Calibrating your pipette has never been so simple
Your dream is coming
true
Calibrate pipette without liquid
Result in less than 30 seconds
No need for a precision balance
ISO 8655 accuracy compliant
Leak measurement
Smart and simple to use
Touch screen
Specific user interface
No training needed
A device you can rely on
High precision sensor
Excellent reproducibility
Accurate
Based on cutting-edge technologies, this affordable
calibrator is simple to use “there is
Designed with a t
functions such as the pipette volume etc. Special attention
was placed on the user interface to s
life while keeping the highest standard in calibration.
Everything is done for you including the statistic, and
provides you with an immediate information.
Patented new measurement principle
ATMOS uses the l
highest precision sensors.
table and reproducible conditions for the pipette
S
calibrations.
Special attention was placed on the design of the pipette
port, and mechanical device.
ouch screen which helps you selecting
atest generation microcontrollers and the
implify your day-to-day
.
CHF 1420.-
(excl. VAT)
Distributed in Switzerland by:
Life Systems Design AG
Pilatusstrasse 41
CH-5630 Muri
Switzerland

QUICK REFERENCE GUIDE
Note : This document is not intended to provide
all safety instruction and complete information.
Please refer to the complete ATMOS operation
manual available at www.unipix-lab.com
1
2
3
Switch ON the instrument
Select pipette size
1000 µL
200 µL
100 µL
20 µL
Select volume
1000 µL
500 µL
200 µL
Calibration
1. Switch on the calibrator
2. Select the pipette size
3. Select the testing volume
Insert Tip
5
4. Insert Tip in the opening and press tightly
5. Press G
6. Press the piston of the pipette
7. Release the piston
8. For calibration: repeat steps 5 to 7 at least 3 times
9. Pressing E
o
nd and the computation will be done automatically
4
The system computes the
9
results, display them on the
screen and provides a direct
feedback
Results
Average =
1000.43
SEM ±
Error %
0.24
0.04
Pipette OK
Restart
w w w . un i p i x – l a b . c o m
Results
µL
µL
Average =
SEM ±
Error %
986.78
0.34
-1.32
Pipette needs revision
Restart
µL
µL