UEi C165 Quick Start Manual

Page 1
GETTING STARTED
FRESH
AIR
PURGE
NAT GAS
C02 8.3%
O2 20.9%
EfG O2++%
Tf 42.4%
Aux
Flue1
Flue 2
Com Test
FRESH AIR
PURGE
TURN
OFF
TIME 5
Prs/Temp
NAT GAS
CO 80P
COa 80P
LOSS O2++%
NAT GAS
C02 8.3%
O2 20.9%
EfG O2++%
PRS/TEMP
PRESS
PRS 0.04i
i= inH20
PRS/TEMP
PRESS
C165 QuiCk Start Guide
TIME 60
Flue 2
Flue1
Com Test
Aux
Status
Menu Room CO
Press
in fresh air and allow to count-
”Power”
down for Fresh Air purge.
Prs/Temp
Exch Test
Rotate selector dial to Flue 1. In fresh air, O2 reading should be 20.9% ±0.3%.
Rotate selector dial to Flue 2. In fresh air, CO reading should be zero (0).
BASIC CO/COMBUSTION ANALYSIS
Tf 42.4%
Ta 0.03m
Flue1
Aux
Status
Menu Room CO
Ta 0.03m
T2 T1
Flue 2
Flue1
Aux
Status
Menu Room CO
Com Test
Prs/Temp
Exch Test
P1 P2
Connect the flue probe thermocouple connector to T1. Connect flue probe to water trap. Connect optional probe to T2.
XAIR O2++%
PRS 0.00i
Flue 2
Flue1
Flue 2
Com Test
Prs/Temp
Exch Test
Status
Menu Room CO
Com Test
Aux
Prs/Temp
Exch Test
Insert Flue probe into stack. Adjust cone so end of the probe is
approximately at the center of the stack.
NOTE: You will have to drill a hole at least 3/8”. Use appropriate sealing method after testing.
DIFFERENTIAL TEMPERATURE TEST
TO
ZERO
Rotate the selector dial to Pressure/Temperature.
Flue 2
Flue1
Aux
Status
Menu Room CO
Com Test
Prs/Temp
Exch Test
Rotate selector dial to Flue 1 for: CO2, O2, Gross efficiency, Flue temperature, Inlet temperature.
Fuel type can be changed via MENU or STATUS setting
TO
ZERO
Flue 2
Flue1
Com Test
Aux
Status
Menu Room CO
Prs/Temp
Exch Test
Press ”Pump” to ZERO (0) pressure sensor.
Rotate selector dial to Flue 2 for: CO (ppm), Losses calculated, Excess air %, pressure reading
Fuel type can be changed via MENU or STATUS setting. Make any adjustments as needed for
proper combustion and wait for ana­lyzer to display change in readings. (Repeat as necessary).
T2 T1
P1 P2
Connect probe to T1. Connect probe to T2. Compatible with any K-Type
thermocouple probe or clamp.
Once complete, remove probe from stack, allow analyzer to purge in fresh air until CO sensor readings return to ZERO (0) and O2 readings reading return to 0% to 21%.
Continue to the next test or power analyzer off by pressing On/Off, if finished.
NOTE: Print and store functions may be used at any time during testing.
PRS 0.04i
i= inH20
T1 167,1F
T2 104.3F
T 62.8F
Flue 2
Flue1
Com Test
Connect thermocouples in test location to start testing
Observe T1, T2 and differential (Delta T).
T1 167,1F
T2 104.3F
T 62.8F
Flue 2
Flue1
Com Test
Aux
Status
Menu Room CO
Prs/Temp
Exch Test
Press “Print” to print results.
Press and Hold to log results.
Page 2
DRAFT & STATIC PRESSURE TEST
PRS/TEMP
PRESS
TO
ZERO
PRS/TEMP
EXCHANGE
TEST
PRS 0.04i
i= inH20
T1 167,1F
T2 104.3F
T 62.8F
Flue 2
Flue1
Com Test
Aux
Status
Prs/Temp
Exch Test
Press “Print” to print results.
Press and Hold to log results.
Flue 2
Flue1
Com Test
Aux
Status
Menu Room CO
Prs/Temp
Exch Test
C165
Rotate the selector dial to
Pressure/Temperature.
PRESS
TO
ZERO
Flue 2
Flue1
Com Test
Aux
Status
Menu Room CO
Prs/Temp
Exch Test
Press ”Pump” to ZERO (0) pressure sensor.
T2 T1
P1 P2
Connect true draft hose to PI. Connect to P2 for differential. Place probe tip in stack to
measure draft. Connect static pressure hose
Place true draft probe tip in the flue to measure draft.
You can also use the combustion probe for measuring pressure.
for measure differential pressure.
HEAT EXCHANGER TEST
There are many methods to test heat exchanger integrity. One of these is to observe Excess Air, O2 AND CO readings both before and after the blower turns on. If the heat exchanger is sealed your O2 and CO readings should remain fairly stable. A breach in the heat exchanger may allow fresh air to be forced into the flue after the blower turns on due to a pressure increase in the plenum. The result may be a rise in the measured O2 in the stack gas and an increase in the Excess Air. In some sealed systems the fresh air drawn in the breach may reduce the combustion available leading to an increase in the CO reading. If either of these situations are present it is probable that there is a problem with the Heat Exchanger which may require additional testing and inspection.
NOTE: Many cracks are invisible to borescopes or the naked eye and only open or separate from pressure or temperature during operation.
TO
START
PRESS
Flue 2
Flue1
Com Test
Aux
Status
Menu Room CO
Rotate selector dial to EXCH Test, press the
“Print”.
BLOWER ON
NAT GAS
CO 0p
O2 20.0%
XAIR 02++%
TIME 60
Flue 2
Flue1
Com Test
Aux
Prs/Temp
The analyzer will wait 60 seconds and then record the Post-Blower values for CO, 02 and Excess Air.
Prs/Temp
Exch Test
SETUP
FUEL TYPE
NAT GAS
PRESS
Flue 2
Flue1
Com Test
Aux
Prs/Temp
Confirm the Fuel Type. Scroll or to switch
a different fuel type.
Press to select fuel type.
LOG No. 10
12:00:00
CO 0p
O2 20.0%
XAIR 02++%
PRINT
Flue 2
Flue1
Com Test
Aux
Prs/Temp
Test results will automati­cally be stored to exchange reports. Report includes both Pre and post Blower test segments.
Press to start printing. See REPORTS section in
this guide.
EXCHANGE
WITH PROBE
IN FRESH
AIR
PRESS
Flue 2
Flue1
Com Test
Aux
Prs/Temp
NOTE: The analyzer will perform a fresh air purge if it did not perform one on start up. Otherwise it will skip past the zeroing step to the Blower Off screen. See manual for details.
Press to continue.
EXCHANGE
ZEROING
TIME 60
Flue 2
Flue1
Com Test
Aux
Prs/Temp
Ensure the analyzer is in fresh air during the countdown. The screen will automatically advance when complete.
BLOWER OFF
NAT GAS
CO 0p
O2 20.0%
XAIR 02++%
PRESS
Flue 2
Flue1
Com Test
Aux
Prs/Temp
Call for heat. Observe and wait for O2
readings to stabilize. After the blower turns on,
press the Button to start the Post-Blower test.
Page 3
O2 20.9%
Status
Exch Test
NAT GAS
Ra 0.0010
CO 80P
C02 8.3%
XAIR 42.4%
PRS 0.03m
CO ROOM TEST
Status
Exch Test
Great for checking ambient CO and back drafting situations
SELECT
ROOM CO
TEST TYPE
GENERAL
Flue 2
Flue1
Com Test
Aux
Status
Menu Room CO
Prs/Temp
Exch Test
No probes or hose connections required for this test.
Rotate selector dial to ROOM CO.
Press ”Print” to start the
ROOM CO
PRESS
TO
START
Press start the test.
CO readings will be logged every 2 minutes for a 30 minute time span.
Readings will be stored once the 30 minute test is completed.
ROOM CO
TEST 1/15
INT. 120s
CO 5p
LIMIT 10p
ALARM 30p
GENERAL test.
AUX SETTING
Programmable Auxiliary screen allows for Tech selectable test parameters to be chosen.
MENU
SETUP
UNITS
SCREEN
REPORT
SERVICE
Flue 2
Flue1
Com Test
Prs/Temp
Aux
Status
Menu Room CO
Exch Test
SCREEN
CONTRAST
BACKLIGHT
AUX
BACK
Flue 2
Flue1
Com Test
AUX
LINE 1
LINE 2
LINE 3
LINE 4
LINE 5
Flue 2
Flue1
Com Test
SCREEN
AUX
LINE 5
PRS 0.02i
ROOM CO
LOG No. 1
TEST 1
CO 5p
GENEAL
PRINT
Flue 2
Flue1
Com Test
Press to print results. Test results will automatically be
stored in ROOM CO reports. See REPORTS section in this guide.
CO 0P
C02 8.3%
XAIR 42.4%
PRS -0.02i
Ta 80.3F
Flue 2
Flue1
Com Test
Aux
Status
Menu Room CO
Prs/Temp
Exch Test
Rotate selector dial to Menu. Scroll or to Screen.
Press ”Print” to select.
REPORTS: PRINTING & VIEWING
Print and Log easily from the following test screens: Aux, Flue 1, Flue 2, Commission Test, Pressure/Temperature, Exchange Test and Room CO
Press ”Print” to start printouts of results from any test screen.
Press during printing to cancel.
HIGH CO ALERT
Flue 2
Flue1
Com Test
Aux
Status
Menu Room CO
Prs/Temp
Exch Test
C165
Scroll or to AUX.
Press to select.
NAT GAS
Ra 0.0010
CO 80P
C02 8.3%
XAIR 42.4%
Press and hold to log current readings.
Align printer above Analyzer.
Scroll or to customize any Line.
Press to select.
[Line 5 shown]
MENU
SETUP
UNITS
SCREEN
REPORT
SERVICE
Flue 2
Flue1
Com Test
Prs/Temp
Aux
To view test results, rotate dial to Menu.
Scroll or to Report Press to select.
Scroll or to select the parameter for this line.
Press to select.
[Pressure shown]
REPORT
AUX
COMBUSTION
COMBUSTION
PRS/TEMP
EXCHANGE
Flue 2
Flue1
Com Test
Prs/Temp
Aux
Status
Exch Test
Scroll or to type of test wanted.
Press to select.
[Exchange shown]
Press to select View.
Rotate selector dial to AUX to view.
Custom selections will now display in AUX mode.
LOG No. 9
12:00:00
CO 0p
02 20.9%
XAIR O2++%
PRINT
Flue 2
Flue1
Com Test
Prs/Temp
Aux
Press to select report log number. When indicator lights blink, use or to scroll through logged reports.
Press to view a selected report log.
At 1000 ppm CO the screen will display HIGH CO and an indicator light will flash and analyzer will beep several times. Above 2000 ppm the purge pump starts.
Page 4
MAIN MENU SUB MENU OPTIONS / COMMENTS
SETUP LANGUAGE English
SET TIME HH:MM:SS format
e.g. 7 am = 07:00:00, 7 pm = 19:00:00
SET DATE MM/DD/YY format
PRINTER KM IRP
KANE IRP-2 WIRELESS (if installed) SERIAL
PASSKEY 1111 (wait 5 secs after entering last digit)
BACK
UNITS FUEL TYPE NAT GAS, TOWN GAS, COKE GAS, PROPANE, BUTANE, LPG, LIGHT OIL, BIO OIL, WOOD
PELLETS, BIO GAS, USER 1 to 5
FUEL ORIGIN UK, FRANCE, SPAIN, N AMERICA, BELGIUM, NETHERLAND
EFFICIENCY GROSS, NET, GROSS COND, NET COND
PRESSURE FILTER: OFF = normal response. ON = slower (damped) response
RESOLUTION: LOW = e.g. 0.00i inH20 resolution. HIGH = displays to an extra decimal place
UNITS: mbar, Pa, PSI, mmHg, hPa, inH2O, mmH20, kPa, psi
TIME: Test units not available in North American market
BACK:
GAS ppm, ppm(n), mg/m3, mg/m3(n), mg/kWh, mg/kWh(n)
TEMP C, F
02 REF Up/down to set value (3% default)
NOx CALC Up/down to set value (5% default)
BACK
SCREEN CONTRAST Factory setting is 14
BACKLIGHT 0 to 300 secs
AUX Enables users to customize the parameters on the AUX display: LINE 1, LINE 2, LINE 3,
LINE 4, LINE 5, LINE 6, BACK
BACK
REPORT AUX Stored AUX tests
VIEW, DEL ALL, BACK
COMBUSTION Stored combustion tests:
VIEW, DEL ALL, BACK
COMMISSION Stored commission tests:
VIEW, DEL ALL, BACK
PRS/TEMP Stored pressure tests:
VIEW, DEL ALL, BACK
EXCH Stored exchange tests:
VIEW, DEL ALL, BACK
ROOM CO Stored room CO tests:
VIEW, DEL ALL, BACK
HEADER LINE 1
LINE 2 BACK
BACK
SERVICE CODE Password protected for authorized service agents only. Leave set to 000000.
www.ueitest.com Email: info@ueitest.com
USA: 1-800-547-5740 • 503-643-6322
CANADA: 1-877-475-0648 • 604-278-4311
Copyright © 2019 Kane USA, Inc. All Rights Reserved L466 0119
Loading...