(1) POWER SWITCH : on/off switch
(2) OUTPUT TERMINAL : positive polarity (+) is red
negative polarity (-) is black
earth and chassis ground is green
(3) DIGITAL VOLT METER : indicates the output voltage
(4) DIGITAL CURRENT METER : indicates the output current
(5) CV LAMP : light when this unit in constant current operation
(6) CC LAMP : lights when this unit is constant current operation
(7) VOLT ADJ (COARSE) KNOB : for the coarse adjustment of the output voltage
(8) VOLT ADJ (FINE) KNOB : for the fine adjustment of the output voltage
(9) AMP ADJ KNOB : for the adjustment of the output current
(10) DC OUTPUT SWITCH : on/off switch for the DC OUTPUT
(11) CC SET SWITCH : current limiting set-up without output terminal shorting orvarious loads, and provide current limit value preview facility
(12) INDEPENDENT/TRACKING SWITCH (TRIPPLE ONLY)
Covering from indepe ndent applying of S1 and S2 outputs to a serial conn ect. In tracking mode , because (-) of S2 and (+) of S1 are internally
connected, it is output in between (+) of S2 and (-) of S1
In case short pins are used in between (_) terminal and ground, eliminate short pins from output terminals of S1 and S2.
WARNING : Before connecting line power to your power supply, make sure that the AC input
voltage is correct for your power source.
TURN-ON CHECKOUT PROCEDURES.
a) Turn A adj. control fully counter clockwise.
b) Set AC power switch push to on position, digital display and CV lamp should light.
c) Turn VOLTAGE controls fully counter clockw ise to ensure that output d ecrease to 0V dc
then fully clockwise to ensure that output voltage increase to maximum output voltage.
d) While depressing CC SET push button, turn the CURRENT contro l fully counter clockw ise
and then fully clockwise to ensure that the current limit value can be set from zero to
maximum rated value.
e) Connect load to output terminals.
CONSTANT VOLTAGE OPERATION
To set up a power supply for a constant voltage operation, proceed as follow;
a) Turn on power supply and adjust V adj. Control for desired output maximum (output
terminal open). CV lamp should light.
b) While depressing CC SET push button, adjust A adj. Control for operation, if a load
change causes the current limit to be exceeded, the power supply will automatically
crossover to constant current mode and output voltage will drop proportionately.
c) Push-on DC OUT push button switch for DC voltage output.
To s et up a power supply for a constant current operation, proceed as follow;
a) Turn A adj. control fully counter clockwise to ensure that output decreases to 0A, and
then on the power supply.
b) Adjust V adj. Control (no load conn ected) fo r maximum ou tput vo ltage allowable (vo ltage
limit), as determined by load conditions.
During actual operation, if a load change causes the voltage limit to be exceeded. The
power supply will automatically crossover to constant voltage operation at the preset
voltage limit and output current will drop proportionately.
c) Adjust A adj. control for desired output current while depressing CC SET button.
d) Push-on DC OUT push button switch for DC voltage output.
CONNECTING LOADS
The output of the supply is isolated from earth ground.
Either output terminal may be grounded or the output can be floated up to 240 volts off ground.
TRACKING OPERATION (TRIPPLE OUTPUT ONLY)
To select a tracking operation output as follows;
a) Set independent and tracking select switch to tracking position.
b) Depressing CC SET push button switch and adjust A adj. control of the each for
maximum output current allowable(current limit), as determined by load conditions.
c) Adjust V adj. controls (COARSE and FINE) of the MASTER fro desired output Voltage.
This time, V adj. controls (COARSE and FINE) of the SLA VE should be set on maximum.
d) Connect load using the negative (black) terminal of the SL AVE supply and the positive
WARNING : The following instructions are for use by qualified personnel only. To avoid
electrical shock, do not perform any servicing other then contained in the operating instructions
unless you are qualified to do so.
Fuse Replacement
If the AC fuse blows, the CV or CC lamp will not light and the power supply will not operate. If
the DC fuse blows, the CV or CC lamp and digital display is light but DC OUTPUT will not
operate.
The fuse should not normally open unless a problem has developed in the unit. Try to
determine and correct the cause of the blown fuse, then replace only with a fuse the correct
rating and type.
The fuse located on the real panel.
line Voltage Conversion
The primary winding of the power transformer is tapped to permit operation from 115 or 230
VAC, 50/60Hz line voltage. Conversion from one line voltage to another is done by AC select
switch on the real panel.
The line voltage to swit ch the unit was factory setted. To convert to a diff erent line voltage,
perform the following procedure;
a) Make sure the power cord is unplugged.
b) Change the AC selects switch to the desired line voltage position.
c) A change in line voltage may also require a corresponding change of fuse value. Install
a) Connect digital multimeter across output terminals of supply and set the DC volt position.
b) Turn on supply and push on DC OUT push button switch.
c) Adjust voltage controls (COARSE and FINE) fully clockwise.
d) Adjust S203 for a reading of rate voltsÝ
e) Adjust S401 until digital multimeter reads exactly maximum rated output voltage
f) Push-off DC O UT push butto n switch and adjust S202 until front panel met er reads exactly
digital multimeter reading.
1.05 on the digital multimeter.
ADJUSTMENT OF THE RATING CURRENT.
a) Connect load resistor and digital multimeter with series and set the range to DC 20Amp
position.
b) Turn on supply.
c) Adjust voltage controls and current control to minimum (fully counter clockwise).
d) Adjust S201 for a reading of 0 Amp on the digital multimeter.
e) Adjust voltage controls and current control to minimum (fully clockwise).
f) Adjust S501 until front panel meter reads exactly digital multimeter reading.
g) Push-on CC SET push button switch.
h) Adjust S204 until digital multimeter exactly the maximum rated output current.