The DMG100 should rarely require recalibration, though it may be necessary to know that y
our gauge
is calibrated properly for full accuracy. The Calibration Test mode assures you that the DMG100 is calibrated to factory specifications. Test the calibration as follows:
1.Turn the DMG100 power off.
2.Important: Expose the DMG100 to atmospheric pressure.
3.Press and hold (do not release) the key for about 5 seconds.
4.The display will show if the instrument is calibrated properly.
5.The display will show if the instrument requires calibration. Please see the Calibration section below.
Calibration
If the Calibration Test indicates recalibration is required, the gauge may be recalibrated. Unlike any other vacuum gauge, the DMG100 can be easily recalibrated to factory specifications without any special equipment, with the following procedure:
1.For best results, clean the sensor with alcohol prior to calibration. Ensure the sensor is completely dry before proceeding.
2.Turn the DMG100 poweroff. If necessary,installafully charged battery into thegauge.
3.Place the dust cap over the vacuum fitting.
4.Hold while pressing . As soon as the power turns on, release and and then
press quickly at least three times. The CALindicator should illuminate, and should show on the Main Numeric Display. If not, turn the power off again and repeat.
5.
Place the DMG100 in a clear Ziploc (resealable zipper storage) bag, press out any extra air, and
seal.
6.Place the bagged DMG100 into a freezer with a temperature of less than –5°C(23°F).
7.Allow the DMG100 to cool to below-2°C (28.4°F). At this point, the alarm will sound and the
display will change to .
8.Remove from freezer and press any key to silence the alarm.
9.Place the DMG100 undisturbed in an area with a room temperature of at least 23°C (73.4°
F) but
no greater than 30°C (86°F).
10.Allow the DMG100 to warm to 20°C(68°F). At this point, the alarm will sound, and will show on the display.
11.Press any key to silence the alarm. The DMG100 is now calibrated to factory specifications.
Note:For accurate calibration, it is necessary to allow the DMG100 to warm slowly. Attempting to accelerate the warming by using a heat source will not provide satisfactory results. During the cooling/warming process, the temperature will be indicated on the Alternate Numeric Display in degrees Celsius. The calibration process may be canceled at any time by turning off the DMG100. The previous calibration will be unchanged.
LCD Display
ItemFunction
AMain Numeric Display
BAlternate Numeric Display
C“Analog” Vacuum Level Bar Graph
Backlight Indicator. Flashing: Temporary, Solid: Always On
Sound Indicator
Battery Level Indicator
Oil Sensor Indicator
Calibration Mode Indicator
Indicates that Temperature is Displayed on the Alternate Numeric Display (during
calibration)
Indicates degrees Celsius on the Alternate Numeric Display (during calibration)
As the battery is depleted, the number of bars displayed decreases, until no bars are left. At this
which the sensor and ba
Approximately every 35 seconds, the gauge will automatically check the pressure and exit Slee
After 1 hour in “Sleep Mode” the DMG100 will automatically turn itself off to further conserve battery
with oil to the extent that it cannot function properly at all, the message
6
Range and Resolution
The DMG100 has a broad vacuum pressure measurement range and very high resolution according to the table below:
Vacuum
RangeResolution
Pascals (Pa)
0
— 3,333.10.1 Pa
Millibar (mBar)
0
— 33.3310.001 mBar
Millitorr (mTorr)
0
— 25,0001 mTorr
mmHg (mmHg)
0
— 25.0000.001 mmHg
Microns (Microns)
0
— 25,0001 Micron
“Analog” Vacuum Level Bar Graph
The DMG100 also displays a Vacuum Level Bar Graph allowing for a quick visual determination of the vacuum level achieved. Each bar corresponds to a vacuum pressure range depending on units according to the following table:
The battery level indicator shows the relative strength of the battery. Four bars indicate full power.
point, it is necessary to replace the batteries (refer to the Battery Installation section above). If the
battery power drops to the point where the DMG100 can no longer function accurately, the alarm will
beep 10 times and the power will turn off automatically. When the battery is replaced, and the
DMG100 is turned back on, the DMG100 will resume operation with all previous settings intact.
Hi-Pressure Indication
If the sensed vacuum pressure exceeds 25,000 Microns (3333.1 Pa, 33.331 mBar, 25,000 mTorr, or
25.000 mmHg), the display will show
.
Sleep Mode
Unlike other Digital Micron Gauges, the DMG100 helps to conserve battery life through advanced
power management. After 5 minutes of displaying
cklight are turned off, thus reducing the load on the battery and extending the
battery life up to 300 hours or more. While in this mode, the display will show
Mode if the pressure is less than 25,000 Microns. The gauge can be manually brought out of Sleep
Mode by pressing any key on the keypad. While in Sleep, the gauge will beep twice every 5 minutes
to remind you that it is on. The DMG100 will not sleep in the Calibration Mode (see the Calibration
section on pg. 8).
, the DMG100 will enter “Sleep Mode” in
Auto Power-Off
power.
.
p
Oil Sensor
It is necessary to prevent oil from being drawn into the Vacuum Sensor. For HVAC/R service, always
recover refrigerant prior to attaching gauge. If possible, always close the blank-off valve on the
vacuum pump prior to turning the vacuum pump off. Over time, oil vapor and other materials may
contaminate the sensor. The DMG100 has a built in Oil Sensor that will detect this condition. If the
OIL indicator on the display illuminates, this is an indication that the vacuum sensor has been
contaminated and is no longer functioning accurately. If the sensor becomes completely saturated
display. Prior to further use of the DMG100, clean the sensor as describe in the “Cleaning the
Vacuum Sensor” on pg. 7.
Units
Change the displayed units by pressing the key on the keypad. The appropriate units indicator
on the LCD display will show Pa, mBar, mTorr, mmHg, or Microns depending on the units set.
The DMG100 has a broad vacuum pressure measurement range and very high resolution according
to the table below:
Vacuum
RangeResolution
Pascals (Pa)
— 3,333.10.1 Pa
0
Millibar (mBar)
Millitorr (mTorr)
mmHg (mmHg)
Microns (Microns)
— 33.3310.001 mBar
0
— 25,0001 mTorr
0
— 25.0000.001 mmHg
0
— 25,0001 Micron
0
“Analog” Vacuum Level Bar Graph
The DMG100 also displays a Vacuum Level Bar Graph allowing for a quick visual determination of
the vacuum level achieved. Each bar corresponds to a vacuum pressure range depending on units
according to the following table:
1.Install the battery as described in the “Battery Installation” section below.
2.Turn the power on by pressing . The display will show
to indicate pressure greater
than 25,000 Microns.
3.Select the desired units by repeatedly pressing .
4.Attach the DMG100 to the system to be evacuated with a high quality vacuum hose. Start the vacuum pump. Read the vacuum level from the Main Display.
5.Turn the power off by pressing and holding.
Battery Installation & Replacement
1.Remove the battery cover from the rear of the DMG100 by compressing tab at the base of the battery cover.
2.If necessary, remove and detach old battery from battery clip.
3.Attach battery clip to new battery and insert into battery compartment. Replace battery cover by aligning tab and snapping back into place.
IMPORTANT: TO PREVENT DAMAGE FROM LEAKING BATTERIES, DO NOT LEAVE A DEAD
BATTERY INSIDE THE DMG100. REMOVE BATTERY IF THE DMG100 IS NOT TO BE USED FOR
AN EXTENDED PERIOD OF TIME.
ItemFunction
Pressto Turn Power On, Press and Hold to Turn PowerOff
Press to Change Display Units
Press and Release to Activate Backlight
Press and Hold to Mute/Un-mute Sound
Backlight
To activate the Backlight temporarily, press once. The indicator on the display will flash, and
the backlight will turn off automatically after 1 minute. To activate the Backlight permanently, press
again. The indicator on the display will show solid. Turn off the backlight by pressing
repeatedly until the indicator turns off. During Sleep Mode, the backlight will turn off to help
conserve battery power, but will turn on again automatically upon resumption of normal operation if it
-on mode.
Sound
The DMG100 has an internal speaker that will emit a beep for each valid key press, and also
functions as an alarm in the Calibration mode (see the Calibration section on pg. 8). It will also beep
every 5 minutes in Sleep Mode. For silent operation, the sound can be muted by pressing and
holding the
the LCD display indicates that the sound is on (not muted).
key. The Calibration mode alarm is not affected by the mute status. The indicator on
Swivel Hook
The DMG100’s removable stainless steel swivel hook enables hanging of the gauge and will allow it
to swivel freely in any direction. The gauge may be operated with or without the hook attached.
When fastening the hook to the gauge, turn it finger tight only. Use of a tool to tighten the hook may
result in damage to the DMG100 case.
Maintenance
The DMG100 should provide many years of service with no maintenance required. When not in use,
the dust cap should remain in place over the sensor port. Clean the plastic enclosure with a damp
(not wet) rag. Mild detergent is acceptable, but use no solvents. Take care not to expose the
vacuum sensor to oil. If the Oil Sensor (see pg. 5) indicates a contaminated sensor, follow the
Sensor Cleaning Procedure below.
Cleaning the Vacuum Sensor
If the vacuum sensor becomes contaminated with oil (as indicated by the Oil Sensor), carefully follow
this procedure:
1.Turn the DMG100 power off.
2.Shake the gauge to remove any large quantities of oil from the sensor.
3.Apply a few drops of rubbing alcohol inside the sensor vacuum port. (DO NOT INSERT ANY OBJECT INTO THE PORT, AS THIS WILL PERMANENTLY DAMAGE THE SENSOR).
4.Place your finger over the port and shake for a few moments.
5.Remove your finger and shake out the alcohol.
6.Repeat steps (3) – (5) at least three times.
7.Allow the sensor to air dry over at least an hour, or pull a vacuum on the sensor to dry it more
quickly.
8.Replace the battery and turn on the gauge. The Oil Indicator should be off. If it is still on, repeat
the cleaning procedure.
9.If full accuracy is desired, perform a calibration cycle (see pg. 8)..
Note: It is important to remove all alcohol vapors from the sensor, either through air-drying or via
vacuum. Any remaining vapors will cause an incorrect vacuum reading.
Place the DMG100 in a clear Ziploc (resealable zipper storage) bag, press out any extra air, and
F) but
Calibration Test
The DMG100 should rarely require recalibration, though it may be necessary to know that y
is calibrated properly for full accuracy. The Calibration Test mode assures you that the DMG100 is
calibrated to factory specifications. Test the calibration as follows:
1.Turn the DMG100 power off.
2.Important: Expose the DMG100 to atmospheric pressure.
3.Press and hold (do not release) the
4.The display will show
5.The display will show
Calibration section below.
key for about 5 seconds.
if the instrument is calibrated properly.
if the instrument requires calibration. Please see the
Calibration
If the Calibration Test indicates recalibration is required, the gauge may be recalibrated. Unlike any
other vacuum gauge, the DMG100 can be easily recalibrated to factory specifications without any
special equipment, with the following procedure:
1.For best results, clean the sensor with alcohol prior to calibration. Ensure the sensor is
completely dry before proceeding.
2.Turn the DMG100 power off. If necessary, install a fully charged battery into the gauge.
3.Place the dust cap over the vacuum fitting.
4.Hold
press quickly at least three times. The CAL indicator should illuminate, and
should show on the Main Numeric Display. If not, turn the power off again and repeat.
5.
seal.
6.Place the bagged DMG100 into a freezer with a temperature of less than –5°C (23°F).
7.Allow the DMG100 to cool to below -2°C (28.4°F). At this point, the alarm will sound and the
display will change to
8.Remove from freezer and press any key to silence the alarm.
9.Place the DMG100 undisturbed in an area with a room temperature of at least 23°C (73.4°
no greater than 30°C (86°F).
10. Allow the DMG100 to warm to 20°C (68°F). At this point, the alarm will sound, and
will show on the display.
11. Press any key to silence the alarm. The DMG100 is now calibrated to factory specifications.
while pressing . As soon as the power turns on, release and and then
.
Note: For accurate calibration, it is necessary to allow the DMG100 to warm slowly. Attempting to
accelerate the warming by using a heat source will not provide satisfactory results. During the
cooling/warming process, the temperature will be indicated on the Alternate Numeric Display in
degrees Celsius. The calibration process may be canceled at any time by turning off the DMG100.
The previous calibration will be unchanged.
A standard alkaline battey may not provide acceptable battery life at temperatures below 0
2
Parts Diagram
Restoring Factory Calibration
The original factory calibration can be restored at any time by the following procedure:
1.Turn the DMG100 power off.
2.Hold
press quickly at least three times. The CAL indicator should illuminate, and
should show on the Main Numeric Display. If not, turn the power off again and repeat.
3.Press
4.The gauge will beep 5 times and return to normal operation. The DMG100 is now reset to the
original factory calibration.
while pressing . As soon as the power turns on, release and and then
five times followed by .
Troubleshooting
Under certain conditions, the display may read or . Please use the table
below to determine and fix the problem:
DisplayModePossible ProblemSolution
Sensor ContaminatedClean the Vacuum Sensor
Normal
Operation
CalibrationSensor Contaminated
Calibration
Ambient Temperature too
Low
Gauged Warmed too
Quickly
Gauge Disturbed During
Calibration
Turn the BluVac off, warm the
vacuum port with your hand,
and turn the BluVac back on.
Clean the Vacuum Sensor
and Restart Calibration
Restart Calibration. Allow the
Gauge to Warm Slowly
Restart Calibration. Leave
the Gauge Undisturbed
During Warming Phase.
Low Temperature Operation
The DMG100 can operate accurately at temperatures as low as 10°F (-12°C). While operating below
freezing (32°F/0°C), the display update rate will slow from 3.5 readings every second to one reading
every two seconds.
For low temperature operation, a 9V Lithium battery is recommended.