UNI-TREND UTD 2025 CL Instructions

Page 1
PN:110401105098X
MAY.2018 REV. 4
Page 2
Introduction
Dear users:
read this Operating Manual carefully and especially its “Safety Notice” before use.
Afer reading it, you are suggested to keep the manual properly. Please keep it with the
instrument together or place it in an accessible location for future use.
Page 3
General Safety Overview
The instrument is designed and manufactured strictly
according to “Safety Requirements of Electronic
Measuring Instruments” (GB4793) and IEC61010-1
Safety Standard . It conforms to “Insulation Overvoltage
Standard” (CAT 600V) and Safety Standard with
Pollution Level You are expected to know the following
safety precaution measures to avoid personal injury or
product damage or any other product connected to it.
Please use the product according to the regulations so
as to avoid the possible damage.
Only personnel with professional training can execute
maintenance procedure to avo
id fire or personal
injury.Appropriate power wires should be used and
special power wires must be approved by the state.
Reliable grounding of products: The product is
grounded by ground conductor of power supply.
Ground conductor must be connected to earth to
avoid electric shock. Please guarantee correct
grounding before connecting input/output terminal of
the product.
Correct connection of oscillograph probe: Earth wire
of probe is the same to earth potential. Please do not
connect earth wire to high voltage.
Review of rated values of all terminals: Please refer to
all rated values a
nd marks onthe product surface to
avoid fire or excessive current shock. Please refer to
product manual before connecting products so as to
know details of rated values.Do not open cover plate
of machine box: It is not allowed to start or operate the
product when opening external cover or panel.
Correct plug-in/out: Please do not plug in/out when
connecting probe or testing conduits to voltage
source.
Operating Manual for UTD2000/3000
Page 4
Operating Manual for UTD2000/3000
Usage of appropriate fuse: It is only allowed to use
fuse type and rated indicator designated by the
product.
To avoid exposed circuits: It is not allowed to touch
exposed connector or element after power supply.
Operation is not allowed if there is any doubt for
product trouble: Please consult qualified maintenance
personnel to inspect the product if there is any trouble
doubt.
It has to maintain appropriate ventilation. It is not
allowed to operate in wet locations. It is not allowed
to operate in combustible and explosive
environment. It has to keep product surface clean
and dry.
Page 5
Terms in the manual: The following terms may be
seen in the manual:
Terms for products: The following terms may be
Danger: It refers to direct injury near the mark.
Warning: It refersto potential injury near the mark.
Note: It refers to potential danger for the product
and other properties.
Warning: Warning statement means the
conditions and behaviors that may endanger lives.
Note: Notice statement means the conditions
and behaviors that may cause other property
damage
Protective
ground terminal
Ground terminal
for chassis
Ground terminal
for testing
Caution!
Refer to manual
High voltage
Safety Terms and Symbols
Product symbol: Following symbols may be shown
on the product surface.
Operating Manual for UTD2000/3000
Page 6
Operating Manual for UTD2000/3000
The manual is about operation information of OPERATING MANUAL FOR UTD2000/3000 digital storage
oscillograph. The manual includes the following chapters
Chapter 1: User Guideline: It is to introduce functions of digital storage oscillograph easily and provide
installation guidelines.
Chapter 2: Instrument Setup:Operation methods of OPERATING MANUAL FOR UTD2000/3000 digital
storage oscillograph.
Chapter 3: Application example: Examples to solve various measurement problems
Chapter 4: System Prompt andTroubleshooting
Chapter 5: Appendix: Technical Indicators
Appendix B
: Fittings of OPERATING MANUAL FOR UTD2000/3000 series digital storage oscillograph
Appendix C: Maintenance and Cleaning Maintenance
Preface
Page 7
Th e manual includes 3 serie s of digital storage oscillograph as follows. Detailed models are shown in the
following for m:
Operating Manual for UTD2000/3000
UTD2 C/UTD3025C
025
UTD2 C/UTD3042C
042
UTD2 C/UTD3062C
062
UTD2 C/UTD3082C
082
UTD2 C/UTD3102C
102
UTD2 C/UTD3152C
152
UTD2 C/UTD3202C
202
UTD2 CE/UTD3042CE
042
UTD2 CE/UTD3062CE
062
UTD2 CE/UTD3082CE
082
UTD2 CE/UTD3102CE
102
UTD2 CE/UTD3152CE
152
UTD2202CE/UTD3202CE
UTD2025CL
UTD2052CL
UTD2052CEX
UTD2072CEX
25MHz
40MHz
60MHz
80MHz
100MHz
150MHz
200MHz
40MHz
60MHz
80MHz
100MHz
150MHz
200MHz
25MHz
50MHz
50MHz
70MHz
250MS/s
500MS/s
500MS/s
500MS/s
500MS/s
500MS/s
500MS/s
1GS/s
1GS/s
1GS/s
1GS/s
1GS/s
1GS/s
1GS/s
1GS/s
250MS/s
500MS/s
UTD2000C/3000C
UTD2000E/3000E
UTD2000 L
UTD2000 EX
UTD2000/3000
Series
Model
Band wi dth
Real-time
UTD2102CEX
100MHz
1GS/s
Page 8
Operating Manual for UTD2000/3000
UTD2000/3000 series digital storage oscillograph
provides simple front panel with clear functions to
users for all basic operations. Scale and position of
various channels can provide intuitional operation
to meet usage habits of traditional instruments.
Users can use it skillfully without learning and
being familiar with operation for a long time. Users
can also press AUTO key directly to accelerate
adjustment and facilitate measurement.
Appropriate waveform and gear setup can be
shown on the instrument panel.
UTD2000/3000 series digital storage oscillograph
also has high-performance indicators and powerful
functions to fulfill measurement tasks quickly
besides easy usage. Quicker signals can be
observed by UTD2000/3000 series digital storage
oscillograph by real-time sampling of 500MS/s (or
1GS/s) and equivalent sampling of 25GS/s (or
50GS/s). Itcan capture and analyze waveform
easily by powerful trigger and analysis capacity.
Users can also observe and analyze signal
problems quickly and clearly by clear LCD display
and mathematical calculation function. We can see
that such ser
ies of digital storage oscillograph can
meet your measurement requirements by following
performance characteristics:
Double analog channels
High-resolution colorful LCD display system with
resolution of 320 240 or 800 480
To support plug and play USB storage instrument
to communicatewith computer
Automatic waveform and status setup
Waveform,setup, bitmap storage, waveform and
setup reoccurrence
●●●●●
× ( × )
Page 9
●●●●●●●●●●●
×,÷)
, ;
.;“ ”;
Elaborate view window extension function for
delicate analysis of waveform details and profile
Automatic measurement of 28 kinds of
waveform parameters
Automatic cursor tracing and measurement
function
Unique waveform recording and playback
functions
Embedded FFT
Multiple mathematical calculation functions
(including +, -, for waveform
Edge, video, pulse width andALTTRIG
functions
Multi-language menu display
Fittings ofUTD2000/3000 series digital storage
oscillograph
2 probes of 1.2m 1:1/10:1 please refer to
manual of probe Fittings of probe details so as
to meet EN61010-
031:2008 standard
1 power wire to meet national standards
1 Operating Manual
1 Product Warranty
USB connecting wire: UT-D06 or UT-D05
Communication control software for
UTD2000/3000 series oscillograph
Operating Manual for UTD2000/3000
Page 10
Item
Elementary understanding ofTrigger System...........................................
Page
...........................................................................................
..........................................................................................
.......................................................................................
.............................................................
Elementary understanding of Vertical System................................................
Elementary understanding ofHorizontal System...........................................
General Safety Overview
Preface
Setup of Vertical System.........................................................................
Setup of Horizontal System......................................................................
Setup of Trigger System...........................................................................
General Inspection
Function Inspection
Probe Compensation
AutomaticSetup of Waveform Display
Chapter 1: User Guideline.............................................................................
......
55899111415162731
Operating Manual for UTD2000/3000
Contents
Chapter 2: Instrument Setup
.......................................................................
1
Page 11
Page
....................................................................................
...........................................................................................
Setup of Auxiliary Functions..........................................................................
....
Automatic Measurement...............................................................................
.....
Cursor Measurement..................................................................................
.......
...................................................................................................
.
.....................................................
..............
...................................................................
Setup of Display System
Storage and Callout
Operation Key
Chapter 3: Application Examples
Example 1: Measurement of Simple Signals
Example 2: Delay Observation of Sine Signals by Circuit..............................
Example 3: Capture of Single Signal.............................................................
Example 4: Reduction of Random Noise for Signals......................................
Example 5:Application Cursor Measurement................................................
Example 6:Application of X-Y Function
Example 7: Trigger of Video Signals
.........................................................
................................................
485055606870727273757678798144
Operating Manual for UTD2000/3000
Item
Setup of Sampling System
.................................................................
Page 12
Item
Page
........................................................
.............................................................................
.....................
......................................................
................................................................................................
.........................................
Examp
Examp : Usage of Upgrading Procedure of U disk
System Prompt Information
Troubleshooting
Appendix A: Technical Indicators.......................................................
Appendix B: Fittings of UTD2000/3000 Series Digital Storage Oscillograph...
Appendix C: Maintenance and Cleaning Maintenance................................
Example 10: Usage of Storage Function
Chapter 4: System Prompt and Troubleshooting
Chapter 5: Technical Indicators............................................................
Example 8: Pass/Fail Detection
Example 9
888991
91
104
10583848588
Operating Manual for UTD2000/3000
Page 13
Operating Manual for UTD2000/3000
Chapter 1: User Guideline
△△△
Function inspection
Elementary understanding of vertical system
Elementary understanding of horizontal system
UTD2000/3000 series digital storage oscillograph is
a small and handy desktop product. Basic test can be
done by providing convenient front panel to facilitate
operation for users.
UTD2000/3000 series are to provide simple front
panel with clear functions to users for basic
operation. Panel also includes knob and function key.
Knob functions are similar to those of other digital
storage oscillograph.
The chapter is to show execution of following steps
General inspection
5 keys in a row in the right side of display panel are
menu operation keys. (They are defined as F1 to F5
keys from top to b
ottom.)You can also set different
options for current menu by them. Other keys are
function keys. You can enter into different function
menus or gain specific function application directly
by them.
Elementary understanding of trigger system
Please know front operation panel firstly after
getting UTD2000/3000 series digital storage
oscillograph. The chapter is to introduce operation
and functions of front panel for
UTD2000/3000products so as to be familiar with
usage within the shortest period.
General inspection
△△△
Automatic setup of waveform display
Probe compensation
1
Page 14
Operating Manual for UTD2000/3000
2
5.7 inch front panel of UNI-T
-
probe compensation
signal output
vertical control
horizontal control
Operation
controls
multi-purpose
knob controller
common
menu
trigger control
EXT TRIG INPUT
analog signal input
interface
Page 15
Operating Manual for UTD2000/3000
3
probe compensation
signal output
vertical control
horizontal control
Operation
controls
multi-purpose
knob controller
common
menu
trigger control
EXT TRIG INPUT
analog signal input
interface
7-inch front panel of UNI-T
Page 16
Operating Manual for UTD2000/3000
4
display of horizontal trigger position
time display of central scale line
vertical scale
coefficient of
display channel
display setup of
main time base
waveform
display window
Menu varies with
different function
keys.
mark of
channel 2
display of trigger status
mark of
channel 1
Interface display diagram of digital storage oscillograph (i.e. UTD2000L series)
Page 17
Operating Manual for UTD2000/3000
5
General inspection
Function inspection
It is suggested to inspect the instrument by
following steps when getting a new UTD2000/3000
series digital storageoscillograph.
1. To check if there is damage due to transportation
or not.
Please replace carton or foaming plastic protection
cushion with severe damage instantly.
2. Please inspect fitting details and shown it in
“Fittings of UTD2000/3000 series digital Storage
Oscillograph” shown in the manual. You can check
if there are deficient fittings or not by referring to it.
Please contact UNI-T distributor or local agent if
there are deficient or damaged fittings.
3. Please ins
pect the entire machine and contact UNI-T
distributor or local agent for damaged appearance,
abnormal operation or failing to pass performance test.
Please pay attention to packages and report to
transportation department and UNI-T distributor if
machine damage is causedby transportation.
1. Connect the power supply
You can supply power to the machine and voltage is AC
100V to AC 240V. Frequency is 45Hz to 440Hz. It has
to press [UTILITY] menu and [F1] for self-correction
after power supply 30 minutesafter operation of hot
machine so that digital storage oscillograph operates
under the
best status. As shown in Figure 1 4, please
enter into next page and press [F1] to call out ex-factory
setup.
UNI-T will arrange maintenance or replacement.
Please carry out once fast function inspection to
check if the machine operates normally or not.
Please operate according to following steps:
Page 18
Warning: Please confirm safety grounding of digital
storage oscillograph to avoid dangers.
2. Signal access of digital storage oscillograph is
double-channel input for UTD2000/3000 series
digital storage oscillograph.There is also 1 EXT
TRIG INPUT channel. Please connect signals
according to following steps:
(As shown in Figure 1-5), connect probe of
digital storage oscillograph to CH1 input terminal
and set attenuation ratio switch on probe to 10×Á
Figure 1 5 Setup of attenuation ratio switch on
probe
-①.
Operating Manual for UTD2000/3000
6
Figure 1-4
power switch
function button
F1
MEASURE
CURSOR
DISPLAY
ACQUIRE
STORAGE
UTILITY
MULTIPURPOS E
SELECT
AUTO
CH1
CH2
F2
F3F4F5
POSITION
SCALE
SCALE
VOLTS DIV
/
VVm
POSITION
SEC DIV
/
SSLEVEL
MATH
FORCE
HELP
MENU
TRIG
MENU
HORI
VERTICAL
CH1
TRIGGER
HORIZONTAL
CH2
SET
Operating Manual for UTD2000/3000
Page 19
Operating Manual for UTD2000/3000
7
②.
In digital storage oscilloscope to set the probe
attenuation factor. The attenuation coefficient of
vertical gear ratio change instruments, which
makes measurement resultsSetup method of
attenuation coefficient of probe is as follows to
show amplitude of measured signals correctly.
Please press
Figure 1-6 Setup of deflection coefficient of probe
on digital storage oscillograph
F4 key for menu display of 10×
Connect probe and grounding clip to
corresponding connection ends then press [AUTO]
button.As shown in Figure 1-7, square wave (1kHz,
about 3V and peak value) can be displayed within
several seconds. It has to inspect CH2 by the same
method then press [OFF] function button so as to
power off CH1. It has to press [CH2] function button
to power on CH2 and repeat the second and third
steps.
③.
Proberation
Figure 1 7
Page 20
Operating Manual for UTD2000/3000
Probe compensation
It needs to adjust it when connecting probe to any
input channel for the first time so that probe can be
appropriate to input channel. Probe without
compensation correction can also bring in
measurement tolerance or error. Please operate
according to following steps for adjustment of
probe compensation:
1. To set probe menu attenuation coefficient to
10× and place probe switch to 10× then connect
digital storage oscillograph probe to CH1. Please
guarantee reliable contact to probe. It has to
connect probe end to signaloutput connector
2ofprobe compensator then connect grounding clip to
earth wire connector of probe compensator. It also
has to power on CH1 and press [AUTO].
3. Adjust variable capacitance on the probe by
screwdriver of non-metallic handle for waveform
display of “insufficient compensation” or
“excessive compensation” shown in above figure
until “correct compensation” for waveform display
on the screen as above.
Warning: Please guarantee perfect insulation
conduits of probe and do not contact metallicpart
of probe when connecting to HV source to avoid
electric shock of probe during HV measurement.
.
Observation of displayed waveform
8
Exceeding
compensation
Correct
compensation
Insufficient
compensation Figure 1 8
Compensation correction of probe
Page 21
Operating Manual for UTD2000/3000
9
Automatic setup of waveform display
UTD2000/3000 series digital storage oscillograph
has automatic setup functions. It can adjust vertical
deflection coefficient,scanning time base and
trigger mode automatically according to input
signal for the most appropriate waveform display.
Frequency of measured signals shall be more than
or equal to 50Hz and duty cycle shall exceed 1%
according to automatic setup requirements.
1. Connect measured signals to signal input
channel.
2. Press [AUTO] button so that digital storage
oscillograph can set vertical deflection coefficient,
scanning time base and trigger mode
automat
ically. They can be adjusted again after
careful observation and automatic setup so that
waveform display can realize the best effect.
As shown in the following figure, there are a series
of keys and knobs in vertical control area. Please
introduce usage of vertical setup gradually as
follows.
Automatic setup
Elementary understanding of vertical system
5.7-inch vertical
control panel of screen
7-inch vertical control
panel of screen
Figure 1 9 Vertical control area on the panel
Page 22
Operating Manual for UTD2000/3000
10
1. Use vertical position knob so that middle display
of waveform signal in the window. Vertical display
position of signal can be controlled by vertical
position knob.
Ground mark moves vertically with waveform when
rotating vertical position knob.
Measurement knack
You can observe difference between waveform
and signal ground to measure DC component
of signal quickly if adopting DC channel
coupling mode. DC components in the signal
will be filtered if adopting AC coupling mode
. You can also display AC component of signal
by higher sensitivity conveniently.
Vertical position of double analog channelsshall be
restored to short-cut key of zero pointnamely,
SET TO ZERO.
The key is to return vertical displacement/horizontal
displacement/hold off position to zero point
(or middle point).
Of which: UTD2000L /2000EX series will trigger
electric level Return to 50% of position.
2 Alter vertical setup and observe alternation of
status information. You can determine variation of any
vertical gear by status bar below waveform window.
Vertical gear of “volt/grid” can be altered by rotating
vertical scale knob to show corresponding variation for
channel gear display which corresponds to status bar.
User can press [CH1], [CH2] , [MATH] and [REF] to
display operation menu, mark, and waveform and
gear status information of corresponding channels.
.
Page 23
Operating Manual for UTD2000/3000
11
Users can also press [OFF] key to close currently
selected channels for UTD2000C/3000C and
UTD2000E/3000E oscillograph. They can also
click on CH1, CH2 and MATH keys twice to close
corresponding channels for UTD2000L/2000EX
oscillograph.
As shown in following figure, there is 1 key and 2
knobs in horizontal control area. Please be familiar
with setup of horizontal time base gradually as
follows:
Elementary understanding of horizontal system
Horizontal control
panel of 5.7-inch
Horizontal control
panel of 7-inch
Figure 1 10 Horizontal control area on the panel
Page 24
Operating Manual for UTD2000/3000
1. Use horizontal SCALE knob to alter horizontal
time base gear then observe alternationof status
information. To rotate horizontal SCALE knob to
alter “s/div” time base gear so as to show
corresponding variation of time base gear display
for corresponding channels of status bar.
Horizontal scanning speed steps from 2ns/div
s/div by way of 1 mode.
Note: Horizontal scanning time base gear varies
for UTD2000/3000 series digital storage
oscillograph due to different models.
﹡~- -
50 2 5
2. Use horizontal POSITION knob to adjust
horizontal position of signals in waveform window.
To use horizontal POSITION knob to control trigger
displacement of signals. Horizontal movement of
waveform with knob can be observed when rotating
horizontal POSITION knob for application for
trigger displacement.
3. Please press [MENU] button for 5.7-inch
oscillograph (and press [HORI MENU] button for 7-
inch oscillograph) to display Zoom menu. Please
press [F3] to open view window extension under
this menu then press [F1] to close view window
extension and return tomain time base. Hold off
time can be set under this menu.
12
Page 25
Operating Manual for UTD2000/3000
13
Trigger point can be restored to vertical middle
point quickly by short-cut key of [SET TO ZERO] so
as to restoring displacement of trigger point to
horizontal zero point short-cut key. Users can also
adjust horizontal position of signal in waveform
window by rotating horizontal knob.
Term explanation
Trigger point: It refers to position of actual trigger
point by comparing with middle point of storage
instrument. Users can move trigger horizontally by
rotating horizontal POSITION knob.
Hold off: It refers to interval to start up trigger
circuit again.
Hold off time can be set by rotating
multi-purpose
knob controller.
As shown in Figure 1 there is 1 knob and 3
(or 2) keys in control area of trigger menu. Please
be familiar with setup of trigger system gradually as
follows:
POSITION
Elementary understanding of trigger system
,
11
5.7-inch screen 7-inch screen
Figure 1-12
Trigger menu
Figure 1 Trigger menu on the panel
11
Page 26
1. To user trigger level knob to alter trigger level
and instruct trigger level line by seeing trigger mark
on the screen so that it moves vertically with knob
rotation. Variation of trigger level in lower part of
screen can be observed when moving trigger level
at the same time.
2. (As shown in Figure 1-12,) TRIGGER MENU can
be used to alter trigger setup.
User can press [F1] key to select“edge” trigger.
n trigger signal amplitude.
i
4.
To press FORCE button: To produce 1 trigger
signal forcefully for mainly application for normal
and single trigger mode.
Operating Manual for UTD2000/3000
14
They can press [F2]key to select“trigger source”
to CH1.
They can press [F3] key to set “gradient” of edge
type to ascending.
They can press [F4 ]key to set “trigger mode” to
automatic. They can also press F5 key to set
“trigger coupling” to AC.
They can press [50%] button for 5.7-inch oscillograph
and press [SET TO ZERO] key for 7-inch oscillograph.
It also has to set vertical middle point of trigger level
2.
Page 27
Operating Manual for UTD2000/3000
15
Chapter 2: Instrument Setup
By now, you have been familiar with operation of
vertical control area, horizontal control area and
trigger system menu for UTD2000/3000 series digital
storage oscillograph preliminarily. You should be
familiar with setup of digital storage oscillograph by
menu operation by introduction shown in previous
chapter. It is suggested to read Chapter 1 if you have
not been familiar with above operation and methods.
The chapter is mainly to introduce following themes
Set trigger systems of [TRIGGER LEVEL],
[MENU], ]50%] and] FORCE];
Set sampling mode of [ACQUIRE];
It is suggested to read the
chapter carefully to know
many measurement functions for UTD2000/3000
and system operation method.
:■■■■■■■■■■
Set vertical systems of [CH1], ]CH2],] MATH],
]REF] , ]OFF] ]VERTICALPOSITION] and
]VERTICAL SCALE];
Set horizontal systems of ]MENU] or ]HOR
MENU], ]HORIZONTAL POSITION] or
[HORIZONTALSCALE];
Set displaymode of [DISPLAY];
Store and call-out for [STORAGE];
Set auxiliary system of [UTILITY];
Automatic measurement of [MEASURE];
Cursor measurement of [CURSOR];
Use execution buttons of [AUTO] and
[RUN/STOP];
Page 28
Operating Manual for UTD2000/3000
Setup of vertical system
CH1 and CH2 channels and setup
Every channel has its own vertical menu. Every
item can be set individually according to different
channels. As shown in Form 2-1, the system can
display operation menu of CH1 or CH2 channel by
pressing CH1 or CH2 function key.
16
110100
1000
×××
×
Setup
Description
Function menu
Coupling
ACDCTo obstruct DC component
of input signal.
To cut off input signals by AC
and DC components of input
signals.
Grounding
Bandwidth
limitONOFF
To limit band width to 20MHz
so as to reduce display noise.
Full band width
Volt/grid
Coarse
tuning
To set vertical deflection
coefficient for coarse tuning
by 1-2-5 system. To refine
within setup scope of coarse
tuning for fine tuning so as to
improve vertical resolution.
Fine
tuning
Probe
To select one value according
to probe attenuation
coefficient to keep correct
vertical deflection coefficient.
There are totally 4 kinds,
namely, 1 , 10 , 100
and 1000 .
× × ×
×
Reverse phase
ON
OFF
To realize reverse direction
function for waveform;
To display waveform normally.
Form 2-1
Page 29
1
Measured signal is a sine signal with DC component
by taking exerting signal to CH1 channel as an
example.
To press F1 to selectAC and AC coupling mode. DC
component contained by measured signal will be
obstructed. Waveform display is shown in following
figure.
Press F1 to DC to pass through DC and AC
components of measured signals of CH1 channel.
Waveform display is also shown in following figure:
Figure 2 2 Synchronous display of DC andAC
components of signal
Press F1 to select grounding and set channels to
grounding mode. DC andAC components
contained by measured signals will be obstructed
.
Waveform display is shown in following figure:
displayed on the screen.
Setup of channel coupling
17
Operating Manual for UTD2000/3000
Figure 2 Jammed DC component of signal
1
Setup of
AC coupling
DCcoupling
setup
Page 30
Operating Manual for UTD2000/3000
(Note: Input signal still shall be connected to
channel circuit under such mode although
waveform is not
Figure 2 3 Synchronous obstruction of DC and
AC components of signal
18
Groundcoupling
setup
2.
Setup of channel bandwidth limit
Input one sine signal of about 40MHz for CH1 as an
example. It has to press [CH1] to open CH1 channel
then press [F2] to set bandwidth limit to OFF.
Channel band width is full band width to pass
through high-frequency components contained by
measured signal. Waveform display is also shown
in following figure.
Figure 2 Waveform display when closing
bandwidth limit
4
Bandwidth
suppression
20MHz
Bandwidthsuppressionicon
Page 31
Operating Manual for UTD2000/3000
19
Please press [F2] to set band width limitation to ON
for great attenuation of high-frequency component
and noise of more than 20MHz contained by
measured signals. Waveform display is also shown
in following figure.
Figure 2 5Waveform display when opening
bandwidth limit
It has to set probe attenuation coefficient in
operation menu of channel to coordinate with
attenuation coefficient setup of probe. Probe
coefficient in channel menu shall be set to 10× if
probe attenuation coefficient is 10:1. It has to
analogize to guarantee correct voltage.
Volt/grid gears of vertical deflection coefficient can
be divided into coarse tuning and fine tuning.
Please refer to following figure for setup and
vertical gear when using 10:1 probe.
3. Setup of multiplying factor for probe
Close
bandwidth
limit
Probe
attenuation
coefficient
Figure 2 Probe attenuation coefficient shown
in channelmenu
6
Page 32
Operating Manual for UTD2000/3000
4.
20
2 5
Regulation setup of vertical volt/grid Setup of reverse phase of waveform
Reverse phase of waveform:
:~5~- -
.
Volt/grid scope is 2mV/div
V/div (or 10V/div), or 1mV/div V/div with step
by way of 1 Deflection coefficient shall be
altered within current vertical gear scope by less
step during fine tuning so as to realize
uninterrupted and continuous adjustment of vertical
deflection coefficient within all vertical gears.
5
.
20
Setup of
fine tuning
Figure 2 Coarse tuning and fine tuning of
vertical deflection coefficient
7
It is to show overturning degree of phase location of
displayed signals. Please refer to Figure 2 for
waveform without reverse phase. Please refer to
Figure 2 for waveform with reverse phase.
89Without
reverse phase
Figure 2 Setup of reverse phase of vertical
channel (or without reverse phase)
8
Page 33
Operating Manual for UTD2000/3000
21
Figure 2 9 Setup of reverse phase of vertical
channel
I. Realization of mathematical calculation
function
Mathematical calculation functions are to show +, -,
, / as well as FFT calculation results for CH1 and
CH2 channel waveform. Menu is shown as follows:
×
Reverse phase
of waveform
Mathematical calculation
Figure 2 Mathematical calculation
10
Page 34
Operating Manual for UTD2000/3000
Analysis of FFT frequency spectrum
Time domain signal can be converted to frequency
domain signal when using mathematical calculation
of fast Fourier transtable (FFT). Following types of
signals can be observed conveniently by using
FFT
Measure harmonic content and distortion within
measurement system;
●●●
Show noise characteristics in DC power;
Analyze vibration;
22
Function Menu
Setup
Description
Type
Math
+ -
, , ,
× ÷
Information
source
CH1
CH2
To set information source 1
as Ch1 channel waveform.
To set information source 1
as Ch2 channel waveform.
+
-
Information source 1
Information source 2
Information source 1
Information source 2
Information source 1
Information source 2
Information source 1
Information source 2
+-×
÷
Operator
Information
source 2
CH1
CH2
To set information source 2
as CH1 channel waveform.
To set information source 2
as CH2 channel waveform.
Form 2-2: Introduction to mathematical calculation
×
÷
Page 35
Form 2-3: FFT menu Description
FFT operation knack:
Signals with DC component or deviation will bring
in error or deviation of FFT waveform component.
AC coupling mode can be selected to reduce DC
component.
Average obtainment mode of digital storage
oscillograph can be set to reduce random noise
and aliasing frequencycomponent of repeated or
single pulse event.
Operating Manual for UTD2000/3000
23
Function
Setup
Description
Type
FFT
FFT mathematical
calculation
Information
source
CH1
CH2
To set CH1and CH2 as
calculation waveform.
Window
Hamming
Blackman
Rectangle
Set Hanning window
function
Set Hamming window
function
Set Blackman window
function
Set Rectangle window
function
Hanning
Vertical unit
Vrms
dBVrm
To set vertical unit to
Vrms or dBVrms.
Page 36
Operating Manual for UTD2000/3000
Selection of FFT window
Mathematical storage oscillograph will carry out FFT conversion for time record with limited length when
assuming continuous repetition of YT waveform.YT waveform will have the same amplitude in beginning
and ending locations for no interrupation. However, high-frequency transient interruption will be caused in
connector due to different waveform amplitudes for non-integral cycle of YTwaveform. Such effect shall be
called as leakage infrequency domain. Therefore, value in forced beginning and ending locations is 0 by
multiplying original waveform by a window function to avoid
leakage. Please refer to following form for
application of window function.
Form 2-4
24
FFT window
Characteristics
The most appropriate measurement content
Rectangle
The best frequency and the poorest
amplitude resolution to be basically
similar to circumstances without window;
Transient or short impulse with basically equivalent
signal level; Sine with equivalent amplitude with
similar frequency; Wide band random noise of wave
spectrum with slow variation;
Hanning
With excellent frequency resolution and
poor amplitude resolution by comparing
with rectangular window;
Sine, cycle and narrow band random noise
Hamming
Frequency resolution of Hamming window
is better than that of Hanning window slightly.
Transient or short impulse with large difference
for signal level;
Blackman
The best amplitude resolution and the
poorest frequency resolution
Mainly used for single-frequency signal to search
for higher order of harmonic wave.
Page 37
Operating Manual for UTD2000/3000
25
Term explanation:
FFT resolution:
Nyquist frequency:
Reference waveform
It is defined as factor of sampling
and calculation point. FFT resolution increases
when sampling rate decreases for fixed calculation
point quantity.
Original waveform can not be
reset until using sampling rate of at least 2f for
waveform with the maximum frequency of “f”. It is
also called as Nyquist guideline. “f” refers to Nyquist
frequency and “2f” refers to Nyquist sampling rate.
Please refer to waveform stored in nonvolatile
memory (NVM) of digital storage oscillograph or
peripheral U disk with name of RefA and RefB as
follows. Please operate according to following
steps to call out or close reference waveform.
5. Please press REF menu key on the front panel
for 5.7-screen oscillograph
Please press [STORAGE]key and enter into the
second page to select call-out for 7-inch
oscillograph (namely, UTD2000L/2000EX series).
2. Please press [RefA] (RefA reference option) and
select informationsource then select its position by
rotating multi-purpose knob in upper p
art of front
panel. 1 10 (or 1 20) can be selected for the
position. Press “call-out” to call out waveform
which is stored inthe position originally after
selecting a certain position such as “1”. Please
plug in U disk if waveform is stored in the U disk.
There are 2 options for magnetic disk as follows:
DSO/USB; Please press [F2] key and select USB.
(The menu can not be activated until plugging in U
disk.) Called out waveform is displayed on the
screen for waveform called out and stored from U
disk. Please press cancel key (F5) Return to
previous menu after waveform call-out.
3. To press [RefB] (RefB r
eference option) and
select
Page 38
information source 2 which participates in
calculation and method is the same to the
secondstep. Please measure and observe related
waveform by UTD2000/3000 series digital storage
oscillograph during actual application. Please
compare current waveform with reference
waveform for analysis. Please press REF key to
display reference waveform menu. Setup is also
shown in following form.
Form 2-5 Selection of storage position
It is allowed to select from 1 to 20 for internal
storage position. Please plug in U disk and press
F2 to select USB magnetic disk for external storage
instrument. Pl
ease refer to STORAGE menu for
waveform storage.
Operating Manual for UTD2000/3000
26
Function menu
Setup
Description
Storage position
1 20
~
1 20
~
refer to positions of
20 groups of waveform
respectively;
There are 200 groups of
waveform positions for
storage to USB;
(Storage position and quantity
vary withmachine model.)
Magnetic disk
DSO
USB
To select internal storage position;
To select external storageposition
(until plugging inU disk);
Close
To close called out waveform;
Callout
To call out selected waveform;
Cancel
Return to previous menu;
Page 39
Operating Manual for UTD2000/3000
27
Setup of horizontal system
Horizontal control knob
Control knob on horizontal surfacecan be used to
alter horizontal scale(or time base) to trigger
horizontal position (or trigger position) shown in
RAM. Vertical middle point toward horizontal
direction on the screen is time reference point for
waveform. Horizontal scale can be altered for
waveform extension or shrinkage by comparing
with screen center. Horizontal position can also
alter by comparing with position of waveform
trigger point.
Horizontal position: To adjust horizontal position of
channel waveform (including mathematical
calculation). Res
olution of suchcontrol key varies
according to time base.
Horizontal scale: To adjust main time base, namely,
s/div. Window width can vary with delay scanning
time base by rotating horizontal scale knob when
opening extension time base. Please refer to
introduction to extension time base for details. (As
shown in the following form), horizontal menu can
be displayed for horizontal controlkey menu.
Form 2-6
Function menu
Setup
Description
Main time base
1.2.To open main time base;
To close view window
extension according to main
time base after opening;
View window
extension
To open extension time base;
Hold off
To regulate inhibition period;
——
——
——
——
——
Page 40
Mark Description:
Trigger horizontal distance of position by
comparing with middle point of view window.
①.②.③
Mark represents position of current waveform
view window inRAM
Mark position of trigger point in RAM.
Mark position of trigger point in current
waveform view window.
Display horizontal time base (or main time
base), namely, s/div.
Term explanation:
Y-T mode:
X-Y mode:
Y axle refers to voltage and X axle
refers to time quantum under this mode.
X axle refers to CH1 voltage and Y axle
refers to CH2 voltage under this mode.
Slow scanning mode: Instrument will enter into
slow scanning sampling mode when setting
horizontal time base control to 100ms/div or less. It
is suggested to set channel coupling to DC when
observing low-frequency signal by applying for
slow scanning mode.
s-div: Time base controlcan expand or compress
waveform for horizontal scale (or time base) unit if
stopping waveform sampling (by using RUN/STOP
key).
Operating Manual for UTD2000/3000
28
Figure 2-11 Interface of horizontal system
Page 41
Operating Manual for UTD2000/3000
29
View window extension
View window extension is to argument a section of
waveform to review image details conveniently.
Setup of view window extension can not be slower
than that of main time base.
As shown in above figure), it can be divided into 2
display areas under extension time base. Original
waveform is displayed for upper part. Selected area
can move horizontally by rotating horizontal
POSITION knob or increase or decrease by rotating
horizontal SCALE knob.
(
Waveform with horizontal extension in originally
selected waveform area is shown in lower part. (As
shown in above figure,) please pay attention that
resolution increases for extension time base by
comparing with main time base. Waveform displayed
by the entire lower party corresponds to area
selected for upper part so users can rotate horizontal
SCALE knob to reduce selected area and increase
extension time base, namely, horizontal extension
multiple of waveform.
The mode must be used for CH1 and CH2 at the
same time.
CH1 voltage will be displayed on X axle and CH2
voltage
Extension
time base
Horizontal extension
of the waveform
Horizontally extended part of the waveform
Figure 2 1 Screen display under view window
extension
2
Page 42
Operating Manual for UTD2000/3000
will be displayed on Y axle after selecting X-Y
display mode.
Note: Digital storage oscillograph can capture
waveform by anysampling speed under normal X-Y
mode. Similarly, it is allowed to adjust sampling
rate, vertical gear of channels and time base gear
under X-Y mode. Default sampling rate is 100MS/s
under X-Y mode. In general, sampling rate can
reduce to gain excellent Lissajou’s figure. Following
functions do not effect during X-Ydisplay.
Automatic testingmode
Cursor measurement mode
Reference or mathematical calculation waveform
View window extension function
Trigger cont
rol
■■■■■
30
Figure 2 Waveform display under X-Ymode
13
Page 43
Operating Manual for UTD2000/3000
31
Setup of trigger system
Trigger can determine data acquisition time and
waveform display for digital storage oscillograph. It
can convert instable display to meaningful
waveform once correct setup. Digital storage
oscillograph will collect sufficient data to map
waveform in the left side of trigger point firstly. It
will acquire data continuously and synchronously
when waiting for trigger conditions. It also will
acquire sufficient data continuously to map
waveform in the right side of trigger point. Trigger
control area of operation panel of digital storage
oscillograph includes trigger level adjustm
ent knob
and trigger menu key of [TRIG MENU];
5.7-inch oscillograph will set 50% of trigger level in
vertical middle point of signal (and realize
UTD2000L /2000EX by press [SETTO ZERO] key).
It has to press forced trigger key of [FORCE].
Trigger level:Trigger level will set signal voltage
which corresponds to trigger point.
50%: To set trigger level to vertical point of trigger
signal amplitude.
FORCE: To produce 1 trigger signal forcefully for
mainly application for normal and single modes
shown in trigger modes.
TRIG MENU: To trigger and set menu key.
Page 44
Operating Manual for UTD2000/3000
32
Trigger control
Trigger mode: Edge, pulse width, video andALTTRIG;
Edge trigger: Trigger when edge of trigger signal reaches a certain given level.
Pulse width trigger: Trigger when pulse width of trigger signal reaches certain set trigger conditions
Video trigger: Carry out field or linetrigger for standard video signals. (UTD2025CL/UTD2000EXseriesis
no such function.)
ALT TRIG: It is appropriate to trigger signal without frequency relevance. Various kinds of trigger menus
are also shown as follows:
Edge trigger:
Edge trigger mode is to trigger threshold value of input signal edge. That is, to tr
igger in ascending/
descending edge of input signal when selecting “edge trigger”.
Form 2-7
Page 45
Operating Manual for UTD2000/3000
33
Function menu
Setup
Description
Type
Edge
Selection of
information source
CH1
CH2
EXT
EXT/5
Municipal power
Alteration
Set CH1 as trigger signal of information source.
et CH2 as trigger signalof information source.
Set EXTTRIG INPUT channel as trigger signal of information source.
Divide outer trigger source by 5 to expand outer trigger level scope.
Set municipal power trigger
ALTTRIG of CH1 and CH2 signals;.
Gradient
Ascend
Descend
Ascend/descend
Set ascending edge trigger for signal;
Set descending edge trigger for signal;
Set ascending/descending edge trigger of signal;
Trigger mode
Automatic
Normal
Single
Set waveformacquisition without detecting trigger conditions;
Set waveformacquisition after meeting trigger conditions;
Set sampling of a waveform and shutdown after detecting once
trigger;
Trigger coupling
ACDCHigh-frequency
inhibition
Low-frequency
inhibition
Obstruct DC component of input signal;
Pass throughAC & DC components of input signal;
Inhibit high-frequency components of more than 80kHz for signal;
Inhibit low-frequency components ofless than 80kHz for signal
Page 46
Operating Manual for UTD2000/3000
Pulse width trigger
Trigger time shall be determined for pulse width trigger according to pulse width. You can capture
abnormal pulse by setting pulse width conditions.
Form 2-8 (the first page)
34
Function menu
Setup
Description
Type
Pulse width
Selection of
information
source
CH1
CH2
EXT
EXT/5
Municipal power
Alteration
To set CH1 as trigger signal of information source.
To set CH2 as trigger signal of information source.
To set EXT TRIG INPUT channel as trigger signal of information source.
To divide outer trigger source by 5 to expand outer trigger level scope.
To set municipal power for trigger.
ALT TRIG of CH1 and CH2 signals
Pulse width
conditions
More than
Less than
Equal to
To trigger when pulse width exceeds set value;
To trigger when pulse width is less than set value;
To trigger when pulse width is equivale
Pulse width
setup
To set pulse width of 20ns to 10s then regulate by multi-purpose knob in
upper part of front panel.
Next page
1/2
Enter into next page.
Page 47
Operating Manual for UTD2000/3000
35
Form 2-9 (the second page)
Function menu
Setup
Description
Type
Pulse width
Trigger polarity
Positive pulse width
Negative pulse width
Trigger mode
To set positive pulse width as trigger signal;
To set negative pulse width as trigger signal;
Automatic
Normal
Single
The system will acquire waveform data automatically
and display scanning base line on the screen without
trigger signal input. It will convert to trigger scanning
automatically when producing trigger signal.
To produce trigger scanning if any trigger signal is produced.
To produce once trigger then stop during trigger signal input.
Trigger coupling
DCACHigh-frequency
inhibition
High-frequency inhibitionTo pass through AC & DC
components of trigger signal.
To obstruct DC components of trigger signal.
To obstruct pass through of high-frequencycomponents of
signal and only low-frequency components can pass through.
Previous page
2/2
Return to previous page.
Page 48
Operating Manual for UTD2000/3000
Video touch
To trigger in field or row with NTSC or PALstandardvideo signals after selecting video trigger for
UTD2025CL oscillograph shown in 5.7-inch or 7-inch screen. Trigger coupling is pre-set to DC. Please
refer to following form for trigger menu.
Form 2-10 Setup of video trigger
36
Function menu
Setup
Description
Type
Selection of
Pulse width
information source
CH1
CH2
EXT
To set CH1 as trigger signal of information source.
To set CH2 as trigger signal of information source.
To set EXT TRIG INPUT channel as trigger signal of
information source.
To set 5 times of attenuation for outer trigger signalas trigger
signal.
ALT TRIG of CH1 and CH2 signals
EXT/5
Alteration
Standard
PAL
NTSC
Video signal to be appropriate to PALsystem;
Video signal to be appropriate to NTSC system;
Synchronization
All rows
Designated row
Odd field
Even field
To set synchronous trigger of video row.
To set synchronous trigger of designated video row for
adjustment by multifunctionalknob in upper part of front panel.
To set synchronous trigger for video odd field.
To set synchronous trigger for video even field.
Page 49
Operating Manual for UTD2000/3000
37
Please refer to Figure 2-14 for screen display when
selecting standard system of PALand row
synchronization as synchronization mode. Please
refer to Figure 2-15 for field synchronization of
synchronization mode.
Figure 2 14 Video trigger: Row synchronization
Figure 2 15 Video trigger: Field synchronization
ALT TRIG mode
ALT TRIGTrigger signal comes from 2 vertical
channels during ALT TRIG. Such trigger mode can be
adopted for synchronous observation of 2 signals
without regard to signal frequency. Please refer to
following figure for trigger alternation waveform
display.
Page 50
Setup of trigger coupling mode
Please enter into trigger setup menu for setup of
trigger coupling mode and gain the most sform
synchronization. Please refer to following form for
menu of trigger coupling mode.
Operating Manual for UTD2000/3000
Form 2-11 Trigger alternation menu
Function menu
Setup
Description
Type
Edge
Set trigger mode to edge
Trigger source
Alternation
ALTTRIG of CH1 and
CH2
Gradient
Ascending
Set trigger gradient to
ascending edge.
Trigger mode
Automatic
Set trigger mode as
automatic one
Trigger coupling
AC
Set trigger coupling
modeto AC mode
Figure 2 Observation of 2 signals with
different frequency by.
16
38
Page 51
Form 2-12
Operating Manual for UTD2000/3000
39
Function menu
Setup
Description
Type
Trigger source
Gradient
Trigger mode
Coupling
ACDCHigh-frequency inhibition
Low-frequencyinhibition
Set pass-through obstruction of DC components.
Set pass-through of all components.
Obstruct pass-through of high-frequency part of signal. Only low-
frequency components can pass through.
Obstruct pass-through of low-frequency part of signal. Only high-
frequency components can pass through.
——
——
——
——
Page 52
Adjustment of hold off time
Adjustment of hold off time can be used to observe complex waveform (such as pulse string series).
Inhibition time refers to waiting period of digital storage oscillograph to start up trigger circuit again.
Digital storage oscillograph will not trigger until inhibition time ends during inhibition period. For instance,
as shown in Figure 2-17, the first pulse among a group of pulse series shall be triggered to set inhibition
time to width of pulse string.
Please refer to following form for hold off menu.
Form 2-13
Operating Manual for UTD2000/3000
40
Function menu
Setup
Description
Main time base
1. To open main time base;
2. To close view windowextension according to main time base after opening
view window extension.
View window extension
To openextension time base.
Hold off
To adjust inhibition time.
——
——
——
——
Page 53
Operating Manual for UTD2000/3000
41
Figure 2—17 Hold off used for synchronization
of complex waveform
Operation Description
1. To select edge, trigger source and gradient shown
in TRIG MENU according to synchronization mode of
normal signal then adjust trigger level for stable
waveform display as practicable as possible.
2. To press HORI MENU key to display horizontal
menu.
3. To adjust multi-purpose knob in upper part of
panel to alter inhibition time for stable waveform
display as practicable as possible.
Term explanations
1. Trigger source:To gain trigger from many kinds of
information sources, namely, input channel(CH1 &
CH2), external trigger (EXT& EXT/5) and municipal
power.
Input channel The most common trigger
information source is optional input channel
Selected one is used as channel of trigger
information source for normal operation regardless
of input display
External trigger: Such trigger information source
can be used for trigger on the third information
channel during synchronous data acquisition on
channels For instance signals from external clock
or circuit to
meas
ure can be used as trigger information source.
EXT and EXT/5 trigger sources shall use external
trigger signals
:
( ) .
.■2
. ,
Trigger
Holdoff
Page 54
connected to EXT TRIG interface. Signal can be
used for EXT and EXT can be used within signal
trigger level scope of -3V to +3V.
Automatic trigger: The system will acquire
waveform data automatically so as to display
scanning base line on the screen without trigger
signal input. It will convert to trigger scanning
automatically for signal synchronization when
producing trigger signal.
It is allowed to select EXT/5 for oscillograph of 5.7-
inch screen. Outer trigger shall be divided by 5 so
that trigger scope can expand to -15V to +15V for
trigger of digital storage oscillograph under large
signal.
Municipalpower: Such trigger mode can be used
to observe signal with regard to municipal power.
For instance, it can provide relationship between
lighting equipment and power supply equipment so
as to gain stable synchronization.
2 Trigger mode: It is to determine behavior mode
of digital storage oscillograph without trigger event.
This digital storage oscillograph can also
provide 3
triggers, namely, automatic, normal and single
trigger modes.
■.■
Operating Manual for UTD2000/3000
42
Note: Without trigger signal is allowed under “
automatic” mode when setting scanning waveform to
5-ms/div or less time base.
Normal trigger: Digital storage oscillograph can
not acquire waveform until meeting trigger conditions
under ordinary trigger mode.Trigger scanning can be
produced when producing trigger signal.
Single trigger: User can press “operation” button
once under single trigger mode for entry and trigger
waiting for digital storage oscillograph. It will sample
and display all acquired waveform then stop after
digital storage oscillograph detects once trigger.
Page 55
Operating Manual for UTD2000/3000
43
3.
4. Pre-trigger/delay trigger: To trigger data acquired
in ahead/after event. Trigger position is often set in
the horizontal center on the screen. You can
observe pre-trigger and delay information of 5div
(or 6div).You can also rotate horizontal position to
regulate horizontal displacement of waveform and
review more pre-trigger details. You can observe
waveform in ahead of trigger by observing pre-
trigger data. For instance, burring reason can be
shown by observation and analysis on pre-trigger
data after capturing burring during circuit startup.
Trigger coupling: Trigger coupling can determine
category of signal component conveyed to trigger
circuit. Coupling types include DC, AC, low-
frequency inhibition and high-frequency inhibition.
Pass through all components of signals under
DC mode.
■■■
Obstruct DC components with signal attenuation
to be less than 400Hz underAC mode.
Obstruct DC components with attenuation of
low-frequency component of less than 80kHz under
“low-frequency inhibition” mode.
Attenuate high-frequency components of more
than 80kHz under “high-frequency inhibition” mode.
Page 56
Operating Manual for UTD2000/3000
Setup of sampling system
As shown in following figure,ACQUIRE in
control area is function key of sampling
system.
Figure 2 18 Function key of sampling system
User can use ACQUIRE key to pop out sampling
setup menu so as to adjust sampling mode by
control button shown in the menu.
44
Page 57
Operating Manual for UTD2000/3000
45
Form 2-14 Sampling menu
Function menu
Setup
Description
Obtainment mode
Sampling
Peak-value detection
Average
Activate ordinary sampling mode.
Activate peak-value detectionmode.
Set average sampling mode and display average times.
Average times
2~256
Set average times for step by multiples of 2, namely, 2, 4, 8, 16,
32, 64, 128 and 256.As shown in Figure 2-18, averagetimes can
alter by pressing left multi-purpose knob.
Sampling mode
Real-time
Equivalent
Set sampling mode to real-time sampling mode.
Set sampling mode to equivalent sampling mode.
Fast acquisition
ON
Acquire by higher screen refreshing rate to reflect dynamic
effect of waveform excellently.
Close fast acquisition.
Page 58
It has to observe waveform display variation by
altering obtainment mode. Please refer to Figure 2-
19 and 2-20 for sampled waveform display if signal
contains large noise when adopting 32 times of
average mode or without adopting average mode.
Operating Manual for UTD2000/3000
46
Figure 2-20 Waveform by adopting 32 times of average mode
Figure 2 Waveform without adopting
19
Note
Please select real-time sampling mode for observation
of single signal.
2. Please select equivalent sampling mode for observation
of high-frequency periodic signal.
3. Please select peak-value detection mode to avoid mixed
signal envelope. Please select average sampling mode with
step of multiples of 2 for average value to reduce random
noise of displayed signal.Average times can be selected
form 2 to 256.
.
1
Page 59
Term explanation
Real-time sampling: It refers to data required for once acquisition.
Equivalent sampling: It refers to repeated sampling mode. It can facilitate careful observation of repeated
periodic signals. It can be adopted to gain horizontal resolution which is higher than real-time sampling
resolution greatly.
Sampling mode: It can be classified into equivalent and real-time modes.
Peak-value detection mode:
Digital storage oscillograph will find out the maximum and minimum input signals within every sampling
interval for waveform display by using such values under suc
h obtainment mode. It can also obtain and
display narrow impulse or they may be omitted for larger noise under such mode. Average mode is as
follows: Digital storage oscillograph may obtain several waveform and calculate average value then
display final waveform under this obtainment mode. Such mode can be adopted to reduce random noise.
Operating Manual for UTD2000/3000
47
Page 60
Operating Manual for UTD2000/3000
Setup of display system
As shown in following figure, DISPLAY in control area is function key of display system
Figure 2 Function key of sampling system
Please press DISPLAY button to pop out setup menu shown in following figure. Display mode can be
adjusted by pressing menu control button.
2 1
48
Page 61
Operating Manual for UTD2000/3000
49
Key points:
Display type: To fill in blank between adjacent sampling points during vector display. Only sampling points
will be displayed for point display.
Function menu
Setup
Descriptions
Display type
Vector point
Only to display sampling points by wiring.
Format
YTXYOperation mode of digital storage oscillograph
X-Y display mode; X input for CH1 and Y input for CH2;
Continuous
Close
1s2s5s
Infinite
Real-time updating of screen waveform;
To update after keeping waveform data shown in the screen for 1s;
To update after keeping waveform data shown in the screen for 2s;
To update after keeping waveform data shown in the screen for 5s;
To always keep display of original waveform data shown in the screen
until this function is inactivated if new data is added into display
continuously;
Waveform brightness
1% - 100%
To set waveform brightness (without this function for
UTD2000C/3000C)
Form 2-15 Display menu
Page 62
Storage and callout
As shown in following figure, STORAGE in MENU
control area is function key of storage system.
Operation steps
Please press STORAGE Enter into type menus.
There are 3 types,namely, waveform, setup and
bitmap.
Please refer to Figure 2-16 for selection of type
waveform and entry into following waveform
storage menu. Please call out REF (reference
waveform shown in the second page) from Section
2 of the chapter after saving waveform.
Form 2-16 Waveform storage menu (the first page)
2.
Operating Manual for UTD2000/3000
50
Figure 2 22 Function key (storage) for sampling system
STORAGE key can be used to display storage setup menu.You
can save oscillograph waveform or setup status to internal
storage area or U disk to call out saved waveform by RefA
(or RefB) or call out setup status by STORAGE key. Waveform
display area of oscillograph can be saved to DSO contents in U
disk in a form of bitmap. Saved bitmap can be read by PC computer.
Page 63
Operating Manual for UTD2000/3000
51
Form 2-16
Function menu
Setup
Description
Type
Waveform
Select waveform storage and callout menu
Information source
CH1
CH2
Select waveform from CH1 channel.
Select waveform from CH2 channel.
Storage position
1 20
~
1 20
~
refer to positions to store 20 groups of waveform.
There are 200 groups of waveform positions for storage to USB.
(Storage position and quantity vary with machine model.)
Save
Save waveform
Next page
1/2
Enter into next page.
Figure 2-23 Waveform storage
Figure 2-24 Waveform storage by U disk
——
——
Page 64
Operating Manual for UTD2000/3000
Form 2-17 Storage menu (the second page)
52
Function menu
Setup
Description
Magnetic disk
DSO
USB
To select internal RAM of digital storage oscillograph then
select external U disk. (This function can not be used until
plugging in U disk.)
Storage depth
Ordinary
Long storage
To set storage depth to ordinary; (Callout only can be done in
REF area when saving data to U disk.)
To set storage depth to long storage. (Note: This function can
not be activated until plugging in U disk. Callout can not be
done until using loading function of computer communication
software or waveform analysisload when saving data to U disk.)
Previous page
2/2
Return to previous page.
——
Page 65
2
.
As shown in Form 2-18, please select setup and enter into setup storage menu.
Figure 2-25 Storage setup
Form 2-18 Setup of storage menu
53
Operating Manual for UTD2000/3000
Function menu
Setup
Description
Setup
To select panel setup menu.
Setupof (storage position)
1 20
Operation setup to save 20 groups of panels: To select by multi-
purpose knob inupper part of front panel. (Position and quantity
vary with machine model
Save
Storage setup
Recall
Callout setup
Page 66
54
3.
As shown in Form 2-19, please select bitmap for entry into bitmap storage menu.
Remark:This function can not be called out until plugging in U disk.
Form 2-19 Bitmap storage menu
Operating Manual for UTD2000/3000
Function menu
Setup
Description
Bitmap
Select bitmap menu
Storage position
1 20
0
200
bitmap data can be saved by selecting multi-purpose knob in
upper part of front panel. (It is 1 to 10 for UTD2000C series.)
Save
Save bitmap data
——
——
Figure 2-26 Bitmap storage
Page 67
Setup of auxiliary functions
As shown in following figure, UTILITY in MENU
control area is function key.
Figure 2 27 Function key (and function) of
sampling
User can press UTILITY key to pop out setup menu
of auxiliary system function.
Form 2-20 (the first page)
55
Operating Manual for UTD2000/3000
Function menu
Setup
Description
Self-correction
Cancel
To carry out self-correction.
To cancel self-correction and
return to previous page.
Execution
By detection
Please refer
to Form 2-23
To set waveform Pass/Fail.
(Only UTD2000E/3000E,
UTD2025C/3025C and
UTD2025CL have this
function.)
Waveform
recording
Please refer
to Form 2-22
Set waveform recording
operation.
Language
Multilanguage
Select interface language.
Page 68
56
Operating Manual for UTD2000/3000
Form 2-21-1 (the second page)
Function menu
Function menu
Setup
Setup
Description
Description
Fast correction
ON/OFF
Activate/inactivate fast correction function of channels within scope
of 2mV/div to 10 mV/div. (Only UTD2025C/3025C has this function.)
Interface style
The first style
The secondstyle
The third style
The fourth style
Set 4 interface styles for digital storage oscillograph.
Gridding brightness
1% - 100%
Adjust gridding brightness of the screen by multi-purpose knob.
UTD2000C/3000C also has no this function.
Next page
2/3
Enter into next page.
Form 2-21-2 (the third page)
——
System information
Display information of current oscillograph system.
Frequency meter
Activate/inactivate function of frequency meter.
The first page
3/3
Return to the first page.
Page 69
Form 2-22 Waveform recording menu
57
Operating Manual for UTD2000/3000
Function menu
Setup
Description
Information source
CH1
CH2
CH1+CH2
Select CH1 as recording signal source.
Select CH2 as recording signal source.
Select ct CH1+CH2 as recording signal source.
Operation
Record by pressing recording key for synchronous display of
recorded quantity in lower part of the screen.
1.
Playback key;
2. Press the key for playback and display of current playback
screen No. in lower right corner of screen. Playback can be
stopped by rotating multi-purpose knob controller in upper part
of panel. However, waveform of a certain screen can be selected
for repeated playback by continuing to rotate it.
3
.
Press to stop then press for playback of all items.
●■■
Stop recording
Save
1 ~ 200
Save newly recorded waveform for U disk plug-in. Storage
position can be selected by multi-purpose knob.
Recall
1 ~ 200
Call out recorded waveform from U disk then select recording
position by multi-purpose knob.
Return
Return to previous level of menu.
——
Page 70
58
Operating Manual for UTD2000/3000
Form 2-23 By detection
Function menu
Setup
Description
Detection status
Open
Close
Open/close information of waveform pass/fail times
Information source
CH1
CH2
MATH
RefA
RefB
Select CH1 as detected signal source.
Select CH2 as detected signal source.
Select MATH as detected signal source.
Select RefA as detected signal source.
Select RefB as detected signal source.
Output conditions
Pass
Pass/Halt
Fail/Halt
Fail
Output concurrency prompt when detecting waveform pass;
Output and pause when detecting waveformpass;
Output and pause when detecting waveformfail;
Output concurrency when detecting waveformfail;
Template setup
Refer to Form
2-24 for details
Enter into template setup menu.
Return
Return to previous menu.
Page 71
59
Key note:Key note:
Self-correction: Self-correction procedure is to
correct measurement error caused by digital
storage oscillograph due to ambient variation. You
can operate the procedure according to demand.
Please open power of digital storage oscillograph
then wait for instrument pre-heatingfor 30 minutes
for more correct calibration. Please press UTILITY
key (as an auxiliary function) for operation
according to prompt shown in the screen.
Language selection: There are many kinds of
languages for UTD2000/3000 series digital storage
oscillograph. Please press UTILITY men
u button
and select appropriatelanguage.
Operating Manual for UTD2000/3000
Form 2-24 Template setup
Function menu
Setup
Description
Template output
Set horizontal and vertical
Pass/Fail tolerance scope
according to adjustment
Horizontal
1-200Pixel
Set horizontal tolerance
scope
Vertical
1-100Pixel
Setvertical tolerance scope
Return
Return to Pass detection
menu
/
/
Page 72
60
Automatic measurement
As shown in following figure, MEASURE is
automatic measurement function key. You can be
familiar with powerful automatic measurement
function of UTD2000/3000 series digital storage
oscillograph gradually by following introduction.
Application example
Measurement menu of such series of oscillograph
can be used to measure 28 kinds of waveform
parameters. User can press MEASURE key Enter
into parameter measurement display menu firstly.
The menu includes 5 areas to display measurement
value at the same time. They correspond to function
keys of F1 to F5 respectively. Please press
corresponding F keys Enter into measurement
category selection menu if measurement category
shall be selected for any area. Measurement
category selection menu can be classified into
voltage and time kinds Enter into voltage ortime
measurement category respec
tively. Please press
F1 to F5 to select measurementcategory then
return to parameter
measurement display menu. In addition,user can
Operating Manual for UTD2000/3000
Figure 2-28 Function key (automatic measurement)
of sampling system
Page 73
press F5 to select “All parameters” to display all
voltage/time measurement parameters. User can
press F2 to select channel to measure. (Channel is
ineffective until opening.) User can press F1 Return
to parameter measurement display menu if not
altering current measurement category. The first
example: Steps are as follows if displaying
measured peak value of CH2 channel in F1 area.
1. To press F1 key Enter into measurement
category selection menu;
2. To press F2 to select CH2;
3. To press F3 to select voltage;
4. To press F5 (shown in the second page, 2/4) to
see that F3 position is “peak value”
5. To press F3 to select “peak value” and return to
parameter measurement display menu
automatically;
Peak value is displayed in F1 area on the
homepage of measurement menu.
The secondexample: Setup of delay measurement:
Delay measurement function is used for interval
between ascending edges of2 information sources,
namely, interval from ascending edge of the first
cycle of a certain information source to that of the
first cycle of additional info
rmation source.
Measurement steps are as follows:
1. To refer to above example and select area
displayed by delay measurement value after
entering into measurement menu (shown in 3/3
page of time category);
2. To select reference information source of CH1
then select delay information sourceof CH2;
3. To select reference information source of
then select delay information sourceof CH2;
4. To press F5 to confirm that delay measurement is
displayed in selected area.
CH1
61
Operating Manual for UTD2000/3000
Page 74
Automatic measurement of voltage parameters
Voltage parameters which can be measured by
UTD2000/3000 series oscillograph automatically
are as follows:
Peak value (Vpp): Voltage from the highest
waveform point to the lowest waveform point;
The maximum value (Vmax): Voltage from the
highest waveform point to GND (ground);
The minimum value (Vmin): Voltage from the lowest
waveform point to GND (ground);
Middle value (Vmid): Halfof sum of waveform top
and bottom voltage;
Amplitude (Vamp): Voltage from waveform top to
bottom;
Middle value (Vmid): Halfof sum of waveform top
and bottom voltage;
Top value (Vtop): Voltage from flat wav
eform top to
GND (ground);
Bottom value (Vbase): Voltage from waveform
bottom to GND (ground);
Overshoot: Ratio of difference of the maximum
waveform value and top value and amplitude;
Preshoot: Ratio of difference of the minimum
waveform value and bottom value and amplitude;
Average value: Average amplitude of signals within
a cycle;
Root-mean-square(Vrms): It is effective value.
Root-mean-squarerefers to DC voltage of
converted energy within 1 cycle according toAC
signal to correspond to equivalent energy.
Operating Manual for UTD2000/3000
62
Page 75
Automatic measurement of time parameters
UTD2000/3000 series oscillograph can measure
signal frequency, cycle, ascending time,
descending time, positive pulse width, negative
pulse width,delay (of 9 combinations), positive duty
cycle, negative duty cycle and 10 time parameters
automatically. Definitions of such time parameters
are as follows:
RiseTime: It refers to period when waveform
amplitude increases from 10% to 90%;
FallTime: It refers to period when waveform
amplitude decreases from 90% to 10%;
Positive pulse width (+Width): It refers to pulse
width of positive pulse under 50% of amplitude;
Negative pulse width
(-Width): It refers to pulse
width of negative pulse under 50% of amplitude;
Delay (ascending edge):Itrefers to delay from
ascending edge to ascending edge;
Delay (descending edge): It refers to delay from
descending edge to descending edge;
Positive duty cycle (+Duty): It refers to ratio of
positive pulse width and cycle;
Negative duty cycle (-Duty): It refers to ratio of
negative pulse width and cycle;
Operation Description: Please press MEASURE
key so that the screen can display 5 measurement
values.As shown in Form 2-25, user can press any
key of F1 to F5 so that the screen enters into
me
asurementselection menu.
Measurement menu
Operating Manual for UTD2000/3000
63
Page 76
Operating Manual for UTD2000/3000
64
Form 2-25
Function menu
Setup
Description
Return
Return to parameter measurement display menu.
Information source
CH1
Select channel of measurement parameters.
CH2
Select channel of measurement parameters.
Voltage kind
Enter into voltage parameter menu.
Time kind
Enter into time parameter menu.
All parameters
Display/close all measurement parameters.
Form 2-26
Function/measurement item
Description
Return
Return to menu shown in Form 2-25.
Preshoot
Return to parameter measurement display menu and
substitute original position parameter after selection.
Amplitude
he same as above
Overshoot
The same as above
Next page (1/4)
Page turning
Of which: Menu included by “voltage kind” is shown in Form 2-26 to 2-29.
Page 77
65
Form 2-27
Function/measurement item
Function/measurement item
Description
Description
Operating Manual for UTD2000/3000
Previous page
Return to previous page.
Average value
Return to parameter measurement display menu and substitute
original position parameter after selection.
Peak value
he same as above
he same as above
Root-mean-square
Next page (2/4)
Page turning
Form 2-28
Previous page
Top value
Return to previous page.
Return to parameter measurement display menu and substitute
original position parameter after selection.
Bottom value
The same as above
The same as above
Middle value
Next page (3/4)
Page turning
Page 78
66
Operating Manual for UTD2000/3000
Function/measurement item
Function/measurement item
Description
Description
Previous page
Return
Return to previous page.
Return to menu shown in Form 2-25.
The maximum value
Frequency
Return to parameter measurement display menu and substitute
original position parameter after selection.
Return to parameter measurement display menu and substitute
original position parameter after selection.
The minimum value
Cycle
The same as above
The same as above
Next page (4/4)
Next page (1/3)
Return to the first page (shown in Form 2-26)
Page turning
Form 2-29
Form 2-30
——
——
Menu included by “time category” is shown in Form 2-30 to Form 2-32.
The same as above
RiseTime
Page 79
67
Operating Manual for UTD2000/3000
Function/measurement item
Function/measurement item
Description
Description
Previous page
Previous page
Return to previous page.
Return to previous page.
Frequency
Delay
Return to parameter measurement display menu and substitute
original position parameter after selection.
Enter into delay option menu (shown in Form 2-32a) after
selection.
Cycle
Positive duty cycle
The same as above
Return to parameter measurement display menu and substitute
original position parameter after selection.
Next page (2/3)
The first page (3/3)
Page turning
Return to the first page (shown in Form 2-30).
Form 2-31
Form 2-32
The same as above
The same as above
RiseTime
Negative duty cycle
Page 80
Cursor measurement
User can press CURSOR button to display
measurement cursor and cursor menu then use
multi-purpose knob controller to alter cursor
position. As shown in following figure, CURSOR in
MENU control area is cursor measurement function
key.
Figure 2-29 Function key (or cursor) of sampling
system
68
Operating Manual for UTD2000/3000
Return to parameter
measurement display menu
and substitute original
position parameter after
selection.
Form 2-32a
Function
Setup
Description
Channel
Channel
CH1/CH2
/MATH
CH1/CH2
/MATH
Select measured channel.
——
——
——
——
Enter
Page 81
User can move cursor for measurement under
CURSOR mode. There are 3 modes, namely, voltage,
time and tracing. V can be measured by pressing
SELECT and COARSE keys on the panel and multi-
purpose knob controller to adjust positions of 2
cursors respectively during voltage measurement.
Similarly, it can bemeasured if selecting time. Cursor
of digital storage oscillograph can also trace signal
variation automatically under tracing mode with
waveform display.
Remark: For 5.7-inch oscillograph
1. Effect of SELECT key is to selectthe cursor.
2. Effect of COARSE key is to adjust movement speed
of cursor.
For 7-inch oscillograph:
User can press multi-function knob key to select the
cursor. Movement speed of cursor can also alter
according to rotation speed.
1. Voltage/time measurement mode: The first or
secondcursor will appear at the same time. Cursor
position on the screen can be adjusted by multi-
purpose knob controller. User can press SELECT key
(or multi-function knob key) to select cursor to adjust.
Displayed reading is also time or voltage between 2
cursors.
2. Tracing mode: H
orizontal and vertical cursors can
intersect to become a crossing cursor. Crossing
cursor can position on the waveform automatically to
adjust horizontal position of crossing cursor on the
waveform by rotating multi-purpose knob controller.
Digital storage oscillograph can also display
coordinate of cursor point at the same time.
3. Measurement value will be displayed in upper
right corner of the screen automatically when
opening cursor function.
Operating Manual for UTD2000/3000
69
Page 82
Usage of operation key
RUN/STOP key is in upper right corner of front
panel of digital storage oscillograph. User can
press the key so that green lamp is on to show
operation status.Red lamp can be on after pressing
the key to show stop status.
Figure 2-30 Operation/stop key
As shown in above figure, automatic setup isto
simplify operation. Digital storage oscillograph can
adjust vertical deflection coefficient and horizontal
time base gear automatically for stable waveform
display on the screen according to waveform
amplitude and frequency when user presses AUTO
key. System setup is shown in following form during
automatic setup:
Automatic setup
Operating Manual for UTD2000/3000
70
Page 83
RUN/STOP: To acquire waveform or stop
acquisition continuously;
The key can be pressed for switchover of waveform
sampling between operation and stop. User can
press the key under operation status so that green
lamp is on. “Auto” can be displayed in upper part of
the screen. Old digital storage oscillograph can
acquire waveform continuously while red lamp is on
by pressing the key under stop status. Digital
storage oscillograph can stop acquisition when
displaying “Stop” in upper part of the screen.
Operating Manual for UTD2000/3000
71
Function
Setup
Obtainment mode
Sampling
Display format
Set to YT
Horizontal position
Automatic adjustment
Adjustment according to signal
frequency
Second/grid
Trigger coupling
AC
Hold off
The minimum value
Trigger level
Set to 50%
Automatic
Trigger mode
Trigger source
Set to CH1; Set to CH2 when
exerting on signal for it if there
is no signal for CH1;
Trigger gradient
Ascending
Trigger type
Edge
All
Vertical band width
Volt/grid
Adjustment according to signal
amplitude
Page 84
Operating Manual for UTD2000/3000
72
Chapter 3: Application Example
The first example: Measurement of simple signal
It has to observe a unknown signal in the circuit for
quick display and measurement of signal frequency
and peak value.
2
1. Please operate according to following steps for
quick signaldisplay:
1. Please operate according to following steps for
quick signaldisplay:
To set probe menu attenuation coefficient to 10*
and set switch on the probe to 10*.
To connect CH1 probe to measured point of
circuit.
To press AUTO button.
Digital storage oscillograph can set automatically
for the best waveform display. You can also adjust
vertical or horizontal gear on above basis so that
waveform display can meet related requirements.
Automatic measurement of signal voltage and
time parameters
Digital storage oscillogra
ph can measure most of
display signals automatically. Please operate
according to followingsteps for measurement of
signal frequency and peak value.
To press MEASURE key to display automatic
measurement menu;
To press F1 to enter into measurement menu
category for selection
To press F3 to select voltage category
To press F5 to turn to 2/4 page then press F3 to
select measurement type of peak value
To press F2 to enter into measurement menu
category for selection then press F to select
time category
①.①②;
;
;⑤4;②
To press F2 to select measurement type of
frequency.
Page 85
Operating Manual for UTD2000/3000
73
The secondexample: Delay observation of sine
wave signal by circuit
The same as above example, please set
attenuation coefficient of probe and digital storage
oscillograph to 10*. Please connect CH1 channel
of digital storage oscillograph to circuit signal input
terminal and connect CH2 channel to output
terminal.
Operation steps:
1. Signal display of CH1 and CH2 channels:
To press AUTO button
To continue to adjust horizontal or vertical gear
so that waveform display can meet your test
requirements.
To press CH1 key to select CH1 then rotate
vertical position knob to adjust vertical position
of CH waveform
To press CH2 key to select CH2 then adjust
vertica
l position of CH2 waveform to facilitate
.
②③1 .
&
Figure 3-1Automatic measurement
Page 86
Operating Manual for UTD2000/3000
74
comparison without waveform overlapping for the
first and second channels.
2. To measure delay of sine signal caused by
circuit then observe waveform variation.
Delay of automatic measurement channel
To press MEASURE button to display automatic
measurement menu;
To press F1 key to enter into measurement menu
category for selection;
To press F4 key to enter into time measurement
parameter list;
To press F5 key twice to enter into 3/3 page;
To press F2 key to select delay measurement;
To press F1 key to select CH1 thenpress F2 key to
select CH2 then press F5 as enter key
You can see delay value from “CH1-CH2 delay” in
F1 area.
Observation of waveformvariation(shown in
following figure)
Figure 3-2: Waveform delay
Page 87
Operating Manual for UTD2000/3000
75
The third example: Capture of single signal
Advantages and characteristics of digital storage
oscillograph include convenient capture of
aperiodic signals such as pulse and burring, etc.It
has to know this signal to set trigger level and
trigger edge to capture a signal signal. For
instance: Trigger level shall be set to about 2V and
set trigger edge to ascending edge trigger if pulse is
a logic signal ofTTL level. User can observe it by
automatic or ordinary trigger mode so as to
determine trigger level and trigger edge for
unknown signals.
Operation steps are as follows:
1. To set attenuation coefficient of probe and CH1
channel like
the previous example;
2. To set triggers;
To press MENU button in trigger control area to
display trigger setup menu
To set trigger type to edge, trigger source to CH1,
gradient to ascend, trigger mode of single mode and
trigger coupling toAC by using menu operation keys
of F1 to F5 respectively under this menu.
To adjust horizontal time base and vertical gear to
appropriate scope
To rotate TRIGGER LEVEL knob to adjust
appropriate trigger level
To press RUN/STOP execution button to wait for
signals which meet trigger conditions. Once
sampling can bedone for display on the screen if a
certain si
gnal reaches set trigger level. Occidental
events can be captured easily by utilizing the
function. For instance: Sudden burring with large
amplitude Please set trigger level to be higher
than normal signal slightly then press
RUN STOP button for waiting The machine will
trigger automatically and record waveform before
and after trigger during burring. Horizontal
POSITION knob in horizontal
①.②③.④.⑤:
/ .
Page 88
Operating Manual for UTD2000/3000
76
control area on the panel can be rotated to alter
horizontal trigger position so as to gain negative
delay trigger with different length to facilitate
observation of waveform before burring.
Figure 3-3 Single signal
You can adjust setup of digital storage oscillograph
for overlapping of random noise to tested signal so
as to filter or reduce noise and avoid disturbance to
ontology signal during measurement. (Please refer
to following figure for waveform.)
Figure 3-4 To reduce random noise of signals
Operation steps are as follows:
1. As shown in previousexample, please set
attenuation c
oefficient of probe and CH1 channel.
The fourth example: To reduce random noise of
the signal
Page 89
77
2. To connect signals for stable waveform display on
digital storage oscillograph.
Please refer to previous example for operation.
Please refer to previous chapter for adjustment of
horizontal time base and vertical gear.
3. To improve trigger by setting trigger coupling.
(1) To press MENU button in trigger area to display
trigger setup menu.
(2) To set trigger coupling to low-frequency
inhibition or high-frequency inhibition.Low-
frequency inhibition is to set a high-pass filter to
filter low-frequency signal component of less than
80kHz and allow pass through of high-frequency
si
gnal component. High- frequency inhibition is to
set a low-pass filter to filter high-frequency signal
component of more than 80kHz and allow pass
through of low-frequency signal component. It can
be set to inhibit low-frequency or high-frequency
noise respectively to gain stable trigger
4. To reduce display noise by setting sampling
modes.
To adopt average sampling mode to eliminate
display of random noise to thinner waveformand
facilitate observation & measurement if waveform is
thickened by overlapping random noise to
measured signal. Signal details can be observed
convenientl
y by reduced random noise after gaining
average value. Detailed operation is as follows
To press ACQUIRE button in panel menu area to
display sampling setup menu To press F1 key as
menu operation key to set obtainment mode to
average status. (As shown in following figure,)
please adjust average times by pressing F2 key as
menu operation key with step from 2 to 256 by
multiples of 2 so that waveform display can meet
observation & test equirements.
①:.
Operating Manual for UTD2000/3000
Page 90
78
Figure 3-5 Inhibited signal noise
To reduce display noise by reducingwaveform
brightness.
Note:Average samplingmode can be used to slow
down updating speed for waveform display as
normal phenomenon.
This digital storageoscillograph can measure 28
kinds of waveform parameters automatically. All
automatic measurement parameters can be measured
by cursor. Waveform time and voltage can be
measured quickly by using cursor.
Please operate according to following steps to
measure 1 staircase voltage of staircase signal:
1. To press CURSOR key to display cursor
measurement menu.
2. To pr
ess F1 key as menu operation key to set
cursor type to voltage.
3. To rotate multi-purpose knob controller to place the
first cursor to 1 staircase of staircase signal.
4. To press SELECT or PUSHSELECT to select the
cursor then rotate multi-purpose knob controller to
place the secondcursor to another staircase of
staircase signal.
The fifth example: Application cursor
measurement
To measure 1 staircase voltage of staircase signal
Operating Manual for UTD2000/3000
Page 91
79
△ , ,
.
V namely pressure difference can be shown
in cursor menu automatically
Please refer to following figure for details.
Note: Please set cursor type to time if onlying
adopting the secondstep asabove for time
measurement by cursor.
Figure 3-6 Voltage difference of signal by cursor
measurement
To review phase location difference of signals of 2
channels.
Example: Phase location variation caused by 1
circuit for tested signals. Please connect digital
storage oscillograph to circuit to monitor
input/output signal of circuits. Please operate
according to followingsteps to review input/o
utput
terminal of circuits by way of X-Y coordinate
diagram.
1. To set probe menu attenuation coefficient to 10*
and set switch on the probe to 10*.
2. To connect CH1 probe to network input then
connect Ch2 probe to network output.
3. To press CH1 and CH2 menu keys and open 2
channels if they have not been displayed.
4. To pressAUTO button.
5. To adjust vertical scale knob for approximate
equivalent display amplitude of 2 loops of signals.
The sixth example: Application of X-Y function
Operating Manual for UTD2000/3000
Page 92
80
6. To press DISPLAY menu key to call out display
control menu.
7. To press F2 to select X-Y. Digital storage
oscillograph will display input/output characteristics of
circuit under Lissa Jous's figure mode.
8. To adjust vertical scale and vertical position knob to
reach the best effect for waveform.
9. (As shown in following figure), please observe and
calculate phase location difference by elliptic
oscillograph figure method.
sinθ =A/B o r C/D. Of which: “θ ” refers t o p hase
location difference angle between channels. Please
refer to above figure for definition ofA, B, C and D.
Therefore,
phase location difference angle is
calculated as follows =±arcsin A B or
±arcsin C D Calculated phase location
difference angle shall be within Quadrant I or IV
(namely, (0~ ) or ( ~ ) if main axle of ellipse
is within Quadrant I or III. Calculated value is within
( ~ ) or ( ~ ) if main axle of ellipse is within
Quadrant II or IV. Additionally, frequency and phase
location relationship between 2 signals can be
calculated according to figure if frequency or phase
location difference of 2 measured signals is integral
multiple.
: ( / )
= ( / ) .
Operating Manual for UTD2000/3000
Signal must be in the
horizontal center
Figure 3-7
3π22π
π2π2ππ3π
2
Page 93
81
10.Form of X Y phase location difference
-
The seventh example: Trigger of video signal
Trigger of video field
Video trigger function can be used for all modes
except UTD2000L/2000EX series. (Of which:
UTD2025CL has video trigger function.)
It has to adopt video trigger to gain stable video
output signal display for observation of a certain
video circuit.
Please operate according to following steps for
trigger in video field.
1. To press[ MENU ]button in trigger control area to
display trigger menu.
2. To press[ F1] key as menu operation key to select
video type.
3. To press[ F2] key as menu operation key to set
trigger source to CH1.
4. To press[ F3] key as menu operation key to select
video standard of PA
L.
Operating Manual for UTD2000/3000
Frequency
ratio of
signal
1:1
Phase location difference
0
°
45
°
90
°
180°270
°
360
°
Page 94
Operating Manual for UTD2000/3000
82
5. To press [F4] key as menu operation key to select
synchronization of odd field or even field.
6. To adjust horizontal time base by rotating
horizontal SCALE knob in horizontal control area so
as to gain clear waveform display.
Figure 3-8 Trigger of video field
Please operate according to following steps for
trigger in video row:
1. To press [MENU] button in trigger control area to
display trigger area.
2. To press[ F1] key as menu operation key to select
video.
3. To press [F2] key as menu operation key to set
trigger source to Ch1.
4. To press[F3] key as menu operation key to select
video standard
of PAL.
5. To press[ F4] key as menu operation key to select
synchronization to designated row.
6. To rotate multi-purpose knob controller for trigger
in any row.
7. To adjust horizontal time base by rotating
horizontal SCALE knob in horizontal control area so
as to gain clear waveform display.
Trigger of video row
Page 95
Operating Manual for UTD2000/3000
83
Figure 3-9 Trigger of video row
It has to check if input signal is within rule scope or
not. Fail refers to out of scope and Pass refers to
within the scope. Pass/Fail signal can be shown by
output port .
1. To press[UTILITY] button then press[F2]Enter into
Pass/Fail menu.
2. Setup of information source: Enter into Pass/Fail
menu then press[F2] to select information source.
3. Template setup:To press[F4] Enter into template
setup menu (Template).To press[F2] and[F3] shown
in Template menu to set horizontal and vertical
tolerance scope by multi-function knob, (namely, 1-
200Pixel toward ho
rizontal direction and 1-100Pixel
toward vertical direction.) Please press F4 finally
Return to Pass/Fail menu.
4. Setup of output conditions:To press[F3] to set
output conditions shown in Pass/Fail menu.
5. To begin detection: To press[F1] toopen
detection status shown in Pass/Fail menu.
The eighth example: Pass/Fail detection
of the rear panel
Rear panel of Pass / Fail output
Pass FailInterface Schematic
/
Page 96
Operating Manual for UTD2000/3000
84
The ninth example: Usage of updating
procedure of U-disk
Updating procedure of U-disk can be used more
conveniently and flexibly. Please operate according to
following steps if using this function:
1. To download procedure documents to update from
Internet then save them in U-disk. (It is suggested to
use U-disk suggested by us.)
2. To close oscillograph and insert U-diskthen open
oscillograph.
3. To turn to confirmation interface of whether to
update it or not if there is only 1 procedure document
in U-disk. Please press[F5] for exit or press F1 to exit
procedure updating. Document selection interace will
appear if there are at least 2 procedure
documents in
U-disk. Please press[F5] key for confirmation (or
press[F1] to exit procedure updating). Please turn to
confirmation interface of whether to update it or not
then press[F5] for updating).
4. To wait for updating for prompt of successful
updating. Pleasepower off oscillograph then open it to
fulfill updating procedure.
Note:
1. Please wait for updating for several minutes.
2. It is not allowed to power off oscillograph or plug
out U-disk during updating to avoid updating failure
or unforeseenable error.
3. Please power off oscillograph then open it for re-
updating after upd
ating failure.
Page 97
Operating Manual for UTD2000/3000
85
The tenth example: Usage of storage function
1. Setup of storage function:
Storage functions of oscillograph include setup
storage, waveform storage, bitmap storageand
specific screen copy function of UTD2000L/2000EX
series of oscillograph.
It has to select display channel, trigger channel,
vertical and horizontal gears of oscillograph for
setup. Please press[STORAGE] key to select
torage type of “setup” by pressing[F1 ]key. As
shown in following figure, it has to select position to
save by multi-function button then press F3 for
saving.
It can be tested directly during next test after saving
setup to reduce re-setup process. UTD2000C series
of oscillog
raph can save 10 kinds of setup while
UTD2000E and UTD2000L/2000EX series of
oscillograph can save 20 kinds of setup.
Figure 3-10 Saving setup interface
Page 98
Operating Manual for UTD2000/3000
86
2. Waveform storage function:
UTD2000 series of oscillographcan provide waveform
storage function. User can press[STORAGE]key and
[F1] key to select storage type of “waveform”. User can
press [F2]key to select channel to save waveform then
press[F5]Enter into next page. Magnetic disk in F1
menu position can designate storage position and
default storage position is digital storage oscillograph
(DSO) without plugging in U-disk. Stored magnetic disk
position can be [DSO] or [USB] after plugging in U-disk.
Default storage depth shown by F2 is ordinary and
savedwaveform only can be displayed by oscillograph.
Stora
ge depth can be selected to long storage to save
“.data” for exportation by waveform analysis software
when plugging in U-disk and selecting magnetic disk
position to [USB]. It is allowed to analyze content data
of every point of saved waveform. User can alsopress
F5 Return toprevious menu then select storage
position by multi-function button then press [F3] for
saving.
Figure 3-12 Waveform storage by U-disk
Figure 3-11 Waveform storage
Page 99
Operating Manual for UTD2000/3000
87
As shown in following figure, saved waveform can
be recalled by short-cut key of [REF] on the panel
for UTD2000C and UTD2000E series of
oscillograph. User can press [F3] shown in the
second page of STORAGE main menu for recall for
UTD2000L/2000EX series of oscillograph.
Figure 3-13 Vertical control panel of 5.7-inch screen
As shown in following figure, user can press[ F1]
menu key of STORAGE main menu to select storage
type of “bitmap” after plugging in U-disk and
successful connection of oscillograph. User can
rotate multi-function knob to select saved bitmap
No. Saved bitmap in U-disk wit
h “.BMP” format can
be read out by computer directly.
Figure 3-14 Bitmap storage
3. Bitmap storage function:
Page 100
Operating Manual for UTD2000/3000
88
4. PrtSc copy & storage function
Introduction to System Prompt
There is 1 PrtSc key in upper left corner of the screen
for UTD2000L/2000EX series of oscillograph. User
can press the key for screen copy after plugging in U-
disk. Displayed image can be saved in U-disk by way
of “.BMP” format.
Figure 3-15 Screen copy
Regulation to the extreme: It prompts that multi-
purpose knob has been regulated to the greatest
extent under current status and continued adjustment
is not allowed. The promptwill display when
regulating vertical deflection coefficient switch, time
base switch, X displacement, vertical displacement
and trigger level to the terminal.
Suc
cessful connection of U-disk: The prompt will
display on the screen after connecting U-disk to
digital storage oscillograph and correct connection.
Plug-out of U-disk: The prompt will display on the
screen when plugging out U-disk from digital
storage oscillograph.Saving: The prompt will
display on the screen with schedule bar below it
when saving waveform
Chapter 4: System Prompt and
Troubleshooting
Loading...