This meter is an intelligent and multi-functional device for various measurement. It features high quality, high
accuracy, high reliability, special high frequency measurement and low cost. The meter is in compliance with IEC
6010 CAT II 1000V and CAT III 600V high voltage standards. Other features of this instrument are listed below:
Multi-display: primary 80000 counts, secondary 80000 counts, bar graph 23 segments
50 measuring functions, with basic DCV, ACV, DCA, ACA, Ω, CAP, Hz, TEMP, diode and Continuity
measuring function, etc.
18 types of frequency, frequency up to 80MHz, 1800 waveform outputs, 0.1%~99% duty cycle.
Simultaneous measure (AC+DC), (AC+Hz), (DC+dBm), (dBm+Hz), (Hz+Duty), (℃+℉).
Auto data update and refresh, auto data hold, auto peak hold.
36-hour dynamic record: MAG, MIN, AVG, MAX-MIN (REL△), (REL%), setting upper & lower limit,
timing measurement
AC measuring adopts highly accurate true RMS measurement, with testing frequency bandwidth and
AC+DC measuring, capable of accurate true RMS measuring of any waveforms in AC range
RS-232 interface
SAFETY INSTRUCTIONS
High voltage GND Dual insulation Refer to manual.
WARNING! To avoid the electric shock and physical injury, and to avoid possible damage to the meter
and the tested equipments, read this operation manual carefully before using this meter, and follow the following
safety guidelines:
Before use, check and make sure that the instrument’s plastic chassis, the test leads and the insulation layer
are intact.
User the meter only as specified in this manual. Otherwise, the protection provided by the meter may be
expired.
Never measure voltage while the test leads are inserted into the current input terminals.
Do not use the meter if it looks damaged
Inspect the leads for damaged insulation or exposed metal, check test lead continuity. Replace damaged
leads.
Disconnect the power and discharge all high-voltage capacitors before testing in resistance, continuity and
diode function.
Be cautious when working at voltage above DC60V or AC42V. Such voltages may cause a shock hazard.
When undertaking measurement, keep your fingers behind the guard’s plant on the test leads or probes.
Select the proper function and range for measurement to avoid damage to the meter.
Disconnect the test leads from the test points before changing to another function.
The specifications are subject to change without notice.
The content of this manual is regarded as correct. If any error or omits is found, please contact with the
manufacturer.
We hereby will not be responsible for the accident and damage caused by improper operation.
The function stated for this User Manual cannot be the reason of special usage.
RANGE: Selects measurement range. The meter is default at auto range mode.
SET:
When the SET button is in operation, the RANGE button is used as a moving up button (
button as a moving down button (
right (
disabled. buttons can be used to enter and adjust the setting values.
Press SET button for at least 2 seconds to start the backlight. Press this button again to turn off the backlight.
The backlight can auto off if this button is not engaged within 30 seconds.
MAX/MIN:
Presses this button to enter the dynamic record mode, with maximum record period of 36 hours.
In the dynamic record mode, the meter automatically records the maximum value (MAX), minimum value
(MIN), difference value (MAX-MIN) and calculate the average value (AVG) of all readings. Press this button
to cycle MAX, MIN, AVG, MAX-MIN on the secondary display.
Pressing this button for at least 2 seconds, the meter returns to auto range.
Timer/RS232:
TIME:
TIMER function is enabled only in the REL△ and MAX/MIN measuring mode.
Press TIMER key to start the secondary display for counting time. Press TIMER key again to turn off the
counting time display.
When the secondary display is display counting time, press key SELECT to enter Beeper setup for setting up
a timer for beeper. Then, press key REL
time. Next press key TIMER to validate the new time data. When the counting time exceeds the preset time,
the beeper sounds.
In the above two modes, press key HOLD to stop timer function and turn off secondary display, but the
preset time remains unchanged.
The time is display in the format of 8.88.88 on the secondary display. The max.counting time is 9.59.59
RS232:
Pressing this button for at least 2 seconds, the RS232 remote control is enabled for PC control and
communication with other instruments. The LCD display “RS232’
Auto power off function is disabled in this RS232 programmable mode.
Pressing this button for at least 2 seconds again to exits this mode and return to normal mode.
HOLD:
Pressing this button, the meter enters auto data hold mode and “A-H” is displayed on the LCD.
The data hold mode allows users to hold the displayed value while the analog bar graph shows the current
reading.
Pressing this button again, the meter enters Peak+ hold mode and a “PH+” appears on the LCD display.
Pressing this button again, the meter enters Peak- hold mode and a “PH-” appears on the LCD display.
Pressing this button for at least 2 seconds, the meter exits HOLD mode and return to normal mode.
). In this case, the RANGE, 2nd VIEW, MAX/MIN and REL△ buttons’ original functions are
), the MAX/MIN as a moving left button ( ), and REL△ as a moving
( ), MIX/MIN ( ), RANGE ( ), 2nd VIEW ( ) to input the
Press this key for at least 2 seconds to return to 606.10Hz, 50% duty output state.
When the SET button is in operation, 2nd VIEW button is using as a moving down button ( ). Pressing this
button moves the setting digit down
REL△:
Pressing this button, the meter enters relative measuring mode and “REL△” appears on the L CD di spl ay.
The relative measuring functions measures the difference between the testing value and the reference value.
The current readings on the secondary display are used as a relative value. The primary display displays the
relative measurement in two modes:
One is: REL△=measuring value – Reference value
The other is: REL%= (REL△/Reference val ue)x 100% (press SELECT button to select REL△ or
Press REL
display.
While the SET button is in operation, the REL
setting digit to the right.
Press REL
Set up reference value for measurement:
2. In every reference value, use RANGE button to select a proper range
3. When SET button is in operation, press SELECT button twice to set up reference value for
4. Use
Press SET button to validate the new setup.
3. Terminal
COM: Common terminal for all measurements
V Ω Hz: Volts, Ohn, Diode, Freq., Temp, and Cap. Measurement and square wave output terminal
mA: Milli ampere current measurement terminal
measurement. At the same time, the
dBm+Hz dBm Hz/ACV/DCV/ACV+DCV
ACmV+Hz ACmV
dBm+Hz dBm Hz/ACmV/DCmV/ACmV+DCmV
REL% mode)
button again, the testing value will be used as reference value and displayed on the secondary
button is used as a moving right button ( ) to move the
for at least 2 seconds to exits reference mode and return to normal mode.
WANRNING: The testing AC voltage must not exceed AC750V!
The measurement of AC voltage has three modes: ACV,
ACV+Hz and dBm.
Set the function key to “
Press SELECT button to select measurement mode.
According to practical demands, press REL
and 2nd VIEW buttons to have relative measuring or
record.
The meter is in auto range mode and will automatically
select a most suitable range. To select a range manually for
ACV measurement, press RANGE button to select a proper
range. There are three ranges to choose from:
8.0000V/80.000V/750.00V
Connect the black test lead to “COM” terminal and the read test lead to “V Ω Hz” terminal.
Touch the test leads to the test points and read the display on both the primary and secondary display.
” position.
, MAX/MIN
NOTE:
1. In dBm measurement mode, the decimal point of dBm is fixed between the second and third digits.
2. When in dBm measurement, the impedance is default at 600Ω. To change the impedance, press RANGE
button to select a proper impedance. The impedance can be chosen from:
the measurement, press RANGE button to select a proper range. There are two ranges to choose from:
Press SELECT
Primary display
Secondary display
Press 2nd VIEW
DCmV
DCmV
ACmV+Hz
ACmV
dBm
dBm
Hz/ ACmV/ DCmV/ DCmV+ ACmV
△
80.000mV/800.00mV
Connect the black test lead to “COM” terminal and the read test lead to “V Ω Hz” terminal.
Touch the test leads to the test points and read the display on both the primary and secondary display.
NOTE:
1. In dBm measurement mode, the decimal point of dBm is fixed between the second and third digits.
2. When in dBm measurement, the impedance is default at 600Ω. To change the impedance, press RANGE
button to select a proper impedance. The impedance can be chosen from:
3. In dBm measurement mode, REL△, MAX/MIN, AVG, A-H functions are disabled.
4. In milli voltage measurement mode, in order to obtain DC+AC function, the input terminal of ADC does not
employ coupling capacitor. Therefore, never apply a voltage over double value of DC or AC voltage of the
rated value of this range.
4.4 AC/DC Milli Current Measurement (ACmA, DCmA)
WANRNING:
To avoid injury and damage to the meter, never attempt an in-circuit current measurement when the
fuse is blown, or when the voltage between open circuit and the ground is 1000V.
To avoid damage to the meter, check the meter’s fuse before proceeding.
Before current measurement, turn of the circuit power supply, discharge the high voltage capacitance.
Under no circumstances, do not test the voltage when the test leads are plugged in “mA” or “20A”
terminals.
Do not place the probes in parallel with a circuit or
component when the leads are plugged into the current
terminals.
The measurement of AC/DC milli current has three modes:
DCmA, ACmA, DCmA+ACmA, and ACmA +Hz
Set the function key to “mA
Press SELECT button to select measurement mode.
According to practical demands, press REL
and 2nd VIEW buttons to have relative measuring or
record.
The meter is in auto range mode and will automatically
select a most suitable range. To select a range manually for the measurement, press RANGE button to select
a proper range. There are two ranges to choose from: 80.000mA/800.00mA
Connect the black test lead to “COM” terminal and the read test lead to “mA” terminal.
Break the circuit path to be tested. Tough the black test leas to the negative of the break and the red test lead
to the positive of the break. Read the display.
After finishing measurement, disconnect the power supply to the circuit and discharge all high voltage
Connect the black test lead to “COM” terminal and the read test lead to “VΩ Hz” terminal.
Touch the probes to the test points and read the display.
Ω Ω/kΩ/MΩ MAX/MIN, REL△ param eters
Continuity mode
Press SELECT button to select “
If the testing points resistance falls below 50Ω. The beeper will sound.
Hi resistance
This function is used to measure the resistance above 80MΩ.
Press SELECT button to select “HiΩ” range. The primary display will display “Hi”.
A single range is 8000.0MΩ. If the testing points resistance falls below 10MΩ and above 8000.0MΩ, “OL”
will appears on the display.
4.7 Capacitance Measurement (CAP)
” range.
WANRNING: To avoid damage to the meter or the testing equipments disconnect circuit power and
discharge all high voltage capacitors before measuring capacitance. Use the DC function to confirm that the
capacitor is discharged.
NOTE:
Some capacitors have polarities. In measuring polarities capacitors, touch the red test lead to the positive
polarity and the black test lead to the negative polarity.
In measuring low capacitance, the resistance of the test leads may cause an error in the test results. To avoid
this error, first short the test leads, next press REL
secondary display displays the resistance of test lead. Measure the to-be tested resistance and the result will
be displayed on the primary display.
Capacitor is capable of storing electric charge. When testing capacitance, only the value on a stable display is the
correct result.
Set the function to “
According to practical demands, press REL
measuring or record.
The meter is in auto range mode and will automatically
select a most suitable range. To select a range manually for
the measurement, press RANGE button to select a proper
The meter is a precise and intelligent instrument. It has been accurately calibrated in the factory before shipment.
Readjustment is recommended only if repairs have been made in a circuit affecting adjustment accuracy or if you
have a reason to believe the unit is out of adjustment.
The following instructions are for use by qualified personnel only. To avoid electrical shock, do not perform any
servicing other than contained in the operating instructions unless you are qualified to do so.
Do not verify the circuit to avoid damaging. Power fuse: 200mA/250V (located in input power socket); Fuse for
measuring current: 800mA/250V (located in the current input terminal), 13A/250V (located the main circuit). The
fuse must be replaced by qualified personnel.
NOTE:
Do not connect the voltage higher than DC1000V or AC 1000V rms.
Do not measure voltage at the Ω range.
When replacing fuse, please take away the test leads from the measuring point and power off at first.
Keep the instrument away from water, dust and shock.
Do not operate the meter in high temperature or strong magnetic place.
Do not use the abrasives or solvents to clean the meter.
7.1 Fuse replacement
NOTE: Please select fuse of the same specification to replace it.
Power fuse: 200mA/250V (located in input power socket); Fuse for measuring current: 800mA/250V (located in
the current input terminal), 13A/250V (located the main circuit).
1. Te s t fuse replacement:
Press the function key to Ω range
Connect a test lead to “VΩHz” terminal, use the pen tip to touch the “mA” or “20A” terminal to test the
resistance of the fuse
If the tested fuse resistance is below 5Ω, it means the fuse is good.
If the tested result is “OL” (over load), the fuse needed to be replaced.
If the fuse is tested as good but the multimeter cannot carry measurement, please send the multimeter for
repair.
2. Fuse replacement
Disconnect the multimeter from AC power source, and take off all test leads.
2A fuse replacement
Press “mA” input jack by finger; take out the fuse jack after turning 90°angle anti-clockwise
Take off the fuse, and replace an equivalent then turn on the fuse jack.
After press the fuse into the fuse jack, turn 90°angle by an inverse hour direction.
13A fuse replacement (must be replaced by qualified personal.)
Screw off the four screws on the button shell by a screwdriver, take off the upper shell.
Take off the fuse, and replace an equivalent then turn on the fuse jack.
Take on the shell, .Screw on the four screws on the button shell by a screwdriver.