U-RIGHT TD-4234 Owner's Manual And Usage Manual

TD-4234
311-4234100-005
Version 2.0, 2010/04
OWNER’S MANUAL
BLOOD GLUCOSE
MONITORING SYSTEM
Thank you for choosing
BLOOD GLUCOSE
MONITORING SYSTEM
Dear U-RIGHT TD-4234 System Owner:
This manual contains important information you must know about the
system. Please read it thoroughly and carefully.
For other questions regarding this system, please contact your local
customer service or place of purchase. For all other questions, please consult your healthcare professional.
TD-4234
IMPORTANT SAFETY INSTRUCTIONS
READ THIS BEFORE USE
The following basic safety precautions should always be taken.
1. Close supervision is necessary when the device is used by, on, or near children, handicapped persons or invalids.
2. Use the device only for the intended use described in this manual.
3. Do not use accessories which are not supplied by the manufacturer.
4. Check that the voltage setting matches the supply voltage.
5. Do not place the equipment in liquid, nor put it where it could fall into liquid. If the equipment becomes wet, unplug it before touching it.
6. Do not let the equipment or its exible cord come into contact with
surfaces which are too hot to touch.
7. Do not use the equipment where aerosol sprays are being used, or
where oxygen is being administered.
8. Do not use device if it is not working properly, or if it has suffered any damage.
9. Before using product to test your blood glucose, read all instruc­tions thoroughly and practice the test. Do all quality control checks as directed and consult with a diabetes healthcare professional.
KEEP THESE INSTRUCTIONS
TABLE OF CONTENTS
IMPORTANT INFORMATION 06
ABOUT ALTERNATIVE SITE TESTING (AST) 07
INTRODUCTION TO THE SYSTEM 09
Intended Use 09
Principle of Measurement 09
Contents of the System 10
Appearance and Key Function of the Meter 11
Indicators for Main Control Mode 12
Test Strip 13
BEFORE USE 14
Battery Recharge 14
Indicators for System Setting Mode 15
Modify Meter Settings 16
Alarm Function 19
Display Self-Checks 20
CONTROL TEST 21
Important Control Solution Information 22
Performing a Control Solution Test 23
TESTING YOUR BLOOD GLUCOSE 26
Testing Your Blood Glucose 27
Expected Test Results 32
COMPARING METER AND LABORATORY RESULTS 33
USING THE METER MEMORY 35
Indicators for Memory Recall Mode 35
Viewing Results on the Meter 37
VIEWING RESULTS ON A PERSONAL COMPUTER 41
TAKING CARE FOR YOUR METER AND TEST STRIPS 43
Cleaning 43
Storage 43
PROBLEM-SOLVING GUIDE 44
Special Message 47
Error Message 48
Operational Problem 49
SPECIFICATIONS 51
IMPORTANT INFORMATION
·Severe dehydration and excessive water loss may cause readings
which are lower than actual values. If you believe you are suffering from severe dehydration, consult a healthcare professional im­mediately.
·If your blood glucose results are lower or higher than usual, and you
do not have symptoms of illness, rst repeat the test. If you have
symptoms or continue to get results higher or lower than usual, follow the treatment advice of your healthcare professional.
·Use only capillary whole blood sample to test your blood glucose. Using other substances will cause wrong results.
·If you are experiencing symptoms that are inconsistent with your
blood glucose test results and you have followed all instructions de­scribed in this owner’s manual, call your healthcare professional.
·Inaccurate results may occur in severely hypotensive individuals or patients in shock. Readings which are lower than actual values
may occur for individuals experiencing a hyperglycemic-hyperosmolar
state, with or without ketosis. Critically ill patients should not be tested with blood glucose meters.
·Please refer to your test strip package insert for additional important information.
Important: AST results serve as references only. Please consult your healthcare professional before you begin AST.
What is AST?
Alternative site testing (AST) refers to when individuals check their blood glucose levels using areas of the body other than the ngertip. This system allows AST at the palm, the forearm, the upper arm, the calf or the thigh with results as reliable as those obtained
from ngertip testing.
What’s the advantage?
Fingertips feel pain more readily because they are full of nerve endings (receptors). Since nerve endings are not so condensed at other body sites, tests performed there will be less painful.
When to use AST?
Food, medication, illness, stress and exercise can affect blood glucose levels. Capillary blood at the ngertip reects these changes
faster than capillary blood at other sites. Therefore, when testing
blood glucose during or immediately after a meal or physical exercise, results from AST and ngertip may be signicantly different.
Alternative Site Testing (AST)
We strongly recommend you do AST in the following intervals:
˙In a pre-meal or fasting state (more than 2 hours since the last meal). ˙Two hours or more after taking insulin. ˙Two hours or more after exercise.
Do NOT use AST if:
˙You think your blood glucose is low. ˙You are unaware of hypoglycemia. ˙You are testing for hyperglycemia. ˙Your AST results do not correspond with the way you feel. ˙Your routine glucose results are often uctuating.
How to increase the accuracy?
Stimulating blood perfusion by rubbing the puncture site prior to
blood extraction has a signicant inuence on the glucose value obtained. Blood from a site that has not been rubbed exhibits a measurably different glucose concentration than blood from the nger. When the puncture site is rubbed prior to blood extraction, the
difference is markedly reduced.
Please follow suggestions below before puncturing the skin:
˙Rub the puncture site approximately 20 seconds before penetration. ˙Use a clear cap (included in the kit) while setting the lancing device.
INTRODUCTION TO THE SYSTEM
Intended Use
The system is intended for use outside the body (in vitro diagnostic use). It should be used only for testing glucose (sugar) and only
with fresh capillary whole blood samples (from the nger, the palm,
the forearm, the upper arm, the calf and the thigh). The system is intended for use in the home and in clinical settings. It should not be used for the diagnosis of diabetes or for the testing of newborns.
AST in this system can be used only during steady-state blood glucose conditions described in the section of “About Alternative Site Testing (AST).”
Principle of Measurement
The test is based on the measurement of electrical current generated by the reaction of glucose with the reagent of the strip. The meter measures the current and displays the corresponding blood glucose level. The strength of the current produced by the reaction depends on the amount of glucose in the blood sample.
10
Contents of the System
The system consists of three main products: the blood glucose meter, test strips, and a control solution. These products have been designed, tested, and proven to work together as a system to produce accurate blood glucose test results. Use only the U-RIGHT TD-4234 test strips and a control solution with the blood glucose meter. Your system includes:
A meter
Lancing device (a clear cap
for AST)
First time user guide
Owner’s manual
Carrying bag
Warranty card
Daily log book
Control solution
Test stirp
Lancet
9 10
10
NOTE
1. Please make sure that all products listed above are included and sealed
in the package before using this system. If you nd any imperfection in our
products, please return the whole system to the place of purchase.
First time user guide
Daily log book
1 3
7
2
8
4 5 6
11
Appearance and Key Function of the Meter
LCD SCREEN guides you through the test using symbols and
simple messages.
TEST SLOT is where you insert the test strip. The meter will turn on
automatically after insertion.
STRIP-EJECTION BUTTON is where the used strip will be
automatically ejected after you push up the button.
UP BUTTON adjusting increment and change main / sub-function.
DOWN BUTTON adjusting decrement and change main /
sub-function.
M BUTTON is used to turn on and navigate through settings.
MINI USB PORT is for data transmission.
7
1
3
7
2
4
5
6
12
Indicators for Main Control Mode
The screen displays several indicators that help you to perform the main functions of this meter. When you start to use the meter, press the M button to turn on the meter and then U-RIGHT logo with date and time appears. Press M button again to enter the Main Control Mode. Press Up or Down button to select the below desired mode:
Glucose Measuring Mode:
Perform your glucose tests.
Memory Recall Mode:
Review your glucose test results in the meter
memory.
System Setting Mode:
Set up the meter such as date, time, the unit of
measurement and memory deletion.
PC Linking Mode:
Connect the meter to PC for data transmission or
recharge the battery through USB adapter.
System Power Off:
Turn off the meter.
Meaning / You can do thisIndictor
13
Test Strip
This system measures the amount of sugar (glucose) in whole blood. Blood is applied to the absorbent hole of the test strip and is automatically drawn into the reaction cell where the reaction takes place.
The test strip consists of the following parts:
Contact Bars
Insert this end of the test strip into the meter.
Push it in rmly until it will go no further.
Test Strip Handle
Hold this part to insert the test strip into the slot.
Conrmation Window
This is where you conrm if enough blood
has been applied to the absorbent hole of the strip.
Absorbent Hole
Apply a drop of blood here, The blood will be sucked automatically.
14
BEFORE USE
Battery Recharge
Your meter comes with built-in rechargeable Li-polymer battery which allows recharging the meter. If the low battery symbol
appears on the screen indicating that the battery is low and it is
time to recharge the battery.
1. Connect the USB adapter to the side of the
meter.
2.Plug the USB adapter to a wall outlet or
connect the other end of USB cable to the
USB port on your personal computer.
3.The recharge symbol shows when bat-
tery is recharging. After the battery is fully
charged, remove the USB cable and then
the meter will return to Main Control Mode.
It will automatically turn off if left idle for 2
minutes.
NOTE
Recharging the battery does not affect the test results stored in memory.
15
Indicators for System Setting Mode
Your meter comes with the time, date, date format, unit of measurement and memory deletion preset. The following indicators for System Setting Mode help you to set your meter easily.
Time and Date Setting:
Set the date and time for the meter
Clock Alarm Setting:
Set the hour and minute of the alarm for the
meter.
Unit of Measurement Setting:
Set the unit of measurement for the meter.
Memory Deletion:
Delete the memory record for the meter.
Exit Mode:
Exit the system setting mode.
Meaning / You can do thisIndictor
12-2514:52
12- 25 14:52
12-2514:52
12-2514:52
Glucose Unit is mg/dL
12-2514:54
16
Follow the below steps to modify the meter settings. In Main Control Mode, press Up or Down button to select System Setting Mode indicator . Press M button to enter System Setting Mode.
Setting the Time and Date
Modify Meter Settings
Step1
Press Up or Down button to select Time and Date Setting indicator. Press M button to enter the setting.
Step2
Press Up or Down button to
increase or decrease the year. Press
M button to continue the setting.
Step3
Press Up or Down button to set date format: MM-DD or DD-MM. Press M button to continue the setting.
Step4
Repeat step 2 and 3 to set the month, date, 12/24 hour, hour and minutes.
After nishing these settings, press
M button to return to System Setting Mode.
14:52
14:52
12-25 14:52
12-25
DateExp : MM-DD
14:52
2009
2009
12-25
12-25
17
Memory Deletion
Step1
Press Up or Down button to select Memory Deletion indicator. Press M button to enter the setting.
Step2
Press Up or Down button to select Y/N to delete memory.
Step3
After nishing the setting, press M but- ton to return to System Setting Mode.
Unit of Measurement Setting
Step1
Press Up or Down button to select Unit of Measurement Setting indicator. Press M button to enter the setting.
Step2
Press Up or Down button to select mg/dL or mmol/L.
Step3
After nishing the setting, press M
button to return to System Setting Mode.
Clear Memory? N Clear Memory? Y
Glucose unit is mg/dL
Glucose unit is mmol
12-25 14:52 12-25 14:52
Glucose unit is mg/dL
12-25 14:52 12-25 14:54
18
1. The milligram per deciliter (mg/dL) is the standard unit in the United States.
The millimole per liter (mmol/L) is the standard unit in Canada. Use of the wrong unit of measure may cause you to misinterpret your blood glucose level, and may lead to incorrect treatment.
2. The time, date, date format, unit of measurement and memory deletion can
ONLY be changed in the setting mode. Therefore, when you perform a glucose testing, those parameters are not possible to be changed.
3. Your meter displays 7-, 14-,-day averages which you can access from the
meter memory. These averages are calculated from results obtained during the 7-, 14-, -day preceding the current date and time settings. When the date and time are changed, the 7-, 14-, averages may change.
4. The meter will turn off automatically without any action in two minutes.
5. You can go back to the previous step by pressing M button for 3 seconds.
6. You can turn back to Main Control Mode to select System Power Off indicator
or press M button for 3 seconds to turn off the meter.
NOTE
Exit Mode
Please select the below screen when you want to exit
System Setting Mode and turn to Main Control Mode.
19
12- 2512:25
Timer 2 is up
10:35 PM
The meter provides four alarms. The alarm function alerts you to per­form your blood glucose regularly.
You can continue to set up alarm 2, 3, and 4 by following previous steps. You can set up all 4 alarms at the same time or just set any one of them.
If the alarm is left beeping for 2 minutes and no actions are taken, the meter will automatically turn off.
Alarm Function
Setting the Alarm
An example if set Alarm 2 at 10:35 PM
The meter automatically turns on at 10:35 PM everyday with beeping at a two seconds period, and with the time blinking. You can directly insert a new strip to perform a blood glu­cose test now or simply press the M button to turn it off.
Step1
Press Up or Down button to select alarm 1 on/off. Then press M button to set the alarm 1.
Step2
The hour will start ashing rst and you can
change the value by pressing Up or Down button.
Then press M button and proceed to the next step.
Step3
The minute will start ashing and you can change
the value by pressing Up or Down button. Then press M button to set the alarm 1 on.
If you chose "off" there will be no alarm set.
In Main Control Mode, press Up or Down button to select Alarm Setting and press M button to en­ter Alarm Setting indicator.
Alarm 1 is on
12:25
Alarm 1 is on
12:25
Alarm 1 is o n
12:25
Alarm 1 is off
--:--
20
Display Self-Checks
The system provides AC/PC modes. Glucose AC is the glucose value before meal; and PC is the glucose value after meal.
While you insert a test strip, the meter turns on automatically and displays a message: “Did you eat any thing within 4 hours?” Please press Up or Down button to select Yes for PC mode or No for AC mode and then press M button to conrm it.
If you select Yes, the screen displays as below:
If you select No, the screen displays as below:
Yes
No.038
M : Confir m
: Change
Did you eat any thing within 4 hours ?
No
No.038
M : Confir m
: Change
Did you eat any thing within 4 hours ?
21
CONTROL TEST
TaiDoc control solutions contain a known amount of glucose that re-
acts with test strips. By comparing your control solution test results with
the expected range printed on the test strip vial label, it is able to check
that the meter and the test strips are working together as a system and that you are performing the test correctly. It is very important that you do this simple check routinely to make sure you get accurate results.
How often should the control solution test be performed?
►When you use this system to test your blood for the first time,
practice the procedure using control solution. When you can do
three tests in a row that are within the expected range, you are ready
to test your blood.
►For routinely check the meter and test strips, perform a single test
for each level of control solution at least once a week.
When should the control solution test be performed?
►When you first get your meter. ►When you begin using a new vial of test strips. ►Whenever you suspect that the meter or test strips are not
working properly.
►When your blood glucose test results are not consistent with how
you feel, or when you think your results are not accurate.
►When your test strips are exposed to extreme environmental
conditions (See Storage section of this manual).
►When you want to practice running the test. ►If you drop the meter.
22
Important Control Solution Information
►Use only TaiDoc control solutions. ►Check the expiration date on the control solution vial. Do not use if
expired.
►Control solution, meter, and test strips should come to room
temperature 68 to 77°F (20 to 25°C) before testing.
►Shake the vial, discard the first drop of control solution, and wipe
off the dispenser tip to ensure a clean sample and an accurate result.
►Use only within 90 days after first opening. Record the discard
date (date opened plus 90 days) on the control solution vial. Discard after 90 days.
►Store the control solution tightly closed at temperatures 36 to
86°F(2 to 30°C). Do not freeze.
NOTE
The control solution range printed on the test strip vial is for control
solution only. It is used to evaluate meter and test strip performance.
It is not a recommended range for your blood glucose level.
23
Performing a Control Solution Test
Step1 Insert Test Strip
Insert a test strip with contact bars end rst and
facing up into the test slot. The meter will turn on automatically and display a message: ►“Did you eat any thing within 4 hours?” Please press the Up or Down button to select Yes/No
and then press M button to conrm it. ►The AC/PC mode with code number will
appear. Press Up or Down button to select QC (Quality Control), the screen will display as following:
Be sure the code number on the display is the same as the code number on the test strip vial.
Yes
M : Confir m
: Change
Did you eat any thing within 4 hours ?
No.038
No.038
WARNING!
Contact bars must be inserted all the way into the meter or you may receive
inaccurate test result.
24
Step2 Obtain Control Solution
Shake the control solution vial well. Remove the cap from the control solution bottle. Place
cap on at surface. Squeeze the vial, discard the rst drop, and wipe off the dispenser tip to
prevent contamination. Squeeze the vial again to produce another drop and place this drop
on the top of cap.
Step3 Apply Control Solution
While holding the meter, move the absorbent hole of the test strip to touch the drop of control solution. Then the drop will be
automatically absorbed. Once the conrmation window lls completely, the meter will begin to
count down.
To avoid contaminating the control solution with the contents of the test strip, you have to place a drop of control solution on a clean surface. Do not directly apply control solution to a strip.
25
Step4 Read and Compare the Result
Compare the result shown on the screen with the range printed on the test strip vial. It should fall within this range.
Out-of-range results
If test results fall outside the range printed on the test strip vial, check “Operating Problem” section in troubleshooting guide and repeat the test. If you continue to get out-of-range results, it means that the system may not be working properly. Do NOT test your blood. Contact the place of purchase for help.
109~165
mg/dL
120
QC
26
WARNING!
To reduce the chance of infection:
˙Never share a lancet or the lancing device.
˙Always use a new, sterile lancet. Lancets are for single use only.
˙Avoid getting hand lotion, oils, dirt, or debris in or on the lancets and the
lancing device.
TESTING YOUR BLOOD GLUCOSE
Be sure to read this section and the test strip package insert found
in the test strip box carefully before testing. Make sure you have all
items needed to test:
Blood Glucose Meter
Test Strip
Lancing Device
Sterile Lancet
Clear Cap (For AST use)
1
3
2
4
5
Cap
Cocking Control
Release Button
Lancet Holder
Ejector
Arrow
Depth Indicator
Adjustable Tip
Lancing device overview:
2 3 41 5
27
Testing Your Blood Glucose
Wash and dry your hands before starting.
► Pull off the cap of the lancing device.
► Insert a lancet into the lancet holder and push down rmly until it is fully
secured.
► Twist the protective disk off the lancet.
► Replace the cap by aligning the arrow
on the cap with the release button.
► Select the depth of penetration by turn-
ing the adjustable tip in either direction so that the arrow on the cap points to the desired depth.
► Pull the cocking control back until it
clicks. You will see a color change in side the release button when it is ready.
Set the lancing device
The Lancing Device offers you 5 depths of skin penetration:
Shallow Deep
Average skinSoft or thin skin Thick or calloused skin
28
Step2 Insert Test Strip Select Glucose Measuring Mode to perform a test.
The meter will display a message:
►“Did you eat any thing within 4
hours?” ►Please press Up or Down button to select and press M button to
conrm. ►Insert the test strip with contact bars end rst and facing up into the test slot.
►The AC mode and PC mode with
code number will appear. Be sure the code number on the display is the same as the code number on the test strip vial.
14:58
12-25
No
M : Confir m
: Change
Did you eat any thing within 4 hours ?
Yes
M : Confir m
: Change
Did you eat any thing within 4 hours ?
No.038
No.038
Blood from sites other than the ngertip
Substitute the lancing device cap with the clear cap for alternative site testing. Pull the cocking control back until it clicks.
When lancing the forearm, upper arm, hand, thigh, or calf, avoid lancing the areas with obvi-
ous veins in order to prevent excess bleeding.
The lancing device is now ready for use. Set aside for later use.
29
NOTE
˙Choose a different spot each time you test. Repeated punctures at the
same spot may cause soreness and calluses.
˙Before you decide to begin AST, please consult your health professional.
˙It is recommended that you discard the rst drop of blood as it might
contain tissue uid and serum, which may affect the test result.
Step3 Get a Drop of Blood
Select the puncture site either in nger or in other parts (AST). Clean the puncture site with 70% using cotton moistened with 70% alcohol and let
it air dry.
Fingertip testing Press the lancing device’s tip rmly against the lower side of your ngertip. Press the release but­ton to prick your nger, then a click indicates that
the puncture is complete.
Sites other than ngertip
Please refer to the section, “Alternative Site Testing”, for possible puncture sites.
After penetration, discard the rst drop of blood
with clean tissue paper or cotton. Then gently squeeze the punctured area to obtain blood. Be careful NOT to smear the blood sample. The volume of blood sample must be at least 0.7 microliter ( actual size).
30
NOTE
˙Do not push your punctured nger (with blood on it) against the test
strip or try to apply a smeared blood sample to the test strip.
˙If you do not apply a blood sample to the test strip within 3 minutes, the
meter will automatically turn off. You must remove and reinsert the test
strip to restart the test procedure.
˙The blood should be completely lled the conrmation window before
the meter begins to count down. If you nd that the conrmation window
is not lled with blood when the meter is counting, NEVER try to add more
blood to the test strip. Discard the test strip and retest with a new one.
˙If you have trouble lling the conrmation window, contact the place of
purchase for help.
Step4 Apply Blood into the Test Strip
Apply your blood to the absorbent hole of the test
strip until the conrmation hole is fully covered. The hourglass symbol will ash when the meter
is working.
Step5 Obtain an Accurate Result in 7 Seconds
The result of your blood glucose test will appear after the meter counts to 0. This reading will automatically be stored in the meter.
mg/dL
AC
73
31
Step6 Eject the Used Test Strip
After nishing the measurement, you can either
take out the used strip by simply pushing up the Strip-Ejection button or remove the test strip directly by your hand. The meter will return to Main Control Mode and turn off automatically without any action in two minutes.
Step7 Remove the Lancet
Always use caution when removing the lancet.
Safely dispose of the used lancet by placing the protective disk on a hard surface, and pushing
the exposed tip into the disk.
Remove the lancet by pulling the cap off rst.
Then slide the ejector forward to remove the used lancet.
WARNING!
The used lancet and the used test strip are potentially biohazardous.
Please discard them carefully according to your local regulations.
32
Expected Test Results
Blood glucose monitoring plays an important role in diabetes control. A long-term study showed that maintaining normal blood glucose levels can reduce the risk of diabetes complications by up to 60%.*1 The results provided by the U-RIGHT TD-4234 system can help you and your healthcare professional monitor and adjust your treatment plan to gain better control of your diabetes.
Source: American Diabetes Association (2008). Clinical Practice Recom-
mendations. Diabetes Care, 31 (Supplement 1): S1–108.
Please consult your doctor to determine a target range that works best for you.
Time of day
Fasting and before meal Less than 100 mg/dL(5.6 mmol/L)
Less than 140 mg/dL(7.8 mmol/L)2 hours after meals
Normal plasma glucose range for people without diabetes
33
COMPARING METER AND LABORATORY RESULTS
The meter provides you with whole blood equivalent results. The result you obtain from your meter may differ somewhat from your laboratory result due to normal variation. Meter results can be affected by factors and conditions that do not affect laboratory results in the same way. (See test strip package insert for typical accuracy and precision data, and for important information on limitations.) To make an accurate comparison between meter and laboratory results, follow the guidelines below.
Before you go to the lab:
˙ Perform a control solution test to make sure that the meter is working properly. ˙ Fast for at least eight hours before doing comparison tests, if possible. ˙ Take your meter with you to the lab.
While at the lab:
Make sure that the samples for both tests (the meter test and the lab test are taken and tested within 15 minutes of each other).
˙ Wash your hands before obtaining a blood sample. ˙ Never use your meter with blood that has been collected in a
gray-top test tube. ˙ Use fresh capillary blood only.
34
You may still have a variation from the result because blood
glucose levels can change signicantly over short periods of time, especially if you have recently eaten , exercised, taken medication, or experienced stress*3. In addition, if you have eaten recently, the blood glucose level from a nger prick can be up to 70 mg/dL higher
than blood drawn from a vein (venous sample) used for a lab test.*
4
Therefore, it is best to fast for eight hours before doing comparison tests. Factors such as the amount of red blood cells in the blood (a
high or low hematocrit) or the loss of body uid (dehydration) may
also cause a meter result to be different from a laboratory result.
*3: Surwit, R.S., and Feinglos, M.N.: Diabetes Forecast (1988), April, 49-51.
*4: Sacks, D.B.: “Carbohydrates. “ Burtis, C.A., and Ashwood, E.R.( ed.),
Tietz Textbook of Clinical Chemistry. Philadelphia: W.B. Saunders
Company (1994), 959.
35
USING THE METER MEMORY
Indicators for Memory Recall Mode
Your meter stores the 400 most recent blood glucose test results along with their respective dates and times in its memory. It also provides you with 7-, 14- averages of your blood glucose test results. The following indicators for Memory Recall Mode help you to review your results stored in the meter easily.
Glucose Results:
View the individual test results in the memory.
7-day Average Result:
View the 7-day average of the test results in the
memory.
14-day Average Result:
View the 14-day average of the test results in
the memory.
Glucose Trend Report for 7-day Average
Results:
View each 7-day average results in glucose
trend report.
Meaning / You can do thisIndictor
14:52
14:52
14:52
14:52
12-25
12-25
12-25
12-25
36
Glucose Trend Report for 14-day Average Re-
sults:
View each 14-day average results in glucose trend
report.
Exit Mode:
Exit the memory recall mode.
Meaning / You can do thisIndictor
14:52
12-25
37
Viewing Results on the Meter
Follow the below steps to review test results in memory. In Main Control Mode, press Up or Down button to select Memory Recall Mode indicator . Press M button to enter Memory Recall Mode.
Glucose Results
Step1
Press Up or Down button to select Glucose Results indicator. Press M button to enter glucose results reviewing.
Step2
Press Up or Down button to view the test results. The latest test result will appear rst. The meter will review up to your last 400 test results in order. When the memory is full, the oldest result is deleted as the newest is added.
Step3
After reviewing the result, press M button to return to Memory Recall Mode.
14:52
14:52
Memory Recall
14:52
mg/dLPC
193
008
8th record
Glucose result
PC mode
Date
Time
Unit of measurement
1
3
2
4 5
6
recall mode
12-25 12-25
12-25
38
7-day/14-day Average Results
Step1
Press Up or Down button to select 7-day Average Result or 14-day Average Result indicator. Press M button to enter day average results reviewing.
Step2
Press Up or Down button to select the category of results you want to view: AC (Before Meal), PC (After Meal).
Press M button to conrm it.
Step3
The day average is calculated from the blood glucose results obtained during the last 7 days or 14 days depends on which Average Result indicator you select.
Step4
After reviewing the result, press M button to return to Memory Recall Mode.
14:52
Memory Recall
14:52
Memory Recall
14:52
14:52
AC mode memory
AC
PC
Glucose result
7-day average
5 test results
Unit of measurement
AC mode
1
3
2
4
5
AC
mg/dL
112
Memory mode 7 days Amount = 0005
recall mode
recall mode
12-25
12-25
12-25
12-25
39
Glucose Trend Report for 7-day/14-day Average Results
Step1
Press Up or Down button to select the following screen. Press M button to enter glucose trend report reviewing.
Step2
Press Up or Down button to select the category of results you want to view: AC (Before Meal), PC (After Meal).
Press M button to conrm it.
Step3
Each 7-day/14-day average result on glucose trend report will display.
Step4
After reviewing the result, press M button to return to Memory Recall Mode.
14:52
Memory Recall
14:52
Memory Recall
AC mode memory
AC
PC
14:52
14:52
Period
7-day/14-day average
Glucose trend graph
1
erage Value is....
221 ‘ 08- 215 ’ 08
600
300
140
80
20
T
3
2
recall mode
recall mode
300
190
110
70
20
12-25 12-25
12-2512-25
40
Exit Mode
Please select the below screen when you want to exit System Setting Mode and turn to Main Control Mode.
NOTE
The meter will turn off automatically without any action in two minutes.
41
VIEWING RESULTS ON A PERSONAL COMPUTER
Results in memory can be transmitted to a personal computer. Health Care System Software and an USB cable are required before installation. The software can be downloaded from the home page of TaiDoc. To learn more about Health Care System Software, please contact the place of parchase for help.
Step1 Install Software
Install Health Care System Software on your Computer by following the instructions provided on TaiDoc's website, http://www.taidoc.com.
Step2 Enter the PC Linking Mode
Press the M button to turn on the meter and then “U-RIGHT” logo, date and time appear. Press M button again to select PC Linking Mode. Press M button again to enter it.
Step3 Connect to Personal Computer
Connect the USB cable to a serial port of your computer. Then, connect the USB cable to the data port of the meter. The indictor will appear on the display, indicating that the meter is ready to transmit data.
14:52
12-25
42
Step4 Transmit Data
Follow the instructions provided in the software to transmit data. Results with respective date and time will be transmitted. Remove the USB cable and then the meter will return to Main Control Mode. It will automatically turn off if left idle for 2 minutes.
NOTE
While the meter is connecting to the PC, it is unable to perform a blood
glucose test.
43
TAKING CARE OF YOUR MONITOR & STRIP
To protect the meter and test strips from dirt, dust and other contaminants, please wash and dry your hands thoroughly before use.
Cleaning
1. To clean the meter’s exterior, wipe with a cloth moistened with
tap water or a mild cleaning agent, then dry the device with a soft
and dry cloth. Do not ush with water.
2. Do not use organic solvents to clean the meter.
Storage
1. Monitor Storage
• Storage condition: -4 to 140°F (-20 to 60°C), below 95% relative
humidity.
• Always store or transport the meter in its original storage case.
• Avoid dropping and strong impact.
• Avoid direct sunlight and humidity.
44
2. Strip Storage
• Storage condition: 4 to 40°C, below 85% relative humidity. Do NOT freeze.
• Store your test strips in their original vial only.
Do not transfer to another container.
• Store test strip packages in a cool, dry place. Keep away from direct sunlight and heat.
• After removing a test strip from the vial, immediately replace the vial cap and close it tightly.
• Touch the test strip with clean and dry hands.
• Use each test strip immediately after removing
it from the vial.
• Write the discard date (the date opened plus 90 days) on the vial label when you rst open
it. Discard remaining test strips 90 days after
rst opening.
• Do not bend, cut, or alter a test strip in any
way.
• Keep the strip vial away from children as the cap and the test strip may be a choking hazard. If swallowed, promptly see a doctor for help.
45
3. Control Solution Storage
• Storage condition: Store the control solution with lid tightly closed at temperatures 2 to 30°C (36 to 86°F). Do NOT freeze.
• Record the discard date (date opened plus 90 days) on the control solution vial. Discard after 90 days.
46
PROBLEM-SOLVING GUIDE
The following is a summary of various display messages. These messages help to identify certain problems but do not always appear in all cases when a problem has occurred. Improper use may cause an inaccurate result without producing an error message. In the event of a problem, refer to the information under the row “ACTION”.
Never try to disassemble the meter under any circumstances. If you encounter any error messages not listed below or if you have followed the actions recommended below but the problem persists unsolved, please contact local customer ser­vice for help.
47
Special Message
Special symbols and messages appear together with your test result.
WHAT IT MEANSMESSAGE WHAT TO DO
Your result is below
measurement limit,
20 mg/dL (1.1 mmol/L).
This indicates hypoglycemia
(low blood glucose.)
You should immediately
treat hypoglycemia as
recommended by your
healthcare professional.
Your result is above
measurement limit,
600 mg/dL (33.3 mmol/L).
This indicates severe
hyperglycemia (high blood
glucose).
You should seek
immediate medical
assistance.
Out of range.
Lo
Out of range.
Hi
48
Error Message
WHAT IT MEANSMESSAGE WHAT TO DO
Strip has been used. Start again with a new
test strip.
No strip is inserted to the
meter during measuring.
Insert a test strip.
The battery is low.
Recharge the battery
immediately.
Appear when
environmental temperature
is below system operation
range.
Meter malfunction.
System operation range
is 10~40°C (50~104°F).
Repeat the test after
the meter and test strip
have reached the above
temperature.
Please contact local
customer service for
assistance.
Charge the battery, please.
None strip.
Temperature...
Strip has been...
Terrible error.
49
2. If the test does not start after applying the sample:
POSSIBLE CAUSE
Insufcient blood sample.
Defective test strip.
Sample applied after automatic
shutdo
Defective meter.
WHAT TO DO
Repeat the test using a new test
strip with larger volume of blood.
Repeat the test with a new test
strip.
Repeat the test with a new test
strip.
Please contact the local customer
service for help.
Operating Problem
1. If the meter does not display a message after inserting a test strip:
POSSIBLE CAUSE
Batteries exhausted.
Test strip inserted upside down or
incompletely.
Defective meter.
WHAT TO DO
Recharge the battery.
Insert the test strip facing up with
the contact bars end rst.
Please contact local customer
service for help.
50
3. If the control solutions test result is out of range:
POSSIBLE CAUSE
Error in performing the test.
Control solution poorly shaken.
Expired or contaminated control
solution.
Control solution that is too warm or
too cold.
Test strip deterioration.
Monitor malfunction.
WHAT TO DO
Read the instructions thoroughly
and repeat the test.
Shake the control solution vigor-
ously and repeat the test again.
Check the expiration date or the dis-
carded date of the control solution.
Place control solution, meter and
test strips in room temperature (68
to 77°F/20 to 25°C) before testing.
Repeat the test with a new test
strip.
Please contact the local customer
service for help.
51
SPECIFICATIONS
Power Source:
built-in rechargeable Li-polymer battery
Dimension & Weight: 84mm(L) x 59mm(W) x 17mm(H), 46g
Memory:
400 measurement results with respective dates and times
Auto electrode inserting detection
Auto sample loading detection
Auto reaction time countdown
Operating Condition:
10 to 40°C, below 85% R.H. (noncondensing)
Meter Storage/Transportation Condition:
-20 to 60°C, below 95% R.H.
Shelf Life:
3 years
Measurement Units: Either in mg/dL or mmol/L
Measurement Range:
20 to 600mg/dL (1.1 to 33.3mmol/L)
The device has been tested to meet the electrical and safety requirements of:
IEC 60601-1, EN 60601-1, IEC 61010-1, EN 61010-1, EN 61010-2-101, EN
60601-1-2, EN 61326.
52
Loading...