Use of Operation Manual
Please read through and understand this Operation Manual before operating the product. After
reading, always keep the manual nearby so that you may refer to it as needed. When moving the
product to another location, be sure to bring the manual as well.
Calibration Notification
We notify that the instruments included in this manual are in compliance with the features and
specifications as stated in this manual. Before shipment, the instrument has been calibrated in
factory. The calibration procedures and standards are compliant to the national regulations and
standards for electronic calibration.
Warranty
We guarantee that the instrument has been passed strict quality check. We warrant our instrument’s
mainframe and accessories in materials within the warranty period of one year. We guarantee the
free spare parts for products which are approved defective in this period. To get repair service,
please contact with your nearest sales and service office. We do not provide any other warranty
items except the one being provided by this summary and the warranty statement. The warranty
items include but not being subjected to the hinted guarantee items related to tradable characteristics
and any particular purpose. We will not take any responsibility in cases regarding to indirect,
particular and ensuing damage, such as modifications to the circuit and functions by the users,
repairing or component replacement by the users, or damage during transportation.
For product improvement, the specifications are subject to change without prior notice. This
manual is meant to be an operation guide to users and has been prepared with the utmost care.
However, if you have any questions or note any errors or omissions, please contact your local
distributor.
I
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 3
Safety Instruction
This chapter contains important safety instructions that you must follow when operating the instrument and when
keeping it in storage. Read the following before any operation to insure your safety and to keep the best condition
for the instrument.
1. Safety Terms and Symbols
The following safety symbols may appear in this manual or on the instrument:
WA RN I N G Identifies conditions or practices that could result in injury or loss of life.
CAUTION
Identifies conditions or practices that could result in damage to the
instrument or to other properties.
2. Safety Guidelines
3. Power Supply
4. Fuse
DANGER High voltage
ATTENTION Refer to the manual
Earth (ground) terminal
Before plugging into local AC mains, check and make sure that the output voltage
is compatible to the load. (It is suggested to disconnect a load before plugging into
local AC mains.
Do not use this instrument near water.
Do not operate or touch this instrument with wet hands.
Do not open the casing of the instrument when it is connected to AC mains.
Do not use the instrument in an atmosphere which contains sulfuric acid mist or
other substances which cause corrosion to metal.
Do not use the instrument in a dusty place or a highly humid place as such will
cause instrument reliability degradation and instrument failures.
Install the instrument in a place where is free from vibration.
Install the instrument in a place where the ambient temperature is in range of
-10~70℃ . Note that the instrument operation may become unstable if it is
operated in an ambient temperature exceeding the range of 0~40℃
AC Input voltage: 110V~120V/220~240V ±10%, 50/60Hz
Connect the protective grounding conductor of the AC power cord to an earth ground
to avoid electrical shock.
Fuse type: 110~120V: T2A /250V, or 220~240V: T1A/250V.
Make sure the correct type of fuse is installed before power up.
Replace the AC fuse with the same type and rating as the original fuse.
Disconnect the power cord before fuse replacement.
Make sure the cause of fuse blowout is fixed before fuse replacement.
I I
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 4
5. Maintenance Inspection
6. Cleaning
Inspect the instrument at regular intervals so that it maintains its initial
performance for a long time.
Check the input power cord for damage of the vinyl cover and overheating of the
plug and cord stopper. Check the terminal screws and binding posts for loosening.
Remove dust from the inside of the casing and ventilation holes of the cover by
using a compressed air of the exhaust air of a vacuum cleaner.
Before cleaning, disconnect the AC mains.
To clean the power supply, use a soft cloth dampened in a solution of mild
detergent and water. Do not spray cleaner directly onto the instrument,
since it may leak into the cabinet and cause damage.
Do not use chemicals containing benzene, benzene, toluene, xylene,
acetone, or similar solvents.
Do not use abrasive cleaners on any portion of the instrument.
III
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 5
Introduction
This operation manual is used for all models of this series of DDS function generator. The last two digits in the
model number represent the maximum frequency of the channel A. There are four models to choose from this
series, with max.frequency range of 10MHz, 20MHz, 40MHz and 60MHz.
This series of DDS function generator uses Direct Digital Synthesis (DDS) technology. Its outstanding
performance and system features make this function generator a perfect solution for your testing requirement. The
simplified and optimized design of the front panel and dual-language (English/Chinese) LCD display interface
make your testing much easier for operation and observation. Additionally, the extendable optional functions can
also improve your system characteristics.
Technical characteristics and system features:
High-accurate Frequency: up to the 10
High Frequency Resolution: full-range resolution is 40 mHz (extendable to 40µHz)
No Range Limitation: no frequency level switch, frequency set directly by digit keypad
No Frequency Transition: momentarily switch to the stable value on phase and amplitude without
transition and distortion
Precise Output Waveform: output waveform is synthesized by the preset function calculation values,
very accurate and less distortion
Multiple Waveforms: output 32 kinds of the pre-stored waveforms
Pulse Characteristics: set accurate pulse duty cycle
Harmonic Wave Characteristics: output the fundamental and harmonic waves with adjustable phase
Sweep Characteristics: frequency sweep and amplitude sweep, free to set the start and stop points
Modulation Characteristics: output FM and AM signals
Shift Keying Characteristics: output FSK, ASK, and PSK signals
Burst Characteristics: output pulse waveform with burst count
Store Characteristics: store 40 sets of the user parameters and recall
Counter Characteristics: count the frequency, period, amplitude RMS value or peak-to-peak value
Operation Mode: key operation for all functions, English/Chinese menu selectable, parameters settings
by numeric keypad or rotary dial
High Reliability: use VLSI components and surface mount technology
Protection: over voltage protection, over current protection, output short circuit protection, reverse
voltage protection
Frequency Testing: optional frequency counter, capable of external frequency signal testing
Power Amplifier: optional part, maximum output power up to 7W
Remote Control: optional USB and RS232 interface
-5
order of magnitude
Check List Optional Parts
● DDS function generator
● Power cord
● Q9 test lead
● Q9 BNC-clip test lead
● Operation manual
● USB software CD (optional)
1
● RS232 software CD (optional)
1
● USB cable (optional)
1
● RS232 cable (optional)
1
1
1
1
1
1
● USB interface
● RS232 interface
● Frequency counter
● Power amplifier
I V
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 6
Contents
Safety Instruction ....................................................................................................................................................... II
Introduction ............................................................................................................................................................. IV
Chapter 1 Quick Starting ....................................................................................................................................... - 1 -
1.1 Preparation ............................................................................................................................................... - 1 -
1.2 Front and Rear Panel ............................................................................................................................... - 2 -
1.3 Description of LCD Display .................................................................................................................... - 3 -
1.4 Key Description ....................................................................................................................................... - 3 -
1.5 Fundamental Operation ........................................................................................................................... - 5 -
Chapter 2 Principle of the Function Generator ...................................................................................................... - 9 -
2.1 Principle Diagram .................................................................................................................................... - 9 -
2.2 Principle of Direct Digital Synthesis ....................................................................................................... - 9 -
2.3 Principle of Operation and Control ........................................................................................................ - 10 -
Chapter 3 Operation Instruction .......................................................................................................................... - 11 -
3.1 Operation Common Rules ..................................................................................................................... - 11 -
3.2 Setup of Channel A ................................................................................................................................ - 12 -
3.3 Setup of Channel B ................................................................................................................................ - 15 -
3.4 Frequency Sweep ................................................................................................................................... - 17 -
3.5 Amplitude Sweep ................................................................................................................................... - 18 -
3.6 Frequency Modulation (FM) ................................................................................................................. - 18 -
3.7 Amplitude Modulation (AM) ................................................................................................................. - 19 -
3.8 Burst Output .......................................................................................................................................... - 20 -
3.9 Frequency Shift Keying (FSK) .............................................................................................................. - 21 -
3.10 Amplitude Shift Keying (ASK) ........................................................................................................... - 22 -
3.11 Phase Shift Keying (PSK) .................................................................................................................... - 22 -
3.12 Measure External Frequency ............................................................................................................... - 23 -
3.13 Parameter Calibration .......................................................................................................................... - 24 -
3.14 System Setup ....................................................................................................................................... - 25 -
Chapter 4 Programmable Interface ...................................................................................................................... - 27 -
4.1 Interface Application ............................................................................................................................. - 27 -
4.2 Interface Selection ................................................................................................................................. - 27 -
4.3 USB Interface ........................................................................................................................................ - 27 -
4.4 RS232 Interface ..........................................................................................................
4
.5 Remote Command ................................................................................................................................. - 29 -
4.6 Application Software ............................................................................................................................. - 31 -
Chapter 5 Optional Parts ...................................................................................................................................... - 33 -
5.1 Programmable Interface ........................................................................................................................ - 33 -
5.2 Frequency Counter ................................................................................................................................ - 33 -
5.3 Power Amplifier .................................................................................................................................... - 33 -
Chapter 6 Technical Specifications ...................................................................................................................... - 35 -
6.1 Output Characteristics of Channel A ..................................................................................................... - 35 -
6.2 Output Characteristics of Channel B ..................................................................................................... - 36 -
6.3 TTL Output Characteristics ................................................................................................................... - 37 -
6.4 Common Characteristics ........................................................................................................................ - 37 -
6.5 Optional Parts Characteristics ................................................................................................................ - 38 -
........................... - 28 -
V
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 7
Chapter 1 Quick Starting
This chapter describes the front and rear panels of the DDS function generator. The brief introduction of the
function generator helps you get familiar with the fundamental operations and functions. The main contents in this
chapter are as the following:
1.1 Preparation
1.1.1 Check the Function Generator and Accessories
Check the function generator and accessory parts and verify all items in the good shape. If the box is broken,
please keep the box until the function generator passes function test.
1.1.2 Connect the Function Generator to Power
Only under the conditions below, you can connect power cable and turn on the power switch.
Voltage: AC220V (1±10%)
AC110V (1±10%) (Pay attention to the position of voltage selection switch)
Frequency: 50Hz (1±5%)
Watts: <45VA
Temperature: 0~40℃ Humidity: <80%
Plug the power cord into an 110V/220V power outlet with ground connection and turn on the function generator.
The function generator starts to initialize - display the instrument name, load the default parameters, display menu
of Channel A frequency, start channel A and B output signal. After initialization, the function generator is in the
normal working mode.
WARNING: For protection from electrical shock, the three-hole power cord with protective earth ground has to
be used.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
-1 -
Page 8
1.2 Front and Rear Panel
1. LCD display 2. Power switch 3. Keypad
4. Channel B output 5. Channel A output 6. Rotary knob
Front panel
Rear Panel
1. TTL output (BNC) 2. Modulation/External signal input (BNC)
3. Power connector with fuse 4. AC110V/220V power selection switch
5. RS232/USB connector
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
-2 -
Page 9
1.3 Description of LCD Display
On the top line of the LCD display, the function and operation option are displayed. In the “CHA Alone” and
“CHB Alone” mode, the current function or the output waveform is shown on the left side and the current option
is shown on the right side. Under each function, there are variety options shown in the table below. The options
with shadow are frequently-used which can be selected by using the shortcut function keys on the front panel.
However, the options without shadow are not common. So first select the corresponding function, then use the key
[Menu] to select them. At the bottom line of the LCD display, the parameters of the current option are displayed.
The table of function and option
Function key CHA Alone (sine) CHB Alone (sine)
CHA frequency CHA offset CHB frequency Start frequency
Peak-to-peak value CHA attenuation Peak-to-peak value Stop frequency
CHA period CHA duty cycle CHB period Sweep type
option
Root mean squared value
Phase
Root mean squared value Sweep mode
Phase Interval time
CHB harmonic wave Single burst
CHB waveform CHA frequency
Function key CHA Amplitude Sweep FM AM CHB Burst
Start amplitude Carrier frequency Carrier frequency CHB frequency
Stop amplitude Carrier amplitude Carrier amplitude CHB amplitude
Sweep type Modulation frequency Modulation frequency Burst count
option
Sweep mode FM deviation AM depth Burst frequency
Interval time Modulation wave Modulation wave
Single burst Signal source Signal source
CHA amplitude
Function key FSK ASK PSK
CHA Frequency
Sweep
Carrier frequency Carrier frequency Carrier frequency
Carrier amplitude Carrier amplitude Carrier amplitude
option
Hop frequency Hop amplitude Hop phase
Interval time Interval time Interval time
Function key Count Calibration System
CHA frequency Calibration switch Interface type
Gate time CHA frequency Remote address
option
Low pass filter Carrier frequency Language setup
Signal source FM deviation Beeper
Software version
1.4 Key Description
There are 26 keys on the front panel. The black word on each key represents the basic function. Simply press the
key to use the basic function. The green word at the upper of key represents the second function of the key. Press
-3 -
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 10
key [Shift] , the sign “↑ ” will be displayed at the lower-right of the LCD. Then press the function key, you can
select the second function. In order to select one of the eleven functions, sequentially press the function key and
key [Menu].
In this section, the basic functions of the 26 keys will be introduced. The selections of green second function and
black function will be introduced in the following chapters.
Numeric keys
Key name Main function Second function Key name Main function Second function
0 Input digit 0 Not available 7 Input digit 7 Not available
1 Input digit 1 Not available 8 Input digit 8 Not available
2 Input digit 2 Not available 9 Input digit 9 Not available
3 Input digit 3 Not available ● Input decimal point Not available
4 Input digit 4 Not available ▬ Input negative sign Not available
5 Input digit 5 Not available ◄ Flash digit left shift Add digit
6 Input digit 6 Not available ► Flash digit right shift Subtract digit
Function keys
Key name Main function Second function Unit function
FREQ/Period Frequency/period selection Sine selection Not available
AMPL/Offset Amplitude/offset selection Square selection Not available
Sweep Sweep selection CHB triangle selection Not available
MOD Modulation selection CHB ramp selection Not available
Menu Menu selection CHB waveform selection Not available
Burst CHB burst selection Duty cycle selection kHz/mVpp
Count Frequency counter selection Phase selection Hz/Vrms
Recall/store Recall/store parameters Harmonic wave selection mHz/mVrms
CAL/SYS Parameter calibration/system setup Reset selection s/%/Time
CHA/CHB CHA Alone/CHB Alone Attenuation selection ms/° /dB/N
Other keys
Key name Main function Others
Output On/Off Signal output on/off Single trigger for sweep and burst function
Shift
Key functions: 26 keys are available on front panel and a “di” sound will be heard when pressed.
Key [Menu] : press key [Menu] without numeric input, users can select all options under the current function.
Most keys have second functions that are marked in green above the keys. To use the second function, key [Shift]
should be pressed first before pressing the function key.
A few keys can also be used as unit keys that marked in black below the keys. To use the unit function, a digit key
should be pressed first before the unit key is pressed.
Key [Shift] : its basic function is for second function shifting. Press key [Shift] , the sign “↑ ” lamp turns on. At
this time, the second function of the function key is validated. Press key [Shift] again, the sign “↑ ” lamp turns off.
At this time, the basic function of the function key is validated. Key [Shift] is also used as unit MHz/Vpp and has
the function for exiting the remote control.
Shift to 2’nd function with other keys
and exit remote control
Unit MHz/Vpp
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
-4 -
Page 11
1.5 Fundamental Operation
This section introduces the fundamental operations on the general requirements through some examples. See
Chapter 3 in “Operation Manual” for more information if you have complex problems or need the complicated
operations.
1.5.1 Function of Channel A
Press key [CHA/CHB] and select “CHA Alone” function.
Set the frequency of the channel A: set frequency as 3.5 kHz.
Press [FREQ] and select “Frequency”, then press keys [3] [.] [5] and the soft key corresponding to [kHz] .
Adjust the frequency of the channel A: press key [] or [ ] to move the cursor left or right, and tune the rotary
knob left or right to add or subtract the digit on cursor, or to decrease or increase the digits continuously by steps,
so as to make coarse or fine adjustment of the frequency. Apply the same procedures for other parameters
adjustments.
Set the period of the channel A: set period at 2.5 ms.
Press key [Period] and select “Period”, then press keys [2] [.] [5] and the soft key corresponding to [ms] .
Set the amplitude of the channel A: set amplitude at 3.2 Vpp.
Press key [AMPL] and select “Amplitude”, then press keys [3] [.] [2] and the soft key corresponding to
[Vpp] .
Set the amplitude format of the channel A: Vrms or Vpp
Press key [AMPL] and select “Amplitude”, then press key [Vrms] or [Vpp].
Select the channel A waveform: select the sine or square wave.
Press keys [Shift] [ ] to select sine wave. Press keys [Shift] [ ] to select square wave.
Set the channel A duty cycle: set the duty cycle at 65%.
Press keys [Shift] [Duty] , then press keys [6] [5] and the soft key corresponding to [%] .
Set the channel A attenuation: select the fixed attenuation 0dB (after power on or reset, the instrument is default as
AUTO attenuation).
Press keys [Shift] [ATT] , then press key [1] and the soft key corresponding to [dB] .
Set the channel A offset: in the attenuation (0dB) mode, set DC offset as -1 Vdc
Press
key [Offset] and select “Offset”, then press keys [-] [1] and the soft key corresponding to [Vdc] .
Recall the stored parameters: recall the 15th stored parameters. (only for the keypad operations)
Press key [Recall] and select recall function, then press keys [1] [5] [No.] .
1.5.2 Function of Channel B
Press key [CHA/CHB] and select “CHB Alone” function.
Set the channel B frequency and amplitude: apply the same setting method as explained in section 1.5.1.
Select the channel B common waveform: select the sine, square, triangle and ramp wave.
Press keys [Shift] [
Press keys [Shift] [
Press keys [Shift] [ B-
Press keys [Shift] [ B-
Select the channel B other waveform: select one of the 32 kinds of waveforms.
Press keys [Shift] [B-Arb] and select “Channel B waveform”. Rotate the knob to select one of the 32 kinds
of waveforms. (For details, please refer to the table of the waveform index number and name in section
3.3.1)
Find Quality Products Online at: sales@GlobalTestSupply.com
] to select sine wave.
] to select square wave.
] to select triangle wave.
] to select ramp wave.
-5 -
www.GlobalTestSupply.com
Page 12
Set the Channel B harmonic wave: set the frequency of channel B as a 3-time harmonic wave of channel A.
Press keys [CHB] [Menu] [Harmonic] [3] [TIME] .
Set the phase of channel A and channel B: set the phase of channel A and channel B as 45° and 90°
respectively.
Press keys [CHA] [Shift] [Phase] [4] [5] [°] .
Press keys [CHB] [Shift] [Phase] [9] [0] [°] .
1.5.3 Frequency Sweep of the Channel A
Press key [Sweep] and select “CHA Sw Fr” function. The instrument will output the frequency sweep signal with
the default parameters at CHA output.
Set the sweep mode: set the Up-Down sweep mode.
Press key [Menu] and select “Sweep mode”, then press keys [2] [No.] .
The setup of other sweep parameters will be introduced in section 3.4.
1.5.4 Amplitude Sweep of the Channel A
Press key [Sweep] and select “CHA Sw Am”. The instrument will output the amplitude sweep signal with the
default parameters at CHA output.
Set the interval time: set the interval time as 0.5s.
Press key [Menu] and select “Interval time”, then press keys [0] [.] [5] [s] .
The setup of other sweep parameters will be introduced in section 3.5.
1.5.5 Frequency Modulation (FM) of the Channel A
Press key [MOD] and select “CHA FM”. The instrument will output the FM signal with the default parameters at
CHA output.
Set the FM deviation: set the FM deviation as 5%.
Press key [Menu] and select “FM Deviation”, then press keys [5] [%] .
The setup of other FM parameters will be introduced in section 3.6.
1.5.6 Amplitude Modulation (AM) of the Channel A
Press key [MOD] and select “CHA AM”. The instrument will output the AM signal with the default parameters at
CHA output.
Set the AM depth: set the AM depth at 50%
Press key [Menu] and select “AM Depth”, next press keys [5] [0] [%] .
The setup of other AM parameters will be introduced in section 3.7.
1.5.7 Burst Output of the Channel B
Press key [Burst] and select “CHB Burst”. The instrument will output the burst count signal with the default
parameters at CHB output.
Set the burst count: set the burst count as 5 cycles.
Press key [Menu] and select “Cycles”, next press keys
The setup of other burst parameters will be introduced in section 3.8.
[5] [CY
CL].
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
-6 -
Page 13
1.5.8 FSK Modulation of the Channel A
Press key [MOD] and select “CHA FSK”. The instrument will output the FSK signal with the default parameters
at CHA output.
Set the hop frequency: set the hop frequency at 1 kHz.
Press key [Menu] and select “Hop Freq”, next press keys [1] [kHz] .
The setup of other FSK parameters will be introduced in section 3.9.
1.5.9 ASK Modulation of the Channel A
Press key [MOD] and select “CHA ASK”. The instrument will output the ASK signal with the default parameters
at CHA output.
Set the carrier amplitude: set the carrier amplitude at 2 Vpp.
Press key [Menu] and select “Carrier Ampl”, next press keys [2] [Vpp] .
The setup of other ASK parameters will be introduced in section 3.10.
1.5.10 PSK Modulation of the Channel A
Press key [MOD] and select “CHA PSK”. The instrument will output the PSK signal with the default parameters
at CHA output.
Set the hop phase: set the hop phase at 180° .
Press key [Menu] and select “Hop Phase”, next press keys [1] [8] [0] [°] .
The setup of other PSK parameters will be introduced in section 3.11.
1.5.11 CHA/CHB Output On/Off
Press key [CHA] , channel A is shown on the LCD display, press [Output] , the output lamp turns to green, the
signal of channel A outputs. Press [Output] again, the output lamp turns off, the instrument closes the signal
output of channel A.
Press key [CHB] , channel B is shown on the LCD display, press [Output], the output lamp turns to red, the signal
of channel B outputs. Press [Output] again, the output lamp turns off, the instrument closes the signal output of
channel B.
After power on, the default state is not having signal output, with the lamp being off. When the output lamp turns
to orange, there are signals output both in channel A and channel B.
1.5.12 System Initialization Status
At power on or pressing keys [Shift] [Reset] , the system initialization status of the instrument is as the
followings:
Channel A:
Waveform: sine wave Frequency: 1.0000000kHz Amplitude: 2.000Vpp
Attenuation: AUTO Period: 1ms Offset: 0.000 Vdc
Pulse duty cycle: 50.0% Square duty cycle: 50.0% Time interval: 10ms
Carrier frequency: 50kHz Start frequency: 500.00000Hz Stop frequency: 50.000000kHz
Modulation frequency: 1.0000000kHz Start amplitude: 10mVpp Stop
AM depth: 100% Carrier amplitude: 2Vpp Sweep mode: Up Sweep
Hop frequency: 5.0000000kHz FM Deviation: 10% Burst count: 3 CYCL
-7 -
amplitude: 2 .000Vpp
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 14
Hop phase: 180° Hop amplitude: 1.000Vpp Burst frequency: 100.0000000Hz
Phase: 0°
Channel B:
Waveform: sine wave Frequency: 1.0000000kHz Amplitude: 2.000Vpp
A harmonic: 1.00 TIME Period: 1ms Phase: 0°
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
-8 -
Page 15
Chapter 2 Principle of the Function Generator
In order to help you better understand the internal operation of the instrument, this chapter describes the basic
signal generator concepts and provides the specific details on the internal operations of the function generator.
2.1 Principle Diagram
2.2 Principle of Direct Digital Synthesis
In order to generate a voltage signal, the traditional analog signal source generally uses the discrete electronic
components to build oscillation circuitries in variety ways. The accuracy of frequency is not only low but also
very unstable. The production and testing procedures are very complicated and time-consuming. Moreover, it is
not convenient for the frequency setting and computer programming. The Direct Digital Synthesis (DDS)
technology is a new signal-generation technique. DDS does not consist of any oscillating components. It applies a
stream of digital data which represents the desired waveform pre-stored in memory to the input of
digital-to-analog converter (DAC). The DAC converts the data and outputs a series of voltage steps approximating
the desired waveform.
For example, to synthesize a sine waveform, first, the function y = sin(x) is digitized, then the digitized value y is
sequentially stored in the waveform memory by the address of x. DDS uses the phase accumulating technique to
control the address of the waveform memory. In every single sampling period, a phase increment is added to the
current content of the phase accumulator. The output frequency of DDS can be modified by changing this phase
increment. The digitized waveform data is read from the waveform memory addressed by the phase accumulator
-9 -
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 16
and converted into an analog voltage through D/A converter and amplifier circuitry. Because the waveform data
are the interval samples, the output of DDS is a stepwise sine wave. Thus, the output waveform from DDS has to
be filtered out the high harmonic components by a low-pass filter. The high accurate reference regulator in D/A
converter assures the amplitude precision and stability of the output waveform.
The amplitude controller is a D/A converter. According to the preset amplitude value, this converter generates a
corresponding analog voltage which multiplies with the output signal and produces a final output signal whose
amplitude is equal to the preset value. The offset controller is also a D/A converter. According to the preset offset
value, this converter generates a corresponding analog voltage which adds with the output signal and produces a
final output signal whose offset is equal to the preset value. The output signal is output at the channel A.
2.3 Principle of Operation and Control
The MCU controls keypad and LCD display through the interface circuitry. When the key is pressed, the MCU
detects the action, finds the code of key, and executes the corresponding program. The LCD display circuit adopts
menu characters to display the working status and parameters of the instrument.
The knob on the front panel can be used to adjust the numeric digit on the cursor position. Every 15-degree
rotation will generate a trigger pulse. Based on the pulse, the MCU can determine the knob left-rotation or
right-rotation. If the rotation is left, the digit on the cursor is subtracted by one. If the rotation is right, the digit is
added by one.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
-10 -
Page 17
Chapter 3 Operation Instruction
In this chapter, you will learn more information about the functions and operations.
3.1 Operation Common Rules
3.1.1 Numeric Keypad Input
If an option is selected, you can input the parameters by using the numeric keys. There are ten numeric keys for
data input. The input method is the shift input from left to right. The data can have only one decimal point. If one
data input has more than one decimal point, only the first one is valid. In the offset mode, the negative sign can be
input too. After input a new value, press unit key to validate it. If the input data has error, there are two ways to
correct it.
Method 1: If the destination side of output signal can receive the wrong signal, press any unit key to terminate the
current operation. Input the correct data, and press the unit key to validate the input.
Method 2: If the destination side of output signal can not receive the wrong signal, the wrong input is not
validated because of no any wrong signal at the output. Reselect the operation, input correct data, and press the
unit key to validate the input.
Although there are a variety of combinations of the decimal point and the unit key for data input, the instrument
always displays the input data in a fixed format. For example, input either 1.5kHz or 1500Hz, after validation, the
instrument will display as 1.5000000kHz.
Different physics value has different unit, such as “Hz” for frequency, “V” for amplitude, “s” for time, and “° ”
for phase. However, when using numeric keys to enter data, the input data will not be effective without the entry
of unit. The instrument will display the corresponding unit when the physics value is selected.
3.1.2 Step Key Input
In practices, a set of frequency or amplitude value with same interval is commonly used. It is complicated and
time consuming to input this kind of data by repeatedly pressing the numeric keypad and unit key. It is very
convenient to use the step input method. Every pressing the corresponding soft key can make the frequency or
amplitude increase or decrease by a step value. The modified data is validated automatically without pressing the
unit key.
For example, to generate a series of frequency with interval 12.5 kHz, press the keys sequentially as following:
Press key [Menu] and select “Step Freq”;
Press keys [1] [2] [.] [5] and the soft key corresponding to [kHz] ;
Press key [CHA] and select “CHA Frequency”;
Press keys [Shift] [ ] to increase the frequency by 12.5 kHz or press keys [Shift] [] to decrease the
frequency by 12.5 kHz. Repeat this operation, a series of waveform with equal interval frequency difference can
be generated. The same procedure can be used for the amplitude operation. This method is only used on the
frequency and amplitude of channel A.
3.1.3 Knob Adjustment
In some applications, it requires to adjust the output signal continuously. Using the rotary knob can complete this
-11 -
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 18
task. On the top of the displayed parameter, there is a triangle-shape cursor. Press key [ ] or [ ] to move the
cursor left or right. Rotate the knob to the right to continuously increase the cursor-located digit by 1 and make the
carry to a higher unit position. Rotate the knob to the left to continuously decrease the cursor-located digit by 1
and make the carry to a lower unit position. When you use the knob to modify a specified data, the modified data
is validated instantaneously without pressing the unit key. A coarse adjustment is made by moving the cursor to
the left; and a fine adjustment is made by moving the cursor to the right.
3.1.4 Input Mode Selection
It is convenient to input the known data through the numeric keypad. It makes the input be set at only one step
without the intermediate transition data. For the partial modification of the input data or at the situation of
monitoring the variation process of the input data, the rotary knob is usually more useful. For a series of equal
interval data input, the step input method is the most efficient. Users can use the different data input methods
according to different situations.
3.2 Setup of Channel A
Press key [CHA/CHB] and select “CHA Alone” function. The upper-left area of LCD displays the waveform
name of channel A.
3.2.1 Set the Frequency of Channel A
Press key [FREQ] to display the current frequency. Input a new frequency data here by numeric keypad or rotary
dial. Press the key corresponding to the frequency units to confirm the new data input. The function generator
outputs the specified frequency at CHA Output.
3.2.2 Set the Period of Channel A
The frequency of the channel A can also be set and displayed in the way of period. Press keys [Period] to display
the current period. Input a new period data here by numeric keypad or rotary dial. Press the key corresponding to
the period time units to confirm the new data input.
However, the instrument still uses the frequency synthesis internally and the input data is just converted to the
correct value when it is input and displayed. Due to the limitation of the lower resolution at low frequency (40μ
Hz), when the period is too long, the function generator can only generate the frequency points with the longer
period interval. Although the preset and displayed period is precise, the actual period of the output signal has the
relatively big variation. Users shall pay attention to this point.
3.2.3 Set the Amplitude of Channel A
Press key [AMPL] and select amplitude of channel A. Input a new amplitude data here by numeric keypad or
rotary dial. The function generator outputs the amplitude signal at the CHA Output.
3.2.4 Amplitude Format
The input and display of the Channel A amplitude value have two formats: Vpp and Vrms. Press keys [ Shift]
[Vpp] after numeric input to input and display amplitude peak-to-peak value. Press keys [Shift] [Vrms] after
numeric input to input and display amplitude true RMS value. The displayed amplitude value varies with the
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
-12 -
Page 19
format conversion simultaneously.
Although the amplitude value has two formats, the instrument still uses the peak-to-peak format internally and
makes the format conversion on the data input and the data display. Due to the limitation of the amplitude
resolution, there is a difference between the format conversions. For example, if the Vpp of sine wave is 1Vpp, the
Vrms conversion value is 0.353 Vrms. If the Vrms of sine wave is 0.353Vrms, the Vpp conversion value is
0.998Vpp. In most of cases, this difference is within the error range. If the square waveform is selected, the
conversion factor is 2. The Vrms format can only be used at the “CHA Alone” mode and the waveform selected as
sine or square wave. For other operations and waveforms, only peak-to-peak format can be used.
3.2.5 Amplitude Attenuation
Press keys [Shift ] [ATT ] to select the attenuation of channel A. After instrument power on or reset, the attenuation
is default in “AUTO” mode. In “AUTO” mode, the instrument automatically selects suitable attenuation ratio
according to preset amplitude values. When the output amplitude is at 2V, 0.2V, or 0.02V, the instrument switches
between attenuation ratios. In this situation, the instrument generates a waveform of high amplitude resolution,
high signal & noise ratio, and low distortion, regardless of signal amplitude values. However, during attenuation
ratio switch, there is instant hopping in the output signal, which is not allowed in some practices. That is why the
instrument has preset with fixed attenuation mode.
Press keys [Shift ] [ATT ]. An attenuation data can be input here by numeric keypad or rotary dial. There are five
ranges of attenuation: 0dB, 20dB, 40dB, 60dB and AUTO, which can be input by numeric key 1, 2, 3, 4 and 0
respectively plus the soft key corresponding to [dB] . The attenuation mode can also be selected by rotary dial.
One step in rotary dial tuning goes to one range. If a fixed attenuation mode is selected, the attenuation is fixed
and does not change according to amplitude value changes, which make continuous variation of the output signal
in full amplitude range. But in range of 0dB, if the signal amplitude value is too small, there will be higher
waveform distortion and the signal & noise ratio can not be so good.
3.2.6 Output Load
The default amplitude value is calibrated when the output is open circuit. The real voltage on the output load is the
default amplitude multiplying the dividing ratio of the load impedance and the output impedance. The output
impedance of the instrument is about 50Ω. When the load impedance is big enough and the voltage dividing ratio
is close to 1, the loss on the output impedance can be ignored. The voltage on the load is close to default
amplitude value. When the load impedance is too small, the loss on the output impedance is significant and can
not be ignored. The voltage on the output load is not same as the default amplitude value. Please pay attention to
this point.
The output of channel A has the over voltage and over current protections. The instrument can not be damaged if
the output is short-circuit for a few minutes or the transient inverse voltage is less than 30V. However, this kind of
operation should avoid in case of potential damage to the instrument.
3.2.7 Amplitude Flatness
If the output frequency is lower than 1MHz, the amplitude and frequency characteristics of the output signal are
very flat. If the output frequency is over 10MHz, due to the characteristics of the output amplitude and load
matching, the amplitude and frequency characteristics of the output signal will not be so good. The maximum
-13 -
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 20
output amplitude is also affected. The higher frequency is, the higher amplitude and the more distortion will be in
output waveform.
3.2.8 Set the Offset of Channel A
Press key [Offset] and select “CHA Offset”. The current offset value is shown on the LCD display. Input a new
offset data here by numeric keypad or rotary dial. Press the key corresponding to the offset units to confirm the
new data input. The function generator outputs the desired signal at the CHA output.
Please note that the sum of half of the signal output amplitude value and the offset value should be less than 10V,
so as to make sure the offset signal peak value is less than ±10V. Otherwise, there will be amplitude distortion.
Besides, when Channel A is at AUTO attenuation, the output offset attenuates as per amplitude attenuation. When
amplitude Vpp value is above about 2V, the actual output offset is the preset offset value. When amplitude Vpp
value is approximately between 0.2V~2V, the actual output offset is 10% of the preset offset value. When the
amplitude Vpp value is approximately less than 0.2V, the actual output offset is 1% of the preset offset value.
When adjusting DC offset value of the signal, the rotary knob is more frequently used than numeric keypad. As a
common rule, tune the rotary knob to the right to increase DC offset level, while to the left to decrease the DC
offset value, regardless of the DC offset value as positive or negative. When the rotary knob tune across zero, the
offset value changes automatically between positive and negative signs.
3.2.9 DC Voltage Output
When the attenuation of channel A is set at 0dB, the output offset is the preset offset value, regardless of
amplitude values. If the amplitude value is set at 0V, the offset value can be set at any value between -10V and
+10V. In this situation, the instrument becomes a DC power source, which can output preset DC voltage signal.
3.2.10 Select the Waveform of Channel A
The channel A can output three kinds of waveforms: sine wave, square wave, and pulse wave. Press shortcut
function key on the front panel to select the corresponding waveform. The selected waveform will be output at
CHA output.
Press keys [Shift] [
Press keys [Shift] [
Press keys [Shift ] [Duty] to select pulse wave.
3.2.11 Set the Duty Cycle of Channel A
Press keys [Shift] [Duty] to select pulse wave. The pulse duty cycle is shown on the LCD display. A new duty
cycle data can be input here by numeric keypad or rotary dial. The duty cycle value can be set at any value
between 0.1%~99.9%. When the converted duty cycle value is less than 0.1%, the duty cycle value is displayed as
0.1%. When the converted duty cycle value is greater than 99.9%, the duty cycle value is displayed as 99.9%.
] to select sine wave.
] to select square wave. The duty cycle of square wave is default to 50%.
3.2.12 Frequency and Amplitude Step Input
In “CHA Alone” mode, it is very convenient to use the step input method to enter a set of frequency or amplitude
value with same interval. For details, please refer to section 3.1.2. This step input method is only used on the
frequency and amplitude of channel A.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
-14 -
Page 21
3.2.13 Set the Phase of Channel A
Press keys [Shift] [Phase] to select “Phase” option. The phase value is shown on the LCD display. Input a new
phase data here by numeric keypad or rotary dial. The phase can be adjusted at any value between 0~360 degree.
The phase resolution is 11.25 degree.
3.2.14 Set the Output Impedance of Channel A
Press key [Menu] and select “Impedance” option. The impedance value is shown on the LCD display. The
function generator is default as high impedance. The impedance value can be set by numeric keypad or rotary dial
and the output impedance is 50Ω.
3.3 Setup of Channel B
Press key [CHA/CHB] and select “CHB Alone” function. The upper-left area of LCD displays the channel B
waveform name or index number.
3.3.1 Select the Waveform of Channel B
There are 32 kinds of waveforms in channel B. Press keys [Shift ] [B Wave] to select “CHB Wave”, the current
waveform name and index number are shown on the bottom of the LCD display. Enter the index number by using
the numeric keypad or rotary knob, press the soft key corresponding to [No.] to confirm the index number input.
Users can use the shortcut function key on the front panel to select the 4 frequently-used waveforms.
Press keys [Shift] [
Press keys [Shift] [
Press keys [Shift] [ B-
Press keys [Shift] [ B-
The function generator outputs the selected waveform at the CHB Output. For 4 frequently-used waveforms, they
are displayed as waveform name in the upper-left portion of the LCD. As to the other 28 waveforms which are not
frequently used, they are displayed as “Arb” in the upper-left portion of the LCD. The index number and name of
the 32-waveform are shown as the table below.
] to select sine wave.
] to select square wave.
] to select triangle wave.
] to select ramp wave.
The Index Number and Waveform Name of Channel B
Index Number Name Index Number Name
00 Sine 16 Exponent
01 Square 17 Logarithm
02 Triang 18 Half round
03 Up ramp 19 Tangent
04 Down ramp 20 Sin(x)/x
05 Pos-pulse 21 Noise
06 Neg-pulse 22 Duty 10%
07 Tri-pulse 23 Duty 90%
08 Up stair 24 Down stair
09 Pos-DC 25 Po-bipulse
10 Neg-DC 26 Ne-bipulse
11 All sine 27 Trapezia
12 Half sine 28 Coscine
13 Limit sine 29 Bidir-SCR
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
-15 -
Page 22
14 Gate sine 30 Cardiogram
15 Square-root 31 Earthquake
3.3.2 Set the frequency of Channel B
Press key [FREQ] and select “CHB Frequency”. Input a new frequency data here by numeric keypad or rotary
dial. There will be corresponding frequency output in CHB output terminal. The displayed units can be MHz, kHz,
Hz, mHz. The resolution is 40 mHz. Same as channel A frequency, channel B frequency can also be set and
displayed in the way of period.
3.3.3 Set the Amplitude of Channel B
Press key [AMPL] and select “CHB Amplitude”. Input a new amplitude data here by numeric keypad or rotary
dial. Press the soft key corresponding to the amplitude units to confirm the new data input. The function generator
outputs the desired signal at CHB Output. The amplitude unit of Channel B can only use Vpp but not Vrms. The
channel B does not have the amplitude attenuation or DC offset functions.
3.3.4 Set the Harmonic Wave
The frequency of channel B can be set and displayed as multiple times of channel A frequency. That is, channel B
signal becomes the multiple N times harmonic of channel A signal.
Press keys [Shift] [Harmonic] to select “CHB Harmonic”. Input harmonic times via numeric keypad or rotary
dial. Press the soft key corresponding to [TIME] to confirm the new data input. Channel B frequency becomes the
set multiple times of channel A frequency. That means, channel B signal becomes the multiple N times harmonic
of channel A signal. At this time, phase of channel A and B signals has stable synchronization. Press keys [Shift]
[Phase] to select “CHB Phase”. Input phase value via numeric keypad or rotary dial. In this mode, the phase
difference of the CHA and CHB signals can be set. If there is a difference between the output phase difference and
the preset phase difference, please adjust the channel B phase. Connect the channel A and B signals to an
oscilloscope to change the phase difference, so that all kinds of stable Lissajou’s figures are generated. If channel
B is not selected, channel A and B will not have synchronization. Although channel B frequency is set at a
multiple time of channel A frequency, channel A and channel B may not able to reach synchronization. Therefore,
to make sure of stable phase synchronization of channel A and B, user must first set up channel A frequency, next
select “B harmonic” and set harmonic time so that channel B frequency can change automatically.
3.3.5 Set the Phase of Channel B
Press keys [Shift] [Phase] to select “Phase” option. The phase value is shown on the LCD display. Input a new
phase data here by numeric keypad or rotary dial. The phase can be adjusted at any value between 0~360 degree.
The phase resolution is 1 degree.
3.3.6 Sum of Channel A and Channel B Signals
After power on the instrument, the signals of channel A and channel B are completely independent. Signals from
the two channels can be summed together.
Press key [Menu] and select “A B ADD” mode. The add switch “On” or “Off” can be set by using the numeric
keypad or rotary dial. The output signal on CHA output becomes the linear sum of channel A and channel B
signal. In this situation, frequency and amplitude of channel A and channel B are still in independent mode and
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
-16 -
Page 23
can be set independently. This function is very useful in the applications of filter experiment and waveform
analysis. To return to independent channels, press add switch “Off” by using the numeric keypad or rotary dial.
The sum of signals of channel A and channel B can only apply in the mode of CHB Alone. In other working mode,
channel A and channel B signals are completely independent.
3.4 Frequency Sweep
Press key [Sweep] and select “CHA Sw Fr” function. “CHA Sw Fr” is shown on the upper-left of LCD. The
function generator outputs the frequency sweep signal at the CHA Output. The sweep mode of output frequency is
step sweep. The output frequency automatically increases or decreases by a step value in a certain interval time.
Start frequency, stop frequency, sweep time, sweep mode and trigger mode are free to be set by the users.
3.4.1 Set the Start and Stop Frequency
The frequency at sweep starting is called start frequency. The frequency at sweep ending is called stop frequency.
Press key [Menu] and select “Start Freq” option. Input a new start frequency data here by numeric keypad or
rotary dial. Press the soft key corresponding to the frequency units to confirm the new data input. To set the stop
frequency, press key [Menu] and select “Stop Freq” option. Input a new stop frequency data here by numeric
keypad or rotary dial.
Note: the stop frequency must be greater than the start frequency. Otherwise the sweep can not operate.
3.4.2 Set the Sweep Direction
The frequency sweep has three directions represented by three index numbers: 0, 1 and 2.
Up Sweep (0 Up Sweep): the output signal sweeps from the start frequency to the stop frequency at the preset
sweep rate.
Down Sweep (1 Down Swe): the output signal sweeps from the stop frequency to the start frequency at the
preset sweep rate.
Up-Down Sweep (2 UpDownS): the output signal sweeps from the start frequency to the stop frequency at the
preset sweep rate, then sweeps from the stop frequency to the start frequency at the same sweep rate.
Press key [Menu] and select “Sweep Direction” option. The instrument displays the index number and name of
the sweep mode. User can input the index number or turn the knob to set the sweep mode.
3.4.3 Select the Sweep Mode
Press key [Menu] to select linear sweep or logarithmic sweep.
Linear sweep: the output signal changes linearly during the duration of sweep.
Logarithmic sweep: the output signal changes logarithmically during the duration of sweep.
3.4.4 Set the Sweep Time
After the start frequency and the stop frequency are set, the sweep time can be determined according to frequency
step. The shorter sweep time is, the bigger frequency step will be. The longer sweep time is, the smaller frequency
step will be. The minimum frequency update time is the software execution time.
Press key [Menu] and select “Sweep Time ” option. Input a new sweep time here by numeric keypad or rotary dial.
Press the soft key corresponding to the time units to confirm the new data input.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
-17 -
Page 24
3.4.5 Single Sweep
After the “CHA Sw Fr” mode is selected, the frequency sweep starts. During the sweep, user can change the
sweep parameter and sweep mode at any time. Press key [Menu] and select “Single Sweep” option, the sweep
process stops at once. The output signal will keep at the stop state without any changes. The current frequency
value is shown on the LCD display. After the sweep process stops, press key [Trigger] once, the sweep process
goes on by one step. The frequency of Channel A will sweep according to the desired sweep mode.
3.4.6 AUTO Sweep
Press key [Menu] and select “CHA Frequency”. Frequency of channel A changes simultaneously according to the
sweep process. Users can have dynamic monitor of the sweep process.
3.5 Amplitude Sweep
Press key [Sweep] and select “CHA Sw Ampl” function. “CHA Sw Ampl” is shown on the upper-left of LCD.
The function generator outputs the amplitude sweep signal at the CHA Output. All operations of sweep parameter,
sweep mode, single sweep, and sweep monitor apply the same operation procedures as explained in section
Channel A frequency sweep.
3.6 Frequency Modulation (FM)
Press key [MOD] and select “CHA FM” option. “CHA FM” is shown on the upper-left of LCD. The function
generator outputs the frequency modulation signal on the CHA Output.
3.6.1 Set the Carrier Frequency
Press key [Menu] and select “Carrier Freq” option. A new carrier frequency data can be input here by numeric
keypad or rotary dial. Press the soft key corresponding to the frequency units to confirm the new data input. In the
frequency modulation mode, the signal of Channel A is used as the carrier signal and the carrier frequency is the
frequency of Channel A. In this mode, the clock signals of DDS synthesizer are switched from the fixed reference
clock to the programmable reference clock, the accuracy and stability of the carrier frequency may decrease.
3.6.2 Set the Modulation Frequency
Press key [Menu] and select “Mod Freq” option. A new modulation frequency data can be input here by numeric
keypad or rotary dial. Press the soft key corresponding to the frequency units to confirm the new data input. In the
frequency modulation mode, the signal of Channel B is the modulation signal and the modulation frequency is the
frequency of Channel B. Generally, the carrier frequency should be at least 10 times higher than the modulation
frequency.
3.6.3 Set the Frequency Modulation Deviation
Press key [Menu] and select “FM Deviation” option. A new frequency modulation deviation data can be input
here by numeric keypad or rotary dial. Press the soft key corresponding to [%] to confirm the new data input.
Deviation shows the variations of the carrier signal frequency during the frequency modulation. It is more
intuitional to show the variation in the way of carrier waveform period variation. Variation in the frequency of
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
-18 -
Page 25
modulated waveform is shown as the equation below:
DEVI% = 100xSHIFT/PERD
DEVI is the modulation frequency deviation;
SHIFT is the peak variation;
PERD is the period of the carrier signal at the zero frequency deviation.
In the modulation demonstration, to have clear observation of the frequency variation, a bigger frequency
deviation should be set. In practices, to reduce the bandwidth of the carrier signal, the frequency deviation is
usually less than 5%.
3.6.4 Set the Modulation Waveform
Since the signal of channel B as the modulation signal, the Channel B waveform is actually the modulated
waveform. Press key [Menu] and select “Mod Wave” option. The index number and name of waveform are
displayed on the LCD. Input the correct index number by numeric keypad or rotary dial to select corresponding
waveform. Press the key corresponding to [No.] to confirm the index number input. The function generator
outputs the desired modulated signal.
3.6.5 External Modulation Source
The frequency modulation can use the external modulation signal. There is a “Modulation In” connector on the
rear panel to import external modulation signal. The external modulation signal frequency should be compatible to
the frequency of carrier signal. The amplitude of external modulation signal should be adjusted according to the
frequency deviation. The bigger the amplitude of external modulation signal is, the bigger the frequency deviation
will be. In the external modulation mode, the frequency deviation should be set to 0 and turn off the internal
modulation signal. Otherwise, it will affect the external modulation process. Similarly, in the internal mode, the
frequency deviation should be set, and the external signal connection should be disconnected. Otherwise, the
internal modulation will be affected.
3.7 Amplitude Modulation (AM)
Press key [MOD] and select “CHA AM” option. “CHA AM” is shown on the upper-left of LCD. The function
generator outputs the amplitude modulation signal on the CHA Output. The operation of modulation parameter
setup is same as the frequency modulation.
3.7.1 Set the Carrier Frequency
Press key [Menu] and select “Carrier Freq” option. Input a new carrier frequency data here by numeric keypad or
rotary dial. Press the soft key corresponding to the frequency units to confirm the new data input. In the amplitude
modulation mode, the signal of Channel A is the carrier signal and the carrier frequency is the frequency of
Channel A.
3.7.2 Set the Modulation Frequency
Press key [Menu] and select “Mod Freq” option. Input a new modulation frequency data here by numeric keypad
or rotary dial. Press the soft key corresponding to the frequency units to confirm the new data input. In the
amplitude modulation mode, the signal of Channel B is the modulation signal and the modulation frequency is the
frequency of Channel B. Generally, the carrier frequency is at least 10 times higher than the modulation
frequency.
-19 -
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 26
3.7.3 Set the Amplitude Modulation Depth
Press [Menu] and select “AM Depth” option. Input a new modulation depth data here by numeric keypad or
rotary dial. Press the soft key corresponding to [%] to confirm the new data input. In the amplitude modulation
mode, the amplitude of the carrier signal varies periodically with the modulation signal. The modulation depth
represents the variation in the carrier signal.
For example, the 100% modulation depth represents that the maximum amplitude of modulation signal is 100% of
the set value, the minimum amplitude of modulation signal is 0% of the set value, i.e. 100% - 0% = 100%. The
0% modulation depth represents that both maximum and minimum amplitude of modulation signal are 50% of the
set value, i.e. 50% - 50% = 0%. Similarly, the 120% depth is 110% - (-10%) = 120%. The modulation depth has
another expression: If the maximum modulation peak-to-peak value is A, the minimum peak-to-peak value is B,
then the modulation depth is expressed as:
DEPTH% = 100× (A-B)/ (A+B)
Actually these two kinds of expressions are the same. However, the second one is easier to understand.
In the amplitude modulation demonstration, in order to clearly observe the amplitude variation, the modulation
depth is set to a relative big value even greater than 100%. But in the practices, in order to avoid the distortion of
the modulation signal, the modulation depth is generally set to less than 50%. This kind of modulation carrier is
called the double-side carrier. Most of the middle-wavelength radio stations use this modulation.
3.7.4 Set the Modulation Waveform
Since the signal of channel B as the modulation signal, the Channel B waveform is actually the modulated
waveform. Press key [Menu] to select “Mod Wave” option. Enter the correct index number by numeric keypad or
rotary dial to select a waveform. Press the key corresponding to [No.] to confirm the index number input. The
function generator outputs the desired modulated signal.
3.7.5 External Modulation
The amplitude modulation can use the external modulation signal. There is a “Modulation In” connector on the
rear panel to import external signal. The frequency of external modulation signal should be compatible to the
carrier frequency. The amplitude of external modulation signal should be adjusted according to amplitude
modulation depth. The bigger amplitude of external modulation signal is, the bigger amplitude modulation depth
will be. In the external modulation mode, the amplitude modulation depth should be set to 0 and turn off the
internal modulation signal. Otherwise, it will affect the external modulation process. Similarly, in the internal
mode, the amplitude modulation depth should be set, and the external signal connection should be disconnected.
Otherwise, the internal modulation will be affected.
3.8 Burst Output
Press key [Burst] and select “CHB Burst” option. “CHB Burst” is displayed in the upper-left portion of the LCD.
The function generator outputs a waveform with a specified number of cycles at the burst frequency continuously
at the CHB Output. Each set has a preset burst count and there is a time interval between each burst.
3.8.1 Set the Frequency of Channel B
Since channel B signal is the burst source for burst output, frequency and amplitude of channel B signal should
firstly be set up. The settings of the frequency and amplitude are introduced in the section 3.3.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
-20 -
Page 27
3.8.2 Set the Burst Count
Press [Menu] and select “Cycles” option. Input a new burst count number here by numeric keypad or rotary dial.
Press the key corresponding to [CYCL] to confirm the new data input. If the burst frequency is fixed, the
maximum value of the burst count is limited. The lower burst frequency is, or we say the longer burst period is,
the bigger the burst count number can be set. Vise visa, the smaller burst count number is. If the burst frequency is
not set, the burst count can be set firstly and then the burst frequency value is adjusted in order to keep an
appropriate time interval between each burst.
3.8.3 Set the Burst Frequency
Press [Menu] and select “Burst Freq” option. Input a new burst frequency data here by numeric keypad or rotary
dial. Press the soft key corresponding to the frequency units to confirm the new data input. The burst frequency is
determined by the Channel B frequency value and the burst count number. The multiple of channel B signal
period value and burst count number is the time needed for a set of burst. In order to keep an appropriate time
interval between each burst, the burst period should be greater than the time of a burst. Otherwise all bursts are
connected to each other and the waveform is not the burst signal anymore.
3.8.4 Burst Start Phase
The start phase of each set of burst waveforms is fixed to 0° .
3.8.5 Set the Single Burst Mode
Press [Menu] and select “Single Burst” option. The function generator turns off the continuous burst mode and
signal output is 0. In the single burst mode, press key [Trigger] once, the burst process executes once and the
function generator outputs one burst waveform according to the preset value of the burst count. If the burst count
number is set to 1, the single burst can be operated manually. If the “Single Burst” option is not selected, the burst
process is back to the continuous burst mode. Burst count can be used to test the dynamic characteristics of the
audio equipment and calibrate the counter.
3.9 Frequency Shift Keying (FSK)
In the digital communication and remote measurement systems, the transmission of the digital signal uses the
Frequency Shift Keying (FSK) or Phase Shift Keying (PSK) method to encode the frequency or phase of the
carrier signal. At the receiving station, the received signal is decoded and recovered to the original digital signal.
Press key [MOD] and select “FSK” option. The upper-left portion of the LCD displays “A_FSK”. The function
generator outputs the FSK signal on the CHA Output. The frequency of the output signal is the alternative
variation of the carrier frequency and the hop frequency. The interval time of alternation is adjustable.
3.9.1 Set the Carrier Frequency
Press [Menu] and select “Carrier Freq” option. Input a new carrier frequency here by numeric keypad or rotary
dial. Press the key corresponding to the frequency units to confirm the new data input. In the FSK mode, the
signal of the channel A is the carrier signal. The carrier frequency is the first frequency value of channel A signal.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
-21 -
Page 28
3.9.2 Set the Hop Frequency
Press [Menu] and select “Hop Freq” option. Input a new hop frequency here by numeric keypad or rotary dial.
Press the key corresponding to the frequency units to confirm the new data input. The hop frequency is the second
frequency value of channel A signal.
3.9.3 Set the Interval Time
Press [Menu] and select “Interval” option. Input a new interval time data here by numeric keypad or rotary dial.
Press the key corresponding to the time units to confirm the new data input.
3.10 Amplitude Shift Keying (ASK)
Press key [MOD] and select “ASK” option. The upper-left portion of the LCD displays “A_ASK”. The function
generator outputs the ASK signal on the CHA Output. The amplitude of output signal is the alternative variation
of the carrier amplitude and the hop amplitude. The interval time of alternation is adjustable.
3.10.1 Set the Carrier Amplitude
Press [Menu] and select “Carrier Ampl” option. Input a new carrier amplitude data here by numeric keypad or
rotary dial. Press the key corresponding to the amplitude units to confirm the new data input. In the ASK mode,
the channel A signal is the carrier signal. The carrier amplitude is the first amplitude value of channel A signal.
3.10.2 Set the Hop Amplitude
Press [Menu] and select “Hop Ampl” option. Input a new hop amplitude data here by numeric keypad or rotary
dial. Press the key corresponding to the amplitude units to confirm the new data input. The hop amplitude is the
second amplitude value of channel A signal. There may be a big difference between carrier amplitude and hop
amplitude. In the ASK process, the generator attenuates according to channel A fixed attenuation, so as to avoid
frequent relay switch during auto attenuation.
3.10.3 Set the Interval Time
Press [Menu] and select “Interval” option. Input a new interval time data here by numeric keypad or rotary dial.
Press the key corresponding to the time units to confirm the new data input.
3.11 Phase Shift Keying (PSK)
Press key [MOD] and select “PSK” function. The upper-left portion of the LCD displays “A_PSK”. The function
generator outputs the PSK signal on the CHA Output. The phase of the output signal is the alternative variation of
the reference phase and the hop phase. The interval time of alternation is adjustable.
3.11.1 Set the Hop Phase
Press [Menu] and select “Hop Phase” option. Input a new hop phase data here by numeric keypad or rotary dial.
Press the key that corresponds to [°] to confirm the new data input. The resolution of the hop phase is 11.25
degree. If users enter the phase value by using the numeric keypad, the instrument will round the input value to
the number of times of 11.25 degree.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
-22 -
Page 29
3.11.2 Set the Interval Time
Press [Menu] and select “Interval ” option. The interval time value is shown on the LCD display. Input a new
interval time here by numeric keypad or rotary dial. Press the key corresponding to the time units to confirm the
new data input.
3.11.3 PSK Observation
Due to the phase variation caused by PSK signal, it is hard to have synchronization on the oscilloscope, and
therefore cannot observe a stable waveform. If the frequency of the channel B is set at the same value of carrier
frequency, and use channel B signals as synchronization burst signals, a stable PSK signal waveform can be
observed.
3.12 Measure External Frequency
This function is available only when the user chooses to install the frequency counter. Press key [Count] to go to
the external measurement mode. The LCD display shows the external measurement interface. In this mode, the
instrument can be used as a frequency counter and measures the frequency value of the external signal.
3.12.1 Frequency and Count Measurement
Press [Menu] and select “Measure Freq” option. Connect the external signal to the “Count In” terminal on the
rear panel. The instrument displays the frequency of measured external signal. The measured external signal can
be periodical signal of the arbitrary waveform. The signal amplitude peak-to-peak value should be
100mVpp~20Vpp.
3.12.2 Set the Gate Time
Press [Menu] and select “Gate Time” option. Input a new gate time data here by numeric keypad or rotary dial.
Press the key corresponding to the time units to confirm the new data input. During frequency measurement, the
tested signal must be continuous. However, the measurement is intermittent. With the preset gate time as a cycle,
the instrument makes sampling of the tested signal, analyzes the testing data and refreshes display. The instrument
usually adopts average periodical measurement. The longer gate time is, the more cycle numbers to be collected
from the testing signal, the more stable the average value will be, the more valid digits to get in testing results, but
the slower tracking to the frequency variation will be, which is suitable to frequency testing requiring a long time
of stability. The shorter gate time is, the less valid digits to get in testing results, the faster tracking to the
frequency variation will be, which is suitable to frequency testing requiring a short time of stability.
3.12.3 Low-pass Filter
In the measurement of the external signal, if the signal frequency is relatively low and has high frequency noise,
the measured data has bigger error and is not stable. Press [Menu] and select “LPF” option. Use the numeric
keypad or rotary dial to change the display to “1 ON” and add a 100 kHz low-pass filter to the input signal. The
low-pass filter filters out the high-frequency noise and makes the measured result much more accurate. If the
frequency of the testing signal is relatively high, the low-pass filter may attenuate the signal, decrease the
sensitivity of the measurement, and even result in the wrong testing data. In this situation, use the numeric keypad
or rotary dial to change the display from “1 ON” to “0 OFF” and remove the 100 kHz low-pass filter. For the low
frequency of square signal, because of steep edge of the signal, the trigger error is not significant. Therefore, the
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
-23 -
Page 30
low pass filter is not necessary here.
3.12.4 Self Check and Demonstration
The instrument is capable of using internal waveform to carry out self check and present demonstration. Use a test
cable to connect the CHA terminal with the “Count In” terminal on the rear panel. Press [Menu] and select
“Measure Freq”. The test result displayed is the current channel A frequency value. Since channel A frequency
synthesizer uses the same clock with frequency tester, the test result does not have the affect of crystal error.
Therefore the test result is highly accurate. However, in the situation of the external frequency measurement, due
to the affect of crystal error, the accuracy and stability may not be as good as those in self checks.
3.13 Parameter Calibration
The parameter errors listed in the technical specification are the factory calibrated specifications. Long-time use or
drastic changes in ambient temperature may enlarge the parameter errors. When the instrument is used as a precise
testing instrument, a calibration should be carried out periodically. Press key [CAL/SYS] and select “User CAL”
option. This function uses the calibration software to calibrate the three parameters through the keypad.
3.13.1 Calibration Code
Press key [Menu] and select “CAL Off” option. On the bottom-left of the LCD, “0 Off” is shown and indicates to
input the calibration code, then enter a calibration code 666666 by numeric keypad and press the soft key
corresponding to [No.] . The “0 Off” changes to “1 On”, which means the instrument is ready for parameter
calibration following the ways below.
3.13.2 Calibrate the Frequency of Channel A
Press key [Menu] and select “CHA Freq” option. The calibration code of channel A frequency is displayed. At
this time, the frequency of channel A is 2MHz. Use a frequency counter of at least 6 digits to test the frequency of
channel A. In the mean time, rotate the knob to change the calibration code to change channel A frequency until
channel A frequency has an accuracy of 10
-6
. Since channel A frequency and Channel B frequency use the same
fixed reference clock, after channel A frequency is calibrated, channel B applies the same accuracy.
3.13.3 Calibrate the Carrier Frequency
In the frequency modulation mode, the carrier frequency uses a programmable clock as its reference clock.
Although the carrier frequency is the frequency of the channel A, it still needs to be calibrated. Press key [Menu]
and select “Carrier Freq” option. The calibration code of carrier frequency is displayed. At this time, the
frequency of channel A is 2MHz. Use a frequency counter to measure the frequency of channel A and adjust the
calibration code by the rotary knob until channel A frequency has an accuracy of 10
-3
.
3.13.4 Calibrate the Frequency Modulation Deviation
Press key [Menu] and select “FM Deviation” option. The calibration code of FM deviation value is displayed. At
this time, the frequency of channel A is 2MHz. Use a modulation meter to measure the FM deviation of channel A
and adjust the calibration code by the rotary knob until the accuracy is calibrated to 10
-3
.
After calibration, press key [Menu] and select “CAL Over” option to exit the calibration mode. The calibration
codes are stored into the nonvolatile memory of the instrument, and will not be lost after power off. The
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
-24 -
Page 31
calibration codes will be refreshed after a second calibration is carried out.
3.14 System Setup
3.14.1 Interface Setup
Press key [SYS] and select “Interface Type” option. The interface type is displayed. Users can select the
USB/RS232 interface by numeric keypad or rotary dial.
3.14.2 Remote Address
Press keys [SYS] [Menu] to select “Remote Address” option. The remote address is displayed and can be
modified by numeric keypad or rotary dial.
3.14.3 Language Setup
Press keys [SYS] [Menu] to select “Language Setup” option. The language is displayed. Users can select Chinese
(simplified) /Chinese (traditional)/English by numeric keypad or rotary dial.
3.14.4 Beeper
Press keys [SYS] [Menu] to select “Beep Setup” option. The Beep On/Off is displayed. Users can select Beep
On/Beep Off by numeric keypad or rotary dial.
3.14.5 Store and Recall Parameters
In some applications, it is required to frequently use a set of certain parameters, such as frequency, amplitude,
offset, waveform, etc. It is very easy and convenient to store and recall these frequently used parameter settings.
The instrument provides 40 index numbers (0-39) for the parameter memories, starting from index number 0 to 39.
Memories of index number 1~39 are available for user defined setting storage, except memory of number 0.
Memory index number 0 is for the storage of default parameters setup. Users also can input recall index number 0
to 39 to recall the stored settings in each memory. Input number 0 recalls the default parameter setting of the
instrument. After the parameters are recalled, the instrument outputs the signal based on the recalled parameters.
After input a set of parameters, press key [Store] and select “Store” option, input the memory index number by
numeric keypad, press the corresponding soft key of [No.] to store this set of parameter setting into the
corresponding memory. Apply the same storage procedure to store up to 40 sets of parameter settings. The
parameter memory uses a non-volatile memory, which can store the settings even the instrument is power off.
If you need to use the stored parameter later, press key [Recall] and select “Recall” option, input the recall index
number and press the soft key corresponding to [No.] . The stored parameter setting is recalled. If press key
[Recall] and select “Recall” option, then press key [0] and the soft key corresponding to [No.] , the default
parameter settings is recalled. It can get the same result to press keys [Shift] [Reset] .
3.14.6 Software Version
Press keys [SYS] [Menu] to select “Software version” option. The software version is displayed.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
-25 -
Page 32
Chapter 4 Programmable Interface
This chapter introduces the usage of the programmable interface. Through the study of this chapter, you will learn
how to setup the testing system and use the testing software to program the instrument.
4.1 Interface Application
Nowadays, the application of computer is prevalent. The traditional test equipment is gradually replaced by the
digital test devices. Most of the manual testing tasks have been replaced by the automatic testing systems based on
the computer control technology, which is an inevitable trend in testing & measuring field. Today almost all
high-end and middle-end test equipments have the communication interfaces. With the interfaces, the test
equipments can be connected to the computer with the communication cables to make a testing system. During the
test, all the devices or equipments in the system can communicate with each other through the interfaces and
cables. Based on the preprogrammed test software, the remote computer can control various devices and make
them work coordinately. For example, first, the remote computer sends a command to the function generator to
request it to output an appropriate signal. Secondly, the computer sends commands to the frequency meter and
voltage meter to measure the frequency and voltage. After collecting all measurement data, the computer
processes and analysis the data, and finally prints out the test report. In this example, all complicated tasks are
completed by the automatic testing system. What the testing engineer needs to do is to write a testing program. A
good testing system can not only save a lot of labor and improve the efficiency, but also provide the reliable
results and reduce the risk of the mistakes and faults caused by the testing engineer. In some situations, the
automatic testing system can even complete the test tasks which cannot be done by the manual test ways.
4.2 Interface Selection
The instrument has two kinds of interfaces.
4.2.1 USB Interface
The USB interface has the advantages of the high transmission rate, high reliability, easy-to-use, and plug-in play.
It is a commonplace on all current computers and devices.
4.2.2 RS232 Interface
The RS232 interface is the serial asynchronous communication interface. It has the characteristics of the long
transmission distance and less number of transmission wire. Most of the current computer has the RS232 interface.
Recently, RS232 interface is gradually being replaced by USB interface because of its low transmission rate and
inconvenient connection.
4.3 USB Interface
The USB protocol used in this instrument meets the standard of USB V1.1.
4.3.1 Interface Pin Assignment
Power Line: +5V and GND;
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
-27 -
Page 33
Signal Line: D+ and D-, three-state bidirectional.
4.3.2 USB Interface Connection
Connect the USB connector on the front panel to the USB connector on the computer with a USB cable.
4.3.3 Operation Instruction
The detailed information is on the CD-ROM shipped with the box.
4.3.4 Environment
The USB adapts the short distance data transmission in the common lab and production environment.
4.4 RS232 Interface
The RS232 used in the instrument meets the standard of EIA-RS232.
4.4.1 Interface Voltage Level
Logic “0”: +5 V to +15 V;
Logic “1”: -5 V to -15 V.
4.4.2 Transmission Format
Each transmission frame consists of eleven bits: 1 start bit (logic “0”), 8 data bits (ASCII code), 1 index bit
(address logic = “1”, data logic = “0”), 1 stop bit (logic “1”).
4.4.3 Transmission Rate
In this instrument, the transmission rate is set to 19200 bits/s and can be reset in the software.
4.4.4 RS232 Interface Connection
Connect the RS232 connector on the rear panel to the COM1 or COM2 connector on the computer with a RS232
cable.
4.4.5 System Construction
In one system, the maximum number of connected instruments is 99 and the total length of connection cable
cannot exceed 100 meters.
4.4.6 Applications
The RS232 adapts the data transmission in the common lab and production environment without heavy electrical
interference.
4.4.7 Address Information
In the programmable mode, the instrument starts to receive the message from computer, checks the index bit, and
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
-28 -
Page 34
determines the received message is either address message or data message. If it is an address message, check
whether the received address matches its own address or not, if not, ignore all following messages and wait for the
next address message. If the address matches, the instrument starts to receive the following data message until the
whole message is received.
4.4.8 Data Information
After receiving the data message, the instrument verifies and stores the data into the memory. If the received
character is the line feed Chr (10), it means that this is the end of the data message. The instrument starts to
execute each operation in this remote command.
4.5 Remote Command
The remote command is a series of ASCII code strings sent by the host computer. The operation of the instrument
is based on the remote command. Every instrument has its own defined command format. In order to control the
instrument successfully, users have to follow the rules when writing the application program.
1 Remote command; 2 Remote command (without digit and unit); 3 Inquiry command
CHA:AFREQ:1.31:MHz CHA:SQUAR CHA:?AFREQ
↑ ↑ ↑ ↑ ↑ ↑ ↑ ↑ ↑ ↑ ↑ ↑ ↑
1 5 2 5 3 5 4 1 5 2 1 6 2
1: Function string 2: Option string 3: Data string 4: Unit string 5: Separate character 6: Question mark
The remote command consists of the function string, the option string, the data string, the unit string, the separate
character, and the question mark. Some commands do not include the data string or the unit string. (Please refer to
the software in the CD ROM attached to the function generator.)
4.5.1 Command String
The instrument remote commands consist of the function string, the option string, the data string, and the unit
string shown in the tables below. Both function and option strings use the uppercase characters. Every definition
is corresponding to the instrument functionality and option. The data string consists of 10 digits from 0 to 9,
decimal point, and negative sign. The unit string consists of the defined uppercase and lowercase characters
depending on the data characteristics. Except the characters in the table, the other characters can not be used.
Otherwise, error occurs.
The Table of the Function Command
Option String Option String
Frequency of channel A CHA Amplitude modulation of channel A AM
Frequency of channel B CHB Burst of channel B BURST
Frequency sweep of channel A FSWP FSK of channel A FSK
Amplitude sweep of channel A ASWP ASK of channel A ASK
Frequency modulation of channel A FM PSK of channel A PSK
External frequency measurement COUNT System SYS
Return to local LOCAL RS232 control RS232
USB control USB
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
-29 -
Page 35
The Table of the Option Command
Option String Option String Option String
CHA frequency AFREQ Interval time INTVL Hop phase HOPP
CHA period APERD Start amplitude STARA Single trigger TRIGG
CHA amplitude AAMPL Stop amplitude STOPA Sine waveform SINE
CHA offset AOFFS Burst count NCYCL Square waveform SQUAR
CHA attenuation AATTE Burst frequency BURSF CHB triangle waveform BTRIG
CHA duty cycle ADUTY Single burst ONCES CHB ramp waveform BRAMP
CHB frequency BFREQ Carrier frequency CARRF Store parameter STORE
CHB amplitude BAMPL Carrier amplitude CARRA Recall parameter RECAL
CHB waveform BWAVE FM deviation DEVIA System reset RESET
CHB harmonic wave BHARM AM depth DEPTH Phase PHASE
Peak-to-peak value VPP Modulation frequency MODUF Signal source SOURCE
Root mean squared value VRMS Modulation waveform MWAVE Beeper BEEP
Start frequency STARF Hop frequency HOPF CHB period BPERD
Stop frequency STOPF Hop amplitude HOPA Gate time STROBE
Sweep type MODEL Low pass filter LPF Measurement frequency MEASF
Sweep mode TYPE Language setup LANG Interface address ADDR
Output switch SWITCH Software version VER
The Table of the Data Command
Data String Data String
Digit 0 1 2 3 4 5 6 7 8 9 Decimal Point ·
Negative Sign -
The Table of the Unit Command
Unit String Unit String
Frequency MHz, kHz, Hz, mHz Phase DEG
Peak-to-peak Value Vpp, mVpp Attenuation dB
Time s, ms Index No.
Harmonic wave TIME Percentage %
Burst count CYCL DC offset Vdc, mVdc
RMS Vrms, mVrms
4.5.2 Character Return
The length of a remote command cannot exceed 60 characters. Every string is ended with a Chr (10), meaning an
end to a data string.
4.5.3 Separate Character
The separate character has to be placed between the function and option strings, between the option and data
strings, between the data and unit strings, and between the function string and question mark.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
-30 -
Page 36
4.5.4 Question Mark
Adding the question mark (?) before the option string will be return to the corresponding value and unit, if no unit,
the instrument will only return to the value.
4.5.5 Data String
The maximum length of the data string is 10 characters.
4.5.6 Control Capability
The remote commands can control all instrument functionalities.
4.5.7 Serial Port Control
In the serial port control mode, users should firstly send the instrument's address and select the corresponding
instrument, next send the remote commands. If the USB interface is used, no need to send the address message.
Select the corresponding device on the computer directly.
4.5.8 Online Operation
First, send the interface select command. For example, send “address + RS232” command to select the RS232
interface. The instrument will switch to the RS232 remote mode. If the address of the instrument is 88, send
“88RS232”. To exit this remote control mode, send “88LOCAL” or press key [Shift] to make the instrument back
to the normal key operation mode. Otherwise, neither the key operation nor the other remote controls can be used.
4.6 Application Software
In the automatic testing system, the application software is a control program which is written by the testing
engineers for the accurate control on the variety of instruments and devices. Users have to read the details of this
guide before starting to write the application program. It is quite easy to program as long as you are familiar with
the manual operations and transfer each manual operation to the program of each remote command gradually. The
application program can be written by a variety of computer languages.
4.6.1 Enter the Program Mode
After power on, the instrument is in the manual operation mode. When the remote command is received, the
instrument enters the remote operation mode. At this time, all the key functions are not available. The instrument
can only be operated by the remote command. To return to the manual operation mode, send the “Address +
LOCAL” (for example “88LOCAL”) command from the computer to tell the instrument back to the manual
operation mode. Or, press key [Shift] to return to manual operation.
4.6.2 Tips for Programming
Users have to pay attention to the program main points below when use the remote command to program.
1: Strictly follow the remote command format, including the uppercase and lowercase characters in the command
code. Otherwise, the instrument will consider the command as a wrong message and ignore the operation.
2: Users should be familiar with the manual operation firstly and make sure the remote command sequence
exactly same as the manual operation. The right sequence of the command should be the function string, the
-31 -
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 37
option string, the data string, and finally the unit string. If the program sequence is not correct or the string
does not match the corresponding functionality, although the instrument executes the code and does not give
any error notice, the result may be wrong. For example, set the frequency of the channel A at 100 kHz, remote
command: 100:AFREQ:kHz:100. This is a string in wrong sequence. Another example, select the No.3
waveform of channel B, remote command: CHA:BWAVE:3No., the command and function are not matching.
4.6.3 Application Example
The application examples of the variety type of the remote command are shown as followings. (No blanks in the
command)
Example1: Channel A, single output, frequency at 1MHz
RS232 mode: 88CHA:AFREQ:1:MHz
USB mode: CHA:AFREQ:1:MHz
Example 2: Channel B, single output, frequency at 1 kHz
RS232 mode: 88CHB:BFREQ:1:kHz
USB mode: CHB:BFREQ:1:kHz
Example 3: Channel A, single output, pulse wave, duty cycle at 25%.
RS232 mode: 88CHA:ADUTY:25%
USB mode: CHA:ADUTY:25%
Example 4: Frequency sweep output, start frequency at 1 kHz.
RS232 mode: 88FSWP:STARF:1:kHz
USB mode: FSWP:STARF:1:kHz
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
-32 -
Page 38
Chapter 5 Optional Parts
This chapter introduces the optional parts of the function generator which are not included as standard.
5.1 Programmable Interface
If users choose to purchase the USB or RS232 interface, the corresponding parts will be installed in the instrument.
The detailed instruction has been described in Chapter 4.
5.2 Frequency Counter
If users choose to purchase the frequency counter, the frequency count board will be installed in the instrument.
The input connector is the “Count In” connector on the rear panel. The detailed instruction has been described in
section 3.12.
5.3 Power Amplifier
If users choose to purchase the power amplifier, the power amplifier board will be installed in the instrument. This
is an independent part of the instrument, with input terminal “power amplifier input” and output terminal “2 time
power amplifier output” located on the rear panel of the instrument. When using the power amplifier, use a
connecting cable to input the signal to the input terminal “power amplifier input”. There will be a
2-time-amplified signal in the output terminal “2 time power amplifier output”. The input signal can be the
channel A signal, or channel B signal, or external signal.
5.3.1 Input Voltage
The power amplifier can amplify the input voltage by 2 times. The maximum output amplitude is 22Vpp.
Therefore, the maximum input amplitude should not exceed 11Vpp. Input amplitude exceeds this limit will cause
distortion.
5.3.2 Frequency Range
The frequency range of power amplifier is 10Hz~150kHz. Amplitude flatness within this frequency range is better
than 3%, and sine distortion is better than 1%. The maximum testing frequency is up to 200kHz.
5.3.3 Output Power
The equation for output power of power amplifier is:
2
P=V
/R
P is the output power, with the unit of W
V is the output amplitude RMS value, with the unit of Vrms
R is the load resistance, with the unit of Ω
The maximum output amplitude is up to 22Vpp (7.8Vrms). The minimum load resistance is 2Ω. As the ambient
temperature increases, the output signal frequency increases too. The lower output signal distortion is required, the
smaller the max.output power is. Generally, the maximum output power can be up to 7W (8Ω) or 1W (50Ω).
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
-33 -
Page 39
5.3.4 Output Protections
The power amplifier has short circuit protection and over heat protection. It may not be burnt out in common
operation. However, to prevent damage to power amplifier’s performance, long time short circuit output should be
avoided. Please also avoid to frequently using the maximum value of the frequency, amplitude and load.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
-34 -
Page 40
Chapter 6 Technical Specifications
6.1 Output Characteristics of Channel A
6.1.1 Waveform Characteristics
Waveform Type: sine, square, pulse, DC
Wave Length: 4 ~ 16000 points
Waveform Amplitude Resolution: 10 bits
Sample Rate: 180MSa/s
Sinusoidal Harmonic Rejection: ≥50dBc (≤1MHz), ≥ 40dBc (1MHz ~ 20MHz), ≥30dBc (20MHz ~ 40MHz)
Sine Wave Total Distortion: ≤0.5 % (20Hz ~ 200kHz)
Pulse Wave and Square Wave: rise or fall time: ≤20ns, overshoot: ≤5%
Square Wave Duty Cycle: 50%
6.1.2 Frequency Characteristics
Frequency Range of Sine Wave: 2kHz ~ the maximum frequency, resolution: 40 mHz
40µHz ~ 2kHz, resolution: 40μHz
Frequency Range of Square Wave: 40µHz~20MHz
Frequency Range of Pulse Wave: 40µHz~10MHz
Frequency Accuracy: ±(5×10
Frequency Stability: ±5×10
-5
+ 40mHz)
-6
/ 3 hours
6.1.3 Pulse Characteristics
Duty Cycle: 0.1%~ 99.9%
6.1.4 Amplitude Characteristics
Amplitude Range: 2mVpp ~ 20Vpp (high impedance)
Resolution: 20mVpp (amplitude > 2V), 2mVpp (amplitude < 2V)
Amplitude Accuracy: ±(1%+2 mVrms) (high impedance, RMS, frequency 1 kHz)
Amplitude Stability: ±0.5%/ 3 hours
Amplitude Flatness: ±5% (frequency<1MHz), ±10% (frequency between 1MHz~10MHz)
±20% (frequency between 10MHz~60MHz)
Output Impedance: 50Ω
Sine Wave Amplitude Setting Range (50Ω)
1mVpp ~ 10Vpp, when output frequency ≤10MHz
1mVpp ~ 5Vpp, when output frequency ≤40MHz
1mVpp ~ 2Vpp, when output frequency ≥ 40MHz
Amplitude Setting Range (high impedance)
2mVpp ~ 20Vpp, when output frequency ≤10MHz
2mVpp ~ 10Vpp, when output frequency ≤40MHz
2mVpp ~ 4Vpp, when output frequency ≥40MHz
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
-35 -
Page 41
6.1.5 Offset Characteristics
Offset Range: ±10 V (high impedance)
Resolution: 20mV
Offset Accuracy: ±(1%+ 20mV)
6.1.6 Sweep Characteristics
Sweep Type: frequency sweep, amplitude sweep
Sweep Range: free to set the start and stop points
Sweep Time: 100ms ~ 600s
Sweep Direction: Up, Down, Up-down
Sweep Mode: linear, logarithmic
Control Mode: auto sweep or manual sweep
6.1.7 Frequency Modulation Characteristics
Modulation signal: Internal or external waveforms
FM Deviation: 0%~20%
6.1.8 Amplitude Modulation Characteristics
Modulation signal: Internal or external waveforms
AM Depth: 0%~120%
6.1.9 Shift Keying Characteristics
FSK: Free to set the hop frequency and the carrier frequency
ASK: Free to set the hop amplitude and the carrier amplitude
PSK: hop phase: 0 ~ 360 degree, Resolution: 11.25 degree
Alternate Rate: 10ms ~ 60s
6.2 Output Characteristics of Channel B
6.2.1 Waveform Characteristics
Wave Type: 32 kinds of waveforms, like sine, square, triangle, sawtooth, ladder etc.
Wave Length: 1024 points
Waveform Amplitude Resolution: 8 bits
Sample Rate: 12.5MSa/s
6.2.2 Frequency Characteristics
Frequency Range: Sine: 10mHz ~ 1MHz, other waveforms: 10mHz ~ 100kHz
Resolution: 10 mHz
Frequency Accuracy: ±(1×10
Find Quality Products Online at: sales@GlobalTestSupply.com
-5
+10 mHz)
-36 -
www.GlobalTestSupply.com
Page 42
6.2.3 Amplitude Characteristics
Amplitude Range: 50mVpp ~ 20Vpp (high impedance)
Resolution: 20mVpp
Output Impedance: 50Ω
6.2.4 Harmonic Characteristics
Channel B frequency is the harmonic wave of channel A.
Harmonic Time: 0.1 ~ 250.0 times, Harmonic Frequency <1MHz
Phase Adjustment: coarse adjustment: 11.25 degree/step, fine adjustment: 2 degree/step
6.2.5 Burst Characteristics
Channel B signal is used as burst signal
Frequency of Channel B: 40mHz ~ 1MHz
Burst Frequency: 10 mHz ~ 50kHz
Burst Count: 1 ~ 65000 cycles
Burst Mode: continuous burst and single burst
6.3 TTL Output Characteristics
Waveform Characteristics: square wave, rise or fall time ≤20ns
Frequency Features: same as sine wave of channel A
Amplitude Characteristics: TTL, CMOS compatible, low level <0.3V, high level >4V
6.4 Common Characteristics
6.4.1 Power Requirement
Power Requirements: AC220V (1±10%)
AC110V (1±10%) (Pay attention to the position of voltage selection switch)
Frequency: 50Hz(1±5%)
Power Consumption: <45VA
6.4.2 Environmental Condition
Temperature: 0 ~ 40℃ Humidity: <80%
6.4.3 Operation Characteristics
Key operation for all functions, menu display, rotary dial adjustment
6.4.4 Display Characteristics
Display: LCD, English, Chinese (simplified), Chinese (traditional)
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
-37 -
Page 43
6.4.5 Dimension and Weight
Chassis Dimensions: 415(D)×295(W)×195(H) mm Weight: 3.5 kg
6.4.6 Manufacturing
Manufacturing Process: Surface mount technology (SMT), large-scale integrated circuits, high reliability and long
service life.
6.5 Optional Parts Characteristics
6.5.1 Remote Interface
USB Universal Serial Bus Interface
RS232 serial interface
6.5.2 Frequency Counter
Frequency Range: 1Hz ~ 200MHz
Input Signal Amplitude: 100mVpp ~ 20Vpp
6.5.3 Power Amplifier
Max.output Power: 7W (8Ω), 1W (50Ω)
Max.output Voltage: 22Vpp
Frequency Bandwidth: 1Hz~200kHz