• Operating Temperature: 32˚ to 122˚F (0˚ to 50˚C)
Storage Temperature: -4˚ to 140˚F (-20˚ to 60˚C)
•
• Operating Humidity: <80%
• Pollution Degree: 2
• Display: 3 5/6 digits 6,000 count
Back light: Yes
•
• Refresh Rate: 3/sec
Over-range: “OL” is displayed
•
• Apo: Auto power off after 30 minutes of use.
• Dimensions: 9.45” x 2.58” x 1.67”
Item Weight: 0.926 lb.
•
• CAT Rating: CATIV 600V
Certifications: cETLus UL 61010-1: 2012, IP42, 6 ft. Drop Protection
•
• Battery Type: (AAA) 4
• Test Leads: Test leads w/ alligator clips & back probes
Use ATL58 test leads w/ ABP3 back probes and
AAC3 alligator clips.
ATL55, ATL57 and ATLTX will not work with DL599
3
Page 4
Important Safety Warnings
WARNING
Read entire Safety Notes section regarding potential hazards and proper
instructions before using this meter. In this manual the word “WARNING” is
used to indicate conditions or actions that may pose physical hazards to the
user. The word “CAUTION” is used to indicate conditions or actions that may
damage this instrument.
WARNING
To ensure safe operation and service of the tester, follow these instructions.
Failure to observe these warnings can result in severe injury or death.
WARNING
• Before each use, verify meter operation by measuring a known voltage or
current.
• Never use the meter on a circuit with voltages that exceed the category
based rating of this meter.
• Do not use this meter during electrical storms or in wet weather.
• Do not use the meter or test leads if they appear damaged.
• Ensure meter leads are fully seated and keep fingers away from the metal
probe contact when making measurements. Always grip the leads behind
the finger guards molded into the probe. For information on test lead shields
instructions on page 19.
• Do not open the meter to replace batteries while the probes are connected.
• Use caution when working with voltages above 60 DC or 25 AC RMS. Such
voltages pose shock hazards.
• To avoid false readings that can lead to electrical shock, replace batteries if
a low battery indicator appears.
• Unless measuring voltage or current, shut off and lockout power before
measuring resistance or capacitance.
• Always adhere to national and local safety codes. Use proper personal
protective equipment (PPE) to prevent shock and arc blast injury where
hazardous live conductors are exposed.
• Always turn off power to a circuit or assembly under test before cutting,
unsoldering or breaking the current path. Even small amounts of current
can be dangerous.
• Always disconnect the live test lead before disconnecting the common test
lead from the circuit.
• In the event of electrical shock, ALWAYS bring the victim to the emergency
room for evaluation, regardless of victim’s apparent recovery. Electrical
shock can cause unstable heart rhythms that may need medical attention.
• If any of the following occur during testing, turn off the power source to the
circuit being tested: arcing, flame, smoke, extreme heat, smell of burning
materials or discoloration or melting of components.
WARNING
Higher voltages and currents require greater awareness of physical safety
hazards. Before connecting the test leads; turn off power to the circuit under
test, set meter to the desired function and range; connect the test leads to
the meter first, then connect to the circuit under test. Reapply power. If an
erroneous reading is observed, disconnect power immediately and recheck all
settings and connections.
WARNING
This meter is designed to provide HVAC/R technicians with the capabilities they
need to diagnose and repair HVAC/R system. Observe all recommended
safety procedures that include proper lockout utilization and use of personal
protective equipment that includes safety glasses, gloves and flame resistant
clothing.
4
Page 5
Symbols
kHz
BT
EF
T
1
T
2
M
k
AC (Alternating current)DC (Direct current)
Negative AC/DC Voltage or Current
Auto-ranging
Overload: Range Exceeded
Low pass Filter
Auto power off ActiveNon-Contact Voltage
Low BatteryHold/Capture Value
Minimum measured
value displayed
Duty CycleHertz/Frequency
VoltageInrush
AmperageOhms/Resistance
DiodeCapacitance
Nanofarad
MicroampsContinuity
Degrees FahrenheitDegrees Celsius
Mega (x106 or 1,000,000)Milli (x10-3 or 0.001)
Kilo (x103 or 1,000)Micro (x10-6 or 0.000001)
Warning or CautionGround
Dangerous Levels
Safe for disconnect from
live conductors
AVG
Average
Temperature input 1
Wireless Connection
Maximum measured
value displayed
Microfarad
µF
Double Insulation
(Protection to Class II)
No reading detected
Electric Field
Temperature input 2
Mega Ohms
Kilo Ohms
Category Definitions
Measurement
Category
II< 10Circuits connected to mains socket
III< 50Mains distributions parts of the building
IV> 50Source of the mains installation in the
Short-Circuit
(typical) kA
a
outlets and similar points in the MAINS
installation
building
Location in the
building installation
5
Page 6
kHz
kHz
kHz
Overview
M
A
600V600A
DL599
Temp
µA
V
SELECT
Low Z
BT
AUTO
B
CAT IV
N
O
A
A
V
P
Q
U
C
D
E
F
G
H
I
J
K
True RMS
30V
T1
T2
TEMP
R
S
T
V
W
S/N
X
L
A. Clamp: Measure inductive AC/DC current. Opens to 1.25” (32.0mm).
B. Conductor Alignment Marks: Use to aid the visual alignment of a
conductor when measuring inductive amperage. Greatest accuracy
is achieved when the conductor inside the clamp is centered at the
intersection of these marks.
C. Worklight: Lights clamp area in dark work environments.
D. Category Max Indicator: Maximum CAT Rating for clamp jaw.
E. NCV Alert Light: Indicates voltage when in NCV (Non Contact Voltage)
mode and High Voltage alert.
F. Clamp Lever: Opens and closes current clamp jaw.
NOTE: The clamp uses a high-tension spring to close the jaw. Do not
allow fingers or objects to become pinched in the base as the jaws close.
G. Rotary Selector Dial: Set Rotary Selector Dial desired function
H. SELECT/Wireless (BT) Button:
• Press to select AC or DC on Low Z setting, AC or DC on voltage
setting, AC or DC on Amps setting.
• Press to activate Ohms, Continuity, Diode, Capacitance on
Ohms/Continuity/Diode/Capacitance setting; AC or DC on Low amps
setting; T1, T2, T1 - T2 on temperature setting.
• Press and Hold to activate Wireless mode.
I. Range/Mode/
Button:
• Press to set manual range desired
• Press and hold Range/Mode/
Button to activate Low filter mode
J. Min/Max/Test Button:
• Press to capture Max reading
• Press again to capture Min reading
• Press again to capture Avg reading
• Press and hold return to Live readings
• Press and hold Min/Max/Test Button and Range/Mode/
activate Motor measurement test mode
6
OPEN
LOCK
Button to
Page 7
Overview (Cont.)
K. Category Max Indicator: Maximum CAT Rating for input jacks.
L. Test Lead Input Jacks: Multifunction and Positive input jacks.
• Multifunction input port used for measuring: AC or DC volts, resistance,
continuity, diode, capacitance, AC or DC µA.
M. Wire Separation Tab/ NCV sensor: Use to isolate an individual wire from
a bundle for testing. NCV sensor detects live voltage.
N. Test Lead Holder
O. Hand Guide: Used as a point of reference for the operator’s safety.
P. Hold/Worklight/ Back light Button:
• Press to hold the reading on the display. Press again to return to live
reading.
• Press and hold to turn on Worklight and Back light. Press and hold
again to turn off.
• Worklight and Backlight turn off after 60 seconds.
Q. Display:
• High contrast dual display with backlit.
• Amps (AC/DC) reading will always display on upper display.
R. LRA Inrush/Zero/˚C/˚F/Hz/Duty Button:
• Press to Zero the DC Amps reading.
• Press to enter LRA Inrush mode (See page 13 for details)
(must be in AC Volts mode first).
• Press again to return to live readings.
• Press and hold to select ˚F or ˚C in temperature setting
S. K-Type Temperature Probe Inputs: T1 (Left) and T2 (Right))
T. Input Jack Lock: Switch to use Temperature or Test lead inputs
U. Built-in versatile magnet to use as a mount or as a strap
V. Battery Cover: Easy access for replacing batteries without breaking
calibration seal.
W. Battery Compartment Latches:
X. Serial Number
Non-Contact Voltage
NCV Sensor in the tip.
CAT IV
600V 600A
DL599
A
Temp
µA
V
SELECT
Low Z
BT
A
AUTO
V
True RMS
30V
T2
T1
TEMP
• Rotate Rotary Selector Dial to NCV position move the tip of the clamp meter
near voltage source.
• Non-Contact Voltage Detection is used to detect power with sensor located
in the tip of the clamp head, indicates positive response with both an
Audible and Visual alert.
• Do not use Non-contact voltage detector to determine if there is current on
the wire. Detection operation could be affected by socket design, insulation
thickness, type or other factors.
• Voltage indicator light may also light when voltage (>AC/DC 30V) is present
on the meter’s input jack or from an external interference such as motors,
flashlights, etc.
Features:
On Voltage
Approx. 24V AC
7
Page 8
Voltage: 1000V AC/DC
CAT IV
600V600A
DL599
A
Temp
High
Voltage
Alert
WARNING
• Use CATIII rated test leads or higher.
• Do not attempt to measure more than 1000V AC/1000V DC.
• Keep hands below line when measuring high current levels.
• Select AC or DC Voltage.
WARNING
• High Voltage indicator will display and audible alert will sound over 600V
AC/DC
• AC/DC and High Voltage indicator will display (without audible alert) over
30V AC/DC
Features:
AC Volts
RangeResolutionAccuracyOverload Protection
600.0mV0.1mV
6.000V0.001V
60.00V0.01V
600.0V0.1V
1000V1V
True RMS: 45Hz to 400Hz
µA
V
SELECT
Low Z
BT
A
AUTO
• Default = Auto Selection
• Press SELECT x1 =
V
• Press SELECT x2 =
• Press RANGE to select resolution
True RMS
30V
T2
T1
TEMP
SELECT
BT
±1.0% + 3dgts1000V RMS
DC Volts
RangeResolutionAccuracyOverload Protection
600.0mV0.1mV
6.000V0.001V
60.00V0.01V
±0.5% + 4dgts
1000V RMS
600.0V0.1V
1000V1V±0.8% + 5dgts
8
Page 9
Frequency (Hz) / Duty Cycle
CAT IV
600V600A
DL599
A
Temp
µA
V
SELECT
Low Z
BT
A
AUTO
V
True RMS
30V
T2
T1
TEMP
Use CAT III rated leads or higher.
Press the SELECT button to select AC voltage, press and hold the button for
Frequency and Duty Cycle modes.
WARNING
Do not attempt to measure more than 1000V AC/1000V DC.
Features:
SELECT
BT
Frequency Measurement - Test lead input
RangeResolutionAccuracyOverload Protection
99.99Hz0.01Hz
999.9Hz0.1Hz
9.999kHz0.001kHz
99.99kHz0.01kHz
Sensitivity: 1.8Vrms
Duty Cycle - Test lead input
RangeAccuracyOverload Protection
1.0 to 99.0%±(0.2% per kHz + 0.1% + 5 dgts) 1000V RMS
• Default = Auto Selection
• Press SELECT x1 =
• Press and hold ZERO x1 =
• Press and hold ZERO x2 =
• Press RANGE to select resolution
0.1% + 3dgts1000V RMS
Frequency Measurement - Jaw input
RangeResolutionAccuracyOverload Protection
999.9Hz0.1Hz
1999Hz1Hz
0.5% + 5dgts600V RMS
Sensitivity: 5A
9
Page 10
dd
600V 600A
DL599
SELECT
BT
AUTO
Low Z (Low Impedance)
CAT IV
600V600A
DL599
A
Temp
µA
V
SELECT
Low Z
BT
A
AUTO
V
True RMS
30V
T2
T1
TEMP
Low Pass Filter
CAT IV
A
µA
Temp
V
Low Z
True RMS
30V
T1
T2
TEMP
A
V
•
With Selector dial in voltage position or
Amps position.
•
Press SELECT x1 =
•
Press RANGE button x1 to enter Manual Range
• Press and hold
Low Pass Filter
• Press and hold
Live readings
•
This function is also activated in Amps.
Features:
Low Z
•
Rotate Selector Dial to Low Z
• Default = Auto Selection
• Press SELECT x1 =
• Press SELECT x2 =
Features:
SELECT
BT
RANGE button to activate
RANGE button to return to
SELECT
BT
10
Page 11
AC/DC Amps <600A Jaw
Single
CAT IV
600V 600A
DL599
A
Temp
µA
V
SELECT
Low Z
BT
AUTO
True RMS
30V
T2
T1
TEMP
• Center wire in guides for best accuracy.
• Opposing currents cancel each other (use line-splitter when necessary).
• Keep hands below guard when measuring high current levels.
• Do not attempt to measure more than 600A AC / 600A DC.
Features:
SELECT
BT
AC Amps Measurement - Jaw input
RangeResolutionAccuracyOverload Protection
60.00A0.01A±2.0% + 5dgts
600.0A0.1A±1.8% + 5dgts
45Hz to 400Hz True RMS
Minimum Current for Clamp Measurement: 0.3A
DC Amps Measurement - Jaw input
RangeResolutionAccuracyOverload Protection
60.00A0.01A±2.0% + 5dgts
600.0A0.1A±1.8% + 5dgts
Minimum Current for Clamp Measurement: 0.2A
Conductor
Only
A
V
• Default = Auto Selection
• Press SELECT x1 =
• Press SELECT x2 =
600V RMS
600V RMS
11
Page 12
A
AC/DC Microamps: <2000µA
CAT IV
600V600A
DL599
A
Temp
µA
V
SELECT
Low Z
BT
A
AUTO
V
True RMS
30V
T2
T1
TEMP
WARNING
• Do not attempt to measure more than 2000µA.
Features:
SELECT
BT
DC Microamps Measurement -Test lead input
RangeResolutionAccuracyOverload Protection
600.0uA0.1uA
2000uA1uA
AC Microamps Measurement -Test lead input
RangeResolutionAccuracyOverload Protection
600.0uA0.1uA
2000uA1uA
45Hz to 400Hz True RMS
• Default = Auto Selection
• Press SELECT x1 =
• Press SELECT x2 =
• Readings show on lower display
±1.2% + 3dgts600V RMS
±1.5% + 3dgts600V RMS
Zero DC Amps
CAT IV
600V 600A
DL599
A
Temp
µA
V
SELECT
Low Z
BT
A
AUTO
V
True RMS
30V
T2
T1
TEMP
Select DC current.
• Press to zero any offset in Amps DC.
• Used to monitor change from present displayed value.
• Required during DC Amps measurement to establish a zero level.
WARNING
Do not use DC Zero mode at amps greater than 600A DC.
• Default = Auto Selection
• Press SELECT x1 =
• Press SELECT x2 =
• Press ZERO = DC Amps Zero
• Press RANGE to select resolution
12
Page 13
LRA Inrush
CAT IV
600V600A
DL599
A
Temp
µA
V
SELECT
Low Z
BT
A
AUTO
V
True RMS
30V
T2
T1
TEMP
• Default = Auto Selection
• Press SELECT x1 =
• Press ZERO = Inrush
The UEi LRA Inrush is programmed to properly capture the starting current for
compressor motors.
• Select AC Amps.
• Select the range capable of capturing the maximum value.
• Press the ZERO button – INRUSH will now be shown on the screen.
• Activate the compressor and read value on the display.
• Press the Zero button to return to live readings.
Phase Rotation
1. Set Dial to V range
2. Press and hold RANGE and MIN/MAX button to enter Phase rotation mode.
Upper display shows “3Pha”
3. Press Min/Max button to enter standby mode
Upper display blinks “L1L2”
NOTE: Each of steps 4 -5 needs to be conducted within 5 seconds. If not, the
display will indicate “oVEr” and the meter will finish the testing.
4. Connect test leads to any two lines (L1 – L2) of 3-Phase voltage lines.
A beep indicates stable voltage is captured, and the Upper display shows
“L1L3”
5. Remove Red test lead from the voltage line
A beep and “L3” blinking on display will indicate test is ready.
6. Connect Red test lead to the third voltage line (L1 – L3)
A beep indicates test is complete
Upper display indicates phase status (Forward or Reverse).
13
Page 14
Voltage Unbalance
1. Set the Dial to V range
2. Press and hold RANGE and MIN/MAX button to enter Phase rotation mode.
Press RANGE button to enter Voltage Unbalance test mode
Upper display indicates “V1V2”
Press Min/Max button to enter Testing standby mode.
“V1V2” will blink on display
NOTE: Steps 3 - 4 needs completed within 30 seconds. If not, the display will
indicate “oVEr” and the meter will finish the testing.
3. Connect test leads to any two lines (V1 – V2) of the 3-Phase voltage lines
A beep indicates stable voltage is captured, and the Upper display shows
“V1V3”
4. Remove Red Test lead from the voltage line
A beep and “V1V3” blinking on display will indicate test is ready.
Connect the Red test lead to the third voltage line (V1 – V3)
5. A beep indicates test is complete
Upper display will indicate test result (PASS or FAIL)
Lower display will indicate the percentage:
PASS % unbalance <2%
FAIL % unbalance > 2%
% V unbalance = (Max deviation from Avg Voltages/Avg Voltage) x 100
Current Unbalance
1. Set the Dial to A range
2. Press and hold RANGE and MIN/MAX to enter Current Unbalance test mode
Lower Display will indicate “A”
3. Press Min/Max button to enter testing standby mode
“-A1-” will blink on display
NOTE: Each of steps 4 – 6 needs to be conducted within 30 seconds. If not,
the display will indicate “oVEr” and the meter will finish the testing.
4. Clamp the jaw around current line (A1)
A beep indicates stable current is captured, and the Lower display shows
“-A2-“
Remove clamp jaw from current line
A beep and “-A2-“ blinking on display will indicate test is ready.
5. Clamp the jaw around current line (A2)
A beep indicates stable current is captured, and the Lower display shows
“-A3-“
Remove clamp jaw from the current line
A beep and “-A3-“ blinking on display will indicate test is ready.
6. Clamp the jaw around current line (A3)
A beep indicates stable current is captured
Upper display indicates test results (PASS or FAIL)
Lower display will indicate the percentage:
PASS % unbalance <10%
FAIL % unbalance > 10%
% I unbalance = (Max deviation from Avg Currents/Avg Currents) x 100
14
Page 15
Resistance: < 60MΩ
k
T
F
1
CAT IV
600V 600A
DL599
A
Temp
µA
V
SELECT
Low Z
BT
A
AUTO
V
True RMS
30V
T2
T1
TEMP
• Default =
• Press RANGE to select resolution
Features:
SELECT
BT
WARNING
• Do not measure resistance on a live circuit.
RangeResolutionAccuracyOverload Protection
600.0Ω0.1Ω
6.000kΩ0.001kΩ
60.00kΩ0.01kΩ
600.0kΩ0.1kΩ
±0.8% + 3dgts
600V RMS
6.000MΩ0.001MΩ
60.00MΩ0.01MΩ±1.2% + 3dgts
Temperature F˚/C˚
CAT IV
600V600A
DL599
A
TEMP
Temp
µA
V
SELECT
Low Z
BT
A
AUTO
V
True RMS
30V
T2
T1
TEMP
Temp
• Default = T1
• Press SELECT x1 = T2
• Press SELECT x2 = T1 - T2
• Press and hold ZERO = F° or C°
• Readings show on lower display
• Disconnect test lead probes from voltage source and meter.
• Move Input Jack Locks to “TEMP” setting.
• Use K-T
ype thermocouple temperature probes only.
• Stated accuracy does not account for thermocouple accuracy.
Features:
SELECT
BT
RangeResolutionAccuracyOverload Protection
-328˚F to 999˚F
(-200˚ to 999˚C)
1000˚F to 2462˚F
(1000˚ to 1350˚C)
0.1°F (0.1˚C)
1°F (1˚C)
±(1.0% +3.6˚F)
±(1.0% + 2.0˚C)
30V RMS
15
Page 16
V
Temperature Calibration
The new DL599 clamp meter offers a digital field temperature calibration
procedure to make the process easier. Here are the steps.
1) Turn on DL599 while holding down RANGE button.
2) When ‘FLD1’ is displayed on LCD, press the ZERO button.
3) After entering Calibration mode, set dial to Temp position.
4) Confirm DL599 is in T1 mode.
5) After immersing T1 temperature probe in ice bath and temperature is
stable, press and hold ZERO button to save.
6) Press SELECT button to switch to T2 mode.
7) After immersing the T2 temperature probe in ice bath and temperature is
stable, press and hold ZERO button to save.
Diode
GOOD DIODE
CAT IV
600V600A
Reverse Bias
Displays "OL"
DL599
A
Temp
µA
V
SELECT
Low Z
BT
AUTO
• Default = Ω
• Press SELECT x2 =
A
V
Forward Bias
Displays approx.
voltage drop
True RMS
30V
T1
T2
TEMP
BAD DIODE
Open Diode
Displays "OL"
Both directions
• Forward voltage drop if forward biased.
• “O.L.” if reverse biased.
Features:
Diode Test
Range Open Circuit VTest CurrentOverload Protection
3.1V< 3.2V DC0.25mA600V RMS
SELECT
BT
or
'0' Both directions
16
Page 17
Capacitance
F
CAT IV
600V 600A
DL599
A
Temp
µA
V
SELECT
Low Z
BT
A
AUTO
True RMS
30V
T2
T1
TEMP
• Default = Ω
V
• Press SELECT x3 =
• Press RANGE to select resolution
Features:
SELECT
BT
RangeResolutionAccuracyOverload Protection
10.00nF
100.0nF
1.000uF
10.00uF
100.0uF
9999uF
0.01nF
0.1nF
0.001uF
2.5% + 5dgts
0.01uF
0.1uF
1uF3.0% + 5dgts
600V RMS
WARNING To avoid damaging the meter or equipment under test, safely
discharge Capacitors before measuring capacitance. Large value capacitors
should be discharged through an appropriate resistance load. Use the DC
Voltage function to confirm the capacitor discharge.
Continuity
CAT IV
600V 600A
DL599
A
Temp
µA
V
SELECT
Low Z
BT
A
AUTO
V
True RMS
30V
T2
T1
TEMP
• Default =
• Press SELECT x1 =
• Buzzer sounds at less than < 40Ω.
WARNING
• Do not measure resistance on a live circuit.
Features:
SELECT
BT
Open Circuit V <1.00VOverload Protection
Threshold Approx. <40Ω600V RMS
17
Page 18
App Overview
• In the App stores Google Play and the App Store, search “DL599”.
• App is compatible with iOS® 12.0 and up and Android™ 6.0 and up.
• If searching for the iPad version, you may have to filter for “iPhone” only.
• Turn on DL599 and launch the DL599 App on device.
• On DL599, Press and hold SELECT / BT. Beep confirms wireless mode
activated. BT shows in display.
• On Device, Press CONNECT
• Devices should pair.
AF
B
C
D
E
A. Menu screen
•
Connection Status
•
Settings screen
•
Info (Manual)
•
UEi App
B. BT connection indication
C. Graphing screen: View live graphing
D. Display: Reading mirrors DL599 display
E. Function Buttons: Operation mirrors DL599 function buttons
•
3-Phase Rotation, Voltage Unbalance and Current Unbalance test
processes must be started from the meter, but can be completed from
the App screen.
F. Log screen
•
Tap to select a log file to view
•
Swipe left to edit (rename or add memo notes) or delete
SUMMARY: Default
•
In the summary screen, the red / button is for
displaying a graph of the JAW measurement value, and the
black / button is for displaying a graph of the Test
lead Input jack measurement value.
• Each graph can be shown or hidden by pressing the button.
SAMPLE: View individual samples from readings
EXPORT: Select a file format and send data
: To return to previous screen
HOME: To exit to main menu
G. Record / Stop Button
H. Sample count: Only visible while logging
G
H
18
Page 19
Test Lead Notes
4mm
Cat IV and CAT II Measurement Locations
4mm
• Ensure the test lead shield is pressed firmly in place. Failure to use the CAT
IV shield increases arc-flash risk.
CAT II Measurement Locations
18mm
• CAT IV shields may be removed for CAT II locations. This will allow testing
on recessed conductors such as standard wall outlets. Take care not to lose
the shields.
WARNING: Test Lead category protections apply only to test leads and
should not be confused with the meter’s specific CAT rating. Observe the
maximum category protection indicated on the meter the test leads are
plugged into.
CAUTION: If the test leads need to be replaced, you must use a new one
which should meet EN 61010-031 standard, rated CATIII 1000V or better.
NOTE: DL599 works with UEi ATL58 test leads, ABP3 back probes and AAC3
alligator clips.
ABP3 Back Probes screw-on to test leads, after removing CATIV shields from
test leads. Back Probes come with covers on probe tips.
FCC INFORMATION
NOTE: This equipment has been tested and found to comply with the limits
for a Class B digital device, pursuant to part 15 of the FCC rules. These limits
are designed to provide reasonable protection against harmful interference
in a residential installation.
This equipment generates, uses and can radiate radio frequency energy and,
if not installed and used in accordance with instructions, may cause harmful
interference to radio communications. However, there is no guarantee that
interference will not occur in any particular installation. If this equipment
does cause harmful interference to radio or television reception, which can
be determined by turning the equipment on and off, the user is encouraged
to try and correct the interference by one or more of the following measures:
• Reorient or relocate the receiving antenna.
• Increase the separation between the equipment and receiver.
• Connect the equipment into an outlet on a circuit different from that to
which the receiver is connected.
• Consult the dealer or an experienced radio/TV technician for help.
CAUTION
manufacturer responsible for compliance could void the user’s authority to
operate the equipment.
(1) this device may not cause interference, and
(2) this device must accept any interference, including interference that may
cause undesired operation of the device.
Changes or modifications not expressly approved by the
19
Page 20
Warranty
LOCK
OPEN
OPEN
The DL599 is warranted to be free from defects in materials and workmanship
for a period of 2 years from the date of purchase. If within the warranty period
your
instrument should become inoperative from such defects, the unit will
be repaired or replaced at UEi’s option. This warranty covers normal use and
does not cover damage which occurs in shipment or failure which results
alteration, tampering, accident, misuse, abuse, neglect or improper
from
maintenance. Batteries and consequential damage resulting from failed
batteries are not covered by warranty.
Any
implied warranties, including but not limited to implied warranties of
merchantability and fitness for a particular purpose, are limited to the express
warranty. UEi shall not be liable for loss of use of the instrument or other
incidental or consequential damages, expenses, or economic loss, or for any
claim or claims for such damage, expenses or economic loss.
Warranty only covers hardware and does not extend to software applications.
purchase receipt or other proof of original purchase date will be required
A
before warranty repairs will be rendered. Instruments out of warranty will be
repaired (when repairable) for a service charge.
This warranty gives you specific legal rights. You may also have other rights,
which vary from state to state.
Battery Replacem
ent
• Rotate Battery Compartment Latches to open
position
• Remove battery cover
• Replace the old batteries with 4 new (AAA)
batteries
• Replace the battery cover
• Rotate Battery Compartment Latches to lock
LOCK
Disposal
CAUTION: This symbol indicates that equipment and its accessories
shall be subject to separate collection and correct disposal.
Cleaning
Periodically clean your meter’s case using a damp cloth. DO NOT use abrasive,
flammable liquids, cleaning solvents, or strong detergents as they may damage the
finish, impair safety, or affect the reliability of the structural components.
Storage
Remove the batteries when instrument is not in use for a prolonged period of time.
Do not expose to high temperatures or humidity. After a period of storage in extreme
conditions exceeding the limits mentioned in the General Specifications section, allow
the instrument to return to normal operating conditions before using it.