Before using this meter, read all safety information carefully.
“WARNING” is used to indicate conditions or actions that may pose physical hazards to the user.
“CAUTION” is used to indicate conditions or actions that may damage this instrument.
WARNING!
This analyzer extracts combustion gases that may be toxic in relatively low concentrations. These gases are exhausted from
the back of the instrument. This instrument must only be used in well-ventilated locations. It must only be used by trained and
competent technicians after due consideration of all the potential hazards.
MENU / SETUP
UpDown Enter
Flue Test
Fuel
Prs
C225
Exch
Test
Temp
Room
CO
Aux
Status
Menu
Rotate dial to Menu
SET DATE
UpDown Enter
Flue Test
Fuel
Prs
C225
Exch
Test
Temp
Room
CO
Aux
Status
Menu
Rotate dial to Menu
-------MENU
SETUP
--------
UpDown Enter
Flue Test
Fuel
Prs
Aux
Exch
Test
SETUP shows press ENTER
“” to proceed.
-------MENU
SETUP
--------
UpDown Enter
Flue Test
Fuel
Prs
Aux
SETUP shows press ENTER
“” to proceed.
Exch
Test
-------SETUP
SET TIME
--------
UpDown Enter
Flue Test
Fuel
Prs
Aux
Exch
Test
SET TIME shows press
ENTER “” to proceed.
--------
SETUP
SET DATE
--------
UpDown Enter
Flue Test
Fuel
Prs
Aux
Press “s“ UP or “t“
DOWN to “SET DATE”
Press ENTER “”
Exch
Test
-------SET TIME
_
09:13:07
HH:MM:SS
UpDown Enter
Flue Test
Fuel
Prs
Aux
Exch
Test
SET TIME and Clock
displays with the cursor
underneath the first digit.
Press “s“ UP or “t“
DOWN to increase or
decrease and press
ENTER “” to set and
proceed to the next digit.
(24 hour format)
-------SET DATE
_
01/01/15
MM/DD/YY
UpDown Enter
Flue Test
Fuel
Prs
Aux
SET DATE displays with
the cursor underneath the
first digit. Press “s“ UP or
“t“ DOWN to increase or
decrease and press
ENTER “” to set and
proceed to next digit.
Exch
Test
-------SET TIME
_
09:13:07
HH:MM:SS
UpDown Enter
Flue Test
Fuel
Prs
Aux
Exch
Test
Press “Send” to complete
setting the time once all
digits are set.
-------SET DATE
_
01/01/15
MM/DD/YY
UpDown Enter
Flue Test
Fuel
Prs
Aux
Press ENTER “” to
complete setting the date
once all digits are set.
NOTE: NOx calculates
estimated total based
on NO reading and this
percentage. Set value
based on compliance
requirements. (Only
available with NO sensor
fitted) (C257)
Exch
Test
NOTE: Only for analyzers
with installed bluetooth
module. This will be the
passkey used to sync to
your computer. The default
passkey is 1111.
Exch
Test
Aux
Press “s“ UP or “t“
DOWN to select between
ENGLISH, FRENCH and
SPANISH. Press ENTER “
” to set.
--------
05 %
--------
NOx will show with the
cursor underneath the first
digit Press “s“ UP or “t“
DOWN to set percentage
Press ENTER “” to
proceed
--------
PASSKEY
1111
--------
Aux
____will show with the cursor underneath the first
digit Press “s“ UP or “t“ DOWN to increase or
decrease and press ENTER “” to proceed to the
next digit.
will have an incomplete
printout on the UEi KMIRP2
(KMIRP).
KMIRP
--------
UpDown Enter
Flue Test
Fuel
Prs
Aux
Press “s“ UP or “t“
DOWN to select between
the KMIRP and IRP-2
infrared printers. Press
ENTER “” to select.
K1g0.350
K1n0.393
K2 11.89
K3 09.83
Aux
Fuel will show with the cursor underneath the first type
and digit Press “s“ UP or “t“ DOWN to increase or
decrease and press ENTER “” to proceed to the
next digit.
If there is no logged test
results the previous screen
will appear.
www.GlobalTestSupply.com
Press “s“ UP or “t“
DOWN to select between
“COMB’N”(combustion),
PRESSURE, EXCH (Heat
exchange integrity test),
TEMP, and ROOM CO
tests. Press ENTER “”
to view.
Press “s“ UP or “t“
DOWN to select “VIEW”
“DEL ALL” or “EXIT”.
Select “View” to display
logs. Select “Del All” to
erase records, or “Exit”
to move to the previous
screen.
To delete records select
“Del All”. A prompt of
“Del All?” with YES shown
prior to erasing records.
Press (SEND) ENTER
“” to continue.
Press and hold “s“ UP to
change upper displayed
parameter or “t“ DOWN
to change the lower
displayed parameter. Press
“s“ UP or “t“ DOWN
briefly to scroll through to
the different logs. (while
maintaining selected
parameter view). LOG ##
is shown on the first line.
Press (SEND) ENTER
“” to print. or press and
hold to return to REPORTS
menu.
7
--------MENU
SETUP
--------
--------MENU
SETUP
--------
--------MENU
SETUP
--------
CONTRAST
MENU / SCREEN
UpDown Enter
Flue Test
Fuel
Prs
C225
Exch
Test
Aux
Status
Menu
Rotate dial to Menu
Temp
Room
CO
-------MENU
SCREEN
--------
UpDown Enter
Flue Test
Fuel
Aux
Press “s“ UP or “t“
DOWN to “SCREEN”
Press ENTER “” to
proceed.
HEADER
UpDown Enter
Flue Test
Fuel
Prs
C225
Exch
Test
Aux
Status
Menu
Rotate dial to Menu
Temp
Room
CO
-------MENU
SCREEN
--------
UpDown Enter
Flue Test
Fuel
Aux
Press “s“ UP or “t“
DOWN to “SCREEN”
Press ENTER “” to
proceed.
SETUP AUXILIARY SCREEN (AUX)
-------MENU
CONTRAST
--------
UpDown Enter
Prs
Exch
Test
Flue Test
Fuel
Prs
Aux
Exch
Test
Press “s“ UP or “t“
DOWN to “CONTRAST”
Press ENTER “” to
proceed.
--------
SCREEN
HEADER
--------
UpDown Enter
Prs
Exch
Test
Press “s“ UP or “t“
DOWN to “HEADER”
Press ENTER “” to
proceed.
Flue Test
Fuel
Prs
Aux
Exch
Test
-------CONTRAST
10
--------
UpDown Enter
Flue Test
Fuel
Prs
Aux
Exch
Test
Press “s“ UP or “t“
DOWN to increase or
decrease screen contrast.
Press ENTER “” to set.
------- HEADER
HEADER1
--------
UpDown Enter
Flue Test
Fuel
Prs
Aux
Press “s“ UP or “t“
DOWN to select desired
header to edit. Press
ENTER “” to proceed.
Exch
Test
--------
_
YOUR COM
PANY NAM
E & ----
UpDown Enter
Flue Test
Fuel
Prs
Aux
The default entry for
HEADER1 will show with
the cursor underneath the
first digit Press “s“ UP or
Exch
Test
“t“ DOWN to increase or
decrease and press ENTER
“” to proceed to the
next digit. Press ENTER
“” to select and exit
when complete.
UpDown Enter
Flue Test
Fuel
Prs
C225
Exch
Test
Aux
Status
Menu
Rotate dial to Menu
-------MENU
Temp
Room
CO
Press “s“ UP or “t“
DOWN to “SCREEN”
Press ENTER “” to
proceed.
SCREEN
--------
UpDown Enter
Flue Test
Fuel
Prs
Aux
Exch
Test
-------SCREEN
AUX
--------
UpDown Enter
Flue Test
Fuel
Prs
Aux
Press “s“ UP or “t“
DOWN to “AUX”
Press ENTER “” to
proceed.
Exch
Test
------- AUX
LINE 1
--------
UpDown Enter
Flue Test
Fuel
Prs
Aux
Press “s“ UP or “t“
DOWN to select desired
line to edit. Press ENTER
“” to proceed.
Exch
Test
------- AUX
LINE 1
NOp_ N/F
UpDown Enter
Flue Test
Fuel
Prs
Aux
The parameter will show
underneath “LINE 1” Press
“s“ UP or “t“ DOWN
to change (See page 2 for
and press ENTER “” to
select and proceed to next
line. (If desired)
BASIC FAST START (OPERATION)
--------MENU
SET TIME
--------
Nat GasUP DOWN TO
CHANGE
O2% 00.0COp 000P 000
TFf 000
1. POWER ON
FRESH
AIR
PURGE
Aux
Status
Menu
UpDownEnter
Flue Test
Fuel
Prs
Exch
Test
Temp
Room
CO
TIME 90
UpDown Enter
Power on in area of fresh air and allow to countdown. On
sealed combustion appliances (ducted inlet) connect flue
probe and power on outside and allow to countdown to set
2. SELECT FUEL
UpDown Enter
Flue Test
Fuel
Aux
Status
Menu
Rotate dial to Fuel.
Prs
Nat Gas
Exch
Test
Temp
Room
CO
Press “s“ UP or “t“ DOWN
to scroll through available
fuels.
UP DOWN
TO
CHANGE
UpDown Enter
Flue Test
Fuel
Prs
Exch
Aux
Test
inlet temperature. (see page 12) Note: Not required if using
inlet probe in T2.
Connect flue probe
thermocouple connector to
T1, and connect flue probe to
water trap as shown above.
Insert Flue Probe in stack.
Adjust the cone so the end of
the probe is approximately at
the center of the stack
(4” stack adjust cone to
aprox. 2” from end of probe.)
Flue Test
Fuel
Prs
C225
Exch
Test
Temp
Room
CO
Aux
Status
Menu
Rotate dial to Flue Test.
Analyzer will display flue test
page 1 (O2, COp, Pressure,
Flue Temp) Press and hold
“s“ UP to see page 2 (CO2,
NO, Efficiency, Excess Air)
Press “Send” to print results
or press and hold “Send” to
log results from either screen.
Connect true draft hose
and probe into P1. Use
both P1 and P2 for
differential. Place probe
tip in stack to measure
draft. Connect static
pressure hose to P1 and
PRESSURE
P 0.00
mBar
10:07:35
UpDown Enter
Flue Test
Fuel
Prs
Aux
Status
Menu
Exch
Test
Temp
Room
CO
Place true draft hose and probe tip in
flue to measure draft. You can also
use the combustion draft probe for
measuring pressure.
P2 to measure differential
pressure. UEi recommends
C225
CO2 H2CO ∆T PRS
INDUSTRIAL EOS ANALYZER
the ASP1 static probe
tips to measure static
pressures.
C225
CO2 H2CO ∆T PRS
INDUSTRIAL EOS ANALYZER
Press “Send” to print results or press
and hold “Send” to log results. (See
page 7)
Note: Pressure readings are also available on “Flue Test” and “Aux” selector positions.
Exch Test (Heat Exchanger Integrity Test)
There are many methods to test heat exchanger integrity. One of these is to observe the Excess Air, O2 and CO readings both before
and after the blower turns on. If the heat exchanger is sealed your O2 and CO readings should remain fairly stable. A breach in the heat
exchanger may allow fresh air to be forced into the flue after the blower turns on due to a pressure increase in the plenum. The result may
be a rise in the measured O2 in the stack gas and an increase in the Excess Air. In some sealed systems the fresh air drawn in through
the breach may reduce the combustion air available leading to an increase in the CO reading. If either of these situations are present it is
probable that there is a problem with the Heat Exchanger which may require additional testing and inspection .
NOTE: Many cracks are invisible to borescopes or the naked eye, and only open or separate from pressure or temperature changes during
operations.
COp 000
O2% 20.9
X 00
PRESS
UpDown Enter
Flue Test
Fuel
Prs
C225
Exch
Test
Temp
Room
CO
Aux
Status
Menu
CO2 H2CO ∆T PRS
INDUSTRIAL EOS ANALYZER
Rotate dial to Exch
Test Rotate test
selector to Exch Test.
Call for heat on the
system. Observe and
wait for O2 readings to
stabilize.
COp 000
O2% 20.9
X 00
PRESS
UpDown Enter
Flue Test
Fuel
Prs
C225
Exch
Test
Temp
Room
CO
Aux
Status
Menu
CO2 H2CO ∆T PRS
INDUSTRIAL EOS ANALYZER
Prior to the blower
turning on, and after
readings have stabilized,
press the SEND button.
This will store the PreBlower test segment.
EXC TEST
TO START
PRESS
UpDown Enter
Flue Test
Fuel
Prs
C225
Exch
Test
Temp
Room
CO
Aux
Status
Menu
CO2 H2CO ∆T PRS
INDUSTRIAL EOS ANALYZER
After the blower turns
on press PUMP to start
the Post- Blower test.
SAMPLE
IN
PROGRESS
TIME 50
UpDown Enter
Flue Test
Fuel
Prs
C225
Exch
Test
Temp
Room
CO
Aux
Status
Menu
CO2 H2CO ∆T PRS
INDUSTRIAL EOS ANALYZER
The meter will wait 60
seconds and then record
the Post-Blower values
for CO, O2 and Excess
Air. Test results will
automatically be stored
to exchange reports.
Report includes both
Pre and Post Blower
Variance
CO 00
O2 00
X 00
UpDown Enter
Flue Test
Fuel
Prs
C225
Exch
Test
Temp
Room
CO
Aux
Status
Menu
CO2 H2CO ∆T PRS
INDUSTRIAL EOS ANALYZER
Results are displayed at
the end of the test and
you can view through
the “MENU” go to the
“REPORT” screen and
select EXCH. Select
“VIEW” and scroll to
desired log. (See page 7)
Press “SEND” to print.
test segments and
differences.
• Oxygen ................................ 3 to 9% O2
• Stack Draft (Water Column Inches) ....... -.02 to -.04wc”
• Overfire Draft (Water Column Inches) .... +0.4 to +0.6wc”
• Carbon Monoxide (parts per million) ........... <100ppm
NOTE: Follow manufacture guidelines for the specific equipment being
serviced.
Typical Excess Air Level
O2% (measured) Excess Air %
Natural gas 3% 16.7%
Light Oil 5% 31%
Coal 8% 62%
15
GENERAL MAINTENANCE
• Re-certify your instrument annually to ensure it meets original performance specifications
• Keep your instrument dry. If it gets wet, wipe dry immediately. Liquids can degrade electronic circuits
• Whenever practical, keep the instrument away from dust and dirt that can cause premature wear
• Although your instrument is built to withstand the rigors of daily use, it can be damaged by severe impacts. Use reasonable caution
when using and storing the meter
PRE TEST CHECKLIST
• Clean particle filter
• Water trap and probe line are empty of water
• Power on and zero
• All hose and thermocouple connections are properly secured
• Flue gas probe is sampling ambient FRESH air
• Water trap is fitted correctly to the instrument
• Flue temperature plug is connected
SETTING INLET TEMPERATURE
• Turn on and zero the analyzer without the flue probe connected to use ambient temperature
• Connect flue probe thermocouple to T1 during zero countdown to store probe tip temperature as inlet (ducted system)
• Connect a thermocouple to T2 to measure inlet during test.
ANALYZER CONNECTIONS
WARNING!
Turning the pump off while the probe is in the flue will leave toxic gases inside
the analyzer. Once data has been printed or copied it is advisable to purge the
unit with fresh air as soon as possible. To do this remove the probe from the flue
and turn ON the pump. Always allow the readings to return to zero (20.9 for O2)
prior to shutting the unit off. The meter will not switch off until the CO reading is
below 20 ppm.
WARNING!
The probe will be hot from flue gases. Remove the probe from the flue and allow
it to cool naturally. Do not immerse the probe in water, as this will be drawn into
the analyzer and damage the pump and sensors. Once the probe is removed from
the flue and the readings have returned to ambient levels hold down “On/Off”
and switch off the analyzer. The instrument will count down from 30 to switch
off. If you pressed “On/Off“ by mistake, pressing “Send“ will return you to normal
operation.
EMPTYING & CLEANING THE IN-LINE WATER TRAP
The in-line water trap should be checked and emptied on a regular basis. Water vapor will condense in
the probe line, which may cause the water trap to fill suddenly if the probe is moved. Care should be
taken at all times.
Carefully remove the rubber plug from the bottom of the water-trap housing. Dispose of the condensate
in a suitable drain, care must be taken as it could be acidic. If condensate spills onto the skin or clothing,
clean off immediately using fresh water, seek medical advice if problems occur. Ensure plug is replaced
before performing combustion tests. Note: O2 reading will be high if the Water Trap Plug is not in place.
CHANGING THE PARTICLE FILTER
This is a very important part of the analyzer and should be changed regularly. It prevents dust and dirt
particles from entering the pump and sensors that will cause damage. The filter MUST be changed when
it appears discolored on the inner surface.
NOTE: Take care when inserting
the temperature probes as the
pins are polarized. Insert with the
smaller pin (+) to the right.
Remove water-trap assembly from the analyzer as shown above. Remove the filter and plastic holder
from the housing. Discard the filter element but keep the holder to fit to the new filter. Clean the inside
of the filter housing with a suitable soft cloth. Fit the holder onto the new filter element and then insert
into the housing. Refit the housing onto the analyzer.
www.GlobalTestSupply.com
REPLACING THE BATTERIES
This meter has been designed for use with both alkaline and rechargeable Nickel Metal Hydride (NiMH) batteries. No other types are
recommended. The analyzer is supplied with 4 “AA” size alkaline batteries. These should be installed into the instrument as shown in the
diagram indicated on the back of the unit.
CAUTION!
Take great care when installing the batteries to observe correct polarity. Always check the meter for operation immediately after installing new batteries.
Using Re-Chargeable Batteries
The battery charger must only be used when NiMH batteries are fitted. Alkaline batteries are not re-chargeable. Attempting to recharge
alkaline batteries may result in damage to the product and may create a fire risk.
Battery Charging
Ensure that you use the correct charger. This unit uses a 9V DC regulated charger. Ensure that the batteries are fitted in the correct manner,
and then charge for at least 16 hours. Subsequent charges should be overnight. NiMH batteries may be charged at any time, even for short
periods to conduct testing.
WARNING!Under NO circumstance should you expose batteries to extreme heat or fire as they may explode and cause injury. Always
dispose of old batteries promptly in a manner consistent with local disposal regulations.
POST TEST
O2% 20.9
COp 000
P 000
TFf 000
UpDown Enter
Remove the probe from the flue and allow the analyzer to purge
with fresh air until readings return to zero. - O2 to 20.9%, CO to
Zero (Be careful as the probe tip will be HOT)
PERIODIC SERVICE
WARNING!
Repair and service of this instrument is to be performed by qualified personnel only. Improper repair or service could result in
physical degradation of the instrument. This could alter the protection from personal injury this meter provides to the operator.
Perform only those maintenance tasks that you are qualified to do.
ANNUAL RE-CERTIFICATION
Drain water trap by unplugging
the drain plug and shake to
get excess water out.
Check particle filter for dirt
and any other sediment and
replace if necessary.
While the CO sensor has an expected life of more than five years in normal use it is recommended that the analyzer is re-certified at least annually,
This is so that long-term drift on the sensor and electronics can be eliminated. Local regulations may require more frequent re-calibration and
users should check with appropriate authorities to ensure they comply with relevant guidelines.
CLEANING
Periodically clean your instruments case using a damp cloth. DO NOT use abrasive, flammable liquids, cleaning solvents, or strong detergents as
they may damage the finish, impair safety, or effect the reliability of the structural components.
Time: Amount of time that the fuel and oxygen are together in the combustion chamber
Temperature: How high the temperature is determines the rate of oxidation, or speed of the combustion
Turbulence: How well the fuel and air are mixed
These three factors are all interrelated, and will move your results along the combustion curves.
COMBUSTION MEASUREMENT TERMS
Other parameters measured include net temperature, draft and efficiency.
Net Temperature
Net temperature is the difference between the combustion air entering the combustion chamber and the flue gas temperature past the heat exchange. This
is used to determine how efficient the system is extracting heat from the combustion process in addition to the performance of the combustion process. On
sealed systems that have ducted inlet air for combustion air, the net temperature must compare this air stream temperature with the flue gases. If the appliance simply uses room air for the combustion air, our analyzers have an internal temperature sensor in the handset, so it will use this temperature when
calculating net temperature. The most accurate results for efficiency are obtained when measuring flue gases at the point where flue temperature (not flame
temperature) is the highest.
Draft
Draft is the difference between the ambient pressure level and the pressure level in the flue.
This is created either by the natural buoyancy of the hot gases created in combustion lifting, or by an inducer fan that assists the flow of flue gases up the
stack. Most combustion equipment will specify the amount of draft that is required for proper operation. Draft helps draw combustion air into the combustion
chamber, and also helps in mixing the fuel and oxygen. Without proper draft, the combustion process can spill poisonous by-products into the space where the
appliance is located. This can be a risk to those in the area, or create a danger to residents or employees working near the combustion equipment.
Efficiency
Efficiency is a measure of how well the fuel is burned to create heat, and how well the generated heat is captured for the intended use.
The information used to create this value are based on the fuels heating value, the heat lost up the flue and the gas components in the flue gas. The original
method to determine efficiency included many manual methods and lookup charts. As an example you would measure the CO2 level and the stack temperature and then reference a slide scale that would give you the relative efficiency number. UEi’s electronic combustion analyzers perform the measurements
on a continuous basis, and can calculate the efficiency as adjustments are being made. Combine this with a printout and you are able to provide a before
and after comparison of the combustion equipment in relatively little time as part of normal servicing. Combustion efficiency is not the same as AFUE
(annual fuel usage efficiency). AFUE is not measurable with any portable flue gas analyzer.
Combustion Efficiency Calculations
This identifies three sources of loss associated with fuel burning:
•Losses due to flue gasses:
Dry Flue gas loss, Moisture and hydrogen,
Sensible heat of water vapor, Unburned gas
• Losses due to refuse:
Combustible in ash, riddling and dust
• Other losses:
Radiation, convection, conduction other unmeasured losses
Net efficiency calculations assume that the energy contained in the water vapor (formed as a product of combustion and from wet fuel) is recovered and the
wet loss term is zero. Gross efficiency calculations assume that the energy contained in the water vapor is not recovered. Since the fuel air mixture is never
consistent there is the possibility of unburned/partially unburned fuel passing through the flue. This is represented by the unburned carbon loss. Losses due to
combustible matter in ashes, riddling, dust and grit, radiation, convection and conduction are not included.
CO Air Free
Certain standards ( ANSI Z21.1) for Carbon Monoxide are stated in terms of air-free. Air-free refers to the concentration of CO in combustion gases undiluted
with flue, or other gases containing little CO. This value is computed using an equation that takes into account the O2 concentration of the flue gas.
• If 5% O2 is measured (O2m) in the flue then the CO gas value will be recalculated as if 0% were measured. The equation for air-free is as follows:: COaf = CO PPM x [(20.9) / (20.9 - O2m)]
• In our example if a reading of 325 PPM were measured then the air-free value would be calculated as follows:
COaf = 325 PPM x [(20.9) / (20.9 - 5)] COaf = 325 PPM x [(20.9) / (15.9)] COaf = 427
We may be given a limit on our gas range by the local authority, which stated that we must not emit more than 400-PPM Carbon Monoxide air-free. In the
example we would be breaking the limit and corrective action should be taken to reduce the level of CO. Air-free values prevent false readings being submitted, e.g. allowing more air into the boiler will increase the oxygen level in the flue and dilute any toxic gas reading. Air-free referencing gives readings as if
they were undiluted.
Nominal range ±64.2 InH2O (160mBar) Max
over-range w/o damage to sensor is ±321InH2O (800 mBar)
Pre-programmed Fuels
Storage Capacity
*2
*2
*2
*2
0 - 60 ppm1ppm±3ppm
61 - 4,000ppm nom1ppm±5% of reading
4,001ppm to 10,000ppm1ppm±10% of reading
20,000ppm max for 15 min1ppmNot specified
0 - 20%0.1% ±0.3%
0 - 21%0.1%±0.3%
0 - 99.9%0.1%±1.0% of reading
0 - 250%0.1%±0.2% of reading
0 - 0.9990.0001±5% of reading
31 - 100 ppm1 ppm±5 ppm > 30ppm
101 - 1000 ppm1 ppm±5% reading >100ppm
±0.6 inH2OMaximum 0.001 InH2O
<10 InH2O
±64.2 inH2O±3% of reading
±0.02 In H2O
Natural gas, L Oil, H Oil, Propane, Wood, Biofuel, Wood, PLUS 3 User defined fuels.
99 Combustion tests, 20 Pressure tests, 20 Heat Exchange tests, 20 Temperature tests
20 Room CO tests
GENERAL SPECIFICATIONS
• Operating Altitude: 2000m (6,561 ft.)
• Storage Altitude: 10,000m (32,808 ft.)
• Operating Temperature: 32°F to 104°F (0°C to 40°C) at 10 - 90% R.H
• Pollution Degree: 2
• Dimensions: Handset: 7.9” (200mm) x 3.5” (90mm) x 1.8” (45mm)
Probe: (L) 11.8” (300mm) x (D) 0.25” (6mm) with 7.8” (200mm)
6ft (3m) neoprene hose
• Weight: 2.2lbs (1kg)
• Certifications: EMCEN 50081-1, EN 50082-1, CE
• Battery Type: 4 x 1.5V AA
• Battery Life: >8Hrs using AA Alkaline
• Power Supply Input: 110V AC Output: 9V DC Regulated
ELECTROMAGNETIC COMPATIBILITY (EMC)
This product has been tested for compliance with the following generic standards: EN 50081-1, EN 50082-1 and is certified to be compliant.
The European Council Directive 89/336/EEC requires that electronic equipment does not generate electromagnetic disturbances that exceed defined levels and
has an adequate level of immunity to enable it to be operated as intended.
Since there are many electrical products in use that pre-date this Directive and may emit electromagnetic radiation in excess of the standards defined in the
Directive there may be occasions where it would be appropriate to check the analyzer prior to use. The following procedure should be adopted.
• Go through the normal start up sequence in the location where the equipment is to be used
• Switch on all localized electrical equipment that might be capable of causing interference
• Check that all readings are as expected (a level of disturbance in the readings is acceptable)
• If not, adjust the position of the instrument to minimize interference or switch off, if possible, the offending equipment for the duration of the test
At the time of writing this manual (April 2015) UEi is not aware of any field based situation where such interference has ever occurred and this advice is only
given to satisfy the requirements of the Directive.
Register your product online at www.ueitest.com and click the PRODUCT REGISTRATION banner. This allows us to notify you when your analyzer has upgrades
available or when recertification is due. We manufacture industry-leading products for professionals like you; please help us to ensure your product is always
fit for purpose by registering your new purchase online - it only takes a few minutes.
IMPORTANT: RE-CERTIFICATION
While the CO sensor has an expected life of more than 5 years in normal use, it is recommended that the analyzer is recertified annually. The C255 analyzer
typically requires an annual recertification each of the first four years and full recertification on the 5th year pending use. This eliminates long term drifting of
the sensor and electronics. Local regulations may require more frequent calibration and users should check with appropriate authorities to ensure compliance.
UEi Test Instruments offers 2 kinds of recertification services.
Annual: Calibration of all installed sensors
Full: Replacement of all applicable sensors
Both services offer:
Full system, 30-point inspection with replacement of any failed parts.
Firmware and/or software upgrades as needed.
Certificate of calibration provided for record keeping.
Recertification work report to document work completed.
Units returned cleaned with paid return freight.
All units are recertified by trained professionals, which eliminate the legal liability of self-installed sensors. In the USA we have two recertification locations,
one in Indianapolis, IN, the second in Portland, OR. For canadian customers we have a recertification service center in Vancouver, BC, which is ISO/IEC
17025:2005 accredited and offers NIST certification services for select products. All recertification services are warranted for one year.
Save time with online recertification check-in by visiting www.ueitest.com and click the SERVICE REQUEST banner and filling out the form.
Our current recertification prices can be found on the SERVICE section of our web site. Up front pricing means NO surprises.
LIMITED WARRANTY
The C255 combustion analyzer is warrantied for five years including sensors.
If within the warranty period your instrument should become inoperative from such defects, the unit will be repaired or replaced at UEi’s option. This warranty
covers normal use and does not cover damage which occurs in shipment or failure which results from alteration, tampering, accident, misuse, abuse, neglect
or improper maintenance (calibration). Batteries and consequential damage resulting from failed batteries are not covered by warranty. Any implied warranties, including but not limited to implied warranties of merchantability and fitness for a particular purpose, are limited to the express warranty. UEi shall not be
liable for loss of use of the instrument or other incidental or consequential damages, expenses, or economic loss, or for any claim or claims for such damage,
expenses or economic loss. A purchase receipt or other proof of original purchase date will be required before warranty repairs will be rendered. Instruments
out of warranty will be repaired (when repairable) for a service charge. Contact UEi for specific warranty and service information. This warranty gives you specific legal rights. You may also have other rights which vary from state to state.