A
B
ON
FLOW
ON/OFF
OFF
high
flow
adjustment
low
VENTURA
OXYGEN
INLET
CPAP
AIR
INTAKE
USER GUIDE
EMERGENCY USE ONLY
C
ventura
high
oxygen
adjustment
low
ventura
TO
patient
not FOR ROUTINE CLINICAL
USE WITHOUT FURTHER
APPROVAL
ALL HOSPITALS MUST
WORK WITH THEIR OXYGEN
engineering TEAMS TO
ASCERTAIN THEIR VIE
OUTFLOW, AND DOWNSTREAM
FLOWS AND PRESSURES TO
SPECIFIC WARD AREAS, BEFORE
DEPLOYING THIS DEVICE
v2.0
ventura
ONLY TO BE USED WITH THE
PATIENT CIRCUIT PROVIDED
USING THE VENTURA CPAP
flow generator
A
b
c
oxygen analyser
(MODELS VARY)
Patient
Interface
1. Turn all controls (A, B, C) OFF (clockwise) and connect
the UCL-Ventura to the 4-bar oxygen supply directly or by
using an extension hose.
2. Connect the ow generator to the patient circuit shown.
3. Turn the FLOW ON/OFF control (A) fully ON
(anticlockwise) and open the FLOW ADJUSTMENT control
(B) (3-4 complete rotations anticlockwise).
4. Increase the oxygen concentration by turning the OXYGEN
ADJUSTMENT control (C) (anticlockwise) to the desired
level.
5. Apply mask to the patient with appropriate straps and
ensure a good t.
6. Alter the FLOW ADJUSTMENT control (B) to ensure
optimal ow to:
• Flutter the mask outlet valve (larger port) throughout
the respiratory cycle (the valve will utter less during
inspiration).
safety peep
valve
(20 cmh 0)
2
PEEP VALVE
(cpap: 5-10 cmh 0)
INSPIRATION
2
POSITION
DURING
flow flag
indicator
INCREASE
FLOW
CORRECT
FLOW
REDUCE
FLOW
• Hold the blue ow indicator ag (where tted) horizontal
or just below horizontal on inspiration (the ag will
remain below horizontal on expiration).
• A virtually continuous ow at the outlet lter should
be felt on the back of a gloved hand throughout the
respiratory cycle.
7. Alter the OXYGEN ADJUSTMENT control (C) if the oxygen
concentration has changed.
If the device becomes damaged it should be
withdrawn from use.