tyco EH9 User Manual

CORCOM Product Guide
LINE
LOAD
E
B
A
D
C
RFI Power Line Filters for UL2601 Healthcare Equipment
H Series
UL Recognized CSA Certified VDE Approved*
H Series
The H series filters were developed specifically for health care applications. These products are suitable for uses in both patient care equipment and non-patient care equipment. Their extremely low leakage current allows compliance with UL2601 and CSA 22.2 health care specifications.
See Appendix C for more information on UL544 health care applications.
The H series features products with a variety of current ratings and case styles. Model 10EH4C additionally incorporates a separate ground circuit inductor to isolate equipment chassis from power line ground at RF frequencies.
For more information on UL2601, see Appendix C.
Note: When using the models with an IEC connector (e.g. 6EH4), remember that the leakage current of the companion line cord, GA400, is more significant than that of the filter – which may disqualify this line cord from use in patient care applications.
H8
E
H9
E
EH4/EH4C
H1
E
H5
E
Specifications
Maximum leakage current,each line-to-ground
@ 120 VAC 60 Hz: 2µA @ 250 VAC 50 Hz: 5
Hipot rating (one minute):
line-to-ground 2250 VDC line-to-line 1450 VDC
Operating frequency: 50/60 Hz Rated voltage (max.): 250 VAC
Minimum insertion loss in dB:
Line-to-ground in 50 ohm circuit
Current Frequency-MHz
Rating .15 .5 1 5 10 30
3A 18 27 30 30 27 18 6A 9 16 20 26 23 18
10A 7 13 15 17 16 14 15A 5 9 11 12 11 9
µA
*15EH4 - VDE rated at 10 amps, 250 VAC
Line Cord
Line Cord No. GA400:
1/2
foot, 3-conductor line cord to mate with all
Electrical Schematic
SVT type 7 models except EH1. 50 micro amps maximum leakage @120 VAC, 60 Hz.
Case Style
EH1
Resistor location for reference only.
Typical dimensions:
Terminals: .250 [ Mounting holes: .188 [
66
Catalog 1654001 Dimensions are in inches and Dimensions are shown for USA Cust. Svc.: 1-800-468-2023 www.corcom.com Revised 05-07 millimeters unless otherwise reference purposes only. CORCOM Prods: 1-847-680-7400 www.tycoelectronics.com
specified. Values in italics are metric equivalents. to change.
Specifications subject
6.35] (5) Holes: .07[1.8] Dia.(4)
4.78] Dia. (2)
Germany: 49-89-6089-0
Slot: .07 x .16 [
1.8 x 4.1]
CORCOM Product Guide
E
D
A
C
B
A
C
B D E
D E
A
C
B
.65
16.5
Max.
C
A
F
F
B
Brown
Green/Yellow
Blue
D E
.140
3.56
.062
1.58
.820
20.8
R
Dia. (2)
1.200
30.48
1.575
40.01
.140
3.56
.234
5.94
.900
22.9
R
Dia. (2)
1.200
30.48
1.575
40.01
Blue Green/Yellow
Brown
B
C
F
F
D
E
A
RFI Power Line Filters for UL2601 Healthcare Equipment (Continued)
H Series
Case Styles Case Dimensions
H3
E
ypical dimensions:
T
ire leads: 4.0 [101.6] Min. 18 AWG Mounting holes: .188 [4.78] Dia. (2)
W
EH4 & EH4C
EH9
Part No. (max) (max) (max) (max) (ref)
3EH1
EH1
6 10EH1 3EH3 6EH3 10EH3
6EH4
6EH5
6EH8
6EH9
10EH4
10EH4C
15EH4
ABCDEF
.25 1.82 0.66 2.125 2.53
2
57.2 46.1 16.7 53.98 64.2
.96 1.82 0.66 2.125 2.53
24.40 46.1 16.7 53.98 64.2
2.20 1.19 0.81 1.575 1.98
55.9 30.2 20.6 40.01 50.3
1.55 1.19 0.85 1.575 1.98 .295
39.4 30.2 21.6 40.01 50.3 7.5
1.56 1.19 0.81 1.575 1.98 .295
39.7 30.2 20.6 40.01 50.3 7.5
1.55 1.19 0.85 1.575 1.98
39.4 30.2 21.6 40.01 50.3
2.62 1.19 0.81 1.575 1.98
66.5 30.2 20.6 40.01 50.3
2.62 1.19 0.81 1.575 1.98
66.5 30.2 20.6 40.01 50.3
2.62 1.19 0.81 1.575 1.98
66.5 30.2 20.6 40.01 50.3
± .015
±
.38
RFI Power Line Filters
Typical dimensions:
Terminals: .250 [ Mounting holes: .132 [
EH5
6.35] (3) Holes: .07 [1.8] Dia.(2) Slot: .07 x .16 [1.8 x 4.1]
3.35] Dia. (2)
3EH1 06EH8 3EH3 06EH9 6EH1 10EH1 6EH3 10EH3 6EH4 10EH4 6EH5 10EH4C
15EH4* Line Cord No. GA400 Insulating Boot No. (EH4, EH4C) FA601
*UL, CSA — 15A; VDE — 10A
Recommended Panel Cutouts
EH8
Part Numbers
Typical dimensions:
Catalog 1654001 Dimensions are in inches and Dimensions are shown for USA Cust. Svc.: 1-800-468-2023 www.corcom.com Revised 05-07 millimeters unless otherwise reference purposes only. CORCOM Prods: 1-847-680-7400 www.tycoelectronics.com
specified. Values in italics are metric equivalents. to change.
Wire leads: 4.0[101.6] Min. 18 AWG
Mounting holes: .132 [
3.35] Dia. (2)
Specifications subject
Note 1: EH4, EH4C and EH8 allow front or back mounting. Note 2: EH5 and EH9 allow back mounting only.
Germany: 49-89-6089-0
Panel Cutout (back mount)Panel Cutout (front mount)
Tolerance ±.005 [0.13]Tolerance ±.005 [0.13]
67