Unitron Z730 series User Manual

Page 1
Distributed By:
Page 2
Z730 STEREO MICROSCOPE
CONTENTS
SAFETY NOTES .................................................................................................................. 3
INTRODUCTION .................................................................................................................. 4
UNPACKING AND COMPONENTS ..................................................................................... 4
COMPONENTS DIAGRAM .................................................................................................. 4
ASSEMBLY .......................................................................................................................... 5
DETAILED ASSEMBLY PROCEDURE ................................................................................ 6
ADJUSMENT AND OPERATION
ADJUSTING BINOCULAR VIEWING HEAD ................................................................... 8
FOCUS ADJUSTMENT ................................................................................................... 9
SPECIFICATIONS .............................................................................................................. 10
TROUBLESHOOTING ........................................................................................................ 11
MAINTENANCE.................................................................................................................. 12
SERVICE ............................................................................................................................ 12
WARRANTY ....................................................................................................................... 12
UNITRON
®
2
Page 3
Z730 STEREO MICROSCOPE
SAFETY NOTES
1. Open the shipping carton carefully to prevent any accessory, i.e. objectives or eyepieces, from
dropping and being damaged.
2. Keep the instrument out of direct sunlight, high temperature or humidity, and dusty environments.
3. If any specimen solutions or other liquids splash onto the stage, objective or any other component,
disconnect the power cord immediately and wipe up the spillage. Otherwise, the instrument may be
damaged.
4. LAMP REPLACEMENT -- CAUTION: the glass envelope of the lamp may be extremely
hot. DO NOT attempt to change the lamp before it is completely cooled or without wearing adequate skin protection.
5. All electrical connectors (power cord) should be inserted into an electrical surge protector to prevent
damage due to voltage fluctuations.
6. FUSE REPLACEMENT
-- For safety when replacing the fuse (ONLY replace with the same size, type and rating of original fuse), be sure the main switch is in the off position, disconnect the power cord from outlet, and replace the fuse. Reconnect the power cord and turn unit on.
7. Confirm that the input voltage indicated on your microscope corresponds to your line voltage. The use of a different input voltage other than indicated will cause severe damage to the microscope. NOTE: Always plug the stereomicroscope power cord into a suitable grounded electrical outlet. A grounded 3-wire cord is provided.
CARE AND MAINTENANCE
1. Do not attempt to disassemble any component including eyepieces, objectives or focusing assembly.
2. Keep the instrument clean; remove dirt and debris regularly. Accumulated dirt on metal surfaces should be cleaned with a damp cloth. More persistent dirt should be removed using a mild soap solution. Do not use organic solvents for cleansing.
3. The outer surface of the optics should be inspected and cleaned periodically using an air stream from an air bulb. If dirt remains on the optical surface, use a soft, lint free cloth or cotton swab dampened with a lens cleaning solution (available at camera stores). All optical lenses should be swabbed using a circular motion. A small amount of absorbent cotton such as a Q-tip or cotton swab, makes a useful tool for cleaning recessed optical surfaces. Avoid using an excessive amount of solvents as this may cause problems with optical coatings or cemented optics or the flowing solvent may pick up grease making cleaning more difficult.
4. Store the instrument in a cool, dry environment. Cover the microscope with the dust cover when not in use.
®
5. UNITRON
microscopes are precision instruments which require periodic servicing to maintain proper performance and to compensate for normal wear. A regular schedule of preventative maintenance by qualified personnel is highly recommended. Your authorized UNITRON can arrange for this service.
UNITRON
®
®
distributor
3
Page 4
Z730 STEREO MICROSCOPE
INTRODUCTION
Congratulations on the purchase of your new UNITRON® microscope. UNITRON® microscopes are engineered and manufactured to the highest quality standards. Your microscope will last a lifetime if
®
used and maintained properly. UNITRON
microscopes are carefully assembled, inspected and tested by our staff of trained technicians in our New York facility. Careful quality control procedures ensure each microscope is of the highest quality prior to shipment.
UNPACKING AND COMPONENTS
Your microscope arrived packed in a molded shipping carton.
Do not discard the carton
: the shipping carton should be retained for reshipment of your microscope if needed. Avoid placing the microscope in dusty surroundings or in high temperature or humid areas as mold and mildew can form. Carefully remove the microscope from the shipping carton and place the microscope on a flat, vibration-free surface.
COMPONENTS DIAGRAM
Binocular Head & Zoom Body
Eyepieces
Stand
Focusing Knob
Diopter Rings
Lock-screw
Stage Plate
Base
NOTE: Stand ordered may not be as pictured above.
UNITRON
®
Power Cord
Brightness Adjustment Knob
4
Page 5
Z730 STEREO MICROSCOPE
ASSEMBLY
The diagram below shows how to assemble the various modules. The numbers indicate the order of assembly.
When assembling the microscope, make sure that all parts are free of dust and dirt, and avoid scratching any parts or touching glass surfaces.
Binocular Head & Zoom Body
Eyepieces
Lock-screw
Power Cord
UNITRON
®
5
Page 6
Z730 STEREO MICROSCOPE
DETAILED ASSEMBLY PROCEDURE
Mounting the Binocular Head (Fig.1)
Loosen the clamping knob③ on the focus mount
and insert the binocular head and zoom body assembly②into the bracket of the focus mount, then tighten the lock screw ③.
Fig. 1
Fig. 2
Mounting the Eyepieces (Fig.1, Fig.2, Fig.3)
Remove the eyepiece dust caps gently insert the two adjustable eyepieces (Fig. 2) into the eyepiece sleeves until they stop,
as shown in Fig.3.
(Fig. 1)and
IMPORTANT
ENVIRONMENT REQUIREMENTS
1 Ambient temperature: 5°C to 40°C (41°F
to 104°F)
2 High temperature will result in moisture
and mildew and could damage instrument.
3 Avoid placing the instrument in a dusty
environment. When not in use, cover the microscope with the dust cover.
Maximum R.H:85%
Fig. 3
UNITRON
®
4 Keep the microscope on a flat and stable
surface.
6
Page 7
Fig. 4
Z730 STEREO MICROSCOPE
For Microscopes Purchased With Illuminated Stands
VOLTAGE CHECK
Confirm that the input voltage indicated on the rear label of the microscope corresponds to your line voltage. The use of a different input voltage than indicated will cause severe damage to your microscope.
Connecting the Power Cord (Fig.4)
Rotate the left and right brightness adjusting switches
position) before connecting the power cord
clockwise until it stops (the off
.
Fig. 5
Insert the power plug ② into the power jack of the microscope; make sure the connection is snug.
Plug the power cord ② into the power supply receptacle.
Adjustable illumination for both top and bottom. Input Rating:100V~240V.
Optional trinocular head is available for camera or imaging accessories.
Video/Photography Accessories Assembly
Thread the C-mount video adapter ④ onto the camera ⑤ and insert into the trinocular port. Secure with lock screw.
UNITRON
®
7
Page 8
Z730 STEREO MICROSCOPE
ADJUSTMENT AND OPERATION
Adjusting Binocular Viewing Head
Fig. 6
Adjusting Interpupillary Distance (Fig.6)
Different users have different interpupillary distances (this distance is between the centers of pupils of each eye). When operators of the microscope change, it will be necessary to adjust interpupillary distance.
While looking through the eyepieces, hold the left and right eyepiece tubes of the binocular assembly①and adjust the eyepiece tubes by opening or closing them until the left and right fields of view coincide completely and you are able to see a complete circle.
Adjusting Diopter Ring (Fig.7)
Set the diopter rings of both eyepieces to “0” (zero) position. (Do this when users change, because different users will have different diopter settings.)
Fig. 7
Place an easy-to-observe specimen on the stage plate, i.e., a coin.
Rotate the zoom knob③to the highest magnification 3x, then turn the focusing knob
to focus the specimen. Rotate the zoom knob③to the lowest
magnification 0.7x, looking only into the left eyepiece, adjust the diopter ring on left eyepiece to focus the specimen. Then repeat procedure for the right eyepiece.
NOTE: The working distance of the microscope is 97mm (the distance between the microscope objective to the top of the specimen), the binocular head bracket⑤can be mounted onto higher or lower positions according to your needs by removing the clamping bolt with the Allen wrench provided.
UNITRON
®
8
Page 9
Focus Adjustment
Fig. 8
Z730 STEREO MICROSCOPE
Adjusting the Rotation Tension of the Focus Adjustment Knob (Fig.8, Fig.9)
To adjust tension, hold both left and right focus adjustment knobs①with both hands, hold the left knob (to prevent it from turning), and rotate the right knob clockwise to increase (tighten) or
counterclockwise to decrease (loosen) the focus knob tension.
After tension adjustment has been completed, always rotate both adjustment knobs in the same direction.
Changing the Magnification
Fig. 9
The zooming knobs located on both sides of the zooming body will change the magnification of the specimen image.
Total Magnification = Magnification of zoom body x magnification of eyepiece (i.e., 0.7 x 10=7x)
NOTE:
If using the optional auxiliary 0.5x objective, then the above magnification would be multiplied by auxiliary objective magnification. (i.e., auxiliary 0.5 x 7=3.5x)
FOR MICROSCOPES PURCHASED WITH ILLUMINATION
Adjustable illumination for both LEDs (top and bottom).
Brightness adjustment: the right brightness adjustment knob③is used to adjust the bottom illumination, and the left brightness adjustment knob④is used to adjust the top illumination. Each knob has its own on/off. Rotate
counterclockwise to increase the power and brightness. Rotating clockwise would decrease the brightness until the power is off.
UNITRON
®
9
Page 10
Z730 STEREO MICROSCOPE
SPECIFICATIONS
Zoom ratio: 4.3 to 1 Objective magnification: 0.7x~3.0x Binocular viewing head: Inclined at 45 or 60 degrees depending on model purchased, interpupillary
distance range: 55~75mm Working distance: 97mm Illumination system: Adjustable illumination for both LEDs (top and bottom)
Eyepiece
Auxiliary Objective
0.5X 180 3.5-15X 51.4-12 5.2-22.5X 40.0-9.4 7-30X 31.4-7.4
0.63X 140 4.4-18.9X 43.9-10.6 6.6-28.3X 31.7-7.4 8.8-37.8X 24.9-5.8 2X 30 14-60X 12.8-3 21-90X 10.0-2.3 28-120X 7.9-1.8
Working
Distance
(mm)
Magnification
97 7-30X 25.7-6 10.5-45X 20.0-4.7 14-60X 15.7-3.7
10X/20 15X/16 20X/12.5
Total
Visual Field
(mm)
Total
Magnification
Visual Field
(mm)
Total
Magnification
Visual Field
(mm)
UNITRON
®
10
Page 11
Z730 STEREO MICROSCOPE
TROUBLESHOOTING
Under certain conditions, performance of this unit may be adversely affected by factors other than defects. If a problem occurs, please review the following list and take remedial action as needed. If you cannot solve the problem after checking the entire list, please contact your local dealer for assistance.
PROBLEM CAUSE SOLUTION PAGE
1.Incomplete binocular vision.
2.Dirt is visible in field of
view
3.The visibility of the image is poor.
4. Specimen image blurs when zoom magnification is changed.
5.Coarse focus adjustment knobs rotate
with too much resistance. 6.Zoom microscope body drops or specimen goes
out of focus during observation.
Interpupillary distance is not correctly adjusted. Diopter adjustment is incomplete. Dirt on specimen Clean specimen 3 – Care &
Dirt on eyepiece Clean eyepiece Dust on objective front
lens. Diopter ring is not
correctly adjusted. Not in complete focus on specimen.
Tension adjustment ring is too tight.
Tension adjustment ring is too loose.
Adjust it correctly 8 – Fig. 6
Complete diopter adjustment
Maintenance
Clean lens surface 3 – Care &
Maintenance
Adjust it correctly. 8 – Fig. 7
Focus specimen correctly at a high magnification. Loosen it properly. 9 – Fig 8
Tighten it properly. 9 – Fig 8
UNITRON
®
11
Page 12
MAINTENANCE
Z730 STEREO MICROSCOPE
Please remember to
never
leave the microscope with eyepieces removed and always protect the
microscope with the dust cover when not in use.
SERVICE
UNITRON® microscopes are precision instruments which require periodic servicing to keep them performing properly and to compensate for normal wear. A regular schedule of preventative maintenance by qualified personnel is highly recommended. Your authorized UNITRON this service. Should unexpected problems be experienced with your instrument, proceed as follows:
®
1. Contact the UNITRON
distributor from whom you purchased the microscope. Some
problems can be resolved simply over the telephone.
2. If it is determined that the microscope should be returned to your UNITRON UNITRON® for warranty repair, pack the instrument in its original Styrofoam shipping carton. If you no longer have this carton, pack the microscope in a crush-resistant carton with a minimum of three inches of a shock absorbing material surrounding it to prevent in-transit damage. The microscope should be wrapped in a plastic bag to prevent Styrofoam dust from damaging the microscope. Always ship the microscope in an upright position;
NEVER SHIP A MICROSCOPE ON ITS SIDE
component should be shipped prepaid and insured.
®
distributor can arrange for
®
distributor or to
. The microscope or
LIMITED MICROSCOPE WARRANTY
This microscope is warranted to be free from defects in material and workmanship for a period of five (5) years for mechanical and optical components and one (1) year for electrical components from the date of invoice to the original (end user) purchaser. This warranty does not cover damage caused in-transit, misuse, neglect, abuse or damage resulting from improper servicing or modification by other then UNITRON® approved service personnel. This warranty does not cover any routine maintenance work or any other work, which is reasonably expected to be performed by the purchaser. Normal wear is excluded from this warranty. No responsibility is assumed for unsatisfactory operating performance due to environmental conditions such as humidity, dust, corrosive chemicals, deposition of oil or other foreign matter, spillage or other conditions beyond the control of Unitron Ltd. This warranty expressly excludes any liability by Unitron Ltd. for consequential loss or damage on any grounds, such as (but not limited to) the non-availability to the End User of the product(s) under warranty or the need to repair work processes. Should any defect in material, workmanship or electronic component occur under this warranty contact your UNITRON® distributor. This warranty is limited to the continental United States of America. All items returned for warranty repair must be sent freight prepaid and insured to Unitron Ltd., 73 Mall Drive, Commack, NY 11725 – USA. All warranty repairs will be returned freight prepaid to any destination within the continental United States of America. For all foreign warranty repairs, return freight charges are the responsibility of the individual/company who returned the merchandise for repair.
UNITRON
®
12
Loading...