Overview
Unpacking Inspection
Safety Information
Rules For Safe Operation
International Electrical Symbols
The Meter Structure
A.The Meter Front Structure
B.The Meter Back and Bottom Structure
Functional Buttons
Display Symbols
Measurement Operation
A. AC Voltage + Frequency Measurement
B. AC Current + AC Voltage Measurement
C. Active Power + Phase Angle Measurement
D. Apparent Power + Reactive Power Measurement
E. Power Factor + Phase Angle Measurement
F. Active Energy + Time Measurement
True RMS Measurement and Average Value Measurement
Specifications
3
3
4
4
6
7
7
8
9
12
14
14
16
18
25
27
29
31
32
1
Model UT232: OPERATING MANUAL
TABLE OF CONTENTS
TITLE PAGE
A. General Specifications
B. Environmental Requirements
Accuracy Specifications
A. AC Voltage
B. Frequency
C. AC Current
D. Active Power
E. Apparent Power
F. Reactive Power
G. Power Factor
H. Phase Angle
I. Active Energy
Maintenance
A. General Service
B. Replacing the Battery
32
33
34
34
34
35
35
36
37
38
38
39
39
39
40
2
Model UT232: OPERATING MANUAL
OverviewUnpacking Inspection
This Operating Manual covers information on safety
and cautions. Please read the relevant information
carefully and observe all the Warnings and Notes
strictly.
Warning
To avoid electric shock or personal injury, read the
ìSafety Informationî and ìRules for Safe Operationî
carefully before using the Meter.
Model UT232 is a three phase intelligent handheld
digital power clamp meter (hereafter referred to as ìthe
Meterî) which has both the features of digital current
meter and also power measurement meter.
The Meter can measure Voltage, Current, Active Power,
Apparent Power, Reactive Power , Power Factor, Phase
Angle, Frequency, Active Energy and etc.
Open the package case and take out the Meter. Check
the following items carefully to see any missing or
damaged part:
Item
1
2
3
4
5
6
7
8
9
10
11
12
13
Description
English Operating Manual
Red Test Lead
Black Test Lead
Blue Test Lead
Yellow Test Lead
Red Alligator Clip
Black Alligator Clip
Blue Alligator Clip
Yellow Alligator Clip
USB Interface Cable
Software
Tool Box
In the event you find any missing or damage, please
contact your dealer immediately.
3
Model UT232: OPERATING MANUAL
Safety Information
This Meter complies with the standards IEC61010: in
pollution degree 2, overvoltage category (CAT. III 600V,
CAT IV 300V) and double insulation.
CAT. III: Distribution level, fixed installation, with smaller
transient overvoltages than CAT. IV
CAT.IV: Primary supply level, overhead lines, cable
systems etc.
Use the Meter only as specified in this operating manual,
otherwise the protection provided by the Meter may be
impaired.
In this manual, a Warning identifies conditions and
actions that pose hazards to the user, or may damage
the Meter or the equipment under test.
A Note identifies the information that user should pay
attention to.
International electrical symbols used on the Meter and
4
in this Operating Manual are explained on page 6.
Rules For Safe Operation
Warning
To avoid possible electric shock or personal injury,
and to avoid possible damage to the Meter or to the
equipment under test, adhere to the following rules:
l
Before using the Meter inspect the case. Do
not use the Meter if it is damaged or the case
(or part of the case) is removed. Look for cracks
or missing plastic. Pay attention to the insulation
around the connectors.
l
Inspect the test leads for damaged insulation
or exposed metal. Replace damaged test leads
with identical model number or electrical
specifications before using the Meter.
l
Do not apply more that the rated voltage, as
marked on the Meter.
l
When measurement has been completed,
disconnect the connection between the test
leads and the circuit under test, remove the
Model UT232: OPERATING MANUAL
testing leads away from the input terminals of
the Meter and turn the Meter power off.
l
Do not carry out the measurement when the
Meterís back case and / or battery door is opened
to avoid electric shock.
l
When the Meter working at an effective voltage
over 30V in AC, special care should be taken.
l
Use the proper terminals and function for your
measurements.
l
Do not use or store the Meter in an environment
of high temperature, humidity, explosive,
inflammable and strong magnetic field. The
performance of the Meter may deteriorate after
dampened.
l
Do not use the Meter if the surface of it is wet
or the userís hands are wet.
l
When using the test leads, keep your fingers
behind the finger guards.
l
Replace the battery as soon as the battery
indicator appears. With a low battery, the
Meter might produce false readings that can
lead to electric shock and personal injury.
l
When opening the battery door, must make sure
the Meter is power off.
l
When servicing the Meter, use only the same
model number or identical electrical
specifications replacement parts.
l
The internal circuit of the Meter shall not be
altered at will to avoid damage of the Meter and
any accident.
l
Soft cloth and mild detergent should be used
to clean the surface of the Meter when servicing.
No abrasive and solvent should be used to
prevent the surface of the Meter from corrosion,
damage and accident.
l
The Meter is suitable for indoor use.
l
Turn the Meter off when it is not in use and take
out the battery when not using for a long time.
l
Constantly check the battery as it may leak
when it has been using for some time, replace
the battery as soon as leaking appears.
A leaking battery will damage the Meter.
5
International Electrical Symbols
AC (Alternating Current)
Grounding
Double Insulated
Warning. Refer to the Operating Manual
Deficiency of Built-In Battery
Danger of High Voltage
Conforms to Standards of European Union
Model UT232: OPERATING MANUAL
6
Model UT232: OPERATING MANUAL
The Meter Structure
A. The Meter Front Structure (see figure 1)
BlackRedBlueYellow
Figure 1
1
Transformer Jaw: designed to pick up the AC
and DC current flowing through the conductor.
It could transfer current to voltage. The tested
conductor must vertically go through the Jaw
center.
2
Hand Guards: to protect userís hand from
touching the dangerous area.
3
Power button
4
Max/ button
5
MENU button
6
LOAD button (recall data)
7
SELECT button (press to select alternate
function and sum of Watts measurement)
8
MIN/ button
9
V3 Input Terminal (Third phase measurement)
10
V2 Input Terminal
(Second phase measurement)
11
V1 Input Terminal (First phase measurement)
12
COM Input Terminal
13
CLEAR button (maximum and minimum)
7
Model UT232: OPERATING MANUAL
14
15
16
17
18
19
20
button (Sum)
SAVE button (data store button)
USB button
LIGHT button (auto display backlight button)
HOLD button
LCD Display
Testing Leads (Red, Black, Blue and Yellow)
8
B. The Meter Back and Bottom Structure
(see figure 2)
Figure 2
1 Infrared slot
2 USB Interface Cable
Model UT232: OPERATING MANUAL
Functional Buttons
Below table indicated for information about the functional
button operations.
Button
POWER
HOLD
LIGHT
MENU
Operation Performed
Press and hold POWER for 1 second to
turn the Meter on.
Press POWER again to turn the Meter off.
Press HOLD to enter the Hold mode in
any mode,
beeps.
Press HOLD again to exit the Hold
mode to return to measurement mode,
the Meter beeps and
Press LIGHT to turn the display backlight
on.
The display backlight will be off
automatically after 30 seconds.
Press MENU to display the following
functions in sequence:
AC Voltage (main display) + Frequency
appears and
disappears.
the Meter
Button
MENU
MAX/
Operation Performed
(secondary display)
AC Current (main display) + AC V oltage
(secondary display)
Active power (main display) + Phase
angle (secondary display)
Apparent power (main display) +
Reactive power (secondary display)
Power factor (main display) + Phase
angle (secondary display)
Active Energy (main display) + Time
(secondary display)
Press to start recording of maximum
value, it valid at voltage, current, active
power and apparent power ranges only.
Press once at LOAD mode, MR icon
shown, the Meter displays the next
stored reading, the left secondary
display showing the index increase one.
Press once at mode, CAL icon shown,
9
Model UT232: OPERATING MANUAL
Button
MAX/
USB
LOAD
10
Operation Performed
the Meter steps through sum of active
power (main display) + sum of reactive
power (secondary display) and sum of
power factor (main display) + sum of
apparent power.
Press USB once to turn the USB
interface on, USB appears and the
Meter beeps.
Press USB again to turn the USB
interface off, USB disappears and the
Meter beeps.
It is invalid at active energy mode
Press once to enter LOAD mode, MR
appears and the Meter beeps.
Press again to exit LOAD mode, MR
disappears and the Meter beeps.
Press and hold LOAD for 1 second to
display the stored data quickly.
At Active power (main display) + Phase
angle (secondary display) mode, press
once button to sum up the current
Button
SAVE
Operation Performed
first phase of 3 phase measurement
result. Then carry out second phase
power measurement.
Press
phase power measurement result, then
carry out third phase power
measurement.
Press
power measurement result, then press
SELECT to enter the measurement of
sum of the watts mode, the display
shows the 3 phase sum of the watts
value.
Press once to store single reading,
MEM appears and the Meter beeps.
Press and hold for over 1 second to
continuous store reading, MEM blinks
and the Meter beeps. The index number
shown on the left secondary display
keep on increasing. Press SAVE again
again to sum up the second
again to sum up the third phase
Model UT232: OPERATING MANUAL
Button
SELECT
MIN /
Operation Performed
to exit. The maximum number of data store
is 99, when it achieves 99, the Meter shows
FUL. Press CLEAR to clear the stored
reading in order to store next reading.
At active power (main display) + phase
angle (secondary display) mode, press
SELECT button to step through first phase,
second phase, third phase and sum of
watts.
At other mode, press SELECT to step
through first phase, second phase and the
third phase.
Press to start recording of minimum value,
it valid at voltage, current, active power
and apparent power ranges only.
Press once at LOAD mode, MR icon
shown, the Meter displays the previous
stored reading, the left secondary display
showing the index decrease one.
Press once at mode, CAL icon shown,
Button
CLEAR
Operation Performed
the Meter steps through sum of active
power (main display) + sum of reactive
power (secondary display) and sum of
power factor (main display) + sum of
apparent power.
At active energy range, press to reset
time the zero, then restart the timing.
At all other ranges, press to clear stored
readings.
11
Model UT232: OPERATING MANUAL
Display Symbols (see figure 3)
Figure 3
12
Number
1
2
3
4
5
6
7
8
9
10
11
Symbol
USB
DC
1
2
3
h
mm
W
s
Hz,
PG,
KVAr
Meaning
Data Output is in progress
Indicator for DC measurement
First phase symbol
Second phase symbol
Third phase symbol
Unit for hour
Unit for minute
Watt: Sum of Watt
The battery is low.
Warning: To avoid false
readings, which could lead to
possible electric shock or
personal injury, replace the
battery as soon as the battery
indicator appears.
Unit for second
Symbol of Unit.
Hz: Hertz.The unit of frequency.
PG: The unit of phase angle
KVAr. The unit of reactive power