Carbon Dioxide Monitor
User Manual
I. Introduction
CO2 is a naturally occurring gas that emitted at great levels by human activity and it is one of several greenhouse gases in our atmosphere. CO 2 only accounts for 0.03% of the fresh air. If you breathe high concentration CO , you may be at risk of carbon dioxide intoxication. Symptoms of
2
carbon dioxide toxicity include headache, irregular heartbeat, nausea, unconsciousness, or even death.
A37 is a CO Monitor that measures CO concentration, temperature and 2 2 humidity. This monitor is widely used in homes, shopping malls, offices, vehicles, stations etc.
Please read carefully and follow all warnings and precautions before using the device.
II. Open box inspection
Open the box and take out the device. Please check whether the following items are deficient or damaged and contact your supplier immediately if they are.
A37 CO monitor -------------------- 1
2
Protective case ----------------------- 1 User manual -------------------------- 1 USB cable ----------------------------- 1
III. Safety information
1. Check the device and its accessories for any damage or anomaly before use. Do not use the device if the cover is damaged, or has any other malfunctions.
2. Please follow the operation instructions.
3. Please do not open the device case and modify the internal circuit.
4. Do not store or use this device in high temperature, high humidity, flammable, explosive and strong electromagnetic environments.
5. Charge the monitor when symbol is displayed.
6. Regularly use soft cloth and mild detergent to clean the case. Do not use abradant or solvent.
7. A37 complies with CE standards.
8. This monitor has ABC (Automatic Baseline Correction) function. Do not use this monitor in closed environment where CO concentration 2 remains relatively high.
9. Use specified adapter to charge the device. The adapter should comply with UL/FCC safety standards.
IV. Structure
1
2
3
4
①: Case LCD screen USB port Functional buttons: : :
V.Symbols
10
1
3
7
12
11
5
4
6 2
8
9
Indication Indication
Audio alarm ON/OFF
1
STEL/TWA/MAX
2
/MIN/AVG modes
3
CO 2 calibration
Timing shutdown
4
Battery status
5
Exit calibration
6
No.No.
7
CO concentration2
CO 2
8
grading emoticon
Relative humidity
9
Date/time
10
Temperature
Data hold
12
concentration
VI. Operation instructions
1. Functional buttons
The following table shows the buttons functions in dif ferent modes.
Modes
OFF
Measure
Calibration
2.Operations
CO 2 measurement Long p re ss to turn on the monitor. It ta kes 120s to stab il ize CO2 concentration. concentration, ambient temperature, relative humidity CO2 will be displayed on the screen. concentration value is refreshed CO2 every 3 seconds.
Note: if >1000ppm, buzzer goes off with red backlight flashing. The CO2 audio alarm’s default setting is OFF, long press to turn on the audio alarm.
1)Concentration grading emoticons
Data Short press MODE to cycle switch between HOLD/MAX/MIN/AVG /TWA/STEL values.
Symbol
HOLD
MAX
MIN
AVG
TWA
STEL
Long press: turn on/off the monitor
Long press: enter calibration mode
Short press: turn on/off the backlight
Short press to switch temperature unit ˚C/˚F
Long press to turn on/off audio alarm
Short press: cycle switch between HOLD/MAX/MIN/AVG/TWA/STEL
Long press: reset to zero
Long press: calibrate the CO2 concentration (outdoor fresh air)
Short press: increase the value
Short press : next parameter (double press to enter date/time calibration)
Long press: save
Short press: decrease the value
Short press: previous parameter
Long press: exit calibration
400ppm ~ 450ppm; 450ppm 700ppm; ~ 700ppm 1000ppm; ~ 1000ppm 2000ppm; ~ 2000ppm 5000ppm; ~ 5000ppm以上;
Indication
Data hold
Maximum value
Minimum value
Average value
Time weighted average
Short-term exposure limit
Functions
Buttons
TWA: The average exposure within the workplace to any hazardous contaminant or agent using the baseline of an 8 hour day or 40 hours week work schedule.
STEL: The acceptable exposure limit to a toxic or irritant substance over a short period of time (time-weighted average), usually 15 minutes.
Auto shutdown
The monitor will automatically shut down if there is no operation for 2 hours.
Backlight
Please refer to Operation Instructions-Functional Buttons
Audio alarm
Please refer to Operation Instructions-Functional Buttons
Temperature unit
Please refer to Operation Instructions-Functional Buttons
Battery
Charge the battery when is displayed on the screen. It takes 2~3 hours to fully charge the battery.
3.Calibration
Date/time
Please refer to Operation Instructions-Functional Buttons to calibrate date/time.
CO2 Note: function has been calibrated before delivery. If the value is CO2
greatly deviated from the standard, you can calibrate it by either methods
below:
1) Automatic calibration
A37 has the built-in self-correcting ABC (Automatic Baseline Correction) algorithm to automatically adjust the base value. a) Turn on the monitor and leave A37 at a place where there is flowing fresh air for 3 days (powered by the adaptor) b) ABC function will automatically adjust and correct the standard value.CO2 Warnings: ABC function is always on, do not use this monitor in closed environment where concentration remains relatively high.CO2
Backlight
Operating environment
Storage environment
Power supply
Weight
Dimension
IX. CO2 concentration and effects (only for reference)
Concentration
350~450ppm
350~1000ppm
1000~2000ppm
2000~5000ppm
>5000ppm
0˚C~50 C, 0%RH~85%RH˚
-20 C~60 C, 0%RH~95%RH˚ ˚
Built-in 3.7V DC rechargeable battery
195g
75mm x 55mm x 130mm
Effect
Fresh air
Normal
Turbid air, sleepy
Headache, irregular heartbeat, nausea
Unconsciousness or even death
X. Maintenance
a.The maintenance and service of the device must be implemented by qualified professionals or the designated departments. b.Regularly use dry cloth to clean the cover. Do not use the detergent containing solvent composition c.To conserve battery, please turn off the device if do not being used for a long time.
2) Manual calibration
a) Turn on the monitor, leave it outdoor for 20 minutes b) Long press until appears on the screen. Calibration takes 10 minutes.
symbol disappears when calibration is done. Note: It is suggested to calibrate the monitor in outdoor environments where there is air flow. Keep device away from animals, plants and human.
VII. Specifications
1.Technical specifications
Function
Temperature
Humidity
*1: Measure environment: 1atm, ambient temperature and humidity *2: It may be influenced by heat emitted by the adaptor
Range
400~5000 ppm
-10.0˚C~50.0˚C
0-99%RH
Resolution
1ppm
0.1˚C
1%RH
2.General specifications
Specifications
Sensor
ABC
Audio alarm
Visual alarm
Display
Low battery indication
Auto power off
Descriptions
NDIR principle
Automatic baseline correction
Buzzer
Red backlight
LCD screen
Accuracy
±(5% of rdg+50ppm) from 400~2000ppm *1
±1˚C *2
(10%RH90%RH) ±5%RH @25˚C