C. A n a l y z e Sa mp l e
8. Rinse a second cuvette with a small amount of the sam ple solution to be tested. Fill
the cuvette half full and wipe it.
9. Put the sample cuvette in the sample compartment. Close the lid.
10. Read the %Transmittance or Absorbance from the digital readout window. Remove
the sample cuvette or test tube.
11. If you are to test the same sample at other wavelengths, repeat steps 3 to 10 for
each wavelength.
12. For each new sample you analyze, repeat steps 2 to 11.
A d d i t i o n a l Feat u r e s of Mo d e l 1 1 0 0 R S
Although operating basically the same as the Model 1100, the 1100RS spectrophotometer
offers two more features (Concentration Mode and Factor Mode) as follows:
Concentration (C) Mode for determining the concentration of
unknown samples.
NOTE: This method should only be used when the relationship between Absorbance and
Concentration is known to be linear. The concentration of the Standard solution used to
calibrate the instrument should be higher than the most concentrated sample.
1. Select the desired wavelength by turning the wavelength control knob.
2. Using the MODE button, select Absorbance (A) mode.
3. Insert the cuvette containing the blank solution.
4. Set 0.000A with the (0A/100%T) control button.
5. Using the MODE button, select C (concentration) mode.
6. Insert a cuvette containing a standard solution of known concentration in the sample
compartment and adjust the digital readout window to the value of the standard by using
the ∧ and ∨ arrows.
7. Press the Enter button.
NOTE: If the reading changes, the factor required is too high (i.e. >1999) to be displayed. In
this case, divide the concentration by 10, re- select the C mode by successive presses on the
Mode button, cycling through the F, %T, and A modes, and follow step 2 above to set the
concentration of the standard to the reduced value.
36-17