UEi CML100 Instruction Manual

CLM100
INSTRUCTION MANUAL
99 Washington Street Melrose, MA 02176
Phone 781-665-1400 Toll Free 1-800-517-8431
Visit us at www.TestEquipmentDepot.com
The CLM100 Cable Length Meter is ideal for measuring cable length on the spool or in existing conduits. Includes test leads, copper calibration bar, an abrasive pad for removing oxidation and carrying case.
Features include
• 3 digit display with enunciators
Measures in FEET (ft) or METERS (m)
• Measures COPPER (CU) or ALUMINUM (AL) wire
• Resistance range for milliohm measurements
• Automatic temperature compensation
• User calibration mode. Calibration standard included
• Auto power off
WARNING!
Make sure wire under test is not energized. Never apply voltage to inputs.
NOTE: Temperature affects accuracy of readings. Please see accuracy specifications. For best results, allow the CLM100 to attain the same ambient temperature as the wire under test. The length of time this will take depends on the ambient temperature. Typically it will take 10 to 15 minutes for the CLM100 to attain equal ambient temperature.
Safety Tips
Before using this meter, read all safety information carefully. In this manual the word "WARNING" is used to indicate conditions or actions that may pose physical hazards to the user. The word "CAUTION" is used to indicate conditions or actions that may damage this instrument.
International Symbols
Operating Instructions
Calibration Procedures
1. Turn the CLM100 on by depressing the green on/off button
2. Insert the black test leads into one pair of input jacks and the red test leads into the other pair of input jacks. (Ignore the gray marking on the banana plug.) This is very important to ensure that the meter works correctly. If black and red test leads are mixed the unit will give a false reading. (Fig 1)
3. Connect the Kelvin clips to the calibration standard. Make sure the clips are as close together as possible. (Fig 2)
NOTE: Make sure the calibration standard is clean. The Kelvin clips must be connected across the diameter of the
standard.
Use the abrasive pad to clean the calibration standard.
NOTE: Make sure clips are as close together as possible. Use the tips of the Kelvin clips when performing measurements on small diameter wire.
4.
Press and hold the “CAL” button until all segments in the
display illuminate. This completes the calibration procedure and will ensure all measurements are as accurate as possible.
5.
Disconnect the leads from the standard.
Measuring Length of Wire
1. Turn the CLM100 on and perform the calibration procedure.
2. Allow the CLM100 to attain the same temperature as the wire under test.
NOTE: Temperature affects accuracy of readings. Please
see accuracy specifications. For best results, allow the CLM100 to attain the same ambient temperature as the wire under test.
WARNING!
Make sure wire under test is not energized. Never apply voltage to inputs.
CLM100-MAN P. 1
Loading...
+ 2 hidden pages