DOOR GASKET & DOOR BELLOWS
Check chamber Door Gasket for any steam leaks, hissing, or water
bubbles at
the Door Bellows. If steam is leaking at the door closing
device, then rotate the Gasket 180˚ to see if the leak follows it. If the
leak follows the Gasket, then replace the Gasket. If the leak does not
follow the Gasket, then replace the Door Bellows.
ISOLATING LEAKS & CORRECTIVE ACTION
HOW TO CHECK THE AIR JET VALVE
Water and steam leaks not only cause damage to the site where the unit is located but also,
will create a low water condition resulting in overheating that could cause major damage to the autoclave.
WATER/STEAM SYSTEM
TROUBLESHOOTING
AIR JET VALVE
Inspect the Air Jet Valve. It should make a slight hissing sound
throughout the STE cycle. If there is excessive hissing, steam, or
water bubbles escaping from the Air Jet Valve, refer to HOW TO
CHECK THE AIR JET VALVE, at the top right hand side of this
page.
Service Tip: Use a dental mirror to help locate the leak.
SAFETY VALVE
Remove the Water Reservoir Fill Cover and visually inspect the Safety
Valve – use a dental mirror to help locate the leak. Confirm that there
is no steam or water drops escaping from the vent holes or threads
of the Safety Valve. If a leakage is observed, replace the Safety Valve.
MULTI-PURPOSE VALVE
Inspect the Multi-Purpose Valve for leakage. Note any leaks at the
three fittings or the valve stem. If none are found, disconnect the
Condensation Coil in the water reservoir at the point where it connects inside the reservoir. Operate the sterilizer in STE mode at
273˚F for 30 minutes on M units
and 15 minutes on MK &
Valueklave 1730 MKV units, and look for any signs of leakage back
into the reservoir from the tubing fitting where the Condensation Coil
was attached. Inspect the water fill tube at the bottom of the reservoir for any signs of steam bubbling back into the reservoir. If any
leakage is noted at either position, repair or replace the MultiPurpose Valve.
Important: Reconnect Condensation Coil before
exhausting Chamber pressure.
CHAMBER & INTERNAL TUBING
Carefully inspect for steam or water bubbles at the Chamber and all
fittings. If a leak is detected at one of the fittings or tubing, tighten
or replace only after the unit has been depressurized and allowed to
cool down.
Visual and audible leaks can be
detected by operating the sterilizer in the normal STE mode with
temperature set at 273º F and the
time set for 30 minutes on M
units and 15 minutes on MK &
Valueklave 1730 MKV units.
Possible points of water/steam
leaks with corrective action, and
order in which they should be checked are noted in #1-5 below.
HOW TO UNCLOG THE MULTI-PURPOSE VALVE DURING A CYCLE
1. Refer to Table A, page 1, and pour the indicated amount of water into
the sterilizer
. Turn the power switch ON.
2. Close and lock the sterilizer door – be sure to make a tight seal and wait
for the heat light to come on.
3. Set the sterilizer to the following settings: Multi-Purpose Valve (MPV) set
to STE position; Timer Knob set to 20 minutes; and, Thermostats Knob to
273˚F (134˚C). Then press power switch to START.
(Note:
With the MPV in
the STE position, heaters will be
ON
, and sterilizer will begin to build pressure.
)
4. When the Chamber pressure reaches 30-31 PSI:
• Turn the Power Switch to the OFF position.
• Turn the MPV to the FILL position. Now the Chamber pressure should
force out debris from the MPV through the Fill Line into the Reservoir.
• When the pressure in the Chamber reaches 0, turn the MPV to the OFF
position, then open the door. Allow the sterilizer to cool.
• Clean out any debris from the inside of the Chamber.
• If the MPV is still clogged, rebuild or replace it.
HOW TO MANUALLY UNCLOG THE MULTI-PURPOSE VALVE
1. Disconnect power from sterilizer. Allow to cool. Remove covers.
2.
At the center port of the Multi-Purpose Valve (MPV), disconnect the fit-
ting. Set MPV to the FILL position.
Service Tip: Drain most of the water
from the Reservoir to prevent excess spillage during this process. This
will also verify that the Drain Tube and Drain Valve are clear.
• If water flows into the Chamber, then the obstruction has been c leared
from MPV and Fill Tube.
•If water does not flow into the Chamber
, check Fill Tube as follows:
Disconnect Fill Tube fitting at bottom of MPV.
- If water flows
, Fill Tube is clear, but MPV must be rebuilt or replaced.
- If water does not flow
, use forced air through Fill Tube and check for
bubbles in Reservoir. If procedure does not clear obstructions, replace
Fill T ube.
Service Tip:When disconnecting Fill Tube, straighten portion
of tube (about 1" lg.) that protrudes into the bottom of Reservoir.
Support Reservoir boss with a wrench.
3. If MPV and Fill Tube are clear, next check Chamber Tube as follows:
Disconnect and remove MPV from Chamber Tube. Use forced air or water
through tube. If procedure does not clear obstructions, replace Chamber
Tube. Also check and clear Chamber Fitting and boss of any obstructions.
4. Check exhaust lines as follows: At the top port of MPV, disconnect
Condensing Coil Tube, see MULTI-PURPOSE VALVE, page 9. Force air
through tube. If the flow is blocked, determine whether Condensing Coil
or Tube is obstructed. Clear obstruction or replace coil and/or tube.
5. If no leaks or obstructions have been found by following the previous
steps, and sterilizer is still experiencing a low water condition resulting in
overheating problems, see HOW TO CHECK THE AIR JET VALVE, above.
1
1
2
2
3
3
4
4
5
5
3
SERVICE TIP
T o help prevent
cl
ogging of the MPV,
install RPI Filters (RPI
Part #MIF062) into
the Water Fill and
Exhaust Lines.
SERVICE TIP
If the Chamber is found to be
defective, tag it as "Out of
Operation". Removal of the
power cord is recommended
until the chamber is replaced.
If a water/steam leak is not related to #1-3 above, disconnect power
from the sterilizer and remove the cover, then carefully remove the insulation blanket. Proceed to #4-5, below.Warning!
Make sure power has
been disconnected prior to removing the cover. When running the sterilizer with the cover removed, the interior of the machine will be very hot
- use extreme caution.
1. Refer to Table A, page 1, and in a
measuring cup, fill it with the amount
of water indicated in the chart for the
corresponding model, then pour the
measured water into the Chamber.
2. Bypass the FILL setting to manually
run sterilizer in STE mode at 273˚F for 30 minutes for M units
and 15
minutes for MK & Valueklave 1730 MKV units. After 30 minutes (or 15
minutes), shut off power , but leave MPV in STE mode until chamber pressure is reduced to 0 PSI and chamber has cooled (approx. 15 minutes).
3. Open Chamber Door, siphon water back into the measuring cup, and
measure the amount of water remaining in the chamber.
- If remaining water is less than 50%
of the original volume and no other
leaks were detected, replace Air Jet Valve.
- If remaining water is greater than 50%
and the pressure did not reach
28 psi within the nominal times (see TABLE D, above), and no
fault
was found within the heating system, then replace the Air Jet Valve.
REMOVING OBSTRUCTIONS
TABLE D
NOMINAL TIMES FOR REACHING 28 PSI
Time from Time from
Model Hot Start Cold Start
M Series 19 min. 24 min.
MK Series & MKV 6 min. 9 min.