UEi Test Instruments DM505 User manual

1000V Digital Multimeter
INSTRUCTION MANUAL
ENGLISH
DM505
DM505
Apo
V
600V
CAT III
1000V
CAT II
RANGE
SEL
A
10A MAX
FUSED
MIN/MAX
mAμA
FUSED
400mA MAX
HOLD
FAST
mA
CAT III
CAT II 1000V
A
MAX
600V
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
TABLE OF CONTENTS
FUNCTIONS
FEATURES .................................................................................................................................................... 3
GENERAL SPECIFICATIONS
CATEGORY DEFINITIONS
IMPORTANT SAFETY WARNINGS
OVERVIEW
SYMBOLS .................................................................................................................................................... 6
OPERATING INSTRUCTIONS
Resistance: <40MΩ ........................................................................................................................... 7
Continuity ........................................................................................................................................... 7
Diode ................................................................................................................................................... 8
Capacitance ......................................................................................................................................... 8
AC/DC Microamps: 6000μA .............................................................................................................. 9
AC/DC Milliamps: 400mA ................................................................................................................. 9
AC/DC Amps: <10A ......................................................................................................................... 10
Test Lead Notes ................................................................................................................................ 10
Battery Replacement ....................................................................................................................... 11
FCC/IC INFORMATION ............................................................................................................................. 11
WARRANTY ................................................................................................................................................ 12
DISPOSAL
CLEANING
STORAGE
................................................................................................................................................. 3
.................................................................................................................... 3
.......................................................................................................................... 3
......................................................................................................... 4
.................................................................................................................................................. 5
AC/DC Voltage: <1000V AC/DC ....................................................................................................6
.................................................................................................................................................. 12
.................................................................................................................................................. 12
.................................................................................................................................................. 12
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
FUNCTIONS
• 1000V AC/DC
• 10A AC/DC
• Resistance 40MΩ
• Diode test
FEATURES
• Auto/Manual ranging
• Auto power off
• Min/Max
• Hold
• 1 ms Fast response Min/Max
• Low battery indicator
GENERAL SPECIFICATIONS
• Operating Temperature: 32˚ to 122˚F (0˚ to 50˚C)
• Storage Temperature: -44˚ to 122˚F (-20˚ to 50˚C)
• Operating Humidity: <75% max.
• Operating Altitude: 6561 ft (2000m)
• Display: 6,000 count
• Backlight: Yes
• Over-range: “OL” is displayed
• Dimensions: 6.49” x 3.3” x 2.17”
• Item Weight: 15.0 oz
• Calibration: Recommended annually
• CAT Rating: CAT III 600V/CAT II 1000V
• Certifications: cELTus UL 61010-1:2012 3rd
CE EN 61010-1:2010 3rd, IEC61010-2-033:2012, Ed.1, EN 61326-1:2013 FCC, RoHS Compliant, TOV protection, IP 42, 6’ Drop protection
• Battery Type: (AA) 2
• Test Leads: CAT III test leads
• Accuracy: ± (% of reading + # of least significant digits)
• Audible continuity
• Capacitance 9999μF
• Milliamps/Microamps
• Test lead holders
• Kick stand
• High resolution backlit display
• Auto selection
• Fused test lead inputs
• Rubber boot
CATEGORY DEFINITIONS
Measurement Category Short-Circuit (typical) kA
II < 10
III < 50 Mains distributions parts of the building
IV > 50 Source of the mains installation in the building
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
a
Location in the building installation
Circuits connected to mains socket outlets and similar points in the MAINS installation
3
IMPORTANT SAFETY WARNINGS
WARNING
Read entire Safety Notes section regarding potential hazards and proper instructions before using this meter. In this manual the word “WARNING” is used to indicate conditions or actions that may pose physical hazards to the user. The word “CAUTION” is used to indicate conditions or actions that may damage this instrument.
WARNING
To ensure safe operation and service of the tester, follow these instructions. Failure to observe these warnings can result in severe injury or death.
WARNING
• Before each use, verify meter operation by measuring a known voltage or current.
• Never use the meter on a circuit with voltages that exceed the category based rating of this meter.
• Do not use this meter during electrical storms or in wet weather.
• Do not use the meter or test leads if they appear damaged.
• Ensure meter leads are fully seated and keep fingers away from the metal probe contact when making measurements. Always grip the leads behind the finger guards molded into the probe.
• Do not open the meter to replace batteries while the probes are connected.
• Use caution when working with voltages above 60 DC or 25 AC RMS. Such voltages pose shock hazards.
• To avoid false readings that can lead to electrical shock, replace batteries if a low battery indicator appears.
• Unless measuring voltage or current, shut off and lockout power before measuring resistance or capacitance.
• Always adhere to national and local safety codes. Use proper personal protective equipment (PPE) to prevent shock and arc blast injury where hazardous live conductors are exposed.
• Always turn off power to a circuit or assembly under test before cutting, unsoldering or breaking the current path. Even small amounts of current can be dangerous.
• Always disconnect the live test lead before disconnecting the common test lead from the circuit.
• In the event of electrical shock, ALWAYS bring the victim to the emergency room for evaluation, regardless of victim’s apparent recovery. Electrical shock can cause unstable heart rhythms that may need medical attention.
• If any of the following occur during testing, turn off the power source to the circuit being tested: arching, flame, smoke, extreme heat, smell of burning materials or discoloration or melting of components.
WARNING
Higher voltages and currents require greater awareness of physical safety hazards. Before connecting the test leads; turn off power to the circuit under test, set meter to the desired function and range; connect the test leads to the meter first, then connect to the circuit under test. Reapply power. If an erroneous reading is observed, disconnect power immediately and recheck all settings and connections.
WARNING
This meter is designed for trade professionals who are familiar with the hazards of their trade. Observe all recommended safety procedures that include proper lockout utilization and use of personal protective equipment that includes safety glasses, gloves and flame resistant clothing.
4
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
OVERVIEW
DM505
A
Apo
V
B
C D E
F
G
A. Apo: Auto power off after 30 minutes of use. Press
and hold the HOLD button while turning the meter on to disable Apo.
B. Digital Backlit Display C. Range Button:
• Press to change from auto ranging to manual ranging.
• Press repeatedly to select proper range.
• Press and hold to return to auto range (AT will be displayed on screen).
D. Min/Max Button:
• Press to enter MAX/MIN mode.
• In the V, μA, mA or Amps functions, either select AC/ DC or change to manual ranging before pressing this button to enter MAX/MIN mode.
• Press repeatedly to alternate between Maximum and Minimum readings.
• Press and hold to return to live readings E. Select Button: Press to select AC,DC or Auto Selection in the following functions: Voltage, μA, mA, Amps. F. Function Dial: Turns on meter and is used to select the function.
RANGE
SEL
A
10A MAX
FUSED
MIN/MAX
mAμA
HOLD
FAST
μA
mA
A
MAX CAT III
400mA
600V
MAX
CAT II
FUSED
1000V
H
I
G. Category Max Indicator: Maximum CAT Rating for
fused input jacks.
• Multifunction input port used for measuring: AC or DC volts, resistance, continuity, diode, Amps, mA, μA.
• Use CATIII test leads or higher.
H. Hold/Fast Button:
• Press to hold the reading on the display. Press again to return to live reading.
• Press to enter Fast MAX/MIN mode in MAX/MIN mode.
• Press again to return to normal MAX/MIN mode. I. Back Light Button: Press to on back light. Press again to turn off. Back light duration is 1 minute.
J. Recess for magnetic hanger K. Test Lead Holders L. Protective Rubber Boot M. Battery Cover (under protective rubber boot) N. Kick Stand O. Serial Number (under kick stand)
J
K
L M
N
O
5
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
SYMBOLS
Negative
Overload: Range Exceeded
Hold/Capture Value
Voltage
A
C/DC Voltage or Current Low Battery
Auto power off Active
Minimum measured value displayed
A
Amps Diode
Maximum measured value displayed
Microfarad
μF
Microamps mA milliamps Ground
nano Farads Ohms/Resistance Dangerous Levels
nF
Continuity Warning or Caution
Fuse Auto-ranging
Auto
Auto Selection
High Voltage Indication
AC/DC Voltage: <1000V AC/DC
DM505
Apo
Apo
AT
RANGE
SEL
A
DM505
V
V
MIN/MAX
HOLD
FAST
μA
mA
A
mAμA
MAX CAT III
400mA
600V
MAX
10A MAX
FUSED
CAT II
FUSED
1000V
WARNING
• Use CATIII rated Test leads or higher.
• Do not attempt to measure more than 1000V AC/DC.
• Do not exceed 25 volts AC or 60 volts DC – RMS at either the common or multifunction input ports as measured from
earth ground.
Features:
RANGE
MIN/MAX
HOLD
SEL
AC VOLTS
Ranges Accuracy Resolution Overload Protection
600 mV to 1000V ±(1.0% +3 dgts) 0.1 mV to 1V 1000V
Minimum sensitivity: 0.5V AC (auto selection mode), Frequency bandwidth: 400Hz
DC VOLTS
Ranges Accuracy Resolution Overload Protection
600 mV to 1000V ±(0.5% +3 dgts) 0.1 mV to 1V 1000V
Minimum sensitivity: 0.5V DC (auto selection mode)
6
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Resistance: <40MΩ
A
μA
mA
A
μA
mA
DM505
Apo
Apo
AT
RANGE
SEL
A
DM505
V
MIN/MAX
HOLD
FAST
μA
mA
A
Features:
mAμA
MAX CAT III
400mA
600V
MAX
10A MAX
FUSED
CAT II
FUSED
1000V
RANGE
WARNING
MIN/MAX
HOLD
• Do not measure resistance on a live circuit.
Ranges Accuracy Resolution Overload Protection
600Ω to 6MΩ ±(0.8% +5 dgts) 0.1Ω to 0.001MΩ 40MΩ ±(1.5% +5 dgts) 0.01MΩ
600V
Continuity
DM505
Apo
Apo
DM505
V
RANGE
MIN/MAX
HOLD
SEL
A
FAST
μA
mA
A
Features:
mAμA
MAX CAT III
400mA
600V
MAX
10A MAX
FUSED
CAT II
FUSED
1000V
• Buzzer sounds at less than 40Ω.
MIN/MAX
WARNING
HOLD
• Do not measure resistance on a live circuit.
Open Circuit Voltage Audible Threshold Overload Protection
Approx.: <1.0V Approx.: 40Ω 600V
7
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Diode
A
μA
mA
A
μA
mA
GOOD DIODE
BAD DIODE
Open Diode
Displays "OL" Both directions
Forward Bias
DM505
DM505
Apo
Apo
V
V
Displays
RANGE
MIN/MAX
approx. voltage drop
HOLD
FAST
SEL
μA
mA
A
• Forward voltage drop if forward biased.
mAμA
A
MAX CAT III
400mA
600V
MAX
CAT II
10A
FUSED
1000V
MAX
FUSED
• “OL” if reverse biased.
Features:
Open Circuit Voltage Overload Protection
Approx.: <3.0V DC 600V
Capacitance
DM505
DM505
Apo
Apo
AT
RANGE
MIN/MAX
SEL
Apo
or
V
Apo
'0' Both directions (shorted)
HOLD
HOLD
FAST
μA
V
nF
mA
A
MIN/MAX
V
mAμA
Features:
RANGE
MIN/MAX
A
HOLD
MAX CAT III
400mA
600V
MAX
CAT II
10A
FUSED
1000V
MAX
FUSED
AC VOLTS
Ranges Accuracy Resolution Overload Protection
10nF to 9999μF ±(3.0% +5 dgts) 0.01nF to 1μF 600V
WARNING
To avoid damaging the meter or euipment under test, safely discharge Capacitors before measuring capacitance. Large value capacitors should be discharged through an appropriate Resistance load. Use the DC Voltage function to confirm the capacitor discharge.
8
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
AC/DC Microamps: 6000μA AC/DC Milliamps: 400mA
A
μA
mA
A
μA
mA
Features:
Apo
Apo
AT
RANGE
SEL
A
10A MAX
FUSED
RANGE
MIN/MAX
mAμA
DM505
DM505
HOLD
FAST
μA
400mA MAX FUSED
MIN/MAX
Apo
V
μ A
mA
A
MAX CAT III 600V CAT II 1000V
HOLD
μA
SEL
Apo
AT
RANGE
A
SEL
DM505
DM505
m
V
A
MIN/MAX
HOLD
FAST
μA
mA
A
mAμA
MAX
CAT III
400mA
600V
MAX
10A MAX
FUSED
CAT II
FUSED
1000V
mA
ACμA
Ranges Accuracy Resolution Overload Protection
600μA to 6000μA ±(1.2% +5 dgts) 0.1μA to 1μA 600mA /1000V Fast Fuse Minimum sensitivity: 50μA AC (auto selection mode only)
DCμA
Ranges Accuracy Resolution Overload Protection
600μA to 6000μA ±(1.0% +3 dgts) 0.1μA to 1μA 600mA /1000V Fast Fuse Minimum sensitivity: 50μA DC (auto selection mode only)
ACmA
Ranges Accuracy Resolution Overload Protection
60mA to 400mA ±(1.2% +5 dgts) 0.01mA to 0.1mA 600mA /1000V Fast Fuse Minimum sensitivity: 5mA AC (auto selection mode only)
DCmA
Ranges Accuracy Resolution Overload Protection
60mA to 400mA ±(1.0% +3 dgts) 0.01mA to 0.1mA 600mA /1000V Fast Fuse Minimum sensitivity: 5mA DC (auto selection mode only)
9
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
AC/DC Amps: <10A
A
μA
mA
DM505
DM505
Apo
Apo
AT
RANGE
SEL
MIN/MAX
V
A
HOLD
FAST
μA
mA
A
WARNING
mAμA
A
MAX CAT III
400mA
600V
MAX
CAT II
10A
FUSED
1000V
MAX
FUSED
• Do not attempt to measure more than 10A AC.
A
Features:
RANGE
MIN/MAX
HOLD
SEL
AC AMPS
Ranges Accuracy Resolution Overload Protection
6A to 10A ±(1.5% +5 dgts) 0.001A to 0.01A 11A/1000V Fast fuse Minimum sensitivity: 500mA AC (auto selection mode only)
CAUTION
: 20A overload for 30 seconds max.
DC AMPS
Ranges Accuracy Resolution Overload Protection
6A to 10A ±(1.2% +5 dgts) 0.001A to 0.01A 11A/1000V Fast fuse
Minimum sensitivity: 500mA DC (auto selection mode only)
CAUTION
: 20A overload for 30 seconds max.
Test Lead Notes
CATIII 1000V Measurement Locations
4mm
• Ensure the test lead shield is pressed firmly in place. Failure to use the CATIV shield increases arc-flash risk.
CATII 1000V Measurement Locations
18mm
• CAT IV shields may be removed for CAT II locations. This will allow testing on recessed conductors such as standard wall outlets. Take care not to lose the shields.
WARNING: Test lead category protections apply only to test leads and should not be confused with the meter’s
specific CAT rating. Observe the maximum category protection indicated on the meter the test leads are plugged into.
10
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Battery Replacement
• When the batteries are too low for safe operation, the Low Battery indicator will display.
AA
AA
S/N:
S/N:
FCC/IC INFORMATION
NOTE: This device complies with Part 15 of the FCC Rules and CAN ICES-3(A).
Operation is subject to the following two conditions: (1) this device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operations.
INFORMATION TO THE USER
This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules. These limits are designedto provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:
• Reorient or relocate the receiving antenna.
• Increase the separation between the equipment and receiver.
• Connect the equipment into an outlet on a circuit different from that to w hich the receiver is connected.
• Consult the dealer or an experienced radio/TV technician for help.
WARNING Any changes or modifications not expressly approved by the manufacturer, could void the user’s authority to
operate equipment.
11
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
WARRANTY
The DM505 is warranted to be free from defects in materials and workmanship for a period of 2 years from the date of purchase. If within the warranty period your instrument should become inoperative from such defects, the unit will be repaired or replaced at UEi’s option. This warranty covers normal use and does not cover damage which occurs in shipment or failure which results from alteration, tampering, accident, misuse, abuse, neglect or improper maintenance. Batteries and consequential damage resulting from failed batteries are not covered by warranty.
Any implied warranties, including but not limited to implied warranties of merchantability and purpose, are limited to the express warranty. UEi shall not be liable for loss of use of the instrument or other incidental or consequential damages, expenses, or economic loss, or for any claim or claims for such damage, expenses or economic loss.
A purchase receipt or other proof of original purchase date will be required before warranty repairs will be rendered. Instruments out of warranty will be repaired (when repairable) for a service charge
This warranty gives you specific legal rights. You may also have other rights, which vary from state to state.
fitness for a particular
DISPOSAL
CAUTION:
correct disposal.
This symbol indicates that equipment and its accessories shall be subject to separate collection and
CLEANING
Periodically clean your meters’ case using a damp cloth. DO NOT use abrasive, flammable liquids, cleaning solvents, or strong detergents as they may damage the finish, impair safety, or affect the reliability of the structural components.
STORAGE
Remove the batteries when instrument is not in use for a prolonged period of time. Do not expose to high temperatures or humidity. After a period of storage in extreme conditions exceeding the limits mentioned in the General Specifications section, allow the instrument to return to normal operating conditions before using it.
Copyright © 2018 Kane USA Inc. All Rights Reserved. 17488 0118
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Loading...