1. Open the shipping carton carefully to prevent any accessory, i.e. objectives or eyepieces, from
dropping and being damaged.
2. Keep the instrument out of direct sunlight, high temperature or humidity, and dusty environments.
3. If any specimen solutions or other liquids splash onto the stage, objective or any other component,
disconnect the power cord immediately and wipe up the spillage. Otherwise, the instrument may be
damaged.
4. LAMP REPLACEMENT -- CAUTION: the glass envelope of the lamp may be extremely
hot. DO NOT attempt to change the lamp before it is completely cooled or without
wearing adequate skin protection.
5. All electrical connectors (power cord) should be inserted into an electrical surge protector to prevent
damage due to voltage fluctuations.
6. FUSE REPLACEMENT -- For safety when replacing the fuse (ONLY replace with the
same size, type and rating of original fuse), be sure the main switch is in the off position,
disconnect the power cord from outlet, and replace the fuse. Reconnect the power cord
and turn unit on.
7. Confirm that the input voltage indicated on your microscope corresponds to your line voltage. The
use of a different input voltage other than indicated will cause severe damage to the microscope.
NOTE: Always plug the stereomicroscope power cord into a suitable grounded electrical outlet. A
grounded 3-wire cord is provided.
CARE AND MAINTENANCE
1. Do not attempt to disassemble any component including eyepieces, objectives or focusing assembly.
2. Keep the instrument clean; remove dirt and debris regularly. Accumulated dirt on metal surfaces
should be cleaned with a damp cloth. More persistent dirt should be removed using a mild soap
solution. Do not use organic solvents for cleansing.
3. The outer surface of the optics should be inspected and cleaned periodically using an air stream from
an air bulb. If dirt remains on the optical surface, use a soft, lint free cloth or cotton swab dampened
with a lens cleaning solution (available at camera stores). All optical lenses should be swabbed using
a circular motion. A small amount of absorbent cotton such as a Q-tip or cotton swab, makes a useful
tool for cleaning recessed optical surfaces. Avoid using an excessive amount of solvents as this may
cause problems with optical coatings or cemented optics or the flowing solvent may pick up grease
making cleaning more difficult.
4. Store the instrument in a cool, dry environment. Cover the microscope with the dust cover when not
in use.
5. UNITRON® microscopes are precision instruments which require periodic servicing to maintain
proper performance and to compensate for normal wear. A regular schedule of preventative
maintenance by qualified personnel is highly recommended. Your authorized UNITRON® distributor
can arrange for this service.
Congratulations on the purchase of your new UNITRON® microscope. UNITRON® microscopes are
engineered and manufactured to the highest quality standards. Your microscope will last a lifetime if
used and maintained properly. UNITRON® microscopes are carefully assembled, inspected and tested
by our staff of trained technicians in our New York facility. Careful quality control procedures ensure each
microscope is of the highest quality prior to shipment.
UNPACKING AND COMPONENTS
Your microscope arrived packed in a molded shipping carton. Do not discard the carton: the shipping
carton should be retained for reshipment of your microscope if needed. Avoid placing the microscope in
dusty surroundings or in high temperature or humid areas as mold and mildew can form. Carefully
remove the microscope from the shipping carton and place the microscope on a flat, vibration-free
surface.
The mechanical stage may be installed in either
side of the stage to enlarge the stage surface.
The x-y controls must be installed on the
opposite side of theauxiliary stage.
Generally, the x-y controls are installed on the
right side for comfortable operation.
Installing The Mechanical Stage
Screw the clamping bolts into to the mechanical
stage, then into the fixed stage from below on
either the right or left side. Tighten the bolts with
a screwdriver until the mechanical stage is
securely attached.
Installing The X-Y Controls
Install the ruler using the same method as the
mechanical stage.
Installing and Replacing the Halogen
UNITRON
®
Bulb (Figures 1-3)
Use only the specified halogen lamp: 6V 30W
1. Pull out the lamp house① (Figure 1)
2. Holding the bulb ① with a piece of gauze or
other protective material, insert the bulb pins
straight and fully into the pin holes③ on the
lamp house. (Figure 2)
CAUTION: The bulb, lamp housing and
nearby parts will be very hot during usage.
3. When replacing the lamp, turn the main
switch to “O (off) and remove the power cord.
Allow the lamp, lamp house and adjacent areas
Connect the power cord, then set the main
switch① on the side of the base to the “on”
position (“-”).
Adjusting The Brightness
(Figure 16)
Rotate the brightness adjustment knob (2)
clockwise to increase the brightness.
Conversely, rotating the knob decreases the
brightness.
Using the lamp at its lowest acceptable
intensity will increase the life of the lamp.
Adjusting the Tension
Adjustment Collar
(Figure 17)
The tension of the coarse focusing knob
(3) has already been properly adjusted.
Use the supplied tension adjustment wrench to
adjust the tension adjustment collar (4).
Rotating the collar clockwise increases the
tension on the coarse focusing knob. A counterclockwise rotation will decrease the tension on
the coarse focus knob.
The coarse focusing knob is too loose if the
nosepiece descends on its own or if the
specimen loses its focus shortly after focusing
with the fine focus knob (5). Adjust the tension
adjustment collar so these do not occur.
1. Make sure the field of view is bright and
evenly illuminated, and has no filament shadow.
If there is filament shadow, adjust the condenser
adjustment knob③ to its proper position.
2. To obtain a clear, sharp image, adjust the
field iris diaphragm① and aperture iris
diaphragm (2) to their proper positions. One or
both diaphragms may need to be adjusted to
obtain the sharpest image.
Aperture Iris Diaphragm: The aperture iris diaphragm is not used to control the brightness of the
image. It controls the aperture angle of the incident rays. The diaphragm is not adjusted properly
until the incident rays just fill the objective lens. At this position, the contrast and sharpness of the
image will be the best. The aperture diaphragm should be re-adjusted when changing objectives.
Field Iris Diaphragm: The field iris diaphragm is used to control the size of the field of view,
reduce the reflected rays and glare, thereby increasing the contrast of the image. Generally, the field
iris diaphragm is at the proper setting when the rays just fill the field of view of the eyepieces.
1. Looking through the right ocular only with your
right eye, use the coarse and fine focusing
adjustment knobs to focus on the specimen.
2. Then with your left eye only, look through the
left ocular. If the image is not sharp, rotate the
diopter adjustment ring① to focus on the
specimen.
Diopter adjustment range: ±5
Adjusting the Interpupillary Distance
(Figure 22)
While observing with both eyes, hold the left and
right prism holders (2 & 3), rotate them around
the axis to adjust the interpupillary distance until
the left and right fields of view coincide to form a
single image.
Interpupillary range:48~75mm
Switching the Light Path
(Figure 23)
Pull out or push in the light path selector lever
(4) to select the light path required.
For the binocular observation, push the lever in
completely .For video or microphotography, pull
the lever out completely.
Observing through binocular position, focus the
specimen until a sharp image is obtained. Then
check the image on the monitor to which the
microscope is connected. If the image is not sharp,
rotate the video adapter (4) until the image is sharp.
Microscope Photography (Figure 26)
Selecting the Light Path (Figure 24)
Follow the instructions on previous page
Installing the Photo Adapter (Figure 26)
1. Loosen the thumb screw ① on the trinocular
viewing tube and remove the dust cap②.
2. Install the photo accessory adapter into the
trinocular port and tighten the thumb screw①.
3. Connect the photo adapter with the camera.
Then insert the adapter into the photo accessory
adapter③.
To avoid disturbing the image focus, place the
camera viewfinder toward the front of the
microscope when installing the camera.
The magnification of photomicrograph =
magnification of objective x magnification of the
photo adapter.
The shutter on some cameras may cause a jarring
impact when photographing through the
microscope. To weaken the impact and obtain a
clear, sharp image, select a longer exposure time.
Focusing
Focus the specimen using binocular observation.
Then, for microphotography, focus the camera.
Adjusting the Color Temperature
When photographing with sunlight film:
1. Mount the blue filter into the filter bracket.
2. Turn the brightness adjustment knob to the
maximum limit. This allows you to obtain sunlight
illumination.
Under certain conditions, performance of this unit may be adversely affected by factors other than
defects. If a problem occurs, please review the following list and take remedial action as needed. If you
cannot solve the problem after checking the entire list, please contact your local dealer for assistance.
OPTICAL
PROBLEM CAUSE CORRECTIVE MEASURE
Darkness at the periphery or uneven
brightness in the field of view
Dirt or dust on the viewfield
Revolving nosepiece not in click
stop position
Dirt or dust on the lens, eyepiece,
condenser, objective, collector lens
or specimen
Poor image quality
Condenser aperture is closed or
open too much
Condenser is positioned too low
IMAGE PROBLEMS
Image moves while focusing Specimen rises from stage surface
Image tinged yellow Blue filter not used
Image is too bright
Revolving nosepiece is not in the
click-stop position
Lamp intensity is too low
Lamp intensity is too high Adjust the light intensity by turning
Turn the nosepiece to click-stop
position so the objective is in the
optical path
Clean the lens
Open or close properly
Position the condenser at the upper
limit
Secure the specimen with the holder
Turn the nosepiece to the click-stop
position
Use daylight blue filter
Adjust the light intensity by turning
the intensity control dial
the intensity control dial
Insufficient brightness
Lamp does not light when switched
on
Slippage of focus when using the
coarse focusing knob
Adjust the light intensity by turning
the intensity control dial
Open to the proper setting
Position the condenser to the upper
limit
No electrical power
Lamp bulb burnt out
Fuse blown out
Tension adjustment is set too low
Check power cord connection
Replace bulb
Replace fuse
Increase the tension on the focusing
knobs
Tension adjustment is set too high Loosen the tension on the focusing
knobs
MEC2 INVERTED METALLURGICAL MICROSCOPE
MAINTENANCE
Please remember to never leave the microscope with eyepieces removed and always protect the
microscope with the dust cover when not in use.
SERVICE
UNITRON® microscopes are precision instruments which require periodic servicing to keep them
performing properly and to compensate for normal wear. A regular schedule of preventative maintenance
by qualified personnel is highly recommended. Your authorized UNITRON® distributor can arrange for
this service. Should unexpected problems be experienced with your instrument, proceed as follows:
1. Contact the UNITRON® distributor from whom you purchased the microscope. Some
problems can be resolved simply over the telephone.
2. If it is determined that the microscope should be returned to your UNITRON® distributor or to
UNITRON® for warranty repair, pack the instrument in its original Styrofoam shipping carton. If you no
longer have this carton, pack the microscope in a crush-resistant carton with a minimum of three inches
of a shock absorbing material surrounding it to prevent in-transit damage. The microscope should be
wrapped in a plastic bag to prevent Styrofoam dust from damaging the microscope. Always ship the
microscope in an upright position; NEVER SHIP A MICROSCOPE ON ITS SIDE. The microscope or
component should be shipped prepaid and insured.
LIMITED MICROSCOPE WARRANTY
This microscope is warranted to be free from defects in material and workmanship for a period of five (5) years for
mechanical and optical components and one (1) year for electrical components from the date of invoice to the
original (end user) purchaser. This warranty does not cover damage caused in-transit, misuse, neglect, abuse or
damage resulting from improper servicing or modification by other then UNITRON® approved service personnel.
This warranty does not cover any routine maintenance work or any other work, which is reasonably expected to be
performed by the purchaser. Normal wear is excluded from this warranty. No responsibility is assumed for
unsatisfactory operating performance due to environmental conditions such as humidity, dust, corrosive chemicals,
deposition of oil or other foreign matter, spillage or other conditions beyond the control of Unitron Ltd. This
warranty expressly excludes any liability by Unitron Ltd. for consequential loss or damage on any grounds, such as
(but not limited to) the non-availability to the End User of the product(s) under warranty or the need to repair work
processes. Should any defect in material, workmanship or electronic component occur under this warranty contact
your UNITRON® distributor or UNITRON® at (631) 543-2000. This warranty is limited to the continental United
States of America. All items returned for warranty repair must be sent freight prepaid and insured to Unitron Ltd.,
73 Mall Drive, Commack, NY 11725 – USA. All warranty repairs will be returned freight prepaid to any destination
within the continental United States of America. For all foreign warranty repairs, return freight charges are the
responsibility of the individual/company who returned the merchandise for repair.