EXAMET-4
MICROSCOPE SERIES
MANUAL
73 Mall Drive, Commack, NY 11725 • 631-543-2000 (P) • 631-589-6975 (F)
www.unitronusa.com • info@unitronusa.com
EXAMET-4 SERIES MICROSCOPE
CONTENTS
SAFETY NOTES .................................................................................................................. 3
CARE AND MAINTENANCE ................................................................................................ 3
INTRODUCTION .................................................................................................................. 4
UNPACKING AND COMPONENTS ..................................................................................... 4
COMPONENTS DIAGRAMS ............................................................................................. 5-8
ASSEMBLY .......................................................................................................................... 9
DETAILED ASSEMBLY PROCEDURE ......................................................................... 10-12
ADJUSMENT AND OPERATION
ILLUMINATION ......................................................................................................................... 13
SELECTING THE LIGHT PATH ................................................................................................13
FOCUSING ...............................................................................................................................14
ADJUSTING THE BRACKET ....................................................................................................14
ADJUSTING THE FOCUSING TENSION .................................................................................14
ADJUSTING THE DIOPTER .....................................................................................................15
ADJUSTING INTERPUPILLARY DISTANCE ............................................................................15
ADJUSTING THE STAGE .........................................................................................................16
ADJUSTING THE FIELD DIAPHRAGM (Reflected) ..................................................................16
ADJUSTING THE APERTURE DIAPHRAGM (Reflected) .........................................................17
ADJUSTING THE OBLIQUE ILLUMINATION (Reflected) .........................................................17
USING THE COLOR FILTER (Reflected) .................................................................................18
USING THE SIMPLE POLARIZER (Reflected) ........................................................................18
CENTERING THE CONDENSER (Transmitted) ......................................................................19
ADJUSTING THE APERTURE DIAPHRAGM(Transmitted) ......................................................20
CENTERING THE FILAMENT (Transmitted) ...........................................................................20
USING THE COLOR FILTER (Transmitted) ..............................................................................21
REPLACING THE FUSE(Transmitted) .....................................................................................21
INSTALLING & USING THE C-MOUNT ADAPTER ..................................................................21
SPECIFICATIONS .............................................................................................................. 22
TROUBLESHOOTING ................................................................................................... 23-24
MAINTENANCE.................................................................................................................. 25
SERVICE ............................................................................................................................ 25
WARRANTY ....................................................................................................................... 25
UNITRON
®
73 Mall Drive, Commack, NY 11725 • 631-543-2000 • www.unitronusa.com 2
EXAMET-4 SERIES MICROSCOPE
SAFETY NOTES
1. Open the shipping carton carefully to prevent any accessory, i.e. objectives or eyepieces, from
dropping and being damaged.
2. Keep the instrument out of direct sunlight, high temperature or humidity, and dusty environments.
3. If any specimen solutions or other liquids splash onto the stage, objective or any other component,
disconnect the power cord immediately and wipe up the spillage. Otherwise, the instrument may be
damaged.
4. LAMP REPLACEMENT -- CAUTION: the glass envelope of the lamp may be extremely
hot. DO NOT attempt to change the lamp before it is completely cooled or without
5. All electrical connectors (power cord) should be inserted into an electrical surge protector to prevent
damage due to voltage fluctuations.
6. FUSE REPLACEMENT -- For safety when replacing the fuse (ONLY replace with the
7. Confirm that the input voltage indicated on your microscope corresponds to your line voltage. The
use of a different input voltage other than indicated will cause severe damage to the microscope.
NOTE: Always plug the stereomicroscope power cord into a suitable grounded electrical outlet. A
grounded 3-wire cord is provided.
wearing adequate skin protection.
same size, type and rating of original fuse), be sure the main switch is in the off position,
disconnect the power cord from outlet, and replace the fuse. Reconnect the power cord
and turn unit on.
CARE AND MAINTENANCE
1. Do not attempt to disassemble any component including eyepieces, objectives or focusing assembly.
2. Keep the instrument clean; remove dirt and debris regularly. Accumulated dirt on metal surfaces
should be cleaned with a damp cloth. More persistent dirt should be removed using a mild soap
solution. Do not use organic solvents for cleansing.
3. The outer surface of the optics should be inspected and cleaned periodically using an air stream from
an air bulb. If dirt remains on the optical surface, use a soft, lint free cloth or cotton swab dampened
with a lens cleaning solution (available at camera stores). All optical lenses should be swabbed using
a circular motion. A small amount of absorbent cotton wound on the end of a tapered stick makes a
useful tool for cleaning recessed optical surfaces. Avoid using an excessive amount of solvents as
this may cause problems with optical coatings or cemented optics or the flowing solvent may pick up
grease making cleaning more difficult.
4. Store the instrument in a cool, dry environment. Cover the microscope with the dust cover when not
in use.
5. UNITRON® microscopes are precision instruments which require periodic servicing to maintain
proper performance and to compensate for normal wear. A regular schedule of preventative
maintenance by qualified personnel is highly recommended. Your authorized UNITRON® distributor
can arrange for this service.
UNITRON
®
73 Mall Drive, Commack, NY 11725 • 631-543-2000 • www.unitronusa.com 3
EXAMET-4 SERIES MICROSCOPE
Reflected
Trinocular
#14250
Camera is Optional
Reflected/
Transmitted
#14251
Reflected/
Transmitted
Trinocular
#14251
Camera is optional
INTRODUCTION
Congratulations on the purchase of your new UNITRON® microscope. UNITRON® microscopes are
engineered and manufactured to the highest quality standards. Your microscope will last a lifetime if
used and maintained properly. UNITRON® microscopes are carefully assembled, inspected and tested
by our staff of trained technicians in our New York facility. Careful quality control procedures ensure each
microscope is of the highest quality prior to shipment.
UNPACKING AND COMPONENTS
Your microscope arrived packed in a molded shipping carton. Do not discard the carton: the shipping
carton should be retained for reshipment of your microscope if needed. Avoid placing the microscope in
dusty surroundings or in high temperature or humid areas as mold and mildew can form. Carefully
remove the microscope from the shipping carton and place the microscope on a flat, vibration-free
surface.
MODELS
UNITRON
®
73 Mall Drive, Commack, NY 11725 • 631-543-2000 • www.unitronusa.com 4
EXAMET-4 SERIES MICROSCOPE
Oblique
Illumination Lever
Variable Light
Intensity Knob
COMPONENTS DIAGRAM 1 – Reflected Model
UNITRON
®
73 Mall Drive, Commack, NY 11725 • 631-543-2000 • www.unitronusa.com 5
EXAMET-4 SERIES MICROSCOPE
Field Diaphragm
Adjustment Lever
Aperture Diaphragm
Adjustment Knob
Aperture Diaphragm
Centering Knob
Field Diaphragm
Centering Lever
Tension
Adjustment
Collar
Beam Splitter
For Camera Port
Lever
COMPONENTS DIAGRAM 2 – Reflected Model
UNITRON
®
73 Mall Drive, Commack, NY 11725 • 631-543-2000 • www.unitronusa.com 6
EXAMET-4 SERIES MICROSCOPE
Oblique
Illumination Lever
Coarse
Adjustment
Limit Knob
Variable Light
Intensity Knob
Field Diaphragm
Adjustment Ring
Condenser Adjustment Knob
COMPONENTS DIAGRAM 3 – Reflected &Transmitted Model
UNITRON
®
73 Mall Drive, Commack, NY 11725 • 631-543-2000 • www.unitronusa.com 7
EXAMET-4 SERIES MICROSCOPE
Field Diaphragm
Adjustment Lever
Aperture Diaphragm
Adjustment Lever
Aperture Diaphragm
Centering Knob
Field Diaphragm
Centering Knob
Variable Light
Intensity Knob
Neutral Density
Control Knob
Tension
Adjustment
Collar
Beam Splitter
For Camera Port
Lever
Condenser Centering
Adjustment Screws
COMPONENTS DIAGRAM 4 – Reflected &Transmitted Model
UNITRON
®
73 Mall Drive, Commack, NY 11725 • 631-543-2000 • www.unitronusa.com 8