u blox 1EIQ24NN, G7C2 User Manual

G7
Technical User Manual
CONTENTS
OVERVIEW
The World’s rst everything detector ........................................................................................... 4
Blackline Safety Network .................................................................................................................... 4
Blackline Live web portal .................................................................................................................... 5
G7 models .................................................................................................................................................. 6
Device comparison ................................................................................................................................. 7
What’s in the box .................................................................................................................................... 8
Hardware details ...................................................................................................................................... 9
INTERACTION
How it works ............................................................................................................................................... 10
OPERATING
Charging ...................................................................................................................................................... 11
Wearing ........................................................................................................................................................ 11
Power on ..................................................................................................................................................... 12
Power o ..................................................................................................................................................... 12
Connectivity light .................................................................................................................................... 13
Viewing G7’s assigned user ................................................................................................................14
GAS DETECTION
Bump test ....................................................................................................................................................15
Calibration .................................................................................................................................................. 17
Bump test and calibration reminder window..........................................................................18
Bump test and calibration lock ........................................................................................................ 18
Zero sensors ................................................................................................................................................19
PID Target gases........................................................................................................................................19
GAS DETECTION FEATURES
Yellow warning alarm gas features ................................................................................................20
Low warning alarm for gas ...................................................................................................... 20
Under limit ..................................................................................................................................... 21
Sensor error ..................................................................................................................................... 21
Calibration ....................................................................................................................................... 21
Bump test .......................................................................................................................................... 21
Pump block ...................................................................................................................................... 21
Red alert gas features ............................................................................................................................ 22
High alert for gas ........................................................................................................................... 22
STEL (short term exposure limit) alert ............................................................................... 22
TWA (time weighted average) alert ....................................................................................23
OL (over limit) alert ...................................................................................................................... 23
CONFIGURATION MODES
Available modes ....................................................................................................................................... 24
Entering a mode ....................................................................................................................................... 24
Exiting a mode .......................................................................................................................................... 25
CARTRIDGES
Gas Cartridge replacement program ............................................................................................ 26
Changing cartridges .............................................................................................................................. 26
Cartridge care ............................................................................................................................................ 27
Cartridge safety precautions ............................................................................................................. 27
Multi-gas pump cartridge ................................................................................................................... 28
Using pump ..................................................................................................................................... 28
Pump options ................................................................................................................................. 29
REALTIME FEATURES
Yellow pending alarm features ........................................................................................................ 30
Potential Fall detected ............................................................................................................... 30
Potential no-motion detected ...............................................................................................31
Check-in request ........................................................................................................................... 31
Yellow warning alarm features .........................................................................................................32
Messages ........................................................................................................................................... 32
Speaker phone ............................................................................................................................... 33
Network connection interruption ....................................................................................... 33
Low battery ...................................................................................................................................... 33
Red alert features ..................................................................................................................................... 34
Fall detected .................................................................................................................................... 34
No-motion detected ................................................................................................................... 34
Missed check-in ............................................................................................................................. 35
SOS alert ............................................................................................................................................. 35
LiveResponse ............................................................................................................................................. 36
Push-to-talk (PTT) .................................................................................................................................... 37
Transmitting ..................................................................................................................................... 37
Receiving ........................................................................................................................................... 37
Available Channels ....................................................................................................................... 37
Changing channels ...................................................................................................................... 38
Changing volume ......................................................................................................................... 38
PTT audio Accessories ................................................................................................................ 39
FIRMWARE UPDATES
Over-the-air (OTA) rmware updates ........................................................................................... 40
SUPPORT
Learn more .................................................................................................................................................. 41
Customer Care ........................................................................................................................................... 41
SPECIFICATIONS
Detailed specications .......................................................................................................................... 42
LEGAL NOTICES AND CERTIFICATIONS
Legal notices .............................................................................................................................................. 44
Intrinsically safe ......................................................................................................................................... 45
OVERVIEW
THE WORLD’S FIRST EVERYTHING DETECTOR
What is G7 Insight?
G7 Insight is a simple solution that bundles industry-leading gas detection with automated compliance and business analytics tools. For the rst time ever, the days of manually collecting data from the eld, reviewing spreadsheets and compiling reports are behind you.
With just a few minutes of connectivity per day, all data automatically recorded by G7 devices in the eld is uploaded to the Blackline Safety Network for automatic compilation and report generation.
What are real-time features?
Real-time safety monitoring leverages location technology, communications and live data to provide complete transparency and control over your safety program. Real-time features utilize G7 devices’ cellular or satellite connectivity, connecting people in the eld with live monitoring teams in real time — ensuring no call for help will ever go unanswered again.
In the event of an safety incident or gas exposure, monitoring personnel can see what has happened, and communicate with workers directly via two-way voice calling or text messaging through their G7 device.
BLACKLINE SAFETY NETWORK
How am I connected?
The Blackline Safety Network is a cloud-hosted system comprised of 2G/3G cellular networks, satellite networks, our Blackline Live™ web portal application, your monitoring account and your G7.
G7 devices require an active service plan in order to connect to the Blackline Safety Network. Depending on your needs and requirements, various service plan options are available such as 24/7 safety monitoring by Blackline’s Safety Operations Center and two-way voice communication. Contact your organization’s safety professional for more information regarding the details of your service plan.
4 || G7 TECHNICAL USER MANUAL
BLACKLINE LIVE WEB PORTAL
What is Blackline Live?
G7 utilizes the cloud-hosted Blackline Live web portal to monitor and manage all your workers and devices, as well as deliver reports and business analytics insights.
With Blackline Live’s real-time alerting and live map with employee locations, you can quickly locate and respond to a worker in distress. Real-time alerts show the employees location on the map with the type of alert, enabling your team to eciently send help.
Blackline Live also allows you to create and customize conguration proles that determine how a device or a group of devices operates in the eld. Similarly, alert proles are set up to determine what contacts should be notied in the event of an incident and what response protocol monitoring personnel will follow to ensure your team gets the help it needs.
Blackline Live keeps track of alert history, calibrations and bump tests, eliminating the need to manually retrieve data logs from the eld.
Blackline Live allows you to tailor user access depending on employee roles: employee, supervisor, administrator and monitoring team. This ensures that everyone has access to the right tools to accommodate their role in a comprehensive monitoring program.
What is Blackline Analytics?
With Blackline Analytics, review data collected from your device eet to make decisions, follow up with your team and ensure everything is running smoothly. Choose from a number of dierent reports and lters to explore your data.
Blackline Analytics is built directly into the Blackline Live portal, allowing anyone with login access to see their own organization’s data, as well as client data. If users have only been given access to specic groups of devices, they will only see data attached to those particular devices.
G7 TECHNICAL USER MANUAL || 5
G7 MODELS
G7
C
G7
X
What G7 model do I have?
There are two G7 models: G7c and G7x. The main dierence between them is how they connect to the Blackline Safety Network.
If you are unsure of which G7 model you have, refer to the logo on the front of your G7.
G7c works anywhere with 2G/3G cellular coverage in over 200 countries to connect you directly to the Blackline Safety Network. Depending on your service plan, your G7c may have two-way voice capabilities.
6 || G7 TECHNICAL USER MANUAL
G7x works in conjunction with Blackline’s
G7 Bridge — a portable satellite base station that keeps you connected in remote locations outside cellular coverage. G7x uses a 900 MHz radio to communicate with G7 Bridge up to 2km away. One G7 Bridge can link up to ve G7x devices to the Blackline Safety Network through Iridium satellite or cellular data.
DEVICE COMPARISON
What features does my G7 have?
G7c and G7x are customized with one of four cartridge types. Cartridge selections include Standard, Single-gas, Multi-gas diusion and Multi-gas pump cartridges. The following comparison chart summarizes the features of each cartridge.
Text messaging
Fall detection
No-motion detection
SOS alert
Check-in timer
Conguration modes
Receive voice calls*
Push-to-talk enabled*
Single gas detection
Multi gas detection
Pump enabled
Low warning alarm for gas
Under limit alarm for gas
High alert for gas
STEL alert for gas
TWA alert for gas
Over limit alert for gas
*G7c devices only
Standard
Single
Multi
(diusion)
l l l l
l l l l
l l l l
l l l l
l l l l
l l l
l l l l
l l l l
l
l l
l l l
l l l
l l l
l l l
l l l
l l l
Multi
(pump)
l
G7 TECHNICAL USER MANUAL || 7
WHAT’S IN THE BOX
Your G7 device comes with the following components:
 G7 personal safety monitoring device  Pre-installed cartridge (Standard, Single-gas or Multi-gas cartridge)  Getting started guide  Optional real-time features guide  Charging system
Removable charging clip
USB cable
USB power adapter
 Certication and support card
If you have a Single-gas or Multi-gas cartridge you will also receive:
 Single-gas or Multi-gas calibration cap (depending on your cartridge)  Calibration gas tubing
8 || G7 TECHNICAL USER MANUAL
HARDWARE DETAILS
OK
Cartridge (Standard, Single-gas or Multi-gas)
Top lights
Pending and warning Alerts
LiveResponse
Connectivity light
LCD screen
Up button
OK button
Power button
Microphone (G7c only)
Metal belt clip
Speaker
Down button
Latch pull
Latch push button
Charging light
Cartridge label
Product label
Charging port
G7 TECHNICAL USER MANUAL || 9
INTERACTION
HOW IT WORKS
Interacting with G7 is easy with its high-visibility LCD display and three-button menu system.
OK button
Press OK to enter the main menu on the LCD screen and to conrm a menu selection.
Latch pull
Pull down the latch to call for help when assistance is required.
Up and down arrow buttons
Press up or down to navigate the menu. Press and hold both simultaneously to mute a yellow warning alarm or red alert.
Latch push button
Push the latch in to check in and tell your G7 that you are safe.
10 || G7 TECHNICAL USER MANUAL
OPERATING
CHARGING
How do I charge my G7?
Insert the micro USB plug into the removable charging clip, then slide the clip onto the charging port at the bottom of your G7. A solid red light at the bottom of the device conrms your G7 is charging. The LCD screen will let you know when it is fully charged, which can take up to four hours.
Blackline recommends that you fully charge your device after every shift.
Charging port
Removable charging clip
Charging cable
NOTE: If you have G7 PowerPack accessory attached to your device, simply slide
the charging clip into PowerPack’s charging port. Both PowerPack and G7 will be charged simultaneously.
WEARING
Where do I wear my device?
G7 monitors you best when clipped to your belt or chest pocket.
G7 TECHNICAL USER MANUAL || 11
POWER ON
How do I turn on my G7c?
Press and hold the power button, and wait for the blinking green connectivity light to turn solid. When connected, the green light will remain solid.
How do I turn on my G7x?
Press and hold the power button on your G7 Bridge to turn it on rst, and wait for the blinking green connectivity light to turn solid, it takes approximately two minutes for G7 Bridge to connect to the Blackline Safety Network. When connected, the green light will remain solid.
Press and hold the power button on your G7x. Wait for the blinking green connectivity light to turn solid. When connected, the green light will remain solid.
NOTE: For best results, ensure your G7 Bridge is turned on before your G7x. If you are having
diculties connecting your G7x to G7 Bridge, please refer to your G7 Bridge manual.
POWER OFF
How do I turn o my G7c?
Press and hold the power button. The device will go into shutdown sequence, once all the lights and vibrations have stopped, you have been logged o from the Blackline Safety Network.
How do I turn o my G7x?
Press and hold the power button of your G7x. The device will go into shutdown sequence, sending your log off status to G7 Bridge.
When suitably powered, G7 Bridge can remain on at all times. When G7 Bridge is o, all connected devices will no longer be monitored. Before powering down, ensure there are no other G7x devices connected. Once conrmed, press and hold the power button on the G7 Bridge. The device will go into shutdown sequence. Your safety is no longer being monitored.
NOTE: For more information, please refer to the G7 Bridge technical user manual.
12 || G7 TECHNICAL USER MANUAL
CONNECTIVITY LIGHT
Are you connected to the Blackline Safety Network?
G7 lets you know your connection status.
Blinking/Solid
Blinking green light
A blinking connectivity light indicates that your G7 is storing data. It is not currently connected to the network and will send data when the light is solid.
Solid green light
A solid connectivity light indicates that G7 is actively transmitting data and is connected to the Blackline Safety Network.
Sound and vibration
Congurable to sound an alarm if connection is lost after 5 minutes.
NOTE: If you are using real-time features, your safety is only monitored when the
connectivity light is solid.
G7 TECHNICAL USER MANUAL || 13
VIEWING G7’S ASSIGNED USER
What is an assigned user?
G7 can be assigned a team member in Blackline Live. From Blackline Live, there is also the option to display the assigned user on the device, which allows you to identify which device belongs to which user. G7 does not have to be assigned a user to function as a safety monitor or gas detector.
Where can I see G7’s assigned user?
When the “display assigned user on device” option is enabled from Blackline Live, G7 will show the assigned team member in three places (as seen in the gures below):
 On startup  On the charging screen  In G7’s menu, under advanced info > user info
On the user info screen, you will also see a team member’s employee ID, if it has been added in Blackline Live.
Startup Charging screen Advanced info > user info
14 || G7 TECHNICAL USER MANUAL
GAS DETECTION
Included with your G7 Single-gas or Multi-gas cartridge is a calibration cap and tubing for bump tests and calibrations. Alternatively, bump tests and calibrations can be completed using a G7 Dock. Only perform bump test and calibrations in a safe environment. For more information and instructions on bumping or calibrating with G7 Dock, refer to the G7 Dock technical user manual.
BUMP TEST
What is a bump test?
It is safe practice to regularly test gas sensors by applying the target gas. G7’s bump test also tests the operation of light, sound and vibration indicators. The bump test schedule depends on your company’s safety policy. G7 automatically communicates bump test data to the Blackline Safety Network, and will remind you when a bump test is overdue. The bump test schedule is congurable.
NOTE: To meet CSA LEL performance standard, you are required to bump test before
each day’s use. Blackline recommends you do not exceed 30 days without a bump test.
How do I bump test?
1. Attach tubing to the calibration cap
2. Ensure the other end of the tubing is attached to a xed ow regulator on the gas tank
3. On your G7, press the OK button to enter the main menu
4. Use the arrow buttons to navigate to gas options, press OK
5. Use the arrow buttons to navigate to bump test, press OK
6. Press the up arrow to continue
7. G7 performs an automatic audio and visual assessment to test vibration and lights
8. You can individually choose which sensors to test. By default, G7 will test all sensors.
9. Attach the calibration cap to your device
10. G7 will begin to count down from 60. Apply the gas within this time window.
11. Turn gas o when prompted on your G7 screen
12. Press the OK button to complete bump test
13. G7 will let you know if the bump test has passed or failed, and when your next
bump test is due
14. Remove the calibration cap and let your G7 sit until the readings stabilize and your
G7 has returned to baseline
NOTE: If you see a bump test fail message on your LCD screen, try the calibration again. If
the error persists, please contact our Customer Care team.
G7 TECHNICAL USER MANUAL || 15
Tubing
Single-gas calibration cap
Tubing
16 || G7 TECHNICAL USER MANUAL
Multi-gas calibration cap
CALIBRATION
What is calibration?
Gas sensors periodically need to be calibrated by applying a known concentration of gas for a set amount of time. This procedure ensures the gas sensor can accurately detect gas levels throughout its operating life. The calibration schedule depends on your company’s safety policy. Blackline recommends not exceeding 180 days without a calibration.
How do I calibrate?
1. Attach tubing to the calibration cap
2. Ensure the other end of the tubing is attached to a xed ow regulator on the gas tank
3. On your G7, press the OK button to enter the main menu
4. Use the arrow buttons to navigate to gas options, press OK
5. Use the arrow buttons to navigate to calibration, press OK
6. Press the up arrow to continue
7. G7 performs an automatic audio and visual assessment to test vibration and lights
8. You can individually choose which sensors to test. By default, G7 will test all sensors.
9. Attach the calibration cap to your device
10. G7 will begin to count down from 60. Apply the gas within this time window and
continue to apply gas for 2 minutes.
11. Turn gas o when prompted on your G7 screen
12. Press the OK button to complete calibration
13. G7 will let you know if the calibration has passed or failed, and when your next
calibration is due
14. Remove the calibration cap and let your G7 sit until the readings stabilize and your
G7 is calibrated
NOTE: If you see a calibration fail message on your LCD screen, try the calibration again. If
the error persists, please contact our Customer Care team.
G7 TECHNICAL USER MANUAL || 17
BUMP TEST & CALIBRATION REMINDER WINDOW
What is the bump test and calibration reminder window?
A bump test and calibration reminder window can be customized from G7’s configuration profile in Blackline Live. This setting will prompt the device to show a bump test or calibration reminder alarm upon startup if one of these tests is due within the configured window of time.
What is the purpose of the bump test and calibration reminder window?
Enabling this feature will help ensure G7 goes into alarm upon startup instead of while it is in use in the field. If you have a daily bump interval your bump test will be due in 24 hours. However, if you start your shift 2 hours early the following day, you will not be prompted to bump your device as 24 hours has not elapsed. When it becomes due, you may already be out in the field without bump test materials. If you have a 2 hour bump test/calibration reminder window, G7 will go into yellow warning alarm for bump test due on startup when the bump is due within this 2 hour window.
BUMP TEST AND CALIBRATION LOCK
What is the bump test and calibration lock?
In the case that an organization wants to make the device unusable when it is out of compliance, the calibration or bump test lock can be enabled from G7’s configuration profile in Blackline Live. This feature locks the device’s screen and all functionality when a bump test or calibration is due on startup. This means the user cannot use their device in any capacity (including gas readings, SOS latch, message sending) until it has been bump tested or calibrated.
What if a bump test or calibration is due during my shift?
If a bump test or calibration comes due during your shift (not on startup), the device will not lock as this is a safety hazard, since the user will not be able to use G7 to call for help. The device will go into a yellow warning alarm for bump test or calibration due, and the message will display in the banner, but the device will remain fully functional.
18 || G7 TECHNICAL USER MANUAL
ZERO SENSORS
What is zeroing?
If G7 is not reading zero and you know you are in an atmosphere with no gas, your gas sensor readings may have shifted. If this happens, it is best to calibrate your sensors. If you are unable to preform a calibration, you can zero your sensors to reset the baseline.
G7 can be congured to automatically or manually zero on start-up when you turn on your device from the conguration prole in Blackline Live.
NOTE: The baseline reading for oxygen is 20.9.
How do I zero sensors?
1. On your G7, press the OK button to open the main menu
2. Press the arrow buttons to navigate to gas options, press OK
3. Use the arrow buttons to navigate to zero sensors, press OK
4. Press the up arrow to begin zeroing. Do not apply any gas.
5. The LCD screen will let you know once zero is complete
NOTE: If you see a zero incomplete message on your LCD screen you may be in an
environment with gas levels, or your cartridge may need replacement.
PID TARGET GASES
What are target gases?
Photoionization (PID) sensors can be used to detect a large range of gases. A target gas refers to the specific gas you are using trying to detect. G7’s readings will be adjusted based on the target gas it is configured to detect.
How do I set G7’s target gas?
G7’s PID sensor’s target gas is set from the configuration profile in Blackline Live. Under the photoionization detector section of the gas sensor settings card, you can choose an existing target gas or set a custom target gas.
Where do I see which target gas G7 is detecting?
The target gas G7 is configured to use can be seen in two places:
 On startup  In the gas options menu, under gas info > VOC target
In both of these places, G7’s screen will display the target gases name, as well as its correction factor.
G7 TECHNICAL USER MANUAL || 19
GAS DETECTION FEATURES
Once you have completed a bump test and calibration, G7 is ready to monitor and notify you of gas exposure. Gas detection alarms will notify you with either a yellow warning alarm or a red alert depending on the gas level measured. All settings are customizable in Blackline Live. Speak to your safety supervisor to learn about how your gas features are congured.
YELLOW WARNING ALARM GAS FEATURES
Rapid Blinking
LOW WARNING ALARM FOR GAS
When does G7 trigger a low yellow warning alarm for gas?
When gas levels reach the low concentration threshold configured by your safety supervisor, G7 will inform you with a yellow warning alarm every two minutes until gas levels are reduced.
NOTE: A G7 with O
oxygen-enriched atmospheres. An oxygen-decient atmosphere poses a risk of insucient oxygen for breathing. An oxygen-enriched atmosphere presents an increased risk of explosion.
20 || G7 TECHNICAL USER MANUAL
sensors will trigger low warning alarms in oxygen-decient and
2
UNDER LIMIT
When does G7 trigger an under limit yellow warning alarm?
If a gas sensor’s baseline shifts, it’s reading will shift accordingly and become unreliable. When this happens, the presence of gas can still be measured, but cannot be converted into an accurate reading. G7 will inform you of such an event with a yellow warning alarm. A calibration will be required to ensure the sensors are accurately detecting gas levels. If you are unable to perform a calibration and are in a clean atmosphere, you can zero your sensors to reset the baseline.
SENSOR ERROR
When does G7 trigger a sensor error yellow warning alarm?
If a gas sensor stops working for any reason, G7 will inform you with a yellow warning alarm. An X on your LCD screen will indicate which sensor or sensors are generating the error message. Power o and restart your G7. If the sensor error warning persists, perform a calibration. If this does not correct the problem, you will need to replace your cartridge.
CALIBRATION
When does G7 trigger a calibration yellow warning alarm?
When gas sensors are due for a calibration, G7 will inform you with a yellow warning alarm.
BUMP TEST
When does G7 trigger a bump test yellow warning alarm?
When gas sensors are due for a bump test, G7 will inform you with a yellow warning alarm.
PUMP BLOCK
When does G7 trigger a pump block yellow warning alarm?
If your pump inlet is blocked, G7 will inform you with a yellow warning alarm.
What do I do in the case of a yellow warning alarm?
Read your G7 screen. Press and hold the up and down arrow buttons at the same time to let your G7 know you have read the message.
Yellow warning alarms are between you and G7, and will not notify monitoring personnel if you are using real-time features.
G7 TECHNICAL USER MANUAL || 21
RED ALERT GAS FEATURES
Rapid Blinking
HIGH ALERT FOR GAS
When does G7 trigger a high alert for gas?
If a gas sensor detects gas levels above the high gas concentration threshold congured by your safety supervisor.
NOTE: A G7 with O
enriched situations.
sensors will trigger red alerts in both oxygen-decient and oxygen-
2
STEL SHORT TERM EXPOSURE LIMIT ALERT
When does G7 trigger a STEL alert?
If a gas sensor detects you have reached the short-term exposure limit configured by your safety supervisor. This limit is the gas concentration that you can be continuously exposed to for a 15-minute time frame without suffering adverse health effects.
22 || G7 TECHNICAL USER MANUAL
TWA TIME WEIGHTED AVERAGE ALERT
When does G7 trigger a TWA alert?
If a toxic sensor detects you have exceeded the average allowable amount of gas during an eight-hour period.
NOTE: The two TWA measuring methods available for use on your G7 are:
OSHA (United States Department of Labor Occupational Safety and Health Administration) or ACGIH (American Conference of Governmental Industrial Hygienists).
OSHA is defined as a rolling average of gas exposure accumulated over an eight-hour period of operation. If the worker is in the field longer, the most recent eight-hour cumulative value is used.
ACGIH is dened as the total accumulated average, from four to 16 hours as congured by your safety supervisor.
OL OVER LIMIT ALERT
When does G7 trigger an OL alert?
If a gas sensor detects an excessive amount of gas and can no longer give you an accurate reading.
NOTE: STEL, LEL and peak readings are reset on power cycle unless configured to
resume readings in G7’s configuration profile in Blackline Live.
What do I do in the case of a red alert?
Evacuate the area and follow your emergency safety protocol. Read the information on your G7 screen. Press and hold the up and down arrow buttons at the same time to mute the sound and vibration. The sound and vibration will return after one minute. This will continue until gas levels or averages have reduced or returned to acceptable concentrations.
For real-time users, red alerts are immediately communicated to monitoring personnel. Muting sound and vibration does not cancel the remote alert sent to monitoring personnel.
G7 TECHNICAL USER MANUAL || 23
CONFIGURATION MODES
Conguration modes are customized in the G7 conguration prole on Blackline Live. Each prole supports up to a total of ve ‘modes’ . These modes allow G7 to temporarily change its behavior for dierent situations, and can be turned on and o through G7’s interface.
AVAILABLE MODES
Normal
This mode is the configuration you determine for everyday operation. G7 will operate using this mode as a default.
Pre-entry
Pre-entry mode is used before entering a space that could potentially contain dangerous gas. This mode can be used with or without a pump cartridge, which will actively draw surrounding air to its sensors and test gas levels.
SCBA
This mode is meant to be used when the device user is wearing a self-contained or supplied air breathing apparatus (SCBA/SABA) and is entering an area that is known to have high gas levels.
Leak check
Leak check mode can be used when checking for gas leaks in a particular area. This mode – like pre-entry mode – can be used with or without a pump cartridge.
High risk
This mode is unique in that it is meant for general high-risk situations, such as an evacuation or travelling through a dangerous area. Unlike the other modes, you will never be timed out and must exit it manually.
Pump run
This mode is unique in that a multi-gas pump cartridge is required to use this mode, and it runs the pump continuously – such as for use in a hole-watch situation. Unlike the other modes, you will never be timed out and must exit it manually.
ENTERING A MODE
How do I enter a mode?
To use a mode, it must first be enabled in G7’s configuration profile on Blackline Live. Entering a mode can be done from G7’s main menu or main status screen.
24 || G7 TECHNICAL USER MANUAL
To enter a mode from the modes menu:
1. Press OK to enter G7’s main menu
2. Use the up and down arrows to navigate to modes
3. Press OK to enter the modes menu
4. Select the mode you wish to enter
5. Conrm that you want to enter the mode by selecting yes
6. G7’s screen will invert and your information banner will display your current mode
To enter a mode from the main status screen:
1. Press the up or down arrow to open G7’s secondary menu
2. Continue to press the up or down arrow until you reach your desired mode
3. Press OK to enter the mode
4. G7’s screen will invert and your information banner will display your current mode
EXITING A MODE
How do I exit a mode?
When you’d like to return the device to normal operation, you will need to exit the configuration mode you are currently in. Exiting a mode can be done from G7’s main menu or main status screen.
To exit from the modes menu:
1. Press OK to enter G7’s main menu
2. Use the up and down arrows to navigate to modes
3. Press OK to enter the modes menu
4. Select normal mode
5. Conrm that you want to return to normal mode by selecting yes
6. G7 will return to normal function
To exit from the main status screen:
1. Press the up or down arrow to open G7’s secondary menu
2. Press OK to exit the mode
3. G7 will return to normal function
What is a mode timeout?
Each configuration mode (excluding normal and high risk mode) has a timeout period. After this time period has elapsed, you will be asked if you would like to continue in this mode. If you select yes, your mode will be extended. If you select no, G7 will return to normal operation. If you do not make a selection within 30 seconds, G7 will automatically return to normal operation. If you have a check-in timer enabled, G7 will immediately ask you to check in.
G7 TECHNICAL USER MANUAL || 25
CARTRIDGES
GAS CARTRIDGE REPLACEMENT PROGRAM
What do I do when I need a new gas cartridge?
If you have an uninterrupted service plan for your G7 gas cartridge, Blackline will replace expired cartridges for you free of charge. To inquire about or request new cartridges, please contact our Customer Care team or your distributor.
CHANGING CARTRIDGES
How do I change my G7 cartridge?
1. Power o G7
2. Using a Phillips #1 screwdriver, remove the screws on each side of the device
3. Pull up on the cartridge
4. Slide a new cartridge onto G7, ensuring the cartridge clicks into place
5. Replace screws into each side of the device
NOTE: Cartridges should always be replaced with a manual (not electric) screwdriver in
order to avoid damage to the device’s plastics.
26 || G7 TECHNICAL USER MANUAL
CARTRIDGE CARE
Sensor contaminants
Gas sensors are susceptible to contamination by a variety of common chemicals, reducing or eliminating their sensitivity. Care should be taken when using silicones, cleaners, solvents and lubricants in close proximity to sensors as exposure may cause permanent damage to the sensor. If a device is exposed to a new chemical or compound, it is best practice to bump test and calibrate units to ensure proper sensor function is maintained.
LEL SENSOR PRECAUTIONS
For safety reasons this equipment must be operated and serviced by qualied personnel only. Read and understand instruction manual completely before operating or servicing.
 With pellistor technology (LEL-P), any rapid up-scale reading followed by declining
or erratic reading may indicate a gas concentration beyond upper scale limit which may be hazardous.
 High o-scale readings may indicate an explosive concentration.  Calibrations must only be performed in areas free of ammable gasses.
Blackline’s NDIR sensor does not detect Hydrogen or Acetylene. Like any gas sensor, please be sure you understand your potential explosive hazards and choose the appropriate sensor technology based on these hazards.
Blackline’s LEL-MPS sensor does not require calibration.
G7 TECHNICAL USER MANUAL || 27
MULTIGAS PUMP CARTRIDGE
USING PUMP
How do I turn pump on?
1. Ensure G7 is equipped with a multi-pump cartridge
and pumped modes, such as pre-entry, leak check or pump run
2. From G7’s main or secondary menu, select the
pumped mode you wish to enter
3. Attach tubing, and perform a block test by
following the instructions on G7’s screen
4. G7’s screen will invert, and the pump icon will show
in the info bar to let you know pump is running
How do I turn pump o?
To turn pump off, enter any non-pumped mode, such as normal, SCBA or high risk mode.
Where do I see my pump details?
When a pump cartridge is attached to G7, you can see the pump status screen by pressing the up or down arrow from G7’s main screen. Here you can see if pump is on or o, your hose length, ow rate and sample time (if enabled).
How do I bump test or calibrate my pumped cartridge?
Manually bump testing and calibrating pump cartridges is done using the same calibration cap and method as G7’s multi-gas cartridge. Please refer to page 14-16 of this manual for step-by-step instructions. G7 cannot be bump tested or calibrated through the pump itself.
Bump testing and calibrating with a G7 Dock requires an updated G7 Dock unit. You can tell if your Dock is an updated unit by checking that it’s unit ID is Dock-P.
Pump on
Hose L: 5m
~300 ml/min.
Sample 75 sec.
1:48
How do I perform a block test?
You will be asked to perform a block test upon entering any pumped mode, such as leak check, pre-entry or pump run. Follow the steps on G7’s screen to complete the block test. If you do not pass the block test, you will not be able to enter the mode.
IMPORTANT: When performing an automatic block test, G7 is in a safe mode where gas
alerts will not be triggered. This prevents triggering false alerts from residual gas in the tubing.
When you are in a pumped mode, you can perform a manual block test at any time by plugging G7’s inlet. This will cause G7 to go into yellow warning alarm, and your screen will let you know that your pump is blocked. Unplug your inlet, and if the warning alarm silences, you know that your equipment is safe to use.
28 || G7 TECHNICAL USER MANUAL
What is ow rate?
Flow rate is the speed at which air is passing over your device’s sensors. In order to provide accurate gas readings, ow rate needs to be above 150 ml/min. A ow rate less than 150 ml/min will result in a pump block yellow warning alarm. By default, Blackline’s pump aims to maintain a ow rate of 300ml/min and G7 will automatically adjust its pump speed to maintain this rate.
PUMP OPTIONS
Settings for pump can be found in G7’s main menu, under settings > pump options.
What is the sample timer?
The sample timer is the amount of time it takes for one air sample to be pumped to your sensors. This calculation is based on your hose length. If sample timer is enabled, the pump status screen will show a countdown and beep once when a sample cycle has completed. The sample cycle will repeat continuously until you turn pump o. If disabled, the sample timer will not show and the device will not beep, and pump will continue to function normally. Blackline recommends a 120 second sample time for 10 feet of tubing, with an additional second per foot of tubing.
How do I enable or disable the sample timer?
1. Press OK to enter G7’s main menu
2. Use the up and down arrows to navigate to settings, press OK
3. Use the up and down arrows to navigate to pump options, press OK
4. The menu will display the current status of your sample timer. Select
sample timer o or sample timer on.
5. Select yes to conrm that you would like to enable or disable the sample timer
What is hose length?
Hose length is an estimate of how long the tubing attached to pump is. This value is customizable from the pump options menu, and factors in to sample time. G7’s multi­gas pump cartridge supports a maximum of 99 feet of tubing.
How do I change hose length?
1. Press OK to enter G7’s main menu
2. Use the up and down arrows to navigate to settings and press OK
3. Use the up and down arrows to navigate to pump options and press OK
4. Select hose length
5. Use the up and down arrows to choose your units, meters or feet, and press OK
6. Use the up and down arrow to enter the first digit of your length and press OK
7. Use the up and down arrow to enter the second digit of your length and press OK
8. Select yes to confirm your length or select edit to make changes to your choice
G7 TECHNICAL USER MANUAL || 29
REALTIME FEATURES
Your G7 can be equipped with many features to monitor your safety. They are categorized into the type of notications they trigger — yellow pending alarms, yellow warning alarms, and red alerts. All are customizable in Blackline Live to best t your needs. Speak to your safety supervisor to learn about how your G7 features are congured.
YELLOW PENDING ALARM FEATURES
Blinking
POTENTIAL FALL DETECTED
What is a potential fall detected?
Your device is constantly monitoring you for falls. If a potential fall is detected, G7 will initiate a yellow pending alarm. The fall detection sensitivity is congurable.
30 || G7 TECHNICAL USER MANUAL
POTENTIAL NOMOTION DETECTED
What is a potential no-motion detected?
Your device is constantly monitoring your motion. It will automatically detect if you do not move within a pre-set duration, and will initiate a yellow pending alarm. The no-motion period and sensitivity are congurable.
CHECKIN REQUEST
What is a check-in?
If enabled, you can congure your device to request periodic check-ins throughout your shift. The check-in countdown is displayed at the top right of the LCD screen. At the end of the timer, your G7 will initiate a yellow pending alarm to conrm you are safe. The check-in timer and pending alarm time are congurable.
NOTE: Your device can be congured to check-in early, before the yellow
pending alarm sounds. If enabled, you can push and hold the red latch button for the duration of three vibrations to reset your check-in timer before the audible alarm. An early check-in cannot be congured if Silent SOS alert is enabled.
What do I do in the case of a yellow pending alarm?
If you are safe, push the red latch button. If you do not push the latch within the congured amount of time, your yellow pending alarm will communicate a red alert to monitoring personnel.
NOTE: G7 can be congured to mute
pending alarms while driving at speeds above 35 km/hr (22 mph) or while charging.
G7 TECHNICAL USER MANUAL || 31
YELLOW WARNING ALARM FEATURES
Rapid Blinking
MESSAGES
How do I receive a message?
Your device can receive messages from monitoring personnel. When there is an incoming message, G7 will inform you with a yellow warning alarm.
How do I send a message?
You can choose from a list of 10 pre-programmed messages to send to monitoring personnel. The messages are customizable from Blackline Live. Press the OK button to enter the main menu, the up or down arrow buttons to navigate the menu, highlight your selection, and the OK button to send.
How do I send a custom message?
At the bottom of the pre-programmed message list is an option to send a 16-character custom message to monitoring personnel. Press the up or down arrow buttons to scroll through the alphabet and numbers, press the OK button to move to the next character, press the OK button again to send.
NOTE: In the conrmation screen of your custom message, you have the
ability to edit the current message by pressing the up arrow button, send the message by pressing the OK button or cancel the message by pressing the down arrow button.
32 || G7 TECHNICAL USER MANUAL
SPEAKER PHONE
How do I use two-way voice calling?
If you have a G7c with a voice-enabled service plan, your speakerphone will automatically answer a call from monitoring personnel. G7 will inform you of an incoming call with a yellow warning alarm and you will hear a beep signifying the two-way voice call has been connected. In a noisy environment, it may be necessary to remove and hold the device near your ear, as you would a two-way radio.
NETWORK CONNECTION INTERRUPTION
How do I know if the connection to my device is lost?
If your device loses connection with the Blackline Safety Network, it will inform you with a yellow warning alarm after 5 minutes. This amount of time is congurable.
LOW BATTERY
How do I know if my device battery is low?
If your battery level goes below 20%, it will inform you with a yellow warning alarm. This percentage level is congurable.
What do I do in the case of a yellow warning alarm?
Read your G7 screen. Press and hold the up and down arrow buttons at the same time to let G7 know you have read the message.
Yellow warning alarms are between you and G7, and will not notify monitoring personnel.
G7 TECHNICAL USER MANUAL || 33
RED ALERT FEATURES
Rapid Blinking
FALL DETECTED
What is fall detected?
If your device detects a fall and you have not responded to the yellow pending alarm, G7 will communicate a red alert to monitoring personnel. The fall detection sensitivity is congurable.
NOMOTION DETECTED
What is no-motion detected?
If you are not moving and you have not responded to the yellow pending alarm, G7 will communicate a red alert to monitoring personnel. The no-motion period and sensitivity are congurable.
34 || G7 TECHNICAL USER MANUAL
MISSED CHECKIN
What is a missed check-in?
If enabled and you are unable to check-in during the yellow pending alarm time, your device will communicate a red alert to monitoring personnel.
SOS ALERT
What is an SOS Alert?
If you require assistance, you can manually send an SOS alert to monitoring personnel and request immediate help to your location by pulling the red latch.
NOTE: Your device can be congured to send a Silent SOS alert to monitoring
personnel without light, sound and vibration.
What do I do in the case of a red alert?
Red alerts are immediately communicated to monitoring personnel. Read the information on your G7 screen. Press and hold the up and down arrow buttons at the same time to mute the sound and vibration. This does not cancel the red alert sent to monitoring personnel.
G7 TECHNICAL USER MANUAL || 35
LiveResponse
When monitoring personnel have acknowledged your red alert, the blue LiveResponse light will shine on your device.
Blinking
This light lets you know that remote monitoring personnel are responding by following your team’s emergency protocol. Once monitoring personnel have resolved the red alert, the blue LiveResponse™ light will shut o.
Depending on your response protocol, a G7c with voice enabled service plan will automatically connect your speaker phone to monitoring personnel.
In the case that a silent SOS alert is sent, the blue LiveResponse light will not shine. Instead, G7 can be congured to vibrate to let you know that remote monitoring personnel are responding by following your team’s emergency protocol.
36 || G7 TECHNICAL USER MANUAL
PUSHTOTALK PTT
If you have a G7c with a PTT service plan and PTT is enabled in your device’s conguration prole, push-to-talk allows you to send and receive voice messages to other G7c device users, similar to a walkie-talkie. Push-to-talk is only available on G7c devices.
TRANSMITTING
How do I send a PTT message?
1. Press and hold the red latch
2. When G7c finishes beeping, continue to hold and begin talking with
the device about 6 inches from your mouth.
NOTE: If you are using an O sensor, be sure that you are talking into G7’s
microphone, not the cartridge, as this could cause gas alarms.
3. When you’re finished talking, release the latch. G7 allows PTT messages
up to 30 seconds in length.
4. G7c will beep once more to let you know it’s done listening
RECEIVING
How do I receive a PTT message?
1. G7c will beep twice to signal an incoming PTT message
2. G7c will play the message
3. G7c will beep once more when the message is done
NOTE: G7c’s screen will display which channel you are transmitting to or receiving from.
AVAILABLE CHANNELS
Channel 0-99
Channels 0 through 99 are available for everyday use. When on a specific channel, you will only be able to communicate with devices that are on the same channel, and will receive transmissions from all call.
All call
All call is a channel where G7c will transmit to all PTT devices in your organization, and will only hear transmissions from all call. This channel is recommended for safety supervisors or managers.
Receive only
The receive only channel only hears transmissions from all call, and can’t transmit to other devices.
G7 TECHNICAL USER MANUAL || 37
CHANGING CHANNELS
How do I change my channel?
Changing your PTT channel is done from the PTT channel menu.
To change to a specific channel number:
1. Press the OK button to enter G7c’s main menu
2. Use the up and down arrows to navigate to PTT channels, and select OK.
3. Select enter channel #
4. Use the up and down arrows to enter the first digit of your channel. For example,
for channel 40, the first digit is 4.
5. Press OK
6. Use the up and down arrow to enter the second digit of your channel. For
example, for channel 40, the second digit is 0.
7. Press OK
8. Select yes to confirm and change G7c’s channel or select edit to make changes
To change to receive only or all call:
1. Press the OK button to enter G7c’s main menu
2. Use the up and down arrows to navigate to PTT channels, and select OK.
3. Select receive only or all call
4. Read the message on the screen
5. Select yes to confirm
NOTE: You can also navigate to the PTT channels menu by pressing the up or down
arrow from the main screen, and pressing OK when the current PTT channel shows.
CHANGING VOLUME
How do I change my push-to-talk call volume?
Changing the volume of incoming calls can be done from G7’s main screen or the PTT channel menu. Changing your PTT volume only affects incoming calls, and won’t change the way G7’s alerts sound.
To change volume from the main screen:
1. Press and hold the up or down arrow
2. Use the up and down arrows to set G7’s volume to the desired level
3. Press OK or wait for the screen to timeout to save
To change volume from the PTT channel menu:
1. Press the OK button to enter G7c’s main menu
2. Use the up and down arrows to navigate to PTT channels, and select OK
3. Use the up and down arrows to navigate to change volume, and select OK
4. Use the up and down arrows to set G7’s volume to the desired level
5. Press OK or wait for the screen to timeout to save
38 || G7 TECHNICAL USER MANUAL
PTT AUDIO ACCESSORIES
G7c is equipped with audio pairing in order to use an audio accessory with push-to-talk. All settings for audio devices can be found in G7’s main menu, under settings > audio pairing.
How do I pair a new audio device?
1. Press OK to enter the main menu
2. Use the up and down arrows to navigate to accessories
3. Navigate to audio pairing, and press OK
4. Use the up and down arrows to navigate to pair new, and press OK
5. Put your audio accessory into pairing mode as per its instructions
6. Select your audio accessory from the list
7. G7 will give you a successful connection message when your device is connected,
and you will see an audio accessory icon in the info bar of G7c’s main screen
How do I reconnect to an audio device?
G7 will remember your audio accessory and pair automatically when both are powered on. In the case that it doesn’t, you can reconnect to your accessory from the audio pairing menu.
1. Press OK to enter the main menu
2. Use the up and down arrows to navigate to settings
3. Navigate to audio pairing, and press OK
4. Use the up and down arrows to navigate to reconnect, and press OK
5. Ensure your accessory is on
6. G7c will reconnect to your audio accessory, and you will see an audio accessory
icon in the info bar of G7c’s main screen when the connection is successful
How do I forget a paired device?
1. Press OK to enter the main menu
2. Use the up and down arrows to navigate to settings
3. Navigate to audio pairing, and press OK
4. Use the up and down arrows to navigate to forget device, and press OK
5. Select yes to confirm that you would like to forget your accessory
NOTE: Audio accessories can only be used for push-to-talk purposes, and cannot be used
for early check-ins or conrming pending alarms.
G7 TECHNICAL USER MANUAL || 39
FIRMWARE UPDATES
OVERTHEAIR OTA FIRMWARE UPDATES
How does G7 get updated?
To oer new features, Blackline Safety periodically releases over-the-air (OTA) rmware updates. Your G7 device will automatically download and install the new rmware.
Blackline Safety will contact you directly for specic information about new updates. If you have any questions, please contact our Customer Care team.
NOTE: OTA rmware updates are only available for G7 Bridge when it is brought into
cellular range. If G7 Bridge cannot be removed from satellite only reception, contact Customer Care to receive a rmware update kit.
40 || G7 TECHNICAL USER MANUAL
SUPPORT
LEARN MORE
Visit support.BlacklineSafety.com to nd support and training materials for G7.
CUSTOMER CARE
For technical support, please contact our Customer Care team.
North America (24 hours)
Toll Free: 1-877-869-7212 | support@blacklinesafety.com
United Kingdom (8am-5pm GMT)
+44 1787 222684 | eusupport@blacklinesafety.com
International (24 hours)
+1-403-451-0327 | support@blacklinesafety.com
G7 TECHNICAL USER MANUAL || 41
SPECIFICATIONS
DETAILED SPECIFICATIONS
Standard G7 safety features
Fall detection and No-motion detection: Tri-axis accelerometer, tri-axis gyro, software processing, congurable sensitivity, congurable time for no-motion
SOS latch: Pull latch to trigger SOS alert Silent SOS: Press and hold latch to trigger SOS alert Low-battery: Congurable threshold Worker check-in: Congurable timer (30 – 180 min, or
o), automatic check-in when driving
Additional gas cartridge features
Under limit Over limit Time-weighted average (TWA) Short-term exposure limit (STEL) High gas alert Low gas alert Bump test and calibration notication Bump test and calibration failure
Size & weight
G7 with Standard Cartridge Size: 64 mm x 124 mm x 27 mm (2.52” x 4.88” x 1.06”) Weight: 162 g (5.7 oz)
G7 with Single-gas Cartridge Size: 64 mm x 128 mm x 27 mm (2.52” x 5.04” x 1.06”) Weight: 167 g (5.9 oz)
G7 with Quad-gas Cartridge Size: 66 mm x 150 mm x 27 mm (2.52” x 5.91” x 1.06”) Weight: 192 g (6.8 oz)
G7 with Multi-gas pump Cartridge Size: 66 mm x 151 mm x 38.5 mm (2.6” x 5.95” x 1.52”) Weight: 238 g (8.4 oz)” Target ow rate: 300ml/min Maximum hose length: 99ft (30.2 m)
User interface
168 by 144 pixel graphical, high contrast, liquid crystal display with front lighting, menu system driven by three­button keypad, power button (on/o), check-in button
(check-in/silent SOS), SOS latch (send SOS alert)
Multi-language support: Yes, EN, FR, ES, NL, DE, IT, PT”
User notication
Green SureSafe® light: Blinking (powered), continuous (connected) Yellow top and front lights: Yellow pending alarm and
yellow warning alarm Red top and front lights: Red alert communicated Blue LiveResponse™ top and front lights: Conrmation that a monitoring team has acknowledged the alert
Alarm Indicators: Speaker, LED lights and vibration motor Speaker sound pressure level: ~95 dB @ 30 cm (~95 dB
@ 11.8”) Voice calling: Speakerphone and phone modes (G7c model only)
G7c wireless radio
Wireless coverage: 100 countries, 200 wireless carriers North America: 2G/3G radio, GSM 850 MHz, PCS 1900
MHz, 3G UMTS bands 2, 5 and 6 International: 2G/3G radio, E-GSM 900 MHz, DSC 1800 MHz, 3G UMTS bands 1 and 8
Antenna: Internal regionally optimised
G7x wireless radio
Works with G7 bridge satellite base station
Radio: 902.0–928.0 MHz, 1 Watt Antenna: Internal Radio link range: 2 km (1.25 mi) real-world
Wireless updates
Device conguration changes: Yes, Device rmware upgrade over-the-air (FOTA): Yes
Location technology
GPS Radio: 48-channel high sensitivity Assisted-GPS: Yes (G7c model only) GPS Accuracy: ~5 m (16 ft) Outdoors Indoor location technology: Blackline Safety location beacons Location update frequency: G7c 5 min, G7x 15 min defaults
Power & battery
Rechargeable Li-ion battery: 1100 mAh Li-ion Battery Life: 18 hours at 20°C (68°F) under normal usage Charge time: 4 hours
Environmental
Storage temperature: -30°C to 60°C (-22°F to 140°F) Operating temperature: -20°C to 55°C (-4°F to 131°F) Charging temperature: 0°C to 45°C (32°F to 113°F) Ingress Protection: Designed to meet IP67
Approvals
G7c: SAR, RoHS, CE, RCM Unit ID 3570xxxxxx
FCC ID: W77G7C2 IC: 8255A-G7C2 Contains FCC ID: XPY1EIQ24NN IC: 8595A-1EIQ24NN or
42 || G7 TECHNICAL USER MANUAL
Unit ID 3567xxxxxx
FCC ID: W77G7C IC: 8255A-G7C Contains FCC ID: XPY1CGM5NNN, IC: 8595A-1CGM5NNN or
Unit ID 3566xxxxxx
Contains FCC ID: XPY1CGM5NNN, IC: 8595A-1CGM5NNN Canada & USA: Class I Division 1 Group A,B,C,D T4; Class I Zone 0 AEx da ia IIC T4; Ex da ia IIC T4 Ga
IECEx: Ex da ia IIC T4 Ga ATEX:Ex da ia IIC T4 Ga LEL: CSA C22.2 No.152; ISA 12.13.01 LEL Pump Cartridge: CSA C22.2 No.152; 0°C ≤ Ta ≤ 40°C;
ANSI/ISA-12.13.01; -10°C ≤ Ta ≤ 40°C
G7x: SAR, RoHS, RCM FCC ID: W77G7X | IC ID: 8255A-G7X Canada & USA: Class I Division 1 Group A,B,C,D T4;
LEL: CSA C22.2 No.152; ISA 12.13.01 LEL Pump Cartridge: CSA C22.2 No.152; 0°C ≤ Ta ≤ 40°C;
ANSI/ISA-12.13.01; -10°C ≤ Ta ≤ 40°C
Warranty
G7: two years limited warranty Cartridges: lifetime with service plan Blackline Complete: three year operating lease with
three year warranty
Blackline Live web application
Cloud-hosted safety monitoring web application is highly customizable for every customer requirement. Includes live map, employee address book, user roles, alert management, device congurations, alert setups and reporting.
Class I Zone 0 AEx da ia IIC T4; Ex da ia IIC T4 Ga IECEx: Ex ib IIC T4 Gb
Gas sensor specications
Gas Sensor type Range Resolution
Ammonia (NH) Electrochemical 0–100 ppm 0.1 ppm
High-range ammonia (NH) Electrochemical 0–500 ppm 1 ppm
Carbon monoxide (CO) Electrochemical 0–500 ppm 1 ppm
High-range carbon monoxide (CO) Electrochemical 0–2000 ppm 5 ppm
Hydrogen resistant carbon monoxide (CO-H) Electrochemical 0–500 ppm 1 ppm
Carbon dioxide (CO) NDIR 0–50,000 ppm 50 ppm
Chlorine (Cl) Electrochemical 0–20 ppm 0.1 ppm
Chlorine dioxide (ClO) Electrochemical 0-2 ppm 0.01 ppm
COSH Electrochemical 0–500 ppm CO, 0–100
Hydrogen (H) Electrochemical 0–40,000 ppm 1% LEL (400 ppm)
Hydrogen cyanide (HCN) Electrochemical 0–30 ppm 0.1 ppm
Hydrogen sulphide (HS) Electrochemical 0–100 ppm 0.1 ppm
High-range hydrogen sulphide (HS) Electrochemical 0-500 ppm 0.5 ppm
LEL-infrared (LEL-IR) NDIR 0–100% LEL 1% LEL
Oxygen (O) Pumped electrochemical 0–25% vol 0.1% vol
Ozone (O) Electrochemical 0-1 ppm 0.01 ppm
Photoionisation (PID) PID 0–6,000 ppm 0.1 ppm (0–100 ppm), 2
Sulphur dioxide (SO) Electrochemical 0–100 ppm 0.1 ppm
ppm HS
1 ppm CO, 0.1 ppm HS
ppm (100–6,000 ppm)
* Check with Blackline for approval status. All specications subject to change.
G7 TECHNICAL USER MANUAL || 43
LEGAL NOTICES AND CERTIFICATIONS
LEGAL NOTICES
Information in this document is subject to change without notice. This document is provided “as is” and Blackline Safety Corp. (“Blackline”) and its aliated companies and partners assume no responsibility for any typographical, technical or other inaccuracies in this document. Blackline reserves the right to periodically change information that is contained in this document. However, Blackline makes no commitment to provide any such changes, updates, enhancements or other additions to this document to you in a timely manner or at all.
Copyright © 2016 Blackline Safety Corp. All rights reserved.
Except as expressly provided herein, no part of this manual may be reproduced, copied, transmitted, disseminated, downloaded, or stored in any storage medium, for any purpose without the express prior written consent of Blackline Safety Corp (“Blackline” ). Blackline hereby grants permission to download a single copy of this manual onto some form of electronic storage medium to be viewed and to print one copy of this manual or any revision hereto, provided that such electronic or printed copy of this manual must contain the complete text of this copyright notice. Further, any unauthorized commercial distribution of this manual or any revision hereto is strictly prohibited.
The Blackline, Alert. Locate. Respond. families of related marks, images and symbols, including Blackline, G7, G7c, G7x, LiveResponse, Loner, Loner IS, Loner IS+, Loner M6, Loner M6i, Loner Mobile, Loner 900, and SureSafe are the exclusive properties and trademarks of Blackline Safety Corp. All other brands, product names, company names, trademarks and service marks are the properties of their respective owners.
Warranty
Your G7 device is warranted against defects in materials and workmanship for up to two years from date of purchase. For further details regarding your Blackline warranty, please refer to your terms and conditions of service.
FCC Compliance
This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC rules. These limits are designed to provide reasonable protection against harmful interference in a residential instal lation.
Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation. Note: the grantee is not responsible for any changes or modications not expressly approved by the party responsible for compliance. Such modications could void the user’s authority to operate the equipment.
This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment o and on, the user is encouraged to try to correct the interference by one or more of the following measures:
Reorient or relocate the receiving antenna.
Increase the separation between the equipment and receiver.
Connect the equipment into an outlet on a circuit dierent from that to which the receiver is connected.
Consult the dealer or an experienced radio/TV technician for further assistance.
RF exposure was tested with the supplied belt clip. Use of third-party accessories may result in non-compliant exposure.
Industry Canada Compliance
This device complies with Industry Canada licence-exempt RSS standard(s). Operation is subject to the following two conditions: (1) this device may not cause interference, and (2) this device must accept any interference, including interference that may cause undesired operation of the device.
RF exposure was tested with the supplied belt clip. Use of third-party accessories may result in non-compliant exposure.
Notication d’Industrie Canada
Ce dispositif est conforme au(x) format(s) RSS libre(s) d’Industrie Canada. Son fonctionnement est assujetti aux deux conditions suivantes : (1) Cet appareil ne peut causer d’interférences nuisibles, et (2) cet appareil doit accepter toute interférence reçue, y compris les interférences pouvant provoquer un mauvais fonctionnement du dispositif.
L’exposition RF a été testée avec le clip de ceinture fourni. L’utilisation d’accessoires tiers peut entraîner une exposition non conforme.
Warning
Do not operate Blackline Safety products where you are not able to safely operate your mobile/cellular phone.
Electrical equipment may be hazardous if misused. Operation of this product, or similar products, must always be supervised by an adult. Do not allow children access to the interior of any electrical product and do not permit them to handle any cables.
Do not operate or store Blackline products outside their specied operating or storage temperatures. Consult the specications section for more information.
Blackline products may contain an internal lithium-ion battery pack. Seek advice from your local electronics recycling authority regarding the disposal of your device. Do not dispose Blackline products in your household trash.
44 || G7 TECHNICAL USER MANUAL
INTRINSICALLY SAFE
Intrinsically Safe
This device is certied Intrinsically Safe for use in Class I Division 1 Groups A,B,C,D T4; Ex da ia IIC T4 Ga; Class I Zone 0 AEx da ia Group IIC T4 Ga hazardous (classied) locations. G7x is certied as Ex ib IIC T4 Gb under IECEx.
Sécurité intrinsèque
Cet appareil est certié à sécurité intrinsèque pour l’usage en classe I division 1 groupe A,B,C,D T4; Ex da ia IIC T4 Ga; classe I zone 0 AEx da ia groupe IIC T4 Ga dans les lieux classés comme dangereux.
CSA: 70098755 UL 60079
Class I Division 1 Groups A,B,C,D; T4 Class I Zone 0 AEx da ia IIC T4 Ga
CAN/CSA C22.2 No. 60079
Ex da ia IIC T4 Ga
-20°C <= Ta <= +55°C
Base unit P/N “G7*-#” (* = c, x, or blank; # = NA, EU or AZ) Gas cartridge: Standard P/N “Z” | Single-gas P/N “S-#” | Multi-gas P/N “Q-####” | Pump Module P/N “P-####” (# = Electro chemical sensor identier or “X” indicating no sensor)
Caution: For safety reasons this equipment must be operated and serviced by qualied personnel only. High o-scale readings may indicate explosive concentration.
The equipment shall only be charged when in the non-hazardous area using a charger specically supplied for use with the unit (for example part number SAW06D-050-1000xx, manufactured by Shenzhen Shi Ying Yuan Electronics Co., Ltd.), approved as SELV or Class 2 equipment against IEC 60950, IEC 61010­1 or an equivalent IEC standard. The maximum voltage and current from the charger shall not exceed 5.625Vdc and 2A respectively.
Standards:
CAN/CSA C22.2 No. 60079-0: 2015 CAN/CSA C22.2 No. 60079-11: 2014 CAN/CSA C22.2 No. 60079-1: 2016 C22.2 No. 152 - M1984 (R2011) UL 913, Eighth Edition UL 60079-0: Sixth Edition UL 60079-11: Sixth Edition UL 60079-1: Seventh Edition
Consult with your organization’s safety professional for further information regarding the topic of intrinsic safety and any policies, procedures, facilities, or locations within facilities that may be related to intrinsic safety.
Blackline Safety | Unit 100, 803 - 24 Avenue SE | Calgary, AB T2G 1P5 | Canada
IECEx/ATEX: IECEx CSA 17.0005; Sira 17ATEX2083X IEC 60079; EN 60079
G7c: Ex da ia IIC T4 Ga G7x: Ex ib IIC T4 Gb
2813 II 1 G
Attention: Pour des raisons de sécurité, cet équipment doit être utilisé, entretenu et réparé uniquement par un personnel qualié. Des lectures supérieures à l’échellepeuvent indiquer des concentration explosives.
L’équipement ne doit être chargé que dans la zone non dangereuse à l’aide d’un chargeur spéciquement fourni pour l’utilisation avec l’appareil (par exemple, la référence SAW06D-050­1000xx, fabriquée par Shenzhen Shi Ying Yuan Electronics Co., Ltd.) SELV ou Classe 2 selon IEC 60950, IEC 61010-1 ou une norme IEC équivalente. La tension et le courant maximum du chargeur ne doivent pas dépasser respectivement
5.625Vdc et 2A.
ANSI/ISA 12.13.01: 2000 EN 60079-0: 2012/A11:2013 EN 60079-11: 2012 EN 60079-1: 2014 IEC 60079-0: 2011 6th Edition IEC 60079-11: 2011 6th Edition IEC 60079-1: 2014 7th Edition
S’il vous plaît consulter professionnel de la sécurité de votre organisation pour de plus amples informations concernant le sujet de la sécurité intrinsèque et les politiques, les procédures, les installations, ou emplacements au sein des établissements qui peuvent être liés à la sécurité intrinsèque.
G7 TECHNICAL USER MANUAL || 45
46 || G7 TECHNICAL USER MANUAL
www.BlacklineSafety.com
0149/R22/2020-02-10
Loading...