WARNING :
uses, to be used as personal safety device or medical applications. Refer to IMPORTANT
SAFETY WARNINGS section for more details.
NOTE: This device complies with Part 15 of the FCC Rules and CAN ICES-3(A). Operation is
subject to the following two conditions: (1) this device may not cause harmful interference.
And (2) this device must accept any interference received, including interference that may
cause undesired operations.
INFORMATION TO THE USER
This equipment has been tested to comply with limits for a Class B digital device, pursuant
to part 15 of the FCC Rules. These limits are designed to provide reasonable protection
against interference in residential installation. This equipment generates, uses, and can
radiate radio frequency energy and, if not installed and used in accordance with the
instructions, may cause harmful interference to radio communications. However, there is no
guarantee that interference will not occur in a particular installation. If this equipment does
cause harmful interference to radio or television reception, which can be determined by
turning the equipment off and on, the user is encouraged to try and correct the interference
by one or more of the following measures:
The COA2 is an ambient CO detector and not intended for in-flue stream
FCC INFORMATION
• Reorient or relocate the receiving antenna
• Increase the separation between the equipment and receiver
• Connect the equipment into an outlet on a circuit different from that which the
receiver is connected
• Consult the dealer or an experienced radio/TV technician for help
WARNING
:
Any changes or modifications not expressly approved by the manufacturer
could void the user’s authority to operate the equipment.
Read entire Safety Notes section regarding potential hazard and proper instructions before
using this tester. In this manual the word “WARNING” is used to indicate conditions
or actions that may pose physical hazards to the user. The word “CAUTION” is used to
indicate conditions or actions that may damage this instrument.
WARNING
• To ensure safe operation and service of the tester, follow these instructions. Failure to
observe these warnings can result in severe injury or death.
• This meter is designed for trade professionals who are familiar with the hazards of their
trade. Observe all recommended safety procedures that include use of personal protective
equipment.
• Carbon monoxide is life threatening even at relatively low concentrations. Be sure you
understand the risks before investigating potential leaks.
• Do not use the tester if it appears damaged.
• Always adhere to national and local safety codes.
• This COA2 is specified to read Carbon Monoxide from 1-999ppm. The app may continue
to read above this threshold. Professional safety precautions are recommended above
35ppm.
• Threshold safety levels are set to industry standards. By customizing these levels to
higher CO ppm thresholds, users risk personal safety. These settings are saved and used
after exiting and relaunching the App. To reset back to industry standards set the
levels to: Green ≤ 9 ppm; Blue ≤ 34 ppm.
• Any changes or modifications not expressly approved by the manufacturer could void the
user’s authority to operate the equipment.
CAUTION
• The COA2 is designed for ambient testing only. DO NOT place the instrument in flue gas
or warm air streams.
• The COA2 is not designed for in-flue detection. Readings more than 999 ppm will damage
the sensor in the COA2, even if readings continue above the 999ppm range.
• Ensure that the sensor vent at the top of the instrument is free from obstruction.
NOTE: A quick instrument check can be performed by the smoke from a blown out match. Let
smoke enter the
vent
. This will cause the indicated CO level to increase.
FEATURES
• Wireless Connectivity
• Visual LED alert
• Audible alert
• Backlight
• Auto power off
• Low battery indication
• Max capture
• Data Hold
• Belt Clip
• CO Sensor Self-Test
GENERAL SPECIFICATIONS
• Operating Temperature: 32° to 104°F (0° to 40°C)
• Storage Temperature: 14° to 122°F (-10° to 50°C)
• Operating Humidity: <80%
• Pollution Degree: 2
• Over-range: “OL” is displayed
• Dimensions: 1.92 x 1.25 x 5.62” (49 x 32 x 143mm)
• Item Weight: 0.35lb
• Calibration: Recommended Annually
• Certification: CE EMC & RED(STD EN 61326-1, EN 60950-1, EN 301 489-1), RoHS, FCC, KC
• Battery Type: (AAA) 3
• Range: 1~999ppm
• Resolution: 1ppm
• Accuracy: ± 3% or ± 2ppm (whichever is greater)
• App Compatibility: iOS®; 11.0 or newer and iPhone 5S or newer
Android™; 6.0 or newer
NOTE:
Requires compatible mobile device and FREE downloadable App “COA2”.
A. CO Sensor Vent: For sensing CO levels (keep clear of obstructions).
B. LED Indicator
• 0 to 1 ppm: White
• 2 to 9 ppm: Green
• 10 to 34 ppm: Blue
• ≥35 ppm: Red
C. Display: View test results, function changes, wireless connection.
D. Max/Back Light Button: Press to view Maximum levels, press again to return to live
reading. Press and hold to turn on back light, press and hold again to turn off back light.
Back light turns off after 2 minutes.
E. Hold/Wireless (BT) Button: Press to hold current readings, press again to return to
live reading. Press and hold to connect wirelessly, press and hold again to disconnect.
F. Power Button
•
Press to power on tester, press and hold to power off tester. The COA2 activates a
self Test and counts down from 50 when powered on. And counts down from 10
when powered off.
•
Press and hold to disable Auto power off, when turning the tester on.
If 25ppm CO (or greater) is detected, the tester cannot be turned off.
G. Belt clip
H. Battery Cover
I. Serial Number
Audible Alert: ≥35ppm
Auto power off: after 30 minutes non-use
BATTERY REPLACEMENT
CONTAINS FCC ID:
QOQBGM13P
R-R-KaA-COA2
R-C-BGT-BGM13P22
S/N:
:3X1.5V LR03 AAA
MADE IN KOREA
• Loosen the 2 captive screws that hold the battery cover and remove battery cover.
• Replace the old batteries with 3 new (AAA) batteries
• Replace the battery cover and tighten screws.
NOTE:
The COA2 will disconnect from the wireless device when batteries are too low to
A. Parts per million screen
B. Hold: Press the HOLD button to hold/capture the reading on the screen.
C. Record: Press the RECORD button to record measurement test results.
D. Max: Press the MAX button to capture the maximum values.
E. Graph: Press the GRAPH button to see a live graph of test readings.
E.
MENU SCREEN
A. Disconnect from Device: Press to disconnect from device
B. Settings:
General
• Button Sound: On/Off
• Button Vibration: On/Off
• Refresh Rate: 100ms to 1000ms
Recording
• Continuous Recording: Record until user stops recording
• Max Samples: Max number of samples in a log file 100 to 1000
Logging Interval: Interval between the recorded adjacent samples
• Minutes: 0 to 59
• Seconds: 1 to 59
C. Info: Press the INFO button to view information about the App, including version.
• Manual: Press the Manual button to view the COA2 manual
D. UEi Apps: Press the UEi Apps button to get to UEi master Apps
E. Logs: Press the LOGS button to view a list of recorded samples
• Title
• Date
• Recording Interval: Set in Menu screen
• Number of Samples: Set in Menu screen
F. Summary Information
• Maximum Value: In ppm
• Average Value: In ppm
• Minimum Value: In ppm
G. Samples
• Date
• Time
• Value
H. Export: Press the EXPORT button to email/export results in either CSV value, Graph
in PNG or Graph in JPEG. Highlight the Log to be exported. Press the Send Data button
The App will not connect with COA2Check batteries in COA2
The App suddenly errors out, loses
connection and is unable to reconnect
The App disconnects from the COA2
while using other Apps and is unable to
reconnect
Phone features not working while COA2
is connected and app is in use.
Check batteries in COA2
Some Android phones disconnect from
the COA2 while in the background. Quit
the App and relaunch. Apple phones
keep connectivity in background mode.
Many devices require the COA2 to be
disconnected to use phone services.
DISPOSAL
CAUTION:
subject to separate collection and correct disposal.
This symbol indicates that equipment and its accessories shall be
CLEANING
Periodically clean your adapter’s case using a damp cloth. DO NOT use abrasive, flammable
liquids, cleaning solvents, or strong detergents as they may damage the finish, impair
safety, or affect the reliability of the structural components.
STORAGE
Remove the batteries when instrument is not in use for a prolonged period of time. Do not
expose to high temperatures or humidity. After a period of storage in extreme conditions
exceeding the limits mentioned in the General Specifications section, allow the instrument to
return to normal operating conditions before using it.
WARRANTY
The COA2 is warranted to be free from defects in materials and workmanship for a period
of one year from the date of purchase. If within the warranty period your instrument should
become inoperative from such defects, the unit will be repaired or replaced at UEi’s option.
This warranty covers normal use and does not cover damage which occurs in shipment
or failure which results from alteration, tampering, accident, misuse, abuse, neglect or
improper maintenance. Batteries and consequential damage resulting from failed batteries
are not covered by warranty.
Any implied warranties, including but not limited to implied warranties of merchantability
and fitness for a particular purpose, are limited to the express warranty. UEi shall not
be liable for loss of use of the instrument or other incidental or consequential damages,
expenses, or economic loss, or for any claim or claims for such damage, expenses or
economic loss.
Warranty only covers hardware and does not extend to software applications.
A purchase receipt or other proof of original purchase date will be required before warranty
repairs will be rendered. Instruments out of warranty will be repaired (when repairable) for
a service charge.