UNI-T UT33B, UT33C, UT33D Operating Manual

Page 1
Page
3 4 5 7
10
11 12 13 13 15 17 20
22 25 27
Title
Overview Unpacking Inspection Safety Information Rules For Safe Operation International Electrical Symbols The Meter structure Functional Buttons Measurement Operation
A. DC Voltage Measurement B. AC Voltage Measurement C. DC Current Measurement D. Measuring Resistance E. Diodes and Continuity
Measurement
F. Temperature Measurement
G. Battery Test
Table of Contents (1)
1
Model UT33B/C/D: OPERATING MANUAL
Page 2
Table of Contents (2)
Title
H. Square Wave Output General Specifications Accuracy Specifications
A. DC Voltage B. AC Current C. DC Current D. Resistance E. Diodes and Continuity
Measurement
F. Temperature G. Battery Test H. Square Wave Output Maintenance A. General Service B. Replacing the Battery C. Replacing the Fuses
Page
28 29 31 31 32 33 34
35
35 36 36 37 37 38 39
2
Model UT33B/C/D: OPERATING MANUAL
Page 3
Overview
This Operating Manual covers information on safety and cautions. Please read the relevant information carefully and observe all the Warnings and Notes strictly.
Warning To avoid electric shock or personal injury , read the “Safety Information” and “Rules for Safety Operation” carefully before using the Meter.
The Model UT33B, UT33C and UT33D Multimeters (hereafter referred as “the Meter”) are 3 1/2 digits with steady operations, fashionable design and highly reliable hand-held measuring instrument. The Meter can measure AC/DC voltage, AC/DC Current, Resistance, Temperature, Diode and Continuity. It is an ideal tool for maintenance. Except where noted, the descriptions and instructions in this Operating Manual apply to all Model UT33B/UT33C/UT33D.
3
Model UT33B/C/D: OPERATING MANUAL
Page 4
Open the package case and take out the Meter. Check the following items carefully to see any missing or damaged part:
In the event you find any missing or damage, please contact your dealer immediately .
Unpacking Inspection
Item
1 2 3 4
Description
Qty
1 piece
1 pair 1 piece 1 piece
English Operating Manual Test Lead Point Contact Temperature Probe (UT33C only) Holster
4
Model UT33B/C/D: OPERATING MANUAL
Page 5
Safety Information(1)
Safety Information
This Meter complies with the standards IEC61010: in pollution degree 2, overvoltage category (CAT I 600V, CAT II 300V) and double insulation.
CAT. I: Signal level, special equipment or parts of equipment, telecommunication, electronic, etc., with smaller transient overvoltages than overvoltages CAT. II.
CAT. II: Local level, appliance, PORTABLE EQUIPMENT etc., with smaller transient overvoltages than CAT. III
Use the Meter only as specified in this operating manual, otherwise the protection provided by the Meter may be impaired.
5
Model UT33B/C/D: OPERATING MANUAL
Page 6
Safety Information(2)
In this manual, a Warning identifies conditions and actions that pose hazards to the user, or may damage the Meter or the equipment under test.A Note identifies the information that user should pay attention on.
International electrical symbols used on the Meter and in this Operating Manual are explained on page 10.
6
Model UT33B/C/D: OPERATING MANUAL
Page 7
l Before using the Meter inspect the case. Do not use the Meter if it is damaged or the case (or part of the case) is removed. Look for cracks or missing plastic. Pay attention to the insulation around the connectors. l Inspect the test leads for damaged insulation or exposed metal. Check the test leads for continuity . Replace damaged test leads with identical model number or electrical specifications before using the Meter. l Do not apply more than the rated voltage, as marked on the Meter, between the terminals or between any terminal and grounding. l The rotary switch should be placed in the right position and no any changeover of range shall be made during measurement is conducted to
Warning To avoid possible electric shock or personal injury, and to avoid possible damage to the Meter or to the equipment under test, adhere to the following rules:
Rules For Safe Operation (1)
7
Model UT33B/C/D: OPERATING MANUAL
Page 8
prevent damage of the Meter. l When the Meter working at an effective voltage over 60V in DC or 42V rms in AC, special care should be taken for there is danger of electric shock.
l Use the proper terminals, function, and range for your measurements. l Do not use or store the Meter in an environment of high temperature, humidity,
explosive, inflammable and strong magnetic field. The performance of the Meter may deteriorate after dampened.
l When using the test leads, keep your fingers behind the finger guards. l Disconnect circuit power and discharge all high-voltage capacitors before
testing resistance, continuity, diodes and current. l Before measuring current, check the Meter’s fuses and turn off power to the circuit before connecting the Meter to the circuit. l Replace the battery as soon as the battery indicator appears. With a low battery, the Meter might produce false readings that can lead to electric shock and personal injury.
Rules For Safe Operation (2)
8
Model UT33B/C/D: OPERATING MANUAL
Page 9
l Remove test leads and temperature probe from the Meter and turn the Meter power off before opening the Meter case. l When servicing the Meter, use only the same model number or identical electrical specifications replacement parts. l The internal circuit of the Meter shall not be altered at will to avoid damage of the Meter and any accident. l Soft cloth and mild detergent should be used to clean the surface of the Meter when servicing. No abrasive and solvent should be used to prevent the surface of the Meter from corrosion, damage and accident.
l The Meter is suitable for indoor use. l Turn the Meter off when it is not in use and take out the battery when not
using for a long time. l Constantly check the battery as it may leak when it has been using for some time, replace the battery as soon as leaking appears. A leaking battery will damage the Meter,
Rules For Safe Operation (3)
9
Model UT33B/C/D: OPERATING MANUAL
Page 10
International Electrical Symbols
Continuity Test
AC Current DC Current
Double Insulated.
Low Battery. Diode. Fuse.
Safety Rules Conforms to Standards of European Union.
Earth Ground
AC or DC
10
Model UT33B/C/D: OPERATING MANUAL
Page 11
The Meter Structure (figure 1)
1) LCD Display
2) HOLD Button
3) Display Backlight Button
4) Rotary Switch
5) COM Input Terminal
6) 10A Input Terminal
7) Other Input Terminals
The Meter Structure
(figure 1)
11
Model UT33B/C/D: OPERATING MANUAL
7
4
3
1
5
6
2
Page 12
Functional Buttons
Below table indicated for information about the functional button operations.
Button Operation Performed
HOLD button
l Press BLUE button once to turn the display backlight on. l Press BLUE button again to turn the display backlight off. l Display backlight will NOT be automatically off unless pressing
the BLUE button
BLUE button
l Press HOLD once to enter hold mode. l Press HOLD again to exit hold mode. l In Hold mode, H is displayed and the present value is shown.
12
Model UT33B/C/D: OPERATING MANUAL
Page 13
Measurement Operation(1)
A. DC Voltage Measurement (see figure 2)
Warning To avoid harms to you or damages to the Meter from electric shock, please do not attempt to measure voltages higher than 500V although readings may be obtained.
The DC Voltage ranges are: 200mV, 2000mV, 20V, 200V and 500V. To measure DC voltage, connect the Meter as follows:
1. Insert the red test lead into theVmAterminal and the black test lead into the COM terminal.
2. Set the rotary switch to an appropriate measurement position in V range.
3. Connect the test leads across with the object being measured. The measured value shows on the display.
(figure 2)
13
Model UT33B/C/D: OPERATING MANUAL
Page 14
Measurement Operation (2)
Note
l If the value of voltage to be measured is unknown, use the maximum measurement
position (500V) and reduce the range step by step until a satisfactory reading is obtained.
l The LCD displays “1” indicating the existing selected range is overload; it is
required to select a higher range in order to obtain a correct reading.
l In each range, the Meter has an input impedance of approx. 10MΩ. This loading
effect can cause measurement errors in high impedance circuits. If the circuit impedance is less than or equal to 10k, the error is negligible (0.1% or less).
l When DC voltage measurement has been completed, disconnect the connection
between the testing leads and the circuit under test.
14
Model UT33B/C/D: OPERATING MANUAL
Page 15
Measurement Operation (3)
B. AC Voltage Measurement (see figure 2)
Warning To avoid harms to you or damages to the Meter from electric shock, please do
not attempt to measure voltages higher than 500Vrms although readings may be obtained.
The AC voltage measurement positions are: 200V and 500V. To measure AC Voltage, connect the Meter as follows:
1. Insert the red test lead into the VmA terminal and the black test lead into the COM terminal.
2. Set the rotary switch to an appropriate measurement position in V range.
3. Connect the test leads across with the object being measured. The measured value shows on the display, which is effective value of sine wave (mean value response).
15
Model UT33B/C/D: OPERATING MANUAL
Page 16
Measurement Operation (4)
Note
l If the value of voltage to be measured is unknown, use the maximum measurement
position (500V) and reduce the range step by step until a satisfactory reading is obtained.
l The LCD displays “1” indicating the existing selected range is overload, it is
required to select a higher range in order to obtain a correct reading.
l In each range, the Meter has an input impedance of approx. 10MΩ. This loading
effect can cause measurement errors in high impedance circuits. If the circuit impedance is less than or equal to 10k, the error is negligible (0.1% or less).
l When AC voltage measurement has been completed, disconnect the connection
between the testing leads and the circuit under test.
16
Model UT33B/C/D: OPERATING MANUAL
Page 17
Measurement Operation (5)
C. DC Current Measurement (see figure 3)
Never attempt an in-circuit current measurement where the voltage between terminals and ground is greater than 60V. If the fuse burns out during measurement, the Meter may be damaged or the operator himself may be hurt. Use proper terminals, function, and range for the measurement. When the testing leads are connected to the current terminals, do not parallel them across any circuit.
Warning
( figure 3)
17
Model UT33B/C/D: OPERATING MANUAL
Page 18
The Model UT33B: the current measurement has 3 measurement positions on the rotary switch: 200µA, 200mA and 10A. The Model UT33C/UT33D: the current measurement has 4 measurement positions on the rotary switch: 2000µA, 20mA, 200mA and 10A
To measure current, do the following:
1. Turn off power to the circuit. Discharge all high-voltage capacitors.
2. Insert the red test lead into the VmA or 10A terminal and the black test lead
into the COM terminal.
3. Set the rotary switch to an appropriate measurement position in A range.
4. Break the current path to be tested. Connect the red test lead to the more positive side of the break and the black test lead to the more negative side of the break.
5. Turn on power to the circuit. The measured value shows on the display.
Measurement Operation (6)
18
Model UT33B/C/D: OPERATING MANUAL
Page 19
Note
l If the value of current to be measured is unknown, use the maximum measurement position (10A)
l and reduce the range step by step until a satisfactory reading is obtained. l When current measurement has been completed, disconnect the connection
between the testing leads l and the circuit under test.
Measurement Operation (7)
19
Model UT33B/C/D: OPERATING MANUAL
Page 20
Measurement Operation (8)
D. Measuring Resistance (see figure 4)
( figure 4)
Warning To avoid damages to the Meter or to the
devices under test, disconnect circuit power and discharge all the high-voltage capacitors before measuring resistance.
The Model UT33B/UT33C: The resistance
measurement positions are: 200, 2000¸, 20k, 200k and 20M The Model UT33D: The resistance measurement positions are: 200, 2000, 20k, 200k,20M and 200M To measure resistance, connect the Meter as follows:
20
Model UT33B/C/D: OPERATING MANUAL
Page 21
1. Insert the red test lead into the VmA terminal and the black test lead into the COM terminal.
2. Set the rotary switch to an appropriate measurement position in range.
3. Connect the test leads across with the object being measured. The measured value shows on the display.
Note
l The test leads can add 0.1to 0.3of error to resistance measurement. To obtain
precision readings in low-resistance measurement, that is the range of 200,
short-circuit the input terminals beforehand and record the reading obtained
(called this reading as X). (X) Is the additional resistance from the test lead.
Then use the equation:
measured resistance value (Y) (X) = precision readings of resistance.
l For high-resistance measurement (>1M), it is normal taking several seconds
to obtain a stable reading.
l When resistance measurement has been completed, disconnect the connection
Measurement Operation (9)
21
Model UT33B/C/D: OPERATING MANUAL
Page 22
between the testing leads and the circuit under test.
Measurement Operation (10)
E. Diodes and Continuity Measurement (see figure 5)
( figure 5)
Testing Diodes
To avoid amages to the Meter or to the devices under test, disconnect circuit power and discharge all the high-voltage capacitors before diodes.
Warning
Use the diode test to check diodes, transistors, and other semiconductor devices. The diode test sends a current through the semiconductor junction, and then measures the voltage drop
22
Model UT33B/C/D: OPERATING MANUAL
Page 23
across the junction. A good silicon junction drops between 0.5V and 0.8V. To test a diode out of a circuit, connect the Meter as follows:
1. Insert the red test lead into the VmA terminal and the black test lead into the COM terminal.
2. Set the rotary switch to (The Model: UT33B) or (The Model: UT33C/ UT33D)
3. For forward voltage drop readings on any semiconductor component, place the red test lead on the component’s anode and place the black test lead on the component’s cathode. The measured value shows on the display.
Note
l In a circuit, a good diode should still produce a forward voltage drop reading of
0.5V to 0.8V; however, the reverse voltage drop reading can vary depending on the resistance of other pathways between the probe tips.
Measurement Operation (11)
23
Model UT33B/C/D: OPERATING MANUAL
Page 24
l Connect the test leads to the proper terminals as said above to avoid error display .
The LCD will display “1 indicating open-circuit for wrong connection. The unit of diode is Volt (V), displaying the positive-connection voltage-drop value.
l When diode testing has been completed, disconnect the connection between
the testing leads and the circuit under test.
Measurement Operation (12)
The Model UT33C/UT33D: Testing for Continuity
To test for continuity, connect the Meter as below:
1. Insert the red test lead into the VmA terminal and the black test lead into the COM terminal.
2. Set the rotary switch to .
3. Connect the test leads across with the object being measured. The buzzer sounds if the resistance of a circuit under test is less than 70Ω.
Note
l The LCD displays “1” indicating the circuit being tested is open.
24
Model UT33B/C/D: OPERATING MANUAL
Page 25
l When continuity testing has been completed, disconnect the connection between
the testing leads and the circuit under test.
Measurement Operation (13)
F. Model UT33C: Temperature Measurement (see figure 6)
To avoid harms to you or damages to the Meter, please do not attempt to input voltages higher than 60V in DC or 30V in AC.
Warning
The Temperature measurement range is -40oC ~1000
o
C
or -40
o
F
~1832
o
F
. To measure
temperature, connect the Meter as follows:
1. Insert the red temperature probe into the
VmA
terminal and the black temperature
probe into the COM temperature.
( figure 6)
25
Model UT33B/C/D: OPERATING MANUAL
Page 26
2. Set the rotary switch to
o
C
or
o
F
.
3. Place the temperature probe to the object being measured. The measured value shows on the display.
Note
l The Meter automatically displays the temperature value inside the Meter when
there is no temperature probe connection.
l The included point contact temperature probe can only be used up to 250
o
C(482oF). For any measurement higher than that, the rod type temperature probe must be used instead.
l When temperature measurement has been completed, disconnect the connection
between the testing leads and the circuit under test.
Measurement Operation (14)
26
Model UT33B/C/D: OPERATING MANUAL
Page 27
Measurement Operation (15)
G. The Model UT33B: Battery Test (see figure 7)
To avoid harms to you or damages to the Meter, please do not attempt to input voltages higher than 60V in DC or 30V in AC.
Warning
( figure 7)
To test the battery, proceed as follows:
1. Insert the red test into the VmA terminal and the black test leads into the COM terminal.
2. Set the rotary switch to an appropriate measurement position in range.
3. Connect the test leads across with the battery being measured ensuring the polarity is correct. The measured value shows on the display, which is the voltage between the cathode and anode of the battery.
27
Model UT33B/C/D: OPERATING MANUAL
Page 28
Note
l When battery testing has been completed, disconnect the connection between
the testing leads and the circuit under test.
Measurement Operation (16)
H. The Model UT33D: Square Wave Output
To avoid damages to the Meter, do not allow output terminals (red test lead) to reach higher than 10V.
Warning
To measure square wave output proceed as follows:
1. Set the rotary switch to OUT.
2. The square wave signal outputs between VmA and COM Terminals.
Note
l The frequency is approx. 50Hz. l The output scope higher than 3Vpp when it is loaded 1MΩ.
28
Model UT33B/C/D: OPERATING MANUAL
Page 29
l When square wave output testing has been completed, disconnect the connection
between the testing leads and the circuit under test.
General Specifications(1)
l Maximum Voltage (including transient overvoltage) between any Terminals
and Grounding: 500V rms.
l Fused Protection for VmA Input Terminal: 315mA, 250V fast type, φ5x20 mm l 10A Terminal: Un-fused. l Range: Manual ranging l Maximum Display: Display: 1999. l Measurement Speed: Updates 2~3 times /second. l Temperature:
Operating: 0oC~40oC (32oF~104 oF).
Storage: -10oC~50oC (14oF~122oF).
l Relative Humidity: 75% @ 0
o
C - 30oC; 50% @ 31 - 40oC.
General Specifications
29
Model UT33B/C/D: OPERATING MANUAL
Page 30
l Altitude:Operating: 2000 m.
Storage: 10000 m.
l Battery Type: One piece of 9V Battery NEDA 1604 or 6F22 or 006P. l Battery Deficiency: Display: . l Negative reading: Display: . l Overloading: Display: 1. l Dimensions (HxWxL): 130 x 73.5 x 35mm. l Weight: Approx. 156g (battery included). l Safety/Compliances: IEC61010 CAT.I 600V overvoltage and double insulation
standard.
l Certification:
General Specifications(2)
30
Model UT33B/C/D: OPERATING MANUAL
Page 31
Accuracy: (a% reading + b digits), guarantee for 1 year. Operating temperature: 23oC 5oC. Relative humidity: <75%. Temperature coefficient: 0.1 x (specified accuracy) / 1oC.
Accuracy Specifications(1)
A. DC Voltage
Remark: Input impedance: 10M.
Range Resolution
200mV
2000mV
20V 200V 500V
Accuracy
UT33B UT33C UT33D
Overload
Protection
100µV
1mV
10mV
100mV
1V
(0.5%+2)
(0.8%+2)
250V DC or AC
500V DC or AC
31
Model UT33B/C/D: OPERATING MANUAL
Page 32
B. AC Voltage
Accuracy Specifications(2)
Remarks:
l Input impedance: approx. 5M. l Displays effective value of sine wave (mean value response). l Frequency response 40Hz ~ 400Hz.
Range Resolution
Accuracy
UT33B UT33C UT33D
Overload
Protection
200V 500V
100mV
1V
(1.2%+10) 500V DC or AC
32
Model UT33B/C/D: OPERATING MANUAL
Page 33
Accuracy Specifications(3)
C. DC Current
Remark:
l At 10A Range: For continuous measurement 10 seconds and interval not less
than 15 minutes.
Range Resolution
200µA
2000µA
20mA
200mA
10A
Accuracy
UT33B UT33C UT33D
Overload
Protection
(1.2%+2)
(2%+5)
0.1µA 1µA
10µA
100µA
10mA
315mA, 250V
fast type fuse:
φ 5x20mm
Un-Fused
(1%+2)
(1%+2)
33
Model UT33B/C/D: OPERATING MANUAL
Page 34
D. Resistance
Accuracy Specifications(4)
Range Resolution
200
2000
20k
200k
20M
200M
Accuracy
UT33B UT33C UT33D
Overload
Protection
0.1 1
10 100 10k
100k
(0.8%+5)
(1%+5)
(0.8%+2)
[5%(reading-10)+10]
250V DC
or AC
34
Model UT33B/C/D: OPERATING MANUAL
Page 35
E. Diodes and Continuity Measurement (Continuity test only for UT33C/UT33D)
F. The Model UT33C: Temperature
1 Buzzer beeps at <70
Range Resolution Remark Overload Protection
1mV
Displays approximate forward
voltage drop: 0.5V~0.8V.
250V DC or AC
Accuracy Specifications(5)
Range Resolution Remark Overload Protection
250V DC or AC
-40oC~150oC
150oC~1000oC
-40
o
F
~302
o
F
302
o
F
~1832
o
F
1
o
C
1
o
F
(1%+3)
(1.5%+15)
(1%+4)
(1.5%+15)
35
Model UT33B/C/D: OPERATING MANUAL
Page 36
G. The Model UT33B: Battery Test
Accuracy Specifications(6)
Range Resolution Internal Resistance
12V
9V
1.5V
10mV 10mV 10mV
H. The Model UT33D: Square Wave Output
Range
Illustration
Approx. output 50Hz square wave signal. As a simple signal source with 47k resistance output.
OUT
Remark:
l No overload protection. l Make sure voltage output of calibrated equipment level is less than 10V to avoid
damages to the meter.
240
1.8k 30
36
Model UT33B/C/D: OPERATING MANUAL
Page 37
This section provides basic maintenance information including battery and fuse replacement instruction.
Maintenance(1)
A. General Service
l Periodically wipe the case with a damp cloth and mild detergent. Do not use
abrasives or solvents.
l To clean the terminals with cotton bar with detergent, as dirt or moisture in the
terminals can affect readings.
l Turn the Meter to OFF position when it is not in use and take out the battery
when not using for a long time.
Do not attempt to repair or service your Meter unless you are qualified to do so and have the relevant calibration, performance test, and service information. To avoid electrical shock or damage to the Meter, do not get water inside the case.
Warning
37
Model UT33B/C/D: OPERATING MANUAL
Page 38
l Do not store the Meter in a place of humidity, high temperature, explosive,
inflammable and strong magnetic field.
B. Replacing the Battery (see figure 8)
T o avoid false readings, which could lead to possible electric shock or personal injury, replace the battery as soon as the battery indicator “ ”appears.
( figure 8)
Warning
Screw
To replace the battery:
1. Disconnect the connection between the testing leads and the circuit under test, and remove the testing leads away from the input terminals of the Meter.
Maintenance(2)
38
Model UT33B/C/D: OPERATING MANUAL
Page 39
2. Turn the Meter to OFF position.
3. Remove the screw from case bottom, and separate the case bottom from the case top.
4. Remove the battery from the battery compartment.
5. Replace the battery with a new 9V battery (NEDA 1604 or 6F22 or 006P).
6. Rejoin the case bottom and case top, and reinstall the screw.
C. Replacing the Fuses (See figure 8)
Maintenance(3)
To avoid electrical shock or arc blast, or personal injury or damage to the Meter, use specified fuses ONLY in accordance with the following procedure.
Warning
To replace the Meter’s fuse:
1. Disconnect the connection between the testing leads and the circuit under test, and remove the testing leads away from the input terminals of the Meter.
2. Turn the Meter to OFF position.
39
Model UT33B/C/D: OPERATING MANUAL
Page 40
3. Remove the screw from case bottom, and separate the case bottom from the case top.
4. Remove the fuse by gently prying one end loose, and then take out the fuse from its bracket.
5. Install ONLY replacement fuses with the identical type and specification as follows and make sure the fuse is fixed firmly in the bracket. 315mA, 250V, fast type, φ 5x20mm.
6. Rejoin the case bottom and case top, and reinstall the screw.
Replacement of the fuses is seldom required. Burning of a fuse always results from improper operation.
** END **
Maintenance(4)
40
Model UT33B/C/D: OPERATING MANUAL
Page 41
41
Model UT33B/C/D: OPERATING MANUAL
Page 42
Copyright 2002 Uni-Trend Group Limited.
All rights reserved. Manufacturer:
Uni-Trend Technology (Dongguan) Limited Dong Fang Da Dao Bei Shan Dong Fang Industrial Development District Hu Men Town, Dongguan City Guang Dong Province China Postal Code: 523 925
Headquarters: Uni-Trend Group Limited Rm901, 9/F, Nanyang Plaza 57 Hung To Road Kwun Tong Kowloon, Hong Kong Tel: (852) 2950 9168 Fax: (852) 2950 9303 Email: info@uni-trend.com http://www.uni-trend.com
42