UNITED PRODUCTS & INSTRUMENTS INC. UNICO 1100, UNICO 1100rs Service Manual

UNICO 1100 SERIES
SPECTROPHOTOMETER
Model 1100 Spectrophotometer
Model 1100RS Spectrophotometer
SERVICE MANUAL
UNITED PRODUCTS & INSTRUMENTS INC.
CONTENTS
1. GENERAL 3
1.1 Introduction 3
1.2 Optical System 4
1.3 Electronic 5
1.4 Specifications 6
1.5 Nomenclature 6-7
2. INSTALLATION 7
3. MAINTENANCE 8
3.1 Working Environment 8
3.2 Replacing Lamps And Adjustments 8
3.3 Wavelength Calibration 8-9
4. SERVICE 9
4.1 Operating Procedure Check 9-10
4.2 Physical 10
4.3 Troubleshooting 11-19
5. DRAWING AND DIAGRAM 20
5.1 PCB Arrangement of Synthetic Electronic System for UNICO 1100 21
5.1.1 Circuit Diagram of Synthetic Electronic System for UNICO 1100 22
5.2 PCB Arrangement of ±15v, +9v Power Supply
23
5.2.1 Circuit Diagram of ±15v, +9v Power Supply
24
5.3 PCB Arrangement of W Lamp Power Supply 25
5.3.1 Circuit Diagram of W Lamp Power Supply 26
5.4 PCB Arrangement of Logarithm Adjustment 27
5.4.1 Circuit Diagram of Logarithm Adjustment 28
5.5 PCB Arrangement of 0%T Adjustment 29
5.5.1 Circuit Diagram of 0%T Adjustment 30
5.6 PCB Arrangement of Amplifier 31
5.6.1 Circuit Diagram of Amplifier 32
6. SCHEMATIC LOYOUT
33
6.1 Schematic Diagram of UNICO 34
7. ELECTRONIC PARTS FOR UNICO 1100 35
7.1 Electronic Parts list for UNICO 1100 35
36-2
1. GENERAL
1.1 INTRODUCTION
The UNICO model 1100 Series spectrophotometers are single beam general purpose instruments designed for use in conventional laboratories. The instruments are ideal for various applications such as: Clinical Chemistries, Biochemistry, Food and Beverage Laboratories, Environmental Protection, Water and Waste Water Labs and other fields of quality control.
The UNICO 1100 Series spectrophotometers cover the visible wavelength range from 335nm to 1000nm. The analytical grating system has 1200 lines/mm for higher dispersion of the spectrum.
36-3
Figure 1-1.1
UNICO Model 1100 Spectrophotometer
Figure 1-1.2
UNICO Model 1100RS Spectrophotometer
36-4
1.2 OPTICAL SYSTEM
As shown in Figure 1-2 the white light coming from the Halogen Lamp is focused by the collecting lens, pass through the Entrance slit of the Monochromator, through a collimating lens to the Analytical grating where it is dispersed into its spectrum, then through the Exit slit of the monochromator, continues through the Sample compartment and finally to the Photo detector.
Figure 1-2
OPTICAL SYSTEM SCHEMATIC DIAGRAM
1. Halogen Lamp 2. Collecting Lens
3. Entrance Slit 4. Collimating Lens
5. Grating 6. Exit Slit
7. Sample 8. Collimating Lens
9. Photodiode
36-5
1.3 ELECTRONIC SYSTEM
The photo detector receives the monochromatic light and generates photocurrent. A direct current signal which is proportional to the amount of light energy to the photo detector. The signal is then amplified and processed. The processed signal is sent to the digital display and also to the external outputs 0 Volt to 1 Volt analog output and an RS-232C output. Figure 1-3-1 and Figure 1-3-2 demonstrates the electronic systems of Model 1100 and Model 1100RS respectively
36-6
36-7
36-8
1.4 SPECIFICATIONS
Table 1-1
MODEL 1100 1100RS
Wavelength Range 335-1000nm 335-1000nm
Spectral Slit Width 20nm 10nm
Wavelength Accuracy
±2nm
±2nm
Wavelength Readability 2nm
Stray Radiant Energy
0.5%T@340nm &400nm
0.5%T@340nm & 400nm
Transmittance 0%T to 125%T 0%T to 125.0%T
Absorbancy 0A to 2A 0A to 2A
Photometric Accuracy
±2.0%T ±1.0%T
Photometric Noise Level
±2.0%T
±1%T
Power Requirement 115-volt, 60Hz, or 230-volt, 50Hz
Size
408mm(W)×308mm(D)×180mm(H)
Weight 5.5kg
Specifications subject to change without notice
1.5 NOMENCLATURE
Figure 1- 4
36-9
Figure 1- 5
UNICO 1100 SPECTROPHOTOMETER AND NOMENCLATURE
2. INSTALLATION
2.1 The instrument should be installed in a clean environment which is free of
corrosive materials, dust and severe vibrations. Furthermore, it should be kept away from any strong electric magnetic fields and high frequency fields.
2.2 LINE VOLTAGE
The power can be selected by means of the built-in switchable power switch to match the line voltage in your country.
36-10
3. MAINTENANCE
3.1 WORKING ENVIRONMENT
The optimum environment for the instruments is as follows
TEMPERATURE 5-35° C HUMIDITY 85%
Install the instrument in a clean non-vibrational area. Avoid direct strong light and air flow. The instrument will give better performance if the power supply stability is at 1% and line frequency less than 1Hz. Always allow the instrument to warm up for 20 minutes prior to taking any readings.
3.2 REPLACING LAMPS AND ADJUSTMENTS
1. Turn off and unplug the instrument.
2. Turn the instrument upside down.
3. Remove the grill plate on bottom of the instrument by removing fixing Screw.
4. Unplug the lamp from the white connector.
5. Insert the new lamp, pushing it in as far as it will go. The lamp filament should face the center of the collecting lens.
CAUTION: DO NOT HANDLE HALOGEN LAMP WITH BARK FINGERS. USE TISSUE OR CLOTH
WHEN HANDLING LAMP.
8. Reinstall the grill plate.
Figure 3-1
Replacing the lamp
3. 3 Wavel e n g t h Calib r a t i o n :
36-11
Normally the UNICO 1100 Series spectrophotometer retains its wavelength calibration indefinitely. However if the instrument receives a severe shock or is abused, use the following methods to check wavelength calibration. Please note that this test requires the UNICO Didymium filter, p/n 1100-110, or the Holmium Oxide filter, p/n 1100-109.
In the filter method, the didymium filter has two distinct absorbance peaks at 529nm and 807nm. The Holmium filter has a distinct peak at 361nm. When the instrument is calibrated properly you will find minimum Transmittance (maximum Absorbance) at the range +2nm from these peaks. Note that the specific Transmittance values are not important as you are only looking for the wavelength where the minimu m transmittance (maximum Absorbance) occurs.
Holmium Oxide Filter Method:
1. Turn instrument on and allow it to warm up for 15 minutes.
2. Select the Absorbance operating mode.
1. Set the wavelength to 350nm.
2. Make sure the cuvette holder is empty and place it in the sample compartment. Close
the sample compartment lid.
3. Set zero Absorbance by pressing the 0A/100%T. Wait a few seconds while the display
flashes ‘BLA’. The reading should then be 0.000A. If not, repeat step 5.
4. Remove the cuvette holder and insert the Holmium filter into it. Place it in the sample
compartment and close the lid.
5. Record the Absorbance reading on the digital display.
6. Advance the wavelength setting by 1nm and repeat steps 4 to 7.
7. Repeat step 8 until the wavelength setting reaches 370nm.
8. Look for the maximum absorbance reading obtained, and this should be found between
359 and 363nm. The wavelength accuracy of the 1100/1100RS is + 2nm.
Didymium Filter Method:
1. Set the Wavelength to 800 nm.
2. Make sure the cuvette holder is empty and place it in the sample compartment. Close
the sample compartment lid.
3. Set zero Abs by pressing the 0A/100%T. Wait a few seconds while the display flashes
‘BLA’. The reading should then be 0.000A. If not, repeat step 3.
36-12
4. Remove the cuvette holder and insert the Didymium filter into it. Place it in the sample
compartment and close the lid.
5. Record the Absorbance reading on the digital display.
6. Advance the wavelength setting by 1nm and repeat steps 2 to 5.
7. Repeat step 6 until the wavelength setting reaches 815nm.
8. Look for the maximum absorbance reading obtained, and this should be found between
805 and 809nm. The wavelength accuracy of the 1100/1100RS is + 2nm.
9. If a “middle” wavelength check is desired, set the wavelength to 522nm (optional)
10. Make sure the cuvette holder is empty and place it in the sample compartment. Close
the sample lid.
11. Set zero Abs by pressing the 0A/100%T button. Wait a few seconds while the display
flashes ‘BLA’. The reading should then be 0.000A. If not repeat step 11.
12. Remove the cuvette holder and insert the Didymium filter into it. Place it in the sample
compartment and close the lid.
13. Record the absorbance reading on the digital display.
14. Advance the wavelength setting by 1nm and repeat steps 10 to 13.
Repeat step 14 until the wavelength setting reaches 536nm. Again, look for the maximu m absorbance reading. It should be between 527 and 531nm.
Absorbance Accuracy Checks
Specification: +2% at 1A (1100), +1% at 1A and 2A (1100RS).
The absorbance accuracy should be checked against a set of neutral density filters accurately calibrated to the NIST standards. Contact your UNICO representative for more information (800-588-9776).
An alternative method using potassium dichromate is described below. Due to the many factors that might affect the results (i.e. temperature, bandpass, weighing and diluting errors), this method is less accurate and should only be used as a guide.
Reference: Johnson E A Potassium Dichromate as an absorbance standard PSG Bulletin 1967, No. 17, page 505
36-13
1. Make up N/100 sulfuric acid as the solvent and use part of it to make a solution
containing 120 +0.5mg/litre of potassium dichromate.
2. Wash out a square cuvette with solvent, and fill with solvent.
3. Put the cuvette in the adapter into the sample compartment and close the lid.
4. Set the wavelength to 350nm.
5. Set the MODE button to A.
6. Set the reading to 0.000A using the 0A/100%T button.
7. Empty the cell. Wash out with dichromate solution, and fill with dichromate solution.
8. Put the cuvette in the adapter into the sample compartment and close the lid.
9. Read the absorbance of the standard from the display. The value should be 1.288 +
0.04A. Refer to the notes above when interpreting the result.
36-14
Stray Light Check
Specification: Less than 0.5%T at 340nm by AST M E 387
A good indication as to whether the stray light level is within specification may be obtained as follows:
1. Set the wavelength to 340nm.
2. Set the MODE switch to %T.
3. With the square cell adapter in the sample compartment, but no cell, close the lid
and press the 0A/100%T button to set the display to 100.0%.
4. Remove the cell adapter from the sample compartment and close the lid. Make a
note of the reading that should be at or near 00.0.
5. Prepare a solution containing 50gm/L of sodium nitrite (NaNO2) in distilled water and
fill a square cuvette with this solution.
6. Insert the cuvette into the adapter and place in the sample compartment. Close the
lid. The display should read<0.5%T. Note that if the reading obtained in step 4 is greater than 00.0, this value should be subtracted from the displayed reading to give the correct reading for the stray light value.
36-15
SERIVCE
4.1 PHYSICAL INSPECTION
Thoroughly inspect the instrument for physical indications that might cause a malfunction. The following categories are usually checked:
Apparent damage, Dirty Optics, Broken wire, loose connection, P.C. Board loose, Broken lands on pc board, any sign of overheated components.
4.2 OPERATING PROCEDURE CHECK
Each time an instrument has a malfunction, the operating procedure should be checked. If necessary such procedures should be repeated to see the patterns that the malfunction occurs.
The basic operating procedure is as follows:
Basi c Opera t i o n :
Simple operating instructions are printed on the front panel of your UNICO 1100/1100RS.
A. Pre pa r e t he sp e ct ro p h o t o m e t e r
1. Turn on the spectrophotometer by pressing the Power Switch (IO). Allow 15 minutes
for the instrument to warm up.
2. Select either the %Transmittance or Absorbance operating mode by pressing the
%T/A selector button (MODE) until the red light for T or A is on.
3. Select the desired wavelength by turning the wavelength control knob
(WAVELENG TH).
B. Pr e p a r e Sa m p le
4. Make a blank reference solution by filling a clean cuvette (or test tube) half full with
distilled or de-ionized water or other specified solvent. Wipe the cuvette with tissue to remove the fingerprints and droplets of liquid.
5. Fit the blank cuvette into the square cuvette adapter and place the adapter in the
sample compartment, aligning the guide mark (if present) with the guide mark at the front of the compartment. Close the lid.
6. Set 0.000A or 100%T with the (0A/100%T) control button. NOTE: This step fulfills
the instruction on the front of the spectrophotometer…(Set Full Scale).
7. Remove the blank cuvette or test tube. Set it aside in the case that you may need to
adjust the (0A/100%T) control button later (i.e. change the wavelength).
36-16
C. A n a l y z e Sa mp l e
8. Rinse a second cuvette with a small amount of the sam ple solution to be tested. Fill
the cuvette half full and wipe it.
9. Put the sample cuvette in the sample compartment. Close the lid.
10. Read the %Transmittance or Absorbance from the digital readout window. Remove
the sample cuvette or test tube.
11. If you are to test the same sample at other wavelengths, repeat steps 3 to 10 for
each wavelength.
12. For each new sample you analyze, repeat steps 2 to 11.
A d d i t i o n a l Feat u r e s of Mo d e l 1 1 0 0 R S
Although operating basically the same as the Model 1100, the 1100RS spectrophotometer offers two more features (Concentration Mode and Factor Mode) as follows:
Concentration (C) Mode for determining the concentration of unknown samples.
NOTE: This method should only be used when the relationship between Absorbance and Concentration is known to be linear. The concentration of the Standard solution used to calibrate the instrument should be higher than the most concentrated sample.
1. Select the desired wavelength by turning the wavelength control knob.
2. Using the MODE button, select Absorbance (A) mode.
3. Insert the cuvette containing the blank solution.
4. Set 0.000A with the (0A/100%T) control button.
5. Using the MODE button, select C (concentration) mode.
6. Insert a cuvette containing a standard solution of known concentration in the sample
compartment and adjust the digital readout window to the value of the standard by using the and arrows.
7. Press the Enter button.
NOTE: If the reading changes, the factor required is too high (i.e. >1999) to be displayed. In this case, divide the concentration by 10, re- select the C mode by successive presses on the Mode button, cycling through the F, %T, and A modes, and follow step 2 above to set the concentration of the standard to the reduced value.
36-17
8. With the standard concentration set, determine the concentration values of samples with
unknown concentration by inserting the sample cuvette into the sample compartment and reading the value direct from the display.
9. To read the value of the multiplier used to convert Abs to Concentration, after measuring
all the samples, change the Mode to Factor (F) and read the multiplier from the display. Keep a record of this value for future use.
Operational Note: if the mode switch is changed to read Factor or Absorbance, the Concentration (C) reading is “frozen”, and cannot be changed. This requires the operator to re- start at step 1.
Factor ( F) Mo d e
This is a special mode for measuring concentration values of unknown samples using a previously determined factor to convert Absorbance readings to concentration.
1. After setting the wavelength, and setting zero Abs on the blank solution, using the
MODE button, select factor (F) mode.
2. Insert a cuvette containing a sam ple.
3. Using the and arrows, set the digital read-out window to the desired value of the
multiplier.
4. Press the ENT button. The spectrophotometer switches to the Concentration (C)
MODE.
Operational Note: If the Concentration of the sample is too high to be displayed, the instrument will not switch to Concentration Mode when the ENT button is pressed. Dilute the sample and multiply the concentration reading by the dilution factor to obtain the original sample concentration.
5. Read the concentration value of the sample direct from the display.
6. Insert a cuvette containing the next sample and read the result. Repeat until all
samples have been measured.
Operational Note: If the MODE switch is changed to A or T, then the concentration reading is “frozen”. This requires the operator to re- start at step 1.
36-18
Mode l 110 0 Spec t r o p h o t o m e t e r :
Mode Indicator: Allows the operator to know the measurement mode currently in use (% Transmittance, Absorbance) .
MODE Button: Switches between %Transmittance and Absorbance operating modes at any time of measurement.
0A/100%T Button: Adjusts digital readout to 100%T or 0.000A when blank reference solution is in sample compartment.
Sam ple Compartment: Insert 10mm test tube and close lid or fit 10mm cuvette adapter and fit adapter (with holder to left of cuvette) into compartment.
Wavelength Control Knob (WAVELENGTH): Selects desired wavelength in nanometers (nm). Wavelength Readout Window: Displays desired wavelength.
36-19
Mode l 110 0 R S Spe ct r o p h o t o m e t e r :
Mode Indicator: Allows the operator to know the measurement mode currently in use (% Transmittance, Absorbance).
MODE Button: Switches between %Transmittance and Absorbance operating modes at any time of measurement.
0A/100%T Button: Adjusts digital readout to 100%T or 0.000A when blank reference solution is in sample compartment.
0%T Button: Wh en in transmittance (T) mode and sample compartment is empty, a flap blocks the beam. Pressing button adjusts readout to 00.0% T.
PRINT Button (ENT): First press after selecting Concentration mode enters reading on digital readout. Other presses cause output of reading to printer. If using UN IC O Application Software, press this button to communicate with software. If operating in Factor (F) Mode, then press to enter factor. This causes the Mode to enter the factor number and change to Concentration mode.
Sam ple Compartment: Insert 10mm test tube and close lid or fit 10mm cuvette adapter and fit adapter (with holder to left of cuvette) into compartment.
36-20
Wavelength Control Knob (WAVELENGTH): Selects desired wavelength in nanometers (nm). Wavelength Readout Window: Displays desired wavelength.
.
36-21
4.3 TROUBLE SHOOTING #1
36-22
Pilot light
Display on?
Check if LCD display
and wiring OK?
Is 115V or 230v
Power on?
Switch function?
PC power Function?
Check the +8V and
–8V On the power
supply board?
Refer to Fig.5.1.1 to check +8V and –8V
power supply
Problem:Turn on, No display
A:Replace CPU board B:Sloving wiring problem or replace socket A2,A3
Line power on
Replace R13-69A switch
Replace PC power
Is diode working fine?
How about wiring?
A:Replace φ3 green diode B:Sloving wiring problem
Y N
Y
Y
Y
Y
N
N
N
N
N
N
N
TROUBLE SHOOTING #2
36-23
Problem:Instrument is on, Display 0.000(%T)or 1---(A)
Halogen lamp on?
Anything in sample comparment blocking the light?
Check lamp OK?
Replace lamp
Check output of lamp Power supply, is it Between +1.6V to 6.2V
Refer Fig.5.3 to check W lamp power supply
Remove it
Is cell holder and Cuvette in position?
Install cell holder cuvette
Shutter operation?
Repair shutter
Halogen lamp peaked?
Amplifier detector Working fine?
Adjust the lamp position(keep shutter open,put a white paper into optical path in sample comparment) at 525nm,you should see bright green light on paper
A:check S1227K photocell B:refer Fig.5.3.1 to check amplifier C:solving wiring problem
N
Y
N
Y
YY
N
Y
N
N
Y
Y
N
Y
TROUBLE SHOOTING #3
36-24
Problem:Turn on,put in cell holder and cuvette ,display “<<10.0”, press “100%T”,display not change
Is the position of halogen lamp
Is the light road out of the position
Adjust the position of halogen lamp,put a piece of white retangle paper into sample comparment(open the shutter),can see a whole,clear,bright facula
Right filter in position?
Adjust the filter
Adjust light road with right way, put apiece of white retangle paper(open the shutter),can see a whole,clear,bright facula
Y
N
N
Y
N
TROUBLE SHOOTING #4
36-25
Problem: Instrument on Display “1— — —”(T%) or show negative at absorbance
Is sample compartment cover closed?
Close the sample compartment
Is cell holder and cuvette in positior
Put the cell holder and cuvette into the sample compatrment
Shutter working
Repair shutter assembly
Press “100%”,after display “100.0” open the sample compartment cover if the display changde?
Is regulate power supply normal
1. Solving wiring problem of J5 socket.
2. replace
IC4(74LS07)、IC5(40 51BE) on the power supply board.
3.
replace power supply
Anplifier working
1. Solving wiring problem.
2. Refer Fig.5.6 to check
amplifier.
N
N
Y
Y
Y
N
N
N
Y
N
36-26
TROUBLE SHOOTING #5
Problem: Instrument on → “100.0” display not stable (T)
Enough warm up time?
Warm up at least 20 min
Groounding OK?
Improuve grounding
Power supply OK
Get rid of possible interference facters
After changed wavelength stable time full
After 5 min press “100%” again
±12V power supply work stable
Replace PC power supply
Regulate power supply working
Refer to Fig5.3.1 to check regulate power supply of halogen lamp
Electric eye working
Replace the same type electric eye (S1227K)
Amplifier working
Check amplifier
N
N
N
N
N
N
N
N
Y
Y
Y
Y
Y
Y
Y
TROUBLE SHOOTING #6
36-27
Problem:measuring recelts incorrect(not stable,repeatability,etc)
Is sample preparation adjust right?
a use right procedures to b prepare samples c make sure no bubble in sample d test only after sample stability e select right wavelength f get rid of possible interference factors
Is sample preparation Installed correctly?
Install the compartment correctly
Are cuvettes used matched?
Is wavelength accurate?
Are %T and 100%T correct?
Is photometric accuracy
Calibrate wavelength
Use matched cuvetter
Diference should be
Within 0.5%T
Use right
procedue to set
N
Y
N
N
N
N
N
Y
Y
Y
Y
36-28
5.Drawing&diagram
5.1 PCB Arrangement of Power Supply
5.1.1 Circuit Diagram of Power Supply
5.2 PCB Arrangement of CPU
5.2.1 Circuit Diagram of CPU
5.3 PCB Arrangement of Amplifier
5.3.1 Circuit Diagram of Amplifier
1. Check if the monochromator light passed onto the detector at right angle and right position (area) of the detector if not, adjust it.
2. Amplifier output voltage (static): 20mV-30mV, put black body to prevent the light from light way, refer to Fig5.3 to adjust the potentiometer W to change the output voltage to 20mV-30mV(measure the voltage between pin3 and pin4).
3. Replace detector SI1227K.
4. Use NBS930D filter to check.
5. If the value of display too big, check the control single of halogen lamp
solving wiring problem of socket J5replace IC4 IC5 on the power supply board replace power supply board
36-29
36-30
36-31
36-32
36-33
36-34
6 Schematic Layout
PART LIST OF UNICO 1100 SERIES SPECTROPHOTOMETERS
Number Description
1
On/Off Switch
2
Pc-power
3
Power socket
4
Wavelength disc
5
Wavelength indicator
6
Part of grating
7
Monochromator
8
Sample compartment
9
Filter drive strap
10
Filter commutation cam
11
Amplifier dark box
12
Amplifier
13
Halogen lamp
14
Junction-box
15
Transistor
16
Radiator
17
Power supply board
36-35
36-36
Loading...