Thank you for purchasing the UEi AQM4 IAQ
meter. The meter measures CO² levels, CO
levels, air temperature, dew point, wet bulb
temperature, and humidity. The AQM4 is an
ideal instrument for indoor air quality (IAQ)
diagnosis.
Poor indoor air quality causes tiredness,
inability to concentrate, and even illness
(ex. Sick Building Syndrome). IAQ monitoring
and surveying especially for CO² levels
and air ventilation has become widely
applied in public offices, classrooms,
factories, hospitals, and hotels. It is also
recommended for regulating standards of
industrial air quality in some countries.
The portable AQM4 uses NDIR (nondispersive infrared) technology for CO²
measurement to ensure reliability and longterm stability. It is useful in verifying HVAC
systems performance and air ventilation
control.
CO is a colorless, odorless, and tasteless
gas which is slightly lighter than air. It is
highly toxic to humans and animals. Carbon
Monoxide is measured using a long-life
electrochemical sensor.
Maximum indoor air quality CO level per
ASHRAE Residential standards 62-1989 for
living area is under 9ppm. It is necessary to
evacuate employees from enclosed spaces
if the CO concentration exceeds 100ppm per
OSHA exposure limit.
unit or increases value in setup.
Cancels data hold function.
Press to manually record the
reading. Selects unit or decreases
value in setup.
Displays MIN, MAX, TWA, STEL
function. Saves and finishes settings
while in sleep mode.
OPERATION
POWER ON/OFF
Press to turn the meter on and off. At
power up, the meter emits a short beep and
begins 30 second warm-up countdown
(Fig. 2), then enters normal mode with
current CO² and real time displays (Fig. 3).
NOTE:
THE REAL TIME DISPLAYS DATE
(YY-MM-DD) AND TIME IN SEQUENCE.
TAKING MEASUREMENT
The meter starts measurements when
powered on and updates readings every
second. In the condition of an operating
environment change, (ex. high to low temp)
it takes 30 seconds for the CO² sensor to
respond and 30 minutes for RH sensor.
NOTE: DO NOT HOLD THE METER IN
CLOSE PROXIMITY TO THE FACE AS
EXHALED BREATH AFFECTS CO² LEVELS.
In normal or hold mode, press key to
record, icon and main display flash for
about 3 seconds. The main display shows
the memory serial number. The AQM4 has
99 locations. (Fig. 15)
Each memory contains all parameters
(CO², CO, TA....%rh), not limited to the
selected parameter on the main display.
If the reading
changes quickly,
press to
freeze the reading
before manually
recording the data.
99 MEMORIES RECALL
In normal or hold mode, press key for
more than 2 seconds until the icon
flashes.
Press or to scroll through the
memory locations. The memory serial
The time displayed in memory recall mode is
the recording time of this memory.
To escape the memory recall mode, press
key to leave and return to normal
display.
ALARM
The meter features an audible alarm to give
warnings when CO concentration exceeds
a pre-set limit. (See Pg 12, P20 for setting
alarm threshold.) It emits beeps (Abt. 80dB)
when CO level goes over the set value and
stops only when the readings fall below the
set value. It beeps again when value goes
over the limit.
The meter can automatically record readings
of CO²/CO/TEMPERATURE/RH for extended
environment monitoring. The memory
capacity is 8,000 points. Users can set up
sampling rate from 1 second to 4 hours, 59
minutes and 59 seconds. (See pg 11, P40)
After sampling rate is selected, press
for 2 seconds under normal mode to start
logging. The icon flashes to indicate
To terminate data logging, press for
2 seconds, until icon stops flashing.
NOTE:
The AQM4 will only record one session. If
is activated all previous data is erased.
Download logged data prior to starting
another session.
AUTO POWER OFF
The meter turns off automatically after
20 minutes of inactivity. To override the
function, hold down and for
2 seconds to turn on the meter until "n" appears. (Fig. 20)
NOTE:
AUTO SLEEP
FUNCTION WILL BE
DISABLED DURING
CALIBRATION MODE
SETUP
When the meter is off,
hold down for
more than 1 second to
enter setup mode. (Fig
21) To exit setup, press
to return
to normal mode.
When P30 and "UNIT" (Fig 25) are displayed
on the LCD, press to go into P31 to
choose the temperature unit.
The set value will display on LCD (Fig 26)
Press or to choose C° or F° and
press to confirm.
-OR- Press to escape and return to
P30.
P40: LOGGING SAMPLING RATE
P40 allows you to set sampling rate of data
logging (Fig. 27). The range is from 1 second
to 4 hours 59 minutes and 59 seconds.
Press to display current setting and
enter edit mode. The display is in the
format [HH. MM. SS.] The hour digits will
be flashing. To change intervals between
to decrease. Press again to
confirm and enter minute setting. Press
to confirm and enter seconds setting.
The AQM4 allows you to calibrate CO² and
RH, for CO calibration please send unit to
UEi. To enter calibration mode, place unit
at suitable calibration site specified in the
following sections.
Press + , then also
simultaneously for 3 seconds to turn on the
meter and enter calibration mode (Fig. 33).
Press or to select parameter to
calibrate. To cancel or exit calibration at any
time, turn meter off.
NOTE:
ENSURE THE BATTERIES ARE AT FULL VOLTAGE
BEFORE THE CALIBRATION TO PREVENT
INTERRUPTION OR FAILED CALIBRATION.
CO² CALIBRATION
The manual calibration is suggested to
be done in sunny outdoor air that is well
ventilated. To begin, enter calibration mode.
Press or to select parameters
for calibration and press enter. The meter
is calibrated at standard 400ppm CO²
concentration at the factory. It is suggested
to manually calibrate regularly to maintain
good accuracy.
NOTE:
WHEN THE ACCURACY BECOMES A CONCERN,
RETURN TO UEi FOR STANDARD CALIBRATION.
CAUTION:
Do not calibrate the meter in the air with
unknown CO² concentration. It could be
calibrated as 400ppm by default which
leads to inaccurate measurements.
Wait 10 minutes until the blinking stops and
the calibration is completed. To cancel the
calibration, turn off the meter at any time.
To begin, enter calibration mode. Press
or to select parameters for calibration
and press enter.
The meters default for calibration of
humidity is 33% and 75% salt solution. The
ambient condition is recommended at 70 ~
75°F (25°C) and stable humidity. To cancel
calibration, turn off the meter.
CAUTION:
Do not calibrate the humidity without
default calibration salts. Contact UEi for
calibration salt or for service
33% Calibration
Plug the sensor probe into 33% salt bottle.
Hold down to enter 33% calibration.
"CAL" and calibrating value (32.8% if at
77°F) are blinking on the LCD.
75% Calibration
Plug sensor probe into 75% salt bottle and
enter calibration mode. In calibration mode,
press to select 75% calibration (Fig. 35),
press to start. "CAL" calibrating value
(75.3% if at 77°F) will blink on the LCD.
Meter will begin calibration and finish in
60 minutes when "CAL" and humidity stop
blinking. To cancel the calibration, turn off
the meter at any time.
return to UEi for standard calibration. It
is necessary to have standard CO gas for
calibration.
CAUTION:
Do not calibrate the meter in the air with
unknown CO concentration. It could be
calibrated as 400ppm by default which
leads to inaccurate measurements.
TROUBLESHOOTING
Will Not Power On
- Press for more than 3 seconds.
- Check that the adaptor is connected
and on.
- Check that battery level is sufficient
and properly installed.
Readings Not Changing
- Check whether data hold function is
activated (HOLD icon is at the left top).
Slow Response
- Check the air flow channels on
the rear. Verify no blockage exists.
Error Messages
EO1/E33: CO² sensor is out of order. Try new
battery first.
E02: The value is under range.
E03: The value is over range
E04: The original data error results in
E11: Retry humidity calibration
E16: Retry CO calibration
E31: Temperature sensor AD damaged
E32: Memory IC damaged.
E33: Humidity sensor or circuit damaged.
PC CONNECTION
The UEi AQM4 can be connected to a PC to
download recorded data.
Plug the USB cable into the socket at the
right side of the meter and the other port to
your PC. Next, install the software on your
PC following these procedures.
1. Insert the CD ROM and run installation.
Select a preferred directory and click
"Next" step by step to finish.
2. After the software installation is
completed, it will run USB driver
installation next.
Start the software and it will detect the
logger connection automatically, indicating
the COM port information at the bottom of
the main screen as well as the COM Port
setting column.
LOGGER SETTINGS
To set up logging plan, click "Setting" icon
and select "Logger". The setting page is
opened.
Users can set up the logger identifier
number. Click OK for setting and exit the
screen to confirm the settings.
To transmit auto recorded data from the
meter, click "Download" icon on the left top
side of the main screen. All auto-recorded
data in the logger will be transmitted. The
raw data with time stamp will display at
the lower part of the screen and the graph
in the middle.
In the Graph display, the Y-axis indicates
Temperature, RH, DP,WBT, CO2, and CO
levels in different line colors. The X-axis
can be switched to show Time or Recorded
Number Stamp.
The following functions help to view the
GRAPH data in more detailed way.
In the view menu, there are 4 tools to
enlarge the Graph data for detailed data
review.
Zoom in: Hold down "Ctrl" key on the PC
keyboard and left click any target point on
the graph to zoom in. Right click at any point
to zoom out.
Zoom Window: Hold down "Ctrl" key and
drag click the left button of the mouse to
select an area on the Graph and the selected
area will be enlarged.
Zoom X-Axis: Hold down "Ctrl" key and
drag click the left button of the mouse to
select an area on the Graph and the X-axis
of the selected area will be enlarged.
Zoom Y-Axis: Hold down "Ctrl" key and
drag click the left button of the mouse to
select an area on the Graph and the Y-axis
of the selected area will be enlarged.
Three icons on the right top side of the main
screen:
Redraw Button: After any review and
zooming on the Graph, click "Redraw" to
reset it to original format.
Clear Button: Click "Clear" to delete all data
and Graph on the window. A warning box