This product conforms with safety product design requirements of UL and CE standards.
UT4000 Series
Digital Storage Oscilloscope
User Manual
Introduction
Dear valued user,
Thank you for purchasing a Uni-Trend instrument. To use your new product correctly,
read this User Manual carefully and completely before operation and pay particular
attention to “Safety Instructions”.
Please keep this User Manual in a safe place after reading it carefully. For easy reference
during future operation, we recommend putting it alongside your Uni-Trend product or in
an easily accessible place.
UNI-T
®
UT4000 User Manual
Copyright Information
z Uni-Trend Technology (Dongguan) Limited. All rights reserved.
z Products of the Company are protected by patents granted and pending in the
People’s Republic of China or other countries.
z The Company reserves the right to change product specifications and prices.
Uni-T all rights reserved. All licensed software products are properties of Uni-Trend
Technology (Dongguan) Limited and its subsidiaries or software providers. They are
protected by the national copyright law and international conventions. Information
contained in this manual supercedes all information in previously published versions.
Uni-T is the registered trademark of Uni-Trend Technology (Dongguan) Limited.
UNI-T®
UT4000 User Manual
If this product is sold or assigned by the original purchaser to a third party within three
years of purchase, the new owner should note that warranty is available for a period of
three years from the day the original purchaser acquired the product from Uni-T or an
authorized dealer. The probe, other accessories and fuses are not covered by warranty.
If any genuine defect is found during the valid warranty period, Uni-T has the option to
repair the defective product without any charge for parts or labor, or replace it with
another product (to be decided by Uni-T). For warranty purposes, Uni-T may use parts,
modules and replacement products that are brand new or repaired to a good-as-new
standard. All old parts, modules and products that are removed during replacement
become properties of Uni-T.
In this User Manual, “customer” means an individual or entity vested with the rights
hereunder. To enjoy the warranty service, the “customer” must report any defect to Uni-T
during the valid warranty period and make appropriate arrangement to allow servicing.
The customer should pack the defective product in a container and deliver it to a
maintenance centre specified by Uni-T. The customer should also prepay all freight cost
and provide a copy of the original sales receipt issued to the original purchaser. If the
product is to be delivered to an address within the country where the maintenance centre
operates, Uni-T shall pay the cost of returning the product to the customer. If the product
is to be delivered any other destination, all freight, custom duty, tax and other costs will
be paid by the customer.
This warranty does not apply to any defect, malfunction or damage caused by accident,
normal wear and tear of mechanical parts, any form of application other than the stated
ones, improper use, improper maintenance or poor maintenance. Under the warranty
terms and conditions, Uni-T has no obligation to provide the following services :
a) Repairing any damage arising from installation, repair or maintenance carried out by
a non Uni-T service representative;
b) Repairing any damage arising from improper use or connection to incompatible
equipment;
c) Repairing any damage or malfunction arising from using a power source not
provided by Uni-T;
d) Servicing a product that has been modified or integrated with other products (such
modification or integration makes repair more time consuming and difficult).
This warranty is made available to this product specifically and supercedes all other
previous warranties, whether express or implied. Uni-T and its dealers will not make any
implied guarantee on the salability or suitability of this product for any specific purpose.
In the event of breach of warranty terms and conditions, repair or replacement of
defective product shall be the only and all remedial measure offered by Uni-T.
Notwithstanding any prior notification of potential damage that is indirect, special,
consequential or inevitable, Uni-T and its dealers shall bear no liability for any such
damage.
UNI-T
®
UT4000 User Manual
General Safety Rules
This unit is designed and manufactured strictly in accordance with GB4793 safety
requirements for electronic testing meters and IEC61010-1 safety standards. It fully meets
CAT II 600V insulation and overvoltage requirements and Grade II anti-pollution safety
standards. To prevent personal injuries and damage of this unit or any other devices
connected to it, please take note of the following safety precautions. To avoid potential
hazards, use this unit strictly as instructed by this User Manual.
Maintenance should only be carried out by a trained professional.
To avoid fire and personal injury :
Use a correct power cable : Use only the specified power cable which is authorized in
the country of use.
Remove the plug correctly : Do not remove the probe or testing cable when they are
connected to power.
Ensure good grounding : This unit is grounded by the ground wire of the power cable.
To avoid electric shock, the grounding conductor must touch the ground. Before
connecting the input or output terminal, ensure the unit is properly grounded.
Connect the probe of the digital storage oscilloscope : The probe ground cable is the
same as ground potential. Do not connect the ground cable to high voltage.
Check the rated values of all terminals : To prevent fire and excessive current shock,
please check all rated values and label data. Read the manual carefully and check the
rated values before connecting the unit.
Do not operate the unit with the chassis cover open : Do not operate this unit when the
outer cover or front panel is open.
Use suitable fuses : Only use specified fuse types and rated specifications.
Avoid exposing circuitry : When power is on, never make contact with exposed adaptor
or components.
When fault is suspected, stop operation : If you suspect a fault, ask a qualified
maintenance professional to carry out inspection.
Maintain good ventilation.
Do not operate in humid condition.
UNI-T®
UT4000 User Manual
Do not operate in combustible and explosive conditions.
Keep the product surface clean and dry.
Safety Messages and Symbols
Safety terminology used in this manual. The following messages may appear in this
manual :
Warning : Warning statements identify conditions or practices that could result in
injury or loss of life.
Caution : Caution statements identify conditions or practices that could result in
damage to this unit or other properties.
Messages on the product : The following messages may appear on the product :
“Danger” means potential damage that is immediate.
“Warning” means potential damage that is not immediate.
“Caution” means possible damage to this product or other properties.
Symbols on the product : The following symbols may appear on the product :
High voltage Caution! Refer to manual Protective ground terminal
Ground terminal for chassis Ground terminal for testing
UNI-T
®
UT4000 User Manual
Preface
This manual provides information on the operation of the UT4000 digital storage
oscilloscope series. Guidance is given in several chapters as follows :
Chapter 1 ⎯ User Guide : Simple guide to oscilloscope functions and installation.
Chapter 2 ⎯ Instrument Setups : Guide to operation of the UT4000 digital storage
oscilloscope series.
Chapter 3 ⎯ Practical Illustrations : Example illustrations are provided to solve
various testing problems.
Chapter 4 ⎯ System Prompts and Trouble-shooting :
Chapter 5 ⎯ Servicing and Support :
Chapter 6 ⎯ Appendixes :
Appendix A : Technical Indicators
Appendix B : Accessories for UT4000 Oscilloscopes
Appendix C : Maintenance and Cleaning
UNI-T®
UT4000 User Manual
UT4000 Introduction
UT4000 oscilloscopes offer user-friendliness, outstanding technical indicators and a host
of advanced features. They are your perfect tools to complete testing tasks swiftly and
efficiently. This manual is a user guide for 7 models of this digital storage oscilloscope
series :
UT4000 oscilloscopes offer a user-friendly front panel with clear indications to allows
access to all basic functions for easy operation. The scaling and position buttons for all
channels are optimally arranged for direct view operation. As design is based on the
controls of traditional instruments, users can use the new units without spending
considerable time in learning and familiarizing with operation. For faster adjustment to
ease testing, there is an [AUTO] key to instantly display the appropriate waveform and
range position.
The performance features listed below will explain why UT4000 series can fully satisfy
your testing and measurement requirements :
z 2GS/c real-time sampling rate (dual channel 1GS/s, single channel 2GS/s) and
z Dual time base function; unrivaled waveform detail observation and analysis
capabilities
z 6k storage depth; 60M equivalent storage depth; 1024k recording length
z 16-channel logic analyzer; 3½ digital multimeter
z Unique envelop sampling feature with direct visual display of carrier wave details
after amplitude modulation
z Quick waveform acquisition and colour change function that visually describes the
strength of waveform
z Scroll display in scan mode for continuous monitoring of signal variations
z HD colour LCD system at 320 x 240 pixel aspect ratio
z Unique XY mode that displays the waveform and Lissajous figure concurrently
z USB drive system software upgrade
z Supports plug-and-play USB storage devices and capable of communicating with a
computer through the USB storage device
z Automatic waveform and status configuration
z Storage of waveforms, setup and bit map; waveforms and setups reproduction
z Automatic measurement of 27 waveform parameters; parameters testing and
customization
z Automatic cursor tracing measurement
z Unique wave recording and replay function
z Built-in FFT
z Multiple waveform mathematics functions (including add, subtract, multiply and
divide)
z Edge, video, pulse width, slew rate and alternate trigger functions
z Visual system help messages
UT4000 accessories :
z 2 x 1.5m, 1:1/10:1 probe. For details refer to the probe instructions. These
accessories conform with EN61010-031: 2002 standards
z Power line conforming to international standards applicable in the country of use
z User Manual
z Product Warranty Card
z USB connecting cable : UT-D04
z UT4000 communication control software (USB-DEVICE)
z 2 x multimeter pen; 2 x current-to-voltage converter module; UT-M03/UT-M04
z UT4000 logic analyzer probe (including clamps and cable, flex cable) : UT-P06
UNI-T®
UT4000 User Manual
Table of Contents
Item
General Safety Rules
Preface
Chapter 1 User Guide 1
General Inspection 5
Functional Check 5
Probe Compensation 8
Automatic Setups for Waveform Display 8
Getting to Know the Vertical System 9
Getting to Know the Horizontal System 10
Getting to Know the Trigger System 12
Chapter 2 Instrument Setups 14
Setting Up the Vertical System 15
Setting Up the Horizontal System 24
Setting Up the Trigger System 27
Setting Up the Sampling System 34
Setting Up the Display System 37
Storage and Recall 39
Setting Up Utility Functions 41
Automatic Measurement 44
Cursor Measurement 47
Using the Run Button 48
Logic Analyzer (LA) Measurement 49
Multimeter Measurement 51
Chapter 3 Practical Example Illustrations 54
Page
UNI-T
UT4000 User Manual
®
Item
Illustration 1 : Measuring simple signals 52
Illustration 2 : Observing the delay caused by a
Illustration 3 : Acquiring single signals 54
Illustration 4 : Reducing random noise of signals 55
Illustration 5 : Using cursors for measurement 56
Illustration 6 : Using the X-Y function 57
Illustration 7 : Video signal triggering 59
Illustration 8 : Pass/fail check 61
Illustration 9 : Quick acquisition 62
Illustration 10: Using the dual time base function 62
Illustration 11: Using the Logic Anaylzer 63
Illustration 12: Using the multimeter 64
Chapter 4 System Prompts and Trouble-shooting 65
Definitions of System Prompts 65
Troubleshooting 65
Chapter 5 Service and Support 67
Warranty 67
Using the USB Upgrade Program 67
Contact Us 74
Chapter 6 Technical Indicators 75
Appendix A : Technical indicators 75
Appendix B : Accessories for UT4000 Oscilloscopes 81
Appendix C : Maintenance and Cleaning 81
Index 83
sine wave signal passes through the circuit
Page
53
UNI-T®
UT4000 User Manual
Chapter 1 ⎯ User Guide
UT4000 Series digital storage oscilloscopes are small and compact portable devices. The
user-friendly front panel enables easy operation for basic testing and measuring tasks.
This chapter provides notes on the following :
U General inspection
U Functional check
U Probe compensation
U Automatic setups for waveform display
U Getting to know the vertical system
U Getting to know the horizontal system
U Getting to know the trigger system
When beginning to use your UT4000 oscilloscope, first familiarize yourself with the
operation front panel. This chapter briefly describes the operation and functions of the
front panel, so you can get started with your UT4000 series digital storage oscilloscope as
quickly as possible.
Your UT4000 oscilloscope comes with a front panel with at-a-glance functions for easy
operation. There are buttons and function keys on the front panel. The functions of knobs
are similar to other oscilloscopes. The row of 5 keys on the right of the display panel are
menu operation keys (designated as [F1] to [F5] from top down). With these keys you can
set up different options of the current menu. The other keys are function keys. You can
use them to enter different function menus or access particular functions directly.
1
UT4000 User Manual
Figure 1-1 Front panel of UT4000 Oscilloscopes
UNI-T
®
2
UNI-T®
UT4000 User Manual
Figure 1-2 Schematic diagram for operating the UT4000 front panel
3
UT4000 User Manual
Figure 1-3 Schematic diagram for the display interface
UNI-T
®
4
UNI-T®
UT4000 User Manual
General Inspection
We suggest checking your new UT4000 oscilloscope in the following steps.
1. Check the unit for possible shipping damages
If the package carton or foam plastic protective lining is seriously damaged, do not
discard until you have carried out an electrical and mechanical check and found the
product and accessories performing satisfactorily.
2. Check the accessories
A checklist of accessories that come with your UT4000 oscilloscope is provided in the
section “UT4000 Accessories” of this user manual. Please check for any missing items
against this list.
If any item is missing or damaged, please contact your UNI-T dealer or our local office.
3. Thorough inspection of the entire unit
If the exterior of the unit is damaged, or it is not operating normally, or it fails to pass the
performance test, please contact your UNI-T dealer or our local office.
Functional Check
Carry out a quick functional check in the following steps to make sure your oscilloscope
is operating normally.
1. Power on the unit
Power on the unit. Power supply voltage is 100-240V AC, 45-440Hz. After connecting to
power, start the self calibration process by pressing the [UTILITY] button (utility
function) once and [F1] twice. Then press [SELECT] to run.
To go to the next page press [UTILITY], [F1], THEN [F5]. To run the factory default
setup, press[F1] then [SELECT], as shown in Figure 1-4.
When the above procedure is complete, press [CH1] to enter the CH1 menu.
5
UNI-T
®
UT4000 User Manual
UTILITY button
Figure 1-4
Warning : To avoid danger, ensure the digital storage oscilloscope is safely grounded.
2. Accessing signals
Your UT4000 oscilloscope has dual input channels and one external trigger input channel.
Please access signals in the following steps :
c Connect the probe of the digital storage oscilloscope to the CH1 input, and set the
attenuation switch of the probe to 10X (Figure 1-5).
Figure 1-5 Setting the attenuation switch
6
UNI-T®
UT4000 User Manual
d You have to set the probe attenuation factor of the oscilloscope. This factor changes the
vertical range multiple to ensure the measurement result correctly reflects the amplitude
of the signal being tested. Set the attenuation factor of the probe as follows : Press [F4]
then [F2] to display 10X on the menu.
Probe ration
Figure 1-6 Setting the probe deflection factor of the oscilloscope
e Connect the probe tip and ground clamp to the corresponding probe compensation
signal terminals. Press [AUTO] and you will see a square wave in the display (about 3V
peak-to-peak at 1kHz) in a few seconds, as shown in Figure 1-7.
4 Repeat these steps to check CH2. Press the [OFF] function button to close CH1, then
press [CH2] function button to open CH2. Repeat steps 2 and 3.
Figure 1-7 Probe compensation signal
7
UNI-T
®
UT4000 User Manual
Probe Compensation
When connecting the probe to any input channel for the first time, perform this
adjustment to match the probe to the channel. Skipping the compensation calibration step
will result in measurement error or fault. Please adjust probe compensation as follows :
1. In the CH1 menu set the probe to 10X. Move the switch on the probe to 10X and
connect the probe to CH1. Connect the probe tip to the probe compensator’s signal
output connector, then connect the ground clamp to the ground wire connector of
the probe compensator. Display CH1 and press [AUTO].
3. If you see an “Undercompensation” or “Overcompensation” waveform display,
adjust the adjustable tab inside the adjustment hole at the tip of the probe with the
adjusting rod in the probe accessory pack, until a “Correct Compensation”
waveform shown in Figure 1-8 is displayed.
Warning : To avoid electric shock when measuring high voltage with the probe,
ensure the probe’s insulation lead is in good condition. Do not touch the metal part
of the probe when connected to HV power.
Automatic Setups for Waveform Display
Your UT4000 oscilloscope features an autoset function. It can automatically adjust the
vertical deflection factor, scanning time base and trigger mode based on the input signal,
until the most appropriate waveform is displayed. The autoset function can only be
operated when the signal to be measured is 50Hz or above and the duty ratio is larger than
1%.
Using the Autoset Function :
1. Connect the signal to be tested to the signal input channel.
2. Press [AUTO]. The oscilloscope will automatically set the vertical deflection factor,
scanning time base and trigger mode. Should you require to make more detailed
check, you can adjust manually after the autoset process until you get the optimal
waveform display.
8
UNI-T®
b
p
UT4000 User Manual
Getting to Know the Vertical System
As shown in the figure below, there are a series of buttons and knobs in the vertical
control zone. The following steps will get you familiar with the use of these controls.
Figure 1-9 Vertical controls
1. Turn the vertical position knob [POSITION]
signal in the centre of the window. The vertical position knob controls the vertical
display position of the signal. When you turn the vertical position knob, the sign
indicating vertical reference of the [GROUND] channel will move up and down with
the waveform.
Measurement Tips
If the channel coupling is DC, you can test the signal’s DC% quickly
the difference between the waveform and signal ground.
In the case of AC coupling, the DC% within the signal will be intercepted. In this
coupling mode you can display the AC% of the signal with higher sensitivity.
Shortcut key for resetting the vertical
zero [SET TO ZERO]
This shortcut key can reset vertical shift, horizontal shift and holdoff to the zero
position (center point).
to enable the waveform to display the
y checking
osition of the dual analog channel to
9
UNI-T
®
UT4000 User Manual
2. Change the vertical setups and check changes of status information. You can
identify changes of any vertical range by reading the status display column at the
lower corner of the waveform window. Turn the vertical VOLTS/DIV
change the vertical VOLT/DIV range. You will find that the range in the current
status column changes accordingly. Press [CH1], [CH2], [MATH] or [REF] and the
screen will show the operation menu, icon, waveform and range status information
of the corresponding channel. Press [OFF] to close the selected channel.
knob to
Getting to Know the Horizontal System
Horizontal controls are shown in the figure below. The following steps will get you
familiar with horizontal time base setups.
Figure 1-10 Horizontal controls
10
UNI-T®
UT4000 User Manual
1. Use the horizontal SEC/DIV knob to change the horizontal time base setup. Turn
the horizontal SEC/DIV
find that the time base range in the current status column of the corresponding
channel changes accordingly. Range of horizontal scanning rate is 5ns~50s/div
(UT4102C), in steps of 1-2-5.
* Note : Minimum time base range of the UT4000 Series varies from model to
model.
2. Use the horizontal [POSITION]
in the waveform window. When the horizontal POSITION
see that the waveform moves horizontally with the knob.
3. Press [MENU] to display the ZOOM and dual time base menu. In this menu press
[F1] to activate window expansion. Then press [F1] again to quit window expansion
and return to the main window. To access dual time base setup, press [F3]. You can
also set the holdoff time with the multifunction control knob [¯].
Shortcut key for resetting the trigger point shift to horizontal zero position
When trigger point has moved from the horizontal centre point considerably by way
of horizontal shift, pressing the shortcut key [SET TO ZERO] can quickly reset the
trigger point to the vertical centre point. You can also turn the knob to make
adjustments.
Definition
Trigger point means the actual trigger point relative to the centre point of the
storage device. When it is at the horizontal centre point, it is also at the centre point
of the storage device. By turning the horizontal knob, you can move the trigger
point horizontally.
Holdoff means the time interval until the trigger circuit is next reactivated. Turn the
multifunction control knob to set the holdoff time. By adjusting the holdoff time,
you can observe combination or complex signals more effectively.
knob to change the SEC/DIV time base range. You will
knob to adjust the horizontal position of the signal
knob is turned, you can
11
UNI-T
®
UT4000 User Manual
Getting to Know the Trigger System
Trigger controls are shown in the figure below. The following steps will get you familiar
with trigger setups.
Figure 1-11 Trigger controls
1. Use the trigger level knob to change the trigger level. The trigger level line on the
screen will move up and down with the knob. While you move the trigger level, you
will find the trigger level value at the bottom of the screen changing accordingly.
2. Open the [TRIGGER MENU] (see Figure 1-12) to change trigger setups. General
setups are as follows :
Press [F1] twice to select the edge as [TRIGGER TYPE].
Press [F2] to select CH1 as [TRIGGER SOURCE] (Select with the multifunction
control knob then press [SELECT] to confirm).
Press [F3] then [F1] to set [TRIGGER COUPLING] as DC.
Press [F4] then [F1] to set [TRIGGER MODE] as AUTO.
Press [F5] to set [EDGE TYPE] as RISING.
12
UNI-T®
UT4000 User Manual
Figure 1-12 Trigger menu
3. Press the [50%] button and set the trigger level at the vertical centre point of the
trigger signal amplitude.
4. Press [FORCE] to generate a compulsory trigger signal that is mainly used in the
normal and single trigger modes.
13
UNI-T
®
UT4000 User Manual
Chapter 2 ⎯ Instrument Setups
You should be familiar with basic operation of the vertical controls, horizontal controls
and trigger system menu of your UT4000 oscilloscope by now. After reading the last
chapter, you should be able to use the menus to set up your digital storage oscilloscope. If
you are still unfamiliar with these basic operation steps and methods, please read Chapter
1.
This chapter will guide you through the following :
Setting up the vertical system ([CH1], [CH2], [MATH], [REF], [OFF], [VERTICAL
POSITION], [VOLTS/DIV])
Setting up the horizontal system ([MENU], [HORIZONTAL POSITION
[SEC/DIV
Setting up the trigger system ([TRIGGER LEVEL
Setting up the sampling method ([ACQUIRE])
Setting up the display mode ([DISPLAY])
Save and transfer ([STORAGE])
Setting up the utility function ([UTILITY])
Automatic measurement ([MEASURE])
Cursor measurement ([CURSOR])
Using the auto setup, run/stop buttons ([AUTO], [RUN/STOP])
Logic analyzer [LA]
Digital multimeter [DMM]
It is recommended that you read this chapter carefully to understand the various
measurement functions and system operation steps of your UT4000 Series Oscilloscope.
]
], [MENU], [50%], [FORCE])
],
14
UNI-T®
UT4000 User Manual
Setting Up the Vertical System
CH1, CH2 and setups
Each channel has its own vertical menu. You should set up each item for each channel
individually. Press the [CH1] or [CH2] function button and the system will display the
operation menu for CH1 or CH2. For explanatory notes please see Table 2-1 below :
Table 2-1 Explanatory notes for channel menu
Function Menu Setup Explanatory Note
Coupling DC
AC
Ground
Bandwidth limit Full bandwidth
20MHz
VOLTS/DIV Coarse tune
Fine tune
Probe 1X
10X
100X
1000X
Invert On
Off
AC and DC quantities of input signal is allowed
to pass.
DC quantities of the input signal are intercepted.
DC level of the channel input terminal upon
equivalent grounding is displayed.
Full bandwidth.
Limit bandwidth to 20MHz to reduce noise
display.
Coarse tune in steps of 1-2-5 to set up the
deflection factor of the vertical system. Fine
tune means further tuning within the coarse tune
setup range to improve the vertical pixel aspect
ratio.
Select one of the values based on the probe
attenuation factor to keep the vertical deflection
factor reading correct. There are four values :
1X, 10X, 100X and 1000X.
Waveform invert function on.
Normal waveform display.
15
UNI-T
®
UT4000 User Manual
1. Setting up channel coupling
Take the example of applying a signal to CH1. The signal being tested is a sine signal that
contains DC %.
Press [F1] then [F2] to select AC. It is now set up as AC coupling. DC quantities of the
signal being tested will be intercepted. The waveform display is as follows :
AC coupling setup
Figure 2-1 DC quantities of the signal are intercepted
Press [F1] twice to select DC. Both DC and AC quantities of the testing signal being
inputted to CH1 can pass through. The waveform display is as follows :
DC coupling setup
Figure 2-2 Both DC and AC quantities of the signal are displayed
16
UNI-T®
UT4000 User Manual
Press [F1] then [F3] to select ground. It is now set up as ground. The waveform display is
as follows :
(Note : In this mode, although waveform is not displayed, the signal remains
connected to the channel circuit)
Ground coupling setup
Figure 2-3 Screen display for the
ground mode
2. Setting up the channel bandwidth
Take the example of applying a signal to CH1. The signal being tested is a signal stacked
with high frequency quantities.
Press [CH1] to turn CH1 on. Next press [F2] then [F1] to set channel bandwidth to full
bandwidth. The signal being measured can pass through even if it contains high frequency
quantities. The waveform display is as follows :
Full bandwidth
Figure 2-4 Waveform display at full bandwidth
17
UNI-T
®
UT4000 User Manual
Press [F2] then [F3]. All high frequency quantities higher than 20MHz in the signal being
tested will be limited. The waveform display is as follows :
Bandwidth limit 20MHZ
Bandwidth limit icon
Figure 2-5 Waveform display when bandwidth limit is on
3. Setting up the probe rate
To match the probe attenuation factor setup, it is necessary to set up the probe attenuation
factor in the channel operation menu accordingly. For example, when the probe
attenuation factor is 10:1, set the probe attenuation factor at 10X in the menu. Apply this
principle to other values to ensure the voltage reading is correct.
The figure below shows the setup and vertical range display when the probe is set at 10:1.
Probe attenuation factor
Vertical range movement
Figure 2-6 Setting up the probe attenuation factor in the channel menu
18
UNI-T®
UT4000 User Manual
4. Vertical VOLTS/DIV adjustment setup
You can adjust the VOLTS/DIV range of the vertical deflection factor either in the coarse
tune mode or fine tune mode. In coarse tune mode, the VOLTS/DIV range is
1mV/div~10V/div. Tuning is in steps of 1-2-5. In fine tune mode, you can change the
deflection factor in even smaller steps within the current vertical range, so as to
continuously adjust the vertical deflection factor within the range of 1mV/div~10V/div
without interruption.
Fine tune setup
Figure 2-7 Coarse tuning and fine tuning the vertical deflection factor
19
UNI-T
®
UT4000 User Manual
5. Wavefor m inversion setup
Waveform inversion : The displayed signal is inverted 180 degrees with respect to the
channel ground level. Figure 2-8 shows the non-inverted waveform. Figure 2-9 shows the
inverted waveform.
Figure 2-8 Inversion setup for vertical channel inversion (non-inverted)
Figure 2-9 Inversion setup for vertical channel inversion (inverted)
Non-inverted
Inverted waveform
20
UNI-T®
UT4000 User Manual
I. Operating Math Functions
Math functions are displays of +, -, ×, ÷, FFT mathematical results and digital filtering
function of CH1 and CH2 waveforms. The menu is as follows :
Math range
Figure 2-10 Math functions
Table 2-2 Explanatory notes for the Math menu
Function Menu
Type Math
Operand 1 CH1
Operator +
Operand 2 CH1
Scaling ratio 1/1
Setup Explanatory Note
To carry out +, -, ×, ÷ functions.
Set operand 1 as CH1 waveform.
CH2
Set operand 1 as CH2 waveform.
Operand 1 + Operand 2
×
÷
Operand 1 - Operand 2
Operand 1 x Operand 2
Operand 1 ÷ Operand 2
Set operand 2 as CH1 waveform.
CH2
Set operand 2 as CH2 waveform.
Scale the waveform by ratio. There are four ratios to
1/10
choose from : 1/1, 1/10, 1/100, 1/1000.
1/100
1/1000
FFT spectrum analysis
By using the FFT (Fast Fourier Transform) algorithm, you can convert time domain
signals (YT) into frequency domain signals. With FFT, you can conveniently observe the
following types of signals :
z Measure the harmonic wave composition and distortion of the system
z Demonstrate the noise characteristics of the DC power
z Analyse oscillation
21
f
b
b
b
p
UNI-T
®
UT4000 User Manual
Table 2-3 Explanatory notes for the FFT menu
Function Menu Setup Explanatory Note
Type FFT To carry out FFT algorithm functions.
Signal source CH1
CH2
Window function Hamming
Blackman
Rectangle
Hanning
How to use FFT functions
Signals with DC quantities or DC offset will cause error or offset of FFT waveform
quantities. To reduce DC quantities, select AC coupling. To reduce random noise and
frequency aliasing resulted by repeated or single pulse event, set the acquiring mode o
your oscilloscope to average acquisition.
Select the FFT Window
Assuming the YT waveform is constantly repeating itself, the oscilloscope will carry out
FFT conversion of time record of a limited length. When this cycle is a whole number, the
YT waveform will have the same amplitude at the start and finish. There is no waveform
interruption. However, if the YT waveform cycle is not a whole number, there will be
different amplitudes at the start and finish, resulting in transient interruption of high
frequency at the connection point. In frequency domain, this is known as leakage. To
avoid leakage, multiply the original waveform by one window function to set the value to
0 for start and finish compulsively. For application of the window function, please see the
table below :
Table 2-4
FFT Window Feature Most Suitable Measurement Item
Rectangle The best frequency
recognition rate, the worst
amplitude recognition rate.
Basically similar to a status
without adding window.
Hanning Frequency recognition rate is
etter than the rectangle
window, but amplitude
recognition rate is poorer.
Hamming Frequency recognition rate is
marginally better than
Hanning window.
Blackman The best amplitude
recognition rate and the
oorest frequency recognition
rate.
Set CH1 as math waveform.
Set CH2 as math waveform.
Set Hamming window function.
Set Rectangle window function.
Set Blackman window function.
Set Hanning window function.
Temporary or fast pulse. Signal level is
generally the same before and after. Equal
sine wave of very similar frequency. There
is broadmoving wave spectrum.
Sine, cyclical and narrownoise.
Temporary or fast pulse. Signal level
varies greatly before and after.
Mainly for single-frequency signals to
search for higher-order harmonic wave.
and random noise with slow
and random
22
UNI-T®
UT4000 User Manual
Definition :
FTT recognition rate : It means the quotient of the sampling and math points. When
math point value is fixed, the sampling rate should be as low as possible relative to the
FFT recognition rate.
Nyquist frequency : To rebuild the original waveform, at least 2f sampling rate should be
used for waveform with a maximum frequency of f. This is known as Nyquist stability
criterion, where f is the Nyquist frequency and 2f is the Nyquist sampling rate.
Digital Filtering Function
Function Menu Setup Explanatory Note
Type Digital filtering Digital filtering
Signal source CH1
CH2
REF A
REF B
Filtering Type Low-pass
High-pass
Band-pass
Minimum
Frequency
Maximum
Frequency
II. Reference Waveform
Display of the saved reference waveforms can be set on or off in the [REF] menu. The
waveforms are saved in the non volatile memory of the oscilloscope or an external USB
device and are identified with the following names : RefA, RefB. To display (recall) or
hide (off) the reference waveforms, take the following steps :
1. Press the [REF] menu button on the front panel.
2. Press [CALL BACK]. Select the signal source by turning the multifunction control
knob on the upper part of the front panel. You can choose from 1 to 10. After
selecting a numeral for saved waveform, e.g. 1, press [SELECT] to recall the
waveform which was originally stored in that position. For instructions on storing or
recalling reference waveforms on the USB device, read “Storage and Recall”.
3. Press RefB (RefB reference option). Select the second signal source for the math
function by repeating step 2.
In actual application, when using your UT4000 to measure and observe such
waveforms, you can compare the current waveform with the reference waveform for
analysis. Press [REF] to display the reference waveform menu. For setup please
refer to Table 2-5.
4. To close the reference waveform, press [OFF] in the vertical zone.
Note : When [AUTO] is pressed after a reference waveform is recalled, that waveform
will remain.
⎯
⎯
Set CH1 as filter target.
Set CH2 as filter target.
Set REF A as filter target.
Set REF B as filter target.
Set the filter to low-pass filtering.
Set the filter to high-pass filtering.
Set the filter to band-pass filtering.
Set the minimum frequency with the multifunction
control knob.
Set the maximum frequency with the multifunction
control knob.
23
UNI-T
®
UT4000 User Manual
Table 2-5 Selecting the storage position
Function Menu Setup Explanatory Note
Reference waveform REF A
REF B
Recall Recall a waveform stored in any of the 10 with the
⎯
Import Recall a reference waveform stored on the USB
To select an internal storage position, choose between 1 and 10. In the case of external
storage device, plug in the USB device, then press [F5], then press [F1] and the
multifunction control knob to select the corresponding file name.
To save a reference waveform on the USB device, see the [STORAGE] menu.
Select the REF A reference option.
Select the REF B reference option.
multifunction control knob, then press [SELECT] to
confirm.
device.
Setting Up the Horizontal System
Horizontal Control Knobs
There are two horizontal control knobs : The SED/DIV
horizontal time base, while the horizontal shift knob is for changing the relative position
of the trigger point on the screen.
Horizontal control knob : Display the horizontal menu (see Table 2-6).
Table 2-6
Function Menu Setup Explanatory Note
Window
⎯
Dual time base Time base mode
Time base graticule
⎯
Holdoff Adjust holdoff time with the multifunction
⎯
Channel
M2 shift
Press [F1] to switch between “main window”
and “window extension”.
Main time base or dual time base.
Select the channel to be displayed : CH1 or
CH2.
Select main time base M1 or delay time base
M2.
Select fine tune or coarse tune shift.
control knob.
knob is for changing the
Figure 2-11 Horizontal system control interface
24
UNI-T®
UT4000 User Manual
Definitions of Horizontal Parameter Icons
c The signal frequency currently selected as the trigger source.
d Position of the current waveform triggering point.
e Trigger level of the current waveform.
f The distance between the triggering position and the horizontal centre point (time).
g Time base of the main time base M1, i.e. SEC/DIV.
d ①
g f e
Figure 2-12 Definitions of horizontal parameter icons
Window Extension
Window extension can be used to zoom in a band of waveform to check image details, as
shown in Figure 2-13.
Figure 2-13 Display with the window extended
In the window extension mode, the display is divided into two zones as shown above. The
upper part displays the original waveform, and the lower part displays the extended
waveform. You select the waveform area to be magnified by turning the horizontal
POSITION
knob and SEC/DIV knob.
25
UNI-T
®
UT4000 User Manual
Dual Time Base Function
The dual time base function is similar to window extension but there is a fundamental
difference. In the extension time base mode you can magnify the waveform 100 times,
whereas in the dual time base mode you can magnify details of the waveform being
observed by thousands of times. In effect, the main time base storage depth is increased
by thousands of times.
Dual time base operation instructions :
Press the horizontal setup menu then press [F3] to enter the dual time base menu. Press
[F1] to activate the dual time base function. To open CH1 or CH2 under the dual time
base mode, select by pressing [F2]. To select M1 or M2 as the activated time base, press
[F3] (M2 is the delayed scanning time base).
Figure 2-14 Dual time base
26
UNI-T®
UT4000 User Manual
Setting up the Trigger System
Triggering decides when the oscilloscope collects data and display waveforms. Once the
trigger is correctly set up, it offers stable display of the waveform being observed. When
beginning to acquire data, the digital storage oscilloscope first collects data required for
pre-trigger. When the trigger condition arises, it will continue to acquire the remaining
data.
The trigger control zone on the operation panel of your digital storage oscilloscope
comprises a trigger level adjustment knob, a trigger menu button [MENU], [50%] for
setting up the trigger level at the vertical centre point of the signal, and a compulsory
trigger button [FORCE].
Trigger level : Trigger level sets the trigger point of the signal being tested.
[50%] : Setting the trigger level at the vertical centre point of the trigger signal
amplitude.
[FORCE] : To generate a compulsory trigger signal. Mainly used in the trigger mode and
“Normal” and “Single” modes.
[MENU] : Button for the trigger setup menu.
Trigger Control
Trigger modes : Edge, pulse, video, slew rate and logic analyzer (LA).
Edge trigger : Trigger is set to occur when the signal is at the rising or falling edge. You
can set the trigger level to change the trigger point’s vertical position on the trigger edge,
i.e. the intersection point of the trigger level line and the signal edge on the screen.
Pulse trigger : When the pulse width of the trigger signal reaches a preset trigger
condition, trigger occurs.
Video trigger : Carry out field or line trigger to standard video signals.
Slew rate trigger : Trigger condition is the signal rising or falling rate.
LA trigger : Trigger condition is the pattern or duration input by the logic analyzer. To
use this trigger mode you must have a LA probe installed and turned on.
Below are notes for various trigger menus.
Edge T rigger : For edge trigger menu setups please see Table 2-7.
Table 2-7
Function Menu Setup Explanatory Note
Type Edge
Signal source
selection
CH1
CH2
EXT
EXT/5
AC
CH1 & CH2
Set CH1 as the signal source trigger signal.
Set CH2 as the signal source trigger signal.
Set the external trigger input channel as the signal
source trigger signal.
Divide the external trigger model by 5.
Set grid trigger.
CH1 and CH2 trigger their respective signals
alternately.
27
UNI-T
®
UT4000 User Manual
Table 2-7 (Con’t)
Function Menu Setup Explanatory Note
D0-D15 Set one of the channels between D0-D15 as the
signal source trigger signal (the LA function must be
enabled).
Trigger coupling DC
AC
L/F Suppression
H/F Suppression
Allow AC and DC quantities of the input signal to
pass.
Intercept DC quantities of the input signal.
Reject low frequency quantities below 80kHz of the
signal.
Reject high frequency quantities above 80kHz of the
signal.
Trigger mode Auto
Normal
Single
Set to sample waveform only if no trigger condition
is detected.
Set to sampling waveform only when trigger
condition is met.
Set to sample waveform once when detecting one
trigger and then stop.
Edge type Fall
Rise
Rise, Fall
Set to trigger at the signal’s falling edge.
Set to trigger at the signal’s rising edge.
Set to trigger at the signal’s rising and falling edges.
Pulse T r igger
Pulse trigger means determining the triggering time based on the pulse width. You can
acquire abnormal pulse by setting the pulse width condition.
Table 2-8
Function Menu Setup Explanatory Note
Type Pulse
Trigger source CH1
CH2
EXT
EXT/5
AC
CH1 & CH2
D0-D15
Trigger coupling DC
AC
L/F Suppression
H/F Suppression
Set CH1 as the signal source trigger signal.
Set CH2 as the signal source trigger signal.
Set the external trigger input channel as the signal source
trigger signal.
Divide the external trigger source by 5 to extend the
trigger level range.
Set grid trigger.
CH1 and CH2 trigger their respective signals alternately.
Set one of the channels between D0-D15 as the signal
source trigger signal (the LA function must be enabled).
Allow AC and DC quantities of the input signal to pass.
Intercept DC quantities of the input signal.
Reject low frequency quantities below 80kHz of the
signal.
Reject high frequency quantities above 80kHz of the
signal.
28
UNI-T®
UT4000 User Manual
Table 2-8 (Con’t)
Function Menu Setup Explanatory Note
Trigger mode Auto
Normal
Single
Pulse width setup See Table 2-9 Set the pulse width.
Table 2-9 Pulse width setup
Pulse width
polarity
Pulse width
condition
Positive
Negative
<
>
=
Pulse width time
Return
⎯
⎯
Video Trigger
By selecting video trigger, you can carry out field or line trigger with NTSC or PAL
standard video signals. Trigger menus are as follows :
Table 2-10 Video trigger setup
Function Menu Setup Explanatory Note
Type Video
Trigger source CH1
CH2
EXT
EXT/5
AC
CH1 & CH2
Video setup See Table 2-10a
Table 2-10a Video setup
Standard trigger PAL
NTSC
Synchronization Odd field
Even field
All lines
Specified lines
Return
⎯
The system automatically acquires waveform data when
there is no trigger signal input. The scan baseline is
shown on the display. When the trigger signal is
generated, it automatically turns to trigger scan.
The system stops acquiring data when there is no trigger
signal. When the trigger signal is generated, trigger scan
occurs.
One trigger will occur when there is an input trigger
signal. Then trigger will stop.
Set the positive pulse width as the trigger signal.
Set the negative pulse width as the trigger signal.
Trigger occurs when pulse width is smaller than the set
value.
Trigger occurs when pulse width is larger than the set
value.
Trigger occurs when pulse width equals to the set value.
Set pulse width to 20ns-10s with the multifunction
control knob on the front panel.
Return to Table 2-8.
Set CH1 as the trigger signal.
Set CH2 as the trigger signal.
Set the external trigger input channel as the trigger
signal.
Attenuate the external trigger source 5 times as the
trigger signal.
Set grid as the trigger signal.
Set CH1 and CH2 as alternate trigger signals.
Suitable for PAL video signals.
Suitable for NTSC video signals.
Set the video odd field to synchronized trigger.
Set the video even field to synchronized trigger.
Set the line signal to synchronize with trigger.
Set synchronized trigger on the specified line and adjust
by turning the multifunction control knob on the upper
front panel.
Return to Table 2-10.
29
UNI-T
®
UT4000 User Manual
When PAL is selected for video and synchronization mode is line, you will see a screen
display as shown in Figure 2-15. When synchronization mode is field, you will see a
screen display as shown in Figure 2-16.
Figure 2-15 Video trigger : line synchronization
Figure 2-16 Video trigger : field synchronization
Slew Rate Trigger
When slew rate trigger is selected, trigger occurs if the slew rate of the signal’s rising or
falling edge is the same as the default value. Below is the trigger menu :
Table 2-11
Function Menu Setup Explanatory Note
Type Slew rate
Signal source CH1
CH2
EXT
EXT/5
AC
CH1 & CH2
Trigger coupling DC
AC
L/F Suppression
H/F Suppression
Trigger mode Auto The system automatically acquires waveform data
Set CH1 as the signal source trigger signal.
Set CH2 as the signal source trigger signal.
Set the external trigger input channel as the signal
source trigger signal.
Divide the external trigger source by 5 times.
Set grid trigger.
CH1 and CH2 trigger their respective signals
alternately.
Allow AC and DC quantities of the trigger signal
to pass.
Intercept DC quantities of the trigger signal.
Reject low frequency quantities and allow only
high frequency quantities to pass.
Reject high frequency quantities and allow only
low frequency quantities to pass.
when there is no trigger signal input. The scan
baseline is shown on the display. When the trigger
signal is generated, it automatically turns to
trigger scan.
30
b
UNI-T®
UT4000 User Manual
Table 2-11 (Con’t)
Trigger mode Normal
Single
Slew rate setup Slew rate type
Slew rate of the
slew condition
Threshold
Return
LA Trigger
First enable the LA function, then select LA trigger. Trigger menu is as follows :
Table 2-12
Function Menu Setup Explanatory Note
Type LA Select LA trigger as the trigger type.
Sub-type Pattern
Duration
Trigger qualifier
Trigger mode Auto
Duration setup Condition
Clock setup Clock source
Back
Alternate Trigger
When alternate trigger is selected, the trigger signal will be present in two vertical
channels. This triggering mode is suitable for observing two signals of unrelated signal
frequencies. The Alternate trigger menu can also be used to compare pulse widths.
⎯
Normal
Single
Time setup
Clock edge
⎯
The system stops acquiring data when there is no
trigger signal. When the trigger signal is
generated, trigger scan occurs.
One trigger will occur when there is an input
trigger signal. Then trigger will stop.
Rise/fall setup.
Set slew rate to : larger than/smaller
than/equalling to with the multifunction control
knob on the front panel.
Low level/high level/High low level setup.
Return to the previous menu.
Select pattern or duration as the sub-type.
Turn the multifunction control knob and [SELECT]
key to set the high and low levels for each of the
16-width values.
The system automatically acquires waveform data
when there is no trigger signal input. The scan
aseline is shown on the display. When the trigger
signal is generated, it automatically turns to trigger
scan.
The system stops acquiring data when there is no
trigger signal. When the trigger signal is generated,
trigger scan occurs.
One trigger will occur when there is an input trigger
signal. Then trigger will stop.
Select above, below or equivalent.
Set a duration that satisfies the “condition” with the
multifunction control knob.
Select the source with the multifunction control
knob.
Select the rising or falling edge of the clock.
Return to the previous menu level.
31
UNI-T
®
UT4000 User Manual
Adjusting the Holdoff Time
You can adjust the holdoff time to observe complicated waveforms (e.g. pulse string
series). Holdoff time means the data acquisition cycle of the digital storage oscilloscope.
It is an exact multiple of the cycle of the complicated waveform being tested, thus
synchronization is possible. For example, you can display one group of pulse series stably
by adjusting holdoff time, as shown in Figure 2-17.
For the holdoff menu please see the table below :
Table 2-13
Function Menu Setup Explanatory Note
Window
⎯
Dual time base Time base mode
Time base graticule
⎯
Holdoff Adjust holdoff time with the multifunction
⎯
Channel
M2 shift
Press [F1] to switch between “main window”
and “window extension”.
Main time base or dual time base.
For the selection of displaying CH1 or CH2.
Selecting main time base M1 or delay time
base M2.
Selecting fine tune or coarse tune shift.
control knob.
Figure 2-17 Use the holdoff function to synchronize complicated waveforms
Operation
1. Follow the normal signal synchronization procedure and select the edge and trigger
source in the trigger [MENU]. Adjust the trigger level to make the waveform display
as stable as possible.
2. Press the horizontal [MENU] key to display the horizontal menu.
3. Adjust the multifunction control knob in the upper front panel. The holdoff time will
change accordingly until the waveform display is stable.
Operation tips : Holdoff time is usually shorter than the “big cycle” time. When
observing the RS232 communication signal, it is easier to observe if the holdoff time is
slightly shorter than the starting edge time of every datum.
32
UNI-T®
UT4000 User Manual
Definitions
1. Trigger source : This means the signal for trigger. Trigger can be obtained from
various sources : Input channel (CH1, CH2), external trigger (EXT, EXT/5), grid,
CH1 and CH2 alternate.
Input Channel : The most common trigger source is input channel (choose
CH1 or CH2). The selected trigger source can operate normally whether the
input is displayed or not.
External Trigger : This type of trigger source can be directly input through
the external trigger input terminal. For example, you can use an external clock
or the signal from a circuit to be tested as the trigger source. Both EXT and
EXT/5 trigger sources use external trigger signals from the EXT TRIG adaptor.
EXT can use the signals directly. You can use EXT within the trigger level
range of –3V and +3V. EXT/5 divide the trigger by 5. As a result, trigger range
is extended to –15V to +15V, enabling the oscilloscope to trigger at a large
signal.
Grid : means the grid power source. This trigger mode is suitable for
observing signals related to the grid ⎯ e.g. the correlation between lighting
equipment and power source equipment ⎯ to achieve stable synchronization.
2. Trigger mode : To cause the digital storage oscilloscope to generate a trigger signal.
This oscilloscope offers three trigger modes for selection : auto, normal and single.
Auto Trigger : The system will sample waveform data automatically when
there is no trigger signal input. The scan baseline is shown on the display.
When the trigger signal is generated, it automatically turns to trigger scan for
signal synchronization.
Note : When time base of the scan range is set to 50ms/div or slower, the
“Auto” mode is a continuous scanning process.
Normal Trigger : In this mode, your oscilloscope samples waveforms only
when triggering conditions are met. The system stops acquiring data and waits
when there is no trigger signal. When the trigger signal is generated, trigger
scan occurs.
Single Trigger : In this mode, you only have to press the “Run” button once
and the oscilloscope will wait for trigger. One sampling will occur and the
acquired waveform will be displayed when the digital storage oscilloscope
detects a trigger. Then trigger will stop.
3. Trigger coupling : Trigger coupling determines which quantities of the signal are
transmitted to the trigger circuit. Coupling modes are DC, AC, low frequency
suppression and high frequency suppression.
DC : Allowing all quantities to pass.
AC : Intercepting DC quantities and attenuating signals under 10Hz.
Low Frequency Suppression : Intercepting DC quantities and attenuating low
frequency quantities under 80kHz.
33
UNI-T
®
UT4000 User Manual
High Frequency Suppression : Attenuating high frequency quantities over
80kHz.
4. Pretrigger/Delayed Trigger : Data sampled before/after triggering.
The trigger position is typically set at the horizontal center of the screen. In this case,
you are able to view 5 divisions of pretrigger waveform and 5 divisions of
post-trigger waveform. By turning the horizontal shift knob [POSITION] clockwise
to adjust horizontal displacement of the waveform, you can see more pretrigger
information, i.e see more waveform details before trigger occurs. For example, you
can detect the glitch that occurs when the circuitry starts. Observation and analysis
of pretrigger data can help you identify the cause of glitch.
Setting Up the Sampling System
As shown below, [ACQUIRE] button in the control zone is the function key for the
sampling system.
Figure 2-18 Function key for the sampling system
34
UNI-T®
UT4000 User Manual
Press the [ACQUIRE] button to pop up the sampling setup menu. You can use this menu
to adjust the sampling mode.
Table 2-14 Sampling menu
Function Menu Setup Explanatory Note
Acquisition mode Normal sampling
Peak detect
Average
Envelop
Average number of
times (when average
sampling is selected)
⎯
Storage depth Standard storage
Quick acquisition On
By changing sampling setup, you can observe the consequent changes in waveform
display. If the signal contains considerable noise, you will see the following displays
when average sampling is not selected and when 32-time average sampling is selected.
For sampling waveform display please see Figure 2-19 and Figure 2-20.
2~256 Set the average number of times in multiples
Long storage
Off
Turn on the ordinary sampling mode.
Turn on the peak detect mode.
Set to average sampling and display the
average number of times.
Use the peak detect mode for every
independent acquisition.
of 2, i.e. 2, 4, 8, 16, 32, 64, 128, 256. To
change the average number of times, use the
multifunction control knob shown in Figure
2-18.
DSO storage.
USB drive storage (long storage is disabled
when the logic analyzer is on and dual time
base is in operation).
Acquire at a high screen refresh rate for more
realistic reflection of waveform status.
Close the quick acquisition function.
Figure 2-19 Waveform without average sampling
35
UNI-T
®
UT4000 User Manual
Figure 2-20 Waveform display when 32-time average sampling is selected
Definitions :
Normal mode : The digital storage oscilloscope acquires signal sample at equal and
regular intervals to reconstruct the waveform.
Peak detect mode : In this acquisition mode, the digital storage oscilloscope identifies
the biggest and smallest values of the input signals at each sampling interval and uses
these values to display the waveform. In effect, it can acquire and display narrow pulse
which would otherwise be omitted in the sampling mode. Noise seems to be more
significant in this mode.
Average mode : The digital storage oscilloscope acquires several waveforms and takes
the average value to display the final waveform. You can use this mode to reduce random
noise.
Envelop mode : Waveforms of different amplitudes are acquired. Sampling points of all
relative triggering points recorded at the same time are calculated, and the maximum
value and minimum value are displayed. In the envelop mode, peak detection is carried
out for every single acquisition.
36
UNI-T®
UT4000 User Manual
Setting Up the Display System
As shown below, the [DISPLAY] button in the control zone is the function key for the
display system.
Figure 2-21 Function key for the sampling system (display).
Press the [DISPLAY] button to pop out the setup menu shown below. You can use this
menu to adjust the display mode.
Table 2-15 Display menu
Function Menu Setup Explanatory Note
Type YT
XY
Format Vector
Dot
Grid Full
Grid
Crosshair
Frame
Duration Auto
1s
2s
5s
Infinite
Intensity 1-32 In “quick acquisition” mode, intensity of the
Operating mode of the oscilloscope.
X-Y is the display mode; CH1 is X input, CH2
is Y input.
Sampling points are linked for display.
Sampling points are directly displayed.
Set the grid display mode of the waveform
zone to full, grid, crosshair or frame.
Waveform on the screen is refreshed at normal
rate.
Waveform on the screen is maintained for 1
second and then refreshed.
Waveform on the screen is maintained for 2
seconds and then refreshed.
Waveform on the screen is maintained for 5
seconds and then refreshed.
Waveform on the screen remains on display.
New data will be added continuously until this
function is disabled.
waveform can be adjusted with the
multifunction control knob.
37
UNI-T
®
UT4000 User Manual
X-Y Mode
This mode is suitable for CH1 and CH2 only. After selecting the X-Y display mode, the
horizontal axis will display the CH1 signal, while the vertical axis will display the CH2
signal.
Figure 2-22 Waveform display in X-Y mode
Caution :
In the normal X-Y mode, you can adjust waveform amplitude with VOLTS/DIV of the
two channels. Adjusting the SEC/DIV time base knob will result in a lissajous figure of
better display quality.
The following functions are disabled in the X-Y display mode :
Automatic measurement mode
Cursor measurement mode
Reference or maths waveform
Vector display type
Horizontal [POSITION] knob
Trigger control
Key points :
Display Type : Vector display fills the space between adjacent sampling points in display.
The dot mode displays only the sampling points.
Refresh Rate : Refresh rate is the number of times the digital storage oscilloscope
refreshes the waveform display per second. The refreshing speed affects the capability to
observe signal movements speedily.
38
UNI-T®
UT4000 User Manual
Definitions
Y-T Mode : In this mode the Y axis indicates voltage and the X axis indicates time.
X-Y Mode : In this mode the X axis indicates CH1 voltage and the Y axis indicates CH2
voltage.
Slow Scanning Mode : When horizontal time base control is set at 50ms/div or slower,
the unit will operate in the slow scan sampling mode. When observing low frequency
signals in slow scanning mode, it is advised to set the channel coupling as DC.
S/div : A horizontal scaling (time base) unit. If waveform sampling is stopped (by
pressing the [RUN/STOP] button), time base control can expand or compress the
waveform.
Storage and Recall
As shown below, the [STORAGE] button in the control zone is the function key for the
storage system.
Figure 2-23 Function key for the sampling system (storage)
Press the [STORAGE] button to display the setup menu. You can use this menu to save
and recall waveforms or setup document stored on the oscilloscope built in memory or a
USB device.
Steps :
Press [STORAGE] to go to the type menu. There are three types to choose from :
waveform, setup and bitmap.
1. Select reference waveform to go to the waveform storage menu shown below (see
Table 2-16). A stored reference waveform can be recalled with the [REF] key (see
“Reference waveform”, page 23). For specific steps please see [REF] operation
instructions
39
N
N
UNI-T®
UT4000 User Manual
Table 2-16 Reference waveform storage menu
Function Menu Setup Explanatory Note
Type Reference waveform Select the waveform storage and recall menu.
Signal source CH1
CH2
Storage position 1~10
⎯
Export Document name
Document format
2. Select setup to enter the setup storage menu. See Table 2-17.
Table 2-17 Storage menu setup
Function Menu Setup Explanatory Note
Setup Save the current front panel setup status.
Save
Call back
Import Document name
Export Document name
OK
⎯
⎯
OK
OK
Select the waveform from CH1.
Select the waveform from CH2.
With the multifunction control knob, set and
select the position in which the reference
waveform is to be stored in the built-in
memory.
Select the default name “UNI T0000” for the
reference waveform file being exported to the
USB device with the multifunction control
knob.
Set the format as : internal or CSV.
Export the reference waveform to the USB
device.
Select a save position with the multifunction
control knob on the front panel (total 10 sets)
and save the setup.
Recall the saved setup status.
ame the reference waveform file being
imported from the USB device with the
multifunction control knob.
If the document exists on the USB device, it
will be imported. Otherwise the system will
indicate “I/O operation failure”.
ame the reference waveform file being
exported to the USB device with the
multifunction control knob.
The current front panel setup status is
exported.
40
UNI-T®
UT4000 User Manual
3. Select bitmap to enter the bitmap storage menu. See Table 2-18.
Note : This function can be recalled and used only when the USB device is inserted.
Table 2-18 Bitmap storage menu
Function Menu Setup Explanatory Note
Bitmap Export the display data to the USB device in
bmp format.
⎯
Export Document name
⎯
⎯
About INTERNAL and CSV formats : You can select Internal or CSV format when you
export the reference waveform to the USB device. In Internal format, it can only be
imported to the oscilloscope in REF and not any other interface. In CSV format, it can be
opened in EXCEL as it is a set of 2-dimensional data of voltage and time correlation.
Setting up Utility Functions
As shown below, the [UTILITY] button in the control zone is the function key for
alternative functions.
Default name of the file being exported to the
OK
USB device is “UNI T0000.BMP”.
Press this key to export to the USB device.
Figure 2-24 Function key for the sampling system (function)
Press the [UTILITY] button to pop up the setup menu for utility functions. See Table
2-19.
41
UNI-T
UT4000 User Manual
Table 2-19 Utility functions menu (1)
Function Menu Setup Explanatory Note
System configuration Auto calibration
System information
Clear information
RTC setup
Factory setup
Interface setup Language
Interface style
Menu display
Grid brightness
Pass check Allow testing
Output
Table 2-19 Utility functions menu (2)
Function Menu Setup Explanatory Note
Signal source
Display message
Run
Stop setup
Template setup
Recording waveform See Table 2-21 Setup for waveform recording.
Run auto system calibration.
Display information like oscilloscope
software and hardware versions.
Clear all saved reference waveforms and
front panel status setups.
Set time and date.
Recall factory setup.
Select one of the languages.
Classic, traditional, modern.
Set the waiting time for automatic menu
shutoff ⎯ 5s, 10s, 20s and manual.
Adjust the brightness of the waveform style
zone with the multifunction control knob.
Turn the testing function on or off (this
function is disabled in the dual time base
and roll operation modes).
Set to output signal to pass or fail. The
signal is output through the PASS/FAIL
interface at the rear of the oscilloscope.
Designate testing type as CH1 or CH2.
Open or close the test result.
Run or stop the “PASS/FAIL” test.
Set the stop testing conditions.
Create a module of test conditions.
®
Next page
1/2
Table 2-20 Utility functions menu (3)
Function Menu Setup Explanatory Note
Frequency counter On
⎯
Next page
2/2
42
⎯
Off
⎯
Go to the next page.
Turn the frequency counter on.
Turn the frequency counter off.
Return to the previous menu.
p
UNI-T®
UT4000 User Manual
Table 2-21 Waveform recording menu
Function Menu Setup Explanatory Note
(F4) Stop recording
f(F3) 1. Playback button.
2. When you press this button the system
lays back and displays the number of the
rewinding screen in the top centre area. By
turning the multifunction control knob on
the upper front panel now, you can
suspend playback. Turn the knob further
and you can select the waveform of a
certain screen to be played back
repeatedly.
3. If you wish to continue full playback,
press [F4] to stop and then press [F3].
4. You can record as many as 100 screens
data (If [AUTO] is pressed after a
waveform is recorded, it will be deleted).
(F5)
Important Points :
Auto Calibration : You can correct measurement errors caused by environmental
changes with the auto calibration function. This process can be run selectively when
necessary. To make calibration more accurate, power on your oscilloscope and allow 20
minutes to warm up. Then press the [UTILIITY] button (help function) and follow
on-screen instructions.
Choose your language : Your UT4000 can be operated in many languages. To select a
display language, press the [UTILITY] menu button and then [LANGUAGE] in
“Interface setup” to choose your desired language.
⎯
Record button. Press this button then the
[MENU ON/OFF] key to record. The number
of screens recorded is displayed in the top
centre area (This function is disabled in dual
time base, average, envelop, roll and long
storage modes).
43
UNI-T
®
UT4000 User Manual
Automatic Measurement
As shown below, the [MEASURE] button is the function key for auto measurement. Read
the following instructions to familiarize with all the powerful automatic measurement
functions of your UT4000.
Figure 2-25 Function keys of the sampling system (auto measurement)
Application Examples
The measurement menu of your oscilloscope is capable of measuring 27 waveform
parameters.
Press [MEASURE] to enter the parameter measurement display menu which has 4
options that correspond to function keys F1~F4. To operate any individual option, press
the assigned function keys [F1~F4] to enter the setup menu for that function.
You can measure voltage or time measurement. You can select with the “Customize
Parameters” option. To enter the menu press [F2] and select the required parameters with
the multifunction control knob, then press [SELECT] to confirm (you can display up to 4
parameters). You can also press [F1] and select “All Parameters” to show all voltage and
time measurement parameters of the current channel. To quit “All Parameters”, press
[CLOSE].
Example 1 : To display the measurement peak-to-peak value of CH2, follow the steps
below :
1. Press [CH2] to select channel 2.
2. Press [MEASURE] to enter the measurement functions menu.
3. Press [F2] to enter the measurement parameters options.
4. Turn the multifunction control knob to the peak-to-peak value.
5. Press [SELECT] to confirm, then press [CLOSE] to exit.
The peak-to-peak value is displayed in the left bottom corner.
44
UNI-T®
UT4000 User Manual
Example 2 : Setup for delayed measurement. You can use the delayed measurement
function to measure the time interval between the rising edge of two signal source, i.e. the
time interval between the rising edge of the first cycle of a certain signal source and the
rising edge of the first cycle of another signal source. Measure as follows :
1. As shown in the previous example, select rising delay in the measurement
parameters menu.
2. Press [SELECT] to display the rising delay parameter.
3. Press [CLOSE] to exit the parameter options.
4. The rising delay parameter is displayed in the left bottom corner.
Automatic measurement of voltage parameters
Your UT4000 can automatically measure the following voltage parameters :
Peak-to-peak value (Vpp) : The voltage value from the highest point to lowest point of the
waveform.
Maximum value (Vmax) : The voltage value from the highest point to ground (GND) of
the waveform.
Minimum value (Vmin) : The voltage value from the lowest point to ground (GND) of the
waveform.
Amplitude value (Vamp) : The voltage value from top to base of the waveform.
Mean (Vmid) : Half of the amplitude.
Top value (Vtop) : The voltage value from the level top to ground (GND) of the
waveform.
Base value (Vbase) : The voltage value from the level base to ground (GND) of the
waveform.
Overshoot : The ratio value of the difference between maximum value and top value to
the amplitude value of the waveform.
Preshoot : The ratio value of the difference between minimum value and base value to the
amplitude value of the waveform.
Average value : Average amplitude of signals within 1 cycle.
Root mean square value (Vrms) : The effective value. Energy generated by AC signal
conversion during 1 cycle with respect to the DC voltage that produces equivalent energy,
i.e. root mean square value.
Automatic measurement of time parameters
Your UT4000 can automatically measure the frequency, cycle, rising time, falling time,
positive pulse width, negative pulse width, delay 1→2 (rising edge), delay 1→2 (falling
edge), positive duty ratio, negative duty ratio and 10 time parameters. Definitions of these
parameters are as follows :
RiseTime : The time taken by the waveform amplitude to rise from 10% to 90%.
FallTime : The time taken by the waveform amplitude to fall from 90% to 10%.
Positive Pulse Width (+Width) : The pulse width when positive pulse is at 50%
amplitude.
Negative Pulse Width (-Width) : The pulse width when negative pulse is at 50%
amplitude.
Rising delay (Rise) : Delayed time of the rising edge of CH1, CH2.
Falling delay (Fall) : Delayed time of the falling edge of CH1, CH2.
Positive duty ratio (+Duty) : Ratio of positive pulse width to cycle.
Negative duty ratio (-Duty) : Ratio of negative pulse width to cycle.
45
UNI-T
®
UT4000 User Manual
Measurement menu
Operation : Press [MEASURE] to display 4 function options, as shown in the table below.
Table 2-22
Function Menu Setup Explanatory Note
All parameters Display all parameters. To close press
[CLOSE].
Customized
parameters
Indicator Visual indication of the measurement value’s
Clear measurement Clear the finished parameters.
Note : Customized parameters are for quick parameter measurement. This device has 27
measurement parameters. In general use, the user is not required to measure all
parameters. Only a few have to be tested. In this case, set the parameters you require as
customized parameters. All such parameters will be displayed on the screen.
Up to 4 parameters can be displayed on
screen.
physical properties by lines.
Figure 2-26 Measurement parameters
46
UNI-T®
UT4000 User Manual
Cursor Measurement
Press [CURSOR] to display the measurement cursor and cursor menu, then adjust the
cursor position by turning the multifunction control knob. As shown in the figure below,
[CURSOR] in the control zone is the function key for cursor measurement.
Figure 2-27 Function key for the sampling system (cursor)
There are two cursor measurement modes to choose from the [CURSOR] menu : voltage
and time. When measuring voltage, press [SELECT] on the front panel and the
multifunction control knob to adjust the positions of the two cursors for ΔV measurement.
Likewise, by selecting time, you can measure Δt. The [CURSOR] menu has two other
modes. In the independent mode, you can adjust the positions of the two cursors
independently. In the tracking mode, you press [SELECT] and multifunction control knob
to move two cursors simultaneously.
1. Measuring voltage/time : Cursor 1 or cursor 2 will appear simultaneously. Adjust
their positions on the screen with the multifunction control knob and select which cursor
to adjust with the [SELECT] button. The displayed reading is the voltage or time value
between the two cursors.
2. When the cursor function is enabled, measurement value is automatically displayed
in the top left corner.
Table 2-23
Function Menu Setup Explanatory Note
Type Time
Voltage
Close
Mode Independent
Tracking
Vertical cursor Second
Hz
Set time as the measurement parameter.
Set voltage as the measurement parameter.
Close the automatic measurement function.
Adjust the difference between two cursors
with the multifunction control knob.
Adjust the difference between two cursors
with the multifunction control knob or fix the
difference and adjust the cursor positions.
Select time as the measurement parameter.
Select frequency as the measurement
parameter.
47
UNI-T
®
UT4000 User Manual
Using the Run Button
There is a [RUN/STOP] button on the top right corner of the front panel. When this
button is pressed and a green indicator lights up, your oscilloscope is in a running status.
When the red light comes on after pressing this button again, it indicates the unit has
stopped operation.
Figure 2-28 RUN/STOP button
Auto Setup :
As shown above, Auto Setup can simplify operation. Press [AUTO] and the oscilloscope
can automatically adjust the vertical deflection factor and horizontal time base range
according to the amplitude and frequency of the waveform, so as to ensure stable
waveform display. When the oscilloscope is in auto setup mode, the system setup is as
shown by Table 2-24 :
Table 2-24 Auto setup
Function Menu Setup
Acquisition mode Adjust to “sampling” or “peak detect”.
Display format Set to YT.
SEC/DIV Adjusted according to signal frequency.
Trigger coupling AC
Trigger level Set at 50%.
Trigger mode Auto
Trigger source Set to CH1 but if there is no signal in CH1 and CH2 applies
a signal, it will set to CH2.
Trigger inclination Rising
Trigger type Edge
Vertical bandwidth Full
VOLT/DIV Adjusted according to amplitude of the signal.
48
UNI-T®
UT4000 User Manual
RUN/STOP : Acquire waveform continuously or stop acquisition.
If you want the oscilloscope to acquire waveform continuously, press [RUN/STOP] once.
Press the button again to stop acquisition. You can use this button to switch between
acquiring and stop acquiring waveform. In the Run mode, a green light comes on and
AUTO appears on the screen. In the STOP mode, a red light comes on and STOP appears
on the screen.
Menu Display ON/OFF: To display or hide the current menu
With the “menu ON/OFF” key, you can display or hide the menu currently on screen.
When the menu is hidden, a vertical yellow band will appear on the right side of the
screen
Logic Analyzer (LA) Measurement
There are 16 LA channels.
Maximum sampling rate : 250Msaps.
Storage depth : 512K.
Maximum input voltage : ±40Vpp.
Minimum voltage swing : 1.2 Vpp.
Trigger types available are : edge, pattern, duration.
The following is the LA operation menu.
Table 2-25
Function/Measurement Item Explanatory Note
LA function LA functions ON/OFF.
Channel setup See Table 2-26.
Threshold type Select threshold type : TTL, CMOS, ECL or any user
defined threshold type.
⎯
Sequence Sequence of channels currently open, from small to
large.
49
UNI-T
®
UT4000 User Manual
Table 2-26
Function Menu Setup Explanatory Note
Channel setup Selection
Move
Display
D7-D0
Next page
D15-D8
Waveform height
Previous page
Return
Caution :
To use the LA function, connect the probe of the logic analyzer (including connector,
interface box, connecting cable for the test point being measured and needle) and insert it
into the LA interface at the rear of the oscilloscope. The LA function is then activated.
When the message “LA module installation failure” appears when the LA function is
enabled, check if the LA probe is well connected.
Select the LA channel.
Move the selected channel.
Display or close the selected channel.
Open D7-D0 channels.
Go to the next page.
Open D15-D8 channels.
Enlarge or restore the designated channel.
Return to the previous page.
Return to the previous menu.
Figure 2-29 Composite display of the logic analyzer and CH1 channel
50
UNI-T®
UT4000 User Manual
Multimeter Measurement
Operation instructions : Open the multimeter function menu (Table 2-27) by pressing the
DMM button.
Resistance Activate the resistance measurement function.
On/off Activate the on/off measurement function. When
Diode
⎯
Window shift Set the horizontal shift or vertical shift of the
Notes :
If it is necessary to test electric current, follow these steps :
Insert the current/voltage converter module (UT-M03 or UT-M04) that comes with the
oscilloscope into the digital multimeter port, and move the switch of the current/voltage
converter so that it is consistent with the selected DMM current. Proceed with testing.
Cautions :
1. Please select a suitable current/voltage converter that fits the measurement range,
and select an appropriate range. If you are not certain about the current being tested,
make an estimation, then select a suitable current/voltage converter and set the range
to maximum.
2. The multimeter cannot be used to measure AC current directly.
Voltage Activate the voltage measurement function.
Maximum measurement value for DC voltage is
400V. Maximum measurement value for AC
voltage is 400V.
Maximum measurement value for resistance is
40MΩ.
on/off measuring is lower than 70Ω, the beeper
will sound.
Activate the diode measurement function.
DC current
Activate the DC current measurement function.
Current range : 4mA/40mA/400mA/4A.
window (with the multifunction control knob).
51
UNI-T
®
UT4000 User Manual
Chapter 3 ⎯ Practical Example Illustrations
Illustration 1 : Measuring simple signals
To observe and measure an unknown signal, and to quickly display and measure the
signal’s frequency and peak-to-peak value.
1. To quickly display this signal, follow the steps below :
cIn the probe menu, set the attenuation factor to 10X and set the switch on the probe
to 10X.
d Connect the CH1 probe to the circuitry to be measured.
e Press [AUTO].
The oscilloscope will carry out auto setup to optimise waveform display. In this
status, you can further adjust the vertical and horizontal range until you get the
desired waveform display.
2. Voltage and time parameters for automatic measurement
Your oscilloscope can automatically measure most display signals. To measure
signal frequency and peak-to-peak value, follow the steps below :
c Press [MEASURE] to display the automatic measurement menu.
d Press [F2] to enter the measurement type selection menu.
e Use the multifunction control knob in the top left corner of the front panel to select
peak-to-peak value, then press [SELECT] to confirm. Next use the multifunction
control knob to select frequency then press [SELECT] to confirm.
fPress [MENU ON/OFF] to exit the selection menu.
Peak-to-peak and frequency measurements will now appear in bottom left corner of the
screen.
Figure 3-1 Automatic measurement
52
UNI-T®
UT4000 User Manual
Illustration 2 : Observing the delay caused by a sine wave signal passes through
the circuit
As in the previous scenario, set the probe attenuation factor of the probe and oscilloscope
channel to 10X. Connect the CH1 to the circuit signal input terminal. Connect CH2 to the
output terminal.
Steps :
1. To display CH1 and CH2 signals :
c Press [AUTO].
d Continue to adjust the horizontal and vertical range until you get the desired
waveform display.
ePress [CH1] to select CH1. Adjust vertical position of the CH1 waveform by turning
the vertical position control knob.
fPress [CH2] to select CH2. In the same way described above, adjust vertical position
of the CH2 waveform so that the waveforms of CH1 and CH2 do not overlap. This
will make observation easier.
2. Observing the delay caused by a sine wave signal passes through the circuit and
observing waveform changes.
c When measuring channel delay automatically, press [MEASURE] to display the
automatic measurement menu. Press [F2] to enter the measurement type selection
menu. Use the multifunction control knob to select “rising delay”. You can see the
delay value in the bottom left corner of the screen.
dObserve waveform changes (see the figure below)
Figure 3-2 Waveform delay
53
UNI-T
®
UT4000 User Manual
Illustration 3 : Acquiring single signals
The advantage and special feature of your digital storage oscilloscopes lie in its ability to
acquire non cyclical signals like pulse and glitch. To acquire a single signal, you must
have transcendental knowledge of that signal to set the trigger level and trigger edge. For
example, if the pulse is a logical signal of TTL level, the trigger level should be set at
about 2V and the trigger edge should be set to rising edge trigger. If you are not certain
about the signal, you can observe it by automatic or normal trigger to determine the
trigger level and trigger edge.
Steps :
1. As in the previous illustration, set the attenuation factor of the probe and CH1.
2. Carry out trigger setup.
c Press [MENU] in the trigger control zone to display the trigger setup menu.
d In this menu, use [F1]~[F5] menus to set the trigger type to edge, set trigger source
to CH1, set trigger coupling to AC, set trigger type to single and set edge type to
rising.
e Adjust horizontal time base and vertical range to an appropriate range.
f Turn the TRIGGER LEVEL
g Press [RUN/STOP] and wait for a signal that meets the trigger condition. If any
signal reaches the set trigger level, the system will sample once and display it on the
screen. By using this function you can easily acquire any occasional event. For
example, when a sudden glitch of relatively big amplitude is acquired : set the
trigger level to just higher than the normal signal level. Press [RUN/STOP] and
begin waiting. When a glitch occurs, the machine will automatically trigger and
record the waveform immediately before and after triggering. By turning the
horizontal POSITION
can change the trigger position horizontally to achieve negative delay trigger of
various lengths for easy observation of waveform occurring before the glitch.
control knob to get the desired level.
knob in the horizontal control zone on the front panel, you
Figure 3-3 Single signal
54
UNI-T®
UT4000 User Manual
Illustration 4 : Reducing random noise of signals
If the signal being measured is stacked with random noise, you can adjust the setups of
your oscilloscope to filter or reduce the noise, so it will not cause interference to the
signal during measurement. (Waveform is shown below)
Figure 3-4 Reducing random noise of signals
Steps :
1. As in the previous illustration, set the attenuation factor of the probe and CH1.
2. Connect the signal to ensure stable display of waveform. See the previous
illustration for operation instruction. See the previous chapter for guidance on
adjusting the horizontal time base and vertical range.
3. Improving trigger by setting trigger coupling.
(1) Press [MENU] in the trigger zone to display the trigger setup menu.
(2) Set trigger coupling to low frequency suppression or high frequency suppression. By
selecting low frequency suppression a high-pass filter is set up. It filters low
frequency signal quantities under 80kHz and allows high frequency signal quantities
to pass through. If you select high frequency suppression a low-pass filter is set up.
It filters high frequency signal quantities higher than 80kHz and allows low
frequency signal quantities to pass through. By setting low frequency suppression or
high frequency suppression, you can suppress low frequency or high frequency noise
respectively and achieve a stable trigger.
4. Reducing display noise by setting the acquisition mode.
cIf the signal being measured is stacked with random noise and the waveform is too
coarse as a result, you can use the average acquisition mode to eliminate random
noise display and reduce the size of waveform for easy observation and
measurement. After getting the mean, random noise is reduced and details of the
signal are clearer. Follow the steps below :
Press [ACQUIRE] in the control zone of the front panel to display the sampling
setup menu. Press [F1] then [F3] to set acquisition mode to ADVERAGE, then turn
the multifunction control knob to adjust the average number of times in multiples of
2, i.e. 2 to 256, until you get the desired waveform display that meets observation
and measurement requirements. (See the figure below)
55
UNI-T
®
UT4000 User Manual
Figure 3-5 Signal noise suppressed
dYou can also lower waveform brightness to reduce display noise.
Caution : In the average acquisition mode the waveform display will update at a
slower rate. This is normal.
Illustration 5 : Using cursors for measurement
Your oscilloscope can measure 27 waveform parameters automatically. All auto
parameters can be measured with the cursors. By using cursors, you can quickly measure
the time and voltage of a waveform.
Measuring the frequency of Sinc first peak
To measure the Sinc frequency of the signal rising edge, follow the steps below :
1. Press [CURSOR] to display the cursor measurement menu.
2. Press [F1] to activate the cursor measurement function.
3. Press [F1] again to set cursor type to time.
4. Press [F3] to set the vertical cursor unit to Hz.
5. Turn the multi-function control knob to set cursor 1 at the Sinc first peak.
6. Press [SELECT] to select cursor 2, then turn the multifunction control knob again to
set cursor 2 at the Sinc second peak.
See the figure below.
Note : When using the cursor to measure voltage, follow step 3 only and set the
cursor type to voltage.
56
UNI-T®
UT4000 User Manual
Figure 3-6 Frequency of the cursor measurement signal
Illustration 6 : Using the X-Y function
To check the phase difference between two channels.
Example : To measure the phase variation when a signal passes through the circuit,
connect your oscilloscope to the circuit and monitor the input and output signals. To
check the circuit’s input and output in the X-Y coordinate mode, follow the steps below :
1. Set the attenuation factor of the probe to 10X. Set the switch of the probe to 10X.
2. Connect the CH1 probe to the input terminal of the network. Connect the CH2 probe
to the output terminal of the network.
3. If the channel is not displayed, press the [CH1] and [CH2] menu button to enable the
two channels.
4. Press [AUTO].
5. Adjust the vertical SCALE
channels to about equal.
6. Press the [DISPLAY] menu key to recall the display menu.
7. Press [F1] to select X-Y. The oscilloscope will display the circuit’s input and output
characteristics in a Lissajous figure.
knob to make the displayed amplitude of the two
57
UNI-T®
UT4000 User Manual
8. Adjust the vertical scale and vertical position knob to achieve the best result of
waveform.
9. Using the elliptic oscilloscope display graph to observe, measure and calculate the
phase difference. (See the figure below)
The signal must be
in the centre
Figure 3-7
If sin θ = A
B D ,
θ is the angle of disparity between the channels. For definitions of A, B, C, D see the
above figure. Calculating with this formula, the angle of disparity is
θ = ±arcsin (A
If the elliptic main axis is within quadrants I and III, the angle of disparity should be
within quadrants I and IV, i.e. inside (0~ π 2 2
If the main axis is within quadrants II and IV, the angle of disparity should be within
quadrants II and III, i.e. inside (π
2 2
Furthermore, if the frequencies and phase differences of two signals being measured are
exact multiples, you can calculate the frequency and phase correlation between the two
signals.
or C
) or θ = ±arcsin (C ).
B D
) or (3π ~ 2π).
~ π) or (π ~ 3π).
58
UNI-T®
UT4000 User Manual
Illustration 7 : V ideo signal triggering
To observe an individual video circuit, use the video trigger function to obtain a stable
display of video output signal.
Video field triggering
To trigger in the video field, follow the steps below :
1. Press [MENU] in the trigger control zone to display the trigger menu.
2. Press menu operation key [F1] then [F3] to set the type to video.
3. Press [F2] then turn the multifunction control knob to set trigger source to CH1.
Next press [SELECT] to confirm.
4. Press [F5] then [F1] to select PAL as the video standard.
5. Press menu operation key [F2] to select odd field or even field synchronization.
6. Turn the SEC/DIV
for a clear waveform display.
knob in the horizontal control zone to adjust horizontal time base
Figure 3-8 Video field triggering
59
UNI-T
®
UT4000 User Manual
Video line triggering
To trigger on the video line, follow the steps below :
1. Press trigger [MENU] button in the trigger control zone to display the trigger menu.
2. Press [F1] then [F3] to set the type to video.
3. Press [F2] then turn the multifunction control knob to set trigger source to CH1.
Next press [SELECT] to confirm.
4. Press [F5] then [F1] to select PAL as the video standard.
5. Press [F2] then [F4] to select specified line synchronization.
6. Use the multifunction control knob to select triggering on any line.
7. Turn the horizontal SEC/DIV
horizontal time base for a clear waveform display.
knob in the horizontal control zone to adjust
Figure 3-9 Video line triggering
60
UNI-T®
UT4000 User Manual
Illustration 8 : Pass/fail check
To check whether the input signal is within the standard range. Fail means outside the
range and pass means within range. You can also output the “pass/fail” signal through the
output terminal.
1. Press [UTILITY] then [F3] to enter the pass/fail menu.
2. To set the signal source : Enter the pass/fail menu and press [F3] for setup.
3. Mask setup : Press next page then [F1] to select “” operation. Press [F3] to enter
the mask setup menu. Press [F1] to select the reference waveform. Press [F3] then
the multifunction control knob to set the maximum horizontal capacitance, then [F4]
and the multifunction control knob to set the vertical capacitance (i.e. horizontal
1-100; vertical 1-100). Finally press [CREATE MASK] and return to the previous
menu.
4. To stop setup : press [F2] to enter the stop setup menu. Press [F1] to select the stop
type. Press [F2] to set the stop condition. Press [F3] then turn the multifunction
control knob to set the threshold.
5. Set the output condition : Press [PREVIOUS PAGE] then [F2] to set the output
condition.
6. Begin checking : Press [F1] to enable check., then press [F4] to open the display
information. (This function is disabled in dual time base and roll operation modes)
Figure 3-10 Stop condition not satisfied
Figure 3-11 Stop condition satisfied
61
UNI-T
®
UT4000 User Manual
Illustration 9 : Quick acquisition
In the [ACQUIRE] menu, [F5] is the key for quick acquisition. To observe the rapidly
changes of a waveform, activate this function. The oscilloscope will depicts waveform
strength with dark and light colour tones.
Figure 3-12 Quick acquisition (OFF)
Figure 3-13 Quick acquisition (ON)
Illustration 10 : Using the dual time base function
In the dual time base mode, you can enjoy unrivaled detailed observation of a waveform.
For example, you can observe the colour synchronized signal of the video signal (clear
observation is impossible with ordinary oscilloscopes).
1. To access a signal in CH1, follow the “Video signal trigger” instructions in “Illustration
7 : Video signal triggering”.
2. Press [MENU] in the horizontal control zone to display the time base setup menu.
3. Press [F3] to enter dual time base setup.
4. Press [F1] to select dual time base, then press [F2] to select CH1. Press [F3] to set
time base graticule to M1. Adjust the M1 time base range with the SEC/DIV
shown in Figure 3-14. Then press [F3] to select M2 and adjust the M2 time base range
with the SEC/DIV
5. Shift the M2 time base with the multifunction control knob to show the parts you
require to observe in detail. When shifting M2, press [F5] to choose coarse or fine
tune.
knob as shown in Figure 3-14.
knob as
62
Figure 3-14 Dual time base function
63
UNI-T®
UT4000 User Manual
Illustration 11 : Using the Logic Analyzer
1. Setting the LA channel :
(1) Insert the LA probe into the interface at the rear of the oscilloscope.
(2) Press the LA function key to access the LA menu.
(3) Press [F1] to activate the LA function, then press [F2] for channel setup.
(4) Press [F1] and then the multifunction control knob to select channel. Press [F3]
to open that channel, and then press [F2] and turn the multifunction control
knob to shift that channel.
(5) Repeat step 4 until all required channels are displayed.
(6) Press [F5] to go to the next page. Press [F2] to set waveform height to “High”
then press [F5] to return to the LA menu.
The above channel selection and shifting operations must be set with the
multifunction control knob. If you have to open more than one channels, press [F4] in
the channel setup menu to open channels D0-D7, then press [F5] to go to the next
page and press [F1] to open channels D8-D15. In the LA menu, you can press [F5] to
automatically arrange the channels on display in sequence.
2. Using LA functions :
Set all LA channels as explained above. Input a signal to one or several channels, then
press [MENU] in the trigger zone. Press [F1] to select trigger type to LA trigger, then
press [F2] to select sub-type. By pressing [F3], you can set the trigger qualifier status.
Press [F4] to select trigger mode and [F5] to set the clock (When the sub-type is
“pattern”) or duration (when sub-type is “duration”). The LA can be used
concurrently with the analog channel.
Figure 3-15 LA channel setup
Figure 3-16 Composite display of the logic analyzer and CH1
64
UNI-T
®
UT4000 User Manual
Illustration 12 : Using the multimeter
The multimeter of this digital storage oscilloscope can measure 5 parameters : voltage,
resistance, on/off, diode and DC current. The following example shows the steps for DC
current measurement :
1. Assuming the size of the current to be measured is unknown, a current-voltage
converter module of 4A range should be used. Connect it to the multimeter port of the
oscilloscope then connect the multimeter pen.
2. Press the [DMM] key and [F1] to open the multimeter function table. Press [F2] then
[F5] to select measurement parameter “DC current’. Press [F5] then [F4] to set current
range to “4A”.
3. Connect the multimeter pen to the circuit being tested.
4. If the signal being tested is relatively small, follow step 2 to select an appropriate
current range and use a suitable current-voltage converter module. Adjust the current
range of the current-voltage converter module so it is consistent with that of the
oscilloscope. Next follow step 3 to measure.
Figure 3-17 Selecting a large range for measurement
Figure 3-18 Selecting a suitable range for measurement
65
UNI-T®
UT4000 User Manual
Chapter 4 ⎯ System Prompts and Trouble-shooting
Definitions of System Prompts
Adjustment at Ultimate Limit : This informs you that the multifunction control knob
has reached its adjustment limit in the current status. No further adjust is possible. When
the vertical deflection factor knob, time base knob, horizontal shift, vertical shift and
trigger level adjustments have reached their ultimate limits, this prompt will appear.
USB Device Installed : After a USB device is plugged into the oscilloscope, this prompt
appears when the connection is valid.
USB Device Unplugged : After a USB key is unplugged from the oscilloscope, this
prompt appears.
Saving : When the oscilloscope is saving a waveform, this prompt is shown on the screen.
A progress bar appears at the bottom.
Waiting : This prompt appears on the screen during auto calibration. A progress bar
appears at the bottom.
I/O Failure : When USB communication fails or no document meeting specific
requirements is found on the USB device, this prompt appears.
Function Disabled : When closing the LA function in the LA menu, this prompt appears
if the LA trigger key in the trigger [MENU] is pressed.
Troubleshooting
1. If the screen of your oscilloscope remains black without any display after powering
on, follow the steps below to find the cause :
c Check the power cable connection and check if there is normal power supply.
d Check if the power switch is properly pressed.
e Restart the unit after performing the above checks.
f If the unit still fails to power on, contact UNI-T to request service.
2. If no waveform is displayed after you have acquired a signal, follow the steps
below to find the cause :
c Check whether the probe is properly connected to the signal lead.
d Check whether the signal lead is properly connected to the BNC (channel
adaptor).
e Check whether the probe is connected properly to the object for measurement.
f Check whether the object for measurement is generating signals (connect the
channel with signals to the questionable channel to find the cause).
g Restart the acquisition process.
66
UNI-T
®
UT4000 User Manual
3. The measurement voltage amplitude value is 10 times larger or smaller than the
actual value : Check if the channel attenuation factor corresponds with the probe
attenuation you have chose.
4. There is waveform display but it is not stable :
c Check the trigger source setup in the trigger menu. See whether it is the same as
the actual signal input channel.
d Check the trigger type : Use edge trigger for ordinary signals and video trigger
for video signals. Stable waveform display is achieved only when the correct
trigger mode is selected.
e Try changing the coupling display to high frequency suppression or low
frequency suppression to filter any high or low frequency noise that is
interfering with triggering.
5. No display after pressing [RUN/STOP] :
c Check if the trigger mode is set to normal or single in the menu, and whether
the level exceeds the waveform range. If so, move the level to the centre or set
trigger mode to AUTO.
d Press [AUTO] to complete the setup.
6. Display speed is slower after average sampling time is enabled :
c If average sampling is carried out for more than 32 times, the display speed will
drop. This is normal.
d You can reduce the intervals of average sampling.
7. Trapezoidal waveform display :
c This is normal. The reason is possibly horizontal time base range is too low. You
can improve horizontal pixel aspect ratio and enhance the display quality by
increasing the horizontal time base.
d If the display type is vector, the connection between sampling dots may cause a
trapezoidal waveform. Set the display type to dot to solve this problem.
67
UNI-T®
UT4000 User Manual
Chapter 5 ⎯ Service and Support
Warranty (UT4000 Series Digital Storage Oscilloscopes)
UNI-T (Uni-Trend Electronic (Shanghai) Limited) warrants that product produced and
marketed by it will be free from defects in materials and workmanship for a period of
three (3) years from the date of shipment by the authorized dealer. If any such product
proves defective during this warranty period, UNI-T will repair the defective product or
provide a replacement in exchange according to detailed provisions of the warranty.
To request maintenance and repair service or a full copy of the warranty, please contact
your nearest UNI-T sales and maintenance office.
Save and except the guarantee given herein or in other applicable warranty, UNI-T makes
no other express or implied guarantee, including but not limited to any implied guarantee
on the product’s tradability and suitability for any specific purpose. Under no
circumstance will UNI-T bear any liability for any indirect, special or subsequent loss.
Using the USB Upgrade Program
The USB upgrade program makes upgrade easier and more flexible. To use this function
follow the steps below :
1. The model name, software version and hardware version of your oscilloscope are
listed in Table 2-19.
2. The software that corresponds with the model and hardware version of this
oscilloscope, i.e the update document, is available for download online. Save the
document on a USB device (preferably a USB device recommended by us).
3. Power off the oscilloscope then insert the USB device. Then power on the
oscilloscope again. The prompt “Please press [F5] to continue or [F1] to cancel”, as
shown below, will appear on the screen.
Figure 5-1 USB upgrade confirmation
68
UNI-T
®
UT4000 User Manual
3. After pressing [F5], you will be prompted to select the upgrade document, as shown
in the figure below. You can use the multifunction control knob to select the upgrade
document you require, then press [F5] to go to the next step. (If there is only one
program upgrade document on the USB device, this prompt will not appear and you
can go straight to the next step)
Figure 5-2 Selection of the USB upgrade document
4. When [F5] is presses as shown in Figure 5-2, the following prompt will appear : Are
you sure you wish to update? If yes, press [F5], otherwise press [F1] to cancel
upgrade.
Figure 5-3 Confirming the USB upgrade document
69
UNI-T®
UT4000 User Manual
5. When the USB document is confirmed the following prompt will appear : “System
upgrade is in progress. Please wait”. (This process may take 3 to 5 minutes)
Figure 5-4 USB upgrade in progress
6. When upgrade is complete, a successful upgrade prompt will appear : “Upgrade is
successfully complete. Please remove USB device, then power off your oscilloscope
and power it back on again to complete the process.” The oscilloscope program is
now updated and system upgrade is complete. In the event of upgrade failure, follow
the steps below.
Figure 5-5 USB upgrade success
70
UNI-T
®
UT4000 User Manual
Caution : If upgrade fails due to power interruption, power off the oscilloscope and repeat
all the USB upgrade steps.
Below are some common causes for upgrade failure :
1. No upgrade document.
If there is no document with the suffix “uts”, the following message will appear
when [F5] is pressed after powering on : “Can’t find the valid update file! Please
ensure there is update file on the USB device then restart the operation.”
Figure 5-6 No valid upgrade file
2. The user cancels the upgrade process.
(1) When the oscilloscope is restarted with the USB device plugged in, you can
cancel upgrade by pressing [F1]. To cancel upgrade and operate your
oscilloscope normally, follow the on-screen instructions : “Remove the USB
device and reboot.”
Figure 5-7 Cancel upgrade (1)
71
UNI-T®
UT4000 User Manual
(2) During the upgrade process, you can cancel upgrade by pressing [F1]. To
cancel upgrade and operate your oscilloscope normally, follow the on-screen
instructions : “Remove the USB device and reboot.”
Figure 5-8 Cancel upgrade (2)
(3) You can still cancel upgrade when the oscilloscope is ready for the process.
Simply press [F1] to return to normal operation. Please follow the on-screen
instructions : “Remove the USB device and reboot.”
Figure 5-9 Cancel upgrade (3)
72
UNI-T
®
UT4000 User Manual
3. The USB device is removed before upgrade begins.
If the USB device is removed when the update document is being read before
commencement of upgrade, the following prompt will appear : “USB device has
been removed! Please reboot.” To resume normal operation, power on the
oscilloscope with the USB device unplugged.
Figure 5-10 USB device is removed before upgrade
4. Incorrect upgrade document
If the upgrade document on the USB device is incorrect, update will fail upon
selecting the upgrade document by pressing [F5] (Figure 5-2). You will see the
prompt : “Incorrect upgrade document”. Save a correct document on the USB
device and repeat the previous steps for upgrade.
Figure 5-11 Incorrect upgrade document
73
UNI-T®
UT4000 User Manual
5. Failure in reading upgrade document
After selecting the upgrade document, press [F5] to read. If the document cannot be
read, you will see the prompt : “Sorry, upgrade has failed. Please contact our
after-sales service department”. It may be that our oscilloscope does not support
your USB device, or document reading is unsuccessful due to other reasons. You
can retry and see if the problem can be solved, or use a USB device we recommend.
If the problem persists, contact our after-sales service department. To resume normal
operation, power off your oscilloscope and then reboot.
Figure 5-12 Reading failure
6. Low software version
If the version of USB document is lower than that of the oscilloscope, upgrade will
fail. You will be prompted that, “Software version too low. Please read the user
manual for further instructions.” This indicates that the software version you have
saved on to the USB device is lower than the existing one on the oscilloscope. You
should acquire the correct version either online or through our after-sales service
department. To see version information, press [UTILITY] ⎯ [SYSTEM
CONFIGURATION] ⎯ [SYSTEM INFORMATION]. Please download the suitable
upgrade document.
Figure 5-13 Low software version
74
UNI-T
®
UT4000 User Manual
7. Software document incompatible with product model
If the USB upgrade document is incompatible with the oscilloscope model, upgrade
will fail. You will be prompted that, “Software and product model incompatible.
Please read the user manual for further instructions.” You should download an
upgrade document which corresponds with your oscilloscope model, e.g. UT4102C,
then repeat the USB upgrade steps.
Figure 5-14 Incompatible software document and product model
Contact Us
If you encounter any inconvenience in using this product, contact Uni-Trend Electronic
(Shanghai) Limited (Uni-T, Inc.) in China directly.
Office hours : 8:00am – 5:30pm Beijing Time Monday to Friday.
Alternatively, email us at infosh@uni-trend.com.cn.
For product support outside China, contact your local UNI-T supplier or sales centre.
Service Support : Many UNI-T products offer optional plans of extended warranty period
or calibration period. For details please contact your local UNI-T supplier or sales centre.
A list of service centres worldwide is available online at
http://www.uni-trend.com.cn.
75
UNI-T®
UT4000 User Manual
Chapter 6 ⎯ Appendixes
Appendix A : Technical Indicators
Unless otherwise specified, all technical specifications apply to probes with a attenuation
setting of 10X and the UT4000 Series Digital Oscilloscopes. To verify that your
oscilloscope meets specifications, it must first meet the following conditions :
The oscilloscope must have been operating continuously for thirty minutes within
the specified operating temperature.
If the operating temperature changes by more than 5°C, you must perform the “Self
Cal” operation, accessible through the System Functions menu. All specifications
are guaranteed unless noted “Typical”.
Technical Indicators
Acquisition
Acquisition modes Real time Equivalent
Acquisition rates 2GS/s 50GS/s
(UT4302C/UT4202C/UT4152C)
Average When all channels have made N acquisitions simultaneously, N is
selectable from 2, 4, 8, 16, 32, 64, 128 to 256.
Input
Input coupling DC, AC, GND
Input impedance 1MΩ±1.5% in parallel with 24pF±3pF.
Probe attenuation factor 1X, 10X, 100X, 1000X.
Maximum input voltage 400V (DC + AC Peak, 1 MΩ input impedance)
(10:1 probe attenuation); 5V (Vrms, BNC).
Time delay between channels (Typical) 150ps
76
UT4000 User Manual
Horizontal
Waveform interpolation Sin (x) / x
Storage depth 24K (Max)
Equivalent storage depth
(dual time base)
Recording length 1024K
Scanning range (s/div) 1ns/div-50s/div (300MHz);
Accuracy of sampling
rate and delay time
Time interval
(ΔT) measurement
accuracy (full bandwidth)
60M
2ns/div-50s/div (200MHz, 150MHz);
5ns/div-50s/div (100MHz, 80MHz, 60MHz, 40MHz);
At 1-2-5 increment.
±50ppm (any time interval ≥1ms)
Single : ± (1 sampling time interval + 50ppm x reading +
0.6ns);
> 16 average values : ± (1 sampling time interval + 50ppm x
reading + 0.4ns).
UNI-T
®
77
UNI-T®
UT4000 User Manual
Vertical
Model UT4302C UT4202C UT4152C UT4102C UT4082C UT4062C UT4042C
A/D converter (A/D) 8-bit resolution, two channels sampling simultaneously
Deflection factor V/div
Range (V/div)
Displacement range ≥ ± 10 div
Analog bandwidth 300MHZ, 200MHz, 150MHz, 100MHz, 80MHz, 60MHz,
Single bandwidth 300MHZ, 200MHz, 150MHz, 100MHz, 80MHz, 60MHz,
Selectable analog bandwidth
limit (Typical)
Low frequency response
(AC coupling, -3dB)
Rise time (at BNC, typical)
DC gain accuracy When vertical sensitivity is 1mV/div : ±5% (sampling or
DC measurement accuracy
(average acquisition mode)
Voltage difference (ΔV)
measurement accuracy
(average acquisition mode)
average acquisition mode);
When vertical sensitivity is 2mV/div : ±4% (sampling or
average acquisition mode);
When vertical sensitivity is 5mV/div~10V/div : ±3% (sampling
or average acquisition mode).
When vertical position is zero and N ≥16 :
± (5% x reading + 0.1 div + 1mV) and 1mV/div is selected;
± (4% x reading + 0.1 div + 1mV) and 2mV/div or 5mV/div is
selected;
± (3% x reading + 0.1 div + 1mV) and 10mV/div~10V/div is
selected.
When vertical position is not zero and N ≥16 :
± (5% x(reading + vertical shift reading) + (1% x vertical shift
reading)] + 0.2div). Set from 1mV/div to 200mV/div plus
2mV;
Setup value > 200mV/div to 10V/div plus 50mV.
Under identical setup and environmental conditions, the
voltage difference (ΔV) between two points of the waveform
after the average of ≥16 waveforms acquired waveforms is
taken : ± (3% x reading + 0.05 div).
78
N
UNI-T
®
UT4000 User Manual
Trigger
Trigger sensitivity
Trigger level range
Trigger level accuracy
(Typical) applied on signals
of ≥20ns rise or fall time
Pretrigger capability
Holdoff range 100ns – 1.5s
Set level to 50% (Typical) Input signal frequency ≥50Hz
Edge Trigger
Edge type Rise, Fall, Rise and Fall
Pulse T r igger
Trigger mode (Greater than, less than, or equal to) positive pulse; (Greater
Pulse width range 20ns – 10s
Video Trigger
Trigger sensitivity (video
trigger, typical)
≤ 1 div
Internal ± 8 div from the centre of the
EXT ± 3V
EXT/5 ± 15V
Internal ± (0.3 div x V/div) (within ± 4
EXT ± (6% default value + 40mV)
EXT/5 ±(6% default value + 200mV)
Adjust the frequency
We recommend setting a one-year frequency
Appendix B : Accessories for UT4000 Oscilloscopes
Standard Accessories :
z Two 1.5m, 1:1 (10:1) passive voltage probes. For details please read the probe
manual. EN61010-031: 2002 standard.
Rating is 150V CAT II when the switch is in the 1X position;
Rating is 300V CAT II when the switch is in the 10X position.
z One international standard power cord.
z One User Manual.
z One Product Warranty Card.
z UT4000 communication control software (USB-Device).
z USB cable : UT-D04.
z Two multimeter pens and two current-voltage convertor modules :
UT-M03/UT-M04.
z UT4000 logic analyzer probe: UT-P06.
Optional Accessories :
z UT4000 LAN module : UT-M01
z UT4000 GRIB module : UT-M02
All accessories (standard and optional) are available by order at your local UNI-T dealers.
UT4000
82
UNI-T
®
UT4000 User Manual
Appendix C : Maintenance and Cleaning
General Care
Do not store or leave the oscilloscope where the LCD display will be exposed to direct
sunlight for long periods of time.
Caution: To avoid damage to the oscilloscope or probes, do not expose them to sprays,
liquids, or solvents.
Cleaning
Inspect the oscilloscope and probes as often as operating conditions require. To clean the
exterior surface, perform the following steps :
1. Remove loose dust on the outside of the oscilloscope and probes with a soft cloth.
When cleaning the LCD, use care to avoid scratching the clear protection film of the
LCD.
2. Use a dampened but not dripping soft cloth to clean the oscilloscope. Remember to
disconnect power. Use a mild cleaner or water. To avoid damaging the product or
probe, do not use abrasive chemical cleaner.
Warning : To avoid short circuit or personal injury due to the presence of moisture,
please ensure the product is completely dry before reconnecting power for operation.