perform a 60 second purge. This should be done in fresh
air.
Rotate Selector Dial to MENU to set up or customize
your settings.
Selected parameter are centered, highlighted with arrow
icons on either side.
Use
Press
and buttons to scroll menu options,
to select and edit a parameter.
Press and buttons to change contents of a field.
Press button to scroll fields.
Press to enter content and move to next field.
STATUS SCREEN
Rotate Selector Dial to STATUS to view parameters. Selected FuelTi = Ambient Temperature
ATM = Atmospheric Pressure (measured in mbar)
CAL = Days remaining until calibration
CO/CO2
O2/Eff
Aux
Status
v3
C
Temp
Exch Test
P
M
E
T
2
O
C
O
C
C163
o
r
e
m
z
y
b
l
u
a
s
n
t
i
A
o
n
Press
and buttons to scroll Status Bar
options.
Time, Pump Status, Battery Status
Fuel Type
Date
Change Fuel Type in STATUS SCREEN
Pressand buttons to the currently selected fuel.
Press and Hold button until arrows indicate fuel
edit mode is activated.
The
and to scroll fuel types.
Press
to select desired fuel.
CO/CO2
O2/Eff
Aux
Status
v3
C
Temp
Exch Test
P
M
E
T
2
O
C
O
C
C163
o
r
e
m
z
y
b
l
u
a
s
n
t
i
A
o
n
BEFORE TESTING
CHECKLIST
o Clean particle filter
o Water trap and probe line are
free of condensate
o All hose and thermocouple
connections are properly
secured
o Water trap is fitted correctly
o Power on and zero
o Flue gas probe is sampling
ambient FRESH air
Remove the rubber boot and red
drain plug. Allow the water to
drain out. Re-insert the rubber
plug and replace boot cover.
COMBUSTION ANALYSIS
T2T1
P2
P1
Remove protective rubber boot.
Remove water trap from analyzer.
Remove particle filter from water
trap and replace with new one.
Reattach trap and boot.
CO/CO2
Temp
O2/Eff
Aux
Status
v3
C
o
Exch Test
P
M
E
T
2
O
C
O
C
C163
r
e
m
z
y
b
l
u
a
s
n
t
i
A
o
n
Insert flue probe into combustion
input. Insert K-type temperature
probe into input T1.
Insert flue probe into stack.
Adjust cone so the probe tip is
approximately at the center of
Rotate function rotary dial
to O2/Eff for Test Screen 1
parameters.
the stack.
NOTE: If there is no hole, check with service and installation instructions of appliance for best testing
method or check with your local authority on best practice
Rotate the function rotary dial
to CO/CO2 for Test Screen 2
parameters.
Exch Test
P
M
E
T
2
O
C
O
C
C163
r
e
m
z
y
b
l
u
a
s
n
t
i
A
o
n
CO/CO2 displays CO (ppm), CO2,
Losses calculated, Excess air %.
Make any adjustments needed
for proper Combustion and wait
for analyzer to display change in
readings. (Repeat as necessary).
CHANGE FUEL TYPE
CO/CO2
Temp
O2/Eff
Aux
Status
v3
C
o
m
b
C163
u
s
t
i
Exch Test
P
M
E
T
2
O
C
O
C
r
e
z
y
l
a
n
A
o
n
Rotate the dial to either STATUS,
AUX, O2/Eff, CO/CO2 or Temp
setting.
Press button until the Fuel
Type displays on the Status Bar.
Press and Hold until two
Press and buttons to
scroll fuel selections.
Press to select fuel.
arrow icons display.
AUXILIARY SCREEN
CO/CO2
Temp
O2/Eff
Aux
Status
v3
C
o
m
When performing a combustion
Test, Rotate function rotary dial
to AUX to view the Auxiliary
Test Screen’s custom parameters.
C163
b
u
Exch Test
P
M
E
T
2
O
C
O
C
r
e
z
y
l
a
s
n
t
i
A
o
n
AUX displays 5 lines of editable
parameters. To edit line values
Press button until Edit?
displays on the Status Bar.
Press and Hold until the
Status Bar reads LINE 1.
Press
and buttons to
scroll parameter selections.
Press
to select a parameter
and advance to the next Line.
Continue until cursor is back to
Status Bar to exit.
The AUX screen allows you to assign any of the following parameters on any display line.
O2, CO, Display Fuel Type, CAL, XAIR, LOSS, Efgc, ATM, Ti, ∆T, T2, T1, CO2, COa, or COn. Pressure(C163 and !64 only). NOX
and No (C164 only)
TEMPERATURE TEST
T2T1
Insert K-type temperature
probes into input T1 and T2.
CO/CO2
Temp
O2/Eff
Aux
Status
v3
C
o
Exch Test
P
M
E
T
2
O
C
O
C
C163
r
e
m
z
y
b
l
u
a
s
n
t
i
A
o
n
Rotate function rotary dial to
TEMP for Temperature Test
Screen.
To Print: Align printer (sold
separately) above Analyzer.
VIEW SAVED LOGS
Rotate the dial to MENU.
Press
is selected between two arrow
icons. Press to select.
button until the LOGS
During testing, Press to start
printing.
Press again to abort.
Screen displays all saved LOGS.
With VIEW selected in the Status
Bar, Press to select.
To Log: During testing, Press and
Hold
to log a report.
To view these Logs, see View Saved Logs section.
Press and buttons to
scroll Test Report categories.
Press to select a category.
Line 1 displays the LOG number
on screen.
To view another LOG, Press
and buttons to NEXT.
Press to advance to next log.
Repeat until you reach desired LOG.
To print the LOG, Press
and buttons until PRINT
is selected between two arrow
icons. Press to select to start
printing.
To EXIT, Press and
buttons until EXIT is selected
between two arrow icons. Press
to EXIT screen.
Repeat until return to MENU screen.
POST TESTING
Remove the probe from the flue and
allow analyzer to purge with fresh
air until readings return to zero. O2
to 20.9% and CO below 20 ppm
CAUTION: probe tip will be HOT.
BATTERIES
The analyzer is fitted with 3
(AA) size rechargeable batteries. Observe correct polarity
whe n replacing.
Always check operation
immediately after battery
replacement.
Do not leave batteries in unit for
extended periods of non-use.
POWERING OFF
When you power off the Analyzer, there is a 10 second purge.
CAUTION
Make sure you do not exceed the analyzer’s operating specifications.
Do not exceed the flue probes maximum temperature (1112˚F)
•
•
Do not exceed the analyzer’s internal temperature range 112°F (50°C)
Do not put the analyzer on a hot surface
•
•
Do not exceed the water trap’s level
Do not let the particle filter become dirty and blocked
•
WARNING
Do not recharge with Alkaline
batteries fitted. Alkaline batteries
are not rechargeable. Attempting
to recharge alkaline batteries
may result in product damage
and create a fire risk.
Time and Date will need to be
reset after changing batteries.
Always dispose batteries using
approved disposal methods that
protect the environment.
WARRANTY
The C162/C163/C164 combustion analyzer is warranted for one year including sensors. If within the
warranty period your instrument should become inoperative from such defects, the unit will be repaired
or replaced at UEi’s option. This warranty covers normal use and does not cover damage which occurs
in shipment or failure which results from alteration, tampering, accident, misuse, abuse, neglect or
improper maintenance (calibration). Batteries and consequential damage resulting from failed batteries
are not covered by warranty. Any implied warranties, including but not limited to implied warranties
of merchantability and fitness for a particular purpose, are limited to the express warranty. UEi shall
not be liable for loss of use of the instrument or other incidental or consequential damages, expenses,
or economic loss, or for any claim or claims for such damage, expenses or economic loss. A purchase
receipt or other proof of original purchase date will be required before warranty repairs will be rendered.
Instruments out of warranty will be repaired (when repairable) for a service charge. Warranty only covers
hardware and does not extend to software applications or accessories. Contact UEi for specifi
and service information.
T
his warranty gives you specific legal rights. You may also have other rights which vary from state to state.
c warranty
ANNUAL RECERTIFICATION SERVICES
Complete care for your C160 series combustion analyzer.
BACKED BY
When you:
Request Recertification or Service Online
Within 1 Year of Purchase or Service
UEi will:
6-Year Warranty: All UEi combustion analyzers have a standard
1-year warranty. Each recertification includes a 30-point inspection
and extends the warranty for 1 more year for up to 6 years from the
date of warranty.
24-hour Service:
and service through UEi Service+ are shipped back on the next
business day.
Free Shipping: UEi Service+ offers free shipping both to and from
our service center. When customers book their recertifi
receive a prepaid UPS Ground shipping label.
Analyzers received for standard recertification
cation, they
Register Online
Registering you analyzer online is quick and easy. Just log in or setup an account, it only takes
a couple of minutes. Once logged in you can register you analyzer by providing some product
information and uploading a proof-of-purchase.
When it’s time to request recertifcation, just log into your account, select the analyzer, select
the service and place your order.
Test Equipment Depot - 800.517.8431 - 5 Commonwealth Ave, MA 01801