Universal Flow Monitors FlowStream OFS User Manual

User Manual
Series: OFS
Firmware Revision 5.79
UNIVERSAL FLOW MONITORS, INC.
1755 East Nine Mile Road
PO Box 249
Hazel Park, MI 48030-0249
TEL (248) 542-9635 FAX (248) 398-4274
http://www.flowmeters.com
FSMAN.6 081214 Page 1
TABLE OF CONTENTS
PROPRIETARY NOTICE ............................................................................................................. 3
NAMEPLATE EXAMPLE ............................................................................................................ 4
GENERAL SPECIFICATIONS .................................................................................................... 5
ELECTRICAL SPECIFICATIONS ............................................................................................... 6
OPERATION ................................................................................................................................. 7
Figure 1. Laminar Flow Illustration ........................................................................................... 7
APPLICATIONS ........................................................................................................................... 7
Using FlowStream at Varying Temperatures ............................................................................. 7
Using FlowStream with Different Gases ................................................................................... 8
Reference Conditions for Mass Flow Measurement .................................................................. 8
WIRING DIAGRAM ..................................................................................................................... 9
Wiring for 4-20mA Transmitters ............................................................................................... 9
Figure 2. Standard Transmitter Wiring ...................................................................................... 9
Figure 3. Required Supply Voltage vs. Loop Resistance ......................................................... 10
Hazardous Environment Wiring............................................................................................... 10
Figure 4. Intrinsically Safe Installation ................................................................................... 11
Wiring for Voltage Output Models .......................................................................................... 12
Figure 5. Voltage Output Wiring ............................................................................................. 12
Wiring for Frequency or Pulse Output Models ........................................................................ 13
Figure 6. Frequency and Pulse Output Wiring ......................................................................... 13
POWER-UP ................................................................................................................................. 14
LCD READOUT .......................................................................................................................... 14
Flow Rate ................................................................................................................................. 15
Totalizer ................................................................................................................................... 15
Pressure .................................................................................................................................... 17
Temperature ............................................................................................................................. 19
USER MENU ............................................................................................................................... 20
Select Response Time .............................................................................................................. 22
Sampling Delay ........................................................................................................................ 23
Signal Averaging ...................................................................................................................... 24
Set High Flow Alarm ............................................................................................................... 26
Set Low Flow Alarm ................................................................................................................ 28
Tare (Re-Zero) the Flowmeter ................................................................................................. 29
Scaling the Output Span ........................................................................................................... 31
Selecting the Gas ...................................................................................................................... 32
Totalizer ON/OFF .................................................................................................................... 33
Pulse Output Setting ................................................................................................................. 34
Checking Battery Voltage ........................................................................................................ 35
SPECIAL FEATURES ................................................................................................................ 36
Factory Reset ............................................................................................................................ 36
Low Supply Voltage Indicator ................................................................................................. 37
Diagnostic Mode ...................................................................................................................... 38
HOW TO ORDER A SINGLE-GAS FLOWMETER ................................................................. 40
DIMENSIONS OF OFS SERIES ................................................................................................ 42
RETURN MATERIAL
AUTHORIZATION
............................................................................... 43
WARRANTY INFORMATION .................................................................................................. 45
FSMAN.6 081214 Page 2
PROPRIETARY NOTICE
The information contained in this publication is derived in part from proprietary and patented data. This information has been prepared for the express purpose of assisting in installation, operation, and maintenance of the instruments described herein. Publication of this information does not convey any rights of use or reproduction other than in connection with the installation, operation and maintenance of the equipment described herein. Universal Flow Monitors, Inc. reserves the right to change the information contained in this publication at any time and without prior notice.
Serial numbers are formatted as YY MM ID 000 YY = year, MM = month, ID = product identifier, 000 through 999 = three-digit sequential number.
FSMAN.6 081214 Page 3
Intrinsically Safe flowmeters must include
“X2A” or “Z2A” in the
Model Code
Firmware revision
Model Code
NAMEPLATE EXAMPLE
The Intrinsically Safe flow transmitters have the following label:
FSMAN.6 081214 Page 4
GENERAL SPECIFICATIONS
Flow Ranges: 500 SCCM full-scale to 1,000 SLPM full-scale 1 SCFH full-scale to 2,100 SCFH full-scale Turndown Ratio: 100:1 Maximum Operating Pressure: 100 PSIG Burst Pressure: 200 PSIG Pressure Effect on Accuracy: Less than 0.03% F.S. / PSI Maximum Operating Temperature: 176 °F (80 ºC) Minimum Operating Temperature: -13 °F (-25 ºC) Temperature Effect on Accuracy: Less than 0.03% F.S./ °F Maximum Pressure Drop: 1.5 PSI at F.S. flow (from inlet port to outlet port) Process Connections: 1/4”-3/8”-1/2”-3/4” NPT female Wetted Parts
Sensors: Glass-filled nylon, alumina-based ceramic, silicon, gold,
epoxy
Flow Body Internals: Anodized aluminum, Viton, stainless steel Enclosure Rating: Type 4 Display: 4-digit LCD digital display, 0.35" high Approvals: CE, CSA, Intrinsic Safety (all classes and divisions) with
proper zener barrier
FSMAN.6 081214 Page 5
ELECTRICAL SPECIFICATIONS
Accuracy (Including Linearity and Repeatability)
Flow: ± 1% of full-scale for flowmeters sized from 15-566 SLPM
(31-1200 SCFH) of Air ± 2% of full-scale for flowmeters sized for lower than 15 SLPM (30 SCFH) of Air ± 2% of full-scale for flowmeters sized for higher than 566 SLPM (1201 SCFH) of Air
Pressure: ± 1 PSI (See Note 1) Temperature: ± 3 °F Totalizer: ± 0.25% of full-scale (in addition to flow accuracy)
Output Signal
Analog: 4-20 mA (2-wire loop powered) 0-5 V, 0-10 V, 1-5 V 0-5 V Bi-directional (2.5 V = 0 flow)
Frequency: 0-1000 Hz, 200-1200 Hz 0-3V signal amplitude
Pulse: 1,250-5,000 pulses/minute, user selectable 0-3V pulse amplitude
2 msec pulse width
Response Time: 5 – 100 msec (step response), user selectable
Alarms: 2 independent open-collector outputs (high/low flow rate)
with corresponding LEDs Open-Collector Rating: 30VDC at 50 mA
Electrical Connection: 4- or 7-conductor shielded pigtail cable Supply Voltage: 10–30 VDC (Standard), 12-24 VDC (Intrinsically Safe)
7.2-9 VDC for battery-operated units (See Note 2) Supply Current: 22 mA @ F.S. flow (includes over-range) for 4-20 mA
loop-powered transmitters
5 mA for voltage, frequency, and pulse outputs
3.5 mA for battery-operated units (See Note 2)
Note 1: Pressure, temperature, and totalizer are only displayed on the LCD. No output signal is available for these parameters. Note 2: Battery-operated units require a standard 9V alkaline battery and will operate for over 100 hours continuously. An On/Off switch allows the user to turn the power off, thus conserving the battery life. These flowmeters have no output signal.
FSMAN.6 081214 Page 6
OPERATION
FlowStream flowmeters accurately measure the mass flow rate of most gases. The flow rate is
determined by measuring the pressure drop across a unique internal restriction, known as Laminar Flow Element (LFE). The restriction is designed such that the gas molecules are forced into moving in parallel paths along the entire length of the passage for the entire range of operation of the device. Unlike other pressure-flow measuring devices, the relationship between pressure drop and flow is linear in laminar flowmeters.
FlowStream mass flowmeters utilize an absolute pressure sensor along with a temperature sensor to compensate for density variations of the gas. When combined with the differential pressure (volumetric flow) output, the mass flow rate of the gas can be determined.
Figure 1. Laminar Flow Illustration
APPLICATIONS
FlowStream flowmeters are designed to work with non-corrosive, non-ionic, clean, dry gases
only. Introduction of liquids to the internal sensors will damage the unit, and the repair is not covered under warranty. Relative humidity of the gas can be as high as 100%, as long as proper installation guarantees that no internal condensation will occur. A 50-micron filter and/or dryer may be required for some applications.
Using FlowStream at Varying Temperatures
Even though FlowStream flowmeters measure true mass flow, rapid variations in ambient and/or gas temperature may affect performance. This is due to the time lag of the internal temperature sensor and the slow heating and cooling of the flowmeter body. It is highly recommended that through proper installation the following two objectives be met:
There be minimal difference between gas temperature and ambient temperature; Rapid temperature variations be avoided.
FSMAN.6 081214 Page 7
The internal temperature sensor is located above the inlet port, very close to where the gas enters the meter. This ensures accurate measurement of the gas temperature. However, because the temperature sensor is embedded inside the flowmeter body, if ambient temperature is different from gas temperature, there would be a discrepancy between what the sensor reads and the true gas temperature. The flowmeter body would track ambient temperature while gas temperature would heat/cool the body at a different rate.
Likewise, if temperature variation is rapid, the flowmeter body may not follow it quickly enough due to the mass of the metal flow chamber, which in turn would result in inaccurate measurement of gas temperature.
For optimal performance, always allow two to four hours from the time the ambient and gas temperatures are stabilized to when the first flowmeter reading is taken.
Using FlowStream with Different Gases
FlowStream flowmeters can easily be used to measure the flow rate of other gases, as long as
the gas compatibility criteria are observed. For example, a flowmeter that is factory-calibrated for air can be used to measure the flow of Argon. (Consult Factory for additional information.)
Reference Conditions for Mass Flow Measurement
Although the correct units for mass are expressed in grams, kilograms, etc., it has become somewhat standard that mass flow rate is specified in SLPM (standard liters per minute), SCFH (standard cubic feet per hour) or other similar units.
This means that the mass flow rate is calculated by normalizing the volumetric flow rate to some standard temperature and pressure (STP). By knowing the gas density at that STP, one can determine the mass flow rate in grams per minute, kilograms per hour, etc. STP is usually specified at sea level conditions; however, no single standard exists for this convention. UFM uses STP of 70° F and 14.7 PSIA.
Note: If used outside the parameters specified in this manual, the proper operation of the flowmeter cannot be guaranteed.
FSMAN.6 081214 Page 8
Wire Color
Function
Red
Supply +
12-30VDC (Standard) 12-24 VDC (Intrinsically Safe)
Black
Supply -
DC GND
Brown
Tare
Short to GND for 5 seconds to tare at zero flow
White
Not Used
Do not connect
Shield
Internally Grounded to Chassis
Blue
High Alarm Output
Open-collector
Orange
Low Alarm Output
Open-collector
Green
Alarm Common
Emitter for both alarms
WIRING DIAGRAM
Wiring for 4-20mA Transmitters
Figure 2 shows a typical 2-wire, 4-20mA transmitter wiring for use with FlowStream
flowmeters. Maximum loop resistance is related to the available supply voltage. Since the flowmeter requires 10 volts to operate, the voltage drop across the loop resistance at full-scale flow output (20mA) must be added in to determine the minimum supply voltage. For example, if a 100-ohm resistor is used to convert the current signal to voltage, the voltage drop across the resistor will be 0.4-2 volts for 4-20 mA, respectively. Minimum required loop voltage in this case is 12 volts, as shown. Likewise, the voltage drop for a 250-ohm resistor is 1-5 volts, requiring a minimum of 15 volts supply. Please refer to the graph in Figure 3.
Figure 2. Standard Transmitter Wiring
(DO NOT Use for Intrinsically Safe Applications)
Note: Shield wire (shown as gray) may be connected to an external chassis ground to improve
electrical noise immunity. However, care must be taken not to connect this ground to signal ground.
FSMAN.6 081214 Page 9
Figure 3. Required Supply Voltage vs. Loop Resistance
Hazardous Environment Wiring
Intrinsically Safe wiring must be installed in accordance with Article 504 of NEC, ANSI NPF 70 and Control Drawing Number 7577, Revision 00 (Figure 4, below). The transmitter approval by the Canadian Standards Association for installation in Hazardous environments is based on installation through an Intrinsic Safety Barrier.
The Transmitter when wired through the I.S. Barrier is suitable for use in:
CLASS I GROUPS A, B, C & D CLASS II GROUPS E, F & G CLASS III HAZARDOUS LOCATIONS
Earth Ground of the I.S. Barrier must be connected to the earth ground of the AC feeder supply. The resistance between Intrinsically Safe ground terminals and A.C. Earth ground must be less than one Ohm. (UFM suggested I.S. Barrier R.Stahl 9001/01-280-075-10, UFM part number
8140).
The Power Supply voltage is limited to 24 VDC Max. The Power Supply Control Unit must not use or be able to generate more than 250 volts. The Maximum Load that can be put on the system is 250 Ohms.
All repairs on the Flow Transmitter should be accomplished at the factory because any substitution of components may impair Intrinsic Safety.
FSMAN.6 081214 Page 10
FSMAN.6 081214 Page 11
Figure 4. Intrinsically Safe Installation
Wire Color
Function
Red
Supply +
15-30VDC
Black
Supply -
DC GND
Brown
Tare
Short to GND for 5 seconds to tare at zero flow
White
Voltage Output
0-5V (or 1-5V) 0-10V (or 2-10V) Bi-directional (0-5V)
Shield
Internally Grounded to Chassis
Blue
High Alarm Output
Open-collector
Orange
Low Alarm Output
Open-collector
Green
Alarm Common
Emitter for both alarms
Wiring for Voltage Output Models
Figure 5 shows the wiring for voltage output models. (Optional wiring is also shown for alarm
outputs.) For 0-5V models, the supply voltage can be 10-30VDC.
Note: Output voltage range must be specified at the time of ordering. They cannot be interchanged by the user.
Figure 5. Voltage Output Wiring
FSMAN.6 081214 Page 12
Wire Color
Function
Red
Supply +
10-30VDC
Black
Supply -
DC GND
Brown
Tare
Short to GND for 5 seconds to tare at zero flow
White
Frequency Output Or Pulse Output
0-3V signal amplitude
Shield
Internally Grounded to Chassis
Blue
High Alarm Output
Open-collector
Orange
Low Alarm Output
Open-collector
Green
Alarm Common
Emitter for both alarms
Wiring for Frequency or Pulse Output Models
Figure 6 shows the wiring for either frequency output or pulse output models. (Optional wiring
is also shown for alarm outputs.)
Note: “Frequency” or “Pulse Output” model must be specified at the time of ordering. They cannot be interchanged by the user.
Figure 6. Frequency and Pulse Output Wiring
FSMAN.6 081214 Page 13
POWER-UP
At power-up, the following appears on the LCD:
Firmware Revision 5.67 or Higher
Electronics Revision
It takes 2 seconds for these messages to be displayed, during which time the output of the flowmeter is clamped at zero flow. After 2 seconds, the output signal starts indicating actual flow.
LCD READOUT
If the flowmeter is configured for High-Speed (HS) response, the LCD only shows “run”. No other parameters can be viewed on the LCD in this mode.
LCD in HS mode:
In Low-Speed (LS) (see Select Response Time) the user can toggle the readout between flow rate, total, pressure, and temperature. When in Run mode, use A2 pushbutton to select. The selection is stored in the internal memory, so if power is removed from the flowmeter it remembers the selection next time it is powered up.
Note: The output signal and alarms always indicate “flow rate”.
FSMAN.6 081214 Page 14
Loading...
+ 32 hidden pages