Ucl Ventura User Manual

A
B
ON
FLOW ON/OFF
OFF
high
flow adjustment
low
VENTURA
OXYGEN
INLET
CPAP
AIR
INTAKE
USER GUIDE
EMERGENCY USE ONLY
C
ventura
high
oxygen adjustment
low
ventura
TO
patient
not FOR ROUTINE CLINICAL USE WITHOUT FURTHER APPROVAL
v2.0
ventura
ONLY TO BE USED WITH THE PATIENT CIRCUIT PROVIDED
USING THE VENTURA CPAP
flow generator
A
b
c
oxygen analyser
(MODELS VARY)
Patient
Interface
1. Turn all controls (A, B, C) OFF (clockwise) and connect the UCL-Ventura to the 4-bar oxygen supply directly or by using an extension hose.
2. Connect the ow generator to the patient circuit shown.
3. Turn the FLOW ON/OFF control (A) fully ON (anticlockwise) and open the FLOW ADJUSTMENT control (B) (3-4 complete rotations anticlockwise).
4. Increase the oxygen concentration by turning the OXYGEN ADJUSTMENT control (C) (anticlockwise) to the desired level.
5. Apply mask to the patient with appropriate straps and
ensure a good t.
6. Alter the FLOW ADJUSTMENT control (B) to ensure
optimal ow to:
• Flutter the mask outlet valve (larger port) throughout
the respiratory cycle (the valve will utter less during
inspiration).
safety peep
valve
(20 cmh 0)
2
PEEP VALVE
(cpap: 5-10 cmh 0)
INSPIRATION
2
POSITION
DURING
flow flag
indicator
INCREASE
FLOW
CORRECT
FLOW
REDUCE
FLOW
Hold the blue ow indicator ag (where tted) horizontal or just below horizontal on inspiration (the ag will remain below horizontal on expiration).
A virtually continuous ow at the outlet lter should be felt on the back of a gloved hand throughout the respiratory cycle.
7. Alter the OXYGEN ADJUSTMENT control (C) if the oxygen concentration has changed.
If the device becomes damaged it should be
withdrawn from use.