Ultrasource Ultravac 500 User Manual

Ultravac® 500
Vacuum Chamber Packaging Machine
EFFECTIVE SERIAL # 2080 +
860052 Revision Q
Date 9/2011
MAIN: 816-753-2150
INTERNET: kochequipment.com
E-MAIL: info@kochequipment.com
TABLE OF CONTENTS
General ...............................................III
Specifications ....................................... IV
SAFETY
Personal Safety ....................................1.1
Food Safety ........................................1.2
General Safety Guidelines ....................1.2
STARTUP
Unpacking .......................................... 2.1
Power Requirements .............................2.1
Grounding Instructions .........................2.1
Vacuum Pump .................................... 2.2
Gas Flush Connection ........................ 2.2
Air-Assist Connection .......................... 2.2
Checking Vacuum Pump Rotation ........ 2.2
OPERATION
Placement of Product ...........................3.1
Operation with Analog Control Panel ... 3.2
Operation with Digital Control Panel .... 3.3
Operator Menu on Digital Panel .......... 3.4
Selecting a New Program .................... 3.5
Sealing with Air-Assist ......................... 3.6
Gas Flush Option ............................... 3.6
Double Seam Seal Option .................. 3.6
MAINTENANCE
Prior to Cleaning ................................. 4.1
Cleaning Recommendations .................4.1
Vacuum Pump Maintenance ................ 4.2
Seal Bar Maintenance ......................... 4.2
Reading the Indicators ........................ 4.3
Troubleshooting ................................. 4.3
Supervisor Menu on Digital Panel ........ 4.5
Opening Chamber Machine ................ 4.8
Changing Vacuum Pump Oil ................4.9
Replacing Potentiometer .......................4.9
Replacing Power Cord .......................4.10
Maintenance Log ..............................4.10
Service Log ....................................... 4.10
SCHEM ATICS
Designation and Function of Controls ....5.1
230 Volt, Single Phase, Analog Panel ... 5.2
230 Volt, 3 Phase, Analog Panel .......... 5.3
380 Volt, 50 Hz, 3 Phase, Analog Panel
.. 5.4
230 Volt, Single Phase, Digital Panel .... 5.5
230 Volt, 3 Phase, Digital Panel ........... 5.6
380 Volt, 50 Hz, 3 Phase, Digital Panel
... 5.7
Non-Gas Flush Pneumatic Diagram ..... 5.8
Gas Flush Pneumatic Diagram ............. 5.9
PARTS
Recommended Spare Parts ...................6 .1
Pump and Chassis .............................. 6.2
Seal Bar ............................................ 6.4
Lid and Hinge System ......................... 6.6
Gas Flush Components....................... 6.8
Main Control - Front Side ...................6.10
Analog Control Panel ........................6 .12
Digital Control Panel .........................6.14
Main Control - Back Side ...................6 .15
Mebner Valve .................................... 6.16
Miscellaneous Machine Parts .............. 6 .17
REFERENCE MANUALS
R5 Series Vacuum Pumps .....................7.1
INDEX .............................................. 8.1
860052
•Q
INTRODUCTION •
PAGE I
MAIN: 816-753-2150
INTERNET: kochequipment.com
E-MAIL: info@kochequipment.com
Ultravac® 500
Vacuum Chamber Packaging Machine
Congratulations on your Ultravac® 500 vacuum packaging machine purchase. This machine was designed to provide years of trouble-free operation and to help in the packaging of your quality food products.
Please read this owner’s manual to gain the maximum benefits of your vacuum packaging machine and its different components.
A note about cleaning: Given all the various ways equipment is used in different environments, we recommend the owner consult sanitation experts on how to properly clean each piece of machinery in their plant and to do bacterial testing to insure that the equipment is cleaned properly.
For Sales, Call:
Phone (816) 753-2150 • Fax (816) 753-4976
Toll-Free (800) 777-5624
For Replacement Parts, Call:
Phone (816) 753-2150 • Fax (816) 561-2854
Toll-Free (800) 746-1723
For Technical Support, Call:
Phone (816) 753-2150 • Fax (816) 531-1447
Toll-Free (800) 777-5624
860052
•Q
©Copyright 2009 Koch Equipment LLC
INTRODUCTION •
PAGE II
MAIN: 816-753-2150
INTERNET: kochequipment.com
E-MAIL: info@kochequipment.com
General
This owner’s manual contains information pertinent to your Ultravac® 500. Basic instructions and maintenance information is provided. Please read carefully. Failure to do so could result in bodily injury and/or damage to the equipment.
Receiving Problems: As in all cases, before signing the bill of lading, be sure all items have been received as listed and there is no damage in shipment. If needed, a claim must be made immediately to the local truck line office and noted on the bill of lading.
Please fill in the information from the bill of lading and the product identification tag.
Model No.
Serial No.
Ship Date:
Owner:
Location:
Electrical service size for your Ultravac® 500 (check one):
230 Volt, 3 phase, 60 Hz
230 Volt, Single phase, 60 Hz
380 Volt, 3 phase, 50 Hz
Please fill in the serial number from the pump identification tags:
Serial No.
860052
•Q
INTRODUCTION •
P A G E III
Specifications
MAIN: 816-753-2150
INTERNET: kochequipment.com
E-MAIL: info@kochequipment.com
Figure 0.1
Length (A): 718mm (28.25-in.) Width (B): 610mm (24-in.) Maximum Height (C): 1334mm (52.5-in.) Working Height (D): 892mm (35.125-in.) Seal Bar Length: 457mm (18-in.) Chamber Length (E): 648mm (25.5-in.) Chamber Width (F): 489mm (19.25-in.) Chamber Height (G): 229mm (9-in.) Between Seal Bars (H): 537mm (21.125-in.) Vacuum Pump: 48m3/h (28cfm) 1.5kW (2-hp)
70m3/h (41cfm) 2.2kW (3-hp) Net Weight: 160-kg, 353-lbs. Electrical Connection: See Power Requirements in Section 2 Capacity: 20-30 seconds per cycle
860052
•Q
INTRODUCTION •
PAGE IV
MAIN: 816-753-2150
DANGER
DANGER
WARNING
WARNING
CAUTION
CAUTION
CAUTION
INTERNET: kochequipment.com
E-MAIL: info@kochequipment.com
SAFETY
Personal Safety
The procedures and guidelines herein must be followed precisely to avoid problems that can result in property damage, personal injury, or death. If you have any questions related to
this information, please contact Koch Equipment Services Inc. at (800) 777-5624.
Hazardous voltage. Disconnect and lockout power before servicing machine or cleaning. Do not remove panels unless power has been disconnected, locked out, and tagged out at risk of electric shock hazard.
Hazardous voltage. Improper connection of the equipment-grounding conductor can result in a risk of electric shock.
Read and understand owner’s manual before using this machine. Failure to follow operating instructions could result in personal injury or damage to equipment.
Explosion hazard. Do not use a gas with an oxygen content greater than 22% with gas flush option.
Blade hazard. Do not remove, install, or replace blades without protective gloves. Perforating knife blade is sharp. Use care when handling.
Blade hazard. Do not remove, install, or replace blades without protective gloves. Precut knife blade is sharp. Use care when handling.
Cleaning agents. Do not get the cleaning agents in eyes, on skin, or on clothing. Always wear rubber gloves, goggles, and protective clothing when contact is likely. Consult product manufacturer for specific details.
Signal words used in classification of potential hazards are defined as follows:
DANGER: Indicates an imminently hazardous situation, which, if not avoided, may result in death or serious injury. WARNING: Indicates a potentially hazardous situation, which, if not avoided, could result in death or serious injury. CAUTION: Indicates a potentially hazardous situation, which, if not avoided, may result in minor or moderate injury. Caution also indicates actions that may cause property damage.
860052
•Q
SAFETY •
P A G E 1.1
MAIN: 816-753-2150
INTERNET: kochequipment.com
E-MAIL: info@kochequipment.com
Food Safety
Food Packaging
While this machine is often used for food packaging and vacuum cooking, there are inherent risks associated with this packaging technique that could cause serious illness or death to the consumer of the food product. If you are using this machine for a food application, you must consult with a reputable food technologist or specialist in vacuum/ modified atmosphere packaging (M.A.P.) to review the safety of your application.
Gas Flush
In order to ensure proper shelf life of the food product packaged in this machine, you must contact a reputable food technologist or specialist in vacuum/modified atmosphere packaging (M.A.P.) to review and develop the appropriate gas mixture for your package, and you must perform quality control and gas analysis on your finished M.A.P. packages.
General Safety Guidelines
Obvious safety guidelines should be observed.
Be sure to turn off power to your packaging machine before any maintenance work is
performed.
Follow approved Lockout/Tagout procedures.
Place machine on a flat, stable surface.
Do not place tools, parts, or other objects on or inside machine while operating.
860052
•Q
SAFETY •
P A G E 1.2
MAIN: 816-753-2150
DANGER
INTERNET: kochequipment.com
E-MAIL: info@kochequipment.com
STARTUP
Unpacking
1. Carefully remove crate from the skid.
2. Remove machine from skid.
3. Wipe down outside of the machine.
Power Requirements
The machine is available with 2 horsepower, or 3 horsepower vacuum pumps.
The owner must supply the correct 3 phase or single phase power source in accordance with the National Electric Code.
Grounding Instructions
The Ultravac® 500 must be grounded. In the event of malfunction or breakdown, grounding provides a path of least resistance for electric current to reduce the risk of electric shock. All 3 phase units are equipped with a equipment-grounding plug and the single phase units are equipped with an equipment-grounding conductor cord. The plug must be plugged into an appropriate outlet that is properly installed and grounded in accordance with all local codes and ordinances.
Hazardous voltage. Improper connection of the equipment-grounding conductor can result in a risk of electric shock.
The equipment-grounding conductor outer surface is green with or without yellow stripes. If repair or replacement of the cord or plug is necessary, do not connect the equipment-grounding conductor to a live terminal. Check with a qualified electrician or Koch Equipment Services Inc. technician if the grounding instructions are not completely understood, or if there is doubt as to whether the unit is properly grounded. Do not modify the plug provided with the unit – if it will not fit the outlet; have a proper outlet installed by a qualified electrician.
Customer Voltage Pump Motor Full Load Amps
230, 60 Hz 1 Phase 13 230, 60 Hz 3 Phase 10 380, 50 Hz 3 Phase 6.3
Supply conductor shall have an ampacity not less than 125% of
the Pump Motor Full Load Amps
860052
•Q
STARTUP •
P A G E 2.1
MAIN: 816-753-2150
DANGER
INTERNET: kochequipment.com
E-MAIL: info@kochequipment.com
Vacuum Pump
It is essential to check the oil level daily and to change the oil after every 500 hours of operation. Read the oil level with the machine turned off. Oil may be added until the level reaches the MAX level shown in the sight glass on the pump. Refer to the pump manual supplied with the machine for details on changing the oil.
NOTE: ALL ULTRAVAC 500 CHAMBER MACHINES ARE SHIPPED WITH OIL IN THE PUMP. ALTHOUGH THE FACTORY RECOMMENDS CHECKING THE OIL LEVEL PRIOR TO FIRST USE!
Gas Flush Connection
The owner must supply a suitable regulator with a range of 0 to 60 p.s.i. We recommend
using food-grade flexible hose with a 1/4-in. I.D. and a maximum length of 15-ft. Maximum
regulator pressure is 15 p.s.i.
OXYGEN ENRICHED USE
KOCH TRAY SEALING AND VACUUM CHAMBER MACHINES ARE NOT DESIGNED FOR USE WITH OXYGEN ENRICHED PROCESS GASES. ANY APPLICATION THAT REQUIRES A PROCESS GAS CONTAINING 25% OR MORE OXYGEN SHOULD CONTACT KOCH
EQUIPMENT SERVICES AT (800) 777-5624 AS RETROFIT OPTIONS ARE REQUIRED.
RUNNING THIS MACHINE WITH AN OXYGEN ENRICHED PROCESS GAS, WITHOUT THE NECESSARY RETROFIT KIT COULD CAUSE SEVERE INJURY, DAMAGE OR DEATH.
KOCH TRAY SEALING AND VACUUM CHAMBER THAT ARE RETROFITTED WITH THE NECESSARY KIT WILL HAVE A DECAL ABOVE OR BESIDE THE GAS SUPPLY CONNECTION INDICATING THE MACHINE’S READINESS TO RUN OXYGEN ENRICHED PROCESS GASES.
Checking Vacuum Pump Rotation
Caution: Check oil level of pump before starting pump (please refer to pump manual). To check the direction of the pump rotation, briefly engage the “POWER ON” switch and observe the motor fan at the end of the pump. The fan should rotate as indicated by the arrow on the fan cover. To correct the rotation, switch any two phases in the plug.
860052
•Q
STARTUP •
P A G E 2.2
MAIN: 816-753-2150
INTERNET: kochequipment.com
E-MAIL: info@kochequipment.com
OPERATION
Placement of Product
For best sealing results, it is important to:
Check the pump oil level daily.
• Select a pouch that fits the product.
• Carefully load the product into the pouch.
• Keep the product and the product residue away from the seal area of the pouch.
• Place the product as far into the pouch as possible.
• Maintain an equal amount of the product above and below the seal bar (see figure 3.1,
below on use of filler plates).
Use Filler Plates for raising the height of the product.
• Lay the pouch flat on the seal area, keeping the pouch free of wrinkles.
• Place the pouch so that the open end is inside the chamber when the lid is closed.
Figure 3 .1
860052
•Q
PRODUCT HALF HEIGHT
EVEN WITH TOP OF
OPERATION •
SEAL BAR
P A G E 3.1
MAIN: 816-753-2150
INTERNET: kochequipment.com
E-MAIL: info@kochequipment.com
Operation with Analog Control Panel
The analog control panel is shown below, figure 3.2.
The range for timed vacuum is 0 to 55 seconds and is controlled by the Vacuum Potentiometer. We suggest an initial setting of “3” on the dial.
Seal impulse is the length of time the seal bar is turned on and can range from 0 to 2 seconds. The impulse time is controlled by the Seal Potentiometer. Koch Equipment recommends an initial setting of “6” on the potentiometer. This setting will vary according to the thickness of the pouch. Thinner pouches will require a lower setting while thicker pouches will require a higher setting.
Experiment with both settings to achieve the best results.
Figure 3.2
POWER SWITC H
VACUUM
POTENTIOMETER
VACUUM LIGHT
(GREEN)
SEAL
POTENTIOMETER
VACUUM
INDICATOR
SEAL LIGHT
(RED)
860052
•Q
OPERATION •
P A G E 3.2
MAIN: 816-753-2150
INTERNET: kochequipment.com
E-MAIL: info@kochequipment.com
Operation with Digital Control Panel
The digital control panel allows the user more options than the standard control panel. The embedded microprocessor controls each sequence of the packaging operation. Settings for the vacuum, gas, and sealing are entered as parameters through the keypad. This allows the user to custom program every step of the packaging process. The precise vacuum and gas pressures are controlled by a pressure based sensor. The vacuum pressure, gas pressure, and seal time are displayed on a large 16-character LCD backlit readout, which is easily readable in all lighting conditions. As each sequence is performed, the real-time pressure level or cycle time is displayed.
The digital front panel can save up to ten pre-programmed routines, which can be retrieved at any time for specific packaging applications. With the supervisor security feature turned on, these programs cannot be inadvertently changed.
The Vacplus option allows the operator to run the pump from 0 to 20 seconds after the set vacuum level is achieved.
The Gas Flush option allows the operator to introduce an inert gas into the chamber after the vacuum stage. This option can be used as a filler to prevent crushing of the product after sealing, as a means to prolong shelf life, or as a means to maintain desirable product appearance.
The digital front panel has an auto stop, which will automatically seal if the preset vacuum is not reached. This feature decreases the cycle time and optimizes the vacuum level of each product.
The digital front panel, which includes the keypad, illuminated display, and microprocessor, use sealed components and is conformal coated in a moisture-proof coating. The digital
front panel meets or exceeds the requirements of NEMA 4. The front of the digital display is
sealed and flush for easy cleaning.
The digital control has both pulsed vacuum and pulsed venting options for fragile product.
The digital control has a maintenance screen for testing valves and a special loop option for multiple vacuum/gas cycles before sealing.
Figure 3.3
860052
•Q
OPERATION •
P A G E 3.3
MAIN: 816-753-2150
THAT WAS RUN WILL BE THE CURRENT PROGRAM SHOWN IN THE WINDOW. TO SET THE OPERATING PARAMETERS FOR THE PROGRAM SHOWN, PRESS THE MENU KEY.
THIS IS THE MAIN MENU SCREEN. WHEN THE MACHINE STARTS UP, THE LAST PROGRAM
NOTE: IF THE SUPERVISOR HAS SET SECURITY ON, THESE SETTINGS CANNOT BE CHANGED
UP AND DOWN ARROW KEYS.
VACUUM IS SET TO % VACUUM USING THE
(FOR THE SPECIFIED NUMBER OF SECONDS) AFTER
THE % VACUUM SET IN THE VACUUM MENU.
THE PRESSURE IN THE CHAMBER HAS REACHED
PUMP TO CONTINUE EVACUATING THE CHAMBER
A SETTING GREATER THAN ZERO ALLOWS THE
VACPLUS MAY BE SET FROM 0 TO 20 SECONDS.
USE THE UP AND DOWN ARROW KEYS TO
THE SEAL SETTING IS IN SECONDS.
CHANGE THE SEAL TIME.
THE RANGE IS 30% TO 99%.
THE RANGE IS 0 TO 2 SECONDS.
THE RANGE IS 0 TO 20 SECONDS.
PRESS THE MENU KEY.
PRESS THE MENU KEY.
PRESS THE MENU KEY.
INTERNET: kochequipment.com
E-MAIL: info@kochequipment.com
Operator Menu on Digital Panel
NOTE: if the supervisor has set security on, these settings cannot be changed.
+
RUNNING PROG: 1
-
+
VACUUM: 98
-
+
VACPLUS: 4.0
-
+
SEAL: 1.2
-
PROGRAM
YES
This is the Main Menu screen. When the machine starts up, the last program that was run will be the current program shown in the window. To set the operating
NO
parameters for the program shown press the MENU Key.
PROGRAM
YES
VACUUM is set to % vacuum using the UP and DOWN arrow keys. The range is 30% to 99%. Press the MENU Key.
NO
PROGRAM
YES
VACPLUS may be set greater than zero to allow the pump to continue evacuating the chamber (for the specified number of seconds) after the pressure in the
NO
chamber has reached the % vacuum set in the vacuum menu. The range is 0 to 20 seconds. Press the MENU Key.
PROGRAM
YES
NO
The SEAL setting is in seconds. Use the UP and DOWN arrow keys to change the seal time. The range is 0 to 2 seconds. Press the MENU Key.
Figure 3.4
860052
•Q
OPERATION •
P A G E 3.4
INTERNET: kochequipment.com
ALSO CORRESPONDS TO THE VACUUM
GAS IS SET TO % GAS USING THE UP AND DOWN ARROW KEYS. THIS VALUE
IF YOUR MACHINE IS EQUIPPED WITH THE GAS
PRESSING THE UP OR DOWN ARROW KEYS.
OPTION, IT CAN BE TURNED ON OR OFF BY
READING INSIDE THE CHAMBER.
WHEN GAS FLUSHING TO A LOW % VACUUM. THE RANGE IS 98% TO 30%.
IT IS RECOMMENDED TO USE PULSE VENT
PRESS THE MENU KEY.
PRESS THE MENU KEY.
WARNING
E-MAIL: info@kochequipment.com
Operator Menu on Digital Panel (continued)
MAIN: 816-753-2150
+
PROGRAM
YES
If the machine is equipped with the gas option, it can be turned ON or OFF by
GAS: YES
pressing the UP or DOWN arrow keys. Press the MENU Key.
-
+
PROGRAM
NO
YES
GAS is set to % gas using the UP and DOWN arrow keys. This value also corresponds to the vacuum reading inside
GAS: 75
the chamber. It is recommended to use Pulse Vent when gas flushing to a low %
-
NO
vacuum. The range is 98% to 30%. Press the MENU Key.
Figure 3.5
Explosion hazard. Do not use a gas with an oxygen content greater than 22% with gas flush option.
Selecting a New Program
RUNNING PROG: 1
ENTER PROGAM: 2
Figure 3.6
860052
•Q
+
PROGRAM
YES
From the Main Menu, use the UP and DOWN arrow keys to select a new program.
-
+
PROGRAM
NO
Press the UP arrow to switch to Program 2 (or any of the 10 programs).
YES
New parameters may be set for Vacuum, Seal Time, and Gas for Program 2 (following the procedures described in the previous example) or simply close the lid on the
-
OPERATION •
NO
machine and run Program 2.
P A G E 3.5
MAIN: 816-753-2150
WARNING
INTERNET: kochequipment.com
E-MAIL: info@kochequipment.com
Sealing with Air-Assist
All machines are equipped with regulators for air-assisted sealing. Set the air pressure
regulator to 20 p.s.i. increasing to a maximum of 40 p.s.i. While a good seal can be
obtained without air-assist, use air-assist when:
• Gas back-flushing above average pressure.
• Using shrink pouches.
• Packaging a product that easily contaminates the seal area of the pouch.
• Trying to overlap pouches.
• Wrinkles cannot be avoided in the seal area.
Using the Double Seam Seal Option.
Gas Flush Option
Gas flushing is the introduction of an inert gas into the chamber after the vacuum stage is finished. Gas can be used as a filler to prevent crushing of the product after sealing, as a means to prolong shelf life, or as a means to maintain desirable product appearance. Commonly used gasses include nitrogen, carbon dioxide, or a mixture of both. Consult your local gas supplier to select the proper gas for your product.
Explosion hazard. Do not use a gas with an oxygen content greater than 22% with gas flush option.
Double Seam Seal Option
All machines are equipped with standard seal bars having a single seal element per seal bar. The Double Seam seal bars have two seal elements per seal bar. We recommend using air-assisted sealing with this option to achieve best results.
860052
•Q
OPERATION •
P A G E 3.6
MAIN: 816-753-2150
DANGER
CAUTION
INTERNET: kochequipment.com
E-MAIL: info@kochequipment.com
MAINTENANCE
Prior to Cleaning
Every environment and application is different; therefore, Koch Equipment LLC cannot provide cleaning instructions to guarantee microbiological sanitation. Koch Equipment requests that the owner of this machine consult with sanitation experts to review the unit working in their particular environment to develop a robust cleaning schedule and methodology, followed by bacterial testing to ensure satisfactory cleaning procedures are followed.
Cleaning Recommendations
Before cleaning the machine, turn power off; disconnect the main power, and lockout the connection.
Hazardous voltage. Disconnect and lockout power before servicing machine or cleaning. Do not remove panels unless power has been disconnected and locked out at risk of electric shock hazard.
Check with the detergent and sanitizer manufacturers that their products are compatible with the listed materials.
Cleaning agents. Do not get the cleaning agents in eyes, on skin, or on clothing. Always wear rubber gloves, goggles, and protective clothing when contact is likely. Consult product manufacturer for specific details.
Never hose down the machine. Damage caused by hosing or high pressure washing is not covered under warranty.
1. Filler Plates: Remove filler plates. The filler plates are made from polyethylene. Clean, sanitize, and dry. High pressure water spray can be used on the filler plates.
2. Lid and Backup Strip: The lid is constructed of acrylic. Use only nonabrasive soap and water. Do not use window sprays or kitchen scouring compounds. The Backup Strip is made of silicone. Clean, sanitize, and dry.
3. Seal Bars: Remove the seal bars by first lifting them up off of the guide rods. Remove the wire connectors from the adapter clips on the seal bar and remove the seal bar from the machine. The seal bars are made of aluminum and phenolic. Clean, sanitize, and dr y.
4. Chamber and Base: The chamber is made of 304 stainless steel and the base is made
from polyurethane. Clean, sanitize, and dry, including under the seal bar bladder (not removable).
5. Clean under the machine.
6. Reinstall the seal bar.
7. Use bacteriological testing to insure cleaning process.
860052
•Q
MAINTENANCE •
P A G E 4.1
MAIN: 816-753-2150
INTERNET: kochequipment.com
E-MAIL: info@kochequipment.com
Vacuum Pump Maintenance
Consult the pump manufacturer’s manual provided with the machine for detailed information.
Seal Bar Maintenance
The following illustrations show replacement of the seal elements for the seal bars.
Step 1. Remove the seal bars from the machine. Pull off the Teflon® tape strip and discard. Clean off any remaining Teflon® tape adhesive using acetone or an equivalent solvent.
Step 2. Using a 2mm Allen wrench, loosen the set screws for the cut-off wire and the seal element on both ends of the seal bar and discard.
Cut-off wire
CUT-OFF WIRE
Seal element
Teflon® tape
TEFLON
SEAL ELEMENT
TAPE
Set screw
SET SCREW
2mm Allen
2mm ALLEN
wrench
Figure 4 .1
WRENCH
Figure 4.2
Step 3. If you are replacing the L-shaped spring retainer and the spring at this time, loosen the set screw and remove the old spring retainer and spring and install the new ones. Adjust the spring retainer to allow a 1/8-in. gap between the spring retainer and the seal bar. This will allow the seal element to remain under tension after tightening the seal element.
860052
Figure 4.3
•Q
Tightening
TIGHTENING
TOOL
tool
MAINTENANCE •
Brass contact
BRASS CONTACT
Slotted tip
SLOTTED TIP SCREWDRIVER
screwdriver
P A G E 4.2
MAIN: 816-753-2150
INTERNET: kochequipment.com
E-MAIL: info@kochequipment.com
Reading the Indicators
If your machine is controlled by a digital control module (with touch pad), this design to aids in troubleshooting. The digital panel has indicator lights mounted on the front below the ON/OFF switch. The indicator lights correspond to the operating devices and should turn on and off in the following sequence when the machine lid is closed:
1. Vacuum Pump [C-1]
2. Gas Flush Valve [SOL-3] (Optional)
3. Seal Bladder Valve [SOL-2] (Stays on until completion of cycle)
4. Seal Impulse Contactor [C-2]
5. Ventilation Valve [SOL-1]
The device should be operating when the light is illuminated. If the lights illuminate in the proper sequence, but there is still a problem, look for the problem in the operating device itself. If the lights do not sequence properly, look for a problem in the control module.
Troubleshooting
Problem Indications Remedy
Machine will not start Green power “ON” light not
lit when switch is turned on
Vacuum pump does not run Make sure that the power
No vacuum When lid is closed, indicator
light (VAC) is “OFF” on the control module
Vacuum not pulling lid down. Check intake screen in vacuum
Longer vacuum cycle times Check intake screen in vacuum
No gas flush (optional) If indicator light (GAS) is lit Check for proper gas pressure
Make sure that the power requirements match those given on the nameplate. Also, check fuse F-2; replace if blown.
requirements match those given on the nameplate
Check lid switches LS-1 for proper adjustment.
pump hose barb for blockage, pieces of bags, labels, bone, etc.
pump hose barb for blockage
going into gas inlet
860052
•Q
If indicator light (GAS) is not lit
MAINTENANCE •
Check for proper operation of gas flush valve (SOL-3)
Check gas flush potentiometer on analog control module or possible defective digital control panel
P A G E 4.3
E-MAIL: info@kochequipment.com
Troubleshooting
Problem Indications Remedy
MAIN: 816-753-2150
INTERNET: kochequipment.com
Chamber not venting (lid will not open)
Lid will not open and red indicator light “VENT” on control module is lit
“VENT” indicator light is not lit
NOTE: Lid can be released by pulling the hose off of the vacuum gauge to remove product.
Improper or no sealing
Seal bladder light on control module is lit but the seal bar does not go up
The seal bar is not heating up even though the red seal light on the front panel comes on
The red seal light on the front panel either does not light for the proper length of time (½ to 1 second) or does not light at all
Check ventilation valve SOL-1 for proper operation
Check cool down potentiometer on analog control module or possible defective digital control panel
Check to make sure that the regulator knob is turned fully clockwise, or, if air-assist is used, set to the recommended pressure
Check seal bladder valve SOL-2 for proper operation
Check seal bar connection points and clips for corrosion and proper tension
Check for broken seal element Check seal bar fuse F-1. Located
on the analog or digital panel. If the machine is 110VAC, 2 seal bar, F-1 is located on the chassis.
Make sure the seal impulse potentiometer is set high enough or check for possible defective digital control panel
NOTE: For proper sealing, three things must occur:
1. The seal bar must come down and place adequate pressure between the seal bar and the backup strip.
2. The seal element must heat up sufficiently to fuse the pouch.
3. The pouch must be allowed to cool for a time to ensure a good “set.”
860052
•Q
MAINTENANCE •
P A G E 4.4
MAIN: 816-753-2150
INTERNET: kochequipment.com
E-MAIL: info@kochequipment.com
Supervisor Menu on Digital Panel
Untrained personnel should not alter any setting in the supervisor menu.
+
PROGRAM
YES
RUNNING PROG: 1
-
NO
The following are parameters that can only be set in the supervisor menu.
+
PROGRAM
YES
PREVAC: 15
-
NO
This is the Main Menu screen. When the machine starts up, the last program that was run will be the current program shown in the window.
Press and hold down both arrow keys for at least 3 seconds to enter the supervisor menu.
NOTE: If your machine is not installed with the precut option, disregard the PREVAC, CUTTIME, and WAITTIME parameters.
PREVAC is the % vacuum that you wish to reach before activating the knife, which makes slits in the bag for precut. Use the up or down arrow keys to change the current setting. The range is from 5% to 20% vacuum. Press the MENU key.
CUTTIME: 1
WAITTIME: 3
Figure 4.4
+
PROGRAM
YES
CUTTIME is the number of seconds that the seal valve is activated. This is the time when the knife actually makes the evacuation slits in the bag/pouch for precut. The range is from 0 to 3
-
NO
seconds. Press the MENU key.
+
PROGRAM
YES
WAITTIME is the number of seconds the machine waits for the knife to retract before resuming the vacuum cycle. The range is from 0 to 6 seconds.
-
NO
Press the MENU key.
860052
•Q
MAINTENANCE •
P A G E 4.5
Supervisor Menu on Digital Panel
MAIN: 816-753-2150
INTERNET: kochequipment.com
E-MAIL: info@kochequipment.com
COOLDOWN: 4.0
SECURITY: NO
+
PROGRAM
YES
COOLDOWN is the time that the machine waits after sealing the bag/pouch before the seal bars retract. This allows time for the actual seal to cool down and take a set.
-
NO
The range is from 1 to 8 seconds. Press the MENU key.
+
PROGRAM
YES
SECURITY can be turned on by selecting yes. If security is on, the operator menu will be read only and the operator will not
-
NO
be able to change any settings. Press the MENU key.
# PROGRAMS: 10
TEST: YES
Figure 4.5
+
PROGRAM
YES
number of programs that may be saved in memory. Each program may have its own set of unique parameters. This range is from 1 to 10.
This is how you specify the maximum
-
NO
Press the MENU key.
If TEST is set to YES (using the up arrow
+
PROGRAM
YES
key), the vacuum valve and vent valve may be operated manually. The up arrow key activates the vacuum valve, the down arrow key activates the vent valve, and pressing both at the same time turns them
-
NO
both off. As you activate either valve, the vacuum pressure is displayed. Press the MENU key.
860052
•Q
MAINTENANCE •
P A G E 4.6
Supervisor Menu on Digital Panel
MAIN: 816-753-2150
INTERNET: kochequipment.com
E-MAIL: info@kochequipment.com
PULSE VAC: YES
PULSE TO: 95
+
PROGRAM
YES
PULSE VAC allows for a more controlled vacuuming process. Rather than simply opening up the vacuum valve for a continuous vacuum the vacuum valve is opened and closed repeatedly until the
-
NO
chamber reaches the set vacuum level. Press the MENU Key.
+
PROGRAM
YES
This is the % vacuum reached by pulsing the vacuum valve before the vacuum valve opens completely to fully evacuate the chamber. The range is 99% to 20%.
-
NO
Press the MENU Key.
PULSE ON: 1.2
Figure 4.6
+
PROGRAM
YES
This is how long the vacuum valve is turned on during the vacuum cycle. The range is from 0 to 3 seconds. Press the MENU Key.
-
NO
860052
•Q
MAINTENANCE •
P A G E 4.7
Supervisor Menu on Digital Panel
DANGER
MAIN: 816-753-2150
INTERNET: kochequipment.com
E-MAIL: info@kochequipment.com
PULSE OFF: 1.0
VAC ERROR: ON
Figure 4.7
+
PROGRAM
YES
This is how long the vacuum valve is turned off during the vacuum cycle. The range is from 0 to 3 seconds. Press the MENU key.
-
NO
VAC ERROR is an option that is turned ON
+
PROGRAM
YES
or OFF. If set OFF, default, the machine will time out if the preset vacuum is not reached in 30 seconds, and the machine with complete the cycle. If set ON and the
-
NO
machine does not reach the preset vacuum in 30 seconds, then the panels will display VAC ERROR and vent the chamber without sealing.
Opening Chamber Machine
1. Turn the power switch to the “OFF” position and disconnect the power cord from the wall receptacle.
Hazardous voltage. Disconnect and lockout power before servicing machine or cleaning. Do not remove panels unless power has been disconnected and locked out at risk of electric shock hazard.
2. Remove the screws located on the back of the machine’s stainless steel cabinet. Lift up on
the stainless panel to access inside the machine.
860052
•Q
MAINTENANCE •
P A G E 4.8
MAIN: 816-753-2150
CAUTION
INTERNET: kochequipment.com
E-MAIL: info@kochequipment.com
Changing Vacuum Pump Oil
1. Open the machine (see above).
2. A trough and hole is located toward the front of the machine, between the pump and
front edge of the chassis. Slide the machine over to the edge of the table with the hole over the edge of the table. Position an oil collection pan underneath this hole.
Tip Over Hazard. Only let the front foot of the chassis hang over the edge of the table.
3. Loosen and remove the Oil Filler Cap with a 1¼-in. open end wrench.
4. Loosen the Oil Drain Plug located to the front and lower middle on the pump with a
1¼-in. open end wrench. Remove the plug slowly to control the flow of oil.
5. With the oil drained, replace the Oil Drain Plug and wipe any excess oil out of the trough
in the chassis.
6. Fill the pump with oil until the level is between the minimum and maximum lines on the
Oil Sight Glass. Replace the Oil Filler Cap.
7. Tilt the machine back down onto the chassis. Plug the power cord back into the wall
receptacle.
8. Cycle the machine a couple of times, unplug, and tilt the machine up to check the Oil
Level. Add oil if necessary. Tilt the machine back down and replace the two screws in the rear of the machine.
Replacing Potentiometer
Refer to Parts, “Analog Control Panel,” for the disassembly of the panel. Remove the potentiometer by unbending the mounting tabs holding the potentiometer to the circuit board. Pull the potentiometer out of the circuit board.
MOUNTING TABS
Figure 4.8
860052
•Q
MAINTENANCE •
P A G E 4.9
MAIN: 816-753-2150
DANGER
INTERNET: kochequipment.com
E-MAIL: info@kochequipment.com
Replacing Power Cord
1. Turn the power switch to the “OFF” position and disconnect the power cord from the wall
receptacle.
Hazardous voltage. Disconnect and lockout power before servicing machine or cleaning. Do not remove panels unless power has been disconnected and locked out at risk of electric shock hazard.
2. Open the machine as described in “Opening Chamber Machine” on page 4.8.
3. Loosen the cord grip nut on the rear of the machine. Remove the cord from the cable
clamps on the bottom of the chamber. Disconnect the cord from the control panel.
4. Route the new power cord through the cord grip on the rear and through the cable
clamps on the bottom of the chamber.
5. Connect the new power cord to the control panel as shown in the table below.
Connect to
Terminal
110 Volt Wire
Color
230 Volt Wire
Color
Terminal 8 Green (Ground) Green/Yellow Terminal 7 White Brown Terminal 6 Black Blue
Maintenance Log
A maintenance log is a journal of all maintenance performed. Each entry includes a date, maintenance performed (details about the type of work done), and technician (who performed the maintenance). The maintenance log is also a place where a schedule is kept for further maintenance.
A maintenance log will clearly show oil changes, daily inspections, Teflon® tape
replacement, and so on. A master copy has been provided on page 4.11, please create a
copy and store in the back of this owner’s manual.
Service Log
A service log is a journal of all service work performed. Each entry includes a date, service provided (details about the type of service), and technician (who performed the service).
A service log will clearly show training provided, frequent wear items, and so on. A master
copy has been provided on page 4.124, please create a copy and store in the back of this
owner’s manual.
860052
•Q
MAINTENANCE •
P A G E 4.10
Loading...
+ 59 hidden pages