Unitron MEC2 User Manual

INVERTED METALLURGICAL
MICROSCOPE
MANUAL
73 Mall Drive, Commack, NY 11725 • 631-543-2000 (P) • 631-589-6975 (F)
www.unitronusa.com • info@unitronusa.com
MEC2 INVERTED METALLURGICAL MICROSCOPE
CONTENTS
SAFETY NOTES .................................................................................................................. 3
CARE AND MAINTENANCE ................................................................................................ 3
INTRODUCTION .................................................................................................................. 4
UNPACKING AND COMPONENTS ..................................................................................... 4
COMPONENT DIAGRAMS ............................................................................................... 4-5
ASSEMBLY .......................................................................................................................... 5
DETAILED ASSEMBLY PROCEDURE ................................................................................ 6
ADJUSMENT AND OPERATION
TURNING ON THE LAMP ............................................................................................. 10
ADJUSTING THE BRIGHTNESS .................................................................................. 10
ADJUSTING THE TENSION ADJUSTMENT COLLAR ................................................. 10
ADJUSTING THE ILLUMINATION ................................................................................ 11
PLACING AND MOVING THE SPECIMEN ................................................................... 12
ADJUSTING THE DIOPTER ......................................................................................... 13
ADJUSTING THE INTERPUPILLARY DISTANCE ....................................................... 13
SWITCHING THE LIGHT PATH .................................................................................... 13
MICROSCOPE VIDEO AND PHOTOGRAPHY ............................................................ 14
SPECIFICATIONS .............................................................................................................. 16
TROUBLESHOOTING ........................................................................................................ 17
MAINTENANCE.................................................................................................................. 18
SERVICE ............................................................................................................................ 18
WARRANTY ....................................................................................................................... 12
UNITRON
®
73 Mall Drive, Commack, NY 11725 • 631-543-2000 • www.unitronusa.com 2
MEC2 INVERTED METALLURGICAL MICROSCOPE
SAFETY NOTES
1. Open the shipping carton carefully to prevent any accessory, i.e. objectives or eyepieces, from dropping and being damaged.
2. Keep the instrument out of direct sunlight, high temperature or humidity, and dusty environments.
3. If any specimen solutions or other liquids splash onto the stage, objective or any other component, disconnect the power cord immediately and wipe up the spillage. Otherwise, the instrument may be damaged.
4. LAMP REPLACEMENT -- CAUTION: the glass envelope of the lamp may be extremely
hot. DO NOT attempt to change the lamp before it is completely cooled or without wearing adequate skin protection.
5. All electrical connectors (power cord) should be inserted into an electrical surge protector to prevent damage due to voltage fluctuations.
6. FUSE REPLACEMENT -- For safety when replacing the fuse (ONLY replace with the
same size, type and rating of original fuse), be sure the main switch is in the off position, disconnect the power cord from outlet, and replace the fuse. Reconnect the power cord and turn unit on.
7. Confirm that the input voltage indicated on your microscope corresponds to your line voltage. The use of a different input voltage other than indicated will cause severe damage to the microscope.
NOTE: Always plug the stereomicroscope power cord into a suitable grounded electrical outlet. A
grounded 3-wire cord is provided.
CARE AND MAINTENANCE
1. Do not attempt to disassemble any component including eyepieces, objectives or focusing assembly.
2. Keep the instrument clean; remove dirt and debris regularly. Accumulated dirt on metal surfaces should be cleaned with a damp cloth. More persistent dirt should be removed using a mild soap solution. Do not use organic solvents for cleansing.
3. The outer surface of the optics should be inspected and cleaned periodically using an air stream from an air bulb. If dirt remains on the optical surface, use a soft, lint free cloth or cotton swab dampened with a lens cleaning solution (available at camera stores). All optical lenses should be swabbed using a circular motion. A small amount of absorbent cotton such as a Q-tip or cotton swab, makes a useful tool for cleaning recessed optical surfaces. Avoid using an excessive amount of solvents as this may cause problems with optical coatings or cemented optics or the flowing solvent may pick up grease making cleaning more difficult.
4. Store the instrument in a cool, dry environment. Cover the microscope with the dust cover when not in use.
5. UNITRON® microscopes are precision instruments which require periodic servicing to maintain proper performance and to compensate for normal wear. A regular schedule of preventative maintenance by qualified personnel is highly recommended. Your authorized UNITRON® distributor can arrange for this service.
UNITRON
®
73 Mall Drive, Commack, NY 11725 • 631-543-2000 • www.unitronusa.com 3
MEC2 INVERTED METALLURGICAL MICROSCOPE
Light Path Selector Lever
On/Off Switch
INTRODUCTION
Congratulations on the purchase of your new UNITRON® microscope. UNITRON® microscopes are engineered and manufactured to the highest quality standards. Your microscope will last a lifetime if used and maintained properly. UNITRON® microscopes are carefully assembled, inspected and tested by our staff of trained technicians in our New York facility. Careful quality control procedures ensure each microscope is of the highest quality prior to shipment.
UNPACKING AND COMPONENTS
Your microscope arrived packed in a molded shipping carton. Do not discard the carton: the shipping carton should be retained for reshipment of your microscope if needed. Avoid placing the microscope in dusty surroundings or in high temperature or humid areas as mold and mildew can form. Carefully remove the microscope from the shipping carton and place the microscope on a flat, vibration-free surface.
COMPONENT DIAGRAMS
Diopter Adjustment Collar
Illumination Intensity Adjustment Dial
Interpupillary Distance Indicator
Tension Adjustment Collar
UNITRON
®
73 Mall Drive, Commack, NY 11725 • 631-543-2000 • www.unitronusa.com 4
Fine Focusing Knob
Coarse Focusing Knob
MEC2 INVERTED METALLURGICAL MICROSCOPE
Analyzer
COMPONENT DIAGRAMS (continued)
Condenser Adjsutment Lever Polarizer
UNITRON
®
73 Mall Drive, Commack, NY 11725 • 631-543-2000 • www.unitronusa.com 5
Aperture
Diaphragm
Field
Diaphragm
Filter
MEC2 INVERTED METALLURGICAL MICROSCOPE
Fig. 1
Fig. 2
Fig. 3
DETAILED ASSEMBLY PROCEDURE
Attaching the Mechanical Stage And X-Y Controls
The mechanical stage may be installed in either side of the stage to enlarge the stage surface. The x-y controls must be installed on the opposite side of theauxiliary stage.
Generally, the x-y controls are installed on the right side for comfortable operation.
Installing The Mechanical Stage
Screw the clamping bolts into to the mechanical stage, then into the fixed stage from below on either the right or left side. Tighten the bolts with a screwdriver until the mechanical stage is securely attached.
Installing The X-Y Controls
Install the ruler using the same method as the mechanical stage.
Installing and Replacing the Halogen
UNITRON
®
Bulb (Figures 1-3
Use only the specified halogen lamp: 6V 30W
1. Pull out the lamp house (Figure 1)
2. Holding the bulb with a piece of gauze or other protective material, insert the bulb pins
straight and fully into the pin holes on the lamp house. (Figure 2)
CAUTION: The bulb, lamp housing and nearby parts will be very hot during usage.
3. When replacing the lamp, turn the main switch to “O (off) and remove the power cord. Allow the lamp, lamp house and adjacent areas
73 Mall Drive, Commack, NY 11725 • 631-543-2000 • www.unitronusa.com 6
to sufficiently cool before handling. The lamp can become very hot and will cause burns.
Fig. 4
Fig. 5
Fig. 6
Fig. 8
Fig. 7
MEC2 INVERTED METALLURGICAL MICROSCOPE
Insert the lamp gently but firmly into the socket.
Do not touch the halogen bulb with bare fingers. Doing so will shorten the service life of the lamp.
If you leave fingerprints on the lamp surface, clean it with a soft cloth or lint free paper.
4. Align the socket with the pins , and the
bolts with the jacks . Then insert the lamphouse cover completely into place.
(Figures 3 & 4)
Installing the Stage Plate (Figure 5)
The stage plate is installed by placing it in the center of fixed stage.
Installing the Eyepieces (Figure 6)
1. Remove the protective caps on the eyepiece tubes.
2. Insert the eyepieces into eyepiece tubes until they are completely inserted.
Installing the Video Adapter
(Figures 7, 8, & 9)
Insert the video adapter tube (Figure 7) into the trinocular viewing head (Figure 8) and tighten the thumb screw (Figure 9 – next page).
UNITRON
®
73 Mall Drive, Commack, NY 11725 • 631-543-2000 • www.unitronusa.com 7
Fig. 9
Fig. 10
Fig. 11
MEC2 INVERTED METALLURGICAL MICROSCOPE
Mounting Filters and Polarizer
(Figure 10)
1. Insert a standard filter(s) into slot (1)
2. For polarized observation, insert the polarizer filter into slot (2).
Standard filters are blue, yellow and green filters
Mounting the Analyzer (Figure 11)
Insert the analyzer (3) into the slot under the trinocular viewing head.
The analyzer may be used alone or with the polarizer.
UNITRON
®
73 Mall Drive, Commack, NY 11725 • 631-543-2000 • www.unitronusa.com 8
Fig. 12
Fig. 13
Fig. 14
MEC2 INVERTED METALLURGICAL MICROSCOPE
Connecting the Power Cord
(Figures 12, 13 & 14)
The electric cord should never be bent, twisted, crushed or subjected to stress.
1. Set the main switchto “O” (off) before connecting the power cord.
2. Insert the power cord into the receptacle②.
3. Insert the BNC cord plug (3) into the BNC connector (4).
Ensure the power cord is connected to a properly grounded electrical outlet.
Replacing the Fuse (Figures 12 & 13)
Ensure the main switchis set to “O” (Off) and the power cord is disconnected before replacing
the fuse. With a flat-head screwdriver, remove the fuse holder (5), replace with a new fuse, then re-install the fuse holder.
Fuse rating: 250V, 500mA
Connecting the Power Cord
(Figure 14)
Insert the power cord (6) into a properly grounded 3-prong electrical outlet (7).
UNITRON
®
73 Mall Drive, Commack, NY 11725 • 631-543-2000 • www.unitronusa.com 9
Fig. 16
Fig. 15
Fig. 17
MEC2 INVERTED METALLURGICAL MICROSCOPE
ADJUSTMENT AND OPERATION
Turning On The Lamp
(Figure 15)
Connect the power cord, then set the main switchon the side of the base to the “on”
position (“”).
Adjusting The Brightness
(Figure 16)
Rotate the brightness adjustment knob (2) clockwise to increase the brightness. Conversely, rotating the knob decreases the brightness.
Using the lamp at its lowest acceptable intensity will increase the life of the lamp.
Adjusting the Tension Adjustment Collar
(Figure 17)
The tension of the coarse focusing knob
(3) has already been properly adjusted.
Use the supplied tension adjustment wrench to adjust the tension adjustment collar (4).
Rotating the collar clockwise increases the tension on the coarse focusing knob. A counter­clockwise rotation will decrease the tension on the coarse focus knob.
The coarse focusing knob is too loose if the nosepiece descends on its own or if the specimen loses its focus shortly after focusing with the fine focus knob (5). Adjust the tension adjustment collar so these do not occur.
UNITRON
®
73 Mall Drive, Commack, NY 11725 • 631-543-2000 • www.unitronusa.com 10
MEC2 INVERTED METALLURGICAL MICROSCOPE
Fig. 18
Adjusting the Illumination
(Figure 18)
1. Make sure the field of view is bright and evenly illuminated, and has no filament shadow. If there is filament shadow, adjust the condenser
adjustment knob to its proper position.
2. To obtain a clear, sharp image, adjust the field iris diaphragmand aperture iris
diaphragm (2) to their proper positions. One or both diaphragms may need to be adjusted to obtain the sharpest image.
Aperture Iris Diaphragm: The aperture iris diaphragm is not used to control the brightness of the image. It controls the aperture angle of the incident rays. The diaphragm is not adjusted properly until the incident rays just fill the objective lens. At this position, the contrast and sharpness of the image will be the best. The aperture diaphragm should be re-adjusted when changing objectives.
Field Iris Diaphragm: The field iris diaphragm is used to control the size of the field of view, reduce the reflected rays and glare, thereby increasing the contrast of the image. Generally, the field iris diaphragm is at the proper setting when the rays just fill the field of view of the eyepieces.
.
UNITRON
®
73 Mall Drive, Commack, NY 11725 • 631-543-2000 • www.unitronusa.com 11
Fig. 19
Fig. 20
MEC2 INVERTED METALLURGICAL MICROSCOPE
Placing and Moving the Specimen
(Figures 19 & 20)
1. Place the slide in the center of the stage and secure with the stage clip.
2. The specimen may be moved to different positions by turning the X-axis knob① and/or the
Y-axis knob.
Travel: X-axis (left-right) direction: 120mm
Y-axis (front-back) direction: 78mm
Use caution when changing the objective to prevent it from touching the stage, stage plate or specimen slide.
UNITRON
®
73 Mall Drive, Commack, NY 11725 • 631-543-2000 • www.unitronusa.com 12
Fig. 21
Fig. 22
Fig. 23
MEC2 INVERTED METALLURGICAL MICROSCOPE
Adjusting the Diopter
(Figure 21)
1. Looking through the right ocular only with your right eye, use the coarse and fine focusing adjustment knobs to focus on the specimen.
2. Then with your left eye only, look through the left ocular. If the image is not sharp, rotate the
diopter adjustment ringto focus on the specimen.
Diopter adjustment range: ±5
Adjusting the Interpupillary Distance
(Figure 22)
While observing with both eyes, hold the left and right prism holders (2 & 3), rotate them around the axis to adjust the interpupillary distance until the left and right fields of view coincide to form a single image.
Interpupillary range4875mm
Switching the Light Path
(Figure 23)
Pull out or push in the light path selector lever (4) to select the light path required.
For the binocular observation, push the lever in completely .For video or microphotography, pull the lever out completely.
UNITRON
®
73 Mall Drive, Commack, NY 11725 • 631-543-2000 • www.unitronusa.com 13
Fig. 24
MEC2 INVERTED METALLURGICAL MICROSCOPE
Light Path
Selecting Lever
Illumination Application
Pushed In 100% for binocular observation Binocular observation
Pulled Out 20% for binocular observation
80% for video or
Binocular observation and video
(CCTV) or microphotography
simultaneously
microphotography
Polarization Observation
1. Insert the polarizer and analyzer as described on page 8 (Figures 10 & 11).
2. The analyzer hand wheel can be rotated from 0 °to 90°. When the field of view becomes the darkest,
the polarization position has been obtained and polarized observation can be performed.
MICROSCOPE VIDEO AND PHOTOGRAPHY
Video Microscopy Selecting the Light Path (Figure 24)
1. Push in the light path selector (1) completely.
2. Using binocular observation, focus your specimen.
3. Pull out the light path selector completely.
Installing the Video Adapter (Figure 25)
1. Loosen the thumb screw on the trinocular
viewing tube, and remove the dust cap.
2. Remove the dust covers on the both ends of the video adapter.
3. Insert the flanged end into the trinocular viewing head and tighten the thumb screw.
4. The threaded video C-mount should be facing upwards.
UNITRON
®
73 Mall Drive, Commack, NY 11725 • 631-543-2000 • www.unitronusa.com 14
Fig. 26
UNITRON
Fig. 25
®
MEC2 INVERTED METALLURGICAL MICROSCOPE
Focusing (Figure 25)
Observing through binocular position, focus the specimen until a sharp image is obtained. Then check the image on the monitor to which the microscope is connected. If the image is not sharp, rotate the video adapter (4) until the image is sharp.
Microscope Photography (Figure 26)
Selecting the Light Path (Figure 24)
Follow the instructions on previous page
Installing the Photo Adapter (Figure 26)
1. Loosen the thumb screw on the trinocular
viewing tube and remove the dust cap.
2. Install the photo accessory adapter into the trinocular port and tighten the thumb screw①.
3. Connect the photo adapter with the camera. Then insert the adapter into the photo accessory
adapter.
To avoid disturbing the image focus, place the camera viewfinder toward the front of the microscope when installing the camera.
The magnification of photomicrograph = magnification of objective x magnification of the photo adapter.
The shutter on some cameras may cause a jarring impact when photographing through the microscope. To weaken the impact and obtain a clear, sharp image, select a longer exposure time.
Focusing
Focus the specimen using binocular observation. Then, for microphotography, focus the camera.
Adjusting the Color Temperature
When photographing with sunlight film:
1. Mount the blue filter into the filter bracket.
2. Turn the brightness adjustment knob to the maximum limit. This allows you to obtain sunlight illumination.
73 Mall Drive, Commack, NY 11725 • 631-543-2000 • www.unitronusa.com 15
SPECIFICATIONS
MEC2 INVERTED METALLURGICAL MICROSCOPE
Optical System
Viewing Tube
Eyepiece
Nosepiece
Objective
Focusing System
Stage
Mechanical Ruler
Illumination
Filters
Infinity Optical System
Siedentopf style; Trinocular Tube; Inclined 30º; Diopter Range: ±5; Interpupillary distance range: 48mm -75mm Standard configuration light distribution: Binocular 100% or Binocular/photo 20%/80%
Extra Wide Field 10x with 20mm field of view with diopter adjustment and eyeguards
Quintuple nosepiece with positive click stops
Plan achromatic objectives: 5x, 10x, 20x, 40x
Coaxial coarse and fine focusing adjustment; Vertical objective movement; Coarse focus: 377mm per roation; Fine focus: 0.2mm per rotation
Area: 160mm x 250mm
X/Y Coaxial Control; Right hand low position Movement range: 120mm x 78mm
6 volt 30 watt halogen; variable intensity; External lamp housing for Kohler illumination
Blue, yellow, green and ground glass
OBJECTIVES
Magnification
5x 0.12 15.4
10x .25 11
20x 0.4 6
40x 0.6 3.7
UNITRON
®
73 Mall Drive, Commack, NY 11725 • 631-543-2000 • www.unitronusa.com 16
Aperture Number
(N.A.)
Working Distance
(mm)
Conjugate Distance
(mm)
8
8
8
8
MEC2 INVERTED METALLURGICAL MICROSCOPE
TROUBLESHOOTING
Under certain conditions, performance of this unit may be adversely affected by factors other than defects. If a problem occurs, please review the following list and take remedial action as needed. If you cannot solve the problem after checking the entire list, please contact your local dealer for assistance.
OPTICAL
PROBLEM CAUSE CORRECTIVE MEASURE
Darkness at the periphery or uneven brightness in the field of view
Dirt or dust on the viewfield
Revolving nosepiece not in click stop position
Dirt or dust on the lens, eyepiece, condenser, objective, collector lens or specimen
Poor image quality
Condenser aperture is closed or open too much
Condenser is positioned too low
IMAGE PROBLEMS
Image moves while focusing Specimen rises from stage surface
Image tinged yellow Blue filter not used
Image is too bright
Revolving nosepiece is not in the click-stop position
Lamp intensity is too low
Lamp intensity is too high Adjust the light intensity by turning
Turn the nosepiece to click-stop position so the objective is in the optical path
Clean the lens
Open or close properly
Position the condenser at the upper limit
Secure the specimen with the holder
Turn the nosepiece to the click-stop position
Use daylight blue filter
Adjust the light intensity by turning the intensity control dial
the intensity control dial
Insufficient brightness
Lamp does not light when switched on
Slippage of focus when using the coarse focusing knob
Fine focus is ineffective
UNITRON
®
73 Mall Drive, Commack, NY 11725 • 631-543-2000 • www.unitronusa.com 17
Lamp intensity is too low
Aperture diaphragm closed too far
Condenser position is too low
Adjust the light intensity by turning the intensity control dial
Open to the proper setting
Position the condenser to the upper limit
No electrical power
Lamp bulb burnt out
Fuse blown out
Tension adjustment is set too low
Check power cord connection
Replace bulb
Replace fuse
Increase the tension on the focusing knobs
Tension adjustment is set too high Loosen the tension on the focusing
knobs
MEC2 INVERTED METALLURGICAL MICROSCOPE
MAINTENANCE
Please remember to never leave the microscope with eyepieces removed and always protect the microscope with the dust cover when not in use.
SERVICE
UNITRON® microscopes are precision instruments which require periodic servicing to keep them performing properly and to compensate for normal wear. A regular schedule of preventative maintenance by qualified personnel is highly recommended. Your authorized UNITRON® distributor can arrange for this service. Should unexpected problems be experienced with your instrument, proceed as follows:
1. Contact the UNITRON® distributor from whom you purchased the microscope. Some
problems can be resolved simply over the telephone.
2. If it is determined that the microscope should be returned to your UNITRON® distributor or to
UNITRON® for warranty repair, pack the instrument in its original Styrofoam shipping carton. If you no longer have this carton, pack the microscope in a crush-resistant carton with a minimum of three inches of a shock absorbing material surrounding it to prevent in-transit damage. The microscope should be wrapped in a plastic bag to prevent Styrofoam dust from damaging the microscope. Always ship the microscope in an upright position; NEVER SHIP A MICROSCOPE ON ITS SIDE. The microscope or component should be shipped prepaid and insured.
LIMITED MICROSCOPE WARRANTY
This microscope is warranted to be free from defects in material and workmanship for a period of five (5) years for mechanical and optical components and one (1) year for electrical components from the date of invoice to the original (end user) purchaser. This warranty does not cover damage caused in-transit, misuse, neglect, abuse or damage resulting from improper servicing or modification by other then UNITRON® approved service personnel. This warranty does not cover any routine maintenance work or any other work, which is reasonably expected to be performed by the purchaser. Normal wear is excluded from this warranty. No responsibility is assumed for unsatisfactory operating performance due to environmental conditions such as humidity, dust, corrosive chemicals, deposition of oil or other foreign matter, spillage or other conditions beyond the control of Unitron Ltd. This warranty expressly excludes any liability by Unitron Ltd. for consequential loss or damage on any grounds, such as (but not limited to) the non-availability to the End User of the product(s) under warranty or the need to repair work processes. Should any defect in material, workmanship or electronic component occur under this warranty contact your UNITRON® distributor or UNITRON® at (631) 543-2000. This warranty is limited to the continental United States of America. All items returned for warranty repair must be sent freight prepaid and insured to Unitron Ltd., 73 Mall Drive, Commack, NY 11725 – USA. All warranty repairs will be returned freight prepaid to any destination within the continental United States of America. For all foreign warranty repairs, return freight charges are the responsibility of the individual/company who returned the merchandise for repair.
UNITRON
®
73 Mall Drive, Commack, NY 11725 • 631-543-2000 • www.unitronusa.com 18
Loading...