 2
NOTES
Page 3

Agilen t 53181A Operating Guide

This guide describes how to use the Agilent 53181A 225 MHz Frequency Counter.
Agilent 531 81A 225 MHz Fr equen cy Counter
Page 4
Copyri ght 1994 - 2003
W
C
W
Agilent Techn olo gies , Inc. All Rights Reserved. Reproduction,
adaptation, or translations without prior written permission is prohibited, except as all owed under the copyright laws.
Print e d: November 2003 Printed in Malaysia
Manual part number 53181-90001

Certification and Warranty

Certification
Agilent Technologies certifies that this product met its published specification at the time of shipment from the factory. Agilent Technologies further certifies that its calibration measurements are traceable to the United States National Institute of Standards and Technology (formerly National Bureau of Standards), to the extent allowed by the Institute’s calibration facility, and to the calibration facilities of other International Standards Organizat ion members.
Warranty
Agilent warrants Agilent hardware, accessories and supplies against defects in materials and workmanship for a period of one year from date of shipment. If Agilent receives notice of such d efects d uring the warranty period, Agilent will, at its option, either repair or replace products which prove to be defective. Replacement products may be either new or like-new.
Agilent warrants that Agilent software will not fail to execute its programming instructions, for the period specified above, due to defects in material and workmanship when properly installed and used. If Agilent receives notice of such defect s during the warran ty period, Agilent will replace software media which does not execute its programming instructions due to such defects.
Safety Con siderations General
This product and related documentation must be reviewed for familiarization with this safety markings and instructions before operation.
Before Cleaning
Disc on ne ct t he product fr o m operating power before cleaning.
Warning Symbols That May Be Used In This Book
Instruction manual symbol; the product will be marked with this symbol when it is necessary for the user to refer to the instruction manual.
Indicates hazardous voltages.
Indicates earth (ground) terminal.
or
Indicates terminal is connected to chassis when such connection is not apparent.
Indicates Alternating current.
Indicates Direct current.
Safety Considerations (contd)
ARNING
BODILY INJUR Y OR DEATH MAY RESULT FROM FAILURE TO HEED A WARNING. DO NOT PROCEED BEYOND A WARNING UNTIL THE INDICATED CONDITIONS ARE FULLY UNDERSTOOD AND MET.
AUTION
Damage to equipment, or in correct measurement data, may result from failure to heed a caution. Do not proceed beyond a CAUTION until the indicated conditions are fully understood and met.
Safety Earth Ground
An uninterruptible safety earth ground must be maintained from the mains power source to the product’s ground circuitry.
ARNING
WHEN MEASURING POWER LINE SIGNALS, BE EXTREMELY CAREFUL AND ALWAYS USE A STEP-DOWN ISOLATION TRANSFORMER WHICH OUTPUT IS COMPATIBLE WITH THE INPUT MEASUREMENT CAPABILITIES OF THIS PRODUCT. THIS PRODUCT’S FRONT AND REAR PANELS ARE TYPCIALLY AT EARTH GROUND. THUS, NEVER TRY
TO MEASURE AC POWER LINE SIGNALS WITHOUT AN ISOLATION TRANSFORMER.
Agilent Techn olo gies , Inc. 7.NC.NL.A.11.03.97.R1.P.CW6FC 900 South Taft Avenue Loveland, Colora do 80537 U .S.A.
Page 5
Warranty (contd)
Agilent does not warrant that the operation of Agilent products will be uninterrupted or error free. If Agilent is unable, within a reasonable time, to repai r or replace any product to a condition as warranted, customer will be entitled to a refund of the purchase price upon prompt return of the prod u ct .
Agilent products may contain remanufactured parts equivalent to new in per formance or may have been subjected to incidental use.
The warranty per iod begins on the date of delivery or on the date of installation if installed by Agilent. If customer schedules or delays Agilent installation more than 30 days after delive ry, warranty begins on the 31st day from delivery.
Warranty does not apply to defects resul ting from (a) improper or inadequate maintenance or calibration, (b) software, interfacing, parts or supplie s not supplie d by Agilent , (c) unauthorized modifi cati on or misus e, (d) operation outside of the published environmental specifications for the product, or (e)improper sit e prepar at i o n or ma intenan ce.
TO THE EXTENT ALL O WED BY LOCAL LAW, THE ABOVE WARRANTIES ARE EXCLUSIVE AND NO OTHER WARRANTY OR CONDITION, WHETHER WRITTEN OR ORAL, IS EXPRESSED OR IMPLIED AND AGILENT SPECIFICALLY DISCLAIMS ANY IMPLIED WARRANTIES OR CONDITIONS OF MERCHANTABILITY, SATISF ACTORY QUALITY, AND FITNESS FOR A PARTICULAR PURPOSE.
Agilent will be liable for damage to tangible property per incident up to the greater of $300,000 or the actual amount paid for the product that is the subject of the claim, and for damages for bodily injury or death, to the extent that all such damages are determined by a court of competent jurisdiction to have been directly caused by a defective Agile nt product.
TO THE EXTEN T AL LO WED BY LOCAL LAW, THE REMEDIES IN THIS WARRA N TY STAT EMEN T ARE CUSTOMER’S SOLE AND EXCLUSIVE REMEDIES. EXCEPT AS I NDICATED ABOVE, IN NO EVENT WILL AGILENT OR ITS SUPPLIERS BE LIABLE FOR LOSS OF DATA OR FOR DIRECT , SPECIAL, INCIDENTAL, CONSEQUENTIAL (INCLUDING LOST PROFIT OR DATA), OR OTHER DAMAGE, WHETHER BASED IN CONTRACT , TORT, OR OTHERWISE.
For consumer transactions in Australia and New Zealand: the warranty terms contained in this statement, except to the extent lawfully permitted, do not exclude, restrict or modify and are in addition to the ma ndatory statutory rights applicable to the sale of this product to you.
Assistance
Product maintenance agreemen ts and other customer assistance agreem ents are available for Agilent Technolo­gies pro du c t s.
For any assistance, contact your nearest Agilent Technologies Sales and Service Office.
Safety Considerations (contd)
WARNING
INSTRUCTIONS FOR ADJUSTMENTS WHILE COVERS ARE REMOVED AND FOR SERVICING ARE FOR USE BY SERVICE­TRAINED PERSONNEL ONLY. TO AVOID DANGEROUS ELECTRIC SHOCK, DO NOT PERFORM SUCH ADJUSTMENTS OR SERVICING UNLESS QUALIFIED TO DO SO.
WARNING
ANY INTERRUPTION OF THE PROTECTIVE GROUNDING CONDUCTOR (INSIDE OR OUTSIDE THE PRODUCT'S CIRCUITRY) OR DISCONNECTING THE PROTEC TIVE EARTH TERMINAL WILL CAUSE A POTENTIAL SHOCK HAZARD THAT COULD RESULT IN PERSONAL INJURY. (GROUNDING ONE CONDUCTOR OF A TWO CONDUCTOR OUTLET IS NOT SUFFICIENT PROTECTION.)
Whenever it is likely that the prote ct i o n ha s be e n impa i r e d , th e instrument must be ma de inoperative and be secured against any unintended operation.
If this instrument is to be energized via an autotransformer (for voltage reduction) make sure the common terminal is connected to the earthed pole terminal (neutral) of the power source.
Instructi ons for adjustments while covers are removed and for servicing are for us e by tra ined- pe r sonn el only. To avoid dangerous electric sho ck, do not per f orm such adjustments or servici ng unless qualified to do so.
For conti nue d pro te cti on ag ain st fire , replace the line fuse(s) of the same current rating and type (for example, normal blow , time delay). D o not use repaired fuses or short circuited fuseholders.
Acoustic Noise Emissions
LpA<47 dB at operator position, at normal operation, tested per EN 27779. All data are the results from type test.
Geräuschemission
LpA<47 dB am Arbeits platz, normaler Betrieb, geprüft n ach EN 27779. Die Angagen beruh e n auf Ergebnissen von Typenprüfungen.
7.NL.A.11.03.97.R1.P.CW6BC
Page 6
I
I
II
DECLARATION OF CONFORMITY
According to ISO/IEC Guide 22 and CEN/CENELEC EN 45014
Manufacturer’s Name: Manufacturer’s Address:
Declares, that the product
Product Name: Model Number: Product Options: This declaration covers all options of the above product.
Conforms with the following European Directives:
The product herewith complies with the requirements of the Low Voltage Directive 73/23/EEC and the EMC Directive 89/336/EEC (including 93/68/EEC) and carries the CE Marking accordingly.
EMC Standard
IEC 61326-1:1997+A1:1998 / EN 61326-1:1997+A1:1998 CISPR 11:1990 / EN 55011:1991 IEC 61000-4-2:1995+A1:1998 / EN 61000-4-2:1995 IEC 61000-4-3:1995 / EN 61000-4-3:1995 IEC 61000-4-4:1995 / EN 61000-4-4:1995 IEC 61000-4-5:1995 / EN 61000-4-5:1995 IEC 61000-4-6:1996 / EN 61000-4-6:1996 IEC 61000-4-11:1994 / EN 61000-4-11:1994
Canada: ICES-001:1998 Australia/New Zealand: AS/NZS 2064.1
Agilent Technologies, Incorporated Santa Clara Site 5301 Stevens Creek Blvd Santa Clara, California 95051
Universal Counter Frequency Counter 53131A, 53132A 53181A
Limit
Group 1 Class A 4kV CD, 8kV AD 3 V/m, 80-1000 MHz
0.5kV signal lines, 1kV power lines
0.5 kV line-line, 1 kV line-ground 3V, 0.15-80 MHz I cycle, 100%
[1]
Safety
Supplemental Information:
[1]
The product was tested in a typical configuration with Agilent Technologies test systems.
Authorized EU-representative: Agilent Technologies Deutschland GmbH, Herrenberger Stra
Revision: A Document No. 53131A
IEC 61010-1:1990+A1:1992+A2:1995 / EN 61010-1:1993+A2:1995 Canada: CSA C22.2 No. 1010.1:1992
July 31, 2001
Date Art Nanawa, Product Regulations Manager
For further information, please contact your local Agilent Technologies sales office, agent or distributor.
β
e 130, D 71034 Böblingen, Germany
Page 7

Contents

In This Guide
Contents and Organization xii Related Documents xiii Types of Service Available if Your Instrument Fails xiv
Standard Repair Services (Worldwide) xiv Express Repair/Perf ormance Calibration Service (USA Only) xiv Repair Instrument Yourself xiv
Repackaging for Shipment xv Description of the 225 MHz Frequency Counter xvi Options xvii
Hardware xvii Support xvii
Accessories Supplied and Available xviii
Accessories Supplied xviii Accessories Available xviii
Agilent 53181A Quick Reference Guide xix
1 Getti ng S tarted
The Front Panel at a Glance 1-2 The Front Panel Indicators at a Glance 1-3 The Front Panel Indicators at a Glance (Cont.) 1-4 The Front Panel Menus at a Glance 1-5 The Front Panel Menus at a Glance (Cont.) 1-6 The Front Panel Menus at a Glance (Cont.) 1-7 The Display Annunciators at a Glance 1-8 The Display Special Characters at a Glance 1-9
The Underscore Placeholder 1-9 The Asterisk Placeholder 1-9
Ope rati ng Guide iii
Page 8
Contents
The Limit Test Graph Characters at a Glance 1-10 The Rear Panel at a Glance 1-11 Making Measurements 1-12
To Measure Frequency 1-14 T o Select Input Coupling and Impedance 1-15
Selecting Input Coupling 1-15 Selecting Input Impedance 1-15
To Set Input Channel Tr igger Level/Sensitivity 1-16
Changing Trigger Mode 1-16 Modifying Input Tr igger Level 1- 17 Selecting Input Trigger Slope 1-18 Selecting Input Sensitivity 1-18 Starting the Measurement 1-19
To Select Scale an d Offset 1-19
Entering th e Scale Value 1-19 Entering the Offset Value 1-20 Displaying the Math Results 1-21 Disabling Math 1-22
To Set Limits of Measurements 1-23
Setting the Upper Limit 1-24 Setting the Lower Limit 1-25 Setting the Count er to Flag and Stop Measuring On Out-of-Limit
Measurements 1-27 Setting the Counter to Flag On Limits But Continue Measuring 1-28 Disabling Limit Testing 1-28 Disabling Math 1-29
T o Perform Statistics on Measurements 1-29
Selecting the Type of Sta tistics (Stats) 1-29 Computing Stats on Filtered Data Only 1-31 Displaying Stats After Filtering Data of Input Signal 1-33 Disabling Stats and Math 1-34
To Control Measurement 1-35
iv Operating Guide
Page 9
Contents
2 Operat i n g Your F reque nc y C o unter
Introduction 2-2
Chapter Sum mary 2-2 Where to Find Some Key W orking Examples 2-3
How this Cou nt er Works for You 2-4 Summary of the Measurement Sequence 2-5 Using the Measurement Control K eys (Run and Stop/Single) 2-6
Overview of the Measurement Control Keys 2-6 To Use the Measu re men t Co n tro l Key s 2-7
Using Entry/Select (Arrow) Keys 2-9
To Use During Numeric Entry 2-9 T o Use When Sequencing Through the Other Meas and Recall
Menus 2-9 To Use During State Ch a ngi ng (O N/OFF, LO/MED/ HI, et c. ) 2-10 To Use on Prompted Help Messages (MATH HELP: NO/YES, PRT
HELP: NO/YES) 2-10
Using the MEASURE Menu Keys 2-11
Overview of the MEASURE Menus 2-11 To Measure Frequency 2-12 To Measure Frequency Ratio (Optional) 2-13 To Measure Period 2-13 To Measure Pos i tive/ Negative Vol t Pe aks 2-14
Using the Gate & External Arm M enu Key 2-15
Overview of Gate/External Ar ming Functions 2-15
TIME Arming 2-15 AUTO Arming 2-15 EXTERNAL Arming 2-15 DIGITS Arming 2-16
To Use the Gate and External Arm 2-17
Example Procedure for Gate and External Arm 2-17
Ope rati ng Guide v
Page 10
Contents
Example Procedure for Changing the Number of Digits of Resolution Displayed for More Precise Measurements 2-19
Using the MATH Menu Keys 2-21
Overview of Scale/Offset Math Menu 2-21 To Use the Scale/ O ffset Ma t h Men u 2-22
Example Procedure for Scale Function 2-22 Example Procedure for Of fset Function 2-23 Example Procedure for Turning Off Math Mode 2-25 Example Procedure for Set Reference Function —Offset From the Last Measurement Value (Seeing Digits Not Visible on the 12- Digit Display) 2-25
Overview of Stati stics (Stats) Menu 2-26 To Use the Stats Menu for Automatic and Continuous Statistical
Analysis 2-27
Example Procedure for Computing S tats 2-27 Example Procedure for Easy Viewing of Stats 2-28 Example Procedure for Filt ering Da ta (Using Limits) During Stats 2-29 Example Procedure for Configur ing SINGLE to Initiate N Measurements 2-30 Example Procedure for Turning Off Stats Mode 2-30
Using the LIMITS Menu Keys 2-31
Overview of Limits Menus 2-31 To Set and Use Au t o m atic L imit Testin g 2-32
Limits Testing Example 1—Flag and Stop Measuring On Limits 2-32
Limits Testing Example 2—Flag On Limits but Continue Measuring 2-34 Limits Testing Example 3—Use Analog Gra ph Display While Adjusting/Alig ning Input Signal 2-34
Limits Testing Example 4—Selecting Filtering Conditions of Stats Computation 2-36
vi Operating Guide
Page 11
Contents
Limits Testing Example 5—Sending the Limit-Detect Output to the RS-232 Serial Port 2-38
Using the DISPLAY Digits Keys 2-39
Overview of the DISPLAY Digits Keys 2-39 To Use the DISPLAY Digits Keys 2-40
Using CHANNEL 1 Input Conditioning Keys 2-41
Overview of Trigger/Sensitivity Menu 2-41 To Use the Trigger/Sensitivity Key to Adjust Counter’s Trigger
Voltage and Sensitivity Level s 2-44 Overview of Input Conditioni ng Toggle Keys 2-47
Using the Save and Recall Menus 2-48
Overview of Save and Recall Functions 2-48 To Use the Save Function 2-49 To Use the Recall Function 2-50 To Unsave a Mea s urement Setup 2-51
Using the Pri nt Menu 2-52
Overview of the Print Menu 2-52 To Use the Print Men u 2-52
Using the Utility Menu 2-53
Overview of the Utility Menu 2-53 To Set the GPIB Ad d ress 2-54
Selecting Operating Mode (Talk/Listen, Talk-Only) 2-54
Setting the GPIB Address 2-54 To Choose the Timebase Source 2-55 To Run the Self-Test Routines 2-56
Overview of the Self-Test Routines 2-56
Example Procedure for Running the Self Test 2-57 To Configure the RS-232 Serial Port for Printing 2-58
Setting the Hardware Pacing 2-58
Setting the Baud Rate 2-59
Setting the Parity 2-59
Setting the Software Pace 2-59
Ope rati ng Guide vii
Page 12
Contents
To Configure the RS-232 Serial Port for Sending Limit-Detect Output 2-60
T o Select the Numerical Convention for the Display 2-61 T o Connect the Counter to a Serial Print er via the RS-232 Port 2-61 T o Connect the Counter to a Printer via GPIB 2-62 T o Select the GPIB Talk-Only Mode for Printing 2-62
Front Panel Display Message s 2- 63
Measurement Result Displa ys 2-63 Power-Up/Self Test Messages 2-64 Menu Messages 2-65 GPIB Messages 2-67
Preset Values After Power-up and *RST 2-68
Preset Values for Functions Accessibl e Via Front Panel or GPIB 2-68
Preset Values for Functions Accessibl e Via GPIB Only 2-72
Common Qu es ti ons 2-74
Why is Stats result not avai lable yet? 2-74 Why won’t printer work? 2-74 Why did Counter stop measuring? 2-74 Why did Counter go to its default state aft er I set up my RS-232
port? 2-74 Counter’s numeric display does not follow the numerical conven tion
for my country. 2-74
3 Specifications
Introduction 3-2
Instrument Inputs 3-2 Instrument Inputs (Continued) 3-3 Time Base 3-4 Measurement Specifications 3-5 Measurement Specific ations (Continued) 3-6 Measurement Specific ations (Continued) 3-12 Measurement Definitions 3-15
viii Operating Guide
Page 13
Index
Contents
Measurement Arming and Processi ng 3-16 General Information 3-17
Ope rati ng Guide ix
Page 14
Contents
x Operating Guide
Page 15

In This Guide

This book is the operating guide for the Agilent 53181A 225 MHz Frequency Counter. It consists of a table of contents, this preface, a quick reference guide, three chapters, and an index.
This preface contains the following information:
Contents and Or ganization
Related Documents
Types of Service Available if Your Instrument Fails
Repackaging for Shipment
Description of the 225 MHz Frequency Counter
•Options
Accessories Supplied and Available
page xii page xiii page xiv page xv page xvi page xvii page xviii
Ope rati ng Guide xi
Page 16
In This Guide
Cont en t s and Organization
The Table of Contents lists the beg inning of each chapter in the gui de, helping you locate information.
The Preface introduces you to the operating guide, the Frequency Counter , and provides general information about the Counter.
The Quick Reference Guide consists of a Menu Tree (tear-out sheet) that serves as a tool to trigger your memory or get you quickly reacquain ted with the instrument, and Menu Roadmaps that illustrate how to navigate through the menus. It is located after this preface.
Chapter 1, “Getting Started,” is a quick start guide that giv es you a brief overvi ew of the Counter’s keys, indicators, menus, displa y, and connectors. Last, a graphical procedure for perfor ming a complete measurement is provide d.
Chapter 2, “Operating Your Frequency Counter,” is an operator’s reference. You are given an overview of each group of front-panel keys, operating functions, and menus followed by a series of exe r cises that guide you through the operation of the Counter.
Chapter 3, “Specifications,” list s the specif ications and characteristics of the Counter.
The Inde x helps you find items easily by arrangi ng information in a sequenc e different from that found in the Table of Contents, disclosing relationships among items.
xii Operating Gui de
Page 17
In This Guide
Related Docum ents
For more information on counters refer to the following S eries 200 Application Notes:
Fundamentals of Ele ctronic Frequency Co unte rs Applica tion Note 200—Agilent part number 02-5952-7506.
Understanding Frequency Counter Specifications Application Note200-4— Agilent part number 02-5952-7522.
Operating Guide xiii
Page 18
In This Guide
Types of Service Av ai la bl e if Your Instrument Fai ls
If your Counter fails wit hin one y ear of origina l purch ase, Agilen t will repai r it free of charge. If your instrument fails after your 1-year warranty expires, Agi lent will repai r it, or you can repair it yourself by ordering the se rvice guide.
There are three type s of repair services:
Standard repair service—if downtime is not critical.
Express Re pair/Performance Calibration Ser vice—if downtim e is critical.
Use the Assembly-Level Service Guide and repair the unit yourself.
Standard Repair Services (Worldwide)
Contact your nearest Agilent Technologies Service Center. They will arrange to have your Counter repaired.
Express Repair/Performance Calibr ation Serv ice (USA O nl y)
If downtime i s crit ical , you c an receiv e yo ur repai red Counter via ove rnig ht shi pment. Just call 1-800-4 03-0801 and ask for Express Repair/Perform ance Calibration Service. When your Counter is r epa ired, it will be returned via overnight shipment.
Repair Instrument Yoursel f
If your Counter 1-year warranty has expired and you choose to repai r the instrument yourself or would like more details on self-test and calibration, use the procedu r es in the Asse mbly-Lev el Service G u ide.
xiv Operating Guide
Page 19
In This Guide
Repackaging for Shipment
For the Express Repa ir/ Performance Calibr ation Ser vic e descri bed above , retu rn your failed Counter to the designated Agil ent Service Center, using the shipping carton of the instr u ment. Agilen t will notify you when your failed instrument has bee n received.
If the instrument is to be shipped to Agilent for service or repair, be sure you do the following:
Attach a tag to the ins trument identifyi ng the owner and indicatin g the re quired service or repair. Include the instrument model number and full serial number.
Place the instrument in its original container (if availa ble) with appropriate packaging mate rial.
Secure the container with strong tape or metal bands.
If the original ship ping contain er is not availa ble, place your unit in a cont ainer which will ensure at least 4 inches of compressible packaging material around all sides of the unit. Use static free packaging materials to avoid additional damage to your unit.
Agilent Technologies suggests that you always insure shipments.
Ope rati ng Guide xv
Page 20
In This Guide
Description of the 225 MHz Frequency Counter
The Agilent 53181A is a frequency counter capable of meas uring frequencies to 225 MHz on Channel 1. Depending on which optional Channel 2 the counter contains, this capability is extended to 1.5 GHz (Option 015), 3.0 GHz (Option 030), 5.0 GHz (Option 050) or 12.5 GHz (Option 124). T he Agilent 53181A has a freque ncy resolution of 10 digits in one second.
The Agilent 53181A provides users with a GPIB measuring spee d of up to 200 measurements per second, and is suitable for bench-top and ATE operation.
The Agilent 53181A basic measurement functions include Frequency, Period, Peak Voltage, and Frequency Ratio (if Channel 2 is ins talled).
The Agilent 53181A includes additio nal measurement functions and features that are designed specifically for manufacturing and service applications:
1, 5, 10 MHz external reference capability—to match cust om er’s house standard,
optional ultra high, high, or mediu m stability oven osci llators for high accuracy needs and le ng t h en ed ca li b r at io n cy cles,
external gating,
full math and statistics,
automatic limit testing,
analog displ ay m ode limit testing, and
SCPI programming capability.
Programmable control is performed via an GPIB. The GPIB and a talk-onl y RS-232C serial port are sta ndard for the Agilent 53181A. The serial port is for printing measured and ana lyzed data on serial printers, or for outputting an out-of-limit signal.
xvi Operating Guide
Page 21
In This Guide
Options
The options available for th e Agilent 53181A 225 MHz Fr equency Counter a re listed following this paragraph. Specif ications for the options are listed in Chap ter 3, “Specifications.” If you’ve purchased an option with the initial order, it will be installed a t the factory and ready for operation at delivery. Refer to the “Retrofitting Optio ns” chapte r in th e A s sembly- L ev el S er v ice Guide f o r ins t r u ct io n s on f ield installa tion of the options.
NOTE The “0s” an d “1s ” in the fo l lo w i n g opt i on numbers ar e numeric char a ct er s (th at is,
they are not lett ers).
Hardware
Medium Stabi lity Oven Timebase, Option 001
DC Power Input, Option 002
High Stabi lity Oven Time bas e, Option 010
Ultra High Stability Oven Timebase, Option 012
1.5 GHz RF Input Channel (Channel 2), Option 015
3.0 GHz RF Input Channel (Channel 2), Option 030
5.0 GHz RF Input Channel (Channel 2), Option 050
12.4 GHz RF Input Channel (Channel 2), Option 124
*
Rear Terminal s
, Option 060
Rack Mount Kit, Option 1CM
Support
5-year Return to Agilent for Repair, Option W50
5-year Return to Agilent for Calibra tion, Option W52
*
Channel 1 will have both front and rear terminals. Option 015 and Option 030 Channel 2 will have a rear
terminal only. Option 050 and Option 124 Channel 2 will have a front terminal only.
Ope rati ng Guide xvii
Page 22
In This Guide
Accessori es Supplied and Availabl e
Accessories Supplied
Power co r d , 2. 3 meters
Accessories Available
Agilent 34161A Accessory Pouch
Agilent 34131A Transit Case
Printer RS-232 Interface cables, Agilent 24542G or Agilent 24542H
GPIB cables, Agilent 10833A/B/ C/D
xviii Operating Guide
Page 23
Agilent 531 81A Quick Re ference Guide
The Quick Reference Guide is designed for exp erienced users of the Agilent 53181A Frequency Counter. It is intended to be used as a tool to trigge r your m emory. If you are using the Agilent 53181A for the first time, Agilent recommends that you at le as t read Chapter 1, “Getting Started,” in the Operating Guide first.
The Quick Reference Guide f ollows this page, and consi sts of the following items:
A Menu Tree which may be torn out of the guide for external use (pages 1 and 2).
Menu Roadmaps which illustrate via key -press sequences how to navigate through the menus under the menu keys (pages 3 through 10). Key-press sequences are provided for the following menu keys:
Freq Ch1 – Freq Ch2 –Other Meas –Gate & ExtArm – Uppr & Lower – Limit Mo de s – Scale & Offset –Stats – Trigger/Sensitivity
Operating Guide xix
Page 24
Preface
xx Operating Guide
Page 25
Agilent 53181A
Frequency Counter
Freq Ch 1
FREQUENCY 1
Freq Ch 2
FREQUENCY 2
Uppr & Lower
UPPR: 0.000000 LOWR: 0.000000
Other Meas
PERIOD
1 RATIO 1 TO 2 RATIO 2 TO 1 VOLT PEAKS 1
Limit Modes
LIM TEST: OFF
LIM TEST: ON
ON FAIL:GO ON
ON FAIL:
STOP
SHOW: NUMBER
GRAPH
SHOW:
GATE:
TIME
GATE: AUTO
TIME: .100 s
Scale & Offset
SCALE:1.000000 OFFS: 0.000000 SET REF: NO
SET REF: YES SET REF: HELP
MATH: OFF
MATH: ON
MATH HELP : NO
MATH HELP: YES
Gate & ExtArm
GATE: EXTERNL
START: POS
START: NEG
STOP: AUTO
STOP: NEG STOP: POS STOP: TIME
TIME: .100 s
Stats
SHOW: MEAS
SHOW: STD DEV SHOW:
MEAN
SHOW:
MAX
SHOW:
MIN N: 100 STATS: OFF
STATS: ON
USE: ALL MEAS
USE: IN
LIMIT ON SINGLE: 1
ON SINGLE: N
GATE: DIGITS
DIGITS: 4
1
Page 26
Agilent 53181A
Frequency Counter
Recall
NO REGISTERS RECALL 0 RECALL 1 RECALL 2 RECALL 3 RECALL 4 RECALL 5
RECALL 20
50 1M
CH 1: 50 OHM
CH 1: 1M OHM
Save & Print
SAVE: UNSAVE: PRINT: OFF
PRINT: ON
PRT HELP: NO
PRT HELP: YES
DC AC
CH 1: DC
CH 1: AC
AUTO TRG: ON LEVEL: 50 PCT
SLOPE: POS
SLOPE: NEG
SENSTVTY: HI
SENSTVTY: LO SENSTVTY:MED
X10 Attenuate
CH 1: X10 ATT
CH 1: X1 ATT
Trigger Sensitivity
AUTO TRG: OFF
LEVEL: 0.000V
CH 1:
CH 1:
100kHz Filter
LP FILT NO FILT
2
Page 27
Agilent 53181A Frequency Counter
Freq Ch 1
MEASURE
Freq Ch 2
Other Meas
Freq Ch 1
Freq Ch 1
FREQUENCY 1
Freq Ch 2
Freq Ch 2
FREQUENCY 2
Other Meas
Other Meas
Other Meas
Other Meas
Other Meas
PERIOD 1
RATIO 1 TO 2
RATIO 2 TO 1
VOLT PEAKS 1
3
Page 28
Agilent 53181A Frequency Counter
Freq Ch 1
FREQUENCY 1
MEASURE
– – – – – – – – – – – –
Gate & ExtArm
Gate & ExtArm
GATE: TIME
GATE: AUTO
GATE
: EXTERNL
Gate & ExtArm
Gate & ExtArm
GATE: DIGITS
STOP:
TIME: .100
Gate &
TIME: .100
ExtArm
TIME: .200
Gate &
START:
ExtArm
START: POS START: NEG
STOP: AUTO STOP: NEG STOP: POS STOP: TIME
TIME: .200
Gate &
DIGITS: 4
ExtArm
Enter
Enter
Gate & ExtArm
Gate & ExtArm
Gate & ExtArm
RUN
RUN
RUN
DIGITS: 5
4
Enter
RUN
Page 29
Agilent 53181A Frequency Counter
Uppr & Lower
LIMITS
Uppr & Lower
Limit Modes
Uppr & Lower
Limit Modes
Limit Modes
UPPR: 0.000000
Uppr &
LOWR: 0.000000
Lower
LIM TEST:
Limit
ON FAIL:GO ON
Modes
Limit
SHOW: NUMBER
Modes
UPPR: 5.100000
LOWR: 4.900000
LIM TEST: OFF LIM TEST: ON
ON FAIL:GO ON ON FAIL: STOP
SHOW: NUMBER SHOW: GRAPH
Enter
Enter
RUN
RUN
RUN
RUN
RUN
5
Page 30
Agilent 53181A Frequency Counter
Scale & Offset
Scale & Offset
MATH
Scale & Offset
Scale & Offset
Scale & Offset
Scale & Offset
Scale & Offset
SCAL:
1.000000
OFFS:
0.000000
SET REF:
SET REF: NO SET REF: YES SET REF: HELP
MATH:
MATH HELP: YES
MATH HELP: NO
SCAL:
OFFS:
2.000000
0.500000
Enter
OFFS:-
Enter
Enter
nnnnnnn
MATH: OFF MATH: ON
Enter
(MEAS X SCALE) + OFFS = RESULT
RUN
RUN
RUN
RUN
6
Page 31
Agilent 53181A Frequency Counter
Stats
Stats
MATH
Stats
SHOW:
SHOW: MEAS SHOW:
STD
DEV
SHOW: MEAN
SHOW: MAX
SHOW: MIN
Stats
Stats
Stats
Stats
N: 100
N: 200
STATS:
STATS: OFF STATS: ON
USE:
USE: ALL MEAS USE: IN LIMIT
ON SINGLE:
ON SINGLE: 1 ON SINGLE: N
Enter
RUN
RUN
Run
Run
7
Page 32
Agilent 53181A Frequency Counter
Trigger Sensitivity
Trigger Sensitivity
CHANNEL 1
Trigger Sensitivity
AUTO TRG:
Enter
Trigger Sensitivity
Trigger Sensitivity
Trigger Sensitivity
Trigger Sensitivity
AUTO TRG: ON
AUTO TRG: OFF
LEVEL: 0.000V
LEVEL: 2.000V
LEVEL: 50 PCT
LEVEL: 80 PCT
SLOPE:
SLOPE: POS SLOPE: NEG
SENSTVTY:
SENSTVTY: HI SENSTVTY: LO SENSTVTY: MED
RUN
RUN
RUN
RUN
RUN
8
Page 33
1
Getting Started
Page 34
Chapter 1 Getting Started
5
9
8
The Front Panel at a Glance
The Front Pane l at a Glanc e
1
1
23
4
67
Remote Error (Flashing)
SRQ
POWER
ON
Utility Menu:
STBY
Hold at power up
53181A
FREQUENCY COUNTER
Period Freq Ch 1 Ch 2 ExtRef
MEASURE LIMITS MATH DISPLAY CHANNEL 1
Freq Ch 1
Other Meas
Recall
Freq Ch 2
Gate & ExtArm
LocalUtility Save &
Print
3 GHz
1 Measurement function menu keys 2 Limits menu keys 3 Math menu keys 4 Sign (+ or ) selecti on toggle key 5 Data Entry/Select (or arrow) keys 6 Enter numeric data (terminate) key. Also
tranverses a menu and exits at the end of a menu.
7 Channel 2 optional RF input
10
Uppr & Lower
Limit Modes
Run
11
Scale & Offset
Stats
Stop/ Single
CHANNEL 2
100 MHz to MAX
Limit
MHz µs Gate
More Digits
Fewer Digits
12
+/–
Enter
Trigger Sensitivity
50 1M
DC AC
X10 Attenuate
Damage Lvl: 5V rms MAX.50
0.1 Hz 225 MHz
Damage Lvl: 5V rms MAX.50
13
100kHz Filter
!
!
8 Utility men u key (Hold during power-up to access
Utility functions.)
9 Recall, Save and Print menu keys
10 Gate and Extern al Arm menu key
11 Measurement control keys 12 Display Digits keys 13 Channel 1 Tr igger/Sens itivity menu key and input
conditioning keys
NOTE It is normal operation for the fan in the Counter to conti nue to run af ter the Count er is
placed in Standby mode. Power to the timebase is continuous to maintain long term measurement reliability, and the f an helps maintain timebase temperature stabilit y.
1-2 Operat ing Guide
Page 35
Chapter 1 Getting Started
The Front Panel Indicators at a Glance
The Fro n t P a nel Indicat ors at a Glance
There are eight different group s of indicators or LEDs. They are li st ed and described in the fo ll o w in g table.
Indicators Description of the In dicators
When the indicator for the Freq Ch 1 or Freq Ch 2 key s is
Freq Ch 1
Other Meas
Scale & Offset
Stats
+/–
Freq Ch 2
Limit Modes
Local Save & Print
Trigger Sensitivity
Enter
lit, it indica te s that t he Freque nc y func tion is ena bled fo r th e corresponding input channel.
When the indicator for the Other Meas key is lit, it simultaneously indicates that the key’s menu and its menu item (for example, PERIOD 1) is enabled.
When these indicators are lit, the key’s “enable” menu item (that is, Limit Modes/LIM TEST, Scale & Offset/MATH, Stats/STATS, and Save & Print/PRINT) is enabled.
When this indicator is lit, it indicates that auto-trigge ring is enabled.
When this indicator flashes, it indicates that the arrow keys can be used to modify or enter data.
The Enter key is used to enter numeric values or parameters. It can also be used to traverse through a menu. Pressi ng the Enter key at the end or bottom of a menu will cause the Counter to exit the menu and return to the measuremen t m o de .
1
Run
Stop/ Single
When one of these indicators is lit, it indicates that the Run or Single function is enabled.
Ope rati ng Guide 1-3
Page 36
Chapter 1 Getting Started
The Front Panel Indicators at a Glance (Cont.)
The Front Panel Ind ic ators at a Glance (Cont.)
1
Indicators Description of the In dicators
When this i ndicator fl ashes, it indicates that the Counter is triggering on the input signal. If the input signal is too high, this indicator remains ON. If the input signal is too low, this indicator is OFF.
50 1M
X10 Attenuate
Remote Error (Flashing)
SRQ
DC AC
100kHz Filter
When one of these in dicators is lit, it indic ates that the adjace nt choice (that is, 50Ω, DC, X10, or 100kHz Filter) is enable d or acti ve. Not e t hat w hen thes e i ndica to rs ar e no t lit , then the other choice (that is, 1MΩ, AC, X1, or no filter) is active.
A lit Remote indicator indicates that the Counter is in remote mode (Note: In the remote mode, the Save & Print key becomes the Local key.)
If (while in remote) an error occurs, the Remote indicator will flash. The indic ator will continue flashing until the controller has read or cleared the error queue, or until the front panel returns to local mode.
An unlit Remote indicator indicates that the Counter is in local mo de .
The SRQ indic ator indicates that the Counter has re quested servi ce from the controller. The SRQ indicat or will remain lit unti l th e con trol le r has rec ogn iz ed t he se rvice req ues t an d seria l polled the Counter, or taken sp ecific action to cancel the request (for exampl e, *CLS command) .
1-4 Operat ing Guide
Page 37
Chapter 1 Getting Started
The Front Panel Menus at a Glance
The Fro n t P anel Menus at a Glance
Freq Ch 1
Other Meas
Gate & ExtArm
1
FREQUENCY 1
Freq Ch 2
FREQUENCY 2
1
Uppr & Lower
UPPR: 0.000000 LOWR: 0.000000
PERIOD
1
1
RATIO 1 TO 2
1
RATIO 2 TO 1 VOLT PEAKS 1
Limit Modes
LIM TEST: OFF
LIM TEST: ON
ON FAIL:GO ON
ON FAIL:
STOP
SHOW: NUMBER
GRAPH
SHOW:
2
GATE:
TIME: .100 s
TIME
Scale & Offset
GATE: AUTO
SCALE:1.000000 OFFS: 0.000000 SET REF: NO
SET REF: YES SET REF: HELP
MATH: OFF
MATH: ON
MATH HELP: NO
MATH HELP: YES
GATE: EXTERNL
START: POS
START: NEG
STOP: AUTO
STOP: NEG STOP: POS STOP: TIME
TIME: .100 s
Stats
SHOW: MEAS
SHOW: STD DEV SHOW:
MEAN
SHOW:
MAX
SHOW:
MIN N: 100 STATS: OFF
STATS: ON
USE: ALL MEAS
LIMIT
USE: IN
ON SINGLE: 1
ON SINGLE: N
GATE: DIGITS
DIGITS: 4
1
These menu items appear only if your Counter contains the optional 1.5 GHz, 3.0 GHz, or 5.0 GHz RF Input (Channel 2).
2
Refer to Chapter 2 in this guide for details on the Gate & ExtArm menu.
Ope rati ng Guide 1-5
Page 38
Chapter 1 Getting Started
The Front Panel Menus at a Glance (Cont.)
The Front Pane l Me nus at a G la nc e (Cont.)
1
Recall
3
NO REGISTERS
4
RECALL 0 RECALL 1 RECALL 2 RECALL 3 RECALL 4 RECALL 5
RECALL 20
50 1M
CH 1: 50 OHM
CH 1: 1M OHM
3
This appears when nothing can be recalled.
4
Only registers which can be recalled will appear in this menu.
5
This menu item only appears if an instrument setup has been saved.
Save & Print
SAVE:
5
UNSAVE: PRINT: OFF
PRINT: ON
PRT HELP: NO
PRT HELP: YES
DC AC
CH 1: DC
CH 1: AC
Trigger Sensitivity
AUTO TRG: ON LEVEL: 50 PCT
SLOPE: POS
SLOPE: NEG
SENSTVTY: HI
SENSTVTY: LO SENSTVTY:MED
X10 Attenuate
CH 1: X10 ATT
CH 1: X1 ATT
AUTO TRG: OFF LEVEL: 0.000V
100kHz
Filter
CH 1:
LP FILT
CH 1:
NO FILT
1-6 Operat ing Guide
Page 39
Chapter 1 Getting Started
The Front Panel Menus at a Glance (Cont.)
The Front Pane l Me nus at a G la nc e (Cont.)
NOTE Turn powe r of f, pre ss and h old R ecall (Utility) key, then pres s POWER key to access
this menu.
POWER
On / Stby
REV: X X X X HP-IB: 3 TIMEBAS:
TIMEBAS: TIMEBAS: EXT
CAL HELP: NO
CAL HELP: YES
TEST LOOP: OFF
TEST LOOP: ON
TST PRINT:OFF
TST PRINT: ON
Utility
Recall
AUTO INT
TEST:
NO
6
TEST:
DISP
TEST:
CPU
TEST: ROM TEST: RAM TEST:
EEPROM
TEST:
HP-IB
TEST: QSPI TEST: FPGA TEST: FR END TEST: MEAS TEST: INTERP TEST:
KEYPAD
TEST:
PRINT
BAUD: 9600
BAUD:
19200
BAUD: 300 BAUD: 1200 BAUD: 2400
PARITY: OFF
PARITY:
EVEN
PARITY: ODD
SW PACE: XON
SW PACE: NONE
DTR: HIGH
DTR: LIMIT DTR:
HW PACE SHOW 9 AS: 9.0
SHOW 9 AS: 9,0
1
6
These menu items appear only if TEST LOOP is OFF.
Ope rati ng Guide 1-7
Page 40
Chapter 1 Getting Started
The Display Annunciators at a Glance
The Display Annunciators at a Glance
1
MHz
µs
Gate
Period Freq Ch 1 Ch 2 Limit ExtRef
Annunciator Indication
Period Counter is set to measure Period. Freq Counter is set to measure Frequency. Ch 1 Counter’s channel 1 is selected to measure an input si gnal. Ch 2 Counter’s channel 2 is selected to measure an input si gnal. Limit Counter is limit testing and the current measurement exceeds the user-
entered limits.
ExtRef Counter is set t o use the signal connected at rear panel Ref In
connect or as the timeba se (TIMEBAS: EXT); orCou nter is set to automatically (TIMEBAS: AUTO) select the timebase and has chosen the signal connected at the rear panel Ref In connector.
Hz The displayed data is in units of Hertz. M The prefix for the unit s of the display ed data is mega (10
µ The prefix for the unit s of the display ed data is micro (10
s The display ed data is in units of seconds. Gate The gate is ope n. Before a mea surement cycle starts, this ann unciator
is OFF, indicating the gat e is clo sed. D ur ing a mea surem ent cyc le , t he annunciator is ON, indicating the gat e is open.
6
).
6
).
1-8 Operat ing Guide
Page 41
Chapter 1 Getting Started
The Display Special Characters at a Glance
The Display Special Characters at a Glance
Freq Ch 2
The Underscore Placeholder
Special C h aracter Description
A placeholder that indicates the digit has been blanked using the Display Digits keys.
1
MHz
Hz
Freq Ch 1
The Asterisk Placehold er
Special C h aracter Description
A placeh older t hat i ndi cates t his digi t is not sign if icant be cau se of the measurement resolution. (The appearance o f this asterisk is determined by the Counter; thus, thes e cannot be removed by the Display Digits keys.)
Ope rati ng Guide 1-9
Page 42
Chapter 1 Getting Started
The Limit Test Graph Characters at a Glance
The Limit Test Graph Characters at a Glance
1
Freq Ch 1
Special C h aracter Description
The colon s represent the lower and upp er limits.
The asterisk represents the last measurement.
1-10 Operating Guide
Page 43
Chapter 1 Getting Started
3
4
5 6 8 9
1 2
7
The Rear Panel at a Glance
The Rear Panel at a Glance
1
WARNING:
1
I N P U T S
WARNING:
1 Rear-panel input connectors (optional) 2 Power modu le (Sen ses inc oming voltage
and automatically selects proper setup.)
3 Fan
NO OPERATOR SERVICEABLE PARTS INSIDE, REFER SERVICING TO SERVICE TRAINED PERSONNEL.
!
60 VA
10 MHz Out
AC LINE:
50/60 Hz
50/60/400 Hz
100 - 120 VAC
200 - 240 VAC
ISM 1-A
FOR LABORATORY USE BY QUALIFIED PERSONNEL FOUR USAGE EN LABORATOIRE PAR PERSONNEL QUALIFIE
OPTIONS
001 MS Oven 010 HS Oven 012 US Oven
To Configure: Hold Recall during turn-on.
Osc Adjust
2
3
Ext Arm
Ref In
FOR CONTINUED FIRE PROTECTION, USE SPECIFIED ~ LINE FUSE.
6 10 MHz Output connector 7 GPIB (IEEE-488.1) interface conn ector 8 Oscillator Adjust potentiometer 9 RS-232 Inte rface or Limi t Output
connector
SERIAL PLATE
RS - 232
4 External Arm input connec tor 5 External Reference In put connector
NOTE It is normal operation for the fan in the Counter to conti nue to run af ter the Count er is
placed in Standby mode. Power to the timebase is continuous to maintain long term measurement reliability, and the f an helps maintain timebase temperature stabilit y.
Ope rati ng Guide 1-11
Page 44
Chapter 1 Getting Started
Making Measurements
Making Measurements
1
One of the first things you will want to do with your Counter is to become a cquainted with its front panel. Therefore, we have written the procedures in this section to familiarize you with some of its controls . Th e followin g procedures are provided:
First you are shown how to turn on the Counter and measure the frequency of a signal applied to the Counter’s input channels.
Second, you are shown how to us e the input coupling, impedance, and trigger/sensitivity keys to set the input conditions of Channel 1 to match the signal being measured.
Third, you are shown how to scale and offset the measurement result.
Fourth, you are shown how to set upper and lower limits for measurements.
Fifth, you are shown how to enable the Counter to compute statis tics (such as standard deviation) and display statistics of measurements.
Last, you are shown how to use the Run and Stop/Single keys to control measurements.
The order of the procedures in this chapter is the recommended ord er for making measurements wi th this Counter.
1-12 Operating Guide
Page 45
Chapter 1 Getting Started
Making Measurements
Study and refer to the following legend, as nee ded, to understand the meaning of the icons which are used throughout this chapt er.
12
34
1 Press key one time and
release
2 Press key two times and
release
3 Repeated key presses 4 Press and hold
5 Result 6 Auto operation 7 Connect signal 8 Disconnect signal
Legend
5
6
1
78
91110
9 Indicator off
10 Indicator on
11 Indicator flashing
Ope rati ng Guide 1-13
Page 46
Chapter 1 Getting Started
Making Measurements
To Measure Fre quency
1
POWER
On / Stby
Period Freq Ch 1 Ch 2 Limit ExtRef
Freq Ch 1
Freq Ch 1
MHz
µs
Gate
Connect (for demonstr ation purposes) the Counter’s rear-panel 10 MHz Out signal to CHANNEL 1 input as shown in the ill ust rated procedure, below.
CHANNEL 1
MHz
Gate
Damage Lvl: 5V rms MAX.50
!
Freq Ch 1
The Counter will automatically display the measured frequency of the input signal.
1-14 Operating Guide
Page 47
Chapter 1 Getting Started
Making Measurements
If you need or want to change CHANNEL 1’s coupling, impedance, and triggering conditions to match the input signal you are trying to measure, the next procedures “To Select Input Coupl ing and Impedance” and “To Set Input Channel Trigger Level/Sensitivity” demonstrate this. Perform these procedures whether or not you want to customize the Counter’s input conditions to measure your signal ; doing this will help you become familiar with the DC/AC,
50Ω/1M, and Trigger/Sensitivity keys.
To Select Input Coupling and Impedance
Selecting Input Coupling
DC AC
Channel 1’s input coupling is now set to dc.
If you want to change the coupl ing back to the default ac coup ling, perform the following s tep.
1
DC AC
Selecting Input Impedance
50 1M
Channel 1’s input impedance is now set to 50Ω.
NOTE The “arrow” keys can also be use d to toggle the state of toggle keys (DC/AC,
50Ω/1MΩ, etc. ) as indica ted by t he fla shing i ndica tor withi n th e arrow ke ys. Ho wever, for simplicity in this procedure, use the corre sponding toggle key to change states.
Ope rati ng Guide 1-15
Page 48
Chapter 1 Getting Started
Making Measurements
If you want to change the input impedance back to the default 1 MΩimpedance, perform the following step.
1
50 1M
Freq Ch 1
MHz
Gate
To Set Input Channel Trigger Level/Sensitivity
Changing Trigger Mode
Trigger Sensitivity
Press any one of these arro w keys to toggle to the next state of Auto Trigger.
Enter
+/–
1-16 Operating Guide
Page 49
Chapter 1 Getting Started
Making Measurements
Modifying Input T r igger Level
Trigger Sensitivity
The leftmost “0” di git in the LEVEL display is highlighted, indicating that if you press the d key once the displayed value will increase to 1.000 volt as shown in the following s tep.
Enter
Channel 1’s trigger level is now set to +1V.
To set the trigger level to −2V, perform the following steps.
1
+/–
Enter
Channel 1’s trigger level is now set to −2V.
Ope rati ng Guide 1-17
Page 50
Chapter 1 Getting Started
Making Measurements
Selecting Input Trigger Slope
1
Trigger Sensitivity
Display 1, 7/13/92
Display 1, 7/13/92
Refer to the sub-s ection titled “To Use the Trigger/Sens itivity Key to Adjus t Counter’s Trigger Voltage and Sensitivity Levels” in Chapter 2 for more information about trigger slope, if needed.
Selecting Input Sensit ivity
Trigger Sensitivity
Display 1, 7/13/92
Display 1, 7/13/92
Display 1, 7/13/92
Refer to the sub-s ection titled “To Use the Trigger/Sens itivity Key to Adjus t Counter’s Trigger Voltage and Sensitivity Levels” in Chapter 2 for more information about sensitivity, if needed.
1-18 Operating Guide
Page 51
Chapter 1 Getting Started
Making Measurements
Starting the Measurement
MATH
Scale & Offset
Stats
Run
Freq Ch 1
MHz
Gate
The Run key initiates repetitiv e meas urements, and is described in the sectio n titled “To Control Measurement” at the end of this chapter.
To Select Scale and Offset
The Scale & Offset key allows for multiplication and a ddition, respectively, of the measurement b y use r -specifi ed cons tant s. Mo difica tion of the di spla yed mea surement by these Math operations is represente d by the following equation:
(Measurement × Scale ) + Off set = Di sp layed Re su lt s
The Scale and Offset Math operations can be used, for example, to subtract systematic errors or display the pe rcentage difference between signals.
Entering the Scale Value
Scale & Offset
1
To demonstrate the Sc ale Math operation, set Scale to 10 as shown in the following steps.
Ope rati ng Guide 1-19
Page 52
Chapter 1 Getting Started
Making Measurements
Press and hold the d key until the value of Scale is 10 as shown in the following step.
1
Enter
NOTE BE SURE to press the Enter key to enter the value of 10.
The Scale is now set to 10, and MATH has been enabled. The Scale & Offset indicator is no w li t to show tha t M ATH is enable d . S in ce MATH is ena bled, th e results are being scaled and offset.
Entering the Offset Value
To demonstrate the Offset Math operation, set the Offset to 1 MHz as shown in the following steps.
Scale & Offset
At this point, pressing the s key will cause the Counter to display the full display of the Offset value as shown in the following step.
1-20 Operating Guide
Page 53
Chapter 1 Getting Started
Making Measurements
Press the s key six more tim es to cause the Counter to dis play your entry in Mega units as shown in the f ollowing step.
The leftmost “0” digit in the OFFSet display is highlighted, indicating that if you press d key once the displayed value will increase to 1 Mega (that is, 1E6) as shown in the fo ll owing st e p.
Enter
NOTE BE SURE to press the Enter key to enter the 1 Mega value.
The Offset is now set to 1 Mega.
Displaying the Math Results
M
1
M
M
Run
Freq Ch 1
M
The Counter displa ys the modifi ed measur ement result s, whic h are based on the scale and offset value s that you selected in the pre vious steps. (For more details and real applications of the Math Scale and Offset ope rations, refer to the appropriate section in Chapter2, “Operating Your Frequency Counter.”)
Ope rati ng Guide 1-21
Page 54
Chapter 1 Getting Started
Making Measurements
Disabling Math
1
Scale & Offset
Display 1, 7/13/92
Display 1, 7/13/92
Note that the Scale & Offset key indicator is now off.
Run
Freq Ch 1
NOTE DO NOT cycle POWER because you will need to use these Scale and Offset values
in the following procedure “To Set Limits of Measurements.” Continue to the following procedure.
MHz
Gate
1-22 Operating Guide
Page 55
Chapter 1 Getting Started
Making Measurements
To Set Limits of Measurements
To demonstrate how Mat h and L im its work together, use the Scale (10) and Offset (1 Mega) values sele ct ed in the previous procedur e “To Select Scale and Offset. ” Enab le Math by performing th e following steps.
Scale & Offset
Display 1, 7/13/92
Display 1, 7/13/92
1
Run
Freq Ch 1
M
The result of this Math operation is a measurement of 101 MHz.
(Measurement × Scale) + O ffset = Result
(10 MHz × 10) + 1 Mega = 101 Mega
Now, set the upper limit to 102 Mega and the lower limit to100 Mega by performing the following proc edures. (Figure 1-1 and Figure 1-2 illustrate the limits settings.)
Ope rati ng Guide 1-23
Page 56
Chapter 1 Getting Started
Making Measurements
Setting the Upper Limit
LIMITS
Uppr &
1
Lower
Limit Modes
Press the s key nine tim es to ca use the Counter to display your entry in hundreds of Mega units as shown in the following step.
The leftmost “0” digit in the UPPR display is highlighted as shown above, indicating that each pre ss of the d key will increase the displayed value.
Uppr & Lower
M
M
Enter
NOTE BE SURE to press the Enter key to enter the 102 Mega value.
1-24 Operating Guide
M
M
M
Page 57
Chapter 1 Getting Started
Making Measurements
1
2
1 102 Mega Upper Limit 2 101 Mega Scale/Offs et Measureme nt
Figure 1-1. 102 Mega Upper Limit Setting
Setting the Lower Limit
Uppr & Lower
Press the arrow keys as shown in the following steps to set the lower limit value. Press the s key nine tim es to ca use the Counter to display your entry in hundreds of Mega units as shown in the following step.
1
Enter
NOTE BE SURE to press the Enter key to enter the 100 Mega value.
Ope rati ng Guide 1-25
M
M
M
Page 58
Chapter 1 Getting Started
Making Measurements
Limits should now be set a s shown in Figure 1-2.
1
1 102 Mega Upper Limit 2 101 Mega Scale/Offs et Measureme nt 3 100 Mega Lower Limit
1 2
3
Figure 1-2. 100 Mega Lower and 102 Mega Upper Limits Settings
Figure 1-3 represents what transpired during this Math and Limits procedure.
1 Raw Me as u r e m en ts 2 Math
3 Limits 4 Measurements (Scale/Offset Results) within
Limits
Figure 1-3. Math and Limits Results
1-26 Operating Guide
Page 59
Chapter 1 Getting Started
Making Measurements
Setting the Counter to Flag and Stop Measuring On Out-of-Limit Measurements
If you want the Counter to stop measuring when the signal exceeds the limits (102 to 100 Mega) that you entered in the previous procedure, perform the followi ng steps to select the STOP choice in the ON FAIL display . (Note that ON FAIL: GO ON is the default state after power-up.)
Limit Modes
Run
Freq Ch 1
M Gate
The current modified measurement of the input sign al applied to CHANNEL 1 is displayed.
Since the Counter is now in the stop-on-fail mode, the Limit annunciator in the display will light and the Counter will stop m aking measurements when a measurement excee ds the limits you set. The measurement result that exceeded the entered limit s will be held on the display for observation.
1
Ope rati ng Guide 1-27
Page 60
Chapter 1 Getting Started
Making Measurements
Setting the Counter to Flag On Limits But Continue Measuring
Perform the following steps to select the GO ON choice in the ON FAIL display if you want the Counter to continue measuring e ven though an measurement result
1
exceeds the limits p r eviously entered.
Limit Modes
Run
Freq Ch 1
M
Gate
The current modified measurement of the input signal applied to CHANNEL 1 is displayed.
Since the Counter is now in the go-on-fail mode, the Limit annunc iator in the displa y will light each time a measurement exceed s the limits you set. However, the Counter will continue to make measurements.
Disabling Limit Testing
Limit Modes
The Counter is now making measurements without limit testing.
1-28 Operating Guide
Page 61
Chapter 1 Getting Started
Making Measurements
Disabling Math
Scale & Offset
Display 1, 7/13/92
Display 1, 7/13/92
Run
Freq Ch 1
MHz
Gate
The Counter is now making measurements without the scale/offset values calculated into the measurem en ts.
To Perform Statistics on Measurements
Selecting the Type of Statistics (Stats)
Suppose you want the Counter to compute and display the standard deviation of the current input data (which is the 10 MHz signal appl ied to CHANNEL 1). Also, you want the Counter to make 20 measurements before it computes the standard deviation. Perform the following steps.
1
MATH
Scale & Offset
Stats
Stats
Ope rati ng Guide 1-29
Page 62
Chapter 1 Getting Started
Making Measurements
Updating the SHOW configura tion caused Stats to be enabled. The Stats indicator is now lit.
1
Stats
Enter
NOTE BE SURE to press the Enter key to enter the value of 20.
1-30 Operating Guide
Page 63
Chapter 1 Getting Started
Making Measurements
The Counter is now set to make sta tistics based on 20 measur ements.
Run
Freq Ch 1
Freq Ch 1
Freq Ch 1
Hz
Gate
Hz
Gate
In this case, the dis played standard devia tion value is computed on all measurements of the 10 MHz signal since no limits were set.
Computing Stats on Filtered Data Only
A special featu re of the Counter allows you to use the upper and lower limits to filter data before stat istical processing or computation as shown in Figure 1-4.
1
2
3
4
1
1 Raw Me as u r e m en ts 2 Limits
3 Filtered data (USE: IN LIMIT) 4 Statistics
Figure 1-4. Filtering Data Before Statistical Computation
Ope rati ng Guide 1-31
Page 64
Chapter 1 Getting Started
Making Measurements
Perfo r m th e follo w in g s te ps to sele ct the IN LI M I T ch o i ce in the USE display if yo u want the Counter to compute statistics on only frequency measurements within the limits you set.
1
Stats
Display 1, 7/13/92
Display 1, 7/13/92
Since the Li mi ts were set to 100 Mega and 102 Mega values that are based on a s cale of 10 and of fset of 1 Mega, you must re-enable Math now to get the measurements to be within the lim its. Perform the following steps.
Scale & Offset
Display 1, 7/13/92
Display 1, 7/13/92
1-32 Operating Guide
Page 65
Chapter 1 Getting Started
Making Measurements
Displaying Stats After Filtering Data of Input Signal
Let’ s ass ume you ha ve set th e upper an d lowe r limit s for the input s ignal, and se lect ed the IN LIMIT (filtering) choice. Now, perform the following steps to display the standard deviation of the filtered measurements. (Note that the first step in the
followin g procedure is optional since you should have already set Stats to show standar d deviation at th e beginning of this Stats procedur e. But, you may want to perform the step anyway to verify that the Counter is displ aying the standard deviation of the measurement.)
Stats
Display 1, 7/13/92
Run
Freq Ch 1
1
Freq Ch 1
Freq Ch 1
Gate
Gate
The standard deviation value shown in the previous illustration is for demonstration purposes. The statistic is computed using only those measurements which fell within the limits you set.
(For more deta ils o n the S ta ts and Li mits fun ct ions, ref er t o the a ppropria te se ction s in Chapter 2, “Operating Your Frequency Counter.”)
Now, disable Math and Sta ts as shown in the following proc edure.
Ope rati ng Guide 1-33
Page 66
Chapter 1 Getting Started
Making Measurements
Disabling S tats and Math
1
Stats
Scale & Offset
Display 1, 7/13/92
Display 1, 7/13/92
Run
Freq Ch 1
MHz
Gate
The Counter is now making and displaying normal measurements (that is, the Counter is not showing statistics or scale/offset results).
1-34 Operating Guide
Page 67
Chapter 1 Getting Started
Making Measurements
To Control Measurement
Run Single/
Stop
Use these two keys to control the meas urement of the Counter. The Run key provide s repetitive measurements, whereas the Stop/Single key allows you to make one measurement.
With the 10 MHz signal still connected to CHANNEL 1, perform the foll owing steps so you can better understand the Run and St op/Single operations.
Run
Freq Ch 1
MHz
Gate
The Counter is now making repetitive measurements (continuously making “live” measurements).
Stop/ Single
Freq Ch 1
MHz
The Counter stopped making measurements. The Gate an nunciator is not lit. Hence, pressing th e Stop/Single key while the Counter is making measurements (in Run) causes the Counter to stop after the measurement in progress is completed. If you press the Stop/Single key again whi le the Counter i s st opped, the Counte r will make a single m e as u rement and th e n st op —the Gate annunciator will light one time, momentarily.
1
If you press the Stop/Single key while the Counter is stopped and when the Stats menu item ON SINGLE is set to N, the Coun ter will make N measurements and then stop. This enables a set of statisti cs to be computed.
While the Counter is still stopped, perf orm the following step.
Run
Freq Ch 1
MHz
Gate
The Counter is making repetitive measurements again.
Ope rati ng Guide 1-35
Page 68
Chapter 1 Getting Started
Making Measurements
1
1-36 Operating Guide
Page 69
2
Operating Your Frequency Counter
Operator’s Reference
Page 70
Chapter 2 Operating Your Frequency Counter
Introduction
Introduction
This is the operator’s reference chapter which contains inform ation and procedures for the front-panel keys, operating functions, and menus of the Agilent 53181A 225 MHz Frequency Counter.
Chapter Summary
How this Counter Works for You page 2-4
2
Summary of the Measurement Sequence page 2-5
Using the Measurement Control Keys (Run and Stop/Single) page 2-6
Using Entry/Select (Arrow) Keys page 2-9
Using the MEASURE Menu Keys page 2-11
Using the Gate & External Arm Menu Key page 2-15
Using the MATH Menu Keys page 2-21
Using the LIMITS Menu Keys page 2-31
Using the DISPLAY Digits Keys page 2-39
Using CHANNEL 1 Input Conditioning Keys page 2-41
Using the Save and Recall Menus page 2-48
Using the Prin t Menu page 2-52
Using the Utility Menu page 2-53
Front Panel Display Messages page 2-63
Preset Values After Power-up and *RST page 2-68
Common Questions page 2-74
2-2 Operat ing Guide
Page 71
Chapter 2 Operating Your Frequency Counter
Introduction
Where to Find Some Key Working Examples
Example Procedure for Gate and External Arm page 2-17
Example Procedure for Changing the Number of Digits of Resolution Displayed for More Preci se Measurements
Example Procedure for Scal e Function page 2-22
Example Procedure for Offs et Function page 2-23
Example Procedure for Set Reference Function—Offset From the Last Measurement Value (Seeing Digits Not Vi sible on the 12­Digit Display)
Example Procedure for Com puting Stats page 2-27
Limits Testing Example 1—Flag and Stop Mea suring On Limits page 2-32
Limits Testing Example 2—Flag On Limits but Continue Measuring
Limits Testing Example 3—Use Analog Graph Display While Adjusting/Al igning Input Signal
Limits Testing Example 4—Selecting Filtering Conditions of Stats Computation
Limits Testing Example 5—Sending the Limit-Detect Output to the RS-232 Serial Port
page 2-19
page 2-25
page 2-34
page 2-34
page 2-36
spage 2-38
2
Ope rati ng Guide 2-3
Page 72
Chapter 2 Operating Your Frequency Counter
How this Counter Works for You
How this Counter Works for You
The following text lists some of the key things the Counter does for you.
The Counter presets the menus to default states and values at power-up (refer to
Table 2-6 for a detailed list of the preset values). Cycling the POWER key presets the Counter.
The Counter’s Utility menu a llows you to select such things as timebase source, GPIB configuration, and RS-232 serial port configur ati on. After your selections,
2
the Counter automati cally stores all these se lections in non-volatile memory (except the timebase source); thus, these settings (except timebase source) will not change when power has been off or after a remote interface reset.
The Counter automatically displays measurement(s) in a couple of seconds after you select a measurement function (e.g., Period).
The Counter accepts your numeric entry for a menu item after you pre s s the Enter key. You MUST p ress th e En t e r ke y to complete num er ic entry.
The Counte r automatically enables Lim its when you enter a num er ic value for upper and/or lower limits, or update any menu item in the Limit Modes menu.
The Counter automatically enables Math after you enter a numeric value for scale and offset.
The Counter perform s Statistical analyses for you, and will enable Stats when you update any item in the Sta ts menu.
The Counter automatically disables Limits, Math, and Stats when you select another measurement function.
The Counter saves measurement setup(s).
The Counter instantly recalls the measurement setup you want to use.
The Counter prints out your measurement and analysis data.
2-4 Operat ing Guide
Page 73
Chapter 2 Operating Your Frequency Counter
Summary of the Measuremen t Sequence
Summary of the Me asurement Sequence
Use Utility menu to:
choose the timebase source.
configure the GPIB if you intend to operate the Counter remotely.
set RS-232 serial port if you intend for the Counter to perform print ing and/or limit-detecting.
Use MEASURE menu keys to select the measurement function.
Use CHANNEL 1 keys to set up input conditioning.
Use Gate & ExtArm key to:
set the gate time and resolution.
•set arming.
Use Scale & Offset key to set up math operations.
Use Uppr & Lower key to set limits.
Use Limit Modes key to set up li mit testing.
Use Stats key to set up statistics and limit fi ltering.
Use Save & Print key to enable or disable printing.
Use Run and Stop/Single keys to control measurements.
2
Ope rati ng Guide 2-5
Page 74
Chapter 2 Operating Your Frequency Counter
Using the Measurement Control Keys (Run and Stop/Single)
Using the Measur ement Control Keys (Run and Stop/Single)
Overview of the Measurement Control Keys
Run Stop/
Single
T wo measur ement con trol key s are provided o n the Cou nte r: Run and Stop/Single. In general, the Run key provides repetitive measurements while the Stop/Single key allows you to make single-shot measurements.
The Run key allows you to:
2
put the Counter into a continuous measurement loop.
exit any measur ement after your selection of another measuremen t function.
abort the current m easurement by starting a new measurement (if already in Run mode or in Single mode with a measurement in progress). This also clears any statistics being proc essed .
The Stop/Single key allows you to:
put the Counter into Single mode (if the Counter was in Run mode) where one measurement is taken with each Stop/Single keypress. (Stats menu item ON SINGLE should be set to 1.)
put the Counter into Single mode (if the Counter was in Run mode) where N measurements are taken with each Stop/Single keypress. (Stats menu item ON SINGLE should be set to N.)
stop (abort) the current measurement (if the Counter was in Single mode).
2-6 Operat ing Guide
Page 75
Chapter 2 Operating Your Frequency Counter
Using the Measurement Control Keys (Run and Stop/Single)
To Use the Measurement Control Keys
The following procedure demonstrates how these keys function.
1 Connect power source to Counter, and turn on Counter.
All segments of the front-panel display will light up while the Counter performs its power-on self-test, and then dashes are displayed. The Counter is now rea dy to measure frequency of a signal applied to CHANNEL 1 input. Note that the Freq and
Ch1 annunciators light.
2 Connect an input signal to CHANNEL 1.
The Counter aut omatically displays the measured freq uency of the input signa l. Note that th e Run key indicator is ON.
3 Press Stop/Single key.
The Counter is put into Single mode and the current meas u rement is completed. One measurement is taken with each press of Stop/ Single key. Note tha t the Stop/Single key indicator is ON.
4 Press Run key.
The Counter aborts any measurement in progres s, and starts making repetitive measurements. That is, the Counter is in a continuous measur ement loop.
5 Press Stop/Single key.
The Counter is put into Single mode.
6 Press Stats key until ON SIN GLE menu item appears, then press any one of the
arrow keys to set ON SING LE menu item to N.
The Counte r is set to take N measurements on each Stop/Single keypress (note: N = 100 at power-up). The state of ON SINGLE menu item affe cts the Stop/Single key operation while statistics are enabled.
2
Observe t hat the Stats indica tor is ON. This indicates statisti cal processing is enabl ed.
Ope rati ng Guide 2-7
Page 76
Chapter 2 Operating Your Frequency Counter
Using the Measurement Control Keys (Run and Stop/Single)
7 Press Stop/Single key.
One set of N frequ ency measur ements is tak en with e ach pre ss of the S top /Single key. One set of stat istics is computed with each press of the Stop/Single ke y.
8 Press Stats key until SHOW me nu item appears.
9 Press any one of the arrow keys to select a stat is tical result to displ ay, and press
Freq Ch 1 key.
The Counter is now displaying one of the statistical results.
2
10 Press Stop/Single key.
One set of N frequ ency measur ements is tak en with e ach pre ss of the S top /Single key. At the completion of the Nth measurement, the Counter displays the s tatistical result selected in step 9.
11 Press Stats key until ON SINGLE menu item appears, then press any one of the
arrow keys to set ON SING LE menu item to 1.
The Counter is set to take one measurement on each Stop/Single keypress.
Disabling the statistics (with the STATS: OFF menu item) is anoth er way to cause the Stop/Single key to initiate one measurement at a time.
2-8 Operat ing Guide
Page 77
+/–
Enter
Chapter 2 Operating Your Frequency Counter
Using Entry/Select (Arrow) Keys
Using Entry/Select (Arrow) Keys
There are six entry/select keys of which four are “arrow” keys. The function of the four arrow keys and the Enter key depends on the Counter’s operating mode (that is, numeric entry, state changing, sequenci ng through choices in a menu, etc.). This section describes how the keys perform in th e different oper ating modes.
To Use During Numeric Entry
Use the s key to move left to selec t adjustable digits (it does not wrap around).
Use the g key to move right to select a djustable digits (it does not wrap around).
Use the d key to increment the sel ected (highlighted) digit of t he displayed value.
Use the f key to decrement the selec ted (highlighted) digit of the displayed value.
•Use the +/− key to change the sign of the numeric value.
•Use the Enter key to complete a numeric entr y. (If a numeric entry is changed and the Enter key is NOT pressed, then the value is not changed.)
2
To Use When Sequencing Through the Other Meas and Recall Menus
Use the d or s key to go back to the previous choice in a funct ion or recall menu. Repeatedly pr essing this key will cycle through (and loop around) the list of choices.
Use the f or g key to go forwa rd or to the next ch oice i n a funct ion or rec all menu . Repeatedly pr essing this key will cycle through (and loop around) the list of choices.
•Use the Enter key to select th e function or execute the r ec all, or wait for the menu to time out for automatic selection/execution.
To Use During State Changing (ON/OFF, LO/MED/HI, et c. )
•Press any of the arrow keys to toggle or cha nge to the next state of the parameters found in the following menus:
Gate & Ex tArm
Ope rati ng Guide 2-9
Page 78
Chapter 2 Operating Your Frequency Counter
Using Entry/Select (Arrow) Keys
Limit Modes –Scale & Offset – Stats – Trigger/Sensitivity –50Ω/1M –DC/AC –X10 Attenuate – 100kHz Filter
2
–Save & Print – Utility Menu
•The +/− key is ignored during the st ate changing operati ng mo de.
T o Use on Prompted Help Messages (MA TH HELP: NO/YES, PRT HELP: NO/YES)
When MATH HELP: YES is di spla yed, press Enter key to displ ay the “(MEAS x SCALE) + OFFS = RESULT” message.
When PRT HELP: YES is displayed, press Enter key to display the “CONFIGURE PORT ON UTILITY MENU; REMEMBER TO SA VE SETUP FIRST” ins truction.
2-10 Operating Guide
Page 79
Chapter 2 Operating Your Frequency Counter
Using the MEASURE Menu Keys
Using the MEASURE Menu Keys
Overview of the MEASURE Menus
MEASURE
Freq Ch 1
Other Meas
Freq Ch 2
Gate & ExtArm
The measurement function menus are divided into three categories:
Frequency measurements on Channel 1.
Frequency measu rem ents on Channel 2 (option al).
Other measurements (period, ratio (optional), and peak voltage).
Table 2-1. Menus of the MEASUREment Keys
Freq Ch 1 key Freq Ch 2 key Other Meas key
FREQUENCY 1 FREQUENCY 2 (optional) PERIOD 1
RATIO 1 TO 2 (optional) RATIO 2 TO 1 (optional) VOLT PEAKS 1
To sequence thro ug h the Other Meas menu, you simply need to repeatedly press the Other Meas key to cycle through (and loop around) the items (or measurement functions) under the key. Each press of a Other Meas key will advance the Counter to the next measurement function in the menu.
NOTE The menus under the MEASURE keys will terminate (select function and return to
measurement result display) after a sh ort timeout.
WARNING When measuring power line freque ncies, be extremely careful and a lway s use a step-
down isolation transformer of which output volta ge is compatible with the input measurement capabilities of this product. Since the shell of the BNC input connector is grounded to the chassis of the instrument, the hot side of a n ac power line could be connected to the BNC shell and create a direct short to ground. Theref ore, NEVER
TRY TO MEASURE AC POWER LINE FREQUENCIES WITHOUT AN ISOLATION TRANSFORMER.
2
Ope rati ng Guide 2-11
Page 80
Chapter 2 Operating Your Frequency Counter
Using the MEASURE Menu Keys
To Measure Fre quency
1 Connect power source to Counter, and turn on Counter.
All segments of the front-panel display will light up while the Counter performs its power-on self-test, and then dashes are displayed. The Counter is now rea dy to measure frequency of a signal applied to CHANNEL 1 input. Note that the Freq and
Ch1 annunciators light.
2 Connect an input signal to CHANNEL 1.
The Counter shoul d autom atica lly dis play the meas ured freq uency of th e input s ig nal.
2
The trigger le vels are set to the 50% points of th e signal. However, you may disable auto triggering, and change the trigger levels and slopes.
T o s et up CHANNEL 1’s coupling, impedanc e, an d tri ggering cond itions to match th e input signal you are trying to measure, refer to the procedures “To Select Input Coupling and Impe dance” and “To Set Input Channel Trigger Level/Sensitivi ty” in Chapter 1, “Getting Started,” or the section titled “Using CHANNEL 1 Input Condition ing Keys” in this chapter.
3 To measure the frequency of a signal applied to the option al CHANNEL 2 input,
press Freq Ch2 key.
FREQUENCY 2 will be momentarily displayed and the Freq and Ch2 annunciators will light. Th e Counter will be ready to measure the fre quency of a signal applied to CHANNEL 2 input.
Note that the Frequency 2 menu item is not avai lable in Counters tha t do not contain the optional Channel 2.
2-12 Operating Guide
Page 81
Chapter 2 Operating Your Frequency Counter
Using the MEASURE Menu Keys
To Measure Frequency Ratio (Optional)
1 Connect signals to be measured to channel s 1 and 2.
2 Press Other Meas key until RA TIO 1 TO 2 is displayed to measure the frequency
ratio between a signal applied to CHANNEL 1 input and a signal applied to the optional CHANNEL 2 input.
RATIO 1 TO 2 will be momentarily displayed and the Freq, Ch1, and Ch2 annunciators will light. The Counte r will be ready to measure the f r equency ratio of a signal applied to CHANNEL 1 in relation to a signal applied to CHANNEL 2 (Ch1/Ch2).
RA TIO 2 TO 1 is also available (Ch2/Ch1).
Note that Rati o 1 to 2 and Rat io 2 to 1 me nu items are not availa ble in Counters that do not contain the optional Channel 2.
Note the result is not scale d by 10 0; it is not a percen tage.
2
The front-panel “arrow” (or entry/select) keys can also be used when sequen cing through the Other Meas menu. Refer to the se ction titled “Using Entry/Select (Arrow) Keys” for details.
To Measure Period
Press Other Meas key until PERIOD 1 is displayed.
PERIOD 1 is momentarily displayed, the Period and Ch1 annunciators light, and the Counter is ready to measure the period of a signal applied to CHANNEL 1.
Ope rati ng Guide 2-13
Page 82
Chapter 2 Operating Your Frequency Counter
Using the MEASURE Menu Keys
To Measure Positive/Negative Volt Peaks
Press Other Meas key until VOLT PEAKS 1 is displayed.
VOLT PEAKS 1 is momentarily displayed, the Ch1 annunciator lights, and the Counter is re ady to m easu re minimu m and maxi mum volta ge peak s of si gnals appli ed to CHANNEL 1.
NOTE In Volt Peaks funct ion, t he Co unter m easu res th e input s i gnal after th e s ign al has be en
conditioned by the input settings (impedance, coupling, attenuation, and filter). One reminder of this is the (AC COUPLED) message, whic h is displayed when Voltage
2
NOTE Arming is not used for Voltage Peaks mea surement s. Also, Limi ts, Math, and Displ ay
Peaks is sele cted and the measurement channel is ac coupled.
The result is corrected for X10 attenuation.
Digits are not available for Voltage Peaks measurements.
2-14 Operating Guide
Page 83
Chapter 2 Operating Your Frequency Counter
Using the Gate & External Arm Menu Key
Using the Gate & External Arm Menu Key
Overview of Gate/External Arming Functions
MEASURE
Freq Ch 1
Other Meas
Freq Ch 2
Gate & ExtArm
Table 2-2 lists the menus for all the arming configurations, an d shows how the Gate and External Arming menu chan ges as a func tion of the arming mode.
The Counter has four arming modes: auto, external, digits and time. Thearming modes are defined in the following paragraphs.
TIME Arming
For frequency, period, and frequency ratio, time arming means you can set the gate time—the length of time for which your signal is measured.
NOTE Gate time and resolution are linked: the longer the gate time, the gr eater the
resolution. With an increase in resolution, you get a larger number of digits on the display.
Time arm ing is dem onstrated in the sub-s ection titled “Example Procedure for Gate and External Arm.”
AUTO Arming
For frequency, period, and frequency ratio, auto arming means make individual measur ements as f ast as poss i b le.
EXTERN AL Arming
2
For frequency, period, and frequency ratio, external arming means you c an start a measurement by using an external signal.
NOTE When external arming mode is enabled, a signal must be connected to the Count er’s
rear-panel Ext Arm connector.
Ope rati ng Guide 2-15
Page 84
Chapter 2 Operating Your Frequency Counter
Using the Gate & External Arm Menu Key
DIGITS Arming
For frequency, period, and frequency ratio, digits ar ming means you can set th e number of digits of resolution that you require your result to provide.
Digits arming is dem o n st r at ed in th e su b- section ti tled “Exa mp le Proced u r e fo r Changing the Number of Digits Displayed.”
*Table 2-2. The Gate & External Arm Key Menus as a Function of the
Arming Mode
Frequency, Period, Ratio
2
Time Arming:
GATE: TIME TIME: <time>
Auto Arming:
GATE: AUTO
Digits Arming:
GATE: DIGITS DIGITS: <digits>
External Armi ng (Refer to page 2-17 or the Menu Roadmap in the Quick R eference Guide, which follows the Preface, for more information.):
GATE: EXTERNL
START: POS
NEG
STOP: AUTO
NEG POS TIME
TIME: <time>
* There is no Gate and External Arm menu available for Voltage Peaks measurements.
2-16 Operating Guide
Page 85
Chapter 2 Operating Your Frequency Counter
Using the Gate & External Arm Menu Key
To Use the Gate and External Arm
Example Procedure for Gate and External Arm
For demonstr ation purposes, you want t o set up the Counter so that it uses an external arm to start a measurement. You want to make a Frequency measurement on a signal applied to channel 1. Also, you want to perform the measurement during an interval of 5 milliseconds as shown in the following figure.
NOTE As shown in the following figure, the Counter is alwa ys synchronized to the input
signal. Arming occurs on the rising edge of external arm (ExtArm), and the measurement gate is synchronized to the trigger event on CHANNEL 1.
CHANNEL 1
Ext Arm
2
Ext Arm
Gate
5 ms
Perform the following procedure to accomplish this task.
1 Turn on the Counter, and press Stop/Single to put the Counter into the single
measurement mode.
2 Connect the signal to channel 1 input, and press Freq Ch 1 key.
3 Connect an external TTL signal to the Counter’s rear-panel Ext Arm connect o r.
4 Press Gate & ExtArm ke y.
GATE: TIME is displayed.
5 Press any one of the arrow keys until GATE: EXTERNL is displayed.
6 Press Gate & ExtArm or Enter key.
Ope rati ng Guide 2-17
Page 86
Chapter 2 Operating Your Frequency Counter
Using the Gate & External Arm Menu Key
START: POS is displ ayed. (If START: NEG is displayed instead, then press any one of the arrow keys to toggl e to the next state—POS).
7 Press Gate & ExtArm or Enter key.
STOP: NEG is displayed.
8 Press any one of the arrow keys until STOP: TIME is displayed.
9 Set the Gate Time to 5 m s by performing the following steps:
2
a. Press Gate & ExtArm or Enter key.
TIME: .100 seconds (the default time) is display ed.
b. Press the appropriate arrow keys until TIME: .00500 s is displayed. c. Press Enter key.
NOTE BE SURE to press the Enter key to complete the numeric entry.
The counter will now gat e for a duration of 5 ms. The 5 ms interval will begin aft er the external edge is detected.
10 Press Stop/Single key to start the measurement.
The measurement is started after the external arm pulse occurs.
2-18 Operating Guide
Page 87
Chapter 2 Operating Your Frequency Counter
Using the Gate & External Arm Menu Key
Example Procedure for Changing the Number of Digit sof Resolution Displayed for More Precise Measurements
1Turn POWER key OFF then ON again (cycle POWER key) to preset the Counter.
2 Connect (for demonstration purposes) the 10 MHz Out signal to the channel 1
input.
3 Press Other Meas key until PERIOD 1 is displayed.
After a few seconds, 0.100,000,000,0 µs is di spla yed. He nce, the de fault number of digits (or resolution) di splays a precise measurement.
4 To demonstrate how to use “digits arming” to change the number of digits
displayed, perform the following steps first.
a. Press Gate & ExtArm key
GATE: TIME is displayed.
2
b. Press any one of the arrow ke ys until AUTO is displayed. c. Press Enter or Run key.
0.100,0 µs is displayed. Note that fewer digits are displayed.
In “auto arming” you produce results quickly. The short gate time produces a lower-resolution result.
Ope rati ng Guide 2-19
Page 88
Chapter 2 Operating Your Frequency Counter
Using the Gate & External Arm Menu Key
5 To cha ng e the disp lay to th e better measurement re solution using digits armin g,
perform the following steps:
a. Press Gate & ExtArm key
GATE: AUTO is displayed.
b. Press any one of the arrow ke ys until DIGITS is display ed. c. Press Gate & ExtArm or Enter key.
DIGITS: 4 is displayed (the default number of dig its )
2
d. To set the number of digits displayed to 8, pr ess the appropriate arrow keys until
DIGITS: 8 is displayed.
e. Press Enter key. f. Press Run key.
0.100,000,000 µs is displayed.
1 2 3 4 5 6 7 8
Note that 8 digits of res olution are displayed (0.100,000,000 µs).
For the convenience of the user, the leading “1” does not count as a digit when 8 digits are selected; thus, a resul t of 0.099,999,999 µs can alternate with a result of 0.100,000,000 µs with a stable display, and no change of settings is required.
2-20 Operating Guide
Page 89
Chapter 2 Operating Your Frequency Counter
Using the MATH Menu Key s
Using the MATH Menu Keys
Note tha t Math and Limits are not availab l e for Voltage Peaks measurements.
MATH
Scale & Offset
Stats
Overview of Scale/Offset Math Menu
The Scale and Offset functions within the Math menu allow you to perform simple mathematical operations on the measurem ent result before it is displayed. The Scale and Offset functions allow for multiplication and addition. Modification of the displayed meas urem ent by these math operations is represented by the following equation:
(Measurement × Scale) + O ffset = Displayed Result
The mat h ope rat io ns c an be us e d, fo r ex amp le, to sub tra ct sys t emat ic erro r s or dis pl ay the percentage difference bet ween signals. Also, the overall math operati on can be disabled and then re-established without having to re-enter constants.
The menu items of the Scale and Offset Math functions allow you to:
enter a desired multiplication factor for a measurement (SCAL:).
enter a desired addition or subtraction val ue for a measu rem ent (OFFS:).
get and display the last measurement val ue, and use it as a negative offset (SET REF: NO or YES).
disable or enable the Math mode (MATH: OFF or ON)—Note that the Math mode is automatically enabled when you enter the scale and/or offse t value(s); thus, the primary function of this menu item is to allow you to turn the Math mode off.
2
get Help, which displays the equation “( MEA S X SCALE) + OFF S = RESULT ” to tell you how the scale and offse t results are computed by the
Counter (MATH HELP: NO or YES).
Ope rati ng Guide 2-21
Page 90
Chapter 2 Operating Your Frequency Counter
Using the MATH Menu Key s
To Use the Scale/Offset Math Menu
Example Procedure for Scale Function
For demonstration purposes, you have a motor tha t has a tachometer attached to its rotating shaft that generates a pulse for every revolution of the shaft.
Your task is to measure the speed or revolutions per minute (rpm) of the tachometer’s output, and make the Counter directly display the measurement in rpm (which is fundamentally a frequency measurement). The equation for this task is :
Frequency (Revolution/Sec) × Scal e ( 60 Sec/Min) =
2
1 Connect the output to channel 1 input, and press Freq Ch 1 key.
Results (Revolution/Min)
Perform the following procedure to accomplish this task.
The frequency of the signal is displayed.
2 Press Scale & Offs et key until SCAL: 1.000000 is displayed.
3 S et th e Sca le to 60 (sin ce there are sixty seco n ds in a minut e) b y perf o rming the
following steps:
a. Press appropriate arrow keys to s et Scale to 60. b. Press Enter key.
NOTE BE SURE to press the Enter key to complete the numeric entry.
The scale is now set to 60 .
When th e scale factor is s et , bo t h sc al e an d offs e t ar e ap p l ie d to th e me asureme nt result. Verify that the offs et value is zero for applica tions requiring only scale by performing the following step.
4 Press Scale & Offs et key, if required, until OFFS: is displayed.
OFFS: 0.000000 should be displayed. If the Offset value is not set to “0.000000”:
2-22 Operating Guide
Page 91
Chapter 2 Operating Your Frequency Counter
Using the MATH Menu Key s
a. Press the appropriate arrow ke ys to set the Offset to “0.000000.” b. Press Enter key.
NOTE BE SURE to press the Enter key to complete the numeric entry.
The offset is now set to 0.
5 Press Run key.
The rpm measurement is displayed. Note that units indicators Hz and s do not ap pear when MATH is ON.
Example Procedure for Off set Function
The offset function can be used to add or subtract a fixed value from measuremen t results. This can be very us eful, for example , wh en tuni ng or alig ning odd freque ncies of a loca l o scillator (LO). The targ et frequency can be entered as a negativ e offset, then the Counter can display the differences between the LO’s frequency and the target frequency. You can then adjust the LO until zero is dis played in the Counter’s display.
2
Your task is to use the offset function to tune or align an LO to the target frequ ency of
222.127 MHz. Perform the following procedure to accomplish this task.
1Cycle POWER key to preset the Counter.
2 Connect the LO to channel 1, and press Freq Ch 1 key to measure the frequency
of the LO.
3 Press Scale & Offs et key until SCAL: is displayed.
4 S et S ca l e t o “1” by pres s in g t he a p propriate ar row k ey s t o ent e r t h e val u e
1.000000 if Scale is no t already set to “1.”
5 Press Enter key.
NOTE BE SURE to press the Enter key to complete the numeric entry.
Ope rati ng Guide 2-23
Page 92
Chapter 2 Operating Your Frequency Counter
Using the MATH Menu Key s
6 Press Scale & Offs et key, if required, until OFFS: is displayed.
OFFS: 0.000000 should be displayed. This is the default value.
7 Set the Offset to -222.127 MHz.
Refer to the sub-section titled “Entering the Offset Value” in Ch apter 1, “Getting Started,” for details on how to perform numeric entry for Offset.
a. Use the appropriate combination of the arrow keys to display a va lue of 222.127
MHz.
2
NOTE BE SURE to press the Enter key to complete numeric entry.
b. Press +/ key, which is located below the arrow keys, to change the pl us sign to
minus. OFFS: -222.127 MHz is displayed.
c. Press Enter key.
8 Press Run key.
The resultant frequency is displayed with -222.127 MHz subtracted from each measurement result.
The Counter displa ys the error or differences from the targeted 222. 127MHz.
For example, the dis play might indica te +000.023 MHz.
9 Adjust the LO as necessary until the Counter displays zero (00.000,000,0).
Your local oscillator is now properly aligned.
Example Procedure for T urning Off Math Mode
1 Press Scale & Offs et key until MATH: ON is displayed.
2 Press any one of the arrow keys.
2-24 Operating Guide
Page 93
Chapter 2 Operating Your Frequency Counter
Using the MATH Menu Key s
MATH: OFF is displayed. Now, Scale and Offset no longer are used to modify the displayed result.
3 Press Run key to return the Counter to its normal operation.
Example Procedure for Set Reference Function—Offset From the Last Measurem ent Value (Seeing Digits Not Visible on the 12-Di g it Disp l ay)
The SET REF: YES menu item uses the negative of the last measurement value, rounded to eleven digit s, to set the OFFSET. It also resets the scale value to 1. 0. Thus, any di fference in the offset (or now the reference value) and the current measurement value is displayed.
Let’s assume you are measuring the frequency of a stable source, but you are only interested in the least significant digits of the measure me nt. Use the SET REF: YES menu item to configure the appropriate OFFSET value.
Your task is to measure a stable sour ce to obtain the changes in the least significant digits.
2
1Cycle POWER key to preset the Counter.
2 Change gate time to 15.0 seconds on GATE menu.
3 To demonstrate SET REF:, connect a 10 MHz sine wave signal to channel 1 of
the Counter, and press Freq Ch 1 key.
(Note that since the gate time is set to 15 seconds, it will take about 15 seconds to for the Counter to dis play the measurement.)
10.000,000,000,1 MHz is displayed.
4 Press Scale & Offs et key until SET REF: is display ed .
5 Press the any one of the arrow keys to select SET REF: YES, and press Enter
key.
OFFS: −10.00000 M is displayed, for example.
Ope rati ng Guide 2-25
Page 94
Chapter 2 Operating Your Frequency Counter
Using the MATH Menu Key s
Note that this value represents the las t m easurement of the input signal rounded to eleven digits . The Counter uses it as the reference offset value, and reverses the sign from + to − .
6 Press Run key.
12* µ (120 µs) is displayed, for example, at the completion of the next me as urement.
This value (−0.00012) is the small difference between the signal being measured and the reference value (OFFS: 10.00000 M) obtained in step 5.
Using the reference al lows you to observe small differences in the
2
13th–15th di gits which would not be obser vable in the 12-digit dis play.
Overview of Statistics (Stats) Menu
The Stats menu allows you to:
select which computed statistics you want displayed
(SHOW: STD DEV , MEAN, MAX, or MIN).
choose between ha ving the actual measu rem ent or statistical result displaye d
(SHOW: MEAS).
select the number of measuremen ts you want to make for each compute d sta tist ic
(N:).
enable or disabl e the Stats mode (STATS: ON or OFF).
enable the Counter to perform statistical computation on only measurements that
are within the user-entered limits; that is, filter data (USE: IN LIMIT), or
enabl e th e Counter to pe r f orm statis ti cal compu t at io n on al l m e asu r ements,
whether they fall in or out of the user-entered limits (USE: ALL MEAS).
configure Stop/Single key to initiate a set of N measurements (ON SINGLE: 1 or N).
2-26 Operating Guide
Page 95
Chapter 2 Operating Your Frequency Counter
Using the MATH Menu Key s
To Use the Stats Menu for Automatic and Continuous Statistical Analysis
Example Procedure for Computing Stats
For demonstration purposes, let’s say you need to know the mean (average) and minimum frequency values of a signal. Also, you want th e Counter to make 20 measurements before it performs these statistical computations.
Perform the following procedure to accomplish this task.
1 Connect the signal to channel 1 input, and press Freq Ch 1 key.
2 Press Stats key until N: is displayed.
N: 100 is displayed (100 is the default value).
3Set N to 20 (since you want the Counter to make 20 measurements before
computing the mean and minimum values) by pressing the appropriate arrow keys.
N: 20 is displayed.
4 Press Enter key.
NOTE BE SURE to press the Enter key to complete numeric entry.
The Counter is now set to make 20 measurements before computing statistics.
5 Press Stats key until SHOW: is displayed.
6 Press any one of the arrow keys until SHOW: MEAN is displayed.
2
Ope rati ng Guide 2-27
Page 96
Chapter 2 Operating Your Frequency Counter
Using the MATH Menu Key s
7 Press Run key.
Immediately after the Run key is pressed, the Counter momentarily displays DOING STATS. When the Counter has made 20 valid measurements, it then displays the mean
frequency value of the input signal.
NOTE The Counter will display updated statistics results every 20 measurements.
8 Press Stop/Single key after you have a mean value displayed.
9 To disp lay the minimum frequency valu e, perform the following:
2
a. Press Stats key. b. Press any one of the arrow ke ys until SHOW: MIN is displa yed. c. Press Freq Ch 1 key.
CAUTION Pressing the Run key now would clear the present statistics results and calculate a
new set of results after “N” measurements.
Example Procedure for Easy Viewing of Stats
The d and f arrow keys provide a short cut to viewing eac h of the stats (that is, standard deviation, mean, maximum, and minimum) when the Counter is already displaying a statistic measurement. These keys allow you to cycle through the different Stats results and automatically view them.
While th e C ou n t er is dis pl a y in g a sta ti st i cs measurement, press ei t her the d or f arrow key to cycle through and display each of the four computed statistics, and the current measuremen t .
NOTE The Display Digits keys (that is, More Digits and Fewer Digits) do NOT affe ct
standard deviation display results.
2-28 Operating Guide
Page 97
Chapter 2 Operating Your Frequency Counter
Using the MATH Menu Key s
Example Procedure for Filteri ng Data (Using Limits) During Stat s
1 Using the procedure in the previous sub-section titled “Example Procedure for
Computing Stats,” set up the Counter to display the minimum frequency value of an input signal.
2 Set the upper and lower limits. (Refer to th e section in Cha pter 1 titled “ToSet
Limits of Measurements” if needed.)
3 Press Stats key until USE: is display ed .
4 Press one of the arrow keys to toggle to IN LIM IT state.
USE: IN LIMIT is displayed.
5 Press Run key.
The Counter will now perform the statistica l computations (that is, standard deviation, mean, maximum, and minimum) on only those measurements that are within the user-entered limits. The Counter then displays the statistics you picked (minimum, in this case).
6 Press Stats key until USE: is display ed .
7 Press any one of the arrow keys to toggle to ALL MEAS state.
USE: ALL MEAS is displ ayed.
8 Press Run key.
The Counter will now perform the statistica l computations (that is, standard deviation, mean, maximum, and minimum) on all measurements, whether they fall in or out of the user-entered limits. The Counter then displays the statistic you picked (minimum, in this case).
2
Ope rati ng Guide 2-29
Page 98
Chapter 2 Operating Your Frequency Counter
Using the MATH Menu Key s
Example Procedure for Configuri ng SI NGLE to Initiate N Measurements
Please re f er to th e p rocedure ti tl ed “To Use the Me asureme nt Co n tr o l Ke y s” on page 2-7 in th is chapt er for an example use o f ON SINGLE.
Example Procedure for T urning Off Stats Mode
1 Press Stats key until STATS: ON is displayed.
2 Press any one of the arrow keys.
STATS : OFF is displayed.
2
3 Press Run key to return the Counter to its normal operation.
2-30 Operating Guide
Page 99
Chapter 2 Operating Your Frequency Counter
Using the LIMITS Menu Keys
Using th e LIMITS Men u K e ys
Note tha t M ath and Limits menus are not available for Voltage Peaks measurements.
LIMITS
Uppr & Lower
Limit Modes
Overview of Limits Menus
The menu items under the Li mi ts keys allow you to:
selec t the desired upper and lower measurement limits (UPPR:, LOWR:).
disable or enable Limit Testing (LIM TEST : OFF or ON)—Note that the Limit
Testing is automatically enabled when you set the upper and lower li mits, or update any me nu i tem in th e Lim it Mode s menu; th us, t he prim ar y functi on o f the LIM TEST menu item is to all ow you to turn Limit Testing off.
set the Counter to stop making measurements when a measurement exceeds the
user-ent ered limits (ON FAIL: STOP). (Stop on limits does not have any effect if the Co unter is in th e s i ng l e measurement mod e.)
enable the Coun ter to continue taking measurements after a measurement
exceeds the user-entered limits (ON FAIL: GO ON).
change the Count er display from numeric to analog graph to quickly visualize changes in the signal (or to quickly see when measurements are in or out of limits) (SHOW: NUMBER or GRAPH).
2
Ope rati ng Guide 2-31
Page 100
Chapter 2 Operating Your Frequency Counter
Using the LIMITS Menu Keys
To Set and Use Automatic Limit Testing
Limits Testing Example 1—Flag and Stop Measurin g On Limits
Automatic limit testing can be used to monito r a test signal to assure that it does not drift outside of a us er-entered set of limits.
For demonstration purposes, you want to first measure the frequency of a 10 MHz signal applied to channel 1 of the Counter. Next, you want to set limits that would cause the Counte r to flag (turn on the Limits annunciator in the front-panel display) and stop making measure me nts if this signal drifts more than ±1 kHz.
2
NOTE After you enter the value 10.001 MHz, BE SURE to press the Enter key to
1 Connect the 10 MHz input signal to channel 1 of the Coun ter, and press Freq Ch
1 key.
Assume that the frequency measure d is 10 MHz. If the drift is not to be larger than ±1 kHz, th en the uppe r limi t sho uld be set t o 10.001 MHz and the l ower lim it to 9.999 MHz. Let’s se t these values in the following st eps.
2 Press Uppr/Lower key until UPPR: is displ ayed.
UPPR: 0.000000 is displayed (0 is the defa ult value).
3Set UPPR to 10.001 MHz. (Refer to the se ction in Chapter 1 titled “To Set Limits of
Measu r em en ts” if need e d. )
compl et e th e ent r y.
4 Press Uppr/Lower key, if required, until LOWR: is displayed.
LOWR: 0.000000 is displayed (0 is the default value).
2-32 Operating Guide
Loading...