Ultrasound Technologies Fetatrack 360 User manual

®
l
fetatrack
fetal monitoring system
360
User Manua

CONTENTS

INTRODUCTION 2 EQUIPMENT 3 SAFETY 4 SPECIFICATIONS 5 CONTROLS AND INDICATORS 7 OPERATING THE FETATRACK 360 9 PAPER LOADING 10 THE PRINTER 11 TRANSDUCER BELTS 12 FETAL HEART RATE MONITORING 13 DIRECT FETAL ECG 15 FETALSAT PULSE OXIMETRY 17 ALARMS 18 UTERINE ACTIVITY MONITORING 19 INTRAUTERINE PRESSURE MONITORING 20 FETAL ACTIVITY MONITORING 22 MATERNAL HEART RATE MONITORING 22 PREVENTATIVE MAINTENANCE 23 MONITOR CHECK-OUT PROCEDURE 24 SETTING UP THE FETATRACK 360 25 WARRANTY 26 GUIDELINES 28 NOTES 29
INTRODUCTION
The FETATRACK 360 Cardiotocograph (available as an Antepartum or Intrapartum monitor) is designed to provide cost effective care throughout pregnancy. Internal or external transducers may be connected to the FETATRACK 360. The functions described below are summarized in the section on EQUIPMENT.
All units include twins ultrasound fetal heart rate measurement using dual Doppler ultrasound transducers and external uterine pressure UA can be monitored using the external (tocodynamometer).
The intrapartum monitor adds fetal ECG (FECG) and has an option of including maternal ECG (MECG) measurement and an intrauterine pressure transducer (IUP) for more accurate uterine pressure measurement.
Both units can add an option for the monitoring of Mater nal pulse oximetry.
The FETATRACK 360is has the addition of the Nonin FetalSat fetal pulse oximetry system.
All parameters are displayed on a large graphic displays which also displays messages as appropriate. Ultrasound signal quality is shown by a change in the flashing heart associated with the fetal rate display, so you can be sure the recording is accurate and reliable.
The information is presented on a high resolution thermal pr inter running at 1, 2 or 3 cm/min, and printed on standard format full width z-fold paper. Event marks, date, time and mode are printed on the paper to complete the record.
All FETATRACK 360s have bradycardia and tachycardia alarm functionality, memory backup and an output socket for remote display of screen information on a local computer, for example in the consultant's office
EQUIPMENT
The FETATRACK 360 is supplie d with the followi ng accessories for each version:
Wide Angle Ultra sound Transducer FECG Leg Lead 1 1 n/a Fetal Scalp Electro des 2 2 n/a External Uterine Pressure Transducer Internal Uterine Pressure Transducer IUP Cable Option Option n/a Maternal Pulse Oximetry Option Option Option Maternal ECG Option Option n/a FetalSat Pulse oximetry 1 N/a N/a FetalSat Dual Sense ecg / FSpO2 Event Marker 1 1 1 Elastic Belt 3 3 3 Belt Buckles 3 3 3 Chart Paper 1 1 1 Ultrasound Coupling Gel
0.25ltr AC line cord 1 1 1 User Instructions 1 1 1
REORDER INFORMATION Part No. Item
0215-0003 Wide Angle Ultrasound Transducer (Blue) M/1/0059 Wide Angle Ultrasound Transducer (Pink) 0211-0003 External Pressure Transducer 9800-0000 Internal Pressure Transducer 0261-0005 FECG Leg Lead LifeTrace Safelink 9021-0002 Transducer Belt Set 9260-0005 Chart Paper 20 BPM/cm 9260-0006 Chart Paper 30 BPM/cm 9000-0005 Coupling Gel 0.25L x 12 bottles 9088-0003 Fetal Scalp Electrode “Safelink” FetalSat leg connector FetalSat dual se nse ecg / FspO2 sensor
360is 360i 360a
2 2 2
1 1 1
Option Option n/a
1 N/a N/a
1 1 1
SAFETY
The FETATRACK 360 is designed to comply with EN60601-1, Class 1 with protective earth via the AC line cord. Connection must be made to ground and supply complying with local standards. For correct operation check that the voltage setting and fuse ratings are correct for the local supply.
The FETATRACK 360 is not explosion proof and must not be used in the presence of flammable anaesthetics.
If in doubt about the correct operation of the FETATRACK 360, fetal condition must be checked immediately by an alternative method.
We would recommend that, to maintain the standard of performance of the FETATRACK 360, whenever possible the monitor is included in a scheduled maintenance scheme.
The following symbols have been used on the front and rear panels of the FETATRACK 360 and are here defined according to EN60101-1.
Alternating Current
Power On (Connection to the mains)
Power Off (Disconnection from the mains)
Type B Equipment
Type CF Equipment
fetatrack 360is issue 1 4
Attention: consult accompanying documents
SPECIFICATIONS
Size: 350 x 370 x 140 mm Weight: 8 kg Voltage: 100-240 Volts AC 50/60 Hz Safety: Complies with EN60101-1
ULTRASOUND Type: Continuous Doppler Operating Frequency 2MHz±2% and 1.8MHz ±2% Power Output 5 mW/sq cm max. SATA Signal Processing Software AUTOCORRELATOR
DIRECT FECG (Intra models only) Signal Range 40µV-5mV Input Noise: <1,4µV p-p CMRR: >125dB
ALARMS Low Level: 60 - 120 BPM, Adjustable in steps of 5 BPM High Level: 160 - 200 BPM, Adjustable in steps of 5 BPM
EXTERNAL UTERINE ACTIVITY Type: Flat faced Tocodynamometer Signal Range: 0 - 110 relative contractions strength
INTERNAL UTERINE ACTIVITY (Intra models only) Intrauterine Pressure: Type optional Signal Range 0 - 100 mmHg
EVENT MARKER Hand held, perso n operate d
PRINTER Type: Thermal Array 8 dots/mm Print Speed: 1,2 and 3 cm/min Paper Size: 145 mm x 15 mtr FHR Sensitivity: 20 or 30 BPM/cm UA Sensitivity: 30/cm IUP Sensitivity: 30mmHg/cm
DISPLAY/PRINTER RANGES FHR Range: 30 - 240 BPM FHR Accuracy: +/- 1 BPM UA Range: 0 - 110 units FspO2 Range: 0 – 100%
OXIMETER Oxygen Saturation 0% to 100% (%FSpO2) Pulse Rate Range 30 to 300 bpm
Accuracy Blood Oxygen Saturation For specification, see footnote “a” Pulse Rate 30 to 300 bpm; ±3 bpm
Displays Graphical Displays 3-digit Pulse Strength Graphic Bargraph
Measurement Wavelengths and Output Power Red 661 nanometers @ 3 mw nominal Infrared 905 nanometers @ 3 mw nominal
“a”. Calibration of the FetalSAT System was accomplished through car efully controlled hypoxia studies in a laboratory animal model to establish the calibration coefficients. For specifications, see FetalSAT Specification Supplement (available upon request).
CONTROLS AND INDICATORS
FRONT PANEL
Set Uterine Pressure to Zero
Ultrasound Volume Decreas
Audio Selector for Ultrasound 1 & 2
Ultrasound Volume Increase
Printer On/Off & Speed Control
e
Maternal & Fetal Electrocardio-graph & Ultrasound 2 Transducers Input
Internal & External Uterine Pressure Transducers Input
Power On Indicator
Ultrasound 1 Transducer Input Socket
Printer Door Release Lever
fetatrack 360is issue 1 7
REAR PANEL
Fuse Compartment
Mains On/Off
Power Inlet
Network or Local Computer Port Pulse Oximeter Input
Event Marker
fetatrack 360is issue 1 8
OPERATING THE FETATRACK 360
The FETATRACK 360 comes to you with a multi-ra nge power supply which can be switched exter nally to suite the local AC line voltage. It will oper ate with an input voltage between 100 and 240 vol ts AC, 50 or 60 Hz.
Connect the AC supply cord and switch the FETATRACK 360 on. The AC input on/off switch is located on the rear of the unit adjacent to the AC line input socket. Observe that the front panel power on indicator (1/0) is illuminated.
The FETATRACK 360 is now operational and the display will indicate the operating mode.