Uni-Trend UT513 User Manual

Model UT513
OPERATING MANUAL
Model UT513: OPERATING MANUAL
Table of Contents
Title Page
Introduction Unpacking the Meter Safety Information International Electrical Symbols Battery Saver (Sleep Mode) Battery Indication The Meter Structure Display Key Functions Measurement Operation A. Measuring Voltages B. Measuring Insulation Resistance a) Continuous Measurement b) Timed Measurement c) Polarization Index (PI) Measurement d) Compare Function The Use of Power Adaptor
5 5 6 8 8 8 9 12 14 17 17 19 20 20 21 22 23
1
Model UT513: OPERATING MANUAL
Title Page
USB Interface Maintenance A. General Service B. Replacing the Battery Specifications Safety and Compliances Physical Specifications General Specifications Feature Summary Detailed Accuracy Specifications A. Voltage Measurement B. Insulation Resistance Measurement
24 25 25 26 27 27 27 28 28 29 29 30
2
Model UT513: OPERATING MANUAL
List of Tables
Table Title Page
1. Unpacking Inspection
2. International Electrical Symbols
3. Battery Indication
4. Meter Front Description
5. Meter Side Description
6. Display Description
7. Key Description
5 8 8 10 11 13 14
3
Model UT513: OPERATING MANUAL
List of Figures
Figure Title Page
1. The Meter Front Structure
2. The Meter Side Structure
3. Display
4. Voltages Measurement
5. Insulation Resistance Measurement
6. The Use of Power Adaptor
7. USB Interface Connection
8. Battery Replacement
9 11 12 17 19 23 24 26
4
Model UT513: OPERATING MANUAL
Introduction
Uni-Trend Model UT513 insulation resistance tester (hereafter, “the Meter”) is a handheld instrument designed primarily to make resistance/ insulation
resistance measurement.
Unpacking the Meter
The Meter includes the following items:
Table 1. Unpacking Inspection
Item
1 2
3
4
5 6 7 8 9
In the event you find any missing or damage, please contact your dealer immediately.
Description
English Operating Manual One plug test lead to one alligator clip (Black colour) One plug test lead to one alligator clip (Green colour) Two plugs test lead to one alligator clip (Red colour)
1.5V Battery (LR14) Tool Box USB Interface Cable Software Power adaptor (input voltage 230V, 50/60Hz, 75mA, output DC14V , 1.0A)(optionally , available at extra cost)
Qty
1 piece 1 piece
1 piece
1 piece
8 pieces 1 piece 1 piece 1 piece 1 piece
5
Model UT513: OPERATING MANUAL
Safety Information
This Meter complies with the standards IEC61010 safety measurement requirement: in pollution degree 2, overvoltage category (CAT. III 600V) and double insulation.
CAT II: Local level, appliance, PORTABLE EQUIPMENT etc., with smaller transient voltage overvoltages than CAT. III
Use the Meter only as specified in this operating manual, otherwise the protection provided by the Meter may be impaired.
Danger identifies conditions and actions that pose hazard(s) to the user.
Warning identifies avoiding electric shock. Caution identifies conditions and actions that may
damage the Meter and carrying out accurate measurement.
6
Operating Caution identifies conditions that user needs to take extra care during operating the Meter
International electrical symbols used on the Meter and in this Operating Manual are explained on page 8.
Danger
Use of instrument in a manual not specifed by the manufactuer may impair safety features/protection provided by the equipment. Read the following safety information carefully before using or servicing the instrument.
Do not apply more than 600V. Do not use the Meter around explosive gas, vapor or dust. Do not use the Meter in a wet environment. When using the test leads, keep your figures away from the lead contacts. Keep your figures behind the finger guards on the leads. Do not use the Meter with any parts or cover removed.
Model UT513: OPERATING MANUAL
When carrying out insulation measurement, do not contact the circuit under test.
Warning
Do not use the Meter if it is damaged or metal part is exposed. Look for cracks or missing plastic. Be careful when working above 33V rms, 46.7V ac rms or 70V DC. Such voltages pose a shock hazard. Discharge all loading of circuit under test after measuring high voltage. Do not change battery when the Meter is in wet environment. Place test leads in proper input terminals. Make sure all the test leads are firmly connected to the Meter’s input terminals. Make sure the Meter is turned off when opening the battery compartment.
Caution
When performing resistance tests, remove all power from the circuit to be measured and discharge all the power.
When servicing the Meter, use only the same model number or identical electrical specifications of test leads and power adaptor. Do not use the Meter if the battery indicator ( ) shows a battery empty condition. Take the battery out from the Meter if it is not used for a long time. Do not use or store the Meter in an environment of high temperature, humidity, explosive, inflammable and strong magnetic field. The performance of the Meter may deteriorate after dampened. Soft cloth and mild detergent should be used to clean the surface of the Meter when servicing. No abrasive and solvent should be used to prevent the surface of the Meter from corrosion, damage and accident. Dry the Meter before storing if it is wet.
7
Model UT513: OPERATING MANUAL
International Electrical Symbols
International symbols on the Meter and in this manual are explained in Table 2.
Table 2. International Electrical Symbols
Risk of electric shock Equipment protected by double or reinforced insulation. DC Measurement AC Measurement Grounding See Manual Empty of Built-In Battery Conforms to Standards of European Union
Battery Saver (Sleep Mode)
The Meter enters the Sleep Mode and blanks the display if there is no button press for 15 minutes. This is done to conserve battery power. The Meter comes out of Sleep Mode when ON/OFF button is pressed and hold for 1 second.
8
Battery Indication
There is a battery indicator shows on the display upper left hand corner. Below Table 3 is the explanation:
Table 3. Battery Indication
Battery Indicator
Battery Voltage
10V or less. It means the battery is empty, don’t use the Meter as it cannot guarantee accuracy. 10V~10.5V. It means the battery is nearly empty, replacing battery is necessary. At this status, the Meter can still do 500V and 1000V output measurement, accuracy will not be affected.
10.6V~11.5V
11.6V or more
Model UT513: OPERATING MANUAL
The Meter Structure
Below Figure 1 and Table 4 shows the Meter front structure and description
Figure 1. The Meter Front Structure
Green Black
Red
9
1
LCD 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16
Scroll Button Emergency Stop Data Clear the Display Backlight Button
Down Button On/Off Button Compare Button Insulation Resistance Button DC Voltages measurement Button Timer Button. AC Voltages measurement Button Test Button USB Button Data Store Button. Data Recall Button
Scroll Button
Table 4. Meter Front Description
17 18
19
20
21
22
Model UT513: OPERATING MANUAL
Up Button LINE: High Voltage output input terminal (two plugs red test lead to one alligator clip) High voltage line shielding input terminal (two plugs red test lead to one alligator clip) GUARD: Grounding protection input terminal (one plug black test lead to one alligator clip) EARTH: High resistance measurement input terminal (one plug green test lead to one alligator clip) Testing leads: Two plugs red test lead to one alligator clip. One plug black test lead to one alligator clip. One plug green test lead to one alligator clip.
10
Model UT513: OPERATING MANUAL
Below Figure 2 and Table 5 shows the Meter side structure and description
Figure 2. The Meter Side Structure
Table 5. Meter Side Description
1 Safety Shutter 2 Power adaptor Input Terminal 3 USB Port
11
Display
Table 6 and Figure 3 describe the display.
Figure 3. Display
Model UT513: OPERATING MANUAL
12
Model UT513: OPERATING MANUAL
Table 6. Display Description
Number
1 2 3 4 5 6 7 8 9 10 11
Meaning
Indicator for DC voltage Indicator for data store full Indicator for clearing Indicator for AC voltage Indicator for timer Step symbol Indicates selected pass/fail compare value Indicates for negative reading Timer 1 symbol Timer 2 symbol Data store is on
Number
12 13 14 15 16 17 18 19 20 21
Meaning
Data recall is on Indicator for polarization index Unit symbols The continuity buzzer is on Compare feature pass Analogue bar graph Risk of electric shock Compare feature fail Indicator for power adaptor Battery life indicator
13
Model UT513: OPERATING MANUAL
Key Functions
ON/OFF
E-STOP
CLEAR /
SAVE
14
Turn on or off the Meter. Press and hold the button for 1 second to turn the Meter on. Press again to turn off the Meter. The Meter default range is 500V insulation resistance continuous measurement when turning on. Emergency stop button. Press this button when the Meter is hang and cannot turn off the power. Press to turn on or off the display backlight Press and hold to clear the stored data Press to store the current measurement value. The maximum number of stored reading is 18. When the stored readings memory is full, the Meter shows FULL and stop storing. Press and hold CLEAR / to clear the stored value in order to store the next measurement
Table 7. Key Description
LOAD
value.
Press once to recall the first stored value. Press again to exit Load feature. Load feature can only be used when there is no high voltage output. When the insulation resistance measurement has no testing voltage output, press to select one voltage range up. Under load mode: press to recall the previous stored value. When the insulation resistance measurement has no testing voltage output, press to select one voltage range down. Under load mode: press to recall the next stored value.
Model UT513: OPERATING MANUAL
Table 7. Key Description
When set the timer duration for the measurement of insulation resistance or polarization index, press to decrement the time. The maximum length of time is 15 minutes and 30 seconds, the Meter will automatically carry out measurement. When compare feature measuring insulation resistance, press to decrement a resistance comparing value. After polarization index measurement, press to display polarization index, TIME 2 insulation resistance value and TIME 1 insulation resistance value in sequence. When set the timer duration for the measurement of insulation resistance or polarization index, press to
USB
increment the time. The maximum length of time is 30 minutes and 30 seconds, the Meter will automatically carry out measurement. When use the compare feature measuring insulation resistance, press to increment a resistance comparing value. After polarization index measurement, press to display polarization index, TIME 2 insulation resistance value and TIME 1 insulation resistance value in sequence. Press once to start the data transferring to the computer via USB, USB symbol shows on the display. Press again to stop the data transferring to the computer via USB, USB symbol disappears.
15
Table 7. Key Description
Model UT513: OPERATING MANUAL
COMP
TIME
TEST
IR
DCV
ACV
16
Set a pass / fail limit for insulation tests. The default value is 10M Pres to step through continuous measurement, timed measurement and polarization index measurement in sequence. Press to stop or start an insulation resistance test Press to initiate insulation resistance measurement Press to initiate DC voltage measurement Pres to initiate AC voltage measurement
Model UT513: OPERATING MANUAL
Measurement Operation
Below section explains how to make measurements. Press and hold ON/OFF to turn on the Meter, press
again to turn off the Meter. The Meter default range is 500V insulation resistance continuous measurement when turning on.
A. Measuring Voltages
Red
Green
Figure 4. Voltages Measurement
17
Operating Catuion
To avoid harms to you or damages to the Meter, please do not attempt to measure voltages higher than 600V or 600V rms, although readings may be obtained. Special care should be taken when measuring high voltage.
To measure voltages, set up the Meter as Figure 4 and do the following:
Press DCV or ACV button to select DC voltage or
1. AC voltage measurement Insert the red and green test lead into the tested
2. circuit. When measuring DC voltage, if the red test lead is
3. negative voltage, “-“ symbol will show on the display.
Note
When voltage measurement has been completed, disconnect the connection between the testing leads and the circuit under test and remove testing leads away from the input terminals of the Meter.
Model UT513: OPERATING MANUAL
18
Model UT513: OPERATING MANUAL
B. Measuring Insulation Resistance
Red
Green
Black
Groud
Figure 5. Insulation Resistance Measurement
Operating Caution When performing insulation resistance tests,
remove all power from the circuit to be measured and discharge all the power. Operating the Meter must be very careful as it outputs dangerous voltage during measurement. Must make sure the tested object is firmed clipped, hands are away from the clips, then press TEST button to put high voltage. Do not short circuit the testing leads during high voltages output or test insulation resistance after high voltages output. This kind of incorrect operating may cause sparking and fire, which damages the Meter and harms to you. Do not measure over 10 seconds when: 500V measure resistance lower than 2M 1000V measure resistance lower than 5M 2500V measure resistance lower than 10M 5000V measure resistance lower than 20M
19
To measure insulation resistance, set up the Meter as Figure 5 and do the following:
Press IR button to select insulation resistance
1. measurement. When there is no testing voltage output, press
2. and button to select voltages of 500V, 1000V, 2500V or 5000V. When performing insulation resistance tests, remove
3. all power from the circuit to be measured and discharged all the power. Insert the red test lead into the LINE input terminal
4. and the black test lead into GUARD input terminal. Connect the red and black alligator clip to the circuit
5. to be measured, negative voltage output from LINE terminal. Choose below insulation resistance measurement
6. mode.
Model UT513: OPERATING MANUAL
Press TIME button to select continuous measurement mode, there is no timer icon on the LCD. Press and hold TEST button for 1 second to carry out continuous measurement. Output insulation resistance testing voltage, TEST button light up, blinks on every 0.5 seconds. Press TEST button to close the insulation resistance measurement voltage when measurement is completed. TEST button lights off, disappears. The LCD shows the current insulation resistance measurement value.
Press TIME button to select timed measurement mode, the LCD displays TIME 1 and symbols. Press and buttons to set the time (00:10~15:00). Within 1 minute, the time increment or decrement by every 10 seconds. Afterward, the time increment or decrement by every 30 seconds.
20
Model UT513: OPERATING MANUAL
Then press and hold TEST button for 2 second to carry out timed measurement. TIME 1 and are displayed and blinked on the LCD on every 0.5 seconds. When the set time is reached, the insulation resistance measurement voltage will be closed and the measurement will be automatically stopped. The LCD displays the insulation resistance reading.
Press TIME button to select timed measurement mode, the LCD displays TIME 1 and symbols. Press and buttons to set the time (00:10~15:00). Within 1 minute, the time increment or decrement by every 10 seconds. Afterward, the time increment or decrement by every 30 seconds. Press TIME button again. TIME 2, PI and symbols appear on the LCD. Press and buttons to set the time (00:15~15:30). Within 1 minute, the time increment or decrement by every 10 seconds. Afterward, the time increment
or decrement by every 30 seconds. Then press and hold TEST button for 2 seconds to carry out timed measurement. TIME 1 and are displayed and blinked on the LCD on every 0.5 seconds before TIME 1 set time is reached. TIME 2 and are displayed and blinked on the LCD on every 0.5 seconds before TIME 2 set time is reached. When the two set time are reached, the insulation resistance measurement voltage will be closed and the measurement will be automatically stopped. The LCD displays the polarization index reading. Press , to set through the polarization index, TIME 2 insulation resistance reading and TIME 2 insulation resistance reading.
Information: PI = 3 minutes ~10 minutes reading / 30 seconds ~1 minute reading
PI Standard
4 or more The best
4~2 Good
2.0~1.0 Warning
1.0 or less Bad
21
Press COMP button to select compare feature. COMP symbol displays on the LCD.. Press and buttons to set the compare value Below is the list in sequence of the compare value: 10M , 20M , 30M , 40M , 50M , 60M , 70M , 80M , 90M ,100M , 200M , 300M , 400M , 500M , 600M , 700M , 800M , 900M ,1G , 2G , 3G , 4G , 5G , 6G , 7G , 8G , 9G , 10G ,20G , 330G , 40G , 50G , 60G , 70G , 80G , 90G , 100G ,200G , 300G , 400G , 500G , 600G , 700G , 800G , 900G Press and hold TEST button for 2 seconds to carry out the measurement. The NG symbol will display if the insulation resistance value is smaller than resistance value. Otherwise GOOD symbol will be displayed.
Model UT513: OPERATING MANUAL
22
Model UT513: OPERATING MANUAL
The Use of Power Adaptor
The use of power adaptor, see figure 6
Open the side safey shutter, then you will see there
1. is a power adaptor input terminal. Make sure the Meter is power off and insert the
2. UT513 power adaptor to the input terminal. It is highly recommeded to take out all the batteries
3. when you are using the power adaptor. Make sure the Meter is power off when you
4. disconnect the UT513 power adaptor from the Meter. It is highly recommeded to use Uni-Trend supplied
5. UT513 power adaptor to avoid dangerous.
Figure 6. The Use of Power Adaptor
23
USB Interface
Connecting the USB interface, see figure 7
Figure 7. USB Interface Connection
Model UT513: OPERATING MANUAL
Install the included software, the installation guide
1. can be seen from the CD. Open the side safety shutter, then you will see there
2. is a USB port. Insert the included USB cable to the Meter’s USB
3. port and the other end to the computer.
24
Model UT513: OPERATING MANUAL
Maintenance
This section provides basic maintenance information including battery replacement instruction.
Warning
Do not attempt to repair or service your Meter unless you are qualified to do so and have the relevant calibration, performance test, and service information.
A. General Service
Periodically wipe the case with a damp cloth and mild detergent. Do not use abrasives or solvents. T o clean the terminals with cotton bar with detergent, as dirt or moisture in the terminals can affect readings. Turn the Meter to OFF when it is not in use. Take out the battery when it is not using for a long time. Do not use or store the Meter in a place of humidity, high temperature, explosive, inflammable and strong magnetic field.
If the Meter is wet, dry it before use.
25
Model UT513: OPERATING MANUAL
B. Replacing the Battery
Figure 8. Battery Replacement
Warning
To avoid electric shock, remove all the test leads from the Meter when replacing the batteries.
26
Operating Caution
Don’t mix to use old and new batteries. Be careful the polarity is correct when installing batteries. Do not use the Meter if the battery indicator ( ) shows a battery empty condition. Do you carry out measuring during the battery compartment is open.
Follow Figure 8 and proceed as follows to replace the battery:
Turn the Meter to OFF and remove all connections from the terminals. Remove the screw from the battery compartment, and separate the battery compartment from the case bottom. Replace with 8pcs of new 1.5V (LR14) batteries. Rejoin the case bottom and battery compartment, and reinstall the screw.
Model UT513: OPERATING MANUAL
Specifications Safety and Compliances
Certification Compliances
IEC 61010 CAT.III 600V overvoltage and double insulation standard
Physical Specifications
Display (LCD)
Operating Temperature Storage Temperature Relative Humidity
Battery Type
Dimensions (HxW xL) Weight
Digital: 9999 counts Analog bar graph. 0 ~40 (32 ~104 )
-20 ~60 (-4 ~152 ) 85% @ 0 ~40 below; 90% @ -20 ~60 :
8pcs of 1.5V (LR14) batteries or power adaptor (input voltage 230V , 50/60Hz, 75mA, input DC14V, 1.0A). Power adaptor is optionally at extra cost. 202 x 155 x 94 mm Approx. 2kg (including battery)
27
General Specifications
Model UT513: OPERATING MANUAL
Range Overloading Battery Indicator Icon Display Current Consumption
Feature Summary
Display Backlight Computer connection Data Logging and Recall Autorange Warning Voltage COMP Measurement
PI Measurement
TIME
28
Auto Display OL on insulation resistance range Display Equips with function and battery indicator icons. Maximum: around 1.0A Average: around 20mA
Bright backlight for clear readings in poorly lighted areas. Via USB interface. 18 The Meter automatically selects best range
and red light will on. Auto release voltage Use the Compare function to set a pass/fail compare level for the insulation measurements. Polarization Index is the ratio of insulation resistance. You can pre-set two point of times and automatically carry out the measurement. To carry out measurement by setting a specified time within 15 minutes.
Model UT513: OPERATING MANUAL
Detailed Accuracy Specifications
Accuracy: ([% of reading] + [number of least significant digits), guarantee for 1 year. Operating temperature: 18 ~28 Relative humidity: 45~75%RH
A. Voltage Measurement
Measurement Range Resolution Accuracy
DC Voltage
30 ~ 600V
1V
(2%+5) Among them 30~100V(50/60Hz) (2%+8)
AC Voltage 30V~600V (50/60Hz)
29
B. Insulation Resistance Measurement
Model UT513: OPERATING MANUAL
Output Voltage Display Range Open Circuit Voltage Test Current Short Circuit Accuracy
500V
0.5M ~20G DC 500V + 20%, -0% 1mA~1.2mA @ 500k
1000V 2M ~40G DC1000V + 20%, -0% 1mA~1.2mA @ 1M
Less than 2.0mA
100k ~ 100M : (3%+5)
100M ~ 10G : (5%+5)
10G ~ 100G : (10%+5)
Above 100G : 20% (Humidity:Below 50%)
2500V 5M ~100G DC 2500V + 20%, -0% 1mA~1.2mA @ 2.5M
5000V 100M ~1000G DC5000V + 20%, -0% 1mA~1.2mA @ 5M
Operating Caution
At any output voltage, when the tested resistance is les than 10M , the testing time cannot exceed 10 seconds continuously.
30
Model UT513: OPERATING MANUAL
*END*
This operating manual is subject to change without notice.