Fuel type can be changed via
MENU or STATUS setting
TO
ZERO
Flue 2
Flue1
Com Test
Aux
Status
MenuRoom CO
Prs/Temp
Exch Test
Press ”Pump” to ZERO
(0) pressure sensor.
Rotate selector dial to Flue 2 for:
CO (ppm), Losses calculated, Excess
air %, pressure reading
Fuel type can be changed via
MENU or STATUS setting.
Make any adjustments as needed for
proper combustion and wait for analyzer to display change in readings.
(Repeat as necessary).
T2 T1
P1 P2
Connect probe to T1.
Connect probe to T2.
Compatible with any K-Type
thermocouple probe or clamp.
Once complete, remove probe from
stack, allow analyzer to purge in
fresh air until CO sensor readings
return to ZERO (0) and O2 readings
reading return to 0% to 21%.
Continue to the next test or power
analyzer off by pressing On/Off, if
finished.
NOTE: Print and store functions may
be used at any time during testing.
PRS 0.04i
i= inH20
T1 167,1F
T2 104.3F
T 62.8F
Flue 2
Flue1
Com Test
Connect thermocouples in
test location to start testing
Observe T1, T2 and
differential (Delta T).
T1 167,1F
T2 104.3F
T 62.8F
Flue 2
Flue1
Com Test
Aux
Status
MenuRoom CO
Prs/Temp
Exch Test
Press “Print” to print
results.
Press and Hold to log
results.
Page 2
DRAFT & STATIC PRESSURE TEST
PRS/TEMP
PRESS
TO
ZERO
PRS/TEMP
EXCHANGE
TEST
PRS 0.04i
i= inH20
T1 167,1F
T2 104.3F
T 62.8F
Flue 2
Flue1
Com Test
Aux
Status
Prs/Temp
Exch Test
Press “Print” to print
results.
Press and Hold to log
results.
Flue 2
Flue1
Com Test
Aux
Status
MenuRoom CO
Prs/Temp
Exch Test
C165
Rotate the selector dial to
Pressure/Temperature.
PRESS
TO
ZERO
Flue 2
Flue1
Com Test
Aux
Status
MenuRoom CO
Prs/Temp
Exch Test
Press ”Pump” to ZERO
(0) pressure sensor.
T2 T1
P1 P2
Connect true draft hose to PI.
Connect to P2 for differential.
Place probe tip in stack to
measure draft.
Connect static pressure hose
Place true draft probe tip in the
flue to measure draft.
You can also use the
combustion probe for
measuring pressure.
for measure differential
pressure.
HEAT EXCHANGER TEST
There are many methods to test heat exchanger integrity. One of these is to observe Excess Air, O2 AND CO readings both before and after the blower turns
on. If the heat exchanger is sealed your O2 and CO readings should remain fairly stable. A breach in the heat exchanger may allow fresh air to be forced into
the flue after the blower turns on due to a pressure increase in the plenum. The result may be a rise in the measured O2 in the stack gas and an increase in the
Excess Air. In some sealed systems the fresh air drawn in the breach may reduce the combustion available leading to an increase in the CO reading. If either of
these situations are present it is probable that there is a problem with the Heat Exchanger which may require additional testing and inspection.
NOTE: Many cracks are invisible to borescopes or the naked eye and only open or separate from pressure or temperature during operation.
TO
START
PRESS
Flue 2
Flue1
Com Test
Aux
Status
MenuRoom CO
Rotate selector dial to
EXCH Test, press the
“Print”.
BLOWER ON
NAT GAS
CO 0p
O2 20.0%
XAIR 02++%
TIME 60
Flue 2
Flue1
Com Test
Aux
Prs/Temp
The analyzer will wait 60
seconds and then record the
Post-Blower values for CO,
02 and Excess Air.
Prs/Temp
Exch Test
SETUP
FUEL TYPE
NAT GAS
PRESS
Flue 2
Flue1
Com Test
Aux
Prs/Temp
Confirm the Fuel Type.
Scroll or to switch
a different fuel type.
Press to select fuel
type.
LOG No. 10
12:00:00
CO 0p
O2 20.0%
XAIR 02++%
PRINT
Flue 2
Flue1
Com Test
Aux
Prs/Temp
Test results will automatically be stored to exchange
reports. Report includes both
Pre and post Blower test
segments.
Press to start printing.
See REPORTS section in
this guide.
EXCHANGE
WITH PROBE
IN FRESH
AIR
PRESS
Flue 2
Flue1
Com Test
Aux
Prs/Temp
NOTE: The analyzer will
perform a fresh air purge
if it did not perform one on
start up. Otherwise it will
skip past the zeroing step to
the Blower Off screen. See
manual for details.
Press to continue.
EXCHANGE
ZEROING
TIME 60
Flue 2
Flue1
Com Test
Aux
Prs/Temp
Ensure the analyzer is
in fresh air during the
countdown. The screen
will automatically advance
when complete.
BLOWER OFF
NAT GAS
CO 0p
O2 20.0%
XAIR 02++%
PRESS
Flue 2
Flue1
Com Test
Aux
Prs/Temp
Call for heat.
Observe and wait for O2
readings to stabilize.
After the blower turns on,
press the Button to
start the Post-Blower test.
Page 3
O2 20.9%
Status
Exch Test
NAT GAS
Ra 0.0010
CO 80P
C02 8.3%
XAIR42.4%
PRS 0.03m
CO ROOM TEST
Status
Exch Test
Great for checking ambient CO and back drafting situations
SELECT
ROOM CO
TEST TYPE
GENERAL
Flue 2
Flue1
Com Test
Aux
Status
MenuRoom CO
Prs/Temp
Exch Test
No probes or hose connections
required for this test.
Rotate selector dial to ROOM CO.
Press ”Print” to start the
ROOM CO
PRESS
TO
START
Press start the test.
CO readings will be logged every 2
minutes for a 30 minute time span.
Readings will be stored once the 30
minute test is completed.
ROOM CO
TEST 1/15
INT. 120s
CO 5p
LIMIT 10p
ALARM 30p
GENERAL test.
AUX SETTING
Programmable Auxiliary screen allows for Tech selectable test parameters to be chosen.
MENU
SETUP
UNITS
SCREEN
REPORT
SERVICE
Flue 2
Flue1
Com Test
Prs/Temp
Aux
Status
MenuRoom CO
Exch Test
SCREEN
CONTRAST
BACKLIGHT
AUX
BACK
Flue 2
Flue1
Com Test
AUX
LINE 1
LINE 2
LINE 3
LINE 4
LINE 5
Flue 2
Flue1
Com Test
SCREEN
AUX
LINE 5
PRS 0.02i
ROOM CO
LOG No. 1
TEST 1
CO 5p
GENEAL
PRINT
Flue 2
Flue1
Com Test
Press to print results.
Test results will automatically be
stored in ROOM CO reports.
See REPORTS section in this guide.
CO 0P
C02 8.3%
XAIR 42.4%
PRS -0.02i
Ta 80.3F
Flue 2
Flue1
Com Test
Aux
Status
MenuRoom CO
Prs/Temp
Exch Test
Rotate selector dial to Menu.
Scroll or to Screen.
Press ”Print” to select.
REPORTS: PRINTING & VIEWING
Print and Log easily from the following test screens: Aux, Flue 1, Flue 2, Commission Test, Pressure/Temperature, Exchange Test and Room CO
Press ”Print” to start
printouts of results from any
test screen.
Press during printing to
cancel.
HIGH CO ALERT
Flue 2
Flue1
Com Test
Aux
Status
MenuRoom CO
Prs/Temp
Exch Test
C165
Scroll or to AUX.
Press to select.
NAT GAS
Ra 0.0010
CO 80P
C02 8.3%
XAIR 42.4%
Press and hold to log
current readings.
Align printer above Analyzer.
Scroll or to customize
any Line.
Press to select.
[Line 5 shown]
MENU
SETUP
UNITS
SCREEN
REPORT
SERVICE
Flue 2
Flue1
Com Test
Prs/Temp
Aux
To view test results,
rotate dial to Menu.
Scrollor to Report
Press to select.
Scroll or to select
the parameter for this line.
Press to select.
[Pressure shown]
REPORT
AUX
COMBUSTION
COMBUSTION
PRS/TEMP
EXCHANGE
Flue 2
Flue1
Com Test
Prs/Temp
Aux
Status
Exch Test
Scroll or to type
of test wanted.
Press to select.
[Exchange shown]
Press to select View.
Rotate selector dial to AUX
to view.
Custom selections will now
display in AUX mode.
LOG No. 9
12:00:00
CO 0p
02 20.9%
XAIR O2++%
PRINT
Flue 2
Flue1
Com Test
Prs/Temp
Aux
Press to select
report log number. When
indicator lights blink,
use or to scroll
through logged reports.
Press to view a
selected report log.
At 1000 ppm CO the screen will display HIGH CO and an indicator light will flash and analyzer will beep several times. Above 2000 ppm the purge pump starts.
Page 4
MAIN MENUSUB MENUOPTIONS / COMMENTS
SETUPLANGUAGEEnglish
SET TIMEHH:MM:SS format
e.g. 7 am = 07:00:00, 7 pm = 19:00:00
SET DATEMM/DD/YY format
PRINTERKM IRP
KANE IRP-2
WIRELESS (if installed)
SERIAL
PASSKEY1111 (wait 5 secs after entering last digit)
BACK
UNITSFUEL TYPENAT GAS, TOWN GAS, COKE GAS, PROPANE, BUTANE, LPG, LIGHT OIL, BIO OIL, WOOD
PELLETS, BIO GAS, USER 1 to 5
FUEL ORIGINUK, FRANCE, SPAIN, N AMERICA, BELGIUM, NETHERLAND
EFFICIENCYGROSS, NET, GROSS COND, NET COND
PRESSUREFILTER: OFF = normal response. ON = slower (damped) response
RESOLUTION: LOW = e.g. 0.00i inH20 resolution. HIGH = displays to an extra decimal place