Unitest 9061 Instruction Manual

TELARIS Fi/RCD-Analyzer
UNITEST
®
Instruction Manual Cat. No. 9061
®
2
English
3
Contents ........................................................................................Page
1.0 Introduction / Scope of Supply ..............................................4
1.1 Model and Type Designation ..................................................4
1.2 Productdescription ................................................................4
2.0 Transport and Storage............................................................5
3.0 Safety Measures ....................................................................5
3.1 Appropriate Usage ................................................................7
4.0 Display / Operation Elements ................................................7
5.0 Carrying out Measurements ..................................................8
5.1 Voltage Measurement ............................................................9
5.2 General Information about RCD Tests ....................................10
5.2.1 Information regarding RCD testing ........................................10
5.2.2 Starting of RCD Test without RCD tripping,
Function Vc/RE ......................................................................11
5.2.3 Starting RCD Test with RCD Tripping, Funtion tRCD/Vc..........12
5.2.4 Starting RCD Test with “RCD Tripping, Function Auto” ..........14
5.2.5 Starting Automatic RCD Test with RCD Tripping ....................16
6.0 Storing Measurement Data ....................................................17
6.1 Infrared Interface, send Measurement Data ..........................18
7.0 Deleting Stored Measurement Data / Display of
All Memory Entries ................................................................18
8.0 Energy Management ..............................................................19
9.0 Maintenance ..........................................................................19
9.1 Cleaning ................................................................................19
10.0 Battery Replacement ............................................................19
11.0 Internal fuse ..........................................................................20
11.1 Indication in the display at defect fuse ..................................20
12.0 Calibration Interval ................................................................20
12.0 Technical Data ......................................................................21
24 month Warranty ..............................................................26
References marked on instrument or in instruc­tion manual:
Warning of a potential danger, comply with in­struction manual.
Reference. Please use utmost attention.
Caution! Dangerous voltage. Danger of electri­cal shock.
Continuous double or reinforced insulation complies with Category II.
Warning of potential danger caused by accu­mulators and batteries.
Conformity symbol, the instrument complies with the valid directives. It complies with the EMC Directive (89/336/EEC), Standards EN 50081 and EN 50082-1 are fulfilled. It also complies with the Low Voltage Directive (73/23/EEC), Standard EN 61010-1 is fulfilled.
The instruction manual contains information and references necessary for safe operation and maintenance of the instrument. Prior to
using the instrument (commissioning / as­sembly) the user is kindly requested to thor­oughly read the instruction manual and com­ply with it in all sections.
Failure to read the instruction manual or to comply with the warnings and references con­tained herein can result in serious bodily injury or instrument damage.
1.0 Introduction / Scope of Supply
You have purchased a high quality measurement in­strument of Ch. BEHA GmbH wich will allow you to carryout measurement over a long time period. The company Ch. BEHA GmbH is a member of the world
-wide operating BEHA Group with its head office in Glottertal/Schwarzwald wich also houses our deve­lopement centre. The BEHA Group is a leading or­ganisation for Test Measurement Instruments.
1.1 Model and Type Designation
The identification label is located on the rear of the instrument. It contains the instrument serial num­ber and product designation. When questions arise regarding the instrument, please always quote product designation and serial number.
1.2 Productdescription
The UNITEST TELARIS FI/RCD Analyzer allows the testing of trip times of standard RCDs and selective RCDs using various residual currents. Additionally, the ground resistance and contact voltage can be measured.
• Testing of trip times of standard RCDs and sel­ective RCDs using various residual currents.
Testing of conventional, selective and all-cur­rent sensitive RC protection switches
Automatic RCD test (all RCD tests are carried out automatically)
Nominal residual currents may be selected: 10, 30, 100, 300, 500, and 1000mA
Automatic phase detection
Various curve shapes: Sine,1/2 wave rectified DC current, DC current
Start of measurement at 0° or 180°
Display of: trip time, trip current, contact voltage, ground resistance
Result assessment, result evaluation
Data memory for up to 100 measurement resutls
Infrared interface (RS232) for easy data transfer
Energy management with integrated auto power off function
Large, easily readable display
Built in compliance with EN 61557 Part 1 and 6, EN 61010-1, DIN VDE 0413 Part 1 and Part 6,
4
Introduction
English
5
Transport and Storage / Safety Measures
Scope of Supply
1 pc UNITEST TELARIS Fi/RCD-Analyzer 1 pc 3 pole test lead 1 pc Test lead with mains plug 1 pc Croccodile clamps 3 pc Test probe 1 pc Protective Holster 1 pc Carrying Case 6 pcs Battery 1.5V, type IEC LR6 (Size AA) 1 pc Instruction manual
2.0 Transport and Storage
Please keep the original packaging for later trans­port, e.g. for calibration. Any transport damage due to faulty packaging will be excluded from warranty claims. In order to avoid instrument damage, we recom­mend that batteries are removed when not using the instrument over a certain period of time. How­ever, should the instrument be contaminated by leaking battery cells, you are kindly requested to re­turn it to the factory for cleaning and inspection. Instruments must be stored in dry and closed areas. In the case of an instrument being transported in ex­treme temperatures, a recovery time of at least 2 hours is required prior to instrument operation.
3.0 Safety Measures
The UNITEST Telaris FI/RCD Analyzer has been de­signed and checked in accordance with the safety regulations for Electronic test and Measurement In­struments EN 61010 and IEC 61010, and left our factory in a safe and perfect condition.
The instruction manual contains information and references necessary for safe operation and maintenance of the instrument.
The respective accident prevention regulations established by the professional association for electrical systems and equipment must be strictly met at all times.
In order to avoid electrical shock, the valid safe­ty and VDE regulations regarding excessive contact voltages must receive the utmost at­tention when working with voltages exceeding 120V (60V) DC or 50V (25V)rms AC. The val­ues in brackets are valid for limited ranges (as for example medicine and agriculture).
Measurements in dangerous proximity of elec­trical installations are only to be executed when instructed by a responsible electrical specialist, and never alone.
If the operator’s safety is no longer guaranteed, the instrument is to be put out of service and protected against use. The safety can no longer be guaranteed if the instrument (or leads):
• shows obvious damage
• does not carry out the desired measurements
• has been stored for too long under un-
favourableconditions
• has been subjected to mechanical stress dur-
ing transport.
Avoid any heating up of the instrument by di­rect sunlight to ensure perfect functioning and long instrument life.
Prior to usage, inspect the instrument and test leads for external damage. Prior to any opera­tion, ensure that connecting leads used and in­struments are in perfect condition.
The instrument may only be used within the op­erating ranges as specified in the technical data section.
3.1 Appropriate Usage
The instrument may only be used under those conditions and for those purposes for which it was built.
When modifying or changing the instrument, the operational safety is no longer guaranteed.
The opening of the instrument for fuse re­placement, for example, may only be carried out by professionals. Prior to opening, the in­strument has to be switched off and discon­nected from any voltages.
Any maintenance and calibration tasks may only be carried out by our repair service staff.
The instrument may not be operated if the bat­tery case is open.
If the instrument is subjected to an extremely high electro-magnetic field, its functioning abil­ity may be impaired.
4.0 Display / Operation Elements
Display
1. Attention, warning symbol
2. Symbol “Limit” (Limitvalue)
3. Excess (breech) of contact voltage limit
4. Socket error
5. Battery status indication
6. Unit display
7. Display for selective RCD
8. Phase position nominal residual current
9. Contact voltage limit
10. Current type
11. Selected nominal residual current
12. Nominal residual current - multiplication factor
13. Symbol for memory entry
14. Measurement data display
15. Symbol for ramp function
16. Temperature overload, overheating of the in­strument
17. AC voltage, display of mains voltage
Operation Elements
18. LCD
19. Key "VC/UL" to select contact voltage limit
20. Nominal residual current selection
21. PE contact electrode
22. Key "Start"
23. Key "Current Type"
24. Key "Phase Position" for nominal residual cur­rent
25. Key "Store
26. Key "Send"
27. Key "Clear"
28. Battery case (instrument rear)
29. Infrared RS232 PC interface
30. Key - "Multiplication Factor" for nominal residu­al current
31. Function selection switch
32. Display key to select individual measurement results
6
Display / Operation Elements
English
7
Carrying Out Measurements
5.0 Carrying out Measurements
FAT (Final Approval Test) measurements have to be carried out in compliance with the ap­propriate applicable standards.
IN:
2.5x
180¡
1
2
3
4
5
6
7
8
9
11
12
13
15
14
16
17
10
IN:
2.5x
180¡
Display
VC / UL
Store
Send
Clear
Vc/RE
tRCD/Vc
tRCD/Vc
18
19 20
21
22 23
24
25
32 20
31
30
29
28
2627
5.1 Voltage Measurement
Prior to usage the test instrument and the test leads have to be tested for correct functioning.
Test leads and test probes may only be touched at the handles provided. The user must not touch the test probes.
The measurement connectors may not be con­nected to external voltages exceeding 300 V AC or DC to avoid any instrument damage.
When changing the measurement function or parameter IN and UL parameters, the TELA­RIS Fi/RCD-Analyzer automatically switches to the voltage measurement function.
Select measurement function using the function
selection switch (31).
Connect test lead /mains connection lead to the
UUT, as described in figure 1 or 2.
Read measurement data on the display.
No data storing possible !
8
Voltage Measurement
IN:
P
IN:
Fig. 1
Fig. 2
English
9
General Information about RCD Tests
5.2 General Information about RCD Tests
The contact voltage UB and the trip time t required by the RCD to disconnect the subsequent current circuit from the mains represent important mea­surement units for the asessment of an RCD.
For this reason, IEC 60364 prescribes that
a) the maximum allowable value for the contact
voltage (25V/50V) may not be exceeded within any system during tripping at nominal residual current.
b) the RCD must trip within a time limit of 300ms.
The task of an residual current device (RCD) con­sists in switching off a system within a defined time period after an error prior the contact voltage reach­es the permissible limit value of 25V / 50V.
The system testing should be started by carrying out a visual inspection, in particular of the protective earth connection.
1. Within the IT system, the protective earth con­ductor doesnot have to be connected with the PEN but with the protective earth connection.
2. The protective earth conductor must be connect­ed to the PEN prior to the RCD within the TN sys­tem.
3. An insulation measurement as described in Sec­tion 5.2 must be performed. In particular it must be proved, that there is no connection between N and PE following the RCD.
4. Proof regarding the low impedance connection of equipotential bonding conductors in compliance with Section 5.4 must be available.
Time-delayed residual current devices trip at nominal residual current within 130…500ms, for double nominal fault current within 60…200 ms. Such RCDs are implemented as main residual current protection devices (please refer to IEC 61008-1) and are marked with the symbol “ S ”.
The contact voltage represents the voltage pre­sent during an insulation error between two si­multaneously touchable components.
At a measuring circuit without probe, available voltages between PE and earth can influence the measurement.
Before using the N-conductor as probe check that all neutral points have low ohm resistance to the main neutral line. A available voltage of the N conductor to the arth can influence the measurement.
The measuring function uses the N-conductor as a probe. Check first the connection between the neu­tral point of the distribution system and earth be­fore the test is started. A possible voltage between the N-conductor and earth may influence the mea­surements.
Leakage currents by preconnected loads can influence the measuring
Attached loads or operating supplies which contains capacitors or circulating machines can elongate the trip time.
Any test and measurement procedures in cir­cuits equipped with residual current devices should only be performed after having con­sulted the operator terminals (data processing systems, material processing, motors, etc.).
Prior to testing, we recommend all loads are switched off as they could falsify the measure­ment result.
The protective earth must be free of external voltage for the RCD test. However if an extra­neous voltage is present, the instrument only indicates the voltage VC having been generat­ed by the measurement. The measurement in­terruption caused by excess of UL by VC is only generated by the actual voltage present be­tween the neutral conductor (N) and the pro­tective earth (PE).
5.2.1 Information regarding RCD testing
The instrument TELARIS RCD Analyzer allows the measurement of trip times for standard RCDs and selective RCDs for different residual currents. Ad­ditionally, the ground resistance and the contact voltage can be measured without RCD tripping. The additional ramp function allows determination of the exact residual current and trip time. The instru­ment is equipped with an automatic RCD function to carry out fully automatic RCD tests.
Table 1 and 2 informs about the type of cur­rent to be selected and about multiplication fac­tors of nominal residual current.The multipli­cation factors may not be selected within all measurement functions and are in these cases disabled for the user.
The nominal residual current selection also al­lows selection of selective RCDs (not possible for RCD TEST I/ta/UB). When selecting the selective RCD, the symbol "S" (7) appears on the display.
To carry out a complete socket test, the contact electrode PE (21) must be touched. If the sym­bol socket error (4) appears, a PE error is pre­sent.
If the instrument is switched off using function selection switch or if the instrument automati­cally switches off due to the automatic power off function, the pre-set nominal residual cur­rent value is maintained. Parameters such as phase position (0°, 180°), current type (AC, DC Plus, DC) and the nominal residual current multiplication factor are reset to the standard (default) values (0°, AC, x1).
If many measurements are carried out inter­rupted only by short breaks, the built-in over­temperature protection of the RCD Analyzer might respond, and the thermometer symbol (15) appears on the display. Subsequent mea­surements can only be carried out, once the in­strument has cooled down and the thermome­ter symbol has gone out, thus avoiding any instrument damage. Any tests at nominal resi­dual currents 10mA / 30mA may be started even if the thermometer symbol is on, as these currents do not lead to a heating up of the in­strument.
If the RCD trips during testing, the socket error symbol (4) is displayed. This symbol is also displayed if the RCD is not in order and an in­correct tripping has been caused!
The "Attention Symbol" indicates that one measure­ment unit has exceeded a limit value. When calling up the measurement data using "Display" key (18), the "Limit" symbol (2) or "UB>UL" (3) appears for the respective measurement unit.
10
Starting of RCD Test
IN:
IN:
IN:
English
11
Starting of RCD Test
5.2.2 Starting of RCD Test without RCD trip-
ping, Function UB/RE
The contact voltage UB and the ground resistance RE are measured during this test by applying a cur­rent amounting to 40 % of the nominal residual cur­rent. The contact voltage displayed is extrapolated to nominal residual current IN or to double the no­minal residual current IN for selective RCDs.
Connect test lead / mains connection cable to
test instrument.
Connect test lead / mains connection cable to
the UUT, as described in figure 1 or 2.
Select desired measurement function Vc/RE
using function selection switch (31).
Set nominal residual current selection (20) to
desired nominal current (depending on RCD).
Select Limitvalue via contact voltage using key
"Vc/UL" (19).
Touch contact electrode PE (21). Note display.
If the symbol "Socket Error" (4) is displayed, a PE error is present.
Press key "Start" (22).
After successful test the display key (32) is used to switch between measurement result.
Display at contact voltage
Display at earth resistance
Measurement results can be saved via pres­sing key “Store” (25).
If the condition of a faulty socket is to be saved, first press key "Start" (22). Upon hearing the "Error Signal" press the key "Store" (25).
If the RCD trip, the "Attention" symbol (1) and "Socket error" symbol (4) are displayed
IN:
Display at RCD has tripped
A possible reason could be:
incorrect selection of test current.
•a residual current is present within the net to be tested, causing the RCD to trip early because it sums up with the test currents.
RCD faulty.
The calculated contact voltage is displayed up to an votage of 70 V
The measurement is interrupted if the contact vol­tage exceeds the set contact voltage limit. "Attenti­on" (1) and "UB>UL" (3) appears on the display.
5.2.3 Starting RCD Test with RCD Tripping, Funtion tRCD/UB
The contact voltage VC is measured during a test, by applying a current amounting to 40 % of the no­minal residual current. The contact voltage display­ed is extrapolated to nominal residual current I∆N or to double the nominal residual current IN for selective RCDs.
Upon the examnation of RCDs a break of 30 sec is had betweeen pre-test and preliminary test.
Connect test lead / mains connection cable to
test instrument.
Connect test lead / mains connection cable to
the UUT, as described in figure 1 or 2.
Select desired measurement function tRCD/UB
using function selection switch (31).
Set nominal residual current selection (20) to
desired nominal current (depending on RCD).
Set phase position using key 0° / 180° (24).Set current type (23) and multiplication factor
using key (30).
12
Starting of RCD Test
IN:
2 x
IN:
IN:
IN:
English
13
Starting of RCD Test
The current types and multiplication factors to be selected at set nominal residual current are listed in tables 1 and 2.
Select contact voltage using key "VC" (19).Touch contact electrode PE (21). Note display.
If the symbol "Socket Error" (4) is displayed, a PE error or other socket error is present.
Press key "Start" (22).
After successful test the display key (32) is used to switch between measurement result tRCD and VC within the measurement value display.
Display at trip time (tRCD)
Display at contact voltage (VC)
The measurement is interrupted if the contact voltage exceeds the set contact voltage limit. "Attention" (1) and "UB>UL" (3) appears on the display.
Display at limit of contact voltage is exceed
If the RCD does not trip within correct tolerance va­lues (also refer to table 3), the "Attention" symbol (1) and "Limit" (2) are displayed besides measure­ment value tRCD. A possible reason could be:
incorrect selection of test current.
•a residual current is present within the net to be
tested, causing the RCD to trip early because it sums up with the test currents.
RCD faulty.
Display at trip time >300 ms
Measurement results may be saved pressing key Store (25).
IN:
IN:
IN:
ms
IN:
If the condition of a faulty socket is to be saved, first press key "Start" (22). Upon hearing the "Error Signal" press the key "Store" (25).
5.2.4 Starting RCD Test with “RCD Tripping, Function Auto”
For this test, the trip current I of the RCD is mea­sured. The contact voltage VC is measured during the preliminary test at a current of 40%IN. The RCD test is resumed after the successful prelimi­nary test at a residual current, rising in steps of 10 % from 40 % IN until maximum 140 % IN. The presently active residual current is indicated on the display and in the event of the RCD tripping the trip time of the RCD is meausured. The contact voltage VC is evaluated for the trip current I.
The method of the rising test current is not ne­cessarily binding in compliance with VDE, ho­wever represents a useful aid during trouble shooting.
Test and measurement instruments within net­works being protected by RCDs should only be used after agreement with the final instrument user (DP systems, process technology, mo­tors, etc.).
If the contact voltage VC already exceeds the pre­set contact voltage limit UL during the preliminary test, the measurement is disabled and the "Attenti­on" symbol (1) and "UB>UL" (3) is displayed.
Connect test lead / mains connection cable to
test instrument.
Connect test lead / mains connection cable to
the UUT, as described in figure 1 or 2.
Select desired measurement function (I
/tRCD/VC) using function selection switch (31).
Set nominal residual current selection (20) to
desired nominal current (depending on RCD).
No possibility to check selective RCDs.
Set phase position using 0°/180° key (24) and
current type by means of key (23).
Select contact voltage using key "VC/UL" (19).Touch contact electrode PE (21). Note display.
If the symbol "Socket Error" (4) appears a PE error is present.
Press key "Start" (22).
After a successful test, it is possible to select between measurement results trip current I<, trip time tRCD, and contact voltage VC in the display (14) using display key (18).
Display trip current
Display trip time
14
Starting of RCD Test
IN:
IN:
IN:
English
15
Starting of RCD Test
Display contact voltage
The measurement results may be saved by pressing key Store (25).
If the condition of a faulty socket is to be saved, first press key "Start" (22). Upon hearing the "Error Signal", press key "Store" (25).
If due to the ramp current the contact voltage is exceeded, the measurement is interrupted and the current is displayed at the level at which the excess occurred.
If trip time tRCD of RCD exceeds 300ms, the mea­surement value for an trip time is displayed as ">300ms". The "Attention" symbol (1) is displayed and for trip time display, the symbol for overload (2) appears.
To make sure that the RCD is tripping correctly, a standard test (tRCD/VC) has to be carried out. Ex- ample: A correctly functioning RCD has a trip time of ">300ms" at 50% IN, whereas the trip time smaller than 300ms. When performing standard test at I∆N.
If the RCD already trips at 50 % IN or not at all at a residual current value higher than 100 % IN, the RCD is faulty or the incorrect nom­inal residual current has been selected. The at­tention symbol (1) is displayed and the symbol for exceeding the limit (2) is indicated togeth­er with the trip current (tRCD) display
IN:
IN:
IN:
IN:
5.2.5 Starting Automatic RCD Test with RCD Tripping
The TELARIS RCD Analyzer is equipped with a fully automatic RCD analysis. As a first step, during a preliminary test, the contact voltage VC is measu­red at a current of 40 % IN and extrapolated to the nominal current. Thereafter, various AC residual currents are generated and both trip time and contact voltage are verified. The user must only make sure to switch on RCD after tripping between the individual test programs. Then, the RCD analy­sis is automatically continued.
Connect test lead / mains connection cable to
test instrument.
Connect test lead / mains connection cable to
the EUT, as described in figure 1 or 2.
Select desired measurement function (Auto)
using function selection switch (31).
Set nominal residual current selection (20) to
desired nominal current (depending on RCD).
Select contact voltage using key "VC/UL" (19).Touch contact electrode PE (21). Note display.
If the symbol "Socket Error" (4) appears a PE error is present.
Press key "Start" (22).
The measurement is interrupted if the contact voltage exceeds the pre-set contact voltage limit. The following symbols appear: "Attention (1), "UB>UL" (3) and "Err
After successful preliminary test the following is displayed (14):
“Pr 1” equals 0.5x IN 0°, RCD must not trip “Pr 2” equals 0.5x IN 180°, RCD must not trip “Pr 3” equals 1x IN 0°, RCD must trip within
300ms (S-Type 130 ... 500ms) (test-time 500ms)
“Pr 4” equals 1x IN 180°, RCD must trip wit-
hin 300ms (S-Type 130 ... 500ms)
“Pr 5” equals 2x IN 0°, RCD must trip within
150ms (S-Type 60 200ms)
“Pr 6” equals 2x IN 180°, RCD must trip wit-
hin 150ms (S-Type 60 ... 200ms)
“*Pr 7” equals 5x I∆N 0°, RCD must trip within
40ms (S-Type 50 ...150ms)
“*Pr 8” equals 5x I∆N 180°, RCD must trip wit
hin 40ms (S-Type 50 ... 150ms)
*function not to be executed at 500mA and 1000mA RCDs
16
Starting of RCD Test
IN: .5x
IN: .5x
English
17
Storing Measurement Data
The RCD trips several times during measurement and has to be switched on again. Thereafter, the measurement is automatically continued and ends with the display "POS" or "Err" on the display (18).
Display at functioning RCD
Display at faulty RCD
"POS" is an abbreviation of positive and signifies that the RCD is functioning correctly. If "Err" (Error) is displayed, the measurement is immediately inter­rupted. "Err" indicates a defective RCD or that the contact voltage has been exceeded. The nominal re­sidual current multiplication factor (12) together with the phase position of nominal residual current (9) indicates in which program the automatic test has been interrupted.
The measurement results may be stored by pres­sing key Store (25). It is advised to store when using this range, as the Report Studio Software enables a detailed decoding of the measurement results.
To interrupt the function AUTO, change the measurement range using the selection switch (31)
Possibility to save.
6.0 Storing Measurement Data
After successful measurement, it is possible to save measurement value by means of "Store” key (25). Alltogether, approximately 100 measurement va­lues can be saved. Proceed as follows:
Carry out measurement.Store measurement result by pressing "Store"
key (25).
After successful storeing a short signal is au-
dible and the symbol "M" (13) is displayed, also, the memory address number briefly appears.
In the event of pressing the Store-key when the memory is full, the “M” Symbol blinks briefly and an error signal is heard
The data in the memory is maintained if in­strument is switched off or during battery re­placement.
Double storing of a measurement is impossi­ble !
Under certain, unintentional circumstances data may be lost or modified during electro­nic saving. CH. BEHA will not be held liable for financial and other loss being caused by data loss, wrong handling or any other rea­son.
We recommand you to save and transmit the measurement results daily to your computer. In case of incidentals outer influences you can lose saved measurement sof the instru­ment.
IN:
5x
180¡
IN:
2 x
6.1 Infrared Interface, send Measurement Data
Additionally, all memorised measurement data may be transferred to the PC (data archive) by means of an Infrared Interface. For this purpose, the interfa­ce adapter and the Software "Report-Studio” (Cat. No. 1206) and Interface (Cat. No. 1157), available as an optional feature, are required.
The data transfer procedure is fairly easy:
Apply the Interface adapter to TELARISConnect adapter cable to PC (COM 2) via 9pole
D-sub plug (RS232).
Call the "Report-Studio” Software (please refer
to user manual "Report-Studio”).
Select menu item "read data” or select the but-
ton "read data from measurement instrument” by mouse click. The message "send data” ap­pears.
Press key "Send" (26).
Data transfer is carried out. After a few seconds all stored data is available for further proces­sing in the PC.
For further details, please refer to user manual "Report-Studio".
7.0 Deleting Stored Measurement Data / Display of All Memory Entries
It is possible to delete data by using key „Clear" (27). Furthermore, it is possible to display the total num­ber of measurement value entries on the display:
Display of Total Number of Measurement Value Entries:
Set measurement range selection switch (31) to
one of the 5 measurements ranges.
Briefly press „Clear” key (27). A short beep sig-
nal is audible.
The total number of all measurement values that
have been saved is now displayed.
Delete Last Stored Measurement Value:
Set measurement range selection switch (31) to
one of the 5 measurement ranges.
Briefly press „Clear” key (27). A short beep sig-
nal is audible and the number of the memorised measurement data is displayed.The total num­ber of all stored measurement data is dis­played.The display disappears after approxi­mately 2 seconds.
Now release the key „Clear” (27) to avoid dele-
tion of all memory entries.
Deleting the last measurement value can be re­peated as often as required, as long as data is available within the memory.
Deleting all Stored Measurement Values:
Set measurement range selection switch (31) to
one of the 5 measurement ranges.
Press „Clear” key (27) for approx. 5 seconds.A beep signals are audible. The number of all
stored measurement values is displayed.
The last stored value is deleted after 2 seconds.
After 5 seconds, all values are deleted.
„0“ appears on the display. The memory is com-
pletely deleted, the „M" symbol in the display disappears.
18
Infrared Interface, send Measurement Data
English
19
Maintenance / Battery Replacement
8.0 Energy Management
Approximately 5 minutes after last key operation, the instrument switches off automatically (auto-power­off). To switch the instrument on again, turn rotary switch from the „OFF" position to the desired func­tion. If the batteries are almost completely discharged, the battery symbol (5) appears. The instrument continu­es functioning without necessarily respecting the specifications. If the batteries continue to be used and discharged, the instrument switches off at a point when the mini­mum battery voltage level is reached and may not be switched on again.
9.0 Maintenance
When using the instrument in compliance with the in­struction manual, no special maintenance is required.
9.1 Cleaning
If the instrument is dirty after daily usage, it is advised to clean it by using a damp cloth and a mild household detergent.
Prior to cleaning, ensure that the instrument is switched off and disconnected from external volt­age supply and any other instruments connected (such as EUT, control instruments, etc.).
Never use acid detergents or dissolvants for clean­ing.
10.0 Battery Replacement
If the battery symbol appears in the display, pro­ceed with battery replacement.
Prior to battery replacement remove the instru­ment from all measurement circuits.
The right order of the batteries is shown on the inside of the battery compartment.
Batteries with reversed polarity can lead to in­strument destruction. Furthermore, they may explode or ignite. Only use batteries as described in the technical data section type 6LR61.
Switch off the instrument.Loosen screws on the instrument rear.Remove battery case cover (by slightly knocking
the instrument in the palm of the hand).
Remove discharged battery.Insert new battery by respecting the correct po-
larity.
Replace battery case cover and tighten screws.Instrument is operational.
Never try to disassemble a battery cell. Never throw a battery into open flames as this could lead to an explosion. Never subject batteries to hu­midity.
Please consider your environment when dis­posing of used batteries. They belong in a rub­bish dump for hazardous waste. Generally speaking you may return your batteries to the point of sale.
The appropriate regulations concerning return, recycling and disposal of used batteries must be respected at all times.
If the instrument is not used over an extended time period, we recommend to remove the bat­tery. In the event of instrument contamination caused by leaking battery cells, the instrument has to be returned for cleaning and verification to our factory.
11.0 Internal fuse
The internal fuse protects the instrument from overload or wrong operation/use. The fuses can trigger, if tensions > 600 V AC/DC are connected to the instrument.
The fuse is inside the instrument, the instrument must be opened for the exchange.
The fuses can only be replaced by returning the instrument to the factory or to an authorized sales
repair service.
Exclusively use fuse of voltage, current and bre­aking capacity values in compliance with the technical data section.
11.1 Indication in the display at defect fuse
If the instrument is correctly connected to the mains and before the measurement displays the symbols attention (1) and UB > UL (3), the inter­nal fuse has blown.
The fuse must be replaced and the measurement in­strument then checked by returning the instrument to the factory or to an authorized sales repair service.
12.0 Calibration Interval
To obtain the specified measurement result accura­cy the instrument should be regularly calibrated by our after sales service. We recommend a calibration interval of one year.
20
Battery Replacement / Calibration Interval
English
21
Technical Data
13.0 Technical Data
13.1 Tables wih test currents Table 1 Test Currents Function VC/RE, tRCD/VC, Auto (mA)
I∆N UB/RE- x0.5 x1 x2 x5
Measurement:
Prestest ~ 10 4 5 3,5 5 10 1410 202820 507050 30 12 15 10,5 15 30 42 30 60 84 60 150 212 150 100 40 50 35 50 100 141 100 200 283 200 500 707 500 300 120 150 105 150 300 424 300 600 849 600 1500 500 200 250 175 250 500 707 500 1000
1000 400 500 1000 2000
Table 2 Ramp Currents Function I<, tRCD/VC (mA)
IN 10 4-14 1,4-20 1-14 30 12-42 4,2 - 59 3-42 100 40-140 14-198 10-140 300 120-420 42-594 30-420 500 200-700 71-990 1000 400-1400
Table 3 RCD Trip Times for standard RCB´s
Fi/RCD-Type Current Type Current Type Test Current prescribed Test Time Standard Fi/RCD I∆NI∆NI∆N Trip Time DIN VDE Analyzer
x0.5 0.5x -
2000ms, I∆N≥100mA: 500ms IEC61008-1
x1 1x <300ms 500ms IEC61008-1 10mA/ x2 2x <150ms 150ms IEC61008-1 30mA/ x5 5x <40ms 40ms IEC61008-1 100mA/ Ramp 0.4x- 1.4x <300ms 300ms IEC61008-1
300mA/ x0.5 0.35x -
2000ms, I∆N≥100mA: 500ms IEC61008-1
500mA/ x1 1.4x <300ms 500ms IEC61008-1 1000mA x2 2.8x <150ms 150ms IEC61008-1
x5 7x <40ms 40ms IEC61008-1
Ramp 0.1x- 1.4x <300ms 300ms IEC61008-1
x0.5 0.5x -
2000ms, I∆N≥100mA: 500ms 0664- Part 100
x1 1x <300ms 500ms
0664- Part 100
x2 2x <150ms 150ms
0664- Part 100
x5 5x <40ms 40ms
0664- Part 100
Ramp 0.1x- 1.4x <300ms 300ms
0664- Part 100
22
Technical Data
Table 4 Trip times for selective RCB´s
Fi/RCD-Type Current Type Current Type Test Current prescribed Test Time Standard Fi/RCD I∆NI∆NI∆N Trip Time Analyzer
x0.5 0.5x - 500ms IEC61008-1
x1 1x
130...500ms 500ms IEC61008-1
100mA/ x2 2x
60...200ms 200ms IEC61008-1
300mA/ x5 5x
50...150ms 150ms IEC61008-1
500mA/ x0.5 0.35x - 500ms IEC61008-1
1000mA x1 1.4x
130...500ms 500ms IEC61008-1
x2 2.8x
60...200ms 200ms IEC61008-1
x5 7x
50...150ms 150ms IEC61008-1
x0.5 0.5x - 500ms 0664- Teil100
x1 1x
130...500ms 500ms 0664- Teil100
x2 2x
60...200ms 200ms 0664- Teil100
x5 5x
50...150ms 150ms 0664- Teil100
Table 5 Trip times for RCB´s with over curent protection
Fi/RCD-Type Current Type Current Type Test Current prescribed Test Time Standard Fi/RCD I∆NI∆NI∆N Trip Time Analyzer
x0.5 0.5x -
2000ms, I∆N≥100mA: 500ms IEC61009-1
x1 1x <300ms 500ms IEC61009-1 10mA/ x2 2x <150ms 150ms IEC61009-1 30mA/ x5 5x <40ms 40ms IEC61009-1 100mA/ Ramp 0.4x- 1.4x <300ms 200ms IEC61009-1
300mA/ x0.5 0.35x -
2000ms, I∆N≥100mA: 500ms IEC61009-1
500mA/ x1 1.4x <300ms 500ms IEC61009-1 1000mA x2 2.8x <150ms 150ms IEC61009-1
x5 7x <40ms 40ms IEC61009-1
Ramp 0.1x- 1.4x <300ms 200ms IEC61009-1
x0.5 0.5x -
2000ms, I∆N≥100mA: 500ms
x1 1x <300ms 500ms
x2 2x <150ms 150ms
x5 5x <40ms 40ms
Ramp 0.1x- 1.4x <300ms 200ms
English
23
Technical Data
Table 6 Trip times for selective RCB´s with over curent protection
Fi/RCD-Type Current Type Current Type Test Current prescribed Test Time Standard Fi/RCD I∆NI∆NI∆N Trip Time Analyzer
x0.5 0.5x - 500ms IEC61009-1 (RCBO) x1 1x 130...500ms 500ms IEC61009-1 (100mA/ x2 2x 60...200ms 200ms IEC61009-1 300mA/ x5 5x 50...150ms 150ms IEC61009-1 500mA/ x0.5 0.35x - 500ms IEC61009-1 1000mA) x1 1.4x 130...500ms 500ms IEC61009-1
x2 2.8x 60...200ms 200ms IEC61009-1
x5 7x 50...150ms 150ms IEC61009-1
x0.5 0.5x - 500ms
x1 1x 130...500ms 500ms
x2 2x 60...200ms 200ms
x5 5x 50...150ms 150ms
Table 7 Trip times for impulse resistent RCB´s
Fi/RCD-Type Current Type Current Type Test Current prescribed Test Time Standard Fi/RCD
I∆NI∆N Trip Time Analyzer
x0.5 0.5x -
2000ms, I∆N≥100mA: 500ms IEC61008-1
x1 1x 10...300ms 500ms IEC61008-1 10mA/ x2 2x 10...150ms 150ms IEC61008-1 30mA/ x5 5x 10...40ms 40ms IEC61008-1 100mA/ Ramp 0.4x- 1.4x 10...200ms 200ms IEC61008-1
300mA/ x0.5 0.35x -
2000ms, I∆N≥100mA: 500ms IEC61008-1
500mA/ x1 1.4x 10...300ms 500ms IEC61008-1 1000mA x2 2.8x 10...150ms 150ms IEC61008-1
x5 7x 10...40ms 40ms IEC61008-1
Ramp 0.1x- 1.4x 10...300ms 200ms IEC61008-1
x0.5 0.5x -
2000ms, I∆N≥100mA: 500ms
x1 1x 10...300ms 500ms
x2 2x 10...150ms 150ms
x5 5x 10...40ms 40ms
Ramp 0.1x- 1.4x 10...300ms 200ms
Table 8
Fi/RCD-Type Current Type Current Type Test Current prescribed Test Time Standard Fi/RCD I∆NI∆NI∆N Trip Time Analyzer
x0.5 0.5x I∆N- 2000ms VDE 661
set to 0.5x I
x1 1x I∆N <200ms 500ms VDE 661
x5 5x I∆N <40ms 40ms VDE 661 10mA/ x0.5 0.35x IN- 2000ms VDE 661
set to 0.5x I
30mA x1 1.4x I∆N <200ms 500ms VDE 661
x5 7x I∆N <40ms 40ms VDE 661
x0.5 0.5x IN- 2000ms
set to 0.5x I
x1 1x I∆N <200ms 2000ms
x5 5x I∆N <40ms 40ms
Table 9
Fi/RCD-Type Current Type Current Type Test Current prescribed Test Time Standard Fi/RCD
I∆NI∆N Trip Time Analyzer
fixed x0.5 0.5x IN- 2000ms VDE 662 set to 0.5x I
(SRCD) x1 1x I∆N <200ms 500ms VDE 662 - (10mA/ x5 5x I∆N <40ms 40ms VDE 662 ­30mA) x0.5 0.35x IN- 2000ms VDE 662
set to 0.5x I
x1 1.4x I∆N <200ms 500ms VDE 662 -
x5 7x I∆N <40ms 40ms VDE 662 -
x0.5 0.5x IN- 2000ms
set to 0.5x I
x1 1x I∆N <200ms 500ms -
x5 5x I∆N <40ms 40ms -
Standardprüfung
General Technical Data
Display ............................................3
1
/
2
digit LCD
Mains nominal voltage ....................230V/240V +10%-15%
Mains nominal frequency ...............50/60Hz +/- 10%
Reference-Range: ............................+17°C to 27°C, max. 70% rel. humidity
Temperature-Range:........................0°C to 40°C
Relative humidity: ............................max. 80%
Height above MSL:...........................up to 2000m
Battery type .....................................6 mignon 1.5V, Type AA, IEC LR6 (no storage batteries)
Current consumption ......................approx. 15mA
Battery life at average usage ...........approx. 2 years
Auto power off ................................after approx. 5 min.
Built in fuse......................................M 1,0A/250V, 5x20 mm
Data memory ..................................approx. 100 measurements
IR/RS232 interface .........................9600 Baud, 1 start bit, 8 data bits, even parity, 1 stop bit
Degree of contamination .................2
Protection .......................................2, protective insulation, IP50
Overvoltage class ............................CAT II / 300V
Dimensions (LxWxH) ......................235 x 105 x 70mm
Weight ............................................650g (batteries included)
24
Technical Data
English
25
Technical Data
Measuring Contact Voltage (UB)
Measurement conditions ................40% I∆N
Measurement time ..........................max 2000 ms,when exceeding limits, immediate cut-off
Measurement range ........................0,6 ... 70,0 V bei I∆N 100, 300, 500, 1000mA
........................................................1...70 V bei IN 10, 30mA
Resolution .....................................0,1 V at I∆N 100, 300, 500, 1000mA
Tolerance ........................................-0%...+10% /+6 Digit
Limits .............................................5V / 50V, contact voltage limit UL to be pre-selected
Measuring Ground Resistance (RE)
Measurement conditions ...............40% I∆N
Measurement time ..........................max. 2000 ms, when exceeding limits, immediate cut-off
Range: ...............................................1...1999 (at IN 100, 300, 500, 1000 mA)
...........................................................0,01...1,99 (at IN 10, 30 mA)
Resolution: ........................................1 (at IN 100, 300, 500, 1000 mA)
...........................................................0,01 k(at IN 10, 30 mA)
Tolerance: .........................................±10%/ ±10 Digit (at IN 100, 300, 500, 1000 mA)
...........................................................±10%/ ±20 Digit (at IN 10, 30 mA) depends of contact voltage
Measuring Trip Time tA:
Measurement range ........................I∆N 10, 30, 100, 300, 500, 1000mA standard
........................................................100, 300, 500, 1000 mA Selektiv
Test currents ..................................I∆N x0.5 / x1 / x2 / x5 (siee table 1)
Test current type AC, .......................intermittent DC residual current
........................................................DC
Phase position................................. 0° / 180°
Test current tolerance .....................0...-10%
Measurement time ..........................see table 3, when exceeding limits, immediate cut-off.
Measurement range ........................2...1999ms
Resolution ......................................1ms
Tolerance ........................................± (2%+2Digit)
Limits .............................................25V / 50V, contact voltage limit UL to be pre-selected
Measuring Trip Current (Ramp Funkction)
Measurement ranges ......................10, 30, 100, 300, 500, 1000mA Standard, see table 2
Test currents .................................40% x I∆N...140% x I∆N, bei AC
........................................................10% x I∆N...140% x I∆N, bei pulsierendem Gleichfehlerstrom bzw. DC
Phase position ................................0° / 180°
Measurement time ..........................see table 3, when exceeding limits, immediate cut-off
Display ............................................10% I∆N...140% I∆N
Resolution ......................................10% I∆N
Tolerance ........................................± 10% I∆N
Display ............................................0...1999ms
Resolution ......................................1ms
Tolerance ........................................± (2%+2Digit)
Display ............................................0...50 V
Resolution ......................................0,1 V bei IN 100, 300, 500, 1000mA
........................................................1V bei I∆N 10, 30mA
Tolerance ........................................-0%...+10% /+6 Digit
Limits .............................................25V / 50V, contact voltage limit UL to be pre-selected
Auto function
Measurement ranges .....................10, 30, 100, 300, 500, 1000mA idem
........................................................100, 300, 500, 1000 mA selective
Test currents ..................................0.5 x I∆N,
........................................................1 x I∆N,
........................................................2 x I∆N,
........................................................5 x I∆N (not at I∆N = 500 / 1000 mA standard or S ),
........................................................automatic selection, also refer to table 1
Phase position ................................0° / 180°, automatic selection
Measurement time ..........................see table 3, when exceeding limits, immediate cut-off
Display ............................................O.K. / bad ( POS / Err )
Limits .............................................25V / 50V, contact voltage limit UL to be pre-selected
Mains Voltage Range
Measurement range ........................3...300V AC
Tolerance ........................................+/-(3%+3 Digit)
Frequency range .............................50/60Hz +/- 10%
Internal resistance ...........................approx. 300k
Overvoltage protection ....................400V AC/DC
24 month Warranty
UNITEST instruments are subject to strict quality control. However, should the instrument function impro­perly during daily use, your are protected by our 24 months warranty (valid only with invoice).We will repair free of charge any defects in workmanship or material, provided the instrument is returned unope­ned and untampered with, i.e. with undamaged warranty label.Any damage due to dropping or incorrect handling are not covered by the warranty. If the instrument shows failure following expiration of warranty, our service department can offer you a quick and economical repair.
Subject to technical changes without notice!
26
Technical Data / 24 month Warranty
27
Die BEHA-Gruppe bestätigt hiermit, dass das erworbene Produkt gemäß den festgeleg­ten Beha-Prüfanweisungen während des Fertigungspro­zesses kalibriert wurde. Alle innerhalb der Beha-Gruppe durchgeführten, qualitätsrele­vanten Tätigkeiten und Prozes­se werden permanent durch ein Qualitätsmanagement­System nach ISO 9000 über­wacht.
Die BEHA-Gruppe bestätigt weiterhin, dass die während der Kalibrierung verwendeten Prüf
­einrichtungen und Instrumente einer permanenten Prüfmittel­überwachung unterliegen. Die Prüfmittel und Instrumente werden in festgelegten Abstän­den mit Normalen kalibriert, deren Kalibrierung auf nationa­le und internationale Standards rückführbar ist.
The BEHA Group confirms herein that the unit you have purchased has been calibrated, during the manufacturing process, in compliance with the test procedures defined by BEHA. All BEHA procedures and quality controls are moni­tored on a permanent basis in compliance with the ISO 9000 Quality Management Stan­dards.
In addition, the BEHA Group confirms that all test equip­ment and instruments used du­ring the calibration process are subject to constant control. All test equipment and instru­ments used are calibrated at determined intervals, using re­ference equipment which has also been calibrated in com­pliance with (and traceable to) the calibration standards of national and international labo­ratories.
Le groupe BEHA déclare que l´appareil auquel ce document fait référence a été calibré au cours de sa fabrication selon les procédures de contrôle dé­finies par BEHA. Toutes ces procédures et contrôles de qualité sont régis par le systé­me de gestion ISO 9000.
Le groupe BEHA déclare par ailleurs que les équipements de contrôle et les instruments uti­lisés au cours du processus de calibrage sont eux-mêmes soumis à un contrôle technique permanent.
Ces mêmes équipements de contrôle sont calibrés réguliè­rement à l´aide d´appareils de référence calibrés selon les di­rectives et normes en vigueur dans les laboratoires de re­cherche nationaux et interna­tionaux.
El grupo BEHA declara que el producto adquirido ha sido ca­librado durante la producción de acuerdo a las instrucciones de test BEHA. Todos los proce­sos y actividades llevados a ca­bo dentro del grupo BEHA en relación con la calidad del pro­ducto son supervisados per­manentemente por el sistema ISO 9000 de control de calidad.
Adicionalmente, el grupo BEHA constata que los equipos e ins­trumentos de prueba utilizados para la calibración también son sometidos a un permanente control. Estos equipos e instru­mentos de prueba son a su vez calibrados en intervalos regu­lares valiéndose de equipos de referencia calibrados de acuer­do a directivas de laboratorios nacionales e internacionales.
®
Connemara Electronics Beha GmbH
Industrial Park Carrigaline, Co. Cork Republic of Ireland Tel. (+353) 21 4919000 Fax (+353) 21 4910010 e-mail: connemara.electronics@connelec.ie
IQ NET
AENOR Spain AFAQ France AIB-Vincotte Inter Belgium APCER Portugal BSI United Kingdom CSIQ Italy
CQS Czech Republic DQS Germany DS Denmark ELOT Greece FCAV Brazil IRAM Argentina JQA Japan
KEMA Netherlands KSA-QA Korea MSZT Hungary NCS Norway NSAI Ireland ÖQS Austria PCBC Poland PSB Singapore
QAS Australia QMI Canada SFS Finland SII Israel JQA Japan SIQ Slovenia SIS-SAQ Sweden SQS Switzerland
IQNet is represented in the USA by the following IQNet members: AFAQ, AIB-Vinçotte Inter, BSI, DQS, KEMA, NSAI and QMI
Qualitätszertifikat • Certificate of Quality
Certificat de Qualité • Certificado de calidad
I.S./ISO 9002/EN 29002
Quality Management System
Quality Management System
ISO 9001
CH. BEHA GmbH
Elektrotechnik - Elektronik
In den Engematten 14
D-79286 Glottertal / Germany
Tel. +49 (0) 7684 / 80 09- 0
Fax +49 (0) 7684 / 8009 - 410
e-mail: info@beha.de
internet: http://www.beha.com
28
CH. BEHA GmbH
Elektronik - Elektrotechnik
In den Engematten 14 79286 Glottertal/ Germany Tel.: +49 (0) 76 84/80 09- 0 Fax: +49 (0) 76 84/80 09-410 Techn. Hotline: +49 (0) 76 84/80 09- 429 internet: http://www.beha.com e-mail: info@beha.de
®
Reg.No. 3335
11/2002 PTEB9061-00-RS
Loading...