Ultrasound Technologies PD1 series, Fetatrack 120 series Service Manual

Ultrasound Technologies Ltd
Service manual
PD1 / FT120 service manual Issue 3 January 2007
©Ultrasound Technologies Ltd, Lodge Way, Portskewett, Caldicot, South Wales NP26 5PS, United Kingdom. T +44 (0) 1291 425425 F: +44 (0) 1291 427093
E: ultratec@doppler.co.uk W: www.doppler.co.uk
Ultrasound Technologies Ltd
Products covered by this service manual
PD1 series pocket dopplers
Product Group
Product version
Product variation
PD1 audio pocket doppler
Babypulse home use doppler
Fetatrack 120 audio doppler
PD1+ fetal doppler with FHR display
PD1v vascular doppler
Fetatrack 120+ fetal doppler with FHR
PD1dwr waterbirth doppler
PD1cv low cost vascular doppler
PD1v flat probe vascular doppler
PD1combi fetal / vascular doppler
V
ascutrack 120
vascular doppler
PD1 / FT120 service manual Issue 3 January 2007
©Ultrasound Technologies Ltd, Lodge Way, Portskewett, Caldicot, South Wales NP26 5PS, United Kingdom. T +44 (0) 1291 425425 F: +44 (0) 1291 427093
E: ultratec@doppler.co.uk W: www.doppler.co.uk
Ultrasound Technologies Ltd
PD1 / FT120 service manual Issue 3 January 2007
©Ultrasound Technologies Ltd, Lodge Way, Portskewett, Caldicot, South Wales NP26 5PS, United Kingdom. T +44 (0) 1291 425425 F: +44 (0) 1291 427093
E: ultratec@doppler.co.uk W: www.doppler.co.uk
Products covered by this service manual
The following products are manufactured by
Ultrasound Technologies ltd,
Lodge Way,
Portskewett,
NP26 5PS,
United Kingdom
Tel:- +44 (0) 1291 425425
Fax: +44 (0) 1291 427093
Email: service@doppler.co.uk
Ultratec PD1 series fetal and vascular dopplers
Rotburg PD1 / PD1+ series fetal dopplers
Sibel PD1/ PD1+ series fetal dopplers
Gima G2000 series fetal dopllers
Gima V2000 seies vascular dopplers
Fetatrack series fetal dopplers
Vascutrack series fetal dopplers
Ultrasound Technologies Ltd
CONTENTS
1. 0General Introduction
1.1.
14
Introduction
1.2.
15
Re Order Information
1.3.
16
Symbols
1.4.
17
CE Marking
1.5.
18
Guidelines for identifying and resolving adverse EMC conditions
1.6.
19
Specification
2.
1
Description of the Instrument
2.1.
20
Audio Unit.
2.2.
21
Audio Unit
2.3.
22
Audio Unit
2.4.
23
Dismantling Instructions
2.5.
24
Audio Unit
2.6.
25
Transducers
3.
2
Circuit Description
3.1.
26
Introduction
3.2.
27
Audio Unit Power Supply
3.3.
28
Battery Low Indicator
3.4.
29
Ultrasound Transducer
3.5.
30
Oscillator and Detector
3.6.
31
Receiver and Detector
3.7.
32
Audio Amplifier
3.8.
33
Velocity Processor
4.
3
Test and Calibration
4.1.
34
Introduction
4.2.
35
Performance Checks
4.3.
36
Test Procedure
4.4.
37
Audio Unit
4.5.
38
Transducer
4.6.
39
Velocity Processor
4.7.
40
Transducer Calibration
5.
4
Fault Finding
5.1.
41
Fault Finding Guidelines
6.
5
Parts Lists
6.1.
42
Audio Unit PCB Assembly
6.2.
43
Audio Unit with heart rate PCB Assembly
6.3.
44
2MHz Transducer assembly
6.4.
45
General Assembly
6.5.
46
+ units General Assembly
7.
6
Drawings and Circuit Diagrams
PD1 / FT120 service manual Issue 3 January 2007
©Ultrasound Technologies Ltd, Lodge Way, Portskewett, Caldicot, South Wales NP26 5PS, United Kingdom. T +44 (0) 1291 425425 F: +44 (0) 1291 427093
E: ultratec@doppler.co.uk W: www.doppler.co.uk
Ultrasound Technologies Ltd
©Ultrasound Technologies Ltd Page 1 of 32 PD1 / FT120 service manual Issue 3 January 2007
1. 7General Introduction
1.1. 47Introduction This service manual is written to support the maintenance and repair of the PD1 Series Doppler detectors. Servicing of this equipment should be performed by a qualified technician, after carefully studying this manual. The drawings and circuit descriptions in this manual are correct as of the date it was prepared, however, however the Company
reserves the right to make changes to improve the operation of the instrument. If your instrument does not exactly match the information in this manual, contact the manufacturer or distributor for revision information.
Inspect the instrument upon receipt, in the unlikely event that the unit has been damaged, notify your distributor or Ultrasound Technologies directly at the following address: -
Ultrasound Technologies Ltd., Lodge Way, Portskewett
48
Caldicot, NP26 5PS South Wales , UK Telephone: +44 (0) 1291 425425 FAX: +44 (0) 1291 427093
49
E-mail: service@doppler.co.uk Retain the packing material for possible future use.
Important Note:
In the unlikely event that the instrument must be returned to Ultrasound Technologies for service or for any other reason, use the same packing material in which the instrument was delivered.
If this is not available, the instrument should be packed in such a way that will provide adequate packing. The Company will not accept any responsibility for loss or damage due to improper use of packing material.
Ultrasound Technologies Ltd
1.1 50Re - Order Information Listed below are the consumables used with the PD1 Series Pocket Dopplers: -
Description Ultrasound Coupling Gel 0.25ltr.
51
MN1604 series Alkaline Battery
1.2
52
Symbols
The following symbols have been used on the PD1 and are here defined according to IEC60601
IEC Symbol 878-02-02
Type B Equipment.
IEC Symbol 348
86
Attention Consult accompanying documents
Headset Output
Volume Control
Serial DATA Out
Rear View of doppler
Showing additional symbols
©Ultrasound Technologies Ltd Page 2 of 32
PD1 / FT120 service manual Issue 3 January 2007
Ultrasound Technologies Ltd
1.3 53CE Mark This product complies with the essential requirements of the European Council Directive 93/42/EEC relating to Medical
Devices.
0120
1.4
54
Guidelines for identifying and resolving adverse EMC conditions
This product is classified as a Class A Group 1 type of product according to EN55011.
Emissions
Care has been taken through the design and manufacturing process to minimise the EM emissions, which may be produced by this equipment. However, in the unlikely event that the unit causes an EM disturbance to adjacent equipment, we suggest that the procedure is carried out ‘out of range’ of the affected equipment.
Immunity
If the user has any doubt regarding the units EM immunity during routine operation, we suggest that the source of EM disturbance is identified and it’s emissions reduced. If there is any doubt regarding the identification and resolution of adverse EM conditions, contact Ultrasound Technologies Ltd directly who will advise on the situation. The address can be found in the front of this manual.
©Ultrasound Technologies Ltd Page 3 of 32
PD1 / FT120 service manual Issue 3 January 2007
Ultrasound Technologies Ltd
1.5 Series specifications
Ultrasound
Frequency (PD1+ / PD1dwr / fetatrack 120+) 2 MHz continuous wave fixed probe
(PD1v / PD1cv / Vascutrack 120) 5 or 8 MHz continuous wave fixed probe
(PD1 combi /PD1+ combi / Combitrack) 2, 3, 5 or 8MHz interchangeable probe Transducer 2 crystal narrow beam. Output power <5mW/cm² SATA (2, 3 MHz fetal probe) <15mW/cm² SATA (5 and 8MHz vascular probe) Audio response 300Hz – 1KHz (fetal) 300Hz – 4KHz (vascular) Audio output 500mW Fetal heart rate (PD1+ / PD1+ combi / PD1dwr / fetatrack 120+) Multipoint real time autocorrelator Fetal Heart Range 50 to 210 bpm
Unit Controls
Keys 1 Key (for unit on/off on audio unit – all models) 1 Key (for unit on/off on probe – fetatrack 120 and vascutrack 120 models) Controls Rotary volume Indicators (PD1 / PD1/ combi / PD1v / fetatrack 120 / vascutrack 120) Yellow battery low LED Indicators (PD1+ / PD1+ combi / PD1dwr / fetatrack 120+)
3 digit FHR LCD display / FHR pulse
icon / Battery Low icon
Power Supply
Battery MN1604 (PP3) 9V Alkaline Manganese Expected battery life (PD1 / PD1v / fetatrack 120 / vascutrack 120) >9 hours Expected battery life (PD1+ / PD1dwr / fetatrack 120+) >8 hours
Outputs
Headset Audio output to optional headset Serial (PD1+ / PD1dwr / fetatrack 120+) RS232 interface to optional UltraTrace software Recorder (PD1v [not PD1v CAT] / vascutrack 120) Voltage output proportional to velocity of
flow
Enclosure
Material ABS Size 150mm by 75mm Weight 295g typically including transducer
Safety
Classification Class 1 Type B - IEC 60601-1
©Ultrasound Technologies Ltd Page 4 of 32 PD1 / FT120 service manual Issue 3 January 2007
Ultrasound Technologies Ltd
2. 8Description of the Instrument
2.1. 55Audio Unit PD1 (Fetatrack 120, Babypulse) The following functions are found in the relevant page of the operating instructions. UNIT ON/OFF On / Off control.
BATTERY LOW LED indicator to show when the battery has reached a point when it requires changing. VOLUME CONTROL Rotary edge potentiometer, which increases or decreases the setting of the volume. HEADSET Audio output for connection to a stereo headset for private listening.
PD1 / Babypulse fetal transducer
Fetatrack 120 fetal transducer
©Ultrasound Technologies Ltd Page 5 of 32 PD1 / FT120 service manual Issue 3 January 2007
Ultrasound Technologies Ltd
2.2. 56Audio Unit PD1+ (PD1dwr / fetatrack 120+ / babypulse +) The following functions are found in the relevant page of the operating instructions.
UNIT ON/OFF On / Off control. VOLUME CONTROL Rotary edge potentiometer, which increases or decreases the setting of the volume. HEADSET Audio output for connection to a stereo headset for private listening.
RS 232 Interface Serial data output, for connection to the optional ULTRATRACE 2 PC software. Connections are
made via the 3.5mm Stereo jack socket.
FHR DISPLAY Custom LCD module displays FHR rate, FH pulse and Battery condition.
PD1+ / Babypulse+ / PD1dwr fetal tr
ansduce
r
Fetatrack 120 + fetal transducer
©Ultrasound Technologies Ltd Page 6 of 32 PD1 / FT120 service manual Issue 3 January 2007
Ultrasound Technologies Ltd
2.3.
57
Audio Unit PD1v
The following functions are found in the relevant page of the operating instructions. UNIT ON/OFF On / Off control.
BATTERY LOW LED indicator to show when the battery has reached a point when it requires changing. VOLUME CONTROL Rotary edge potentiometer, which increases or decreases the setting of the volume. HEADSET Audio output for connection to a stereo headset for private listening.
©Ultrasound Technologies Ltd Page 7 of 32 PD1 / FT120 service manual Issue 3 January 2007
Ultrasound Technologies Ltd
2.4. 58Dismantling Instructions
Special Note: -The instrument must not be dismantled when in use on a patient, ensure all accessories are disconnected.
2.5. 59Audio Unit The Audio unit houses the main circuit board. To remove the circuit board, turn the unit onto it's front and
remove the three M4 countersunk screws. The rear panel will now come away revealing the PCB. Remove the two M3 screws securing the PCB.
©Ultrasound Technologies Ltd Page 8 of 32 PD1 / FT120 service manual Issue 3 January 2007
Ultrasound Technologies Ltd
The transducers, loudspeaker and battery connections are ‘hard wired’ to the PCB. De-solder the connections at J4 (retractile cable) and J3 (loudspeaker), the PCB can then be pulled clear of the case.
©Ultrasound Technologies Ltd Page 9 of 32 PD1 / FT120 service manual Issue 3 January 2007
Ultrasound Technologies Ltd
60
2.6 Transducers
The transducers are a ‘sealed unit’. However, it is possible to gain access to the 2 MHz assembly by removing the two blanking grommets and M3 countersink screws. Removing the screws will allow the assembly to be split along its length. Gently prise the assembly apart.
The circuit board is bonded and keyed to the faceplate assembly using a high strength cyno-acrylic adhesive.
Circuit board removal is carried out by carefully breaking the bond and de-soldering the crystal wires and retractile cable connections (see diagram below).
©Ultrasound Technologies Ltd Page 10 of 32 PD1 / FT120 service manual Issue 3 January 2007
Ultrasound Technologies Ltd
3.0 Circuit Description
3.1 61Introduction
The circuit has been divided into functional blocks with each block being described separately, the blocks are as follows:
Audio Unit Power Supply
Battery Low Indicator
Ultrasound Transducer
Oscillator and Detector
Oscillator and Transmitter Amplifier
Receiver and Detector
Audio Amplifier
Controls and Indicators
Unit ON / OFF
Battery Low
3.2
62
Audio Unit Power Supply
The unit operates from a single 9-Volt dry battery. It is recommended that only alkaline cells be used. The audio circuit is fed directly from the battery. D1 provides circuit protection from battery reversal.
The unit is turned on by the membrane switch, which is mounted on the front of the unit. Closing the switch, grounds the CLK input of U1 pin 3, the output on pin 2 switches on Q2 allowing current to flow into the various circuits.
The unit will remain switched on for approximately 5 minutes (set by C4 and R5) after which it turns off automatically, unless the user forces the unit off by de-pressing the ON/OFF switch.
3.3
63
Battery Low Indicator
U3 is a voltage detector, when the voltage on pin 1 drops below the threshold set by R10 and R11 (7.3V), the output state changes forcing the LED (D2) to illuminate.
3.4
64
Ultrasound Transducer
The ultrasound transducer operates on the pulse wave Doppler principle. There are a number of transducer frequencies suitable for different applications, however the basic operating principles are identical.
Each transducer consists of a pair of piezo ceramic crystals, each crystal pair is arranged as a transmitter and receiver, the ultrasonic output beam is focussed through a lens or faceplate.
©Ultrasound Technologies Ltd Page 11 of 32 PD1 / FT120 service manual Issue 3 January 2007
Ultrasound Technologies Ltd
With the exception of the 2MHz Fetal transducer, the electronics are housed in the main audio unit.
The oscillator and detector are built up of four discrete sections. These are the master oscillator, transmitter amplifier, receiver amplifier and detector.
These operate to produce a pulse wave ultrasound signal that is passed to the transmitting crystal in the transducer.
The signal is then reflected from moving interfaces within the body to the receiver crystal in the transducer, amplified and then detected so the audio Doppler shift of that moving interface can be heard audibly and / or converted into a velocity signal.
3.5
65
Oscillator and Transmitter Amplifier
Field effect transistor Q2, with L1, C16, C17 and associated components form a Colpitts oscillator. This oscillator runs at a nominal frequency of 2, 5 or 8MHz producing a pulse wave of amplitude of approximately 5V Pk.
The signal is then fed to output transistor Q3 that drives the transmitter crystal in the transducer. The output power is set by VR1. The signal is fed to the transducer via a tuned transformer L2 (C23), the output impedance of which is set correctly to match the transducer crystal impedance. The output drive signal is nominally 1.5V Pk.
3.6
66
Receiver and Detector
The reflected Doppler signal is fed via a resonant transformer L4 (C25) to the gate of Q5, the drain of this FET connects to the source of Q4 to form a cascode amplifier the drain of which contains the resonant circuit L3,(C21).
From the drain of Q4 the amplitude complex of the received signal is detected by passing the signal through diode D2 with the high frequency signals being filtered by R12 and C15.
The raw low frequency complex is then amplified and filtered by U1 where its associated components form a band pass filter amplifier with a bandwidth of 150Hz to 1KHz for the obstetrics or 300Hz to 4KHz in the vascular transducer.
This signal is passed to the audio unit via the retractile cable.
3.7
67
Audio Amplifier
The audio signal is routed via the retractile cable to J4 pin 4 on the audio circuit board. The signal passes through the potentiometer VR1 to the audio amplifier U2, where it is amplified and output to the loudspeaker connected to J3.
3.8
68
Velocity Processor (PD1V Model Only)
The audio signal is fed from the band pass amplifier (U5) through the threshold control VR3 to the V to F converter (U3).
The conversion factor is 0.5V/ KHz with fine adjustment provided by VR2. The output of this circuit is then scaled by R4, 6 to a suitable voltage level for most ECG
machines and recorders.
©Ultrasound Technologies Ltd Page 12 of 32 PD1 / FT120 service manual Issue 3 January 2007
Ultrasound Technologies Ltd
4.0 9Test and Calibration
4.1
69
Introduction
The following sections detail tests to ensure that the unit is operating within specification. These tests may be performed in whole or part, however, if any repairs are carried out to the power supply circuits then it is recommended that the whole test / calibration procedure is undertaken.
4.2
70
Performance Checks
The following procedure is intended to provide a means of determining the functional status of the unit. It should be included as part of a preventative maintenance plan and should be performed on a regular basis at least once a year.
Ensure that a full capacity alkaline cell is fitted. Switch the unit ON and increase the volume to maximum. The battery low indicator will flash
ON momentarily and then extinguish. Place the transducer with ultrasound gel on the palm of the hand over the radial artery, a
clear pulse will be heard. Check the unit for signs of damage, particularly the transducer cable near the point of entry to
the unit and transducer body.
4.3
71
Equipment
Oscilloscope 2 channel 50 MHz bandwidth minimum resolution 5mV/cm DC Power Supply 0 – 30VDC @1A Digital Multimeter 4 Digit measuring 1mV, 1mA, 0.1ohm Frequency Counter 0 - 10MHz resolution at 2MHz 1KHz. Signal Generator 10Hz to 10MHz 1mV to 10V Sinewave
4.4
72
Audio Unit Set the power supply to 9VDC output and connect to battery fling leads. Switch ON the power supply.
Switch the board under test ON by shorting pins 1 and 2 of connector J1. Check the battery low LED flashes momentarily.
Check the current being drawn by the board under test is less than 50ma. Using the DVM, measure the DC voltage at the Drain of Q2 (junction of C8 & C11) this will be
between 8.80 and 8.60 V DC. Carefully decrease the power supply output, check and record the voltage reading at which
point the Battery low LED illuminates permanently. This will be between 7.0 and 7.5 V DC
©Ultrasound Technologies Ltd Page 13 of 32 PD1 / FT120 service manual Issue 3 January 2007
Ultrasound Technologies Ltd
Using a pair of tweezers, Introduce noise into pin 4 of J4. Rotate the volume control through its complete range of travel and observe the audio output from the speaker increases and decreases accordingly.
4.5
73
Transducer
87
Transmitter
Set the DC Power Supply to 9Vdc, solder the +ve lead to pin 1 of J1 and the –ve lead to J1 pin 2 on the board under test.
Switch ON the power supply, check the current being drawn by the board under test is less than 50ma.
Using the DVM, measure the DC voltage at the junction of C8 and R7, this will be between 5.2 and 5.8V DC.
Using the oscilloscope connected to the pad TxSig, adjust L1 so that the signal is set to
2.000MHz. Adjust L2 to set the output to 1.5V pk/pk.
88
Receiver
Connect the oscilloscope to the junction of D2 and R12, adjust both L4 and L3 to maximise the DC signal.
Move the oscilloscope to J1 pin 4, carefully stroke the transducer faceplate and observe a low frequency signal is displayed on the oscilloscope. The measured output will be a DC voltage of greater than 5.5V.
4.6
74
Velocity Processor
With the oscilloscope set to AC input, adjust the output signal from audio generator to a frequency of 1kHz and an amplitude of 15mV + 1mV pk-pk. Connect this signal to the junction of C22 and U5 pin 1.
Set frequency to 10KHz and attach oscilloscope to pin 7 IC3, adjust VR3 until the signal just changes state from 0V to 5V. The signal will be rough DC.
Using a multimeter, place the +ve probe to the junction of R18, R19 and C19, adjust VR4 until the until the DC level measures 5V.
©Ultrasound Technologies Ltd Page 14 of 32 PD1 / FT120 service manual Issue 3 January 2007
Ultrasound Technologies Ltd
5.0 10Fault finding guidelines
This section is an aid to trouble shooting and should be used in conjunction with the relevant circuit diagrams found at the rear of this manual. In each case the re calibration must be carried out after any repair.
The following tables list some of the symptoms and the relevant circuit areas. NOTE Do not, under any circumstances attempt to repair the instrument whilst it is connected to
a patient.
Symptom Suspect Circuit Check
Unit does not switch ON Remote switch on front panel J1 pin 1 goes low when
switch is depressed
U1 ON/OFF Control Output goes low when switch
is depressed Battery J1 pin 1 > 7V Battery Low LED on permanently
Battery condition J1 pin 1 > 7V
Replace battery U3 Comparator See test / calibration section Excessive battery drain U1 / Q2 ON/OFF Control Q2 must switch when switch
is depressed.
Change battery Ultrasound OK but no audio Transducer position Re position transducer Crackling operation of volume
control VR1
Replace volume control Broken conductor in cable Check audio signal on pin 3
of U2 Headphone socket SK2 Audio signal on J3 pin 1
Replace loudspeaker Unit unstable Ultrasound transducer Check output drive into
crystal is 1.5V pk/pk.
Check received signal in
transducer J1 pin 4, signal
should be a low frequency
audio signal approx. 2Vpk/pk.
Replace transducer.
©Ultrasound Technologies Ltd Page 15 of 32 PD1 / FT120 service manual Issue 3 January 2007
Ultrasound Technologies Ltd
6.0 11Parts list
The following section contains parts lists for the following items: -
PD1+ Audio PCB assembly
PD1 Audio PCB assembly
PD1 2MHz Probe assembly
PD1 Final assembly
PD1+ Final assembly
6.1
75
PD1+ Audio PCB Assembly
Device Value Device Type Circuit Reference Quantity
Capacitor 470uF Electrolytic Axial 10V Panasonic SU
ECEB1AU471
C10 1
Capacitor 100uF Min Electrolytic 16V
5mm
Nippon Chemi-con SRA / Samsung SSM
C2,9,11,12,13,18,19,21,27,28,29,30,32,33 ,34,35,37
17
Capacitor 10uF tant 20V case size A AVX / Kemet case size A C5,15,16 3
Capacitor 0u47F tant 25V case size A AVX / Kemet case size A C7 1
Capacitor 0805 100nF Mono Ceramic 0805 series Z5U dielectric C1,6,8,14,17 5
Capacitor 0805 47nF Mono Ceramic 0805 series X7R dielectric C26 1
Capacitor 0805 10nF Mono Ceramic 0805 series X7R dielectric C4,36 2
Capacitor 0805 1nF Mono Ceramic 0805 series COG dielectric C20,31 2
Capacitor 0805 22pF Mono Ceramic 0805 series COG dielectric C23,24 2
Capacitor 1206 1uF Z5U Ceramic 1206 series Z5U dielectric C3,22,25 3
IC Analogue CD4013D Philips/ Harris/National U2 1
IC Analogue ZM33064Z Zetex U4 1
IC Analogue LM386M-1 National U3 1
IC Analogue LM324M National U7 1
IC PSU ICL7660CBA Maxim U1 1
IC PSU ZR78L05G Zetex U11 1
IC Digital CD74HC4066D Philips/ Harris/National U9 1
IC Digital MAX1243BESA Maxim U6 1
IC Digital MAX202ECSA Maxim U10 1
IC Digital TSC87251G1-A16CB Temec U8 1
©Ultrasound Technologies Ltd Page 16 of 32 PD1 / FT120 service manual Issue 3 January 2007
Ultrasound Technologies Ltd
IC Digital ICM7211AMIM44 Harris U5 1
Diode 10BQ040 IR D1 1
Diode BAS16 D2,5 2
Crystal 12MHz 86SMX X1 1
Transistor RFD8P05SM Harris Q1 1
Resistors 1% 0805 10R Samsung/Phillips/ROHM R5,6 2
Resistors 1% 0805 100R Samsung/Phillips/ROHM R3 1
Resistors 1% 0805 2K2 Samsung/Phillips/ROHM R7 1
Resistors 1% 0805 4K7 Samsung/Phillips/ROHM R8 1
Resistors 1% 0805 10K Samsung/Phillips/ROHM R9,13,15,16,18,19,20,21,22,23,24,25,26,2
8,29
15
Resistors 1% 0805 22K Samsung/Phillips/ROHM R17 1
Resistors 1% 0805 47K Samsung/Phillips/ROHM R12 1
Resistors 1% 0805 68K Samsung/Phillips/ROHM R11,14 2
Resistors 1% 0805 100K Samsung/Phillips/ROHM R10,27,31 3
Resistors 1% 0805 330K Samsung/Phillips/ROHM R30 1
Resistors 1% 0805 1M Samsung/Phillips/ROHM R1,2,4 3
LCD module PD1D107 1
Volume Control 10K Log Tsubuya 161P-N2, A10K VR1 1
Volume Control Knob Tsubuya 1KF-4B Black Knob VR1 knob 1
Connector Jack socket Hosiden HSJ 1501-010010 P1,2 2
Connector PP3 battery connector 1
Connector Molex 3 way ZIF Molex 39-51-4032 J2 1
PCB PD1 plus audio board PD1D109 1
©Ultrasound Technologies Ltd Page 17 of 32 PD1 / FT120 service manual Issue 3 January 2007
Ultrasound Technologies Ltd
6.2
76
PD1 Audio PCB Assembly
Device
12
Value
Device Type Circuit Reference
Quantity
Capacitor 470uF Electrolytic Axial 10V Panasonic SU
ECEB1AU471
C7 1
Capacitor 100uF Min Electrolytic 16V
5mm
Nippon Chemi-con SRA / Samsung SSM
C1,8,9,11,12 5
Capacitor 10uF tant 20V AVX / Kemet case size A C13,14 2
Capacitor 0u47F tant 25V AVX / Kemet case size A C4,5 2
Capacitor 1206 100nF Mono Ceramic 1206 series Z5U dielectric C6,8,9,10,15 5
Capacitor 1206 10nF Mono Ceramic 1206 series X7R dielectric C3 1
Capacitor 1206 1uF Z5U Ceramic 1206 series Z5U dielectric C2 1
IC Analogue CD4013D Philips/ Harris/National U1 1
IC Analogue ZM33064Z Zetex U3 1
IC Analogue LM386M-1 National U2 1
Diodes 10BQ040 IR D1 1
Diode LED Yellow Temic TLLY4401 D2 1
Transistor RFD8P05SM Harris Q2 1
Transistor BSS138 Fairchild Semiconductor Q1 1
Resistors 1% 1206 10R Samsung/Phillips/ROHM R8,9 2
Resistors 1% 1206 1K Samsung/Phillips/ROHM R2 1
Resistors 1% 1206 2K2 Samsung/Phillips/ROHM R10 1
Resistors 1% 1206 4K7 Samsung/Phillips/ROHM R11,12 2
Resistors 1% 1206 100K Samsung/Phillips/ROHM R4 1
Resistors 1% 1206 1M Samsung/Phillips/ROHM R3,5,6,7 4
Resistors 1% 1206 3M3 Samsung/Phillips/ROHM R1 1
Volume Control 10K Log Tsubaya 161P-N2, A10K VR1 1
Volume Control Knob Tsubuya 1KF-4B Black
Knob
VR1 knob 1
Connector Jack socket Hosiden HSJ 1501-010010 P1,2 2
Connector PP3 battery connector 1
Connector Molex 3 way ZIF Molex 39-51-3032 J2 1
PCB PD1D103 PD1 Audio PCB 1
©Ultrasound Technologies Ltd Page 18 of 32 PD1 / FT120 service manual Issue 3 January 2007
Ultrasound Technologies Ltd
6.3
77
PD1 2MHz Transducer Assembly
Description
85
Value
Device Type Circuit Reference
Quantity
Capacitor 100uF Min Electrolytic 16V
5mm
Nippon Chemi-con SRA / Samsung SSM
C5,8,12 3
Capacitor 1u0F tant 35V case size A C4,11 2
Capacitor 100nF Mono Ceramic 0805 series Z5U dielectric C1,2,3,7,10,26,22,24 8
Capacitor 10nF Mono Ceramic 0805 series X7R dielectric C18 1
Capacitor 22nF Mono Ceramic 0805 series X7R dielectric C6,13 2
Capacitor 1n0F COG Ceramic 0805 series COG dielectric C9,15,23 3
Capacitor 220pF COG Ceramic 0805 series COG dielectric C19,20,25 3
Capacitor 330pF COG Ceramic 0805 series COG dielectric C17 1
Capacitor 470pF COG Ceramic 0805 series COG dielectric C21 1
Capacitor 680pF COG Ceramic 0805 series COG dielectric C16 1
IC Analogue TL072CD SGS / TI U1 1
Diodes BZX84C6V2 D1 1
Diodes BAT54 D2 1
Transistor MMBF4416LT1 Motorola Q2,4,5 3
Transistor BC848B Q1,3 2
Inductor Toco 0838 L4 1
Inductor Toco 0842 L3 1
Inductor Toco 0841 L1 1
Inductor Toco 0876 L2 1
Resistors 1% 0805 10R Samsung/Phillips/ROHM R17,16 2
Resistors 1% 0805 22R Samsung/Phillips/ROHM R7 1
Resistors 1% 0805 100R Samsung/Phillips/ROHM VR1 1
Resistors 1% 0805 1K Samsung/Phillips/ROHM R3,8 2
Resistors 1% 0805 1K5 Samsung/Phillips/ROHM R15 1
Resistors 1% 0805 2K2 Samsung/Phillips/ROHM R1,2,11 3
Resistors 1% 0805 27K Samsung/Phillips/ROHM R14 1
Resistors 1% 0805 100K Samsung/Phillips/ROHM R4,5,6,9,10,12,18,19 8
PCB PD1D110 PD1 probe board 1
Piezo Ceramic Crystal 2MHz Resonant Frequency 2
©Ultrasound Technologies Ltd Page 19 of 32 PD1 / FT120 service manual Issue 3 January 2007
Loading...
+ 52 hidden pages