Failure and error messages1.
Failure encountered
Possible cause Corrective action
Display Err - 300 Excessive cuff pressure. Cuff
inflated to maximum due to
movement of the arm or body.
- Repeat measurement
- Do not move arm
- Do not talk
Display Err - 1 Measurement error. Measure-
ment could not be carried out
due to movement of the arm
or body.
- Repeat measurement
- Do not move arm
- Do not talk
Display Err - 2 Inflation error, pressure built up
too quickly or too slowly.
Check seating of air connector,
repeat measurement.
Fault was recognised. - Repeat measurement
- Do not move arm
- Do not talk
Irregular heartbeat, disruptive
movements, shaking, wobbling (objects), breathing in
deeply, etc. Irregular heartbeat,
disruptive movements, shaking,
wobbling (objects), breathing in
deeply, etc.
Repeat measurement after
3-5 minutes rest. Possible factors
(depending on the severity of
the arrhythmia) affecting the
measurement results must be
discussed with your doctor.
Measured values
too high
Was the necessary rest observed
prior to measurement?
Repeat measurement after a
break of approx. 3-5 mins.
- Do not move arm
- Do not talk
Unusual measured
values
Movement or talking during
measurement, resting time not
observed, feet possibly crossed,
smoking or coffee cunsumption.
Check conditions and repeat measurement. Observe instructions for
use P. EN-38.
Failure encountered
Possible cause Corrective action
Display 0
0
Start/Stop button accidentally
pressed while battery was
being changed.
Switch unit off and then on again
with Start/Stop button.
Measurement was
interrupted and
cuff first deflated
and then inflated
Unit detects a fault or that
diastolic value is too low.
Unit shows no fault. Repeat
measurement.
Movement during measurement
If necessary, interrupt measurement, measure again after a break
of 5 minutes.
No display after
unit is switched on
Batteries inserted incorrectly? Check position of batteries.
Batteries flat? Change batteries.
Battery compartment dirty? Clean battery compartment.
Dashes in display
after Memory button is pressed
No measured values stored. No display when no measured
values are stored.
Batteries flat? Change batteries.
Measurement
interrupted
Batteries flat? Change batteries.
Measured values
extremely high
or low
Was the correct measurement
position maintained?
Repeat measurement, ensuring
correct body posture. Do not move
your arm and do not speak.