UPRtek MK350N PREMIUM User Manual

Page 1
uSpectrum PC Software
Download File
www.uprtek.com Membership Login Support Download Center
uFlicker PC Software
Download File
www.uprtek.com Membership Login Support Download Center
MK350N Premium
Spectrometer
Original / Authorized Agent Stamp
Product Serial Number
Purchase Date
52-72-00015-0000
TEL:+886-37-580-885
FAX:+886-37-580-398
Website:www.uprtek.com
Address:No.38, Keyi St., Zhunan Township, Miaoli County 35059, Taiwan, R.O.C
User Manual
Page 2
Contents
1
Introduction
1.1 Packing Contents
1.2 Appearance Introduction
1.3 Annual Product Calibration
1.4 Product Notes and Precautions
2
Preparations before Use
2.1 Preparing Before Use
2.2 Basic Measurement
2.3 System Reset
3
System Overview
3.1 Main Page
3.2 Measurement Modes Introduction
3.3 Measurement Introduction
3.4 Measurement Setting Introduction
3.5 Continuous Measurement Introduction
3.6 Customizing BASIC Mode List
3.7 Instruction of Spectrum Mode
3.8 Switch Between CIE1931 & CIE1976
3.9 TM-30 Mode Introduction
3.10 BROWSER Mode Introduction
3.11 FLICKER Mode Introduction
3.12 FREQUENCY Mode Introduction
3.13 Compare Mode Introduction
3.14 Transferring Data with PC Connection
3.15 Connecting with Mobile APP
3.16 Connecting with uSpectrum
3.17 Connecting with uFlicker
......................................................... 04
............................................. 05
............................... 06
............................ 07
................................ 08
...................................... 09
........................................ 13
.................................................... 14
................................................. 18
......................................................... 19
.............................. 22
................................ 30
........................... 32
................................ 39
...................... 07
.
.................. 21
................. 23
......... 26
.......................... 27
........................ 28
............... 29
........................ 31
.................... 34
......................... 35
.......... 36
.......................... 37
.......................... 38
4
Specification
4.1 Product Specification
4.2 Product Appearance and Dimension
5
Appendices
5.1 Product Warranty Terms
5.2 Measurem
5.3 Q&A
To get more information related to operation, firmware upgrade and warranty terms; online application for correction and Repair service, please visit www.uprtek.com to download the complete version.
......................................................... 40
........................................ 41
........................................................... 46
.................................. 47
....................................................... 50
.................................................................. 59
............... 43
- 1 - - 2 -
Page 3
Introduction
1.1 Packing Contents
1.2 Appearance Introduction
1.3 Annual Product Calibration
1.4 Product Notes and Precautions
- 3 - - 4 -
1
Page 4
1.1 Packing Contents 1.2 Appearance Introduction
Please check MK350N Premium packing before use, contact your agent if there are missing items.
MK350N Premium
Spectrometer
Screen wiper
SD card
Protection Bag Case
Strap Neck Strap
USB Cable
( It is used to connect with PC software.)
Battery
Stand Bracket
Power Adaptor
3.5” Touch Screen
Battery
Power Jack
Mini USB
Strap Hole
SD Card Slot
Reset Key
Power Key
Sensor Cover
Capture Key
MK350N Premium
User Manual
User manual
Warranty card
- 5 -
WiFi Wing
wireless remote control card
( It is used to connect with APP.)
Battery Cover
If the system crashes, please press the power key for 3 seconds to turn off the system. If the problem still can’t be fixed, please use a pin to press the reset key to turn off the system.
- 6 -
Page 5
1.3 Annual Product Calibration
As the product is a high-precision instrument, please use it cautiously. To ensure the accuracy of measurements, annual calibration is recommended. Please consult the agent or the customer service department for the calibration service.
1.4
Product Notes and Precautions
MK350N Premium Spectrometer is a high-precision instrument. Please
1.
unpack with care. Any vibration or collision may cause instrument damage. If the product doesn’t work normally or needs repair, please don’t attempt any repairs. All repairs must be performed by the authorized customer service agent.
Most LCD screens have a very small and inconsequential defective pixel
2.
rate (usually less than 0.1%). This results in occasional pinpoints of white or other colors but will not affect the accuracy of measurements.
Precautions / Warnings
Please read the following precautions to avoid fire, excessive heat, chemical leakage and explosion :
◆ Do not disassemble or modify the battery.
◆ Do not expose the battery to heat (fire) or water/moisture.
◆When disposing used/old batteries, wrap with insulation tape to shield the battery from
electrical contact with metallic objects, which might ignite a fire or explosion.
◆If the unit is plugged into the power adapter and the battery seems to be overheating, or if
there is smoke or peculiar odors emanating from the unit, unplug immediately to avoid the
possibility of fire.
◆However, do not touch the cables if there is heat emanating from near the cables as melted
or deformed cables could expose wiring and result in burns or electric shock.
◆Do not use cloth or anything to wrap or cover the equipment while charging – this could
cause the unit to overheat, melting the casing or causing fire.
◆If the unit is accidently immersed in water, or if moisture has seeped inside, or metal objects
have dropped into the casing, immediately remove the battery to avoid fire or electric shock.
◆Do not operate or store the battery in high-temperature environments-it will cause battery
leakage and/or shorten the life of the battery .
◆Do not use paint thinner, benzene or other organic solvents to clean the equipment – this
may damage the exterior finish or touch screen, and may even ignite fire.
- 7 - - 8 -
Preparations before Use
2
2.1 Preparing Before Use
2.2 Basic Measurement
2.3 System Reset
Page 6
2.1 Preparing Before Use
2.1 使用前準備
2.1 Preparing Before Use
Battery installation:
Step1. Hold MK350N Premium, then press the battery cover.
Step2. Press the battery cover down to remove it.
Step3. Install the battery after removing battery cover.
1
2
※ Press the battery cover down then you can remove the
cover smoothly from MK350N Premium.
1. Please charge the battery for 6 hours at first time usage.
2. To avoid any interruption while measuring, please check battery status before use.
3. Please purchase a new battery if it gets low easily after being fully charged, which means battery life comes to an end.
4. Usage time will depends on battery life, MK350N Premium battery can operate 5 hours after being fully charged.
Battery charging:
1. Plug power charging cable into power jack to start charging.
2. Power Off Status: Users can check power key, it will light in red while
charging and light off after being fully charged.
3. Power On Status: Users can check MK350N Premium screen, it will appear
a lightning symbol to indicate charging, and lightning symbol will
disappear after being fully charged.
3
※ The power key will light
2
up in red, and light off after fully charged.
1
3
※ Battery capacity status
while charging.
※ Lightning symbol will appear
when charging and disappear after being fully charged.
- 9 - - 10 -
Page 7
2.1 使用前準備 2.1 使用前準備
2.1 使用前準備
2.1 Preparing Before Use2.1 Preparing Before Use
Installing the SD Card:
※ Insert the memory card
according to the direction.
Remove the SD Card:
1
※ Press to remove the SD card
Strap installation:
Step1. Unscrew the strap.
Step2. Tie the strap to the strap hole of MK350N Premium.
Step3. Tie the strap to the sensor cover.
Step4. Screw the strap.
Power on
Press the Power button for 1 second and then release it.
※ When the power is on, the Power
button will turn on the green indicator.
2
When the power is off, the green indicator will be off.
Dark calibration
When the system is powered on and the power key is in green, the screen will show the Dark Calibration.
Do you want to do dark calibration?” Please select “Yes” .
When the screen shows “Dark calibration completed”,
3
please press “OK” to enter the main page.
After the sensor cover is put on, please select “OK”.
1
2
- 11 - - 12 -
4
It is recommended performing dark calibration every time
when it is powered on.
Page 8
2.2 Basic Measurement
2.1 使用前準備 2.1 使用前準備
2.1 Preparing Before Use
2.1 使用前準備
Set date and time
After the dark calibration is completed, please set date and time before measurement.
Go to the next page.
Select “Option
”.
Select “Date” and
”.
Time
4 5
After the setting is finished, please press “Yes”.
After the setting is finished, please press “Yes”.
Measure
Select “Basic” to enter measurement page.
Point the sensor head at the target light source.
Please press the power key for 3 seconds to turn off the system.
- 13- - 14 -
Press Capture key at appropriate distance.
Beep sound
while measuring
The measurement data will be shown on the screen after beep sound.
Page 9
2.2 Basic Measurement
System Reset
2.3
Save measurement data
Press Save key on the lower right of the screen, and select “Yes” key.
EXCEL file name:
ESPDYYYY_MMDD_HHMMSS
Fixed (Year) (Month) (Date) (Time)
.xls
▲Can save the original data completely.
The measurement data are saved in the memory card successfully. Please record the file name if necessary.
JPG file name:
IMGYYYY_MMDD_HHMMSS Fixed (Year) (Month) (Date) (Time)
.JPG
Users can press the reset key to turn off the system if system crashes.
Reset Key
※ Use a pen to press the reset key
to turn off the system.
- 15- - 16 -
Page 10
System Overview
3
3.1 Main Page
3.2 Measurement Modes Introduction
3.3 Measurement Introduction
3.4 Measurement Setting Introduction
3.5 Continuous Measurement Introduction
3.6 Customizing BASIC Mode List
3.7 Instruction of Spectrum Mode
3.8 Switch Between CIE1931 & CIE1976
3.9 TM-30 Mode Introduction
3.10 BROWSER Mode Introduction
- 17- - 18 -
3.11 FLICKER Mode Introduction
3.12 FREQUENCY Mode Introduction
3.13 Compare Mode Introduction
3.14 Transferring Data with PC Connection
3.15 Connecting with Mobile APP
3.16 Connecting with uSpectrum
3.17 Connecting with uFlicker
Page 11
Main Page Main Page
3.1 3.1
The interface of the product is touch screen operation. By touching the
screen with your figure, you may select the target menu or option.
The sub menus of each measurement mode are as shown below.
BASIC
SPECTRUM
TM-30
Next page
The interface of the product is touch screen operation. By touching the
screen with your figure, you may select the target menu or option.
The sub menus of each measurement mode are as shown below.
FLICKER
FREQUENCY
Back
page
CIE 1931
CRI
COMPARE
- 19- - 20 -
Page 12
3.2 3.3
Measurement Modes Introduction
Measurement Introduction
Interface introduction:
Means button
1 2 3 4
8
5 6 7
1. Measuring Mode - Basic
2. Exposure Mode -Automatic /Manual
3. Measuring Function –Single operation
/continuous operation . For continuous measurement,
please refer to 3.5.
4. Battery Power
5. Measurement Setting - please refer to 3.4.
6. Back to Main Page
7. Store the measured data- please refer to 3.3.
8. Show measured data – It shows different data based on the
measuring mode. You may make selections by yourself.
- 21 - - 22 -
Save measured data:
Press save button to save file
Please insert SD card
Confirm
Success Failed Failed
Please change SD card
Page 13
3.4 3.4
Measurement Setting Introduction Measurement Setting Introduction
Please go to measurement page to do measurement settings.
Save format
Keyboard
Measurement page
Capture function
Operation sound
Sampling time
Flicker resolution
R
Frequency domain (start)
One time
2.0Hz
Integration mode
Integration time
Long time exposure
Previous
page
measurement page
Frequency domain (end)
Off
4K
0Hz
Dark calibration
Auto mode
Back to
JPG
On
Off
Next page
300Hz
Long time exposure Setting:
Long time exposure
Capture Function Setting:
On
Off
Yes No
Sampling time Setting: Flicker resolution Setting:
Sampling time
4K
Yes No
Flicker resolution
R
2.0Hz
Yes No
Integration Time Setting:Integration Mode Setting:Save Format Setting:
Operation Sound Setting:
+ / -:Adjust key Yes / No:End of the setting
- 23 - - 24 -
Page 14
3.4 Measurement Setting Introduction 3.5
Dark calibration Setting:Frequency domain(start) : Frequency domain(end) :
Continuous Measurement Introduction
7 8 9
4 5 6
1 2 3
0 BS
Keyboard Setting:
Keyboard Instruction:
OK
0 1 2 3 4
5 6 7 8 9
A B C D E F
G H I J K L
M N O P Q R
S T
U V W X
Y
Z
DEL OK
Delete the input English
letter or number.
7 8 9
4 5 6
1 2 3
0 BS
Keypad ON: The keypad interface will show up when saving file, which allows you to rename the file.
Keypad OFF: It will skip the keypad interface when saving files, and name the file automatically.
OK
Key in the number or
English letter.
Select it to show the
input value on the
upper side.
Basic mode
(start measurement)
press capture key
MK350N Premium will process
Step1. Select the continuous capture
Step2. Press the capture key
Step3. Processing the continuous measurement
Step4. Press capture key again to stop continuous measurement
Confirm the input value and save it,
then return to the interface.
1. Continuous measurement has no beep sound after press capture key.
2. Users cannot save the measurement data while processing continuous measurement.
3. Adjust the integration time is only allowed in the manual mode.
- 25 - - 26 -
Select Continuous
※ The capture symbol
is rotating when MK350N Premium is processing continuous measurement.
measurement with
interval of 3 seconds.
Basic mode
(after measurement)
Press the capture key again to
see the measurement results.
Page 15
Customizing BASIC Mode List
3.6 3.7 Instruction of Spectrum Mode
The 4 items on the Basic list can be customized with different
units of measure (e.g. CQS, fc, R9, TLCI).
Interface introduction:
CRI CCT LUX
Means button
1 2 3 4
Rg
Previous
Y
For more information about Measurement Index, please refer to 5.2.
Back to
page
BASIC mode
Next
page
Step1. Tap the box of measuring unit
Step2. Select an item from the list
Step3. The new data item will be displayed
5 6 7
1. Measuring Mode - SPECTRUM
2. Exposure Mode -Automatic /Manual
3. Measuring Function –Single operation
/continuous operation . For continuous measurement,
please refer to 3.5.
4. Battery Power
5. Measurement Setting - please refer to 3.4.
6. Back to Main Page
7. Store the measured data- please refer to 3.3.
- 27 - - 28 -
Page 16
Switch Between CIE1931&CIE1976
3.8 3.9
Touch the graph to switch the table between CIE1931 and CIE1976.
TM-30 Mode Introduction
TM-30-15 provides equations and direction for calculating Fidelity Index (Rf) and gamut (increase or decrease in chroma) through a Gamut Index (Rg), including the spectral reflectance functions for 99 Color Evaluation Samples (CES).
Rf:Fidelity Index
Rf takes a maximum value of 100. If Rf is 100, it is a match to the reference source (true to the “real color”, similar to CRI Ra 100). If color shifts occur, the arrows get longer and Rf decreases (to a minimum value of zero).
Rg:Gamut Index
Rg=100 means that the test light source doesn’t change the hue or saturation of the CES compared to the reference source.
Rg>100 indicates that the test
light source increases the saturation of the CES producing colors that are more vivid.
Rg<100 indicates that the test
light source decreases the saturation of the CES producing colors that are less saturated.
red line:test source
gray line:reference source
Color Vector Graphic:one of the outputs of TM-30.
This graph can tell us how, on average, various colors are distorted.
1. outward arrows mean more saturated colors
2. inward arrows mean duller colors
3. sideways arrows mean a hue shift
4. no arrow means that the color isn’t distorted
- 29 - - 30 -
Page 17
BROWSER Mode Introduction
3.10 3.11 FLICKER Mode Introduction
The Browser (on Home Screen) allows you to review historical
data that was previously saved to the SD card.
Interface introduction:
Means button
1 2 3 4
Previous
Back to
page
main manu
CCT
CRI
Return to the
previous folder
Next page
LUX
Back to
file list
Step1. Press the “BROWSER”icon
Step2. A file browser will show the files on the SD card. If you select an excel file, a review menu will be displayed.
Step3. Press any of the icons to review the data.
Step4. Displaying the data of excel file .
- 31 - - 32 -
5 6 7
1. Measuring Mode - FLICKER
2. Exposure Mode -Automatic /Manual
3. Measuring Function –Single operation
/continuous operation . For continuous measurement,
please refer to 3.5.
4. Battery Power
5. Measurement Setting - please refer to 3.4.
6. Back to Main Page
7. Store the measured data- please refer to 3.3.
Page 18
3.12 FREQUENCY Mode Introduction 3.12 FREQUENCY Mode Introduction
Touch the graph to switch the table between Frequency.
Interface introduction:
Means button
1 2 3 4
5 6 7
1. Measuring Mode - FREQUENCY
2. Exposure Mode -Automatic /Manual
3. Measuring Function –Single operation
/continuous operation . For continuous measurement,
please refer to 3.5.
4. Battery Power
5. Measurement Setting - please refer to 3.4.
6. Back to Main Page
7. Store the measured data- please refer to 3.3.
More about FFT: Refer to Page 37~ Page 40 of Flicker Handbook.
More about Light Wave: Refer to Page 25~ Page 28 of Flicker Handbook.
- 33 - - 34 -
Page 19
3.13 Compare Mode Introduction
與手機APP連線
Transferring Data with PC Connection
3.14
Users can use the real-time measurement or select
the historical measurement data to show the measured
light source data with parallel display. We offer four
modes that allow you to compare the measurement of
the source data: BASIC, SPECTRUM, CIE 1931 and CIE 1976.
CIE 1931
Select the "light source" mode you need.
After the measurement is completed, the measured light source data are compared by parallel display.
Storage the measurement data:
Inside the package has a USB cable, connect the mini USB to MK350N Premium
and plug the USB to your PC then you can read the data saved in SD
After connected USB cable, MK350N Premium will display as below.
card.
Insert SD card
- 36 -- 35 -
Page 20
Connecting with Mobile APP
3.15
與手機APP連線
Connecting with uSpectrum
3.16
WiFi Wing wireless remote control card
Insert WiFi Wing card to the system. At IOS or Android platform, download MK350N Premium APP and install it on your mobile. Then you can connect it with system for the measurement of wireless control.
WiFi Wing
wireless remote control card
( It is used to connect with APP.)
※For more operation, please refer to Wing user manual.
Please visit UPRtek official site, http://www.uprtek.com
Install APP
Download and install the MK350N Premium APP on your Mobile.
Select WiFi UPRtek_Wing.
Measure
→Support→Download Center.
MK350N Premium
UPRtek_Wing
Install uSpectrum PC software
Please visit UPRtek official site, http://www.uprtek.com→Support→
Download Center, Download and install uSpectrum software on the PC. Then you can connect it with the system for USB control of measurement.
uSpectrum
Install file
Set USB PC connection mode
Select “PC connection
Select “Option
”.
Select “USB mode”.
and press Yes.
Measure
USB connectionWiFi connection
uSpectrum PC software
- 38 -- 37 -
Page 21
Connecting with uFlicker
3.17
Install uFlicker PC software
Please visit UPRtek official site, http://www.uprtek.com→Support→
Download Center, Download and install uFlicker software on the PC. Then you can connect it with the system for USB control of measurement.
uFlicker
Install file
Set USB PC connection mode
Select “Option
Measure
”.
Select “USB mode”.
USB connection
- 39 -
Select “PC connection and press Yes.
uFlicker PC software
Specification
4
4.1 Product Specification
4.2 Product Appearance and Dimension
- 40 -
Page 22
Spectrum
Sensor
CMOS Linear Image Sensor
Wavelength Range
380 to 780 nm
Wavelength Data Increment
1 nm
Spectral Bandwidth
Approximately 12 nm (Half Bandwidth)
Wavelength Reproducibility
± 1 nm (Input source must be a stable light source.)
Measurement Range
5 to 100,000 lx
Illuminance Accuracy
± 2.5%
0.2% (100 ~ 100,000 lx)
0.5% (5 ~ 100 lx)
x y: ± 0.0025 (5 to 100 lx)
Color Repeatability
± 0.0005 in CIE 1931 x,y
CCT Accuracy
± 2%
CRI Accuracy @ Ra
± 1.5%
Stray Light
-25 dB max. (Input the 550nm monochromatic light and measure the stray light ratio at 550nm ± 40nm.)
Integration Time Range
100us to 1,000 ms
Digital Resolution
16 bits
Flicker
Measurement Range
5 to 100,000 lx
Sampling Rate
100k sample/sec
Frequency Range
5 to 50k Hz
Frequency Resolution
2, 4, 8, 16, 32 Hz
Flicker Accuracy
± 5%
Feature
Capture Function
One time / Continuous
Standalone Mode / WiFi Mode (It can be connected to mobile phones and tablets.) /
USB Mode (MSC Mode (MSC- Mass Storage Class.) + PC connection)
Integration Mode
Auto/Manual
1. Basic Mode
2. Spectrum Mode
3. CRI Mode
4. CIE 1931 / 1976 Chromaticity Mode
5. TM-30-15 Mode
6. Flicker Mode
7. Frequency Mode
8. Browser Mode
9. Compare Mode
1. Illuminance (LUX) / Foot Candle (fc)
2. Correlated Color Temperature (CCT)
3. CIE Chromaticity Coordinates
(3) CIE 1931 XYZ Value
4. x , y , u' , v'
5. Delta uv (Duv)
6. Dominant Wavelength (λd)
7. Excitation Purity
8. Color Rendering Index (CRI, Ra) / R1 to R15
9. Color Quality Scale (CQS)
10. Television Lighting Consistency Index (TLCI)
11. TM-30-15 (Rf, Rg, Color Vector Graphic)
12. Flicker Frequency
13. Percent Flicker
14. Flicker Index
15. Stroboscopic Effect Visibility Measure (SVM)
16. Spectral Power Distribution (SPD) mW/m2
17. Peak Wavelength (λp)
18. Peak Wavelength Value (λpV)
19. Intergration Time (I-Time)
20. Scotopic and Photopic Ratio (S/P)
System Configurations
Display
3.5" 320X240 Resistive Touch LCD
Max. Files
68,000 Files @ 8GB SD Card (Excel + JPG)
Battery Operation Time
5 hours / Fully Charged
Power
Adapter; 2500 mAh (3.7V Rechargeable Li-ion Battery)
SD Card (SD2.0,SDHC/up to 32G) /
WiFi SD Card compatible with iOS and Android
Data Format
Compatible Excel / JPG
Dimensions
147.5 x 78 x 24 mm (H x W x D)
Storage Temperature /
Humidity
-10 to 40 ,
relative humidity 70% or less without condensation
English / Traditional Chinese / Simplified Chinese / Japanese / Spanish / German / French
/ Italian / Russian
Cosine Correction
Product Specification Product Specification
4.1 4.1
(1) CIE 1931 x,y Coordinates
(2) CIE 1976 u',v' Coordinates
Illuminance Repeatability (2σ)
Color Accuracy
Color Repeatability (2σ)
Operation Mode
Measuring Modes
Illuminant A @ 2,856 K at 20,000 lx(Temperature 23±2
and relative humidity 50% or
less.)
x y: ± 0.002 (100 to 100,000 lx)
x y: 0.0002 (500 to 100,000 lx)
x y: 0.0004 (30 to 500 lx)
x y: 0.001 (5 to 30 lx)
- 41 - - 42 -
Measuring Capabilities
Data Output Interface
Weight (with Battery) 225 g ± 10 g
Operating Temperature /
Humidity
Display languages
The company reserves the right to change product specifications at any time without prior notice.
Mini USB Port (USB 2.0) /
0 to 35 ,
relative humidity 70% or less without condensation
Page 23
Product Appearance and Dimension Product Appearance and Dimension
4.2 4.2
Back Support Installation Diagram
Unit: mm Unit: mm
- 43 - - 44 -
Page 24
Appendices
5.1 Product Warranty Terms
5.2 Measurement Indexes
5.3 Q&A
- 45 - - 46 -
5
Page 25
5.1 Product Warranty Terms 5.1 Product Warranty Terms
Warranty Principles
UPRTEK provides the service of repairing or changing equivalent products for the customer in case of the material or functional defect and fault during the product warranty period.
1.Changing new product:
If the consumer finds any functional defect or fault or finds any part missing within 7 days after purchasing the product, it should consult the original seller and report to the vendor immediately. Moreover, it should leave the contact information, so the customer service personnel of the original vendor will check it out and generate the number of changing new product for the customer. In this case, the original vendor provides the service of “changing new product” within 7 days.
The customer should return the product to the original vendor within 30 days after getting the repair number. As for the international customer, it permits the flexibility of logistical time and expands the time for returning the product to the original vendor.
Remarks: It requires complete package when returning the product to the original vendor. There should be no part missing or scratch on the surface unless the part is found missing when the product is delivered. Under this circum­stance, the original vendor reserves the ultimate right to determine whether to change new product.
2. Repair service:
If it exceeds the period for changing new product (seven days), the functional non-compliance or defect found on the product should go through RMA workflow. When any product needs to be returned to the original vendor for repair, it should consult the business service personnel via Email, fax or phone call to apply for repair number before returning the product to the original vendor for repair service.
After the original vendor receives the returned product, the internal engineer of the Company will check it initially and confirms the causes for product defect. If it is within the warranty period and it is the functional fault of the product, it should follow the general repair workflow. However, if the engineer checks it is human damage rather than functional fault, it is inapplicable to the warranty term.
Remarks: To avoid damage during the product transporta­tion, we strongly recommend choosing international express service and protecting the product carefully.
Limited Warranty
The warranty term is inapplicable to the damage caused by unnatural or external factors, such as the following circumstances:
1.When the fault is caused by the natural disaster and improper human operation rather than the product itself;
2.When the product is repaired or disassembled by others rather than the technician authorized by the company;
3.When the warranty volume label or disassembly-proof volume label is modified, damaged or gone;
4.When the product serial number is wrong, damaged or unclear.
Exemption from Liabilities
UPRTEK is not liable to the product defect or damage caused by any factor during the transportation of sending MK350N Premium series product for repair. It is recom­mended taking out the storage device, packing and transporting the product properly on your own before sending the product for repair.
UPRTEK is not liable to the compensation for the operation loss, expected cost loss, data loss caused by or related to the usage of the product of the Company, as well as any other indirect, accident or derivative loss or damage provided it is permitted by the laws.
- 47 - - 48 -
Page 26
5.1 5.1
Product Warranty Terms Product Warranty Terms
Applicable Subjects for Warranty Terms Service after Warranty Period
UPRTEK’s warranty service terms are only applicable to the consumers who purchase the company products through formal or legal sales channels.
Warranty Period
1.Host of MK350N Premium series product: UPRTEK provides two years of warranty service for all MK350N Premium series products, with free calibration service for one time.
2.Product supplies and related accessories: The product suppliers and the related accessories such as battery, tripod and neck strap, are excluded from the warranty service.
Services Provided by Authorized Agent or Distributor
The qualifies agent or distributor may receive or deliver the host of MK350N Premium series product for basic inspec­tion, so as to confirm whether the device needs to be returned to the original vendor for further calibration service.
URL of qualified agents or distributors: www.uprtek.com.
The basic inspection service provided for the consumer is charged based on the announcement made by UPRTEK.
Other Notes
UPRTEK doesn’t produce or manufacture all materials and parts of the product. If the target material and part is discontinued within the warranty period, UPRTEK is entitled to replace it with equivalent alternative to finish repair.
If any damage or fault occurs to MK350N Premium series product after two years of warranty period, the consumer may still return the product to UPRTEK for repair. However, the repair service should be charged based on the product fault or damage condition. If the engineer finds the following situations, it is suggested purchasing new product:
The MK350N Premium series product or part is not available anymore;
The functions of the device can’t be recovered due to the damage caused by water, strong collision, serious contamination or corrosion.
The product is deformed due to falling or strong collision, and the functions can’t be recovered even if the main parts are replaced.
The product is aging or used in adverse environment, so many parts don’t work normally. In this case, it has to be changed.
The part is not available even if it is it the warranty period.
Sending for Repair
Consumer may send the product back to the original vendor RMA repair service through either of the following two methods:
The consumer sends the product to UPRTEK’s agent or distributor which will return it to the original vendor for repair.
The consumer sends the product to UPRTEK for repair directly.
- 49 - - 50 -
Page 27
5.2 Measurement Indexes 5.2 Measurement Indexes
Abbreviation Full Name Unit Description
CCT Corelated
Color
Temperature
CRI (Ra) Color
Rendering
Index
R1、R2
….R15
Rendering
Index
(kelvin)
The color temperature is
K
the color radiated by a black-body radiator under different temperatures. CCI has the color that is the closest to the ideal black-body radiator.
As defined by CIE, R1~R8
---
represent the value of eight standard colors, while CRI(Ra) is the average value of R1~ R8. The value 100 indicates the best quality of light source, while the value 0 indicates the worst quality of light source.
It represents the quality
---
of light source, with the indexes corresponding to 15 standard colors, including: R1: light grey-red; R2: dark grey-yellow; R3: saturated yellow-green; R4: middle yellow-green; R5: light
next page
LUX Illuminance lx
λp
(LambdaP)
λpV
(LambdaPV)
λd
(LambdaD)
Peak wavelength
Peak wavelength value
Dominant wavelength
I-Time Integration
time
nm
mW/m
nm
ms
yellow-green; R6: light blue; R7: light purple-blue; R8: light red-purple; R9: saturated red; R10: saturated yellow; R11: saturated yellow; R12: saturated yello; R13: white skin color; R14: Leaf green; and R15: yellow skin color.
It is the light flux received by each unit area.
It is the wavelength with the highest power in the measured spectrum.
It is the highest power in
2
the measured spectrum, in the unit: mW/m2.
The dominant wavelength is used to express the color of the measured light. It could be hybrided by the spectrum color of the wavelength and the standard illuminant E(x,y =
0.333, 0.333).
The integration time measured by the spectrometer.
- 51 - - 52 -
Page 28
5.2 Measurement Indexes 5.2 Measurement Indexes
Abbreviation Full Name Unit Description
x
y
X
CIE1931
color
coordinate
The color coordinate was built by CIE in 1931.
---
Y
Z
u'
v'
CIE1976
color
The color coordinate was built by CIE in 1976.
---
coordinate
Duv
(Δuv)
Δx
(deltax)
(dx)
CIE1960 uv
color
coordinate
difference
CIE1931 x
color
coordinate
difference
It is the uv distance between CIE1960 coordinate and the Planck’s blackbody radiation with the same color temperature. The value close to 0 indicates the color temperature and color are closer to that of the
---
blackbody radiation. The positive value indicates it is above the blackbody radiation, while the negative value indicates it is below the blackbody radiation.
It is the x difference between CIE1931
---
coordinate and the Planck’s blackbody radiation with the same color temperature.
Δy
(deltay)
(dy)
Δu'
(deltau)
(du)
Δv'
(deltav)
(dv)
CIE1931 y color coordinate difference
CIE1976 u' color coordinate difference
CIE1976 v' color coordinate difference
fc fcFootcandle
Purity %Color purity
PPFD
(PPF)
Photosynthetic
Photon Flux
Density
It is the y difference between CIE1931 coordinate and the Planck’s blackbody radiation
---
with the same color temperature.
It is the u' difference between CIE1976 coordinate and the Planck’s blackbody radiation
---
with the same color temperature.
It is the v' difference between CIE1976 coordinate and the Planck’s blackbody radiation
---
with the same color temperature.
It is based on the unit of
2
lm/ft .
It is the percent of the dominant wavelength in the standard illuminant. The closer the color purity is to 100%, the closer it is to the dominant wavelength.
It is the Photosynthetic
umol/
Photon Flux Density defined in 400~700nm.
2
(sec m )
next page
- 53 - - 54 -
Page 29
5.2 5.2
Measurement Indexes Measurement Indexes
S/P
(SP-ratio)
GAI Gamut
Scotopic
vision/
photopic
vision ratio
Area Index
The sensitivity of human eyes
varies with the light of
different wavelength. It can
get the data such as
illuminance based on the
brightness function. However,
the brightness function is
different for human eyes
under high brightness and
low brightness. Therefore, it
---
defines the ratio of photopic
vision under high brightness
and scotopic vision under low
brightness, which is called S/P
ratio.
The Gamut Area Index is the
ratio between the area of the
eight standard colors (R1~R8)
on CIE1976 coordinate
measured by CRI and the
area measured with the
---
standard illuminant E(x,y =
0.333, 0.333). The higher the
value is, the more saturated
and bright the color is.
TLCI
W/m
BIN
SDCM
2
Television
Lighting
Consistency
Index
Irradiance
---
Standard
Deviation
Color
Matching
The parameter is defined by
referring to European
Broadcasting Union (EBU),
which is used to evaluate the
---
light source under for
television lighting.
It is the irradiance within the
2
W/m
range of the wavelength
specified in the specification.
To classify the chromaticity of
LED, ENERGY STAR combines
with ANSI C78.377 standard
to classify various classes and
make uniform BIN types. So
---
the user may distinguish the
chromaticity of LED products
easily.
SDCM has the same meaning
with that of Macadam Ellipse.
The color within Macadam
Ellipse (SDCM=1) drawn on
the coordinate can’t be
---
identified by human eyes.
- 55 - - 56 -
Page 30
5.2 5.2
Measurement Indexes Measurement Indexes
TRANSMIT Transmittance
Rf TM30 Rf
Rg TM30 Rg
The transmittance is defined
as (the radiant flux
transmitted/the radiant flux
received)x100 %, based on
%
which the wavelength
response can be obtained.
It is a rating method proposed
by IES for the performance of
light source. Different from 8
standard colors, it defines 99
standard colors to present the
performance of various
wavelengths. Rf represents the
similarity between the target
light source and the reference
light source under the standard
colors. The value 100 represents
it is identical, while the value 0
represents great difference. Rg
represents the saturation
---
difference between the target
light source and the reference
light source under the standard
colors. The value 100 represents
the saturation is the same, while
the value larger than 100
represents saturation is higher,
and the value smaller than 100
represents saturation is lower.
Moreover, it could identify the
difference more obviously
through the color vector
diagram and the color tolerance
diagram.
Findex
(Flicker Idx)
Fpercent
(Flicker Pct)
SVM
Freq
(Frequency)
Flicker index
Flicker
percent
Stroboscopic Effect visibility Measure
Flicker
Frequency
Flicker CycleTwave
MiredMired
Flicker index= Area above the
mean value /area below the
mean value. Higher value
represents obvious flicker.
---
Flicker percent=(Maximum
value-minimum value)/
( Maximum value + minimum
%
value). Higher value
represents obvious flicker.
SVM is an index used to
quantify the flicker effect.
---
Higher value represents
obvious flicker effect.
It is the flicker rate of the
measured light, or the
Hz
frequency shown per second.
It is the countdown of the
flicker frequecny of the
ms
measured light, or the time of
one flicker.
Mired = 1000000/CCT, is the
deviation unit of the color
-1
MK
temperature of the filter.
- 57 - - 58 -
Page 31
Q&A
5.3
The following situations are not fault. Please double check it before
requesting for repair. If it doesn’t work normally after the inspection,
it could be caused by fault of the device. In this case, please take out
the battery and contact the seller or the vendor for repair.
State Inspection item Solution
Power-on failure
(No display)
Check whether the
battery is installed
correctly.
Check whether the
battery runs out of power.
Check whether the
battery connector is dirty.
Please don’t install the battery
inversely, or push it in violently,
which may destroy the battery
spring.
Please charge the battery for 6
hours for the first use. (Please refer
to 2.1)
Please wipe it with dry cloth.
Item
Why I feel the card is stuck when inserting and unplugging it?
What is the function of dark calibration?
Do I need to perform dark calibration at every startup.
Why my SD card can’t be used on
MK350N Premium?
- 59 - - 60 -
Solution
It is the anti-shedding function designed specially. It prevents SD card shedding in case of collision under external force, which may cause write-in error.
The dark calibration is also known as zero calibration. As the temperature change of the environment affects the measurement results, the user is suggested performing dark calibration before measurement, so as to improve the measurement accuracy. (As for the calibration instruction, please refer to 2.6).
The SD card should be formatted into FAT before being inserted into MK350N Premium. Moreover, the memory of SD card is required to be 1G or higher.
To get more information related to operation, firmware upgrade and warranty terms; online application for correction and Repair service, please visit www.uprtek.com to download the complete version.