UEi Test Instruments AQM4 User manual

HOLD
MIN MAX AVG
CAL
REC Re call
INSTRUCTION MANUAL
TA
CO
2
DP
CO
AM PM
WBT
O
C
O
%rh
AQM4
Indoor Air Quality Meter
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
INTRODUCTION
Thank you for purchasing the UEi AQM4 IAQ meter. The meter measures CO² levels, CO levels, air temperature, dew point, wet bulb temperature, and humidity. The AQM4 is an ideal instrument for indoor air quality (IAQ) diagnosis.
Poor indoor air quality causes tiredness, inability to concentrate, and even illness (ex. Sick Building Syndrome). IAQ monitoring and surveying especially for CO² levels and air ventilation has become widely applied in public offices, classrooms, factories, hospitals, and hotels. It is also recommended for regulating standards of industrial air quality in some countries.
The portable AQM4 uses NDIR (non­dispersive infrared) technology for CO² measurement to ensure reliability and long­term stability. It is useful in verifying HVAC systems performance and air ventilation control.
CO is a colorless, odorless, and tasteless gas which is slightly lighter than air. It is highly toxic to humans and animals. Carbon Monoxide is measured using a long-life electrochemical sensor.
Maximum indoor air quality CO level per ASHRAE Residential standards 62-1989 for living area is under 9ppm. It is necessary to evacuate employees from enclosed spaces if the CO concentration exceeds 100ppm per OSHA exposure limit.
1
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
FEATURES
• Large LCD display with blue backlight for
use in dark areas
• One touch to display CO² / dew point /
temperature / wet bulb temperature / air temperature / humidity / CO in sequence
• Designed with NDIR waveguide
technology CO² sensor
• Programmable CO level alarm
• Extended drift compensation on
CO² sensor
• Audible alarm (~ 80db) threshold setting
• Max and Min functions included (TWA and STEL for CO² only)
• Mini USB to PC download function
for analysis
• 99 points manual recording
• 8000 points auto recording
• Review 99 points manual record
from meter
• Hold function freezes current readings
• Housing designed to help air ventilation
for quick and accurate response.
• Easy to manually calibrate CO² to fresh air
(380-420ppm)
MATERIAL SUPPLIED
This package contains:
• Meter
• 4 AA batteries
• Mini USB cable and software CD
• Operation manual
• Hard carrying case
• 33% calibration salt
• 75% calibration salt
• AC Power Adapter
2
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
POWER SUPPLY
The meter is powered by either 4 AA batteries or a DC adaptor (9V/1A output.)
Install the batteries into the battery compartment on the rear of the instrument. Verify correct polarity and solid contact.
When an adaptor is used, it will cut off
the power supply from the batteries. The adaptor can not be used as a battery charger.
When battery voltage gets low, will appear on the LCD (Fig. 1.) The CO² sensor
accuracy is compromised under low batteries, replace with fresh batteries or connect to an adaptor.
LCD DISPLAY
PRIMARY LCD
SECONDARY LCD
3
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
SYMBOLS
Primary LCD Measured and calculated
values
Secondary LCD Real time clock
CO² Carbon dioxide reading CO Carbon monoxide reading
HOLD Readings are frozen
MIN/MAX Minimum/maximum readings
STEL Short-term exposure limit (15
minutes weighted average)* TWA Time weighted average (8 hours)*
Low battery indicator
DP Dew point temperature TA Air temperature
WBT Wet bulb temperature %RH Percentage of relative humidity °C/°F Celsius/Fahrenheit scale CAL In calibration status
REC In manual/automatic logging
RECALL In manual records recall mode
*CO² Only
KEYPAD
Turns the meter on or off. Press and hold to enter setup mode while meter is off. Press while turning the meter on to disable auto power off.
Exits setup/recall page. Press and hold to start automatic logging.
Press to switch display mode. Press and hold to enter memory recall mode.
4
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
KEYPAD CONT.
Freezes the current readings. Selects
unit or increases value in setup. Cancels data hold function.
Press to manually record the reading. Selects unit or decreases value in setup.
Displays MIN, MAX, TWA, STEL
function. Saves and finishes settings while in sleep mode.
OPERATION
POWER ON/OFF
Press to turn the meter on and off. At power up, the meter emits a short beep and begins 30 second warm-up countdown
(Fig. 2), then enters normal mode with
current CO² and real time displays (Fig. 3).
NOTE:
THE REAL TIME DISPLAYS DATE (YY-MM-DD) AND TIME IN SEQUENCE.
TAKING MEASUREMENT
The meter starts measurements when powered on and updates readings every second. In the condition of an operating environment change, (ex. high to low temp) it takes 30 seconds for the CO² sensor to respond and 30 minutes for RH sensor.
NOTE: DO NOT HOLD THE METER IN CLOSE PROXIMITY TO THE FACE AS EXHALED BREATH AFFECTS CO² LEVELS.
5
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
CO² (CARBON DIOXIDE)
Press to switch the mode to CO². CO² concentration will display in ppm on primary
display (Fig. 4). The lower display shows the
real time clock.
AIR, (DP, WBT, TA ) TEMPERATURES
Press to switch to temperature display
(Fig. 5). In the AQM4, dew point temperature
and wet bulb temperature are also available
by repeated pressing of (Fig 6). The
lower display shows the real time clock.
HUMIDITY
Press to switch to humidity display
(Fig. 7). The lower display shows the real
time clock.
CO (CARBON MONOXIDE)
Press to switch the mode to CO. CO concentration will display in ppm on primary
display (Fig. 8). The lower display shows the
real time clock.
6
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
DATA HOLD
In normal display mode, press to freeze
the readings, the "HOLD" icon is displayed
on the left top of the LCD (Fig. 9). All current
readings are kept unchanged. Press to cycle through all held readings.
While in "HOLD" mode, STEL and TWA will continue updating every 5 minutes.
BACKLIGHT
The backlight will be activated for 10 seconds by pressing any key.
MIN, MAX, STEL, TWA
This meter allows you to check the minimum
and maximum, STEL and TWA values
calculated since power on. Under normal mode, press to see the minimum,
maximum, STEL and TWA and real time
sequence. Each press of , displays
MIN, MAX, STEL, TWA in sequence and returns to current value (Fig 10 - 13). STEL and TWA values are calculated for CO² only.
If the meter is turned on for less than
15 minutes, the STEL value will be the
weighted average of readings taken since
power on. This is also true for TWA values
displayed prior to 8 hours. It takes at least
5 minutes to calculate STEL and TWA. The display shows "----" during the first 5
minutes from power on.
7
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
In the MAX/MIN/STEL/TWA modes, it
shows the corresponding readings on the main display and the real time clock on the lower display.
In this mode, press to switch the
displayed parameters (Fig 14).
The displayed sequence in this mode is MIN
MAX STEL TWA. Press key to return to
live readings.
"HOLD", "Manually Record", Record" and "Recall" functions are all available when meter is in MIN/MAX/AVG mode.
To clear the MIN/MAX/AVG since power on, restart the meter or press for more than 2 seconds to clear previous data and restart.
NOTE:
TWA & STEL ARE FOR CO2 ONLY. MIN/MAX ARE FOR ALL
PARAMETERS
8
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
MANUALLY RECORDING
The meter features 99 manual memory locations.
In normal or hold mode, press key to record, icon and main display flash for about 3 seconds. The main display shows the memory serial number. The AQM4 has
99 locations. (Fig. 15)
Each memory contains all parameters
(CO², CO, TA....%rh), not limited to the
selected parameter on the main display.
If the reading changes quickly, press to
freeze the reading
before manually recording the data.
99 MEMORIES RECALL
In normal or hold mode, press key for more than 2 seconds until the icon flashes.
Press or to scroll through the memory locations. The memory serial
number displays on the main LCD first and
then readings stored in that location.
(Fig. 16 & 17)
9
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Press key to switch the display
parameter. (Fig. 18)
The time displayed in memory recall mode is the recording time of this memory.
To escape the memory recall mode, press key to leave and return to normal display.
ALARM
The meter features an audible alarm to give warnings when CO concentration exceeds a pre-set limit. (See Pg 12, P20 for setting alarm threshold.) It emits beeps (Abt. 80dB) when CO level goes over the set value and stops only when the readings fall below the set value. It beeps again when value goes over the limit.
The meter can automatically record readings of CO²/CO/TEMPERATURE/RH for extended environment monitoring. The memory capacity is 8,000 points. Users can set up
sampling rate from 1 second to 4 hours, 59 minutes and 59 seconds. (See pg 11, P40)
After sampling rate is selected, press for 2 seconds under normal mode to start logging. The icon flashes to indicate
the logging status and LCD main display
shows the real time CO² measured value.
Lower display is the real time clock. (Fig. 19)
10
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
To terminate data logging, press for 2 seconds, until icon stops flashing.
NOTE:
The AQM4 will only record one session. If is activated all previous data is erased. Download logged data prior to starting another session.
AUTO POWER OFF
The meter turns off automatically after 20 minutes of inactivity. To override the function, hold down and for 2 seconds to turn on the meter until "n" appears. (Fig. 20)
NOTE:
AUTO SLEEP FUNCTION WILL BE DISABLED DURING CALIBRATION MODE
SETUP
When the meter is off,
hold down for more than 1 second to
enter setup mode. (Fig
21) To exit setup, press to return to normal mode.
11
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Press or to scroll through set-up
When the preferred alarm value is set, press
options from the following: P10 - 99 Memory Clear P20 - CO Alarm P30 - Temperature Units
P40 - Logging Sample Rate P50 - Pressure Compensation
P60 - Real Time Clock
P10: 99 MEMORIES CLEAR
P10 and "CLR" (Fig. 21) are displayed on the LCD, press to enter into P11 and clear
or keep all the manual records. The current
setting will be blinking on the LCD. (Fig. 22)
Press or to choose NO or YES
and press to confirm.
-OR- Press to escape and return to P10
P20: CO ALARM
When P20 and "ALAR" (Fig. 23) are displayed on LCD, press to enter into
P21 for setting CO alarm threshold. The
current set value will be blinking on the LCD (Fig 24). Press to increase or
to decrease the value. The selectable alarm
limits are 25~200ppm in 5 ppm increments.
to save the setting
-OR- Press without saving and return to P20.
12
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
P30 TEMPERATURE UNIT
logged readings, press to increase and
When P30 and "UNIT" (Fig 25) are displayed on the LCD, press to go into P31 to
choose the temperature unit.
The set value will display on LCD (Fig 26)
Press or to choose C° or F° and
press to confirm.
-OR- Press to escape and return to P30.
P40: LOGGING SAMPLING RATE
P40 allows you to set sampling rate of data
logging (Fig. 27). The range is from 1 second to 4 hours 59 minutes and 59 seconds.
Press to display current setting and enter edit mode. The display is in the format [HH. MM. SS.] The hour digits will be flashing. To change intervals between
to decrease. Press again to confirm and enter minute setting. Press to confirm and enter seconds setting.
(Fig 28) Press to confirm the rate
setting.
-OR- Press to escape and return to P40.
13
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
P50 PRESSURE COMPENSATION
year digits blinking. Press to increase or
When P50 and "PRES" (Fig. 29) are displayed on the LCD, press to set pressure
compensation value for CO² measurement.
The current set will flash on LCD. (Fig. 30)
The barometric pressure unit is kPa.
Press to increase or to decrease barometric pressure.
-OR- Press to escape and return to P50.
NOTE: ONLY ENTER THIS MODE IF CURRENT BAROMETER PRESSURE IS KNOWN. MULTIPLY IN.
Hg BY 3.39 TO OBTAIN kPa VALUE.
P60: REAL TIME CLOCK
Enter P60 for setting the real time clock of this meter. Press to enter into P61 to set the time format as 12 hour or 24 hour. The
current set will flash on LCD. Press
or to change the format and press to confirm and continue to set the date
and time. (Fig 32)
-OR- Press to escape and return to P60.
NOTE: DATE FORMAT IS YY:MM:DD.
The date will show on the lower display with
to decrease digits. Press to confirm and continue to next setting. Repeat above steps to complete the year/month/day/
hour/minute/second setting. (Fig. 32)
14
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
CALIBRATION MODE
The AQM4 allows you to calibrate CO² and RH, for CO calibration please send unit to UEi. To enter calibration mode, place unit at suitable calibration site specified in the following sections. Press + , then also simultaneously for 3 seconds to turn on the
meter and enter calibration mode (Fig. 33).
Press or to select parameter to calibrate. To cancel or exit calibration at any time, turn meter off.
NOTE:
ENSURE THE BATTERIES ARE AT FULL VOLTAGE BEFORE THE CALIBRATION TO PREVENT INTERRUPTION OR FAILED CALIBRATION.
CO² CALIBRATION
The manual calibration is suggested to be done in sunny outdoor air that is well ventilated. To begin, enter calibration mode. Press or to select parameters for calibration and press enter. The meter is calibrated at standard 400ppm CO² concentration at the factory. It is suggested to manually calibrate regularly to maintain good accuracy.
NOTE:
WHEN THE ACCURACY BECOMES A CONCERN, RETURN TO UEi FOR STANDARD CALIBRATION.
CAUTION: Do not calibrate the meter in the air with unknown CO² concentration. It could be calibrated as 400ppm by default which leads to inaccurate measurements.
Wait 10 minutes until the blinking stops and
the calibration is completed. To cancel the calibration, turn off the meter at any time.
15
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
RH CALIBRATION
To begin, enter calibration mode. Press or to select parameters for calibration and press enter. The meters default for calibration of
humidity is 33% and 75% salt solution. The ambient condition is recommended at 70 ~ 75°F (25°C) and stable humidity. To cancel
calibration, turn off the meter.
CAUTION: Do not calibrate the humidity without default calibration salts. Contact UEi for calibration salt or for service
33% Calibration
Plug the sensor probe into 33% salt bottle. Hold down to enter 33% calibration. "CAL" and calibrating value (32.8% if at 77°F) are blinking on the LCD.
75% Calibration
Plug sensor probe into 75% salt bottle and
enter calibration mode. In calibration mode,
press to select 75% calibration (Fig. 35),
press to start. "CAL" calibrating value
(75.3% if at 77°F) will blink on the LCD.
Meter will begin calibration and finish in 60 minutes when "CAL" and humidity stop blinking. To cancel the calibration, turn off the meter at any time.
16
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
CO CALIBRATION
When the accuracy becomes a concern,
return to UEi for standard calibration. It is necessary to have standard CO gas for calibration.
CAUTION:
Do not calibrate the meter in the air with unknown CO concentration. It could be calibrated as 400ppm by default which leads to inaccurate measurements.
TROUBLESHOOTING
Will Not Power On
- Press for more than 3 seconds.
- Check that the adaptor is connected and on.
- Check that battery level is sufficient and properly installed.
Readings Not Changing
- Check whether data hold function is activated (HOLD icon is at the left top).
Slow Response
- Check the air flow channels on the rear. Verify no blockage exists.
Error Messages
EO1/E33: CO² sensor is out of order. Try new battery first. E02: The value is under range. E03: The value is over range E04: The original data error results in
this error (DP, WB)
17
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Error Messages Cont.
E11: Retry humidity calibration E16: Retry CO calibration E31: Temperature sensor AD damaged E32: Memory IC damaged. E33: Humidity sensor or circuit damaged.
PC CONNECTION
The UEi AQM4 can be connected to a PC to download recorded data.
Plug the USB cable into the socket at the right side of the meter and the other port to your PC. Next, install the software on your PC following these procedures.
1. Insert the CD ROM and run installation. Select a preferred directory and click "Next" step by step to finish.
2. After the software installation is completed, it will run USB driver installation next.
18
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
AUTO CONNECT
Start the software and it will detect the logger connection automatically, indicating the COM port information at the bottom of the main screen as well as the COM Port setting column.
LOGGER SETTINGS
To set up logging plan, click "Setting" icon
and select "Logger". The setting page is
opened.
Users can set up the logger identifier number. Click OK for setting and exit the screen to confirm the settings.
19
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
DATA TRANSMISSION
To transmit auto recorded data from the meter, click "Download" icon on the left top side of the main screen. All auto-recorded data in the logger will be transmitted. The raw data with time stamp will display at the lower part of the screen and the graph in the middle.
In the Graph display, the Y-axis indicates Temperature, RH, DP,WBT, CO2, and CO
levels in different line colors. The X-axis can be switched to show Time or Recorded Number Stamp.
20
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
VIEW MENU
The following functions help to view the GRAPH data in more detailed way.
In the view menu, there are 4 tools to enlarge the Graph data for detailed data review.
Zoom in: Hold down "Ctrl" key on the PC keyboard and left click any target point on
the graph to zoom in. Right click at any point to zoom out.
Zoom Window: Hold down "Ctrl" key and drag click the left button of the mouse to select an area on the Graph and the selected area will be enlarged.
Zoom X-Axis: Hold down "Ctrl" key and drag click the left button of the mouse to select an area on the Graph and the X-axis of the selected area will be enlarged.
Zoom Y-Axis: Hold down "Ctrl" key and drag click the left button of the mouse to
select an area on the Graph and the Y-axis
of the selected area will be enlarged.
Three icons on the right top side of the main screen:
Redraw Button: After any review and
zooming on the Graph, click "Redraw" to
reset it to original format.
Clear Button: Click "Clear" to delete all data and Graph on the window. A warning box
shows up for confirmation. Press "Yes" and
all data will be cleared from the PC.
21
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Cursor Button: Switch cursor ON and it shows the raw data with the cursor. Click anywhere on the graph and the digital data
will be displayed. Switch OFF to disable the
function.
DATA PROCESSING The downloaded data can be Saved or Printed via "File" processing function.
Load: To retrieve saved files, just click
"Load" and select a desired file and it will be loaded in the main screen with the file details at the left top side.
22
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
SPECIFICATION
Measuring Range
CO
²
CO 0 ~ 1000ppm
Temperature -20 ~ 60˚C
Relative Humidity 0.1% ~ 99.9%
Dew Point Temp. -20.0 ~ 59.9˚C
Wet Bulb Temp -5.0 ~ 59.9˚C
Resolution 1ppm, 0.1˚C/˚F, 0.1%RH
Accuracy
CO
²
CO ±10ppm for less than
Temperature ±0.6˚C/+0/9˚F
Relative Humidity ±3% (at 25˚C, 10~90%)
Response Time
CO
²
CO <60 seconds
Tair <2 mins
Relative Humidity <10 mins
LCD/Meter Size (mm) 26(H) x 44 (L),
Operating Condition -20 to 50˚C (CO² Sensor)
Storage Condition -20 ~ 60˚C, 10 ~ 90% RH
Power Supply AA x 4pcs or 9V Adaptor
Battery Life >24 hours
Weight
0 ~ 9999ppm
-5 ~ 140˚F
±30ppm ± 5% of reading (0 ~ 5000ppm)
100ppm
±10% of reading for 101~500ppm ±20% of reading for 501ppm
and above
±5% (at 25˚C, other range)
<30 seconds
(90% step change)
(90% step change)
(90% step change)
(90% step change)
205(L) x 70 (W) x 56(H)
0 to 500˚C (CO Sensor)
-20 to 60˚C
(Alkaline Battery)
200g
23
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
CO² LEVELS AND GUIDELINES
REGULATORY EXPOSURE LIMIT
ASHRAE Standard 62.1-2004:
CO² concentration in occupied building should not exceed 1000ppm.
Building Bulletin 101 (BB101): 1500ppm
UK standards for schools state that CO² averaged over the whole day (i.e. 9am to
3:30pm) should not exceed 1500ppm.
OSHA: 5000ppm
Time weighted average over five 8-hour
work days should not exceed 5000ppm.
Mak: Germany, Japan, Australia, UK:
5000ppm
Eight hours weighted average in
occupational exposure limit is 5000ppm.
(See Ashrae 62.1-2004 Table B-1)
NON-ENFORCED REFERENCE LEVELS
NIOSH Recommends 250-350ppm: normal outdoor ambient
concentrations
600ppm: minimal air quality complaints
600 - 1000ppm: less clearly interpreted
1000ppm: indicates inadequate ventilation;
complaints such as headaches, fatigue, eye/ throat irritation will be more widespread. 1000ppm should be used as an maximum limit for indoor levels.
24
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
CO LEVELS
ppm Symptoms and Applicable Standards
Normal background levels
0-1
Maximum indoor air quality level:
9
Maximum allowable concentration per ASHRAE Residential Standards 62-1989 for living area.
Maximum limit 8 hours of continuous
25
exposure per California OSHA workplace standards
Maximum 8 hours average exposure
35
level per US OSHA workplace standards
Maximum concentration for
50
continuous exposure for any 8 hours average level per OSHA standards
Evacuate employees from continuous
100
exposure in any 8 hour average level per OSHA standards.
Mild headache, fatigue, nausea, and
200
dizziness within 2-3 hours
Frontal headache, life threatening
400
after 3 hours. Maximum concentration in flue gas per the US EPA and AGA standards
Dizziness, nausea, convulsions, death
800
within 2-3 hours
Nausea within 20 minutes, death
1600
within 2-3 hours.
25
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com